ME COMPACT 9_Service_Manual_ENG_Ver1.1
ME COMPACT 9_Service_Manual_ENG_Ver1.1
SERVICE MANUAL
Version 1.1
www.medical-econet.com
Service Manual of COMPACT 9
Revision History
Edition Date Page Comment
Table of Contents
Section 1 General Information .................................................................................................. 5
1-1. Introduction................................................................................................................ 5
1-2. Features of the equipment........................................................................................ 5
1-3. Equipment description .............................................................................................. 6
1-4. System installation ................................................................................................... 13
1-5. Terms of warranty .................................................................................................... 15
1-6. Specification ............................................................................................................ 16
Section 2 Troubleshooting ....................................................................................................... 21
2-1. Introduction.............................................................................................................. 21
2-2. Trouble Shooting Guide .......................................................................................... 22
Section 3 Technical Description.............................................................................................. 37
3-1. Introduction.............................................................................................................. 37
3-2. COMPACT 9 Hardware Overview........................................................................... 37
3-3. Power Supply ........................................................................................................... 38
3-4. Main digital Board ................................................................................................... 40
3-5. Analog1 Board......................................................................................................... 41
3-6. Analog2 Board......................................................................................................... 42
3-7. Analog3 Board......................................................................................................... 43
3-8. Noninvasive Blood Pressure Assembly .................................................................. 43
Section 4 Check Out ............................................................................................................... 44
4-1. Visual Inspection ..................................................................................................... 45
4-2. Cleaning................................................................................................................... 45
4-3. Battery Maintenance .............................................................................................. 46
4-4. Electrical Safety Tests .............................................................................................. 48
4-5. ECG Test ................................................................................................................... 57
4-6. Respiration Test ........................................................................................................ 59
4-7. Temperature Test ..................................................................................................... 60
4-8. Cardiac Output Test (option) .................................................................................. 60
4-9. Invasive Blood Pressure Test (option)..................................................................... 61
4-10. Pulse Oximetry Test for SpO2 Oximeter ................................................................. 62
4-11. Noninvasive Blood Pressure Test ............................................................................ 63
4-12. NIBP Calibration ....................................................................................................... 64
4-13. ECG Calibration ....................................................................................................... 66
4-14. IBP Calibration ......................................................................................................... 66
4-15. End tidal CO2 Test (Option) .................................................................................... 66
▣ Front view
⑤ Function Key
⑦Power Key
▣ Rear view
① Alarm Lamp
② Nurse Call
③ RJ45 LAN Port
④ VGA Output
⑤ DC Power Input Port
⑥Wireless LAN Power
⑦Protective Ground Terminal
PRINTER
USB / SDCARD
SLOT
EXTENSION
BATTERY SLOT
① Printer
② USB/SD Card Slot
③ Extension Battery Slot
CSM
CONNECTOR
ECG
CONNECTOR
NIBP
CONNECTOR
SpO2
CONNECTOR
TEMPERATURE
CONNECTOR
EtCO2
CONNECTOR
IBP
CONNECTOR
C.O.
CONNECTOR
① CSM Connector
② ECG Connector
③ NIBP Connector
④ SpO2 Connector
⑤ Temperature Connector
⑥ EtCO2 Connector
⑦ IBP Connector
⑧ C.O. Coonector
▣ Buttons
▣ Indicator Lights
▣ Rotary Key
Installation precautions
When installing COMPACT 9 system, take the following precautions.
- The equipment should be operated within 10∼40℃ temperature and
30∼85%humidity.
- Check the power cord connection and probe should be handled with care.
- Do not plug several power cords in one electric outlet.
- Main unit should be installed on an even, level place.
- Check that the equipment is properly grounded to avoid noise.
- The entire system settings are stored on the internal memory even when system
power is turned off and on again
- Avoid excessive shock or vibration that could result in equipment damage.
- Avoid using in the vicinity of dust and flammable material.
The equipment operates when the power adaptor end connect into the COMPACT 9
power adptor. Replace adapter of the equipment if power is not turned on despite that
power adpator is connected and battery power is in normal condition. Contact service
dept. of Medical ECONET Co., Ltd. if the equipment is not operated after replacing adaptor.
Product information
Manufacture : Bridgepower corp.
Model Name: BPM050S18F02
Input power: 100~240V 1.2A
Output power: 18V ,3.2A
Caution
This equipment must only be connected to a supply mains with protected earth.
Noise or distortion of signals using non-off-the-shelf products rather than adapters supplied by
our company may be caused. Also dangerous to the safety related isolation and protection by
using the adapter supplied by our company in order to give
This illustration shows a system where both the medical electrical equipment and thenon-
medical electrical equipment are situated at the patient’s bedside.
For any IT equipment (IEC60950) operated in patient vicinity ensure that the equipment
leakage current does not exceed the limits described in IEC 60601-1. Use a separation
device to ensure compliance. After installation of IT equipment in patient vicinity, an
1-7. Specification
Display, Resolution 12.1” color TFT, 800 x 600 pixels
Dimension, Weight 322(W) x 250(H) x 224.8(D) mm, Approx. 4.5kg
Parameter ECG, Heart Rate, Respiration Rate, SpO2, Pulse Rate, Systolic BP,
Diastolic BP, Mean BP, 2 x Temperature, 2 x IBP, EtCO2, FiCO2,
Airway Respiration Rate
Trace 6 waveforms : 2*ECG, SpO2, RR or EtCO2, 2*IBP
Sweep speed : 6.25, 12.5, 25, 50 mm/sec
Indicators Categorized alarms (3 priority levels), Visual alarm lamp handle
QRS beep & SpO2 pulse beep, Percent(%) SpO2 pitch tone
Battery status, External power LED, Touch screen, Rotary knob
Interfaces DC input connector : 12 to 18VDC, 2.5A
Defibrillator Sync. Output :
- Signal Level : 0 to 5V pulse
- Pulse width : 100 ± 10 ms
LAN digital output for transferring data, Nurse call system
connection -0.3A at 125VAC – 1A at 24VDC
DC output : 5VDC, 1A Max
USB Barcode Scanner, USB & SD memory data storage
Battery Rechargeable Li-ion battery, 1hours for continuous working
Thermal Printer Speed : 25, 50mm/sec, Paper width : 58mm
(option)
Data Storage 168hours trends, 20cases of 10sec alarm waveform
Language English, French, Spanish, Italian, Germany, Chinese, Russian,
Czech, Bulgarian, Portuguese, Romanian, Hungarian, Turkish,
Polish,Korean
ECG Performance
Lead type 3-lead, 5-lead, 10-lead(option)
Lead Selection 3-lead : I, II, III
5-lead : I, II, III, aVR, aVL, aVF, V
10-lead: I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
ECG waveforms 3-lead : 1 channel
5-lead : 2/7 channels
10-lead: 12 channels
Heart Rate Range Adult : 30 – 300 bpm
NIBP Performance
Method Oscillometry with linear deflation
Operation Mode Manual/Automatic/Continuous
Measurement range Adult Pressure : 20 to 260 mmHg
Pediatric Pressure : 20 to 160 mmHg
Neonate Pressure : 20 to 130 mmHg
Accuracy Meets accuracy requirements of ANSI/AAMI SP10:1992 and 2002
Temperature Performance
Measurement range 15 to 45℃ (59 to 113℉)
Accuracy ±1℃
Compatibility YSI Series 400 temperature probes
Section 2 Troubleshooting
2-1. Introduction
A list of system error messages and troubleshooting information for common problems
you may encounter while using the monitor and its accessories is given in the manual for
Use. This chapter provides a guide for qualified service personnel for troubleshooting
problems that cannot be resolved by the user.
Caution
Warning
High Voltage - Voltages dangerous to life are present in the
instrument when it is connected to the mains power supply. Do not
perform any disassembly procedures with power applied to the
instrument. Failure to adhere to this warning could cause serious
injury or death.
Problems
Before starting any detailed troubleshooting, complete a thorough visual inspection of the
following.
All cable connections secure?
Devices properly powered?
Connected to a proper power source?
switch to AC from battery power. processor PCB to recognize. Replace the processor/power
management assembly.
Missing segments in the graph Print head may be dirty or Clean the printhead. Refer
data. defective.
to “Clean the Print Head”
Perform a graph test. If
problem persists, replace the
printer assembly.
Optional Alarm Light Indicator Cable may be loose or Restore the connection.
The red or yellow lights do not disconnected. Replace the alarm light
light on boot up of the patient LEDs are burned out. assembly or PCB.
monitor.
Video Display Possible burned-out pixels. Run the display tests in the
There are bars/strips of pixels boot loader.
missing on the display in Replace the display
rows/columns. Or only one assembly.
row/column of pixels on the No display
display is missing or never
turned on. Display may be in standby Press the POWER button. If
The remaining portion of the mode. display still does not appear
display functions properly. Backlight inverted may be within 20 seconds, replace the
defective. display
assembly.
Replace back light inverter.
Network Connector The patient monitor is Connect the network cable
No network connection when connected to a BM Central to the remote station’s internet
a station. The patient monitor’s connector.
network cable is connected to Network Connector is inactive Enable network in boot code.
the when it is connected to the
patient monitor’s Network remote station.
Connector.
Network is disabled in boot
code.
with these patient signals, a function of data acquisition, which is generated by the analog PCB.
1) Power is off
Does the
Change the
power input
Yes power input A
cable break
cable
down?
No
A
Does the
Change the
power switch
Yes power switch A
cable short-
cable
circuit?
Normal
Yes Use
No operation
Does the
Change the B
main power No
Yes main power A
connector
connector
break down?
No
No Yes
Yes
Ask the
manufacturer
about it, then
take action
LCD screen is
blank although the
power is on
No
Change the
Broken 34Pin Normal
Yes 34Pin shield A Yes Use
shield cable? operation
cable
No No
Is the LCD
module Change the B
Yes A
unsatisfactory LCD Module
?
No
Ask the
manufacturer
about it, then
take action
4) Noise in ECG
- Gel is dry
- Electrodes does not stick well to skin
The patient’s skin Yes Apply ECG gel or water to the area
is extremely dry of contact then use the electrode
No
No
No
5) SpO2 malfunction
No
No
6) Temp malfunction
No
7) NIBP malfunction
No
The patient moved Yes Make sure the patient stay still
while measuring during diagnosis
No
No
No
Does the
Does the
paper come Change the A
No motor work No
out while motor
properly?
printing?
Yes
Have you
Connet the
connected the A
No motor power
motor power
cable correctly
cable correctly?
Yes Yes A
Normal
Yes Use
operation
Does the Have you
signal come Yes connected Connect the
A
out on the No the TPH No THP power
paper while powor cable cable correctly
printing? correctly
Yes B
Is the TPH
power cable Change the TPH A
Yes No
is power cable
satisfactory?
Yes
Ask the
manufacturer
about it, then
take action
10) ECG wave does not come out or it's very small
Is the patient
Reconnect the
cable
No patient cable A
connecter
carefully
correctly?
A
Is the
electrode Cleanse the
connected to electrode with
Yes A
the patient alcohol or water,
cable then it use
correctly? Normal
No Yes Use
operation
Is the
electrode Reconnected the
connected to electrode to the
No A
the patient patient cable
cable properly
correctly? B
Yes
Is the patient
Change the
cable short- Yes A
patient cable
circuited?
No
Control the
Is the ECG
volume by setting
signal adjusted No A
a gain on the
properly?
main board
Yes
Ask the
manufacturer
about it, then take
action
Yes
No
Warning
Under normal conditions, after switching off the unit all internal
capacitors will discharge to safe levels within ten seconds. Under
fault conditions however, high voltage charges may be held for
much longer and suitable precautions should be taken before
handling a disconnected or removed Power Supply Module.
The COMPACT 9 is connected between the local source of AC mains and the double-
fused IEC mains input connector on the adaptor module, using the supply lead provided.
The unit will operate from AC supplies in the range 100 to 120 volts or 220 to 240 volts without
adjustment. Note however, that the supply input fuses located in the inlet connector should
be rated at 1.5 Ampere for supplies in the range 100 to 120 volts, 0.75 Ampere for 200 to 240
volts.
Fig 3.2 Block Diagram: Power Supply Part in the Power Board
The various regulated output voltages produced by the Power Supply part are routed to
the Power Board for use on the board and also for distribution to other boards and
assemblies. As shown in Fig 3.2, Four voltage regulator devices are fitted to the power board;
two of these are powered from the +18 V supply, 7.5 V regulator for the printer motor and
printer head. The remaining Voltage provide the +5Vdigi and two +5Vanlog supplies from
the +12V supply.
Microprocessor Section:
- S3C2440 ( ARM9TDMI ) microprocessor.
- Flash containing all system software and a boot program allowing
- Software upgrades via the Service Connector( USB HOST , JTAG ).
- Flash memory, backed to store short time device and user settings etc.
- A real-time clock, battery-backed and power-fail protected. The processor has
access to all clock registers.
- A watchdog function which monitors the operation of the microprocessor and
restarts the system if it is not served at a constant rate.
Service Interface: A RS232 serial interface is provided for use during production
and service. Connection is via a DSUB 9pin on the rear cover.
- Serial interface for option and software upgrades and Debugging.
- Settings and configuration via an external computer.
Printer interface :
- The Parallel( Main Digital CPU to Printer CPU) for the thermal array data.
- A heating timer to control the heating pulse for the thermal array. The voltage
from a thermistor on the thermal array is A/D converted and the CPU programs
the heating timer with a value which compensates the ambient temperature.
- The recorder sensing hardware consists of a reflective light sensor that detects
the black marks on the paper and paper out.
- +7.5V is blocked by capacitor and supplied to the thermal print head.
ARM7 core
DSP
(algorithm,
I/O processor)
DAC
HOST
Frequency
Medical ECONET recommends that you perform the checkout procedures:
Upon receipt of the device.
Every 12 months thereafter.
Each time the main enclosure is disassembled or a circuit board is removed, tested, or replaced.
4-2. Cleaning
Precautions
Use one of the following approved solutions:
- Cidex solution, or
- Sodium hypochlorite bleach (diluted), or
- Mild soap (diluted)
- Lint-free cloth
- Dust Remover (compressed air)
To avoid damage to the equipment surfaces, never use the following cleaning agents:
- organic solvents,
- ammonia-based solutions,
- acetone solution,
- alcohol-based cleaning agents,
- Betadine solution,
- a wax containing a cleaning substance, or
- abrasive cleaning agents.
Clean the Display
Use a soft, clean, lint-free cloth dampened with a glass cleaner to clean the display.
Caution
To avoid getting liquid into connector openings, do not spray
glass cleaning or general cleaning solutions directly onto the
product’s surface.
The battery is charged whenever the patient monitor is connected to an AC power source.
When you store the battery inside of a patient monitor that is connected to an power adapter
source, the battery will selfdischarge to less than 90% of its Full Charge Capacity after
approximately two weeks (depending upon the temperature of the battery). At this time, the
patient monitor will automatically recharge the battery to 100% of its Full Charge Capacity.
Warning
PATIENT RISK HAZARD — Never condition a battery while the
patient monitor is connected to a patient. Serious injury or death
could result.
Complete the following steps to manually condition the battery using the patient monitor:
1. Remove all batteries from the patient monitor.
2. Insert one battery into Slot of the patient monitor.
3. Disconnect the patient monitor from the AC adaptor power source.
4. Access the patient monitor’s Battery Status window to monitor the
battery’s Charge Level.
5. Allow the battery to discharge to less than a 90% Charge Level.
6. Re-connect the patient monitor to the AC adaptor power source and fully
charge the battery.
7. Disconnect the patient monitor from the AC adaptor power source and allow the patient
monitor to run until it displays the BATTERY LOW message, or until the patient monitor shuts down.
8. Re-connect the patient monitor to the AC adaptor power source. Allow the battery to fully
charge to complete the conditioning cycle.
How to Store the Battery
- Condition the battery before storing it. Conditioning the battery to 100% capacity
increases the storage shelf-life of the battery.
- Store the battery outside of the patient monitor at a temperature between 20°C to 25°C
(68°F to 77°F).
- When the battery is stored inside a patient monitor that is powered by an AC power
source, the battery cell temperature increases by 10°C to 15°C (18°F to 27°F) above the
room’s ambient temperature. This reduces the life of the battery.
- When the battery is stored inside a patient monitor that is continuously powered by an AC
power source and is not powered by battery on a regular basis, the life of the battery may
be less than 12 months. Medical ECONET recommends that you remove the battery and
store it near the patient monitor until it is needed for transport.
- Re-condition a stored battery every six months to increase the storage shelf-life.
How to Replace the Batteries
1. Open the battery doors on the left side of the patient monitor, along the bottom.
2. Remove the faulty batteries.
3. Replace with a new battery. The patient monitor supports up to two exchangeable lithium-ion
batteries. Install the battery with the connection pins facing down and inserted first.
4. Close the battery covers.
5. Verify that the patient monitor operates correctly.
a. Confirm that the Battery with a battery icon displays in the upper right corner of the patient
monitor.
b. Verify that the Battery LEDS illuminate amber.
Warning
EXPLOSION HAZARD — DO NOT incinerate the battery or store at
high temperatures. Serious injury or death could result.
Test Conditions
Electrical safety tests may be performed under normal ambient conditions of temperature,
humidity, and pressure.
Test Equipment
The recommended test equipment required to perform electrical safety tests is listed below.
Item Specification
Leakage Current Tester Equivalent to the circuits shown
Digital Multimeter (DMM) AC volts, ohtm
Ground Bond Tester 0 – 1ohm
ECG Test Body All leads together
SpO2 Test Body %spo2, Pulse rate
LINE
NEUTRAL
GROUND
For international power outlets, refer to the internal standards agencies of that particular country.
If other than normal polarity and ground is indicated, corrective action must be taken before
proceeding. The results of the following tests will be meaningless unless a properly wired power
outlet is used.
Use this procedure to verify that the AC power outlet that the patient monitor is plugged into is
correctly wired. This is accomplished by performing a:
- voltage measurement between all three connections of the power outlet, and
- ground-to-neutral loop resistance measurement.
1. Use a digital multimeter (DMM) to measure the voltage between the three connections.
a. Select the AC voltage scale on the DMM.
b. Measure the voltage from line to neutral, line to ground, and neutral to ground and make sure
these are correct. With a correctly wired power outlet used in korea, the following readings
should be obtained:
Line to neutral: 220 VAC
Line to ground: 220 VAC
Neutral to ground: < 3 VAC
Readings other than these indicate improper wiring. Have the power outlet checked by an
electrician.
2. Use a DMM to measure the ground-to-neutral loop resistance.
Caution
DO NOT check the ground-neutral loop resistance unless the
power outlet is correctly wired.
GROUND pin
Perform the test method below that is required by your Country/Local governing safety
organization.
2. Connect the power cord of the device under test to the power receptacle on the leakage
tester.
3. Apply power to the device under test.
4. Read the current leakage indicated on DMM.
5. Set the polarity switch on the leakage tester to RVS (reverse).
6. Read the current leakage indicated on DMM.
NOTE
If either reading is greater than the appropriate specification below, the device under test fails.
Contact Medical ECONET Technical Support.
- 300 µA (0.3 volts on the DMM), and the device under test is powered from 100-120 V/50-60
Hz
- 300 µA (0.3 volts on the DMM), and the device under test is powered from a centered-
tapped 200-240 V/50-60 Hz, single phase circuit
- 500 µA (0.5 volts on the DMM), and the device under test is powered from a non-center-
tapped, 200-240 V/50-60 Hz, single phase circuit
NOTE
Center-tapped and non-center-tapped supply circuits produce different leakage currents and the
UL and IEC limits are different.
7. Repeat steps 1 through 7 with NEUTRAL open.
2. Connect an ECG test body to the ECG/RESP connector of the device under test.
3. Set the power switch of the device to ON.
4. Read the leakage current indicated on the DMM.
5. Change the leakage tester polarity switch to the RVS position.
6. Read the leakage current indicated on the DMM.
NOTE
If either reading is greater than 50 µA (0.05 volts on the DMM), the device fails this test. Contact
Medical ECONET Technical Support.
7. Change the GND switch to the Closed position.
8. Read the leakage current indicated on the DMM.
9. Change the leakage current switch to the RVS position.
10. Read the leakage current indicated on the DMM.
NOTE
If either reading is greater than 10 µA (0.01 volts on the DMM), the device fails this test. Contact
Medical ECONET Technical Support.
11. Set the power switch of the device to OFF.
NOTE
The AAMI and IEC single fault condition (ground open) is 50 µA, whereas the normal condition
(ground closed) is less.
12. Repeat the steps in this procedure using the appropriate MEDICAL ECONET or SPO2 Test Body.
Connect the SPO2 Test Body to the Green SPO2 connector of the device under test.
Patient (Sink) Leakage Current Test (Mains Voltage on the Applied Part)
This procedure only applies to Class I (grounded/earthed) equipment, and measures the leakage
current from a mains voltage source into the ECG/RESP connector.
1. Configure the leakage tester like the circuit shown below with GND switch CLOSED and polarity
switch NORM.
Warning
Shock hazard. The following step causes high voltage at the test
body. Do not touch the test body.
- 10 µA (0.01 volts on the DMM) at 120 VAC using the test body.
- 20 µA (0.02 volts on the DMM) at 240 VAC using the test body.
- 50 µA (0.05 volts on the DMM) at 120-240 VAC using the ECG cable.
NOTE
The 10 and 20 µA limits are based on internal design standards. The 50 µA limit is common to all
standards. AAMI ES-1 standard requires using the patient cable.
6. Set the power switch on the device to OFF.
7. Repeat the steps in this procedure using the appropriate MEDICAL ECONET or SPO2 Test Body.
Connect the SPO2 Test Body to the green SPO2 connector of the device under test.
- lead III automatically displays in place of lead II in the top trace position.
13. Reconnect the RA leadwire to the patient simulator.
14. Inject a 1-millivolt calibration signal using the patient simulator and start a manual graph.
15. Observe that the calibration pulse properly displays and graphs.
16. This completes the 5 Lead ECG test. Leave the ECG patient cable connected to ECG/RESP and
continue to the next steps of these checkout procedures.
6. Verify correct IT readings appear on the patient monitor display, as shown in the table above.
7. Disconnect the CO cable adaptor from the CO connector of the patient monitor. This completes
the CO tests.
N 5. Observe a reading of 200/200 (200) ± 4 mmHg (or 240 ± 4 mmHg if using the Marq II or MarqIII) in
the BP parameter window on the patient monitor display.
6. Change the patient simulator BP output to WAVE or to a 120/80 wave form (simulated BP
waveform).
7. Observe a distortion-free BP waveform and a reading of approximately 120/80 (93) in the BP
parameter window on the patient monitor display.
8. Disconnect the BP simulator cable from the BP connector of the patient monitor.
9. Complete steps for each enabled BP connector.
NOTE
A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor.
However, it can be used to demonstrate that a particular pulse oximeter monitor
reproduces a calibration curve that has been independently demonstrated to fulfill a
particular accuracy specification.
Procedures
1. Attach the digital manometer, noninvasive blood pressure (NIBP)cuff, tees and tubing, as shown
in the illustration below, to the NIBPconnector of the patient monitor.
2. Set the digital manometer power switch to the on position.
3. Set the digital manometer range switch to 200 mmHg.
Using the Trim Knob control, access the Service Mode menu starting from the Main Menu.
1. Select MORE MENU > SETUP > MAKER SERVICE.
2. Enter password using the Trim Knob control to select the day and month from patient monitor
screen with leading zeros.
3. Select CALIBRATE > NIBP > CHECK CAL.
The text on the menu item changes from CAL OFF to CAL IN PROGRESS.
Verify the readings in the NIBP parameter window on the patient monitor display and readings on
the digital manometer are equal (± 1mmHg) for at least one full minute. If the readings are not
equal for at least one full minute, the NBP circuit requires calibration.
4. Select CAL IN PROGRESS > STOP.
1
The pneumatic control circuit of the patient monitor vents air pressure in the pneumatic circuit of
the patient monitor to atmosphere and causes the NBP cuff to deflate.
5. Remove the NBP test setup apparatus from the patient monitor. The NBP tests are complete.
1
6. It is recommended that the performance be checked every 2 years and after maintenance and
repair, by utilizing the manometer mode.
Warning
When the NBP cuff is used in this procedure, it must be tightly
wrapped around a rigid cylinder or pipe. DO NOT put the NIBP
cuff around a human arm during the calibration procedures due
to the potential for injury.
Calibration Procedure
Using the Trim Knob control, access the Service Mode menu starting from the Main Menu.
10. Unplug the patient monitor from AC power source and remove the test apparatus from the
patient monitor.
ECG or BP Calibration
To complete the ECG or BP calibration, connect a voltmeter to the patient monitor.
1. Attach the analog output cable (pn 2000633-001) to the patient monitor.
2. Connect a 5.5 digit precision voltmeter to the port pin to be calibrated (If ECG: Pin 7 = ECG
signal, Pin 3 = ground. If BP: Pin 6 = BP, Pin 5 = ground.).
- Lead I:0.5 mV
- Lead II:1 mV
- Lead III:0.5 mV
- Lead V:–0.5 mV
5. It may be necessary to run a graph to accurately measure the cal pulses. Perform these steps to
graph all four waveforms.
a. From the main menu, select GRAPH & ALARMS.
b. Select GRAPH CONTROL from the menu.
c. Set the graphed waveforms for the ECG leads:
d. Press the Graph Go/Stop (Print on Dash 5000) front panel
control on the patient monitor to start and stop a manual graph.
e. Compare the printed graph with the sample shown below.
f. Change the patient simulator output from cal pulses to an 80bpm ECG waveform.
The displayed ECG waveforms should be similar to those shown in the figure below.
If this is the case, the ECG functions of the analog2 PCB, as well as communication between the
acquisition and processor PCBs function as designed.
1
3. Test the patient simulator and ECG patient cables on a working patient monitor to verify the ECG
signal.
4. If the ECG signal, patient simulator and ECG patient cables are good, the acquisition PCB is
suspect and you need to replace it.
11. Verify a stable heart rate display of approximately 80 bpm. Verify the pacemaker spikes are
again at the same amplitude.
12. Disable the pacemaker detection function of the patient monitor.
Setup IBP
Connect the Blood Pressure output of the patient simulator to each enabled IBP patient connector
on the side panel of the patient monitor.
Zero-Reference Enabled BP Input
Properly zero-reference each enabled BP input.
1. Set the patient simulator IBP output to 0 mmHg
2. Press the ZERO ALL (FUNCTION) key on the patient monitor.
Respiration Functions
Connect the Multifunction simulator and appropriate patient cables to the ECG/RESP side panel
connector on the patient monitor.
1. Adjust the patient simulator to output a respiration waveform using the following settings:
- Rate BPM – 30
- Baseline Impedance Ohms – 750,
- ∆R Ohms – 2.0.
2. Enable the respiration function of the patient monitor.
a. Select SETUP from the main menu display on the patient monitor,
b. Select PARAMETERS ON/OFF from the patient monitor setup menu.
c. Turn and push the Trim Knob control to scroll to and select RR in the parameters on/off pop-up
window.
d. Toggle and select ON in the RR line of the parameters on/off popup window.
3. Verify the following:
- Respiration rate is displayed and accurate.
- Respiration waveform is displayed and noise-free.
Respiration Functions Work Properly on Patient Simulator but not on Actual Patient
Refer to the Operator’s Manual for detailed information regarding patient preparation relative to
respiration monitoring functions.
Achieving optimum results for respiration waveforms and accurate respiration rate detection by the
patient monitor requires proper preparation for ECG electrode placement on the patient. Following
is an example of a noisy respiration waveform which is usually due to bad patient preparation.
NOTE
With patients that exhibit excessively high baseline chest
impedance, proper respiration monitoring can be extremely difficult,
if not impossible.
Noninvasive Blood Pressure Functions
Perform the noninvasive blood pressure (NIBP) Checkout Procedure found in the Maintenance
chapter. This procedure determines whether or not the NIBP functions of the patient monitor are
working as designed or whether the patient monitor requires NIBP calibration.
If, after performing the prescribed checkout procedure, it is determined that there are potential
problems that NIBP calibration does not cure, try the following:
1. If calibration is unsuccessful and cannot be properly performed, there could be leaks in the
pneumatic circuit plumbing. The following steps assist you in determining this.
The NIBP cuff and tubing is the easiest area to inspect for leaks and is also the most likely area for
failure in this regard. Closely inspect these items for cracks or leaks. Test the NIBP cuff and
tubing on a working patient monitor to determine the source of the problem.
If the NIBP cuff and tubing are determined to be good after testing them on a working patient
monitor, the leaks are probably internal to the patient monitor. Disassemble the patient monitor
and check inspect all internal tubing and connections in the pneumatic circuit plumbing.
2. If no leaks are found after performing the previous step, the NIBP pump assembly is suspect. Swap
the NIBP pump assembly with one from a working patient monitor and/or replace as necessary.
Warning
•Before attempting to open or disassemble the monitor, disconnect
it from the AC mains supply.
•Energized circuits are accessible with the covers open. Do not work
on the monitor with the covers open and AC power connected.
Only qualified service personnel should open or disassemble the
monitor.
6-1. Introduction
Remember to store all screws and parts in a safe place for later refitting.
① Nipper
② Screwdriver(4mm X 100mm)
③ Screwdriver(6mm X 200mm)
6-5. Disassembly
38) Removing the printer board and digital board cable of equipment
6-6. Reassembly
7-2. Accessory
Basic Accessory
Item Code
Picture Description
(Part No.)
Adult
Reusable type
Option Accessory
Item Code
Picture Description
(Part No.)
152600-
1 Multi ECG 3-Lead Button EN Cable For USA Option
023900
152600-
2 Multi ECG 3-Lead Button EU Cable For Europe Option
024200
152600-
3 Multi ECG 10-Lead Button EN Cable For USA
025000
152600-
4 Multi ECG 10-Lead Button EU Cable For Europe
025300
Child
Option
Option
Temperature sensor - E
152600-
15 Surface/Skin type
006800
Option
Temperature sensor - E
120108-
16 Rectal type
042600
Option
120108-
17 Single line monitoring kit. I-MX9604A Option
044400
120108-
21 Double line monitoring kit. I-MX9602A Option
044300
120108- X 2 EA
23
044200 I-MX960
Option
120108-
26 Single line monitoring kit. I-MX9504T Option
044100
Option
(1 per box)Option
CABLE:백색
1
2
3.백색(RA) 470-CSM-PAT-CABLE
42 N/A CABLE:녹색
2.녹색(LL
)
3S PLUG 4P
CABLE:흑색
1.흑색(LA)
COMPACT 9 SE CSM Patient Cable
cable
25)
54 N/A Sampling line with male luer lock connector, 2.0m (Box
of 25)
56 N/A 774HF75
57 N/A M3L9FHSI
2 RXD-Input
3 TXD-Output
4 NC
5 Ground-Isolated 0V reference
6 NC
7 NC
8 NC
9 NC
2) VGA
The 15-way D-type connector is the D-SUB interface connector. The pin connections are
given in Table below.
2 GREEN
3 BLUE
4,9,11 NC
12 CRT_SCL
13 HRTC
14 VRTC
15 CRT_SDA
3) USB
The 6-way A-type connector is a USB interface connector. The pin connections are given
in Table below.
2 Data interface D-
3 Data interface D+
4 Ground-Isolated 0V reference
4) SD Card
The 13-way connector is the SD card interface connector. The pin connections are given
in Table below.
5 Clock signal
7 Data signal 0
8 Data signal 1
9 Data signal 2
10 Write protect
12 Ground-Isolated 0V reference
13 Ground-Isolated 0V reference
4) LAN
The 12-way connector is the LAN interface connector. The pin connections are given in
Table below.
2 Data interface D+
3 Ground-Isolated 0V reference
4 Data interface D-
5 Data interface D-
6 Ground-Isolated 0V reference
7 Data interface D+
8 NC
9 NC
10 NC
Basic Board
SMPS
Connector Connection details
Connector Pin
Pin
1 1 Ground-Isolated 0V reference
2 2 LA CARRIER
3 3 LL CARRIER
4 4 RA CARRIER
5 5 RA MEA
6 6 LA MEA
7 7 LL MEA
8 8 RA ECG
9 9 LA ECG
10 10 LL ECG
11 11 V1 ECG
12 12 RL ECG
13 13 V2 ECG
14 14 V3 ECG
15 15 V4 ECG
16 16 V5 ECG
17 17 V6 ECG
18 18 CABLE DETECTION
19 19 Ground-Isolated 0V reference
20 20 Ground-Isolated 0V reference
Basic Board
SMPS
Connector Connection details
Connector Pin
Pin
1 1 Ground
2 2 NC
3 3 Ground
4 4 5VCC
5 5 ETCO2 Tx
6 6 ETCO2 Rx
7 7 CO2 RESET
8 8 NC
Basic Board
SMPS
Connector Connection details
Connector Pin
Pin
1 12 Ground-Isolated 0V reference
2 11 Ground-Isolated 0V reference
3 10 Ground-Isolated 0V reference
Basic Board
SMPS
Connector Connection details
Connector Pin
Pin
1 1 Ground
2 NC
3 NC
4 2 12VCC
Basic Board
SMPS
Connector Connection details
Connector Pin
Pin
1 1 VALVE ON2
2 2 POWER ON SIGNAL0
3 3 POWER ON SIGNAL1
4 4 Voltage SW signal
5 5 VALVE ON0
6 6 VALVE ON1
7 7 MOTOR ON
8 8 RxD0
9 9 TxD0
10 10 L RxD-
11 11 L RxD+
12 12 L TxD-
13 13 L TxD+
14 14 PORT0
15 15 PORT1
16 16 PORT2
17 17 PORT3
22 22 Timer output 0
23 23 RxA
24 24 TxA
25 25 RxB
26 26 TxB
27 27 RxD1
28 28 TxD1
29 29 PORT4
30 30 RxD2
31 31 TxD2
32 32 RxC
33 33 TxC
34 34 RxD
35 35 TxD
36 36 N.C.
Basic Board
SMPS
Connector Connection details
Connector Pin
Pin
4 11 Voltatge SW Signal
5 10 PHASE A signal
6 9 PHASE B signal
7 8 Rotary SW signal
8 7 Switch 0 signal
9 6 Switch 1 signal
10 5 Switch 2 signal
11 4 Switch 3 signal
14 1 Ground-Isolated 0V reference
Basic Board
SMPS
Connector Connection details
Connector Pin
Pin
2 2 USB D- SIGNAL
3 3 USB D+ SIGNAL
4 4 Ground-Isolated 0V reference
Basic Board
SMPS
Connector Connection details
Connector Pin
Pin
7 2 Ground-Isolated 0V reference
8 1 Ground-Isolated 0V reference
Basic Board
SMPS
Connector Connection details
Connector Pin
Pin
7 2 Ground-Isolated 0V reference
8 1 Ground-Isolated 0V reference
Basic Board
SMPS
Connector Connection details
Connector Pin
Pin
2 NC
3 1 Ground-Isolated 0V reference
4 3 Ground-Isolated 0V reference
Basic Board
SMPS
Connector Connection details
Connector Pin
Pin
2 2 Battery Ground
Basic Board
SMPS
Connector Connection details
Connector Pin
Pin
1 NC
2 NC
4 1 Ground-Isolated 0V reference
Basic Board
SMPS
Connector Connection details
Connector Pin
Pin
2 5 RxA
3 4 TxA
4 3 RxB
5 2 TxB
6 1 Ground-Isolated 0V reference
Basic Board
SMPS
Connector Connection details
Connector Pin
Pin
1 1 Black lead
2 2 GREEN lead
3 3 White lead
4 NC
Basic Board
SMPS
Connector Connection details
Connector Pin
Pin
3 3 RxB
4 2 TxA
5 1 Ground-Isolated 0V reference
<TOP>
<BOTTOM>
9-2. Analog1 Board(PCB-400-ANA1-C-3)
<TOP>
<BOTTOM>
17) Analog2 Board-7ch (PCB-460-ECG07-2)
<TOP>
<BOTTOM>
9-3. Analog2 Board-12ch (PCB-460-ECG12-2)
<TOP>
<BOTTOM>
9-4. Analog3 Board (PCB-470-ANA3-1)
<TOP>
<BOTTOM>
<TOP>
<BOTTOM>
<TOP>
<BOTTOM>
<TOP>
<BOTTOM>
<TOP>
<BOTTOM>
<TOP>
<BOTTOM>
<TOP>
<BOTTOM>
<TOP>
<BOTTOM>
<TOP>
<BOTTOM>
<TOP>
<BOTTOM>
<TOP>
<BOTTOM>