OPERATION MANUAL
TABLE OF CONTENTS
1. FEATURES..............................................................................................................
1
2. SPECIFICATIONS...................................................................................................
2
3. FRONT PANEL DESCRIPTION..........................................................................
6
4. MEASURING PROCEDURE....................................................................................
7
4-1 Sound Pressure Level Measurements (SPL)......................... 7
4-2 Integrating Sound Level Meter Measurements (Leq)......................................... 8
4-3 Data Hold..........................................................................................................
10
4-4 Data Record ( Max., Min. reading )....................................................................
10
4-5 LCD Backlight ON/OFF……………………………………….. 11
5. DATALOGGER........................................................................................................
12
5-1 Preparation before execute datalogger function................................................
12
5-2 Auto Datalogger ( Set sampling time ≧ 1 second )..........................................12
5-3 Manual Datalogger ( Set sampling time = 0 second )....................................... 14
5-4 Check time and sampling time information........................................................
14
5-5 SD Card Data structure.....................................................................................
15
6. Saving data from the SD card to the computer.....................................................
16
7. ADVANCED SETTING.............................................................................................
18
7-1 Set clock time ( Year/Month/Date, Hour/Minute/ Second )................................. 18
7-2 Set sampling time ............................................................................................
19
7-3 Auto power OFF management ........................................................................
19
7-4 Set beeper Sound ON/OFF...............................................................................
20
7-5 Decimal point of SD card setting........................................................ 20
7-6 SD memory card format...................................................................................
21
7-7 Frequency weighting to A or C setting............................................................
21
8. POWER SUPPLY FROM DC ADAPTER................................................................. 22
9. BATTERY REPLACEMENT.....................................................................................
23
10. RS232 PC SERIAL INTERFACE............................................................................
23
11.CALIBRATION
FREQUENCY WEIGHTING
.......................................................................................................
CHARACTERISTICS 25
12. FREQUENCY WEIGHTING CHARACTERISTICS
OF A & C NETWORKS..........................................................................................
25
13. TIME WEIGHTING (FAST & SLOW)
CHARACTERISTICS.............................................................................................
26
14. PATENT.................................................................................................................
26
1. FEATURES
* Frequency and Time weighting are designed to meet IEC 61672 class 2.
* The meter with programmable integrating time provides precise linearity
over a wide range (100dB) and displays Leq and SPL measurements.
* A & C weighting networks comply with standards.
* 0.5" standard microphone head.
* Time weighting (Fast & Slow) dynamic characteristic modes.
* Build External calibration VR.
* Auto range & Manual range selection.
* Condenser microphone for high accuracy & long-term stability.
* Memory function to store the Max. & Min. value.
* Hold and Peak Hold functions.
* Real time SD memory card Datalogger, it Built-in Clock
and Calendar, real time data recorder, sampling time set
from 1 second to 3600 seconds.
* Manual datalogger is available ( set the sampling
time to 0 second ), during execute the manual datalogger
function, it can set the different position ( location ) No.
( position 1 to position 99 ).
* Innovation and easy operation, computer is not need to
setup extra software, after execute datalogger, just take
away the SD card from the meter and plug in the SD card
into the computer, it can download the all the measured
value with the time information ( year/month/date/
hour/minute/second ) to the Excel directly, then user can
make the further data or graphic analysis by themselves.
* SD card capacity : 1 GB to 16 GB.
* LCD with green light backlight, easy reading.
* Can default auto power off or manual power off.
* Data hold, record max. and min. reading.
* Microcomputer circuit, high accuracy.
* Power by UM3/AA ( 1.5 V ) x 6 batteries or DC 9V adapter.
* RS232/USB PC COMPUTER interface.
* Heavy duty & compact housing case.
* Leq, SPL, MAX, and MIN measurements.
1
2. SPECIFICATIONS
Circuit Custom one-chip of microprocessor LSI
circuit.
Display LCD size : 51 mm x 30 mm
LCD with green backlight ( ON/OFF ).
Measurement SPL : Sound pressure level
Type Leq : Equivalent Continuous Noise Level
Measurement 30 - 130 dB.
Range
Resolution 0.1 dB.
Function dB ( A & C frequency weighting ),
Time weighting ( Fast, Slow ),
Peak hold, Data hold
Record ( Max., Min. ).
Accuracy Characteristics of " A " frequency
(23 ±5 ℃) weighting network meet ANSI S1.4-2014 /
IEC 61672 -1 : 2013 class 2
Under 94 dB input signal, the accuracy
are :
31.5 Hz ±3.0 dB
63 Hz ±2.0 dB
125 Hz ±1.5 dB
250 Hz ±1.5 dB
500 Hz ±1.5 dB
1K Hz ±1.0 dB
2K Hz ±2.0 dB
4K Hz ±3.0 dB
8 K Hz ±5.0 dB
Remark :
The above spec. are tested under the
environment RF Field Strength less
than 3 V/M & frequency less than 30
MHz only.
2
Frequency Characteristics of A & C.
Weighting A weighting :
Network The characteristic is simulated as "Human
Ear Listing" response. Typical, if making
the environmental sound level
measurement, always select to A
weighting.
C weighting
The characteristic is near the "FLAT"
response. Typical, it is suitable for
checking the noise of machinery (Q.C.
check) & knowing the sound pressure
level of the tested equipment.
Time Fast - t = 125 ms
weighting * "Fast" range is simulated the human ear
(Fast & Slow) response time weighting.
Slow - t = 1 s
* "Slow" range is easy to get the average
values of vibration sound level.
* Proposal to test FAST & SLOW Function ,Please use
the manual range mode
Data hold To freeze the measurement value.
Peak hold To keep the peak ( max. ) measurement
value.
Range selector Auto range : 30 to 130 dB.
Manual range :
2 range, 30 to 80 dB,
80 to 130 dB, 50 dB on each step,
with over & under range indicating.
Frequency 31.5 to 8,000 Hz.
Microphone Electric condenser microphone.
type
Microphone Out size, 12.7 mm DIA. ( 1/2 inch).
size
Data error no. ≦ 0.1 % no. Of total saved data typically.
3
Calibration VR Build in external calibration VR, easy to
calibrate on 94 dB level by screw driver.
* Calibrated via external SOUND
CALIBRATOR ( SC-942, optional ).
Calibrator B & K (Bruel & kjaer), MULTIFUNCTION
ACOUSTIC CALIBRATOR Type 4226.
Datalogger Auto 1 second to 3600 seconds
Sampling Time @ Sampling time can set to 1 second,
Setting range but memory data may loss.
Manual Push the data logger button
once will save data one time.
@ Set the sampling time to
0 second.
@ Manual mode, can also select the
1 to 99 position ( Location ) no.
Memory Card SD memory card. 1 GB to 16 GB.
Advanced * Set clock time ( Year/Month/Date,Hour/Minute/ Second )
setting * Set sampling time
* Auto power OFF management
* Set beep Sound ON/OFF
* Decimal point of SD card setting
* SD memory card Format
* Frequency weighting to A or C setting
Over Indication Show " - - - - ".
Data Hold Freeze the display reading.
Memory Recall Maximum & Minimum value.
Sampling Time Approx. 1 second.
of Display
Data Output RS 232/USB PC computer interface.
* Connect the optional RS232 cable
UPCB-02 will get the RS232 plug.
* Connect the optional USB cable
USB-01 will get the USB plug.
AC output AC 0.5 Vrms corresponding to each
range step.
* Output impedance : 600 ohm.
4
Power off Auto shut off saves battery life or
manual off by push button.
Operating 0 to 50 ℃.
Temperature
Operating Less than 85% R.H.
Humidity
Power Supply * Alkaline
AC/DC power
or heavy
adapter
duty DCis optional.
1.5 V battery
( UM3, AA ) x 6 PCs, or equivalent.
* DC
AC/DC
9V adapter
power adapter
input. ( is
AC/DC
optional.
power
adapter is optional ).
Power Current Normal operation ( w/o SD card save
data and LCD Backlight is OFF) :
Approx. DC 8.2 mA.
When SD card save the data but and
LCD Backlight is OFF) :
Approx. DC 34 mA.
* If
AC/DC
LCD backlight
power adapter
on, theispower
optional.
consumption will increase approx.
6 mA.
Weight 323 g/0.71 LB.
Dimension 245 x 68 x 45 mm.
( 9.6 x 2.7x 1.8 inch ).
Accessories * Instruction manual......................................... 1 PC
Included * Hard carrying case, CA-06............................ 1 PC
Optional * Sound calibrator ( 94 dB ), SC-941.
Accessories * Sound calibrator ( 94/114 dB ), SC-942.
* Sound wind shield ball, SB-01
* SD Card ( 4 GB )
* USB cable, USB-01.
* RS232 cable, UPCB-02.
* Data Acquisition software,
SW-U801-WIN.
* AC to DC 9V adapter.
* Soft carrying case, CA-05A.
5
3. FRONT PANEL DESCRIPTION
Fig. 1
6
4. MEASURING PROCEDURE
4-1 Sound Pressure Level Measurements (SPL)
7
pressing and holding the " ▲ Button "( 3-5, Fig. 1 ) again
range will enter to the manual range ( range 1,range 2 )
and auto range in sequence ,There are still 2 manual ranges.
for your choice :
Note :
a. If select the function of " Fast " time weighting, the
display will show the unit of " FAST".
b. If select the function of " Slow " time weighting, the
display will show the unit of " SLOW ".
c. If select the function of " PEAK FAST " time weighting, the
display will show the unit of " PEAK FAST " and will show
maxmun reading .
2) Select the Leq mode by pressing and holding the " INTEG
TIME button" ( 3-20 Fig 1 )>2 sec. select to the Leq MODEL .
8
3) Frequency weighting select
How to select the frequency weighting ( A or C), the procedures
please refer to chapter 7-7 page 21 .
Remark:
if want to do Leq data logger , Press" Enter button"( 3-8 Fig . 1 ) > 1.5
seconds , into data logger function , then to execute the following actions.
9
6) When Integrating Time set to zero ,push " buttom "
( 3-19 Fig . 1) once, meter to begin integral measurement,At the
same time meter will Show " ► " Symbol at the display The run
time counter will zero and begin counting While running the
meter will Show " ► "Symbol at the display.
When the “ button"( 3-19 Fig . 1 ) is pressed again,
the measurement and Integrating time counter will automatically
stop , and the display will show and hold Integrating
reading and show " ■ " Symbol, at the same time " ► "
Symbol Disappear.
During the measurement, press the " Hold Button " ( 3-3,
Fig. 1 ) once will hold the measured value & the LCD will
display a " HOLD " symbol.
Press the " Hold Button " once again will release the data
hold function.
10
b) Press the " REC Button " ( 3-4, Fig. 1 ) again, the
" REC MIN " symbol along with the minimum value
will appear on the display.
11
5. DATALOGGER
5-1 Preparation before execute datalogger function
a. Insert the SD card
Prepare a " SD memory card " ( 1 GB to 16 GB, optional ),
insert the SD card into the " SD card socket " ( 3-18, Fig. 1).
The front panel of the SD card should face against the
the down case.
b. SD card Format
If SD card just the first time use into the meter, it
recommend to make the " SD card Format " at first. ,
please refer chapter 7-6 ( page 21 ).
*If the SD card exist the trouble during format by the meter,
use the computer to reformat again can fix the problem.
c. Time setting
If the meter is used at first time, it should to adjust the
clock time exactly, please refer chapter 7-1 ( page 18 ).
d. Decimal format setting
The numerical data structure of SD card is
default used the " . " as the decimal, for
example "20.6" "1000.53" . But in certain
countries ( Europe ...) is used the " , " as the
decimal point, for example " 20, 6 "
"1000,53". Under such situation, it should
change the Decimal character at first, details
of setting the Decimal point, refer to Chapter
7-5, page 20.
Press the " ENTER Button ( 3-8, Fig. 1 ) over 2 seconds into
datalogger mode , The LCD will show the text "LOGGER"
and flashing , at the same time the measuring data along
the time information will be saved into the memory circuit.
12
Remark :
How to set the sampling time, refer to Chapter 7-2,
page 19.
Remark :
If press the " ENTER Button " ( 3-8, Fig. 1 ) once again
will execute the Datalogger again, the text of " LOGGER "
will flashing .
During pause the Datalogger, press the " ENTER Button "
( 3-8, Fig. 1) continuously at least two seconds, the " LOGGER "
indicator will be disappeared and finish the Datalogger.
Remark :
When the battery is under the low battery condition
( show the low battery indicator ), the Datalogger
function is disable.
13
5-3 Manual Datalogger ( Set sampling time = 0
second )
Press the " ENTER Button ( 3-4, Fig. 1 ) at least 1.5 seconds , the LCD will
show the text " LOGGER and P = no ", then press the " ENTER Button "
( 3-8, Fig. 1 ) once, the " LOGGER " will flashing once, at the
same time the measuring data along the time information,
will be saved into the memory circuit.
Remark :
* Upper Display will show the Position/Location no.
( P = 1 - 99 ) .
* During execute the Manual Datalogger,
press the " ▲ Button " ( 3-5, Fig. 1) or
" ▼ Button " ( 3-6, Fig. 1 ) to set the measuring
position ( 1 to 99, for example room 1 to room 99 )
to identify the measurement location.
14
5-5 SD Card Data structure
1) When the SD card is used into the meter, the SD card
When the first time, the SD card is used into the meter,
the SD card will generate a folder :
SLC01
15
6. Saving data from the SD card
to the computer ( EXCEL software )
16
EXCEL graphic screen ( for example )
17
7. ADVANCED SETTING
Under do not execute the Datalogger function,
press the " TIME Button " ( 3-7, Fig. 1 ) continuously at
least 1.5 seconds will enter the " Advanced Setting " mode.
then press the "TIME Button " (3-7, Fig. 1 ) once a while
in sequence to select the seven main function, the lower
display will show :
1) Use the " ▲ Button " ( 3-5, Fig. 1 ) or " ▼ Button "
( 3-6, Fig. 1 ) to adjust the value ( Setting start from
Year value ). After the desired value is set, press the
" Enter Button " ( 3-8, Fig. 1 ) once will going to
next value adjustment ( for example, first setting
value is Year then next to adjust Month, Date, Hour,
Minute, Second value ).
18
2) After set all the time value ( Year, Month, Date, Hour,
Minute, Second ),
the screen will jump to " Sampling time "
setting screen ( Chapter 7-2 ).
Remark :
After the time value is setting, the internal clock will
run precisely even Power is off ( The battery is under
normal condition, no low battery condition ).
1) Use the " ▲ Button " ( 3-5, Fig. 1 ) or " ▼ Button " (
3-6, Fig. 1 ) to adjust the value ( 0~3600 sec.)
Remark :
If select the sampling time to " 0 second ", it is ready
for manual Datalogger.
1) Use the " ▲ Button " ( 3-5, Fig. 1 ) or " ▼ Button "
( 3-6, Fig. 1 ) to select the upper text to " YES " or
" NO ".
2) After select the upper text to " YES " or " NO ", press the
" ENTER Button " ( 3-8, Fig. 1 ) will save the setting
function with default.
19
7-4 Set beeper sound ON/OFF
1) Use the " ▲ Button " ( 3-5, Fig. 1 ) or " ▼ Button "
( 3-6, Fig. 1 ) to select the upper text to " YES " or
" NO ".
2) After select the upper text to " YES " or " NO ", press the
" ENTER Button " ( 3-8, Fig. 1 ) will save the setting
function with default.
Remark :
After execute the datalogger function, the buzzer sound
will off automatically to prevent any interference of
the measurement.
1) Use the " ▲ Button " ( 3-5, Fig. 1 ) or " ▼ Button "
( 3-6, Fig. 1 ) to select the upper value to " USA " or
" EURO ".
20
USA - Use " . " as the Decimal point with default.
EURO - Use " , " as the Decimal point with default.
2) After select the upper text to " USA " or " EURO ",
press the " ENTER Button " ( 3-8, Fig. 1 ) will save the
setting function with default.
1) Use the " ▲ Button " ( 3-5, Fig. 1 ) or " ▼ Button " (
3-6, Fig. 1 ) to select the upper text to "YES " or
" NO ".
2) If select the upper to " YES ", press the " ENTER Button
" ( 3-8, Fig. 1 ) once again, the Display will show text
" YES ENTER " to confirm again, if make sure to do the
SD memory card format, then press " Enter Button "
once will format the SD memory clear all the existing
data that already saving into the SD card.
1) Use the " ▲ Button " ( 3-5, Fig. 1 ) or " ▼ Button " (
3-6, Fig. 1 ) to select the lower text to "A " or
" C ".
21
2) After select the lower text to " A " or " C ",
press the " ENTER Button " ( 3-8, Fig. 1 ) will save the
setting function with default.
Note :
a. The characteristic table of A, C weighting, please
ref. page 25.
b. The characteristic of A weighting is simulated as
the " Human Ear Listening " response.
Typically always select the A weighting when makes
environmental sound level measurement.
c. The C weighting characteristic is near the " FLAT "
response. Typically it is suitable for checking the
noise of machinery ( Q.C. check ) & knowing the
real sound level of the tested equipment.
22
9. BATTERY REPLACEMENT
1) When the left corner of LCD display show " ", it
is necessary to replace the battery. However, in-spec.
measurement may still be made for several hours after
low battery indicator appears before the instrument
become inaccurate.
2) Loose the " Battery Cover Screws " ( 3-15, Fig. 1 )
and take away the " Battery Cover " ( 3-17, Fig. 1 )
from the instrument and remove the battery.
3) Replace with DC 1.5 V battery ( UM3, AA,
Alkaline/heavy duty ) x 6 PCs, and reinstate the cover.
4) Make sure the battery cover is secured after changing
the battery.
Meter PC
(9W 'D" Connector)
Center Pin............................................
Pin 4
(3.5 mm jack plug)
Ground/shield...........................................
Pin 2
2.2 K
resistor
Pin 5
23
The 16 digits data stream will be displayed in the
following format :
24
11. CALIBRATION
1) Prepare the optional " SOUND CALIBRATOR ", such as
" SC-941 " or " SC-942 " ( set range to 94.0 dB ). Power
on the Sound Calibrator & plug calibrator output socket
into the " Microphone " head ( 3-9, Fig. 1 ) of the
Sound Level meter.
2) Select manual range to "80- 130 dB ".
3) Select " Time Weighting " at " Fast " position.
4) Select " A " weighting.
5) Adjust the " Calibration VR " ( 3-11, Fig. 1 ) carefully
with a " - " screw driver until the display reading value
within " 94 +/- 0.2 " dB.
25
13. TIME WEIGHTING ( FAST / SLOW )
CHARACTERISTICS
` Max. response Tolerance
Charac. ref. continuous ( IEC 61672 Class 2 )
signal
F ( Fast ) - 1.0 dB + 1.0 dB , - 2.0 dB
S ( Slow ) - 4.1 dB ±2.0 dB
14. PATENT