0% found this document useful (0 votes)
36 views17 pages

HGY2028型贴片电导率仪说明书 zh-CN en

The HGY2028 online conductivity meter is a highly intelligent device designed for continuous monitoring of conductivity in various water types, featuring a titanium electrode for durability and accuracy. It includes a large LCD display with a Chinese menu for easy operation, automatic range conversion, and temperature compensation, along with data storage capabilities. The document provides detailed instructions on installation, operation, maintenance, and technical specifications of the instrument.

Uploaded by

Hassan Jatoi
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
36 views17 pages

HGY2028型贴片电导率仪说明书 zh-CN en

The HGY2028 online conductivity meter is a highly intelligent device designed for continuous monitoring of conductivity in various water types, featuring a titanium electrode for durability and accuracy. It includes a large LCD display with a Chinese menu for easy operation, automatic range conversion, and temperature compensation, along with data storage capabilities. The document provides detailed instructions on installation, operation, maintenance, and technical specifications of the instrument.

Uploaded by

Hassan Jatoi
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 17

Translated from Chinese (Simplified) to English - www.onlinedoctranslator.

com

--

------------------------------------- ---------------------------------------

HGY2028 online conductivity meter

make

use

explain

bright

Book

(new version 3.0)


Nuclear Industry Beijing Research Institute of Chemical Industry and Metallurgy

Beijing Borisai Technology Co., Ltd.

Headrecord
one,Overview……………………………………………………………………………………..1

two,How the instrument works………………………………………………………………………..1

three,Technical specifications of the instrument………………………………………………………………………..2

Four,Installation and use of instruments………………………………………………………………..3

1.Unboxing……………………………………………………………………………………3

2.Installation of secondary instruments……………………………………………………………………3

3.Sensor installation……………………………………………………………………5

4.Connection between sensor and secondary meter……………………………………………………………………5

5.Power on………………………………………………………………………………………………6

five,Detailed explanation of instrument operation……………………………………………………………………………………..7

1.Display and function keys……………………………………………………………………7

2.Menu operation…………………………………………………………………………… 8

(1) historical data………………………………………………………………….8

(2) hardware test………………………………………………………………….9

(3) Parameter settings………………………………………………………………….10

six,Analog calibration of instruments……………………………………………………………………………………..13

seven,Instrument maintenance……………………………………………………………………………………..14

eight,Completeness……………………………………………………………………………………..14

Nine,Ordering and after-sales service…………………………………………………………………15


HGY2028Model Online Conductivity Meter Instructions

I. Overview
HGY2028The online conductivity meter is a highly intelligent online continuous monitor. The complete set of instruments consists of electrodes, measuring cells

It consists of a secondary instrument and is connected with a shielded cable. It can be widely used in boiler feed water, steam condensate water and pure water conductivity.

Continuous monitoring of rate values.

The main features of this instrument are as follows: the electrode part is made of titanium material, which is corrosion-resistant and has a long service life, ensuring measurement

The value is accurate and reliable; the secondary instrument adopts a large-screen LCD display, which uses a Chinese menu to guide operations and simultaneously displays on the screen.

Display conductivity value, output current, temperature, time and instrument status.

The instrument functions are as follows:

1.Fully intelligent: microprocessor is used to complete conductivity measurement and temperature compensation, without functional switches and adjustment knobs

2.The instrument adopts a large-screen LCD display, which uses a Chinese menu to guide operations, and displays the measured conductivity value,

Parameters such as temperature, output current, time and instrument status.

3.Automatic range conversion: Automatic range conversion is achieved within the measurement range covered by the electrode.

4.It has automatic timing calibration function to ensure accurate and stable meter measurement.

5.The instrument has upper and lower limit alarm output functions and4~20mAIsolated current output, and its output lower limit and output full

The degree can be set arbitrarily within the measuring range of the instrument.

6.Automatic temperature compensation: in0~60Automatic temperature compensation within the ℃ range, for (ultra)pure water in power plants

displayed25The conductivity value is automatically temperature compensated at ℃.

7.Historical data query: If the secondary instrument is pressed every10Measurement data is automatically stored once every minute and can be stored continuously for a period of time.

Conductivity data for more than 10 years. In the data query state, the measured conductivity value at a specific point and at a specific time is obtained.

8.Hardware test function: Through this function, you can test whether the upper and lower limit alarm relays operate normally.4-20mAseparated

Check whether the current is correct, and by setting the current output, the manual current source function can be realized to facilitate detection and recording.

instruments and automatic control devices.

9.Backlight function: can be used in dimly lit or completely dark environments.

10.Anti-program crash function: ensures that the instrument will not crash.

2. Working principle of the instrument

1.Measurement principle: In order to avoid electrode polarization, the instrument uses a highly stable square wave voltage signal to be added to the electrode and flow through

The current of the electrode is proportional to the conductivity of the solution being measured. The secondary meter converts the current from the high impedance op amp to AC

After compressing the signal, a DC voltage signal reflecting the conductivity value is obtained after program-controlled signal amplification, full-wave rectification and filtering, and microprocessing

The device switches through a program-controlled electronic switch, alternately samples the temperature signal and the conductivity signal, and obtains it after calculation and temperature compensation.

The solution being tested is in25The conductivity value at ℃ and the temperature value of the solution at that time.

1
HGY2028Model Online Conductivity Meter Instructions

2.Temperature compensation principle: The conductivity of electrolyte solution is greatly affected by changes in solution temperature, and temperature compensation must be performed.

The temperature coefficient for a weak electrolyte solution is2.00%℃, the greater the concentration, the smaller the temperature coefficient. For (ultra)pure water temperature

Coefficients are generally not used2.00%℃, while by the user in0.00~9.99%Any setting within the range. For specific setting methods, see "Parameters

Set the temperature coefficient in "Data Setting".

3.Instrument measurement circuit block diagram:

Conductivity
electricity source LCD Monitor
sensor

4-20mA
Analog measurement A/D Microcontroller D/A V /I
amplifying circuit Convert system Convert Change Change

temperature
superior Alarm contact
Control circuit Drive circuit
sensor

operate

keyboard

three,Technical specifications of the instrument

1.Measuring range: Conductivity:0.002~30.00μS/cm (match0.01electrode), automatic range conversion;

0.020~300.0μS/cm (match0.1electrode), automatic range conversion;

0.200~3000μS/cm(match1.0electrode), automatic range conversion;

temperature:0~60.0℃

2.Basic error: ±1FS; Temperature: ±0.5℃

3.Measured water sample temperature:5℃~50℃; Automatic temperature compensation range:0~60℃ (in25℃ as the basis)

4.Measured water sample flow rate:50~400ml/min; Measured water sample pressure: no greater than0.5MPa

5.Isolated output:4~20 mA (Load resistance≤500Ω) Communication Interface:RS 485;

6.Number of data stores:1months (1point/5minute) ; Data storage time after continuous power failure: greater than6Year

7.power supply:AC 220V±22V,50Hz±1 Hz Power consumption:10vA

8.Upper and lower limit alarm contact capacity:3A240VAC, 6A28VDC

9.Secondary instrument dimensions:145(Length) ×145(width) ×150(deep)mm;

2
HGY2028Model Online Conductivity Meter Instructions

10.Secondary instrument installation method: panel-mounted or wall-mounted. Instrument weight:1.2kg

Panel mounting opening dimensions:138×138mm;

Wall-mounted installation can be installed with the optional mounting bracket produced by our company.

11.Water inlet and outlet pipe connection: hard pipe or hose, pipe diameter ø6(Inside)xø8(outside)mm.

12.Working conditions:a)Ambient temperature:15℃~35℃;

b)Environmental relative humidity: no greater than85%;

c)No vibration, no corrosive gas, no direct sunlight.

4. Installation and use of instruments

HGY2028The online conductivity meter consists of a sensor and a secondary instrument, which are connected with a special shielded cable. two

The secondary instrument adopts an internationally accepted standard chassis, and the required connections are led out from the terminal blocks. The sensor consists of a conductivity electrode and a flow-through electrode

The conductive cell is composed of0.01,0.10and1.00Three options are available, and users can choose according to the measurement range of the medium being measured.

1.Unboxing

After unpacking the instrument, you should check the contents of the box with the packing list. If there is any shortage, please contact the manufacturer in time.

In this case, you can wait for the manufacturer to unpack it during installation and debugging. If there is obvious damage please notify the shipping department and manufacturer immediately.

2.Installation of secondary instruments

The secondary instrument is an internationally accepted standard chassis with a fully enclosed plastic shell structure. The instrument can be installed on a panel.

Optional or wall-mounted. Its appearance and dimensions are shown in the figure.1:

Figure 1 Dimensional drawing of secondary table

3
HGY2028Model Online Conductivity Meter Instructions

(1) Dashboard-mounted installation

When installing the secondary dial, the dimensions of the instrument panel opening are shown in the figure.2, use the special fixing clip provided when installing the upper plate.

Hold the meter and turn the screws that tighten the fixing card respectively. The fixing card will gradually tighten the dial until it is satisfied.

Figure 2 Dimensional drawing of secondary dial-mounted installation openings (unitmm)

(2) Installation of instrument wall-mounted

When the secondary instrument is wall-mounted, you can choose a special mounting bracket produced by our company for installation. First, bring the random

Connect the special fixing card to the installation bracket, then use the fixing card to clamp the instrument and fix it, and finally fix the installation bracket to the wall.

Can. Its wall-mounted installation instructions3:

Installation hole center distance

4
HGY2028Model Online Conductivity Meter Instructions

Figure 3 Schematic diagram of secondary meter wall-mounted installation (unitmm)

The closer the distance between the secondary meter and the sensor is, the better. Generally speaking, it should not exceed10meters to facilitate plumbing and electrical connections;

There should be a certain space around the instrument to facilitate the operation and maintenance of the instrument; it is best to fix the secondary meter at the best line of sight and

Install the instrument in a clean, dry, well-ventilated, and vibration-free place, and there should be no corrosive gas or steam around it. Electrode and two

Do not lay the connecting cable of the sub-meter in close parallel with the power cord to avoid adverse effects on the signal. The meter is a small signal

Precision instruments for measurement must be well grounded. The power cord must be a three-wire system. The ground terminal must be reliably grounded, otherwise

May cause signal instability.

3.Sensor installation

The sensor is composed of conductivity electrodes (the cell constants are:0.01,0.1and1.0)and measuring pool. The measurement cell uses flow

The general structure is made of stainless steel, which can shield harmful interference. Its appearance structure and installation method are shown in the figure.4.

Figure 4: Sensor appearance structure and installation dimensions

picture4The lower section is the water inlet, the upper section is the water outlet, and the outer diameter of the inlet and outlet pipe is φ8, the available outer diameter is φ8hose, too

Available outer diameter is φ8Hard pipe connection to meet the requirements of different users. The sensor is very easy to install, just connect the flow cell to

Just use nuts to fix it on the dial (stand) at the water outlet.

In addition, according to user needs, the sensor can also be made into a recessed installation method.

4.Connection between sensor and secondary meter

The sensor and the secondary instrument are connected with a special cable. The length of the cable before leaving the factory is generally10rice. Special use

The length of the cable can be lengthened appropriately (the user can notify the manufacturer in advance, the maximum length is100rice) . However, during on-site installation, the electricity

The length of the cable should be as short as possible to reduce external interference caused by excessive cable length. There is a label on the terminal of the cable

Just insert the small clip of the secondary instrument into the corresponding terminal of the secondary instrument and tighten the fastening screw. Terminals inside the instrument

5
HGY2028Model Online Conductivity Meter Instructions

See the picture5:

type conductivity meter

switching power supply

+5V
land
upper limit 1

Lower limit 1

+12V
upper limit 2

Alarm Output
- 12V
Lower limit 2

upper limit 3

AC220V50Hz
Lower limit 3

conductivity electrode Connect to switching power supply


L

land
!
land
land

Warming (+)
Electrode (inner)

Warming (-)

+12V
Electrode (outside)

- 12

+5

Digital output Analog output

-+ +-+-
4-20 4-20
one two

Figure 5 Wiring terminal diagram inside the instrument

5.Power on

After the installation is completed according to the above installation method, turn on the power, and the instrument will automatically display the following main menu:

historical data

hardware test

Parameter settings

About10~60The meter automatically enters the measurement state after seconds.

When the meter is powered on and measuring, if the LCD display is not clear, you can open the front panel of the meter and adjust the circuit board appropriately.

Upper left corner potentiometerRP2(10K), just make the LCD display clear;

In addition, before putting the instrument into production operation, please set the "Parameter settings"Parameters under ", such as: according to the power distribution

Set the “electrode constant” of the instrument; set the “temperature coefficient” of the instrument’s temperature compensation (Factory setting is2.00%, users can root

Modify this parameter according to different water quality); set "alarm upper (lower) limit", "Output full scale", "output lower limit" and other parameters. Tool

6
HGY2028Model Online Conductivity Meter Instructions

For body setting method, please refer to "Parameter settings"item.

5. Detailed explanation of instrument operation

1.Display and function keys

The schematic diagram of the display and panel layout after the instrument enters the measurement state is shown in the figure.6:

EM HGY2028 online conductivity meter

2003-06-06 18:20:36

1.000us/cm
normal

23.0C 4.8mA

ESC
OK

Figure 6: Schematic diagram of instrument display and panel layout

The panel function keys are defined as follows:

, keys are direction keys. Press them to move the cursor up or down in the menu item to the item you want to select.

on the project. The selected item is displayed in reverse video; when entering data modification, , are used to increase or decrease the value respectively.

The key is a digit selection key. When modifying data, it is used to select the digit of the value that needs to be modified. Under the selected digit there will be

The cursor flashes.

" ESC" key is used to exit the current state. In the measurement state, press " ESC” key will enter the main menu. Enter
After any level menu, press " ESC” key will return to the previous menu. Pressing it continuously will return to the measurement state.
OK
The " " key can be used to enter the selected menu, or to confirm the setting value after the parameter setting is completed.

After entering the measurement state, the meter displays as follows: the conductivity value is the main display, and the dimensions are:16×24mm;Temperature, lose

Output current, year, month, day and hour are secondary displays, the dimensions are:8×8mm;The status prompt line display size is:16×16mm,

The following information can be prompted: ① Normal ② Upper limit alarm ③ Lower limit alarm ④ Over-range (Note: The meter displays when prompted for over-range

The indication value has lost its meaning)

7
HGY2028Model Online Conductivity Meter Instructions

In the measurement state, press " ESC” key to enter the main menu.

2.Menu operation

HGY2000The series instruments have many functions. Due to the use of a classified menu structure, the function keys on the panel are basically

There is no multiplicity in this book, and the operation is relatively simple. You can follow the prompts on the screen.

ESC
After pressing the " " key in the measurement state, the main menu will be entered, as shown below:

historical data

hardware test

Parameter settings

use , key to move the cursor to select an item (the selected item is displayed in reverse video) and press " OK ” key to enter the
corresponding submenu. according to" ESC” key to return to the measurement state.
The menu structure has multiple layers. You can enter and exit layer by layer. After entering a certain level of submenu, each time you press " ESC"
key to return to the previous menu until returning to the measurement state.

Once the meter exits the measurement state and enters the menu, the internal scheduled timer will start and automatically record the time of exiting the measurement.

length of time, exceeding5It will automatically return to the measurement state after a few minutes to avoid inadvertent operation and failure after menu operation.

There is a complete return to the measurement state, resulting in the inability to measure for a long time.

(1)historical data

Select "Historical Data" in the main menu and press " OK After pressing the ” key, the screen displays:

Query data

Select "Query Data" and press " OK ” button, the screen displays the recently recorded10data, the recording time interval
for5Minutes/times, displayed as follows:

06-05-28 09:15 1.682


06-05-28 09:20 1.682
06-05-28 09:25 1.682
06-05-28 09:30 1.684
06-05-28 09:35 1.682
06-05-28 09:40 1.682
06-05-28 09:45 1.682
06-05-28 09:50 1.682
06-05-28 09:55 1.682
06-05-28 10:00 1.682
8
HGY2028Model Online Conductivity Meter Instructions

Continue to press " OK ” key, the screen will prompt you to enter the query date, as shown below:

expiration date06moon05day

Inquiry date:06moon05day

use , , key to modify the date to be queried, and then press " OK ” key, the screen will display the corresponding date
data within. use , key to flip the screen to display the data of the upper and lower time periods.

In the Query Data menu, press OK "key or enter the query date and press " OK ” button, if the query date is within

No data, the screen will display:

no data

Continue to press " OK " key, the screen will prompt you to enter the query date again and press " OK ” key to query the newly entered date
data, press " ESC” key will return to the previous menu.
instrument per5Automatically stores measurement data once every minute, which can store approximately300days of data, the previous data will be

The movement is covered.

(2)hardware test

Select "Hardware Test" under the main menu and press " OK ” key, the screen displays:

Upper limit alarm close

Lower limit alarm close

Current output 04mA

Test voltage

use , key to select an item and press " OK ” button, you will enter the corresponding submenu.
Select "Upper Limit Alarm" or "Lower Limit Alarm" and press " OK ” key, the corresponding alarm status is selected (displayed in reverse),

9
HGY2028Model Online Conductivity Meter Instructions

use or The key can change the switch status of the upper (lower) limit alarm, and the corresponding relay is activated or released. Press " ESC"
Press the key to cancel the upper (lower) limit alarm test function.

OK
After selecting "Current Output" and pressing the " " key, the test "Current Output" function is selected (the output current value is displayed in reverse) ,
use The or key can change the output current value so that the instrument has a manual current source function. designed this function

The purpose is: first, to allow the user to check the accuracy of the instrument’s output current within the full output range; second, to communicate with the recorder or slave computer

When connected, you can check whether the recorder and slave computer are sampling correctly to identify responsibilities, and even manually control subsequent adjustments.

device. PressESC
the " " key to exit the "current output" test function.

Select "Test Voltage" and pressOK


the " " key, use or button can test the reference voltage signal, temperature

degree voltage signal, conductivity analog voltage signal (×1, ×10, ×100 three levels). This function is particularly useful during instrument simulation calibration.

ESC
Don't use it. Press the " " key to exit the test voltage function.

(3) Parameter setting

Select "Parameter Settings" under the main menu and press " OK After pressing the ” key, the screen displays:

Please enter password

9999

In order to prevent misoperation by unrelated personnel, operations that affect the operation of the instrument, such as parameter modification and other functions,HGYSeries instrument

Tables are protected by passwords, and users can only enter after checking them; operations that have no impact on system operation can be

There is no password verification.

use, and After entering the password, press " OK ” button, if the password is correct, go to the next step.
Otherwise, "Wrong password" will be prompted. At this time, continue to press " OK "key will return to the "Please enter password" state.

2028
The factory password of this instrument is: .HGYThe password of the series instrument is the same as the number of the table. If the user feels it is necessary

Change the password and modify the old password. For specific modification methods, see the "Password Modification" item in "Parameter Settings".

OK
Press factory password2028After completing the input, press the " " key, the screen will display:

Alarm upper limit 020.0

Alarm lower limit 000.0

Output full scale 020.0

Output lower limit 000.0

Cell constant 0.0100

10
HGY2028Model Online Conductivity Meter Instructions

use or You can select the corresponding item. When the highlighted cursor is on the last item, continue pressing key, the screen will

The next screen of "Parameter Settings" is displayed.

Temperature Coefficient 0.000%


time correction

temperature setting automatic

manual temperature 25.0


dynamic compensation ×1.000

Use the same or You can select the corresponding item. When the highlighted cursor is on the last item, continue pressing key, screen

The next screen of "Parameter Settings" will be displayed.

change Password

In the above three-screen "Parameter Setting" menu, use , key to select the corresponding entry. according to" OK” key, that is
You can enter the modification and settings of the corresponding items.

Select "Parameter Settings" in the menuAlarm upper limit" (or"Alarm lower limit","Enter fullness","Output lower limit")

OK
Submenu. After pressing the " " key, it will enter the modification of the corresponding parameters. The first digit of this parameter starts to flash. Use the keys to select

Select the modification bit of the parameter (the selection bit cursor flashes), use the and keys to modify the selection bit, and the modification is completed. according to" " OK
key to complete the setting of this parameter. (This parameter needs to be set to be valid within the measurement range of the instrument)

OK
choose"Cell constant”, after pressing the “ ” key, it will enter the “Cell Constant” modification. Press the mark on the measuring cell shell.

The electrode constant, use the keys to select the modification bit (the cursor of the selection bit flashes), use the and keys to modify the selection bit, and follow the measurement

OK
After inputting the calibrated electrode constant on the measuring cell, continue to press the " " key to complete the modification of "electrode constant".

OK
choose"Temperature Coefficient", after pressing the " " key, it will enter the modification of "temperature coefficient", use the key to select modification

bit (select the bit cursor flashes), use , key to modify the selection bit. After inputting the temperature coefficient, continue to press the " " key. OK
That is, the modification of the temperature coefficient is completed. When the "temperature setting" of the instrument is automatic, the instrument will automatically adjust according to the entered temperature coefficient.

Dynamic temperature compensation. When the instrument leaves the factory, the temperature coefficient is set to2.000%. As we all know, the temperature compensation of pure water and ultrapure water

The problem is very complex and closely related to the properties and concentration of conductive ions in the water. so,HGY2028type

The temperature coefficient of the online conductivity meter can be determined by the user0.000-9.999set between %. We recommend: Conductivity in water samples

rate less than1μs/cmWhen, the temperature coefficient selection4.55%; less than10μs/cmwhen, select4%; less than100μs/cmhour,

choose3%;Exceed100μs/cmwhen, select2.10%, higher conductivity depends on the temperature coefficient of the main ions in the water

11
HGY2028Model Online Conductivity Meter Instructions

Make settings.

choose"time correction",according to" OK "After pressing the key, the year, month and day correction will be entered. Press Select correction bit year, month,

Day, use the and keys to modify the selected digit. After modifying the year, month and day, continue to press " OK ” key to enter hour and minute repair
OK
Yes, the correction method is the same as the year, month and day correction. After the correction is completed, continue to press the " " key to complete the time correction.

choose"temperature setting”, after pressing the “Confirm” key, "Temperature setting" status is displayed in reverse, use or Keys are interchangeable

Change the temperature setting status to "Auto" or "Manual". When "Auto" is selected, press " OK ”, that is, the completion temperature

Set to automatic state, at this time the instrument will measure according to automatic temperature compensation; when selected to "manual" state, press " OK "
After pressing the key, the temperature is set to manual state. At this time, the instrument will measure according to manual temperature compensation. The purpose is to satisfy the instrument

The instrument can still continue to operate when the temperature measurement circuit fails. After the temperature is set to manual state, a message will appear behind the temperature display.

A "hand" symbol.(Under normal circumstances, the meter temperature is set to automatic state)

OK
choose"manual temperature", after pressing the " " key, it will enter the "manual temperature" modification. The first digit of the parameter starts to flash.

to move, use the keys to select the modification bit of the parameter (the selection bit cursor flashes), use the and keys to modify the selection bit, modify

OK
complete. Press the " " key to complete the "Manual Temperature" setting.

OK
choose"dynamic compensation”, after pressing the “ ” key, it will enter the setting of dynamic compensation value. Use Key selection parameter modification

Change the position (the cursor of the selected position flashes). Use the and keys to modify the selected position. After modification, press " OK "Press to complete" action

"Status Compensation" setting. (When there is a systematic deviation between the instrument measurement value and the standard instrument measurement value, you can modify the "dynamic compensation"

parameters to eliminate the deviation of the instrument. If the meter display is too large, set "dynamic compensation" to less than 1; if the meter display

"Dynamic
On the small side, compensation" is set to greater than 1; if the instrument displays no error, "Dynamic compensation" is set equal to 1, usually

The lower "Dynamic Compensation" value is set equal to 1)

OK
choose"change Password”, after pressing the “ ” key, it will enter the “Please enter the password9999", the password is four digits, use

Press to select the modified bit (the cursor of the selected bit flashes). Use the and keys to modify the selected bits. After entering the new password, press " OK"
After pressing the key, the password modification is completed. Next time you enter "Parameter Settings", you need to enter the newly set password before entering. If you use

If you have forgotten your newly set password, please contact us.

12
HGY2028Model Online Conductivity Meter Instructions

6. Simulation calibration of instruments

If the user wants to judge whether the secondary instrument is good or bad, he can remove the connecting cable between the secondary instrument and the sensor and connect it to the conductivity electrode.

Connect resistance boxes to the temperature compensation terminals for simulation verification.

First, set the "Temperature Coefficient" under the "Parameter Setting" menu of the secondary instrument to0.000%, at this time the resistance box at both ends of the electrode

readingRThe relationship with the displayed conductivity is:

S=1000000×K/R
In the formula:R-Resistance box resistance, unit is Ω

S - conductivity value, unit is μs/cm

K - Conductivity cell constant (according to the different measuring range of the instrument, conductivity cells with different cell constants should be selected.

The pole matches the secondary instrument. At this time, the "Cell Constant" in the "Parameter Settings" of the secondary table

The set value should be consistent with the cell constant calibrated on the measuring cell casing. The setting value should be

0.0100, 0.1000 and 1.0000 respectively)

Correspondence table between resistance box and conductivity (The temperature coefficient is0.000%or temperature manually as25℃)

Conductivity (μs/cm) Conductivity (μs/cm) Conductivity (μs/cm)


Resistor box value (Ω) (Cell constant (Cell constant (Cell constant
=0.0100) =0.1000) =1.0000)
333 30.00 300.0 3000

500 20.00 200.0 2000

1000 10.00 100.0 1000

2000 5.000 50.00 500.0

5000 2.000 20.00 200.0

10000 1.000 10.00 100.0

30000 0.333 3.333 33.33

40000 0.250 2.500 25.00

50000 0.200 2.000 20.00

13
HGY2028Model Online Conductivity Meter Instructions

The above table shows the resistance box value and the theoretical conductivity value displayed by the secondary meter. If it is found that the actual conductivity displayed by the secondary meter is different from the

There is a large error in the theoretical value in the above table, which can be judged as a secondary instrument failure.

The instrument adopts two-wire system for temperature measurement.NTC—Negative temperature coefficient thermistor acts as a temperature sensor. Use a resistor box

When simulating the temperature calibration, please connect the resistance box to both ends of the temperature compensation resistor, and set the temperature to the automatic measurement state. At this time, the resistance

The box value and secondary instrument display temperature should comply with the following table:

(
Correspondence table between resistance box and displayedTemperature
temperature is set to automatic state)

7352 4481 2813 2252 1814 1199


Resistor box value
811.4 560.3
(Ω)

0.0 10.0 20.0 25.0 30.0 40.0 50.0 60.0


nominal temperature

(℃)

Error≤±0.5℃ is qualified.

Note: After the simulation calibration is completed, please restore the "parameter settings" of the instrument before putting it into operation, such as "temperature coefficient", "electrode normal

number", "temperature set to automatic", etc.

7. Instrument maintenance

The secondary instrument does not require routine maintenance. If an obvious fault occurs, in principle, do not open the secondary instrument and repair it yourself.

The conductivity cell and conductivity electrode should be cleaned in time if they are found to be contaminated. If there is oil pollution, it is necessary to use a solution dipped in analytical grade tetrachlorine.

Wipe carefully with a cotton ball of charcoal, acetone or absolute ethanol; you can also use 50% warm detergent (or household soap powder) to clean.

Scrub with a nylon brush, and then rinse the inside of the electrode with distilled water to ensure that there is no oil deposit on the inner and outer electrode surfaces. Remember to use your fingers

Touch the conductivity electrode. For dirt deposits with strong adhesion, soak them in about 2% hydrochloric acid solution for 3 to 5 minutes, and then

Wash repeatedly with distilled water until there is no obvious dirt on the electrode surface.

8. Complete set

1. HGY2028 electronic unit......................one set

2. Sensor (flow cell + conductivity electrode) ........................ one

3. Special cable (connected to sensor)............8 meters

4. ф8mm connecting hose........................ 2 meters

5. Installation accessories and fuses......................1 bag

14
HGY2028Model Online Conductivity Meter Instructions

6. Instruction Manual........................One copy

7. Certificate of conformity........................................a copy

8. Packing list........................................a copy

9. Ordering and after-sales service

Since the measuring range of the instrument is different, conductivity electrodes with different electrode constants need to be equipped, so the user needs to provide the instrument measurement when ordering.

Range, indicate conductivity cell constant. If you have other special requirements, please indicate.

The warranty period of this instrument is one year. If the failure is not caused by human factors within one year, our company will be responsible for repairs. Specific contact party

The formula is as follows:

Nuclear Industry Beijing Research Institute of Chemical Industry and Metallurgy

Beijing Borisai Technology Co., Ltd.

Unit address: No. 145, Jiushkeshu, Tongzhou District, Beijing

Correspondence address: Beijing 234 Mailbox Instrument Control Company

Zip code; 101149

Service hotline: 010-51674210

Telephone/Fax: 010-51674497 010-81544820

Email:[email protected]

Website: www.bricem.com.cn

15

You might also like