I. Overview
HGY2028The online conductivity meter is a highly intelligent online continuous monitor. The complete set of instruments consists of electrodes, measuring cells
It consists of a secondary instrument and is connected with a shielded cable. It can be widely used in boiler feed water, steam condensate water and pure water conductivity.
The main features of this instrument are as follows: the electrode part is made of titanium material, which is corrosion-resistant and has a long service life, ensuring measurement
The value is accurate and reliable; the secondary instrument adopts a large-screen LCD display, which uses a Chinese menu to guide operations and simultaneously displays on the screen.
Display conductivity value, output current, temperature, time and instrument status.
1.Fully intelligent: microprocessor is used to complete conductivity measurement and temperature compensation, without functional switches and adjustment knobs
2.The instrument adopts a large-screen LCD display, which uses a Chinese menu to guide operations, and displays the measured conductivity value,
3.Automatic range conversion: Automatic range conversion is achieved within the measurement range covered by the electrode.
4.It has automatic timing calibration function to ensure accurate and stable meter measurement.
5.The instrument has upper and lower limit alarm output functions and4~20mAIsolated current output, and its output lower limit and output full
The degree can be set arbitrarily within the measuring range of the instrument.
6.Automatic temperature compensation: in0~60Automatic temperature compensation within the ℃ range, for (ultra)pure water in power plants
7.Historical data query: If the secondary instrument is pressed every10Measurement data is automatically stored once every minute and can be stored continuously for a period of time.
Conductivity data for more than 10 years. In the data query state, the measured conductivity value at a specific point and at a specific time is obtained.
8.Hardware test function: Through this function, you can test whether the upper and lower limit alarm relays operate normally.4-20mAseparated
Check whether the current is correct, and by setting the current output, the manual current source function can be realized to facilitate detection and recording.
10.Anti-program crash function: ensures that the instrument will not crash.
1.Measurement principle: In order to avoid electrode polarization, the instrument uses a highly stable square wave voltage signal to be added to the electrode and flow through
The current of the electrode is proportional to the conductivity of the solution being measured. The secondary meter converts the current from the high impedance op amp to AC
After compressing the signal, a DC voltage signal reflecting the conductivity value is obtained after program-controlled signal amplification, full-wave rectification and filtering, and microprocessing
The device switches through a program-controlled electronic switch, alternately samples the temperature signal and the conductivity signal, and obtains it after calculation and temperature compensation.
The solution being tested is in25The conductivity value at ℃ and the temperature value of the solution at that time.
1
HGY2028Model Online Conductivity Meter Instructions
2.Temperature compensation principle: The conductivity of electrolyte solution is greatly affected by changes in solution temperature, and temperature compensation must be performed.
The temperature coefficient for a weak electrolyte solution is2.00%℃, the greater the concentration, the smaller the temperature coefficient. For (ultra)pure water temperature
Coefficients are generally not used2.00%℃, while by the user in0.00~9.99%Any setting within the range. For specific setting methods, see "Parameters
Conductivity
electricity source LCD Monitor
sensor
4-20mA
Analog measurement A/D Microcontroller D/A V /I
amplifying circuit Convert system Convert Change Change
temperature
superior Alarm contact
Control circuit Drive circuit
sensor
operate
keyboard
temperature:0~60.0℃
3.Measured water sample temperature:5℃~50℃; Automatic temperature compensation range:0~60℃ (in25℃ as the basis)
4.Measured water sample flow rate:50~400ml/min; Measured water sample pressure: no greater than0.5MPa
6.Number of data stores:1months (1point/5minute) ; Data storage time after continuous power failure: greater than6Year
2
HGY2028Model Online Conductivity Meter Instructions
Wall-mounted installation can be installed with the optional mounting bracket produced by our company.
11.Water inlet and outlet pipe connection: hard pipe or hose, pipe diameter ø6(Inside)xø8(outside)mm.
HGY2028The online conductivity meter consists of a sensor and a secondary instrument, which are connected with a special shielded cable. two
The secondary instrument adopts an internationally accepted standard chassis, and the required connections are led out from the terminal blocks. The sensor consists of a conductivity electrode and a flow-through electrode
The conductive cell is composed of0.01,0.10and1.00Three options are available, and users can choose according to the measurement range of the medium being measured.
1.Unboxing
After unpacking the instrument, you should check the contents of the box with the packing list. If there is any shortage, please contact the manufacturer in time.
In this case, you can wait for the manufacturer to unpack it during installation and debugging. If there is obvious damage please notify the shipping department and manufacturer immediately.
The secondary instrument is an internationally accepted standard chassis with a fully enclosed plastic shell structure. The instrument can be installed on a panel.
Optional or wall-mounted. Its appearance and dimensions are shown in the figure.1:
3
HGY2028Model Online Conductivity Meter Instructions
When installing the secondary dial, the dimensions of the instrument panel opening are shown in the figure.2, use the special fixing clip provided when installing the upper plate.
Hold the meter and turn the screws that tighten the fixing card respectively. The fixing card will gradually tighten the dial until it is satisfied.
When the secondary instrument is wall-mounted, you can choose a special mounting bracket produced by our company for installation. First, bring the random
Connect the special fixing card to the installation bracket, then use the fixing card to clamp the instrument and fix it, and finally fix the installation bracket to the wall.
4
HGY2028Model Online Conductivity Meter Instructions
The closer the distance between the secondary meter and the sensor is, the better. Generally speaking, it should not exceed10meters to facilitate plumbing and electrical connections;
There should be a certain space around the instrument to facilitate the operation and maintenance of the instrument; it is best to fix the secondary meter at the best line of sight and
Install the instrument in a clean, dry, well-ventilated, and vibration-free place, and there should be no corrosive gas or steam around it. Electrode and two
Do not lay the connecting cable of the sub-meter in close parallel with the power cord to avoid adverse effects on the signal. The meter is a small signal
Precision instruments for measurement must be well grounded. The power cord must be a three-wire system. The ground terminal must be reliably grounded, otherwise
3.Sensor installation
The sensor is composed of conductivity electrodes (the cell constants are:0.01,0.1and1.0)and measuring pool. The measurement cell uses flow
The general structure is made of stainless steel, which can shield harmful interference. Its appearance structure and installation method are shown in the figure.4.
picture4The lower section is the water inlet, the upper section is the water outlet, and the outer diameter of the inlet and outlet pipe is φ8, the available outer diameter is φ8hose, too
Available outer diameter is φ8Hard pipe connection to meet the requirements of different users. The sensor is very easy to install, just connect the flow cell to
Just use nuts to fix it on the dial (stand) at the water outlet.
In addition, according to user needs, the sensor can also be made into a recessed installation method.
The sensor and the secondary instrument are connected with a special cable. The length of the cable before leaving the factory is generally10rice. Special use
The length of the cable can be lengthened appropriately (the user can notify the manufacturer in advance, the maximum length is100rice) . However, during on-site installation, the electricity
The length of the cable should be as short as possible to reduce external interference caused by excessive cable length. There is a label on the terminal of the cable
Just insert the small clip of the secondary instrument into the corresponding terminal of the secondary instrument and tighten the fastening screw. Terminals inside the instrument
5
HGY2028Model Online Conductivity Meter Instructions
+5V
land
upper limit 1
Lower limit 1
+12V
upper limit 2
Alarm Output
- 12V
Lower limit 2
upper limit 3
AC220V50Hz
Lower limit 3
land
!
land
land
Warming (+)
Electrode (inner)
Warming (-)
+12V
Electrode (outside)
- 12
+5
-+ +-+-
4-20 4-20
one two
5.Power on
After the installation is completed according to the above installation method, turn on the power, and the instrument will automatically display the following main menu:
historical data
hardware test
Parameter settings
When the meter is powered on and measuring, if the LCD display is not clear, you can open the front panel of the meter and adjust the circuit board appropriately.
Upper left corner potentiometerRP2(10K), just make the LCD display clear;
In addition, before putting the instrument into production operation, please set the "Parameter settings"Parameters under ", such as: according to the power distribution
Set the “electrode constant” of the instrument; set the “temperature coefficient” of the instrument’s temperature compensation (Factory setting is2.00%, users can root
Modify this parameter according to different water quality); set "alarm upper (lower) limit", "Output full scale", "output lower limit" and other parameters. Tool
6
HGY2028Model Online Conductivity Meter Instructions
The schematic diagram of the display and panel layout after the instrument enters the measurement state is shown in the figure.6:
2003-06-06 18:20:36
1.000us/cm
normal
23.0C 4.8mA
ESC
OK
, keys are direction keys. Press them to move the cursor up or down in the menu item to the item you want to select.
on the project. The selected item is displayed in reverse video; when entering data modification, , are used to increase or decrease the value respectively.
The key is a digit selection key. When modifying data, it is used to select the digit of the value that needs to be modified. Under the selected digit there will be
" ESC" key is used to exit the current state. In the measurement state, press " ESC” key will enter the main menu. Enter
After any level menu, press " ESC” key will return to the previous menu. Pressing it continuously will return to the measurement state.
OK
The " " key can be used to enter the selected menu, or to confirm the setting value after the parameter setting is completed.
After entering the measurement state, the meter displays as follows: the conductivity value is the main display, and the dimensions are:16×24mm;Temperature, lose
Output current, year, month, day and hour are secondary displays, the dimensions are:8×8mm;The status prompt line display size is:16×16mm,
The following information can be prompted: ① Normal ② Upper limit alarm ③ Lower limit alarm ④ Over-range (Note: The meter displays when prompted for over-range
7
HGY2028Model Online Conductivity Meter Instructions
In the measurement state, press " ESC” key to enter the main menu.
2.Menu operation
HGY2000The series instruments have many functions. Due to the use of a classified menu structure, the function keys on the panel are basically
There is no multiplicity in this book, and the operation is relatively simple. You can follow the prompts on the screen.
ESC
After pressing the " " key in the measurement state, the main menu will be entered, as shown below:
historical data
hardware test
Parameter settings
use , key to move the cursor to select an item (the selected item is displayed in reverse video) and press " OK ” key to enter the
corresponding submenu. according to" ESC” key to return to the measurement state.
The menu structure has multiple layers. You can enter and exit layer by layer. After entering a certain level of submenu, each time you press " ESC"
key to return to the previous menu until returning to the measurement state.
Once the meter exits the measurement state and enters the menu, the internal scheduled timer will start and automatically record the time of exiting the measurement.
length of time, exceeding5It will automatically return to the measurement state after a few minutes to avoid inadvertent operation and failure after menu operation.
There is a complete return to the measurement state, resulting in the inability to measure for a long time.
(1)historical data
Select "Historical Data" in the main menu and press " OK After pressing the ” key, the screen displays:
Query data
Select "Query Data" and press " OK ” button, the screen displays the recently recorded10data, the recording time interval
for5Minutes/times, displayed as follows:
Continue to press " OK ” key, the screen will prompt you to enter the query date, as shown below:
expiration date06moon05day
Inquiry date:06moon05day
use , , key to modify the date to be queried, and then press " OK ” key, the screen will display the corresponding date
data within. use , key to flip the screen to display the data of the upper and lower time periods.
In the Query Data menu, press OK "key or enter the query date and press " OK ” button, if the query date is within
no data
Continue to press " OK " key, the screen will prompt you to enter the query date again and press " OK ” key to query the newly entered date
data, press " ESC” key will return to the previous menu.
instrument per5Automatically stores measurement data once every minute, which can store approximately300days of data, the previous data will be
(2)hardware test
Select "Hardware Test" under the main menu and press " OK ” key, the screen displays:
Test voltage
use , key to select an item and press " OK ” button, you will enter the corresponding submenu.
Select "Upper Limit Alarm" or "Lower Limit Alarm" and press " OK ” key, the corresponding alarm status is selected (displayed in reverse),
9
HGY2028Model Online Conductivity Meter Instructions
use or The key can change the switch status of the upper (lower) limit alarm, and the corresponding relay is activated or released. Press " ESC"
Press the key to cancel the upper (lower) limit alarm test function.
OK
After selecting "Current Output" and pressing the " " key, the test "Current Output" function is selected (the output current value is displayed in reverse) ,
use The or key can change the output current value so that the instrument has a manual current source function. designed this function
The purpose is: first, to allow the user to check the accuracy of the instrument’s output current within the full output range; second, to communicate with the recorder or slave computer
When connected, you can check whether the recorder and slave computer are sampling correctly to identify responsibilities, and even manually control subsequent adjustments.
device. PressESC
the " " key to exit the "current output" test function.
degree voltage signal, conductivity analog voltage signal (×1, ×10, ×100 three levels). This function is particularly useful during instrument simulation calibration.
ESC
Don't use it. Press the " " key to exit the test voltage function.
Select "Parameter Settings" under the main menu and press " OK After pressing the ” key, the screen displays:
9999
In order to prevent misoperation by unrelated personnel, operations that affect the operation of the instrument, such as parameter modification and other functions,HGYSeries instrument
Tables are protected by passwords, and users can only enter after checking them; operations that have no impact on system operation can be
use, and After entering the password, press " OK ” button, if the password is correct, go to the next step.
Otherwise, "Wrong password" will be prompted. At this time, continue to press " OK "key will return to the "Please enter password" state.
2028
The factory password of this instrument is: .HGYThe password of the series instrument is the same as the number of the table. If the user feels it is necessary
Change the password and modify the old password. For specific modification methods, see the "Password Modification" item in "Parameter Settings".
OK
Press factory password2028After completing the input, press the " " key, the screen will display:
10
HGY2028Model Online Conductivity Meter Instructions
use or You can select the corresponding item. When the highlighted cursor is on the last item, continue pressing key, the screen will
Use the same or You can select the corresponding item. When the highlighted cursor is on the last item, continue pressing key, screen
change Password
In the above three-screen "Parameter Setting" menu, use , key to select the corresponding entry. according to" OK” key, that is
You can enter the modification and settings of the corresponding items.
Select "Parameter Settings" in the menuAlarm upper limit" (or"Alarm lower limit","Enter fullness","Output lower limit")
OK
Submenu. After pressing the " " key, it will enter the modification of the corresponding parameters. The first digit of this parameter starts to flash. Use the keys to select
Select the modification bit of the parameter (the selection bit cursor flashes), use the and keys to modify the selection bit, and the modification is completed. according to" " OK
key to complete the setting of this parameter. (This parameter needs to be set to be valid within the measurement range of the instrument)
OK
choose"Cell constant”, after pressing the “ ” key, it will enter the “Cell Constant” modification. Press the mark on the measuring cell shell.
The electrode constant, use the keys to select the modification bit (the cursor of the selection bit flashes), use the and keys to modify the selection bit, and follow the measurement
OK
After inputting the calibrated electrode constant on the measuring cell, continue to press the " " key to complete the modification of "electrode constant".
OK
choose"Temperature Coefficient", after pressing the " " key, it will enter the modification of "temperature coefficient", use the key to select modification
bit (select the bit cursor flashes), use , key to modify the selection bit. After inputting the temperature coefficient, continue to press the " " key. OK
That is, the modification of the temperature coefficient is completed. When the "temperature setting" of the instrument is automatic, the instrument will automatically adjust according to the entered temperature coefficient.
Dynamic temperature compensation. When the instrument leaves the factory, the temperature coefficient is set to2.000%. As we all know, the temperature compensation of pure water and ultrapure water
The problem is very complex and closely related to the properties and concentration of conductive ions in the water. so,HGY2028type
The temperature coefficient of the online conductivity meter can be determined by the user0.000-9.999set between %. We recommend: Conductivity in water samples
rate less than1μs/cmWhen, the temperature coefficient selection4.55%; less than10μs/cmwhen, select4%; less than100μs/cmhour,
choose3%;Exceed100μs/cmwhen, select2.10%, higher conductivity depends on the temperature coefficient of the main ions in the water
11
HGY2028Model Online Conductivity Meter Instructions
Make settings.
choose"time correction",according to" OK "After pressing the key, the year, month and day correction will be entered. Press Select correction bit year, month,
Day, use the and keys to modify the selected digit. After modifying the year, month and day, continue to press " OK ” key to enter hour and minute repair
OK
Yes, the correction method is the same as the year, month and day correction. After the correction is completed, continue to press the " " key to complete the time correction.
choose"temperature setting”, after pressing the “Confirm” key, "Temperature setting" status is displayed in reverse, use or Keys are interchangeable
Change the temperature setting status to "Auto" or "Manual". When "Auto" is selected, press " OK ”, that is, the completion temperature
Set to automatic state, at this time the instrument will measure according to automatic temperature compensation; when selected to "manual" state, press " OK "
After pressing the key, the temperature is set to manual state. At this time, the instrument will measure according to manual temperature compensation. The purpose is to satisfy the instrument
The instrument can still continue to operate when the temperature measurement circuit fails. After the temperature is set to manual state, a message will appear behind the temperature display.
A "hand" symbol.(Under normal circumstances, the meter temperature is set to automatic state)
OK
choose"manual temperature", after pressing the " " key, it will enter the "manual temperature" modification. The first digit of the parameter starts to flash.
to move, use the keys to select the modification bit of the parameter (the selection bit cursor flashes), use the and keys to modify the selection bit, modify
OK
complete. Press the " " key to complete the "Manual Temperature" setting.
OK
choose"dynamic compensation”, after pressing the “ ” key, it will enter the setting of dynamic compensation value. Use Key selection parameter modification
Change the position (the cursor of the selected position flashes). Use the and keys to modify the selected position. After modification, press " OK "Press to complete" action
"Status Compensation" setting. (When there is a systematic deviation between the instrument measurement value and the standard instrument measurement value, you can modify the "dynamic compensation"
parameters to eliminate the deviation of the instrument. If the meter display is too large, set "dynamic compensation" to less than 1; if the meter display
"Dynamic
On the small side, compensation" is set to greater than 1; if the instrument displays no error, "Dynamic compensation" is set equal to 1, usually
OK
choose"change Password”, after pressing the “ ” key, it will enter the “Please enter the password9999", the password is four digits, use
Press to select the modified bit (the cursor of the selected bit flashes). Use the and keys to modify the selected bits. After entering the new password, press " OK"
After pressing the key, the password modification is completed. Next time you enter "Parameter Settings", you need to enter the newly set password before entering. If you use
If you have forgotten your newly set password, please contact us.
12
HGY2028Model Online Conductivity Meter Instructions
If the user wants to judge whether the secondary instrument is good or bad, he can remove the connecting cable between the secondary instrument and the sensor and connect it to the conductivity electrode.
Connect resistance boxes to the temperature compensation terminals for simulation verification.
First, set the "Temperature Coefficient" under the "Parameter Setting" menu of the secondary instrument to0.000%, at this time the resistance box at both ends of the electrode
S=1000000×K/R
In the formula:R-Resistance box resistance, unit is Ω
K - Conductivity cell constant (according to the different measuring range of the instrument, conductivity cells with different cell constants should be selected.
The pole matches the secondary instrument. At this time, the "Cell Constant" in the "Parameter Settings" of the secondary table
The set value should be consistent with the cell constant calibrated on the measuring cell casing. The setting value should be
Correspondence table between resistance box and conductivity (The temperature coefficient is0.000%or temperature manually as25℃)
13
HGY2028Model Online Conductivity Meter Instructions
The above table shows the resistance box value and the theoretical conductivity value displayed by the secondary meter. If it is found that the actual conductivity displayed by the secondary meter is different from the
There is a large error in the theoretical value in the above table, which can be judged as a secondary instrument failure.
The instrument adopts two-wire system for temperature measurement.NTC—Negative temperature coefficient thermistor acts as a temperature sensor. Use a resistor box
When simulating the temperature calibration, please connect the resistance box to both ends of the temperature compensation resistor, and set the temperature to the automatic measurement state. At this time, the resistance
The box value and secondary instrument display temperature should comply with the following table:
(
Correspondence table between resistance box and displayedTemperature
temperature is set to automatic state)
(℃)
Error≤±0.5℃ is qualified.
Note: After the simulation calibration is completed, please restore the "parameter settings" of the instrument before putting it into operation, such as "temperature coefficient", "electrode normal
7. Instrument maintenance
The secondary instrument does not require routine maintenance. If an obvious fault occurs, in principle, do not open the secondary instrument and repair it yourself.
The conductivity cell and conductivity electrode should be cleaned in time if they are found to be contaminated. If there is oil pollution, it is necessary to use a solution dipped in analytical grade tetrachlorine.
Wipe carefully with a cotton ball of charcoal, acetone or absolute ethanol; you can also use 50% warm detergent (or household soap powder) to clean.
Scrub with a nylon brush, and then rinse the inside of the electrode with distilled water to ensure that there is no oil deposit on the inner and outer electrode surfaces. Remember to use your fingers
Touch the conductivity electrode. For dirt deposits with strong adhesion, soak them in about 2% hydrochloric acid solution for 3 to 5 minutes, and then
Wash repeatedly with distilled water until there is no obvious dirt on the electrode surface.
8. Complete set
14
HGY2028Model Online Conductivity Meter Instructions
Since the measuring range of the instrument is different, conductivity electrodes with different electrode constants need to be equipped, so the user needs to provide the instrument measurement when ordering.
Range, indicate conductivity cell constant. If you have other special requirements, please indicate.
The warranty period of this instrument is one year. If the failure is not caused by human factors within one year, our company will be responsible for repairs. Specific contact party
Email:[email protected]