0% found this document useful (0 votes)
62 views19 pages

DS 500&DS 502 Documentation Folder T6195 3

The document outlines the specifications and diagrams for measuring instruments related to project number 12200503, including block diagrams, exploded views, electronics details, load cell specifications, and sealing positions. It includes detailed information on various components such as main boards and load cells, along with their characteristics and performance metrics. The type examination confirms that the documentation aligns with legal regulations and does not deviate from the stated specifications.

Uploaded by

Djura Elmer
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
62 views19 pages

DS 500&DS 502 Documentation Folder T6195 3

The document outlines the specifications and diagrams for measuring instruments related to project number 12200503, including block diagrams, exploded views, electronics details, load cell specifications, and sealing positions. It includes detailed information on various components such as main boards and load cells, along with their characteristics and performance metrics. The type examination confirms that the documentation aligns with legal regulations and does not deviate from the stated specifications.

Uploaded by

Djura Elmer
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 19

Documentation

folder
Number T6195-3
Project number 12200503
Page 1 of 1

Number Pages Description Remark

Block diagrams
6195/4-01 1 DS-500 block diagram -
6195/4-02 1 DS-502 block diagram -

Exploded views
6195/4-03 1 DS-500 exploded view -
6195/4-04 1 DS-502 exploded view -

Electronics
6195/4-05 1 TPB-2857 main board -
6195/4-06 1 PL2857; IC list on the board -
6195/4-07 1 TPB-2858 A/D board -
6195/4-08 1 PL2858; IC list on the board -
6195/4-09 1 STB-2158 mainboard & A/D board -
6195/4-10 1 PL2158; IC list on the board -
6195/4-11 1 STB-2158 mainboard -
6195/4-12 1 PLS2158V1.0; IC list on the board -
6195/4-13 1 STB-2120- A/D board -
6195/4-14 1 PLS2120; IC list on the board -

Load cell
6195/4-15 1 DS-500 load cell assembly -
6195/4-16 2 Load cell type J specification sheet -

Sealing
6195/4-17 1 DS-500 sealing position -

The measuring instrument for which the type examination is granted, is described in the above
mentioned documentation as far as:
(a) this documentation is not in conflict with legal regulations or the intent of this type examination
(b) this type examination does not deviate from the stated documentation.
500
Doc no 6195/4-01
Page 1 of 1
Drawing no. DS-502-BD1

Doc no 6195/4-02
Page 1 of 1
Doc no 6195/4-03
Page 1 of 1
Doc no 6195/4-04
Page 1 of 1
Doc no 6195/4-05
Page 1 of 1
Doc no 6195/4-06
Page 1 of 1
Doc no 6195/4-07
Page 1 of 1
Doc no 6195/4-08
Page 1 of 1
o
o

Doc no 6195/4-09
Page 1 of 1
IC list on the board

Drawing number: PLS2158

STB-2158 Main board & A/D Board

IC Number IC type

U1 IC 1232

U2 IC 2985

U3 IC R5F21254 or R5F21255 or R5F21256 or

R5F21257

U4 IC 9793

Doc no 6195/4-10
Page 1 of 1
Doc no 6195/4-11
Page 1 of 1
IC list on the board

Drawing number: PLS2158V1.0

STB-2158 Main board

IC Number IC type

U1 Not used

U2 IC 2985

U3 IC R5F21254 or R5F21255 or R5F21256 or

R5F21257

U4 IC 9793

Doc no 6195/4-12
Page 1 of 1
IC list on the board

Drawing number: PLS2120

STB-2120 A/D BOARD

IC Number IC type

U1 IC 1232

Doc no 6195/4-13
Page 1 of 1
o
o

Doc no 6195/4-14
Page 1 of 1
Doc no 6195/4-15
Page 1 of 1
Load cell specification sheet
Drawing Nr LC-JS 1/2
1. Type J Type Load Cell
off-center / single point

2. Accuracy Class C3

3. Characteristics
a) Maximum Capacity (EMax) 2.3kg , 4.5kg , 9kg , 22.5kg , 45kg
b) Rated Output (R.O.) 1.5mV/V 7.5%
c) Non-linearity Over 0~1/6 of R.O.
d) Hysteresis within 0.01% of R.O.
e) Repeatability Over 1/6~4/6 of R.O.
f) Output error due to temp. within 0.02% of R.O.
(-10 ~ 40 ) Over 4/6~1 of R.O.
g) Cumulative error ( c ~ f ) within 0.03% of R.O.

h) Creep within 0.03% R.O.


Note Output value difference between after loading and 30 minutes after loading.
i) Zero return within 0.01% of R.O.
Note Zero point difference between before loading and after removing a 30 min.
Applied load.
j) Zero shift Due temp. within 0.02% of R.O. per 5

4. Temperature range
a) Range compensated (Operating ) -10 ~ 40
b) Range allowable (storge ) -20 ~ 50

5. Electric factors
a) Applied voltage recommended 5~12 VDC
b) Applied voltage allowable 20 VDC
c) Input impedance 1100 50 ohm
d) Output impedance 1000 10 ohm
e) Insulation resistance 500 Mohm or more
(50 VDC between bridge / earth)
f) Zero balance within 50% of R.O.
g) Output / Input signals
-Applied voltage ( ) red wire
-Applied voltage ( ) white wire
-Output voltage ( ) green wire
-Output voltage ( ) yellow wire
-Earth black wire

Doc no 6195/4-16
Page 1 of 2
Load cell specification sheet
Drawing Nr LC-JS 2/2

6. Mechanical factors
a Safe overload 200 %
b Durability within 0.03% of R.O.
(Output change in between before and after 1 million time loading test of rated
capacity)

7. Protection against moisture


a Protection Butyl rubber sheet
b Moisture compensation 0 ~95 % RH

Doc no 6195/4-16
Page 2 of 2
Doc no 6195/4-17