Copia de Man-553 Ventilador de Traslado Hamilton Medical c2
Copia de Man-553 Ventilador de Traslado Hamilton Medical c2
160001
0197
Service Manual
HAMILTON-C2
August 2018
624165/07
© 2018 Hamilton Medical AG. All rights reserved. Printed in Manufacturer
Switzerland.
Hamilton Medical AG
No part of this publication may be reproduced, stored in a Via Crusch 8, CH-7402 Bonaduz,
database or retrieval system, or transmitted in any form or Switzerland
by any means, electronic, mechanical, or by photocopying, Phone: (+41) 58 610 10 20
recording, or otherwise, without prior written permission of Fax: (+41) 58 610 00 20
Hamilton Medical AG. [email protected]
www.hamilton-medical.com
This document may be revised, replaced, or made obsolete
by other documents by Hamilton Medical AG at any time
and without notice. Ensure that you have the most current Distributor in USA
applicable version of this document; if in doubt, contact the
Hamilton Medical, Inc.
technical support department of Hamilton Medical AG,
4990 Energy Way, P.O. Box 30008
Switzerland. While the information set forth herein is
Reno, NV 89520
believed to be accurate, it is not a substitute for the exercise
Phone: (775) 858-3200
of professional judgment.
Toll-free: (800) 426-6331
Nothing in this document shall limit or restrict in any way Fax: (775) 856-5621
Hamilton Medical AG’s right to revise or otherwise change [email protected]
or modify the equipment (including its software) described
herein, without notice. In the absence of an express, written
agreement
agreeme nt to the contrary, Hamilton Medical AG has no
obligation to furnish any such revisions, changes, or modifi-
cations to the owner or user of the equipment (including
software) described herein.
The equipment must be operated, serviced, or upgraded
only by trained professionals. Hamilton Medical AG’s sole
responsibility with respect to the equipment and its use is as
stated in the limited warranty provided in the device Opera-
tor’s Manual.
4 English | 624165/07
Table of Contents
tsT
n
fC
leo
ab
Chapter
Chapter 1 Conventio
Conventions...........
ns..................
...............
...............
..............
..............
...............
...............
..............
..............
...............
...............
..............
..............
...............
.......... 13
1.1 Notes,
Notes, Cautions
Cautions and Warnings
Warnings ......
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
.........
... 14
1.2 Typograph
Typographic
ic Conve
Conventions..........
ntions................
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
........ 14
1.3 Expressio
Expressions.......
ns.............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
..........
.... 15
1.4 Foreword
Foreword ......
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
........ 16
Chapter
Chapter 2 HAMILTON-
HAMILTON-C2
C2 Overview
Overview .......
..............
..............
..............
...............
...............
..............
..............
...............
...............
..............
..............
..............
.........
.. 19
2.1 Front Component
Componentss Overview
Overview ......
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
........ 20
2.2 Rear Componen
Components
ts Overview
Overview .....
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
........
.. 21
2.3 Interacti
Interaction
on Panel
Panel Internal
Internal Components
Components Overvie
Overview
w ......
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...... 22
2.4 Ventilatio
Ventilation
n Unit Internal
Internal Components
Components Overview..Overview........ ............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...... 23
2.4.1 Top Section
Section .....
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
........
.. 24
2.4.2 Middle Secti
Section
on ......
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
........
.. 25
2.4.3 Bottom
Bottom Sec
Section...
tion.........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
..........
.... 26
Chapter
Chapter 3 Pneumatic
Pneumatics:
s: Overview
Overview and Theory
Theory of Operatio
Operation.........
n................
..............
..............
...............
...............
..............
..............
.........
.. 27
3.1 Overview.......
Overview............
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
........ 28
3.2 Blowerr Module Overview..
Blowe Overview.......
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...... 29
3.3 Oxygen
Oxygen Mixer
Mixer Block Assemb
Assembly
ly Overview
Overview ......
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
........ 30
3.4 Inspirator
Inspiratoryy Valv
Valvee Over
Overview....
view..........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
........ 31
3.5 Ambient
Ambient Valv
Valvee Over
Overview
view......
...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
........
.. 32
3.6 Expiratory
Expiratory Valve Over
Overview
view.....
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
........ 33
3.7 Patient
Patient Flow Sensor Overvie
Overview
w ......
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...... 34
3.8 Oxygen
Oxygen Sensor Overview
Overview ......
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
........ 35
3.9 Principal
Principal Gas Flow Overview..........
Overview................
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
.........
... 36
3.9.1 Component
Comp onentss that
that Manage
Manage the
the Princip
Principal al Gas
Gas Flow..........
Flow...............
...........
............
............
...........
...........
............
............
...........
...........
..........
.... 37
3.9.2 Principal
Principal Gas
Gas Flow ......
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
.........
... 38
3.9.3 Component
Comp onentss for the Ambient
Ambient Stat Statee Gas
Gas Flow..............
Flow....................
...........
...........
............
............
............
............
............
...........
...........
...........
..... 39
3.9.4 Component
Comp onentss for the Over-
Over-pres
pressure
sure Relie
Relieff Gas
Gas Flow
Flow ......
...........
...........
............
............
...........
...........
............
............
...........
...........
...........
..... 40
3.9.5 Component
Comp onentss for the
the Nebulize
Nebulizerr Gas Flow....
Flow.........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
........
41
3.10 Venti
Ventilatio
lation
n Contr
Control
ol Princ
Principle
iple .....
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
........
.. 42
3.10.1 Pres
Pressure
sure Sens
Sensor
or Asse
Assembly.........
mbly...............
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
...........
..... 43
3.10.2 Flow Sensors
Sensors used for Flow Measurem
Measurement
ent Overview
Overview ......
...........
...........
............
............
...........
...........
............
............
...........
...........
..........
.... 44
3.10.3 Flow Restrictor
Restrictorss used for Flow Reduction
Reduction Overview.......
Overview.............
............
............
............
...........
...........
............
............
...........
...........
..........
.... 45
3.11 Gas Rinse Flow
Flows........
s..............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
........ 46
3.12 Measurem
Measurements
ents of Pneu
Pneumat
matic
ic Sens
Sensors.......
ors............
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...... 47
3.13 Component
Componentss that Measure Gas Flow and Pres
Pressure.
sure.......
............
............
............
............
...........
...........
............
............
...........
...........
............
............
........
.. 48
Chapter
Chapter 4 Electroni
Electronics:
cs: Component
Component Function
Functionss Overview.........
Overview................
..............
...............
...............
..............
..............
...............
..............
...... 49
4.1 Overview.......
Overview............
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
........ 50
4.2 Interacti
Interaction
on Panel Electronics
Electronics Components.
Components......
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
...........
..... 51
4.3 Ventilatio
Ventilation
n Unit Electronic
Electronicss Components
Components ..... ...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
........ 52
4.3.1 Mainboard
Mainb oard MSP160200..
MSP160200........
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
..........
.... 52
4.3.2 Mainboard
Mainboard MSP160382..
MSP160382........
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
.........
.... 53
4.3.3 Mainboard
Mainboard Components..
Components........
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
.........
.... 54
4.3.4
4.3.4 Voltage
Voltage Distri
Distribution
bution and Test
Test Point
Pointss ......
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
..........
..... 56
4.3.5
4.3.5 Powerr Supply
Powe Supply Comp
Component
onentss Overvi
Overview
ew (MSP
(MSP160600
160600 Revis Revision
ion 02)
02) ......
............
............
...........
...........
............
............
...........
........
... 58
Chapter
Chapter 5 Lithium
Lithium Ion Battery
Battery ........
...............
..............
..............
..............
...............
...............
..............
..............
...............
...............
..............
..............
...............
..........
.. 59
5.1 Rechargea
Rechargeable
ble Lithium Ion Batte
Battery...
ry.........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
.........
... 60
5.1.1 Introduction
Introductio n ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
...... 60
5.1.2 Replacing
Repla cing the Battery........
Battery..............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
........
... 62
5.1.3 Storage
Storage Hints.
Hints.......
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
.........
.... 62
5.1.4 Backup
Backup Battery
Battery .....
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
........
... 62
5.1.5 Battery
Battery Life
Life Expecta
Expectancy
ncy / Li-Ion
Li-Ion Batter
Batteryy Maintena
Maintenance.......
nce.............
...........
...........
............
............
...........
...........
............
............
...........
........
... 62
5.1.6 Lithium
Lithium Ion Batter
Batteryy Maintena
Maintenance
nce ......
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
.........
.... 63
5.2 Battery
Battery Charger / Calib
Calibrator....
rator.........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
...... 64
5.2.1 Battery
Battery Charger
Charger / Calibr
Calibrator
ator Instal
Installatio
lationn ......
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
..........
..... 65
5.2.2 Battery
Battery Charging
Charging .....
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
..........
..... 65
5.2.3 Battery
Battery Recharge
Recharge Time.............
Time..................
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
.......
.. 65
5.3 Evaluatio
Evaluation
n of the Battery
Battery State
State of Healt
Health
h ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
........ 66
5.4 Battery
Battery Re-c
Re-calibr
alibration
ation ......
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
........ 68
5.4.1 General
Gene ral Informat
Information
ion ......
............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
..........
..... 69
5.4.2 Re-calibr
Re-calibration
ation Time.............
Time...................
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
...... 70
5.4.3 Re-calibr
Re-calibration
ation Descript
Description
ion......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
.........
.... 71
5.4.4
5.4.4 Recharge
Rech argeable
able Lithiu
Lithium
m Ion
Ion Battery
Battery CapaCapacitycity ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
...... 72
5.5 Lithium
Lithium Ion Battery
Battery Handling
Handling Precautions....
Precautions..........
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
........
.. 73
Chapter
Chapter 6 Preventive
Preventive Maintenan
Maintenance
ce and Test
Testing....
ing............
...............
..............
..............
...............
...............
..............
..............
..............
...............
.......... 75
6.1 Overview.......
Overview.............
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
........ 76
6.1.1 Introductio
Introduction n ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
...... 76
6.1.2 Checking
Chec king the
the Softwar
Softwaree Level
Level ......
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
.........
.... 76
6.1.3 Itemss Required
Item Required for
for Prevent
Preventive
ive Mainten
Maintenance ance andand Testing....
Testing..........
...........
...........
............
............
............
............
............
...........
...........
...... 77
6.1.4
6.1.4 Overview
Over view of Preven
Preventive
tive Maint
Maintenan
enance ce and
and Testing
Testing ......
............
............
...........
...........
............
............
...........
...........
............
............
...........
.......
.. 77
6.2 Hospital
Hospital Preventive
Preventive Maintenan
Maintenance
ce ......
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
........ 78
6.3 Engineer
Engineer Preventive
Preventive Maintena
Maintenance
nce ......
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
...... 79
Chapter
Chapter 7 Electrica
Electricall Safety ........
...............
..............
..............
...............
...............
..............
..............
...............
...............
..............
..............
...............
...............
..............
....... 81
7.1 Electrica
Electricall Safe
Safety
ty Over
Overview
view ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
........ 82
7.2 Electrica
Electricall Safe
Safety
ty Test
Tests.........
s...............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
...........
..... 82
7.2.1 Preparat
Prep aration
ion for Test
Tests......
s............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
..........
..... 82
7.2.2 Perform
Perform the Tests
Tests ......
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
.........
.... 83
7.2.3 Device
Devi ce Type ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
.......
.. 83
7.2.4 Internal
Internal Resistanc
Resistancee Checks..........
Checks................
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
.........
.... 84
7.2.5 Electrica
Electricall Safety
Safety Tests
Tests - IEC 62353..........
62353................
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
..........
..... 87
Chapter
Chapter 8 Service
Service Softw
Software.....
are.............
...............
..............
..............
...............
...............
..............
..............
...............
...............
..............
..............
...............
...............
......... 91
8.1 Introductio
Introduction
n ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
.........
... 92
6 English | 624165/07
Table of Contents
8.2 Functions
Functions of the Service Softwar
Softwaree ......
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...... 93
8.3 Structure
Structure of the Service
Service Software.....
Software...........
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
........ 94
8.4 Starting
Starting the Serv
Service
ice Software.......
Software.............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...... 95
8.5 Service
Service Software
Software Screen
Screen Layout ......
............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
........
.. 96
8.6 Making
Making Scre
Screensho
enshots
ts ......
............
...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
........
.. 97
8.7 Ventilator
Ventilator Info ......
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...... 98
8.16.1 Pressure..
Pressure........
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
...........
.....
196
8.16.2 Flow...............
Flow.....................
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
.........
...198
8.16.3 O2 Mixe
Mixerr ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
...........
.....200
8.16.4 Tightness.....
Tightness..........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
.......202
.
8.16.5 Alarming
Alarming ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
...........
.....204
8.17 Sensor
Sensor Data....
Data..........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
......207
8.17.1 Sensor
Sensor 1 ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
........208
8.17.2 Sensor
Sensor 2 ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
........209
8.17.3 Sensor
Sensor 3 ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
........210
8.17.4 Sensor
Sensor Stat
Status....
us..........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
........211
8.18 Log/Config
Log/Config Files...
Files.........
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
........212
8.18.3
8.18.3 Export
Export (Inst
(Instrume
rument
nt Repor
Reportt and Events)...............
Events).....................
............
............
...........
...........
............
............
...........
...........
............
............
...........
.........
.... 215
8.19 Software
Software Upda
Updates.....
tes...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
........218
8.19.1
8.19.1 Software
Software Update
Update from Version < 2.2.0 ......
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
......218
8.19.2 Software
Software Update
Update from
from Version
Version 2.x to 2.2.5 or Higher
Higher ......
............
...........
...........
............
............
...........
...........
............
............
...........
.......
.. 221
8.20 Pre-opera
Pre-operationa
tionall Chec
Checks.......
ks.............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
.........
...226
8.21 General Tests....................................................
Tests.......................................................................................................................................
................................................................................... 227
Chapter
Chapter 9 Technical
Technical Faults....................
Faults...........................
..............
..............
...............
...............
..............
..............
...............
...............
..............
..............
...............
..........
..243
9.1 Technica
Technicall Fault
Faultss Over
Overview........
view.............
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
...........
..... 244
9.2 Troublesho
Troubleshooting
oting.....
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
.........
... 246
8 English | 624165/07
Table of Contents
12.3 Trolley
Trolley PN 160150.......
160150.............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
...........
.....336
12.4 Major Components
Components of the Interaction
Interaction Pane
Panel......
l............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
........
.. 337
12.5 Major Components
Components of the Ventilation
Ventilation Unit ......
...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
.........
... 338
12.5.1
12.5.1 Ventilatio
Ventilation
n Unit Cove
Covers........
rs..............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
..........
.....339
12.5.2
12.5.2 Ventilatio
Ventilation
n Unit Top Sec
Section
tion ......
...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
..........
.....340
12.5.3
12.5.3 Ventilatio
Ventilation
n Unit Middle Sec
Section
tion .....
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
......341
12.5.4
12.5.4 Venti
Ventilatio
lation
n Unit Botto
Bottom m Sec
Section...
tion.........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
........
...342
342
12.6 Inter
Interacti
action
on Pane
Panell Part
Partss ......
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
......343
12.6.1
12.6.1 Interacti
Interaction
on Pane
Panell Cove
Covers
rs ......
............
...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
..........
.....343
12.6.2
12.6.2 Backlight
Backlight Drive
Driverr Boar
Board
d ......
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
........
... 344
12.6.3
12.6.3 Front Pane
Panell Board........
Board..............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
.......
.. 345
12.6.4
12.6.4 LCD Displa
Displayy and Touc
Touch
h Scre
Screen....
en..........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
.........
.... 346
12.6.5
12.6.5 P&T Contr
Control
ol Knob Enco
Encoder..........
der................
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
........
... 347
12.7 Ventilatio
Ventilation
n Unit Part
Parts........
s..............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
......348
12.7.1
12.7.1 Top Cove
Coverr ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
.........
....348
12.7.2
12.7.2 Front Cove
Coverr ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
.......
..349
12.7.3
12.7.3 Rear Cove
Coverr ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
........
...350
12.7.4
12.7.4 Ventilatio
Ventilation
n Unit Base ......
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
.........
....
351
12.7.5
12.7.5 Backup
Backup Batte
Battery
ry Pac
Packk Compa
Compartme
rtment.....
nt...........
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
.......
.. 352
12.7.6
12.7.6 Expiratory
Expiratory Valv
Valvee Asse
Assembly
mbly ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
........
... 353
12.7.7
12.7.7 Mixerr Block Asse
Mixe Assembly.........
mbly...............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
......354
12.8 Detailed Parts Descriptions.......................................................
Descriptions...................................................................................................................
............................................................ 355
12.8.1
12.8.1 Tubing
Tubing and Fitti
Fittings.......
ngs.............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
......355
12.8.2
12.8.2 Clamps
Clamps and Fast
Fastener
eners..........
s................
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
..........
.....357
12.8.3
12.8.3 Stickers
Stickers and Labe
Labels
ls ......
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
.......
.. 358
12.8.4
12.8.4 Pneumati
Pneumaticc Part
Partss and Asse
Assemblie
mbliess .....
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
..........
.....359
12.8.5
12.8.5 Flow Sens
Sensors
ors ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
..........
.....361
12.8.6
12.8.6 Assemble
Assembled
d Comp
Component
onents..........
s...............
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
.......
..362
12.8.7
12.8.7 Colored
Colored Metal Rings and Contr
Controls
ols ......
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
.......
..
363
12.8.8
12.8.8 Rubberr Sea
Rubbe Seals
ls and Grom
Grommets.
mets.......
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
.........
.... 363
12.8.9
12.8.9 Metall Brac
Meta Brackets
kets and Fram
Framee Comp
Component
onents......
s............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
..........
.....364
12.8.10 Electrical/Ele
Electrical/Electronic
ctronic Cables.......................................................................................................... 366
12.8.11 Electronic Printed Circuit Boards .................................................................................................369
12.8.12 Electrical/Ele
Electrical/Electronic
ctronic Devices............................................................................
Devices.........................................................................................................
............................. 372
12.8.13 External Covers and External Hardware...........................................................
Hardware.......................................................................................
............................ 374
12.8.14 Screws..............................................................
Screws........................................................................................................................................
.......................................................................... 376
12.8.15 Standoffs.................................................................................
Standoffs....................................................................................................................................
................................................... 379
12.8.16 Washers ...............................................................................................................................
.....................................................................................................................................
...... 380
12.8.17 Screw Nuts ...........................................................................................................................
.................................................................................................................................
...... 380
Chapter
Chapter 13 Sche
Schematic
maticss .......
...............
...............
..............
..............
...............
...............
..............
..............
...............
...............
..............
..............
...............
...............
..............
.......381
13.1 Block Diag
Diagram........
ram..............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
.........
...382
10 English | 624165/07
Table of Contents
Chapter
Chapter 14 Softw
Software
are Revis
Revisions
ions,, Featu
Features
res and Comp
Compatib
atibilit
ilityy .......
...............
...............
..............
..............
...............
...............
..............
..........
...385
14.1 Introductio
Introduction
n .....
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
.........
...386
14.2 Software
Software Vers
Version
ion Inform
Informatio
ation
n ......
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
......386
Chapter
Chapter 15 Hardw
Hardware
are Revi
Revision
sions,
s, Featu
Features
res and Comp
Compatibi
atibility
lity ........
...............
..............
..............
...............
...............
..............
..............
.........
..397
15.1 Introductio
Introduction
n .....
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
.........
...398
15.2 Comp
Component
onent Histo
History
ry and Impro
Improvem vements.
ents.......
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
........
..398
398
15.2.1 Inter
Interacti
action
on Pane
Panel......
l............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
...........
.....398
15.2.2 Ventilatio
Ventilation
n Unit ......
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
........402
Chapter
Chapter 16 Glos
Glossary
sary ........
...............
..............
..............
...............
...............
..............
..............
..............
...............
...............
..............
..............
...............
...............
..............
...........
....415
16.1 Definition
Definitionss of Expr
Expressio
ession.....
n..........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
..........
....416
16.2 Alarm Codes .....
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
........
..421
16.2.1 Patient
Patient Alarms - 100000 Alarm Codes
Codes ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
........
.. 421
16.2.2 Technical
Technical Alarms
Alarms - 200000 Alarm Codes .....
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
..........
.... 423
16.2.3 Technical
Technical Failure
Failure Ending
Ending in Safety
Safety Mode - 300000
300000 Alarm Codes
Codes ......
............
...........
...........
............
............
...........
...........
.........
... 429
16.2.4 Technical
Technical Failur
Failuree Ending
Ending in Ambient
Ambient Mode - 400000
400000 Alarm
Alarm Codes
Codes ......
...........
...........
............
............
...........
...........
...........
.....431
16.2.5 500000 Alarm Code
Codess ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
..........
.... 433
12 English | 624165/07
1
Conventions
13
1 Conventions
WARNING
A WARNING alerts the user to the possibility of injury, death, or other serious adverse reactions associated
associated with the use or
misuse of the device.
CAUTION
NOTICE
• The vent
ventila
ilator
tor case
case
• A tou
touch
ch ke
keyy
• A printed
printed circu
circuit
it board
board
Number sequence 1. Step one in a sequence Organizes the performance of actions into
2. Step two in a sequence steps.
Blue text 1.2 If you are using a PDF file to view this, you
can hyperlink to the reference by clicking on
these items.
14 English | 624165/07
Expressions 1
1.3 Expressions
Table 1-2. Expressions
Expression Example Explanation
Activate Activate LED ON/OFF. Using the P&T Control Knob, you must first
select the LED ON/OFF Button on the venti-
lator's screen, and then press the P&T Con-
trol Knob. The button on the screen
changes its appearance, so that it looks
“pressed”. It now performs its function
(turning the LED ON in this case). Some-
times you are told to “activate and set” a
field. In this case you first activate the field,
and then turn the P&T Control Knob to set a
value.
Deactivate Deactivate LED ON/OFF. With LED ON/OFF still selected and acti-
vated, you must press the P&T Control Knob
again. The button on the screen changes its
appearance, so that it looks “unpressed”. It
stops performing its function (turning the
LED OFF in this case).
Select Select LED ON/OFF. Touch the Touch Screen to select the LED
ON/OFF Button.
De-select De-select LED ON/OFF. Touch the Touch Screen to deselect the LED
ON/OFF Button.
Pressure Patient pressure is 80 mbar. Pressure refers to the amount of pressure
above ambient pressure. If the patient pres-
sure (Ppat) is 80 mbar, it means the pressure
is 80 mbar above the ambient (room) pres-
sure.
Upgrade Timplement
his kit or so
afnew
twarefunction.
enables an upgrade to An
to aupgrade is the are
device. There addition
threeof newtofunctions
ways perform
an upgrade:
• Add a hardwar
hardwaree item that
that offers addi-
addi-
tional functions.
• Upgrade
Upgrade to a higher softwar
softwaree revision.
revision.
Depending on how important is the soft-
ware upgrade, it is either marked by an
increment on the first or on the second
digit (e.g. 1.2.0 to 1.3.0 or 1.2.0 to
2.0.0).
• Upgra
Upgrade
de to a higher
higher type of softwa
software.
re.
* With this kit, you can update or upgrade Unless otherwise stated, a syntax variable
from software version. (“wild card”) indicates the use of any alpha-
numeric character.
1.4 Foreword
WARNING
To prevent possible patient injury, disconnect the patient from the ventilator before you start service or maintenance.
Service the HAMILTON-C2 only as described in this manual, using only parts approved or supplied by Hamilton Medical.
Incorrect parts, components or assemblies could result in patient injury. See available Spare Parts Section 12
12..
Hamilton Medical does not offer any warranty on incorrectly used parts.
NOTICE
If you have questions about testing or any part of this manual, contact Hamilton Medical AG Technical Support.
Information about operating the HAMILTON-C2. See the HAMILTON-C2 Operator’s Manual for operating instructions.
This HAMILTON-C2 Service Manual covers:
16 English | 624165/07
Foreword 1
HAMILTON-C2 Overview Section 2 This section explains the theory behind the You should fully understand this section.
HAMILTON-C2.
Pneumatics: Overview and Theory of This section explains each component and the You should be able to name and explain the
Operation Section 3 gas flows, flow measurements, and pressure functions of all the major components.
measurements in the pneumatic circuits.
Electronics: Component Functions This section explains the basic functions of the You should be able to identify all circuit
Overview Section 4 printed circuit boards, Printed circuit boards boards, and know where they are positioned
are not repaired in the field. in the HAMILTON-C2.
Lithium Ion Battery Section 5 This section explains the use, care and mainte- You have to understand the safety concerns
nance of the Lithium Ion Battery Pack. and hazards, and know how to perform
charging and calibration of the Lithium Ion
Battery Pack.
Hospital Preventive Maintenance Sec- This section gives the Hospital Preventive You should be able to perform this mainte-
tion 6.2 Maintenance details. nance and ascertain if this maintenance is
being performed regularly.
Engineer Preventive Maintenance Sec- This section gives Engineer Preventive Mainte- You have to be able to perform all the tasks in
tion 6.3 nance details. this section.
Electrical Safety Overview Section 7.1 This section lists further test's you must per- You have to be able to use this
t his section to test
form on the HAMILTON-C2 before you start the HAMILTON-C2.
the Service Software test's.
Service Software Section 8 This section explains how you perform the You have to know how to perform all the
test's that are built into the software of the appropriate test's.
HAMILTON-C2.
Technical Faults Section 9.1 This section gives an overview of the alarm You should know how to interpret technical
indications. faults to isolated faulty components.
Components Removal/Assembly Sec- This section explains how to remove and You have to be able to use this
t his section to
tion 10 assemble each major component. make repairs.
Maintenance Tools and Test Equipment This section lists the equipment you require to Check this section to make sure you have the
Section 11 work on the HAMILTON-C2. correct tools and test equipment.
Spare Parts Section 12 Information resource. You only require this section when you must
order spare parts.
Schematics Section 13 This section includes many of the schematics You are sometimes directed to this section
produced by Hamilton Medical for internal when reading in other parts of the manual.
use.
Software Revisions, Features and Com- This section informs about the software ver- You only require this section when the soft-
patibility Section 14.1 sions that were introduced over time. ware must be updated or when the update is
recommended.
Hardware Revisions, Features and Com- This section informs about the hardware com- You only require this section when parts need
patibility Section 15 ponents that have changed over time and to be replaced.
about its compatibility.
Glossary Section 16 Glossary contains Definitions of Expressions Helps you to understand the ventilator and the
and Alarm Codes alarms.
Test Report Section 17.1 Test Report pages for the Service Software Complete the report when using the service
section. software tests.
18 English | 624165/07
2
HAMILTON-C2 Overview
19
2 HAMILTON-C2 Overview
E Oxygen Se
Sensor an
and Co
Cover (O
(Oxygen Se
Sensor no
not sh
shown) K Quick Lo
Lock to
to re
remove HA
HAMILTON-C2 fr
from th
the Tro
Trollley
20 English | 624165/07
Rear Components Overview 2
B 10.4” TFT Display with Backlight D P&T (Press and Turn) Control Knob Encoder
F Back
Backliligh
ghtt Inv
Inver
erte
terr Boa
Board
rd (O
(Obs
bsol
olet
ete)
e)
22 English | 624165/07
Ventilation Unit Internal Components Overview 2
B Middle Section
24 English | 624165/07
Middle Section 2
Figure 2-9. Ventilation Unit Internal Components Middle Section Rear Overview
E HEPA Filter La
Latch
E Ambient Va
Valve
Figure 2-11. Ventilation Unit Internal Components Bottom Section Rear Overview
26 English | 624165/07
3
Pneumatics: Overview and Theory of
Operation
3.1 Overview........
Overview.. ...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
........ 28
3.2 Blowerr Module Ove
Blowe Overvie
rvieww ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
........ 29
3.3 Oxygen
Oxyg en Mixer
Mixer Block Asse Assembly
mbly OveOverview......
rview...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
.........
... 30
3.4 Inspirator
Inspiratoryy Valv
Valvee Ove
Overvie
rvieww ......
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
...........
..... 31
3.5 Ambient
Ambie nt Valv
Valvee Overv
Overview iew ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
.........
... 32
3.6 Expiratory
Expiratory Valv
Valvee Overv
Overview iew ......
...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...... 33
3.7 Patient
Patient Flow
Flow Sensor
Sensor Overvi
Overview.....
ew...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
........ 34
3.8 Oxygen
Oxyg en Sensor
Sensor Overv
Overview........
iew.............
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...... 35
3.9 Principal
Principal Gas
Gas Flow
Flow Overvie
Overview..........
w................
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
...........
..... 36
3.10 Ventilatio
Venti lation
n Contr
Control ol Princ
Principle.......
iple.............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
........ 42
3.11 Gas Rinse Flow
Flowss ......
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
..........
.... 46
3.12 Measurem
Meas urements
ents of PneuPneumati
maticc Sens
Sensors ors ......
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...... 47
3.13 Component
Comp onentss that Measure
Measure Gas Gas Flow and Pressur Pressuree ......
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
..........
.... 48
27
3 Pneumatics: Overview and Theory
Theory of Operation
3.1 Overview
WARNING
Repair of parts, components or assemblies is not permitted by Hamilton Medical AG, as incorrectly repaired parts, compo-
nents or assemblies could result in patient injury. See the available Spare Parts Section 12
This section introduces all the major pneumatic components in the ventilator. In addition, the gas flows and theory of opera-
tion are explained.
NOTICE
The figures in this section show exploded and transparent views of the ventilator components. The exploded diagrams may
not always show components in their correct positions.
Filter Air
Filter O2
Filter Air
Sensor QO2
HEPA
Sensor Pfilter
Mixer valve O2
Noise dampening
Nebulizer
valve
Sensor TBlower
Sensor Pambient
Blower
Heat exchanger
Sensor
Sensor Pvent_control
Paw
Inspiratory valve
Filter Sensor
Sensor Rinseflow Pvent_monitor
Pflowsensor
Oxygen sensor
Sensor Qvent
Autozero
valve Autozero
valve
Autozero
valve
Ambient valve
Standard
rinse flow Internal expiratory
Standard
rinse flow valve
Nebulizer
Expiratory
proximal flow valve cover
28 English | 624165/07
Blower Module Overview 3
A Shaped
Shaped Tub
Tubee from
from the
the QO2
QO2 Flow
Flow Sens
Sensor
or to
to the Blo
Blowe
werr Mod-
Mod- E High
High Pr
Pres
essu
sure
re Ox
Oxyg
ygen
en Co
Cont
ntro
roll Va
Valv
lvee
ule
B QO2 Flow Sensor F Outlet to supply Oxygen to the Front Panel Nebulizer Con-
nection
C Oxygen Mixer Block Assembly G High pressure Oxygen DISS or NIST Connection
30 English | 624165/07
Inspiratory Valve Overview 3
A The Ai
Air/O
r/Oxyg
xygen
en Mixt
Mixture
ure fro
from
m the
the Inspi
Inspirat
ratory
ory Val
Valve
ve to the D Amb
mbie
ient
nt Va
Valv
lvee Sol
Solen
enoi
oid
d
Patient Breathing Circuit
C Ambi
Ambien
entt Va
Valv
lvee Se
Seal
alin
ing
g Me
Memb
mbra
rane
ne
32 English | 624165/07
Expiratory Valve Overview 3
WARNING
Never attach a Spirometer
Spirometer or any other device or tube to the exhaust port of the Expiratory Valve. This will cause the venti-
lator to lose full control of PEEP/CPA
PEEP/CPAP.
P.
The Expiratory Valve enables gas to escape from the Patient Breathing Circuit in a controlled manner, allowing the patient to
exhale.
The Expiratory Valve:
• Maintains PEEP/CPAP
PEEP/CPAP if required
required - to do this, the action of the Expiratory Valve
Valve is closely synchronized
synchronized with the action of
the Inspiratory Valve
• In Ambient
Ambient State
State,, the Expirat
Expiratory
ory Valve
Valve is open
open
A Expiratory Valve Assembly E Expired gas from the Patient Breathing Circuit
D Inlet
Inlet conn
connect
ection
ion fro
from
m the
the Patie
Patient
nt Brea
Breathi
thing
ng Circ
Circuit
uit
A Blue Tube
Blue Tube - Proxi
Proximal
mal sid
sidee meas
measure
ured
d Pflows
Pflowsens
ensor
or Press
Pressure
ure D Two-way
Two-w ay Gas
Gas Flow
Flow con
connec
nectio
tion
n on the Dis
Distal
tal sid
sidee of the
the
and Paw Pressure Patient Flow Sensor
B Clea
Clearr Tub
Tubee - Dis
ista
tall si
side mea
easu
sure
ress Pfl
Pflow
owse
sens
nsor
or Pre
resssu
sure
re E Two-
o-w
way GaGass Fl
Flow co
con
nne
nect
ctio
ion
n on
on the
the Pr
Pro
oxi
xim
mal si
side
de of th
thee
Patient Flow Sensor
C Vari
Variab
able
le Or
Orif
ific
icee Me
Memb
mbra
rane
ne
NOTICE
Ventilation is not totally dependent on the Patient Flow Sensor. If the Patient Flow Sensor malfunctions, patient ventilation
continues. The ventilator has an internal flow measurem
measurement
ent at the Inspiratory Valve for Inspiratory Flow.
34 English | 624165/07
Oxygen Sensor Overview 3
WARNING
The ventilator should never be used for ventilating a patient without some means of monitoring the oxygen content in the
gas mixture delivered to the patient. The ventilator can not be operated without an Oxygen Sensor installed.
The Oxygen Sensor (A, see Figure 3-8 Oxygen Sensor Overview)
Overview) is attached to the front left side of the ventilator behind the
Oxygen Sensor Cover (B, see Figure 3-8 Oxygen Sensor Overview). It is used to monitor the oxygen concentration in the gases
delivered
required. to the patient. The Oxygen Sensor performs only a monitoring function, and can be disabled by the user, if
NOTICE
Hamilton Medical Oxygen Sensors are available from Hamilton Medical AG only.
The Hamilton Medical Oxygen Sensor produces a voltage between 11 and 13mV at 21% O2 which changes with oxygen
concentration. Each unit comprises a teflon-bonded gold cathode and a lead anode, submerged in a liquid electrolyte solu-
tion. When oxygen diffuses through the flouropolymer membrane, the electrochemical
electrochemical reduction of oxygen on the cathode
and the corresponding oxidation of the anode generates an electrical current that is proportional to the concentration of oxy-
gen.
• Room air
air and oxygen
oxygen mixture
mixture in the Blower
Blower Module (D regulati
regulation
on point 1).
• Room air and
and oxygen mixture
mixture through the
the Inspiratory Valve and Patient Flow Sensor (E - regulation point 2).
• Expired gas
gas from the patient through the Expiratory Valve (F - regulation point 3).
Expired gases from the patient are allowed to escape into the room.
The following sections provides details of the components that manage the “Principal Gas Flow”.
NOTICE
High-pressure-oxygen:
High-pressure-oxygen: Maximum Pressure 600 kPa / Maximum Flow 200 l/min
Low-pressure-oxygen:
Low-pressure-oxygen: Maximum Pressure 600 kPa / Maximum Flow 15 l/min
36 English | 624165/07
Components that Manage the Principal Gas Flow 3
C Mixer Bl
Block As
Assembly Measures and Re
Regulates th
the flow of th
the high pressure ox
oxygen (B
(B, se
see Figure 3-10 Com-
ponents that Manage the Principal Gas Flow) into the Blower Module.
D Blower Module Compresses the air and oxygen mixture with a Turbine Blower.
E Insp
Inspir
irat
ator
oryy Val
Valve
ve an
and
d Qve
Qvent
nt Fl
Flow
ow Se
Sen-
n- Measures and Regulates the flow of the air/oxygen mixture to the patient and the pres-
sor sure during inspiration. It regulates the Base Flow during expiration.
F Patient Br
Breathing Ci
Circ
rcu
uit Supplies ai
air/oxygen ga
gas mi
mixture to th
the pa
patient, an
and re
removes ex
exhalation ga
gases fr
from th
the
patient.
G Patient Fl
Flow Sensor The Pa
Patient Fl
Flow Sensor mo
monitors the flow of ga
gases to
to and fr
from the patient’s ai
airway.
H Patient The patient breathes in and out (inhales and exhales) with the assistance of the ventila-
tor.
I Expiratory Valve Regulates the flow of expiratory gases from the patient. The Expiratory Valve is nearly
closed during inspiration and regulates PEEP during expiration.
C HEPA Filter Assembly H Air/Oxygen Mixture into the Patient Breathing Circuit
D Air and oxygen mixed in the Blower Module I Expired gas to the Expiratory Valve
E Air/Ox
Air/Oxyge
ygen
n Mixtu
Mixture
re flow
flowss throu
through
gh the
the Turb
Turbine
ine Blo
Blowe
werr
38 English | 624165/07
Components for the Ambient State Gas Flow 3
The one-way function of the Expiratory Valve make sure the patient does not inhale
i nhale his exhaled CO2 (rebreathing).
D Into the Patient Breathing Circuit H From the Patient Breathing Circuit
C Exce
Excess
ss pr
pres
essu
sure
re to th
thee ro
room
om
40 English | 624165/07
Components for the Nebulizer Gas Flow 3
B Tubing fro
Tubing from
m Oxyge
Oxygen
n Mixer
Mixer Bloc
Blockk Asse
Assembl
mblyy to the
the Nebu
Nebuliz
lizer
er D High
High Pr
Pres
essu
sure
re Ox
Oxyg
ygen
en In
Inle
lett
Front Panel connection
On
has the othereffect
a direct hand,on
the Expiratory
the gas in theValve manages
Patient the Circuit
Breathing outflow(back
of the gas. It applies
pressure). a known
It never forcecloses
completely to thethe
membrane
outlet of which
the
Patient Breathing Circuit during ventilation, so the membrane is floating all the time.
The correct ventilation pressure is given by the close collaboration of the Inspiratory Valve, that applies the pressure into the
Patient Breathing Circuit and the Expiratory Valve that stabilizes it by a controlled release of the gas.
42 English | 624165/07
Pressure Sensor Assembly 3
A Pres
Pressu
sure
re Se
Sens
nsor
or As
Asse
semb
mbly
ly E Paw Pressure Sensor
• Provides components for pressure
pressure measurements • A measurement
measurement of the pressure in the Patient Breathing
Breathing
Circuit as measured at the Proximal Flow Sensor, in the
proximal chamber attached to the blue (patient side) pres-
sure-sensing tube
B Pven
Pvent_
t_mo
moni
nito
torr Pres
Pressu
sure
re Sen
Senso
sorr F Pven
Pvent_
t_mo
moni
nito
torr Aut
Autoz
ozer
ero
o Val
Valve
ve
• A sensor which
which measures
measures the pressure at the patient
patient out- • Switches the Pvent_monitor
Pvent_monitor air way pressure
pressure measured at
let, used for monitoring the Patient Flow Sensor to allow for offset re-calculation
to compensate for sensor drift for Pvent_control and
Pvent_monitor
C Pven
Pvent_
t_co
cont
ntro
roll Pres
Pressu
sure
re Sen
Senso
sorr G Dis
ista
tall Aut
utoz
ozeero Va
Valv
lvee
• A sensor which
which measures
measures the pressure at the patient
patient out- • Switches the Pflowsensor
Pflowsensor Pressure Sensor to Ambient Air
Air
let, used for controlling to allow for offset re-calculation to compensate for sensor
drift
D Pflows
Pflowsen
enso
sorr Pres
Pressu
sure
re Sen
Senso
sorr H Pro
roxi
xima
mall Au
Auto
toze
zero
ro Val
Valve
ve
• A measurement
measurement of the pressure difference between
between the • Switches the Pflowsensor
Pflowsensor and Paw Pressure
Pressure Sensor
Sensor to
proximal and distal chambers of the Patient Proximal Flow Ambient Air to allow for offset re-calculation to compen-
Sensor sate for sensor drift
B The Qve
Qvent
nt Flow
Flow Sen
Sensor
sor mea
measur
sures
es the
the flow
flow of the
the Air/
Air/Oxy
Oxygen
gen
Mixture into the Patient Breathing Circuit
44 English | 624165/07
Flow Restrictors used for Flow Reduction Overview 3
A A Restr
Restrict
ictive
ive “Or
“Orifi
ifice”
ce” for
for the
the Oxygen
Oxygen Flow
Flow to the
the Nebul
Nebulize
izerr C A Restri
Restricti
ctive
ve “Scre
“Screen”
en” for
for the Pati
Patient
ent Flow
Flow Sens
Sensor
or Rinse
Rinse Flow
Flow
Valve supply from the Inspiratory Valve
B Two pre
precis
cisely
ely matc
matched
hed “Or
“Orifi
ifice”
ce” Rest
Restric
rictor
torss for the
the Pati
Patient
ent
Flow Sensor Rinse Flow
Figure 3-18. Components that generate the Patient Flow Sensor Rinse Flow
B Inspiratory Valve outlet to QVent Flow Sensor F Rinse Flow Assembly (including flow restrictor)
C Tube outlet to
to th
thee Oxygen Sensor Mo
Mounting Bl
Block G Front pa
panel co
connections fo
forr th
the Patient Pr
Proximal Flow Sensor
D Tube
Tube to the
the Ri
Rins
nsee Flo
Flow
w Ass
Assem
embl
blyy
46 English | 624165/07
Measurements of Pneumatic Sensors 3
Pvent_monitor Pressure Sensor Assembly Measures pressure at the Inspiratory Valve outlet for monitoring
Paw Pressure Sensor Assembly Measures pressure at the Proximal Side of the Patient Flow Sensor
Pfilter Blower Module Measures pressure after the filter; used to indicate if the HEPA Filter
needs to be replaced
Qvent Ventilation Flow Sensor Measures the air/oxygen flow in the Patient Breathing Circuit.
QO2 Oxygen Flow Sensor Measures the oxygen flow into the Blower Module
B Pven
Pvent_
t_mo
moni
nito
torr Pres
Pressu
sure
re Sen
Senso
sorr E Paw Pressure Sensor
• Monitor the pressure in the Patient
Patient Breathing Circuit • Measures the pressure on the
the proximal side
side of the Patient
Flow Sensor
C Pven
Pvent_
t_co
cont
ntro
roll Pr
Pres
essu
sure
re Se
Sens
nsor
or
• Used to control
control the
the Inspirato
Inspiratory
ry Valve
NOTICE
During the Autozero procedure, which occurs in a fraction of a second, the patient is ventilated normally. Since parameters
such as expired tidal volume, resistance, compliance, and PEEP cannot be measured or monitored, the ventilator uses the
most recently measured values until the procedure is complete.
48 English | 624165/07
4
Electronics: Component Functions Overview
4.1 Overview........
Overview.. ...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
........ 50
4.2 Interacti
Interaction
on Pane
Panell Elec
Electroni
tronics cs Comp
Component onentss .....
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...... 51
4.3 Ventilatio
Venti lation
n Unit
Unit Elec
Electronic
tronicss Compon
Components.......
ents............
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...... 52
49
4 Electronics: Component Functions Overview
4.1 Overview
WARNING
Service the ventilator only as described in this manual, using only parts approved or supplied by Hamilton Medical AG.
Incorrectly repaired parts, components or assemblies could result in patient injury. See available Spare Parts Section 12
12..
This section introduces the major electronic components in the ventilator. Where you require more information, cross-refer-
ences direct you to other parts of this manual.
This section does not include a theory of operation, because engineers do not require a detailed knowledge of board-level
electronics to service and maintain the ventilator. All electronic failures are dealt with by replacing complete circuit boards.
Any repairs at a lower level than the spare parts listed in the Spare Parts Section 12 is not permitted
NOTICE
Always send defective Printed Circuit Boards to Hamilton Medical AG with a completed Returned Goods ID Tag.
Refer to PN 699138 ky2help User Guide, available as a download from the Hamilton Medical AG Partner-net.
50 English | 624165/07
Interaction Panel Electronics Components 4
A Front Pa
Panel Bo
Board E Loud
Loudsp
spea
eake
kerr an
and
d Mi
Micr
crop
opho
hone
ne
• Provides 7 Hard Keys for the operator
operator with LED Indicators
Indicators • Audi
Audible
ble indica
indication
tion of alarm
alarmss
• Inte
Interface
rfacess the P&T Control
Control Knob Encode
Encoderr • Mi
Micr
crop
opho
hone
ne::
• LED Colo
Colors
rs avail
available
able:: – The Microph
Microphone
one monitor
monitorss the operati
operation
on of the
– Red Loudspeaker
B Back
Backliligh
ghtt Inv
Inver
erte
terr Boa
Board
rd (O
(Obs
bsol
olet
ete)
e) F USB Socket
• Converts
Converts 5 VDC (Input
(Input Voltage)
Voltage) to 1100 VAC for the Dis-
Dis- • Us
Used
ed to
to::
play Backlight – Down
Download
load Softw
Software
are from
from a USB Stick
– Red - High
High Priority
Priority Alarms
Alarms and
and Technica
Technicall Faults
Faults – -16 Encode
Encoderr position
positionss from
from center
center
– +16 Encod
Encoder
er positi
positions
ons from cente
centerr
– Switch
Switch activatio
activation
n when the
the P&T
P&T Control
Control Knob
Knob is
depressed
D Graphicc LCD
Graphi LCD TFT
TFT VGA
VGA 10.4"
10.4" NLT
NLT Disp
Display
lay wi
with
th Front
Front Pane
Panell
and LED Backlight
• Graph
Graphical
ical User
User Interfa
Interface
ce (GUI)
(GUI)
C J1 − Ethernet Connection (for internal use only) Q P13 − Cable to Nebulizer Valve (PN 160400)
D J2 − RS232 Connection R J14 − FFC Cable (PN 160355) to Pressure Sensor Assembly
(MSP160300)
E P25 − Cable from Power Supply for Fan Supply S P16 − FDC Cable (PN 160357) to Display
F P3 − Cab
Cable
le (PN
(PN 160
16053
539)
9) fro
from
m Powe
Powerr Supp
Supply
ly 24
24 VDC
VDC T J17 − FFC
FFC Cable
Cable (PN 160
160356
356)) to Key Pa
Panel
nel for P&
P&TT and
and Key
Key
(MSP160600) Panel
G J4 − FFC Cable (PN 160351) to Battery Data U J18 − FFC Cable (PN 160356) to Key Panel for Backlight,
I P6 − Cable (PN 160370) to Battery Power W P20 − Cable (PN 160354) to Oxygen Sensor (PN 396200)
J P60 − Opti
Options
ons Slo
Slott for
for Commu
Communic
nicati
ation
on Boa
Board
rd (for
(for CO2
CO2,, X J21 − FFC
FFC Cable
Cable (PN 160
160353
353)) to Blow
Blower
er Modu
Module
le Temp
Tempera
era--
(SpO2 not implemented in HAMILTON-C2) and Nurse Call ture Sensor
Options)
L J9 − FFC Cable (PN 160347) to Filter Pressure Board Z P23 − Cable to Blower Module (MSP160250)
M J8 − FFC
FFC Cable
Cable (PN 1603
160359)
59) to
to Pressu
Pressure
re Senso
Sensorr Board
Board Binar
Binaryy AA P28
P28 − Cab
Cable
le to
to Fan
Fan 12V
12V (PN
(PN 1603
160346
46))
Valves
N P10 − Cable
Cable (PN 160
160387
387)) to QO2 Flo
Flow
w Senso
Sensorr (TSI,
(TSI,
MSP399124 or Sensirion, MSP161657)
52 English | 624165/07
Mainboard MSP160382 4
C J1 − Ethernet Connection (for internal use only) Q J14 − FFC Cable (PN 160355) to Pressure Sensor Assem-
bly (MSP160300)
E P3 − Cab
Cable
le (P
(PN
N 16
1605
0539
39)) fro
from
m Po
Powe
werr Sup
Suppl
plyy 24
24VD
VDC
C S P16
P16 − FD
FDC
C Ca
Cabl
blee (P
(PN
N 160
16038
386)
6) to Di
Disp
spla
layy
(MSP160600)
F J4 − FFC Cable (PN 160351) to Battery Data T J17 − FFC Cable (PN 160356) to Key Panel for P&T
Encoder and Key Panel
G P5 − Cable (PN 160372) to DC Input U J18 − FFC Cable (PN 160356) to Key Panel for Backlight,
Speaker and Microphone
H P6 − Cable (PN 160370) to Battery Power V P19 − Cable (PN 160387) to Qvent Flow Sensor (TSI,
MSP399123 or Sensirion, MSP161657)
I P60 − Op
P60 Optition
onss Slo
Slott for
for Co
Comm
mmun
unic
icat
atio
ion
n Boa
Board
rd (f
(for
or CO
CO2,
2, W P20
P20 − Cab
Cable
le (PN
(PN 16
1603
0354
54)) to Oxy
Oxyge
gen
n Sens
Sensor
or (PN
(PN 396
39620
200)
0)
SpO2 (Not Implemented in HAMILTON-C2) and Nurse
Call Options)
J P7 − Cable to O2 Proportional Valve X J21 − FFC Cable (PN 160353) to Blower Module Temper-
ature Sensor Board
K J9 − FFC Cable (PN 160347) to Filter Pressure Board Y J22 − FFC Cable to Inspiratory Valve (MSP160230)
L J8 − FFC
FFC Ca
Cabl
blee (PN
(PN 16
1603
0359
59)) to
to Pre
Press
ssur
uree Sen
Senso
sorr Boa
Board
rd Z J23
J23 − Ca
Cabl
blee to
to Blo
Blowe
werr Mod
Modul
ulee (MS
(MSP1
P160
6025
250)
0)
Binary Valves
M P10 − Ca
P10 Cabl
blee (PN
(PN 16
1603
0387
87)) to
to QO2
QO2 Fl
Flow
ow Se
Sens
nsor
or (T
(TSI
SI,, AA P28
P28 − Ca
Cabl
blee to
to Fan
Fan 12
12V
V (PN
(PN 16
1603
0346
46))
MSP399124 or Sensirion, MSP161657)
N J11
J11 − FF
FFC
C Cab
Cable
le to Ex
Expi
pira
rato
tory
ry Va
Valv
lvee (MS
(MSP1
P160
6024
240)
0)
Mainboard Serves as the Carrier Board for the Processor Board, Power Management System
and a Connector Board for all Sensors, Valves and Periphery Components
Controls • Blower
• ON
ON/O
/OFF
FF Ke
Keyy
• Ha
Hard
rd Key
Key and
and Lamps
Lamps
• Ke
Keyy Pan
Panel
el
• TF
TFTT Dis
Displ
play
ay
• Amb
Ambien
ientt Val
Valve
ve
• Exp
Expira
irator
toryy Valve
Valve
• Mi
Mixe
xerr Valv
Valves
es
• Ne
Nebul
bulize
izerr Valve
Valve
• Emer
Emergency
gency Statu
Statuss Indicato
Indicatorr
Interf
Interface
acess with
with the foll
followi
owing
ng Venti
Ventilat
lation
ion Unit
Unit boar
boards:
ds: • Pow
Power
er Suppl
Supplyy
• Pres
Pressure
sure Sens
Sensor
or Assemb
Assembly
ly
• Fi
Filt
lter
er Boa
Board
rd
• Sm
Smart
art Bat
Batter
teries
ies
Sensor Monitoring: • Bl
B lower Temperature
• Instr
Instrumen
umentt Temp
Temperatu
erature
re
• Pve
Pvent_
nt_con
contro
troll
• Pve
Pvent_
nt_mon
monito
itorr
• Pf
Pflo
lows
wsen
enso
sorr
• Pa
Pamb
mbie
ient
nt
• Qve
Qvent
nt Flow
Flow Sen
Sensor
sor
• QO2 Flo
Flow
w Sens
Sensor
or
• Paw
• Oxy
Oxygen
gen Se
Senso
nsorr
Inpu
Inputt Vol
Volta
tage
ge su
supp
pplilied
ed fro
from
m the
the Po
Powe
werr Sup
Suppl
ply:
y: • 24 VD
VDC
C ±5%
±5% 6.
6.6A
6A/1
/150
50W
W (R
(Ran
ange
ge 22
22.8
.8V.
V...
..25
25.2
.2V)
V)
54 English | 624165/07
Mainboard Components 4
Output Voltages created by Switching Regulators on 3.3 VDC • Qvent Flow Sensor
the Ventilation Unit Mainboard: • QO2 Flo
Flow
w Sens
Sensor
or
• Pre
Pressu
ssure
re Senso
Sensors
rs
• P&
P&TT Contr
Control
ol Knob
Knob
• ES
ESM
M Mo
Modu
dule
le
• Blow
Blower
er Tempera
Temperature
ture Senso
Sensorr
• HEPA Filte
Filterr Temperature
Temperature Sensor
Sensor
+24V_PS (from power 24 VDC 22.8 - 25.2 VDC 7A at 22.8 V; 5.7 at 25.2 150W
supply)
+24V_DcIn (from 20 V 12-24 VDC 11.0 - 27.0 VDC 9.1A at 11 V; 5A at 20V 100W
100W DC Input)
7.5A at >20 V; 5.6A at 150W
27V (Battery starts charg-
ing from DC Input if volt-
age is above 20 V)
Battery 1 (Primary Bat- 14.4 VDC 12.0 - 16.8 VDC 6A at 16.8 V; 8.3A at 100W
tery) 12V
Battery 2 (Optional Bat- 14.4 VDC 12.0 - 16.8 VDC 6A at 16.8 V; 8.3A at 100W
tery) 12V
Signal Name Voltage Level Voltage Range Pin P in Signal Name Voltage Level Voltage Range
GND o o GND
+5V_Usb 5.0 VDC 4.75 - 5.25 V o o +24V_DcIn 12-24 VDC 11.0 - 27.0 VDC
+5V 5.0 VDC 4.60 - 5.40 V o o +24V_PS 24 VDC 22.8 - 25.2 VDC
+3V3_Ooc 3.3 VDC 3.00 - 4.20 V o o +14V4_Bat2 14.4 VDC 12.0 - 16.8 VDC
+3V_Ref 3.0 VDC 2.91 - 3.09 V o o +14V4_Bat1 14.4 VDC 12.0 - 16.8 VDC
GND o 1 GND
P44
English | 624165/07
56
Pin Signal Name Voltage Level Voltage Range Pin Signal Name Voltage Level Voltage Range
o GND o GND
o SoundLevel 1.15 (if silent) 1.11 - 1.19 o TO2 1.25 (measured at 1.15 - 1.35
25°C, when the
oxygen sensor is
disconnected)
o GND 1 GND
P42 P41
– 10
100
0 to
to 240
240 VA
VAC
C
– 50 to 60 Hz
• Out
Output
put Vol
Voltag
tage:
e:
– +2
+24
4 VD
VDC ±5%
±5%
B Powe
Powerr Inl
nlet
et Con
onn
nec
ecto
torr
• Cabl
Cablee from AC Power
Power Inlet
Inlet
5
Lithium Ion Battery
5.1 Rechargeable
Recharge able Lithiu
Lithiumm Ion Batte
Batteryry ......
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
........ 60
5.2 Battery
Batte ry Char
Charger
ger / Calibr
Calibrator
ator .....
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
..........
.... 64
5.3 Evaluatio
Evaluationn of the Batte
Batteryry State
State of Health
Health......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
........ 66
5.4 Battery
Batte ry Re-cal
Re-calibrat
ibration
ion ......
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
........ 68
5.5 Lithium
Lithium Ion Batt
Battery
ery Hand
Handling
ling Prec
Precautio
autions ns ......
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...... 73
59
WARNING
It is mandatory that the HAMILTON-C2 is operated with at least one battery installed.
NOTICE
5.1.1 Introduction
Two backup batteries, one standard and one optional, are used the HAMILTON-C2 when the primary power source is low or
unavailable.
When the primary power source (either AC mains power or a DC power supply) fails, the ventilator automatically
automatically switches to
backup battery operation with no interruption in ventilation. An alarm sounds to signal the switchover. To reset the alarm,
silence the alarm to confirm notification of the power system change.
If the optional battery (battery 2) is available and adequately charged, the ventilator switches to this battery first. When bat-
tery 2 is depleted or unavailable, the ventilator switches to the standard battery (battery 1).
The batteries power the ventilator until the primary power source is adequate or the battery is depleted.
As a further safeguard, the HAMILTON-C2 provides a low battery alarm. It also has a capacitor-drive
capacitor-driven
n backup buzzer that
sounds continuously for at least 2 minutes when battery power is completely lost.
The ventilator recharges the batteries whenever the ventilator is connected to either AC or >20VDC, with or without the
ventilator power switched on. The battery charge indicator (see Figure 5-1 Power Source Symbols and Battery Charge Indica-
tor) lights to show that the batteries are being charged.
The power source symbols in the bottom right-hand corner of the screen show the available power sources. A frame around
a symbol indicates the current ventilator power source. Green indicates the level of battery charge.
60 English | 624165/07
Introduction 5
WARNING
When a battery has reached its end-of-life, the capacity is reduced and the battery weakens. The use of a ventilator with
weak batteries can cause the device to report incorrect fuel gauge information, leading to a possible unexpected ventilation
stop if a power failure occurs.
The batteries are hot-swappable; that is, they can be replaced while the ventilator is operating. Replace with a newly charged
battery. For battery replacement refer to the Backup Battery Pack Removal/Assembly Section 10.4.1
10.4.1..
CAUTION
It is mandatory that the HAMILTON-C2 is operated with at least one battery installed.
NOTICE
- The backup battery is intended for short-term use only. It is not intended to be a primary power source.
- Hamilton Medical recommends that the ventilators batteries be fully charged before you ventilate a patient. If the batteries
are not fully charged and the AC power fails, always pay close attention to the level of battery charge.
5.1.6.3 Disposal
• Regulations vary country by country.
country. Dispose of the batteries
batteries in accordance
accordance with
with local regulations.
regulations.
NOTICE
When the battery reaches its end-of-life, the capacity is reduced and the battery becomes unreliable. The use of a ventilator
with such batteries may cause wrong fuel gauge information leading to an unexpected ventilation stop when a power fail
occurs!
Hamilton Medical | HAMILTON-C2 Service Manual 63
WARNING
• Do not expose the charger or power
power supply to water or conductive liquids, this is not a sealed case.
case.
• Do not open the charger
charger or power supply case. There are no user serviceable
serviceable parts inside.
• Do not cover the fan exhaust or obstruct
obstruct the airflow, as this will cause overheating.
overheating. Use only the manufacturer’s
manufacturer’s 24V
2.5A power supply and observe terminal polarity
CAUTION
During re-calibration, the battery connector and base of the charger may become warm.
The Battery Charger / Calibrator (PN 369104, see Figure 5-2 Battery Charger / Calibrator, please make sure you are using a
charger with Revision 07 or higher) is a standalone desktop battery charger with the added ability to re-calibrate the fuel
gauge on smart batteries. It works with all Hamilton Medical smart batteries.
Package contents:
• One deskto
desktop
p Battery
Battery Charger
Charger / Calibrat
Calibrator
or (PN 369104)
369104)
• Three plastic
plastic spacers to modify the battery recess
recess to suit your battery size
• One 24V 2.5A
2.5A AC:DC
AC:DC Power
Power Supply,
Supply, universa
universall mains input
input
• One
One ma
main
inss ca
cabl
blee
– European chargers
chargers (CH5000E) are packed
packed with a European 2-pin Mains Power
Power Cord
– UK chargers
chargers (CH5000U) are packed
packed with a UK 3-pin Mains Power Cord
– North American
American chargers (CH5000A) are packed with
with a US 3-pin Mains Power
Power Cord
• In
Inst
stru
ruct
ctio
ion
n manua
manuall
NOTICE
Chargers shipped with medical grade power supplies carry the suffix “M” – eg. CH5000AM.
64 English | 624165/07
NOTICE
Full
Full Ch
Char
arge
ge Capa
paci
city
ty FC
FCC
C [mA
[mAh]
h] See th
thee te
tech
chni
nica
call sta
state
te or re
read
ad th
thee Go into service software, technical state, the value is
value in the exported and printed shown on the bottom.
instrument report.
Calendar age AGE [Y] See printed on battery. The battery’s date of manufacture is stamped on top
of the battery connector (IEyywwvv). A coding of e.g.
IE144514 means Year 2014, Week 45, Version 1.4,
SN 67, 6.8Ah
Cycles CYC [1] See the technical state or read the Go into service software, technical
t echnical state, value shown
value in the exported and printed on the bottom.
instrument report.
Error ERR (%) See the technical state or read the Go into service software, technical state, the value is
value in the exported and printed shown on the bottom.
instrument report.
To recalibrate the battery, press the calibrate button on the front of the charger.
The blue calibration LED will flash to indicate that the battery is undergoing the re-calibration cycle. There may be a short
delay before the calibration begins. During calibration the discharge resistors will heat up and the fan will operate to main-
tain the temperature within acceptable limits.
At the end of the procedure the blue calibration LED will illuminate indicating a fully charged, fully calibrated battery.
NOTICE
No action is necessary if only a recharge is required. The charger will automatically begin to charge the battery.
The most common cause of calibration failure is overheating of the battery during discharge. Keep the charger away from
direct sunlight or heat sources.
CAUTION
During recalibration, the battery connector and base of the charger may become warm.
68 English | 624165/07
General Information 5
NOTICE
No additional action is required for re-charging the battery. The process starts automatically.
The blue calibration LED will flash to indicate the re-calibration cycle. There may be a short delay before the calibration
begins. During calibration the discharge
di scharge resistors will heat up and the fan will operate to maintain temperature within accept-
able limits.
At the end of this procedure the blue LED will stay constant indicating a fully charged, fully calibrated battery.
CAUTION
The most common cause of calibration failure is overheating of the battery during discharge. Please keep the charger away
from direct sunlight or heat sources.
Hamilton Medical | HAMILTON-C2 Service Manual 69
Calibration begins each time the re-calibration button is pressed. Do not press the re-calibration button part way through the
re-calibration cycle.
The re-calibration cycle is as follows:
1. Dis
Discha
charge
rge of
of any res
residu
idual
al capa
capacit
city.
y.
2. A calibra
calibration
tion charge
charge is deliver
delivered
ed to
to the batt
battery.
ery.
3. Dis
Discha
charge
rge of
of calibr
calibrati
ation
on capac
capacity
ity..
4. The batt
battery
ery is give
given
n a regul
regular
ar char
charge.
ge.
Re-calibration Description 5
The battery can directly display the capacity information. The battery capacity is displayed as the RSoC (Relative State of
Charge). Each LED segment represents 25 percent of the full charge capacity. The LED pattern definition is given in the table
below. The LED’s illuminate for 4 seconds following switch activation. If the battery voltage is too low, there will be no LED
indication.
1 2 3 4
WARNING
• Before using lithium ion batteries
batteries for the first time, carefully study this section, including
including all Cautions and Warnings.
Warnings.
• Keep this
this section
section for future
future reference
reference..
WARNING
• Do not use the battery for a purpose other than those specified.
specified. Otherwise,
Otherwise, its performance will be compromised
compromised and/or
its service life will be shortened. Depending on the equipment in which the battery is used, excessively high current can
flow through the battery, possibly damaging it and leading to acid leakage, overheating, smoke emission, bursting and/
or ignition.
• Do not use the battery in combination with primary
primary batteries (such as dry-cell
dry-cell batteries) or batteries
batteries of different capaci-
ties or brands. Otherwise, the battery can be overdischarged during use, or overcharged during recharging, abnormal
chemical reactions may occur, possibly leading to acid leakage, overheating,
overheating, smoke emission, bursting and/or ignition.
• Do not use or leave the battery near
near a heat source such as a fire or a heater
heater (+80ºC or higher). If the resin separator
should be damaged due to overheating, internal short-circuiting may occur to the battery, possibly leading to acid leak-
age, smoke emission, bursting and/or ignition.
• Do not put the battery
battery into a microwave
microwave oven or pressurized
pressurized container. Rapid
Rapid heating or disrupted sealing
sealing can lead to
acid leakage, overheating, smoke emission, bursting and/or ignition.
• Do not discard the battery into fire or heat it. Otherwise,
Otherwise, its insulation can melt down, its gas release
release vent or safety fea-
fea-
tures will be damaged and/or its electrolyte can ignite, possibly leading to acid leakage, overheating, smoke emission,
bursting and/or ignition on it.
high current and voltage and, as a result, abnormal chemical reactions can occur in it, possibly leading to acid leakage,
overheating, smoke emission, bursting and/or ignition.
• To recharge the battery,
battery, use the battery charger
charger specifically designed for the purpose
purpose and observe the recharging
recharging condi-
tions, refer to 5.2
5.2.. A recharging operation under non-conforming recharging conditions (higher temperature and larger
voltage/current than specified, modified battery charger, etc.) can cause the battery to be overcharged, or charged with
extremely high current. An abnormal chemical reaction can occur in the battery, possibly leading to acid leakage, over-
heating, smoke emission, bursting and/or ignition.
• If recharging operation fails to complete
complete even when a specified
specified recharging time has elapsed,
elapsed, immediately stop further
further
recharging. Otherwise, acid leakage, overheating,
overheating, smoke emission, bursting and/or ignition can occur.
CAUTION
• If you find rust, a bad odor, overheating and/or
and/or other irregularities when using the battery for the first time, return
return it to
Hamilton Medical AG.
• If acid leaking from the battery comes into contact with your skin or clothing, immediately wash it away with running
water. Otherwise, skin inflammation can occur.
• The battery incorporates built-in
built-in safety devices. Do not use it in a location where static electricity (greater than the manu-
facturer’s guarantee) may be present. Otherwise, the safety devices can be damaged, possibly leading to acid leakage,
overheating, smoke emission, bursting and/or ignition.
• Do not use or subject the battery to intense
intense sun or above 65°C. Otherwise, acid leakage, overheating and/or smoke
emission can occur. Also, its guaranteed performance will be lost and/or its service life will be shortened.
• The guaranteed recharging temperature
temperature range is 0°C to +45ºC. A recharging operation
operation outside this temperature
temperature range
can lead to acid leakage and/or overheating of the battery and may cause damage to it.
• Store the battery in a location where children cannot reach
reach it. Also, make sure that no child takes the
the battery out of the
the
battery charger or equipment.
74 English | 624165/07
6
Preventive Maintenance and Testing
6.1 Overview........
Overview.. ...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
........ 76
6.2 Hospital
Hospital Preve
Preventive
ntive Main
Maintena
tenance........
nce..............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...... 78
6.3 Engineer
Engineer Prev
Preventiv
entivee Mainten
Maintenance......
ance............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...... 79
75
6.1 Overview
WARNING
− To prevent disease transmission, you must use personal protective equipment when handling contaminated bacterial
bacterial fil-
ters or other patient accessories. Refer to the HAMILTON-C2 Operator’s
Operator’s Manual for instructions on sterilizing patient system
parts.
− Device is potentially contaminated
− The device must be cleaned and disinfected to prevent the spread of infections and germs
CAUTION
Make sure to take full ESD (ElectroStatic Discharge) precautions before handling any electronic parts, or before opening the
HAMILTON-C2. For more information refer to:
− Electrical Safety Overview Section 7.1
− Standard Tools Section 11.2
− Special Tools Section 11.3
− Test Equipment Section 11.4
− You must complete a service training course for the HAMILTON-C2 with Hamilton Medical before undertaking the main-
tenance and testing procedures described in this manual.
6.1.1 Introduction
All Preventive Maintenance and Testing must be performed:
• Aft
After
er repla
replacin
cing
g any comp
compone
onent
nt
• Once a year or once every
every 5000 operating hours,
hours, whichever
whichever comes first
To perform Preventive Maintenance,
Maintenance, perform all the steps shown in the Overview of Preventive Maintenance
Maintenance and Testing Sec-
tion 6.1.4
6.1.4..
• Blow
Blowerer Module
Module (MSP1
(MSP160250
60250))
• Lithi
Lithium
um Ion Battery
Battery Pack(s)
Pack(s) (MSP369106)
(MSP369106)
• Battery Charger / Calibrator
Calibrator (PN 369104,
369104, Rev 07 or higher)
higher)
• Oxyge
Oxygen
n Sensor
Sensor (PN 396200
396200))
• HPO Inlet
Inlet Filter
Filter Kit
Kit (PN 160497)
160497)
• Complete Patient
Patient Breathing
Breathing Circuit (Adult, PN 260086):
– Test Lung
Lung with ET-Tube
ET-Tube (PN 151815)
151815) and Adapte
Adapterr (PN 281420)
281420)
– Ins
Inspir
pirato
atory
ry Filter
Filter (PN
(PN 279204)
279204)
– Short Sili
Silicone
cone Tube (PN 26010
260100)
0)
NOTICE! For more details of parts, refer to the Spare Parts Section
12..
12
NOTICE! For more details of Test Equipment, refer to the Test Equipment Section
11.4..
11.4
1. Perfor
Performm (or
(or co
conf
nfir
irm
m itit has
has be
been
en pe
perf
rfor
orme
med)
d) th
thee Hos
Hos-- Hospital Preventive Maintenance Sec- 5-15 min.
pital Preventive Maintenance tion 6.2
2. Perf
rfo
orm the Engineer Preventive Maintenance Engineer Preventive Maintenance Sec- 10 min.
tion 6.3
4. Perform the Electrical Safety Tests Electrical Safety Tests Section 7.2 10 min.
6. Finish
Finish th
thee te
test
stin
ing
g by co
comp
mple
leti
ting
ng th
thee ta
task
skss do
docu
cu-- The HAMILTON-C2 Operator’s Manual 10 min.
mented in the Tests, Calibrations and Utilities Section or local-language equivalent
of the HAMILTON-C2 Operator’s Manual
Hamilton Medical | HAMILTON-C2 Service Manual 77
• Perfo
Perform
rm the
the tasks
tasks you
yourse
rself
lf
• Tra
Train
in staff
staff how to perform
perform these task
taskss
Between patients and according to hos- Patient Breathing Circuit (including Replace with sterilized or new single use parts. Run
pital policy Mask, Inspiratory Filter, Flow Sensor, the Tightness Test and the Flow Sensor Calibration as
Nebulizer Jar, Expiratory Valve and shown in the HAMILTON-C2 Operator’s Manual.
Membrane)
Every 2 days or according to hospital Patient Breathing Circuit Empty any water from breathing tubes or water
policy traps. Inspect parts for damage. Replace as necessary.
Every month (or more often, if Air Intake Dust Filter and Fan Filter Set Check for dust and lint. If needed, clean or replace as
WARNING
This section is not a stand-alone, independent part of the manual. Perform the tasks detailed here only as a part of, and as
instructed. Refer to the Preventive Maintenance and Testing Overview Section 6.1 and Hospital Preventive Maintenance Sec-
tion 6.2
6.2..
Pre-
Pre-Op
Oper
erat
atio
iona
nall Che
Check
ckss Perfor
Perform
m the
the Pr
Pre-
e-op
oper
erat
atio
iona
nall Che
Check
ckss as
as sho
shown
wn in th
thee HAM
HAMIL
ILTO
TON-
N-C2
C2
Operator’s Manual > Running the Pre-operational Check.
Alarm Tests Perform the alarm checks as shown in the HAMILTON-C2 Opera-
tor’s Manual > Alarm Tests.
HPO Inlet Filter Replace as described in the High Pressure Oxygen Connector
Removal/Assembly Section 10.4.4
Removal/Assembly 10.4.4..
Oxygen Sensor Replace if depleted as shown in the HAMILTON-C2 Operator’s
Manual Section > Replacing the Oxygen Sensor.
Eve
very
ry 5 ye
years (3
(30,
0,00
000
0 Hou
Hours
rs)) LCD
LCD Dis
ispl
plaay Bac
Backklig
ight
ht Replac
Repl acee th
the LC
LCD Di
Disp
spla
layy Ba
Back
ckli
ligh
ghtt as
as de
desc
scri
ribe
bed
d (R
(Ref
efeer to
to LC
LCD Bac
Back-
k-
light Removal/Assembly Section 10.3.8
10.3.8).).
When ebt or blower timer is ≥ Blower Module Replace the Blower Module when the Blower Timer has reached
100%. 100%(dynamic lifetime surveillance,
surveillance, typically after 8 years). Also
replace
the nextthe Blower reaches
6 months Module 100%
when the prediction
(Refer of Module
to Blower Blower Timer in
Removal/Assembly
Removal/Assem bly Section 10.4.10
10.4.10).
).
Hamilton Medical | HAMILTON-C2 Service Manual 79
7
Electrical Safety
7 Electrical Safety
WARNING
Electrical Safety Tests detailed
detailed in this section must be performed as part of, or as instructed in the Overview of Preventive
Maintenance and Testing Section 6.1.46.1.4..
In addition, to comply with IEC 62353, the Electrical Safety Test's must be performed after:
− The Power Supply is replaced
− The Mainboard is replaced
− Removing any ground contact from the Ventilator
− When performing Preventive Maintenance
− Any repair
− Installation
7.2.1
• Mul
ulti
tim
Preparation
mete
terr
for Tests
• Sa
Safe
fety
ty An
Anal
alyz
yzer
er
• Ele
Electr
ctrica
icall Safety
Safety Test
Test Cabl
Cablee
7 Electrical Safety
7 Electrical Safety
Specifications
Table 7-2. IEC 62353 Specifications for Class II Type B and BF Devices
Setup Setup for Type B Device (no Communi- Setup for Type BF Device (with Com-
cation Board Installed) Section 7.2.5.1 munication Board for Capnography)
Section 7.2.5.2
* refer to IEC62353
Applied Parts of the medical device, which are designed to come into physical contact with the patient or parts that are likely
to be brought into contact with the patient.
Type B Applied Parts
Type B Applied Parts are those parts, which are usually Earth reference
referenced.
d. Type B is used for applied parts that are generally
not conductive and can be immediately released
released from the patient.
Type BF Applied Parts
Type BF Applied Parts are generally for devices that have conductive contact with the patient, or having medium or long term
contact with the patient. For example, using capnostat and/or SPO2 finger probe on a patient. This applied part complies
with a higher degree of protection against electric shock than Type B applied part.
Hamilton Medical | HAMILTON-C2 Service Manual 87
7 Electrical Safety
NOTICE
For detailed information regarding the use of your Electrical Safety Tester, please check the Users Manual for the device.
Figure 7-7. Setup for Type B Device (no Communication Board Installed)
A AC Power Ca
Cable In
Inlet (R
(RIGEL 288 Safety Analyzer) E Electrical Safety Te
Test Ca
Cable (PN 161618)
B RIGEL 288
RIGEL 288 Safet
Safetyy Analy
Analyzer
zer (Han
(Handhe
dheld
ld Medic
Medical
al Elec
Electri
trical
cal F Pow
oweer Co
Cord (H
(HAM
AMIILT
LTON
ON-C
-C2)
2)
Safety Tester)
C 4mm Ear
Earth
th Bond
Bond Pro
Probe
be Sock
Socket
et (gre
(green)
en) (RI
(RIGEL
GEL 288 Saf
Safety
ety G USB Port
Analyzer)
D EUTT Sock
EU Socket
et (RI
(RIGE
GELL 288
288 Safe
Safety
ty Ana
Analy
lyze
zer)
r)
88 English | 624165/07
NOTICE
For detailed information regarding the use of your Electrical Safety Tester, please check the Users Manual for the device.
2. Connect
Type the (with
BF Device RIGELCommunication
288 Safety Analyzer
Boardand
forthe HAMILTON-C2 to each other as shown below (see Figure 7-8 Setup for
Capnography).
3. Turn on the HAMILTON-C2 (Standby mode) for the electrical safety tests.
Figure 7-8. Setup for Type BF Device (with Installed Communication Board for Capnography)
B RIGEL 288
RIGEL 288 Safet
Safetyy Analyz
Analyzer
er (Hand
(Handhel
held
d Medic
Medical
al Elec
Electri
trical
cal H Power Plug
Safety Tester)
C 4 mm Eart
Earth
h Bond
Bond Prob
Probee Socke
Sockett (green
(green)) (RIGEL
(RIGEL 288 Sa
Safet
fetyy I USB Port
Analyzer)
D EUT Socket (RIGEL 288 Safety Analyzer) J CO2 Sensor Electrical Safety Test Cable (PN 159171)
F Electrical Safety Test Cable (PN 161618) L AP-Box 331A700 (RIGEL 288 Safety Analyzer)
Hamilton Medical | HAMILTON-C2 Service Manual 89
7 Electrical Safety
90 English | 624165/07
8
Service Software
8.1 Introduction
Introductio n .....
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
...........
..... 92
8.2 Functions
Funct ions of the
the Service
Service Soft
Software..
ware........
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
...........
..... 93
8.3 Structure
Struc ture of the
the Servi
Servicece Softw
Softwareare ......
............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...... 94
8.4 Starting
Starting the Serv
Serviceice Soft
Software
ware .....
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
........ 95
8.5 Service
Serv ice Softw
Softwareare Scre
Screenen Layout...........
Layout.................
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
...........
..... 96
8.6 Making
Maki ng Scree
Screenshot
nshotss ......
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
..........
.... 97
8.7 Ventilator Info..........
Ventilator Info................
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
.........
... 98
8.8 Technica
Tech nicall State...............
State.....................
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
.........
... 99
8.9 Service
Serv ice Time
Timer.......
r.............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
........105
8.10 Real Time ClocClockk (RTC) ......
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
......108
108
8.11 Battery Data.........................................................................................................................................
Data.........................................................................................................................................109 109
8.12 Tests Overview
Overview..................................................................................................................
.....................................................................................................................................
...................110
110
8.13 Testss / Calib
Test Calibratio
ration n Scre
Screenen ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
........111
8.14 Adjustments / Calibration.....................................................................................................................
Calibration.....................................................................................................................112 112
8.15 Component Tests.................................................................................................................................
Tests.................................................................................................................................142 142
8.16 System Test..........................................................................................................................................
Test..........................................................................................................................................194194
8.17 Sensor
Sens or Data .....
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
...........
.....207
207
8.18 Log/Config
Log/C onfig Files ......
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
..........
....212
212
8.19 Software
Softw are Upda
Updates tes ......
............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
......218
218
8.20 Pre-opera
Pre-operationa
tionall Chec
Checks ks ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
...........
.....226
226
8.21 General
Gene ral Tes
Teststs .....
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
.........
...227
227
8.22 CO2 Sens
Sensor or Accu
Accuracyracy Chec
Checkk .....
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
.........
...230
230
8.23 Nurse Call Functional Check.................................................................................................................
Check.................................................................................................................240 240
8.24 Final Test
Testss ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
........242
8.25 Set the Service Timer............................................................................................................................
Timer............................................................................................................................242 242
8.26 Checkk the Batt
Chec Battery
ery Data ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
...........
.....242
242
91
8 Service Software
8.1 Introduction
WARNING
− Read the Preventive Maintenance
Maintenance and Testing Overview Section 6.1 6.1,, before performing any of the tests in this section.
− If one of the tests indicates that you must replace a part, do so immediately and update the Service Software
Software Modify Tab
(see the Modify Tab Section 8.8.3
8.8.3)) and then repeat the complete series of tests. See contents in this section.
CAUTION
To prevent patient or ventilator contamination, always use a bacterial filter between the HAMILTON-C2 and the Inspiratory
Limb of the Patient Breathing Circuit.
− For troubleshooting see knowledge base
− Record all results on the HAMILTON-C2 Summary of Test Report (see the Summary of Test Report Section 17.1)
− Confirm that the Technical State is updated and that the device has been restarted after Technical State modifications
(see the Modify Tab Section 8.8.3). Always use an Adult Breathing Circuit, if a circuit is required during the tests.
tests.
NOTICE
This section describes each of the units containing the HAMILTON-C2 Service Software.
Software.
Before starting, be sure that you are familiar with Typographic Conventions Section 1.2 and Expressions Section 1.3
1.3..
The HAMILTON-C2 Summary
Summary of Test Report is the standard form to be used and must be completed each time the Service
Software is performed. If you do not have a suitable form, you can photocopy and use the form named HAMILTON-C2 Sum-
mary of Test Report (see the Summary of Test Report Section 17.1
17.1)) at the back of this manual.
92 English | 624165/07
8 Service Software
8.3 Structure of
of the Service Software
Software
Figure 8-1. Service Software Structure
94 English | 624165/07
To start the Service Software, you must place the HAMILTON-C2 into the Service Software Mode.
8 Service Software
3b..
3b Opti
Option
onal
al Me
Menu
nu Ta
Tab
b Lay
Layer
er 2
96 English | 624165/07
Making Screenshots 8
The print screen function saves a JPG file of the current ventilator screen to a USB Memory Stick (B).
The USB Memory Stick can be removed when the Print Screen Key LED is no longer lit.
NOTICE
8 Service Software
Technical State 8
8 Service Software
• Flo
Flow
w Sens
Sensor
or AIR
AIR (Qv
(Qvent
ent))
• Fl
Flow
ow Sen
Senso
sorr O2 (QO
(QO2)
2)
• Ox
Oxyg
ygen
en Se
Sens
nsor
or
• Co
Comm
mmun
unic
icat
atio
ion
n Bo
Boar
ard
d
3. In case the technical state is corrupt, or after upgrading from SW version < 2.1.2, the Clean-up Button will appear.
4. In case the technical state cannot be read out from the EEprom of a component, the unit will fail in self-test and alarm
with “Technical state failed”. Other Technical faults will appear as after effect.
5. Record the Hardware Version Information on the HAMILTON-C2 Summary
Summary of Test Report Section 17.1
17.1..
100 English | 624165/07
2. Record the Software Version Information on the HAMILTON-C2 Summary of Test Report Section 17.1
17.1..
Hamilton Medical | HAMILTON-C2 Service Manual 101
8 Service Software
Always update the technical state when an assembly as listed below has been replaced.
Restart the device prior to performing the service software tests and calibrations.
The Modify Tab allows updating information when a part has been replaced.
Modify Tab 8
NOTICE
Make sure to select the correct Part Number. Cross check with the label of the Spare Part / Installation Guide of the MSP
Spare Part.
Hamilton Medical | HAMILTON-C2 Service Manual 103
8 Service Software
To allow you to export/import the technical state of the device from a USB Memory Stick.
By exporting, the service software creates a directory, if it does not exist, on the USB Memory Stick.
The naming convention of this directory is:
INSTRUMENT-sn#device_serial_number:
INSTRUMENT-sn#device_se rial_number: /myUSBDevice/C2-sn-1010/
/myUSBDevice/C2-sn-1010/ the file which is exported is a .csv (comma separated file)
file containing the device technical state.
It has the same naming convention as the directory created on the USB Memory Stick:
e.g.: C2-sn-1010.csv
For importing values to the device, be sure that there is a directory following the previous given naming convention and
a .csv file having the proper name according to the device serial number.
The technical state can be exported and imported to/from a USB Memory Stick. Ensure the USB Memory Stick is connected to
the ventilator, then touch the corresponding button.
NOTICE
Insert USB Memory Stick is visible, as long as there is no memory stick inserted.
NOTICE
The format of the file is .csv (comma separated values) and can be edited with a text editor (e.g. Notepad, Wordpad). Do
not use Excel to modify the .csv file.
104 English | 624165/07
Service Timer 8
3. Record the Operating hours and the Service timer hours on the HAMILTON-C2 Summary of Test Report Section 17.1
17.1..
4. After successful service, reset the service counter by touching the Reset Button. To enable the reset function, select the
Alarm limit Timer first.
5. The Alarm limit Timer Button allows the set number of hours between service intervals to be changed.
8 Service Software
6. Touch the Alarm limit Timer Button or rotate the P&T Control Knob until the Alarm Limit Rotary Screen Button is high-
lighted, then press the P&T Control Knob.
7. The number of hours can be changed by rotating the P&T Control Knob.
8. Touch the Alarm limit Timer Button again or press the P&T Control Knob to save the new value.
9. Touch Reset.
10. After resetting the service timer, the current operating hours will be added to the set alarm limit hours.
106 English | 624165/07
The expected operation time (100%) is calculated based on the operating hours, temperature and rotation speed of the
blower. When 100% is reached, the alarm Blower service required appears and the Blower Module needs to be replaced.
Replace the Blower Module when the blower timer 100%. Also replace the Blower Module when the Prediction Blower
Timer in the Next 6 Months reaches 100%. See prediction calculation formula in the table below.
After replacement of the Blower Module, enter the Serial Number and Revision Number of the replacement blower in the
Technical State (see the Modify Tab 8.8.3
8.8.3).
). This will reset the Blower timer.
3. Record the Blower timer % on the HAMILTON-C2 Summary of Test Report Section 17.1
17.1..
Hamilton Medical | HAMILTON-C2 Service Manual 107
8 Service Software
2. Touch the Screen Button(s) that need to be changed, or rotate the P&T Control Knob until the desired button is high-
lighted, then press the P&T Control Knob.
3. Change the number displayed by rotating the P&T Control Knob.
4. Touch the Screen Button again or press the P&T Control Knob to keep the new value.
5. When all selections are complete, touch the Set Button.
Battery Data 8
8 Service Software
User Interface Tab Page No 2115 User Interface Tab Section 8.15.2
8 Service Software
Calibration Tab 8
8 Service Software
Adj/Calib > Calibration > Touch Screen Allows calibrating the coordinates of the touch controller (4–point • To
Touc
uch
h Scre
Screen
en
(Page No 2321) calibration).
Touch Screen Section 8.14.1.1
Adj/Calib > Calibration > Inspiratory Allows aligning the gain and offset for voltage-offset of the inspira- • Ins
Inspir
pirato
atory
ry Valve
Valve
Valve (Page No 2321) tory valve.
Inspiratory Valve Section 8.14.1.2 Sets Offset 1 (low pressure) at 20 mbar blower pressure and Offset
2 (high pressure) at 60 mbar blower pressure when required Qvent
Flow is reached.
Adj/Calib > Calibration > Pressure (Page Allows adjusting the pressure sensor gain using an external pressure Pressure Sensor:
No 2341) measurement as reference. • Pve
Pvent_
nt_con
contro
troll
Pressure Section 8.14.1.3 • Pvent_
Pvent_mon
monito
itorr and
Paw
Adj/Calib > Calibration > Exp. Valve Before calibration starts, an autozero of Pvent_control, Pvent_moni- • Exp
Expira
irator
toryy Valve
Valve
(Page No 2343) tor, Paw, Pflow-sensor and IexpValve is performed. If the autozero
Expiratory Valve Section 8.14.1.4 procedure fails, a failure will be displayed and calibration cannot be
started.
During calibration, gain and offset for the valve characteristics of
the expiratory valve will be aligned.
Adj/Calib > Calibration > O2 Cell (Page The Oxygen Sensor calibration is separated into two
t wo parts: • Oxy
Oxygen
gen Sen
Sensor
sor
No 2346) • Offset calibration. Calibrates the offset voltage of the amplifier
Oxygen Sensor Calibration Section circuit.
8.14.1.5 • Gain calibration. During this 2-min calibration of the Oxygen Sen-
sor, the ventilator delivers an increased oxygen concentration (if
oxygen is connected in the high pressure mode) or 21% oxygen
(if oxygen is connected in the low pressure mode or discon-
nected). It tests the Oxygen Sensor and resets the calibration
points specific to the Oxygen Sensor in use.
Adj/Calib > Calibration > Flow Sensor This calibration checks and resets the calibration point specific to • Pat
Patien
ientt Flow Sens
Sensor
or
(Page No 2347) the flow sensor in use.
Patient Flow Sensor Calibration Section The Patient Flow Sensor is calibrated in both directions.
8.14.1.6
114 English | 624165/07
Touch Screen 8
8 Service Software
4. You are instructed to touch the cross located on the top left of the screen with your finger.
5. Next, you are instructed to touch the cross located on the top right, then bottom right and then bottom left of the screen
with your finger.
6. Next, you are instructed to touch the Test Button in the middle of the screen with your finger.
Touch Screen 8
7. The Touch Screen Calibration is complete. Touch Apply to accept the new Touch Screen Calibration or Cancel.
8. The touch screen calibration is complete indicated by Touch screen calibration values saved. on the screen.
8 Service Software
NOTICE
− Confirm that the Technical State is updated with the current Inspiratory
I nspiratory Valve serial number AND revision number, refer
to Modify Tab 8.8.3
8.8.3..
− The HAMILTON-C2 needs a warm-up period. Make sure it was running for at least 20 minutes in the Ventilation Soft-
ware.
The Inspiratory Valve has a flow-voltage-hysteresis, in that the valve opening characteris
characteristic
tic is different from the valve closing
characteristic.
characteristic. The valve needs to be calibrated with either the opening process or the closing process. Based on the design
and pressure control mechanism, the Inspiratory valve is calibrated with the closing characteristic.
The required flow adjustment depends on the Inspiratory Valve revision:
• MS
MSP16
P16023
0230/0
0/00
0 –> 12 l/min
l/min (Qvent
(Qvent))
• MSP
MSP160230
160230/01
/01 and
and higher
higher –> 18 l/min
l/min (Qvent)
(Qvent)
The Technical State needs to contain the correct Inspiratory Valve revision to ensure proper control function
Inspiratory valve calibration:
Inspiratory Valve 8
5. The Inspiratory Valve Calibration is complete indicated by Inspiration valve calibration successful displayed on the screen.
Hamilton Medical | HAMILTON-C2 Service Manual 119
8 Service Software
8.14.1.3 Pressure
Allows adjusting the Pressure Sensor Gain using an external pressure measurement
measurement as a reference.
1. Touch the Pressure Button.
7. As soon as you touch the Start Button, the Ventilator does an autozero procedure to prevent wrong calibrations due to
remaining pressure inside the Patient Breathing Circuit.
Pressure 8
8. The Values on the screen should equal the value on the Pressure Gauge.
9. If adjustment is necessary, adjust the Gain by using the P&T Control Knob. Make the adjustment with the P&T Control
Knob, then, press the P&T Control Knob for the change in Gain to be applied.
10. After completion, touch the Save Button to save the changes.
NOTICE
If the test cannot be performed, check the whole setup for leaks. Furthermore, there should be almost no flow from the
expiration outlet. If you perform this test more than once, please close, then re-open the Pressure Calibration Window.
Hamilton Medical | HAMILTON-C2 Service Manual 121
8 Service Software
2. Connect a 30cm tube with a filter (PN 279204) as shown on the screen.
3. Touch the Start Button.
4. The Ventilator performs an autozero procedure to prevent wrong calibrations due to remaining pressure inside the limb.
122 English | 624165/07
Expiratory Valve 8
5. The calibration runs automatically indicated by the Pvent_monitor and Iexp. values changing during the calibration process.
6. The successful finish of the Expiratory Valve will be indicated with following message: Calibration successfully finished.
NOTICE
If the test takes more than 3 minutes or if calibration is not OK, replace the membrane and perform the test again.
Hamilton Medical | HAMILTON-C2 Service Manual 123
8 Service Software
NOTICE
With Mainboard (MSP160200) Revisions 1-5, perform the test as shown below.
With Mainboard (MSP160200) Revisions 6 and higher and Mainboard (MSP160382), perform the Oxygen Sensor Calibra-
tion as shown in the Oxygen Sensor Calibration with Mainboard MSP160200 Revisions 6 and Higher Section 8.14.1.5.2
8.14.1.5.2..
3. Disconnect the Oxygen Sensor Cable from the Oxygen Sensor and connect the Oxygen Sensor Calibration Tool (PN
160367).
124 English | 624165/07
5. The Offset Calibration runs automatically until Offset calibration completed appears on the screen.
6. Connect the Oxygen Sensor Cable to the Oxygen Sensor.
7. Touch the Start Button to begin the Gain Calibration.
8 Service Software
8. The Gain Calibration runs automatically indicated by Oxygen cell calibration running... displayed on the screen.
NOTICE
For Mainboard (MSP160200) Revisions 6 and higher and Mainboard (MSP160382) Revisions 1-4.
3. Disconnect the Oxygen Sensor Cable from the Oxygen Sensor. Make sure the Oxygen Sensor connector is freely sus-
pended.
4. Touch the Start button.
8 Service Software
8. The Gain Calibration runs automatically indicated by Oxygen cell calibration running... displayed on the screen.
8 Service Software
NOTICE
For Mainboard (MSP160382) Revisions 5 and higher: The calibration is complete when OK is displayed on the screen.
4. The Gain Calibration runs automatically indicated by Oxygen cell calibration running... indicated on the screen.
5. The calibration is complete when OK is displayed on the screen.
130 English | 624165/07
Flow Sensor 8
2. Connect the tubing system and turn the Patient Flow Sensor. The Patient Flow Sensor has to be inverted for the Patient
Flow Sensor Calibration and an adapter is required.
8 Service Software
Flow Sensor 8
7. The successful finish of the Patient Flow Sensor Calibration will be indicated with following message: Flow sensor calibra-
tion successful.
8 Service Software
Pressure 8
8.14.2.1 Pressure
1. Touch the Pressure Button.
2. The screen displays the Pvent_control, Pvent_monitor and Paw Pressure Sensors Offset and Gain values.
Hamilton Medical | HAMILTON-C2 Service Manual 135
8 Service Software
NOTICE
Valves 8
8.14.2.3 Valves
1. Touch the Valves Button.
8 Service Software
Rinsp 8
8.14.2.5 Rinsp
1. Touch the Rinsp Button
8 Service Software
CAUTION
If calibration is not possible due to a corrupted calibration file, touch the Default set. Tab. Therefore, all steps of the Service
Software have to be run again.
2. The Default Set. Tab provides a method to reset or set the values to a pre-determine
pre-determined
d Factory Default Setting. You can
choose between setting the default values of only the Flow Sensor or of All settings.
3. To reset the values, touch the Set Button.
4. When the Set Button is selected, choose between Flow sensor and All settings.
6. You must touch the Confirm Button to reset to the Default Settings.
Hamilton Medical | HAMILTON-C2 Service Manual 141
8 Service Software
Component Tests 8
Adj/Calib > Calibration > Alarm System This test allows activating the components related to • To
Touc
uch
h Scre
Screen
en
(Page No 2102) the alarm system, it also tests the loudness levels. The • Alarm
Alarm lamp ligh
lights
ts red or yello
yellow,
w,
Alarm Section 8.15.1.1 test result needs to be confirmed manually. depending
Adj/Calib > Calibration > Alarm Monitor This test activates the software watchdog. The test • Watc
Watchdog
hdog (soft
(software
ware tasks
tasks))
Adj/Calib > Calibration > User Interface With this test, the User Interface related components • P&
P&TT Contr
Control
ol Knob
Knob
Tab (Page No 2115) can be activated in order to check the interaction of • Ha
Hard
rd Keys
Keys + LED’s
LED’s
User Interface Tab Section 8.15.2 the Hard Keys with the GUI. Hard Key combinations
• Back
Backligh
lightt (day/night
(day/night brightnes
brightness)
s)
can also be tested. Dimming the screen and alarm
lamp is tested using the Day/Night Button. The P&T
Control Knob has 16 steps. The test is to see that all
16 steps register with the ventilator.
Adj/Calib > Calibration > Autozero The test repeats the Autozero sequence 5 times • Au
Auto
toze
zero
ro of:
of:
(Page No 2109) under a defined pressure. – Paw Pre
Pressu
ssure
re Se
Senso
nsorr
Autozero Section 8.15.3.2 – Qaw Pro
Proxim
ximal
al Flow
Flow Sens
Sensor
or
– Pv
Pven
ent_
t_mo
moni
nito
torr
– Pv
Pven
ent_
t_co
cont
ntro
roll
Adj/Calib > Calibration > Blower Flow Several flows (ml/s) will be set as target to the blower. • Blo
low
wer
(Page No 2104) The achieved blower speed in rpm measured by the
Blower Flow Section 8.15.3.3 Hall Sensor should be within the defined tolerance.
Adj/Calib > Calibration > Blower Pres- Several pressures are set as target to the blower. The • Blo
low
wer
sure (Page No 2105) achieved pressure measured by the Pvent_monitor
Blower Pressure Section 8.15.3.4 Pressure Sensor should be within the given tolerance.
8 Service Software
Adj/Calib > Calibration > Expiratory This test checks the Expiratory Valve. Several pres- • Exp
Expira
irator
toryy Valve
Valve
Valve (Page No 2111) sures are set as target for a constant flow Qinsp to
Expiratory Valve Section 8.15.3.5 the blower. The pressures are measured by Paw and
have to be within the defined tolerance
Adj/Calib > Calibration > Inspiratory This test checks the Inspiratory Valve tightness and • Ins
Inspir
pirato
atory
ry Valve
Valve
Valve (Page No 2107) flow control with Qinsp 0, 3, 18 and 150 l/min as
Inspiratory Valve Section 8.15.3.6 well as pressure measurement Pvent_monitor with
Pinsp of 2, 10, 25 and 40 mbar.
Adj/Calib > Calibration > O2 Input (Page For this test, the ventilator needs to be connected to • O2 Mixer
Mixer Asse
Assembl
mblyy
No 2112) high pressure oxygen. Several flows will be set as a
O2 Input Section 8.15.4.1 target for the O2 Mixer Valve. The oxygen flow is
measured by the internal QO2 Flow Sensor and has
to be within a certain tolerance. In a second test the
leakage is tested.
Adj/Calib > Calibration > Nebulizer For this test, the ventilator needs to be connected to • Neb
Nebuli
ulizer
zer Val
Valve
ve
Valve (Page No 2116) high pressure oxygen. The Patient Flow Sensor is used
Nebulizer Valve Section 8.15.4.2 to measure the generated flow when the Nebulizer
Valve is in a closed or opened state. The test results
successfully if the measured Qaw matches to the
defined tolerances.
Adj/Calib > Calibration > Ambient Valve This test checks the ambient valve with a blower pres- • Amb
Ambie
ient
nt Val
Valve
ve
(Page No 2108) sure of 20 and 60 mbar in off, closed and active
Ambient Valve Section 8.15.4.3 opened state. The result is measured with
Pvent_monitor Pressure Sensor.
Adj/Calib > Calibration > Proximal Flow This test consists of three parts: • Ri
Rins
nsee Fl
Flow
ow
(Page No 2110) • Paw Pres
Pressure
sure Sens
Sensor
or
- Rinse Flow Test:
Proximal Flow Section 8.15.4.4 • Qaw Proxim
Proximal
al Flow
Flow Sensor
Sensor
This test requires checking the rinse flow manually by
immersing the 2 flow tubes in a glass of water as
depicted. The number of bubbles has to be approxi-
mately equal on each outlet.
- Proximal Pressure:
This test allows checking the proximal pressure of the
Patient Flow Sensor. Two different pressures are
applied consecutively by the blower. The achieved
pressure is measured by the Paw Pressure Sensor and
has to be within the tolerance.
- Proximal Flow:
A certain flow is set as target to the blower, con-
trolled by the internal Qvent Flow Sensor. The flow
through the proximal side of the Patient Flow Sensor
is measured by the internal differential Qaw Proximal
Flow Sensor and has to be within a certain tolerance.
Adj/Calib > Calibration > Air Entry This test allows checking the Pfilter Pressure Sensor • HE
HEPA
PA Fi
Filt
lter
er
(Page No 2117) with and without an obstructed filter. • Pfil
Pfilter
ter Pressu
Pressure
re Senso
Sensorr
Air Entry Section 8.15.4.5
144 English | 624165/07
Electronics Tab 8
This loudness setting also sets the loudness for the System Test, Alarming. To check the alarm sound on High-, Medium-,
Low-Priority Alarms with the loudness range 1–10, set the loudness here, then perform the System Test, Alarming (see
Alarming Section 8.16.5
8.16.5).
).
8 Service Software
Alarm System 8
4. The speaker volume can be adjusted with the screen adjustment knob. Check that the volume changes to minimum and
maximum.
5. Indicate on the screen if the speaker is operating properly by touching OK or NOT OK.
8 Service Software
6. When the OK Button is touched, the speaker will switch OFF, the Speaker Start Button will switch to Start and OK will be
indicated on the screen.
Alarm System 8
9. Indicate on the screen if the Yellow lamp is operating properly by touching OK or NOT OK.
10. When the OK Button is touched, the Yellow Lamp will switch OFF, the Yellow Lamp Start Button will switch to Start and
OK will be indicated on the screen.
8 Service Software
Alarm System 8
14. When the OK Button is touched, the Red Lamp will switch OFF, the Red Lamp Start Button will switch to Start and OK will
be indicated on the screen.
15. After completion of the Alarm System Tests, the results are indicated on the screen.
Hamilton Medical | HAMILTON-C2 Service Manual 151
8 Service Software
Alarm Monitor 1 8
3. Indicate on the screen if the Alarm Light and the Alarm Silence LED are blinking by touching OK or NOT OK.
NOTICE
With a Front Panel Board Revision 6 and higher, only the Alarm Silence LED will blink.
4. Record the results on the HAMILTON-C2 Summary of Test Report Section Summary of Test Report.
Hamilton Medical | HAMILTON-C2 Service Manual 153
8 Service Software
5. Next, indicate on the screen if the Alarm Light and the Alarm Silence LED are ON by touching OK or NOT OK.
NOTICE
With a Front Panel Board Revision 6 and higher, only the Alarm Silence LED will be ON.
NOTICE
With a Front Panel Board Revision 6 and higher, only the Alarm Silence LED will be ON.
Alarm Monitor 1 8
8. Next, indicate on the screen if the Alarm Light is OFF by touching OK or NOT OK.
8 Service Software
12. Indicate on the screen if the Expiratory Valve Plunger can be moved by hand by touching OK or NOT OK.
Alarm Monitor 1 8
13. Indicate on the screen if the Alarm Silence LED is ON and the Buzzer sounds by touching OK or NOT OK.
WARNING
Be careful when stopping the cooling fan by hand.
14. Next, remove the Rear Cover and Fan Filter, then physically STOP the cooling fan by hand until the alarm is shown on the
screen.
8 Service Software
15. After the Fan Failure Alarm is displayed, release the fan.
16. Verify that the message Test completed is displayed on the screen.
Alarm Monitor 2 8
• The Ala
Alarm
rm LED
LED sho
should
uld bli
blink
nk
• The Buz
Buzzer
zer sho
should
uld sou
sound
nd
Hamilton Medical | HAMILTON-C2 Service Manual 159
8 Service Software
3. Touch the OK or NOT OK Button and reboot the ventilator into the Service Software.
Software.
4. Test the P&T Control Knob Encoder by selecting the P&T Control Knob on the screen and rotating the knob. Observe there
are 16 steps in 1 full turn of the P&T Control Knob.
8 Service Software
5. Press each button on the Interaction Panel and observe the reaction to the button pressed on the screen.
6. Touch the Night/Day Button to switch between the two display brightness levels.
7. Confirm with OK/NOT OK.
Pneumatics 1 Tab 8
8 Service Software
Binary Valve 8
4. The test runs automatically indicated by Component test binary valves running... displayed on the screen.
5. The Component test binary valves is complete when OK is indicated on the screen. This means that the internal sensor
checks are ok.
8 Service Software
8.15.3.2 Autozero
The test repeats the Autozero sequence 5 times under a defined pressure.
1. Touch the Autozero Button.
Autozero 8
3. The Paw Pressure Sensor and Qaw Proximal Flow Sensor will Autozero.
4. The Paw Pressure Sensor and Qaw Proximal Flow Sensor Autozero adjustment is complete when OK is indicated on the
screen.
8 Service Software
Blower Flow 8
4. The test runs automatically indicated by Component test blower flow running... displayed on the screen.
5. The test is complete when Component test blower flow completed. is displayed on the screen and the results are displayed
with OK or NOT OK.
NOTICE
The Blower Pressure values are in mbar and the Flow values are in ml/minute.
Hamilton Medical | HAMILTON-C2 Service Manual 169
8 Service Software
2. Disconnect the Inspiration Tube and seal the Patient Outlet with a stopper.
3. Touch the Start button.
Blower Pressure 8
4. The test runs automatically indicated by Component test blower pressure running... displayed on the screen.
5. The test is complete when Component test blower flow completed. is displayed on the screen and the results are displayed
with OK or NOT OK.
NOTICE
8 Service Software
NOTICE
During this test, the Technical Event: 231001 may appear. This failure message can be ignored.
This test checks the Expiratory Valve. Several pressures are set as target for a constant flow Qinsp to the blower. The pres-
sures are measured by Paw and have to be within the defined tolerance.
1. Touch the Exp. valve Button.
Expiratory Valve 8
5. The test is complete when Expiration valve pressure test completed is displayed on the screen and the result is displayed
with OK or Not OK.
6. If the pressure test or leakage test fails, the values which are out of tolerance are displayed at the bottom of the page.
8 Service Software
2. Connect an Adult 22 mm Tube (PN 260100) and Filter (PN 279211) to the ventilator and attach the Flow Analyzer (as
shown in the picture on the screen).
3. Touch the Start Button.
Inspiratory Valve 8
5. The Inspiratory Valve Leakage Test is complete when OK is indicated on the screen. If the pressure tests fails, the out of tol-
erance pressure is displayed.
8 Service Software
7. Pass through the Flow Control Test with the values of 0 l/min., 3 l/min., 18 l/min. and 150 l/min. Only verify the 18 l/min.
value with the external Flow Analyzer. It must be within a tolerance of ± 8% (16.5 - 19.5 l/min).
8. The test runs automatica
automatically
lly indicated by Flow control in progress on the screen.
9. The Inspiratory Valve Flow Control Test is complete when OK is indicated on the screen.
10. Connect the tubing as shown with a 7mm ET Tube.
11. Touch the Start Button to continue to the Inspiratory Valve Pressure Control Test.
Inspiratory Valve 8
12. The test runs automatically indicated by Pressure control in progress on the screen.
13. The Inspiratory Valve Pressure Control Test is complete when OK is indicated on the screen.
8 Service Software
O2 Input 8
8.15.4.1 O2 Input
For this test, the ventilator needs to be connected to high pressure oxygen. Several flows will be set as a target for the O2
Mixer Valve. The oxygen flow is measured by the internal QO2 Flow Sensor and has to be within a certain tolerance. In a
second test the leakage is tested.
1. Touch the O2 input Button.
2. Connect the ventilator to high pressure oxygen (Minimum 2.8 bar to Maximum 6.0 bar (41 - 86 psi)).
3. Disconnect the Inspiration Tube.
4. Touch the Start Button to begin the O2 Input Flow and Leakage Tests.
8 Service Software
O2 Input 8
8 Service Software
Nebulizer Valve 8
8 Service Software
NOTICE
In software version 2.0.6 and earlier, during this test, the Technical Event: 231001 may appear. This failure message can be
ignored.
This test checks the Ambient Valve with a blower pressure of 20 and 60 mbar in off, closed and active opened state. The
result is measured with the Pvent_monitor Pressure Sensor.
1. Touch the Ambi valve Button.
2. Connect the tube system as shown and close the Flow Sensor Outlet.
3. Touch the Start Button.
Ambient Valve 8
4. The test runs automatically indicated by Test ambient valve power off in progress displayed on the screen.
5. The Ambient Valve Power-Off Test is complete when OK is indicated on the screen.
6. The next test runs automatically indicated by Test ambient valve Closed in progress displayed on the screen.
Figure 8-146. Ambient Valve Tests Step 2, Possible Technical Event (only with SW 2.0.6 or lower)
7. The Ambient Valve Closed Test is complete when OK is indicated on the screen.
8. The next test runs automatically indicated by Test ambient valve Active open in progress displayed on the screen.
9. The Ambient Valve Active Open Test is complete when OK is indicated on the screen.
8 Service Software
Proximal Flow 8
8 Service Software
Proximal Flow 8
11. Touch the Start button to begin the Proximal Flow Test.
12. The test runs automatically indicated by Proximal flow in progress... displayed on the screen.
8 Service Software
13. The Proximal Flow Test is complete when OK is indicated on the screen.
Air Entry 8
NOTICE
If the test is NOT OK, restart the HAMILTON-C2 Service Software and perform the test again.
This test allows checking the Filter Pressure Sensor Pfilter with and without an obstructed filter.
8 Service Software
4. The default HEPA Filter Test is completed if OK is stated behind the description.
5. Obstruct the HEPA Filter with a piece of paper (A) with a small hole.
Air Entry 8
6. Connect the obstructed HEPA Filter and touch the Next Button. The pressure must be higher than 2 mbar as shown.
7. The test is complete when OK is indicated on the screen.
8 Service Software
System Test 8
System Test > Pressure (Page No This test requires three different pressures to be applied and to ver- • Pve
Pvent_
nt_con
contro
troll
2201) ify the values (Pvent_control, Pvent_monitor and Paw) are within • Pve
Pvent_
nt_mon
monito
itorr
Pressure Section 8.16.1 the given tolerance range (shown in a green or red background).
• Paw
System Test > Flow (Page No This test requires a flow to be applied and to verify the internal • Qve
Qvent
nt Flow
Flow Sens
Sensor
or
2202) Qvent Flow Sensor and the Qaw Proximal Flow Sensor values are • Qaw Proxi
Proximal
mal Flow Sens
Sensor
or
Flow Section 8.16.2 within the given tolerance range (shown in a green or red back-
ground).
System Test > O2 Mixer (Page No This test requires a flow at different oxygen concentrations to be • O2 Mixe
Mixerr Assem
Assembly
bly
2203) applied to verify the internal QO2 Flow Sensor, Qvent Flow Sensor, • QO2 Flo
Flow
w Sens
Sensor
or
O2 Mixer Section 8.16.3 the Patient Flow Sensor and the Oxygen Sensor values are within
• Ox
Oxyg
ygen
en Sens
Sensor
or
the given tolerance range (shown in a green or red background).
System Test > Tightness (Page This test checks for leakage in the Patient Breathing Circuit and • Patie
Patient
nt Breathi
Breathing
ng Circuit
Circuit
No 2204) determines the circuit's compliance compensation factor. The venti-
Tightness Section 8.16.4 lator is pressurized to a certain pressure. The Patient Breathing Cir-
cuit is considered tight if this pressure can be maintained. If there is
a leak, the pressure drops in proportion to the size of the leak.
System Test > Alarming (Page No This test allows activation of the 3 different alarm priorities and • Ala
Alarm
rm Prior
Prioriti
ities
es
2205) combinations by pressing the corresponding button and to verify
Alarming Section 8.16.5 the correct alarm lamp and sound has been generated.
Hamilton Medical | HAMILTON-C2 Service Manual 195
8 Service Software
8.16.1 Pressure
This test requires three different pressures to be applied and to verify the values (Pvent_control, Pvent_monitor
Pvent_monitor and Paw) are
within the given tolerance range (shown in a green or red background).
1. Touch the Pressure Button.
Pressure 8
6. All values should read 5 mbar ±1.0, 25 mbar ±1.2 and 50 mbar ±2.5.
7. The test is complete when the Status OK is shown.
8 Service Software
8.16.2 Flow
This test requires a flow to be applied and to verify the internal Qvent Flow Sensor and the Qaw Proximal Flow Sensor values
values
are within the given tolerance range (shown in a green or red background).
1. Touch the Flow Button.
Flow 8
8 Service Software
8.16.3 O2 Mixer
This test requires a flow at different oxygen concentrations to be applied to verify the internal QO2 Flow Sensor, Qvent Flow
Sensor, the Patient Flow Sensor and the Oxygen Sensor values are within the given tolerance range (shown in a green or red
background).
1. Connect the ventilator to high pressure oxygen.
2. Touch the O2 mixer Button.
O2 Mixer 8
8 Service Software
8.16.4 Tightness
NOTICE
During this test the Technical Event: 231001 may appear. You can ignore this fault message.
This test checks for leakage in the Patient Breathing Circuit and determines the circuit's compliance compensation factor. The
ventilator is pressurized to a certain pressure. The Patient Breathing Circuit is considered tight if this pressure can be main-
tained. If there is a leak, the pressure drops in proportion to the size of the leak.
1. Touch the Tightness Button.
Tightness 8
3. The test runs automatically indicated by Tightness tubing test in progress... indicated on the screen.
8 Service Software
8.16.5 Alarming
NOTICE
During this test, Technical Events and Technical Faults will be displayed. This is a needed behavior and does not need any
action.
This test allows activation of the 3 different alarm priorities and combinations by pressing the corresponding button and to
verify the correct alarm lamp and sound has been generated.
1. Touch the Alarming Button.
Alarming 8
2. Touch the Alarm High, Alarm Medium and Alarm Low Buttons to generate alarm(s).
3. Check if the correct Alarm Lamp and Alarm Sounds are generated.
4. Check the Alarm Priority by combining the Alarms.
5. Touch the Alarm High Button.
6. Select the Alarm High and Alarm Medium Buttons and verify if the alarm priority is working correctly.
8 Service Software
Sensor Data 8
8 Service Software
8.17.1 Sensor 1
Touch the Sensor 1 Tab to display the sensors values and status.
Sensor 2 8
8.17.2 Sensor 2
Touch the Sensor 2 Tab to display the sensors values and status.
8 Service Software
8.17.3 Sensor 3
Touch the Sensor 3 Tab to display the sensors value and status.
Sensor Status 8
8 Service Software
How to export Event Log, Service Log, Config File and Instrument Report.
Event Log 8
2. The Event Log displays the Technical Events which have occurred with the date, time, device affected and a description of
the Technical Event.
3. To view another section of the Event Log listing, rotate the P&T Control Knob to select the slider on the right side of the
screen.
4. Press the P&T Control Knob to activate the slider, then rotate the P&T Control Knob to move the slider up or down.
Hamilton Medical | HAMILTON-C2 Service Manual 213
8 Service Software
2. The Service Log displays the test and calibration results with the date, time and a description of the results.
3. To view another section of the Service Log listing, rotate the P&T Control Knob to select the slider on the right side of the
screen.
4. Press the P&T Control Knob to activate the slider, then rotate the P&T Control Knob to move the slider up or down.
214 English | 624165/07
8 Service Software
If you have to send log files to Hamilton Medical for further investigation, please send all files which are created in the spe-
cific folder (error.log, eventLog.txt, serviceLog.txt,
serviceLog.txt, etc.).
8 Service Software
WARNING
After a software update, the ventilator must be restarted and the complete Service Software Calibration and Test's must be
performed.
Software downgrade
downgrade is not allowed/possible, because this will corrupt the unit and the ESM Module will need to be
exchanged.
To update the HAMILTON-C2 from Software Version 1.x.x to Software Version 2.2.0 or higher, a migration step is required.
For further information see the HAMILTON-C2 Software Upgrade Guide (PN 624200).
1. The following files are required:
• C2
C2_V
_V2.
2.2.
2.0-
0-Mi
Migr
grat
atio
ion.
n.ta
tarr This is re
This requ
quir
ired
ed as a mig
migra
rati
tion
on st
step
ep fr
from
om 1.
1.x.
x.xx to
to 2.2
2.2.0
.0 si
sinc
ncee the
the so
soft
ftwa
ware
re st
stru
ruct
ctur
uree has
has be
been
en
changed.
• C2_
C2_EM0
EM01A_
1A_V2.
V2.2.5
2.5.ta
.tarr or highe
higherr This is the
This the late
latest
st Softw
Software
are to
to be inst
install
alled
ed afte
afterr the Migr
Migrati
ation
on Soft
Softwar
ware.
e. For
For the late
latest
st softw
software
are
check the Partner-net.
3. The Software Update Screen allows update or upgrade of the HAMILTON-C2 Software.
218 English | 624165/07
5. Insert the USB Memory Stick containing the migration and upgrade version as listed in the table above.
If a correct file is available on the USB Memory Stick, the message appears indicating the update version, otherwise, the mes-
sage No Update tarball file is available... will be displayed. In case there is more than one software update stored on the USB
Memory Stick, please ensure to choose the correct version.
6. If the USB Memory Stick content is not displayed directly, refresh this page by closing and entering this page again.
8 Service Software
11. Restart the HAMILTON-C2. The unit starts up automatically in the Service Software.
12. For further update of the HAMILTON-C2 Software, see the Software Update from Version 2.x to 2.2.5 or Higher Section
8.19.2..
8.19.2
220 English | 624165/07
Version
Software Update from Version 2.x to 2.2.5 or Higher 8
WARNING
After a software update, the ventilator must be restarted and the complete Service Software Calibration and Test's must be
performed.
NOTICE
2. The Software Update Screen allows update or upgrade of the HAMILTON-C2 Software.
Hamilton Medical | HAMILTON-C2 Service Manual 221
8 Service Software
4. If the USB Memory Stick is not inserted into the USB Connector or no software is available on the USB Memory Stick, the
message No Update tarball file is available... is displayed.
Version
Software Update from Version 2.x to 2.2.5 or Higher 8
5. If the correct file is available on the USB Memory Stick, the message appears indicating the update version.
8 Service Software
6. To begin the software update, touch the Start Button. With software version 2.2.0 or higher, the software which is to be
updated can be selected from the scroll down menu (shown on the picture below).
Version
Software Update from Version 2.x to 2.2.5 or Higher 8
8. After the software update, the message update successfully finished is displayed on the screen.
9. Restart the device and do not terminate the device during the first start up after performing a software update!
10. Perform the complete Calibration and Tests as described in the Service Software Section 8
Hamilton Medical | HAMILTON-C2 Service Manual 225
8 Service Software
General Tests 8
All devices connected to the HAMILTON-C2 must be for medical use and meet the requirements of standard IEC 60601-1.
1. Connect
Connect the HAMILT
HAMILTON-C
ON-C2
2 to the AC Mains
Mains Power
Power and to the
the DC Power
Power Supply.
Supply.
• If no DC Power Supply
Supply is available, disconnect
disconnect the AC Mains
Mains Power and verify
verify that the HAMILTON-C2
HAMILTON-C2 switched
switched to internal
Battery Power. Verify that the alarm speaker sounds correctly for at least one minute.
CAUTION
The buzzer is powered by the gold caps (super capacitors) on the Mainboard. Make sure the unit is switched on (Stand-by
mode or Service Mode) for at least 20 minutes before the test in order that the gold caps are fully charged.
Replace the Mainboard (MSP160382) if the buzzer sounds for less than 2 minutes.
Hamilton Medical | HAMILTON-C2 Service Manual 227
8 Service Software
E. Fan Check
F. Buzzer Test
– Verif
Verifyy that the
the buzzer
buzzer sounds
sounds for at least
least 2 minutes
minutes..
• If the buzzer sounds
sounds for less
less than 2 minutes, replace the Mainboard
Mainboard (MSP160382).
G. Communication Interface RS232 Test
General Tests 8
1. Run HAMILTON-Da
HAMILTON-Datalogger.exe
talogger.exe (available as download from the Hamilton Medical Partner-net).
The Welcome window appears.
2. Click > ENTER
NOTICE
In addition, for all connections you may order USB to RS232 Converter (PN 396187, in case your computer has only an USB
Interface).
Important: The USB to RS232 Converter needs an additional driver to work properly. It is enclosed on the CD in the pack-
age.
Hamilton Medical | HAMILTON-C2 Service Manual 229
8 Service Software
8 Service Software
5. Activate the CO2 Hardware Option: Utilities > Configuration > Options > HW options
9. Connect the Patient Breathing Circuit; Test Lung, CO2 Sensor, CO2 Adapters, CO2 Calibration Gas and CO2 Gas Regula-
tor as shown.
8 Service Software
10. Select the System Tab, Tests & calib, then CO2 Sensor to start the CO2 Sensor Calibration.
11 Start ventilation with standard settings. Select the PCO2 curve for displaying on the screen (see Figure 8-218 CO2 Sensor
Accuracy Check).
12. Open the CO2 Regulator slightly so that the display shows a stable CO2 curve (see Figure 8-218 CO2 Sensor Accuracy
Check).
NOTICE
Only a small regulated gas flow is required to achieve a stable curve. To achieve a stable CO2 reading, ensure that the CO2
Sensor and CO2 Adapters are upright.
13. To get the PCO2 value in mmHg, press the Freeze Button (e.g. 34mmHg as shown).
8 Service Software
Figure 8-220. Acceptable Range for PCO2 at Atmospheric Pressure (17.5 to 22.5 °C)
236 English | 624165/07
Figure 8-221. Acceptable Range for PCO2 at Atmospheric Pressure (22.5 to 27.5 °C)
Hamilton Medical | HAMILTON-C2 Service Manual 237
8 Service Software
Figure 8-222. Acceptable Range for PCO2 at Atmospheric Pressure (27.5 to 32.5 °C)
238 English | 624165/07
Figure 8-223. Acceptable Range for PCO2 at Atmospheric Pressure (32.5 to 37.5 °C)
Hamilton Medical | HAMILTON-C2 Service Manual 239
8 Service Software
6 — —
3. Check the three relays with a multimeter by measuring the resistance during ventilation, alarm off and alarm on between
the following wires:
• Enter Service software > Tests/Calibration > System test > Alarming (See Figure 8-225 Alarm On/Off Relay Check).
• Activ
Activate/
ate/deac
deactivat
tivatee alarm
alarm (e.g.
(e.g. Alr. High).
High).
8 Service Software
Depending on how frequent you use the ventilator, you can replace the battery if the State of Health (SoH) is < 40%.
Refer to the Evaluation of the Battery State of Health Section 5.3 for details on how to read the battery data.
1. Che
Checkck the
the Batte
Battery
ry State
State of
of Health
Health (SoH)
(SoH)..
2. Repl
Replac
acee the
the ba
batt
tter
eryy if:
if:
– the State
State of Health
Health (SoH
(SoH)) is < 20%
– the cell voltages are permanently
permanently asymmetrical,
asymmetrical, that is, the difference between
between cells voltage
voltage is more than
than 100 mV.
3. Cal
Calibr
ibrate
ate the bat
batter
teryy ifif the
the Error is ≥ 5%.
4. Recor
Record
d the batter
batteryy data on the Summ
Summary
ary of Test
Test Report
Report Sectio
Section
n 17.1
17.1..
242 English | 624165/07
9
Technical Faults
9 Technical Faults
If more than one alarm is active, the associated alarm messages alternate in the message bar.
Alarm Indications
High-priority Red, with alarm Red A sequence of 5 beeps, repeated until The patient’s safety is compromised.
alarm message the alarm is reset. If the audible alarm is The patient needs immediate attention.
not silenced during the first minute, the
continuous-tone buzzer also sounds.
Low-
Low-pr
prio
iori
rity
ty alar
alarm
m Yellow
Yellow,, with
with Yell
Yellow
ow Two se
Two sequ
queenc
nces
es of be
beeeps
ps.. Thi
Thiss is no
not Operator awareness is required.
alarm message repeated.
Technical fa
fault Red, wi
with Safety Red Same as for high-priority alarm, if tech- The ventilator enters safety mode, or, if
ventilation: xxxxxx nically possible. At the minimum a con- it cannot safely ventilate, the ambient
or Technical fault: tinuous buzzer tone. The buzzer cannot state. Provide alternative ventilation.
xxxxxx be silenced. Turn off the ventilator. Have the venti-
lator serviced.
Technical Failure Ending in Safety Mode - 300000 Alarm Codes Sec- Technical Failures resulting in the ventilator switching to the Safety
tion 16.2.3 Mode
Technical Failure Ending in Ambient Mode - 400000 Alarm Codes Technical Failures resulting in the ventilator switching to the Ambi-
Section 16.2.4 ent Mode
500000 Alarm Codes Section 16.2.5 System Failures; invisible TF for user, no action required. For internal
use.
244 English | 624165/07
• Patient Alarms: (Alarm code 100000) Alarms related to : Pressure, Flow, Volume, Apnea
– Hig
High
h Prio
Priorit
rityy Alar
Alarms
ms
– Med
Medium
ium Pr
Prior
iority
ity Ala
Alarms
rms
– Low Pri
Priori
ority
ty Ala
Alarm
rmss
• Technical Alarms: (Alarm code 200000) Alarms related to: Temperat
Temperature,
ure, Battery, Valves, Sensors
– Hig
High
h Prio
Priorit
rityy Alar
Alarms
ms
– Med
Medium
ium PrPrior
iority
ity Ala
Alarms
rms
– Low Pri
Priori
ority
ty Ala
Alarm
rmss
Technical Faults Technical Faults end up in one of the following states:
9 Technical Faults
9.2 Troubleshooting
For troubleshooting, refer to document number 624485 "Troubleshooting Platform C" available for download from the
Hamilton Medical Partner-net.
Follow suggested troubleshooting steps in order as described.
Follow these steps when troubleshooting:
1. Confir
Confirm
m that
that the test confi
configurat
guration
ion is
is correc
correct.
t.
2. When replacing
replacing components,
components, repeat the applicable
applicable test following each
each replacement.
replacement. If the test passes successfully,
repeat all test's in order (see the Service Software Section 8).
3. If a problem cannot be resolved, enter
enter a Customer
Customer Event Report (CER) on the Hamilton
Hamilton Medical AG on the Online Help
Desk (https://hamilton.ky2help.com).
(https://hamilton.ky2help.com). Export the Event Log and Instrument Report for failure analysis. Check regularly the
knowledge base of the ky2help for the latest information and troubleshooting.
246 English | 624165/07
10
Components Removal/Assembly
10 Components Removal/Ass
Removal/Assembly
embly
WARNING
You must perform the repairs detailed in this section only as instructed in the Preventive Maintenance and Testing Overview
Section 6.1
6.1,, Hospital Preventive Maintenance
Maintenance Section 6.2 and Engineer Preventive Maintenance Section 6.3
6.3..
CAUTION
After performing Maintenance or Replacement of a component or module,
module, perform the necessary Service Software checks,
calibrations and Safety Tests to ensure the HAMILTON-C2 is performing properly.
Hamilton Medical AG does not permit repairs to parts that are supplied as an assembly. For example:
− You have to replace the Blower Module as a complete assembly
− You have to replace the Expiratory Valve as a complete assembly
− You have to replace the Mixer Block as complete assembly
− You have to replace Hamilton Medical AG printed circuit boards only
Make sure to take full ESD (electrostatic discharge) precautions before opening the HAMILTON-C2. Refer to the ESD (Elec-
trostatic Discharge ) Protection Section 11.3.3
11.3.3..
When the HAMILTON-C2 is switched ON, even when not connected to the external Mains Power, the Battery Pack(s) supply
power. Therefore, a short circuit is possible when the HAMILTON-C2
HAMILTON-C2 is switched ON.
NOTICE
Always send defective parts, components or assemblies to Hamilton Medical AG with a completed Return Goods Authoriza-
tion (RGA) request.
Before making any repairs, remove from the HAMILTON-C2:
− External Mains Power Supply and Batteries
− Oxygen Supply
To prevent unintentional disconnection of the Power Cord, make sure it is well seated into the ventilator’s socket and
secured with the Power Cord Security Clamp.
The HAMILTON-C2 does not require protective earth grounding, because it is a class II device, as classified according to IEC
60601-1.
Secure the Power Cord with the Security Clamp as shown in the picture below.
Two different Security Clamps are available. Use the suitable Security Clamp as listed in the table below.
To replace the Power Cord, release the screw (PN 420724) of the Security Clamp.
When reconnecting the Power Cord, make sure it is well seated into the ventilator’s socket and secured with the Power Cord
Security Clamp.
10 Components Removal/Ass
Removal/Assembly
embly
The Interaction Panel must be tilted completely to the front to remove the Top Cover.
To remove the Ventilation Unit Top Cover:
1. Lift the Handle of the Ventilation Unit.
2. Remove 6 Torx screws from the Top Cover of the Ventilation Unit:
• A: 2 Torx
Torx scr
screw
ewss (PN
(PN 42068
420682)
2)
• B: 4 Torx
Torx scr
screw
ewss (PN
(PN 420642
420642))
3. Remove the Ventilation Unit Top Cover (C).
WARNING
Torx screws (A) cannot be any longer than the specified size (M3x6), otherwise, damage can be caused to the Interaction
Panel Board.
10 Components Removal/Ass
Removal/Assembly
embly
NOTICE
The 2 Flexible Flat Cables (D) can more easily be disconnected from the Interaction Panel Board, but can also be discon-
nected from the Ventilation Unit Mainboard. The ferrite cores for these 2 cables are secured to the top of the front cover,
and are not intended to be removed with the cables.
4. Remove 4 Torx screws (B, PN 420727) from the Hinges of the Interaction Panel.
5. Remove the Interaction Panel (A) from the Ventilation Unit.
6. Place aside in a safe location.
10 Components Removal/Ass
Removal/Assembly
embly
WARNING
When replacing the GRAPHIC LCD TFT VGA 10.4” NEC Display (Obsolete, PN 380031), order the Display Upgrade Kit
(MSP380046). The DC/AC Backlight Inverter Boards (Obsolete, PN 396197 or PN 396314) are not compatible with the latest
GRAPHIC LCD TFT VGA 10.4" NLT Display. Use the Backlight Driver Board (PN 396314) if the Front Panel Boards is below
Revision 08. Front Panel Boards at Revision 08 and higher have integrated Backlight Driver Components (see the Front Panel
Board Removal/Assembly Section 10.3.6 for configuration instructions).
WARNING
The Backlight LED Driver Components are integrated into Front Panel Boards with Revision 08 and higher, compatible with
the latest GRAPHIC LCD TFT VGA 10.4" NLT Display. The installation of an External Backlight Driver Board is not necessary
when you configure Jumper P14 to pins 1 and 2 on the new Front Panel Board. An Extension Cable (PN 161569) is used to
bridge the connections for the removed Backlight Driver Board (see instructions below).
CAUTION
Update the Technical State (refer to Modify Tab Section 8.8.3).
10 Components Removal/Ass
Removal/Assembly
embly
6. Remove 5 Torx screws (C, PN 420724) and 2 Torx screws (D, PN 420641) from the Interaction Panel Board.
7. Remove the Interaction Panel Board (E).
9. Connect Jumper P14 to Pins 1 and 2 on the new Front Panel Board if no external Backlight Driver Board is installed. Con-
nect the Extension Cable (PN 161569) from the Backlight Connector to P13 on the Front Panel Board.
Pins 1 and 2 if no External Backlight Driver Board is Installed
10. Connect Jumper P14 to Pins 2 and 3 when using an external Backlight Driver Board with the new Front Panel Board.
11. Connect the Backlight Driver Board between the Backlight Cable and P7 Connector on the Front Panel Board.
10 Components Removal/Ass
Removal/Assembly
embly
WARNING
When replacing the GRAPHIC LCD TFT VGA 10.4” NEC Display (Obsolete, PN 380027 or PN 380031), order the Display
Upgrade Kit (MSP380046).
10 Components Removal/Ass
Removal/Assembly
embly
CAUTION
The P&T Control Knob must be positioned to allow free movement. Attach so there is a 3 mm gap between the back of the
Knob and the face of the Interaction Panel. Test the Knob actions for both Press and Turn.
CAUTION
Do not remove the P&T Control Knob for maintenance. Removal of the P&T Control Knob can damage the P&T Control
Knob Encoder. Only remove if necessary to replace a non-functioning P&T Control Knob Encoder.
To remove the P&T (Press and Turn) Knob Encoder (PN 372036):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly
Removal/Assembly Section
10.3.1).
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
10.3.2).
3. Remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assembly
Removal/Assembly Section 10.3.3
10.3.3).
).
4. Remove the Interaction Panel Large Rear Cover (refer to the Interaction Panel Large Rear Cover Removal/Assembly Section
Section
10.3.4).
10.3.4).
5. Disconnect the Cable from the Front Panel Board.
6. Remove the P&T Control Knob (A, PN 159449) from the front of the Interaction Panel.
7. Remove the Hex Nut (B) and Washer (C) from the P&T Control Knob Encoder (D).
8. Remove the P&T Control Knob Encoder (D) and Large Washer (E).
10 Components Removal/Ass
Removal/Assembly
embly
CAUTION
It is mandatory that the ventilator is operated with at least one battery installed.
NOTICE
− The Backup Battery Compartment provides space for 2 Backup Battery Packs. The primary Battery Pack is positioned into
the left compartment. An Optional 2nd Battery Pack is positioned into the right compartment.
− A Battery Lock is used to secure the Battery Release Latch. The Battery Release Latch must be in the latched position for
the Battery Door to close.
2. Open the Battery Lock (B) by rotating the Battery Lock counter-clockwise with a screwdriver.
3. Raise the Battery Release Latch (C) to release the Battery Pack (D) from the Battery Compartment.
4. Remove the Battery Pack (D).
10 Components Removal/Ass
Removal/Assembly
embly
5. If a second Battery Pack is used, release the Battery Release Latch (E) and remove the Battery Pack (F) from the Battery
Compartment.
10 Components Removal/Ass
Removal/Assembly
embly
3. Use the Handle (B) to pull the HEPA Filter from the holder.
4. Remove the HEPA Filter (C).
10 Components Removal/Ass
Removal/Assembly
embly
It is not necessary to remove the Interaction Panel from the Front Cover.
9. Depress the Trolley Release Front Latch to remove the Front Cover (F).
Figure 10-23. Front and Rear Covers Removal
CAUTION
Update the Technical State (refer to the Modify Tab 8.8.3).
NOTICE
Export the Technical State (Instrument Report) before removing the Ventilation Unit ESM (refer to the Export Section 8.18.3
8.18.3))
because, after removing the Ventilation Unit ESM, the Technical State will be lost.
10 Components Removal/Ass
Removal/Assembly
embly
CAUTION
Take care not to damage the Top Foam Section Latches used to secure the Ventilation Unit Mainboard.
NOTICE
Export the Technical State (Instrument Report) before removing the Ventilation Unit Mainboard (refer to the Export Section
8.18.3)) because, after removing the Ventilation Unit Mainboard the Technical State will be lost.
8.18.3
CAUTION
Update the Technical State (refer to the Modify Tab 8.8.3). Import the Technical State you have exported before and
change the revision and serial number of the replacement Mainboard.
3. Optional,
bly if needed,
Section 10.3.3
10.3.3).
). remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assem-
4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly
Removal/Assembly Section
10.4.5).
10.4.5).
5. Disconnect all cabling connections from the Ventilation Unit Mainboard.
6. Remove the Ventilation Unit Mainboard (A) from the Top Foam Section.
7. Remove the ESM Holder (PN 257080) and LCD Cable Lock Holder (PN 362051).
10 Components Removal/Ass
Removal/Assembly
embly
6. Lift the Top Foam Section (A) from the Ventilation Unit.
CAUTION
Be careful not to bend the FFC Pins when re-assembling the Blower Module.
NOTICE
The air flow of the Fan is directed into the device. When the power is on, hold a piece of paper against the Fan opening. If
the Fan is installed correctly the paper, will be drawn against the Fan.
To remove the Cooling Fan which includes the small board (PN 160346):
1. Remove the Top Cover from the Ventilation Unit (refer to the Ventilation Unit Top Cover Removal/Assembly
Removal/Assembly Section
10.3.1).
10.3.1).
2. Remove the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
10.3.2).
3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assem-
bly Section 10.3.3
10.3.3).
).
4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section
10.4.5).
10.4.5).
5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8
10.4.8).
).
6. Lift the Cooling Fan (A) from the Middle Foam Section (B) of the Ventilation Unit.
10 Components Removal/Ass
Removal/Assembly
embly
CAUTION
Update the Technical State (refer to the Modify Tab 8.8.3).
NOTICE
If Software > 2.2.3 is installed, you have to set the Blower Timer manually, after replacement
replacement of the Blower Module
6. Lift the Blower Module (A) from the Middle Foam Section (B) of the Ventilation Unit.
CAUTION
Update the Technical State (refer to the Modify Tab 8.8.3).
10 Components Removal/Ass
Removal/Assembly
embly
CAUTION
Be sure that all cables are free from the Middle Foam Section.
10 Components Removal/Ass
Removal/Assembly
embly
CAUTION
Update the Technical State (refer to the Modify Tab 8.8.3).
NOTICE
11. Lift the Pressure Sensor Assembly (A) slightly to remove the 2 Tubings (B) from the bottom of the Pressure Sensor Assem-
Assem-
bly.
10 Components Removal/Ass
Removal/Assembly
embly
CAUTION
Update the Technical State (refer to the Modify Tab 8.8.3).
12. Lift the Inspiratory Valve (B), Qvent Flow Sensor (C) and Ambient Valve (D) as an assembly from the Bottom Foam of the
Ventilation Unit.
13. Disconnect the Inspiratory Valve (B) from the Qvent Flow Sensor (C).
14. Assemble in the reverse order of removal.
Hamilton Medical | HAMILTON-C2 Service Manual 281
10 Components Removal/Ass
Removal/Assembly
embly
CAUTION
Update the Technical State (refer to the Modify Tab 8.8.3).
12. Lift the Inspiratory Valve (B), Qvent Flow Sensor (C) and Ambient Valve (D) as an assembly from the Bottom Foam of the
Ventilation Unit.
13. Disconnect the Inspiratory Valve (B) from the Qvent Flow Sensor (C).
14. Not shown; Remove a metal bracket (PN 160741) attached with 2x screws (PN 420636) secures the Qvent Flow Sensor to
the Ambient Valve.
15. Disconnect the Qvent Flow Sensor (C) from the Ambient Valve (D).
16. Remove the Qvent Flow Sensor (C).
17. Assemble in the reverse order of removal.
Hamilton Medical | HAMILTON-C2 Service Manual 283
10 Components Removal/Ass
Removal/Assembly
embly
CAUTION
Update the Technical State (refer to the Modify Tab 8.8.3).
12. Lift the Inspiratory Valve (B), Qvent Flow Sensor (C) and Ambient Valve (D) as an assembly from the Bottom Foam of the
Ventilation Unit ().
13. Not shown; Remove a metal bracket (PN 160741) attached with 2x screws (PN 420636) secure the Qvent Flow Sensor to
the Ambient Valve.
14. Disconnect the Qvent Flow Sensor (C) from the Ambient Valve (D).
15. Remove the Ambient Valve (D).
16. Assemble in the reverse order of removal.
Hamilton Medical | HAMILTON-C2 Service Manual 285
10 Components Removal/Ass
Removal/Assembly
embly
11. Lift the Oxygen Sensor Block (C) from the Bottom Foam.
12. Remove the Oxygen Sensor Block (C).
10 Components Removal/Ass
Removal/Assembly
embly
2. Remove
10.3.2). the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assem-
bly Section 10.3.3
10.3.3).
).
4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly
Removal/Assembly Section
10.4.5).
10.4.5).
5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8
10.4.8).
).
6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly
Removal/Assembly Section 10.4.9
10.4.9).
).
7. Remove the Blower Module (refer to the Blower Module Removal/Asse
Removal/Assembly
mbly Section 10.4.10
10.4.10).
).
8. Remove the Power Supply (refer to the Power Supply Removal/Assembly Section 10.4.11
10.4.11).
).
9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly Section 10.4.12
10.4.12).
).
10. Remove the top tubing (A) from the Oxygen Sensor Block (refer to the Oxygen Sensor Block Removal/Assembly
Removal/Assembly Section
10.4.17).
10.4.17).
11. Remove the Rinse Flow Tube (B) from the Inspiratory Valve.
12. Tubings (C and E) should be already disconnected from the Pressure Sensor
Sensor Board.
13. Remove the Nebulizer Tube (D).
14. Disconnect the Qvent Flow Sensor cable and remove from the mount on the Rinse Flow Block.
15. Lift the Rinse Flow Block (F) from the Bottom Foam.
Hamilton Medical | HAMILTON-C2 Service Manual 289
10 Components Removal/Ass
Removal/Assembly
embly
CAUTION
Update the Technical State (refer to the Modify Tab 8.8.3).
8. Remove the Power Supply (refer to the Power Supply Removal/Assembly Section
Section 10.4.11
10.4.11).
).
9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly
Removal/Assembly Section 10.4.12
10.4.12).
).
10. Lift the Expiratory Valve Assembly (A) from the Bottom Foam Section of the Ventilation Unit.
10 Components Removal/Ass
Removal/Assembly
embly
3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assem-
bly 10.3.3
10.3.3).
).
4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly
Removal/Assembly
10.4.5).
10.4.5).
5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly 10.4.8
10.4.8).
).
6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly
Removal/Assembly 10.4.9
10.4.9).
).
7. Remove the Blower Module (refer to the Blower Module Removal/Asse
Removal/Assembly
mbly 10.4.10
10.4.10).
).
8. Remove the Power Supply (refer to the Power Supply Removal/Assembly 10.4.11
10.4.11).
).
9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly 10.4.12
10.4.12).
).
10. Remove 2 Torx screws (C, PN 420699) to remove the Cable Clamp (B, PN 160405) from the rear of the Battery Compart-
ment (A).
Battery Compartme
Compartment
nt Removal/Ass
Removal/Assembly
embly 10
10 Components Removal/Ass
Removal/Assembly
embly
CAUTION
Update the Technical State (refer to the Modify Tab 8.8.3). Also update the Technical State of the QO2 Flow Sensor.
10 Components Removal/Ass
Removal/Assembly
embly
2. Remove
). the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
10.3.2
3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assem-
bly Section 10.3.3
10.3.3).
).
4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly
Removal/Assembly Section
10.4.5).
10.4.5).
5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8
10.4.8).
).
6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly
Removal/Assembly Section 10.4.9
10.4.9).
).
7. Remove the Blower Module (refer to the Blower Module Removal/Asse
Removal/Assembly
mbly Section 10.4.10
10.4.10).
).
8. Remove the Power Supply (refer to the Power Supply Removal/Assembly Section 10.4.11
10.4.11).
).
9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly Section 10.4.12
10.4.12).
).
10. Remove the Oxygen Mixer Block Assembly (refer to the Oxygen Mixer Block Assembly Removal/Assembly Section
Section
10.4.21).
10.4.21).
11. Remove 2 Phillips screws (A) from the Nebulizer Valve (B).
12. Remove the Nebulizer Valve (B).
2. Remove
). the Interaction Panel Small Rear Cover (refer to the Interaction Panel Small Rear Cover Removal/Assembly Section
10.3.2).
10.3.2
3. Optional, if needed, remove the Interaction Panel from the Ventilation Unit (refer to the Interaction Panel Removal/Assem-
bly Section 10.3.3
10.3.3).
).
4. Remove the Front and Rear Covers from the Ventilation Unit (refer to the Front and Rear Covers Removal/Assembly Section
10.4.5).
10.4.5).
5. Remove the Top Foam from the Ventilation Unit (refer to the Top Foam Removal/Assembly Section 10.4.8
10.4.8).
).
6. Remove the Cooling Fan (refer to the Cooling Fan Removal/Assembly Section 10.4.9
10.4.9).
).
7. Remove the Blower Module (refer to the Blower Module Removal/Asse
Removal/Assembly
mbly Section 10.4.10
10.4.10).
).
8. Remove the Power Supply (refer to the Power Supply Removal/Assembly Section
Section 10.4.11
10.4.11).
).
9. Remove the Middle Foam (refer to the Middle Foam Removal/Assembly
Removal/Assembly Section 10.4.12
10.4.12).
).
10. Remove the Oxygen Mixer Block Assembly (refer to the Oxygen Mixer Block Assembly Removal/Assembly Section
10.4.21).
10.4.21).
11. Remove the Bottom Foam (A) from the Ventilation Unit Base Frame.
10 Components Removal/Ass
Removal/Assembly
embly
10 Components Removal/Ass
Removal/Assembly
embly
10 Components Removal/Ass
Removal/Assembly
embly
2. The Cable will route on the side of the Middle and Top Foam Sections when assembled.
3. It will then be connected to the Ventilation Unit Mainboard when assembled.
302 English | 624165/07
2. The Cable will route on the side of the Middle and Top Foam Sections when assembled.
3. It will then be connected to the Ventilation Unit Mainboard when assembled.
Hamilton Medical | HAMILTON-C2 Service Manual 303
10 Components Removal/Ass
Removal/Assembly
embly
2. The Expiratory Valve FFC (D) (part of the Expiratory Valve, no part number) is routed under a channel in the Bottom Foam
Section to the side.
10 Components Removal/Ass
Removal/Assembly
embly
2. Connect the Proximal Flow Sensor Tubing (B, from Tubing Set MSP160476) and the Distal Flow Sensor Tubing (D, from
Tubing Set MSP160476) to the Rinse Flow Block.
3. Connect the Nebulizer Tubing (C, PN 160409) from the Mixer Block Fitting and the Nebulizer Connector to the Rinse Flow
Block.
5. The other tubing ends (B and D) will connect to the Pressure Sensor Assembly when assembled.
assembled.
306 English | 624165/07
10 Components Removal/Ass
Removal/Assembly
embly
10.5.10 Inspiratory Valve, Qvent Flow Sensor and Ambient Valve Assembly
CAUTION
Update the Technical State (refer to the Modify Tab 8.8.3).
NOTICE
The Ambient Valve cable (F) is part of the Ambient Valve Assembly and has no part number. The Inspiratory Valve FFC (G) is
part of the Inspiratory Valve and has no part number.
1. Pre-assemble the Inspiratory Valve (A, MSP160230), Qvent Flow Sensor (B, MSP161657 (Sensirion) or MSP399124 (TSI)),
Ambient Valve (C, MSP160290), Shaped Tube (D, PN 160223) and Patient Connection Tube (E, PN 160295).
2. Not shown; a metal bracket (PN 160741) attached with 2x screws (PN 420636) secures the Qvent Flow Sensor to the
Ambient Valve.
Figure 10-64. Inspiratory Valve, Qvent Flow Sensor and Ambient Valve Assembly Step 1
3. Insert the complete assembly (I) into the Bottom Foam Section (J).
Hamilton Medical | HAMILTON-C2 Service Manual 309
10 Components Removal/Ass
Removal/Assembly
embly
Figure 10-65. Inspiratory Valve, Qvent Flow Sensor and Ambient Valve Assembly Step 2
4. Connect the tubing (K) from the Front Panel Connector Block to the Inspiratory Valve Assembly.
Assembly.
5. Place the Qvent Flow Sensor cable into the Cable Holder (L) located on top of the Front Panel Connector Block.
6. The cable will route on the side of the Middle and Top Foam Sections when assembled.
7. It will then be connected to the Ventilation Unit Mainboard when assembled.
310 English | 624165/07
8. The Inspiratory Valve FFC (M) (part of the Inspiratory Valve, no part number) is routed under a channel in the Bottom Foam
Section to the side.
Figure 10-66. Inspiratory Valve, Qvent Flow Sensor and Ambient Valve Assembly Step 3
9. The cable will route on the side of the Bottom, Middle and Top Foam Sections when assembled.
10. It will then be connected to the Ventilation Unit Mainboard when assembled.
11. The cable from the Ambient Valve Assembly (part of the Ambient Valve, no part number) is routed through a channel (N)
in the Bottom Foam Section, over the Shaped Tube (O) and through another channel (P) to the side.
Figure 10-67. Inspiratory Valve, Qvent Flow Sensor and Ambient Valve Assembly Step 4
12. The cable will route on the side of the Middle and Top Foam Sections when assembled.
13. It will then be connected to the Ventilation Unit Mainboard when assembled.
10 Components Removal/Ass
Removal/Assembly
embly
2. Connect the Oxygen Sensor Cable (D, PN 160354) to the Oxygen Sensor (C, PN 396200).
3. The Cable will route on the side of the Middle and Top Foam Sections when assembled.
4. It will then be connected to the Ventilation Unit Mainboard when assembled.
5. Connect the tubing (E, PN 160475) from the side of the Oxygen Sensor Block to the Inspiratory Valve.
10 Components Removal/Ass
Removal/Assembly
embly
2. Assemble the Tube Flange (C, PN 160287) and screw onto the Shaped Tubes from the QO2 Flow Sensor.
NOTICE
NOTICE
10 Components Removal/Ass
Removal/Assembly
embly
NOTICE
Observe the air flow direction of the Fan. Flow direction is into the ventilator.
CAUTION
Before installing the Blower Module, ensure that the P-Filter cable will not be damaged.
CAUTION
Update the Technical State (refer to the Modify Tab Section Modify Tab).
NOTICE
1. Insert the Blower Module (A, MSP160250) into the Middle Foam Section (B).
10 Components Removal/Ass
Removal/Assembly
embly
2. Position the Blower Cable (C) (part of the Blower Module, no part number) and the Temperature Sensor FFC (D, PN
160353) to the side.
3. The cables will route on the side of the Top Foam Section when assembled.
4. The cables will then be connected to the Ventilation Unit Mainboard when assembled.
5. Position the Filter Pressure Sensor Cable (E) into the cutout of the Middle Foam Section.
CAUTION
Update the Technical State (refer to the Modify Tab Section 8.8.3).
1. Position the Top Foam (A, PN 160239) onto the Middle Foam (B).
Figure 10-78. Top Foam and Mainboard Assembly Step 1
2. Insert the Mainboard (C, MSP160200 or MSP160382) onto the Top Foam.
Figure 10-79. Top Foam and Mainboard Assembly (Front View) Step 2
Hamilton Medical | HAMILTON-C2 Service Manual 319
10 Components Removal/Ass
Removal/Assembly
embly
3. Route the Fan 12V Cable (D) into the side of the Top Foam and plug into Connector P28 on the Mainboard.
Figure 10-80. Top Foam and Mainboard Assembly (Left Side View) Step 3
4. Route the Blower Cable (E) into the side of the Top Foam and plug into Connector P23 on the Mainboard.
5. Route the Temperature Sensor FFC (F) into the side of the Top Foam and plug into Connector J21 on the Mainboard.
6. Route the Inspiratory Valve FFC (G) into the Bottom, Middle and Top Foam and plug into Connector J22 on the Main-
board.
7. Route the Oxygen Sensor Cable (H) into the front of the Middle and Top Foam and plug into Connector P20 on the Main-
board.
Figure 10-81. Top Foam and Mainboard Assembly (Front View) Step 4
8. Route the Qvent Flow Sensor Cable (I) into the front of the Middle and Top Foam and plug into Connector P19 on the
Mainboard.
9. Route the Pressure Sensor Board FFC (J) into the front of the Middle and Top Foam and plug into Connector J17 on the
Mainboard.
10. Route the Nebulizer Valve Cable (K) into the side of the Middle and Top Foam and plug into Connector P13 on the Main-
board.
Figure 10-82. Top Foam and Mainboard Assembly (Right Side View) Step 5
11. Route the Ambient Valve Cable (L) into the side of the Middle and Top Foam and plug into Connector P12 on the Main-
board.
12. Route the Expiratory Valve FFC (M) into the side of the Bottom, Middle and Top Foam and plug into Connector J11 on
the Mainboard.
13. Route the O2 Valve Cable (N) into the side of the Middle and Top Foam and plug into Connector P7 on the Mainboard.
14. Route the Filter Pressure Sensor FFC (O) into the side of the Middle and Top Foam and plug into Connector J9 on the
Mainboard.
15. Route the Binary Valve FFC (P) from the Pressure Sensor Assembly into the side of the Middle and Top Foam and plug
into Connector J8 on the Mainboard.
16. Route the O2 Flow Sensor Cable (Q) into the side of the Middle and Top Foam and plug into Connector P10 on the
Mainboard.
Hamilton Medical | HAMILTON-C2 Service Manual 321
10 Components Removal/Ass
Removal/Assembly
embly
17. Route the Battery Power Cable (R) into the rear of the Middle and Top Foam and plug into Connector P6 on the Main-
board.
Figure 10-83. Top Foam and Mainboard Assembly (Rear View) Step 6
18. Route the DC Input Cable (S) into the rear of the Middle and Top Foam and plug into Connector P5 on the Mainboard.
19. Route the Battery Data FFC (T) into the rear of the Middle and Top Foam and plug into Connector J4 on the Mainboard.
20. Route the Fan Supply Cable (U) into the rear of the Middle and Top Foam and plug into Connector P25 on the Main-
board.
21. Route the Power Supply Cable (V) into the rear of the Middle and Top Foam and plug into Connector P3 on the Main-
board.
22. Assemble the Front and Rear Covers (refer to the 10.4.5
10.4.5).
).
23. Assemble the Interaction Panel (refer to the Interaction Panel Removal/Assembly Section 10.3.3
10.3.3).
).
24. Assemble the Ventilation Unit Top Cover (refer to the 10.3.1
10.3.1).
).
322 English | 624165/07
11
Maintenance Tools and Test Equipment
11.1 Overview..............................................................................................................................................324
Overview.............................................................................................................................................. 324
11.2 Standard Tools ..................................................................................................................
.....................................................................................................................................
...................324
324
11.3 Special
Spec ial Tools .....
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
..........
....325
325
11.4 Test Equipment
Equipment....................................................................................................................................
....................................................................................................................................328
328
11.5 Calibratio
Calibration
n of Test Equip
Equipment..........
ment................
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
........332
323
11.1 Overview
Standard Tools, Special Tools, ElectroStatic Discharge (ESD) protection and Test Equipment detailed in the following sections
are required to carry out:
• The Preventive
Preventive Maintenance
Maintenance procedures
procedures in the Engineer Preventive
Preventive Maintenance
Maintenance Section
Section 6.3
• The Tests
Tests Functions
Functions in the Servic
Servicee Software
Software Section
Section 8
• The C
Component
omponent Replacement
Replacementss in the Components
Components Removal/Asse
Removal/Assembly
mbly Section
Section 10
Special Tools 11
• 11-3 shows a wrist band and connecting cable complete with the crocodile clip that must be attached to the work surface
Figure 11-3. ESD Wrist Strap and Cable (PN 500009)
Contains:
• 5003
500308
08 TSI
TSI Flow
Flow mete
meterr
• 5000
500085
85 TSI-Flow
TSI-Flow meter
meter Battery
Battery box
box
• 500086
500086 TSI-Flow
TSI-Flow meter
meter Soft
Soft carrying
carrying
case
• 279204
279204 Bacteria
Bacteria filter
filter 260100
260100 Sili-
Sili-
cone tube 30cm 22F
- Digital Voltmeter
(With red and black crocodile
cr ocodile clips)
Test Equipment 11
Test Equipment 11
Digital Vo
Voltmeter (D
(DVM) As re
recommended by
by th
the ma
manufacturer. As re
recommended by
by ma
manufacturer.
12
Spare Parts
12.1 Introduction
Introductio n to Major Comp
Componentonentss ......
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
..........
....334
334
12.2 Trolley PN 160170................................................................................................................................
160170................................................................................................................................335335
12.3 Trolley PN 160150................................................................................................................................
160150................................................................................................................................336336
12.4 Major Comp
Component
onentss of the Inter
Interacti
action
on Pane
Panell .....
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
.........
...337
337
12.5 Major Comp
Component
onentss of the Vent
Ventilatio
ilationn Unit........
Unit..............
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
........338
12.6 Interaction Panel Parts..........................................................................................................................
Parts..........................................................................................................................343343
12.7 Ventilatio
Venti lation
n Unit Part
Partss ......
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
............
............
............
...........
...........
............
............
........348
12.8 Detailed
Deta iled Part
Partss Desc
Descriptio
riptions
ns ......
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
...........
.....355
355
333
12 Spare Parts
NOTICE
− For information about consumables (such as tubing) and complete assemblies (such as the Patient Tubing Support Arm),
see the Hamilton Medical Product Catalog (PN 689060). It can be located on the Hamilton Medical AG Partner-net.
− Some photographs shown in later sections may not be displayed to scale.
− Some Spare Parts have a prefix MSP (Medical Spare Part). If you exchange a MSP Part, you usually have to update the
Technical State of the device with the new Serial Number and Revision Number.
A Ven
Venti
tila
lati
tion
on Uni
nitt (P
(PN
N 160
6000
001)
1) B Int
nter
eraact
ctio
ion
n Pane
nell (P
(PN
N 16
160
032
325
5)
See the Ventilation Unit Parts Section 12.7 See the Interaction Panel Parts Section 12.6
334 English | 624165/07
Trolley PN 160170 12
12 Spare Parts
A LCD Di
Display (P
(PN 38
380046) E Inte
Intera
ract
ctio
ion
n Pan
Panel
el Sm
Smal
alll Re
Rear
ar Co
Cove
verr (PN
(PN 16
1604
0402
02))
See the LCD Display and Touch Screen Section 12.6.4 See the Interaction Panel Covers Section 12.6.1
C P&TT Con
P& Contr
trol
ol Kn
Knob
ob Co
Comp
mple
lete
te (M
(MSP
SP16
1603
0328
28)) G Fro
Front Pan
anel
el Bo
Boar
ard
d (M
(MSP
SP16
1601
019
96)
See the P&T Control Knob Section 12.6.5 See the Front Panel Board Section 12.6.3
Hamilton Medical | HAMILTON-C2 Service Manual 337
12 Spare Parts
A Ven
Venti
tila
lati
tion
on Uni
Unitt Cove
Covers
rs C Vent
Ventililat
atio
ion
n Un
Unit
it Mi
Midd
ddle
le Se
Sect
ctio
ion
n
See the Ventilation Unit Covers Section 12.5.1 See the Ventilation Unit Middle Section 12.5.3
B Vent
Ventil
ilat
atio
ion
n Uni
Unitt Top
Top Se
Sect
ctio
ion
n D Vent
Ventililat
atio
ion
n Un
Unit
it Bo
Botto
ttom
m Se
Sect
ctio
ion
n
See the Ventilation Unit Top Section 12.5.2 See the Ventilation Unit Bottom Section 12.5.4
338 English | 624165/07
A Vent
Ventil
ilat
atio
ion
n Uni
Unitt Filt
Filter
er Co
Cove
verr (PN
(PN 160
16034
343)
3) D Vent
Ventil
ilat
ator
or Un
Unit
it To
Top
p Cov
Cover
er (P
(PN
N 160
16060
601)
1)
See the External Covers and External Hardware Section See the External Covers and External Hardware Section
External Covers and External Hardware External Covers and External Hardware
B Comp
Co mple
lete
te Rea
earr Cov
Coveer (MS
(MSP1
P160
604
481
81)) E Front Cover (M
(MSP160480)
See the External Covers and External Hardware Section See the External Covers and External Hardware Section
External Covers and External Hardware External Covers and External Hardware
C Batte
terry Cover (PN 160506) F Ventilation Unit Base
See the External Covers and External Hardware Section See the Metal Brackets and Frame Components Section
External Covers and External Hardware Metal Brackets and Frame Components
Hamilton Medical | HAMILTON-C2 Service Manual 339
12 Spare Parts
A Mainboard
rd--C2 (M
(MSP160382) C Top Foam (PN 160239)
See the Electronic Printed Circuit Boards Section 12.8.11 See the Metal Brackets and Frame
Fr ame Components Section
12.8.9
B Embe
Embedd
dded
ed Sy
Syst
stem
em Mo
Modu
dule
le (E
(ESM
SM)) (MS
(MSP1
P160
6020
206)
6)
See the Electronic Printed Circuit Boards Section 12.8.11
340 English | 624165/07
C Middle Fo
Foam (P
(PN 16
160238)
See the Metal Brackets and Frame Components Section
12.8.9
Hamilton Medical | HAMILTON-C2 Service Manual 341
12 Spare Parts
A Batt
Batteery Com
omp
par
artm
tmen
entt (PN
(PN 16
1603
030
02) G Pres
Pressu
sure
re Se
Sens
nsor
or As
Asse
semb
mbly
ly (M
(MSP
SP16
1603
0300
00))
See the Backup Battery Pack Compartment Section 12.7.5 See the Electronic Printed Circuit Boards Section 12.8.11
B Insp
Inspir
irat
ator
oryy Val
Valve
ve Co
Comp
mple
lete
te (M
(MSP
SP16
1602
0230
30)) H Mixe
Mixerr Blo
Block
ck As
Asse
semb
mbly
ly (M
(MSP
SP16
1606
0608
08 or MS
MSP1
P160
6022
226)
6)
See the Assembled Components Section 12.8.6 See the Mixer Block Assembly Section 12.7.7
C Qventt Flow
Qven Flow Sen
Senso
sorr (MSP
(MSP16
1616
1657
57,, Sens
Sensir
irio
ion)
n) or
or (MSP
(MSP39
3991
9124
24,, I QO2 Flow Sensor
TSI) See the Pneumatic Parts and Assemblies Section 12.8.4
See the Flow Restrictors and Flow Sensors Section 12.8.5
D Oxyg
Oxygen
en Se
Sens
nsor
or HA
HAMI
MILT
LTON
ON-C
-C2
2 (co
(code
ded)
d) (P
(PN
N 396
39620
200)
0) J Expi
Expira
rato
tory
ry Va
Valv
lvee As
Asse
semb
mbly
ly (M
(MSP
SP16
1602
0240
40))
See the Electrical/Electronic Devices Section 12.8.12 See the Expiratory Valve Assembly Section 12.7.6
E Rin
Rinse Fl
Flow
ow Ass
sseembl
blyy (M
(MSSP1
P160
604
498
98)) K Ambi
Ambien
entt Va
Valv
lvee Co
Comp
mple
lete
te (M
(MSP
SP16
1602
0290
90))
See the Assembled Components Section 12.8.6 See the Assembled Components Section 12.8.6
F Bottom Fo
Foam (P
(PN 16
160237)
See the Metal Brackets and Frame Components Section
12.8.9
A Inte
Intera
ract
ctio
ion
n Pan
Panel
el La
Larg
rgee Rea
Rearr Cov
Cover
er (P
(PN
N 160
16032
326)
6) D Inte
Intera
ract
ctio
ion
n Pan
Panel
el Sm
Smal
alll Rea
Rearr Cov
Cover
er (P
(PN
N 160
16040
402)
2)
See the External Covers and External Hardware Section See the External Covers and External Hardware Section
12.8.13 12.8.13
12 Spare Parts
A DC/AC
DC/A C Inv
Inver
erte
terr Boar
Board
d (PN
(PN 3961
396197
97 (Ob
(Obso
sole
lete
te)) or PN 396
39622
229
9 D 2x Screws (P
(PN 420663)
(Obsolete) See the Screws Section 12.8.14
Backlight LED Driver Board (PN 396314, no longer available)
See the Electronic Printed Circuit Boards Section 12.8.11
C 2x Standoffs (PN 257076) Bypass Cable (PN 161569) for Front Panel Board Revision 08
See the Standoffs Section 12.8.15 or higher for the LED Backlights
See the Electrical/Electronic Cables Section 12.8.10
NOTICE
Please check compatibility for HAMILTON-C2 Serial Number <1676, see Interaction Panel Section 15.2.1
344 English | 624165/07
A Fron
Frontt Pane
nell Boa
oard
rd (M
(MSSP16
160
019
196)
6) E FFC
FFC Ca
Cabl
blee to Ke
Keyy Pa
Pane
nell (P
(PN
N 16
1603
035
56)
See the Electronic Printed Circuit Boards Section 12.8.11 See the Electrical/Electronic Cables Section 12.8.10
D Bypass
Bypa ss Ca
Cabl
blee (PN
(PN 16
1615
1569
69)) to LE
LED
D Bac
Backl
klig
ight
htss
See the Electrical/Electronic Cables Section 12.8.10
NOTICE
12 Spare Parts
A Display Fr
Front (M
(MSP160362) I Display Gasket (PN 160341)
See the Electronic Printed Circuit Boards Section 12.8.11 See the Metal Brackets and Frame
Fr ame Components Section
12.8.9
B Disp
Displa
layy Mo
Moun
unti
ting
ng Br
Brac
acke
kett (P
(PN
N 16
1604
0404
04)) J FDC Ca
Cable to
to Di
Display (P
(PN 16
160357)
See the Metal Brackets and Frame Components Section See the Electrical/Electronic Cables Section 12.8.10
12.8.9
C 4x Sc
Screws (PN 42
420724) K USB Cover (P
(PN 341548)
See the Screws Section 12.8.14 See the Clamps and Fasteners Section 12.8.2
E Graphi
Graphicc LCD
LCD TF
TFTT VGA
VGA 10
10.4
.4”
” NEC
NEC (P
(PN
N 380
38002
027
7 or
or 380
38003
031
1 M 2x Fe
Ferr
rriite Slee
eeve
vess (P
(PN 393
39311
113
3)
(Obsolete))
Graphic LCD TFT VGA 10.4" NLT (PN 380046, new)
See the Electronic Printed Circuit Boards Section 12.8.11
H 1x St
Standoff (P
(PN 257038) ESD Isolation Sticker on Monitor (PN 160467)
See the Standoffs Section 12.8.15
NOTICE
Refer to KB ID 3315 if you have to exchange an old version of LCD Display (PN 380027 or PN380031), or an old version of
DC/AC Inverter Board (PN 396197 or PN 396229)
Please see Interaction Panel Section 15.2.1 for Display compatibility.
A P&TT Con
P& Contr
trol
ol Kn
Knob
ob co
comp
mple
lete
te (M
(MSP
SP16
1603
0328
28)) C Washer (PN 409908)
See the Colored Metal Rings and Controls Section 12.8.7 See the Washers Section 12.8.16
B P&TT Co
P& Cont
ntro
roll Kn
Knob
ob En
Enco
code
derr (P
(PN
N 37
3720
2036
36))
See the Electrical/Electronic Devices Section 12.8.12
Hamilton Medical | HAMILTON-C2 Service Manual 347
12 Spare Parts
A Vent
Ventil
ilat
ator
or Un
Unit
it To
Top
p Cov
Cover
er (P
(PN
N 160
16060
601)
1) C 2x Screws (P
(PN 420682)
See the External Covers and External Hardware Section See the Screws Section 12.8.14
12.8.13
Front Cover 12
A Vent
Ventililat
atio
ion
n Unit
Unit Fro
Front
nt Cov
Cover
er wit
with
h Labe
Labell (MSP
(MSP16
1604
0480
80)) G Inte
Intera
ract
ctio
ion
n Pan
Panel
el Hi
Hing
ngee Typ
Typee B Ri
Righ
ghtt (PN
(PN 28
2817
1782
82))
See the External Covers and External Hardware Section See the Metal Brackets and Frame Components Section
12.8.13 12.8.9
C Inte
Intera
ract
ctio
ion
n Pan
Panel
el Hi
Hing
ngee Typ
Typee A Le
Left
ft (P
(PN
N 281
28178
781)
1) I 2x Screws (PN 420657)
See the Metal Brackets and Frame Components Section See the Screws Section 12.8.14
12.8.9
12 Spare Parts
A P-Cl
P-Clam
amp
p / Sc
Scre
rew
w (PN
(PN 36
3610
1061
61 / PN
PN 420
42072
724
4) G Ethern
rneet Co
Cover (P
(PN 16
160509)
See the Clamps and Fasteners Section 12.8.2 See the Clamps and Fasteners Section 12.8.2
C Stickers (P
(PN 16
160020) I Vent
Ventil
ilat
ator
or Uni
Unitt Rear
Rear Co
Cove
verr comp
comple
lete
te (M
(MSP
SP16
1604
0481
81))
See the Stickers and Labels Section 12.8.3 See the External Covers and External Hardware Section
12.8.13
D Blan
Blankk Cov
Cover
er fo
forr Com
Commu
muni
nica
cati
tion
on Bo
Boar
ard
d (PN
(PN 16
1604
0431
31)) J 4x Screws (PN 420667)
See the External Covers and External Hardware Section See the Screws Section 12.8.14
12.8.13
E 2x Screws (PN 420730) K Cable Support (PN 160408)
See the Screws Section 12.8.14 See the External Covers and External Hardware Section
12.8.13
A Base Pl
Plate (M
(MSP160228) D Base Frame (PN 160483)
See the Metal Brackets and Frame Components Section See the Metal Brackets and Frame Components Section
12.8.9 12.8.9
12 Spare Parts
A Batt
Batteery Com
omp
par
artm
tmen
entt (PN
(PN 16
1603
030
02) B Battery LI-ION (M
(MSP369106)
See the Electrical/Electronic Devices Section 12.8.12 See the Electrical/Electronic Devices Section 12.8.12
NOTICE
A Expi
Expira
rato
tory
ry Va
Valv
lvee Com
Compl
plet
etee (MS
(MSP1
P160
6024
240)
0) C Expi
Expira
rato
tory
ry Va
Valv
lvee Sil
Silic
icon
onee Mem
Membr
bran
anee (PN
(PN 16
1605
0500
00))
See the Assembled Components Section 12.8.6 See the Rubber Seals and Grommets Section 12.8.8
B Expi
Expira
rato
tory
ry Va
Valv
lvee Sea
Seall wit
with
h Mem
Membr
bran
anee (PN
(PN 16
1602
0245
45))
See the Assembled Components Section 12.8.6
Hamilton Medical | HAMILTON-C2 Service Manual 353
12 Spare Parts
A O2 Mi
Mixer As
Assembly Co
Complete G O2 - NI
NIST Con
Conne
nect
ctor
or (P
(PN
N 160
1604
471
71))
(MSP160226 with TSI QO2 Flow Sensor) See the Tubings and Fittings Section 12.8.1
(MSP160608 with Sensirion QO2 Flow Sensor)
See the Pneumatic Parts and Assemblies Section 12.8.4
C Mixer Block Cover Plate (PN 160459) I Nebulizer Valve (PN 160400)
See the Pneumatic Parts and Assemblies Section 12.8.4
E O2 - DIS
ISSS Con
onne
nect
ctor
or (P
(PN
N 16
160
047
470
0)
See the Tubings and Fittings Section 12.8.1
Polyurethane Tubing O2 blue 4x6 (PN 160474) (see Tubing and Fittings Section 12.8.1
12.8.1).
).
354 English | 624165/07
12 Spare Parts
MSP160498 TUBING SE
SET (N
(Nebulizer Ri
Rinse Fl
Flow)
356 English | 624165/07
12 Spare Parts
160020 STICKERS
255358 STICKER
MSP1
MSP160
6022
226
6 O2 MI
MIXE
XER
R ASS
ASSEM
EMBL
BLY
Y wi
with
th TS
TSII FLO
FLOW
W SE
SENS
NSOR
OR (c
(com
ompl
plet
ete)
e)
• Use Cable (PN 160387) from Oxygen Flow Sensor to Mainboard
Mainboard
MSP160382
• Use Cable (PN 160373) from Oxygen Flow Sensor to Mainboard
Mainboard
MSP160200
Includes:
• Mixe
Mixerr Assemb
Assembly
ly Compl
Complete
ete
• TSI O2
O2 Flow Sensor
Sensor with
with Cable
Cable
• Nebu
Nebulize
lizerr Valve
Valve with
with Cable
Cable
• Req
Requir
uired
ed Scre
Screws
ws
• Ca
Cabl
blee Tie
Tie
NOTICE! Only compatible when the TSI Qvent Flow Sensor MSP399123
is installed.
MSP1
MSP160
6060
608
8 O2 MIX
MIXER
ER ASS
ASSEM
EMBLBLY
Y with
with SEN
SENSI
SIRI
RION
ON FlO
FlOW
W SEN
SENSO
SOR
R (com
(compl
plet
ete)
e)
NOTICE! Installed in the C2 from SN 9216
Includes:
• Mixe
Mixerr Assemb
Assembly
ly Compl
Complete
ete
• Sens
Sensirio
irion
n O2 Flow Sensor
Sensor with
with Cable
Cable
• Nebu
Nebulize
lizerr Valve
Valve with
with Cable
Cable
• Req
Requir
uired
ed Scre
Screws
ws
• Ca
Cabl
blee Tie
Tie
• O2 Con
Conne
nectctor
or
• DC Cabl
Cablee
12 Spare Parts
Flow Sensors 12
CAUTION
Always use Flow Sensors from the same brand in one ventilator (Sensirion or TSI).
MSP161657 SENSIRION FL
FLOW SE
SENSOR AI
AIR (Q
(Qvent)
12 Spare Parts
MSP160230 INSPIRATORY VA
VALVE CO
COMPLETE
160500 EXPIRATORY VA
VALVE SI
SILICONE ME
MEMBRANE 5 PIECES
Hamilton Medical | HAMILTON-C2 Service Manual 363
12 Spare Parts
12 Spare Parts
12.8.10 Electrical/Elect
Electrical/Electronic
ronic Cables
Table 12-10. Electrical/Ele
Electrical/Electronic
ctronic Cables
Part Number Description Photo
Electrical/Electronic
Electrical/Electronic Cables 12
160358 CABLE TO BA
BACKLIGHT DC/AC INVERTER BOARD (PN 39
396197 ol
old or PN
PN
396229 new)
161569 CABLE TO
TO GR
GRAPHIC LC
LCD TF
TFT VG
VGA 10
10.4” NL
NLT DI
DISPLAY BA
BACKLIGHT LE
LED's
(Bypass for the Backlight Driver Board)
160359 FFC CABLE TO BINARY VALVES
12 Spare Parts
MSP160200 MAINBOARD-C2
Includes:
• Disp
Display
lay Cable
Cable Holder
Holder (PN
(PN 160457)
160457)
• Dist
Distance
ance Holde
Holderr 6mm (PN
(PN 362051)
362051)
MSP160382 MAINBOARD-C3
(used in the Hamilton-C2)
Includes:
• Disp
Display
lay Cable
Cable Holder
Holder (PN
(PN 160457)
160457)
• Dist
Distance
ance Holde
Holderr 6mm (PN
(PN 362051)
362051)
• 2x Cables
Cables to Flow Sensor (QO2
(QO2 and Qvent)
Qvent) (PN 160387)
12 Spare Parts
Includes:
• To
Touc
uch
h Scre
Screen
en
• Ke
Keyy Pan
Panel
el
• Disp
Display
lay Gaske
Gaskett (PN 16034
160341)
1)
• Symb
Symbol
ol Insert
Insert (PN
(PN 160376)
160376)
NOTICE! Obsolete
NOTICE! Obsolete
For replacement, refer to the Interaction Panel Display Section 15.2.1.1 for
compatibility.
160140 EXTENDED CO
COMMUNICATION BOBOARD CO
CO2, SP
SPO2 (N
(Not im
implemented in
in
HAMILTON-C2) AND NURSE CALL
12 Spare Parts
12.8.12 Electrical/Elect
Electrical/Electronic
ronic Devices
Table 12-12. Electrical/Ele
Electrical/Electronic
ctronic Devices
Part Number Description Photo
160302 BATTERY CO
COMPARTMENT
Includes:
• PN 160370
160370 Cable
Cable to Batter
Batteryy Power
Power
• PN 160351
160351 FFC
FFC Cable
Cable to Battery
Battery Data
Data
161532 REPLACEMENT SP
SPEAKER
MSP369106 BATTERY LI
LI-ION 14.4V/6.8Ah
Electrical/Ele
Electrical/Electronic
ctronic Devices 12
12 Spare Parts
MSP
SP16
1604
0480
80 VEN
VENTIL
ILA
ATOR UNI
NITT FR
FRON
ONTT CO
COVE
VER
R wit
ith
h La
Lab
bel
MSP
SP16
1604
0481
81 VEN
VENTIL
ILA
ATOR UNI
NITT RE
REAR CO
COVE
VER
R CO
COMP
MPLLET
ETEE (se
(seee 12.7.3
12.7.3))
12 Spare Parts
12.8.14 Screws
Table 12-14. Screws
Part Number Description Photo
Screws 12
12 Spare Parts
420710 TORX FI
FILLISTER HE
HEAD SC
SCREW 25
25x8 D1
D1=2.5, L=
L=8 (G
(Galvanized bl
blue)
420720 TORX FI
FILLISTER HE
HEAD SC
SCREW I-I-6R M3
M3x10 (G
(Galvanized bl
blue st
steel)
420724 TORX FILLISTER HEAD SCREW 30x8
Standoffs 12
12.8.15 Standoffs
Table 12-15. Standoffs
Part Number Description Photo
12 Spare Parts
12.8.16 Washers
Table 12-16. Washers
Part Number Description Photo
13
Schematics
13 Schematics
Wiring Diagram 13
13 Schematics
384 English | 624165/07
14
Software Revisions, Features and
Compatibility
14.1 Introduction
Introduction .....
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
...........
.....386
386
14.2 Software Version Information...............................................................................................................
Information...............................................................................................................386 386
385
14.1 Introduction
NOTICE
For actual information and latest software, log in to Hamilton Medical AG Partner-net. There are the latest information
about upgrading and updating a HAMILTON-C2.
This section gives an overview of all published HAMILTON-C2 Software. It includes information about new features intro-
duced with software upgrades and software updates information about compatibility between software of different versions.
The improvement list is not meant to be complete and covers only the most relevant changes.
WARNING
The (*) indicates an obsolete software version. Mandatory update must be performed.
• Battery
Battery SoH added
added to the Instrument
Instrument Report
Report
– The Battery State of Health (SoH)
(SoH) information
information is added
added to the instrument
instrument
report in the new software
• FP
FPGA
GA up
upda
date
te
– A new internal Field Programmable
Programmable Gate Array (FPGA) clockclock distribution
distribution
improves the reliability of the integrated circuit boards
• Blow
Blower
er time
time can be updated
updated
– The blower
blower time
time can be
be set again
again as it was before
before
• Extend log files CPU temperature and
and POST will be logged in the new software
• Wrong 'positive
'positive oxygen supply failed'
failed' alarm solved
• Replace
Replace Oxyge
Oxygen
n Sensor
Sensor
• Suc
Suctio
tionin
ning
g man
maneuv
euver
er
• Tec
Techni
hnical
cal err
error
or
• Techn
Technical
ical state fail
failed
ed
• Tou
Touch
ch not funct
function
ional
al
• Unk
Unknow
nown
n part numbe
numberr
• Rele
Release
ase valve defe
defective
ctive
Replaced the numerical code with text and added troubleshooting information to
the Operator’s Manual for the following technical alarms:
• Batte
Battery
ry communicat
communication
ion error
error (244011)
(244011)
• Batte
Battery
ry totally
totally discharged
discharged (444001
(444001))
• Blow
Blower
er faul
faultt (43100
(431001)
1)
• Buzze
Buzzerr defecti
defective
ve (24300
(243004)
4)
• Louds
Loudspeak
peaker
er defective
defective (24300
(243003)
3)
• Optio
Options
ns not
not found
found (249001
(249001))
• Self test fail
failed
ed (485002
(485002))
• Blo
Blower
wer lif
lifeti
etime
me
Ventilation mode:
• The ambient and the safety mode is not displayed any more as regular TF
xxx’xxx In case of a TF leading to ambient mode "Ventilation cancelled"
cancelled" will be
displayed.
• In case of a TF leading
leading to safety mode "Safety ventilation"
ventilation" will be displayed
displayed
• Trend time
time has been
been extended
extended to 72
72 hours
• Suctioning tool is now displayed in the options list
list
• Occurrence of watchdog failures
failures has been reduced
reduced again
• Known technical faults during suctioning
suctioning maneuver
maneuver do not occur
occur anymore
• GUI "black
"black holes"
holes" problem
problem solved
solved
• RTC time will
will not be lost
lost during ventila
ventilation
tion
• Technical event: 243005 (loudspeaker
(loudspeaker sounds continual)
continual) will only be checked
checked
during start-up
• Technical event: 232037 (Unknown alarm)
alarm) during the CO2 calibration will
will not
appear anymore
• The sequence
sequence of the component test
test has been improved
improved
• Ser
Servic
vicee softwar
software:
e:
– Tech
Technica
nicall state now displays
displays all hardwa
hardware
re revisions
revisions in one window
window
– Pressure calibration
calibration procedure
procedure is adapted to the new expiratory valve
valve cover
with the metal inlay: With the expiratory valve cover (revision 02) the
pressure inside the breathing circuit can no longer be adjusted to 50 mbar
(51 cmH2O), because the new covers are significantly tighter than the
previous version. Therefore the pressure will be calibrated by comparing
the external reference (pressure gauge) with the internal measurement
displayed on screen (Pvent_control). Furthermore the saving of the gain
value (adjustable with turn knob on screen) will be disabled if the
difference between the Pvent_control and Pvent_monitor resp. Pressure
sensor Paw value is more than 3 mbar.
– Technical fault and event during performing the service software
software will be
suppressed
– Pneumatics 1 -> Exp. valve test: The test limiting
limiting values could be enlarged
– The sequen
sequence
ce of the componen
componentt test has been
been improved
improved
• The software
advice version
version
to updated 2.0.3 in
systems is the
still considered to be safe.
safe.
field with software However,
version 2.0.4 we
for would
cus-
tomers ventilating their patients intensively with DuoPAP or APRV.
• seve
several
ral other,
other, minor
minor adjustme
adjustments
nts
NOTICE! The software 1.1.4 (1.0.5 for Japan) MUST be installed to all devices
in the field according to the information you received in the partner letter April
6th, 2010.
• Autozero procedure:
procedure: Due to internal
internal investigation
investigation the autozero
autozero interval takes
takes
place more frequently to improve the
t he flow measurement performance.
• Event Log: All technical
technical faults generated
generated while performing
performing the Service Software
Software
tests will no longer be registered in the user event log as numerical codes.
• Syste
System
m test O2 mixer
mixer impleme
implemented
nted
• Syste
System
m test flow
flow implem
implemented
ented
15
Hardware Revisions, Features and
Compatibility
15.1 Introduction
Introduction .....
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
...........
.....398
398
15.2 Component
Comp onent Histo
History ry and Impro
Improvem vementsents ......
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
..........
....398
398
397
15.1 Introduction
This section brings together information found in other parts of the service manual concerning hardware components that
have changed over time. The section also provides additional information including.
Hamilton ventilator serial numbers and the dates of changes.
WARNING
If the Interaction Panel Display (PN 380027 or PN 380031) or the Backlight DC/AC Inverter (PN 396197 or PN 396229)
needs replacement
replacement and the Front Panel Board is Revision 01 to 07, please order the following parts:
1x Display (MSP380046)
1x Backlight LED Driver Board (PN 396314)
1x Cable (PN 161569)
2x Ferrite Plus (PN 393133)
2x Ferrite Clip (PN 393134)
WARNING
If the Interaction Panel Display (PN 380027 or PN 380031) or the Backlight DC/AC Inverter (PN 396197 or PN 396229)
needs replacement
replacement and the Front Panel Board is Revision 08 or higher, please order the following parts:
1x Display (MSP380046)
1x Backlight Extension Cable (PN 161569)
1x Cable (PN 161569)
2x Ferrite Plus (PN 393133)
2x Ferrite Clip (PN 393134)
Set the Front Panel Board Jumper P14 to pins 1 and 2.
398 English | 624165/07
00 P1st
N 3generation
80027 LCDwith
TFTFlorescent
VGA 10.4” NEC Grap(PN
Backlights hic Display,
380030)
Used in HAMILTON-C2 from SN1001 to SN1675
NOTICE! Obsolete
NOTICE! Obsolete
00
01 Initial release.
03 Speaker loudness.
Remove Foil Cover Sticker from the Interaction Panel
Cover (if still attached) after Rev > 03 installation.
04 Improvement of
of th
the TF
TF 24
243003 pr
problem an
and lo
loud-
speaker loudness and sound quality.
05 Improvement of
of lo
loudspeaker lilifetime.
Speaker no longer soldered, but with connector.
07 Not implemented.
08 Integrated Ba
Backlight LE
LED Co
Components fo for th
the Gra
Grap
phic
LCD TFT VGA 10.4” NLT Display (PN 380046) with
LED backlights. With this Front Panel Board, the
external Backlight LED Driver Board (PN 396314) is no
longer required. An Extension Cable (PN 161569) is
required as a bypass for the Backlight Driver Board.
AC Backlight Inverter
Inverter Board (PN 396229),
396229), if available.
400 English | 624165/07
NOTICE! Obsolete
NOTICE! Obsolete
01 PGraphic
N 3963LCD
14 BaTFT
ckliVGA
ght LE10.4”
D DrivNLT
er BDisplay
oard forBacklight
the
th
LED's.
00
01
02
04
05
07 New layout.
Blower end stage robustness.
Used in HAMILTON-C2 from SN2141 to SN2422.
08 New firmware.
Used in HAMILTON-C2 from SN 2423 to SN 2784.
09 Qvent Fl
Flow Sensor pr
precision im
improved.
EMC immunity to 20 v/m increased.
Used in HAMILTON-C2 from SN2785 to SN3050.
10 Not implemented.
13 EMC im
improvements an
and bu
buzzer ad
adjustability.
Neonatal implemented.
Used in HAMILTON-C2 from SN4456.
NOTICE
The Oxygen Sensor Calibration Tool (PN 160367) is no longer required for Mainboard Revisions 06 and higher.
Mainboard (MSP160382) 15
NOTICE
The Mainboard MSP160382 was previously used in the HAMILTON-C3 only. In the HAMILTON-C2, it is assembled in devices
with SN > 10’000. It is compatible from older devices as long as software version 2.2.0 or higher is installed.
Table 15-5. Mainboard (MSP160382)
00
01
02
03
04
06 New Po
PoL vo
voltage co
converters fo
for im
impro
rovved EM
EMI be
behav-
ior.
Oxygen Sensor calibration without disconnection.
RS232 monitoring interface bit rate up to 38.2 kBaud
for Hamilton block protocol.
07 Not implemented.
08 Incre
reaased ES
ESD im
immunity ag
against co
contact di
discharge to
to
fulfill the new standard for electromagnetic compati-
bility (EMC) IEC 60601-1-2 (2014) 4th edition.
Improved polarity protection on DC Input.
New gold caps with improved lifetime.
09
NOTICE
If Mainboard (MSP160382) replaces Mainboard (MSP160200), make sure to roll up the excess Fan Supply Cable.
00
01 First official release.
02 New layout.
New Firmware.
SW 1.0.1.
Ready for SW Upgrade with USB memory stick.
05 SW 1.0.5.
No more loss of Technical State due to SW upgrade,
not used components dropped.
06 EMC improvements.
07 Pre-installed la
latest so
software 2.
2.x.x.
08 Layout correction.
09 Pre-installed SW 2.2.1.
10 Pre-installed SW 2.2.3.
11 Cloc
Clockk ge
gene
nera
rato
tors
rs re
rep
pla
lace
ced
d due to di
disc
sco
ont
ntin
inue
ued
d pa
part
rts.
s.
BGA components underfilled to improve mechanical
robustness.
00
Flow Sensors 15
CAUTION
Always use Flow Sensors from the same brand in one ventilator (Sensirion or TSI).
00
01
02
03
00
01
02
03
00
01
Hamilton Medical | HAMILTON-C2 Service Manual 405
01 Features:
• Pla
Plasti
sticc Disk
Disk (A)
(A)
• With
Without
out Turbule
Turbulence
nce Filter
Filter (A)
(A)
• Manufactured with
with the small O2 connector seat
seat (B)
(see Mounting Plate Section 15.2.2.7
15.2.2.7))
• Manufactur
Manufactured ed without
without O2 Inlet
Inlet Filter (C) (see
(see O2
Inlet Filter Section 15.2.2.8
15.2.2.8))
02 Features:
• Sin
Sinter
ter dis
diskk (A)
(A)
• Equi
Equipped
pped with
with turbulenc
turbulencee screen
screen (B)
• Manufactured with
with the small O2 connector seat
seat (B)
from SN1590 to SN2154 (see Mounting Plate Sec-
tion 15.2.2.7
15.2.2.7))
• Manufactur
Manufactured ed without
without O2 Inlet
Inlet Filter (C) (see
(see O2
Inlet Filter Section 15.2.2.8
15.2.2.8))
03 Features:
• The QO2 Flow
Flow Sensor
Sensor (MSP161657
(MSP161657)) is part of the
mixer assembly (A, MSP160226)
• Manufactur
Manufactured
ed with the small
small O2 connector
connector seat
seat
(A1) from SN1590 to SN1625
• Manufactur
Manufactured
ed with the large
large O2 connector
connector seat
seat (B)
from SN1626
• Or
Orif
ific
icee (C)
(C)
• Turbu
Turbulenc
lencee scree
screen
n (D)
(D)
In HAMILTON C2 from SN1590 to SN2154.
04 Adaption of
of th
the ma
manufacturing pr
process.
In HAMILTON-C2 from SN 2155.
05 Intr
Introd
oduc
ucti
tio
on of an In
Injject
ctio
ion
n Mo
Mollde
ded
d Con
onn
nec
ecti
tion
on Pl
Plaate
te..
In HAMILTON-C2 from SN 5587.
01 The mo
mounting plplate wi
with the lalarge O2
O2 Co
Connector se
seat
(right image) is installed in all HAMILTON-C2 from
SN1626.
408 English | 624165/07
NOTICE
A number of devices with Mixer Block Assembly (MSP160226) REV 02 (SN 1590 - 1625) are not equipped with the large O2
connector seat. For these devices, it is necessary to order the Mounting Kit (PN 160496) together with the DISS or NIST con-
nector.
Hamilton Medical | HAMILTON-C2 Service Manual 409
02 Assembly pr
process improved.
05 Replacement of
of fifilter pre
presssure bo
board in
introduced.
00
01
02
03
04
00 First of
official re
release in
in HA
HAMILTON-C2 un
until SN
SN 30
3004.
01 Housing an
and flflow ch
characteristic im
improved in
in
HAMILTON-C2 since SN 3005.
02 Stick-s
-sllip ef
effect im
improved in
in HA
HAMILTON-C
-C2
2 si
since SN
SN
4233.
Neonatal implemented.
00
01 Neonatal im
implemented.
00
01 First off
offiicial re
release in
in HA
HAMILTON-C2 fr
from SN
SN 10
1000.
00 First release.
01 To prevent an incre
reaase of battery temperatures >
54°C 5, vent holes (A) are drilled on the top of the
Battery Compartment.
01
NOTICE! Obsolete
Old Version
00 PN 160506 New Version
If the broken Battery Cover is the Old Version (PN
160344), order the complete Rear Cover
(MSP160481), (see the Rear Cover Section 15.2.2.20
15.2.2.20).
).
New Version
00
01 To pr
prevent da
damage of of th
the de
device, th
the cy
cylinder bo
bolt (A
(A)
was extended and locked against removing with a
retaining ring (B) inside.
The top cover can be replaced with PN 160601.
This improvement is implemented from HAMILTON-
C2 SN1271.
16
Glossary
16 Glossary
Alar
Alarm
m Buf
Buffe
ferr An ar
area
ea of me
memo
mory
ry co
cont
ntai
aini
ning
ng de
deta
tail
ilss of
of the
the tw
twen
enty
ty mo
most
st re
rece
cent
nt al
alar
arms
ms.. The
The mo
most
st re
rece
cent
nt si
sixx of
of the
the tw
twen
enty
ty
alarms can be accessed by activating the alarm symbol at the bottom left of the screen when the HAMILTON-
C2 is in normal operating mode.
Alarm
Alarm Status
Status Indica
Indicator
tor An area
area at the bottom
bottom of
of the Touch
Touch Screen
Screen of the HAMI
HAMILTON-
LTON-C2
C2 which
which displa
displays
ys the alarm statu
statuss and powe
powerr
status.
Ambi
Ambien
entt Sta
State
te A sta
state
te th
that
at th
thee HAM
HAMIL
ILTO
TON-
N-C2
C2 us
uses
es wh
when
en it ca
cann
nnot
ot fu
func
ncti
tion
on no
norm
rmal
ally
ly be
beca
caus
usee of
of an
an int
inter
erna
nall or
or ext
exter
erna
nall fau
fault
lt..
The Ambient State is often associated with Technical Faults, but can also be caused by such things as air or
oxygen supply. It is always accompanied by a high-priority patient alarm, sounded by the loudspeaker or
buzzer.
Ambient
Ambient State
State Gas
Gas Flow If the HAMILT
HAMILTON-C2
ON-C2 is in the
the Ambient
Ambient State,
State, the patie
patient
nt breathes
breathes unassi
unassisted.
sted.
Audi
Audibl
blee Ala
Alarm
rm An al
alar
arm
m sou
sound
nded
ed wh
when
en th
ther
eree is
is a Te
Tech
chni
nica
call Fau
Fault
lt or ot
othe
herr erro
errorr con
condi
diti
tion
on.. An
An aud
audib
ible
le al
alar
arm
m is
is gen
gener
erat
ated
ed.. A
buzzer is used as a backup if the audible alarm through the loudspeaker does not function.
Aut
utoz
ozeero A me
meth
thod
od to aut
utom
omat
atiica
call
llyy adj
adju
ust fo
forr ele
elect
ctro
roni
nicc dri
drift
ft of a de
devi
vicce due
due to te
temp
mper
eraatu
ture
re an
and
d env
enviiro
ronm
nmen
enta
tall con
cond
di-
tions.
Basic Input Output Sys- The basic input output system for the central processing unit (CPU). This is held on an EPROM or EEPROM
tem (BIOS) mounted on the ESM.
Buzzer A piezoelectrtriic de
device mounted on
on the Control Board, us
used by
by th
the HAMILTON-C2 toto so
sound so
some alarms. ItIt
functions independently of the loudspeaker and the power supply, and typically indicates a high-priority tech-
nical faults that causes the HAMILTON-C2 to go into the ambient state. Despite its name, the buzzer makes a
high frequency sound.
NOTICE! Despite its name, the buzzer makes a high frequency sound.
Commun
Communica
icatio
tion
n Board
Board A commun
communica
icatio
tion
n interf
interface
ace for
for the CO2,
CO2, SpO2
SpO2 and
and Nurse
Nurse Call
Call func
functio
tions.
ns.
Communica
Communication
tion Interfa
Interface
ce The Etherne
Ethernett Port on the rear
rear of the HAMILT
HAMILTON-C2
ON-C2..
Config
Configura
uratio
tion
n Menu
Menu A scree
screen
n used
used to
to selec
selectt venti
ventilat
lation
ion and oth
other
er oper
operati
ation
on para
parame
meter
terss of the
the HAMI
HAMILTO
LTON-C
N-C2.
2.
Cont
Contro
roll Kno
Knob
b Used to se
Used sele
lect
ct an op
opti
tion
on on th
thee LCD
LCD sc
scre
reen
en by ro
rota
tati
ting
ng an
and
d can
can be us
used
ed to ac
acti
tiva
vate
te a sel
selec
ecti
tion
on on th
thee LCD
LCD
screen by pressing.
Definitions of Expression 16
De-a
De-act
ctiv
ivat
atee An ac
acti
tion
on on th
thee To
Touc
uch
h Sc
Scre
reen
en or th
thee P&
P&TT Co
Cont
ntro
roll Kn
Knob
ob to cl
clea
earr th
thee se
sele
lect
ctio
ion
n of a fu
func
ncti
tion
on or ac
acti
tion
on..
De-s
-seele
lect
ct An act
actio
ion
n on
on th
thee Tou
ouch
ch Sc
Scre
reeen or
or th
the P&
P&T Co
Con
ntr
tro
ol Kn
Knob to cl
cleear th
the se
sele
lect
ctio
ion
n of
of a fu
fun
nct
ctio
ion
n or
or act
actio
ion.
n.
DISS Diameter index safety standard, a standard connector used for connection of oxygen to the HAMILTON-C2,
designed so the oxygen supply cannot be connected incorrectly.
dP Fl
Flow
ow Sen
Senso
sorr Measur
Meas ures
es th
thee diff
differ
eren
enti
tial
al pre
press
ssur
uree (pre
(press
ssur
uree diffe
differe
renc
nce)
e) acr
acros
osss the
the flow
flow sen
senso
sor.
r. The
The mea
measu
sure
reme
ment
nt is
is perf
perfor
orme
med
d
by a pressure sensor inside the HAMILTON-C2, and is used to calculate airway gas flow.
Elect
Electric
rical
al Saf
Safety
ety Tes
Tests
ts A set
set of elec
electri
trical
cal tes
tests
ts used
used to chec
checkk the
the safe
safety
ty of
of a dev
device
ice..
Emergency Buzzer Alarm An alarm that sounds when a Technical Faults occurs that is serious enough
enough to cause
cause the HAMILTON-C2 to
switch to Ambient Mode.
In Ambient Mode, all valves switch to
t o their unpowered position, and the HAMILTON-C2 is completely passive.
NOTICE! The buzzer makes a high frequency sound. It functions independently of the loudspeaker.
Eve
vent
nt Lo
Log
g A rec
recor
ord
d of
of mos
mostt act
actiivi
vity
ty in th
the HA
HAMI
MILT
LTON
ON-C
-C2
2. Thi
Thiss inc
inclu
lude
dess use
userr ac
acti
tion
onss and
and in
inte
tern
rnaal act
actiivi
vity
ty suc
uch
h as
as:
• Cal
Calibr
ibrati
ation
on Results
Results
• Al
Alaarm
rmss
• Tec
Techni
hnical
cal Fau
Faults
lts
• Con
Contro
trols
ls Setti
Settings
ngs
• Configuration, Serial Numbers, Revision Numbers
• Swi
Switch
tch-on
-on Tim
Times
es
The Event Log always holds a maximum of 1,000 events. Switching OFF and ON the HAMILTON-C2 does not
delete the storage of the Event Log.
You can access a subset of the event log suited for clinical use by activating the Event Log symbol in the moni-
toring menu in normal operating mode.
Expiratory Valve Block The plastic connector used to attach the Patient Breathing Circuit on the side of the HAMILTON-C2.
Connection
Fan Filter A filter used to capture dust from the Cooling Fan.
Flow Restrictor A device that restricts flow of a gas.
Front Pa
Panel Ke
Keys Control bu
butt
tto
ons at
at th
the fr
front of
of th
the Fro
Fron
nt Pa
Panel.
Fron
Frontt Pan
Panel
el Part of th
Part thee HAM
HAMIL
ILTO
TON-
N-C2
C2 co
cont
ntai
aini
ning
ng th
thee use
userr int
inter
erfa
face
ce fo
forr int
inter
erac
acti
tion
on wi
with
th th
thee ven
venti
tila
lato
tor;
r; LC
LCD
D Dis
Displ
play
ay an
and
d
Hard Keys.
Galv
Galvan
anic
ic Ox
Oxyg
ygen
en Se
Sens
nsor
or Seee Oxygen Sensor
Se
Gol
old
d Cap
Capss On
One
e pro
whenprovi
novide
dess pow
power power
er fo
forrisem
source em erge
ergenc
ncyyfrom
provided buzze
buzzer
r al
alar
mainsarm
m for 2ormin
for
power mbattery.
inut
utes
es and one fo
forr the
the re
real
al ti
tim
me clo
clock
ck fo
forr 3 mon
onth
thss
16 Glossary
Hott Swa
Ho Swapp
ppab
able
le A ter
term
m use
used
d to
to ind
indic
icat
atee a de
devi
vice
ce ca
can
n be
be dis
disco
conn
nnec
ecte
ted
d wit
witho
hout
ut re
remo
movi
ving
ng po
powe
wer.
r.
Intern
Internal
al Tempe
Temperat
rature
ure The HAMI
HAMILTO
LTON-C
N-C2
2 contai
contains
ns devic
devices
es to meas
measure
ure the
the inter
internal
nal oper
operati
ating
ng tempe
temperat
rature
ures.
s.
Light Emitting Diode Light Emitting Diode
(LED)
Loud
Loudsspe
peak
akeer A sp
speeaker us
useed to ind
ndiica
cate
te alar
arm
m co
con
ndi
diti
tion
onss fr
from
om th
thee ve
ven
nti
tillat
atiion uni
nit.
t.
Maiin Pow
Ma Poweer Sw
Switc
tch
h Pow
oweers th
thee HA
HAMI
MILT
LTO
ON-
N-C
C2 ON
ON and
and OF
OFF.
F.
Metr
Metron
on EST
EST Te
Test
ster
er Auto
Automa
mate
ted
d devi
device
ce us
used
ed to per
perfo
form
rm the
the el
elec
ectr
tric
ical
al sa
safe
fety
ty tes
test.
t.
Micr
Microp
opho
hone
ne In th
thee HAM
HAMIL
ILTO
TON-
N-C2
C2,, a mic
micro
roph
phon
onee is
is pos
posit
itio
ione
ned
d nea
nearr the
the lo
loud
udsp
spea
eake
kerr as
as a fee
feedb
dbac
ackk cir
circu
cuit
it to de
dete
term
rmin
inee tha
thatt
the loudspeaker is operating properly.
Nebuli
Nebulizer
zer Conn
Connect
ection
ion Connec
Connectio
tion
n on the side
side of
of the HAMIL
HAMILTON
TON-C2
-C2 to
to attach
attach tubin
tubing
g for the
the Nebuli
Nebulizer
zer outp
output.
ut.
Neb
ebul
uliz
izeer Va
Valv
lvee Valv
Valvee us
used to
to co
cont
ntro
roll th
the flflow of oxy
oxyge
gen
n to
to the
the Ne
Nebu
buli
lize
zerr Ja
Jar.
NIST Non-interchangeable screw thread, a standard connector used for connection of oxygen to th
thee HAMILTON-
O2 Cell For O2 Cell, O2 Sensor or Oxygen Cell, see the Oxygen Sensor.
Orific
Orificee Flow
Flow Rest
Restric
rictor
tor A devic
devicee with
with a fixe
fixed
d openi
opening
ng whic
which
h restr
restrict
ictss the flo
flow
w of a gas
gas..
Oriifi
Or fice
ce Tub
ubee Use
sed
d in
in the
the te
test
st co
con
nfi
figu
gura
rati
tion
on to pe
perf
rfo
orm adj
djus
ustm
tmen
ents
ts in the
the se
serv
rviice so
soft
ftw
war
aree.
Oxyg
Oxygen
en Se
Sens
nsor
or A sm
smal
all,
l, re
repl
plac
acea
eabl
ble,
e, pl
plas
asti
ticc un
unit
it us
used
ed by th
thee HA
HAMI
MILT
LTON
ON-C
-C2
2 to me
meas
asur
uree ox
oxyg
ygen
en co
conc
ncen
entra
trati
tion
on..
The Oxygen Sensor reacts to the presence of oxygen, producing a voltage output in proportion to the oxygen
concentration.
The Oxygen Sensor must be replaced after a period of service, when it can no longer be calibrated. Typically,
about one year.
Oxygen Sensor Calibra- A procedure which supplies a controlled flow of air, oxygen and air/oxygen mixture to the Oxygen Sensor for
tion calibration.
Pambient A sensor which measures the ambient pressure or the room pressure.
Pati
Patien
entt Alar
Alarms
ms An al
alar
arm
m ind
indic
icat
atin
ing
g that
that th
ther
eree is
is a pro
probl
blem
em or pot
poten
enti
tial
al pr
prob
oble
lem
m in
in vent
ventililat
atin
ing
g the
the pa
pati
tien
ent.
t.Th
Ther
eree are
are th
thre
reee lev
lev--
els of patient alarms: high, medium and low. They are indicated by beeps on the loudspeaker, messages on
the display and the alarm lamp.
Patient
Patient Breathing
Breathing Circuit
Circuit A Patient
Patient Breathing
Breathing Circuit
Circuit carries the air/oxyg
air/oxygen
en mixture
mixture to the patient and carries
carries the expired
expired (exhaust)
(exhaust) air
from the patient. It may include Tubings, Mask, Inspiratory Filter, Flow Sensor, Nebulizer Jar, Expiratory Valve
Cover and Membrane.
Paw A sensor which measures the pressure at the proximal side of the Patient Flow Sensor.
Pfilter A sensor wh
which me
measures the pressure af
after th
the HE
HEPA Filter; us
used to in
indicate if th
the HEPA fifilter ne
needs to
to be
replaced.
Pfl
flo
ows
wsen
enso
sorr A se
sens
nsor
or whi
hich
ch mea
easu
sure
ress the
the di
diff
ffeere
ren
nti
tiaal pre
presssu
sure
re acr
cros
osss the
the pa
pati
tieent airw
rwaay fl
flow se
sens
nsor
or..
Port
Portab
able
le Ta
Tank
nk Oxyg
Oxygen
en ta
tank
nkss use
used
d for
for su
supp
pply
ly in a por
porta
tabl
blee env
envir
iron
onme
ment
nt wh
when
en th
thee fac
facil
ilit
ityy sou
sourc
rcee is
is not
not av
avai
aila
labl
ble.
e.
Press and Turn Control A button on the front of the front panel used to select and activate screen functions.
Knob (P&T)
Press and Turn Encoder An electrical device which provides input signals for control of screen functions.
(P&T Encoder)
Pressure Gauge A me
mechanical de
device us
used to
to me
measure ga
gas pr
pressure.
Pres
Pressu
sure
re Re
Regu
gula
lato
torr A dev
devic
icee tha
thatt reg
regul
ulat
ates
es th
thee pre
press
ssur
uree and
and re
rest
stri
rict
ctss the
the fl
flow
ow of a gas
gas..
Pressure Sensor Assembly Provides components for pressure measurements used to monitor ventilation.
ventilation.
Definitions of Expression 16
Preventive Maintenance A term used for periodic maintenance of a device with specific planned tasks and items to check, adjust and
(PM) replace.
Princi
Principal
pal Gas Flo
Flow
w The mai
mainn gas
gas flo
flow
w throu
through
gh the HAM
HAMILT
ILTON-
ON-C2
C2 fro
from
m the
the HEP
HEPAA Filte
Filterr for
for air
air and
and the oxy
oxygen
gen inl
inlets
ets to the
the
patient, and then from the patient through the expiratory valve.
Pven
Pvent_
t_co
cont
ntro
roll A se
sens
nsor
or wh
whic
ich
h me
meas
asur
ures
es th
thee pr
pres
essu
sure
re at th
thee pa
pati
tien
entt ou
outl
tlet
et,, us
used
ed fo
forr co
cont
ntro
roll
llin
ing.
g.
Pven
Pvent_
t_mo
moni
nito
torr A sen
senso
sorr whi
which
ch me
meas
asur
ures
es th
thee pre
press
ssur
uree at
at the
the pa
pati
tien
entt out
outle
let,
t, us
used
ed fo
forr mon
monit
itor
orin
ing.
g.
QO2 Fl
Flow
ow Sens
nso
or Use
sed
d to
to me
measu
sure
re th
thee oxy
oxyg
gen fl
flo
ow int
into
o th
the Blo
Blow
wer Mo
Modu
dulle.
Qven
Qventt Flow
Flow Se
Sens
nsor
or Used
Used to me
meas
asur
uree the
the ai
air/
r/ox
oxyg
ygen
en fl
flow
ow in
into
to the
the Pa
Pati
tien
entt Brea
Breath
thin
ing
g Cir
Circu
cuit
it..
Rins
Rinsee Flo
Flow
w A ver
veryy sma
small
ll,, con
conti
tinu
nuou
ouss flo
flow
w of
of gas
gas th
thro
roug
ugh
h bot
both
h the
the bl
blue
ue (p
(pat
atie
ient
nt si
side
de;; pro
proxi
xima
mal)
l) an
and
d cle
clear
ar (v
(ven
enti
tila
lato
torr sid
side;
e;
distal) tubes to the Patient Flow Sensor.
The flow minimizes the possibility of tube blockage, and hinders the potential migration of bacteria and
viruses from the patient’s expired gases through the tubes, towards the pressure sensors inside the ventilator.
Safe
Safety
ty Mod
Modee An em
emer
erge
genc
ncyy sta
state
te tha
thatt ens
ensur
ures
es a bas
basic
ic mi
minu
nute
te ven
venti
tila
lati
tion
on wh
whil
ilee giv
givin
ing
g the
the use
userr time
time fo
forr corr
correc
ecti
tive
ve ac
acti
tion
onss in
in
case of some technical fault alarms. The default inspiratory pressure is maintained, the expiratory valve opens
as needed to switch system pressure levels between PEEP and inspiratory pressure, and patient sensing is non-
functional.
Select An ac
action on
on the touchscreen or th
the P&T control knob to
to choose a function or
or ac
action.
Serv
Servic
icee Soft
Softwa
ware
re The softw
The softwar
aree used
used to
to perfo
perform
rm test
tests,
s, adj
adjus
ustm
tmen
ents
ts and
and cal
calib
ibra
rati
tion
onss of the
the HAMI
HAMILT
LTON
ON-C
-C2.
2. Als
Also
o enab
enable
less view
viewin
ing
g
and exporting data from the Event Log. The HAMILTON-C2 cannot be used for patient ventilation when in
service software mode.
NOTICE! The HAMILTON-C2 cannot be used for patient ventilation when in Service Software Mode.
Service
Service Softw
Software
are Mode The opera
operationa
tionall status
status of the
the HAMILT
HAMILTON-C2
ON-C2 when in the
the servic
servicee softwa
software.
re.
Smart
Smart Bat
Batter
teryy Pack
Pack Li-Ion
Li-Ion bat
batter
tery,
y, hot swa
swappa
ppable
ble whi
which
ch mon
monito
itors
rs and com
commun
munica
icates
tes the bat
batter
teryy cond
conditi
ition.
on.
SoH The State of Health (SoH) is a "measurement" that refl fleects the general condition of a battery and its ability to
deliver the specified performance compared with a fresh battery. It takes into account such factors as charge
acceptance, internal resistance, voltage and self-discharge. It is a measure of the long term capability of the
battery and gives an "indication" not an absolute measurement, of how much of the available "lifetime
energy throughput" of the battery has been consumed, and how much is remaining. Using the automotive
analogy, it can be compared to the "odometer"display function which indicates the number of miles traveled
since the vehicle was new.
Tech
Techni
nica
call Even
Events
ts A very
very min
minor
or fau
fault
lt or
or even
eventt reco
record
rded
ed by
by the
the HAMI
HAMILT
LTON
ON-C
-C2
2 in th
thee Even
Eventt Log
Log for
for use
use only
only by
by softw
softwar
aree deve
develo
lop-
p-
ers.
Tech
Techni
nica
call Faul
Faults
ts An ala
alarm
rm con
condi
diti
tion
on ind
indic
icat
atin
ing
g a maj
major
or mal
malfu
func
ncti
tion
on of
of the
the HAMI
HAMILT
LTON
ON-C
-C2.
2. (Th
(This
is con
contr
tras
asts
ts wit
with
h an al
alar
arm,
m, tha
thatt
indicates a problem with the status of a patient.) Technical Faults are intended to alert users and engineers of
the need for intervention, and are recorded in the Event Log.
Test Config
Configurati
uration
on (TC) A particula
particularr configurati
configuration
on of devices
devices and tubing
tubingss use to perform
perform the tests/
tests/calib
calibratio
rations
ns in the service
service softwa
software.
re.
Touc
Touch
h Scr
Scree
een
n An in
inte
tera
ract
ctio
ion
n met
metho
hod
d whe
where
re th
thee LCD
LCD di
disp
spla
layy scr
scree
een
n can
can be to
touc
uche
hed
d wit
with
h a fin
finge
gerr to ac
acti
tiva
vate
te or sel
selec
ectt a fun
func-
c-
tion.
Trolley A pa
part of
of th
the HA
HAMILTON-C2 on
on wh
which th
the ve
ventilator is
is mo
mounte
ted
d fo
for tr
transport wi
within th
the cu
customer lo
location.
Upd
pdat
atee An upd
pdaate is an im
impr
prov
oveemen
entt to an exi
xissti
ting
ng fu
func
ncti
tion
on.. An up
upd
dat
atee no
norm
rmaall
llyy in
invo
vollve
vess on
only
ly sof
oftw
twar
aree. A so
soft
ftw
ware
update is generally a revision number increment in a digit after the decimal point: for example, from 3.2 to
3.3.
Upg
pgra
rade
de An upg
pgra
rade
de is th
thee add
ddiiti
tion
on of ne
new
w fu
fun
nct
ctiion
onss to a dev
eviice
ce.. The
here
re are th
thre
reee ways to pe
perf
rfo
orm an upg
pgra
rade
de::
• Add a hardware item
item that offers additional
additional functions
• Upgrade to a higher
higher software revision indicated
indicated by a higher
higher value before the decimal point: for example
example
01.03 to 02.00
• Upgra
Upgrade
de to a highe
higherr type of softwa
software
re
16 Glossary
Variable Orifice Mem- A variable flow restrictor as used in the Patient Flow Sensor.
brane
Alarm Codes 16
Breath Monitoring
133004 paBPG_CheckPatientInterface
16 Glossary
Gas Delivery
231026 taGD_expValveCoverInvalid
231044 taGD_calibrateExpValveOK
231045 taGD_calibrateExpValveError
16 Glossary
16 Glossary
241002 taVMC_tightnessTestReleaseValveDefect
Alarming
Power Management
244001 taPM_externalPowerLoss Loss of external power
255019 taIVS_PercentageMinVolHighFineFailed
255020 taIVS_PercentageMinVolHighLimitFineWrong
255021 taIVS_PercentageMinVolLowFineFailed
255022 taIVS_PercentageMinVolLowLimitFineWrong
255023 taIVS_CtrlOszillationFiO2
255024 taIVS_CtrlOszillationPEEP
16 Glossary
255025 taIVS_CtrlOszillationPercMinVol
Ventilation GUI
Service GUI
16 Glossary
Ventilation GUI
GUI Library
16 Glossary
Do not consult
workers Hamilton Medical
or Hamilton-Medical Technical
Trained SupportEngineers.
Field Service concerning these technical faults. They have no significance for hospital
Hamilton Medical | HAMILTON-C2 Service Manual 433
16 Glossary
434 English | 624165/07
17
Test Report
17 Test Report
HAMILTON-C2
Serial Number: SN __ __ __ __ __ __
Inventory Number: __ __ __ __ __ __ __
Software Version: __ __ __ __ __ __ __
Air Intake Dust Filter cleaned or replaced? See the Operator's Manual Section
10.3.1.
Batteries calibrated (% error > 5%)? — See the Service Manual Section 5.
Batteries replaced if State of Health (SoH) < 20% — See the Service Manual Section 5.
Blower Module replaced (ebt ≥ 100%)? — See the Service Manual Section 8.9.2
and Section 10.4.10
10.4.10..
All marks, symbols, and labels related to safety are com- — See the Operator's Manual Section 1.4.
plete?
Visually inspect the unit for damage and clean the unit if — —
necessary.
AC Battery Test successfully performed? — See the Service Manual Section 8.21
8.21..
DC Battery Test successfully performed? See the Service Manual Section 8.21
8.21..
Power (battery) loss simulated: Ambient Mode test success- — See the Service Manual Section 8.21
8.21..
fully performed?
Buzzer Sound Test: Does the buzzer sounds for at least 2 — See the Service Manual Section 8.21
8.21..
minutes?
RS232 Test performed successfully? See the Service Manual Section 8.21
8.21..
Nurse Call functional test performed successfully? See the Service Manual Section 8.23
8.23..
Pre-operational checks performed in accordance with the — See the Operator's Manual Section 3.2.
procedures in the Operator's Manual.
Alarm tests performed in accordance with the procedures — See the Operator's Manual Section 3.5.
in the Operator's Manual?
Electrical safety tests performed successfully? — See the Service Manual Section 7.
Instrument report download performed and archived? — — See the Service Manual Section 8.24
8.24..
Event Log download performed and archived? — See the Service Manual Section 8.24
8.24..
Ventilator Info: Technical State, Service, and RTC (Real Time — See the Service Manual Section 8.7
8.7..
Clock) checked and each status is OK?
Adjustments / Calibration: Touch Screen, Insp. Valve, Pres- — See the Service Manual Section 8.14
8.14..
sure, Exp. Valve, Oxygen Sensor, Flow Sensor calibrated
and each status is OK?
Component Tests: Electronics, User I/F, Pneumatics 1, and — See the Service Manual Section 8.15
8.15..
Pneumatics 2 checked and each status is OK?
System Tests: Pressure, Flow, O2 mixer, Tightness, Alarming — See the Service Manual Section 8.16
8.16..
checked and each status is OK?
Flow Analyzer __ __ __ __ __ __ __ SN __ __ __ __ __ __ __ __
Pressure Gauge __ __ __ __ __ __ __ SN __ __ __ __ __ __ __ __
Remarks
17 Test Report
438 English | 624165/07
18
Environmental Requirements
18 Environmental Requirements
Storage:
-20°C to 60°C (-4°F to 140°F), in original packaging
-15°C to 60°C (5°F to
t o 140°F) otherwise
19
Document History
19.1 Document
Document Histo
History
ry ......
............
............
............
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
...........
...........
............
............
......442
442
441
19 Document History