Science made smarter
Service Manual
AT235
Impedance Audiometer
Item No 8533743 – D-0132907 C – 2023/09
Table of Contents
1 Introduction ...................................................................................................................................................... 1
1.1 About this manual .................................................................................................................................... 1
1.2 About warnings and cautions .................................................................................................................. 1
1.3 General information ................................................................................................................................. 1
2 Maintenance and Disassembly ....................................................................................................................... 2
2.1 General cleaning and maintenance procedure ....................................................................................... 2
2.2 Maintenance of the pump module ........................................................................................................... 3
2.3 Disassembling the Cabinet ...................................................................................................................... 6
2.3.1 Mainboard .................................................................................................................................... 7
2.3.2 Rotary dial.................................................................................................................................... 8
2.3.3 Display ......................................................................................................................................... 9
2.3.4 Nice to know ................................................................................................................................ 9
3 Calibration Procedure ................................................................................................................................... 10
3.1 Service, adjustment and repair .............................................................................................................. 10
3.1.1 Hardware calibration – requirements ........................................................................................ 10
3.2 Equipment requirements ....................................................................................................................... 11
3.2.1 Equipment requirements for impedance calibrations ................................................................ 11
3.2.2 Equipment requirements for audiometer calibrations ................................................................ 11
3.3 Calibration of test equipment ................................................................................................................. 12
3.3.1 Calibration mode entering ......................................................................................................... 12
3.3.2 Checklist for calibration ............................................................................................................. 13
3.4 Installing the Diagnostic Suite ............................................................................................................... 14
3.5 Main startup ........................................................................................................................................... 16
3.6 Saving the calibration data .................................................................................................................... 17
4 Calibration of Transducers ........................................................................................................................... 19
4.1 Calibration standard for Impedance ...................................................................................................... 19
4.1.1 Transducer setup ....................................................................................................................... 19
4.1.2 Probe tone speaker calibration .................................................................................................. 20
4.1.3 Transducer calibration ............................................................................................................... 21
4.1.4 Volume calibration ..................................................................................................................... 22
4.1.5 Pressure calibration ................................................................................................................... 23
4.1.6 Calibration standard for Audiometer .......................................................................................... 24
4.2 Selecting a transducer / how-to-guide ................................................................................................... 25
4.2.1 Transducer types and equipment requirements for calibration ................................................. 27
5 Testing the Tympanometer Functionality ................................................................................................... 29
5.1 Pressure test .......................................................................................................................................... 29
5.2 Diode test ............................................................................................................................................... 30
5.3 Volume test ............................................................................................................................................ 31
5.4 Attenuator test ....................................................................................................................................... 33
6 Appendix A – Technical Specifications .......................................................................................................... I
6.1 Calibration properties.............................................................................................................................. III
6.2 Reference values for stimulus calibration ................................................................................................ V
6.3 Reference equivalent threshold values for transducers ......................................................................... VI
6.4 Survey of reference and max. hearing level Tone Audiometer ............................................................. VII
6.5 Survey of reference and max. hearing level Tympanometer.................................................................. IX
7 Appendix B – Parts & Accessories ................................................................................................................ X
8 Appendix C – Firmware Update Procedure ................................................................................................. XV
9 Appendix D – update history ...................................................................................................................... XVII
Service Manual
AT235
1 Introduction
1.1 About this manual
This manual is for our Impedance Audiometer AT235, firmware version 1.42. The contents of the manual cover
instrument related data, such as disassembly, calibration, and pump module maintenance. Technical
specifications, schematics, parts & accessories list along with the firmware update procedure can be found as
appendices in the back of the manual.
1.2 About warnings and cautions
Where applicable, the below warning, caution and notice symbols are used throughout the manual, indicating the
level of attention required for a given action:
WARNING The WARNING label identifies conditions or practices that may
present danger to the patient and/or user.
CAUTION
The CAUTION label identifies conditions or practices that could
result in damage to the equipment.
NOTICE is used to address practices not related to personal
NOTICE injury.
1.3 General information
We continuously strive to improve our products and their performance, hence the specifications in this service
manual are subject to change without further notice.
The performance and specifications of our products can only be guaranteed if technical maintenance is
conducted routinely every year. Technical maintenance should be carried out by qualified personnel authorized
by Interacoustics.
We are happy to receive any inquiries about our products. Our contact details are:
Interacoustics A/S Phone: +45 6371 3555
Audiometer Allé 1
5500 Middelfart Mail:
[email protected] Denmark Web: www.interacoustics.com
1
Service Manual
AT235
2 Maintenance and Disassembly
2.1 General cleaning and maintenance procedure
CAUTION
• Before cleaning the device, always make sure that the device is switched off and disconnected from
the power supply
• Use a soft cloth lightly dampened with cleaning solution to clean all exposed surfaces
• Do not allow liquid to come into contact with the metal parts inside the earphones / headphones
• Do not autoclave, sterilize or immerse the instrument or accessory in any fluid
• Do not use hard or pointed objects to clean any part of the instrument or accessory
• Do not let parts that have been in contact with fluids dry before cleaning
• Rubber ear-tips or foam ear-tips are single use components
Recommended cleaning and disinfection solutions:
• Warm water with mild, nonabrasive cleaning solution (soap)
Procedure
• Clean the instrument by wiping outer case with a lint free cloth lightly dampened in cleaning solution
• Clean cushions and patient hand switch and other parts with a lint free cloth lightly dampened in
cleaning solution
• Make sure not to get any moisture in the speaker portion of the earphones and similar parts
2
Service Manual
AT235
2.2 Maintenance of the pump module
This chapter contains a thorough introduction and guide to the AT235. Carried out by a technician authorized by
Interacoustics, the most common action would be the pump module maintenance with the purpose of detecting
any deterioration on the pump tube. We recommend that pump module maintenance is conducted every 2-3
years to ensure proper function of the device.
The chapter includes other common actions, except from calibration which is thoroughly described in Chapter 3
and Chapter 4.
TOOLS
A Torx T8 & T10
A flatsided screwdriver
An unbrako Hex Key 1.5
Nut screwdriver 5
Cross-point screwdriver
The pump module should be checked and if applicable maintained on a regular basis. We have ensured direct
access to the module with the purpose of avoiding total disassembly of the instrument for pump module
maintenance.
By removing all 4 screws in the access panel, This allows the module to be taken out, inspected
direct access to the pump is enabled. and replaced if necessary (e.g., due to
deterioration).
Fig.1 Fig.2
3
Service Manual
AT235
The panel can now be carefully removed by lifting The pump module which is attached to the access
it up in the top end and releasing the bottom from panel can now be flipped over. Please note that it is
its holder. still attached with a cable and a tube.
Fig. 3 Fig. 4
Releasing the module from the instrument allows By sliding the black lock to the right allowing release
easier access. This is done by carefully releasing of the cable, the module is no longer attached to the
the tube on the right hand side. instrument.
Fig. 5 Fig. 6
The silicone tube is easily removed from the It is now possible to open the pump arm to the left,
pump outlet by pulling it apart gently. Pushing the and the module is ready to be de-attached.
pump arm slightly, the spring is ready to be
unhooked.
Fig. 7 Fig. 8
4
Service Manual
AT235
Remove the screw with an Unbrako key 1.5. Use a flat bladed screwdriver to carefully open the
lock mechanism holding the wheel in place.
Fig. 9 Fig. 10
Once unlocked, the wheel with the pump tube can
be lifted up and off of the module. It is now ready
for replacement.
NOTICE The wheel may have a tight fit and
thus require some effort to remove.
Fig. 11
5
Service Manual
AT235
2.3 Disassembling the Cabinet
To disassemble the instrument, the top must be The panels are now loose for dismantling the parts.
separated from the bottom cabinet. The Be careful whilst separating as the parts are still
cabinets are separated by removing the attached with cables on the inside. Please note that
indicated screws with a Torx T10. the screen is not fully attached anymore.
Fig. 12 Fig. 13
By tilting the top, the cabinet is open for service. Carefully, free the ribbon cable from the cabinet.
Fig. 14 Fig. 15
After releasing the ribbon cable and the display
cable, the bottom and top cabinets are separated.
Fig. 16
6
Service Manual
AT235
2.3.1 Mainboard
This chapter describes how the mainboard is removed.
The probe connection board is released by
unfastening the indicated points:
The white tube should be released at the
indicated end and removed through the loops.
The orange cable and the pump module cable
are released by sliding their respective lock
mechanisms to the right.
NOTICE There are 4 screws holding the probe
connection board in place. Two on the inside
and two on the outer rear panel.
It is possible to change the probe connection
Fig. 17 board without removing the main board
from the bottom cabinet.
Make sure that no tubes and cables are attached The main board can now be removed from the
to the mainboard. Use the lock mechanism for cabinet by pulling it towards the cabinet front.
releasing the main board from the frame.
Fig. 18 Fig. 19
7
Service Manual
AT235
2.3.2 Rotary dial
For dismantling the rotary dial, remove the cover With a screwdriver, remove the 3 screws holding the
by flipping it off. cover in place.
Fig. 20 Fig. 21
Now, the 3 black screws holding the dial in place The rotary dial can now be removed from the top
can also be removed with a screwdriver. cabinet.
Fig. 22 Fig. 23
8
Service Manual
AT235
2.3.3 Display
To fully release the display from the cabinet, By unlocking the snaps for the display cover,
remove the screw from the rear panel cover and access is enabled when opening the panel.
the 2 screws on the instrument, with a cross-point
screwdriver.
Fig. 24 Fig. 25
Remove the strip of kapton tape and then the
display power cable. The display is now loose from
the device.
NOTICE The display is attached to the bottom
cabinet, hence removal of the screws on
the bottom cabinet is required before
disassembly of the display is possible.
Please refer to Disassembling the Cabinet
Fig. 26
2.3.4 Nice to know
Access to the SD card is possible when removing
the panel and the screw holding it in place.
We highly recommend use of SD card types
authorized by Interacoustics, as we otherwise
cannot guarantee safe storage of the data.
NOTICE Replacement of the SD card protocol
overwrites present data and user settings.
This, however, does not affect the
calibration of the device.
Fig. 27
9
Service Manual
AT235
3 Calibration Procedure
This chapter contains calibration standards for AT235.
3.1 Service, adjustment and repair
This instrument can be serviced, adjusted and repaired without losing the validity of the CE-marking, provided the
measuring equipment fulfills the below requirements, the adjustment procedures are followed, and staff have the
necessary qualifications as approved by Interacoustics.
Before any attempt is made to calibrate AT235, it must have been powered on for 3-5 minutes in an ambient
temperature that is both stable and comfortable.
Annual calibration (ANSI/ASA S3.6)
The calibration procedure will validate all relevant performance requirements given
in ANSI/ASA S3.6. After the audiometric equipment has been submitted for
calibration, a calibration certificate showing the values recorded must be filled out,
signed and dated by the technician, demonstrating that the instrument conforms to
all relevant requirements given in this standard.
CAUTION It is considered good practice to perform annual checks of the audiometric test room’s noise level as part
of the annual calibration using the applicable table per ANSI/ASA S3.1 American National Standard
Maximum Permissible Ambient Noise Levels for Audiometric Test Rooms to ensure that audiometer
performance will not be affected by any changes in the audiometric test room’s noise levels.
Annual calibration (IEC 60645-1)
A basic calibration must be performed by a competent laboratory. The procedure should be such that, after the
audiometric equipment has been submitted for a basic calibration, it should meet the relevant requirements given
in IEC 60645-1.
When the instrument is returned after basic calibration, it should be checked in accordance with procedures outlined
in calibration checklist (Refer to section 3.3.2).
3.1.1 Hardware calibration – requirements
The hardware calibration is typically only performed during the production of the device and therefore does not
require adjustment in the field, except in certain circumstances. Refer to Appendices - Technical specifications for
further information.
10
Service Manual
AT235
3.2 Equipment requirements
3.2.1 Equipment requirements for impedance calibrations
To ensure proper calibration of audiometric equipment, test instruments must be reliable, stable and their
calibration certified.
The minimum requirements for measuring equipment are:
1) A measuring amplifier with input for condenser microphone or a sound level meter, fulfilling Type 1 of the
IEC 60651 requirements.
2) An acoustics coupler, 6cc and 2cc fulfilling the requirements of IEC 60318-3 and IEC 60318-5.
3) An occluded ear simulator/artificial ear fulfilling the requirements of IEC 60318-4 and IEC 60318-1.
4) A manometer with a range from at least +300 to -600 daPa and an accuracy of ± 5% of reading or ± 5
daPa, whichever is the greatest. Including a syringe (manual pump).
5) CAT55 calibration cavities (0.2/0.5 and 2.0/5.0ml).
6) We recommend the use of an acoustic calibrator for the control of the complete measuring chain.
7) A general purpose frequency counter.
8) A general purpose digital multimeter.
9) A general purpose oscilloscope for the purpose of tracing and monitoring signals.
10) An oscilloscope or peak-peak voltage measurement equipment.
3.2.2 Equipment requirements for audiometer calibrations
1) A Type 1 sound level meter – Octave band filter set (1/3 octave, with range from 125 Hz to 16000 Hz)
must meet Type 1 sound level meter requirements (ANSI S1.4). Sound level meters should be checked
before use by means of a calibrator.
2) A system to measure:
• Frequency
• Rise time
• Fall time
• Duration
• Overshoot
• THD (Total Harmonic Distortion)
3) A Piston phone or sound level meter calibrator.
• An ear simulator ANSI/ASA S3.7 or IEC 60318-1, Part 1: Electroacoustics - Simulators of human
head and ear Part 1: Ear simulator for the measurement of supra-aural and circumaural earphones.
• Acoustic Coupler – NBS–9A (6cc) ANSI/ASA S3.7, or IEC 60318-3, Part 3: Electroacoustics –
Simulators of human head and ear – Part 3: Acoustic coupler for the calibration of supra-aural
earphones used in audiometry.
• IEC 60318-5, Electroacoustics – Simulators of human head and ear – Part 5: 2 cm³ coupler for the
measurement of hearing aids and earphones coupled to the ear by means of ear inserts.
4) Specification / calibration information for all microphones, filters, SLM, and couplers showing correction
values required for maintaining accuracy.
5) Other couplers and adapters as stated or specified by the manufacturer of the audiometer.
6) Calibration reports/sheets that demonstrate that the audiometer meets the requirements and record the
results of measurement during the calibration.
11
Service Manual
AT235
3.3 Calibration of test equipment
All test equipment used for the calibration of audiometers is to be calibrated annually. Calibration of all test
equipment is to be performed by a laboratory which is traceable to the National Institute of Standards and
Technology (NIST) or equivalent. Test equipment calibration documents are to be kept on file.
3.3.1 Calibration mode entering
Connection between the instrument and a computer, should be conducted as described in Installing the
Diagnostic Suite.The calibration program is activated by starting the program and following the on-screen
instructions. The program is found via the below path:
C:\Program Files (x86)\Interacoustics\Diagnostic Calibration Tools\CalibrationAD629AC40AT235AA222.exe
12
Service Manual
AT235
3.3.2 Checklist for calibration 1
Use this list to ensure no aspect of the calibration is forgotten.
MECHANICAL INSPECTION PASS / FAIL
Plug - verify that plugs appear intact and undamaged
Cables - verify that the cable insulation appears intact
Display - verify that display is intact and undamaged. The tilt function should also work
properly
Keyboard - verify proper tactility function of keys
Rotary - verify that the dial functions without hinder
Headband - check proper tension
Headphone cushions - make sure that the cushions are functional
Probe tip - verify that the probe tip is intact and undamaged. If damaged, replacement
of the tip is recommended. Check that the probe tip is not blocked in any way
Probe "handle" - verify that it is undamaged
Masking devices - check that they are intact
Headset - check that it is intact
ELECTRIC TEST 2
Instrument indicators - upon switch-on of the instrument, make sure that the light
indicators are properly illuminated
Key test – check all keys incl. the rotary button for functionality from Patient response /
Keyboard (the on/off key is exempted)
Patient response – connect and check unit functionality from Patient response /
Keyboard
Probe switch button - verify that instrument and the probe interact properly using the
button/LED test function
Probe indicator - using the Diode test 3, verify proper diode function
Linearity of attenuator- test according to Attenuator test
Puretone - by means of an oscilloscope, make sure that the sinus tone is clean
Routing - verify that signals function with the remaining output options
Unwanted noise - control whether there is unwanted noise on any of the chosen
transducers
Crosstalk - verify that crosstalk is less than -70 dB
USB host - verify USB functionality by printing on a MPT3 printer
Tone switch - can be verified from the audiometer mode by tapping the enter button
lightly
ACOUSTIC TEST
Puretone - by means of an oscilloscope connected to the coupler amplifier output OR a
hearing test. Making sure that the sinus tone is clean, test at various frequencies,
especially at low frequencies
Calibration - verify that all units are compatible and correctly calibrated
Volume measurement - follow description in Volume test
1 Post calibration and before operation, the instrument should be checked for normal function
2 Conducted with the calibration utility
3 Under the tympanometer tab
13
Service Manual
AT235
3.4 Installing the Diagnostic Suite
The Diagnostic Suite is available in the USB stick. However, the free Diagnostics Suite upgrade package and the
Diagnostic Calibration Tools are available from the Interacoustics extranet.
The upgrade package enables upgrade from previous versions of the software.
The calibration tools contain the software required for the annual calibration of the following instruments:
AC40 AD629 AT235 AA222
Prior to the first time the instrument is connected to the computer, a USB driver for the purpose of storing data
has to be installed on the computer. In the below section, a guide to proper installation of the software can be
found.
Insert the Diagnostic Suite USB into the designated drive on the computer. Select DiagnosticSuiteSetup
Fig. 28
Select Install from the welcome page and allow the computer to install the software.
Fig.29
14
Service Manual
AT235
The computer now installs the software.
Fig. 30
When the installation is completed, the Installer can be shut down by selecting Close
Tip for use: create a
shortcut to the program
or pin it to the task bar for
easier access
Fig. 31
The driver has now been successfully installed and the AT235 is ready for connection to the computer.
NOTICE: Using operating systems where Microsoft have discontinued software and security support will increase
the risk of viruses and malware, which may result in breakdowns, data loss and data theft and misuse.
Interacoustics A/S cannot be held liable for your data. Some Interacoustics A/S products support or may work
with operating systems unsupported by Microsoft. Interacoustics A/S recommends you to always use Microsoft
supported operating systems that are kept fully security updated.
15
Service Manual
AT235
3.5 Main startup
Select
C:\Program Files (x86)\Interacoustics\Diagnostic Calibration Tools\Calibration AD629 AC40 AT235 AA222.exe
The startup screen should look as below with the Connect button highlighted to indicate correct connection.
If the Connect button is fully highlighted, the instrument is not properly connected. In this case shut down the
application, open it and Connect again. Make sure that the instrument is connected correctly to a power source,
and to the computer via a USB cable.
Fig. 28
16
Service Manual
AT235
3.6 Saving the calibration data
There are two ways of saving the calibration data in the programme, and the only difference between the two is
that method 1 is easier and ensures a continuous “saving flow”, should the computer or the equipment shut down
in an untimely manner, for instance caused by a power outage.
Method 1.
Following each action simply save by selecting the Save button, as exemplified below.
Save can be used at any point to save the present values. All values in all screens are saved.
Save & Close saves and exits the application. In this case, a dialog box with the message “Calibration data
successfully written to hardware” pops up when the action has been successfully completed.
Fig. 33
17
Service Manual
AT235
Method 2.
After a given action, go to the Main menu and select ‘Write to hardware’ as illustrated below. A dialog box with
the message Calibration data successfully written to hardware pops up when the action is successfully
completed.
Fig. 29
18
Service Manual
AT235
4 Calibration of Transducers
4.1 Calibration standard for Impedance
4.1.1 Transducer setup
Select Transducer setup to upload the calibration values to the contra phone connected to the device.
NOTICE Switching between different transducers requires a new calibration of the transducer in question.
The probe contains calibration data for the Probe tone, IPSI and volume.
NOTICE Present calibration data will be over-written if saved! In this case, a total recalibration will be necessary
for correct functionality of the selected item.
Fig. 30
19
Service Manual
AT235
4.1.2 Probe tone speaker calibration
Select Probe Tone Speaker for calibration.
Connect the probe to a 2 cc coupler with corresponding preamplifier and RMS-meter.
Follow the on-screen instructions.
NOTICE It is essential to click ‘Calculate Gain’
after each frequency level calibration.
Fig. 31
20
Service Manual
AT235
4.1.3 Transducer calibration
Select Transducer calibration to calibrate the transducers.
Select the relevant transducer and follow the on-screen instructions.
Upload defaults can be used when calibrating a new transducer in order to have a slide starting point closer to
target.
Output damping is equal to attenuator settings.
Fig. 32
21
Service Manual
AT235
4.1.4 Volume calibration
Select Volume Calibration.
A CAT55 cavity set should be used. Background noise should be kept to a minimum.
Follow the on-screen instructions.
NOTICE Height above sea level must be set before the volume calibration is conducted.
Verify that the result is successful when the calibration is completed in order to
avoid problems later in the process.
Fig. 33
22
Service Manual
AT235
4.1.5 Pressure calibration
Select Pressure Calibration. Follow the on-screen instructions.
The Safety Valve itself is NOT calibrated; this is only a verification of the sensor. The Safety Valve calibration lies
in the hardware and is therefore verified in the production process.
Fig. 39
23
Service Manual
AT235
4.1.6 Calibration standard for Audiometer
This chapter contains calibration standards for the Audiometer. Included here is an explanatory guide and a
complete set-up of how to calibrate each transducer type.
Tone standard is selected via AUD and ‘Standard’. The options are IEC or ANSI.
Calibration method is selected via AUD and ‘Calibration method’. The options are SPL mode or HL mode.
SPL mode
In SPL mode, the attenuator is adjusted according to the standard correction for each stimuli and frequency. In
that way the target level can be the same SPL level for all stimuli, and the microphone correction is the only thing
that should be added.
HL mode
In HL mode, the attenuator values are rounded and the target level in SPL varies and will be similar to the values
known from older HL-calibrated devices. The standard correction will be reflected as the difference between the
given HL and SPL level. Microphone correction should be added to the target level.
Fig. 40
24
Service Manual
AT235
4.2 Selecting a transducer / how-to-guide
The following procedure is applicable for all transducer selections. Here, we have exemplified the procedure with
the transducer type DD45, the standard transducer for this instrument.
Please note that following this section, a list with the specific equipment requirements for each suitable
transducer type can be found as well as photos of the correct equipment setup. The procedure described below
is applicable for all the listed transducers.
Example:
Select Transducer Setup.
Select Headset phone as type DD45.
Fig. 41
25
Service Manual
AT235
CALIBRATION OF DD45
Equipment needed for DD45 calibration
• Measuring amplifier
• Acoustical IEC 60318-3 (IEC 303) coupler
• Frequency counter
• Correction values for coupler
From the AUD menu, select Transducer Calibration.
Fig. 42
Place the transducer on the 6cc IEC 60318-3 (IEC 303) coupler
Adjust the slider until the calibration level is obtained on the measurement amplifier
(i.e. 90 dB SPL).
Fig 43
Check the frequency accuracy with a frequency counter. The accuracy has to be better than ±1%.
When the intended calibration level is obtained, use ‘Enter’ to get the next frequency.
Repeat until all stimuli are calibrated, for both the left and right transducer.
NOTICE Do not forget to add or subtract the correction for your microphone
26
Service Manual
AT235
4.2.1 Transducer types and equipment requirements for calibration
Conduct the same procedure exemplified in Selecting a transducer / how-to-guide can be applied to all the
transducers mentioned below.
TDH39
Equipment needed for TDH39 calibration
• Measuring amplifier
• Acoustical IEC 60318-3 (IEC 303) coupler
• Frequency counter
• Correction values for coupler
From the AUD menu, select Transducer Calibration.
Place the transducer on the 6cc IEC 60318-3 (IEC 303) coupler Fig. 44
(fig. 44)
EAR3A
Equipment needed for calibration of EAR3A
• Measuring amplifier
• Acoustical IEC 60318-5 (IEC 126) coupler
• Frequency counter
• Correction values for coupler
• 30° adaptor cable
From the AUD menu, select Transducer Calibration.
Place the transducer on the 2cc IEC 60318-5 (IEC 126) coupler Fig. 45
(fig. 45)
27
Service Manual
AT235
CIR 33 4
Equipment needed for calibrating the CIR 33
• Measuring amplifier
• Acoustical IEC 60318-5 (IEC 126) coupler, fitted with eartip
10
• Frequency counter
• Correction values for coupler
From the AUD menu, select Transducer Calibration.
Place the transducer on the 2cc IEC 60318-5 (IEC 126) coupler
(fig. 46) Fig. 46
DD65 V2 5
Equipment required:
• Measuring amplifier
• Acoustical IEC 60318-3 (IEC 303) coupler
• Frequency counter
• Correction values for coupler
• Spacer for HDA200 to coupler IEC 60318-3 (IEC 303)
From the AUD menu, select Transducer Calibration.
Place the transducer on the 6cc IEC 60318-3 (IEC 303) coupler
(fig. 47)
Fig. 47
4 As of 2020, CIR33 is discontinued as a AD629 accessory. Hence, this information is applicable for previously purchased transducers only
5 Please note that this set-up does not refer to DD45 or DD65 v1
28
Service Manual
AT235
5 Testing the Tympanometer Functionality
This chapter explains how to test the tympanometer.
5.1 Pressure test
Test maximum pressure limit (red dotted line)
Verify that the actual pressure does not cross the red dotted line at 600 daPa.
While at the current pressure, watch the pressure value in the bottom of the pressure test view. Make sure that the
pressure does not drop more than 5 daPa (0.5 daPa/sec) during the next 10 seconds.
Release pressure by manually removing the probe from its cavity.
Set target pressure to -1000 daPa and verify that the actual pressure does not drop below the red dotted line at -
800 daPa.
While at the current pressure, watch the pressure value in the bottom of the pressure test view. Make sure that the
pressure does not drop more than 5 daPa (0.5 daPa/sec) during the next 10 seconds.
Release pressure by manually removing the probe from its cavity.
Test minimum pressure limit (green dotted line)
Place the probe in a 2ml hard wall cavity and select the slowest pump speed. Set target pressure to 1000 daPa
and verify that the actual pressure does not cross the green dotted line at 500 daPa.
Set target pressure to -1000 daPa and verify that the actual pressure does not drop below the green dotted line at
-700 daPa.
Select Test and then Pressure Test.
Place the probe in the 2cc cavity.
Reading
the graphic
________
.. represents the criteria chosen
with the slider and selected with
’Go to target’.
________
.. represents the actual
values of the chosen criteria.
…. …. ….
.. represent the maximum and
minimum required pressure
levels.
Fig. 34
29
Service Manual
AT235
5.2 Diode test
Select Diode.
Operation of the multicolor probe status indication LED is now enabled for testing.
This LED is situated at the rear of the probe body
Each of the seven colors becomes selectable with the purpose of verifying properly functioning LED glows.
Select Test and then Diode Test.
Check the functionality for all color options.
Fig. 49
30
Service Manual
AT235
5.3 Volume test
Select Volume test.
A dynamic readout of the different probe tones volumes is displayed.
NOTICE At higher probe tone frequencies, larger volume
capacities cannot be measured correctly
Maximum volume for probe tone frequencies:
Frequency Hz 226 678 800 1000
4.23 3.39
Max. cavity volume 8.0 ml 5.0 ml
ml ml
Select Test and then Volume Test.
Fig. 50
31
Service Manual
AT235
Control whether the tone 226 Hz is audible in the probe tip.
Place probe tip in a 2 cc cavity and check that the colored bar is changing color from blue to green.
The volume shows 2 ml (1.90 to 2.10 ml).
Fig 51
32
Service Manual
AT235
5.4 Attenuator test
This test screen enables a manual check of all the input/output channels of the instrument. The screen is divided
into three sections, and must be used with utmost caution, as very high sound intensities can be reached.
CAUTION Very high sound intensities can
be reached.
1. Output signals
In the signal output section, each of the output attenuators are controllable, enabling routing of the sound
to a specific output (i.e. headphones, probe speaker etc.).
• Tone switch ON/OFF
• Pulse ON/OFF
• Stimuli frequency
The attenuator itself is divided into three sections (we refer to the block diagram for more information).
Test points 5, 6 and 7 enable testing of each part of the attenuator.
• Max. output = damping 0 - Min. output = damping 120
• Fixed attenuator level 0 = no damping - Level 25 = add 25 dB attenuation/damping.
• +20 dB attenuation, ON/OFF, ON = add 20dB on top of 4 + 5
When damping is 20 dB or less the +20 box should be ticked. When damping is exactly 20 dB,
ticking the +20 box is optional. Above 20 dB damping, the box should remain empty.
2. Input Signals
In the signal input section, the functionality of an input can be verified. Via the dropdown menu the input
is selected. The slider enables control of the input attenuator’s amplification and the input level data on
the VU-meter.
3. Monitor / Talkback levels
This section verifies the functionality of the monitor / Talkback functionality. By changing position, control
of the levels is enabled.
Fig. 35
33
Service Manual
AT235
As illustrated below, make sure that the values are as follows:
0 dB = 3 V ± 0,3 V
This particular example is measured on the left output.
Fig. 53
34
Service Manual
AT235
Now check at -60 dB = 3mV ± 0,3 mV
This particular example is measured on the left output.
Fig. 54
Repeat for the second attenuator using the Headset1Right output, making sure that the output corresponds with
the selected.
35
Service Manual
AT235
6 Appendix A – Technical Specifications
General
Medical CE-mark: The CE-mark in combination with MD symbol indicates that Interacoustics A/S meets the
requirements of the Medical Device Regulation (EU) 2017/745.
Approval of the quality system is made by TÜV – identification no. 0123.
Standards: Safety: IEC 60601-1, Class I, Type B applied parts
EMC: IEC 60601-1-2+AMD1:2020
Impedance: IEC 60645-5 (2004)/ANSI S3.39 (2012), Type 1
Audiometer: IEC60645-1 (2012)/ANSI S3.6 (2010), Type 4
Operation environment: Temperature: 15° – 35 °C
Relative humidity: 30 – 90%
Ambient pressure: 98 kPa – 104 kPa
Warm-up time: 1 minute
Display 10 inch high resolution color display 1024x600
Transport & storage: Storage temperature: 0° – 50°C
Transport temperature: -20° – 50 °C
Relative humidity: 10 – 95%
Internal storage 500 clients and 50.000 sessions
Internal battery CR2032 3 V, 230 mAh, Li. Not serviceable by user
PC control: USB: Input/output for computer communication
Data can be sent to and saved on the PC and stored in
OtoAccess™ (Diagnostic Suite sync module is required).
Thermal printer Type: MPT-III Thermal MPT-III printer with recording paper in rolls Print
(Optional): on command via USB
Use only specified power supply unit UE60 type
Power supply UES65: Input: 100-240 VAC 50-60 Hz, 2.5 A
Output: 24.0 VDC
Dimensions H x W x L: 29 x 38 x 7,5 cm
AT235 weight 2.5 kg
Impedance measuring system
Probe tone: Frequency: 226 Hz, 678 Hz, 800 Hz, 1000 Hz; pure tones; ±1%
Level: 85 dB SPL (≈ 69 dB HL) ±1.5 dB
Air pressure: Control: Automatic.
Indicator: Measured value is displayed on the graphical display
Range: -600 to +400 daPa. ±5%
Pressure limitation: -750 daPa and +550 daPa
Pump speed: Automatic, Fast 300 daPa/s, Medium 200 daPa/s, Slow 100
daPa/s, Very slow 50 daPa/s.
Compliance: Range: 0.1 to 8.0 ml at 226 Hz probe tone (Ear volume: 0.1 to 8.0
ml) and 0.1 to 15 mmho at 678, 800 and 1000 Hz probe
tone. All ±5%
Test types: Tympanometry: Automatic, where the start and stop pressure can be user-
programmed in the setup function.
Manual control of all functions
Eustachian tube function 1 - Williams test
Non perforated eardrum:
Eustachain tube function 2 Toynbee test
- Perforated eardrum:
Eustachian tube function 3 Continuous sensitive impedance measurement
- Patulous Eustachian tube:
Reflex functions
Signal sources: Tone - Contra, Reflex: 250, 500, 1000, 2000, 3000, 4000, 6000, 8000 Hz, Wide
Band, High and Low pass
Tone - Ipsi, Reflex: 500, 1000, 2000, 3000, 4000 Hz wide band, high and low
pass
NB noise – Contra, Reflex: 250, 500, 1000, 2000, 3000, 4000, 6000, 8000 Hz
I
Service Manual
AT235
NB noise – Ipsi, Reflex: 1000, 2000, 3000, 4000 Hz
Stimulus duration: 750 ms
Reflex acceptance: Adjustable between 2% and 6%, or 0.05 – 0.15 ml change
of ear canal volume
Intervals: Down to 1 dB step size.
Intensity max: 90, 100, 120 dBHL.
Outputs: Contra Earphone: TDH39 earphone, DD45 earphone, CIR insert and/or
EARtone 3A insert for Reflex measurements.
Ipsi Earphone: Probe earphone incorporated in the probe system for
Reflex measurements
Probe connection: Connection of the electrical and air system to the probe
Test types: Manual reflex: Manual control of all functions.
Automated Reflex Single intensities
Reflex growth
Reflex decay: Automatic, 10 dB above threshold and manually controlled
with stimulus durations of 10
Reflex latency: Automated, first 300 ms from stimulus start
Audiometer functions
Signals: Frequencies Hz: Intensities dB HL:
125 -10 to 70
250 -10 to 90
500 -10 to 100
1000 -10 to 100
2000 -10 to 100
3000 -10 to 100
4000 -10 to 100
6000 -10 to 100
8000 -10 to 90
Test types Auto threshold determination (Modified Hughson Westlake)
Auto testing: duration 1-2 s adjusted in 0.1 s intervals
II
Service Manual
AT235
6.1 Calibration properties
Calibrated transducers: Contralateral earphone: Telephonics TDH39/DD45 with a static force of 4.5 N 0.5
N and/or EARtone 3A and/or CIR insert phone
Probe system: Ipsilateral Earphone: is integrated in the probe system
Probe frequency transmitter and receiver and pressure
transducer is integrated in the probe system
Accuracy: General: Generally the instrument is made and calibrated to be
within and better than the tolerances required in the
specified standards:
Reflex Frequencies: ±1%
Contralateral reflex and 3 dB for 250 to 4000 Hz and 5 dB for 6000 to 8000 Hz
audiometer tone levels:
Ipsilateral reflex tone 5 dB for 500 to 2000 Hz and +5/-10 dB for 3000 to 4000
levels: Hz
Pressure measurement: 5% or 10 daPa, whichever is greater
Compliance
measurement: 5% or 0.1 ml, whichever is greater
Stimulus presentation Reflexes: ON-OFF ratio ≥ 70 dB
control: Rise time = 20 ms
Fall time = 20 ms
A weighted SPL in Off = 31 dB
Impedance calibration properties
Probe tone: Frequencies: 226 Hz 1%, 678 Hz 1%, 800 Hz 1%, 1000 Hz 1%
Level: 85 dB SPL 1.5 dB measured in an IEC 60318-5
acoustic coupler. The level is constant for all volumes in
the measurement range.
Distortion: Max 1% THD
Compliance: Range: 0.1 to 8.0 ml
Temperature -0.003 ml/C
dependence:
Pressure dependence: -0.00020 ml/daPa
Reflex sensitivity: 0.001 ml is the lowest detectable volume change
Reflex artifact level: ≥95 dB SPL (measured in the 711 coupler, 0.2 ml, 0.5 ml,
2.0 ml & 5.0 ml hard walled cavities)
Temporal reflex Initial latency = 35 ms (5 ms)
characteristics: Rise time = 42 ms (5 ms)
(IEC60645-5 clause Terminal latency = 23 ms (5 ms)
5.1.6) Fall time = 44 ms (5 ms)
Overshoot = max. 1%
Undershoot = max. 1%
Pressure: Range: Values between -600 to +400 daPa can be selected in the
setup.
Safety limits: -750 daPa and +550 daPa, 50 daPa
Barometric pressure: The barometer pressure Admittance can vary inside: ± 4%
chances influence on the The pressure accuracy is: ±10 daPa or 10%, whichever is
impedance measurement greater
in the specified range
(97300 – 105300
Pascal).
Height above sea level: The pressure sensor used, is a differential/gauge type, which means, it measure the
pressure difference and therefore not affected of the height above sea level.
Probe tones 0 meters 500 meters 1000 meters 2000 meters 4000 meters
226 Hz 1.0 mmho 1.06 mmho 1.13 mmho 1.28 mmho 1.65 mmho
678 Hz 3.0 mmho 3.19 mmho 3.40 mmho 3.85 mmho 4.95 mmho
800 Hz 3.54 mmho 3.77 mmho 4.01 mmho 4.55 mmho 5.84 mmho
1000 Hz 4.42 mmho 4.71 mmho 5.01 mmho 5.68 mmho 7.30 mmho
III
Service Manual
AT235
The pressure accuracy is: ±10 daPa or 10%, whichever is greater.
To minimize the influence of temperature, barometer pressure, humidity and height
above sea level, it always recommended to calibrate the unit in the local positions.
Temperature The temperature has no theoretic impact on the impedance calculation, but the
temperature has influence on the electronic circuits. This temperature influence for
the standard specified temperature range (15-35 °C) is inside:
Admittance can vary inside: ± 5%, ± 0.1 cm3, ±10-9 m3/Pa·s, whichever is greater.
Reflex calibration standards and spectral properties:
General: Specifications for stimulus and audiometer signals are made to follow IEC 60645-5
Contralateral earphone: Pure tone: ISO 389-1 for TDH39 and ISO 389-2 for CIR.
Wide Band noise (WB): Interacoustics Standard
Spectral properties: As “Broad band noise” specified in IEC 60645-5, but with
500 Hz as lower cut-off frequency.
Low Pass noise (LP): Interacoustics Standard
Spectral properties: Uniform from 500 Hz to 1600 Hz, 5 dB re. 1000 Hz level
High Pass noise (HP): Interacoustics Standard
Spectral properties: Uniform from 1600 Hz to 10KHz, 5 dB re. 1000 Hz level
Ipsilateral Earphone: Pure tone: Interacoustics Standard
Wide Band noise (WB): Interacoustics Standard
Spectral properties: As “Broad band noise” specified in IEC 60645-5, but with
500 Hz as lower cut-off frequency.
Low Pass noise (LP): Interacoustics Standard
Spectral properties: Uniform from 500 Hz to 1600 Hz, 10 dB re. 1000 Hz level
High Pass noise (HP): Interacoustics Standard
Spectral properties: Uniform from 1600 Hz to 4000 Hz, 10 dB re. 1000 Hz
level
General about levels: The actual sound pressure level at the eardrum will
depend on the volume of the ear.
The risk of artifacts at higher stimulus levels in reflex measurements are minor and will not activate the reflex
detection system
IV
Service Manual
AT235
6.2 Reference values for stimulus calibration
Freq. Reference Equivalent Threshold Sound Level Variation of Ipsi Sound
(RETSPL) stimulus levels attenuation
[dB re. 20 µPa] for different values for
volumes of the TDH39/DD45
ear canal earphones using
Relative to the MX41/AR or
calibration PN51 cushion
performed on an [dB]
IEC 126 coupler
[dB]
0.5 ml 1 ml
(Interacoustics
(Interacoustics
Interacoustics
Interacoustics
Interacoustics
(ISO 8798)
ISO 389-1
ISO 389-2
ISO 389-4
Standard)
Standard)
Standard
Standard
[Hz] TDH39 EARtone DD65 v2 DD45 Probe NB Standard
3A / IP30 Stimulus
Correction
Values
125 45 26 30,5 47.5 41 4 3
250 25.5 14 17 27 24.5 4 5
500 11.5 5.5 8 13 9.5 4 9.7 5.3 7
1000 7 0 4,5 6 6.5 6 9.7 5.3 15
1500 6.5 2 2,5 8 5 6 21 (1600 Hz)
2000 9 3 2,5 8 12 6 11.7 3.9 26
3000 10 3.5 2 8 11 6 -0.8 -0.5 31 (3150 Hz)
4000 9.5 5.5 9,5 9 3.5 5 -1.6 -0.8 32
6000 15.5 2 21 20.5 3 5 26 (6300 Hz)
8000 13 0 21 12 -5 5 24
RETSPL
WB -8 -5 -8 -8 -5 7.5 3.2
LP -6 -7 -6 -6 -7 8.0 3.6
HP -10 -8 -10 -10 -8 3.9 1.4
V
Service Manual
AT235
6.3 Reference equivalent threshold values for transducers
Impedance - frequencies and intensity ranges
AT235 Maximums IMP
TDH39 DD65v2 EARtone 3A / IP30 IPSI DD45
Center Reading Reading Reading Reading Reading
Freq. Tone NB Tone NB Tone NB Tone NB Tone NB
[Hz] [dB HL] [dB HL] [dB HL] [dB HL] [dB HL] [dB HL] [dB HL] [dB HL] [dB HL] [dB HL]
125 80 65 85 70 100 85 70 60 75 60
250 100 85 100 85 110 100 85 75 100 85
500 120 100 110 95 115 105 100 85 120 100
750 120 105 115 100 120 110 100 85 120 105
1000 120 105 115 100 120 110 105 90 120 105
1500 120 105 115 100 120 110 110 90 120 100
2000 120 105 115 95 120 110 105 90 120 100
3000 120 105 115 100 120 110 95 90 120 105
4000 120 105 110 95 120 105 100 85 120 105
6000 120 100 100 85 105 100 85 80 110 90
8000 105 95 95 70 90 85 80 75 105 95
10000
WB - 115 - 115 - 115 - 95 - 120
LP - 120 - 115 - 120 - 100 - 120
HP - 115 - 115 - 120 - 95 - 120
Table 1
VI
Service Manual
AT235
6.4 Survey of reference and max. hearing level Tone Audiometer
ANSI DD45 IEC DD45
Coupler: IEC 60318-3 1998 (6ccm) Coupler: IEC 60318-3 1998 (6ccm)
Tone Audiometer Tone Audiometer
Tone Narrow Band Noise Tone Narrow Band Noise
PTB – DTU 2010 ISO 389-4 1994 PTB – DTU 2010 ISO 389-4 1994
Frequency RETSPL Max. HL RETSPL Max. HL Frequency RETSPL Max. HL RETSPL Max. HL
125 47.5 85 51.5 65 125 47.5 85 51.5 65
250 27.0 105 31.0 85 250 27.0 105 31.0 85
500 13.0 120 17.0 100 500 13.0 120 17.0 100
750 6.5 120 11.5 105 750 6.5 120 11.5 105
1000 6.0 120 12.0 105 1000 6.0 120 12.0 105
1500 8.0 120 14.0 105 1500 8.0 120 14.0 105
2000 8.0 120 14.0 105 2000 8.0 120 14.0 105
3000 8.0 120 14.0 105 3000 8.0 120 14.0 105
4000 9.0 120 14.0 105 4000 9.0 120 14.0 105
6000 20.5 110 25.5 95 6000 20.5 110 25.5 95
8000 12.0 105 17.0 95 8000 12.0 105 17.0 95
White Noise 0 120 White Noise 0 120
Table 2 Table 3
ANSI TDH39 IEC TDH39
Coupler: ANSI S3.7-1995 (NBS-9A) / Coupler: IEC 60318-3 1998 (6ccm)
IEC 60318-3 1998 (6ccm)
Tone Audiometer Tone Audiometer
Tone Narrow Band Noise Tone Narrow Band Noise
ANSI S3.6-2010 ANSI S3.6-2010 ISO 389-1 1998 ISO 389-4 1994
Frequency RETSPL Max. HL RETSPL Max. HL Frequency RETSPL Max. HL RETSPL Max. HL
125 45.0 85 49.0 70 125 45.0 85 49.0 70
250 25.5 105 29.5 90 250 25.5 105 29.5 90
500 11.5 120 15.5 105 500 11.5 120 15.5 105
750 8.0 120 13.0 105 750 7.5 120 12.5 105
1000 7.0 120 13.0 105 1000 7.0 120 13.0 105
1500 6.5 120 12.5 105 1500 6.5 120 12.5 105
2000 9.0 120 15.0 105 2000 9.0 120 15.0 105
3000 10.0 120 16.0 105 3000 10.0 120 16.0 105
4000 9.5 120 14.5 105 4000 9.5 120 14.5 105
6000 15.5 120 20.5 100 6000 15.5 120 20.5 100
8000 13.0 105 18.0 95 8000 13.0 105 18.0 95
White Noise 0.0 120 White Noise 0.0 120
Table 4 Table 5
ANSI EAR 3A / IP30 IEC EAR 3A / IP30
Coupler: ANSI S3.7-1995 (HA-2 with 5mm rigid Tube) Coupler: IEC 60318-5 2006 2ccm
Tone Audiometer Tone Audiometer
Tone Narrow Band Noise Tone Narrow Band Noise
ANSI S3.6-2010 ANSI S3.6-2010 ISO 389-2 1994 ISO 389-4 1994
Frequency RETSPL Max. HL RETSPL Max. HL Frequency RETSPL Max. HL RETSPL Max. HL
125 26.0 90 30.0 85 125 26.0 90 30.0 85
250 14.0 105 18.0 100 250 14.0 105 18.0 100
500 5.5 110 9.5 105 500 5.5 110 9.5 105
750 2.0 115 7.0 110 750 2.0 115 7.0 110
1000 0.0 120 6.0 110 1000 0.0 120 6.0 110
1500 2.0 120 8.0 110 1500 2.0 120 8.0 110
2000 3.0 120 9.0 110 2000 3.0 120 9.0 110
3000 3.5 120 9.5 110 3000 3.5 120 9.5 110
4000 5.5 115 10.5 105 4000 5.5 115 10.5 105
6000 2.0 100 7.0 95 6000 2.0 100 7.0 95
8000 0.0 90 5.0 90 8000 0.0 90 5.0 90
White noise 0.0 110 White Noise 0.0 110
Table 6 Table 7
VII
Service Manual
AT235
ANSI CIR 33 IEC CIR 33
Coupler ANSI S3.7-1995 (HA-2) Coupler IEC 60318-5 2006 2ccm
Tone Audiometer Tone Audiometer
Tone Narrow Band Noise Tone Narrow Band Noise
ANSI S3.6-2010 ANSI S3.6-2010 ISO 389-2 1994 ISO 389-4 1994
Frequency RETSPL Max. HL RETSPL Max. HL Frequency RETSPL Max. HL RETSPL Max. HL
125 26.0 90 30.0 85 125 26.0 90 30.0 85
250 14.0 105 18.0 100 250 14.0 105 18.0 100
500 5.5 110 9.5 105 500 5.5 110 9.5 105
750 2.0 115 7.0 110 750 2.0 115 7.0 110
1000 0.0 120 6.0 110 1000 0.0 120 6.0 110
1500 2.0 120 8.0 110 1500 2.0 120 8.0 110
2000 3.0 120 9.0 110 2000 3.0 120 9.0 110
3000 3.5 120 9.5 110 3000 3.5 120 9.5 110
4000 5.5 115 10.5 105 4000 5.5 115 10.5 105
6000 2.0 100 7.0 95 6000 2.0 100 7.0 95
8000 0.0 90 5.0 90 8000 0.0 90 5.0 90
White Noise 0.0 110 White Noise 0.0 110
Table 8 Table 9
VIII
Service Manual
AT235
6.5 Survey of reference and max. hearing level Tympanometer
DD45 Contra TDH39 Contra
Center Att. Reading Center Att. Reading
Frequency setting Tone NB Frequency setting Tone NB
[Hz] [dB HL [dB HL] [dB HL] [Hz] [dB HL [dB HL] [dB HL]
125 85 75 60 125 85 80 65
250 85 100 85 250 85 100 85
500 85 115 100 500 85 115 100
750 85 120 105 750 85 120 105
1000 85 120 105 1000 85 120 105
1500 85 120 100 1500 85 120 105
2000 85 115 100 2000 85 120 105
3000 85 125 105 3000 85 120 105
4000 85 115 105 4000 85 120 105
6000 85 110 90 6000 85 120 100
8000 75/70 105 95 8000 75/70 105 95
WB 85 - 120 WB 85 - 115
LP 85 - 120 LP 85 - 120
HP 85 - 120 HP 85 - 115
Table 10 Table 11
IOW IPSI Sp. IP30 Contra / EAR 3A Contra
Center Att. Reading Center Att. Reading
Frequency Setting Tone NB Frequency setting Tone NB
[Hz] [dB HL [dB HL] [dB HL] [Hz] [dB HL [dB HL] [dB HL]
125 85 70 60 125 85 100 85
250 85 85 75 250 85 110 100
500 85 100 85 500 85 115 105
750 85 100 85 750 85 120 110
1000 85 105 90 1000 85 120 110
1500 85 110 90 1500 85 120 110
2000 85 105 90 2000 85 120 110
3000 85 95 90 3000 85 120 110
4000 85 100 85 4000 85 120 105
6000 85 85 80 6000 85 105 100
8000 75/70 80 75 8000 75/70 90 85
WB 85 - 95 WB 85 - 115
LP 85 - 100 LP 85 - 120
HP 85 - 95 HP 85 - 120
Table 12 Table 13
CIR 33 Contra
Center Att. Reading
Frequency setting Tone NB
[Hz] [dB HL [dB HL] [dB HL]
125 85 90 70
250 85 100 85
500 85 110 100
750 85 110 105
1000 85 115 105
1500 85 115 105
2000 85 115 105
3000 85 115 105
4000 85 110 100
6000 85 95 95
8000 80/85 75 80
WB 85 - 115
LP 85 - 115
HP 85 - 115
Table 14
IX
Service Manual
AT235
7 Appendix B – Parts & Accessories
ACCESSORIES PART NUMBER DESCRIPTION
8502562 Daily check cavity
8011296 CAT50
8108148 Diagnostic probe system 1077
8108189 Clinical probe system 1077
8503373 TDH39P contra headset P3045 neutral
8106919 DD45 contra headset P3045
8520328 DD65v2 headset
8103323 IP30 insert phone 10ohm single contra
8106339 DD45 audiometric headset
8106355 TDH39 audiometric headset HBA headband
8101884 IP30 insert phone 10ohm set
X
Service Manual
AT235
ACCESSORIES PART NUMBER DESCRIPTION
8517507 RadioEar Circumaural Headband cable
8517499 RadioEar Cushions for Circumaural Headbands
8105761 Strap -key kit
8106038 Probe floss kit
8105696 Printer kit 1077
8106795 Wall mounting kit 1077
8103838 APS3 patient response switch
8106790 Diagnostic probe holder 1077
XI
Service Manual
AT235
CONSUMABLES PART NUMBER DESCRIPTION
8500580 Kit eartip adaptor, 6pcs
8500830 Eartip adaptor cleaning tool
8011348 Probe cleaning tool
Sanibel Ear Tip Kit
Eartips of various sizes (8511445)
Types: umbrella, mushroom, flanged
Single Use Eartip Box (8515491)
TPR MPT-III thermal paper 8500610
Foam pad HBC39 headset. 10 pcs (1 bag) 8002802
XII
Service Manual
AT235
SPARE PARTS PART NUMBER DESCRIPTION
8108046 Pump tube kit 1077
8107762 Probe tip kit 1077 incl. gasket
8004140 Cable for pump 24 pins
8000656 Pump module
8102693 Gasket
8527201 Mainboard with FW Replacement Kit AT235 1043
8108143 PCA keyboard 1077
8108144 Bottom 1077
8108145 PCA probe connection board
8106874 Rotary encoder 1077
8102178 Knob plastic 1077
XIII
Service Manual
AT235
SPARE PARTS PART NUMBER DESCRIPTION
8102071 Display cable 1077
8004158 Cable for keyboard 40 pins
8101484 Rotary dial cover 1077
8108146 Top cabinet
8108147 Display 1077
8524210 Power supply 24V 60W UES65-240250SPA3
8507211 Contra headband P3045
XIV
Service Manual
AT235
8 Appendix C – Firmware Update Procedure
New firmware will be made available on the Interacoustics website. After downloading the zip file follow the
below instruction:
1. Unzip the firmware archive on a new FAT32 formatted USB stick
2. Insert the USB in the unpowered instrument
3. Turn on the instrument and wait for Tone test
4. Enter Common Setting to 1.5 and below OR Instrument setting for 1.6 or higher using the Setup
Button
5. Press the function key named Install, and follow the instructions on the instrument display
6. Wait for the instrument to return to Tone test
CAUTION Restarting the instrument after installing the firmware may be necessary.
Do not shut down the instrument or remove the USB stick while updating.
AT235 Stand Alone Unit - Print Logo Update
AT235 direct printout allows results to be printed directly after testing (via a supported USB printer – please
contact Interacoustics customer service for a list of supported PC printers if in doubt). The printout logo can be
configured via the audiometer itself (see below) or downloaded via the Diagnostic Suite (in the General Setup).
1. Open the program Paint which can be found in the accessories folder.
2. Open Image Properties using the shortcut Ctrl + E.
3. Set Width to 945 and Height to 190 as shown. Click OK.
4. Edit the image and the company data to fit inside the set area.
5. Save the created file as PrintLogo.bmp.
6. Zip the PrintLogo.bmp file to the following name format Update_user.logo.bin.
The update_user.logo.bin file is now ready to be used
7. Insert a USB thumb drive with at least 32MB in total size
8. Go to My Computer and right click on the USB thumb drive and select Format
This action will erase everything on your USB
CAUTION thumb drive!
XV
Service Manual
AT235
9. Ensure that FAT32 is selected as your file system - Leave other settings as listed.
10. Click start - depending on the size of your thumb drive this may take a moment. When the formatting
is complete you will receive a pop-up indicating successful formatting.
11. Copy the update_user.logo.bin file to the formatted thumb drive.
12. It is very important that this file is the ONLY FILE on this USB thumb drive.
13. With the audiometer switched off, insert the thumb drive into any available USB port.
14. Then switch the instrument on and push the Temp/Setup Button from the Tone test screen.
15. Enter Common Settings using the Setup/Tests Button.
16. To the question “Do you want to install?” select Yes.
17. After the installation is completed, select Back to get to the testing screen.
XVI
Service Manual
AT235
9 Appendix D – update history
The following modifications have been made to this instrument and/or service manual:
DATE ACTION REMARKS
2023/09 Revision of the service Part list revised (8104678 removed)
manual
2023/03 Revision of the service Update history added
manual Fax no. removed
Block diagram removed
XVII