G7110 G7111 G5654 Iso Quat Pumps - SerMa en SD 29000179
G7110 G7111 G5654 Iso Quat Pumps - SerMa en SD 29000179
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual
Notices
Document Information Warranty Safety Notices
Document No: SD-29000179 Rev. D The material contained in this document
Edition: 07/2023 is provided “as is,” and is subject to being C AU T I O N
changed, without notice, in future edi-
A CAUTION notice denotes a
Copyright tions. Further, to the maximum extent
permitted by applicable law, Agilent dis- hazard. It calls attention to an
© Agilent Technologies, Inc. 2016-2023 claims all warranties, either express or operating procedure, practice, or
No part of this manual may be repro- implied, with regard to this manual and the like that, if not correctly per-
duced in any form or by any means any information contained herein, includ- formed or adhered to, could result
(including electronic storage and retrieval ing but not limited to the implied warran- in damage to the product or loss
or translation into a foreign language) ties of merchantability and fitness for a
of important data. Do not proceed
without prior agreement and written con- particular purpose. Agilent shall not be lia-
sent from Agilent Technologies, Inc. as ble for errors or for incidental or conse- beyond a CAUTION notice until
governed by United States and interna- quential damages in connection with the the indicated conditions are fully
tional copyright laws. furnishing, use, or performance of this understood and met.
document or of any information con-
tained herein. Should Agilent and the user WAR N IN G
Agilent Technologies have a separate written agreement with
Hewlett-Packard-Strasse 8 warranty terms covering the material in A WARNING notice denotes a
76337 Waldbronn, Germany this document that conflict with these hazard. It calls attention to an
terms, the warranty terms in the separate operating procedure, practice, or
agreement shall control.
the like that, if not correctly per-
Technology Licenses formed or adhered to, could
result in personal injury or death.
The hardware and/or software described Do not proceed beyond a WARN-
in this document are furnished under a
ING notice until the indicated
license and may be used or copied only in
accordance with the terms of such conditions are fully understood
license. and met.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual
In This Guide...
1 Introduction
This chapter gives an introduction to the module, instrument overview, and internal
connectors.
5 Optimizing Performance
This chapter gives hints on how to optimize the performance or use additional
devices.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 3
7 Error Information
This chapter describes the meaning of error messages, and provides information
on probable causes and suggested actions how to recover from error conditions.
11 Identifying Cables
This chapter provides information on cables used with the Agilent 1200 Infinity
Series modules.
12 Hardware Information
This chapter describes the pump in more detail on hardware and electronics.
13 LAN Configuration
This chapter provides information on connecting the module to the Agilent
ChemStation PC.
14 Appendix
This chapter provides addition information on safety, legal, and web.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 4
Contents
1 Introduction 10
Introduction to the Pump 11
G7110B Isocratic Pump 12
G7111A Quaternary Pump VL 14
G7111B Quaternary Pump 16
G5654A Bio-inert Pump 18
Overview of the Hydraulic Path 20
Leak and Waste Handling 27
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 5
4 Using the Pump 69
Magnets 70
Turn on/off 71
Status Indicators 73
Best Practices 74
Leak and Waste Handling 76
Hints for Successful Use of the Pump 78
Setting up the Pump with the G4208A Instant Pilot 79
Setting up the Pump with the Instrument Control Interface 80
Solvent Information 88
Algae Growth in HPLC Systems 98
Prevent Blocking of Solvent Filters 99
Normal Phase Applications 100
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 6
9 Maintenance and Repair 147
Introduction to Maintenance and Repair 149
Cautions and Warnings 150
Overview of Maintenance and Repair 152
Cleaning the Module 153
Remove and Install Doors 154
Checking and Replacing the Solvent Filter 156
Exchange the Passive Inlet Valve 157
Exchange the Outlet Valve 159
Exchange the Purge Valve Frit or the Purge Valve 161
Replace the O-Ring on the Purge Valve 164
Remove the Pump Head Assembly 166
Maintenance of a Pump Head Without Seal Wash Option 168
Maintenance of a Pump Head with Seal Wash Option 171
Reinstall the Pump Head Assembly 176
Seal Wear-in Procedure 179
Exchange the Multi-Channel Gradient Valve (MCGV) 180
Replace the Solvent Selection Valve (Isocratic Pump) 183
Exchange the Active Inlet Valve (AIV) or its Cartridge 185
Exchange the Seal Wash Cartridge 187
Replace Leak Handling System Parts 188
Replacing the Module Firmware 189
Introduction to Repair 190
Overview of Repairing the Pump 191
Exchanging Internal Parts 193
Remove the Top Cover and Foam 194
Exchange the Pump Mainboard 196
Replace the Fusion Board 202
Changing the Type and Serial Number 203
Exchange the Damper 208
Exchange the Pump Drive 212
Exchange the Fan 214
Replace Power Supply and Cables 216
Replace the Power Switch 219
Replace the Degasser 221
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 7
Replace Parts Inside the 4-Channel-Degasser 223
Replace Parts Inside the 2-Channel-Degasser 230
Replace the Leak Panel 236
Replace the Leak Sensor 238
Replace the Status LED 239
Install an Active Seal Wash Upgrade Kit (Quaternary Pump) 241
Assemble the Main Cover 244
Install the Top Cover and Foam 245
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 8
11 Identifying Cables 281
Cable Overview 282
Analog Cables 284
Remote Cables 286
CAN Cable 290
Agilent Module to PC 291
Agilent 1200 Module to Printer 292
14 Appendix 336
General Safety Information 337
Waste Electrical and Electronic Equipment (WEEE) Directive 343
Radio Interference 344
Sound Emission 345
Agilent Technologies on Internet 346
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 9
1 Introduction
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 10
1 Introduction
Introduction to the Pump
The Pumps comprise one pump head unit with an optional solvent cabinet, a
vacuum degasser (optional for the isocratic pump), and a multi-channel gradient
valve (MCGV) for quaternary pumps, or an optional solvent selection valve (SSV)
for isocratic pumps. The combination of the pump head assembly and
high-speed proportioning MCGV with a possible connection up to four different
solvents in quaternary pumps provides gradient generation by low pressure
mixing. The combination of the pump head assembly and SSV in isocratic pump
provides opportunity to make a choice between three different solvents. A
solvent cabinet provides enough space for eight one-liter bottles. The Active Seal
Wash (ASW) and Active Inlet Valve (AIV) options are available for quaternary
pumps when concentrated buffer solutions are used. The Bio-inert pump is
optimized for these conditions, thus is always equipped with ASW and AIV.
Degasser
Outlet Valve
Purge Valve
Inlet Valve
MCGV
Pump Head
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 11
1 Introduction
G7110B Isocratic Pump
Product Description
The Agilent 1260 Infinity II Isocratic Pump is ideal for demanding QA/QC tasks
and routine applications. It maintains virtually pulse-free and stable solvent flows.
The dual floating, precise servo-controlled pistons in the delivery mechanism
adjust the stroke volume according to your chosen flow rate.
The optional integrated degasser and solvent selection valve offers increased
ease-of-use and method flexibility.
The pump can deliver a broad range of pressures up to a maximum 600 bar,
giving you the flexibility to use small particle size columns, longer columns or
alternative higher viscosity solvents.
The Agilent 1260 Infinity II Isocratic Pump is the ideal pump for GPC/SEC
applications where run-to-run and day-to-day precision in retention times is
crucial.
Status LED
Purge valve
Power switch
Leak drain
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 12
1 Introduction
G7110B Isocratic Pump
Features
• Higher resolution and increased throughput with the new standard 600 bar
pressure for modern LC columns.
• Low operating and maintenance costs - due to excellent instrument uptime.
• Future proof - your investment is protected since you can upgrade to a
quaternary pump if required.
• Versatile wide flow range – from 0.2 mL/min to 10 mL/min for analytical and
semi-preparative separations and purifications.
• Fast and easy maintenance - front access to all maintenance parts means you
can change components quickly without having to disassemble the
instrument.
• Longevity and low operating costs - robust materials (stainless steel, titanium,
gold, ruby, sapphire, ceramics, PEEK and PTFE) withstand the most
demanding applications and keep your pump functioning for years. Our
optional seal wash for high-salt mobile phases keeps maintenance to a
minimum and reduces operating costs.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 13
1 Introduction
G7111A Quaternary Pump VL
Product Description
The Agilent 1260 Infinity II Quaternary Pump VL has an extended power range,
delivering pressures up to 400 bar. At these high pressures you can use smaller
particle size columns and get higher resolution and faster separations. It
maintains virtually pulse-free, well mixed and stable solvent flows. Its dual
floating, precise servo-controlled pistons adjust the stroke volume according to
your chosen flow rate.
Offering access to up to four solvents, the Agilent 1260 Infinity II Quaternary
Pump VL provides the greatest flexibility in automated solvent blending and is
recommended for a wide range of research and routine applications (especially
in the food, environmental and pharmaceutical sectors) as well as for method
development.
Status indicator
Degasser
Purge valve
MCGV
Power switch
Leak drain
Figure 3 Overview of the pump
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 14
1 Introduction
G7111A Quaternary Pump VL
Features
• Higher resolution, faster separations - the maximum pressure of 400 bar
allows you to use smaller particle size columns, longer columns and/or higher
viscous solvents as an alternative to acetonitrile.
• Analytical and semi-preparative purifications - the wide flow range (up to
10 mL/min) perfectly supports standard and semi-preparative applications.
The perfect choice for 4.6 mm and 3.0 mm id columns.
• Quicker experiments - the pump offers convenient access to up to four
solvents for isocratic or gradient analysis which can really speed up method
development, preparation of mobile phases and system flushing.
• Easy, fast and secure transfer of methods from HPLC to sub-2-micron
columns.
• Integrated degasser
• Longevity and low operating costs - robust materials (stainless steel, titanium,
gold, ruby, sapphire, ceramics, PEEK and PTFE) withstand the most
demanding applications and keep your pump functioning for years. Our
optional seal wash for high-salt mobile phases keeps maintenance to a
minimum and reduces operating costs.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 15
1 Introduction
G7111B Quaternary Pump
Product Description
The Agilent 1260 Infinity II Quaternary Pump has an extended power range,
delivering pressures up to 600 bar. At these high pressures you can use smaller
particle size columns and get higher resolution and faster separations. It
maintains virtually pulse-free, well mixed and stable solvent flows. Its dual
floating, precise servo-controlled pistons adjust the stroke volume according to
your chosen flow rate.
Offering access to up to four solvents, the Agilent 1260 Infinity II Quaternary
Pump provides the greatest flexibility in automated solvent blending and is
recommended for a wide range of research and routine applications (especially
in the food, environmental and pharmaceutical sectors) as well as for method
development.
Status indicator
Degasser
Purge valve
MCGV
Power switch
Leak drain
Figure 4 Overview of the pump
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 16
1 Introduction
G7111B Quaternary Pump
Features
• Higher resolution, faster separations - the maximum pressure of 600 bar
allows you to use smaller particle size columns, longer columns and/or higher
viscous solvents as an alternative to acetonitrile.
• Analytical and semi-preparative purifications - the wide flow range (up to
10 mL/min) perfectly supports standard and semi-preparative applications.
The perfect choice for 4.6 mm and 3.0 mm id columns.
• Quicker experiments - the pump offers convenient access to up to four
solvents for isocratic or gradient analysis which can really speed up method
development, preparation of mobile phases and system flushing.
• Easy, fast and secure transfer of methods from HPLC to sub-2-micron
columns.
• Integrated degasser
• Longevity and low operating costs - robust materials (stainless steel, titanium,
gold, ruby, sapphire, ceramics, PEEK and PTFE) withstand the most
demanding applications and keep your pump functioning for years. Our
optional seal wash for high-salt mobile phases keeps maintenance to a
minimum and reduces operating costs.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 17
1 Introduction
G5654A Bio-inert Pump
Product Description
The 1260 Infinity II Bio-Inert Pump is the pump of choice for all your biological
and extreme pH applications. The titanium-based pump offers highest corrosion
resistance against high salt concentration (2 M) and offers a handling of a wide
pH range (1 – 13, short term 14). It features a pressure range of up to 600 bar
and a flow rate up to 10 mL/min (at 200 bar). Which enables the use of almost
any column: conventional, sub-2 µm particle, or superficially porous columns.
Together with the Agilent Bio-HPLC column portfolio for SEC and IEX, the highest
resolution per time is achieved for protein and NBE characterization. The Agilent
Buffer Advisor software allows fast pH scouting and easy buffer/solvent
preparation in ion exchange chromatography.
Status indicator
Degasser
Purge valve
MCGV
Power switch
Leak drain
Figure 5 Overview of the Bio-inert Pump
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 18
1 Introduction
G5654A Bio-inert Pump
Features
• Titanium pump - for highest corrosion resistance against high salt
concentrations and harsh cleaning procedures
• Steel and iron-free wetted parts in solvent delivery unit - ensure the integrity of
bio-molecules and minimize unwanted surface interactions while increasing
column life-time
• High salt tolerance (2 M) and wide pH range (1 – 13, short term 14) - for
increased flexibility
• Flow rates up to 10 mL/min - enable power ranges from lowest pressure for
traditional bio-purification columns up to high pressure STM analytical
bio-columns
• Active seal wash and quaternary solvent blending included
• Comes with Agilent Buffer Advisor software - for faster pH scouting and easy
buffer/solvent preparation in ion exchange chromatography
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 19
1 Introduction
Overview of the Hydraulic Path
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 20
1 Introduction
Overview of the Hydraulic Path
Hydraulic Path
to sampling unit and column
Seal wash
pump Damper
Purge valve
Outlet
Inlet valve valve
to waste
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 21
1 Introduction
Overview of the Hydraulic Path
Seal wash
pump Damper
Degasser
Purge valve
Outlet
Inlet valve valve
to waste
MCGV
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 22
1 Introduction
Overview of the Hydraulic Path
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 23
1 Introduction
Overview of the Hydraulic Path
Damper
Piston 1
Piston 2
Gear
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 24
1 Introduction
Overview of the Hydraulic Path
When turned on, the pump runs through an initialization procedure to determine
the upper dead position of the first piston. The first piston moves slowly upwards
into the mechanical stop of the pump chamber and from there it moves back for
a defined distance. The controller stores this piston position in memory. After
this initialization the pump starts operation with the set parameters. The inlet
valve opens and the down-moving piston draws solvent into the first pump
chamber. At the same time, the second piston moves upwards delivering to the
system. After a controller-defined stroke length that depends on the flow rate the
drive motor is stopped and the inlet valve closes. The motor direction is reversed
and moves the first piston up until it reaches the stored upper limit and at the
same time the second piston moves downwards. Then the sequence starts
again moving the pistons up and down between the two limits. During the up
movement of the first piston, the solvent in the pump chamber is pressed
through the outlet valve into the second pump chamber. The second piston
draws in half of the volume displaced by the first piston and the remaining half
volume is directly delivered to the system. During the drawing stroke of the first
piston, the second piston delivers the drawn volume to the system.
Quaternary pumps: For solvent compositions from the solvent bottles A, B, C, D
the controller divides the length of the intake stroke in certain fractions in which
the gradient valve connects the specified solvent channel to the pump input.
For specifications, see:
• isocratic pump: “Performance Specifications Agilent 1260 Infinity II Isocratic
Pump (G7110B)” on page 40
• quaternary pumps:
• “Performance Specifications Agilent 1260 Infinity II Quaternary Pump
(G7111B)” on page 44
• “Performance Specifications Agilent 1260 Infinity II Quaternary Pump VL
(G7111A)” on page 42
• Bio-inert pump: “Performance Specifications Agilent 1260 Infinity II Bio-inert
Pump (G5654A)” on page 46
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 25
1 Introduction
Overview of the Hydraulic Path
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 26
1 Introduction
Leak and Waste Handling
The Agilent InfinityLab LC Series has been designed for safe leak and waste
handling. It is important that all security concepts are understood and
instructions are carefully followed.
The solvent cabinet is designed to store a maximum volume of 8 L solvent. The
maximum volume for an individual bottle stored in the solvent cabinet should not
exceed 2 L. For details, see the usage guideline for the Agilent Infinity II Solvent
Cabinets (a printed copy of the guideline has been shipped with the solvent
cabinet, electronic copies are available on the Internet).
All leak plane outlets are situated in a consistent position so that all Infinity and
Infinity II modules can be stacked on top of each other. Waste tubes are guided
through a channel on the right hand side of the instrument, keeping the front
access clear from tubes.
The leak plane provides leak management by catching all internal liquid leaks,
guiding them to the leak sensor for leak detection, and passing them on to the
next module below, if the leak sensor fails. The leak sensor in the leak plane
stops the running system as soon as the leak detection level is reached.
Solvent and condensate is guided through the waste channel into the waste
container:
• from the detector's flow cell outlet
• from the Multisampler needle wash port
• from the Sample Cooler or Sample Thermostat (condensate)
• from the pump's Seal Wash Sensor (if applicable)
• from the pump's Purge Valve or Multipurpose Valve
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 27
1 Introduction
Leak and Waste Handling
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 28
1 Introduction
Leak and Waste Handling
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 29
1 Introduction
Leak and Waste Handling
The waste tube connected to the leak plane outlet on each of the bottom
instruments guides the solvent to a suitable waste container.
Leak Sensor
Solvent incompatibility
C AU T I O N
The solvent DMF (dimethylformamide) leads to corrosion of the leak sensor.
The material of the leak sensor, PVDF (polyvinylidene fluoride), is incompatible
with DMF.
Do not use DMF as mobile phase.
Check the leak sensor regularly for corrosion.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 30
1 Introduction
Leak and Waste Handling
Waste Concept
1 Agilent recommends using the 6 L waste can with 1 Stay Safe cap GL45 with
4 ports (5043-1221) for optimal and safe waste disposal. If you decide to use
your own waste solution, make sure that the tubes don't immerse in the liquid.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 31
2 Site Requirements and Specifications
Site Requirements 33
Physical Specifications 36
Physical Specifications Agilent 1260 Infinity II Isocratic Pump (G7110B) 36
Physical Specifications Agilent 1260 Infinity II Quaternary Pump VL (G7111A) 37
Physical Specifications Agilent 1260 Infinity II Quaternary Pump (G7111B) 38
Physical Specifications Agilent 1260 Infinity II Bio-inert Pump (G5654A) 39
Performance Specifications 40
Performance Specifications Agilent 1260 Infinity II Isocratic Pump (G7110B) 40
Performance Specifications Agilent 1260 Infinity II Quaternary Pump VL (G7111A) 42
Performance Specifications Agilent 1260 Infinity II Quaternary Pump (G7111B) 44
Performance Specifications Agilent 1260 Infinity II Bio-inert Pump (G5654A) 46
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 32
2 Site Requirements and Specifications
Site Requirements
Site Requirements
Power Considerations
The module power supply has wide ranging capability. It accepts any line voltage
in the range described in Table 1 on page 36. Consequently there is no voltage
selector in the rear of the module. There are also no externally accessible fuses,
because automatic electronic fuses are implemented in the power supply.
Module is partially energized when switched off, as long as the power cord
WAR N IN G
is plugged in.
Repair work at the module can lead to personal injuries, e.g. shock hazard,
when the cover is opened and the module is connected to power.
Make sure that it is always possible to access the power plug.
Remove the power cable from the instrument before opening the cover.
Do not connect the power cable to the Instrument while the covers are
removed.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 33
2 Site Requirements and Specifications
Site Requirements
Power Cords
Country-specific power cords are available for the module. The female end of all
power cords is identical. It plugs into the power-input socket at the rear. The male
end of each power cord is different and designed to match the wall socket of a
particular country or region.
Agilent makes sure that your instrument is shipped with the power cord that is
suitable for your particular country or region.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 34
2 Site Requirements and Specifications
Site Requirements
Bench Space
The module dimensions and weight (see Table 1 on page 36) allow you to place
the module on almost any desk or laboratory bench. It needs an additional
2.5 cm (1.0 inches) of space on either side and approximately 8 cm (3.1 inches)
in the rear for air circulation and electric connections.
If the bench shall carry a complete HPLC system, make sure that the bench is
designed to bear the weight of all modules.
The module should be operated in a horizontal position.
Agilent recommends that you install the HPLC instrument in the InfinityLab Flex
NOTE Bench rack. This option helps to save bench space as all modules can be placed
into one single stack. It also allows to easily relocate the instrument to another
lab.
Condensation
Condensation within the module
C AU T I O N
Condensation can damage the system electronics.
Do not store, ship or use your module under conditions where temperature
fluctuations could cause condensation within the module.
If your module was shipped in cold weather, leave it in its box and allow it to
warm slowly to room temperature to avoid condensation.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 35
2 Site Requirements and Specifications
Physical Specifications
Physical Specifications
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 36
2 Site Requirements and Specifications
Physical Specifications
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 37
2 Site Requirements and Specifications
Physical Specifications
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 38
2 Site Requirements and Specifications
Physical Specifications
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 39
2 Site Requirements and Specifications
Performance Specifications
Performance Specifications
Flow precision ≤0.07 % RSD, or <0.02 min SD, Based on retention time at
whichever is greater constant temperature
Pressure pulsation < 2 % amplitude (typically < 1.3 %), At 1 mL/min isopropanol, at all
or < 0.3 MPa (3 bar, 44 psi), pressures > 1 MPa
whichever is greater (10 bar, 145 psi)
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 40
2 Site Requirements and Specifications
Performance Specifications
Instrument Control LC & CE Drivers A.02.14 or above For details about supported
Instrument Control Framework software versions refer to the
(ICF) A.02.04 or above compatibility matrix of the LC &
InfinityLab LC Companion CE Drivers version in use
(G7108AA) with firmware D.07.25
or above
Instant Pilot (G4208A) with
firmware B.02.20 or above
Lab Advisor B.02.08 or above
For use with flow rates below 500 µL/min a vacuum degasser is required.
NOTE
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 41
2 Site Requirements and Specifications
Performance Specifications
Flow precision ≤0.07 % RSD, or ≤0.02 min SD, based on retention time at
whichever is greater constant temperature
Pressure pulsation < 2 % amplitude (typically < 1.3 %), At 1 mL/min isopropanol, at all
or < 0.3 MPa (3 bar, 44 psi), pressures > 1 MPa
whichever is greater (10 bar, 145 psi)
Delay volume 600 – 900 µL, dependent on back Measured with water at
pressure 1 mL/min
(water/water with tracer)
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 42
2 Site Requirements and Specifications
Performance Specifications
Composition precision < 0.2 % RSD or < 0.04 min SD, At 1 mL/min; based on
whichever is greater retention time at constant
temperature
Instrument Control LC & CE Drivers A.02.14 or above For details about supported
Instrument Control Framework software versions refer to the
(ICF) A.02.04 or above compatibility matrix of the LC &
InfinityLab LC Companion CE Drivers version in use
(G7108AA) with firmware D.07.25
or above
Instant Pilot (G4208A) with
firmware B.02.20 or above
Lab Advisor software B.02.08 or
above
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 43
2 Site Requirements and Specifications
Performance Specifications
Flow precision ≤0.07 % RSD, or ≤0.02 min SD, Based on retention time at
whichever is greater constant temperature
Pressure pulsation < 2 % amplitude (typically < 1.3 %), At 1 mL/min isopropanol, at all
or < 0.3 MPa (3 bar, 44 psi), pressures > 1 MPa
whichever is greater (10 bar, 145 psi)
Delay volume 600 – 900 µL, dependent on back Measured with water at
pressure 1 mL/min
(water/water with tracer)
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 44
2 Site Requirements and Specifications
Performance Specifications
Composition precision < 0.2 % RSD or < 0.04 min SD, At 1 mL/min; based on
whichever is greater retention time at constant
temperature
Instrument Control LC & CE Drivers A.02.14 or above For details about supported
Instrument Control Framework software versions refer to the
(ICF) A.02.04 or above compatibility matrix of the LC &
InfinityLab LC Companion CE Drivers version in use
(G7108AA) with firmware D.07.25
or above
Instant Pilot (G4208A) with
firmware B.02.20 or above
Lab Advisor software B.02.08 or
above
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 45
2 Site Requirements and Specifications
Performance Specifications
Type Specification
Hydraulic system Dual piston in series pump with proprietary servo-controlled variable stroke
drive, floating pistons and active inlet valve, integrated 4-channel degassing
unit
Settable flow range 0.001 – 10 mL/min, in 0.001 mL/min increments
Flow precision < 0.07 % RSD, or <0.02 min SD whichever is greater, based on retention time
at constant room temperature
Flow accuracy ± 1 % or 10 µL/min whichever is greater
Pressure Operating range up to 60 MPa (600 bar, 8700 psi) up to 5 mL/min
Operating range up to 20 MPa (200 bar, 2950 psi) up to 10 mL/min
Pressure pulsation < 2 % amplitude (typically < 1.3 %), at 1 mL/min isopropanol, at all pressures
> 1 MPa (10 bar)
Compressibility User-selectable, based on mobile phase compressibility
compensation
Recommended pH 1 – 13, short term 141
range
Gradient formation Low pressure quaternary mixing/gradient capability using proprietary
high-speed proportioning valve
Delay volume 600 – 900 µL, dependent on back pressure.
Settable composition 0 – 100 % in 0.1 % increments
range
Composition < 0.2 % RSD or < 0.04 min SD, whichever is greater, at 1 mL/min; based on
precision retention time at constant room temperature
Instrument Control Lab Advisor B.02.08 or above
LC and CE Drivers A.02.14 or above
For details about supported software versions refer to the compatibility
matrix of your version of the LC and CE Drivers
Local Control Agilent Instant Pilot (G4208A) B.02.20 or above
Communications Controller-area network (CAN), USB
Extended Remote Interface: ready, start, stop and shut-down signals
Materials in flowpath Titanium, Gold, Platinum-Iridium, Sapphire, PEEK, PTFE, Ruby, Ceramic, FEP,
PFA
Active seal wash Included
1 For solvent compatibility, refer to section "Solvent information" in the manual for the Agilent 1260 Infinity
Bio-inert LC System.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 46
3 Installing the Pump
This chapter gives information about the preferred stack setup for your system
and the installation of your module.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 47
3 Installing the Pump
Unpacking the Pump
If the delivery packaging shows signs of external damage, please call your Agilent
Technologies sales and service office immediately. Inform your service
representative that the instrument may have been damaged during shipment.
Delivery Checklist
Ensure all parts and materials have been delivered with the pump. For checking
the completeness of your specific shipment, please use the list included in your
shipment. To aid in parts identification, please refer to chapter Parts and
Materials for Maintenance. Please report missing or damaged parts to your local
Agilent Technologies sales and service office.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 48
3 Installing the Pump
Unpacking the Pump
Accessory Kits
Accessory Kit G7111-68755
The Accessory Kit (G7111-68755) contains the following items:
# p/n Description
2 5043-1013 Tubing Clip IF-II
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 49
3 Installing the Pump
Unpacking the Pump
# p/n Description
2 5043-1013 Tubing Clip IF-II
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 50
3 Installing the Pump
Optimizing the Stack Configuration
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 51
3 Installing the Pump
Optimizing the Stack Configuration
Solvent cabinet
Shelves
Detector
Column
compartment
Handle grip
Sampler
Pump
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 52
3 Installing the Pump
Optimizing the Stack Configuration
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 53
3 Installing the Pump
Optimizing the Stack Configuration
Solvent cabinet
Detector
Column
compartment
Sampler
Pump
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 54
3 Installing the Pump
Optimizing the Stack Configuration
Solvent cabinet
Sampler
Column
compartment
Detector
Pump
Figure 15 Single stack configuration (bench installation, example shows a Vialsampler with
optional ICC installed)
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 55
3 Installing the Pump
Optimizing the Stack Configuration
Solvent cabinet
Detector
MCT
Vialsampler
Pump
Figure 16 Single stack configuration (bench installation, example shows a Vialsampler coupled
with a standalone column compartment)
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 56
3 Installing the Pump
Optimizing the Stack Configuration
Column compartment
Detectors
Solvent cabinet
Sampler
Pump
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 57
3 Installing the Pump
Optimizing the Stack Configuration
Column compartment
Detectors
Solvent cabinet
Sampler
Pump
Figure 18 Two stack configuration (bench installation, example shows a Vialsampler coupled
with a standalone column compartment)
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 58
3 Installing the Pump
Installing the Pump
For other cables see text below and “Cable Overview” on page 282.
Module is partially energized when switched off, as long as the power cord
WAR N IN G
is plugged in.
Repair work at the module can lead to personal injuries, e.g. shock hazard,
when the cover is opened and the module is connected to power.
Make sure that it is always possible to access the power plug.
Remove the power cable from the instrument before opening the cover.
Do not connect the power cable to the Instrument while the covers are
removed.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 59
3 Installing the Pump
Installing the Pump
Status indicator
Power switch
Serial number
Figure 19 Front of pump
3 Connect the power cable to the power connector at the rear of the module.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 60
3 Installing the Pump
Installing the Pump
4 Connect the required interface cables to the pump, see “Connecting Modules
and Control Software” on page 62.
Configuration switch
USB-Mini-Port
LAN
ERI
CAN
Power socket
5 Connect all capillaries, solvent tubes and waste tubing (see “Flow
Connections of the Pump” on page 63).
6 Press the power switch to turn on the module.
The power switch stays pressed in and a green indicator lamp in the power
NOTE switch is on when the module is turned on. When the line power switch stands
out and the green light is off, the module is turned off.
The pump was shipped with default configuration settings. To change these
NOTE settings, see “Setting the 6-bit Configuration Switch” on page 317.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 61
3 Installing the Pump
Connecting Modules and Control Software
Connecting Modules
1 Place the individual modules in a stack configuration as shown in “Optimizing
the Stack Configuration” on page 51.
2 Ensure the power switches on the front of the modules are OFF (switches
stand out).
3 Plug a CAN cable into the CAN connector at the rear of the respective module
(except vacuum degasser).
4 Connect the CAN cable to the CAN connector of the next module.
5 Press in the power switches to turn on the modules.
1 Ensure the power switches on the front of the modules in the stack are OFF
(switches stand out).
2 Connect the LAN enabled module with a LAN cable to the data system.
3 Plug the CAN connector of the Instant Pilot into any available CAN port of the
system.
4 Connect the CAN cable to the CAN connector of one of the modules.
5 Press in the power switches to turn on the modules.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 62
3 Installing the Pump
Flow Connections of the Pump
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 63
3 Installing the Pump
Priming the System
Initial Priming
When Before a degasser or solvent tubing can be used, it is necessary to prime the system. Isopropanol is
recommended as priming solvent due to its miscibility with nearly all HPLC solvents and its excellent
wetting properties.
Preparations Connect all modules hydraulically as described in the respective module manuals.
Fill each solvent bottle with 100 mL isopropanol
Switch the system on
The purge tool of the LabAdvisor or Instrument Utilities can be used to purge the
NOTE pump automatically.
If the pump is not able to aspirate the solvent from the bottles, use a syringe to
NOTE draw the solvent manually through tubing and degasser.
When priming the vacuum degasser with a syringe, the solvent is drawn through
NOTE the degasser tubes very quickly. The solvent at the degasser outlet will therefore
not be fully degassed. Pump for approximately 10 minutes at your desired flow
rate before starting an analysis. This will allow the vacuum degasser to properly
degas the solvent in the degasser tubes.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 64
3 Installing the Pump
Priming the System
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 65
3 Installing the Pump
Priming the System
Regular Priming
When When the pumping system has been turned off for a certain time (for example, overnight) air will
rediffuse into the solvent channel between the vacuum degasser and the pump. If solvents
containing volatile components are left in the degasser without flow for a prolonged period, there will
be a slight loss of the volatile components.
The purge tool of the Lab Advisor can be used for purging the pump.
NOTE
1 Open the purge valve of your pump by turning it counterclockwise and set the
flow rate to 5 mL/min.
2 Flush the vacuum degasser and all tubes with at least 10 mL of solvent.
3 Repeat step 1 and 2 for the other channel(s) of the pump.
4 Set the required composition and flow rate for your application and close the
purge valve.
5 Pump for approximately 10 minutes before starting your application.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 66
3 Installing the Pump
Priming the System
Changing Solvents
When When the solvent of a channel is to be replaced by another solvent that is not compatible (solvents
are immiscible or one solvent contains a buffer), it is necessary to follow the procedure below to
prevent clogging of the pump by salt precipitation or residual liquid droplets in parts of the system.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 67
3 Installing the Pump
Priming the System
To clean the system when using HPLC grade water Best solvent to re-dissolve
buffers buffer crystals
After changing aqueous solvents HPLC grade water Best solvent to re-dissolve
buffer crystals
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 68
4 Using the Pump
Magnets 70
Turn on/off 71
Status Indicators 73
Best Practices 74
Daily / Weekly tasks 74
Power up / Shut-down the pump 74
Prepare the pump 75
How to deal with solvents 75
Select channels for Multi-Channel Gradient Valve (MCGV) 75
Leak and Waste Handling 76
Waste Concept 77
Hints for Successful Use of the Pump 78
Setting up the Pump with the G4208A Instant Pilot 79
Setting up the Pump with the Instrument Control Interface 80
Overview 80
Instrument Configuration 80
The Pump User Interface (Dashboard Panel) 82
Control Settings 84
Method Parameter Settings 85
Solvent Information 88
Materials in Flow Path 88
Bio-inert Materials 90
General Information about Solvent/Material Compatibility 92
Algae Growth in HPLC Systems 98
How to Prevent and/or Reduce the Algae Problem 98
Prevent Blocking of Solvent Filters 99
Normal Phase Applications 100
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 69
4 Using the Pump
Magnets
Magnets
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 70
4 Using the Pump
Turn on/off
Turn on/off
1 2
Power switch: On
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 71
4 Using the Pump
Turn on/off
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 72
4 Using the Pump
Status Indicators
Status Indicators
Status indicators
1. Idle
2. Run mode
3. Not-ready. Waiting for a specific pre-run condition to be reached or
completed.
4. Error mode - interrupts the analysis and requires attention (for example, a
leak or defective internal components).
5. Resident mode (blinking) - for example, during update of main firmware.
6. Bootloader mode (fast blinking). Try to re-boot the module or try a
cold-start. Then try a firmware update.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 73
4 Using the Pump
Best Practices
Best Practices
Weekly Tasks
• Change seal wash solvent (10 % isopropanol in water) and bottle.
• If applications with salts were used, flush all channels with water and remove
possible salt deposits manually.
• Inspect solvent filters for dirt or blockages. Exchange if no flow is coming out
of the solvent line when removed from the degasser inlet.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 74
4 Using the Pump
Best Practices
Seal wash
Seal Wash guarantees a maximum seal life time. Use Seal Wash:
• When using buffers with elevated salt concentrations
• When using volatile solvents with non-volatile additives
Contaminated seal wash solvent
C AU T I O N
Do not recycle seal wash solvent to avoid contamination.
Weekly exchange seal wash solvent.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 75
4 Using the Pump
Leak and Waste Handling
For details, see the usage guideline for the solvent cabinet. A printed copy of the
NOTE guideline has been shipped with the solvent cabinet, electronic copies are
available in the Agilent Information Center or via the Internet.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 76
4 Using the Pump
Leak and Waste Handling
Waste Concept
1 Agilent recommends using the 6 L waste can with 1 Stay Safe cap GL45 with
4 ports (5043-1221) for optimal and safe waste disposal. If you decide to use
your own waste solution, make sure that the tubes don't immerse in the liquid.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 77
4 Using the Pump
Hints for Successful Use of the Pump
• Always place the solvent cabinet with the solvent bottles on top of the
quaternary pump (or at a higher level).
• When using salt solutions and organic solvents in the quaternary pump it is
recommended to connect the salt solution to one of the bottom gradient
valve ports and the organic solvent to one of the upper gradient valve ports. It
is best to have the organic channel directly above the salt solution channel.
Regular flushing of all MCGV channels with water is recommended to remove
all possible salt deposits in the valve ports.
• Before operating the quaternary pump, flush the pump and vacuum degasser,
see “Regular Priming” on page 66. This is especially recommended if it has
been turned off for some time (for example, overnight) and volatile solvent
mixtures are used in the channels.
• Prevent blocking of solvent inlet filters. Never use the pump without solvent
inlet filter. Prevent the growth of algae, see “Prevent Blocking of Solvent
Filters” on page 99).
• Regularly check the purge valve frit and column frit. A blocked purge valve frit
can be identified by a black or yellow surface, deposits or by a pressure
greater than 10 bar, when pumping distilled water at a rate of 5 mL/min with
an open purge valve.
• When using the quaternary pump at low flow rates (for example, 0.2 mL/min)
check all 1/16-inch fittings for any signs of leaks.
• Whenever exchanging the pump seals the purge valve frit should be
exchanged, too.
• When using buffers or other salt solutions, flush the system with water before
switching it off. The seal wash option should be used when salt
concentrations of 0.1 M or higher will be used for long time periods.
• Check the pump pistons for scratches when changing the piston seals.
Scratched pistons will cause micro leaks and will decrease the lifetime of the
seal.
• Pressurize the system according to the wear in procedure after changing the
piston seals (see “Maintenance of a Pump Head Without Seal Wash
Option” on page 168).
• Consider recommendations given in the solvent information section, see
“Solvent Information” on page 88.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 78
4 Using the Pump
Setting up the Pump with the G4208A Instant Pilot
Generic operation of the G4208A Instant Pilot is covered in the Agilent Instant
Pilot G4208A User's Guide (G4208-90006). Details about setting up module
specific parameters can be found in the Instant Pilot online help.
The pump parameters are described in detail in “Overview” on page 80.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 79
4 Using the Pump
Setting up the Pump with the Instrument Control Interface
Overview
The instrument control interface offers the parameters described in the following
sections, and can usually be accessed through Agilent instrument control
software. For details, please see manuals and online help of respective user
interfaces.
Instrument Configuration
Use the Instrument Configuration dialog box to examine and, if necessary, modify
your instrument configuration. The Configurable Modules panel contains a list of
all modules available for configuration. The Selected Modules panel contains the
list of configured modules.
Auto Configuration: Under Communication settings, select either the Host Name
option or the IP address option and enter the appropriate value for the host
computer to enable automatic detection of the hardware configuration. The
system configures the instrument automatically with no further manual
configuration necessary.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 80
4 Using the Pump
Setting up the Pump with the Instrument Control Interface
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 81
4 Using the Pump
Setting up the Pump with the Instrument Control Interface
The current solvent flow rate (in mL/min) is displayed above the pressure display.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 82
4 Using the Pump
Setting up the Pump with the Instrument Control Interface
Instrument Signals
The following pump signals are displayed:
Table 10 Pump signals (isocratic pump)
Pressure The current pump pressure (in bar, psi or MPa, see “Instrument
Configuration” on page 80).
Pressure The current pump pressure (in bar, psi or MPa, see “Instrument
Configuration” on page 80).
Composition A:B The contributions of channels A and B to the current solvent composition.
Composition C:D The contributions of channels C and D to the current solvent composition.
Context Menu
The context menu of the dashboard panel contains the following commands:
Set Error Method Sets the method that is loaded if an error occurs to the method
that is currently available in the hardware.
Identify Device Causes the LED on the front of the module to blink for a few
seconds.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 83
4 Using the Pump
Setting up the Pump with the Instrument Control Interface
Control Settings
The pump control parameters are in three sections:
• Pump
• Seal Wash
• Automatic Turn On
Table 12 Pump control parameters
Pump Enables you to switch the pump On, Off or to a Standby condition.
In the Standby condition, the pump motor is still active, and when the pump
is switched on again, does not need to be re-initialized.
Seal Wash The seal wash can be set up to be run once ( Single wash) or periodically (
Periodic).
• Off: no seal wash is used.
• Single wash: the seal will be purged for a specified time.
• Periodic: a periodic wash will be applied for a defined period in minutes.
The option is available only when the pump has seal wash capability. The
seal wash capability is detected by the module itself. If seal wash is
installed, it is recommended to use it in order to increase the primary seal
lifetime.
Automatic Turn On Module can be turned on at a specified date/time. This feature can only be
used if the module power switch is turned on.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 84
4 Using the Pump
Setting up the Pump with the Instrument Control Interface
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 85
4 Using the Pump
Setting up the Pump with the Instrument Control Interface
Advanced Settings
The pump advanced method setup parameters are in four sections:
• Minimum Stroke
• Compressibility
• Maximum Flow Gradient
• Primary Channel (Quaternary Pump only)
Table 14 Advanced method parameters
Minimum Stroke 20 – 100 µL The Stroke Volume is used for optimizing between performance of the
module and seal life time. For performance a low stroke volume is
beneficial, as it divides disturbances into smaller packages, but a larger
volume is extending the life time of the pump seals.
If Automatic is activated, the pump tries to achieve an optimized stroke
volume for best performance and lowest ripple.
Compressibility 40·10-6 – 150·10-6. The compressibility of the mobile phase has an effect on the performance
Default compressibility of the pump. For best flow accuracy and mixing performance, you can set
value: 100·10-6 the parameter according to the mobile phase being used.
Maximum Flow 0.1 – 100 mL/min/min You can set a limit on the rate of change of the solvent flow to protect your
Gradient Default value: analytical column.
100 mL/min/min
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 86
4 Using the Pump
Setting up the Pump with the Instrument Control Interface
Timetable Settings
Use the Timetable to program changes in the pump parameters during the
analysis by entering a time in the Time field and appropriate values in the
following fields of the timetable. Changes in flow rate occur linearly from either
time zero or the time of the last defined change; other parameters change
instantaneously at the time defined in the timetable.
Show Advanced Timetable toggles the timetable display between standard mode
and advanced mode.
The following parameters can be changed:
• Change Contacts
• Change Flow
• Change Max. Pressure Limit
• Change Solvent Composition - You can only use solvents, which have been
enabled in the solvents section.
• Function centric view - This checkbox allows you displaying parameter
changes instead of a time table.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 87
4 Using the Pump
Solvent Information
Solvent Information
Part Materials
Degasser chamber TFE/PDD Copolymer, PFA (internal tubings), PEEK (inlets), FEP
(tubings), ETFE (fittings)
Pistons Sapphire
Piston seals/wash seals PTFE, SST (reversed phase) or UHMW-PE, SST (normal phase)
Capillaries/fittings SST
Tubings PTFE
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 88
4 Using the Pump
Solvent Information
Part Materials
Degasser chamber TFE/PDD Copolymer, PFA (internal tubings), PEEK (inlets), FEP
(tubings), ETFE (fittings)
Active inlet valve Platinum-iridium, titanium, gold, sapphire, ruby, ceramic, PTFE
Pistons Sapphire
Capillaries/fittings Titanium
Tubings PTFE
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 89
4 Using the Pump
Solvent Information
Bio-inert Materials
For the Bio-inert LC system, Agilent Technologies uses highest-quality materials
in the flow path (also referred to as wetted parts), which are widely accepted by
life science scientists, as they are known for optimum inertness to biological
samples and ensure best compatibility with common samples and solvents over
a wide pH range. Explicitly, the complete flow path is free of stainless steel and
free of other alloys containing metals such as iron, nickel, cobalt, chromium,
molybdenum, or copper, which can interfere with biological samples. The flow
downstream of the sample introduction contains no metals whatsoever.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 90
4 Using the Pump
Solvent Information
Module Materials
Agilent 1260 Infinity II Bio-inert Analytical Fraction Collector PEEK, ceramic, PTFE
(G5664B)
Standard flow cell bio-inert, 10 mm, 13 µL, 120 bar ( 12 MPa) for MWD/DAD, PEEK, ceramic, sapphire, PTFE
includes 0890-1763 – 0.18 x 1500 mm PEEK capillary and 5063-6591 – PEEK
fittings (G5615-60022)
(for Agilent 1260 Infinity II DAD G7115A, and MWD G7165A)
Bio-inert flow cell, 8 µL, 20 bar (pH 1–12), includes and PEEK fittings PEEK, fused silica, PTFE
(G5615-60005)
(for Agilent 1260 Infinity II FLD G7121A/B)
Bio-inert Valve heads (G4235A, G5631A, G5632A, G5639A) PEEK, ceramic (Al2O3 based)
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 91
4 Using the Pump
Solvent Information
Disclaimer
Subsequent data was collected from external resources and is meant as a
reference. Agilent cannot guarantee the correctness and completeness of such
information. Data is based on compatibility libraries, which are not specific for
estimating the long-term life time under specific but highly variable conditions of
UHPLC systems, solvents, solvent mixtures, and samples. Information also
cannot be generalized due to catalytic effects of impurities like metal ions,
complexing agents, oxygen etc. Apart from pure chemical corrosion, other
effects like electro corrosion, electrostatic charging (especially for nonconductive
organic solvents), swelling of polymer parts etc. need to be considered. Most
data available refers to room temperature (typically 20 – 25 °C, 68 – 77 °F). If
corrosion is possible, it usually accelerates at higher temperatures. If in doubt,
please consult technical literature on chemical compatibility of materials.
MP35N
MP35N is a nonmagnetic, nickel-cobalt-chromium-molybdenum alloy
demonstrating excellent corrosion resistance (for example, against nitric and
sulfuric acids, sodium hydroxide, and seawater) over a wide range of
concentrations and temperatures. In addition, this alloy shows exceptional
resistance to high-temperature oxidation. Due to excellent chemical resistance
and toughness, the alloy is used in diverse applications: dental products, medical
devices, nonmagnetic electrical components, chemical and food processing
equipment, marine equipment. Treatment of MP35N alloy samples with 10 %
NaCl in HCl (pH 2.0) does not reveal any detectable corrosion. MP35N also
demonstrates excellent corrosion resistance in a humid environment. Although
the influence of a broad variety of solvents and conditions has been tested, users
should keep in mind that multiple factors can affect corrosion rates, such as
temperature, concentration, pH, impurities, stress, surface finish, and dissimilar
metal contacts.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 92
4 Using the Pump
Solvent Information
PEEK
PEEK (Polyether-Ether Ketones) combines excellent properties regarding
biocompatibility, chemical resistance, mechanical and thermal stability. PEEK is
therefore the material of choice for UHPLC and biochemical instrumentation.
It is stable in the specified pH range (for the Bio-Inert LC system: pH 1 – 13, see
bio-inert module manuals for details), and inert to many common solvents.
There are still some known incompatibilities with chemicals such as chloroform,
methylene chloride, THF, DMSO, strong acids (nitric acid > 10 %, sulfuric acid >
10 %, sulfonic acids, trichloroacetic acid), halogens or aqueous halogen
solutions, phenol and derivatives (cresols, salicylic acid, and so on).
When used above room temperature, PEEK is sensitive to bases and various
organic solvents, which can cause it to swell. Under such conditions, normal
PEEK capillaries are sensitive to high pressure. Therefore, Agilent uses stainless
steel clad PEEK capillaries in bio-inert systems. The use of stainless steel clad
PEEK capillaries keeps the flow path free of steel and ensures pressure stability
up to 600 bar. If in doubt, consult the available literature about the chemical
compatibility of PEEK.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 93
4 Using the Pump
Solvent Information
Polyimide
Agilent uses semi-crystalline polyimide for rotor seals in valves and needle seats
in autosamplers. One supplier of polyimide is DuPont, which brands polyimide as
Vespel, which is also used by Agilent.
Polyimide is stable in a pH range between 1 and 10 and in most organic solvents.
It is incompatible with concentrated mineral acids (e.g. sulphuric acid), glacial
acetic acid, DMSO and THF. It is also degraded by nucleophilic substances like
ammonia (e.g. ammonium salts in basic conditions) or acetates.
Polyethylene (PE)
Agilent uses UHMW (ultra-high molecular weight)-PE/PTFE blends for yellow
piston and wash seals, which are used in 1290 Infinity pumps, 1290 Infinity II
pumps, the G7104C and for normal phase applications in 1260 Infinity pumps.
Polyethylene has a good stability for most common inorganic solvents including
acids and bases in a pH range of 1 to 12.5. It is compatible with many organic
solvents used in chromatographic systems like methanol, acetonitrile and
isopropanol. It has limited stability with aliphatic, aromatic and halogenated
hydrocarbons, THF, phenol and derivatives, concentrated acids and bases. For
normal phase applications, the maximum pressure should be limited to 200 bar.
Tantalum (Ta)
Tantalum is inert to most common HPLC solvents and almost all acids except
fluoric acid and acids with free sulfur trioxide. It can be corroded by strong bases
(e.g. hydroxide solutions > 10 %, diethylamine). It is not recommended for the use
with fluoric acid and fluorides.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 94
4 Using the Pump
Solvent Information
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 95
4 Using the Pump
Solvent Information
Titanium (Ti)
Titanium is highly resistant to oxidizing acids (for example, nitric, perchloric and
hypochlorous acid) over a wide range of concentrations and temperatures. This
is due to a thin oxide layer on the surface, which is stabilized by oxidizing
compounds. Non-oxidizing acids (for example, hydrochloric, sulfuric and
phosphoric acid) can cause slight corrosion, which increases with acid
concentration and temperature. For example, the corrosion rate with 3 % HCl
(about pH 0.1) at room temperature is about 13 μm/year. At room temperature,
titanium is resistant to concentrations of about 5 % sulfuric acid (about pH 0.3).
Addition of nitric acid to hydrochloric or sulfuric acids significantly reduces
corrosion rates. Titanium is sensitive to acidic metal chlorides like FeCl3 or CuCl2.
Titanium is subject to corrosion in anhydrous methanol, which can be avoided by
adding a small amount of water (about 3 %). Slight corrosion is possible with
ammonia > 10 %.
Gold
Gold is inert to all common HPLC solvents, acids, and bases within the specified
pH range. It can be corroded by complexing cyanides and concentrated acids like
aqua regia.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 96
4 Using the Pump
Solvent Information
Platinum/Iridium
Platinum/Iridium is inert to almost all common acids, bases, and solvents. There
are no documented incompatibilities for HPLC applications.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 97
4 Using the Pump
Algae Growth in HPLC Systems
The presence of algae in HPLC systems can cause many problems that may be
incorrectly diagnosed as instrument or application problems. Algae grow in
aqueous media, preferably in a pH range from 4 to 8. Their growth is accelerated
by buffers, for example phosphate or acetate. Since algae grow through
photosynthesis, light will also stimulate their growth. Even in distilled water
small-sized algae grow after some time.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 98
4 Using the Pump
Prevent Blocking of Solvent Filters
Contaminated solvents or algae growth in the solvent bottle will reduce the
lifetime of the solvent filter and will influence the performance of the module.
This is especially true for aqueous solvents or phosphate buffers (pH 4 to 7). The
following suggestions will prolong lifetime of the solvent filter and will maintain
the performance of the module.
• Use a sterile, if possible amber, solvent bottle to slow down algae growth.
• Filter solvents through filters or membranes that remove algae.
• Exchange solvents every two days or refilter.
• If the application permits add 0.0001 – 0.001 M sodium azide to the solvent.
• Place a layer of argon on top of your solvent.
• Avoid exposure of the solvent bottle to direct sunlight.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 99
4 Using the Pump
Normal Phase Applications
Current passive inlet valves and outlet ball valves used with 1260 and 1290
Infinity pumps do not work well with applications using non-polar solvents as for
normal phase applications (e.g. hexane, heptane and CO2). With such
applications, pressure drops could be observed. They are a result of particles
electrostatically charging up in insulating solvents and sticking to the balls inside
the valves, such that the valves do not close properly any more after some time
of use (can be hours).
For normal phase applications, a second type of valves is available, which has a
design based on the existing one for 1260 and 1290 Infinity valves. These valves
use a new material for valve balls, which is a conductive ceramic and replaces
non-conductive ruby balls. The balls do not charge up electrostatically and show
good performance in normal phase.
The valves are marked with N for non-polar or normal phase.
Agilent recommends using these valves for (and only for) normal phase
applications.
Corrosion of valves
C AU T I O N
Normal phase balls/valves corrode quickly in aqueous solutions and acids (at
or below pH 7).
Do not use normal phase valves in applications running with aqueous
solutions.
The N-Valves have been tested successfully in using hexane at pressures below
100 bar; heptane can be used as a substitute for neurotoxic hexane.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 100
4 Using the Pump
Normal Phase Applications
Inlet valves 1260 Infinity Inlet Valve Type N 1290 Infinity Inlet Valve Type N
(G1312-60166) (G4220-60122)
1290 Infinity Quat Inlet Valve
Type N (G4204-60122)
Outlet valves Outlet Valve Type N/SFC 1290 Infinity Outlet Valve Type N
(G1312-60167) (G4220-60128)
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 101
5 Optimizing Performance
This chapter gives hints on how to optimize the performance or use additional
devices.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 102
5 Optimizing Performance
Using the Degasser
The quaternary pump has a built-in degasser, which should always be included to
the flow path.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 103
5 Optimizing Performance
Operational Hints for the Vacuum Degasser
If you are using the vacuum degasser for the first time, if the vacuum degasser
was switched off for any length of time (for example, overnight), or if the vacuum
degasser chambers are empty, you have to prime the vacuum degasser before
running an analysis. Priming is usually done by pumping at a high flow rate
(3 – 5 mL/min). Alternatively, a syringe can be used to draw the solvent through
the (empty) degasser if the pump does not aspirate the solvent by itself.
For details see “Initial Priming” on page 64.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 104
5 Optimizing Performance
Operational Hints for the Multi Channel Gradient Valve (MCGV)
In a mixture of salt solutions and organic solvent the salt solution might be well
dissolved in the organic solvent without showing precipitations. However in the
mixing point of the gradient valve, at the boundary between the two solvents,
micro precipitation is possible. Gravity forces the salt particles to fall down.
Normally the A channel of the valve is used for the aqueous/salt solution and the
B channel of the pump is used for the organic solvent. If used in this
configuration the salt will fall back into the aqueous solution and will be
dissolved. When using the pump in a different configuration (e.g., D - salt solution,
A - organic solvent) the salt can fall into the port of the organic solvent and may
lead to performance problems.
Precipitations formed during the mixing of buffers and organic solvents which do
NOTE not dissolve salts may cause a loss of pump performance (flow/retention time
stability), a blockage or internal leak of the pump. Avoid the use of such solvent
combinations, as they can cause irreproducible chromatographic results.
The use of an Inline filter (G1311-60006) can avoid or reduce such effects by
filtering crystals and re-dissolving them over time.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 105
5 Optimizing Performance
When to use the Seal Wash Function
Highly concentrated buffer solutions will reduce the lifetime of the seals and
pistons in your pump. The seal wash function allows to maintain the seal lifetime
by flushing the back side of the seal with a wash solvent.
The seal wash function is strongly recommended when buffer concentrations of
0.1 M or higher will be used for long time periods in the pump.
Active seal wash can be ordered as factory installed option or can be upgraded
later by ordering the field upgrade kit G1398A.
The seal wash function comprises a support ring, secondary seal, gasket and
seal holder for both piston sides. Place a wash bottle filled with 90 % water / 10 %
isopropanol above the pump in the solvent cabinet and guide the SW waste into
the waste bottle underneath the LC stack. The peristaltic pump moves a flow
through the pump head removing all possible buffer crystals from the back of the
pump seal. This mixture prevents growth of algae or bacteria in the wash bottle
and reduces the surface tension of the water.
For information on the installation of the active seal wash function refer to “Install
an Active Seal Wash Upgrade Kit (Quaternary Pump)” on page 241.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 106
5 Optimizing Performance
Choosing the Right Pump Seals
The standard seal for the pump can be used for most applications. However
applications that use normal phase solvents (for example, hexane) are not suited
for the standard seal and require a different seal when used for a longer time in
the pump.
For applications that use normal phase solvents (for example, hexane) we
recommend using polyethylene pump seals (PE seals (pack of 2) (0905-1420))
and Wash Seal PE (0905-1718). For normal phase applications, these seals have
less abrasion compared to the standard seals.
Polyethylene seals have a limited pressure range of 0 – 200 bar. When used
NOTE above 200 bar their lifetime is reduced significantly.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 107
5 Optimizing Performance
Optimize the Compressibility Compensation Setting
The compressibility compensation default setting is 100 × 10-6 /bar for the
pump. This setting represents an average value. Under normal conditions the
default setting typically reduces the pressure pulsation to values below 1 % of
system pressure that are sufficient for most applications and for all gradient
analyses. The compressibility settings can be optimized by using the values for
the various solvents described in Table 19 on page 109. If the solvent in use is
not listed in the compressibility tables, when using isocratic mixtures of solvents
and if the default settings are not sufficient for your application the following
procedure can be used to optimize the compressibility settings.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 108
5 Optimizing Performance
Optimize the Compressibility Compensation Setting
Acetone 126
Acetonitrile 115
Benzene 95
Chloroform1 100
Cyclohexane 118
Ethanol 114
Heptane 120
Hexane 150
Isobutanol 100
Isopropanol 100
Methanol 120
1-Propanol 100
Toluene 87
Water 46
1 Please check section "Solvent Information" for compatibility to your specific LC system.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 109
6 Troubleshooting and Diagnostics
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 110
6 Troubleshooting and Diagnostics
User Interfaces
User Interfaces
Depending on the user interface, the available tests vary. Some descriptions are
only available in the service manual.
Table 20 Test functions available vs. user interface
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 111
6 Troubleshooting and Diagnostics
Agilent Lab Advisor Software
The Agilent Lab Advisor Software (basic license, shipped with an Agilent LC
pump) is a standalone product that can be used with or without a
chromatographic data system. Agilent Lab Advisor helps to manage the lab for
high-quality chromatographic results by providing a detailed system overview of
all connected analytical instruments with instrument status, Early Maintenance
Feedback counters (EMF), instrument configuration information, and diagnostic
tests. With the push of a button, a detailed diagnostic report can be generated.
Upon request, the user can send this report to Agilent for a significantly improved
troubleshooting and repair process.
The Agilent Lab Advisor software is available in two versions:
• Lab Advisor Basic
• Lab Advisor Advanced
Lab Advisor Basic is included with every Agilent 1200 Infinity Series and Agilent
InfinityLab LC Series instrument.
The Lab Advisor Advanced features can be unlocked by purchasing a license key,
and include real-time monitoring of instrument actuals, all various instrument
signals, and state machines. In addition, all diagnostic test results, calibration
results, and acquired signal data can be uploaded to a shared network folder. The
Review Client included in Lab Advisor Advanced allows to load and examine the
uploaded data no matter on which instrument it was generated. This makes Data
Sharing an ideal tool for internal support groups and users who want to track the
instrument history of their analytical systems.
The optional Agilent Maintenance Wizard Add-on provides an easy-to-use,
step-by-step multimedia guide for performing preventive maintenance on Agilent
1200 Infinity LC Series instrument.
The tests and diagnostic features that are provided by the Agilent Lab Advisor
software may differ from the descriptions in this manual. For details, refer to the
Agilent Lab Advisor software help files.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 112
7 Error Information
This chapter describes the meaning of error messages, and provides information
on probable causes and suggested actions how to recover from error conditions.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 113
7 Error Information
What Are Error Messages
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 114
7 Error Information
General Error Messages
General error messages are generic to all Agilent series HPLC modules and may
show up on other modules as well.
Timeout
Error ID: 0062
The timeout threshold was exceeded.
1 The analysis was completed successfully, Check the logbook for the occurrence and source
and the timeout function switched off the of a not-ready condition. Restart the analysis
module as requested. where required.
2 A not-ready condition was present during a Check the logbook for the occurrence and source
sequence or multiple-injection run for a of a not-ready condition. Restart the analysis
period longer than the timeout threshold. where required.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 115
7 Error Information
General Error Messages
Shutdown
Error ID: 0063
An external instrument has generated a shutdown signal on the remote line.
The module continually monitors the remote input connectors for status signals.
A LOW signal input on pin 6 of the Enhanced Remote Interface (ERI) connector
generates the error message.
1 Leak detected in another module with a CAN Fix the leak in the external instrument before
connection to the system. restarting the module.
2 Leak detected in an external instrument with Fix the leak in the external instrument before
a remote connection to the system. restarting the module.
4 The degasser failed to generate sufficient Check the vacuum degasser for an error
vacuum for solvent degassing. condition. Refer to the Service Manual for the
degasser or the pump that has the degasser
built-in.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 116
7 Error Information
General Error Messages
Remote Timeout
Error ID: 0070
A not-ready condition is still present on the remote input. When an analysis is
started, the system expects all not-ready conditions (for example, a not-ready
condition during detector balance) to switch to run conditions within one minute
of starting the analysis. If a not-ready condition is still present on the remote line
after one minute the error message is generated.
1 Not-ready condition in one of the Ensure the instrument showing the not-ready
instruments connected to the remote line. condition is installed correctly, and is set up
correctly for analysis.
3 Defective components in the instrument Check the instrument for defects (refer to the
showing the not-ready condition. instrument’s documentation).
1 CAN cable disconnected. • Ensure all the CAN cables are connected
correctly.
• Ensure all CAN cables are installed correctly.
3 Defective mainboard in another module. Switch off the system. Restart the system, and
determine which module or modules are not
recognized by the system.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 117
7 Error Information
General Error Messages
Leak
Error ID: 0064
A leak was detected in the module.
The signals from the two temperature sensors (leak sensor and board-mounted
temperature-compensation sensor) are used by the leak algorithm to determine
whether a leak is present. When a leak occurs, the leak sensor is cooled by the
solvent. This changes the resistance of the leak sensor which is sensed by the
leak sensor circuit on the mainboard.
3 Loose or leaking purge valve, inlet valve, or Ensure pump components are seated correctly.
outlet valve. If there are still signs of a leak, exchange the
appropriate seal (purge valve, inlet valve, outlet
valve).
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 118
7 Error Information
General Error Messages
1 Leak sensor not connected to the power Ensure the leak sensor is connected correctly.
switch board.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 119
7 Error Information
General Error Messages
2 Leak sensor incorrectly routed, being • Make sure the leak sensor is installed
pinched by a metal component. correctly.
• Correct the routing of the cable.
• If cable defective, exchange the leak sensor.
4 Cable or contact problem. • Inspect the flat ribbon cable from the power
switch board to the mainboard.
• Inspect the contacts on the flat ribbon
cable's ends.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 120
7 Error Information
General Error Messages
1 Loose connection between the power switch Check the ribbon cable between the power
board and the mainboard. switch board and the mainboard.
2 Loose connection between the power switch Check the ribbon cable between the power
board and the mainboard. switch board and the mainboard.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 121
7 Error Information
General Error Messages
Fan Failed
Error ID: 0068
The cooling fan in the module has failed.
The hall sensor on the fan shaft is used by the mainboard to monitor the fan
speed. If the fan speed falls below a certain limit for a certain length of time, the
error message is generated.
This limit is given by 2 revolutions/second for longer than 5 seconds.
Depending on the module, assemblies (e.g. the lamp in the detector) are turned
off to assure that the module does not overheat inside.
4 Improperly positioned cables or wires Ensure the fan is not mechanically blocked.
obstructing fan blades.
ERI Messages
Error ID: 11120 (+5 V), 11121 (+25 V)
The ERI (Enhanced Remote Interface) provides two error events related to over
current situations on the +5 V and +24 V lines.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 122
7 Error Information
Module Error Messages
1 Volume in bottle below specified volume. Refill bottles and reset solvent counters.
1 Upper pressure limit set too low. Ensure the upper pressure limit is set to a value
suitable for the analysis.
2 Blockage in the flowpath (after the damper). Check for blockage in the flow path. The
following components are particularly subject to
blockage: inline filter frit, needle (autosampler),
seat capillary (autosampler), sample loop
(autosampler), column frits and capillaries with
small internal diameters (e.g. 50 µm ID).
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 123
7 Error Information
Module Error Messages
1 Lower pressure limit set too high. Ensure the lower pressure limit is set to a value
suitable for the analysis.
2 Air bubbles in the mobile phase. • Make sure that the degasser is in flow path
and works correctly. Purge the module.
• Ensure solvent inlet filters are not blocked.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 124
7 Error Information
Module Error Messages
MCGV Fuse
Error ID: 2043
Valve Fuse 0: Channels A and B
Valve Fuse 1: Channels C and D
The gradient valve in the quaternary pump has drawn excessive current causing
the electronic fuse to open.
AIV Fuse
Error ID: 2044
The active-inlet valve in the module has drawn excessive current causing the
inlet-valve electronic fuse to open.
1 Defective active inlet valve. Restart the module. If the error message appears
again, exchange the active inlet valve.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 125
7 Error Information
Module Error Messages
2 Connection cable (inside instrument) not Ensure the connection cable is connected
connected. correctly.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 126
7 Error Information
Module Error Messages
Motor-Drive Power
Error ID: 2041, 2042
The current drawn by the pump motor exceeded the maximum limit.
Blockages in the flow path are usually detected by the pressure sensor in the
damper, which result in the pump switching off when the upper pressure limit is
exceeded. If a blockage occurs before the damper, the pressure increase cannot
be detected by the pressure sensor and the module will continue to pump. As
pressure increases, the pump drive draws more current. When the current
reaches the maximum limit, the module is switched off, and the error message is
generated.
1 Flow path blockage in front of the damper. Ensure the capillaries and frits between the
pump head and damper inlet are free from
blockage.
2 Blocked (passive or active) inlet valve. Exchange the (passive or active) inlet valve.
4 High friction (partial mechanical blockage) in Remove the pump-head assembly. Ensure there
the pump drive assembly. is no mechanical blockage of the pump-head
assembly or pump drive assembly.
5 Defective pump drive assembly. • Remove the pump head assembly. Ensure
there is no mechanical blockage of the pump
head assembly or pump drive assembly.
• Exchange the pump drive assembly.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 127
7 Error Information
Module Error Messages
1 Mechanical blockage of the module. Remove the pump-head assembly. Ensure there
is no mechanical blockage of the pump-head
assembly or pump drive assembly.
2 Disconnected or defective cable. Ensure the pump assembly cables are not
damaged or dirty. Make sure the cables are
connected securely to the mainboard.
3 Blocked (passive or active) inlet valve. Exchange the (passive or active) inlet valve.
4 Defective pump drive assembly. • Remove the pump head assembly. Ensure
there is no mechanical blockage of the pump
head assembly or pump drive assembly.
• Exchange the pump drive assembly.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 128
7 Error Information
Module Error Messages
1 Pump head not installed correctly (screws Install the pump head correctly. Ensure nothing
not secured, or pump head not seated (e.g. capillary) is trapped between the pump head
correctly). and body.
Index Limit
Error ID: 2203, 2213
The time required by the piston to reach the encoder index position was too short
(pump).
During initialization, the first piston is moved to the mechanical stop. After
reaching the mechanical stop, the piston reverses direction until the encoder
index position is reached. If the index position is reached too fast, the error
message is generated.
1 Irregular or sticking drive movement. Remove the pump head, and examine the seals,
pistons, and internal components for signs of
wear, contamination or damage. Exchange
components as required.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 129
7 Error Information
Module Error Messages
Index Adjustment
Error ID: 2204, 2214
The encoder index position in the module is out of adjustment.
During initialization, the first piston is moved to the mechanical stop. After
reaching the mechanical stop, the piston reverses direction until the encoder
index position is reached. If the time to reach the index position is too long, the
error message is generated.
1 Irregular or sticking drive movement. Remove the pump head, and examine the seals,
pistons, and internal components for signs of
wear, contamination or damage. Exchange
components as required.
Index Missing
Error ID: 2205, 2215, 2505
The encoder index position in the module was not found during initialization.
During initialization, the first piston is moved to the mechanical stop. After
reaching the mechanical stop, the piston reverses direction until the encoder
index position is reached. If the index position is not recognized within a defined
time, the error message is generated.
1 Disconnected or defective encoder cable. Ensure the encoder cables are not damaged or
dirty. Make sure the cables are connected
securely to the mainboard.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 130
7 Error Information
Module Error Messages
Initialization Failed
Error ID: 2207, 2217
The module failed to initialize successfully within the maximum time window.
A maximum time is assigned for the complete pump-initialization cycle. If the
time is exceeded before initialization is complete, the error message is
generated.
1 Blocked (passive or active) inlet valve. Exchange the (passive or active) inlet valve.
2 Defective pump drive assembly. • Remove the pump head assembly. Ensure
there is no mechanical blockage of the pump
head assembly or pump drive assembly.
• Exchange the pump drive assembly.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 131
7 Error Information
Module Error Messages
Wait Timeout
Error ID: 2053
When running certain tests in the diagnostics mode or other special applications,
the pump must wait for the pistons to reach a specific position, or must wait for a
certain pressure or flow to be reached. Each action or state must be completed
within the timeout period, otherwise the error message is generated.
Possible Reasons for a Wait Timeout:
· Pressure not reached.
· Pump channel A did not reach the delivery phase.
· Pump channel B did not reach the delivery phase.
· Pump channel A did not reach the take-in phase.
· Pump channel B did not reach the take-in phase.
· Solvent volume not delivered within the specified time.
2 Leak at fittings, purge valve, active inlet valve, • Ensure pump components are seated
outlet valve or piston seals. correctly. If there are still signs of a leak,
exchange the appropriate seal (purge valve,
active inlet valve, outlet valve, piston seal).
• Exchange defective capillaries.
3 Flow changed after starting test. Ensure correct operating condition for the
special application in use.
4 Defective pump drive assembly. Exchange the defective pump drive assembly.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 132
7 Error Information
Module Error Messages
2 Degasser sensor not connected to Connect the degasser sensor to the mainboard.
mainboard.
1 Liquid in degasser tubing. Wait for evaporation of liquid. Then replace the
tubing, degasser chamber, and vacuum pump.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 133
7 Error Information
Module Error Messages
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 134
8 Test Functions and Calibration
Introduction 136
System Pressure Test 137
Running the Test 139
Evaluating the Results 140
Potential Causes of System Pressure Test Failure 141
Leak Rate Test 142
Running the Test 144
Evaluating the Results 145
Potential Causes of Leak Rate Test Failure 146
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 135
8 Test Functions and Calibration
Introduction
Introduction
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 136
8 Test Functions and Calibration
System Pressure Test
Introduction
The System Pressure Test is used for checking the tightness of the LC system
and identifying leaks between the pump and a position in the flow path following
the pump blocked by a blank nut.
System requirements
Minimum software revisions:
• Lab Advisor B.02.08
Minimum firmware revision: D.07.01.
Test Principle
A solvent can be chosen from available solvent channels and a maximum
pressure can be defined at which the test will be run. In contrast to older
revisions of this test, any solvent can be used.
Before the test, the pump and system are flushed with solvent in order to remove
air bubbles, as air bubbles are compressed during the test and therefore would
appear as leaks. Using a degasser is highly recommended. Then the flow path is
blocked by a blank nut at any position between the purge valve and the column
outlet.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 137
8 Test Functions and Calibration
System Pressure Test
pressure
2
1
time
flow
leak rate
time
In the first phase of the test, the pump delivers flow at a rate of 200 µL/min until a
pressure of 50 bar below the defined maximum pressure is reached. In the
second phase, the pump delivers a small flow which is increased stepwise. If
there is a leak in the system, the pressure will drop initially, as the low flow cannot
compensate the leak flow. As soon as the pump flow rate exceeds the leak flow
rate, the pressure will increase again and the test is stopped at about 20 bar
below the maximum pressure. The point in phase 2, where the lowest pressure is
reached and stays constant for a short time corresponds to the leak rate, that is
provided as a test result. A leak rate smaller than 3 µL/min is good enough for
operating the pump reliably.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 138
8 Test Functions and Calibration
System Pressure Test
Make sure to release the pressure by slowly opening the purge valve when the
NOTE test has finished.
“Evaluating the Results” on page 140 describes the evaluation and interpretation
of the System Pressure Test results. For detailed instructions refer to the Agilent
Lab Advisor software.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 139
8 Test Functions and Calibration
System Pressure Test
Often it is only a damaged blank nut itself (poorly shaped from overtightening)
NOTE that causes a failure of the test. Before investigating on any other possible
sources of failure make sure that the blank nut you are using is in good condition
and properly tightened!
If the test fails again, insert the blank nut at the outlet of the previous module in
the stack (e.g. autosampler outlet if TCC has been tested before), and repeat the
test. Exclude each module one by one to determine which module is leaking.
If the pump is determined to be the source of the leak, run the Pump Leak Rate
Test.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 140
8 Test Functions and Calibration
System Pressure Test
3 Pump: Damaged pump seals or pistons. Run the Leak Rate Test to confirm the leak.
4 Loose purge valve. Tighten the purge valve nut (14 mm wrench).
7 Autosampler: Damaged metering seal or Exchange the metering seal. Check the piston for
piston. scratches. Exchange the piston if required.
9 Column compartment: Loose or leaky fitting. Tighten or exchange the fitting or capillary.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 141
8 Test Functions and Calibration
Leak Rate Test
Introduction
The Leak Rate Test is used for verifying the internal tightness of the pump and
helps identifying parts which may have caused a leak.
System requirements
Minimum software revisions:
• Lab Advisor B.02.08
Minimum firmware revisions:
• D.07.01
Test Principle
A solvent can be chosen from available solvent channels and a maximum target
pressure can be defined at which the test will be run. Typically, this is the
maximum pressure specified for the pump. The test can be run with any solvent
compatible to the pump.
Before the test, the pump is flushed with solvent in order to remove air bubbles,
as air bubbles are compressed during the test and therefore would appear as
leaks. Using a degasser is highly recommended.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 142
8 Test Functions and Calibration
Leak Rate Test
piston 2
at rear position
piston 1
at rear position
piston 2
delivers
piston 1
delivers
Initially, the pressure is increased to about 100 bar below the target pressure,
which has been set for the test.
Then piston 1 is brought to its rear position. An increasing flow is delivered by
piston 1. In case of a leak, the pressure will drop initially as long as the flow rate
delivered by the piston is lower than the leak rate. As soon as the flow rate of the
piston exceeds the leak rate, the measured pressure will increase again.
Therefore the minimum pressure of that curve segment corresponds to the flow
and leak rate at that time and the leak rate is measured. Compare to the
description of the system pressure test ( “System Pressure Test” on page 137).
Subsequently, piston 2 is moved to its rear position, then piston 2 delivers and the
measurement is done as described for piston 1.
For a binary pump (G1312B/C, G7112B, K1312B, G4782A), the test is run for both
pump heads for channels A and B.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 143
8 Test Functions and Calibration
Leak Rate Test
Make sure to release the pressure by slowly opening the purge valve when the
NOTE test has finished.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 144
8 Test Functions and Calibration
Leak Rate Test
to column
Damper
Outlet valve
Purge valve
Seals
Piston 2
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 145
8 Test Functions and Calibration
Leak Rate Test
6 Leak at piston 2 or seal in chamber 2 Inspect piston, replace piston and/or seal
Primary Leak
If a leak is found for movement of piston 1 (primary leak), any leak described for
piston movement 2 will cause a failure for piston 1 as well, as the liquid can move
through the outlet valve to chamber 2. Such cases need to be identified as
described before. Additionally, following causes are possible:
1 Leak at piston 1 or seal in chamber 1 Inspect piston, replace piston and/or seal
2 Leak at inlet valve Replace inlet valve or inlet valve cartridge (AIV
only)
1 Leak at outlet valve Replace the part which has failed and re-run the
test.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 146
9 Maintenance and Repair
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 147
Changing the Type and Serial Number 203
Recover Instructions 205
Exchange the Damper 208
Exchange the Pump Drive 212
Exchange the Fan 214
Replace Power Supply and Cables 216
Replace the Power Switch 219
Replace the Degasser 221
Replace Parts Inside the 4-Channel-Degasser 223
Replace a Degasser Chamber 225
Replace the Degasser Pump 226
Replace the Sensor Board Assembly 227
Replace the Degasser Tubing Kit 229
Replace Parts Inside the 2-Channel-Degasser 230
Replace a Degasser Chamber 231
Replace the Degasser Pump 232
Replace the Sensor Board Assembly 234
Replace the Degasser Tubing Kit 235
Replace the Leak Panel 236
Replace the Leak Sensor 238
Replace the Status LED 239
Install an Active Seal Wash Upgrade Kit (Quaternary Pump) 241
Assemble the Main Cover 244
Install the Top Cover and Foam 245
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 148
9 Maintenance and Repair
Introduction to Maintenance and Repair
The module is designed for easy repair. The most frequent repairs such as piston
seal change and purge valve frit change can be done from the front of the
module with the module in place in the system stack.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 149
9 Maintenance and Repair
Cautions and Warnings
The module is partially energized when switched off, as long as the power
WAR N IN G
cord is plugged in.
Repair work at the module can lead to personal injuries, e.g. electrical
shock, when the cover is opened and the module is connected to power.
Always unplug the power cable before opening the cover.
Do not connect the power cable to the instrument while the covers are
removed.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 150
9 Maintenance and Repair
Cautions and Warnings
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 151
9 Maintenance and Repair
Overview of Maintenance and Repair
The following pages describe maintenance (simple repairs) of the pump that can
be carried out without opening the main cover.
Table 21 Simple repair procedures
“Checking and Replacing the Solvent If solvent filter is blocked Gradient performance problems, intermittent
Filter” on page 156 pressure fluctuations
“Exchange the Passive Inlet Valve” on If internally leaking Pressure ripple unstable, run Leak Rate Test for
page 157 verification
“Exchange the Active Inlet Valve (AIV) or its If internally leaking Pressure ripple unstable, run Leak Rate Test for
Cartridge” on page 185 verification
“Exchange the Outlet Valve” on page 159 If internally leaking Pressure ripple unstable, run Leak Rate Test for
verification
“Exchange the Purge Valve Frit or the Purge If internally leaking Solvent dripping out of waste outlet when valve
Valve” on page 161 closed
“Exchange the Purge Valve Frit or the Purge If the frit shows indication A pressure drop of > 10 bar across the frit (at a
Valve” on page 161 of contamination or water flow of 5 mL/min with open purge valve)
blockage indicates blockage
“Maintenance of a Pump Head Without Seal If pump performance Leaks at lower pump head side, unstable
Wash Option” on page 168 indicates seal wear retention times, pressure ripple unstable - run
Leak Rate Test for verification
Exchange pistons, see “Maintenance of a If scratched Seal life time shorter than usual - check pistons
Pump Head Without Seal Wash Option” on while changing the seals
page 168
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 152
9 Maintenance and Repair
Cleaning the Module
To keep the module case clean, use a soft cloth slightly dampened with water, or
a solution of water and mild detergent. Avoid using organic solvents for cleaning
purposes. They can cause damage to plastic parts.
Liquid dripping into the electronic compartment of your module can cause
WAR N IN G
shock hazard and damage the module
Do not use an excessively damp cloth during cleaning.
Drain all solvent lines before opening any connections in the flow path.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 153
9 Maintenance and Repair
Remove and Install Doors
The figures shown in this procedure exemplarily show the Infinity II Multisampler
NOTE module.
The principle of how to remove and/or install doors works in the same way for all
Infinity II modules.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 154
9 Maintenance and Repair
Remove and Install Doors
1 Press the release buttons and pull the front door out. 2 For the Installation of the front door. Insert the hinges
into their guides and move the door in until the release
buttons click into their final position.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 155
9 Maintenance and Repair
Checking and Replacing the Solvent Filter
A functional solvent filter is essential for a good pump performance and for
protecting the LC system.
Small particles can permanently block the capillaries and valves of the module.
C AU T I O N
Damage of the module.
Always filter solvents.
Never use the module without solvent inlet filter.
If the filter is in good condition the solvent will freely drip out of the solvent tube
NOTE (hydrostatic pressure). If the solvent filter is partly blocked only very little solvent
will drip out of the solvent tube.
1 Remove the solvent filter from the inlet filter adapter and replace it by a new
one.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 156
9 Maintenance and Repair
Exchange the Passive Inlet Valve
1 Disconnect the solvent inlet tube from the inlet valve. 2 Using a 14 mm wrench loosen the passive inlet valve
and remove the valve from the pump head.
NOTE
Beware of leaking solvents due to hydrostatic pressure.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 157
9 Maintenance and Repair
Exchange the Passive Inlet Valve
3 Insert the new valve into the pump head and tighten the 4 Reconnect the solvent inlet tube to the passive inlet
valve using a torque wrench (approx. 10 Nm). valve.
approx.
10 Nm
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 158
9 Maintenance and Repair
Exchange the Outlet Valve
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 159
9 Maintenance and Repair
Exchange the Outlet Valve
1 Using a 1/4 inch wrench disconnect the valve capillary 2 Using the 14 mm wrench loosen the valve and remove it
from the outlet valve. from the pump body.
NOTE
Do not disassemble the outlet valve, as this can damage the
valve.
3 Reinstall the outlet valve and tighten it using a torque 4 Reconnect the valve capillary to the outlet valve.
wrench set to approx. 10 Nm.
approx.
10 Nm
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 160
9 Maintenance and Repair
Exchange the Purge Valve Frit or the Purge Valve
When • Frit – when piston seals are exchanged or when contaminated or blocked (pressure drop of
> 10 bar across the frit at a flow rate of 5 mL/min of water with purge valve opened)
• Purge valve – if internally leaking
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 161
9 Maintenance and Repair
Exchange the Purge Valve Frit or the Purge Valve
1 Using a 1/4 inch wrench disconnect the pump outlet 2 Using the 14 mm wrench unscrew the purge valve and
capillary from the purge valve. Disconnect the waste remove it.
tube.
NOTE
Beware of leaking solvents due to hydrostatic pressure.
3 Remove the seal cap from the purge valve. 4 Using a pair of tweezers or a toothpick remove the frit.
Seal cap
Frit
5 Place a new frit into the purge valve with the orientation 6 Reinstall the seal cap including the gold seal.
of the frit as shown below (slit in frit points to the front).
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 162
9 Maintenance and Repair
Exchange the Purge Valve Frit or the Purge Valve
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 163
9 Maintenance and Repair
Replace the O-Ring on the Purge Valve
1 Disassemble the purge valve. 2 Remove the old o-ring from the purge valve.
3 Clean the purge valve parts.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 164
9 Maintenance and Repair
Replace the O-Ring on the Purge Valve
4 Place the new o-ring on the Screw Purge Valve. 5 Place the screw with o-ring on the Purge Valve Body.
6 Place the mounting ring on the screw and push down 7 Push the screw up and guide the o-ring into the gap.
the o-ring.
8 Place a new frit into the purge valve with the orientation 9 Install the purge valve to the pump. Make sure not to
of the frit as shown below (slit in frit points to the front). turn the purge valve body when the screw is fixed to the
Reinstall the seal cap including the gold seal. pump. The o-ring will take damage.
NOTE
Before reinstallation always check the gold seal in the seal
cap. A deformed seal cap should be exchanged.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 165
9 Maintenance and Repair
Remove the Pump Head Assembly
Preparations • Switch off the pump at the main power switch and unplug the power cable.
• Use an optional solvent shutoff valve or lift up solvent filters in solvent reservoirs for avoiding
leakages.
• Open the doors of the module.
• If an active inlet valve is installed, disconnect the active inlet valve cable.
1 Using a 1/4 inch wrench disconnect the valve capillary 2 Disconnect the solvent inlet tube from the inlet valve.
from the outlet valve.
NOTE
Beware of leaking solvents due to hydrostatic pressure.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 166
9 Maintenance and Repair
Remove the Pump Head Assembly
3 Using a 1/4 inch wrench disconnect the pump outlet 4 If applicable, remove tubings from the seal wash
capillary from the purge valve. Disconnect the waste support rings.
tube.
NOTE
Beware of leaking solvents due to hydrostatic pressure.
5 Remove the capillary at the bottom of the pump head. 6 Using a 4 mm hexagonal key, stepwise loosen the two
pump head screws.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 167
9 Maintenance and Repair
Maintenance of a Pump Head Without Seal Wash Option
1 Place the pump head on a flat surface. Loosen the lock 2 While holding the lower half of the assembly (piston
screw (two turns). housing), carefully pull the pump housing away from the
piston housing.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 168
9 Maintenance and Repair
Maintenance of a Pump Head Without Seal Wash Option
3 Remove the support rings from the piston housing. 4 Lift the housing away from the pistons.
5 Check the piston surface and remove any deposits or 6 Using the steel side of the insert tool, carefully remove
layers: clean the piston surface with abrasive paper and the seal from the pump housing.
rinse with 2-propanol. Replace piston if scratched.
Piston Surface
7 Using the plastic side of the insert tool, insert new seals 8 Place the support rings on the piston housing. Note the
into the pump head. correct position of the pins.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 169
9 Maintenance and Repair
Maintenance of a Pump Head Without Seal Wash Option
9 Place the pump housing onto the piston housing. 10 Insert the pistons and carefully press them into the
seals.
11 Tighten the lock screw. 12 Install the pump head, see “Reinstall the Pump Head
Assembly” on page 176.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 170
9 Maintenance and Repair
Maintenance of a Pump Head with Seal Wash Option
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 171
9 Maintenance and Repair
Maintenance of a Pump Head with Seal Wash Option
1 Place the pump head on a flat surface. Loosen the lock 2 While holding the lower half of the assembly (piston
screw (two turns). housing), carefully pull the pump housing away from the
piston housing.
3 Remove the seal holder and the seal wash support rings 4 Remove the seal holder from the support ring
from the piston housing. assembly.
5 Lift the housing away from the pistons. 6 Check the piston surface and remove any deposits or
layers: clean the piston surface with abrasive paper and
rinse with 2-propanol. Replace piston if scratched.
Piston Surface
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 172
9 Maintenance and Repair
Maintenance of a Pump Head with Seal Wash Option
7 Using the steel side of the insert tool, carefully remove 8 Using the plastic side of the insert tool, insert new seals
the seal from the pump housing. into the pump head.
9 Using the steel side of the insert tool, remove the seal 10 Using the plastic side of the insert tool, press the new
wash gasket and the wash seal from the support ring. wash seal (spring pointing upwards) into the recess of
the support ring.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 173
9 Maintenance and Repair
Maintenance of a Pump Head with Seal Wash Option
cut end
13 G5654A only: Place a seal wash gasket in the recess of 14 Place the support rings on the piston housing. Note the
the support ring. Use a matching orientation of gasket correct position of the pins on the support ring.
and support ring. Put the seal holder on top of the
gasket.
Backup seal
Seal holder
Seal holder
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 174
9 Maintenance and Repair
Maintenance of a Pump Head with Seal Wash Option
15 Place the pump housing onto the piston housing. 16 Insert the pistons and carefully press them into the
seals.
17 Tighten the lock screw. 18 Install the pump head, see “Reinstall the Pump Head
Assembly” on page 176.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 175
9 Maintenance and Repair
Reinstall the Pump Head Assembly
1 If needed, apply a small amount of grease on the back 2 Slide the pump head assembly onto the pump drive.
of the screws. Normally, the grease added during
manufacturing is sufficient for a long time.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 176
9 Maintenance and Repair
Reinstall the Pump Head Assembly
3 Insert the pump head screws and tighten them 4 Install the capillary at the bottom of the pump head.
stepwise. First tighten the left screw, then the right
screw, with a torque of approximately 5 Nm.
approx.
5 Nm
5 Reconnect the valve capillary to the outlet valve. 6 Reconnect the solvent inlet tube to the passive inlet
valve.
7 Reconnect the outlet capillary and the waste tubing. 8 If installed, reconnect the active inlet valve cable to its
connector.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 177
9 Maintenance and Repair
Reinstall the Pump Head Assembly
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 178
9 Maintenance and Repair
Seal Wear-in Procedure
1 Place a bottle with 100 mL of isopropanol in the solvent cabinet and put a
tubing (including bottle head assembly) into the bottle.
2 If an active inlet valve is installed, screw the PEEK adapter 1/4-28 to 10-32
(0100-1847) to the AIV and connect the inlet tube from the bottle head directly
to it.
3 Connect the Restriction capillary (5022-2159) to the purge valve. Connect its
other end to a waste container.
4 Open the purge valve and purge the system for 5 min with isopropanol at a
flow rate of 2 mL/min.
5 Close the purge valve and set the flow to a rate adequate to achieve a
pressure of 350 bar. Pump 15 min at this pressure to wear in the seals. The
pressure can be monitored using your instrument control software or tool.
6 Turn OFF the pump, slowly open the purge valve to release the pressure from
the system, disconnect the restriction capillary and reinstall the bottle with the
solvent for your application.
7 Rinse your system with the solvent used for your next application.
8 Replace the purge valve frit, see “Exchange the Purge Valve Frit or the Purge
Valve” on page 161.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 179
9 Maintenance and Repair
Exchange the Multi-Channel Gradient Valve (MCGV)
The life time of the multi-channel gradient valve can be increased by regularly
NOTE flushing the valve, especially when using buffers. If using buffers, flush all
channels of the valve with water to prevent precipitation of the buffer, otherwise
salt crystals could drop into an unused channel and form plugs that may cause
leaks of that channel. Such leaks will interfere with the general performance of
the valve. When using buffers in combination with organic solvents, it is
recommended to connect the aqueous solutions/buffers to one of the bottom
ports (A and D) and the organic solvent to one of the upper gradient valve ports.
It is best to have the organic channel directly above the buffer channel (e.g., A -
buffer, B - organic solvent).
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 180
9 Maintenance and Repair
Exchange the Multi-Channel Gradient Valve (MCGV)
1 Disconnect the solvent tubes from the MCGV. 2 Disconnect the connecting tube from the MCGV.
3 Press the lower sides of the cover to unclip it. Remove 4 Disconnect the MCGV cable from its connector.
the cover.
5 Unscrew the two screws and remove them. 6 Remove the MCGV.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 181
9 Maintenance and Repair
Exchange the Multi-Channel Gradient Valve (MCGV)
7 Place the new MCGV into position. Make sure that 8 Tighten the two screws.
channel A of the MCGV is put at the bottom-right
position.
11 Reconnect the tube from the inlet valve to the middle 12 Connect solvent tubes for channels A-D from the MCGV
position of the MCGV. to the degasser outlets.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 182
9 Maintenance and Repair
Replace the Solvent Selection Valve (Isocratic Pump)
When If leaking internally (crossflow between the ports), or if one of the channels is blocked
1 Lift solvent bottles out of the solvent cabinet and place 2 Disconnect all tubings from the solvent selection valve.
them on the table. Disconnect the solvent tubes from
the solvent selection valve and empty the tubes into the
bottles. Place the bottles back into the solvent cabinet.
3 Using a screwdriver, loosen the holding screws of the 4 Carefully disconnect the valve cable at the SSV
valve holder. connector on the Z-Panel.
5 Using a screwdriver, loosen the holding screws of the
degasser front cover.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 183
9 Maintenance and Repair
Replace the Solvent Selection Valve (Isocratic Pump)
6 Carefully pull the valve with a cable out of the pump 7 Exchange the defective solvent selection valve.
mainframe.
8 Tighten the screws of the degasser front cover and the 9 Reconnect all tubings to the solvent selection valve.
valve holder.
NOTE
After an exchange of the valve it may be required to pump
several mL of solvent before the flow stabilizes at a
pressure ripple as low as it used to be when the system was
still working properly.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 184
9 Maintenance and Repair
Exchange the Active Inlet Valve (AIV) or its Cartridge
Preparations • Switch off pump at the main power switch and unplug the power cable.
• Use an optional solvent shutoff valve or lift up solvent filters in solvent reservoirs for avoiding
leakages.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 185
9 Maintenance and Repair
Exchange the Active Inlet Valve (AIV) or its Cartridge
4 Using a 14 mm wrench loosen the active inlet valve and remove the valve
from the pump head.
5 Using a pair of tweezers remove the valve cartridge from the actuator
assembly.
6 Before inserting the new valve cartridge clean the area in the actuator
assembly. Flush the cartridge area thoroughly with alcohol.
7 Insert a new cartridge into the actuator assembly (make sure the valve
cartridge is completely inserted into the actuator assembly).
8 Insert the new valve into the pump head. Using the 14 mm wrench turn the
nut until it is hand tight.
9 Position the valve such that the solvent inlet tube connection points towards
the front.
10 Using the 14 mm wrench tighten the nut by turning the valve into its final
position (not more than a quarter turn).
11 Reconnect the adapter at the active inlet valve.
12 Reconnect the solvent inlet tube to the adapter. Reconnect the active inlet
valve cable to the connector in the Z-panel.
13 Purge the system with 30 mL of solvent in order to achieve a low pressure
ripple.
See “Regular Priming” on page 66
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 186
9 Maintenance and Repair
Exchange the Seal Wash Cartridge
1 Remove the wash solvent tubings from the support ring 2 Unclip the peristaltic pump cartridge from the module
outlet and from the adapter leading to the waste bottle. housing and remove it.
3 Put the new peristaltic pump cartridge onto the rod of 4 Connect the peristaltic pump tubes to the support rings
the pump motor and push the plastic clips into the outlet and to the adapter leading to the waste bottle.
module housing.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 187
9 Maintenance and Repair
Replace Leak Handling System Parts
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 188
9 Maintenance and Repair
Replacing the Module Firmware
Preparations Read update documentation provided with the Firmware Update Tool.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 189
9 Maintenance and Repair
Introduction to Repair
Introduction to Repair
Repairs may require the exchange of defective internal parts. Exchange of these
parts requires removing the module from the stack, removing the covers, and
disassembling the module.
Repair and service activities should only be performed by trained and authorized
NOTE service personnel.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 190
9 Maintenance and Repair
Overview of Repairing the Pump
Figure 22 on page 191 shows the main repair assemblies of the isocratic pump,
Figure 23 on page 192 shows the main repair assemblies of the quaternary
pump. The pump head and its parts require normal maintenance (for example,
seal exchange) and can be accessed from the front (simple repairs). Replacing
internal parts will require to remove the pump from its stack and to open the top
cover.
11
10
9 12
13
8
14
7
6
15
5
4
3
2
1
Figure 22 Overview of Main Assemblies (Isocratic Pump)
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 191
9 Maintenance and Repair
Overview of Repairing the Pump
6 9, 12
10
5
11
4
13
3
9, 12 Damper and Damper Leak Panel, “Exchange the Damper” on page 208
13 Active Seal Wash, “Install an Active Seal Wash Upgrade Kit (Quaternary Pump)” on
page 241
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 192
9 Maintenance and Repair
Exchanging Internal Parts
The procedures in this section describe how to exchange internal parts. You
must remove the pump from the stack in order to open the main cover.
Table 22 Exchanging internal parts
“Exchange the Pump Mainboard” on page 196 If defective Certain error messages are
generated
“Install an Active Seal Wash Upgrade Kit (Quaternary Pump)” on If seals show indication of Leaks at lower pump head
page 241 leaks side, loss of wash solvent
“Exchange the Fan” on page 214 If defective Fan not running; error
message “Fan Failed” is
generated
“Exchange the Pump Drive” on page 212 If mechanically blocked or Certain error messages are
defective generated
“Replace the Leak Sensor” on page 238 If defective Leak messages without leak
in the leak pan
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 193
9 Maintenance and Repair
Remove the Top Cover and Foam
Preparations • Switch off pump at the main power switch and unplug the power cable.
• Disconnect the solvent inlet and outlet tubes from the pump. Beware of leaking solvents due to
hydrostatic flow.
• Remove the pump from the stack.
• Remove the doors of the module.
Module is partially energized when switched off, as long as the power cord
WAR N IN G
is plugged in.
Repair work at the module can lead to personal injuries, e.g. shock hazard,
when the cover is opened and the module is connected to power.
Make sure that it is always possible to access the power plug.
Remove the power cable from the instrument before opening the cover.
Do not connect the power cable to the Instrument while the covers are
removed.
1 Release the clips of the cover at the rear of the module. 2 Lift the cover up and slide it towards the rear.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 194
9 Maintenance and Repair
Remove the Top Cover and Foam
3 Unscrew the screws at the rear end of the top plate. 4 Slide the plate towards the front and remove it.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 195
9 Maintenance and Repair
Exchange the Pump Mainboard
Preparations • Turn off the pump, disconnect all cables and remove the pump from the stack.
• Remove the top cover and foam (see “Remove the Top Cover and Foam” on page 194).
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 196
9 Maintenance and Repair
Exchange the Pump Mainboard
The pump mainboard carries the Fusion board. Both boards can be ordered and
NOTE replaced separately.
2 Remove the connector screws and the grounding screw, which fixes the
board to the module housing.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 197
9 Maintenance and Repair
Exchange the Pump Mainboard
If you want to exchange other parts as well, continue with that work first.
NOTE
4 Lift the FUSION Board carefully from the mainboard. Depending on the
module there might be two white plastic spacers that fix the board in position.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 198
9 Maintenance and Repair
Exchange the Pump Mainboard
6 On the new fusion board check the configuration switch S1 setting, see
“Setting the 6-bit Configuration Switch” on page 317.
An incorrect switch setting (e.g., TEST/BOOT) may cause the module to turn into
NOTE a non-operational mode (yellow or red flashing status light). In such a case turn
off the module, re-set the address switches, and turn on the module again.
7 Install the new board. Make sure that the board is fitted correctly in the board
recess holes at the rear panel.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 199
9 Maintenance and Repair
Exchange the Pump Mainboard
11 12 13 14
10
9
8
7
4 15
3 16
2
17
1
4 Fan 13 LAN
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 200
9 Maintenance and Repair
Exchange the Pump Mainboard
10 Replace foam and top covers, see “Install the Top Cover and Foam” on
page 245.
11 Reinstall the pump in the stack. Reconnect all cables. Turn on the pump. If
pump status indicator turns red continue with the section “Serial Number and
Type Change” on page 203 and “Setting the 6-bit Configuration Switch” on
page 317.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 201
9 Maintenance and Repair
Replace the Fusion Board
1 Lift the FUSION Board carefully from the mainboard. 2 Insert the (new) FUSION board on the two connectors.
Depending on the module there might be two white
plastic spacers that fix the board in position.
Next Steps:
3 On the new fusion board check the configuration switch S1 setting, see “Setting the 6-bit Configuration Switch” on
page 317.
NOTE
An incorrect switch setting (e.g., TEST/BOOT) may cause the module to turn into a non-operational mode (yellow or red
flashing status light). In such a case turn off the module, re-set the address switches, and turn on the module again.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 202
9 Maintenance and Repair
Changing the Type and Serial Number
With firmware A.06.02/B.01.02 and above, the entry is checked against the board
NOTE revision, which inhibits setting wrong types for a board. The caution above for
multiple usage boards still applies.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 203
9 Maintenance and Repair
Changing the Type and Serial Number
Some Agilent modules require the correct mainboard version to match the type.
NOTE
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 204
9 Maintenance and Repair
Changing the Type and Serial Number
Recover Instructions
Overview
The following situations may come up where the instrument is no longer usable
due to
• an incorrect type (product number) entry after the replacement of a
mainboard of the module.
• load of wrong firmware based on the wrong type.
With firmware A.06.02/B.01.02 and above, the entry is checked against the board
NOTE revision.
User Interface incorrect but valid type incorrect but valid type incorrect and invalid type
Example Conditions correct type = G1315B correct type = G1315B correct type = G1315B
entered type = G1314B entered type = G1314B entered type = G1319B
plus wrong firmware from
G1314B
Agilent Diagnostic shows the incorrect product shows the incorrect product shows the incorrect product
Software (preferred number number number
tool) access to the module is access to the module is access to the module is
possible as described in possible as described in possible as described in
“Recover with Agilent Lab “Recover with Agilent Lab “Recover with Agilent Lab
Advisor (Type Only)” on Advisor (Type Only)” on Advisor (Type Only)” on
page 206 page 206 page 206
If wrong firmware has been
loaded in addition, only the
Agilent Lab Advisor Software
can revert to correct product
number as described in
“Recover with Agilent Lab
Advisor (Type & Firmware)” on
page 207
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 205
9 Maintenance and Repair
Changing the Type and Serial Number
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 206
9 Maintenance and Repair
Changing the Type and Serial Number
1 To set up the module in Lab Advisor, connect the module (or the system) in
Lab Advisor.
2 Select Instrument Control on the left panel.
3 Open the Controls menu of the respective module.
4 Expand the Conversions sections of the menu.
5 Click Transfer to Resident FW.
6 Open a connection to the module (or via system).
7 Select Service & Diagnostics on the left navigation panel.
8 Select the module of interest in the middle panel.
9 Select Board Check and Change and press Run in the right bottom corner.
10 In the type field enter G1315B and press Apply.
11 Close the Lab Advisor Software.
12 Switch off and on the changed LC module.
13 Open Lab Advisor and connect to the module (or the system)
14 Set module's configuration back to Main Mode using Lab Advisor (Instrument
Control > Controls > Conversions).
15 Update firmware if required.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 207
9 Maintenance and Repair
Exchange the Damper
Preparations • Switch off the pump at the main power switch, remove all cables and tubings, remove the pump
from the stack.
• Remove the MCGV Cover (see “Exchange the Multi-Channel Gradient Valve (MCGV)” on
page 180)
• Remove the top cover and top foam section (see “Remove the Top Cover and Foam” on
page 194).
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 208
9 Maintenance and Repair
Exchange the Damper
Wrong damper
C AU T I O N
Using a wrong damper will cause a wrong regulation of the pump and wrong
measurement results.
Make sure to use the correct damper type. The 600 bar damper for the
Isocratic and Quaternary Pump can be identified by two rings and a yellow
mark. The 400 bar damper for the Quaternary Pump VL does not have rings
and the mark. The 600 bar damper of the G1312B or G7112B Binary Pump
does not have a pressure sensor.
1 Disconnect the two damper capillaries. 2 Disconnect the damper connector at the mainboard.
Lift the damper out of its position.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 209
9 Maintenance and Repair
Exchange the Damper
3 Loosen the screw of the degasser cover and the 4 4 Remove the damper leak panel.
screws of the Z-panel and take it out of the instrument.
5 Make sure that the pressure sensor cable of the damper 6 Reinstall the leak panel.
is wound around the pressure sensor as indicated (2
windings).
7 Install the Z-panel and fix it with the 4 screws. Insert the 8 Place the new damper into the recess and connect it to
lower end of the degasser cover to the module housing. the mainboard. The damper needs to be aligned such
Then fix it with a screw to the Z-panel. that both capillary connections are on a vertical line.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 210
9 Maintenance and Repair
Exchange the Damper
9 Do not confuse the connectors for the seal wash pump 10 Reconnect the two damper capillaries.
and the pressure sensor, as this will destroy the
mainboard!
PS
SW
Next Steps:
11 Replace the MCGV Cover (see “Exchange the Multi-Channel Gradient Valve (MCGV)” on page 180).
12 Replace the top foam section, metal cover and top cover (see “Install the Top Cover and Foam” on page 245). Replace
the pump on stack, reconnect all tubings and cables.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 211
9 Maintenance and Repair
Exchange the Pump Drive
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 212
9 Maintenance and Repair
Exchange the Pump Drive
1 Loosen the screw of the degasser cover and the 4 2 Disconnect the pump drive cables from the mainboard
screws of the Z-panel and take it out of the instrument. and lift the drive out of the foam.
3 Place the new pump drive into the recess in the foam 4 Install the Z-panel and fix it with the 4 screws. Insert the
and connect the cables to the connectors on the lower end of the degasser cover to the module housing.
mainboard. Then fix it with a screw to the Z-panel.
Next Steps:
5 Install the top cover and top foam section (see “Install the Top Cover and Foam” on page 245).
6 Install the MCGV cover (see “Exchange the Multi-Channel Gradient Valve (MCGV)” on page 180).
7 Reinstall the pump head (see “Reinstall the Pump Head Assembly” on page 176).
8 Install the doors, replace the pump on stack, reconnect all tubings and cables.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 213
9 Maintenance and Repair
Exchange the Fan
Preparations • Switch off the pump at the main power switch, remove all cables and tubings, remove the pump
from the stack.
• Remove the doors.
• Remove the top cover and top foam section (see “Remove the Top Cover and Foam” on
page 194).
1 Disconnect the fan connector from the mainboard. 2 Remove the mainboard, see “Exchange the Pump
Mainboard” on page 196.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 214
9 Maintenance and Repair
Exchange the Fan
3 Push the fan to the left and lift it out of the foam part. 4 Slide the new fan into its position.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 215
9 Maintenance and Repair
Replace Power Supply and Cables
Preparations • Remove the doors, see “Remove and Install Doors” on page 154.
• Remove the MCGV, see “Exchange the Multi-Channel Gradient Valve (MCGV)” on page 180.
• Remove the leak interface, see “Replace Leak Handling System Parts” on page 188.
• Remove the top covers and foams, see “Remove the Top Cover and Foam” on page 194.
• Remove the mainboard, see “Exchange the Pump Mainboard” on page 196.
• Remove the degasser, see “Replace the Degasser” on page 221.
• Remove the pump drive, see “Exchange the Pump Drive” on page 212.
• Remove the fan, see “Exchange the Fan” on page 214.
1 Remove the bottom foam. 2 Unplug the power switch cable from the power switch
connector (1), and remove the adhesive stripes (2).
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 216
9 Maintenance and Repair
Replace Power Supply and Cables
3 Open grounding screws at the rear of the instrument, 4 Remove the power supply.
which fix the power supply to the instrument housing.
5 Check that the Power Switch is in the inner (pressed) 6 Ensure that the PWS-CONTROL Cable is connected to
position. If not, use a screw driver and press on it. the Power Supply.
NOTE
If the switch is in the outer (released) position the Power
Supply is in the OFF-mode and the Power Switch in the front
will not work!
7 Install the Power Supply into the instrument housing. 8 Fix the power supply to the instrument housing with the
grounding screws.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 217
9 Maintenance and Repair
Replace Power Supply and Cables
9 Plug the new power switch cable into the power switch 10 Install the bottom foam.
connector (1) (contacts upward), and fix the cable with
the adhesive stripes (2) on the bottom.
2
1
Installation Activities
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 218
9 Maintenance and Repair
Replace the Power Switch
1 Place the module on its left side and locate the Power 2 While pressing with a flat screw driver into the slit on
Switch. the bottom [1] push the levers [2] into the panel.
3 Place the module in its normal position on the bench. 5 Slide out the Power Switch completely.
4 Remove the Leak Panel, see “Replace the Leak
Panel” on page 236.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 219
9 Maintenance and Repair
Replace the Power Switch
6 Disconnect the Leak Sensor Cable [1] and the Power 7 Reconnect the Power Switch cable [1] and the Leak
Switch Cable [2] (just pull the flat cable out of the Sensor cable [2] to the new Power Switch.
connector).
2 1 1 2
8 Slide the power switch into its final position (until it Next Steps:
clicks).
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 220
9 Maintenance and Repair
Replace the Degasser
1 Remove solvent tubings from the degasser inlets and 2 Open the screw of the degasser cover and remove the
outlets. cover.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 221
9 Maintenance and Repair
Replace the Degasser
3 Pull out the degasser for some centimeters and remove 4 Reconnect the cable.
the cable.
5 Slide in the degasser. 6 Install the degasser cover, tighten the screw of the
degasser cover.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 222
9 Maintenance and Repair
Replace Parts Inside the 4-Channel-Degasser
Preparations Remove the degasser from the pump, see “Replace the Degasser” on page 221.
Clip
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 223
9 Maintenance and Repair
Replace Parts Inside the 4-Channel-Degasser
Connector
4 Reinstall the degasser into the pump and reconnect the degasser connector,
see “Replace the Degasser” on page 221.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 224
9 Maintenance and Repair
Replace Parts Inside the 4-Channel-Degasser
Preparations • Remove the degasser, see “Replace the Degasser” on page 221.
• Open the degasser cover, see “Replace Parts Inside the 4-Channel-Degasser” on page 223.
1 Disconnect the tubing from the chamber. 2 Remove the screw at the front of the degasser housing
and remove the degasser chamber.
Degasser chamber
Tubing
Screws
3 Install the new degasser chamber, tighten the screw at the front of the degasser and reconnect the tubing to the new
chamber.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 225
9 Maintenance and Repair
Replace Parts Inside the 4-Channel-Degasser
Preparations • Remove the degasser, see “Replace the Degasser” on page 221.
• Open the degasser cover, see “Replace Parts Inside the 4-Channel-Degasser” on page 223.
1 For replacing the degasser pump, disconnect both 2 Remove the three screws at the bottom of the degasser
tubings and the cable to the pressure sensor. and remove the pump.
Degasser pump
3 For installing the new pump, fix the screws at the 4 Connect the cable to the pressure sensor board.
bottom of the degasser. Connect the tubing from the
chambers to the inlet of the pump and the drain tubing
to the holder at the bottom of the degasser.
Drain outlet
Inlet
Tubing holder
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 226
9 Maintenance and Repair
Replace Parts Inside the 4-Channel-Degasser
Preparations • Remove the degasser, see “Replace the Degasser” on page 221.
• Open the degasser cover, see “Replace Parts Inside the 4-Channel-Degasser” on page 223.
1 For replacing the sensor board assembly, remove the 2 Then open both screws at the front of the degasser.
cable to the vacuum pump. Disconnect the tubing from
the pressure sensor.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 227
9 Maintenance and Repair
Replace Parts Inside the 4-Channel-Degasser
3 Open both screws next to the pressure sensor, lift up 4 Tighten the two screws next to the pressure sensor and
the board and replace it by the new one. the two screws at the front of the degasser.
Connector
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 228
9 Maintenance and Repair
Replace Parts Inside the 4-Channel-Degasser
Preparations • Remove the degasser, see “Replace the Degasser” on page 221.
• Open the degasser cover, see “Replace Parts Inside the 4-Channel-Degasser” on page 223.
1 For replacing the degasser tubing kit, remove tubings 2 Connect the new tubing from the chambers to the inlet
from the vacuum pump, degasser chambers and the of the pump and the new drain tubing to the holder at
pressure sensor. the bottom of the degasser.
Drain outlet
Inlet
Tubing holder
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 229
9 Maintenance and Repair
Replace Parts Inside the 2-Channel-Degasser
Preparations Remove the degasser from the pump, see “Replace the Degasser” on page 221.
1 Open the degasser cover by lifting it up from the clips. 2 Replace a degasser chamber (see “Replace a Degasser
Chamber” on page 231)
OR
replace the degasser pump (see “Replace the Degasser
Pump” on page 232)
OR
replace the sensor board assembly (see “Replace the
Sensor Board Assembly” on page 234)
OR
replace the degasser tubing kit (see “Replace the Degasser
Clip Tubing Kit” on page 235).
3 Close the degasser cover. Note the correct orientation 4 Reinstall the degasser into the pump and reconnect the
of the cover with respect to the connector. degasser connector, see “Replace the Degasser” on
page 221.
Connector
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 230
9 Maintenance and Repair
Replace Parts Inside the 2-Channel-Degasser
Preparations • Remove the degasser, see “Replace the Degasser” on page 221.
• Open the degasser cover, see “Replace Parts Inside the 2-Channel-Degasser” on page 230.
1 Disconnect the tubing from the chamber. 2 Remove the screw at the front of the degasser housing.
5 Tighten the screw at the front of the degasser. 6 Reconnect the tubing to the new chamber.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 231
9 Maintenance and Repair
Replace Parts Inside the 2-Channel-Degasser
Preparations • Remove the degasser, see “Replace the Degasser” on page 221.
• Open the degasser cover, see “Replace Parts Inside the 2-Channel-Degasser” on page 230.
1 For replacing the degasser pump, disconnect both 2 Remove the three screws at the bottom of the
tubings and the cable to the mainboard. degasser.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 232
9 Maintenance and Repair
Replace Parts Inside the 2-Channel-Degasser
5 Fix the screws at the bottom of the degasser. 6 Connect the tubing from the chambers to the inlet of
the pump and the drain tubing to the holder at the
bottom of the degasser. Connect the cable to the
mainboard.
Drain Outlet
Inlet
Tubing holder
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 233
9 Maintenance and Repair
Replace Parts Inside the 2-Channel-Degasser
Preparations • Remove the degasser, see “Replace the Degasser” on page 221.
• Open the degasser cover, see “Replace Parts Inside the 2-Channel-Degasser” on page 230.
1 For replacing the sensor board assembly, remove the 2 Open the clips next to the pressure sensor and lift up
cable from the mainboard. Disconnect the tubing from the board.
the pressure sensor.
3 Install the sensor board assembly. 4 Connect the cable to the mainboard.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 234
9 Maintenance and Repair
Replace Parts Inside the 2-Channel-Degasser
Preparations • Remove the degasser, see “Replace the Degasser” on page 221.
• Open the degasser cover, see “Replace Parts Inside the 2-Channel-Degasser” on page 230.
1 For replacing the degasser tubing kit, remove tubings 2 Connect the new tubing from the chambers to the inlet
from the vacuum pump, degasser chambers and the of the pump and the new drain tubing to the holder at
pressure sensor. the bottom of the degasser.
Drain outlet
Inlet
Tubing holder
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 235
9 Maintenance and Repair
Replace the Leak Panel
When If the leak panel, main power switch or leak sensor are damaged
3 Remove the sealing lip. 4 Remove or replace the leak sensor as described in
“Replace the Leak Sensor” on page 238
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 236
9 Maintenance and Repair
Replace the Leak Panel
5 Install the sealing lip. 6 Insert the new leak panel and push it to the bottom of
the module bottom panel. Make sure to keep the sealing
lip in the correct position.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 237
9 Maintenance and Repair
Replace the Leak Sensor
Preparations Remove the leak panel, see “Replace the Leak Panel” on page 236.
1 Remove the leak sensor and disconnect the leak sensor 2 Replace the leak sensor.
cable from the power switch connector.
3 Route the leak sensor cable to the left and reconnect it 4 Reinsert the leak panel, see “Replace the Leak Panel” on
to the power switch board. page 236.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 238
9 Maintenance and Repair
Replace the Status LED
Preparations • Remove the doors, see “Remove and Install Doors” on page 154.
• Remove the leak interface, see “Replace Leak Handling System Parts” on page 188.
• Remove the top covers and foam, see “Remove the Top Cover and Foam” on page 194.
1 The Status LED cable is connected on the mainboard to 2 Disconnect the cable from the mainboard
connector STATUS_LED [1] and pasted to the right (STATUS_LED).
sheet metal wall [2].
3 Pull the cable from the side wall starting from the rear. 4 Lift the Status LED to the top.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 239
9 Maintenance and Repair
Replace the Status LED
5 Insert the new Status LED. 6 Route the cable to the rear and fix it on the wall (there
are adhesive stripes on the back of the cable).
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 240
9 Maintenance and Repair
Install an Active Seal Wash Upgrade Kit (Quaternary Pump)
When This kit is recommended when using buffers or other salt solutions as solvents. It is used for
cleaning pump head pistons and seals.
1 Place the pump head on a flat surface. Loosen the lock 2 While holding the lower half of the assembly (piston
screw (two turns). housing), carefully pull the pump housing away from the
piston housing.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 241
9 Maintenance and Repair
Install an Active Seal Wash Upgrade Kit (Quaternary Pump)
3 Lift the housing away from the pistons. 4 Place the support rings with seal wash seals on the
piston housing. Note the correct position of the pins on
the support ring.
5 Place the pump housing onto the piston housing. 6 Insert the pistons and carefully press them into the
seals.
7 Tighten the lock screw. 8 If needed, apply a small amount of grease to the pump
head screws. Slide the pump head assembly onto the
pump drive. Tighten the pump head screws stepwise
with increasing torque (5 Nm). See “Reinstall the Pump
Head Assembly” on page 176.
9 Reconnect all capillaries, tubes, and (if installed) the
active inlet valve cable to its connector.
10 Fix the seal wash pump motor to the Z-panel of the
pump housing using two Pozidriv screws and two
shock mounts.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 242
9 Maintenance and Repair
Install an Active Seal Wash Upgrade Kit (Quaternary Pump)
11 Put the new peristaltic pump cartridge onto the rod of 12 Connect the cable to the pump mainboard.
the pump motor and push the plastic clips into the
module housing.
SW
13 Do not confuse the connectors for the seal wash pump 14 Install the foam and top cover, see “Install the Top
and the pressure sensor, as this will destroy the Cover and Foam” on page 245.
mainboard!
Connector
pressure sensor
Connector
seal wash pump
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 243
9 Maintenance and Repair
Assemble the Main Cover
Wrong assembled
C AU T I O N
In case you insert the left or right side in the opposite position, you may not be
able to remove the side from the top part.
Take care not to mix up left and right side.
1 Place the top part on the bench and insert the left and right side into the top
part.
Front
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 244
9 Maintenance and Repair
Install the Top Cover and Foam
Preparations Make sure that after your repair all assemblies, cables, capillaries and connectors are located in their
correct place.
1 Install the top foam section. 2 Install the top plate and slide it towards the rear.
3 Fix the screws at the rear end of the top plate. 4 Replace the cover.
Next Steps:
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 245
10 Parts for Maintenance and Repair
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 246
10 Parts for Maintenance and Repair
Accessory Kits
Accessory Kits
# p/n Description
2 5043-1013 Tubing Clip IF-II
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 247
10 Parts for Maintenance and Repair
Accessory Kits
# p/n Description
2 5043-1013 Tubing Clip IF-II
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 248
10 Parts for Maintenance and Repair
Pump Head Assembly Without Seal Wash
6 5063-6589 Piston seal PTFE, carbon filled, black (pack of 2), default
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 249
10 Parts for Maintenance and Repair
Pump Head Assembly Without Seal Wash
The Pump Head 1200 SL without Seal Wash (G1312-60056) includes items 1-7,
10 and 12.
For piston seals, see “Choosing the Right Pump Seals” on page 107.
1
9
10
2
3
4
5
6
12
7
8
11
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 250
10 Parts for Maintenance and Repair
Pump Head Assembly with Seal Wash Option
Pump head assembly with seal wash (G1312-60045) and Pump Head with Seal
Wash Function (Bio-inert) (G5611-60165) contain:
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 251
10 Parts for Maintenance and Repair
Pump Head Assembly with Seal Wash Option
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 252
10 Parts for Maintenance and Repair
Pump Head Assembly with Seal Wash Option
The Pump head assembly with seal wash (G1312-60045) includes items 1-8, 10
and 13.
The 1220/1260 Seal Wash Pump Assembly (G1311-60161) does not include
NOTE screws. For installation, two screws (Screws for Seal Wash Pump Motor
(0515-1508)) are needed.
15
14
1
10 11 2
3
4
5
6
7
13
8
9
12
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 253
10 Parts for Maintenance and Repair
Outlet Valve
Outlet Valve
p/n Description
G1312-60067 Outlet valve
for G7110B, G7111A, and G7111B
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 254
10 Parts for Maintenance and Repair
Passive Inlet Valve
p/n Description
G1312-60066 Passive inlet valve 1220/1260
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 255
10 Parts for Maintenance and Repair
Purge Valve Assembly
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 256
10 Parts for Maintenance and Repair
Active Inlet Valve Assembly
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 257
10 Parts for Maintenance and Repair
Active Inlet Valve Assembly
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 258
10 Parts for Maintenance and Repair
1260 Infinity II Max Uptime Kit
This kit will be sold optionally with the pump and is not essential for operation.
You can reorder the individual parts.
# p/n Description
1 01018-22707 PTFE frits (pack of 5)
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 259
10 Parts for Maintenance and Repair
1260 Infinity II Max Uptime Kit Bio
This kit will be sold optionally with the pump and is not essential for operation.
You can reorder the individual parts.
# p/n Description
1 5043-0915 Fitting mounting tool
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 260
10 Parts for Maintenance and Repair
1260 Infinity II Low Dispersion Kit MCT
This kit will be sold optionally with the pump and is not essential for operation.
You can reorder the individual parts.
The 1260 Infinity II Low Dispersion Kit (G7111-68708) contains:
p/n Description
5500-1201 Capillary ST 0.12 mm x 105 mm SL
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 261
10 Parts for Maintenance and Repair
HPLC System Tool Kit
5023-2504
(Hex Driver SW-4 slitted)
5023-3138 5023-2503
(Reversible Screwdriver) (Hex Driver
SW-5 slitted)
5023-2502
(Hex Driver SW-6,
35/ ¼” slitted)
5023-3088
8710-1924
8710-2409
8710-0510
8710-0510
5023-2500
8710-1534
8710-2394
(Hex Key 9/64“, 15cm)
Box with:
9301-0411 (Syringe, Plastic) 5023-3089
9301-1337 (Syringe Adapter) Torx Key Set (T8,T9,T10,T15,T20,T25)
0100-1710 (Mounting tool for flangeless nut)
0100-1681 (Adapter luer/barb) 5023-2499
01018-23702 (Seal Insert tool) (Hex Key Set)
5067-6127 (Blank Nut V) 5043-1361
5023-2653 (Hex Key 3/32“) (Hex Key Set Driver)
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 262
10 Parts for Maintenance and Repair
Solvent Cabinet
Solvent Cabinet
p/n Description
5067-6871 Solvent Cabinet Kit
For details refer to: Usage Guideline for the Solvent Cabinet (01200-90150)
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 263
10 Parts for Maintenance and Repair
Bottle Head Assembly
p/n Description
5063-6598 Ferrules with lock ring (10/Pk)
Wire marker
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 264
10 Parts for Maintenance and Repair
Hydraulic Path of the Quaternary Pump
4 1 G7111-81601 Capillary
outlet valve to damper
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 265
10 Parts for Maintenance and Repair
Hydraulic Path of the Quaternary Pump
5
4
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 266
10 Parts for Maintenance and Repair
Hydraulic Path of the Isocratic Pump
3 G7111-81601 Capillary
outlet valve to damper
1
Figure 31 Hydraulic Flow Path of the Isocratic Pump
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 267
10 Parts for Maintenance and Repair
Overview of Main Assemblies
6 9, 12
10
5
11
4
13
3
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 268
10 Parts for Maintenance and Repair
Overview of Main Assemblies
6 5067-6199 Fan 92 mm x 92 mm
7 G7111-65800 Mainboard 1260 Iso, Quat & Bio Pump 600 bar
10 5067-6214 Status LED Assembly includes Status LED Holder and cable
(The cable has adhesive stripes at the back)
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 269
10 Parts for Maintenance and Repair
Overview of Main Assemblies
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 270
10 Parts for Maintenance and Repair
Overview of Main Assemblies
9 5067-6199 Fan 92 mm x 92 mm
10 G7111-65800 Mainboard 1260 Iso, Quat & Bio Pump 600 bar
13 5067-6214 Status LED Assembly includes Status LED Holder and cable
(The cable has adhesive stripes at the back)
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 271
10 Parts for Maintenance and Repair
Overview of Main Assemblies
11
10
9 12
13
8
14
7
6
15
5
4
3
2
1
Figure 34 Overview of Main Assemblies (Isocratic Pump)
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 272
10 Parts for Maintenance and Repair
Overview of Main Assemblies
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 273
10 Parts for Maintenance and Repair
Degasser
Degasser
Four-Channel Degasser
Item p/n Description
G1311-60070 Degasser 4 Channels for Quaternary Pump
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 274
10 Parts for Maintenance and Repair
Degasser
4
5
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 275
10 Parts for Maintenance and Repair
Cover Parts
Cover Parts
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 276
10 Parts for Maintenance and Repair
Sheet Metal Kit
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 277
10 Parts for Maintenance and Repair
Foam Parts
Foam Parts
# p/n Description
1 G7111-68702 Foam kit (includes upper and lower foam)
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 278
10 Parts for Maintenance and Repair
Power, Status LED and Illumination LED
4 5067-6214 Status LED Assembly includes Status LED Holder and cable
(The cable has adhesive stripes at the back)
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 279
10 Parts for Maintenance and Repair
Leak Parts
Leak Parts
5067-6214 Status LED Assembly includes Status LED Holder and cable
(The cable has adhesive stripes at the back)
3
7
2 8
9
1
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 280
11 Identifying Cables
This chapter provides information on cables used with the Agilent 1200 Infinity
Series modules.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 281
11 Identifying Cables
Cable Overview
Cable Overview
Never use cables other than the ones supplied by Agilent Technologies to ensure
NOTE proper functionality and compliance with safety or EMC regulations.
Analog cables
p/n Description
35900-60750 Agilent 35900A A/D converter
Remote cables
p/n Description
5188-8029 ERI
to general purpose
CAN cables
p/n Description
5181-1516 CAN cable, Agilent module to module, 0.5 m
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 282
11 Identifying Cables
Cable Overview
LAN cables
p/n Description
5023-0203 Cross-over network cable, shielded, 3 m (for point to point connection)
RS-232 cables
(not for p/n Description
FUSION board)
RS232-61601 RS-232 cable, 2.5 m
Instrument to PC, 9-to-9 pin (female). This cable has special pin-out,
and is not compatible with connecting printers and plotters. It is also
called "Null Modem Cable" with full handshaking where the wiring is
made between pins 1-1, 2-3, 3-2, 4-6, 5-5, 6-4, 7-8, 8-7, 9-9.
USB cables
p/n Description
5188-8050 USB A M-USB Mini B 3 m (PC-Module)
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 283
11 Identifying Cables
Analog Cables
Analog Cables
1 Not connected
2 Shield Analog -
3 Center Analog +
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 284
11 Identifying Cables
Analog Cables
1 Not connected
2 Black Analog -
3 Red Analog +
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 285
11 Identifying Cables
Remote Cables
Remote Cables
10 violet DGND
12 red-blue PGND
13 white-green PGND
NC yellow-brown
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 286
11 Identifying Cables
Remote Cables
10 GND 1
2 Stop 8 Low
3 Ready 7 High
5 Power on 6 High
4 Future 5
7 Start 3 Low
8 Prepare 2 Low
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 287
11 Identifying Cables
Remote Cables
p/n 5188-8057 Pin (ERI) Signal Pin (APG) Active (TTL) Pin (RJ45)
10 GND 1 5
1 Start 9 High
Request
2 Stop 8 High
3 Ready 7 High
4 Fraction 5 High 4
Trigger
5 Power on 6 High
7 Start 3 High
8 Prepare 2 High
Ground Cable NC
Shielding
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 288
11 Identifying Cables
Remote Cables
p/n 5061-3378 Pin 35900 A/D Pin Agilent Signal Name Active
module (TTL)
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 289
11 Identifying Cables
CAN Cable
CAN Cable
p/n Description
5181-1516 CAN cable, Agilent module to module, 0.5 m
LAN Cables
p/n Description
5023-0203 Cross-over network cable, shielded, 3 m (for point to point connection)
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 290
11 Identifying Cables
Agilent Module to PC
Agilent Module to PC
p/n Description
RS232-61601 RS-232 cable, 2.5 m
Instrument to PC, 9-to-9 pin (female). This cable has special pin-out,
and is not compatible with connecting printers and plotters. It is also
called "Null Modem Cable" with full handshaking where the wiring is
made between pins 1-1, 2-3, 3-2, 4-6, 5-5, 6-4, 7-8, 8-7, 9-9.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 291
11 Identifying Cables
Agilent 1200 Module to Printer
To connect a USB Flash Drive use a USB OTG cable with Mini-B plug and A
socket.
p/n Description
5188-8050 USB A M-USB Mini B 3 m (PC-Module)
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 292
12 Hardware Information
This chapter describes the pump in more detail on hardware and electronics.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 293
12 Hardware Information
The Low-Pressure Pump Mainboard (LPM)
Block Diagrams
control signal
data/ address
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 294
12 Hardware Information
The Low-Pressure Pump Mainboard (LPM)
ASIC
Active inlet
Valve drive
valve
Current Motor
control control
1
Data/
Pump head Drive address
Motor J
Control
Encoder
Processor
Motor drive
Multiplexer
Leak sensor
Amplifier & J ambient measurement
Fan
Drive
Sensor
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 295
12 Hardware Information
The Low-Pressure Pump Mainboard (LPM)
Hardware +36 V
Motor J16
+24 V switch
Motor drive
Encoder J17
+/-15V
Processor
Inlet +36V
J19 Electronic
AIV Z-panel valvedrive fuse
MCGV +36 V
MCGV J30 Electronic Hardware
Z-panel fuse switch
driveC&D +24 V
+36 V
MCGV Electronic
driveA&B fuse
+24 V J25
Fan drive Fan
Active seal P7 Seal wash +24 V
wash driver J24
Damper
+/-15V Damper
converter
J3 analog
output
+36 V
LPM
+24 V power J9 Power
supply supply
+/-15V
+5V
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 296
12 Hardware Information
The Low-Pressure Pump Mainboard (LPM)
FUSION Board
The FUSION board holds the CORE-electronics for all Infinity II modules that use
the FUSION board (at introduction G71xx detectors and Multisamplers).
The FUSION board is the same for all Infinity II mainboards that host a FUSION
board.
FUSION board
Mainboard
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 297
12 Hardware Information
The Low-Pressure Pump Mainboard (LPM)
Features
• Configuration Switch (6 dip switches), see “Fusion Configuration Switch”.
• CAN (2 x) for inter module communication
• LAN (Local Area Network) , supports
• a PC with control software (e.g. Agilent ChemStation or Lab Advisor)
• USB (Universal Serial Bus) - replaces RS232, see “USB (Universal Serial Bus)”.
It supports:
• a PC with control software (e.g. Agilent ChemStation or Lab Advisor)
• USB Flash Disk (e.g. Agilent Feature Dongle).
• The USB connector provides the same functionally as the LAN interface. It
can be used by:
• Agilent Open Lab CDS (C.01.XX and later)
• Agilent Lab Advisor (B.02.06 and later)
• Firmware (Main and Resident)
• Module type will be read from the mainboard
Connectors
CAN
Configuration switch
LAN
USB
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 298
12 Hardware Information
The Low-Pressure Pump Mainboard (LPM)
Replacement Information
• The FUSION board is not part of the mainboard.
• It’s recommended to have one spare board in the service equipment. Can be
reused for all FUSION type modules.
• Can be replaced individually, see “Troubleshooting FUSION Board” (from other
module or as individual part)
• A new FUSION board contains just a resident firmware.
• During installation both, resident and main firmware (from the same firmware
set) should be installed.
• If the FUSION board is new or was used with a different module, it will turn the
module into resident mode; see “Replacement of Main and FUSION Board”.
• Depending on the Module Type (e.g. G7167 Multisampler) special actions are
required for initial operation (beside of setting type and serial number and
firmware update); see “Replacement of Main and FUSION Board”.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 299
12 Hardware Information
The Low-Pressure Pump Mainboard (LPM)
Description
This interface contains 8 individual programmable input/output pins. In addition
it provides 24 V power and 5 V power and a serial data line to detect and
recognize further add-ons that could be connected to this interface. This way the
interface can support various additional devices like sensors, triggers (in and out)
and small controllers, etc.
ERI
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 300
12 Hardware Information
The Low-Pressure Pump Mainboard (LPM)
Leak Sensor
The module has a PTC (Positive Temperature Coefficient) for the leak detection
and an NTC (Negative Temperature Coefficient) for the ambient temperature
compensation. A leak would cool down the PTC and this change in resistance
would generate a leak signal. The NTC is located on the power switch assembly.
Solvent incompatibility
C AU T I O N
The solvent DMF (dimethylformamide) leads to corrosion of the leak sensor.
The material of the leak sensor, PVDF (polyvinylidene fluoride), is incompatible
with DMF.
Do not use DMF as mobile phase.
Check the leak sensor regularly for corrosion.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 301
12 Hardware Information
Firmware Description
Firmware Description
Resident System
This resident section of the firmware is identical for all Agilent
1100/1200/1220/1260/1290 series modules. Its properties are:
• the complete communication capabilities (CAN, LAN, USB and RS- 232)
• memory management
• ability to update the firmware of the 'main system'
Main System
Its properties are:
• the complete communication capabilities (CAN, LAN, USB and RS- 232)
• memory management
• ability to update the firmware of the 'resident system'
In addition the main system comprises the instrument functions that are divided
into common functions like
• run synchronization through APG/ERI remote,
• error handling,
• diagnostic functions,
• or module specific functions like
• internal events such as lamp control, filter movements,
• raw data collection and conversion to absorbance.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 302
12 Hardware Information
Firmware Description
Firmware Updates
Firmware updates can be done with the Agilent Lab Advisor software with files
on the hard disk (latest version should be used).
Required tools, firmware and documentation are available from the Agilent web:
https://www.agilent.com/en-us/firmwareDownload?whid=69761
The file naming conventions are:
PPPP_RVVV_XXX.dlb, where
• PPPP is the product number, for example, 1315B for the G1315B DAD,
• R the firmware revision, for example, A for G1315B or B for the G1315C DAD,
• VVV is the revision number, for example 650 is revision 6.50,
• XXX is the build number of the firmware.
For instructions on firmware updates refer to section Replacing Firmware in
chapter "Maintenance" or use the documentation provided with the Firmware
Update Tools.
Update of main system can be done in the resident system only. Update of the
NOTE resident system can be done in the main system only.
Main and resident firmware must be from the same set.
Main FW update
Resident System Main System
Resident FW Update
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 303
12 Hardware Information
Firmware Description
The firmware update tools, firmware and documentation are available from the
Agilent web.
• https://www.agilent.com/en-us/firmwareDownload?whid=69761
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 304
12 Hardware Information
The Main Power Supply Assembly
The main power supply comprises a closed assembly and must not be
disassembled further for safety reasons. In case of a defect, the entire power
supply needs to be replaced.
The power supply provides all DC voltages used in the module. The line voltage
can vary in a range from 100 – 240 volts AC ± 10 % and needs no manual setting.
MPS
+36V (not used)
+36V
supply
+5V
switching +5V / 3A
regulator
To disconnect the instrument from line, unplug the power cord. The power
NOTE supply still uses some power, even if the power switch on the front panel is
turned off.
No accessible hardware fuse is needed because the main power supply is safe
against any short circuits or overload conditions on the output lines. When
overload conditions occur, the power supply turns off all output voltages. Turning
the line power off and on again resets the power supply to normal operation if the
cause of the overload condition has been removed.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 305
12 Hardware Information
The Main Power Supply Assembly
An over-temperature sensor in the main power supply is used to turn off output
voltages if the temperature exceeds the acceptable limit (for example, if the
cooling fan of the instrument fails). To reset the main power supply to normal
operating conditions, turn the instrument off, wait until it is approximately at
ambient temperature and turn the instrument on again.
The following table gives the specifications of the main power supply.
Table 26 Power Supply Specifications (Standard)
Pin 2 AGND
Pin 5 PGND
Pin 6 PGND
Pin 11 DGND
Pin 12 + 5 VDC
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 306
12 Hardware Information
Electrical Connections
Electrical Connections
Electrical Connections
• The CAN bus is a serial bus with high-speed data transfer. The two
connectors for the CAN bus are used for internal module data transfer and
synchronization.
• The ERI/REMOTE connector may be used in combination with other analytical
instruments from Agilent Technologies if you want to use features such as
start, stop, common shutdown, prepare, and so on.
• With the appropriate software, the LAN connector may be used to control the
module from a computer through a LAN connection. This connector is
activated and can be configured with the configuration switch.
• With the appropriate software, the USB connector may be used to control the
module from a computer through a USB connection.
• The power input socket accepts a line voltage of 100 – 240 VAC ± 10 % with a
line frequency of 50 or 60 Hz. Maximum power consumption varies by
module. There is no voltage selector on your module because the power
supply has wide-ranging capability. There are no externally accessible fuses
because automatic electronic fuses are implemented in the power supply.
Never use cables other than the ones supplied by Agilent Technologies to ensure
NOTE proper functionality and compliance with safety or EMC regulations.
Electric Shock
Electric shock due to insufficient insulation of connected instruments
WAR N IN G
Personal injury or damage to the instrument
Any other instruments connected to this instrument shall be approved to
a suitable safety standard and must include reinforced insulation from
the mains.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 307
12 Hardware Information
Electrical Connections
Configuration switch
USB-Mini-Port
LAN
ERI
CAN
Power socket
CCXZZ00000 Format
CC Country of manufacturing
• DE = Germany
• JP = Japan
• CN = China
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 308
12 Hardware Information
Interfaces
Interfaces
Pumps
Samplers
Detectors
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 309
12 Hardware Information
Interfaces
Fraction Collectors
Others
G1170A 2 No No No No No
G7122A No No No Yes No A
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 310
12 Hardware Information
Interfaces
Overview Interfaces
CAN
The CAN is inter-module communication interface. It is a 2-wire serial bus
system supporting high speed data communication and real-time requirement.
LAN
The modules have either an interface slot for a LAN card (e.g. Agilent G1369B/C
LAN Interface) or they have an on-board LAN interface (e.g. detectors G1315C/D
DAD and G1365C/D MWD). This interface allows the control of the
module/system via a PC with the appropriate control software. Some modules
have neither on-board LAN nor an interface slot for a LAN card (e.g. G1170A
Valve Drive or G4227A Flexible Cube). These are hosted modules and require a
Host module with firmware B.06.40 or later or with additional G1369C LAN Card.
USB
The USB interface replaces the RS-232 Serial interface in new FUSION generation
modules. For details on USB refer to “USB (Universal Serial Bus)” on page 316.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 311
12 Hardware Information
Interfaces
Remote (ERI)
The ERI (Enhanced Remote Interface) connector may be used in combination
with other analytical instruments from Agilent Technologies if you want to use
features as common shut down, prepare, and so on.
It allows easy connection between single instruments or systems to ensure
coordinated analysis with simple coupling requirements.
The subminiature D connector is used. The module provides one remote
connector which is inputs/outputs (wired- or technique).
To provide maximum safety within a distributed analysis system, one line is
dedicated to SHUT DOWN the system’s critical parts in case any module detects
a serious problem. To detect whether all participating modules are switched on
or properly powered, one line is defined to summarize the POWER ON state of all
connected modules. Control of analysis is maintained by signal readiness READY
for next analysis, followed by START of run and optional STOP of run triggered on
the respective lines. In addition PREPARE and START REQUEST may be issued.
The signal levels are defined as:
• standard TTL levels (0 V is logic true, + 5.0 V is false),
• fan-out is 10,
• input load is 2.2 kOhm against + 5.0 V, and
• output are open collector type, inputs/outputs (wired- or technique).
All common TTL circuits operate with a 5 V power supply. A TTL signal is defined
NOTE as "low" or L when between 0 V and 0.8 V and "high" or H when between 2.0 V
and 5.0 V (with respect to the ground terminal).
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 312
12 Hardware Information
Interfaces
1 START REQUEST (L) Request to start injection cycle (for example, by start key on any
module). Receiver is the autosampler.
2 STOP (L) Request to reach system ready state as soon as possible (for
example, stop run, abort or finish and stop injection). Receiver is any
module performing run-time controlled activities.
3 READY (H) System is ready for next analysis. Receiver is any sequence
controller.
4 POWER ON (H) All modules connected to system are switched on. Receiver is
any module relying on operation of others.
5 Not used
6 SHUT DOWN (L) System has serious problem (for example, leak: stops pump).
Receiver is any module capable to reduce safety risk.
Special Interfaces
There is no special interface for this module.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 313
12 Hardware Information
Interfaces
ERI Description
The ERI interface contains eight individual programmable input/output pins. In
addition, it provides 24 V power and 5 V power and a serial data line to detect and
recognize further add-ons that could be connected to this interface. This way the
interface can support various additional devices like sensors, triggers (in and out)
and small controllers, etc.
ERI
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 314
12 Hardware Information
Interfaces
1 IO 1 (START
REQUEST)
2 IO 2 (STOP)
3 IO 3 (READY)
4 IO 4 (POWER ON)
5 IO 5 (NOT USED)
6 IO 6 (SHUT DOWN)
7 IO 7 (START)
8 IO 8 (PREPARE)
9 1 wire DATA
10 DGND
11 +5 V ERI out
12 PGND
13 PGND
IO (Input/Output) Lines
• Eight generic bi-directional channels (input or output).
• Same as the APG Remote.
• Devices like valves, relays, ADCs, DACs, controllers can be
supported/controlled.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 315
12 Hardware Information
Interfaces
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 316
12 Hardware Information
Setting the 6-bit Configuration Switch
The 6-bit configuration switch is located at the rear of the module with FUSION
electronics. Switch settings provide configuration parameters for LAN and
instrument specific initialization procedures.
All modules with FUSION electronics:
• Default is ALL switches DOWN (best settings).
• Default IP address for LAN 192.168.254.11
• For specific LAN modes switches 4-5 must be set as required.
• For boot resident/cold start modes switches 1+2 or 6 must be UP.
Configuration switch
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 317
12 Hardware Information
Setting the 6-bit Configuration Switch
Mode Function/Setting
1 When selecting mode COM, settings are stored to non-volatile memory. When selecting mode Test, COM settings are taken from
non-volatile memory.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 318
12 Hardware Information
Setting the 6-bit Configuration Switch
Special Settings
Boot-Resident/Main
Firmware update procedures may require this mode in case of firmware loading
errors (main/resident firmware part).
If you use the following switch settings and power the instrument up again, the
instrument firmware stays in the resident/main mode. In resident mode, it is not
operable as a module. It only uses basic functions of the operating system for
example, for communication. In this mode the main firmware can be loaded
(using update utilities).
Loss of data
C AU T I O N
Forced cold start erases all methods and data stored in the non-volatile
memory. Exceptions are calibration settings, diagnosis and repair log books
which will not be erased.
Save your methods and data before executing a forced cold start.
If you use the following switch settings and power the instrument up again, it will
start as described above.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 319
12 Hardware Information
Setting the 6-bit Configuration Switch
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 320
12 Hardware Information
Instrument Layout
Instrument Layout
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 321
12 Hardware Information
Early Maintenance Feedback
EMF Counters
EMF counters increment with use and can be assigned a maximum limit which
provides visual feedback in the user interface when the limit is exceeded. Some
counters can be reset to zero after the required maintenance procedure.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 322
13 LAN Configuration
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 323
13 LAN Configuration
What You Have to Do First
This chapter is generic and may show figures that differ from your module. The
NOTE functionality is the same.
1 Note the MAC (Media Access Control) address for further reference. The MAC
or hardware address of the LAN interfaces is a world wide unique identifier.
No other network device will have the same hardware address. The MAC
address can be found on a label at the rear of the module underneath the
configuration switch (see Figure 53 on page 324).
Figure 52 MAC-Label
LAN interface
MAC label
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 324
13 LAN Configuration
TCP/IP parameter configuration
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 325
13 LAN Configuration
Configuration Switches
Configuration Switches
Configuration switch
The module is shipped with all switches set to OFF, as shown above.
To perform any LAN configuration, SW1 and SW2 must be set to OFF.
NOTE
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 326
13 LAN Configuration
Initialization Mode Selection
0 0 0 1 0 0 Use DHCP
Using Stored
When initialization mode Using Stored is selected, the parameters are taken from
the non-volatile memory of the module. The TCP/IP connection will be
established using these parameters. The parameters were configured previously
by one of the described methods.
Non-Volatile
RAM
Active
Parameter
Stored
Parameter
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 327
13 LAN Configuration
Initialization Mode Selection
Using Default
When Using Default is selected, the factory default parameters are taken instead.
These parameters enable a TCP/IP connection to the LAN interface without
further configuration, see Table 32 on page 328.
Default Active
Parameter Parameter
Using the default address in your local area network may result in network
NOTE problems. Take care and change it to a valid address immediately.
IP address: 192.168.254.11
Since the default IP address is a so-called local address, it will not be routed by
any network device. Thus, the PC and the module must reside in the same
subnet.
The user may open a Telnet session using the default IP address and change the
parameters stored in the non-volatile memory of the module. He may then close
the session, select the initialization mode Using Stored, power-on again and
establish the TCP/IP connection using the new parameters.
When the module is wired to the PC directly (e.g. using a cross-over cable or a
local hub), separated from the local area network, the user may simply keep the
default parameters to establish the TCP/IP connection.
In the Using Default mode, the parameters stored in the memory of the module
NOTE are not cleared automatically. If not changed by the user, they are still available,
when switching back to the mode Using Stored.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 328
13 LAN Configuration
Dynamic Host Configuration Protocol (DHCP)
DHCP Active
Server Parameter
1 It may take some time until the DHCP server has updated the DNS server with
NOTE the hostname information.
2 It may be necessary to fully qualify the hostname with the DNS suffix, e.g.
0030d3177321.country.company.com.
3 The DHCP server may reject the hostname proposed by the card and assign a
name following local naming conventions.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 329
13 LAN Configuration
Dynamic Host Configuration Protocol (DHCP)
Setup (DHCP)
The DHCP functionality is available on all Agilent HPLC modules with on-board LAN Interface or LAN
Interface Card G1369C, and “B”-firmware (B.06.40 or above) or modules with "D"-firmware. All
modules should use latest firmware from the same set.
1 Note the MAC address of the LAN interface (provided with G1369C LAN
Interface Card or mainboard). This MAC address is on a label on the card or at
the rear of the mainboard, for example, 0030d3177321.
On the Local Controller the MAC address can be found under Details in the
LAN section.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 330
13 LAN Configuration
Dynamic Host Configuration Protocol (DHCP)
2 Set the configuration switch to DHCP either on the G1369C LAN Interface
Card or the mainboard of above mentioned modules.
Table 33 G1369C LAN Interface Card (configuration switch on the card)
SW 4 SW 5 SW 6 SW 7 SW 8 Initialization Mode
Table 34 LC Modules with 8-bit configuration switch (B-firmware) (configuration switch at rear
of the instrument)
SW 6 SW 7 SW 8 Initialization Mode
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 331
13 LAN Configuration
Manual Configuration
Manual Configuration
Manual configuration only alters the set of parameters stored in the non-volatile
memory of the module. It never affects the currently active parameters.
Therefore, manual configuration can be done at any time. A power cycle is
mandatory to make the stored parameters become the active parameters, given
that the initialization mode selection switches are allowing it.
TELNET
Session
Non-Volatile
RAM
Stored
Parameter
Control
Module
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 332
13 LAN Configuration
Manual Configuration
With Telnet
Whenever a TCP/IP connection to the module is possible (TCP/IP parameters set
by any method), the parameters may be altered by opening a Telnet session.
1 Open the system (DOS) prompt window by clicking on Windows START
button and select “Run...”. Type “cmd” and press OK.
2 Type the following at the system (DOS) prompt:
• c:\>telnet <IP address> or
• c:\>telnet <host name>
where <IP address> may be the assigned address from a Bootp cycle, a
configuration session with the Handheld Controller, or the default IP address
(see “Configuration Switches” on page 326).
When the connection was established successfully, the module responds
with the following:
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 333
13 LAN Configuration
Manual Configuration
3 Type
? and press enter to see the available commands.
Value Description
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 334
13 LAN Configuration
Manual Configuration
5 Use the “/” and press Enter to list the current settings.
6 Change the IP address (in this example 192.168.254.12) and type “/” to list
current settings.
change of IP setting to
Initialization mode is Using Stored
If the Initialization Mode Switch is changed now to “Using Stored” mode, the
NOTE instrument will take the stored settings when the module is re-booted. In the
example above it would be 192.168.254.12.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 335
14 Appendix
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 336
14 Appendix
General Safety Information
Safety Standards
This is a Safety Class I instrument (provided with terminal for protective earthing)
and has been manufactured and tested according to international safety
standards.
General
Do not use this product in any manner not specified by the manufacturer. The
protective features of this product may be impaired if it is used in a manner not
specified in the operation instructions.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 337
14 Appendix
General Safety Information
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 338
14 Appendix
General Safety Information
In Case of Damage
Damage to the module
WAR N IN G
Personal injury (for example electrical shock, intoxication)
Instruments that appear damaged or defective should be made
inoperative and secured against unintended operation until they can be
repaired by qualified service personnel.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 339
14 Appendix
General Safety Information
Solvents
Toxic, flammable and hazardous solvents, samples and reagents
WAR N IN G
The handling of solvents, samples and reagents can hold health and safety
risks.
When working with these substances observe appropriate safety
procedures (for example by wearing goggles, safety gloves and
protective clothing) as described in the material handling and safety
data sheet supplied by the vendor, and follow good laboratory practice.
Do not use solvents with an auto-ignition temperature below 200 °C
(392 °F). Do not use solvents with a boiling point below 56 °C (133 °F).
Avoid high vapor concentrations. Keep the solvent temperature at least
40 °C (72 °F) below the boiling point of the solvent used. This includes
the solvent temperature in the sample compartment. For the solvents
methanol and ethanol keep the solvent temperature at least 25 °C
(45 °F) below the boiling point.
Do not operate the instrument in an explosive atmosphere.
Do not use solvents of ignition Class IIC according IEC 60079-20-1 (for
example, carbon disulfide).
Reduce the volume of substances to the minimum required for the
analysis.
Never exceed the maximum permissible volume of solvents (8 L) in the
solvent cabinet. Do not use bottles that exceed the maximum
permissible volume as specified in the usage guideline for solvent
cabinet.
Ground the waste container.
Regularly check the filling level of the waste container. The residual free
volume in the waste container must be large enough to collect the waste
liquid.
To achieve maximal safety, regularly check the tubing for correct
installation.
For details, see the usage guideline for the solvent cabinet. A printed copy of the
NOTE guideline has been shipped with the solvent cabinet, electronic copies are
available in the Agilent Information Center or via the Internet.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 340
14 Appendix
General Safety Information
Safety Symbols
Table 36 Symbols
The apparatus is marked with this symbol when the user shall refer to the
instruction manual in order to protect risk of harm to the operator and to
protect the apparatus against damage.
The apparatus is marked with this symbol when hot surfaces are available
and the user should not touch it when heated up.
Flammable Material
For Sample Thermostat which uses flammable refrigerant consult Agilent
Information Center / User Manual before attempting to install or service this
equipment. All safety precautions must be followed.
Manufacturing date.
Pacemaker
Magnets could affect the functioning of pacemakers and implanted heart
defibrillators.
A pacemaker could switch into test mode and cause illness. A heart
defibrillator may stop working. If you wear these devices keep at least
55 mm distance to magnets. Warn others who wear these devices from
getting too close to magnets.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 341
14 Appendix
General Safety Information
Table 36 Symbols
Magnetic field
Magnets produce a far-reaching, strong magnetic field. They could damage
TVs and laptops, computer hard drives, credit and ATM cards, data storage
media, mechanical watches, hearing aids and speakers. Keep magnets at
least 25 mm away from devices and objects that could be damaged by
strong magnetic fields.
A WARNING
WAR N IN G
alerts you to situations that could cause physical injury or death.
Do not proceed beyond a warning until you have fully understood and
met the indicated conditions.
A CAUTION
C AU T I O N
alerts you to situations that could cause loss of data, or damage of equipment.
Do not proceed beyond a caution until you have fully understood and met
the indicated conditions.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 342
14 Appendix
Waste Electrical and Electronic Equipment (WEEE) Directive
This product complies with the European WEEE Directive marking requirements.
The affixed label indicates that you must not discard this electrical/electronic
product in domestic household waste.
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 343
14 Appendix
Radio Interference
Radio Interference
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 344
14 Appendix
Sound Emission
Sound Emission
Sound pressure
Sound pressure Lp <70 db(A) according to DIN EN ISO 7779
Schalldruckpegel
Schalldruckpegel Lp <70 db(A) nach DIN EN ISO 7779
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 345
14 Appendix
Agilent Technologies on Internet
For the latest information on products and services visit our worldwide web site
on the Internet at:
https://www.agilent.com
InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 346
In This Book
www.agilent.com
Agilent Technologies Inc. 2016-2023
Edition: 07/2023
Document No: SD-29000179 Rev. D