0% found this document useful (0 votes)
2K views347 pages

G7110 G7111 G5654 Iso Quat Pumps - SerMa en SD 29000179

Uploaded by

infinity1200
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
2K views347 pages

G7110 G7111 G5654 Iso Quat Pumps - SerMa en SD 29000179

Uploaded by

infinity1200
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 347

Agilent InfinityLab LC Series

1260 Infinity II Isocratic- and Quaternary Pumps


Service Manual

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual
Notices
Document Information Warranty Safety Notices
Document No: SD-29000179 Rev. D The material contained in this document
Edition: 07/2023 is provided “as is,” and is subject to being C AU T I O N
changed, without notice, in future edi-
A CAUTION notice denotes a
Copyright tions. Further, to the maximum extent
permitted by applicable law, Agilent dis- hazard. It calls attention to an
© Agilent Technologies, Inc. 2016-2023 claims all warranties, either express or operating procedure, practice, or
No part of this manual may be repro- implied, with regard to this manual and the like that, if not correctly per-
duced in any form or by any means any information contained herein, includ- formed or adhered to, could result
(including electronic storage and retrieval ing but not limited to the implied warran- in damage to the product or loss
or translation into a foreign language) ties of merchantability and fitness for a
of important data. Do not proceed
without prior agreement and written con- particular purpose. Agilent shall not be lia-
sent from Agilent Technologies, Inc. as ble for errors or for incidental or conse- beyond a CAUTION notice until
governed by United States and interna- quential damages in connection with the the indicated conditions are fully
tional copyright laws. furnishing, use, or performance of this understood and met.
document or of any information con-
tained herein. Should Agilent and the user WAR N IN G
Agilent Technologies have a separate written agreement with
Hewlett-Packard-Strasse 8 warranty terms covering the material in A WARNING notice denotes a
76337 Waldbronn, Germany this document that conflict with these hazard. It calls attention to an
terms, the warranty terms in the separate operating procedure, practice, or
agreement shall control.
the like that, if not correctly per-
Technology Licenses formed or adhered to, could
result in personal injury or death.
The hardware and/or software described Do not proceed beyond a WARN-
in this document are furnished under a
ING notice until the indicated
license and may be used or copied only in
accordance with the terms of such conditions are fully understood
license. and met.

Restricted Rights Legend


U.S. Government Restricted Rights. Soft-
ware and technical data rights granted to
the federal government include only
those rights customarily provided to end
user customers. Agilent provides this
customary commercial license in Soft-
ware and technical data pursuant to FAR
12.211 (Technical Data) and 12.212
(Computer Software) and, for the Depart-
ment of Defense, DFARS 252.227-7015
(Technical Data - Commercial Items) and
DFARS 227.7202-3 (Rights in Commer-
cial Computer Software or Computer
Software Documentation).

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual
In This Guide...

This manual covers:


• Agilent 1260 Infinity II Isocratic Pump (G7110B),
• Agilent 1260 Infinity II Quaternary Pump VL (G7111A),
• Agilent 1260 Infinity II Quaternary Pump (G7111B), and
• Agilent 1260 Infinity II Bio-inert Pump (G5654A).

1 Introduction
This chapter gives an introduction to the module, instrument overview, and internal
connectors.

2 Site Requirements and Specifications


This chapter provides information on environmental requirements, physical, and
performance specifications.

3 Installing the Pump


This chapter gives information about the preferred stack setup for your system
and the installation of your module.

4 Using the Pump


This chapter provides information for optimized usage of the module.

5 Optimizing Performance
This chapter gives hints on how to optimize the performance or use additional
devices.

6 Troubleshooting and Diagnostics


This chapter gives an overview about the troubleshooting and diagnostic features
and the different user interfaces.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 3
7 Error Information
This chapter describes the meaning of error messages, and provides information
on probable causes and suggested actions how to recover from error conditions.

8 Test Functions and Calibration


This chapter describes the tests for the module.

9 Maintenance and Repair


This chapter describes the maintenance and repair of the module.

10 Parts for Maintenance and Repair


This chapter provides information on parts for maintenance and repair.

11 Identifying Cables
This chapter provides information on cables used with the Agilent 1200 Infinity
Series modules.

12 Hardware Information
This chapter describes the pump in more detail on hardware and electronics.

13 LAN Configuration
This chapter provides information on connecting the module to the Agilent
ChemStation PC.

14 Appendix
This chapter provides addition information on safety, legal, and web.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 4
Contents

1 Introduction 10
Introduction to the Pump 11
G7110B Isocratic Pump 12
G7111A Quaternary Pump VL 14
G7111B Quaternary Pump 16
G5654A Bio-inert Pump 18
Overview of the Hydraulic Path 20
Leak and Waste Handling 27

2 Site Requirements and Specifications 32


Site Requirements 33
Physical Specifications 36
Performance Specifications 40

3 Installing the Pump 47


Unpacking the Pump 48
Optimizing the Stack Configuration 51
Installing the Pump 59
Connecting Modules and Control Software 62
Flow Connections of the Pump 63
Priming the System 64

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 5
4 Using the Pump 69
Magnets 70
Turn on/off 71
Status Indicators 73
Best Practices 74
Leak and Waste Handling 76
Hints for Successful Use of the Pump 78
Setting up the Pump with the G4208A Instant Pilot 79
Setting up the Pump with the Instrument Control Interface 80
Solvent Information 88
Algae Growth in HPLC Systems 98
Prevent Blocking of Solvent Filters 99
Normal Phase Applications 100

5 Optimizing Performance 102


Using the Degasser 103
Operational Hints for the Vacuum Degasser 104
Operational Hints for the Multi Channel Gradient Valve (MCGV) 105
When to use the Seal Wash Function 106
Choosing the Right Pump Seals 107
Optimize the Compressibility Compensation Setting 108

6 Troubleshooting and Diagnostics 110


User Interfaces 111
Agilent Lab Advisor Software 112

7 Error Information 113


What Are Error Messages 114
General Error Messages 115
Module Error Messages 123

8 Test Functions and Calibration 135


Introduction 136
System Pressure Test 137
Leak Rate Test 142

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 6
9 Maintenance and Repair 147
Introduction to Maintenance and Repair 149
Cautions and Warnings 150
Overview of Maintenance and Repair 152
Cleaning the Module 153
Remove and Install Doors 154
Checking and Replacing the Solvent Filter 156
Exchange the Passive Inlet Valve 157
Exchange the Outlet Valve 159
Exchange the Purge Valve Frit or the Purge Valve 161
Replace the O-Ring on the Purge Valve 164
Remove the Pump Head Assembly 166
Maintenance of a Pump Head Without Seal Wash Option 168
Maintenance of a Pump Head with Seal Wash Option 171
Reinstall the Pump Head Assembly 176
Seal Wear-in Procedure 179
Exchange the Multi-Channel Gradient Valve (MCGV) 180
Replace the Solvent Selection Valve (Isocratic Pump) 183
Exchange the Active Inlet Valve (AIV) or its Cartridge 185
Exchange the Seal Wash Cartridge 187
Replace Leak Handling System Parts 188
Replacing the Module Firmware 189
Introduction to Repair 190
Overview of Repairing the Pump 191
Exchanging Internal Parts 193
Remove the Top Cover and Foam 194
Exchange the Pump Mainboard 196
Replace the Fusion Board 202
Changing the Type and Serial Number 203
Exchange the Damper 208
Exchange the Pump Drive 212
Exchange the Fan 214
Replace Power Supply and Cables 216
Replace the Power Switch 219
Replace the Degasser 221

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 7
Replace Parts Inside the 4-Channel-Degasser 223
Replace Parts Inside the 2-Channel-Degasser 230
Replace the Leak Panel 236
Replace the Leak Sensor 238
Replace the Status LED 239
Install an Active Seal Wash Upgrade Kit (Quaternary Pump) 241
Assemble the Main Cover 244
Install the Top Cover and Foam 245

10 Parts for Maintenance and Repair 246


Accessory Kits 247
Pump Head Assembly Without Seal Wash 249
Pump Head Assembly with Seal Wash Option 251
Outlet Valve 254
Passive Inlet Valve 255
Purge Valve Assembly 256
Active Inlet Valve Assembly 257
1260 Infinity II Max Uptime Kit 259
1260 Infinity II Max Uptime Kit Bio 260
1260 Infinity II Low Dispersion Kit MCT 261
HPLC System Tool Kit 262
Solvent Cabinet 263
Bottle Head Assembly 264
Hydraulic Path of the Quaternary Pump 265
Hydraulic Path of the Isocratic Pump 267
Overview of Main Assemblies 268
Degasser 274
Cover Parts 276
Sheet Metal Kit 277
Foam Parts 278
Power, Status LED and Illumination LED 279
Leak Parts 280

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 8
11 Identifying Cables 281
Cable Overview 282
Analog Cables 284
Remote Cables 286
CAN Cable 290
Agilent Module to PC 291
Agilent 1200 Module to Printer 292

12 Hardware Information 293


The Low-Pressure Pump Mainboard (LPM) 294
Firmware Description 302
The Main Power Supply Assembly 305
Electrical Connections 307
Interfaces 309
Setting the 6-bit Configuration Switch 317
Instrument Layout 321
Early Maintenance Feedback 322

13 LAN Configuration 323


What You Have to Do First 324
TCP/IP parameter configuration 325
Configuration Switches 326
Initialization Mode Selection 327
Dynamic Host Configuration Protocol (DHCP) 329
Manual Configuration 332

14 Appendix 336
General Safety Information 337
Waste Electrical and Electronic Equipment (WEEE) Directive 343
Radio Interference 344
Sound Emission 345
Agilent Technologies on Internet 346

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 9
1 Introduction

Introduction to the Pump 11


G7110B Isocratic Pump 12
Product Description 12
Features 13
G7111A Quaternary Pump VL 14
Product Description 14
Features 15
G7111B Quaternary Pump 16
Product Description 16
Features 17
G5654A Bio-inert Pump 18
Product Description 18
Features 19
Overview of the Hydraulic Path 20
Hydraulic Path 21
How Does the Pump Work? 23
How Does Compressibility Compensation Work? 26
How Does Variable Stroke Volume Work? 26
Leak and Waste Handling 27
Leak Sensor 30
Waste Concept 31

This chapter gives an introduction to the module, instrument overview, and


internal connectors.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 10
1 Introduction
Introduction to the Pump

Introduction to the Pump

The Pumps comprise one pump head unit with an optional solvent cabinet, a
vacuum degasser (optional for the isocratic pump), and a multi-channel gradient
valve (MCGV) for quaternary pumps, or an optional solvent selection valve (SSV)
for isocratic pumps. The combination of the pump head assembly and
high-speed proportioning MCGV with a possible connection up to four different
solvents in quaternary pumps provides gradient generation by low pressure
mixing. The combination of the pump head assembly and SSV in isocratic pump
provides opportunity to make a choice between three different solvents. A
solvent cabinet provides enough space for eight one-liter bottles. The Active Seal
Wash (ASW) and Active Inlet Valve (AIV) options are available for quaternary
pumps when concentrated buffer solutions are used. The Bio-inert pump is
optimized for these conditions, thus is always equipped with ASW and AIV.

Degasser

Outlet Valve

Purge Valve

Inlet Valve

MCGV

Pump Head

Figure 1 Overview of the pump

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 11
1 Introduction
G7110B Isocratic Pump

G7110B Isocratic Pump

Product Description
The Agilent 1260 Infinity II Isocratic Pump is ideal for demanding QA/QC tasks
and routine applications. It maintains virtually pulse-free and stable solvent flows.
The dual floating, precise servo-controlled pistons in the delivery mechanism
adjust the stroke volume according to your chosen flow rate.
The optional integrated degasser and solvent selection valve offers increased
ease-of-use and method flexibility.
The pump can deliver a broad range of pressures up to a maximum 600 bar,
giving you the flexibility to use small particle size columns, longer columns or
alternative higher viscosity solvents.
The Agilent 1260 Infinity II Isocratic Pump is the ideal pump for GPC/SEC
applications where run-to-run and day-to-day precision in retention times is
crucial.

Status LED

Purge valve

Power switch

Leak drain

Figure 2 Overview of the pump

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 12
1 Introduction
G7110B Isocratic Pump

Features

• Higher resolution and increased throughput with the new standard 600 bar
pressure for modern LC columns.
• Low operating and maintenance costs - due to excellent instrument uptime.
• Future proof - your investment is protected since you can upgrade to a
quaternary pump if required.
• Versatile wide flow range – from 0.2 mL/min to 10 mL/min for analytical and
semi-preparative separations and purifications.
• Fast and easy maintenance - front access to all maintenance parts means you
can change components quickly without having to disassemble the
instrument.
• Longevity and low operating costs - robust materials (stainless steel, titanium,
gold, ruby, sapphire, ceramics, PEEK and PTFE) withstand the most
demanding applications and keep your pump functioning for years. Our
optional seal wash for high-salt mobile phases keeps maintenance to a
minimum and reduces operating costs.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 13
1 Introduction
G7111A Quaternary Pump VL

G7111A Quaternary Pump VL

Product Description
The Agilent 1260 Infinity II Quaternary Pump VL has an extended power range,
delivering pressures up to 400 bar. At these high pressures you can use smaller
particle size columns and get higher resolution and faster separations. It
maintains virtually pulse-free, well mixed and stable solvent flows. Its dual
floating, precise servo-controlled pistons adjust the stroke volume according to
your chosen flow rate.
Offering access to up to four solvents, the Agilent 1260 Infinity II Quaternary
Pump VL provides the greatest flexibility in automated solvent blending and is
recommended for a wide range of research and routine applications (especially
in the food, environmental and pharmaceutical sectors) as well as for method
development.

Status indicator

Degasser

Purge valve

MCGV

Power switch

Leak drain
Figure 3 Overview of the pump

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 14
1 Introduction
G7111A Quaternary Pump VL

Features
• Higher resolution, faster separations - the maximum pressure of 400 bar
allows you to use smaller particle size columns, longer columns and/or higher
viscous solvents as an alternative to acetonitrile.
• Analytical and semi-preparative purifications - the wide flow range (up to
10 mL/min) perfectly supports standard and semi-preparative applications.
The perfect choice for 4.6 mm and 3.0 mm id columns.
• Quicker experiments - the pump offers convenient access to up to four
solvents for isocratic or gradient analysis which can really speed up method
development, preparation of mobile phases and system flushing.
• Easy, fast and secure transfer of methods from HPLC to sub-2-micron
columns.
• Integrated degasser
• Longevity and low operating costs - robust materials (stainless steel, titanium,
gold, ruby, sapphire, ceramics, PEEK and PTFE) withstand the most
demanding applications and keep your pump functioning for years. Our
optional seal wash for high-salt mobile phases keeps maintenance to a
minimum and reduces operating costs.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 15
1 Introduction
G7111B Quaternary Pump

G7111B Quaternary Pump

Product Description
The Agilent 1260 Infinity II Quaternary Pump has an extended power range,
delivering pressures up to 600 bar. At these high pressures you can use smaller
particle size columns and get higher resolution and faster separations. It
maintains virtually pulse-free, well mixed and stable solvent flows. Its dual
floating, precise servo-controlled pistons adjust the stroke volume according to
your chosen flow rate.
Offering access to up to four solvents, the Agilent 1260 Infinity II Quaternary
Pump provides the greatest flexibility in automated solvent blending and is
recommended for a wide range of research and routine applications (especially
in the food, environmental and pharmaceutical sectors) as well as for method
development.

Status indicator

Degasser

Purge valve

MCGV

Power switch

Leak drain
Figure 4 Overview of the pump

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 16
1 Introduction
G7111B Quaternary Pump

Features
• Higher resolution, faster separations - the maximum pressure of 600 bar
allows you to use smaller particle size columns, longer columns and/or higher
viscous solvents as an alternative to acetonitrile.
• Analytical and semi-preparative purifications - the wide flow range (up to
10 mL/min) perfectly supports standard and semi-preparative applications.
The perfect choice for 4.6 mm and 3.0 mm id columns.
• Quicker experiments - the pump offers convenient access to up to four
solvents for isocratic or gradient analysis which can really speed up method
development, preparation of mobile phases and system flushing.
• Easy, fast and secure transfer of methods from HPLC to sub-2-micron
columns.
• Integrated degasser
• Longevity and low operating costs - robust materials (stainless steel, titanium,
gold, ruby, sapphire, ceramics, PEEK and PTFE) withstand the most
demanding applications and keep your pump functioning for years. Our
optional seal wash for high-salt mobile phases keeps maintenance to a
minimum and reduces operating costs.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 17
1 Introduction
G5654A Bio-inert Pump

G5654A Bio-inert Pump

Product Description
The 1260 Infinity II Bio-Inert Pump is the pump of choice for all your biological
and extreme pH applications. The titanium-based pump offers highest corrosion
resistance against high salt concentration (2 M) and offers a handling of a wide
pH range (1 – 13, short term 14). It features a pressure range of up to 600 bar
and a flow rate up to 10 mL/min (at 200 bar). Which enables the use of almost
any column: conventional, sub-2 µm particle, or superficially porous columns.
Together with the Agilent Bio-HPLC column portfolio for SEC and IEX, the highest
resolution per time is achieved for protein and NBE characterization. The Agilent
Buffer Advisor software allows fast pH scouting and easy buffer/solvent
preparation in ion exchange chromatography.

Status indicator

Degasser

Purge valve

MCGV

Power switch

Leak drain
Figure 5 Overview of the Bio-inert Pump

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 18
1 Introduction
G5654A Bio-inert Pump

Features
• Titanium pump - for highest corrosion resistance against high salt
concentrations and harsh cleaning procedures
• Steel and iron-free wetted parts in solvent delivery unit - ensure the integrity of
bio-molecules and minimize unwanted surface interactions while increasing
column life-time
• High salt tolerance (2 M) and wide pH range (1 – 13, short term 14) - for
increased flexibility
• Flow rates up to 10 mL/min - enable power ranges from lowest pressure for
traditional bio-purification columns up to high pressure STM analytical
bio-columns
• Active seal wash and quaternary solvent blending included
• Comes with Agilent Buffer Advisor software - for faster pH scouting and easy
buffer/solvent preparation in ion exchange chromatography

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 19
1 Introduction
Overview of the Hydraulic Path

Overview of the Hydraulic Path

The Low Pressure Mixing Pumps are based on a two-channel, dual-piston


in-series design which comprises all essential functions that a solvent delivery
system has to fulfill. Metering of solvent and delivery to the high-pressure side
are performed by one pump assembly which can generate pressure up to
600 bar.
In the quaternary pumps, degassing of the solvents is done in a built-in vacuum
degasser. Solvent compositions are generated on the low-pressure side by an
MCGV.
The pump assembly includes a pump head with an inlet valve and an outlet valve.
A damping unit is connected between the two piston chambers. A purge valve
including a PTFE frit is fitted at the pump outlet for convenient priming of the
pump head.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 20
1 Introduction
Overview of the Hydraulic Path

Hydraulic Path
to sampling unit and column

Seal wash
pump Damper

Purge valve
Outlet
Inlet valve valve
to waste

Figure 6 Hydraulic path of the isocratic pump

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 21
1 Introduction
Overview of the Hydraulic Path

to sampling unit and column

Seal wash
pump Damper
Degasser

Purge valve
Outlet
Inlet valve valve
to waste

MCGV

Figure 7 Hydraulic Path of the Quaternary Pump and Bio-inert Pump

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 22
1 Introduction
Overview of the Hydraulic Path

How Does the Pump Work?


In the quaternary pump, the liquid runs from the solvent reservoir through the
degasser to the MCGV and from there to the inlet valve.
For the isocratic pump, the solvent bottle is directly connected to the inlet valve.
The pump assembly comprises two substantially identical piston/chamber units.
Both piston/chamber units comprise a ball-screw drive and a pump head with
one sapphire piston for reciprocating movement in it.
A servo-controlled variable reluctance motor drives the two ball screw drives in
opposite directions. The gears for the ball-screw drives have different
circumferences (ratio 2:1) allowing the first piston to move at twice the speed of
the second piston. The solvent enters the pump head close to the bottom limit
and leaves it at its top. The outer diameter of the piston is smaller than the inner
diameter of the pump head chamber allowing the solvent to fill the gap in
between. The first piston has a stroke volume in the range of 20 – 100 µL
depending on the flow rate. The microprocessor controls all flow rates in a range
of 1 µL/min – 10 mL/min. The inlet of the first pumping unit is connected to the
inlet valve.
The outlet of the first piston/chamber unit is connected through the outlet valve
and the damping unit to the inlet of the second piston/chamber unit. The outlet of
the purge valve assembly is then connected to the following chromatographic
system.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 23
1 Introduction
Overview of the Hydraulic Path

Damper

Pump chamber 1 Pump chamber 2


Purge valve
To column
Outlet
valve
Inlet valve
To waste
From solvent bottle/
degasser/SSV or MCGV Seal

Piston 1
Piston 2

Ball screw drive

Gear

Motor with encoder

Figure 8 Principle of the Pump

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 24
1 Introduction
Overview of the Hydraulic Path

When turned on, the pump runs through an initialization procedure to determine
the upper dead position of the first piston. The first piston moves slowly upwards
into the mechanical stop of the pump chamber and from there it moves back for
a defined distance. The controller stores this piston position in memory. After
this initialization the pump starts operation with the set parameters. The inlet
valve opens and the down-moving piston draws solvent into the first pump
chamber. At the same time, the second piston moves upwards delivering to the
system. After a controller-defined stroke length that depends on the flow rate the
drive motor is stopped and the inlet valve closes. The motor direction is reversed
and moves the first piston up until it reaches the stored upper limit and at the
same time the second piston moves downwards. Then the sequence starts
again moving the pistons up and down between the two limits. During the up
movement of the first piston, the solvent in the pump chamber is pressed
through the outlet valve into the second pump chamber. The second piston
draws in half of the volume displaced by the first piston and the remaining half
volume is directly delivered to the system. During the drawing stroke of the first
piston, the second piston delivers the drawn volume to the system.
Quaternary pumps: For solvent compositions from the solvent bottles A, B, C, D
the controller divides the length of the intake stroke in certain fractions in which
the gradient valve connects the specified solvent channel to the pump input.
For specifications, see:
• isocratic pump: “Performance Specifications Agilent 1260 Infinity II Isocratic
Pump (G7110B)” on page 40
• quaternary pumps:
• “Performance Specifications Agilent 1260 Infinity II Quaternary Pump
(G7111B)” on page 44
• “Performance Specifications Agilent 1260 Infinity II Quaternary Pump VL
(G7111A)” on page 42
• Bio-inert pump: “Performance Specifications Agilent 1260 Infinity II Bio-inert
Pump (G5654A)” on page 46

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 25
1 Introduction
Overview of the Hydraulic Path

How Does Compressibility Compensation Work?


The compressibility of the solvents in use will affect retention-time stability when
the backpressure in the system changes (for example, aging of column). To
minimize this effect, the pump provides a compressibility compensation feature,
which optimizes the flow stability according to the solvent type. The
compressibility compensation is set to a default value and can be changed
through the user interface.
Without a compressibility compensation, the following will happen during a
stroke of the first piston. The pressure in the piston chamber increases and the
volume in the chamber will be compressed depending on backpressure and
solvent type. The volume displaced into the system will be reduced by the
compressed volume.
With a compressibility value set, the processor calculates a compensation
volume that depends on the backpressure of the system and the selected
compressibility. This compensation volume will be added to the normal stroke
volume and compensates the previously described loss of volume during the
delivery stroke of the first piston.

How Does Variable Stroke Volume Work?


Due to the compression of the pump-chamber volume each piston stroke of the
pump will generate a small pressure pulsation, influencing the flow stability of the
pump. The amplitude of the pressure pulsation depends mainly on the stroke
volume and the compressibility compensation for the solvent in use. Small stroke
volumes generate pressure pulsations of smaller amplitude than higher stroke
volumes at the same flow rate. In addition, the frequency of the pressure
pulsations is higher. This decreases the influence of flow pulsations on
quantitative results.
In gradient mode, smaller stroke volumes result in a lower flow ripple and will
improve composition ripple.
The module uses a processor-controlled spindle system for driving its pistons.
The normal stroke volume is optimized for the selected flow rate. Small flow
rates use a small stroke volume while higher flow rates use a higher stroke
volume.
By default, the stroke volume for the pump is set to AUTO mode. This means that
the stroke is optimized for the flow rate in use. A change to larger stroke volumes
is possible but not recommended.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 26
1 Introduction
Leak and Waste Handling

Leak and Waste Handling

The Agilent InfinityLab LC Series has been designed for safe leak and waste
handling. It is important that all security concepts are understood and
instructions are carefully followed.
The solvent cabinet is designed to store a maximum volume of 8 L solvent. The
maximum volume for an individual bottle stored in the solvent cabinet should not
exceed 2 L. For details, see the usage guideline for the Agilent Infinity II Solvent
Cabinets (a printed copy of the guideline has been shipped with the solvent
cabinet, electronic copies are available on the Internet).
All leak plane outlets are situated in a consistent position so that all Infinity and
Infinity II modules can be stacked on top of each other. Waste tubes are guided
through a channel on the right hand side of the instrument, keeping the front
access clear from tubes.
The leak plane provides leak management by catching all internal liquid leaks,
guiding them to the leak sensor for leak detection, and passing them on to the
next module below, if the leak sensor fails. The leak sensor in the leak plane
stops the running system as soon as the leak detection level is reached.
Solvent and condensate is guided through the waste channel into the waste
container:
• from the detector's flow cell outlet
• from the Multisampler needle wash port
• from the Sample Cooler or Sample Thermostat (condensate)
• from the pump's Seal Wash Sensor (if applicable)
• from the pump's Purge Valve or Multipurpose Valve

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 27
1 Introduction
Leak and Waste Handling

Figure 9 Infinity II Leak Waste Concept (Flex Bench installation)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 28
1 Introduction
Leak and Waste Handling

Figure 10 Infinity II Single Stack Leak Waste Concept (bench installation)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 29
1 Introduction
Leak and Waste Handling

Figure 11 Infinity II Two Stack Leak Waste Concept (bench installation)

The waste tube connected to the leak plane outlet on each of the bottom
instruments guides the solvent to a suitable waste container.

Leak Sensor
Solvent incompatibility
C AU T I O N
The solvent DMF (dimethylformamide) leads to corrosion of the leak sensor.
The material of the leak sensor, PVDF (polyvinylidene fluoride), is incompatible
with DMF.
 Do not use DMF as mobile phase.
 Check the leak sensor regularly for corrosion.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 30
1 Introduction
Leak and Waste Handling

Waste Concept
1 Agilent recommends using the 6 L waste can with 1 Stay Safe cap GL45 with
4 ports (5043-1221) for optimal and safe waste disposal. If you decide to use
your own waste solution, make sure that the tubes don't immerse in the liquid.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 31
2 Site Requirements and Specifications

Site Requirements 33
Physical Specifications 36
Physical Specifications Agilent 1260 Infinity II Isocratic Pump (G7110B) 36
Physical Specifications Agilent 1260 Infinity II Quaternary Pump VL (G7111A) 37
Physical Specifications Agilent 1260 Infinity II Quaternary Pump (G7111B) 38
Physical Specifications Agilent 1260 Infinity II Bio-inert Pump (G5654A) 39
Performance Specifications 40
Performance Specifications Agilent 1260 Infinity II Isocratic Pump (G7110B) 40
Performance Specifications Agilent 1260 Infinity II Quaternary Pump VL (G7111A) 42
Performance Specifications Agilent 1260 Infinity II Quaternary Pump (G7111B) 44
Performance Specifications Agilent 1260 Infinity II Bio-inert Pump (G5654A) 46

This chapter provides information on environmental requirements, physical, and


performance specifications.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 32
2 Site Requirements and Specifications
Site Requirements

Site Requirements

A suitable environment is important to ensure optimal performance of the


instrument.

Power Considerations
The module power supply has wide ranging capability. It accepts any line voltage
in the range described in Table 1 on page 36. Consequently there is no voltage
selector in the rear of the module. There are also no externally accessible fuses,
because automatic electronic fuses are implemented in the power supply.

Incorrect line voltage at the module


WAR N IN G
Shock hazard or damage of your instrument can result if the devices are
connected to line voltage higher than specified.
 Connect your module to the specified line voltage.

Module is partially energized when switched off, as long as the power cord
WAR N IN G
is plugged in.
Repair work at the module can lead to personal injuries, e.g. shock hazard,
when the cover is opened and the module is connected to power.
 Make sure that it is always possible to access the power plug.
 Remove the power cable from the instrument before opening the cover.
 Do not connect the power cable to the Instrument while the covers are
removed.

Inaccessible power plug.


WAR N IN G
In case of emergency it must be possible to disconnect the instrument
from the power line at any time.
 Make sure the power connector of the instrument can be easily reached
and unplugged.
 Provide sufficient space behind the power socket of the instrument to
unplug the cable.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 33
2 Site Requirements and Specifications
Site Requirements

Power Cords
Country-specific power cords are available for the module. The female end of all
power cords is identical. It plugs into the power-input socket at the rear. The male
end of each power cord is different and designed to match the wall socket of a
particular country or region.
Agilent makes sure that your instrument is shipped with the power cord that is
suitable for your particular country or region.

Unintended use of power cords


WAR N IN G
Using power cords for unintended purposes can lead to personal injury or
damage of electronic equipment.
 Never use a power cord other than the one that Agilent shipped with this
instrument.
 Never use the power cords that Agilent Technologies supplies with this
instrument for any other equipment.
 Never use cables other than the ones supplied by Agilent Technologies
to ensure proper functionality and compliance with safety or EMC
regulations.

Absence of ground connection


WAR N IN G
The absence of ground connection can lead to electric shock or short
circuit.
 Never operate your instrumentation from a power outlet that has no
ground connection.

Electrical shock hazard


WAR N IN G
Solvents may damage electrical cables.
 Prevent electrical cables from getting in contact with solvents.
 Exchange electrical cables after contact with solvents.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 34
2 Site Requirements and Specifications
Site Requirements

Bench Space
The module dimensions and weight (see Table 1 on page 36) allow you to place
the module on almost any desk or laboratory bench. It needs an additional
2.5 cm (1.0 inches) of space on either side and approximately 8 cm (3.1 inches)
in the rear for air circulation and electric connections.
If the bench shall carry a complete HPLC system, make sure that the bench is
designed to bear the weight of all modules.
The module should be operated in a horizontal position.

Agilent recommends that you install the HPLC instrument in the InfinityLab Flex
NOTE Bench rack. This option helps to save bench space as all modules can be placed
into one single stack. It also allows to easily relocate the instrument to another
lab.

Condensation
Condensation within the module
C AU T I O N
Condensation can damage the system electronics.
 Do not store, ship or use your module under conditions where temperature
fluctuations could cause condensation within the module.
 If your module was shipped in cold weather, leave it in its box and allow it to
warm slowly to room temperature to avoid condensation.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 35
2 Site Requirements and Specifications
Physical Specifications

Physical Specifications

Physical Specifications Agilent 1260 Infinity II


Isocratic Pump (G7110B)
Table 1 Physical Specifications G7110B

Type Specification Comments

Weight 14.5 kg (32 lbs)

Dimensions 180 x 396 x 436 mm


(height × width × depth) (7.1 x 15.6 x 17.2 inches)

Line voltage 100 – 240 V~, ± 10 % Wide-ranging capability

Line frequency 50 or 60 Hz, ± 5 %

Power consumption 80 VA, 65 W

Ambient operating 4–55 °C (39–131 °F)


temperature

Ambient non-operating -40 – 70 °C (-40 – 158 °F)


temperature

Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing

Operating altitude Up to 3000 m (9842 ft)

Safety standards: Overvoltage category II, For indoor use only


IEC, EN, CSA, UL Pollution degree 2

ISM Classification ISM Group 1 Class B According to CISPR 11

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 36
2 Site Requirements and Specifications
Physical Specifications

Physical Specifications Agilent 1260 Infinity II


Quaternary Pump VL (G7111A)
Table 2 Physical Specifications G7111A

Type Specification Comments

Weight 14.5 kg (32 lbs)

Dimensions 180 x 396 x 436 mm


(height × width × depth) (7.1 x 15.6 x 17.2 inches)

Line voltage 100 – 240 V~, ± 10 % Wide-ranging capability

Line frequency 50 or 60 Hz, ± 5 %

Power consumption 80 VA, 65 W

Ambient operating 4–55 °C (39–131 °F)


temperature

Ambient non-operating -40 – 70 °C (-40 – 158 °F)


temperature

Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing

Operating altitude Up to 3000 m (9842 ft)

Safety standards: Overvoltage category II, For indoor use only


IEC, EN, CSA, UL Pollution degree 2

ISM Classification ISM Group 1 Class B According to CISPR 11

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 37
2 Site Requirements and Specifications
Physical Specifications

Physical Specifications Agilent 1260 Infinity II


Quaternary Pump (G7111B)
Table 3 Physical Specifications G7111B

Type Specification Comments

Weight 14.5 kg (32 lbs)

Dimensions 180 x 396 x 436 mm


(height × width × depth) (7.1 x 15.6 x 17.2 inches)

Line voltage 100 – 240 V~, ± 10 % Wide-ranging capability

Line frequency 50 or 60 Hz, ± 5 %

Power consumption 80 VA, 65 W

Ambient operating 4–55 °C (39–131 °F)


temperature

Ambient non-operating -40 – 70 °C (-40 – 158 °F)


temperature

Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing

Operating altitude Up to 3000 m (9842 ft)

Safety standards: Overvoltage category II, For indoor use only


IEC, EN, CSA, UL Pollution degree 2

ISM Classification ISM Group 1 Class B According to CISPR 11

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 38
2 Site Requirements and Specifications
Physical Specifications

Physical Specifications Agilent 1260 Infinity II


Bio-inert Pump (G5654A)
Table 4 Physical Specifications G5654A

Type Specification Comments

Weight 14.7 kg (32 lbs)

Dimensions 180 x 396 x 436 mm


(height × width × depth) (7.1 x 15.6 x 17.2 inches)

Line voltage 100 – 240 V~, ± 10 % Wide-ranging capability

Line frequency 50 or 60 Hz, ± 5 %

Power consumption 80 VA, 65 W

Ambient operating 4–55 °C (39–131 °F)


temperature

Ambient non-operating -40 – 70 °C (-40 – 158 °F)


temperature

Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing

Operating altitude Up to 3000 m (9842 ft)

Non-operating altitude Up to 4600 m (15092 ft) For storing the module

Safety standards: Overvoltage category II, For indoor use only


IEC, EN, CSA, UL Pollution degree 2

ISM Classification ISM Group 1 Class B According to CISPR 11

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 39
2 Site Requirements and Specifications
Performance Specifications

Performance Specifications

Performance Specifications Agilent 1260 Infinity II


Isocratic Pump (G7110B)
Table 5 Performance Specifications G7110B

Type Specification Comments

Hydraulic system Dual piston in series pump with


servo-controlled variable stroke
drive, power transmission by gears
and ball screws, floating pistons

Flow range Settable: 0.001 – 10 mL/min Set points in 0.001 mL/min


Recommended: 0.2 – 10.0 mL/min increments
For use with flow rates below
500 μL/min a vacuum
degasser is required.

Flow precision ≤0.07 % RSD, or <0.02 min SD, Based on retention time at
whichever is greater constant temperature

Flow accuracy ± 1 % or 10 µL/min, whichever is Pumping degassed H2O at


greater 10 MPa (100 bar, 1450 psi)

Pressure operating range Up to 60 MPa (600 bar, 8702 psi)


up to 5 mL/min
Up to 20 MPa (200 bar, 2901 psi)
up to 10 mL/min

Pressure pulsation < 2 % amplitude (typically < 1.3 %), At 1 mL/min isopropanol, at all
or < 0.3 MPa (3 bar, 44 psi), pressures > 1 MPa
whichever is greater (10 bar, 145 psi)

Compressibility User-selectable, based on mobile


compensation phase compressibility

Recommended pH range 1.0 – 12.5 Solvents with pH < 2.3 should


not contain acids which attack
stainless steel

Integrated degassing unit Optional


Number of channels: 2
Internal volume per channel: 1.5 mL

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 40
2 Site Requirements and Specifications
Performance Specifications

Table 5 Performance Specifications G7110B

Type Specification Comments

Instrument Control LC & CE Drivers A.02.14 or above For details about supported
Instrument Control Framework software versions refer to the
(ICF) A.02.04 or above compatibility matrix of the LC &
InfinityLab LC Companion CE Drivers version in use
(G7108AA) with firmware D.07.25
or above
Instant Pilot (G4208A) with
firmware B.02.20 or above
Lab Advisor B.02.08 or above

Communication Controller Area Network (CAN)


Enhanced Remote Interface (ERI)
Local Area Network (LAN)
Universal Serial Bus (USB)

Safety and maintenance Extensive diagnostics, error


detection and display through
Agilent Lab Advisor, leak detection,
safe leak handling, leak output
signal for shutdown of the pumping
system.
Low voltage in major maintenance
areas.

GLP features Early maintenance feedback (EMF)


for continuous tracking of
instrument usage in terms of seal
wear and volume of pumped
mobile phase with pre-defined and
user settable limits and feedback
messages. Electronic records of
maintenance and errors

Housing All materials are recyclable Safety Data Sheet in SAP:


4093-3577 Rev. A

For use with flow rates below 500 µL/min a vacuum degasser is required.
NOTE

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 41
2 Site Requirements and Specifications
Performance Specifications

Performance Specifications Agilent 1260 Infinity II


Quaternary Pump VL (G7111A)
Table 6 Performance Specifications G7111A

Type Specification Comments

Hydraulic system Dual piston in series pump with


servo-controlled variable stroke
drive, power transmission by gears
and ball screws, floating pistons

Flow range Settable: 0.001 – 10 mL/min Set points in 0.001 mL/min


Recommended: 0.2 – 10.0 mL/min increments

Flow precision ≤0.07 % RSD, or ≤0.02 min SD, based on retention time at
whichever is greater constant temperature

Flow accuracy ± 1 % or 10 µL/min whichever is Pumping degassed H2O at


greater 10 MPa (100 bar, 1450 psi)

Pressure operating range Up to 40 MPa (400 bar, 5802 psi)


up to 5 mL/min
Up to 20 MPa (200 bar, 2901 psi)
up to 10 mL/min

Pressure pulsation < 2 % amplitude (typically < 1.3 %), At 1 mL/min isopropanol, at all
or < 0.3 MPa (3 bar, 44 psi), pressures > 1 MPa
whichever is greater (10 bar, 145 psi)

Compressibility User-selectable, based on mobile


compensation phase compressibility

Recommended pH range 1.0 – 12.5 Solvents with pH < 2.3 should


not contain acids which attack
stainless steel

Gradient formation Low pressure quaternary


mixing/gradient capability using
proprietary high-speed
proportioning valve

Delay volume 600 – 900 µL, dependent on back Measured with water at
pressure 1 mL/min
(water/water with tracer)

Composition range Settable: 0 – 100 % In 0.1 % increments


Recommended: 5 – 95 % or
10 μL/min per channel, whichever
is greater

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 42
2 Site Requirements and Specifications
Performance Specifications

Table 6 Performance Specifications G7111A

Type Specification Comments

Composition precision < 0.2 % RSD or < 0.04 min SD, At 1 mL/min; based on
whichever is greater retention time at constant
temperature

Integrated degassing unit Number of channels: 4


Internal volume per channel: 1.5 mL

Instrument Control LC & CE Drivers A.02.14 or above For details about supported
Instrument Control Framework software versions refer to the
(ICF) A.02.04 or above compatibility matrix of the LC &
InfinityLab LC Companion CE Drivers version in use
(G7108AA) with firmware D.07.25
or above
Instant Pilot (G4208A) with
firmware B.02.20 or above
Lab Advisor software B.02.08 or
above

Communication Controller Area Network (CAN)


Local Area Network (LAN)
Enhanced Remote Interface (ERI)
Universal Serial Bus (USB)

Safety and maintenance Extensive diagnostics, error


detection and display through
Agilent Lab Advisor, leak detection,
safe leak handling, leak output
signal for shutdown of the pumping
system.
Low voltage in major maintenance
areas.

GLP features Early maintenance feedback (EMF)


for continuous tracking of
instrument usage in terms of seal
wear and volume of pumped
mobile phase with pre-defined and
user settable limits and feedback
messages. Electronic records of
maintenance and errors.

Housing All materials are recyclable

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 43
2 Site Requirements and Specifications
Performance Specifications

Performance Specifications Agilent 1260 Infinity II


Quaternary Pump (G7111B)
Table 7 Performance Specifications G7111B

Type Specification Comments

Hydraulic system Dual piston in series pump with


servo-controlled variable stroke
drive, power transmission by gears
and ball screws, floating pistons

Flow range Settable: 0.001 – 10 mL/min Set points in 0.001 mL/min


Recommended: 0.2 – 10.0 mL/min increments

Flow precision ≤0.07 % RSD, or ≤0.02 min SD, Based on retention time at
whichever is greater constant temperature

Flow accuracy ± 1 % or 10 µL/min, whichever is Pumping degassed H2O at


greater 10 MPa (100 bar, 1450 psi)

Pressure operating range Up to 60 MPa (600 bar, 8702 psi)


up to 5 mL/min
Up to 20 MPa (200 bar, 2901 psi)
up to 10 mL/min

Pressure pulsation < 2 % amplitude (typically < 1.3 %), At 1 mL/min isopropanol, at all
or < 0.3 MPa (3 bar, 44 psi), pressures > 1 MPa
whichever is greater (10 bar, 145 psi)

Compressibility User-selectable, based on mobile


compensation phase compressibility

Recommended pH range 1.0 – 12.5 Solvents with pH < 2.3 should


not contain acids which attack
stainless steel

Gradient formation Low pressure quaternary


mixing/gradient capability using
proprietary high-speed
proportioning valve

Delay volume 600 – 900 µL, dependent on back Measured with water at
pressure 1 mL/min
(water/water with tracer)

Composition range Settable: 0 – 100 % In 0.1 % increments


Recommended: 5 – 95 % or
10 μL/min per channel, whichever
is greater, under recommended
flow rate conditions

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 44
2 Site Requirements and Specifications
Performance Specifications

Table 7 Performance Specifications G7111B

Type Specification Comments

Composition precision < 0.2 % RSD or < 0.04 min SD, At 1 mL/min; based on
whichever is greater retention time at constant
temperature

Integrated degassing unit Number of channels: 4


Internal volume per channel: 1.5 mL

Instrument Control LC & CE Drivers A.02.14 or above For details about supported
Instrument Control Framework software versions refer to the
(ICF) A.02.04 or above compatibility matrix of the LC &
InfinityLab LC Companion CE Drivers version in use
(G7108AA) with firmware D.07.25
or above
Instant Pilot (G4208A) with
firmware B.02.20 or above
Lab Advisor software B.02.08 or
above

Communication Controller Area Network (CAN)


Local Area Network (LAN)
Extended Remote Interface (ERI)
Universal Serial Bus (USB)

Safety and maintenance Extensive diagnostics, error


detection and display through
Agilent Lab Advisor, leak detection,
safe leak handling, leak output
signal for shutdown of the pumping
system.
Low voltage in major maintenance
areas.

GLP features Early maintenance feedback (EMF)


for continuous tracking of
instrument usage in terms of seal
wear and volume of pumped
mobile phase with pre-defined and
user settable limits and feedback
messages. Electronic records of
maintenance and errors

Housing All materials are recyclable

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 45
2 Site Requirements and Specifications
Performance Specifications

Performance Specifications Agilent 1260 Infinity II


Bio-inert Pump (G5654A)
Table 8 Performance Specifications G5654A

Type Specification
Hydraulic system Dual piston in series pump with proprietary servo-controlled variable stroke
drive, floating pistons and active inlet valve, integrated 4-channel degassing
unit
Settable flow range 0.001 – 10 mL/min, in 0.001 mL/min increments
Flow precision < 0.07 % RSD, or <0.02 min SD whichever is greater, based on retention time
at constant room temperature
Flow accuracy ± 1 % or 10 µL/min whichever is greater
Pressure Operating range up to 60 MPa (600 bar, 8700 psi) up to 5 mL/min
Operating range up to 20 MPa (200 bar, 2950 psi) up to 10 mL/min
Pressure pulsation < 2 % amplitude (typically < 1.3 %), at 1 mL/min isopropanol, at all pressures
> 1 MPa (10 bar)
Compressibility User-selectable, based on mobile phase compressibility
compensation
Recommended pH 1 – 13, short term 141
range
Gradient formation Low pressure quaternary mixing/gradient capability using proprietary
high-speed proportioning valve
Delay volume 600 – 900 µL, dependent on back pressure.
Settable composition 0 – 100 % in 0.1 % increments
range
Composition < 0.2 % RSD or < 0.04 min SD, whichever is greater, at 1 mL/min; based on
precision retention time at constant room temperature
Instrument Control Lab Advisor B.02.08 or above
LC and CE Drivers A.02.14 or above
For details about supported software versions refer to the compatibility
matrix of your version of the LC and CE Drivers
Local Control Agilent Instant Pilot (G4208A) B.02.20 or above
Communications Controller-area network (CAN), USB
Extended Remote Interface: ready, start, stop and shut-down signals
Materials in flowpath Titanium, Gold, Platinum-Iridium, Sapphire, PEEK, PTFE, Ruby, Ceramic, FEP,
PFA
Active seal wash Included
1 For solvent compatibility, refer to section "Solvent information" in the manual for the Agilent 1260 Infinity
Bio-inert LC System.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 46
3 Installing the Pump

Unpacking the Pump 48


Delivery Checklist 48
Accessory Kits 49
Optimizing the Stack Configuration 51
Agilent InfinityLab Flex Bench Family 51
One Stack Configuration 54
Two Stack Configuration 57
Installing the Pump 59
Connecting Modules and Control Software 62
Connecting Modules 62
Connecting Control Software and/or G4208A Instant Pilot 62
Flow Connections of the Pump 63
Priming the System 64
Initial Priming 64
Regular Priming 66
Changing Solvents 67

This chapter gives information about the preferred stack setup for your system
and the installation of your module.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 47
3 Installing the Pump
Unpacking the Pump

Unpacking the Pump

If the delivery packaging shows signs of external damage, please call your Agilent
Technologies sales and service office immediately. Inform your service
representative that the instrument may have been damaged during shipment.

"Defective on arrival" problems


C AU T I O N
If there are signs of damage, please do not attempt to install the module.
Inspection by Agilent is required to evaluate if the instrument is in good
condition or damaged.
 Notify your Agilent sales and service office about the damage.
 An Agilent service representative will inspect the instrument at your site and
initiate appropriate actions.

Delivery Checklist
Ensure all parts and materials have been delivered with the pump. For checking
the completeness of your specific shipment, please use the list included in your
shipment. To aid in parts identification, please refer to chapter Parts and
Materials for Maintenance. Please report missing or damaged parts to your local
Agilent Technologies sales and service office.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 48
3 Installing the Pump
Unpacking the Pump

Accessory Kits
Accessory Kit G7111-68755
The Accessory Kit (G7111-68755) contains the following items:

# p/n Description
2 5043-1013 Tubing Clip IF-II

1 5181-1519 CAN cable, Agilent module to module, 1 m

1 5500-1246 Capillary ST 0.17 mm x 500 mm SI/SI

1 5500-1217 Capillary, ST, 0.17 mm x 900 mm SI/SX

3 5063-6527 Tubing, Silicon Rubber, 1.2 m, ID/OD 6/9 mm

1 G1311-90107 Algae note

3 5500-1169 Y Tube Connector ID 6.4

3 5500-1155 Tube Connector, 90 degree, ID 6.4

1 5043-1372 Tubing Connector Leak 3-1

2 5043-1373 Tubing Connector Leak Cap

2 0890-1195 Flexible sleeving 1.45 mm/2.5 mm, PTFE

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 49
3 Installing the Pump
Unpacking the Pump

Accessory Kit G5654-68755

For bio-inert modules use bio-inert parts only!

The Accessory Kit (G5654-68755) contains the following items:

# p/n Description
2 5043-1013 Tubing Clip IF-II

1 5181-1519 CAN cable, Agilent module to module, 1 m

1 G5611-60502 Capillary Ti 0.17 mm x 900 mm, L (Bio-inert)

1 5500-1264 Capillary Ti 0.17 mm x 500 mm, SL/SLV

3 5063-6527 Tubing, Silicon Rubber, 1.2 m, ID/OD 6/9 mm

1 G1311-90107 Algae note

3 5500-1155 Tube Connector, 90 degree, ID 6.4

2 5043-1373 Tubing Connector Leak Cap

1 5043-1372 Tubing Connector Leak 3-1

3 5500-1169 Y Tube Connector ID 6.4

2 0890-1195 Flexible sleeving 1.45 mm/2.5 mm, PTFE

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 50
3 Installing the Pump
Optimizing the Stack Configuration

Optimizing the Stack Configuration

To ensure safe operation and optimum performance of an Agilent InfinityLab LC


System, Agilent Technologies prescribe stack configurations. The following
configurations are possible:
• InfinityLab Flex Bench
• Single Stack (maximal 4 modules, in a bench rack or directly on the bench)
• InfinityLab Benchtop, providing more flexibility
• Two Stacks (in a bench rack or directly on the bench)

Agilent InfinityLab Flex Bench Family


Agilent A-Line LC Flex Bench
Agilent recommends using the InfinityLab Flex Bench for all Agilent LC systems.
Main features:
• Increases flexibility in the lab
• Safe moving of LC
• Easy stack customization
• Included waste management

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 51
3 Installing the Pump
Optimizing the Stack Configuration

Solvent cabinet
Shelves

Detector

Column
compartment
Handle grip

Sampler

Pump

Container for waste bottles

Wheel locking brake

Figure 12 Agilent InfinityLab Flex Bench

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 52
3 Installing the Pump
Optimizing the Stack Configuration

Agilent InfinityLab Benchtop

Figure 13 Agilent InfinityLab Benchtop

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 53
3 Installing the Pump
Optimizing the Stack Configuration

One Stack Configuration


Ensure optimum performance by stacking the modules as shown exemplarily in
Figure 14 on page 54. This configuration optimizes the flow path for minimum
delay volume and minimizes the bench space required.

Solvent cabinet

Detector

Column
compartment

Sampler

Pump

Figure 14 Single stack configuration (bench installation, example shows a Multisampler)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 54
3 Installing the Pump
Optimizing the Stack Configuration

One Stack Configuration (Vialsampler with Integrated Column


Compartment)
Ensure optimum performance by stacking the modules as shown exemplarily in
Figure 15 on page 55 and Figure 16 on page 56 for a set up with and without an
ICC, respectively. This configuration optimizes the flow path for minimum delay
volume and minimizes the bench space required.

Solvent cabinet

Sampler

Column
compartment

Detector

Pump

Figure 15 Single stack configuration (bench installation, example shows a Vialsampler with
optional ICC installed)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 55
3 Installing the Pump
Optimizing the Stack Configuration

Solvent cabinet

Detector

MCT

Vialsampler

Pump

Figure 16 Single stack configuration (bench installation, example shows a Vialsampler coupled
with a standalone column compartment)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 56
3 Installing the Pump
Optimizing the Stack Configuration

Two Stack Configuration


To avoid excessive height of the stack (for example when using the system in
combination with an additional detector), it is recommended to form two stacks.

Column compartment

Detectors

Solvent cabinet

Sampler

Pump

Figure 17 Two stack configuration (bench installation, example shows a Multisampler)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 57
3 Installing the Pump
Optimizing the Stack Configuration

Vialsampler Two Stack Configurations

Column compartment

Detectors

Solvent cabinet

Sampler

Pump

Figure 18 Two stack configuration (bench installation, example shows a Vialsampler coupled
with a standalone column compartment)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 58
3 Installing the Pump
Installing the Pump

Installing the Pump

Parts required # p/n Description


1 Pump
1 Data System
and/or
1 G4208A Instant Pilot
1 Power cord

For other cables see text below and “Cable Overview” on page 282.

Preparations • Locate bench space.


• Provide power connections.
• Unpack the module.

Module is partially energized when switched off, as long as the power cord
WAR N IN G
is plugged in.
Repair work at the module can lead to personal injuries, e.g. shock hazard,
when the cover is opened and the module is connected to power.
 Make sure that it is always possible to access the power plug.
 Remove the power cable from the instrument before opening the cover.
 Do not connect the power cable to the Instrument while the covers are
removed.

"Defective on arrival" problems


C AU T I O N
If there are signs of damage, please do not attempt to install the module.
Inspection by Agilent is required to evaluate if the instrument is in good
condition or damaged.
 Notify your Agilent sales and service office about the damage.
 An Agilent service representative will inspect the instrument at your site and
initiate appropriate actions.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 59
3 Installing the Pump
Installing the Pump

1 Place the module on the bench in a horizontal position.


2 Ensure the power switch on the front of the module is OFF (switch stands
out).

Status indicator

Power switch
Serial number
Figure 19 Front of pump

3 Connect the power cable to the power connector at the rear of the module.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 60
3 Installing the Pump
Installing the Pump

4 Connect the required interface cables to the pump, see “Connecting Modules
and Control Software” on page 62.

Configuration switch

USB-Mini-Port

LAN

ERI

CAN

Power socket

Figure 20 Rear view of the pump – electrical connections and label

5 Connect all capillaries, solvent tubes and waste tubing (see “Flow
Connections of the Pump” on page 63).
6 Press the power switch to turn on the module.

The power switch stays pressed in and a green indicator lamp in the power
NOTE switch is on when the module is turned on. When the line power switch stands
out and the green light is off, the module is turned off.

7 Purge the pump (see “Initial Priming” on page 64).

The pump was shipped with default configuration settings. To change these
NOTE settings, see “Setting the 6-bit Configuration Switch” on page 317.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 61
3 Installing the Pump
Connecting Modules and Control Software

Connecting Modules and Control Software

Use of unsupplied cables


WAR N IN G
Using cables not supplied by Agilent Technologies can lead to damage of
the electronic components or personal injury.
 Never use cables other than the ones supplied by Agilent Technologies
to ensure proper functionality and compliance with safety or EMC
regulations.

Connecting Modules
1 Place the individual modules in a stack configuration as shown in “Optimizing
the Stack Configuration” on page 51.
2 Ensure the power switches on the front of the modules are OFF (switches
stand out).
3 Plug a CAN cable into the CAN connector at the rear of the respective module
(except vacuum degasser).
4 Connect the CAN cable to the CAN connector of the next module.
5 Press in the power switches to turn on the modules.

Connecting Control Software and/or G4208A Instant


Pilot
Usually the detector is producing the most data in the stack, followed by the
NOTE pump, and it is therefore highly recommended to use either of these modules for
the LAN connection.

1 Ensure the power switches on the front of the modules in the stack are OFF
(switches stand out).
2 Connect the LAN enabled module with a LAN cable to the data system.
3 Plug the CAN connector of the Instant Pilot into any available CAN port of the
system.
4 Connect the CAN cable to the CAN connector of one of the modules.
5 Press in the power switches to turn on the modules.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 62
3 Installing the Pump
Flow Connections of the Pump

Flow Connections of the Pump

Tools required p/n Description


8710-0510 Open-end wrench 1/4 — 5/16 inch

Parts required Description


Other modules
Parts from accessory kits

Preparations Pump is installed in the LC system

Toxic, flammable and hazardous solvents, samples and reagents


WAR N IN G
The handling of solvents, samples and reagents can hold health and safety
risks.
 When working with these substances observe appropriate safety
procedures (for example by wearing goggles, safety gloves and
protective clothing) as described in the material handling and safety
data sheet supplied by the vendor, and follow good laboratory practice.
 The volume of substances should be reduced to the minimum required
for the analysis.
 Do not operate the instrument in an explosive atmosphere.

1 Open the doors.


2 Place the solvent cabinet on top of the stack.
3 Put the bottle-head assemblies into empty solvent reservoirs and place the
bottle in the solvent cabinet.
4 Connect the inlet tubes from the bottle-head assemblies to the inlet
connectors A to D at the right hand side of the vacuum degasser. Fix the
tubes in the tube clips of the pump.
5 Using a piece of sanding paper connect the waste tubing to the purge valve
and place it into your waste system.
6 Connect the waste tube to the waste outlet of the pump leak handling system.
7 Connect the pump outlet capillary (pump to injection device) to the outlet of
the purge valve.
8 Fill solvent reservoirs with your mobile phase.
9 Prime your system before first use (see “Initial Priming” on page 64).

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 63
3 Installing the Pump
Priming the System

Priming the System

Initial Priming

When Before a degasser or solvent tubing can be used, it is necessary to prime the system. Isopropanol is
recommended as priming solvent due to its miscibility with nearly all HPLC solvents and its excellent
wetting properties.

Parts required # Description


1 Isopropanol

Preparations Connect all modules hydraulically as described in the respective module manuals.
Fill each solvent bottle with 100 mL isopropanol
Switch the system on

When opening capillary or tube fittings, solvents may leak out.


WAR N IN G
The handling of toxic and hazardous solvents and reagents can carry
health risks.
 Observe appropriate safety procedures (for example, wear goggles,
safety gloves and protective clothing) as described in the material
handling and safety data sheet supplied by the solvent vendor,
especially when toxic or hazardous solvents are used.

The purge tool of the LabAdvisor or Instrument Utilities can be used to purge the
NOTE pump automatically.

If the pump is not able to aspirate the solvent from the bottles, use a syringe to
NOTE draw the solvent manually through tubing and degasser.

When priming the vacuum degasser with a syringe, the solvent is drawn through
NOTE the degasser tubes very quickly. The solvent at the degasser outlet will therefore
not be fully degassed. Pump for approximately 10 minutes at your desired flow
rate before starting an analysis. This will allow the vacuum degasser to properly
degas the solvent in the degasser tubes.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 64
3 Installing the Pump
Priming the System

1 Open the purge valve of the pump


2 Set the flow rate to 5 mL/min.
3 Select channel A.
4 Turn the flow on
5 Observe if the solvent in the tubing of channel A is advancing towards the
pump. If not, disconnect the solvent tubing from the MCGV, attach a syringe
with a syringe adapter and pull the liquid through the degasser. Reattach the
tubing to the MCGV.
6 Pump 30 mL isopropanol to remove residual air bubbles.
7 Switch to the next solvent channel and repeat steps 5 and 6 until all channels
have been purged.
8 Turn the flow off and close the purge valve.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 65
3 Installing the Pump
Priming the System

Regular Priming

When When the pumping system has been turned off for a certain time (for example, overnight) air will
rediffuse into the solvent channel between the vacuum degasser and the pump. If solvents
containing volatile components are left in the degasser without flow for a prolonged period, there will
be a slight loss of the volatile components.

Preparations Switch the system on

The purge tool of the Lab Advisor can be used for purging the pump.
NOTE

1 Open the purge valve of your pump by turning it counterclockwise and set the
flow rate to 5 mL/min.
2 Flush the vacuum degasser and all tubes with at least 10 mL of solvent.
3 Repeat step 1 and 2 for the other channel(s) of the pump.
4 Set the required composition and flow rate for your application and close the
purge valve.
5 Pump for approximately 10 minutes before starting your application.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 66
3 Installing the Pump
Priming the System

Changing Solvents

When When the solvent of a channel is to be replaced by another solvent that is not compatible (solvents
are immiscible or one solvent contains a buffer), it is necessary to follow the procedure below to
prevent clogging of the pump by salt precipitation or residual liquid droplets in parts of the system.

Parts required # p/n Description


1 Purging solvent(s), see Table 9 on page 68
1 5022-2184 Union ZDV

Preparations Remove the column and replace it by a ZDV fitting


Prepare bottles with appropriate intermediate solvents (see Table 9 on page 68)

1 If the channel is not filled with buffer, proceed to step 4.


2 Place the solvent intake filter into a bottle of water.
3 Flush the channel at a flow rate suitable for the installed tubing (typically
3 – 5 mL/min) for 10 min.
4 Modify the flow path of your system as required for your application.
Buffer salt of aqueous buffers may precipitate in residual isopropanol.
C AU T I O N
Capillaries and filter may be clogged by precipitating salt.
 Flush solvent lines containing high concentration of salts first with water
before introducing organic solvent.
 Always remove aqueous buffer before performing steps 5 to 7.

5 Replace the solvent bottle by a bottle of isopropanol.


6 Flush the channel at a flow rate suitable for the installed tubing (typically
3 – 5 mL/min) for 5 min.
7 Swap the bottle of isopropanol with a bottle of solvent for your application.
8 Repeat steps 1 to 7 for the other channel(s) of the pump.
9 Install the desired column, set the required composition and flow rate for your
application and equilibrate the system for approx. 10 minutes prior to starting
a run.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 67
3 Installing the Pump
Priming the System

Table 9 Choice of Priming Solvents for Different Purposes

Activity Solvent Comments

After an installation Isopropanol Best solvent to flush air out of


When switching between reverse Isopropanol the system
phase and normal phase (both Miscible with almost all
times) solvents

After an installation Ethanol or methanol Alternative to isopropanol


(second choice) if no
isopropanol is available

To clean the system when using HPLC grade water Best solvent to re-dissolve
buffers buffer crystals
After changing aqueous solvents HPLC grade water Best solvent to re-dissolve
buffer crystals

After the installation of normal Hexane + 5 % isopropanol Good wetting properties


phase seals (PE seals (pack of 2)
(0905-1420))

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 68
4 Using the Pump

Magnets 70
Turn on/off 71
Status Indicators 73
Best Practices 74
Daily / Weekly tasks 74
Power up / Shut-down the pump 74
Prepare the pump 75
How to deal with solvents 75
Select channels for Multi-Channel Gradient Valve (MCGV) 75
Leak and Waste Handling 76
Waste Concept 77
Hints for Successful Use of the Pump 78
Setting up the Pump with the G4208A Instant Pilot 79
Setting up the Pump with the Instrument Control Interface 80
Overview 80
Instrument Configuration 80
The Pump User Interface (Dashboard Panel) 82
Control Settings 84
Method Parameter Settings 85
Solvent Information 88
Materials in Flow Path 88
Bio-inert Materials 90
General Information about Solvent/Material Compatibility 92
Algae Growth in HPLC Systems 98
How to Prevent and/or Reduce the Algae Problem 98
Prevent Blocking of Solvent Filters 99
Normal Phase Applications 100

This chapter provides information for optimized usage of the module.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 69
4 Using the Pump
Magnets

Magnets

1 Magnets in doors of pumps, autosamplers, detectors, and fraction collectors.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 70
4 Using the Pump
Turn on/off

Turn on/off

This procedure exemplarily shows an arbitrary LC stack configuration.

1 2

Power switch: On

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 71
4 Using the Pump
Turn on/off

3 Turn instrument On/Off with the control software. 4

Power switch: Off


5

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 72
4 Using the Pump
Status Indicators

Status Indicators

This procedure exemplarily shows an arbitrary LC stack configuration.


1 The module status indicator indicates one of six possible module conditions:

Status indicators
1. Idle
2. Run mode
3. Not-ready. Waiting for a specific pre-run condition to be reached or
completed.
4. Error mode - interrupts the analysis and requires attention (for example, a
leak or defective internal components).
5. Resident mode (blinking) - for example, during update of main firmware.
6. Bootloader mode (fast blinking). Try to re-boot the module or try a
cold-start. Then try a firmware update.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 73
4 Using the Pump
Best Practices

Best Practices

Daily / Weekly tasks


Daily tasks
• Replace mobile phase based on water/buffer.
• Replace organic mobile phase latest every second day.
• Check seal wash solvent.

Weekly Tasks
• Change seal wash solvent (10 % isopropanol in water) and bottle.
• If applications with salts were used, flush all channels with water and remove
possible salt deposits manually.
• Inspect solvent filters for dirt or blockages. Exchange if no flow is coming out
of the solvent line when removed from the degasser inlet.

Power up / Shut-down the pump


Power up the pump
• Use new or different mobile phase (as required).
• Purge pump heads with 2.5 – 3 mL/min for 5 min.
• Stabilize the system by running for 10 – 20 min.

Long-term shut-down of the system


• Flush system with water to remove buffer.
• Remove all samples from the sampler and store according to good laboratory
practice.
• Use recommended solvents to store the system.
• Power off the system.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 74
4 Using the Pump
Best Practices

Prepare the pump


Purge
Use the Purge function to:
• fill the pump,
• exchange a solvent,
• remove air bubbles in tubes and pump heads.

Seal wash
Seal Wash guarantees a maximum seal life time. Use Seal Wash:
• When using buffers with elevated salt concentrations
• When using volatile solvents with non-volatile additives
Contaminated seal wash solvent
C AU T I O N
 Do not recycle seal wash solvent to avoid contamination.
 Weekly exchange seal wash solvent.

How to deal with solvents


• Use clean bottles only.
• Exchange water-based solvents daily.
• Select solvent volume to be used up within 1 – 2 days.
• Use only HPLC-grade solvents and water filtered through 0.2 µm filters.
• Label bottles correctly with bottle content, and filling date / expiry date.
• Use solvent inlet filters.
• Reduce risk of algae growth: use brown bottles for aqueous solvents, avoid
direct sunlight.

Select channels for Multi-Channel Gradient Valve


(MCGV)
• Use lower channels (A and/or D) for buffer solutions.
• Regularly flush all MCGV channels with water to remove possible salt
deposits.
• Check compatibility of buffers and organic solvents to avoid precipitation.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 75
4 Using the Pump
Leak and Waste Handling

Leak and Waste Handling

Toxic, flammable and hazardous solvents, samples and reagents


WAR N IN G
The handling of solvents, samples and reagents can hold health and safety
risks.
 When working with these substances observe appropriate safety
procedures (for example by wearing goggles, safety gloves and protective
clothing) as described in the material handling and safety data sheet
supplied by the vendor, and follow good laboratory practice.
 Do not use solvents with an auto-ignition temperature below 200 °C
(392 °F). Do not use solvents with a boiling point below 56 °C (133 °F).
 Avoid high vapor concentrations. Keep the solvent temperature at least
40 °C (72 °F) below the boiling point of the solvent used. This includes the
solvent temperature in the sample compartment. For the solvents
methanol and ethanol keep the solvent temperature at least 25 °C (45 °F)
below the boiling point.
 Do not operate the instrument in an explosive atmosphere.
 Do not use solvents of ignition Class IIC according IEC 60079-20-1 (for
example, carbon disulfide).
 Reduce the volume of substances to the minimum required for the
analysis.
 Never exceed the maximum permissible volume of solvents (8 L) in the
solvent cabinet. Do not use bottles that exceed the maximum permissible
volume as specified in the usage guideline for solvent cabinet.
 Ground the waste container.
 Regularly check the filling level of the waste container. The residual free
volume in the waste container must be large enough to collect the waste
liquid.
 To achieve maximal safety, regularly check the tubing for correct
installation.

For details, see the usage guideline for the solvent cabinet. A printed copy of the
NOTE guideline has been shipped with the solvent cabinet, electronic copies are
available in the Agilent Information Center or via the Internet.

For details on correct installation, see separate installation documentation.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 76
4 Using the Pump
Leak and Waste Handling

Waste Concept
1 Agilent recommends using the 6 L waste can with 1 Stay Safe cap GL45 with
4 ports (5043-1221) for optimal and safe waste disposal. If you decide to use
your own waste solution, make sure that the tubes don't immerse in the liquid.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 77
4 Using the Pump
Hints for Successful Use of the Pump

Hints for Successful Use of the Pump

• Always place the solvent cabinet with the solvent bottles on top of the
quaternary pump (or at a higher level).
• When using salt solutions and organic solvents in the quaternary pump it is
recommended to connect the salt solution to one of the bottom gradient
valve ports and the organic solvent to one of the upper gradient valve ports. It
is best to have the organic channel directly above the salt solution channel.
Regular flushing of all MCGV channels with water is recommended to remove
all possible salt deposits in the valve ports.
• Before operating the quaternary pump, flush the pump and vacuum degasser,
see “Regular Priming” on page 66. This is especially recommended if it has
been turned off for some time (for example, overnight) and volatile solvent
mixtures are used in the channels.
• Prevent blocking of solvent inlet filters. Never use the pump without solvent
inlet filter. Prevent the growth of algae, see “Prevent Blocking of Solvent
Filters” on page 99).
• Regularly check the purge valve frit and column frit. A blocked purge valve frit
can be identified by a black or yellow surface, deposits or by a pressure
greater than 10 bar, when pumping distilled water at a rate of 5 mL/min with
an open purge valve.
• When using the quaternary pump at low flow rates (for example, 0.2 mL/min)
check all 1/16-inch fittings for any signs of leaks.
• Whenever exchanging the pump seals the purge valve frit should be
exchanged, too.
• When using buffers or other salt solutions, flush the system with water before
switching it off. The seal wash option should be used when salt
concentrations of 0.1 M or higher will be used for long time periods.
• Check the pump pistons for scratches when changing the piston seals.
Scratched pistons will cause micro leaks and will decrease the lifetime of the
seal.
• Pressurize the system according to the wear in procedure after changing the
piston seals (see “Maintenance of a Pump Head Without Seal Wash
Option” on page 168).
• Consider recommendations given in the solvent information section, see
“Solvent Information” on page 88.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 78
4 Using the Pump
Setting up the Pump with the G4208A Instant Pilot

Setting up the Pump with the G4208A Instant


Pilot

Generic operation of the G4208A Instant Pilot is covered in the Agilent Instant
Pilot G4208A User's Guide (G4208-90006). Details about setting up module
specific parameters can be found in the Instant Pilot online help.
The pump parameters are described in detail in “Overview” on page 80.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 79
4 Using the Pump
Setting up the Pump with the Instrument Control Interface

Setting up the Pump with the Instrument Control


Interface

Overview
The instrument control interface offers the parameters described in the following
sections, and can usually be accessed through Agilent instrument control
software. For details, please see manuals and online help of respective user
interfaces.

Instrument Configuration
Use the Instrument Configuration dialog box to examine and, if necessary, modify
your instrument configuration. The Configurable Modules panel contains a list of
all modules available for configuration. The Selected Modules panel contains the
list of configured modules.
Auto Configuration: Under Communication settings, select either the Host Name
option or the IP address option and enter the appropriate value for the host
computer to enable automatic detection of the hardware configuration. The
system configures the instrument automatically with no further manual
configuration necessary.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 80
4 Using the Pump
Setting up the Pump with the Instrument Control Interface

The Quaternary Pump configuration parameters are in two sections:


• Communication
• Options
Communication: The parameters in this dialog box are detected automatically
during autoconfiguration.
• Device name,
• Type ID,
• Serial number,
• Firmware revision,
• Button Connection settings
Options:
• Pressure Unit:
select the pressure units from the drop-down list (bar, psi or MPa).
• Seal wash installed:
This check box is marked to indicate that an optional seal wash has been
detected during autoconfiguration.
Please refer to the online help of your user interface for more detailed
information.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 81
4 Using the Pump
Setting up the Pump with the Instrument Control Interface

The Pump User Interface (Dashboard Panel)


Module Graphic
The items in the pump graphic have the following meaning and function:

The level of solvent in the bottle is denoted by the gray area.


When the solvent level falls below the specified volume, the area
turns yellow. When the bottle is empty, the area turns red.
Clicking the solvent bottle displays the Bottle Fillings dialog box.
The tooltip for the bottle shows the solvent name.

Indicates that the ISET option is installed but with no active


method (light gray) or installed and active (dark gray).

The pressure setpoints. The red line shows the current


maximum pressure limit; the gray area shows the current
pressure (also shown as text).

The current solvent flow rate (in mL/min) is displayed above the pressure display.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 82
4 Using the Pump
Setting up the Pump with the Instrument Control Interface

Instrument Signals
The following pump signals are displayed:
Table 10 Pump signals (isocratic pump)

Flow The current solvent flow rate (in mL/min).

Pressure The current pump pressure (in bar, psi or MPa, see “Instrument
Configuration” on page 80).

Ripple The current ripple (in %).

Pressure Limit The current maximum pressure limit.

Table 11 Pump signals (quaternary pump)

Flow The current solvent flow rate (in mL/min).

Pressure The current pump pressure (in bar, psi or MPa, see “Instrument
Configuration” on page 80).

Ripple The current ripple (in %).

Pressure Limit The current maximum pressure limit.

Composition A:B The contributions of channels A and B to the current solvent composition.

Composition C:D The contributions of channels C and D to the current solvent composition.

Context Menu
The context menu of the dashboard panel contains the following commands:

Control Displays the pump's Control dialog box.

Method Displays the pump's Method Setup dialog box.

Set Error Method Sets the method that is loaded if an error occurs to the method
that is currently available in the hardware.

Identify Device Causes the LED on the front of the module to blink for a few
seconds.

Switch Pump On/Off Toggles the status of the pump, on or off.

Bottle Fillings Displays the Bottle Fillings dialog box.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 83
4 Using the Pump
Setting up the Pump with the Instrument Control Interface

Control Settings
The pump control parameters are in three sections:
• Pump
• Seal Wash
• Automatic Turn On
Table 12 Pump control parameters

Parameter Limits Description

Pump Enables you to switch the pump On, Off or to a Standby condition.
In the Standby condition, the pump motor is still active, and when the pump
is switched on again, does not need to be re-initialized.

Seal Wash The seal wash can be set up to be run once ( Single wash) or periodically (
Periodic).
• Off: no seal wash is used.
• Single wash: the seal will be purged for a specified time.
• Periodic: a periodic wash will be applied for a defined period in minutes.
The option is available only when the pump has seal wash capability. The
seal wash capability is detected by the module itself. If seal wash is
installed, it is recommended to use it in order to increase the primary seal
lifetime.

Automatic Turn On Module can be turned on at a specified date/time. This feature can only be
used if the module power switch is turned on.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 84
4 Using the Pump
Setting up the Pump with the Instrument Control Interface

Method Parameter Settings


The pump method setup parameters are in eight sections:
• Flow
• Solvent(s)
• Stoptime
• Posttime
• Pressure Limits
• Timetable
• Advanced
Table 13 Method parameters

Parameter Limits Description


Flow ≤200 bar: The flow is the rate of movement of eluent along the column. It is important
0.0 – 10.00 mL/min in steps that the flow rate is kept constant to ensure precise retention time, and peak
of 0.001 measurements. Variations in flow rate can occur as a result of the following:
> 200 bar: 0.0 – 5.00 mL/min partial failure of the pumping system, air in the pumping system, a change in
in steps of 0.001. the mobile phase viscosity, or a temperature change.
Solvent(s) Isocratic Pump: The text box allows you to type a brief description of the
solvent.
Quaternary Pump: You can set the percentages of solvents B, C, and D to
any value from 0 through 100 %. Solvent A always delivers the remaining
volume: 100 - (%B + %C + %D). The check boxes allow you to turn on the
solvent channels (checked) or off (cleared). The text boxes allow you to type
a brief description of each of the solvents.
Stoptime 0.01 – 99999 minor As The stoptime sets a time limit for your analysis. After the stoptime, all
Injector/No Limit (an infinite gradients are stopped and the pump parameters return to their initial values.
run time). The pump can be used as a stoptime master for the complete analytical
system. The pump also stops the detectors if they have a No Limit stoptime
setting. If no limit is given, a method will have to be stopped manually.
Posttime 0.01 – 99999 minor Off Your instrument remains in a not ready state during the posttime to delay
( 0.0 min). the start of the next analysis. You can use the Posttime to allow your column
to equilibrate after changes in solvent composition (for example after
gradient elution).
Pressure Limits Max: 600 bar (8700 psi) for Sets the maximum and minimum pressure limits for the pump.
flow rates up to 5 mL/min. • Max is the maximum pressure limit at which the pump will switch itself
200 bar (2900 psi) for flow off, protecting the analytical system against overpressure.
rates > 5 mL/min. • Min is the minimum limit at which the pump will switch itself off. For
Min: any value between 0 and example, if any solvent reservoir is empty, this setting prevents system
the upper pressure limit damage by pumping air.
setting.
Timetable See “Timetable Settings” on page 87
Advanced See “Advanced Settings” on page 86

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 85
4 Using the Pump
Setting up the Pump with the Instrument Control Interface

Advanced Settings
The pump advanced method setup parameters are in four sections:
• Minimum Stroke
• Compressibility
• Maximum Flow Gradient
• Primary Channel (Quaternary Pump only)
Table 14 Advanced method parameters

Parameter Limits Description

Minimum Stroke 20 – 100 µL The Stroke Volume is used for optimizing between performance of the
module and seal life time. For performance a low stroke volume is
beneficial, as it divides disturbances into smaller packages, but a larger
volume is extending the life time of the pump seals.
If Automatic is activated, the pump tries to achieve an optimized stroke
volume for best performance and lowest ripple.

Compressibility 40·10-6 – 150·10-6. The compressibility of the mobile phase has an effect on the performance
Default compressibility of the pump. For best flow accuracy and mixing performance, you can set
value: 100·10-6 the parameter according to the mobile phase being used.

Maximum Flow 0.1 – 100 mL/min/min You can set a limit on the rate of change of the solvent flow to protect your
Gradient Default value: analytical column.
100 mL/min/min

Primary Channel Using Automatic is recommended.


(Quaternary pump The primary channel can be specified as A to D for optimizing highly specific
only) methods. It is split up to deliver the first and last solvent package created by
the MCGV in order to optimize composition precision. The primary channel
does not change during a gradient, as long as the channel is used. Using
Automatic chooses the channel with the highest percentage at start
conditions before a gradient.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 86
4 Using the Pump
Setting up the Pump with the Instrument Control Interface

Timetable Settings
Use the Timetable to program changes in the pump parameters during the
analysis by entering a time in the Time field and appropriate values in the
following fields of the timetable. Changes in flow rate occur linearly from either
time zero or the time of the last defined change; other parameters change
instantaneously at the time defined in the timetable.
Show Advanced Timetable toggles the timetable display between standard mode
and advanced mode.
The following parameters can be changed:
• Change Contacts
• Change Flow
• Change Max. Pressure Limit
• Change Solvent Composition - You can only use solvents, which have been
enabled in the solvents section.
• Function centric view - This checkbox allows you displaying parameter
changes instead of a time table.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 87
4 Using the Pump
Solvent Information

Solvent Information

Observe the following recommendations on the use of solvents.


• Follow the recommendations for avoiding the growth of algae, see “Algae
Growth in HPLC Systems” on page 98
• Small particles can permanently block capillaries and valves. Therefore,
always filter solvents through 0.22 µm filters.
• Avoid or minimize the use of solvents that may corrode parts in the flow path.
Consider specifications for the pH range given for different materials such as
flow cells, valve materials etc. and recommendations in subsequent sections.

Materials in Flow Path


Following materials are used in the flow path of this module:
Table 15 Materials in flow path (G7110B, G7111A, G7111B)

Part Materials

Degasser chamber TFE/PDD Copolymer, PFA (internal tubings), PEEK (inlets), FEP
(tubings), ETFE (fittings)

MCGV SST, PTFE, PEEK, FEP, PFA, ceramic, sapphire, ruby

Passive inlet valve SST, gold, sapphire, ruby, ceramic, PTFE

Active inlet valve SST, gold, sapphire, ruby, ceramic, PTFE

Outlet valve SST, gold, ruby, ZrO2-based ceramic, tantalum

Adapter SST, gold

Pump head (body) SST

Pistons Sapphire

Piston seals/wash seals PTFE, SST (reversed phase) or UHMW-PE, SST (normal phase)

Pressure sensor SST

Purge valve SST, gold, PTFE, ceramic

Damping unit SST, gold

Capillaries/fittings SST

Tubings PTFE

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 88
4 Using the Pump
Solvent Information

Table 16 Materials in flow path (G5654A)

Part Materials

Degasser chamber TFE/PDD Copolymer, PFA (internal tubings), PEEK (inlets), FEP
(tubings), ETFE (fittings)

MCGV Platinum-iridium, titanium, PTFE, PEEK, FEP, PFA, ceramic, sapphire,


ruby

Active inlet valve Platinum-iridium, titanium, gold, sapphire, ruby, ceramic, PTFE

Outlet valve Titanium, gold, ruby, ZrO2-based ceramic, tantalum

Adapter Titanium, gold

Pump head (body) Titanium

Pistons Sapphire

Piston seals/wash seals PTFE, gold

Pressure sensor Titanium

Purge valve Titanium, gold, PTFE, ceramic

Damping unit Titanium, gold

Capillaries/fittings Titanium

Tubings PTFE

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 89
4 Using the Pump
Solvent Information

Bio-inert Materials
For the Bio-inert LC system, Agilent Technologies uses highest-quality materials
in the flow path (also referred to as wetted parts), which are widely accepted by
life science scientists, as they are known for optimum inertness to biological
samples and ensure best compatibility with common samples and solvents over
a wide pH range. Explicitly, the complete flow path is free of stainless steel and
free of other alloys containing metals such as iron, nickel, cobalt, chromium,
molybdenum, or copper, which can interfere with biological samples. The flow
downstream of the sample introduction contains no metals whatsoever.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 90
4 Using the Pump
Solvent Information

Table 17 Used Bio-inert materials

Module Materials

Agilent 1260 Infinity II Bio-inert Pump Titanium, gold, platinum-iridium, ceramic,


(G5654A) ruby, PTFE, PEEK

Agilent 1260 Infinity II Bio-inert Multisampler Upstream of sample introduction:


(G5668A) • Titanium, gold, PTFE, PEEK, ceramic
Downstream of sample introduction:
• PEEK, ceramic

Agilent 1260 Infinity II Bio-inert Manual Injector PEEK, ceramic


(G5628A)

Agilent 1260 Infinity II Bio-inert Analytical Fraction Collector PEEK, ceramic, PTFE
(G5664B)

Bio-inert Flow Cells:

Standard flow cell bio-inert, 10 mm, 13 µL, 120 bar ( 12 MPa) for MWD/DAD, PEEK, ceramic, sapphire, PTFE
includes 0890-1763 – 0.18 x 1500 mm PEEK capillary and 5063-6591 – PEEK
fittings (G5615-60022)
(for Agilent 1260 Infinity II DAD G7115A, and MWD G7165A)

Bio-inert flow cell, 8 µL, 20 bar (pH 1–12), includes and PEEK fittings PEEK, fused silica, PTFE
(G5615-60005)
(for Agilent 1260 Infinity II FLD G7121A/B)

Bio-inert Heat Exchangers, Valves and Capillaries:

Quick Connect Heat Exchanger Bio-inert (G7116-60041) PEEK (steel-cladded)


(for Agilent 1260 Infinity II Multicolumn Thermostat G7116A)

Bio-inert Valve heads (G4235A, G5631A, G5632A, G5639A) PEEK, ceramic (Al2O3 based)

Bio-inert Connection capillaries Upstream of sample introduction:


• Titanium
Downstream of sample introduction:
• Agilent uses stainless-steel-cladded
PEEK capillaries, which keep the flow
path free of steel and provide pressure
stability up to 600 bar.

To ensure optimum biocompatibility of your Agilent 1260 Infinity II Bio-inert LC


NOTE system, do not include non-inert standard modules or parts to the flow path. Do
not use any parts that are not labeled as Agilent “Bio-inert”. For solvent
compatibility of these materials, see “General Information about Solvent/Material
Compatibility” on page 92.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 91
4 Using the Pump
Solvent Information

General Information about Solvent/Material


Compatibility
Materials in the flow path are carefully selected based on Agilent’s experiences in
developing highest-quality instruments for HPLC analysis over several decades.
These materials exhibit excellent robustness under typical HPLC conditions. For
any special condition, please consult the material information section or contact
Agilent.

Disclaimer
Subsequent data was collected from external resources and is meant as a
reference. Agilent cannot guarantee the correctness and completeness of such
information. Data is based on compatibility libraries, which are not specific for
estimating the long-term life time under specific but highly variable conditions of
UHPLC systems, solvents, solvent mixtures, and samples. Information also
cannot be generalized due to catalytic effects of impurities like metal ions,
complexing agents, oxygen etc. Apart from pure chemical corrosion, other
effects like electro corrosion, electrostatic charging (especially for nonconductive
organic solvents), swelling of polymer parts etc. need to be considered. Most
data available refers to room temperature (typically 20 – 25 °C, 68 – 77 °F). If
corrosion is possible, it usually accelerates at higher temperatures. If in doubt,
please consult technical literature on chemical compatibility of materials.

MP35N
MP35N is a nonmagnetic, nickel-cobalt-chromium-molybdenum alloy
demonstrating excellent corrosion resistance (for example, against nitric and
sulfuric acids, sodium hydroxide, and seawater) over a wide range of
concentrations and temperatures. In addition, this alloy shows exceptional
resistance to high-temperature oxidation. Due to excellent chemical resistance
and toughness, the alloy is used in diverse applications: dental products, medical
devices, nonmagnetic electrical components, chemical and food processing
equipment, marine equipment. Treatment of MP35N alloy samples with 10 %
NaCl in HCl (pH 2.0) does not reveal any detectable corrosion. MP35N also
demonstrates excellent corrosion resistance in a humid environment. Although
the influence of a broad variety of solvents and conditions has been tested, users
should keep in mind that multiple factors can affect corrosion rates, such as
temperature, concentration, pH, impurities, stress, surface finish, and dissimilar
metal contacts.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 92
4 Using the Pump
Solvent Information

Polyphenylene Sulfide (PPS)


Polyphenylene sulfide has outstanding stability even at elevated temperatures. It
is resistant to dilute solutions of most inorganic acids, but it can be attacked by
some organic compounds and oxidizing reagents. Nonoxidizing inorganic acids,
such as sulfuric acid and phosphoric acid, have little effect on polyphenylene
sulfide, but at high concentrations and temperatures, they can still cause material
damage. Nonoxidizing organic chemicals generally have little effect on
polyphenylene sulfide stability, but amines, aromatic compounds, and
halogenated compounds may cause some swelling and softening over extended
periods of time at elevated temperatures. Strong oxidizing acids, such as nitric
acid (> 0.1 %), hydrogen halides (> 0.1 %), peroxy acids (> 1 %), or chlorosulfuric
acid degrade polyphenylene sulfide. It is not recommended to use polyphenylene
sulfide with oxidizing material, such as sodium hypochlorite and hydrogen
peroxide. However, under mild environmental conditions, at low concentrations
and for short exposure times, polyphenylene sulfide can withstand these
chemicals, for example, as ingredients of common disinfectant solutions.

PEEK
PEEK (Polyether-Ether Ketones) combines excellent properties regarding
biocompatibility, chemical resistance, mechanical and thermal stability. PEEK is
therefore the material of choice for UHPLC and biochemical instrumentation.
It is stable in the specified pH range (for the Bio-Inert LC system: pH 1 – 13, see
bio-inert module manuals for details), and inert to many common solvents.
There are still some known incompatibilities with chemicals such as chloroform,
methylene chloride, THF, DMSO, strong acids (nitric acid > 10 %, sulfuric acid >
10 %, sulfonic acids, trichloroacetic acid), halogens or aqueous halogen
solutions, phenol and derivatives (cresols, salicylic acid, and so on).
When used above room temperature, PEEK is sensitive to bases and various
organic solvents, which can cause it to swell. Under such conditions, normal
PEEK capillaries are sensitive to high pressure. Therefore, Agilent uses stainless
steel clad PEEK capillaries in bio-inert systems. The use of stainless steel clad
PEEK capillaries keeps the flow path free of steel and ensures pressure stability
up to 600 bar. If in doubt, consult the available literature about the chemical
compatibility of PEEK.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 93
4 Using the Pump
Solvent Information

Polyimide
Agilent uses semi-crystalline polyimide for rotor seals in valves and needle seats
in autosamplers. One supplier of polyimide is DuPont, which brands polyimide as
Vespel, which is also used by Agilent.
Polyimide is stable in a pH range between 1 and 10 and in most organic solvents.
It is incompatible with concentrated mineral acids (e.g. sulphuric acid), glacial
acetic acid, DMSO and THF. It is also degraded by nucleophilic substances like
ammonia (e.g. ammonium salts in basic conditions) or acetates.

Polyethylene (PE)
Agilent uses UHMW (ultra-high molecular weight)-PE/PTFE blends for yellow
piston and wash seals, which are used in 1290 Infinity pumps, 1290 Infinity II
pumps, the G7104C and for normal phase applications in 1260 Infinity pumps.
Polyethylene has a good stability for most common inorganic solvents including
acids and bases in a pH range of 1 to 12.5. It is compatible with many organic
solvents used in chromatographic systems like methanol, acetonitrile and
isopropanol. It has limited stability with aliphatic, aromatic and halogenated
hydrocarbons, THF, phenol and derivatives, concentrated acids and bases. For
normal phase applications, the maximum pressure should be limited to 200 bar.

Tantalum (Ta)
Tantalum is inert to most common HPLC solvents and almost all acids except
fluoric acid and acids with free sulfur trioxide. It can be corroded by strong bases
(e.g. hydroxide solutions > 10 %, diethylamine). It is not recommended for the use
with fluoric acid and fluorides.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 94
4 Using the Pump
Solvent Information

Stainless Steel (SST)


Stainless steel is inert against many common solvents. It is stable in the
presence of acids and bases in a pH range of 1 to 12.5. It can be corroded by
acids below pH 2.3. It can also corrode in following solvents:
• Solutions of alkali halides, their respective acids (for example, lithium iodide,
potassium chloride) and aqueous solutions of halogens.
• High concentrations of inorganic acids like nitric acid, sulfuric acid, and
organic solvents especially at higher temperatures (replace, if your
chromatography method allows, by phosphoric acid or phosphate buffer,
which are less corrosive against stainless steel).
• Halogenated solvents or mixtures, which form radicals and/or acids, for
example:
2 CHCl3 + O2→ 2 COCl2 + 2 HCl
This reaction, in which stainless steel probably acts as a catalyst, occurs
quickly with dried chloroform if the drying process removes the stabilizing
alcohol.
• Chromatographic grade ethers, which can contain peroxides (for example,
THF, dioxane, diisopropyl ether). Such ethers should be filtered through dry
aluminum oxide, which adsorbs the peroxides.
• Solutions of organic acids (acetic acid, formic acid, and so on) in organic
solvents. For example, a 1 % solution of acetic acid in methanol will attack
steel.
• Solutions containing strong complexing agents (for example, EDTA,
ethylenediaminetetraacetic acid).
• Mixtures of carbon tetrachloride with isopropanol or THF.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 95
4 Using the Pump
Solvent Information

Titanium (Ti)
Titanium is highly resistant to oxidizing acids (for example, nitric, perchloric and
hypochlorous acid) over a wide range of concentrations and temperatures. This
is due to a thin oxide layer on the surface, which is stabilized by oxidizing
compounds. Non-oxidizing acids (for example, hydrochloric, sulfuric and
phosphoric acid) can cause slight corrosion, which increases with acid
concentration and temperature. For example, the corrosion rate with 3 % HCl
(about pH 0.1) at room temperature is about 13 μm/year. At room temperature,
titanium is resistant to concentrations of about 5 % sulfuric acid (about pH 0.3).
Addition of nitric acid to hydrochloric or sulfuric acids significantly reduces
corrosion rates. Titanium is sensitive to acidic metal chlorides like FeCl3 or CuCl2.
Titanium is subject to corrosion in anhydrous methanol, which can be avoided by
adding a small amount of water (about 3 %). Slight corrosion is possible with
ammonia > 10 %.

Diamond-Like Carbon (DLC)


Diamond-Like Carbon is inert to almost all common acids, bases, and solvents.
There are no documented incompatibilities for HPLC applications.

Fused silica and Quartz (SiO2)


Fused silica is used in Max Light Cartridges. Quartz is used for classical flow cell
windows. It is inert against all common solvents and acids except hydrofluoric
acid and acidic solvents containing fluorides. It is corroded by strong bases and
should not be used above pH 12 at room temperature. The corrosion of flow cell
windows can negatively affect measurement results. For a pH greater than 12,
the use of flow cells with sapphire windows is recommended.

Gold
Gold is inert to all common HPLC solvents, acids, and bases within the specified
pH range. It can be corroded by complexing cyanides and concentrated acids like
aqua regia.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 96
4 Using the Pump
Solvent Information

Zirconium Oxide (ZrO2)


Zirconium Oxide is inert to almost all common acids, bases, and solvents. There
are no documented incompatibilities for HPLC applications.

Platinum/Iridium
Platinum/Iridium is inert to almost all common acids, bases, and solvents. There
are no documented incompatibilities for HPLC applications.

Fluorinated polymers (PTFE, PFA, FEP, FFKM, PVDF)


Fluorinated polymers like PTFE (polytetrafluorethylene), PFA (perfluoroalkoxy),
and FEP (fluorinated ethylene propylene) are inert to almost all common acids,
bases, and solvents. FFKM is perfluorinated rubber, which is also resistant to
most chemicals. As an elastomer, it may swell in some organic solvents like
halogenated hydrocarbons.
TFE/PDD copolymer tubings, which are used in all Agilent degassers except
G1322A/G7122A, are not compatible with fluorinated solvents like Freon,
Fluorinert, or Vertrel. They have limited life time in the presence of
hexafluoroisopropanol (HFIP). To ensure the longest possible life with HFIP, it is
best to dedicate a particular chamber to this solvent, not to switch solvents, and
not to let dry out the chamber. For optimizing the life of the pressure sensor, do
not leave HFIP in the chamber when the unit is off.
The tubing of the leak sensor is made of PVDF (polyvinylidene fluoride), which is
incompatible with the solvent DMF (dimethylformamide).

Sapphire, Ruby, and Al2O3-based ceramics


Sapphire, ruby, and ceramics based on aluminum oxide Al2O3 are inert to almost
all common acids, bases, and solvents. There are no documented
incompatibilities for HPLC applications.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 97
4 Using the Pump
Algae Growth in HPLC Systems

Algae Growth in HPLC Systems

The presence of algae in HPLC systems can cause many problems that may be
incorrectly diagnosed as instrument or application problems. Algae grow in
aqueous media, preferably in a pH range from 4 to 8. Their growth is accelerated
by buffers, for example phosphate or acetate. Since algae grow through
photosynthesis, light will also stimulate their growth. Even in distilled water
small-sized algae grow after some time.

Instrumental Problems Associated With Algae


Algae deposit and grow everywhere within the HPLC system, causing the
following problems:
• Blocked solvent filters, or deposits on inlet or outlet valves, resulting in
unstable flow, composition or gradient problems, or a complete failure of the
pump.
• Plugging of small-pore, high-pressure solvent filters, usually placed before the
injector, resulting in high system pressure.
• Blockage of PTFE frits, leading to increased system pressure.
• Plugging of column filters, giving high system pressure.
• Dirty flow cell windows of detectors, resulting in higher noise levels (since the
detector is the last module in the flow path, this problem is less common).

How to Prevent and/or Reduce the Algae Problem


• Always use freshly prepared solvents, especially use demineralized water,
which was filtered through 0.2 µm filters.
• Never leave mobile phase in the instrument for several days without flow.
• Always discard old mobile phase.
• Use the amber solvent bottle (Solvent bottle, amber, 1000 mL (9301-6526))
supplied with the instrument for your aqueous mobile phase.
• If possible add a few mg/L sodium azide or a few percent organic solvent to
the aqueous mobile phase.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 98
4 Using the Pump
Prevent Blocking of Solvent Filters

Prevent Blocking of Solvent Filters

Contaminated solvents or algae growth in the solvent bottle will reduce the
lifetime of the solvent filter and will influence the performance of the module.
This is especially true for aqueous solvents or phosphate buffers (pH 4 to 7). The
following suggestions will prolong lifetime of the solvent filter and will maintain
the performance of the module.
• Use a sterile, if possible amber, solvent bottle to slow down algae growth.
• Filter solvents through filters or membranes that remove algae.
• Exchange solvents every two days or refilter.
• If the application permits add 0.0001 – 0.001 M sodium azide to the solvent.
• Place a layer of argon on top of your solvent.
• Avoid exposure of the solvent bottle to direct sunlight.

Never use the system without solvent filter installed.


NOTE

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 99
4 Using the Pump
Normal Phase Applications

Normal Phase Applications

Current passive inlet valves and outlet ball valves used with 1260 and 1290
Infinity pumps do not work well with applications using non-polar solvents as for
normal phase applications (e.g. hexane, heptane and CO2). With such
applications, pressure drops could be observed. They are a result of particles
electrostatically charging up in insulating solvents and sticking to the balls inside
the valves, such that the valves do not close properly any more after some time
of use (can be hours).
For normal phase applications, a second type of valves is available, which has a
design based on the existing one for 1260 and 1290 Infinity valves. These valves
use a new material for valve balls, which is a conductive ceramic and replaces
non-conductive ruby balls. The balls do not charge up electrostatically and show
good performance in normal phase.
The valves are marked with N for non-polar or normal phase.
Agilent recommends using these valves for (and only for) normal phase
applications.

Corrosion of valves
C AU T I O N
Normal phase balls/valves corrode quickly in aqueous solutions and acids (at
or below pH 7).
 Do not use normal phase valves in applications running with aqueous
solutions.

The N-Valves have been tested successfully in using hexane at pressures below
100 bar; heptane can be used as a substitute for neurotoxic hexane.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 100
4 Using the Pump
Normal Phase Applications

Seals for Normal Phase Applications


For running normal phase applications on 1200 Infinity Series pumps, yellow PE
seals are required, which exist as piston seals and wash seals. Seal wash is very
uncommon for normal phase applications (no buffers needed), but wash seals
are needed for seal wash pump heads.
1290 Infinity pumps use PE seals by default. In combination with ceramic
pistons, PE seals are used for both reversed phase (1200 bar) and normal phase
applications.
1260 Infinity pumps use sapphire pistons and black PTFE piston and wash seals
by default (600 bar). Such PTFE seals create small wear particles in normal
phase applications, which can clog valves and other parts in the flow path.
PE seals have a limited life time when used with normal phase solvents and
sapphire pistons. Agilent recommends a maximum pressure of 200 bar for this
combination, which shall also be applied for pressure tests.

Choice of Normal Phase Valves and Seals


Table 18 Recommended valves and seals for normal phase applications

1260 Infinity 1290 Infinity

Inlet valves 1260 Infinity Inlet Valve Type N 1290 Infinity Inlet Valve Type N
(G1312-60166) (G4220-60122)
1290 Infinity Quat Inlet Valve
Type N (G4204-60122)

Outlet valves Outlet Valve Type N/SFC 1290 Infinity Outlet Valve Type N
(G1312-60167) (G4220-60128)

Seals PE seals (pack of 2)


(0905-1420)
Wash Seal PE (0905-1718)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 101
5 Optimizing Performance

Using the Degasser 103


Operational Hints for the Vacuum Degasser 104
Operational Hints for the Multi Channel Gradient Valve (MCGV) 105
When to use the Seal Wash Function 106
Choosing the Right Pump Seals 107
Optimize the Compressibility Compensation Setting 108

This chapter gives hints on how to optimize the performance or use additional
devices.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 102
5 Optimizing Performance
Using the Degasser

Using the Degasser

The quaternary pump has a built-in degasser, which should always be included to
the flow path.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 103
5 Optimizing Performance
Operational Hints for the Vacuum Degasser

Operational Hints for the Vacuum Degasser

If you are using the vacuum degasser for the first time, if the vacuum degasser
was switched off for any length of time (for example, overnight), or if the vacuum
degasser chambers are empty, you have to prime the vacuum degasser before
running an analysis. Priming is usually done by pumping at a high flow rate
(3 – 5 mL/min). Alternatively, a syringe can be used to draw the solvent through
the (empty) degasser if the pump does not aspirate the solvent by itself.
For details see “Initial Priming” on page 64.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 104
5 Optimizing Performance
Operational Hints for the Multi Channel Gradient Valve (MCGV)

Operational Hints for the Multi Channel Gradient


Valve (MCGV)

In a mixture of salt solutions and organic solvent the salt solution might be well
dissolved in the organic solvent without showing precipitations. However in the
mixing point of the gradient valve, at the boundary between the two solvents,
micro precipitation is possible. Gravity forces the salt particles to fall down.
Normally the A channel of the valve is used for the aqueous/salt solution and the
B channel of the pump is used for the organic solvent. If used in this
configuration the salt will fall back into the aqueous solution and will be
dissolved. When using the pump in a different configuration (e.g., D - salt solution,
A - organic solvent) the salt can fall into the port of the organic solvent and may
lead to performance problems.

When using salt solutions and organic solvents it is recommended to connect


NOTE the salt solution to one of the bottom ports of the MCGV and the organic solvent
to one of the upper gradient valve ports. It is best to have the organic channel
directly above the salt solution channel. Regular flushing with water of all MCGV
channels is recommended to remove all possible salt deposits in the valve ports.

Precipitations formed during the mixing of buffers and organic solvents which do
NOTE not dissolve salts may cause a loss of pump performance (flow/retention time
stability), a blockage or internal leak of the pump. Avoid the use of such solvent
combinations, as they can cause irreproducible chromatographic results.
The use of an Inline filter (G1311-60006) can avoid or reduce such effects by
filtering crystals and re-dissolving them over time.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 105
5 Optimizing Performance
When to use the Seal Wash Function

When to use the Seal Wash Function

Highly concentrated buffer solutions will reduce the lifetime of the seals and
pistons in your pump. The seal wash function allows to maintain the seal lifetime
by flushing the back side of the seal with a wash solvent.
The seal wash function is strongly recommended when buffer concentrations of
0.1 M or higher will be used for long time periods in the pump.
Active seal wash can be ordered as factory installed option or can be upgraded
later by ordering the field upgrade kit G1398A.
The seal wash function comprises a support ring, secondary seal, gasket and
seal holder for both piston sides. Place a wash bottle filled with 90 % water / 10 %
isopropanol above the pump in the solvent cabinet and guide the SW waste into
the waste bottle underneath the LC stack. The peristaltic pump moves a flow
through the pump head removing all possible buffer crystals from the back of the
pump seal. This mixture prevents growth of algae or bacteria in the wash bottle
and reduces the surface tension of the water.
For information on the installation of the active seal wash function refer to “Install
an Active Seal Wash Upgrade Kit (Quaternary Pump)” on page 241.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 106
5 Optimizing Performance
Choosing the Right Pump Seals

Choosing the Right Pump Seals

The standard seal for the pump can be used for most applications. However
applications that use normal phase solvents (for example, hexane) are not suited
for the standard seal and require a different seal when used for a longer time in
the pump.
For applications that use normal phase solvents (for example, hexane) we
recommend using polyethylene pump seals (PE seals (pack of 2) (0905-1420))
and Wash Seal PE (0905-1718). For normal phase applications, these seals have
less abrasion compared to the standard seals.

Polyethylene seals have a limited pressure range of 0 – 200 bar. When used
NOTE above 200 bar their lifetime is reduced significantly.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 107
5 Optimizing Performance
Optimize the Compressibility Compensation Setting

Optimize the Compressibility Compensation


Setting

The compressibility compensation default setting is 100 × 10-6 /bar for the
pump. This setting represents an average value. Under normal conditions the
default setting typically reduces the pressure pulsation to values below 1 % of
system pressure that are sufficient for most applications and for all gradient
analyses. The compressibility settings can be optimized by using the values for
the various solvents described in Table 19 on page 109. If the solvent in use is
not listed in the compressibility tables, when using isocratic mixtures of solvents
and if the default settings are not sufficient for your application the following
procedure can be used to optimize the compressibility settings.

When using mixtures of solvents it is not possible to calculate the


NOTE compressibility of the mixture by interpolating the compressibility values of the
pure solvents used in that mixture or by applying any other calculation. In these
cases the following empirical procedure has to be applied to optimize your
compressibility setting.
Inappropriate settings would mainly affect retention times of peaks eluted at the
beginning of a gradient. Therefore optimize settings for the solvent at the
beginning of the gradient. For mixtures containing up to 50 % water, use
compressibility settings of water.

1 Start the pump with the required flow rate.


2 Before starting the optimization procedure, the flow must be stable. Check the
tightness of the system with the pressure test.
3 Your pump must be connected to a data system or Instant Pilot with which
the pressure and %-ripple can be monitored, or connect an external
measurement device to the analog pressure output (see “Electrical
Connections” on page 307)
4 Starting with a compressibility setting of 40·10-6 /bar increase the value in
steps of 10. Re-zero the signal display as required. The compressibility
compensation setting that generates the smallest pressure ripple is the
optimum value for your solvent composition. If ChemStation (classic view) is
used and the ripple shown is positive, then the compressibility setting should
be decreased. If it is negative, it should be increased.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 108
5 Optimizing Performance
Optimize the Compressibility Compensation Setting

Table 19 Solvent Compressibility

Solvent (pure) Compressibility (1·10-6 /bar)

Acetone 126

Acetonitrile 115

Benzene 95

Carbon tetrachloride1 110

Chloroform1 100

Cyclohexane 118

Ethanol 114

Ethyl acetate 104

Heptane 120

Hexane 150

Isobutanol 100

Isopropanol 100

Methanol 120

1-Propanol 100

Toluene 87

Water 46

1 Please check section "Solvent Information" for compatibility to your specific LC system.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 109
6 Troubleshooting and Diagnostics

User Interfaces 111


Agilent Lab Advisor Software 112

This chapter gives an overview about the troubleshooting and diagnostic


features and the different user interfaces.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 110
6 Troubleshooting and Diagnostics
User Interfaces

User Interfaces

Depending on the user interface, the available tests vary. Some descriptions are
only available in the service manual.
Table 20 Test functions available vs. user interface

Test Instant Pilot G4208A Agilent Lab Advisor

System Pressure Test Yes Yes

Pump Leak Rate Test No Yes

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 111
6 Troubleshooting and Diagnostics
Agilent Lab Advisor Software

Agilent Lab Advisor Software

The Agilent Lab Advisor Software (basic license, shipped with an Agilent LC
pump) is a standalone product that can be used with or without a
chromatographic data system. Agilent Lab Advisor helps to manage the lab for
high-quality chromatographic results by providing a detailed system overview of
all connected analytical instruments with instrument status, Early Maintenance
Feedback counters (EMF), instrument configuration information, and diagnostic
tests. With the push of a button, a detailed diagnostic report can be generated.
Upon request, the user can send this report to Agilent for a significantly improved
troubleshooting and repair process.
The Agilent Lab Advisor software is available in two versions:
• Lab Advisor Basic
• Lab Advisor Advanced
Lab Advisor Basic is included with every Agilent 1200 Infinity Series and Agilent
InfinityLab LC Series instrument.
The Lab Advisor Advanced features can be unlocked by purchasing a license key,
and include real-time monitoring of instrument actuals, all various instrument
signals, and state machines. In addition, all diagnostic test results, calibration
results, and acquired signal data can be uploaded to a shared network folder. The
Review Client included in Lab Advisor Advanced allows to load and examine the
uploaded data no matter on which instrument it was generated. This makes Data
Sharing an ideal tool for internal support groups and users who want to track the
instrument history of their analytical systems.
The optional Agilent Maintenance Wizard Add-on provides an easy-to-use,
step-by-step multimedia guide for performing preventive maintenance on Agilent
1200 Infinity LC Series instrument.
The tests and diagnostic features that are provided by the Agilent Lab Advisor
software may differ from the descriptions in this manual. For details, refer to the
Agilent Lab Advisor software help files.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 112
7 Error Information

What Are Error Messages 114


General Error Messages 115
Timeout 115
Shutdown 116
Remote Timeout 117
Lost CAN Partner 117
Leak 118
Leak Sensor Open 119
Leak Sensor Short 120
Compensation Sensor Open 121
Compensation Sensor Short 121
Fan Failed 122
ERI Messages 122
Module Error Messages 123
Solvent Zero Counter 123
Pressure Above Upper Limit 123
Pressure Below Lower Limit 124
Pressure Signal Missing 124
MCGV Fuse 125
AIV Fuse 125
Valve Failed (MCGV) 126
Motor-Drive Power 127
Servo Restart Failed 128
Pump Head Missing 129
Index Limit 129
Index Adjustment 130
Index Missing 130
Initialization Failed 131
Wait Timeout 132
Degasser: signal fail 133
Degasser: Vacuum cannot be maintained 133
Degasser: limit not reached 134
SSV Protection circuit triggered 134

This chapter describes the meaning of error messages, and provides information
on probable causes and suggested actions how to recover from error conditions.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 113
7 Error Information
What Are Error Messages

What Are Error Messages

Error messages are displayed in the user interface when an electronic,


mechanical, or hydraulic (flow path) failure occurs which requires attention
before the analysis can be continued (for example, repair, or exchange of
consumables is necessary). In the event of such a failure, the red status indicator
at the front of the module is switched on, and an entry is written into the module
logbook.
If an error occurs outside a method run, other modules will not be informed about
this error. If it occurs within a method run, all connected modules will get a
notification, all LEDs get red and the run will be stopped. Depending on the
module type, this stop is implemented differently. For example, for a pump the
flow will be stopped for safety reasons. For a detector, the lamp will stay on in
order to avoid equilibration time. Depending on the error type, the next run can
only be started, if the error has been resolved, for example liquid from a leak has
been dried. Errors for presumably single time events can be recovered by
switching on the system in the user interface.
Special handling is done in case of a leak. As a leak is a potential safety issue and
may have occurred at a different module from where it has been observed, a leak
always causes a shutdown of all modules, even outside a method run.
In all cases, error propagation is done via the CAN bus or via an APG/ERI remote
cable (see documentation for the APG/ERI interface).

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 114
7 Error Information
General Error Messages

General Error Messages

General error messages are generic to all Agilent series HPLC modules and may
show up on other modules as well.

Timeout
Error ID: 0062
The timeout threshold was exceeded.

Probable cause Suggested actions

1 The analysis was completed successfully, Check the logbook for the occurrence and source
and the timeout function switched off the of a not-ready condition. Restart the analysis
module as requested. where required.

2 A not-ready condition was present during a Check the logbook for the occurrence and source
sequence or multiple-injection run for a of a not-ready condition. Restart the analysis
period longer than the timeout threshold. where required.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 115
7 Error Information
General Error Messages

Shutdown
Error ID: 0063
An external instrument has generated a shutdown signal on the remote line.
The module continually monitors the remote input connectors for status signals.
A LOW signal input on pin 6 of the Enhanced Remote Interface (ERI) connector
generates the error message.

Probable cause Suggested actions

1 Leak detected in another module with a CAN Fix the leak in the external instrument before
connection to the system. restarting the module.

2 Leak detected in an external instrument with Fix the leak in the external instrument before
a remote connection to the system. restarting the module.

3 Shut-down in an external instrument with a Check external instruments for a shut-down


remote connection to the system. condition.

4 The degasser failed to generate sufficient Check the vacuum degasser for an error
vacuum for solvent degassing. condition. Refer to the Service Manual for the
degasser or the pump that has the degasser
built-in.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 116
7 Error Information
General Error Messages

Remote Timeout
Error ID: 0070
A not-ready condition is still present on the remote input. When an analysis is
started, the system expects all not-ready conditions (for example, a not-ready
condition during detector balance) to switch to run conditions within one minute
of starting the analysis. If a not-ready condition is still present on the remote line
after one minute the error message is generated.

Probable cause Suggested actions

1 Not-ready condition in one of the Ensure the instrument showing the not-ready
instruments connected to the remote line. condition is installed correctly, and is set up
correctly for analysis.

2 Defective remote cable. Exchange the remote cable.

3 Defective components in the instrument Check the instrument for defects (refer to the
showing the not-ready condition. instrument’s documentation).

Lost CAN Partner


Error ID: 0071
During an analysis, the internal synchronization or communication between one
or more of the modules in the system has failed.
The system processors continually monitor the system configuration. If one or
more of the modules is no longer recognized as being connected to the system,
the error message is generated.

Probable cause Suggested actions

1 CAN cable disconnected. • Ensure all the CAN cables are connected
correctly.
• Ensure all CAN cables are installed correctly.

2 Defective CAN cable. Exchange the CAN cable.

3 Defective mainboard in another module. Switch off the system. Restart the system, and
determine which module or modules are not
recognized by the system.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 117
7 Error Information
General Error Messages

Leak
Error ID: 0064
A leak was detected in the module.
The signals from the two temperature sensors (leak sensor and board-mounted
temperature-compensation sensor) are used by the leak algorithm to determine
whether a leak is present. When a leak occurs, the leak sensor is cooled by the
solvent. This changes the resistance of the leak sensor which is sensed by the
leak sensor circuit on the mainboard.

Probable cause Suggested actions

1 Loose fittings. Ensure all fittings are tight.

2 Broken capillary. Exchange defective capillaries.

3 Loose or leaking purge valve, inlet valve, or Ensure pump components are seated correctly.
outlet valve. If there are still signs of a leak, exchange the
appropriate seal (purge valve, inlet valve, outlet
valve).

4 Defective pump seals. Exchange the pump seals.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 118
7 Error Information
General Error Messages

Leak Sensor Open


Error ID: 0083
The leak sensor in the module has failed (open circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak sensor current to
change within defined limits. If the current falls outside the lower limit, the error
message is generated.

Probable cause Suggested actions

1 Leak sensor not connected to the power Ensure the leak sensor is connected correctly.
switch board.

2 Defective leak sensor. Exchange the leak sensor.

3 Leak sensor incorrectly routed, being Exchange the leak sensor.


pinched by a metal component.

4 Power switch assembly defective. Exchange the power switch assembly.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 119
7 Error Information
General Error Messages

Leak Sensor Short


Error ID: 0082
The leak sensor in the module has failed (short circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak sensor current to
change within defined limits. If the current increases above the upper limit, the
error message is generated.

Probable cause Suggested actions

1 Defective leak sensor. Exchange the leak sensor.

2 Leak sensor incorrectly routed, being • Make sure the leak sensor is installed
pinched by a metal component. correctly.
• Correct the routing of the cable.
• If cable defective, exchange the leak sensor.

3 Power switch assembly defective. Exchange the power switch assembly.

4 Cable or contact problem. • Inspect the flat ribbon cable from the power
switch board to the mainboard.
• Inspect the contacts on the flat ribbon
cable's ends.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 120
7 Error Information
General Error Messages

Compensation Sensor Open


Error ID: 0081
The ambient-compensation sensor (NTC) on the power switch board in the
module has failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the
power switch board is dependent on ambient temperature. The change in
resistance is used by the leak circuit to compensate for ambient temperature
changes. If the resistance across the sensor increases above the upper limit, the
error message is generated.

Probable cause Suggested actions

1 Loose connection between the power switch Check the ribbon cable between the power
board and the mainboard. switch board and the mainboard.

2 Defective power switch assembly. Exchange the power switch assembly.

Compensation Sensor Short


Error ID: 0080
The ambient-compensation sensor (NTC) on the power switch board in the
module has failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the
power switch board is dependent on ambient temperature. The change in
resistance is used by the leak circuit to compensate for ambient temperature
changes. If the resistance across the sensor falls below the lower limit, the error
message is generated.

Probable cause Suggested actions

1 Defective power switch assembly. Exchange the power switch assembly.

2 Loose connection between the power switch Check the ribbon cable between the power
board and the mainboard. switch board and the mainboard.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 121
7 Error Information
General Error Messages

Fan Failed
Error ID: 0068
The cooling fan in the module has failed.
The hall sensor on the fan shaft is used by the mainboard to monitor the fan
speed. If the fan speed falls below a certain limit for a certain length of time, the
error message is generated.
This limit is given by 2 revolutions/second for longer than 5 seconds.
Depending on the module, assemblies (e.g. the lamp in the detector) are turned
off to assure that the module does not overheat inside.

Probable cause Suggested actions

1 Fan cable disconnected. Ensure the fan is connected correctly.

2 Defective fan. Exchange fan.

3 Defective mainboard. Exchange the mainboard.

4 Improperly positioned cables or wires Ensure the fan is not mechanically blocked.
obstructing fan blades.

ERI Messages
Error ID: 11120 (+5 V), 11121 (+25 V)
The ERI (Enhanced Remote Interface) provides two error events related to over
current situations on the +5 V and +24 V lines.

Probable cause Suggested actions

1 The load on the ERI is too high. Reduce the load.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 122
7 Error Information
Module Error Messages

Module Error Messages

These errors are pump specific.

Solvent Zero Counter


Error ID: 2055, 2524
Pump firmware version A.02.32 and higher allow to set solvent bottle fillings in
the data system. If the volume level in the bottle falls below the specified value
the error message appears when the feature is configured accordingly.

Probable cause Suggested actions

1 Volume in bottle below specified volume. Refill bottles and reset solvent counters.

2 Incorrect setting. Make sure the limits are set correctly.

Pressure Above Upper Limit


Error ID: 2014, 2500
The system pressure has exceeded the upper pressure limit.

Probable cause Suggested actions

1 Upper pressure limit set too low. Ensure the upper pressure limit is set to a value
suitable for the analysis.

2 Blockage in the flowpath (after the damper). Check for blockage in the flow path. The
following components are particularly subject to
blockage: inline filter frit, needle (autosampler),
seat capillary (autosampler), sample loop
(autosampler), column frits and capillaries with
small internal diameters (e.g. 50 µm ID).

3 Defective damper. Exchange the damper.

4 Defective mainboard. Exchange the mainboard.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 123
7 Error Information
Module Error Messages

Pressure Below Lower Limit


Error ID: 2015, 2501
The system pressure has fallen below the lower pressure limit.

Probable cause Suggested actions

1 Lower pressure limit set too high. Ensure the lower pressure limit is set to a value
suitable for the analysis.

2 Air bubbles in the mobile phase. • Make sure that the degasser is in flow path
and works correctly. Purge the module.
• Ensure solvent inlet filters are not blocked.

3 Leak. • Inspect the pump head, capillaries and


fittings for signs of a leak.
• Purge the module. Run a pressure test to
determine whether the seals or other module
components are defective.

4 Defective damper. Exchange the damper.

5 Defective mainboard. Exchange the mainboard.

Pressure Signal Missing


Error ID: 2016
The pressure signal is missing.
The pressure signal must be within a specific voltage range. If the pressure signal
is missing, the processor detects a voltage of approximately -120 mV across the
damper connector.

Probable cause Suggested actions

1 Damper disconnected. Ensure the damper is connected correctly to the


mainboard.

2 Defective damper. Exchange the damper.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 124
7 Error Information
Module Error Messages

MCGV Fuse
Error ID: 2043
Valve Fuse 0: Channels A and B
Valve Fuse 1: Channels C and D
The gradient valve in the quaternary pump has drawn excessive current causing
the electronic fuse to open.

Probable cause Suggested actions

1 Defective gradient valve. Restart the quaternary pump. If the error


message appears again, exchange the gradient
valve.

2 Defective connection cable (front panel to Exchange the connection cable.


mainboard).

3 Defective mainboard. Exchange the mainboard.

AIV Fuse
Error ID: 2044
The active-inlet valve in the module has drawn excessive current causing the
inlet-valve electronic fuse to open.

Probable cause Suggested actions

1 Defective active inlet valve. Restart the module. If the error message appears
again, exchange the active inlet valve.

2 Defective connection cable (front panel to Exchange the connection cable.


mainboard).

3 Defective mainboard. Exchange the mainboard.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 125
7 Error Information
Module Error Messages

Valve Failed (MCGV)


Error ID: 2040
Valve 0 Failed: valve A
Valve 1 Failed: valve B
Valve 2 Failed: valve C
Valve 3 Failed:valve D
One of the valves of the multi-channel gradient valve has failed to switch
correctly.
The processor monitors the valve voltage before and after each switching cycle.
If the voltages are outside expected limits, the error message is generated.

Probable cause Suggested actions

1 Gradient valve disconnected. Ensure the gradient valve is connected correctly.

2 Connection cable (inside instrument) not Ensure the connection cable is connected
connected. correctly.

3 Connection cable (inside instrument) Exchange the connection cable.


defective.

4 Gradient valve defective. Exchange the gradient valve.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 126
7 Error Information
Module Error Messages

Motor-Drive Power
Error ID: 2041, 2042
The current drawn by the pump motor exceeded the maximum limit.
Blockages in the flow path are usually detected by the pressure sensor in the
damper, which result in the pump switching off when the upper pressure limit is
exceeded. If a blockage occurs before the damper, the pressure increase cannot
be detected by the pressure sensor and the module will continue to pump. As
pressure increases, the pump drive draws more current. When the current
reaches the maximum limit, the module is switched off, and the error message is
generated.

Probable cause Suggested actions

1 Flow path blockage in front of the damper. Ensure the capillaries and frits between the
pump head and damper inlet are free from
blockage.

2 Blocked (passive or active) inlet valve. Exchange the (passive or active) inlet valve.

3 Blocked outlet valve. Exchange the outlet valve.

4 High friction (partial mechanical blockage) in Remove the pump-head assembly. Ensure there
the pump drive assembly. is no mechanical blockage of the pump-head
assembly or pump drive assembly.

5 Defective pump drive assembly. • Remove the pump head assembly. Ensure
there is no mechanical blockage of the pump
head assembly or pump drive assembly.
• Exchange the pump drive assembly.

6 Defective mainboard. Exchange the mainboard.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 127
7 Error Information
Module Error Messages

Servo Restart Failed


Error ID: 2201, 2211
The pump motor in the module was unable to move into the correct position for
restarting.
When the module is switched on, the first step is to switch on the C phase of the
variable reluctance motor. The rotor should move to one of the C positions. The C
position is required for the servo to be able to take control of the phase
sequencing with the commutator. If the rotor is unable to move, or if the C
position cannot be reached, the error message is generated.

Probable cause Suggested actions

1 Mechanical blockage of the module. Remove the pump-head assembly. Ensure there
is no mechanical blockage of the pump-head
assembly or pump drive assembly.

2 Disconnected or defective cable. Ensure the pump assembly cables are not
damaged or dirty. Make sure the cables are
connected securely to the mainboard.

3 Blocked (passive or active) inlet valve. Exchange the (passive or active) inlet valve.

4 Defective pump drive assembly. • Remove the pump head assembly. Ensure
there is no mechanical blockage of the pump
head assembly or pump drive assembly.
• Exchange the pump drive assembly.

5 Defective mainboard. Exchange the mainboard.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 128
7 Error Information
Module Error Messages

Pump Head Missing


Error ID: 2202, 2212
The pump-head end stop in the pump was not found.
When the pump restarts, the metering drive moves forward to the mechanical
end stop. Normally, the end stop is reached within 20 s, indicated by an increase
in motor current. If the end point is not found within 20 s, the error message is
generated.

Probable cause Suggested actions

1 Pump head not installed correctly (screws Install the pump head correctly. Ensure nothing
not secured, or pump head not seated (e.g. capillary) is trapped between the pump head
correctly). and body.

2 Broken piston. Exchange the piston.

Index Limit
Error ID: 2203, 2213
The time required by the piston to reach the encoder index position was too short
(pump).
During initialization, the first piston is moved to the mechanical stop. After
reaching the mechanical stop, the piston reverses direction until the encoder
index position is reached. If the index position is reached too fast, the error
message is generated.

Probable cause Suggested actions

1 Irregular or sticking drive movement. Remove the pump head, and examine the seals,
pistons, and internal components for signs of
wear, contamination or damage. Exchange
components as required.

2 Defective pump drive assembly. Exchange the pump drive assembly.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 129
7 Error Information
Module Error Messages

Index Adjustment
Error ID: 2204, 2214
The encoder index position in the module is out of adjustment.
During initialization, the first piston is moved to the mechanical stop. After
reaching the mechanical stop, the piston reverses direction until the encoder
index position is reached. If the time to reach the index position is too long, the
error message is generated.

Probable cause Suggested actions

1 Irregular or sticking drive movement. Remove the pump head, and examine the seals,
pistons, and internal components for signs of
wear, contamination or damage. Exchange
components as required.

2 Defective pump drive assembly. Exchange the pump drive assembly.

Index Missing
Error ID: 2205, 2215, 2505
The encoder index position in the module was not found during initialization.
During initialization, the first piston is moved to the mechanical stop. After
reaching the mechanical stop, the piston reverses direction until the encoder
index position is reached. If the index position is not recognized within a defined
time, the error message is generated.

Probable cause Suggested actions

1 Disconnected or defective encoder cable. Ensure the encoder cables are not damaged or
dirty. Make sure the cables are connected
securely to the mainboard.

2 Defective pump drive assembly. Exchange the pump drive assembly.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 130
7 Error Information
Module Error Messages

Initialization Failed
Error ID: 2207, 2217
The module failed to initialize successfully within the maximum time window.
A maximum time is assigned for the complete pump-initialization cycle. If the
time is exceeded before initialization is complete, the error message is
generated.

Probable cause Suggested actions

1 Blocked (passive or active) inlet valve. Exchange the (passive or active) inlet valve.

2 Defective pump drive assembly. • Remove the pump head assembly. Ensure
there is no mechanical blockage of the pump
head assembly or pump drive assembly.
• Exchange the pump drive assembly.

3 Defective mainboard. Exchange the mainboard.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 131
7 Error Information
Module Error Messages

Wait Timeout
Error ID: 2053
When running certain tests in the diagnostics mode or other special applications,
the pump must wait for the pistons to reach a specific position, or must wait for a
certain pressure or flow to be reached. Each action or state must be completed
within the timeout period, otherwise the error message is generated.
Possible Reasons for a Wait Timeout:
· Pressure not reached.
· Pump channel A did not reach the delivery phase.
· Pump channel B did not reach the delivery phase.
· Pump channel A did not reach the take-in phase.
· Pump channel B did not reach the take-in phase.
· Solvent volume not delivered within the specified time.

Probable cause Suggested actions

1 Purge valve open. Ensure that purge valve is closed.

2 Leak at fittings, purge valve, active inlet valve, • Ensure pump components are seated
outlet valve or piston seals. correctly. If there are still signs of a leak,
exchange the appropriate seal (purge valve,
active inlet valve, outlet valve, piston seal).
• Exchange defective capillaries.

3 Flow changed after starting test. Ensure correct operating condition for the
special application in use.

4 Defective pump drive assembly. Exchange the defective pump drive assembly.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 132
7 Error Information
Module Error Messages

Degasser: signal fail


Error ID: 2243, 8016
The pump board gets no or wrong pressure signals from the built-in degasser:
• No valid pressure signal is shown during startup of the degasser
• Measured pressure is higher than 32000 hPa

Probable cause Suggested actions

1 Degasser sensor defect. Replace the degasser sensor.

2 Degasser sensor not connected to Connect the degasser sensor to the mainboard.
mainboard.

Degasser: Vacuum cannot be maintained


Error ID: 8014
Pressure in degasser vacuum chamber exceeds 180 mbar limit during normal
operation.

Probable cause Suggested actions

1 Liquid in degasser tubing. Wait for evaporation of liquid. Then replace the
tubing, degasser chamber, and vacuum pump.

2 Leak in degasser tubing or chamber. • Replace tubing.


• Replace degasser chamber.

3 Degasser vacuum pump defective. Replace vacuum pump.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 133
7 Error Information
Module Error Messages

Degasser: limit not reached


Error ID: 2244, 8053
Degasser is not ready within 8 min after startup, because the pressure inside the
vacuum chamber exceeds 120 mbar.

Probable cause Suggested actions

1 Liquid in degasser tubing. • Wait for evaporation of liquid.


• Replace tubing.
• Replace degasser chamber.
• Replace vacuum pump.

2 Leak in degasser tubing or chamber. • Replace tubing.


• Replace degasser chamber.

3 Degasser vacuum pump defect. Replace vacuum pump.

SSV Protection circuit triggered


Error ID: 2049
The protection circuit of SSV has switched OFF the valve. During initialization, the
processor checks the presence of the SSV connector. If a defective valve or a
wrong connection is detected the error message is generated.

Probable cause Suggested actions

1 Wrong valve connected. Check if the correct valve is connected to the


correct port.

2 Defective Solvent Selection Valve. Change the Solvent Selection Valve.

3 Defective mainboard. Replace mainboard.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 134
8 Test Functions and Calibration

Introduction 136
System Pressure Test 137
Running the Test 139
Evaluating the Results 140
Potential Causes of System Pressure Test Failure 141
Leak Rate Test 142
Running the Test 144
Evaluating the Results 145
Potential Causes of Leak Rate Test Failure 146

This chapter describes the tests for the module.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 135
8 Test Functions and Calibration
Introduction

Introduction

Following tests are available in Lab Advisor:


• System Pressure Test
• Leak Rate Test
All four pumps are available with LabAdvisor B.02.08.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 136
8 Test Functions and Calibration
System Pressure Test

System Pressure Test

Introduction
The System Pressure Test is used for checking the tightness of the LC system
and identifying leaks between the pump and a position in the flow path following
the pump blocked by a blank nut.

System requirements
Minimum software revisions:
• Lab Advisor B.02.08
Minimum firmware revision: D.07.01.

Test Principle
A solvent can be chosen from available solvent channels and a maximum
pressure can be defined at which the test will be run. In contrast to older
revisions of this test, any solvent can be used.
Before the test, the pump and system are flushed with solvent in order to remove
air bubbles, as air bubbles are compressed during the test and therefore would
appear as leaks. Using a degasser is highly recommended. Then the flow path is
blocked by a blank nut at any position between the purge valve and the column
outlet.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 137
8 Test Functions and Calibration
System Pressure Test

pressure

2
1

time

flow

leak rate

time

In the first phase of the test, the pump delivers flow at a rate of 200 µL/min until a
pressure of 50 bar below the defined maximum pressure is reached. In the
second phase, the pump delivers a small flow which is increased stepwise. If
there is a leak in the system, the pressure will drop initially, as the low flow cannot
compensate the leak flow. As soon as the pump flow rate exceeds the leak flow
rate, the pressure will increase again and the test is stopped at about 20 bar
below the maximum pressure. The point in phase 2, where the lowest pressure is
reached and stays constant for a short time corresponds to the leak rate, that is
provided as a test result. A leak rate smaller than 3 µL/min is good enough for
operating the pump reliably.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 138
8 Test Functions and Calibration
System Pressure Test

Running the Test

Parts required p/n Description


5067-6127 Blank Nut SL

Damage to pressure sensitive parts


C AU T I O N
Even columns that are suitable for high pressures are sensitive to pressure
drops that occur during this test.
 Do not include any pressure sensitive parts to the flow path and choose a
maximum pressure that is compatible to your system. For example, do not
include columns, a standard pressure flow cell (up to 20 bar) or a 400 bar
autosampler to a 600 bar pressure test.

Running the test from the Agilent Lab Advisor


1 Select the System Pressure Test from the Test Selection menu.
2 Start the test and follow the instructions.

Make sure to release the pressure by slowly opening the purge valve when the
NOTE test has finished.

“Evaluating the Results” on page 140 describes the evaluation and interpretation
of the System Pressure Test results. For detailed instructions refer to the Agilent
Lab Advisor software.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 139
8 Test Functions and Calibration
System Pressure Test

Evaluating the Results


The test fails, if the leak rate between pump and blank nut is higher than the limit
of 5 µL/min.
If the System Pressure Test fails:
• Ensure that all fittings between the pump and the blank nut are tight.
• Repeat the test.

Often it is only a damaged blank nut itself (poorly shaped from overtightening)
NOTE that causes a failure of the test. Before investigating on any other possible
sources of failure make sure that the blank nut you are using is in good condition
and properly tightened!

If the test fails again, insert the blank nut at the outlet of the previous module in
the stack (e.g. autosampler outlet if TCC has been tested before), and repeat the
test. Exclude each module one by one to determine which module is leaking.
If the pump is determined to be the source of the leak, run the Pump Leak Rate
Test.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 140
8 Test Functions and Calibration
System Pressure Test

Potential Causes of System Pressure Test Failure


System Pressure Test failed
The test will fail, if the sum of all leaks in the system (pump, autosampler or
column compartment and connections) exceeds the test limit. After isolating and
fixing the cause of the leak, repeat the System Pressure Test to confirm the
system is pressure tight.

Probable cause Suggested actions

1 Purge valve open. Close the purge valve.

2 Loose or leaky fittings. Tighten the fitting or exchange the capillary.

3 Pump: Damaged pump seals or pistons. Run the Leak Rate Test to confirm the leak.

4 Loose purge valve. Tighten the purge valve nut (14 mm wrench).

5 Autosampler: Loose or leaky fitting. Tighten or exchange the fitting or capillary.

6 Autosampler: Rotor seal (injection valve). Exchange the rotor seal.

7 Autosampler: Damaged metering seal or Exchange the metering seal. Check the piston for
piston. scratches. Exchange the piston if required.

8 Autosampler: Needle seat. Exchange the needle seat.

9 Column compartment: Loose or leaky fitting. Tighten or exchange the fitting or capillary.

10 Column compartment: Rotor seal in optional Exchange the rotor seal.


valve.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 141
8 Test Functions and Calibration
Leak Rate Test

Leak Rate Test

Introduction
The Leak Rate Test is used for verifying the internal tightness of the pump and
helps identifying parts which may have caused a leak.

System requirements
Minimum software revisions:
• Lab Advisor B.02.08
Minimum firmware revisions:
• D.07.01

Test Principle
A solvent can be chosen from available solvent channels and a maximum target
pressure can be defined at which the test will be run. Typically, this is the
maximum pressure specified for the pump. The test can be run with any solvent
compatible to the pump.
Before the test, the pump is flushed with solvent in order to remove air bubbles,
as air bubbles are compressed during the test and therefore would appear as
leaks. Using a degasser is highly recommended.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 142
8 Test Functions and Calibration
Leak Rate Test

piston 2
at rear position
piston 1
at rear position

piston 2
delivers
piston 1
delivers

about 100 bar


below target
pressure

Initially, the pressure is increased to about 100 bar below the target pressure,
which has been set for the test.
Then piston 1 is brought to its rear position. An increasing flow is delivered by
piston 1. In case of a leak, the pressure will drop initially as long as the flow rate
delivered by the piston is lower than the leak rate. As soon as the flow rate of the
piston exceeds the leak rate, the measured pressure will increase again.
Therefore the minimum pressure of that curve segment corresponds to the flow
and leak rate at that time and the leak rate is measured. Compare to the
description of the system pressure test ( “System Pressure Test” on page 137).
Subsequently, piston 2 is moved to its rear position, then piston 2 delivers and the
measurement is done as described for piston 1.
For a binary pump (G1312B/C, G7112B, K1312B, G4782A), the test is run for both
pump heads for channels A and B.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 143
8 Test Functions and Calibration
Leak Rate Test

Running the Test

Parts required p/n Description


5067-6127 Blank Nut SL

Running the test from the Agilent Lab Advisor


1 Select the Leak Rate Test from the Test Selection menu.
2 Start the test and follow the instructions.

Make sure to release the pressure by slowly opening the purge valve when the
NOTE test has finished.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 144
8 Test Functions and Calibration
Leak Rate Test

Evaluating the Results


Results of the leak rate test are the leak rates measured for pistons 1 and 2 as
described for the test principle. If any of the leak rates exceeds 3 µL/min, the test
will fail.

to column

Damper

Outlet valve

Purge valve

Inlet valve to waste

Seals

Pump chamber 1 Pump chamber 2


Piston 1

Piston 2

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 145
8 Test Functions and Calibration
Leak Rate Test

Potential Causes of Leak Rate Test Failure


Secondary Leak
If a leak is found for movement of piston 2 (secondary leak), the following
reasons are possible:

Probable cause Suggested actions

1 System not flushed properly Flush system for several minutes

2 Degassing efficiency is low Check degasser performance

3 Purge valve not closed or defect Check purge valve

4 Blank nut not installed tightly Tighten or replace blank nut

5 Outlet valve leaking (read below) Replace outlet valve

6 Leak at piston 2 or seal in chamber 2 Inspect piston, replace piston and/or seal

Primary Leak
If a leak is found for movement of piston 1 (primary leak), any leak described for
piston movement 2 will cause a failure for piston 1 as well, as the liquid can move
through the outlet valve to chamber 2. Such cases need to be identified as
described before. Additionally, following causes are possible:

Probable cause Suggested actions

1 Leak at piston 1 or seal in chamber 1 Inspect piston, replace piston and/or seal

2 Leak at inlet valve Replace inlet valve or inlet valve cartridge (AIV
only)

Internal Outlet Valve Leak


A leak of the outlet valve will be identified separately (internal outlet valve leak) by
calculating the difference between leak rate 1 and leak rate 2. If the second leak
rate is higher than the first one, this is due to a flow back through the outlet valve.

Probable cause Suggested actions

1 Leak at outlet valve Replace the part which has failed and re-run the
test.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 146
9 Maintenance and Repair

Introduction to Maintenance and Repair 149


Cautions and Warnings 150
Overview of Maintenance and Repair 152
Cleaning the Module 153
Remove and Install Doors 154
Checking and Replacing the Solvent Filter 156
Exchange the Passive Inlet Valve 157
Exchange the Outlet Valve 159
Exchange the Purge Valve Frit or the Purge Valve 161
Replace the O-Ring on the Purge Valve 164
Remove the Pump Head Assembly 166
Maintenance of a Pump Head Without Seal Wash Option 168
Maintenance of a Pump Head with Seal Wash Option 171
Reinstall the Pump Head Assembly 176
Seal Wear-in Procedure 179
Exchange the Multi-Channel Gradient Valve (MCGV) 180
Replace the Solvent Selection Valve (Isocratic Pump) 183
Exchange the Active Inlet Valve (AIV) or its Cartridge 185
Exchange the Seal Wash Cartridge 187
Replace Leak Handling System Parts 188
Replacing the Module Firmware 189
Introduction to Repair 190
Overview of Repairing the Pump 191
Exchanging Internal Parts 193
Remove the Top Cover and Foam 194
Exchange the Pump Mainboard 196
Replace the Fusion Board 202

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 147
Changing the Type and Serial Number 203
Recover Instructions 205
Exchange the Damper 208
Exchange the Pump Drive 212
Exchange the Fan 214
Replace Power Supply and Cables 216
Replace the Power Switch 219
Replace the Degasser 221
Replace Parts Inside the 4-Channel-Degasser 223
Replace a Degasser Chamber 225
Replace the Degasser Pump 226
Replace the Sensor Board Assembly 227
Replace the Degasser Tubing Kit 229
Replace Parts Inside the 2-Channel-Degasser 230
Replace a Degasser Chamber 231
Replace the Degasser Pump 232
Replace the Sensor Board Assembly 234
Replace the Degasser Tubing Kit 235
Replace the Leak Panel 236
Replace the Leak Sensor 238
Replace the Status LED 239
Install an Active Seal Wash Upgrade Kit (Quaternary Pump) 241
Assemble the Main Cover 244
Install the Top Cover and Foam 245

This chapter describes the maintenance and repair of the module.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 148
9 Maintenance and Repair
Introduction to Maintenance and Repair

Introduction to Maintenance and Repair

The module is designed for easy repair. The most frequent repairs such as piston
seal change and purge valve frit change can be done from the front of the
module with the module in place in the system stack.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 149
9 Maintenance and Repair
Cautions and Warnings

Cautions and Warnings

The module is partially energized when switched off, as long as the power
WAR N IN G
cord is plugged in.
Repair work at the module can lead to personal injuries, e.g. electrical
shock, when the cover is opened and the module is connected to power.
 Always unplug the power cable before opening the cover.
 Do not connect the power cable to the instrument while the covers are
removed.

Sharp metal edges


WAR N IN G
Sharp-edged parts of the equipment may cause injuries.
 To prevent personal injury, be careful when getting in contact with sharp
metal areas.

Toxic, flammable and hazardous solvents, samples and reagents


WAR N IN G
The handling of solvents, samples and reagents can hold health and safety
risks.
 When working with these substances observe appropriate safety
procedures (for example by wearing goggles, safety gloves and
protective clothing) as described in the material handling and safety
data sheet supplied by the vendor, and follow good laboratory practice.
 The volume of substances should be reduced to the minimum required
for the analysis.
 Do not operate the instrument in an explosive atmosphere.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 150
9 Maintenance and Repair
Cautions and Warnings

Electronic boards and components are sensitive to electrostatic discharge


C AU T I O N (ESD).
ESD can damage electronic boards and components.
 Be sure to hold the board by the edges, and do not touch the electrical
components. Always use ESD protection (for example, an ESD wrist strap)
when handling electronic boards and components.

Safety standards for external equipment


C AU T I O N
 If you connect external equipment to the instrument, make sure that you
only use accessory units tested and approved according to the safety
standards appropriate for the type of external equipment.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 151
9 Maintenance and Repair
Overview of Maintenance and Repair

Overview of Maintenance and Repair

The following pages describe maintenance (simple repairs) of the pump that can
be carried out without opening the main cover.
Table 21 Simple repair procedures

Procedure Typical Frequency Notes

“Checking and Replacing the Solvent If solvent filter is blocked Gradient performance problems, intermittent
Filter” on page 156 pressure fluctuations

“Exchange the Passive Inlet Valve” on If internally leaking Pressure ripple unstable, run Leak Rate Test for
page 157 verification

“Exchange the Active Inlet Valve (AIV) or its If internally leaking Pressure ripple unstable, run Leak Rate Test for
Cartridge” on page 185 verification

“Exchange the Outlet Valve” on page 159 If internally leaking Pressure ripple unstable, run Leak Rate Test for
verification

“Exchange the Purge Valve Frit or the Purge If internally leaking Solvent dripping out of waste outlet when valve
Valve” on page 161 closed

“Exchange the Purge Valve Frit or the Purge If the frit shows indication A pressure drop of > 10 bar across the frit (at a
Valve” on page 161 of contamination or water flow of 5 mL/min with open purge valve)
blockage indicates blockage

“Maintenance of a Pump Head Without Seal If pump performance Leaks at lower pump head side, unstable
Wash Option” on page 168 indicates seal wear retention times, pressure ripple unstable - run
Leak Rate Test for verification

Exchange pistons, see “Maintenance of a If scratched Seal life time shorter than usual - check pistons
Pump Head Without Seal Wash Option” on while changing the seals
page 168

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 152
9 Maintenance and Repair
Cleaning the Module

Cleaning the Module

To keep the module case clean, use a soft cloth slightly dampened with water, or
a solution of water and mild detergent. Avoid using organic solvents for cleaning
purposes. They can cause damage to plastic parts.

Liquid dripping into the electronic compartment of your module can cause
WAR N IN G
shock hazard and damage the module
 Do not use an excessively damp cloth during cleaning.
 Drain all solvent lines before opening any connections in the flow path.

A solution of 70 % isopropanol and 30 % water might be used if the surface of the


NOTE module needs to be disinfected.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 153
9 Maintenance and Repair
Remove and Install Doors

Remove and Install Doors

Parts required p/n Description


5067-5746 Door Assembly Infinity 180 Right
5067-5745 Door Assembly Infinity 180 Left

The figures shown in this procedure exemplarily show the Infinity II Multisampler
NOTE module.
The principle of how to remove and/or install doors works in the same way for all
Infinity II modules.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 154
9 Maintenance and Repair
Remove and Install Doors

1 Press the release buttons and pull the front door out. 2 For the Installation of the front door. Insert the hinges
into their guides and move the door in until the release
buttons click into their final position.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 155
9 Maintenance and Repair
Checking and Replacing the Solvent Filter

Checking and Replacing the Solvent Filter

A functional solvent filter is essential for a good pump performance and for
protecting the LC system.

When If solvent filter is blocked.

Parts required p/n Description


5041-2168 Solvent inlet filter, 20 µm pore size

See “Bottle Head Assembly” on page 264 for related parts.

Small particles can permanently block the capillaries and valves of the module.
C AU T I O N
Damage of the module.
 Always filter solvents.
 Never use the module without solvent inlet filter.

If the filter is in good condition the solvent will freely drip out of the solvent tube
NOTE (hydrostatic pressure). If the solvent filter is partly blocked only very little solvent
will drip out of the solvent tube.

1 Remove the solvent filter from the inlet filter adapter and replace it by a new
one.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 156
9 Maintenance and Repair
Exchange the Passive Inlet Valve

Exchange the Passive Inlet Valve

When If internally leaking (backflow)

Tools required p/n Description


8710-1924 Open-end wrench 14 mm
Pair of tweezers
5067-5688 Torque wrench 1 – 25 Nm with 14 mm wrench

Parts required # p/n Description


1 G1312-60066 Passive inlet valve 1220/1260
OR 1 G1312-60166 1260 Infinity Inlet Valve Type N

Preparations • Switch off the pump at the main power switch.


• Open the doors of the module.

This chapter is not applicable for G5654A Bio-inert pumps!


NOTE

1 Disconnect the solvent inlet tube from the inlet valve. 2 Using a 14 mm wrench loosen the passive inlet valve
and remove the valve from the pump head.
NOTE
Beware of leaking solvents due to hydrostatic pressure.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 157
9 Maintenance and Repair
Exchange the Passive Inlet Valve

3 Insert the new valve into the pump head and tighten the 4 Reconnect the solvent inlet tube to the passive inlet
valve using a torque wrench (approx. 10 Nm). valve.

approx.
10 Nm

5 Close the doors.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 158
9 Maintenance and Repair
Exchange the Outlet Valve

Exchange the Outlet Valve

For bio-inert modules use bio-inert parts only!

When If internally leaking

Tools required p/n Description


8710-0510 Open-end wrench 1/4 — 5/16 inch
8710-1924 Open-end wrench 14 mm
5067-5688 Torque wrench 1 – 25 Nm with 14 mm wrench

Parts required # p/n Description


1 G1312-60067 Outlet valve
OR 1 G1312-60167 Outlet valve (type N/SFC)
OR 1 G5611-60067 Outlet Valve (Bio-Inert)

Preparations • Switch off the pump at the main power switch.


• Open the doors of the module.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 159
9 Maintenance and Repair
Exchange the Outlet Valve

1 Using a 1/4 inch wrench disconnect the valve capillary 2 Using the 14 mm wrench loosen the valve and remove it
from the outlet valve. from the pump body.

NOTE
Do not disassemble the outlet valve, as this can damage the
valve.
3 Reinstall the outlet valve and tighten it using a torque 4 Reconnect the valve capillary to the outlet valve.
wrench set to approx. 10 Nm.

approx.
10 Nm

5 Close the doors.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 160
9 Maintenance and Repair
Exchange the Purge Valve Frit or the Purge Valve

Exchange the Purge Valve Frit or the Purge Valve

For bio-inert modules use bio-inert parts only!

When • Frit – when piston seals are exchanged or when contaminated or blocked (pressure drop of
> 10 bar across the frit at a flow rate of 5 mL/min of water with purge valve opened)
• Purge valve – if internally leaking

Tools required p/n Description


8710-0510 Open-end wrench 1/4 — 5/16 inch
8710-1924 Open-end wrench 14 mm
Pair of tweezers
OR Toothpick

Parts required # p/n Description


1 G7111-60061 Purge valve
for G7110B, G7111A, and G7111B
1 G5611-60064 Purge valve
for G5654A (Bio-inert)
1 01018-22707 PTFE frits (pack of 5)
1 5067-4728 Seal cap (OPTIONAL)

Preparations • Switch off the pump at the main power switch.


• Open the doors of the module.
• Use an optional solvent shutoff valve or lift up solvent filters in solvent reservoirs for avoiding
leakages.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 161
9 Maintenance and Repair
Exchange the Purge Valve Frit or the Purge Valve

1 Using a 1/4 inch wrench disconnect the pump outlet 2 Using the 14 mm wrench unscrew the purge valve and
capillary from the purge valve. Disconnect the waste remove it.
tube.

NOTE
Beware of leaking solvents due to hydrostatic pressure.

3 Remove the seal cap from the purge valve. 4 Using a pair of tweezers or a toothpick remove the frit.

Seal cap

Frit

5 Place a new frit into the purge valve with the orientation 6 Reinstall the seal cap including the gold seal.
of the frit as shown below (slit in frit points to the front).

Seal cap NOTE


Before reinstallation always check the gold seal in the seal
Frit cap. A deformed seal cap should be exchanged.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 162
9 Maintenance and Repair
Exchange the Purge Valve Frit or the Purge Valve

8 Reconnect the outlet capillary and the waste tubing.


C AU T I O N
Damage to the purge valve

 Do not lift the pump using the purge valve as a handle,


it might get leaky.

 Do not try to turn the purge valve into the correct


position when already fixed to the pump. The rubber
o-ring might break.

 Anticipate the correct position of the connections


before tightening the valve.
7 Insert the purge valve into the pump head and fix it with
a 14 mm wrench.

9 Close the doors.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 163
9 Maintenance and Repair
Replace the O-Ring on the Purge Valve

Replace the O-Ring on the Purge Valve

When If the original o-ring is damaged and needs to be replaced

Tools required p/n Description


8710-0510 Open-end wrench 1/4 — 5/16 inch
8710-1924 Open-end wrench 14 mm
Pair of tweezers
OR Toothpick

Parts required # p/n Description


1 5067-6595 1260 PV O-ring FKM 5/pack
1 01018-22707 PTFE frits (pack of 5) (OPTIONAL)
1 5067-4728 Seal cap (OPTIONAL)

Preparations • Switch off pump at the main power switch.


• Open the doors of the module.
• Use an optional solvent shutoff valve or lift up solvent filters in solvent reservoirs for avoiding
leakages.
• Remove the purge valve from the pump head.

1 Disassemble the purge valve. 2 Remove the old o-ring from the purge valve.
3 Clean the purge valve parts.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 164
9 Maintenance and Repair
Replace the O-Ring on the Purge Valve

4 Place the new o-ring on the Screw Purge Valve. 5 Place the screw with o-ring on the Purge Valve Body.

6 Place the mounting ring on the screw and push down 7 Push the screw up and guide the o-ring into the gap.
the o-ring.

8 Place a new frit into the purge valve with the orientation 9 Install the purge valve to the pump. Make sure not to
of the frit as shown below (slit in frit points to the front). turn the purge valve body when the screw is fixed to the
Reinstall the seal cap including the gold seal. pump. The o-ring will take damage.

NOTE
Before reinstallation always check the gold seal in the seal
cap. A deformed seal cap should be exchanged.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 165
9 Maintenance and Repair
Remove the Pump Head Assembly

Remove the Pump Head Assembly

When • Exchanging the seals


• Exchanging the pistons
• Exchanging seals of the seal wash function

Tools required p/n Description


8710-0510 Open-end wrench 1/4 — 5/16 inch
5023-2524 Hex Key Set + Driver

Preparations • Switch off the pump at the main power switch and unplug the power cable.
• Use an optional solvent shutoff valve or lift up solvent filters in solvent reservoirs for avoiding
leakages.
• Open the doors of the module.
• If an active inlet valve is installed, disconnect the active inlet valve cable.

Damage of the pump drive


C AU T I O N
Starting the pump when the pump head is removed may damage the pump
drive.
 Never start the pump when the pump head is removed.

1 Using a 1/4 inch wrench disconnect the valve capillary 2 Disconnect the solvent inlet tube from the inlet valve.
from the outlet valve.
NOTE
Beware of leaking solvents due to hydrostatic pressure.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 166
9 Maintenance and Repair
Remove the Pump Head Assembly

3 Using a 1/4 inch wrench disconnect the pump outlet 4 If applicable, remove tubings from the seal wash
capillary from the purge valve. Disconnect the waste support rings.
tube.

NOTE
Beware of leaking solvents due to hydrostatic pressure.

5 Remove the capillary at the bottom of the pump head. 6 Using a 4 mm hexagonal key, stepwise loosen the two
pump head screws.

7 Remove the pump head from the pump drive.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 167
9 Maintenance and Repair
Maintenance of a Pump Head Without Seal Wash Option

Maintenance of a Pump Head Without Seal


Wash Option

When In case of maintenance or pump head internal leaks.

Tools required Description


Wrench 1/4 inch
Hexagonal key, 4 mm

Parts required # p/n Description


1 01018-23702 Insert tool
1 5063-6589 Piston seal PTFE, carbon filled, black (pack of 2), default
OR 1 0905-1420 PE seals (pack of 2)
1 5063-6586 Sapphire piston, 100 µL
1 8660-0852 Abrasive mesh

Preparations • Switch off the pump at the main power switch.


• Open the doors of the module.
• Remove the Pump Head Assembly.

1 Place the pump head on a flat surface. Loosen the lock 2 While holding the lower half of the assembly (piston
screw (two turns). housing), carefully pull the pump housing away from the
piston housing.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 168
9 Maintenance and Repair
Maintenance of a Pump Head Without Seal Wash Option

3 Remove the support rings from the piston housing. 4 Lift the housing away from the pistons.

5 Check the piston surface and remove any deposits or 6 Using the steel side of the insert tool, carefully remove
layers: clean the piston surface with abrasive paper and the seal from the pump housing.
rinse with 2-propanol. Replace piston if scratched.

Piston Surface

7 Using the plastic side of the insert tool, insert new seals 8 Place the support rings on the piston housing. Note the
into the pump head. correct position of the pins.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 169
9 Maintenance and Repair
Maintenance of a Pump Head Without Seal Wash Option

9 Place the pump housing onto the piston housing. 10 Insert the pistons and carefully press them into the
seals.

11 Tighten the lock screw. 12 Install the pump head, see “Reinstall the Pump Head
Assembly” on page 176.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 170
9 Maintenance and Repair
Maintenance of a Pump Head with Seal Wash Option

Maintenance of a Pump Head with Seal Wash


Option

For bio-inert modules use bio-inert parts only!

When In case of maintenance or pump head internal leaks.

Tools required Description


Wrench 1/4 inch
Hexagonal key, 4 mm

Parts required # p/n Description


1 01018-23702 Insert tool
1 0905-1175 Wash seal (PTFE)
OR 1 0905-1718 Wash Seal PE
OR 1 0905-1731 Bio-Inert Wash Seal
1 5062-2484 Gasket, seal wash (pack of 6)
1 5063-6589 Piston seal PTFE, carbon filled, black (pack of 2), default
OR 1 G5611-21503 Piston Seal PTFE (Bio-inert)
1 0905-1420 PE seals (pack of 2)
1 5063-6586 Sapphire piston, 100 µL
OR 1 5067-4695 Sapphire piston (Bio-inert)
1 8660-0852 Abrasive mesh

Preparations • Switch off the pump at the main power switch.


• Open the doors of the module.
• Use an optional solvent shutoff valve or lift up solvent filters for avoiding leakages.
• Remove the Pump Head Assembly.
• Remove the wash solvent tubings from the support ring inlet and outlet.

G5654A only: Seal Holder (Bio-inert) (G5611-60016) consists of two separate


NOTE parts: seal holder and backup ring. Make sure to install the backup ring in the
correct way as described below in order to avoid damaging the piston seals.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 171
9 Maintenance and Repair
Maintenance of a Pump Head with Seal Wash Option

1 Place the pump head on a flat surface. Loosen the lock 2 While holding the lower half of the assembly (piston
screw (two turns). housing), carefully pull the pump housing away from the
piston housing.

3 Remove the seal holder and the seal wash support rings 4 Remove the seal holder from the support ring
from the piston housing. assembly.

5 Lift the housing away from the pistons. 6 Check the piston surface and remove any deposits or
layers: clean the piston surface with abrasive paper and
rinse with 2-propanol. Replace piston if scratched.

Piston Surface

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 172
9 Maintenance and Repair
Maintenance of a Pump Head with Seal Wash Option

7 Using the steel side of the insert tool, carefully remove 8 Using the plastic side of the insert tool, insert new seals
the seal from the pump housing. into the pump head.

9 Using the steel side of the insert tool, remove the seal 10 Using the plastic side of the insert tool, press the new
wash gasket and the wash seal from the support ring. wash seal (spring pointing upwards) into the recess of
the support ring.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 173
9 Maintenance and Repair
Maintenance of a Pump Head with Seal Wash Option

11 Place a seal wash gasket in the recess of the support


ring. Use a matching orientation of gasket and support C AU T I O N
ring. Put the seal holder on top of the gasket. G5654A only: Seal life time
Wrongly inserting the backup ring will strongly reduce the
life time of the seals and the pump head parts.

 Note the correct installation of the backup ring inside


the support ring.
12 G5654A only: If the backup seal has been removed from
the seal holder, insert it in the correct orientation. Please
ensure that the cut end faces the seal holder.

cut end

13 G5654A only: Place a seal wash gasket in the recess of 14 Place the support rings on the piston housing. Note the
the support ring. Use a matching orientation of gasket correct position of the pins on the support ring.
and support ring. Put the seal holder on top of the
gasket.

Backup seal
Seal holder

Seal wash gasket

Seal holder

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 174
9 Maintenance and Repair
Maintenance of a Pump Head with Seal Wash Option

15 Place the pump housing onto the piston housing. 16 Insert the pistons and carefully press them into the
seals.

17 Tighten the lock screw. 18 Install the pump head, see “Reinstall the Pump Head
Assembly” on page 176.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 175
9 Maintenance and Repair
Reinstall the Pump Head Assembly

Reinstall the Pump Head Assembly

When When reassembling the pump

Tools required p/n Description


5023-2524 Hex Key Set + Driver
5067-5688 Torque wrench 1 – 25 Nm with 14 mm wrench
G4220-20013 4 mm hex bit
G4220-20015 Adapter ¼ in square to hex

Parts required # p/n Description


1 79846-65501 Pump head grease

Damage of the pump drive


C AU T I O N
Starting the pump when the pump head is removed may damage the pump
drive.
 Never start the pump when the pump head is removed.

1 If needed, apply a small amount of grease on the back 2 Slide the pump head assembly onto the pump drive.
of the screws. Normally, the grease added during
manufacturing is sufficient for a long time.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 176
9 Maintenance and Repair
Reinstall the Pump Head Assembly

3 Insert the pump head screws and tighten them 4 Install the capillary at the bottom of the pump head.
stepwise. First tighten the left screw, then the right
screw, with a torque of approximately 5 Nm.

approx.
5 Nm

5 Reconnect the valve capillary to the outlet valve. 6 Reconnect the solvent inlet tube to the passive inlet
valve.

7 Reconnect the outlet capillary and the waste tubing. 8 If installed, reconnect the active inlet valve cable to its
connector.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 177
9 Maintenance and Repair
Reinstall the Pump Head Assembly

9 Close the doors. Next Steps:

10 If a standard seal has been installed, run the seal


wear-in procedure, which includes a replacement of the
purge valve frit (“Seal Wear-in Procedure” on page 179).
11 For the normal phase seal, the purge valve frit should be
replaced, see “Exchange the Purge Valve Frit or the
Purge Valve” on page 161.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 178
9 Maintenance and Repair
Seal Wear-in Procedure

Seal Wear-in Procedure

Parts required p/n Description


0100-1847 Adapter AIV to solvent inlet tubes
5022-2159 Restriction capillary

Before replacing your solvent by isopropanol or replacing isopropanol by your


NOTE solvent, consider solvent miscibility. For example, do not directly switch from
buffers to isopropanol and vice versa.

1 Place a bottle with 100 mL of isopropanol in the solvent cabinet and put a
tubing (including bottle head assembly) into the bottle.
2 If an active inlet valve is installed, screw the PEEK adapter 1/4-28 to 10-32
(0100-1847) to the AIV and connect the inlet tube from the bottle head directly
to it.
3 Connect the Restriction capillary (5022-2159) to the purge valve. Connect its
other end to a waste container.
4 Open the purge valve and purge the system for 5 min with isopropanol at a
flow rate of 2 mL/min.
5 Close the purge valve and set the flow to a rate adequate to achieve a
pressure of 350 bar. Pump 15 min at this pressure to wear in the seals. The
pressure can be monitored using your instrument control software or tool.
6 Turn OFF the pump, slowly open the purge valve to release the pressure from
the system, disconnect the restriction capillary and reinstall the bottle with the
solvent for your application.
7 Rinse your system with the solvent used for your next application.
8 Replace the purge valve frit, see “Exchange the Purge Valve Frit or the Purge
Valve” on page 161.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 179
9 Maintenance and Repair
Exchange the Multi-Channel Gradient Valve (MCGV)

Exchange the Multi-Channel Gradient Valve


(MCGV)

For bio-inert modules use bio-inert parts only!

Tools required p/n Description


8710-0899 Screwdriver Pozidrive Shaft

Parts required # p/n Description


1 G1311-67701 Multi channel gradient valve (MCGV)
OR 1 G5611-67701 Multi-Channel Gradient Valve (MCGV) (Bio-inert)

Preparations • Switch off the pump at the main power switch.


• Open the doors of the module.
• Use an optional solvent shutoff valve or lift up solvent filters in solvent reservoirs for avoiding
leakages.

The life time of the multi-channel gradient valve can be increased by regularly
NOTE flushing the valve, especially when using buffers. If using buffers, flush all
channels of the valve with water to prevent precipitation of the buffer, otherwise
salt crystals could drop into an unused channel and form plugs that may cause
leaks of that channel. Such leaks will interfere with the general performance of
the valve. When using buffers in combination with organic solvents, it is
recommended to connect the aqueous solutions/buffers to one of the bottom
ports (A and D) and the organic solvent to one of the upper gradient valve ports.
It is best to have the organic channel directly above the buffer channel (e.g., A -
buffer, B - organic solvent).

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 180
9 Maintenance and Repair
Exchange the Multi-Channel Gradient Valve (MCGV)

1 Disconnect the solvent tubes from the MCGV. 2 Disconnect the connecting tube from the MCGV.

3 Press the lower sides of the cover to unclip it. Remove 4 Disconnect the MCGV cable from its connector.
the cover.

5 Unscrew the two screws and remove them. 6 Remove the MCGV.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 181
9 Maintenance and Repair
Exchange the Multi-Channel Gradient Valve (MCGV)

7 Place the new MCGV into position. Make sure that 8 Tighten the two screws.
channel A of the MCGV is put at the bottom-right
position.

9 Connect the cable to its connector. 10 Install the MCGV cover.

11 Reconnect the tube from the inlet valve to the middle 12 Connect solvent tubes for channels A-D from the MCGV
position of the MCGV. to the degasser outlets.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 182
9 Maintenance and Repair
Replace the Solvent Selection Valve (Isocratic Pump)

Replace the Solvent Selection Valve (Isocratic


Pump)

When If leaking internally (crossflow between the ports), or if one of the channels is blocked

Tools required p/n Description


8710-0899 Screwdriver Pozidrive Shaft

Parts required p/n Description


5067-5895 Solvent selection valve

Preparations Switch off the pump at the main power switch.

1 Lift solvent bottles out of the solvent cabinet and place 2 Disconnect all tubings from the solvent selection valve.
them on the table. Disconnect the solvent tubes from
the solvent selection valve and empty the tubes into the
bottles. Place the bottles back into the solvent cabinet.

3 Using a screwdriver, loosen the holding screws of the 4 Carefully disconnect the valve cable at the SSV
valve holder. connector on the Z-Panel.
5 Using a screwdriver, loosen the holding screws of the
degasser front cover.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 183
9 Maintenance and Repair
Replace the Solvent Selection Valve (Isocratic Pump)

6 Carefully pull the valve with a cable out of the pump 7 Exchange the defective solvent selection valve.
mainframe.

8 Tighten the screws of the degasser front cover and the 9 Reconnect all tubings to the solvent selection valve.
valve holder.

NOTE
After an exchange of the valve it may be required to pump
several mL of solvent before the flow stabilizes at a
pressure ripple as low as it used to be when the system was
still working properly.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 184
9 Maintenance and Repair
Exchange the Active Inlet Valve (AIV) or its Cartridge

Exchange the Active Inlet Valve (AIV) or its


Cartridge

For bio-inert modules use bio-inert parts only!

When If internally leaking (backflow)

Tools required Description


Wrench, 14 mm
Pair of tweezers

Parts required p/n Description


G5699A Active Inlet Valve Upgrade Kit
G1312-60025 Active inlet valve body, without cartridge
G1312-60020 Cartridge for active inlet valve 600 bar
G1311-67304 Connecting tube
G1311-61601 Active Inlet Valve Cable Assembly
0100-2298 Adapter, PEEK int. 1/4-28 to ext. 10-32
G5611-60025 Active Inlet Valve (AIV, Bio-inert), without cartridge
G5611-60020 Cartridge for AIV (Bio-inert)

Preparations • Switch off pump at the main power switch and unplug the power cable.
• Use an optional solvent shutoff valve or lift up solvent filters in solvent reservoirs for avoiding
leakages.

1 Unplug the active inlet valve cable from the connector.


2 Disconnect the solvent inlet tube from the inlet valve (be aware that solvent
may leak out of the tube due to hydrostatic flow).
3 Unscrew the adapter from the active inlet valve.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 185
9 Maintenance and Repair
Exchange the Active Inlet Valve (AIV) or its Cartridge

4 Using a 14 mm wrench loosen the active inlet valve and remove the valve
from the pump head.

Active inlet valve cartridge

Active inlet valve

Figure 21 Active Inlet Valve Assembly

5 Using a pair of tweezers remove the valve cartridge from the actuator
assembly.
6 Before inserting the new valve cartridge clean the area in the actuator
assembly. Flush the cartridge area thoroughly with alcohol.
7 Insert a new cartridge into the actuator assembly (make sure the valve
cartridge is completely inserted into the actuator assembly).
8 Insert the new valve into the pump head. Using the 14 mm wrench turn the
nut until it is hand tight.
9 Position the valve such that the solvent inlet tube connection points towards
the front.
10 Using the 14 mm wrench tighten the nut by turning the valve into its final
position (not more than a quarter turn).
11 Reconnect the adapter at the active inlet valve.
12 Reconnect the solvent inlet tube to the adapter. Reconnect the active inlet
valve cable to the connector in the Z-panel.
13 Purge the system with 30 mL of solvent in order to achieve a low pressure
ripple.
See “Regular Priming” on page 66

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 186
9 Maintenance and Repair
Exchange the Seal Wash Cartridge

Exchange the Seal Wash Cartridge

Parts required p/n Description


5065-4445 Peristaltic pump with PharMed tubing

Preparations • Switch off pump at the main power switch.


• Open the doors.

1 Remove the wash solvent tubings from the support ring 2 Unclip the peristaltic pump cartridge from the module
outlet and from the adapter leading to the waste bottle. housing and remove it.

3 Put the new peristaltic pump cartridge onto the rod of 4 Connect the peristaltic pump tubes to the support rings
the pump motor and push the plastic clips into the outlet and to the adapter leading to the waste bottle.
module housing.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 187
9 Maintenance and Repair
Replace Leak Handling System Parts

Replace Leak Handling System Parts

Parts required p/n Description


5043-0856 Leak Adapter
5063-6527 Tubing, Silicon Rubber, 1.2 m, ID/OD 6/9 mm
approximately 85 mm required

1 Open the doors.


2 Press the Leak Adapter down and remove it together with the tubing.
3 Install the Leak Adapter by pressing it into the Main Cover.
4 Insert the Tubing (approximately 85 mm required for replacement) between
Leak Adapter outlet and Leak Panel.
5 Close the doors.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 188
9 Maintenance and Repair
Replacing the Module Firmware

Replacing the Module Firmware

When The installation of newer firmware might be necessary


• if a newer version solves problems of older versions or
• to keep all systems on the same (validated) revision.
The installation of older firmware might be necessary
• to keep all systems on the same (validated) revision or
• if a new module with newer firmware is added to a system or
• if third party control software requires a special version.

Tools required Description


Agilent Lab Advisor software

Parts required # Description


1 Firmware, tools and documentation from Agilent web site

Preparations Read update documentation provided with the Firmware Update Tool.

To upgrade/downgrade the module’s firmware carry out the following steps:


1 Download the required module firmware, the latest FW Update Tool and the
documentation from the Agilent web.
https://www.agilent.com/en-us/firmwareDownload?whid=69761
2 For loading the firmware into the module follow the instructions in the
documentation.
Module Specific Information
There is no specific information for this module.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 189
9 Maintenance and Repair
Introduction to Repair

Introduction to Repair

Repairs may require the exchange of defective internal parts. Exchange of these
parts requires removing the module from the stack, removing the covers, and
disassembling the module.

Repair and service activities should only be performed by trained and authorized
NOTE service personnel.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 190
9 Maintenance and Repair
Overview of Repairing the Pump

Overview of Repairing the Pump

Figure 22 on page 191 shows the main repair assemblies of the isocratic pump,
Figure 23 on page 192 shows the main repair assemblies of the quaternary
pump. The pump head and its parts require normal maintenance (for example,
seal exchange) and can be accessed from the front (simple repairs). Replacing
internal parts will require to remove the pump from its stack and to open the top
cover.

11

10

9 12

13
8
14
7

6
15
5
4
3
2

1
Figure 22 Overview of Main Assemblies (Isocratic Pump)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 191
9 Maintenance and Repair
Overview of Repairing the Pump

6 9, 12

10

5
11
4
13
3

Figure 23 Overview of Main Assemblies (Quaternary Pump)

1 Power switch, “Replace the Power Switch” on page 219

2 Leak panel, “Replace the Leak Panel” on page 236

3 Leak sensor, “Replace the Leak Sensor” on page 238

4 Degasser, “Replace the Degasser” on page 221

5 Power supply, “Replace Power Supply and Cables” on page 216


(not shown)

6 Fan, “Exchange the Fan” on page 214


(not shown)

7 Pump mainboard, “Exchange the Pump Mainboard” on page 196

8 FUSION board, “Replace the Fusion Board” on page 202

9, 12 Damper and Damper Leak Panel, “Exchange the Damper” on page 208

10 Status LED, “Replace the Status LED” on page 239

11 Pump drive, “Exchange the Pump Drive” on page 212

13 Active Seal Wash, “Install an Active Seal Wash Upgrade Kit (Quaternary Pump)” on
page 241

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 192
9 Maintenance and Repair
Exchanging Internal Parts

Exchanging Internal Parts

The procedures in this section describe how to exchange internal parts. You
must remove the pump from the stack in order to open the main cover.
Table 22 Exchanging internal parts

Procedure Typical Frequency Notes

“Exchange the Pump Mainboard” on page 196 If defective Certain error messages are
generated

“Install an Active Seal Wash Upgrade Kit (Quaternary Pump)” on If seals show indication of Leaks at lower pump head
page 241 leaks side, loss of wash solvent

“Exchange the Damper” on page 208 If leaking or defective No pressure output

“Exchange the Fan” on page 214 If defective Fan not running; error
message “Fan Failed” is
generated

“Exchange the Pump Drive” on page 212 If mechanically blocked or Certain error messages are
defective generated

“Replace Power Supply and Cables” on page 216 If defective

“Replace the Leak Sensor” on page 238 If defective Leak messages without leak
in the leak pan

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 193
9 Maintenance and Repair
Remove the Top Cover and Foam

Remove the Top Cover and Foam

Tools required p/n Description


8710-0899 Screwdriver Pozidrive Shaft

Preparations • Switch off pump at the main power switch and unplug the power cable.
• Disconnect the solvent inlet and outlet tubes from the pump. Beware of leaking solvents due to
hydrostatic flow.
• Remove the pump from the stack.
• Remove the doors of the module.

Module is partially energized when switched off, as long as the power cord
WAR N IN G
is plugged in.
Repair work at the module can lead to personal injuries, e.g. shock hazard,
when the cover is opened and the module is connected to power.
 Make sure that it is always possible to access the power plug.
 Remove the power cable from the instrument before opening the cover.
 Do not connect the power cable to the Instrument while the covers are
removed.

1 Release the clips of the cover at the rear of the module. 2 Lift the cover up and slide it towards the rear.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 194
9 Maintenance and Repair
Remove the Top Cover and Foam

3 Unscrew the screws at the rear end of the top plate. 4 Slide the plate towards the front and remove it.

5 Remove the top foam.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 195
9 Maintenance and Repair
Exchange the Pump Mainboard

Exchange the Pump Mainboard

When Board defective

Tools required Description


Wrench, 5 mm
Screwdriver, Torx T20

Parts required p/n Description


G7111-65810 Mainboard 1260 Infinity II Quaternary Pump VL 400 bar
G7111-65800 Mainboard 1260 Iso, Quat & Bio Pump 600 bar

Preparations • Turn off the pump, disconnect all cables and remove the pump from the stack.
• Remove the top cover and foam (see “Remove the Top Cover and Foam” on page 194).

Wrong connection on mainboard


C AU T I O N
Interchanging the connectors for seal wash pump and pressure sensor will
destroy the power supply (more likely) or the mainboard.
 Connect the seal wash pump cable and pressure sensor cable to the correct
connectors on the mainboard.
 In case of damage, check the -15 V output of the power supply connected to
the mainboard to find out which part has got damaged.

Damage to the Fusion Board


C AU T I O N
 Always remove the fusion and mainboard together.
 Only separate the fusion board from the mainboard outside of the module.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 196
9 Maintenance and Repair
Exchange the Pump Mainboard

1 Disconnect all connectors from the mainboard.

When removing connectors, counterhold on the board on a near connector with


NOTE one finger.

The pump mainboard carries the Fusion board. Both boards can be ordered and
NOTE replaced separately.

2 Remove the connector screws and the grounding screw, which fixes the
board to the module housing.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 197
9 Maintenance and Repair
Exchange the Pump Mainboard

3 Remove the boards. Place the boards on an ESD kit.

If you want to exchange other parts as well, continue with that work first.
NOTE

4 Lift the FUSION Board carefully from the mainboard. Depending on the
module there might be two white plastic spacers that fix the board in position.

5 Insert the (new) FUSION board on the two connectors.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 198
9 Maintenance and Repair
Exchange the Pump Mainboard

6 On the new fusion board check the configuration switch S1 setting, see
“Setting the 6-bit Configuration Switch” on page 317.

An incorrect switch setting (e.g., TEST/BOOT) may cause the module to turn into
NOTE a non-operational mode (yellow or red flashing status light). In such a case turn
off the module, re-set the address switches, and turn on the module again.

7 Install the new board. Make sure that the board is fitted correctly in the board
recess holes at the rear panel.

8 Reconnect the connectors to the mainboard.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 199
9 Maintenance and Repair
Exchange the Pump Mainboard

11 12 13 14
10

9
8
7

4 15

3 16

2
17
1

Figure 24 Mainboard connectors

Table 23 Identifying connectors on the pump mainboard


Connector Description Connector Description

1 Pump drive motor 10 Degasser pressure sensor

2 Pump drive encoder 11 CAN connector

3 AIV (optional) 12 6-bit configuration switch

4 Fan 13 LAN

5 MCGV (quaternary pump 14 USB


only)

6 Seal wash pump 15 Power control

7 Degasser pump motor 16 Power supply

8 Power switch board 17 Status LED

9 Pump pressure sensor ERI (not shown; located on


mainboard below the fusion board)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 200
9 Maintenance and Repair
Exchange the Pump Mainboard

9 Fix the connector screws and the grounding screw.

10 Replace foam and top covers, see “Install the Top Cover and Foam” on
page 245.
11 Reinstall the pump in the stack. Reconnect all cables. Turn on the pump. If
pump status indicator turns red continue with the section “Serial Number and
Type Change” on page 203 and “Setting the 6-bit Configuration Switch” on
page 317.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 201
9 Maintenance and Repair
Replace the Fusion Board

Replace the Fusion Board

The FUSION Board sits on two connectors on the mainboard.

Parts required p/n Description


G4200-65880 Fusion Board

1 Lift the FUSION Board carefully from the mainboard. 2 Insert the (new) FUSION board on the two connectors.
Depending on the module there might be two white
plastic spacers that fix the board in position.

Next Steps:

3 On the new fusion board check the configuration switch S1 setting, see “Setting the 6-bit Configuration Switch” on
page 317.

NOTE
An incorrect switch setting (e.g., TEST/BOOT) may cause the module to turn into a non-operational mode (yellow or red
flashing status light). In such a case turn off the module, re-set the address switches, and turn on the module again.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 202
9 Maintenance and Repair
Changing the Type and Serial Number

Changing the Type and Serial Number

Serial Number and Type Change


When the mainboard or the fusion board has to be replaced, the new board does
not have a serial number. For modules with multiple usage boards, e.g. for some
autosamplers and pumps, the module type has to be set as well. Use the
information from the serial number plate of your module.
Keep in mind that
• the changes become active after a power cycle of the module.
• the information from the serial number plate of the module is used.
• the exact type (product number) is used.

Unusable module or mainboard


C AU T I O N
If the type (product number) and/or serial number are entered incorrectly, the
module may become unusable or is not recognized in a system.
The mainboard may become irreversibly unusable.
 Enter the correct type (product number) and/or serial number.
 All letters must be CAPITAL.

For further information proceed to “Overview” on page 205.

With firmware A.06.02/B.01.02 and above, the entry is checked against the board
NOTE revision, which inhibits setting wrong types for a board. The caution above for
multiple usage boards still applies.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 203
9 Maintenance and Repair
Changing the Type and Serial Number

Using the Agilent Lab Advisor


Prerequisites The Agilent Lab Advisor Software must be configured in FSE Mode to have
access to the function Board Check and Change.
Close other user interfaces.

1 Start the Lab Advisor Software.


2 Select Service & Diagnostics on the left navigation panel.
3 Select Board Check and Change in the list on the right panel and click Run in
the right bottom corner.
4 Change the field Type and/or Serial as required (use capital letters only).

Some Agilent modules require the correct mainboard version to match the type.
NOTE

5 Press the button Apply to complete the action.


A message opens: "The module type was changed to XXXXXX. Perform the
following steps now... Close this application. Switch off and on the changed
LC module. Update firmware if required."
6 Close the Lab Advisor Software.
7 Power cycle the module.
8 Restart the User Interface.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 204
9 Maintenance and Repair
Changing the Type and Serial Number

Recover Instructions
Overview
The following situations may come up where the instrument is no longer usable
due to
• an incorrect type (product number) entry after the replacement of a
mainboard of the module.
• load of wrong firmware based on the wrong type.

With firmware A.06.02/B.01.02 and above, the entry is checked against the board
NOTE revision.

The wrong type (product number) could be


• incorrect, but a valid module number
• incorrect and invalid module number (any name)
Based on above, the User Interfaces respond differently.
Table 24 Recover From Wrong Type

User Interface incorrect but valid type incorrect but valid type incorrect and invalid type

Example Conditions correct type = G1315B correct type = G1315B correct type = G1315B
entered type = G1314B entered type = G1314B entered type = G1319B
plus wrong firmware from
G1314B

Agilent Diagnostic shows the incorrect product shows the incorrect product shows the incorrect product
Software (preferred number number number
tool) access to the module is access to the module is access to the module is
possible as described in possible as described in possible as described in
“Recover with Agilent Lab “Recover with Agilent Lab “Recover with Agilent Lab
Advisor (Type Only)” on Advisor (Type Only)” on Advisor (Type Only)” on
page 206 page 206 page 206
If wrong firmware has been
loaded in addition, only the
Agilent Lab Advisor Software
can revert to correct product
number as described in
“Recover with Agilent Lab
Advisor (Type & Firmware)” on
page 207

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 205
9 Maintenance and Repair
Changing the Type and Serial Number

Recover with Agilent Lab Advisor (Type Only)

Prerequisites The Lab Advisor Software is used in FSE mode.


If no LAN connection is possible use RS-232.
The example uses G1315B as correct type.

1 Open a connection to the module (or system).


The module will be listed with the wrong product number (type).
2 Select Service & Diagnostics on the left navigation panel.
3 Select the module of interest in the middle panel.
4 Select Board Check and Change in the right window and press Run in the right
bottom corner.
5 In the type field enter G1315B and press Apply.
6 Close the Lab Advisor software.
7 Power cycle the module, restart Lab Advisor and reconnect to the LC stack.
Verify that type and serial number have been updated correctly.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 206
9 Maintenance and Repair
Changing the Type and Serial Number

Recover with Agilent Lab Advisor (Type & Firmware)

Prerequisites Agilent Lab Advisor Software in FSE mode.


If no LAN connection is possible use RS-232 or USB.
The example uses G1315B as correct type.
The module must be configured to "Stay in Resident Mode" (module boots in
resident mode - flashing status LED).

1 To set up the module in Lab Advisor, connect the module (or the system) in
Lab Advisor.
2 Select Instrument Control on the left panel.
3 Open the Controls menu of the respective module.
4 Expand the Conversions sections of the menu.
5 Click Transfer to Resident FW.
6 Open a connection to the module (or via system).
7 Select Service & Diagnostics on the left navigation panel.
8 Select the module of interest in the middle panel.
9 Select Board Check and Change and press Run in the right bottom corner.
10 In the type field enter G1315B and press Apply.
11 Close the Lab Advisor Software.
12 Switch off and on the changed LC module.
13 Open Lab Advisor and connect to the module (or the system)
14 Set module's configuration back to Main Mode using Lab Advisor (Instrument
Control > Controls > Conversions).
15 Update firmware if required.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 207
9 Maintenance and Repair
Exchange the Damper

Exchange the Damper

For bio-inert modules use bio-inert parts only!

When No pressure output or when leaking

Tools required p/n Description


8710-0899 Screwdriver Pozidrive Shaft
Wrench, 1/4 inch

Parts required # p/n Description


1 G5611-60065 Damper (Bio-inert)
1 G1311-60065 Damper Isocratic/Quaternary Pump 600 bar
1 79835-60005 Damper Quaternary Pump 400 bar
1 G1311-44162 Damper Leak Panel
1 G7111-81600 Capillary
damper to inlet pump chamber
1 G5654-81600 Capillary, Ti, damper to inlet valve
1 G7111-81601 Capillary
outlet valve to damper
1 G5654-81601 Capillary, Ti, outlet valve to damper

Preparations • Switch off the pump at the main power switch, remove all cables and tubings, remove the pump
from the stack.
• Remove the MCGV Cover (see “Exchange the Multi-Channel Gradient Valve (MCGV)” on
page 180)
• Remove the top cover and top foam section (see “Remove the Top Cover and Foam” on
page 194).

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 208
9 Maintenance and Repair
Exchange the Damper

Wrong damper
C AU T I O N
Using a wrong damper will cause a wrong regulation of the pump and wrong
measurement results.
 Make sure to use the correct damper type. The 600 bar damper for the
Isocratic and Quaternary Pump can be identified by two rings and a yellow
mark. The 400 bar damper for the Quaternary Pump VL does not have rings
and the mark. The 600 bar damper of the G1312B or G7112B Binary Pump
does not have a pressure sensor.

Wrong connection on mainboard


C AU T I O N
Interchanging the connectors for seal wash pump and pressure sensor will
destroy the power supply (more likely) or the mainboard.
 Connect the seal wash pump cable and pressure sensor cable to the correct
connectors on the mainboard.
 In case of damage, check the -15 V output of the power supply connected to
the mainboard to find out which part has got damaged.

1 Disconnect the two damper capillaries. 2 Disconnect the damper connector at the mainboard.
Lift the damper out of its position.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 209
9 Maintenance and Repair
Exchange the Damper

3 Loosen the screw of the degasser cover and the 4 4 Remove the damper leak panel.
screws of the Z-panel and take it out of the instrument.

5 Make sure that the pressure sensor cable of the damper 6 Reinstall the leak panel.
is wound around the pressure sensor as indicated (2
windings).

7 Install the Z-panel and fix it with the 4 screws. Insert the 8 Place the new damper into the recess and connect it to
lower end of the degasser cover to the module housing. the mainboard. The damper needs to be aligned such
Then fix it with a screw to the Z-panel. that both capillary connections are on a vertical line.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 210
9 Maintenance and Repair
Exchange the Damper

9 Do not confuse the connectors for the seal wash pump 10 Reconnect the two damper capillaries.
and the pressure sensor, as this will destroy the
mainboard!

PS

SW

Next Steps:

11 Replace the MCGV Cover (see “Exchange the Multi-Channel Gradient Valve (MCGV)” on page 180).
12 Replace the top foam section, metal cover and top cover (see “Install the Top Cover and Foam” on page 245). Replace
the pump on stack, reconnect all tubings and cables.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 211
9 Maintenance and Repair
Exchange the Pump Drive

Exchange the Pump Drive

When Error message: Motor Drive Power

Tools required p/n Description


8710-0899 Screwdriver Pozidrive Shaft
Wrench, 1/4 inch
Hexagonal key, 4 mm

Parts required # p/n Description


1 G1311-60001 Pump drive

Preparations • Switch off the pump at the main power switch.


• Remove all cables and tubings.
• Remove the pump from the stack.
• Remove the doors.
• Remove the MCGV cover (see “Exchange the Multi-Channel Gradient Valve (MCGV)” on
page 180).
• Remove the pump head (see “Remove the Pump Head Assembly” on page 166).
• Remove the top cover and top foam section (see “Remove the Top Cover and Foam” on
page 194).

Damage of the pump drive


C AU T I O N
Starting the pump when the pump head is removed may damage the pump
drive.
 Never start the pump when the pump head is removed.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 212
9 Maintenance and Repair
Exchange the Pump Drive

1 Loosen the screw of the degasser cover and the 4 2 Disconnect the pump drive cables from the mainboard
screws of the Z-panel and take it out of the instrument. and lift the drive out of the foam.

3 Place the new pump drive into the recess in the foam 4 Install the Z-panel and fix it with the 4 screws. Insert the
and connect the cables to the connectors on the lower end of the degasser cover to the module housing.
mainboard. Then fix it with a screw to the Z-panel.

Next Steps:

5 Install the top cover and top foam section (see “Install the Top Cover and Foam” on page 245).
6 Install the MCGV cover (see “Exchange the Multi-Channel Gradient Valve (MCGV)” on page 180).
7 Reinstall the pump head (see “Reinstall the Pump Head Assembly” on page 176).
8 Install the doors, replace the pump on stack, reconnect all tubings and cables.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 213
9 Maintenance and Repair
Exchange the Fan

Exchange the Fan

When Fan not running

Tools required p/n Description


8710-0899 Screwdriver Pozidrive Shaft

Parts required # p/n Description


1 5067-6199 Fan 92 mm x 92 mm

Preparations • Switch off the pump at the main power switch, remove all cables and tubings, remove the pump
from the stack.
• Remove the doors.
• Remove the top cover and top foam section (see “Remove the Top Cover and Foam” on
page 194).

Wrong fan orientation


C AU T I O N
Installing the fan in the wrong orientation will not cool the instrument
efficiently. As a consequence, electronic components may be damaged.
 Make sure that the air flow is directed as indicated by installing the fan
such, that the arrow on the fan points to the same direction as the arrrow on
the foam.

1 Disconnect the fan connector from the mainboard. 2 Remove the mainboard, see “Exchange the Pump
Mainboard” on page 196.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 214
9 Maintenance and Repair
Exchange the Fan

3 Push the fan to the left and lift it out of the foam part. 4 Slide the new fan into its position.

5 Connect the fan on the mainboard. Next Steps:

6 Reinstall the mainboard, see “Exchange the Pump


Mainboard” on page 196.
7 Reinstall top foam and top covers, see “Install the Top
Cover and Foam” on page 245.
8 Install the doors.
9 Replace the pump on stack, reconnect all tubings and
cables.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 215
9 Maintenance and Repair
Replace Power Supply and Cables

Replace Power Supply and Cables

Tools required Description


Screwdriver, Torx T20

Parts required p/n Description


0950-5274 Power supply
5067-5909 PWS Control Cable
5067-5974 Power Switch Cable

Preparations • Remove the doors, see “Remove and Install Doors” on page 154.
• Remove the MCGV, see “Exchange the Multi-Channel Gradient Valve (MCGV)” on page 180.
• Remove the leak interface, see “Replace Leak Handling System Parts” on page 188.
• Remove the top covers and foams, see “Remove the Top Cover and Foam” on page 194.
• Remove the mainboard, see “Exchange the Pump Mainboard” on page 196.
• Remove the degasser, see “Replace the Degasser” on page 221.
• Remove the pump drive, see “Exchange the Pump Drive” on page 212.
• Remove the fan, see “Exchange the Fan” on page 214.

1 Remove the bottom foam. 2 Unplug the power switch cable from the power switch
connector (1), and remove the adhesive stripes (2).

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 216
9 Maintenance and Repair
Replace Power Supply and Cables

3 Open grounding screws at the rear of the instrument, 4 Remove the power supply.
which fix the power supply to the instrument housing.

5 Check that the Power Switch is in the inner (pressed) 6 Ensure that the PWS-CONTROL Cable is connected to
position. If not, use a screw driver and press on it. the Power Supply.

NOTE
If the switch is in the outer (released) position the Power
Supply is in the OFF-mode and the Power Switch in the front
will not work!
7 Install the Power Supply into the instrument housing. 8 Fix the power supply to the instrument housing with the
grounding screws.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 217
9 Maintenance and Repair
Replace Power Supply and Cables

9 Plug the new power switch cable into the power switch 10 Install the bottom foam.
connector (1) (contacts upward), and fix the cable with
the adhesive stripes (2) on the bottom.

2
1

Installation Activities

1 Install the Fan, see “Exchange the Fan” on page 214.


2 Install the Pump Drive, see “Exchange the Pump Drive” on page 212.
3 Install the Degasser, see “Replace the Degasser” on page 221.
4 Install the mainboard, see “Exchange the Pump Mainboard” on page 196.
5 Install the foam and top cover, see “Install the Top Cover and Foam” on page 245.
6 Install the Leak Interface, see “Replace Leak Handling System Parts” on page 188.
7 Install the MCGV, see “Exchange the Multi-Channel Gradient Valve (MCGV)” on page 180.
8 Install the Doors, see “Remove and Install Doors” on page 154.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 218
9 Maintenance and Repair
Replace the Power Switch

Replace the Power Switch

Parts required p/n Description


5067-5712 Power Switch Assembly Infinity II

Preparations • Remove the module from the stack.


• Remove the doors, see “Remove and Install Doors” on page 154.
• Remove the leak interface, see “Replace Leak Handling System Parts” on page 188.

1 Place the module on its left side and locate the Power 2 While pressing with a flat screw driver into the slit on
Switch. the bottom [1] push the levers [2] into the panel.

3 Place the module in its normal position on the bench. 5 Slide out the Power Switch completely.
4 Remove the Leak Panel, see “Replace the Leak
Panel” on page 236.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 219
9 Maintenance and Repair
Replace the Power Switch

6 Disconnect the Leak Sensor Cable [1] and the Power 7 Reconnect the Power Switch cable [1] and the Leak
Switch Cable [2] (just pull the flat cable out of the Sensor cable [2] to the new Power Switch.
connector).

2 1 1 2

8 Slide the power switch into its final position (until it Next Steps:
clicks).

9 Install the Leak Panel, see “Replace the Leak Panel” on


page 236.
10 Install the Leak Interface, see “Replace Leak Handling
System Parts” on page 188.
11 Install the Doors, see “Remove and Install Doors” on
page 154.
12 Install the module in the stack.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 220
9 Maintenance and Repair
Replace the Degasser

Replace the Degasser

When In case of a degasser failure

Tools required p/n Description


8710-0899 Screwdriver Pozidrive Shaft

Parts required # p/n Description


1 G1311-60070 Degasser 4 Channels for Quaternary Pump
OR 1 G4280-60070 Degassing unit
2 channels, for Isocratic Pump

This procedure exemplarily shows the replacement of a four-channel degasser.


NOTE
In principle, replacing a two-channel degasser is done in the same way. The
pump inlet tubing is disconnected from the solvent selection valve and the
degasser front cover is removed together with the solvent selection valve.

1 Remove solvent tubings from the degasser inlets and 2 Open the screw of the degasser cover and remove the
outlets. cover.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 221
9 Maintenance and Repair
Replace the Degasser

3 Pull out the degasser for some centimeters and remove 4 Reconnect the cable.
the cable.

5 Slide in the degasser. 6 Install the degasser cover, tighten the screw of the
degasser cover.

7 Reconnect the solvent tubings to the degasser inlets


and outlets.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 222
9 Maintenance and Repair
Replace Parts Inside the 4-Channel-Degasser

Replace Parts Inside the 4-Channel-Degasser

Tools required p/n Description


8710-0899 Screwdriver Pozidrive Shaft

Parts required p/n Description


5067-4734 Vacuum Chamber
G1311-60007 Vacuum Pump for Quaternary Pump Degasser
G1311-67302 Tubing Kit for Quaternary Pump Degasser
G1311-67007 Housing Assembly for Quaternary Pump Degasser
G1311-67600 Sensor Board Assembly for Quaternary Pump Degasser
G1311-61605 Degasser Cable Assembly

Preparations Remove the degasser from the pump, see “Replace the Degasser” on page 221.

1 Open the degasser cover by lifting it up from the clips.

Clip

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 223
9 Maintenance and Repair
Replace Parts Inside the 4-Channel-Degasser

2 Replace a degasser chamber (see “Replace a Degasser Chamber” on


page 225)
OR
replace the degasser pump (see “Replace the Degasser Pump” on page 226)
OR
replace the sensor board assembly (see “Replace the Sensor Board
Assembly” on page 227)
OR
replace the degasser tubing kit (see “Replace the Degasser Tubing Kit” on
page 229).
3 Close the degasser cover. Note the correct orientation of the cover with
respect to the connector.

Connector

4 Reinstall the degasser into the pump and reconnect the degasser connector,
see “Replace the Degasser” on page 221.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 224
9 Maintenance and Repair
Replace Parts Inside the 4-Channel-Degasser

Replace a Degasser Chamber

Parts required p/n Description


5067-4734 Vacuum Chamber

Preparations • Remove the degasser, see “Replace the Degasser” on page 221.
• Open the degasser cover, see “Replace Parts Inside the 4-Channel-Degasser” on page 223.

1 Disconnect the tubing from the chamber. 2 Remove the screw at the front of the degasser housing
and remove the degasser chamber.

Degasser chamber

Tubing
Screws

3 Install the new degasser chamber, tighten the screw at the front of the degasser and reconnect the tubing to the new
chamber.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 225
9 Maintenance and Repair
Replace Parts Inside the 4-Channel-Degasser

Replace the Degasser Pump

Parts required p/n Description


G1311-60007 Vacuum Pump for Quaternary Pump Degasser

Preparations • Remove the degasser, see “Replace the Degasser” on page 221.
• Open the degasser cover, see “Replace Parts Inside the 4-Channel-Degasser” on page 223.

1 For replacing the degasser pump, disconnect both 2 Remove the three screws at the bottom of the degasser
tubings and the cable to the pressure sensor. and remove the pump.

Degasser pump

3 For installing the new pump, fix the screws at the 4 Connect the cable to the pressure sensor board.
bottom of the degasser. Connect the tubing from the
chambers to the inlet of the pump and the drain tubing
to the holder at the bottom of the degasser.

Drain outlet

Inlet

Tubing holder

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 226
9 Maintenance and Repair
Replace Parts Inside the 4-Channel-Degasser

Replace the Sensor Board Assembly

Parts required p/n Description


G1311-67600 Sensor Board Assembly for Quaternary Pump Degasser

Preparations • Remove the degasser, see “Replace the Degasser” on page 221.
• Open the degasser cover, see “Replace Parts Inside the 4-Channel-Degasser” on page 223.

1 For replacing the sensor board assembly, remove the 2 Then open both screws at the front of the degasser.
cable to the vacuum pump. Disconnect the tubing from
the pressure sensor.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 227
9 Maintenance and Repair
Replace Parts Inside the 4-Channel-Degasser

3 Open both screws next to the pressure sensor, lift up 4 Tighten the two screws next to the pressure sensor and
the board and replace it by the new one. the two screws at the front of the degasser.

Connector

Please note the correct orientation of the cover with


respect to the connector.
5 Connect the cable to the vacuum pump.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 228
9 Maintenance and Repair
Replace Parts Inside the 4-Channel-Degasser

Replace the Degasser Tubing Kit

Parts required p/n Description


G1311-67302 Tubing Kit for Quaternary Pump Degasser

Preparations • Remove the degasser, see “Replace the Degasser” on page 221.
• Open the degasser cover, see “Replace Parts Inside the 4-Channel-Degasser” on page 223.

1 For replacing the degasser tubing kit, remove tubings 2 Connect the new tubing from the chambers to the inlet
from the vacuum pump, degasser chambers and the of the pump and the new drain tubing to the holder at
pressure sensor. the bottom of the degasser.

Drain outlet

Inlet

Tubing holder

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 229
9 Maintenance and Repair
Replace Parts Inside the 2-Channel-Degasser

Replace Parts Inside the 2-Channel-Degasser

Tools required p/n Description


8710-0899 Screwdriver Pozidrive Shaft

Parts required p/n Description


G4280-60009 Pressure sensor assembly
G4280-60006 Micro Vacuum Pump
5067-4734 Vacuum Chamber
G4280-67300 Vacuum Tubing Kit
G4280-60007 Degasser Housing Assembly

Preparations Remove the degasser from the pump, see “Replace the Degasser” on page 221.

1 Open the degasser cover by lifting it up from the clips. 2 Replace a degasser chamber (see “Replace a Degasser
Chamber” on page 231)
OR
replace the degasser pump (see “Replace the Degasser
Pump” on page 232)
OR
replace the sensor board assembly (see “Replace the
Sensor Board Assembly” on page 234)
OR
replace the degasser tubing kit (see “Replace the Degasser
Clip Tubing Kit” on page 235).

3 Close the degasser cover. Note the correct orientation 4 Reinstall the degasser into the pump and reconnect the
of the cover with respect to the connector. degasser connector, see “Replace the Degasser” on
page 221.

Connector

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 230
9 Maintenance and Repair
Replace Parts Inside the 2-Channel-Degasser

Replace a Degasser Chamber

Parts required p/n Description


5067-4734 Vacuum Chamber

Preparations • Remove the degasser, see “Replace the Degasser” on page 221.
• Open the degasser cover, see “Replace Parts Inside the 2-Channel-Degasser” on page 230.

1 Disconnect the tubing from the chamber. 2 Remove the screw at the front of the degasser housing.

3 Remove the degasser chamber. 4 Install the new degasser chamber.

5 Tighten the screw at the front of the degasser. 6 Reconnect the tubing to the new chamber.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 231
9 Maintenance and Repair
Replace Parts Inside the 2-Channel-Degasser

Replace the Degasser Pump

Parts required p/n Description


G4280-60006 Micro Vacuum Pump

Preparations • Remove the degasser, see “Replace the Degasser” on page 221.
• Open the degasser cover, see “Replace Parts Inside the 2-Channel-Degasser” on page 230.

1 For replacing the degasser pump, disconnect both 2 Remove the three screws at the bottom of the
tubings and the cable to the mainboard. degasser.

3 Remove the pump. 4 Install the new pump.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 232
9 Maintenance and Repair
Replace Parts Inside the 2-Channel-Degasser

5 Fix the screws at the bottom of the degasser. 6 Connect the tubing from the chambers to the inlet of
the pump and the drain tubing to the holder at the
bottom of the degasser. Connect the cable to the
mainboard.

Drain Outlet

Inlet

Tubing holder

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 233
9 Maintenance and Repair
Replace Parts Inside the 2-Channel-Degasser

Replace the Sensor Board Assembly

Parts required p/n Description


G4280-60009 Pressure sensor assembly

Preparations • Remove the degasser, see “Replace the Degasser” on page 221.
• Open the degasser cover, see “Replace Parts Inside the 2-Channel-Degasser” on page 230.

1 For replacing the sensor board assembly, remove the 2 Open the clips next to the pressure sensor and lift up
cable from the mainboard. Disconnect the tubing from the board.
the pressure sensor.

3 Install the sensor board assembly. 4 Connect the cable to the mainboard.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 234
9 Maintenance and Repair
Replace Parts Inside the 2-Channel-Degasser

Replace the Degasser Tubing Kit

Parts required p/n Description


G4280-67300 Vacuum Tubing Kit

Preparations • Remove the degasser, see “Replace the Degasser” on page 221.
• Open the degasser cover, see “Replace Parts Inside the 2-Channel-Degasser” on page 230.

1 For replacing the degasser tubing kit, remove tubings 2 Connect the new tubing from the chambers to the inlet
from the vacuum pump, degasser chambers and the of the pump and the new drain tubing to the holder at
pressure sensor. the bottom of the degasser.

Drain outlet

Inlet
Tubing holder

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 235
9 Maintenance and Repair
Replace the Leak Panel

Replace the Leak Panel

When If the leak panel, main power switch or leak sensor are damaged

Tools required p/n Description


Flat head screwdriver
5023-2501 Screwdriver Torx-T10

Parts required p/n Description


5043-1053 Leak Plane
0515-2529 Screw Tapping PAN-HD-TORX T10 3x8 ST-ZN
G7111-40100 Leak Seal

Preparations • Switch off the pump at the main power switch.


• Remove the MCGV cover, see “Exchange the Multi-Channel Gradient Valve (MCGV)” on
page 180.
• Remove the pump head, see “Remove the Pump Head Assembly” on page 166.

1 Unscrew the Leak Panel screws. 2 Remove the leak panel.

3 Remove the sealing lip. 4 Remove or replace the leak sensor as described in
“Replace the Leak Sensor” on page 238

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 236
9 Maintenance and Repair
Replace the Leak Panel

5 Install the sealing lip. 6 Insert the new leak panel and push it to the bottom of
the module bottom panel. Make sure to keep the sealing
lip in the correct position.

7 Fix the Leak Panel screws.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 237
9 Maintenance and Repair
Replace the Leak Sensor

Replace the Leak Sensor

Tools required p/n Description


Screwdriver, flat-head
8710-0899 Screwdriver Pozidrive Shaft

Parts required p/n Description


5061-3356 Leak Sensor Assembly

Preparations Remove the leak panel, see “Replace the Leak Panel” on page 236.

1 Remove the leak sensor and disconnect the leak sensor 2 Replace the leak sensor.
cable from the power switch connector.

3 Route the leak sensor cable to the left and reconnect it 4 Reinsert the leak panel, see “Replace the Leak Panel” on
to the power switch board. page 236.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 238
9 Maintenance and Repair
Replace the Status LED

Replace the Status LED

Parts required p/n Description


5067-6214 Status LED Assembly includes Status LED Holder and cable
(The cable has adhesive stripes at the back)

Preparations • Remove the doors, see “Remove and Install Doors” on page 154.
• Remove the leak interface, see “Replace Leak Handling System Parts” on page 188.
• Remove the top covers and foam, see “Remove the Top Cover and Foam” on page 194.

1 The Status LED cable is connected on the mainboard to 2 Disconnect the cable from the mainboard
connector STATUS_LED [1] and pasted to the right (STATUS_LED).
sheet metal wall [2].

3 Pull the cable from the side wall starting from the rear. 4 Lift the Status LED to the top.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 239
9 Maintenance and Repair
Replace the Status LED

5 Insert the new Status LED. 6 Route the cable to the rear and fix it on the wall (there
are adhesive stripes on the back of the cable).

7 Reconnect the cable to the mainboard (STATUS_LED). Next Steps:


The cable’s contact side should be facing to the right
(wall).
8 Install the foam and top cover, see “Install the Top
Cover and Foam” on page 245.
9 Install the Leak Interface, see “Replace Leak Handling
System Parts” on page 188.
10 Install the Doors, see “Remove and Install Doors” on
page 154.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 240
9 Maintenance and Repair
Install an Active Seal Wash Upgrade Kit (Quaternary Pump)

Install an Active Seal Wash Upgrade Kit


(Quaternary Pump)

When This kit is recommended when using buffers or other salt solutions as solvents. It is used for
cleaning pump head pistons and seals.

Tools required p/n Description


Hexagonal key, 4 mm
8710-0899 Screwdriver Pozidrive Shaft
Wrench 1/4 inch

Parts required p/n Description


G1398A Active Seal Wash Option Kit

Preparations • Switch off the pump.


• Remove the doors.
• Remove tubing and cable connections to other modules.
• Remove pump module from the stack.
• Remove the pump head, see “Remove the Pump Head Assembly” on page 166.
• Remove top cover and foam, see “Remove the Top Cover and Foam” on page 194.

1 Place the pump head on a flat surface. Loosen the lock 2 While holding the lower half of the assembly (piston
screw (two turns). housing), carefully pull the pump housing away from the
piston housing.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 241
9 Maintenance and Repair
Install an Active Seal Wash Upgrade Kit (Quaternary Pump)

3 Lift the housing away from the pistons. 4 Place the support rings with seal wash seals on the
piston housing. Note the correct position of the pins on
the support ring.

5 Place the pump housing onto the piston housing. 6 Insert the pistons and carefully press them into the
seals.

7 Tighten the lock screw. 8 If needed, apply a small amount of grease to the pump
head screws. Slide the pump head assembly onto the
pump drive. Tighten the pump head screws stepwise
with increasing torque (5 Nm). See “Reinstall the Pump
Head Assembly” on page 176.
9 Reconnect all capillaries, tubes, and (if installed) the
active inlet valve cable to its connector.
10 Fix the seal wash pump motor to the Z-panel of the
pump housing using two Pozidriv screws and two
shock mounts.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 242
9 Maintenance and Repair
Install an Active Seal Wash Upgrade Kit (Quaternary Pump)

11 Put the new peristaltic pump cartridge onto the rod of 12 Connect the cable to the pump mainboard.
the pump motor and push the plastic clips into the
module housing.

SW

13 Do not confuse the connectors for the seal wash pump 14 Install the foam and top cover, see “Install the Top
and the pressure sensor, as this will destroy the Cover and Foam” on page 245.
mainboard!

Connector
pressure sensor

Connector
seal wash pump

15 Connect the seal wash tubes to the nozzles of the


support rings. Connect the peristaltic pump tubes to the
support rings outlet and to the adapter leading to the
waste bottle.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 243
9 Maintenance and Repair
Assemble the Main Cover

Assemble the Main Cover

When If cover is broken

Parts required p/n Description


G7104-68713 Cabinet Kit 180 Infinity II
(includes sides, bottom, top, leak adapter top and status indicator insert)

The plastics kit contains all parts, but it is not assembled.


NOTE

Wrong assembled
C AU T I O N
In case you insert the left or right side in the opposite position, you may not be
able to remove the side from the top part.
 Take care not to mix up left and right side.

1 Place the top part on the bench and insert the left and right side into the top
part.

Front

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 244
9 Maintenance and Repair
Install the Top Cover and Foam

Install the Top Cover and Foam

Tools required p/n Description


8710-0899 Screwdriver Pozidrive Shaft

Preparations Make sure that after your repair all assemblies, cables, capillaries and connectors are located in their
correct place.

1 Install the top foam section. 2 Install the top plate and slide it towards the rear.

3 Fix the screws at the rear end of the top plate. 4 Replace the cover.

Next Steps:

5 Install the leak funnel and the waste tube.


6 Place the pump back to its position in the stack, place the solvent cabinet back in place and re-connect all tubes and
capillaries.
7 Install the doors and reconnect the power cable to the pump. Switch the modules on.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 245
10 Parts for Maintenance and Repair

Accessory Kits 247


Pump Head Assembly Without Seal Wash 249
Pump Head Assembly with Seal Wash Option 251
Outlet Valve 254
Passive Inlet Valve 255
Purge Valve Assembly 256
Active Inlet Valve Assembly 257
1260 Infinity II Max Uptime Kit 259
1260 Infinity II Max Uptime Kit Bio 260
1260 Infinity II Low Dispersion Kit MCT 261
HPLC System Tool Kit 262
Solvent Cabinet 263
Bottle Head Assembly 264
Hydraulic Path of the Quaternary Pump 265
Hydraulic Path of the Isocratic Pump 267
Overview of Main Assemblies 268
Overview of Main Assemblies (Quaternary Pump) 268
Overview of Main Assemblies (Isocratic Pump) 271
Degasser 274
Four-Channel Degasser 274
Two-Channel Degasser (Isocratic Pump) 275
Cover Parts 276
Sheet Metal Kit 277
Foam Parts 278
Power, Status LED and Illumination LED 279
Leak Parts 280

This chapter provides information on parts for maintenance and repair.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 246
10 Parts for Maintenance and Repair
Accessory Kits

Accessory Kits

Accessory Kit G7111-68755


The Accessory Kit (G7111-68755) contains the following items:

# p/n Description
2 5043-1013 Tubing Clip IF-II

1 5181-1519 CAN cable, Agilent module to module, 1 m

1 5500-1246 Capillary ST 0.17 mm x 500 mm SI/SI

1 5500-1217 Capillary, ST, 0.17 mm x 900 mm SI/SX

3 5063-6527 Tubing, Silicon Rubber, 1.2 m, ID/OD 6/9 mm

1 G1311-90107 Algae note

3 5500-1169 Y Tube Connector ID 6.4

3 5500-1155 Tube Connector, 90 degree, ID 6.4

1 5043-1372 Tubing Connector Leak 3-1

2 5043-1373 Tubing Connector Leak Cap

2 0890-1195 Flexible sleeving 1.45 mm/2.5 mm, PTFE

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 247
10 Parts for Maintenance and Repair
Accessory Kits

Accessory Kit G5654-68755


For bio-inert modules use bio-inert parts only!

The Accessory Kit (G5654-68755) contains the following items:

# p/n Description
2 5043-1013 Tubing Clip IF-II

1 5181-1519 CAN cable, Agilent module to module, 1 m

1 G5611-60502 Capillary Ti 0.17 mm x 900 mm, L (Bio-inert)

1 5500-1264 Capillary Ti 0.17 mm x 500 mm, SL/SLV

3 5063-6527 Tubing, Silicon Rubber, 1.2 m, ID/OD 6/9 mm

1 G1311-90107 Algae note

3 5500-1155 Tube Connector, 90 degree, ID 6.4

2 5043-1373 Tubing Connector Leak Cap

1 5043-1372 Tubing Connector Leak 3-1

3 5500-1169 Y Tube Connector ID 6.4

2 0890-1195 Flexible sleeving 1.45 mm/2.5 mm, PTFE

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 248
10 Parts for Maintenance and Repair
Pump Head Assembly Without Seal Wash

Pump Head Assembly Without Seal Wash

The Pump Head 1200 SL without Seal Wash (G1312-60056) contains:

Item p/n Description


1 5063-6586 Sapphire piston, 100 µL

2 G1311-60002 Piston housing

3 5067-1560 Support Ring SL, no seal wash

4 5062-2484 Gasket, seal wash (pack of 6)

5 5042-8952 Seal holder

6 5063-6589 Piston seal PTFE, carbon filled, black (pack of 2), default

OR 0905-1420 PE seals (pack of 2)

7 G1311-25200 Pump chamber housing

8 G1312-60066 Passive inlet valve 1220/1260

OR G1312-60166 1260 Infinity Inlet Valve Type N

G1312-60025 Active inlet valve body, without cartridge (OPTIONAL)

G1312-60020 Cartridge for active inlet valve 600 bar


for G7110B and G7111B (OPTIONAL)

5062-8562 Active Inlet Valve Cartridge (400 bar)


for G7111A (OPTIONAL)

9 G1312-60067 Outlet valve

OR G1312-60167 Outlet valve (type N/SFC)

10 5042-1303 Lock screw

11 G7111-60061 Purge valve


for G7110B, G7111A, and G7111B

12 0515-2118 Pump head screw (M5, 60 mm)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 249
10 Parts for Maintenance and Repair
Pump Head Assembly Without Seal Wash

The Pump Head 1200 SL without Seal Wash (G1312-60056) includes items 1-7,
10 and 12.
For piston seals, see “Choosing the Right Pump Seals” on page 107.

1
9

10
2
3
4
5
6
12
7

8
11

Figure 25 Pump head assembly without seal wash option

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 250
10 Parts for Maintenance and Repair
Pump Head Assembly with Seal Wash Option

Pump Head Assembly with Seal Wash Option

For bio-inert modules use bio-inert parts only!

Pump head assembly with seal wash (G1312-60045) and Pump Head with Seal
Wash Function (Bio-inert) (G5611-60165) contain:

Item p/n Description


1 5063-6586 Sapphire piston, 100 µL

OR 5067-4695 Sapphire piston (Bio-inert)

2 G1311-60002 Piston housing

OR G1312-60062 Piston housing (incl. spring)


(for Bio-inert pump head)

3 01018-60027 Support ring seal wash


(OPTIONAL)

OR G5611-63010 Support Ring with Seal Wash Function (Bio-inert)

01018-23702 Insert tool

4 0905-1175 Wash seal (PTFE)

OR 0905-1718 Wash Seal PE

OR 0905-1731 Bio-Inert Wash Seal

5 5062-2484 Gasket, seal wash (pack of 6)

6 5042-8952 Seal holder

OR G5611-60016 Seal Holder (Bio-inert)


(including: Bio-inert seal keeper (G5611-26210), Backup Ring for
Seal Holder (G4220-24013))

G1398A Active Seal Wash Option Upgrade


(includes service)

G1398-60000 Active Seal Wash Option Kit

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 251
10 Parts for Maintenance and Repair
Pump Head Assembly with Seal Wash Option

Item p/n Description


7 5063-6589 Piston seal PTFE, carbon filled, black (pack of 2), default

OR 0905-1420 PE seals (pack of 2)

OR G5611-21503 Piston Seal PTFE (Bio-inert)

5065-9978 Tubing, 1 mm i.d., 3 mm o.d., silicone, 5 m


for seal wash option

8 G1311-25200 Pump chamber housing

OR G5611-25260 Pump Chamber Housing (Bio-inert)

9 G1312-60066 Passive inlet valve 1220/1260

OR G1312-60166 1260 Infinity Inlet Valve Type N

G1312-60025 Active inlet valve body, without cartridge (OPTIONAL)

G5611-60025 Active Inlet Valve (AIV, Bio-inert), without cartridge

G1312-60020 Cartridge for active inlet valve 600 bar


for G7110B, and G7111B (OPTIONAL)

5062-8562 Active Inlet Valve Cartridge (400 bar)


for G7111A (OPTIONAL)

G5611-60020 Cartridge for AIV (Bio-inert)

10 5042-1303 Lock screw

11 G1312-60067 Outlet valve

OR G1312-60167 Outlet valve (type N/SFC)

OR G5611-60067 Outlet Valve (Bio-Inert)

12 G7111-60061 Purge valve


for G7110B, G7111A, and G7111B

G5611-60064 Purge valve


for G5654A

13 0515-2118 Pump head screw (M5, 60 mm)

14 5065-4445 Peristaltic pump with PharMed tubing

15 G1311-60161 1220/1260 Seal Wash Pump Assembly


see note below

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 252
10 Parts for Maintenance and Repair
Pump Head Assembly with Seal Wash Option

The Pump head assembly with seal wash (G1312-60045) includes items 1-8, 10
and 13.

The 1220/1260 Seal Wash Pump Assembly (G1311-60161) does not include
NOTE screws. For installation, two screws (Screws for Seal Wash Pump Motor
(0515-1508)) are needed.

15

14
1

10 11 2

3
4
5
6
7
13
8
9
12

Figure 26 Pump Head with Seal Wash Option

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 253
10 Parts for Maintenance and Repair
Outlet Valve

Outlet Valve

p/n Description
G1312-60067 Outlet valve
for G7110B, G7111A, and G7111B

G5611-60067 Outlet Valve (Bio-Inert)


for G5654A

G1312-60167 Outlet Valve Type N/SFC


for G7110B, G7111A, and G7111B

Figure 27 Outlet valve

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 254
10 Parts for Maintenance and Repair
Passive Inlet Valve

Passive Inlet Valve

p/n Description
G1312-60066 Passive inlet valve 1220/1260

G1312-60166 1260 Infinity Inlet Valve Type N

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 255
10 Parts for Maintenance and Repair
Purge Valve Assembly

Purge Valve Assembly

For bio-inert modules use bio-inert parts only!

Item p/n Description


1 G5611-60064 Purge valve
for G5654A

1 G7111-60061 Purge valve


for G7110B, G7111A, and G7111B

2 01018-22707 PTFE frits (pack of 5)

3 5067-4728 Seal cap

Figure 28 Purge Valve Assembly

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 256
10 Parts for Maintenance and Repair
Active Inlet Valve Assembly

Active Inlet Valve Assembly

For bio-inert modules use bio-inert parts only!

Item p/n Description


G5699A Active Inlet Valve Upgrade Kit
includes service and the parts listed below

1 G1312-60025 Active inlet valve body, without cartridge

OR G5611-60025 Active Inlet Valve (AIV, Bio-inert), without cartridge

2 G1312-60020 Cartridge for active inlet valve 600 bar

OR 5062-8562 Active Inlet Valve Cartridge (400 bar)

OR G5611-60020 Cartridge for AIV (Bio-inert)

G1311-67304 Connecting tube


(needed for quaternary pump only)

G1311-61601 Active Inlet Valve Cable Assembly

0100-2298 Adapter, PEEK int. 1/4-28 to ext. 10-32


(needed for isocratic pump only)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 257
10 Parts for Maintenance and Repair
Active Inlet Valve Assembly

Active inlet valve cartridge

Active inlet valve

Figure 29 Active Inlet Valve Assembly

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 258
10 Parts for Maintenance and Repair
1260 Infinity II Max Uptime Kit

1260 Infinity II Max Uptime Kit

This kit will be sold optionally with the pump and is not essential for operation.
You can reorder the individual parts.

# p/n Description
1 01018-22707 PTFE frits (pack of 5)

1 0890-1763 Capillary PEEK 0.18 mm x 1.5 m

3 5022-2184 Union, stand LC flow, no fitting

2 5041-2168 Solvent inlet filter, 20 µm pore size

1 5065-4426 Colored finger-tight PEEK fittings, 10/PK

3 5500-1193 InfinityLab Quick Turn Capillary ST 0.17 mm x 105 mm, long


socket
Quick Turn

1 5500-1217 Capillary, ST, 0.17 mm x 900 mm SI/SX

2 5500-1246 Capillary ST 0.17 mm x 500 mm SI/SI

2 5500-1250 Capillary, ST, 0.17 mm x 120 mm SL/SL, long socket

1 8710-1930 Plastic tubing cutter

1 8710-2391 Rheotool socket wrench ¼ inch

1 5500-1191 InfinityLab Quick Turn Capillary ST 0.12 mm x 280 mm, long


socket

6 G7167-68703 Fitting Intermediate Kit

1 G7111-90120 Technical Note 1260 Infinity II Max Uptime Kit

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 259
10 Parts for Maintenance and Repair
1260 Infinity II Max Uptime Kit Bio

1260 Infinity II Max Uptime Kit Bio

This kit will be sold optionally with the pump and is not essential for operation.
You can reorder the individual parts.

# p/n Description
1 5043-0915 Fitting mounting tool

1 01018-22707 PTFE frits (pack of 5)

1 0890-1763 PEEK tubing 0.18 mm x 1500 mm

3 5067-4741 ZDV union (Bio-inert)

2 5041-2168 Solvent inlet filter, 20 µm pore size

1 5065-4426 Colored finger-tight PEEK fittings, 10/PK

1 G5611-60502 Capillary Ti 0.17 mm x 900 mm, L (Bio-inert)

1 5500-1264 Capillary Ti 0.17 mm x 500 mm, SL/SLV

1 G5667-81005 Capillary PK/ST 0.17 mm x 500 mm, RLO/RLO (Bio-inert)

1 8710-1930 Plastic tubing cutter

1 8710-2391 Rheotool socket wrench ¼ inch

1 G5654-90130 Technical Note 1260 Infinity II Max Uptime Kit Bio

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 260
10 Parts for Maintenance and Repair
1260 Infinity II Low Dispersion Kit MCT

1260 Infinity II Low Dispersion Kit MCT

This kit will be sold optionally with the pump and is not essential for operation.
You can reorder the individual parts.
The 1260 Infinity II Low Dispersion Kit (G7111-68708) contains:

p/n Description
5500-1201 Capillary ST 0.12 mm x 105 mm SL

5500-1157 Capillary ST 0.12 mm x 500 mm SL/S

G7116-60015 Quick Connect Heat Exchanger Standard

G7167-68703 Fitting Intermediate Kit

G7111-90130 Technical Note 1260 Infinity II Low Dispersion Kit MCT

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 261
10 Parts for Maintenance and Repair
HPLC System Tool Kit

HPLC System Tool Kit

5023-2504
(Hex Driver SW-4 slitted)
5023-3138 5023-2503
(Reversible Screwdriver) (Hex Driver
SW-5 slitted)
5023-2502
(Hex Driver SW-6,
35/ ¼” slitted)

5023-3088
8710-1924
8710-2409
8710-0510
8710-0510
5023-2500
8710-1534

8710-2394
(Hex Key 9/64“, 15cm)

Box with:
9301-0411 (Syringe, Plastic) 5023-3089
9301-1337 (Syringe Adapter) Torx Key Set (T8,T9,T10,T15,T20,T25)
0100-1710 (Mounting tool for flangeless nut)
0100-1681 (Adapter luer/barb) 5023-2499
01018-23702 (Seal Insert tool) (Hex Key Set)
5067-6127 (Blank Nut V) 5043-1361
5023-2653 (Hex Key 3/32“) (Hex Key Set Driver)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 262
10 Parts for Maintenance and Repair
Solvent Cabinet

Solvent Cabinet

p/n Description
5067-6871 Solvent Cabinet Kit

For details refer to: Usage Guideline for the Solvent Cabinet (01200-90150)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 263
10 Parts for Maintenance and Repair
Bottle Head Assembly

Bottle Head Assembly

The Bottle Head Assembly (G7120-60007) contains:

p/n Description
5063-6598 Ferrules with lock ring (10/Pk)

5063-6599 Tube screw (10/Pk)

Wire marker

5062-2483 Tube PTFE 1.5 mm x 5 m, 3 mm od

5062-8517 Inlet filter adapter (4/Pk)

5041-2168 Solvent inlet filter, 20 µm pore size

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 264
10 Parts for Maintenance and Repair
Hydraulic Path of the Quaternary Pump

Hydraulic Path of the Quaternary Pump

For bio-inert modules use bio-inert parts only!

Item # p/n Description


1 1 G7111-81600 Capillary
damper to inlet pump chamber

OR 1 G5654-81600 Capillary, Ti, damper to inlet valve


for G5654A

1 G7120-60007 Bottle Head Assembly


(not shown)

2 1 5067-5378 Connecting tube, MCGV to PIV

OR 1 G1311-67304 Connecting tube


MCGV to AIV

3 4 G7111-60100 Solvent Tubes including labels


for connection degasser to MCGV

4 1 G7111-81601 Capillary
outlet valve to damper

OR 1 G5654-81601 Capillary, Ti, outlet valve to damper


for G5654A

5 1 5062-2461 Waste tube, 5 m (reorder pack)

6 1 5500-1217 Capillary, ST, 0.17 mm x 900 mm SI/SX


Pump to Vialsampler with ICC
OR 1 5500-1246 Capillary ST 0.17 mm x 500 mm SI/SI

OR 1 5500-1264 Capillary Ti 0.17 mm x 500 mm, SL/SLV


for G5654A

1 0100-1847 PEEK adapter 1/4-28 to 10-32


(Adapter AIV to solvent inlet tubes) (not shown)

1 G1311-60006 Inline filter


(not shown) (OPTIONAL)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 265
10 Parts for Maintenance and Repair
Hydraulic Path of the Quaternary Pump

5
4

Figure 30 Hydraulic Flow Path of the Quaternary Pump

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 266
10 Parts for Maintenance and Repair
Hydraulic Path of the Isocratic Pump

Hydraulic Path of the Isocratic Pump

Item p/n Description


1 G7111-81600 Capillary
damper to inlet pump chamber

2 5500-1217 Capillary, ST, 0.17 mm x 900 mm SI/SX


Pump to Vialsampler with ICC

5500-1246 Capillary ST 0.17 mm x 500 mm SI/SI

3 G7111-81601 Capillary
outlet valve to damper

4 5062-2461 Waste tube, 5 m (reorder pack)

G7120-60007 Bottle Head Assembly


(not shown)

G1311-60006 Inline filter


(not shown) (OPTIONAL)

1
Figure 31 Hydraulic Flow Path of the Isocratic Pump

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 267
10 Parts for Maintenance and Repair
Overview of Main Assemblies

Overview of Main Assemblies

Overview of Main Assemblies (Quaternary Pump)


For bio-inert modules use bio-inert parts only!

6 9, 12

10

5
11
4
13
3

Figure 32 Overview of Main Assemblies (Quaternary Pump)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 268
10 Parts for Maintenance and Repair
Overview of Main Assemblies

Repair parts - pump housing and main assemblies (front view)

Item p/n Description


1 5067-5712 Power Switch Assembly Infinity II

2 5043-1053 Leak Plane

3 5061-3356 Leak Sensor Assembly

4 G1311-60070 Degasser 4 Channels for Quaternary Pump

5 0950-5274 Power supply

6 5067-6199 Fan 92 mm x 92 mm

7 G7111-65800 Mainboard 1260 Iso, Quat & Bio Pump 600 bar

OR G7111-65810 Mainboard 1260 Infinity II Quaternary Pump VL 400 bar

8 G4200-65880 Fusion Board

9 G1311-60065 Damper Isocratic/Quaternary Pump 600 bar

OR 79835-60005 Damper Quaternary Pump 400 bar

OR G5611-60065 Damper (Bio-inert)

10 5067-6214 Status LED Assembly includes Status LED Holder and cable
(The cable has adhesive stripes at the back)

11 G1311-60001 Pump drive

12 G1311-44162 Damper Leak Panel

13 G1398A Active Seal Wash Option Kit

G1311-61601 Active Inlet Valve Cable Assembly


(not shown) (OPTIONAL)

G1311-61600 Cable MCGV to mainboard


(not shown)

G1311-67701 Multi channel gradient valve (MCGV)

OR G5611-67701 Multi-Channel Gradient Valve (MCGV) (Bio-inert)

5041-8365 Blank plug


for MCGV

0515-0906 Screw for MCGV (M4, 40 mm)

1520-0401 Rubber holder for MCGV

G1311-44161 MCGV cover

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 269
10 Parts for Maintenance and Repair
Overview of Main Assemblies

Figure 33 Overview of Main Assemblies (Rear View)

Repair parts - pump housing and main assemblies (rear view)

Item p/n Description


1 0950-5274 Power supply

2 5067-5406 Screw Tx20, M4 x 8 mm


Plus washer, to fix power supply at rear panel

3 0515-0924 Grounding screw board to housing

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 270
10 Parts for Maintenance and Repair
Overview of Main Assemblies

Overview of Main Assemblies (Isocratic Pump)


Item p/n Description
1 5067-5712 Power Switch Assembly Infinity II

2 5043-1053 Leak Plane

3 5061-3356 Leak Sensor Assembly

4 5067-4661 Tubing kit 270 mm, 2/pk


(for re-ordering) Connection between Degassing unit and Passive Inlet
Valve

5 G4220-60035 Tubing kit 140 mm, 2/pk


(for re-ordering) Connection between Solvent Selection Valve and
Degassing unit

6 G4280-60070 Degassing unit

7 5067-5895 Solvent selection valve

8 0950-5274 Power supply

9 5067-6199 Fan 92 mm x 92 mm

10 G7111-65800 Mainboard 1260 Iso, Quat & Bio Pump 600 bar

11 G4200-65880 Fusion Board

12 G1311-60065 Damper Isocratic/Quaternary Pump 600 bar

13 5067-6214 Status LED Assembly includes Status LED Holder and cable
(The cable has adhesive stripes at the back)

14 G1311-60001 Pump drive

15 G1311-44162 Damper Leak Panel

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 271
10 Parts for Maintenance and Repair
Overview of Main Assemblies

11

10

9 12

13
8
14
7

6
15
5
4
3
2

1
Figure 34 Overview of Main Assemblies (Isocratic Pump)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 272
10 Parts for Maintenance and Repair
Overview of Main Assemblies

Item p/n Description


1 0950-5274 Power supply

2 5067-5406 Screw Tx20, M4 x 8 mm

3 0515-0924 Grounding screw board to housing

Figure 35 Overview of Main Assemblies (Rear View)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 273
10 Parts for Maintenance and Repair
Degasser

Degasser

Four-Channel Degasser
Item p/n Description
G1311-60070 Degasser 4 Channels for Quaternary Pump

1 G1311-67007 Housing Assembly for Quaternary Pump Degasser

2 5067-4734 Vacuum Chamber

3 G1311-67302 Tubing Kit for Quaternary Pump Degasser

4 G1311-60007 Vacuum Pump for Quaternary Pump Degasser

5 G1311-67600 Sensor Board Assembly for Quaternary Pump Degasser

G1311-61605 Degasser Cable Assembly


Connection between Degassing unit and Pump mainboard (not
shown)

Figure 36 Degasser Parts

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 274
10 Parts for Maintenance and Repair
Degasser

Two-Channel Degasser (Isocratic Pump)


Item p/n Description
G4280-60070 Degassing unit

1 G4280-60007 Degasser Housing Assembly

2 G4280-60009 Pressure sensor assembly

3 5067-4734 Vacuum Chamber

4 G4280-67300 Vacuum Tubing Kit

5 G4280-60006 Micro Vacuum Pump

4
5

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 275
10 Parts for Maintenance and Repair
Cover Parts

Cover Parts

Item p/n Description


1 5067-5746 Door Assembly Infinity 180 Right

2 5067-5745 Door Assembly Infinity 180 Left

3 5043-1354 Name Plate 1290 Infinity 2

4 G7104-68713 Cabinet Kit 180 Infinity II


(includes sides, bottom, top, leak adapter top and status indicator
insert)

Figure 37 Cover Parts

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 276
10 Parts for Maintenance and Repair
Sheet Metal Kit

Sheet Metal Kit

Item p/n Description


1 G7111-04301 Degasser Front Cover for 1260 Infinity II Quaternary Pump
including Metal cover screw with spring (5022-2112)

OR G7111-04302 Degasser Front Cover for 1260 Infinity II Isocratic Pump


including Metal cover screw with spring (5022-2112)
(not shown)

2 G7111-68701 Sheet Metal Kit, including Top, base and Z-Panel

Figure 38 Sheet Metal Kit

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 277
10 Parts for Maintenance and Repair
Foam Parts

Foam Parts

# p/n Description
1 G7111-68702 Foam kit (includes upper and lower foam)

6 1520-0404 Bushing for pump drive, rear


(not shown)

1 G7111-44000 Transport Foam


(not shown)

Figure 39 Foam Parts

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 278
10 Parts for Maintenance and Repair
Power, Status LED and Illumination LED

Power, Status LED and Illumination LED

Item p/n Description


1 0950-5274 Power supply

2 5067-5974 Power Switch Cable

3 5067-5712 Power Switch Assembly Infinity II

4 5067-6214 Status LED Assembly includes Status LED Holder and cable
(The cable has adhesive stripes at the back)

5067-5406 Screw Tx20, M4 x 8 mm

5067-5909 PWS Control Cable

Status LED Holder

Figure 40 Power, Status LED and Illumination LED

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 279
10 Parts for Maintenance and Repair
Leak Parts

Leak Parts

Item p/n Description


1 5043-1053 Leak Plane

2,3,4 5067-5712 Power Switch Assembly Infinity II

5 5067-5974 Power Switch Cable

6,7 5043-0856 Leak Adapter

8 5061-3356 Leak Sensor Assembly

9 G7111-40100 Leak Seal

5067-6214 Status LED Assembly includes Status LED Holder and cable
(The cable has adhesive stripes at the back)

3
7

2 8

9
1

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 280
11 Identifying Cables

Cable Overview 282


Analog Cables 284
Remote Cables 286
CAN Cable 290
Agilent Module to PC 291
Agilent 1200 Module to Printer 292

This chapter provides information on cables used with the Agilent 1200 Infinity
Series modules.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 281
11 Identifying Cables
Cable Overview

Cable Overview

Never use cables other than the ones supplied by Agilent Technologies to ensure
NOTE proper functionality and compliance with safety or EMC regulations.

Analog cables
p/n Description
35900-60750 Agilent 35900A A/D converter

01046-60105 Analog cable (BNC to general purpose, spade lugs)

Remote cables
p/n Description
5188-8029 ERI
to general purpose

5188-8044 Remote Cable ERI – ERI

5188-8045 Remote Cable APG – ERI

5188-8059 ERI-Extension-Cable 1.2 m

5061-3378 Remote Cable


to 35900 A/D converter

01046-60201 Agilent module to general purpose

5188-8057 Fraction Collection ERI remote Y-cable

CAN cables
p/n Description
5181-1516 CAN cable, Agilent module to module, 0.5 m

5181-1519 CAN cable, Agilent module to module, 1 m

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 282
11 Identifying Cables
Cable Overview

LAN cables
p/n Description
5023-0203 Cross-over network cable, shielded, 3 m (for point to point connection)

5023-0202 Twisted pair network cable, shielded, 7 m (for point to point


connection)

RS-232 cables
(not for p/n Description
FUSION board)
RS232-61601 RS-232 cable, 2.5 m
Instrument to PC, 9-to-9 pin (female). This cable has special pin-out,
and is not compatible with connecting printers and plotters. It is also
called "Null Modem Cable" with full handshaking where the wiring is
made between pins 1-1, 2-3, 3-2, 4-6, 5-5, 6-4, 7-8, 8-7, 9-9.

5181-1561 RS-232 cable, 8 m

USB cables
p/n Description
5188-8050 USB A M-USB Mini B 3 m (PC-Module)

5188-8049 USB A F-USB Mini B M OTG (Module to Flash Drive)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 283
11 Identifying Cables
Analog Cables

Analog Cables

One end of these cables provides a BNC connector to be connected to Agilent


modules. The other end depends on the instrument to which connection is being
made.

Agilent Module to 35900 A/D converters

p/n 35900-60750 35900 Pin Agilent Signal Name


module

1 Not connected

2 Shield Analog -

3 Center Analog +

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 284
11 Identifying Cables
Analog Cables

Agilent Module to BNC Connector

p/n 8120-1840 Pin BNC Pin Agilent Signal Name


module

Shield Shield Analog -

Center Center Analog +

Agilent Module to General Purpose

p/n 01046-60105 Pin Pin Agilent Signal Name


module

1 Not connected

2 Black Analog -

3 Red Analog +

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 285
11 Identifying Cables
Remote Cables

Remote Cables

ERI (Enhanced Remote Interface)


• 5188-8029 ERI to general purpose (D-Sub 15 pin male - open end)
• 5188-8044 ERI to ERI (D_Sub 15 pin male - male)
• 5188-8059 ERI-Extension-Cable 1.2 m (D-Sub15 pin male / female)

p/n 5188-8029 pin Color code Enhanced Classic Active (TTL)


Remote Remote

1 white IO1 START Low


REQUEST

2 brown IO2 STOP Low

3 green IO3 READY High

4 yellow IO4 PEAK DETECT Low

5 grey IO5 POWER ON High

6 pink IO6 SHUT DOWN Low

7 blue IO7 START Low

8 red IO8 PREPARE Low

9 black 1wire DATA

10 violet DGND

11 grey-pink +5V ERI out

12 red-blue PGND

13 white-green PGND

14 brown-green +24V ERI out

15 white-yellow +24V ERI out

NC yellow-brown

Configuration is different with old firmware revisions.


NOTE
The configuration for IO4 and IO5 is swapped for modules with firmware lower
than D.07.10.
Peak Detection is used for LCMS systems connected with the Fraction Collection
NOTE Remote Y-Cable (5188-8057).

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 286
11 Identifying Cables
Remote Cables

• 5188-8045 ERI to APG (Connector D_Subminiature 15 pin (ERI), Connector


D_Subminiature 9 pin (APG))

p/n 5188-8045 Pin (ERI) Signal Pin (APG) Active (TTL)

10 GND 1

1 Start Request 9 Low

2 Stop 8 Low

3 Ready 7 High

5 Power on 6 High

4 Future 5

6 Shut Down 4 Low

7 Start 3 Low

8 Prepare 2 Low

Ground Cable Shielding NC

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 287
11 Identifying Cables
Remote Cables

• 5188-8057 ERI to APG and RJ45 (Connector D_Subminiature 15 pin (ERI),


Connector D_Subminiature 9 pin (APG), Connector plug Cat5e (RJ45))
Table 25 5188-8057 ERI to APG and RJ45

p/n 5188-8057 Pin (ERI) Signal Pin (APG) Active (TTL) Pin (RJ45)

10 GND 1 5

1 Start 9 High
Request

2 Stop 8 High

3 Ready 7 High

4 Fraction 5 High 4
Trigger

5 Power on 6 High

6 Shut Down 4 High

7 Start 3 High

8 Prepare 2 High

Ground Cable NC
Shielding

One end of these cables provides an Agilent Technologies APG (Analytical


Products Group) remote connector to be connected to Agilent modules. The
other end depends on the instrument to be connected to.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 288
11 Identifying Cables
Remote Cables

Agilent Module to Agilent 35900 A/D Converters

p/n 5061-3378 Pin 35900 A/D Pin Agilent Signal Name Active
module (TTL)

1 - White 1 - White Digital ground

2 - Brown 2 - Brown Prepare run Low

3 - Gray 3 - Gray Start Low

4 - Blue 4 - Blue Shut down Low

5 - Pink 5 - Pink Not connected

6 - Yellow 6 - Yellow Power on High

7 - Red 7 - Red Ready High

8 - Green 8 - Green Stop Low

9 - Black 9 - Black Start request Low

Agilent Module to General Purpose

p/n 01046-60201 Wire Color Pin Agilent Signal Name Active


module (TTL)

White 1 Digital ground

Brown 2 Prepare run Low

Gray 3 Start Low

Blue 4 Shut down Low

Pink 5 Not connected

Yellow 6 Power on High

Red 7 Ready High

Green 8 Stop Low

Black 9 Start request Low

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 289
11 Identifying Cables
CAN Cable

CAN Cable

Both ends of this cable provide a modular plug to be connected to


Agilent modules CAN or LAN connectors.
CAN Cables

p/n Description
5181-1516 CAN cable, Agilent module to module, 0.5 m

5181-1519 CAN cable, Agilent module to module, 1 m

LAN Cables

p/n Description
5023-0203 Cross-over network cable, shielded, 3 m (for point to point connection)

5023-0202 Twisted pair network cable, shielded, 7 m (for point to point


connection)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 290
11 Identifying Cables
Agilent Module to PC

Agilent Module to PC

p/n Description
RS232-61601 RS-232 cable, 2.5 m
Instrument to PC, 9-to-9 pin (female). This cable has special pin-out,
and is not compatible with connecting printers and plotters. It is also
called "Null Modem Cable" with full handshaking where the wiring is
made between pins 1-1, 2-3, 3-2, 4-6, 5-5, 6-4, 7-8, 8-7, 9-9.

5181-1561 RS-232 cable, 8 m

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 291
11 Identifying Cables
Agilent 1200 Module to Printer

Agilent 1200 Module to Printer

To connect a USB Flash Drive use a USB OTG cable with Mini-B plug and A
socket.

p/n Description
5188-8050 USB A M-USB Mini B 3 m (PC-Module)

5188-8049 USB A F-USB Mini B M OTG (Module to Flash Drive)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 292
12 Hardware Information

The Low-Pressure Pump Mainboard (LPM) 294


Block Diagrams 294
FUSION Board 297
ERI (Enhanced Remote Interface) 300
Leak Sensor 301
Firmware Description 302
The Main Power Supply Assembly 305
Electrical Connections 307
Rear View of the Module 308
Serial Number Information 308
Interfaces 309
Overview Interfaces 311
ERI (Enhanced Remote Interface) 314
USB (Universal Serial Bus) 316
Setting the 6-bit Configuration Switch 317
Special Settings 319
Instrument Layout 321
Early Maintenance Feedback 322

This chapter describes the pump in more detail on hardware and electronics.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 293
12 Hardware Information
The Low-Pressure Pump Mainboard (LPM)

The Low-Pressure Pump Mainboard (LPM)

Block Diagrams
control signal
data/ address

diagnose signals (digital)


diagnose signals (analog)
temperature of hybrid

Figure 41 Block Diagram Signal Explanations

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 294
12 Hardware Information
The Low-Pressure Pump Mainboard (LPM)

MCGV 4 valve drives

ASIC
Active inlet
Valve drive
valve

Current Motor
control control
1
Data/
Pump head Drive address
Motor J
Control
Encoder

Processor
Motor drive

Multiplexer

Damper Pressure Filter & A/D converter


sensor amplifier

Integrator Analog Offset


output correction pressure converter

Leak sensor
Amplifier & J ambient measurement
Fan
Drive

Sensor

Figure 42 Block diagram low pressure mainboard (LPM)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 295
12 Hardware Information
The Low-Pressure Pump Mainboard (LPM)

Hardware +36 V
Motor J16
+24 V switch
Motor drive
Encoder J17
+/-15V
Processor

Inlet +36V
J19 Electronic
AIV Z-panel valvedrive fuse

MCGV +36 V
MCGV J30 Electronic Hardware
Z-panel fuse switch
driveC&D +24 V

+36 V
MCGV Electronic
driveA&B fuse

+15 V Leak J23 Leak


converter sensor

+24 V J25
Fan drive Fan
Active seal P7 Seal wash +24 V
wash driver J24
Damper
+/-15V Damper
converter
J3 analog
output

+36 V
LPM
+24 V power J9 Power
supply supply
+/-15V
+5V

Figure 43 Interconnection diagram low pressure mainboard (LPM)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 296
12 Hardware Information
The Low-Pressure Pump Mainboard (LPM)

FUSION Board
The FUSION board holds the CORE-electronics for all Infinity II modules that use
the FUSION board (at introduction G71xx detectors and Multisamplers).
The FUSION board is the same for all Infinity II mainboards that host a FUSION
board.

FUSION board

Mainboard

Figure 44 Position of FUSION board on mainboard

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 297
12 Hardware Information
The Low-Pressure Pump Mainboard (LPM)

Features
• Configuration Switch (6 dip switches), see “Fusion Configuration Switch”.
• CAN (2 x) for inter module communication
• LAN (Local Area Network) , supports
• a PC with control software (e.g. Agilent ChemStation or Lab Advisor)
• USB (Universal Serial Bus) - replaces RS232, see “USB (Universal Serial Bus)”.
It supports:
• a PC with control software (e.g. Agilent ChemStation or Lab Advisor)
• USB Flash Disk (e.g. Agilent Feature Dongle).
• The USB connector provides the same functionally as the LAN interface. It
can be used by:
• Agilent Open Lab CDS (C.01.XX and later)
• Agilent Lab Advisor (B.02.06 and later)
• Firmware (Main and Resident)
• Module type will be read from the mainboard

Connectors

CAN

Configuration switch

LAN

USB

Figure 45 Connectors of the FUSION board

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 298
12 Hardware Information
The Low-Pressure Pump Mainboard (LPM)

Replacement Information
• The FUSION board is not part of the mainboard.
• It’s recommended to have one spare board in the service equipment. Can be
reused for all FUSION type modules.
• Can be replaced individually, see “Troubleshooting FUSION Board” (from other
module or as individual part)
• A new FUSION board contains just a resident firmware.
• During installation both, resident and main firmware (from the same firmware
set) should be installed.
• If the FUSION board is new or was used with a different module, it will turn the
module into resident mode; see “Replacement of Main and FUSION Board”.
• Depending on the Module Type (e.g. G7167 Multisampler) special actions are
required for initial operation (beside of setting type and serial number and
firmware update); see “Replacement of Main and FUSION Board”.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 299
12 Hardware Information
The Low-Pressure Pump Mainboard (LPM)

ERI (Enhanced Remote Interface)


ERI replaces the AGP Remote Interface used in the HP 1090/1040/1050/1100
NOTE HPLC systems and Agilent 1100/1200/1200 Infinity HPLC modules. All new
1200 Infinity II products using the FUSION core electronics use ERI. This
interface is already used in the Agilent Universal Interface Box 2 (UIB2).

Description
This interface contains 8 individual programmable input/output pins. In addition
it provides 24 V power and 5 V power and a serial data line to detect and
recognize further add-ons that could be connected to this interface. This way the
interface can support various additional devices like sensors, triggers (in and out)
and small controllers, etc.

ERI

Figure 46 Location of the ERI interface (example shows a G7114A/B VWD)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 300
12 Hardware Information
The Low-Pressure Pump Mainboard (LPM)

Leak Sensor
The module has a PTC (Positive Temperature Coefficient) for the leak detection
and an NTC (Negative Temperature Coefficient) for the ambient temperature
compensation. A leak would cool down the PTC and this change in resistance
would generate a leak signal. The NTC is located on the power switch assembly.

Solvent incompatibility
C AU T I O N
The solvent DMF (dimethylformamide) leads to corrosion of the leak sensor.
The material of the leak sensor, PVDF (polyvinylidene fluoride), is incompatible
with DMF.
 Do not use DMF as mobile phase.
 Check the leak sensor regularly for corrosion.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 301
12 Hardware Information
Firmware Description

Firmware Description

The firmware of the instrument consists of two independent sections:


• a non-instrument specific section, called resident system
• an instrument specific section, called main system

Resident System
This resident section of the firmware is identical for all Agilent
1100/1200/1220/1260/1290 series modules. Its properties are:
• the complete communication capabilities (CAN, LAN, USB and RS- 232)
• memory management
• ability to update the firmware of the 'main system'

Main System
Its properties are:
• the complete communication capabilities (CAN, LAN, USB and RS- 232)
• memory management
• ability to update the firmware of the 'resident system'
In addition the main system comprises the instrument functions that are divided
into common functions like
• run synchronization through APG/ERI remote,
• error handling,
• diagnostic functions,
• or module specific functions like
• internal events such as lamp control, filter movements,
• raw data collection and conversion to absorbance.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 302
12 Hardware Information
Firmware Description

Firmware Updates
Firmware updates can be done with the Agilent Lab Advisor software with files
on the hard disk (latest version should be used).
Required tools, firmware and documentation are available from the Agilent web:
https://www.agilent.com/en-us/firmwareDownload?whid=69761
The file naming conventions are:
PPPP_RVVV_XXX.dlb, where
• PPPP is the product number, for example, 1315B for the G1315B DAD,
• R the firmware revision, for example, A for G1315B or B for the G1315C DAD,
• VVV is the revision number, for example 650 is revision 6.50,
• XXX is the build number of the firmware.
For instructions on firmware updates refer to section Replacing Firmware in
chapter "Maintenance" or use the documentation provided with the Firmware
Update Tools.

Update of main system can be done in the resident system only. Update of the
NOTE resident system can be done in the main system only.
Main and resident firmware must be from the same set.

Main FW update
Resident System Main System

Resident FW Update

Figure 47 Firmware Update Mechanism

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 303
12 Hardware Information
Firmware Description

Some modules are limited in downgrading due to their mainboard version or


NOTE their initial firmware revision. For example, a G1315C DAD SL cannot be
downgraded below firmware revision B.01.02 or to a A.xx.xx.
Some modules can be re-branded (e.g. G1314C to G1314B) to allow operation in
specific control software environments. In this case, the feature set of the target
type is used and the feature set of the original one is lost. After re-branding (e.g.
from G1314B to G1314C), the original feature set is available again.
All this specific information is described in the documentation provided with the
firmware update tools.

The firmware update tools, firmware and documentation are available from the
Agilent web.
• https://www.agilent.com/en-us/firmwareDownload?whid=69761

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 304
12 Hardware Information
The Main Power Supply Assembly

The Main Power Supply Assembly

The main power supply comprises a closed assembly and must not be
disassembled further for safety reasons. In case of a defect, the entire power
supply needs to be replaced.
The power supply provides all DC voltages used in the module. The line voltage
can vary in a range from 100 – 240 volts AC ± 10 % and needs no manual setting.

MPS
+36V (not used)
+36V
supply

line primary +24V


+24V / 4.5A
voltage circuits supply

+5V
switching +5V / 3A
regulator

linear +15V / .3A


voltage
regulators -15V / .3A

Figure 48 Block Diagram of Standard Power Supply

To disconnect the instrument from line, unplug the power cord. The power
NOTE supply still uses some power, even if the power switch on the front panel is
turned off.

No accessible hardware fuse is needed because the main power supply is safe
against any short circuits or overload conditions on the output lines. When
overload conditions occur, the power supply turns off all output voltages. Turning
the line power off and on again resets the power supply to normal operation if the
cause of the overload condition has been removed.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 305
12 Hardware Information
The Main Power Supply Assembly

An over-temperature sensor in the main power supply is used to turn off output
voltages if the temperature exceeds the acceptable limit (for example, if the
cooling fan of the instrument fails). To reset the main power supply to normal
operating conditions, turn the instrument off, wait until it is approximately at
ambient temperature and turn the instrument on again.
The following table gives the specifications of the main power supply.
Table 26 Power Supply Specifications (Standard)

Maximum power 160 VA / 130 W Continuous output

Line Input 100 – 240 volts AC ± 10 %, line Wide ranging


frequency of 50/60 Hz

Pin 1 Power Fail error message

Pin 2 AGND

Pin 3 -15 VDC

Pin 4 +15 VDC

Pin 5 PGND

Pin 6 PGND

Pin 7 +24 VDC

Pin 8 +24 VDC

Pin 9 +36 VDC

Pin 10 +36 VDC

Pin 11 DGND

Pin 12 + 5 VDC

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 306
12 Hardware Information
Electrical Connections

Electrical Connections

Electrical Connections
• The CAN bus is a serial bus with high-speed data transfer. The two
connectors for the CAN bus are used for internal module data transfer and
synchronization.
• The ERI/REMOTE connector may be used in combination with other analytical
instruments from Agilent Technologies if you want to use features such as
start, stop, common shutdown, prepare, and so on.
• With the appropriate software, the LAN connector may be used to control the
module from a computer through a LAN connection. This connector is
activated and can be configured with the configuration switch.
• With the appropriate software, the USB connector may be used to control the
module from a computer through a USB connection.
• The power input socket accepts a line voltage of 100 – 240 VAC ± 10 % with a
line frequency of 50 or 60 Hz. Maximum power consumption varies by
module. There is no voltage selector on your module because the power
supply has wide-ranging capability. There are no externally accessible fuses
because automatic electronic fuses are implemented in the power supply.

Never use cables other than the ones supplied by Agilent Technologies to ensure
NOTE proper functionality and compliance with safety or EMC regulations.

Electric Shock
Electric shock due to insufficient insulation of connected instruments
WAR N IN G
Personal injury or damage to the instrument
 Any other instruments connected to this instrument shall be approved to
a suitable safety standard and must include reinforced insulation from
the mains.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 307
12 Hardware Information
Electrical Connections

Rear View of the Module

Configuration switch

USB-Mini-Port

LAN

ERI

CAN

Power socket

Figure 49 Rear view of the pump – electrical connections and label

Serial Number Information


The serial number information on the instrument labels provide the following
information:

CCXZZ00000 Format

CC Country of manufacturing
• DE = Germany
• JP = Japan
• CN = China

X Alphabetic character A-Z (used by manufacturing)

ZZ Alpha-numeric code 0-9, A-Z, where each combination


unambiguously denotes a module (there can be more than one
code for the same module)

00000 Serial number

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 308
12 Hardware Information
Interfaces

Interfaces

The Agilent InfinityLab LC Series modules provide the following interfaces:


Table 27 Agilent InfinityLab LC Series Interfaces

Module CAN USB LAN RS-232 Analog APG (A) Special


(on-board) / ERI (E)

Pumps

G7104A/C 2 No Yes Yes 1 A

G7110B 2 Yes Yes No No E

G7111A/B, G5654A 2 Yes Yes No No E

G7112B 2 Yes Yes No No E

G7120A, G7132A 2 No Yes Yes 1 A

G7161A/B 2 Yes Yes No No E

Samplers

G7129A/B/C 2 Yes Yes No No E

G7167A/B/C, G7137A, 2 Yes Yes No No E


G5668A, G3167A

G7157A 2 Yes Yes No No E

Detectors

G7114A/B 2 Yes Yes No 1 E

G7115A 2 Yes Yes No 1 E

G7117A/B/C 2 Yes Yes No 1 E

G7121A/B 2 Yes Yes No 1 E

G7162A/B 2 Yes Yes No 1 E

G7165A 2 Yes Yes No 1 E

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 309
12 Hardware Information
Interfaces

Table 27 Agilent InfinityLab LC Series Interfaces

Module CAN USB LAN RS-232 Analog APG (A) Special


(on-board) / ERI (E)

Fraction Collectors

G7158B 2 Yes Yes No No E

G7159B 2 Yes Yes No No E

G7166A 2 No No No No No Requires a host module


with on-board LAN with
minimum FW B.06.40 or
C.06.40, or with
additional G1369C LAN
Card

G1364E/F, G5664B 2 Yes Yes No No E THERMOSTAT for


G1330B

Others

G1170A 2 No No No No No

G7116A/B 2 No No No No No Requires a host module


with on-board LAN or
with additional G1369C
LAN Card.

G7122A No No No Yes No A

G7170B 2 No No No No No Requires a host module


with on-board LAN with
minimum FW B.06.40 or
C.06.40, or with
additional G1369C LAN
Card

The detector (DAD/MWD/FLD/VWD/RID) is the preferred access point for


NOTE control via LAN. The inter-module communication is done via CAN.

• CAN connectors as interface to other modules


• LAN connector as interface to the control software
• RS-232C as interface to a computer
• USB (Universal Series Bus) as interface to a computer
• REMOTE connector as interface to other Agilent products
• Analog output connector(s) for signal output

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 310
12 Hardware Information
Interfaces

Overview Interfaces
CAN
The CAN is inter-module communication interface. It is a 2-wire serial bus
system supporting high speed data communication and real-time requirement.

LAN
The modules have either an interface slot for a LAN card (e.g. Agilent G1369B/C
LAN Interface) or they have an on-board LAN interface (e.g. detectors G1315C/D
DAD and G1365C/D MWD). This interface allows the control of the
module/system via a PC with the appropriate control software. Some modules
have neither on-board LAN nor an interface slot for a LAN card (e.g. G1170A
Valve Drive or G4227A Flexible Cube). These are hosted modules and require a
Host module with firmware B.06.40 or later or with additional G1369C LAN Card.

If an Agilent detector (DAD/MWD/FLD/VWD/RID) is in the system, the LAN


NOTE should be connected to the DAD/MWD/FLD/VWD/RID (due to higher data load).
If no Agilent detector is part of the system, the LAN interface should be installed
in the pump or autosampler.

USB
The USB interface replaces the RS-232 Serial interface in new FUSION generation
modules. For details on USB refer to “USB (Universal Serial Bus)” on page 316.

Analog Signal Output


The analog signal output can be distributed to a recording device. For details
refer to the description of the module’s mainboard.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 311
12 Hardware Information
Interfaces

Remote (ERI)
The ERI (Enhanced Remote Interface) connector may be used in combination
with other analytical instruments from Agilent Technologies if you want to use
features as common shut down, prepare, and so on.
It allows easy connection between single instruments or systems to ensure
coordinated analysis with simple coupling requirements.
The subminiature D connector is used. The module provides one remote
connector which is inputs/outputs (wired- or technique).
To provide maximum safety within a distributed analysis system, one line is
dedicated to SHUT DOWN the system’s critical parts in case any module detects
a serious problem. To detect whether all participating modules are switched on
or properly powered, one line is defined to summarize the POWER ON state of all
connected modules. Control of analysis is maintained by signal readiness READY
for next analysis, followed by START of run and optional STOP of run triggered on
the respective lines. In addition PREPARE and START REQUEST may be issued.
The signal levels are defined as:
• standard TTL levels (0 V is logic true, + 5.0 V is false),
• fan-out is 10,
• input load is 2.2 kOhm against + 5.0 V, and
• output are open collector type, inputs/outputs (wired- or technique).

All common TTL circuits operate with a 5 V power supply. A TTL signal is defined
NOTE as "low" or L when between 0 V and 0.8 V and "high" or H when between 2.0 V
and 5.0 V (with respect to the ground terminal).

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 312
12 Hardware Information
Interfaces

Table 28 ERI signal distribution

Pin Signal Description

1 START REQUEST (L) Request to start injection cycle (for example, by start key on any
module). Receiver is the autosampler.

2 STOP (L) Request to reach system ready state as soon as possible (for
example, stop run, abort or finish and stop injection). Receiver is any
module performing run-time controlled activities.

3 READY (H) System is ready for next analysis. Receiver is any sequence
controller.

4 POWER ON (H) All modules connected to system are switched on. Receiver is
any module relying on operation of others.

5 Not used

6 SHUT DOWN (L) System has serious problem (for example, leak: stops pump).
Receiver is any module capable to reduce safety risk.

7 START (L) Request to start run / timetable. Receiver is any module


performing run-time controlled activities.

8 PREPARE (L) Request to prepare for analysis (for example, calibration,


detector lamp on). Receiver is any module performing pre-analysis
activities.

Special Interfaces
There is no special interface for this module.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 313
12 Hardware Information
Interfaces

ERI (Enhanced Remote Interface)


ERI replaces the AGP Remote Interface that is used in the HP
1090/1040/1050/1100 HPLC systems and Agilent 1100/1200/1200 Infinity
HPLC modules. All new InfinityLab LC Series products using the FUSION core
electronics use ERI. This interface is already used in the Agilent Universal
Interface Box 2 (UIB2)

ERI Description
The ERI interface contains eight individual programmable input/output pins. In
addition, it provides 24 V power and 5 V power and a serial data line to detect and
recognize further add-ons that could be connected to this interface. This way the
interface can support various additional devices like sensors, triggers (in and out)
and small controllers, etc.

ERI

Figure 50 Location of the ERI interface (example shows a G7114A/B VWD)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 314
12 Hardware Information
Interfaces

Pin Enhanced Remote

1 IO 1 (START
REQUEST)

2 IO 2 (STOP)

3 IO 3 (READY)

4 IO 4 (POWER ON)

5 IO 5 (NOT USED)

6 IO 6 (SHUT DOWN)

7 IO 7 (START)

8 IO 8 (PREPARE)

9 1 wire DATA

10 DGND

11 +5 V ERI out

12 PGND

13 PGND

14 +24 V ERI out

15 +24 V ERI out

IO (Input/Output) Lines
• Eight generic bi-directional channels (input or output).
• Same as the APG Remote.
• Devices like valves, relays, ADCs, DACs, controllers can be
supported/controlled.

1-Wire Data (Future Use)


This serial line can be used to read out an EPROM or write into an EPROM of a
connected ERI-device. The firmware can detect the connected type of device
automatically and update information in the device (if required).

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 315
12 Hardware Information
Interfaces

5V Distribution (Future Use)


• Available directly after turning on the hosting module (assures that the
firmware can detect certain basic functionality of the device).
• For digital circuits or similar.
• Provides 500 mA maximum.
• Short-circuit proof with automatic switch off (by firmware).

24V Distribution (Future Use)


• Available by firmware command (defined turn on/off).
• For devices that need higher power
• Class 0: 0.5 A maximum (12 W)
• Class 1: 1.0 A maximum (24 W)
• Class 2: 2.0 A maximum (48 W)
• Class depends on hosting module’s internal power overhead.
• If a connected device requires more power the firmware detects this
(overcurrent detection) and provides the information to the user interface.
• Fuse used for safety protection (on board).
• Short circuit will be detected through hardware.

USB (Universal Serial Bus)


USB (Universal Serial Bus) - replaces RS232, supports:
• a PC with control software (for example Agilent Lab Advisor)
• USB Flash Disk

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 316
12 Hardware Information
Setting the 6-bit Configuration Switch

Setting the 6-bit Configuration Switch

The 6-bit configuration switch is located at the rear of the module with FUSION
electronics. Switch settings provide configuration parameters for LAN and
instrument specific initialization procedures.
All modules with FUSION electronics:
• Default is ALL switches DOWN (best settings).
• Default IP address for LAN 192.168.254.11
• For specific LAN modes switches 4-5 must be set as required.
• For boot resident/cold start modes switches 1+2 or 6 must be UP.

Configuration switch

Figure 51 Location of Configuration switch (example shows a G7114A/B VWD)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 317
12 Hardware Information
Setting the 6-bit Configuration Switch

Table 29 6-bit Configuration Switch

Mode Function/Setting

Switch 1 Switch 2 Switch 3 Switch 4 Switch 5 Switch 6

COM1 0 n.a.2 n.a. LAN Init Mode n.a.

Use Default IP Address3 0 0 0 0 0

Use Stored IP Address 0 0 0 1 0

Use DHCP to request IP Address4 0 0 1 0 0

Test 1 System n.a. n.a. n.a. ColdStart

Boot Main System / Keep Data 0 0 0 0 0

Boot Resident System / Keep Data 1 0 0 0 0

Boot Main System / Revert to


0 0 0 0 1
Default Data

Boot Resident System / Revert to


1 0 0 0 1
Default Data

1 When selecting mode COM, settings are stored to non-volatile memory. When selecting mode Test, COM settings are taken from
non-volatile memory.

2 not assigned - Always keep these switches on position ‘0’ (off)

3 Default IP Address is 192.168.254.11

4 Host Name will be the MAC address.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 318
12 Hardware Information
Setting the 6-bit Configuration Switch

Special Settings
Boot-Resident/Main
Firmware update procedures may require this mode in case of firmware loading
errors (main/resident firmware part).
If you use the following switch settings and power the instrument up again, the
instrument firmware stays in the resident/main mode. In resident mode, it is not
operable as a module. It only uses basic functions of the operating system for
example, for communication. In this mode the main firmware can be loaded
(using update utilities).

Forced Cold Start


A forced cold start can be used to bring the module into a defined mode with
default parameter settings.
• Boot Main System / Revert to Default Data
The instrument will boot to main mode and changes to the module’s default
parameter. May be also required to load resident firmware into the module.
• Boot Resident System / Revert to Default Data
The instrument will boot to resident mode and changes to the module’s
default parameter. May be also required to load main firmware into the
module.

Loss of data
C AU T I O N
Forced cold start erases all methods and data stored in the non-volatile
memory. Exceptions are calibration settings, diagnosis and repair log books
which will not be erased.
 Save your methods and data before executing a forced cold start.

If you use the following switch settings and power the instrument up again, it will
start as described above.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 319
12 Hardware Information
Setting the 6-bit Configuration Switch

Table 30 Boot Resident / Forced Coldstart

SW1 SW2 SW3 SW4 SW5 SW6 Init Mode

1 0 0 0 0 0 Boot Main System / Keep Data

1 1 0 0 0 0 Boot Resident System / Keep Data

1 0 0 0 0 1 Boot Main System / Revert to Default Data

1 1 0 0 0 1 Boot Resident System / Revert to Default Data

Note: The setting '0' (down) is essential.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 320
12 Hardware Information
Instrument Layout

Instrument Layout

The industrial design of the module incorporates several innovative features. It


uses Agilent’s E-PAC concept for the packaging of electronics and mechanical
assemblies. This concept is based upon the use of expanded polypropylene
(EPP) layers of foam plastic spacers in which the mechanical and electronic
boards components of the module are placed. This pack is then housed in a
metal inner cabinet which is enclosed by a plastic external cabinet. The
advantages of this packaging technology are:
• virtual elimination of fixing screws, bolts or ties, reducing the number of
components and increasing the speed of assembly/disassembly,
• the plastic layers have air channels molded into them so that cooling air can
be guided exactly to the required locations,
• the plastic layers help cushion the electronic and mechanical parts from
physical shock, and
• the metal inner cabinet shields the internal electronics from electromagnetic
interference and also helps to reduce or eliminate radio frequency emissions
from the instrument itself.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 321
12 Hardware Information
Early Maintenance Feedback

Early Maintenance Feedback

Maintenance requires the exchange of components that are subject to wear or


stress. Ideally, the frequency at which components are exchanged should be
based on the intensity of use of the module and the analytical conditions, and not
on a predefined time interval. The early maintenance feedback (EMF) feature
monitors the use of specific components in the instrument, and provides
feedback when the user-selectable limits have been exceeded. The visual
feedback in the user interface provides an indication that maintenance
procedures should be scheduled.

EMF Counters
EMF counters increment with use and can be assigned a maximum limit which
provides visual feedback in the user interface when the limit is exceeded. Some
counters can be reset to zero after the required maintenance procedure.

Using the EMF Counters


The user-settable EMF limits for the EMF Counters enable the early maintenance
feedback to be adapted to specific user requirements. The useful maintenance
cycle is dependent on the requirements for use. Therefore, the definition of the
maximum limits need to be determined based on the specific operating
conditions of the instrument.

Setting the EMF Limits


The setting of the EMF limits must be optimized over one or two maintenance
cycles. Initially the default EMF limits should be set. When instrument
performance indicates maintenance is necessary, take note of the values
displayed by the EMF counters. Enter these values (or values slightly less than the
displayed values) as EMF limits, and then reset the EMF counters to zero. The
next time the EMF counters exceed the new EMF limits, the EMF flag will be
displayed, providing a reminder that maintenance needs to be scheduled.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 322
13 LAN Configuration

What You Have to Do First 324


TCP/IP parameter configuration 325
Configuration Switches 326
Initialization Mode Selection 327
Dynamic Host Configuration Protocol (DHCP) 329
General Information (DHCP) 329
Setup (DHCP) 330
Manual Configuration 332
With Telnet 333

This chapter provides information on connecting the module to the Agilent


ChemStation PC.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 323
13 LAN Configuration
What You Have to Do First

What You Have to Do First

The module has an on- board LAN communication interface.

This chapter is generic and may show figures that differ from your module. The
NOTE functionality is the same.

1 Note the MAC (Media Access Control) address for further reference. The MAC
or hardware address of the LAN interfaces is a world wide unique identifier.
No other network device will have the same hardware address. The MAC
address can be found on a label at the rear of the module underneath the
configuration switch (see Figure 53 on page 324).

Part number of the detector mainboard


Revision Code, Vendor, Year and Week of assembly
MAC address
Country of Origin

Figure 52 MAC-Label

2 Connect the instrument's LAN interface (see Figure 53 on page 324) to


• the PC network card using a crossover network cable (point-to-point) or
• a hub or switch using a standard LAN cable.

LAN interface

MAC label

Figure 53 Location of LAN interfaces and MAC label

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 324
13 LAN Configuration
TCP/IP parameter configuration

TCP/IP parameter configuration

To operate properly in a network environment, the LAN interface must be


configured with valid TCP/IP network parameters. These parameters are:
• IP address
• Subnet Mask
• Default Gateway
The TCP/IP parameters can be configured by the following methods:
• by automatically requesting the parameters from a network-based DHCP
Server (using the so-called Dynamic Host Configuration Protocol). This mode
requires a LAN-onboard Module or a G1369C LAN Interface card, see “Setup
(DHCP)” on page 330
• by manually setting the parameters using Telnet
• by manually setting the parameters using the Local Controller
The LAN interface differentiates between several initialization modes. The
initialization mode (short form ‘init mode’) defines how to determine the active
TCP/IP parameters after power-on. The parameters may be derived non-volatile
memory or initialized with known default values. The initialization mode is
selected by the configuration switch, see Table 31 on page 327.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 325
13 LAN Configuration
Configuration Switches

Configuration Switches

The configuration switch can be accessed at the rear of the module.

Configuration switch

Figure 54 Location of Configuration switch (example shows a G7114A/B VWD)

The module is shipped with all switches set to OFF, as shown above.

To perform any LAN configuration, SW1 and SW2 must be set to OFF.
NOTE

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 326
13 LAN Configuration
Initialization Mode Selection

Initialization Mode Selection

The following initialization (init) modes are selectable:


Table 31 Initialization Mode Switches

SW1 SW2 SW3 SW4 SW5 SW6 Init Mode

0 0 0 0 0 0 Use Default IP Address

0 0 0 0 1 0 Use Stored IP Address

0 0 0 1 0 0 Use DHCP

Note: The setting ‘0’ (down) is essential.

Default IP address for LAN is 192.168.254.11.


DHCP address is the module’s LAN MAC address.

Using Stored
When initialization mode Using Stored is selected, the parameters are taken from
the non-volatile memory of the module. The TCP/IP connection will be
established using these parameters. The parameters were configured previously
by one of the described methods.

Non-Volatile
RAM
Active
Parameter
Stored
Parameter

Figure 55 Using Stored (Principle)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 327
13 LAN Configuration
Initialization Mode Selection

Using Default
When Using Default is selected, the factory default parameters are taken instead.
These parameters enable a TCP/IP connection to the LAN interface without
further configuration, see Table 32 on page 328.

Default Active
Parameter Parameter

Figure 56 Using Default (Principle)

Using the default address in your local area network may result in network
NOTE problems. Take care and change it to a valid address immediately.

Table 32 Using Default Parameters

IP address: 192.168.254.11

Subnet Mask: 255.255.255.0

Default Gateway not specified

Since the default IP address is a so-called local address, it will not be routed by
any network device. Thus, the PC and the module must reside in the same
subnet.
The user may open a Telnet session using the default IP address and change the
parameters stored in the non-volatile memory of the module. He may then close
the session, select the initialization mode Using Stored, power-on again and
establish the TCP/IP connection using the new parameters.
When the module is wired to the PC directly (e.g. using a cross-over cable or a
local hub), separated from the local area network, the user may simply keep the
default parameters to establish the TCP/IP connection.

In the Using Default mode, the parameters stored in the memory of the module
NOTE are not cleared automatically. If not changed by the user, they are still available,
when switching back to the mode Using Stored.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 328
13 LAN Configuration
Dynamic Host Configuration Protocol (DHCP)

Dynamic Host Configuration Protocol (DHCP)

General Information (DHCP)


The Dynamic Host Configuration Protocol (DHCP) is an auto configuration
protocol used on IP networks. The DHCP functionality is available on all Agilent
HPLC modules with on-board LAN Interface or LAN Interface Card G1369C, and
“B”-firmware (B.06.40 or above) or modules with "D"-firmware. All modules
should use latest firmware from the same set.
When the initialization mode “DHCP” is selected, the card tries to download the
parameters from a DHCP Server. The parameters obtained become the active
parameters immediately. They are not stored to the non-volatile memory of the
card.
Besides requesting the network parameters, the card also submits its hostname
to the DHCP Server. The hostname equals the MAC address of the card, e.g.
0030d3177321. It is the DHCP server's responsibility to forward the
hostname/address information to the Domain Name Server. The card does not
offer any services for hostname resolution (e.g. NetBIOS).

DHCP Active
Server Parameter

Figure 57 DHCP (Principle)

1 It may take some time until the DHCP server has updated the DNS server with
NOTE the hostname information.
2 It may be necessary to fully qualify the hostname with the DNS suffix, e.g.
0030d3177321.country.company.com.
3 The DHCP server may reject the hostname proposed by the card and assign a
name following local naming conventions.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 329
13 LAN Configuration
Dynamic Host Configuration Protocol (DHCP)

Setup (DHCP)

The DHCP functionality is available on all Agilent HPLC modules with on-board LAN Interface or LAN
Interface Card G1369C, and “B”-firmware (B.06.40 or above) or modules with "D"-firmware. All
modules should use latest firmware from the same set.

1 Note the MAC address of the LAN interface (provided with G1369C LAN
Interface Card or mainboard). This MAC address is on a label on the card or at
the rear of the mainboard, for example, 0030d3177321.
On the Local Controller the MAC address can be found under Details in the
LAN section.

Figure 58 LAN Setting on Instant Pilot

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 330
13 LAN Configuration
Dynamic Host Configuration Protocol (DHCP)

2 Set the configuration switch to DHCP either on the G1369C LAN Interface
Card or the mainboard of above mentioned modules.
Table 33 G1369C LAN Interface Card (configuration switch on the card)

SW 4 SW 5 SW 6 SW 7 SW 8 Initialization Mode

ON OFF OFF OFF OFF DHCP

Table 34 LC Modules with 8-bit configuration switch (B-firmware) (configuration switch at rear
of the instrument)

SW 6 SW 7 SW 8 Initialization Mode

ON OFF OFF DHCP

3 Turn on the module that hosts the LAN interface.


4 Configure your Control Software (e.g. OpenLAB CDS ChemStation Edition, Lab
Advisor, Firmware Update Tool) and use MAC address as host name, e.g.
0030d3177321.
The LC system should become visible in the control software (see Note in
section “General Information (DHCP)” on page 329).

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 331
13 LAN Configuration
Manual Configuration

Manual Configuration

Manual configuration only alters the set of parameters stored in the non-volatile
memory of the module. It never affects the currently active parameters.
Therefore, manual configuration can be done at any time. A power cycle is
mandatory to make the stored parameters become the active parameters, given
that the initialization mode selection switches are allowing it.

TELNET
Session
Non-Volatile
RAM

Stored
Parameter

Control
Module

Figure 59 Manual Configuration (Principle)

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 332
13 LAN Configuration
Manual Configuration

With Telnet
Whenever a TCP/IP connection to the module is possible (TCP/IP parameters set
by any method), the parameters may be altered by opening a Telnet session.
1 Open the system (DOS) prompt window by clicking on Windows START
button and select “Run...”. Type “cmd” and press OK.
2 Type the following at the system (DOS) prompt:
• c:\>telnet <IP address> or
• c:\>telnet <host name>

Figure 60 Telnet - Starting a session

where <IP address> may be the assigned address from a Bootp cycle, a
configuration session with the Handheld Controller, or the default IP address
(see “Configuration Switches” on page 326).
When the connection was established successfully, the module responds
with the following:

Figure 61 A connection to the module is made

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 333
13 LAN Configuration
Manual Configuration

3 Type
? and press enter to see the available commands.

Figure 62 Telnet Commands

Table 35 Telnet Commands

Value Description

? displays syntax and descriptions of commands

/ displays current LAN settings

ip <x.x.x.x> sets new ip address

sm <x.x.x.x> sets new subnet mask

gw <x.x.x.x> sets new default gateway

exit exits shell and saves all changes

4 To change a parameter follows the style:


• parameter value, for example:
ip 134.40.28.56
Then press [Enter], where parameter refers to the configuration parameter
you are defining, and value refers to the definitions you are assigning to that
parameter. Each parameter entry is followed by a carriage return.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 334
13 LAN Configuration
Manual Configuration

5 Use the “/” and press Enter to list the current settings.

information about the LAN interface


MAC address, initialization mode
Initialization mode is Using Stored
active TCP/IP settings

TCP/IP status - here ready


connected to PC with controller software (e.g. Agilent
Figure 63 Telnet - Current settings in "Using Stored" mode ChemStation), here not connected

6 Change the IP address (in this example 192.168.254.12) and type “/” to list
current settings.

change of IP setting to
Initialization mode is Using Stored

active TCP/IP settings

stored TCP/IP settings in non-volatile memory

connected to PC with controller software (e.g. Agilent


ChemStation), here not connected

Figure 64 Telnet - Change IP settings

7 When you have finished typing the configuration parameters, type


exit and press Enter to exit with storing parameters.

Figure 65 Closing the Telnet Session

If the Initialization Mode Switch is changed now to “Using Stored” mode, the
NOTE instrument will take the stored settings when the module is re-booted. In the
example above it would be 192.168.254.12.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 335
14 Appendix

General Safety Information 337


General Safety Information 337
Safety Standards 337
General 337
Before Applying Power 338
Ground the Instrument 338
Do Not Operate in an Explosive Atmosphere 339
Do Not Remove the Instrument Cover 339
Do Not Modify the Instrument 339
In Case of Damage 339
Solvents 340
Safety Symbols 341
Waste Electrical and Electronic Equipment (WEEE) Directive 343
Radio Interference 344
Sound Emission 345
Agilent Technologies on Internet 346

This chapter provides addition information on safety, legal, and web.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 336
14 Appendix
General Safety Information

General Safety Information

General Safety Information


The following general safety precautions must be observed during all phases of
operation, service, and repair of this instrument. Failure to comply with these
precautions or with specific warnings elsewhere in this manual violates safety
standards of design, manufacture, and intended use of the instrument. Agilent
Technologies assumes no liability for the customer’s failure to comply with these
requirements.

Ensure the proper usage of the equipment.


WAR N IN G
The protection provided by the equipment may be impaired.
 The operator of this instrument is advised to use the equipment in a
manner as specified in this manual.

Safety Standards
This is a Safety Class I instrument (provided with terminal for protective earthing)
and has been manufactured and tested according to international safety
standards.

General
Do not use this product in any manner not specified by the manufacturer. The
protective features of this product may be impaired if it is used in a manner not
specified in the operation instructions.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 337
14 Appendix
General Safety Information

Before Applying Power


Wrong voltage range, frequency or cabling
WAR N IN G
Personal injury or damage to the instrument
 Verify that the voltage range and frequency of your power distribution
matches to the power specification of the individual instrument.
 Never use cables other than the ones supplied by Agilent Technologies
to ensure proper functionality and compliance with safety or EMC
regulations.
 Make all connections to the unit before applying power.

Note the instrument's external markings described under “Safety Symbols” on


NOTE page 341.

Ground the Instrument


Missing electrical ground
WAR N IN G
Electrical shock
 If your product is provided with a grounding type power plug, the
instrument chassis and cover must be connected to an electrical ground
to minimize shock hazard.
 The ground pin must be firmly connected to an electrical ground (safety
ground) terminal at the power outlet. Any interruption of the protective
(grounding) conductor or disconnection of the protective earth terminal
will cause a potential shock hazard that could result in personal injury.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 338
14 Appendix
General Safety Information

Do Not Operate in an Explosive Atmosphere


Presence of flammable gases or fumes
WAR N IN G
Explosion hazard
 Do not operate the instrument in the presence of flammable gases or
fumes.

Do Not Remove the Instrument Cover


Instrument covers removed
WAR N IN G
Electrical shock
 Do Not Remove the Instrument Cover
 Only Agilent authorized personnel are allowed to remove instrument
covers. Always disconnect the power cables and any external circuits
before removing the instrument cover.

Do Not Modify the Instrument


Do not install substitute parts or perform any unauthorized modification to the
product. Return the product to an Agilent Sales and Service Office for service and
repair to ensure that safety features are maintained.

In Case of Damage
Damage to the module
WAR N IN G
Personal injury (for example electrical shock, intoxication)
 Instruments that appear damaged or defective should be made
inoperative and secured against unintended operation until they can be
repaired by qualified service personnel.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 339
14 Appendix
General Safety Information

Solvents
Toxic, flammable and hazardous solvents, samples and reagents
WAR N IN G
The handling of solvents, samples and reagents can hold health and safety
risks.
 When working with these substances observe appropriate safety
procedures (for example by wearing goggles, safety gloves and
protective clothing) as described in the material handling and safety
data sheet supplied by the vendor, and follow good laboratory practice.
 Do not use solvents with an auto-ignition temperature below 200 °C
(392 °F). Do not use solvents with a boiling point below 56 °C (133 °F).
 Avoid high vapor concentrations. Keep the solvent temperature at least
40 °C (72 °F) below the boiling point of the solvent used. This includes
the solvent temperature in the sample compartment. For the solvents
methanol and ethanol keep the solvent temperature at least 25 °C
(45 °F) below the boiling point.
 Do not operate the instrument in an explosive atmosphere.
 Do not use solvents of ignition Class IIC according IEC 60079-20-1 (for
example, carbon disulfide).
 Reduce the volume of substances to the minimum required for the
analysis.
 Never exceed the maximum permissible volume of solvents (8 L) in the
solvent cabinet. Do not use bottles that exceed the maximum
permissible volume as specified in the usage guideline for solvent
cabinet.
 Ground the waste container.
 Regularly check the filling level of the waste container. The residual free
volume in the waste container must be large enough to collect the waste
liquid.
 To achieve maximal safety, regularly check the tubing for correct
installation.

For details, see the usage guideline for the solvent cabinet. A printed copy of the
NOTE guideline has been shipped with the solvent cabinet, electronic copies are
available in the Agilent Information Center or via the Internet.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 340
14 Appendix
General Safety Information

Safety Symbols
Table 36 Symbols

The apparatus is marked with this symbol when the user shall refer to the
instruction manual in order to protect risk of harm to the operator and to
protect the apparatus against damage.

Indicates dangerous voltages.

Indicates a protected ground terminal.

The apparatus is marked with this symbol when hot surfaces are available
and the user should not touch it when heated up.

Sample Cooler unit is designed as vapor-compression refrigeration system.


Contains fluorinated greenhouse gas (refrigerant) according to the Kyoto
protocol.
For specifications of refrigerant, charge capacity, carbon dioxide equivalent
(CDE), and global warming potential (GWP) see instrument label.

Flammable Material
For Sample Thermostat which uses flammable refrigerant consult Agilent
Information Center / User Manual before attempting to install or service this
equipment. All safety precautions must be followed.

Confirms that a manufactured product complies with all applicable


European Community directives. The European Declaration of Conformity is
available at:
http://regulations.corporate.agilent.com/DoC/search.htm

Manufacturing date.

Power symbol indicates On/Off.


The apparatus is not completely disconnected from the mains supply when
the power switch is in the Off position

Pacemaker
Magnets could affect the functioning of pacemakers and implanted heart
defibrillators.
A pacemaker could switch into test mode and cause illness. A heart
defibrillator may stop working. If you wear these devices keep at least
55 mm distance to magnets. Warn others who wear these devices from
getting too close to magnets.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 341
14 Appendix
General Safety Information

Table 36 Symbols

Magnetic field
Magnets produce a far-reaching, strong magnetic field. They could damage
TVs and laptops, computer hard drives, credit and ATM cards, data storage
media, mechanical watches, hearing aids and speakers. Keep magnets at
least 25 mm away from devices and objects that could be damaged by
strong magnetic fields.

Indicates a pinching or crushing hazard

Indicates a piercing or cutting hazard.

A WARNING
WAR N IN G
alerts you to situations that could cause physical injury or death.
 Do not proceed beyond a warning until you have fully understood and
met the indicated conditions.

A CAUTION
C AU T I O N
alerts you to situations that could cause loss of data, or damage of equipment.
 Do not proceed beyond a caution until you have fully understood and met
the indicated conditions.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 342
14 Appendix
Waste Electrical and Electronic Equipment (WEEE) Directive

Waste Electrical and Electronic Equipment


(WEEE) Directive

This product complies with the European WEEE Directive marking requirements.
The affixed label indicates that you must not discard this electrical/electronic
product in domestic household waste.

Do not dispose of in domestic household waste


NOTE
To return unwanted products, contact your local Agilent office, or see
https://www.agilent.com for more information.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 343
14 Appendix
Radio Interference

Radio Interference

Cables supplied by Agilent Technologies are screened to provide optimized


protection against radio interference. All cables are in compliance with safety or
EMC regulations.

Test and Measurement


If test and measurement equipment is operated with unscreened cables, or used
for measurements on open set-ups, the user has to assure that under operating
conditions the radio interference limits are still met within the premises.

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 344
14 Appendix
Sound Emission

Sound Emission

Sound pressure
Sound pressure Lp <70 db(A) according to DIN EN ISO 7779

Schalldruckpegel
Schalldruckpegel Lp <70 db(A) nach DIN EN ISO 7779

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 345
14 Appendix
Agilent Technologies on Internet

Agilent Technologies on Internet

For the latest information on products and services visit our worldwide web site
on the Internet at:
https://www.agilent.com

InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps Service Manual 346
In This Book

This manual contains technical information


about the Agilent 1260 Infinity II Isocratic
Pump (G7110B), the Agilent 1260 Infinity II
Quaternary Pump VL (G7111A), the Agilent
1260 Infinity II Quaternary Pump (G7111B),
and the Agilent 1260 Infinity II Bio-inert Pump
(G5654A).
The manual describes the following:
• introduction,
• site requirements and specifications,
• installing the pump,
• using the pump,
• optimizing performance,
• troubleshooting and diagnostics,
• maintenance and repair,
• parts and materials for maintenance and
repair,
• identifying cables,
• hardware information,
• appendix.

www.agilent.com
Agilent Technologies Inc. 2016-2023
Edition: 07/2023
Document No: SD-29000179 Rev. D

You might also like