Nanodyne Replacement Illuminator for Leica M655 / M400E Surgical Microscope - Installation: Included Items.
Additional Items Included But Not Shown:
PN 10337 Hex Key 2mm short arm (to attach
Nanodyne illuminator)
PN 11709 Hex Key 1.5mm long arm (for M2
latch screws)
B B
PN 11115 SHCS M2 x 0.4 x 4mm (2 pcs) (for
attaching spring latch to new illuminator base)
PN 11661 Leica M655 illuminator Assy
The illuminator
may be powered
by plugging the
cable into the
power supply
provided, or into a
suitable USB port
on a computer or
other device.
PN 11163 Power Supply 5V 2.1A and PN
10734 Power Cable 1.35mm ID/3.5mm
OD x USB A 6 ft.
A A
Power supplies
PN 10362 M3 x 0.5 x 6mm are subject to
Button Head Cap Screw substitution PN 11000 Power Cord Extension 1.35mm
(4 pcs) without notice ID x 3.5mm OD RA plug to jack 6 foot.
due to availability (use if needed for more length)
issues and Wayne Bonin
changes in 3/19/2020
regulations. PN 11708 Leica M655 / M400E REV
illuminator Installation 1
Copyright 2020 Nanodyne Measurement Systems. Document authorized for installation of Nanodyne equipment only. www.nano-dyne.com Wayne Bonin 651-323-8592 SHEET 1 OF 4
4 3 2 1
4 3 2 1
Nanodyne Replacement Illuminator for Leica M655 / M400E Surgical Microscope - Installation: Step 1. Remove OEM illuminator.
B B
(1) Original microscope, view 1. (2) Original microscope, view 2. (3) Lamp power cable. (4) OEM lamp house from back side. (5) Back cover lifted, lamp holder visible.
1) Remove the AC power cord. It will not be
used. (not shown here)
2) Disconnect the lamp power cable at the
power supply. (photo 3)
3) Disconnect the lamp power cord at the lamp.
(photo 6) Either disassemble the unit as required
to loosen the connections at the terminal block
(red circle), or cut the wires as they are no longer
needed.
4) Remove the lamp holder by removing the 4
A screws holding the lamp assembly to the bottom A
of the housing. (arrow, photo 6) Use a fillips
screwdriver.
Wayne Bonin 3/19/2020
PN 11708 Leica M655 / M400E REV
(6) Power cord, red circle. Mounting screws, (7) OEM lamp holder removed from lamp house. (8) Close up of latch (to be removed). Circled in red, illuminator Installation 1
1 of 4, red arrow. photo at left.
SHEET 2 OF 4
4 3 2 1
Copyright 2020 Nanodyne Measurement Systems. Document authorized for installation of Nanodyne equipment only. www.nano-dyne.com Wayne Bonin 651-323-8592
4 3 2 1
Nanodyne Replacement Illuminator for Leica M655 / M400E Surgical Microscope - Installation: Step 2. Transfer Latch to Nanodyne illuminator.
B B
(1) Nanodyne illuminator assembly as received. (2) Installing OEM latch on illuminator assembly. (3) Close up of OEM latch with new socket cap screws.
1) Remove the latch from the original illuminator (see photos 7, 8 2) Install the latch on the Nanodyne illuminator (photos 2 and 3, this
on sheet 2). page).
A A
Unscrew the 2 screws using a small flat blade screwdriver (not Use the brass washers from the old unit, with the two M2 cap
supplied). screws supplied. Use the 1.5mm hex key to tighten the screws.
Save the 2 brass washers. The slots in the latch are a lot bigger than the screws. Try installing
it with the latch slots centered around the screws. then adjust if
We supply 2 replacement screws that are much easier to install. needed to properly engage the cover (photo 4, sheet 2).
Wayne Bonin 3/19/2020
You don't need to use the original screws unless you want to prove
you have the hands of a brain surgeon. The long arm on the hex key is only provided to clear the heatsinks PN 11708 Leica M655 / M400E REV
on the illuminator. Do not use the long arm to over-tighten the illuminator Installation 1
screws or you could strip the threads in the baseplate.
SHEET 3 OF 4
4 3 2 1
Copyright 2020 Nanodyne Measurement Systems. Document authorized for installation of Nanodyne equipment only. www.nano-dyne.com Wayne Bonin 651-323-8592
4 3 2 1
Nanodyne Replacement Illuminator for Leica M655 / M400E Surgical Microscope - Installation: Step 3. Attach illuminator and Connect Power Supply.
B B
PN 10362 M3 x 0.5 x 6mm
Button Head Cap Screw
(4 pcs)
2) After installing the illuminator, connect the power
supply to an AC outlet, connect the USB end of the power
cord to the power supply and the other end to the
A 1) Attach the illuminator to the bottom of the lamp house A
illuminator as shown above. Use the extension cord if 6
just like the original was. Use the 4 screws provided, and
feet is not enough.
the 2mm hex key.
All intensity control is by the knob shown above. It has a
(no photo - we did not have the microscope).
tactile click feel in the off position. There is no other switch.
Wayne Bonin 3/19/2020
PN 11708 Leica M655 / M400E REV
illuminator Installation 1