Aa300 Manual 02 19
Aa300 Manual 02 19
1
Table of Contents
1 Safety 3
4 Components 18
5 Installation 20
6 Warranty 25
6.1 Limitation of warranty and liability 26
2
1. SAFETY
3
1.1 Safety Practices
This document describes the general safety practices and precautions that must be
observed when operating the Airflow Monitor.
This advice is intended to supplement, not supersede, the normal safety codes in the user’s
country. The information provided does not cover every safety procedure that should be
followed. Ultimately, maintenance of a safe laboratory environment is the responsibility of
the user and the user’s organisation.
Please consult all documentation supplied with the Airflow Monitor before starting to work.
Carefully read the safety information in this document and in the other documentation
supplied. When setting up the equipment or performing analysis or maintenance
procedures, strictly follow the instructions provided.
Within this User Guide WARNINGS are used to highlight information or instructions that
must be followed in order to avoid personal injury to yourself or other people in the vicinity,
eg. switch off the mains voltage before any maintenance.
WARNINGS appear as below:
Switch off the mains voltage and remove the power supply
before maintenance.
WARNING
1.3 Precautions
The following precautions must be observed when using the Airflow Monitor and associated
systems:
• Be sure that the voltage of the Airflow Monitor equipment corresponds to the voltage
available where it is to be installed.
• Never remove the side or back panels of the Airflow Monitor without first shutting
down the equipment and disconnecting the power supply.
4
1.4 General Operating Conditions
The Airflow Monitor and equipment have been designed and tested in accordance with the
safety requirements of the International Electrotechnical Commission (IEC). The Airflow
Monitor conforms to IEC61010-1 (Safety Requirements for electrical equipment for
measurement, control and laboratory use) as it applies to IEC Class 1 (earthed) appliances,
and therefore meets the requirements of EC directive 73/23/EEC.
If possible, avoid any adjustment, maintenance or repair to the eqipment whilst covers are
open or it is operative. However, if any adjustment, maintenance or repair is necessary while
the covers are open, this must be done by a skilled person who is aware of the hazards
involved.
Whenever circumstances arise that mean an Airflow Monitor may be unsafe, make it
inoperative. In particular, an Airflow Monitor may be unsafe if it:
The Airflow Monitor should only be used under the following conditions:
• Indoors.
• In ambient temperatures between 5°C and 40°C.
• With relative humidity below 80% for temperatures up to 31°C, decreasing linearly to
50% relative humidity at 40°C.
5
1.6 Electrical Safety
The Airflow Monitor and associated equipment are designed to protect the user from
potential electrical hazards. This section describes some recommended electrical safety
practices.
WARNING Even when the power switch is off, high voltages can still be present
– capacitors within the equipment may still be charged even if the
equipment has been disconnected from all live voltage sources.
The Airflow Monitor and associated equipment must be correctly connected to a suitable
electrical supply. The supply must have a correctly installed protective conductor (earth or
ground) and must be installed and checked by a qualified electrician before initial power up .
If it is possible that the equipment is no longer electrically safe for use, make the equipment
inoperative and secure it against any unauthorised or unintentional operation.
6
The electrical safety of the equipment is likely to be impaired if:
• It has any signs of visible damage.
• If it has been subjected to prolonged storage in unfavourable conditions.
• If it has been subjected to severe stress during transportation.
EC Directive
The Airflow Monitor System is designed and tested to meet the requirements of the EC
directive 89/336/EEC and 93/68/EEC and complies with the EMC standard EN61326 (EMC
standard for electrical equipment for measurement, control and laboratory use) and
EN55011 (ISM) Class A (RF emissions).
FCC Rules and Regulations
The Airflow Monitor System is classified as a digital device used exclusively as industrial,
commercial or medical test equipment. It is exempt from the technical standards specified in
Part 15 of the FCC Rules and Regulations based on Section 15.103 (c).
Warning labels attached to the equipment draw attention to specific hazards - refer to this
guide and other documentation provided with the equipment for more details concerning
potential hazards and any precautions or other actions that must be taken.
7
2. Overview of the Airflow Monitor
8
2.1 Operator Display Panel
LED Indicators
9
2.2 General Description
If the Sash Alarm System option is ordered there will also be a sash micro switch or proximity switch.
If the Remote Airflow Sensor is ordered there will be remote sensor provided
Operator Features --- the alarm has the following operator features :-
Safe LED - Green LED (Not flashing) will be displayed if the airflow is greater than the
Low air alarm point.
Alarm LED - Red LED (Not flashing) will be displayed if the airflow is lower than the
Low air alarm point.
Visor Open - Red LED (Flashing) will be displayed when the Visor is open.
Audible Alarm -- the Audible alarm will sound ( can be muted ) in the Air Fail and the
Sash High alarm condition
Night Set-back -- when the Night Setback input is activated the Audible alarm will be muted
and the Green LED will flash on/off
ENTER --- the alarm has an Enter button -- this is multi-functional as follows :-
Press Enter momentarily when Low Air alarm is sounding will mute the alarm
Press Enter momentarily when Visor Open alarm is sounding will mute
the alarm and initiate a repeat timer that will re-sound the alarm if the Visor is
not closed before the end of the time period
Note - The Audible Alarm can be permanently disabled or re-enabled by pressing and holding the
SET button for approximately 10 seconds, the monitor will give 3 rapid beeps to confirm the alarm has
been disabled or re-enabled. When the audible alarm is disabled the Green LED will flash when the
Airflow is safe to indicate that the alarm is permanently muted.
10
2.3 External Connections
Input 1 --- volt free relay input – ( close contact to activate the input )
This input is configured as :-
NIGHT SETBACK
Visor Open Input --- a. Connection point for Visor micro switch. ( Switch contact to
close and remain closed in Visor Open condition )
11
3. Functions & Operation
12
3.1 Alarm Configuration / Calibration
The alarm has two modes for the calibration of the airflow alarm point. These two modes are selectable via a
‘jumper connection’ on the back of the alarm PCB.
a. With Jumper J2 connected - The airflow is reduced to the alarm point mechanically and this airflow is
‘captured’ by the alarm. The airflow is then restored to the normal operating value and the Low Air alarm will
activate if the airflow subsequently falls to the alarm point
b. With JumperJ2 not connected - This is a two point calibration method. The airflow is set to the normal
operating value and this value is ‘captured’ by the alarm and taken to be 100%. The airflow is then increased
to one and a half time the normal operating value and this value is ‘captured’ by the alarm and taken to be
150%. The Low Air alarm will then activate if the airflow subsequently falls to 80% of the normal operating
value.
See ‘Quick Start Installation’ below for details of the Calibration procedures
3.2 Start Up
1. The 12V DC power is applied to the airflow sensor and the alarm then performs a self-test on the functions ,
LEDs and audible alarm ( approx 2 sec ) and then initiates a delay timer of 30 secs to allow the airflow sensor
to stabilize.
2. During the whole of the 30 sec period all alarms and relay outputs are inhibited and the Red & Green LEDs
will be permanently ON.
3. At the end of the 30 sec delay the unit performs one of two options :-
a. If the alarm calibration has been previously completed – the unit goes to
normal operating mode ( Run )
b. If the unit has not been calibrated -- the Red & Green LEDs will flash on/off and
the audible alarm will be muted. It is then possible to press the Enter button for
5 secs and go into the calibration mode – (See ‘Quick Start Installation’ below for
details of the Calibration procedures
13
3.3 Events / Actions
Safe airflow
Low airflow
• Airflow below alarm level for longer than the low air delay time ( 5 secs)
• Red LED on ( Not flashing )
• Audible alarm sounds ( ‘Beep’ on/off every 1 sec ) -- can be muted via Enter pushbutton
• Low air relay R1 operates.
Reset : -- when airflow rises above Low air level for longer than the low air to
safe air delay time ( 2 secs ) the Low air alarm resets automatically
Visor Open
• When the input configured as Visor Open is activated ( Micro switch or Proximity switch )
• Red LED on ( Flashing )
• Audible alarm sounds ( ‘Beep’ on/off every 1 sec )
• Audible can be muted via Enter pushbutton -- this silences the alarm and initiates a repeat timer ( factory set
to 5 mins ) After the delay time the alarm re-sounds ( and can be re-muted). During this time the Red LED
flashes on / off.
Night set-back
• The connection and each element of the airflow sensor are monitored at all times. In the event of a problem
with the sensor the audible alarm will sound using a different
and distinctive tone best described as ‘ a modulating siren effect’
The audible alarm can only be silenced by re-connecting the sensor ( if a remote sensor is being used ) or switching
the power to the unit off
14
3.4 Quick Start Installation
Follow the instructions below for installing and commissioning the unit. :-
Before connecting the unit, it is important to decide on the type of calibration that is to be used on the installation.
The two methods are described in section 1.2 Alarm Configuration / Calibration above.
1. Fit the alarm to the Fume Hood using the cut-out details provided with the unit
--- see page 11
2. Plug in the power adapter to a Mains AC power socket and connect the flying lead to the alarm unit --- see
typical connection diagram on page 13 & 14
3 Power up the unit and wait at least 30 secs while the sensor temperature stabilizes.
During this time the Red & Green LEDs will both be on ( not flashing )
If the unit has not been previously calibrated the Red & Green LEDs will begin to flash
on/off at the end of the 30 sec start up time delay but the audible alarm will not sound.
If the alarm has been calibrated it will go into normal operation.
5. Calibration :-
b. Press and hold the Enter button for 5 secs to go into the calibration mode. This is indicated by both Red
and Green LEDs flashing on/off together with the audible alarm sounding ( ‘Beep on/off 4 times every 1
sec ).
c. To initiate the alarm point calibration press and hold the ENTER and the SET buttons at the same time.
The unit will then sample the airflow for a 5 sec period during which time the GREEN LED goes off and
the RED LED flashes on/off. The audible alarm continues to sound during this period and if the sampling
is successful will give a two tone beep at the end of the period and the unit will then go automatically into
the RUN mode.
If the buttons are released during the sampling period or if the airflow is fluctuating more than the pre-set
value the audible alarm will give a lower frequency buzzing sound for a short period and then go back
into the calibration mode. If this occurs re- press the ENTER and SET buttons to repeat the airflow
sampling.
d. When complete re-set the airflow to the normal value and the unit will go to the Safe
running condition with the GREEN LED on.
The unit will now function and go into the alarm condition if the Fume Hood air velocity falls below the alarm
value.
15
Calibration continued:-
b. Press and hold the Enter button for 5 secs to go into the Calibration mode. This is
indicated by both Red and Green LEDS flashing on/off together with the audible
alarm sounding ( ‘Beep’ on/off 4 times every 1 sec ).
c. To initiate the normal airflow (100%) calibration press and hold the ENTER and the
SET button at the same time. The unit will then sample the normal airflow for a
5 sec period during which time the Green LED goes off and the Red LED flashes
on/off. The audible alarm continues to sound during this period and if the sampling
is successful will give a two tone beep at the end of the period and the unit will
then go automatically into the higher airflow calibration mode.
If the buttons are released during the sampling period or if the airflow is fluctuating
more than the pre-set value the audible alarm will give a lower frequency buzzing
sound for a short period and then go back into the calibration mode. If this occurs
re-press the ENTER and SET buttons to repeat the airflow sampling.
d. The higher airflow calibration mode is indicated by both Red and Green LEDS
flashing on/off together with the audible alarm sounding ( ‘Beep’ on/off for 2 secs
every 10 secs ).
Adjust the fan speed until the face velocity rises to one and a half times the
normal operating value used in a. above using a calibrated instrument to check
the value
To initiate the higher airflow (150%) calibration press and hold the ENTER and the
SET button at the same time. The unit will then sample the higher airflow for a
5 sec period during which time the Green LED goes off and the Red LED flashes
on/off. The audible alarm continues to sound during this period and if the sampling
is successful will give a two tone beep at the end of the period and the unit will
then go automatically into the Run mode.
If the buttons are released during the sampling period or if the airflow is fluctuating
more than the pre-set value the audible alarm will give a lower frequency buzzing
sound for a short period and then go back into the calibration mode. If this occurs
re-press the ENTER and SET buttons to repeat the airflow sampling.
The unit will now function and go into the ALARM condition if the Fume Hood air velocity falls below 80% of
the normal operating value.
16
3.5 Calibration Notes
The two calibration methods are intended to give the installer two options for calibrating the alarm.
Using the alarm point capture method it is necessary to adjust the air velocity on the Fume Hood using a
mechanical damper ( or fan speed controller ) to the desired alarm point and this is sampled by the unit.
It is necessary to re-adjust the air velocity back to the normal operating value. This method produces a
very accurate alarm point at a fixed value. This method of calibration is ideal when the fan speed can be
adjusted to the low air alarm value.
The second method of two point capture is slightly more involved but does not require the fan speed to
be adjusted to the alarm point level, however the fan speed will need to be adjusted to give a higher
value sample.
For fan fail alarm point calibration single point calibration is ideal. It is recommended that the fan is set to
give an alarm velocity of 0.3 m/sec ( 60 fpm) minimum and 0.5 m/sec (100 fpm) maximum. This
considers a normal operating velocity of approximately 0.75 m/sec (150 fpm).
Using the two point capture method is ideal for low air alarm calibration where the alarm point is required
to indicate a low operating velocity – typically 80% of the normal operating velocity. The normal velocity
calibration point should be the normal operating velocity and the high velocity calibration point should be
150% of the normal operating velocity. For example - a normal operating velocity of 150 fpm would give
a low calibration point of 150 fpm and a high calibration point of 225 fpm.
This would give a low air alarm point of 120 fpm.
Take time when measuring the air velocities for the calibration procedure to allow for the velocities across the
airflow sensor to stabilize. If the velocities are changing or are turbulent during the sampling period the alarm will
detect this and ask you to repeat the sample.
17
4. Components
18
4.1 Airflow Monitor components
--
--
--
TEL
-
19
5. Installation
20
5.1 Installation of standard components
The following section outlines the installation of the various components of the Airflow monitor.
As the size and format of individual Fume Hoods varies considerably, specific instructions are not
possible, though the principles outlined below should remain valid in all cases.
The airflow monitor can be mounted on either side of the Fume Hood
1. Fit the AA300 monitor to the Fume Hood using the cut-out details provided with the unit. Ensure
that the monitor fits into the cut out before securing with the fixing screws.
.
2. Fit the airflow sensor hose to the Fume Hood using the cut out and installation details provided.
21
5.2 Airflow Monitor Installation.
22
5.3 Airflow Sensor installation.
AA300
AFA500
Mounted on front
Mounted on
front
of side of
post
side post
Adaptor through
Inner side wall
Adapter through
Inner side wall
Front of of
Front
sideside
postpost
INSTALLATION NOTES :-
The back connection spigot of the sensor is designed to accept a 1” OD tube which should be connected to
the inner chamber of the fume hood. (This tube and fittings is known as the " vent kit " )
The ideal position for the end of the 1” tube for most fume hoods is 4” back from the sash glass
and 4” higher than the normal sash opening height through the inner side
23
5.4 Typical connection diagram
--
--
--
TEL
-
24
6. Warranty
25
6.1 LIMITATION OF WARRANTY AND LIABILITY
Seller warrants that this product, under normal use and service as described in the operator’s manual shall be free from defects in
workmanship and material for a period of twelve (12) months, or the length of time specified in the operator’s manual, from the date of
shipment to the customer. This limited warranty is subject to the following exclusion: -
a. Batteries and certain other components when indicated in specifications are warranted for a period of 90 days
from the date of shipment to the customer.
b. With respect to any repair services rendered, Seller warrants that the parts repaired or replaced will be free
from defects in workmanship and material, under normal use, for a period of 90 days from the date of shipment
to the customer
c. Seller does not provide any warranty on finished goods manufactured by others. Only the original
manufacturer’s warranty applies.
d. Unless specifically authorized in a separate writing by Seller, Seller makes no warranty with respect to, and
shall have no liability in connection with, any goods which are incorporated into other products or equipment by
the Buyer. All goods returned under warranty shall be at the Buyer’s risk of loss, Seller’s factory prepaid, and
will be returned at Seller’s risk of loss, Buyer’s factory prepaid.
The foregoing is IN LIEU OF all other warranties and is subject to the conditions and LIMITATIONS stated herein. NO OTHER
EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE.
THE EXCLUSIVE REMEDY OF THE USER OR PURCHASER, AND THE LIMIT OF LIABILITY OF SELLER FOR ANY AND
ALL LOSSES, INJURIES, OR DAMAGES IN CONNECTION WITH THIS PRODUCT (INCLUDING CLAIMS BASED ON
CONTRACT NEGLIGENCE, STRICT LIABILITY, OTHER TORT, OR OTHERWISE) SHALL BE THE RETURN OF THE
PRODUCT TO THE FACTORY OR DESIGNATED LOCATION AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE
OPTION OF THE SELLER, THE REPAIR OR REPLACEMENT OF THE PRODUCT. IN NO EVENT SHALL SELLER BE
LIABLE FOR ANY SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES.SELLER SHALL NOT BE RESPONSIBLE FOR
INSTALLATION, DISMANTLING, REASSEMBLY OR REINSTALLATION COSTS OR CHARGES. NO ACTION,
REGARDLESS OF FORM, MAY BE BROUGHT AGAINST THE SELLER MORE THAN ONE YEAR AFTER THE CAUSE OF
ACTION HAS ACCRUED.
The purchaser and all users are deemed to have accepted the terms of this LIMITATION OF WARRANTY AND LIABILITY,
which contains the complete and exclusive limited warranty of Seller. This LIMITATION OF WARRANTY AND LIABILITY may
not be amended or modified nor may any of its terms be waived except by a writing signed by an authorized representative of
the Seller.
26