Brochure Product Gas Analysis Thermal Conductivity Gas Analyzer ZAF4 21C1 E 0031 en

The ZAF-4 Thermal Conductivity Gas Analyzer is a compact, flame-proof device designed for measuring gas concentrations such as H2, CO2, and CH4, utilizing thermal conductivity principles. It features an easy-to-read LCD, automatic calibration options, and PC communication capabilities via RS232C (MODBUS). The analyzer is suitable for various applications, including semiconductor equipment and hydrogen generation, with a range of customizable outputs and alarm functions.

Uploaded by

r.alikarami
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
53 views6 pages

Brochure Product Gas Analysis Thermal Conductivity Gas Analyzer ZAF4 21C1 E 0031 en

The ZAF-4 Thermal Conductivity Gas Analyzer is a compact, flame-proof device designed for measuring gas concentrations such as H2, CO2, and CH4, utilizing thermal conductivity principles. It features an easy-to-read LCD, automatic calibration options, and PC communication capabilities via RS232C (MODBUS). The analyzer is suitable for various applications, including semiconductor equipment and hydrogen generation, with a range of customizable outputs and alarm functions.

Uploaded by

r.alikarami
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 6

THERMAL CONDUCTIVITY

GAS ANALYZER
(Type: ZAF-4)

Flame-proof type
(NESPI Exd C T6Gb)

Standard type

21C1-E-0031
Easy-to-read, large LCD.

Operation is facilitated with the aid of guidance in English.

PC communication available. Compact size mountable on panel.


(Unit mm)
Mountable in same
Personal Computer panel cutout as
heretofore.

240
RS232C (MODBUS) for
measurement value,
setting range read/write,
device status output. 169
192 Mass 5kg.

Automatically calibrates zero/span. (option) Computes and corrects influence by other gases. (option)
Type ZAF main frame
Sample Gas aspirator Flow checker Membrane filter
gas Corrected output signal
H2 measurement
Exhaust
Mist filter 0.4L/min
Peltier gas cooler 4 to 20mA DC, etc.
Span calibration signal

Zero calibration signal


Drain pot CO2 measurement signal
Solenoid valve
(1 to 5V DC)
CO2 measurement
Pressure regurator

Span gas Zero gas

Gas concentration alarm output. (option) Two range. (option)


A range selected by key on front panel or by external contact closure
Alarm output (1a relay contact) command.
selectable for: Range ratio Up to 1:10 for H2 or He measurement,
Up to 1:5 for Ar, CH4 or CO2 measurement.
High limit
Low limit
High-high limit Linear measurement output signal.
Low-low limit
Dispenses with external linearizer.

Main applications Thermal Conductivity Ratio of Gases Measuring Principle Sample gas

This thermal conductivity gas analyzer measures gas


H2 concentration measurement for Gases
Comparative thermal conductivity (0℃)
when replacing thermal conductivity of air concentration by utilizing the different thermal
semiconductor equipment. (2.41 x 10-2 w/(m.k) with 1 conductivities of 2 gas components. In the detector,
H2 concentration measurement for Sulfur dioxide gas SO2 there are reference and measuring chambers in each of
which a thin platinum wire is stretched. The reference
Measuring chamber

CO2
Reference chamber

hydrogen generator. Carbon dioxide gas chamber is filled with reference gas and through the
H2 concentration measurement for Argon Ar measuring chamber, sample gas is flowed. Each
kiln. Carbon monoxide CO platinum wire composes a bridge circuit in combination
with an external fixed resistor, and it is heated by
Steam (100˚C) H 2O
Ar, He or CH4 concentration measurement flowing a constant current. When there is a change in
Air the concentration of the component under
for gas generation plant.
Nitrogen N2 measurement, the thermal conductivity of sample gas
He concentration measurement for O2 will change to affect the temperature of the platinum
Oxygen
super-conducting equipment. Methane CH4
wire in the measuring chamber. The resulting thermal
change is taken out as a change in electric resistance,
Ar concentration measurement for air Hydrogen H2 according to which the concentration of measured gas
separation plant. is calculated.
Specifications Standard type
Standard Specifications Optional Specifications
Measuring principle Measurement of thermal conductivity 5 SPST relay contact outputs
Measurable component He,Ar,H2,CH4,CO2 Relay contact capacity; 220V AC/2A (resistive load)
Measurable range As specified for particular type. Isolated with relay between contacts, and between contacts and internal circuit.
Max. 5 functions are selectable among those listed below.
4 to 20mA DC, 0 to 1V DC, 0 to 10mV DC
Output signal Non-isolated output(Any one-output signal specifiable in CODE SYMBOLS)
Zero-side solenoid valve drive output for automatic calibration
Span-side solenoid valve drive output for automatic calibration
Allowable load resistance 550 max. (in 4 to 20mA DC output) Suction pump OFF output in automatic calibration
Output resistance 100k (in 0 to 1V DC or 0 to 10mV DC output) Upper limit (1point) concentration alarm output
Display unit LCD with backlight Lower limit (1point) concentration alarm output
Relay contact output Upper/lower limit (1point) concentration alarm output
Display of measured value Max. 4 digits
Upper limit (1point) and lower limit (1point) concentration alarm output
Display language English (Total 2 points)
Output signal holding In both manual and automatic calibrations, output value just before calibration can be held. 2-step upper limit (1point at each step) concentration alarm output
Power supply 100 to 240V AC, 50/60Hz, Approx. 50VA (Total 2 points)
Warm-up time At least 30min 2-step lower limit (1point at each step) concentration alarm output
(Total 2 points)
Ambient temperature 5 to 45 Analyzer error or automatic calibration error alarm output
Ambient humidity Less than 90% RH (condensation unallowable) Calibrating status output
Storage conditions 2060, less than 95% RH (condensation unallowable) Range identification output (for 2 range type only)
Mounting Flush mounting on panel
3 non-voltage contact inputs
External dimensions(HWD) 240192213mm ON; 0 V, OFF; 5V DC, current at ON; 5mA
Mass Approx. 5kg Isolated with photo coupler between inputs and internal circuit.
Finish color Off-white (equivalent to 10Y7.5/0.5) Contact input
Not isolated between contact inputs.
Housing Steel-plate case, indoor type The following actions can be input.
Remote holding of measured value output
Material of gas- JIS SUS304, platinum, platinum iridium, silver,fluororubber, epoxy resin, Remote range changeover (only with 2-range meter)
contacting parts nickel, tin Remote start of automatic calibration
Gas inlet/outlet, purge port Rc1/4 or NPT1/4 (whichever specified)
Analog input for H2 meter interference correction (1 to 5V DC)
Purge gas flow rate Approx. 1L/min (as required) Either CO2 or CH4 component of an external gas analyzer is to be input.
Interference gas
Applied standard CE mark (Option) measured value input Adjustment is required at Fuji Electric's factory.
Details of measurement gas will be checked when receiving an order.
Performance Zero and span calibrations are automatically carried out at the predetermined
Repeatability 1% of F.S. Automatic calibration intervals.
function Calibration gases are flowed sequentially by driving the externally installed
Zero pointWithin 2% of full scale/week (H 2 meter, reference gas N2)
solenoid valves.
Drift SpanWithin 2% of full scale/week (H 2 meter, reference gas N2)
RS-232C (9-pin D-sub output)
Standard within 60sec (at flow rate 0.4L/min) Half duplex, asynchronous
Response speed High speed within 10sec (at flow rate 1L/min), allowed only for H2 meter
(90% response) Communicating function MODBUSTM protocol, communication speed 9600 bps
(reference gas N2) Contents of communication: Reading/writing of measured concentration
values and various set values, and output of device status

Standard Gas Measurement Conditions


Temperature 0 to 50
Installation Conditions
The analyzer should not be exposed to direct sunlight or radiation from a hot object.
Gas flow rate Constant at 0.40.05 L/min
A place subjected to heavy vibrations should be avoided. A location with clean
Dust Less than 100 g/Nm3 with a particle size of 0.3 m max.
atmosphere should be selected.
Pressure 10kPa max. Before measuring combustible gases, the existing gases should be purged from
Mist,Corrosive gas Unallowable the analyzer using air or N2.
Moisture Below saturation at 2 When the analyzer is installed outdoors, it should be sheltered with a housing or
Standard gases
Zero gas: same as reference gas cover to protect it from rain and wind.
Span gas: Concentration within 90 to 100% of measuring range
for calibration Concentration beyond 100% is inapplicable.

Outline Diagram (Standard type) (Unit: mm)


193
169 24 External terminal(option)
Panel 140
192 M3.5 screw (43) Rc1/4 or NPT1/4 (whichever specified)
1.6 t 10 t
INLET
(35)

GAS ANALYZER
1 11
2 12 INLET

3 13
4 14
5 15
6 16
130

7 17
240

(248)
(268)

8 18
240
230
230

9 19
1020
OUTLET
MODE ESC ZERO

ENT SPAN

1 2 3 4 5 PURGE Rc1/4 or NPT1/4 (whichever specified)


OUTLET

Rc1/4 or NPT1/4 (whichever specified)


182 Purging port
192
RS232C/Dsub 9p
External terminal
Front view Side view Rear view (whichever specified)
(standard)

Panel Cutout (Unit: mm) Connection Diagram


L Power supply 1 11 DO1
1 Contact
100 to 240V AC,
184
+2
0 2 N 50/60Hz 2 12 DO1 output 1 RS232C communication (option)
3 E Grounding terminal AIN 3 13 DO2 Pin Terminal Signal
Interference correcting Contact
number
4 input 1 to 5V DC AIN 4 14 DO2 output 2
Measured value 1 5
output (as specified) 2 Recive Data
5 Remote output DI1 5 15 DO3 Contact
4 to 20mA DC holding output 3
(M3.5screw) DI1 6 16 DO3 3 Transmit Data
+2
232 0

0 to 1V DC
Remote range DI2 7 17 DO4 Contact
0 to 10mV DC changeover 6 9 5 Signal GND
DI2 8 18 DO4 output 4
D-Sub9P
Remote automatic DI3 9 19 DO5 Contact male Other NC
calibration start DI3 output 5
10 20 DO5
Panel cutout
(M3.5 screw)
Specifications Flame-proof type
Standard Specifications Optional Specifications
Measuring principle Measurement of thermal conductivity 5 SPST relay contact outputs
Measurable component He,Ar,H2,CH4,CO2 Relay contact capacity; 220V AC/2A (resistive load)
Measurable range As specified for particular type. Isolated with relay between contacts, and between contacts and internal circuit.
Max. 5 functions are selectable among those listed below.
4 to 20mA DC, 0 to 1V DC, 0 to 10mV DC
Output signal Non-isolated output(Any one-output signal specifiable in CODE SYMBOLS)
Zero-side solenoid valve drive output for automatic calibration
Span-side solenoid valve drive output for automatic calibration
Allowable load resistance 550 max. (in 4 to 20mA DC output) Suction pump OFF output in automatic calibration
Output resistance 100k (in 0 to 1V DC or 0 to 10mV DC output) Upper limit (1point) concentration alarm output
Display unit LCD with backlight Lower limit (1point) concentration alarm output
Relay contact output Upper/lower limit (1point) concentration alarm output
Display of measured value Max. 4 digits
Upper limit (1point) and lower limit (1point) concentration alarm
Display language English, Japanese, Chinese output (Total 2 points)
Output signal holding In both manual and automatic calibrations, output value just before calibration can be held. 2-step upper limit (1point at each step) concentration alarm output
Power supply 100 to 240V AC, 50/60Hz, Approx. 50VA (Total 2 points)
Warm-up time At least 30min 2-step lower limit (1point at each step) concentration alarm output
(Total 2 points)
Ambient temperature 5 to 45 Analyzer error or automatic calibration error alarm output
Ambient humidity Less than 90% RH (condensation unallowable) Calibrating status output
Storage conditions 2060, less than 95% RH (condensation unallowable) Range identification output (for 2 range type only)
Mounting Mounted flush on panel
3 non-voltage contact inputs
External dimensions(HWD) 470354211mm ON; 0 V, OFF; 5V DC, current at ON; 5mA
Mass Approx. 22kg Isolated with photo coupler between inputs and internal circuit.
Finish color Off-white (equivalent to 10Y7.5/0.5) Contact input
Not isolated between contact inputs.
Housing Steel-plate case, indoor type The following actions can be input.
Remote holding of measured value output
Material of gas- JIS SUS304, platinum, platinum iridium, silver,fluororubber, epoxy resin, Remote range changeover (only with 2-range meter)
contacting parts nickel, tin Remote start of automatic calibration
Gas inlet/outlet, purge port Rc1/2 or NPT1/2 or G1/2
Analog input for H2 meter interference correction (1 to 5V DC)
Ex. standard NEPSI (Exd C T6Gb) Either CO2 or CH4 component of an external gas analyzer is to be input.
Interference gas
measured value input Adjustment is required at Fuji Electric's factory.
Performance Details of measurement gas will be checked when receiving an order.

Repeatability 1% of F.S. Zero and span calibrations are automatically carried out at the
Zero pointWithin 2% of full scale/week (H 2 meter, reference gas N2)
Automatic calibration predetermined intervals.
Drift SpanWithin 2% of full scale/week (H 2 meter, reference gas N2)
function Calibration gases are flowed sequentially by driving the externally
installed solenoid valves.
Standard within 60sec (at flow rate 0.4L/min)
Response speed High speed within 10sec (at flow rate 1L/min), allowed only for H2 meter
RS-232C (9-pin D-sub output)
(90% response) (reference gas N2)
Half duplex, asynchronous
Communicating function MODBUSTM protocol, communication speed 9600 bps
Contents of communication: Reading/writing of measured concentration
Standard Gas Measurement Conditions values and various set values, and output of device status

Temperature
Gas flow rate
0 to 50
Constant at 0.40.05 L/min
Installation Conditions
The analyzer should not be exposed to direct sunlight or radiation from a hot
Dust Less than 100 g/Nm3 with a particle size of 0.3 m max.
object.
Pressure 10kPa max.
A place subjected to heavy vibrations should be avoided. A location with
Mist,Corrosive gas Unallowable
clean atmosphere should be selected.
Moisture Below saturation at 2 When the analyzer is installed outdoors, it should be sheltered with a housing
Standard gases
Zero gas: same as reference gas or cover to protect it from rain and wind.
Span gas: Concentration within 90 to 100% of measuring range
for calibration Concentration beyond 100% is inapplicable.

Outline Diagram (Flame-proof type) (Unit: mm) Dimensions of


mounting (Unit: mm)
Gas inlet Rc1/2, NPT1/2, G1/2
72 (to order) For communicaton For contact input
Mounting hole 4-φ14

60 60
ZAF
GAS ANALYZER
186

60 60 60
40
96

402
402
442

362

Made in Japan

or power supply
GAS ANALYZER

For other
TYPE ZAFEK4A4-PYAYYACA-3N
H2 0 50, %
OUTPUT 4-20mADC
POWER 100-240VAC 50/60Hz 50VA
SER.NO.
MFD 2012-01
Fuji Electric Co.,Ltd.

GAS ANALYZER
MODEL ZAFE
Ex.PROOF Exd Ⅱ BT6
AMB.TEMP 60℃
SUPPLY 60V AC
For output For contact output
( )
90℃
TC
174

Made in Japan Fuji Electric Co.,Ltd.

/CAUTION
2

Please do a carrying task and installed work by


two people or more. Because this is heavy load.
43

Conduct conn.G1/2 286


19 87.5
4−φ14
16
Name plate
φ25.5

17 211 286
27.3 326
Gound terminal M6 M20 Pitch 1.5 Note) φ8 to φ10 cable is suitble
Cable gland for flame proof
Gas outlet Rc1/2, NPT1/2, G1/2 (to order) (to order) Cable gland

Connection Diagram
(Standard) (Option) 13th digit A and 14th digit A, C, E
1 11 DO1 RS232C communication (option)
Measured value Contact output 1
102
OUT PUT

output (as specified) 2 12 DO1 Pin Terminal Signal


number
4 to 20mA DC, 0 to 1V DC Interference correcting AIN+ 3 13 DO2
101

0 to 10mV DC Contact output 2 1 5


input 1 to 5V DC AIN– 4 14 DO2 2 Recive Data

E Grounding terminal DI1 5 15 DO3


Remote output holding
PE

Contact output 3 3 Transmit Data


DI1 6 16 DO3
SOURCE

N DI2 7 17 DO4 6 9 5 Signal GND


N

Remote range changeover


V

Power supply Contact output 4


DI2 8 18 DO4 D-Sub9P
100 to 240V AC, male Other NC
Remote automatic DI3 9 19 DO5
POWER

50/60Hz
L Contact output 5
AC

calibration start
L

DI3 10 20 DO5
(M3.5 screw) (M3.5 screw)
Code Symbols (Standard type)
4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
4 A Y Y
Digit Description
4 <Construction>
Standard type 3
CE mark 4
5 <Measured component>
H2 K
Ar L
He M
CH4 E
CO2 (reference gas Ar unallowable) A
Other Z
6 <Reference gas> (Note 1)
N2 4
Air (incompatible with H2/CH4 measurement) 5
O2 (incompatible with H2/CH4 measurement) 6
Other Z
7 <Connection port size>
Rc1/4 0
NPT1/4 1
8 <Revision No.> 4
9 <Measuring range (1st range)>
0 to 3% (H2) Q
0 to 5% (H2 ,He) L
0 to 10% (H2, He, Ar, CO2)
0 to 20%
M
N EC Directive Compliance
0 to 30% V The product conforms to the requirements of the Low Voltage Directive
0 to 50% P
0 to 80% T 2006/95/EC and EMC directive 89/336/EEC (as amended by Directive
0 to 100% J 92/31/EEC), both as amended by Directive 93/68/EEC.
100 to 90% (H2, He, Ar) 9 It conforms to following standards for product safety and electromagnetic
100 to 80% (H2, He, Ar,CH2) 8 compatibility ;
Other Z
10 <Measuring range (2nd range)> (Note 2) EN61010-1:2010,EN62311:2008
None Y Safety requirements for electrical equipment for
0 to 5% (H2 , He) L measurement, control and laboratory use.
0 to 10% (H2, He, Ar) M
0 to 20% (H2, He, Ar, CO2) N “Installation Category II”
0 to 30% V “Pollution Degree 2”
0 to 50% P “Altitude up to 2187 yard [2,000m]”
0 to 80% T EN61326-1:2006, EN61326-2-3:2006,
0 to 100% J EN61000-3-2:2006, A1:2009,A2:2009
Other Z EN61000-3-3:2008
11 <Measured value output>
A Electrical equipment for measurement, control and
DC4 to 20 mA
DC0 to 1V B laboratory use — EMC requirements.
DC4 to 20 mA + RS-232C communication C
DC 0 to 1 V + RS-232C communication D
E Input/output contact specifications 14th digit:A 14th digit:C 14th digit:E
DC0 to 10mV
12 — (Note 3) A Automatic calibration Concentration alarm Contact output selection
13 <H2 meter interference corrective calculation> (Note 4) related related (Note 7)
None Y
Provided A Automatic Zero gas valve drive (DO1)
14 <Input/output contacts> calibration Span gas valve drive (DO2)
None Y related Suction pump OFF in automatic calibration (DO3)
Automatic calibration A
Upper limit (1 point) concentration alarm
Concentration alarm See table below. C
Contact output selection Lower limit (1 point) concentration alarm Any one
Any one
15 <Indication> Upper/lower limit (1 point as a set) alarm
alarm
Japanese J Concen- concentration alarm settable on
settable on
English E Contact tration Upper limit (1 point) and lower limit screen
output alarm (1 point) concentration alarm screen
16 <Response speed> (DO1, 2)
related 2-step upper limit (1 point each) 2 Point
Standard response A 2 Point
(NO)
High-speed response (Note 5) B concentration alarm (NO)
contact
17 — Y 2-step lower limit (1 point each) contact
concentration alarm
18 — Y
Note Calibration status (DO4) (DO4)
Reference gas refers to gas other than the component to be
measured in sample gas. Other Range information (2-range meter)(Note3) (DO3)
("Z" must be specified when interference gas is to be contained.) Analyzer error or automatic calibration error (DO5) (DO5)
Note The ratio of maximum range to the first range is as given below. Remote automatic calibration start (Note4) (DI3) (DI3) (DI3)
Contact Remote range changeover (2-range meter)(Note5) (DI2)
For CO2, Ar or CH4 measurement : 1st range x 5 (times) input
(DI2) (DI2)
For He or H2 measurement : 1st range x 10 (times) A range from 0 to ...% Remote measured value output holding (Note6) (DI1) (DI1) (DI1)
cannot be combined with that from 100 to ...%.
1st range < 2nd range (Note 1) Mark: Normally Open (NO) contact
Note Specify Y if linearization in the 12th digit is not reguired. (Note 2) Mark: Normally Closed (NC) contact, after turning on power supply
(Note 3) Low range: Contacts close, High range: Contacts open
Note A CO2 or CH4 meter needs to be prepared separately. (Note 4) When contacts open 1.5 sec after their closure, automatic calibration starts.
A reverse range such as 100 to 0% cannot be specified. (Note 5) Contacts closed: Low range, Contacts open: High range
Input signal is 1 to 5 V DC. (Note 6) Contacts closed: Holding, Contacts open: Holding canceled
Adjustment is required at Fuji Electric’s factory. (Note 7) Up to 5 contact outputs can be set.
Details of measurement gas will be checked when receiving an order.
Reverse range such as 100% to 0% cannot be specified.
Cannot be specified if high-speed response is selected.
SCOPE OF DELIVERY
Note High-speed response is for H2 meter used for reference gas N2 only. Analyzer main unit
Panel mounting brackets (1 set)
2 power fuses (250 V AC, 1 A)
Instruction Manual

ITEMS TO BE PREPARED SEPARATELY


Gas sampling equipment, standard gas, receiving instrument, etc.
With interference corrective calculation: CO or CO2 gas analyzer

ORDERING INFORMATION
1. Analyzer type
2. Gas component to be measured
3. Measuring range
4. Gas component other than measured
4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
Code Symbols (Flame-proof type) E 4 A N Gas inlet Mounting hole (4−φ14)
(Rc1/2, NPT1/2,
Digit Description
4 <Construction> G1/2(to order))
Hazardous location E
5 <Measured component>
H2 K
Ar L
He M
CH4 E
CO2 (reference gas Ar unallowable) A
Other Z
6 <Reference gas> (Note 1)
N2 4
Air (incompatible with H2/CH4 measurement) 5
O2 (incompatible with H2/CH4 measurement) 6
Other Z
7 <Connection port size>
G1/2 A
Rc1/2 B
NPT1/2 C
8 <Revision No.> 4 Operation key
9 <Measuring range (1st range)>
0 to 3% (H2) Q
0 to 5% (H2 ,He) L Gas outlet For contact input/output
0 to 10% (H2, He, Ar, CO2) M (Rc1/2, NPT1/2,
0 to 20% N For power supply For communicaton (RS232C)
G1/2(to order))
0 to 30% V 100 to 240V AC
0 to 50% P 50/60Hz For output
0 to 80% T 4 to 20mA DC,
0 to 100% J 0 to 1V DC, 0 to10mV DC
100 to 90% (H2, He, Ar) 9
100 to 80% (H2, He, Ar,CH2) 8
Other Z
10 <Measuring range (2nd range)> (Note 2)
Input/output contact specifications 14th digit:A 14th digit:C 14th digit:E
None Y
0 to 5% (H2 , He) L Automatic calibration Concentration alarm Contact output selection
0 to 10% (H2, He, Ar) M related related (Note 7)
0 to 20% (H2, He, Ar, CO2) N
0 to 30% V Automatic Zero gas valve drive (DO1)
0 to 50% P calibration Span gas valve drive (DO2)
0 to 80% T related Suction pump OFF in automatic calibration (DO3)
0 to 100% J
Upper limit (1 point) concentration alarm
Other Z
11 <Measured value output> Lower limit (1 point) concentration alarm Any one
Any one
DC4 to 20 mA A Upper/lower limit (1 point as a set) alarm
alarm
DC0 to 1V B Concen- concentration alarm settable on
settable on
DC4 to 20 mA + RS-232C communication C Contact tration Upper limit (1 point) and lower limit screen
output alarm (1 point) concentration alarm screen
DC 0 to 1 V + RS-232C communication D (DO1, 2)
related 2-step upper limit (1 point each) 2 Point
DC0 to 10mV E 2 Point
(NO)
12 — A concentration alarm (NO)
contact
13 <H2 meter interference corrective calculation> (Note 3) 2-step lower limit (1 point each) contact
None Y concentration alarm
Provided A Calibration status (DO4) (DO4)
14 <Input/output contacts>
Other Range information (2-range meter)(Note3) (DO3)
None Y
Analyzer error or automatic calibration error (DO5) (DO5)
Automatic calibration A
Concentration alarm See table below. C Remote automatic calibration start (Note4) (DI3) (DI3) (DI3)
Contact output selection Contact Remote range changeover (2-range meter)(Note5) (DI2) (DI2) (DI2)
15 <Indication> input
Remote measured value output holding (Note6) (DI1) (DI1) (DI1)
Japanese J
English E
Chinese C (Note 1) Mark: Normally Open (NO) contact
16 <Response speed> (Note 2) Mark: Normally Closed (NC) contact, after turning on power supply
Standard response A (Note 3) Low range: Contacts close, High range: Contacts open
High-speed response (Note 4) B (Note 4) When contacts open 1.5 sec after their closure, automatic calibration starts.
17 <Number of cable gland> (Note 5) Contacts closed: Low range, Contacts open: High range
None 0 (Note 6) Contacts closed: Holding, Contacts open: Holding canceled
3 3
4 4 (Note 7) Up to 5 contact outputs can be set.
5 5
6 6
7 7 SCOPE OF DELIVERY
8 Z
18 <Ex. Standard> Analyzer main unit
NEPSI N Panel mounting brackets (1 set)
2 power fuses (250 V AC, 1 A)
Note Reference gas refers to gas other than the component to be measured in Opner1
sample gas.
("Z" must be specified when interference gas is to be contained.) Wrench1
Instruction Manual
Note The ratio of maximum range to the first range is as given below.
For CO2, Ar or CH4 measurement : 1st range x 5 (times)
For He or H2 measurement : 1st range x 10 (times) A range from 0 to ...% ITEMS TO BE PREPARED SEPARATELY
cannot be combined with that from 100 to ...%. Gas sampling equipment, standard gas, receiving instrument, etc.
1st range < 2nd range
With interference corrective calculation: CO or CO2 gas analyzer
Note A CO2 or CH4 meter needs to be prepared separately.
A reverse range such as 100 to 0% cannot be specified.
Input signal is 1 to 5 V DC. ORDERING INFORMATION
Adjustment is required at Fuji Electric’s factory.
Details of measurement gas will be checked when receiving an order. 1. Analyzer type
Reverse range such as 100% to 0% cannot be specified. 2. Gas component to be measured
Cannot be specified if high-speed response is selected. 3. Measuring range
Note High-speed response is for H2 meter used for reference gas N2 only. 4. Gas component other than measured

International Sales Div.


Sales Group
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fujielectric.com
Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425
http://www.fujielectric.com/products/instruments/

Information in this catalog is subject to change without notice. Printed in Japan 2013-8/30FOLS

You might also like