W044 - Webinar Presentation - Click to View
W044 - Webinar Presentation - Click to View
Same lab
ISO 17043
INTRODUCTION
Both PT and ILC aim to monitor and improve the quality and accuracy of
laboratory testing, but they have some distinct differences:
Aspect Proficiency Testing (PT) Interlaboratory Comparisons (ILCs)
Structure Highly standardized and formal Flexible and can be informal
Managed by an external provider Can be self-organized or facilitated by
Oversight
(ISO/IEC 17043 accredited) peers
(c) is the evaluation criteria for the competency of a Proficiency Testing Scheme provider –
accredited to ISO/IEC 17043 Conformity assessment – General requirements for the
competence of proficiency testing providers.
SELECTING A PT PROVIDER
When selecting a PT provider, the laboratory should consider the type of
target value offered. Target values are:
a) independently set by a reference method, or
b) set by overall consensus data, and/or
c) set by method peer group consensus data, or
d) set by a panel of experts.
“Target values” refers to the desired or expected result that laboratories aim
to achieve when participating in an PT program. It serves as a benchmark
against which the laboratory's own test results are compared to assess the
accuracy and reliability of their testing processes.
SELECTING A PT PROVIDER
The different types of target values mentioned in the clause represent how
these benchmarks are established:
1. Independently set by a reference method:
• The target values are determined by a well-established reference method,
which is considered the gold standard for the particular test or assay.
• Laboratories aim to achieve results that closely match those obtained by
the reference method.
2. Set by overall consensus data:
• Target values may also be established based on consensus data derived
from a large group of laboratories performing the same test.
• This consensus reflects the typical or expected range of results obtained
by these laboratories.
SELECTING A PT PROVIDER
3. Set by method peer group consensus data:
• Some PT programs set target values based on consensus data specific to
a peer group of laboratories using the same testing method or platform.
• This ensures that the benchmarks are tailored to the performance
characteristics of the method being used.
4. Set by a panel of experts:
• In certain cases, target values may be established by a panel of experts in
the field who review available evidence and provide recommendations
based on their expertise.
• These experts may consider factors such as clinical significance and
technical feasibility when setting target values.
SELECTING A PT PROVIDER
Other factors to consider:
• Do any PT schemes exist for the various areas of technical competence?
• Is the frequency of rounds per year sufficient?
• Is the selected PT scheme independent of any manufacturing or marketing
interests in equipment, reagents or calibrators in its field of operation?
• Regulatory requirements / import permits required.
• Technical support provided by PT provider. Consider the ease of access to
PT provider contacts, including timely and adequate responses to queries.
• Costs of participation.
ISO/IEC 17025:2017 REQUIREMENTS
ISO/IEC 17025:2017 Clause 7.7.1:
The laboratory shall have a procedure for monitoring the validity of results.
The resulting data shall be recorded in such a way that trends are
detectable and, where practicable, statistical techniques shall be applied to
review the results.
ISO/IEC 17025:2017 REQUIREMENTS
ISO/IEC 17025:2017 Clause 7.7.1 (continued):
Implementation:
• Obtain information from the selected PT Scheme regarding the enrollment
process and participation requirements. Document this information in the
laboratory’s internal procedures. This should include all relevant details
related to the handling of PT samples, submission of results, receipt of
results, and related activities.
• If the laboratory participates in ILC activities, documented procedures
should also cover all the above-stated requirements.
• Describe how and where results are recorded and describe the statistical
analysis tools and acceptance criteria used to evaluate the results.
• Consider additional accreditation requirements related to PT participation.
ISO/IEC 17025:2017 REQUIREMENTS
ISO/IEC 17025:2017 Clause 7.7.1 (continued):
This monitoring shall be planned and reviewed and shall include, where
appropriate, but not be limited to:
a) use of reference materials or quality control materials;
b) use of alternative instrumentation that has been calibrated to provide
traceable results;
c) functional check(s) of measuring and testing equipment;
d) use of check or working standards with control charts, where applicable;
e) intermediate checks on measuring equipment;
ISO/IEC 17025:2017 REQUIREMENTS
ISO/IEC 17025:2017 Clause 7.7.1 (continued):
This monitoring shall be planned and reviewed and shall include, where
appropriate, but not be limited to:
f) replicate tests or calibrations using the same or different methods;
g) retesting or recalibration of retained items;
h) correlation of results for different characteristics of an item;
i) review of reported results;
j) intralaboratory comparisons;
k) testing of blind sample(s).
ISO/IEC 17025:2017 REQUIREMENTS
ISO/IEC 17025:2017 Clause 7.7.2:
The laboratory shall monitor its performance by comparison with results of
other laboratories, where available and appropriate. This monitoring shall be
planned and reviewed and shall include, but not be limited to, either or both
of the following:
a) participation in proficiency testing;
b) participation in interlaboratory comparisons other than proficiency
testing.
• Laboratories should establish clear criteria for assessing the acceptability of PT/ILC
results. These criteria may include allowable limits of deviation from target values,
precision requirements, or other performance indicators relevant to the specific tests
being assessed.
COMMON PROBLEMS ENCOUNTERED
• Failure to have a comprehensive PT/ILC plan.
• Failure to keep the PT/ILC plan up to date.
• Failure to adhere to the PT/ILC plan.
• Not using ISO 17043 accredited schemes when they are available.
• Participation with another laboratory and each laboratory draws up its own report with
different conclusions. (i.e., no procedure implemented before ILC).
• Inappropriate choice of matrix or reference material.
• Trying to pass off a PT/ILC for one test method as being adequate as competence for
another method.
• Failing to prepare a testing protocol before commencing the PT/ILC which means that
different laboratories test at different points and report differently, and statistical evaluation
is not possible).
• The PT/ILC report does not cover the required items listed in the accreditation
documents.
PT / ILC FREQUENCY
The laboratory should define the level and frequency of its participation
after careful analysis of its other quality assurance (QA) measures to ensure
the validity of the results (especially those that are able to disclose, quantify
and follow the development of bias of a stated magnitude).
PT samples must be analysed and fall within the working range of the test
method. For example:
PT value
Working range:
10 ppm 1000 ppm
PERSONNEL COMPETENCY
Establishing personnel test method competency can generally be
divided into 3 parts:
a) Initial and on-going training as per a documented training program.
b) Witnessing / observation of the performance of a test method to
determine that the activity is being performed correctly.
c) Participating in PT/ILC to determine that the results obtained are
accurate.
Negative %Bias Indicates that the laboratory’s results are lower than the target value.
ACCEPTANCE CRITERIA: Z-SCORE
Z<1 The data point is within one standard deviation of the mean.
The data point is one or more standard deviations away from the mean,
Z≥1
indicating it is relatively unusual.
ACCEPTANCE CRITERIA: Z-SCORE
Formula for Z-score:
The z-score for a data point 𝑋 is calculated using the formula:
ACCEPTANCE CRITERIA: Z-SCORE
Calculation:
• After all participating laboratories have submitted their results for a
particular test, the mean (𝜇) and standard deviation (𝜎) of these results are
calculated.
• Each laboratory's result is then converted to a z-score using the above
formula.
ACCEPTANCE CRITERIA: Z-SCORE
Simply put:
Performance evaluation: Think of the z-scores as a QC chart in your
• -2 ≤ z ≤ 2: The result is generally laboratory and the z-score represents the number
considered acceptable and indicates of standard deviations your result is from the true
value.
satisfactory performance.
3 3sd
2 2sd
• 2 < z < 3 or -2 > z > -3: The result is a
1 1sd
warning signal, suggesting that the
Z-score
0
laboratory should review its procedures
and check for potential issues. -1 1sd
-2 2sd
-3 3sd
It may be possible that after a thorough investigation, the origin of the poor
performance is not identified. A single poor performance could then be
attributed to random error. If it is a repeated poor performance, then the
analytical process should be questioned.
This webinar is sponsored by NSI Lab Solutions
THANK YOU
www.michemdynamics.com