Spirometer OG
Spirometer OG
Owner’s Guide
This document was, as far as possible, accurate at the time of release. However,
changes may have been made to the software and hardware it describes since
then. ADInstruments Pty Ltd reserves the right to alter specifications as
required. Late-breaking information may be supplied separately.
Trademarks of ADInstruments
PowerLab®, LabTutor® and MacLab® are registered trademarks of ADInstruments
Pty Ltd. The names of specific recording units, such as PowerLab 8/30, are
trademarks of ADInstruments Pty Ltd. LabChart, Chart and Scope (application
programs) are trademarks of ADInstruments Pty Ltd.
Other Trademarks
Apple, Mac and Macintosh are registered trademarks of Apple Computer, Inc.
All rights reserved. No part of this document may be reproduced by any means
without the prior written permission of ADInstruments Pty Ltd.
Web: www.adinstruments.com
Technical Support: [email protected]
Documentation: [email protected]
Safety Notes 5
1 Overview 13
How to Use This Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Checking the Front-end . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Front-end Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
The Front-end . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
The Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
The Status Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
The Spirometer Input Pipes . . . . . . . . . . . . . . . . . . . . . . . . . 16
The Back Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
I2C Input and Output Sockets . . . . . . . . . . . . . . . . . . . . . . . . 16
The Analog Out Socket . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Equipment and Technique . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Using the Spirometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Fitting the Flow Head. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Calibrating the Flow Head . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Reducing Drift . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
2 Setting Up 23
Connecting to the PowerLab. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Single Front-ends . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Multiple Front-ends . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Using LabChart and Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
The Front-end Driver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Front-end Self-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
The Spirometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Signal Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Setting the Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Filtering. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Inverting the Signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Offset Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Contents iii
A Technical Aspects 31
Spirometer Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
B Troubleshooting 35
Problems and Solutions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
C Specifications 39
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Control Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Physical Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Index 41
• is a medical device,
• may be interpreted as a medical device, or
• is safe to be used as a medical device.
Safety Notes 5
Safety Symbols
Devices manufactured by ADInstruments that are designed for direct
connection to humans are tested to IEC 601-1:1998 (including
amendments 1 and 2) and 60601-1-2, and carry one or more of the
safety symbols below. These symbols appear next to those inputs and
output connectors that can be directly connected to human subjects.
!
BF symbol: Body- CF symbol: Cardiac- Warning symbol: ‘see
protected equipment protected equipment documentation’
Safety Notes 7
• Do not use faulty cables, or those that have exhibited intermittent
faults.
• Do not attempt to measure or record the Isolated Stimulator
waveform while connected to a subject using a PowerLab input or
any other piece of equipment that does not carry the appropriate
safety symbol (see Safety Symbols above).
Always check the status indicator on the front panel. It will always
flash green each time the stimulator delivers a current pulse. A yellow
flash indicates an ‘out-of-compliance’ (OOC) condition that may be
due to the electrode contact drying up. Always ensure that there is
good electrode contact at all times. Electrodes that are left on a subject
for some time need to be checked for dry contacts. An electrode
impedance meter can be used for this task.
Safety Notes 9
• Many electrical systems (particularly those in metal enclosures)
depend upon the presence of a protective earth for electrical safety.
This is generally provided from the power outlet through a power
cord, but may also be supplied as a dedicated safety earth
conductor. Power cords should never be modified so as to remove
the earth connection. The integrity of the protective earth
connection between each piece of equipment and the protective
earth should be verified regularly by qualified personnel.
• Avoid using multiple portable socket-outlets (such as power
boards) where possible as they provide an inherently less safe
environment with respect to electrical hazards. Individual
connection of each piece of equipment to fixed mains socket-
outlets is the preferred means of connection.
If multiple portable socket outlets are used, they are subject to the
following constraints:
If the PowerLab system is found not to comply with such testing you
should contact your PowerLab representative to arrange for the
equipment to be checked and serviced. Do not attempt to service the
device yourself.
Environment
Electronic components are susceptible to corrosive substances and
atmospheres, and must be kept away from laboratory chemicals.
Storage Conditions
Operating Conditions
Disposal
Safety Notes 11
12 Spirometer Owner’s Guide
1 Overview
Chapter 1 Overview 13
How to Use This Guide
This owner’s guide describes how to set up and begin using your
Spirometer. The chapters give an overview of front-ends in general
and the Spirometer in particular, and discuss how to connect the
hardware, perform a simple power-up test, and use the front-end with
some ADInstruments programs. The appendices provide technical
information about the Spirometer, and take a look at some potential
problems and their solutions.
At the end of this guide, you’ll find an index. Technical terms that are
not defined in the glossary of terms included with the owner’s guide
for your PowerLab, or in the guide that came with your computer, are
defined as they appear.
Front-end Basics
The PowerLab system consists of a recording unit and application
programs that run on the computer to which the unit is connected. It
is an integrated system of hardware and software designed to record,
display, and analyze experimental data. Your Spirometer is one of a
family of front-ends meant for use with your PowerLab system.
The Front-end
The Spirometer is a precision differential pressure transducer for
measuring respiratory variables, such as inspiration and expiration
flows and tidal volumes. It measures differential pressure across fine
gauze mounted in a flow head. With a flow head of a suitable size, the
Spirometer can be used with a variety of creatures, from small animals
such as mice and rats, to large animals and humans. Accessories such
as flow heads (various sizes), tubing, and calibration syringes are
available, and can be purchased separately.
Input pipes
Status indicator
Chapter 1 Overview 15
The Status Indicator
WARNING Connections are made to the Spirometer using two pipes on the front
The Spirometer input is panel. These are physical connections for airflow, not electrical ones.
sensitive. Do not blow into Two flexible plastic tubes (3 mm internal diameter, 5 mm external
or apply high air flows to diameter) connect the input pipes on the Spirometer to the connection
the Spirometer input, as
this may damage the
pipes on the flow head. The input pipes carry a warning symbol (see
internal transducer margin).
Figure 1–2
The back panel of the
Spirometer
Signal output to PowerLab
The flow head itself contains no electronic parts, and is simply a tube
with a wire mesh placed across it. Two pipes, one on either side of the
mesh, allow the pressure difference to be measured by a high-
precision differential pressure transducer in the Spirometer itself,
when connected with plastic tubing.
The Spirometer can support several sizes of flow head, each with
differing maximum flows, but all with the same connection to the
Spirometer. Standard flow heads, obtainable separately, are:
Chapter 1 Overview 17
Fitting the Flow Head
To connect the flow head to the Spirometer, simply push the ends of
the two connection tubes firmly over the flow head pipes and over the
input pipes on the front panel of the Spirometer. In some cases you
may find that the tubes are difficult to fit because they are too tight. If
so, dip the ends of the tubes into some boiling water to soften the
plastic to make it easier to push the tubes onto the pipes.
Any leakage from the joint will affect the precision of the flow
readings, so ensure that the tubes are pushed on firmly. The flow head
is washable and able to be cold-sterilized, and should be dried gently
before use. Care should be taken to ensure that condensation does not
block the tubing connecting the flow head to the Spirometer. To avoid
problems, the flow head should be turned so that the tubing connects
at the top, not at the bottom.
Figure 1–3
A flow head, with the pipes
in the correct position:
upright
Before using the flow head, you will probably want to calibrate the
Spirometer to read in terms of flow (L/s rather than V). There are two
ways of doing this: using an approximate conversion factor, or
injecting a known volume and integrating. The Spirometry Extension
for LabChart (available on both Windows and Macintosh platforms)
can be used to assist with either of these methods.
Set up a new channel for Volume from the Channel Settings dialog.
Display the integral of the flow signal in this channel by choosing
Integral… from the Channel Function pop-up menu and selecting Flow
as the source channel, using a standard integral with no reset.
Chapter 1 Overview 19
Figure 1–4
The flow signal integrated
for a single injection.
The value of the integral at
the cursor position is 3L
–76.79 mV.s ----------------------------- = 39.1 L/s/V
0.07679 V.s
Figure 1–5
Setting units to calibrate
the Spirometer channel
Finally, if you are using the Spirometry Extension, you can apply drift
correction after recording, provided that your recording meets certain
conditions. For more details on drift correction, please refer to the
Spirometry Extension User’s Guide (Macintosh) or the Help Center
(Windows).
Chapter 1 Overview 21
22 Spirometer Owner’s Guide
2 Setting Up
Chapter 2 Setting Up 23
Connecting to the PowerLab
To connect a front-end to the PowerLab, first ensure that the
PowerLab is turned off. Failure to do this may damage the PowerLab,
the front-end, or both.
Single Front-ends
Connect the I2C output of the PowerLab to the I2C input of the front-
end using the I2C cable provided.
Check that the plugs for the I2C bus are screwed in firmly. Check the
BNC cable for firm connections as well. Loose connectors can cause
erratic front-end behavior, or may cause the front-end to fail to work
at all. The BNC cable can be tucked under the front-end to keep it out
of the way if desired.
Chapter 2 Setting Up 25
If the application fails to find a front-end attached, the normal text
remains. If you were expecting a connected front-end, close the
program, turn the PowerLab off and check the connections. Then
restart the PowerLab and program. Note: do not leave the PowerLab
on while checking the connections, as doing so may damage the
PowerLab, the front-end, or both.
There are several front-end drivers for the various front-ends made by
ADInstruments (a driver is a piece of software the computer uses to
control a peripheral device). For example the Bridge Amplifier front-
end driver is used with the Spirometer and GP Amp; and the Stimulus
Isolator front-end driver is used with the Stimulus Isolator.
Front-end Self-test
If the indicator lights correctly, the front-end has been found by the
PowerLab and is working properly, and you can close the application
or carry on as appropriate. If the indicator doesn’t light, check your
cable connections and repeat the procedure. If this does not solve the
problem, contact your ADInstruments representative.
Amplitude
axis
Click this
to remove
the offset
for the
Spirometer
Signal Display
The signal at the channel input is displayed so you can see the effect
of changing the settings – data is not recorded while setting things up.
Slowly changing signals are represented quite accurately, whereas
quickly changing signals are displayed as a solid dark area showing
only the envelope (shape) of the signal from minimum and maximum
recorded values. The average signal value is displayed at the top of
the display area: the offset is displayed when the Spirometer is not
zeroed, and may indicate a problem if it is large.
You can stop the signal scrolling by clicking the Pause button at the
bottom left (Macintosh) or top right (Windows) of the data display
area. This changes to the Scroll button on the Macintosh. Click the
Scroll button to start scrolling again.
Chapter 2 Setting Up 27
Shift and stretch the vertical Amplitude axis, by clicking and dragging
it in various ways, to make the best use of the available display area.
It functions the same as the Amplitude axis of the Chart Window,
controls are identical and any change is applied to the Chart Window.
The Range pop-up menu lets you select the input range or sensitivity
of the channel. Changing the range in the Spirometer dialog is
equivalent to changing it in the Chart or Scope window. The available
ranges are 500 mV, 200 mV, 100 mV, 50 mV and 20 mV.
Filtering
Offset Adjustment
Units
Click Units… to open the Units Conversion dialog, with which you can
set the units for a channel and, using waveform measurements,
calibrate the channel. The waveform in the data display area of the
Spirometer dialog is transferred to the data display area of the Units
Conversion dialog. (Use the Pause button to capture a specific signal.)
This units conversion only applies to subsequently recorded signals,
so it is more limited than choosing Units Conversion… from the
Channel Function pop-up menu, as it does not allow the conversion of
individual blocks or pages of data.
Chapter 2 Setting Up 29
30 Spirometer Owner’s Guide
A P P E N D I X
A Technical
Aspects
100Hz
100 Hz
Signal
Transducer
Signal
Transducer
Pressure
Pressure Pressure
Output
Output Pressure
1010Hz
Hz Gain Inputs
Inputs
11Hz
Hz
B
B Troubleshooting
This appendix describes most of the common problems that can occur
when using the Spirometer with your PowerLab recording unit. It
covers how these problems are caused, and what you can do to
alleviate them. If the solutions here do not work, earlier chapters, the
LabChart Help Center, and the guide to your PowerLab may contain
possible solutions. If none of the solutions here or elsewhere are of
help, then consult your ADInstruments representative.
Appendix B Troubleshooting 35
Problems and Solutions
The status indicators fail to light when the software is started, or the
front-end commands and so on do not appear where they should
The I2C cable or the BNC-to-BNC cable from the front-end to the
PowerLab is not connected, has been connected incorrectly (to the
wrong input or output, for instance), or is loose.
• Turn everything off. Check to see that all cables are firmly seated
and screwed in. The BNC cable from the Signal Output of the
Spirometer must be connected to a positive input on the
PowerLab. Make sure the input is the same channel from which
you expect to use the front-end in the software. Start up again to
see if this has fixed the problem.
• Replace the cable and try again. Immediately label all cables
proved faulty so that you don’t use them again by accident.
• This is the least likely event. If the front-end will not work
properly after the previous measures, then try using it on another
PowerLab. If the same problems recur with a second PowerLab,
the front-end may be faulty. Contact your ADInstruments
representative to arrange for repairs.
• Replace the cable and try again. Immediately label all cables
proved faulty so that you don’t use them again by accident.
• This is the least likely event. If the front-end will not work
properly after the previous measures, then try using it on another
PowerLab. If the same problems recur with a second PowerLab,
the front-end may be faulty. Contact your ADInstruments
representative to arrange for repairs.
The trace will not zero properly when using the automatic or manual
zeroing controls
• Make sure that the apparatus is kept still and that no varying
signal is applied during auto-zeroing.
The signal from the flow head is beyond the range of the Spirometer’s
zeroing circuitry.
Appendix B Troubleshooting 37
The signal is noisy at lower ranges
This is usually the amplified noise from the transducer and its
associated circuitry, not a fault as such.
• Ensure that both the tubing and gauze is clean and is free from
condensation, otherwise dry it. Make sure that the flow head is
used with the tubes upwards.
C
C Input
Specifications
Appendix C Specifications 39
Repeatability: ± 0.25% full scale
Filters
Low-pass filtering: 1, 10 or 100 Hz (software-selectable) using
fourth-order Bessel filter
Control Port
I2C port: Provides control and power. Interface
communications rate of ~50 kbits/s.
Physical Configuration
Dimensions (h × w × d): 50 mm × 76 mm × 260 mm
(1.96" × 3.0" × 10.2")
A I
ADInstruments programs 25–29 I2C bus 15, 16, 31, 32
analog output 17
automatic zeroing 28
L
B LabChart 25–29
LabChart extensions
back panel 16–17 Spirometry 19, 20–21
block diagram 32
M
C
maintenance 10
calibrating flow heads 19–21
checking the front-end 14
cleaning 10
O
connections offset adjustment 28–29
multiple front-ends 25
single front-end 24
to the PowerLab 24–25
P
PowerLab system 14
D problems and solutions 35–38
Index 41
Spirometer
accessories 18
dialog 27
software 27–29
Spirometry Extension 19, 20–21
status indicator 16
storage 10
T
technical specifications 39–40
U
user modification voids warranty 31
using ADInstruments programs 25–29
using LabChart 25–29
using Scope 25
using this guide 14
Z
Zero button 28
zeroing
automatic 28
Spirometer 28–29