The document provides installation and startup instructions for the DICOM MPPS + SC (ExamWorks+) option for Siemens imaging systems, including FLUOROSPOT T.O.P., POLYTRON T.O.P., and AXIOM Artis. It outlines system requirements, necessary tools, and safety information, emphasizing that installation should be performed by qualified personnel. Additionally, it details configuration steps for integrating ExamWorks+ into the imaging systems and managing DICOM data effectively.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
15 views29 pages

AX41-pdf .814.01.02.02

The document provides installation and startup instructions for the DICOM MPPS + SC (ExamWorks+) option for Siemens imaging systems, including FLUOROSPOT T.O.P., POLYTRON T.O.P., and AXIOM Artis. It outlines system requirements, necessary tools, and safety information, emphasizing that installation should be performed by qualified personnel. Additionally, it details configuration steps for integrating ExamWorks+ into the imaging systems and managing DICOM data effectively.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 29

DICOM MPPS + SC

Installation and Startup


DICOM MPPS + SC
AXIOM Artis Systems and Systems with FLUOROSPOT T.O.P., POLY‐
TRON T.O.P.,
2 Document Version / Disclaimer / Copyright

Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents cor‐
respond to the version at the time of system delivery and/or printout. Versions to hard‐
copy documentation are not automatically distributed.
Please contact your local Siemens office to order a current version or refer to our website
http://www.healthcare.siemens.com.

Disclaimer
Siemens provides this documentation “as is” without the assumption of any liability under
any theory of law.
The content described herein requires superior understanding of our equipment and may
only be performed by qualified personnel who are specially trained for such installation
and/or service.

Copyright
“© Siemens, 2001” refers to the copyright of a Siemens entity such as:
Siemens Healthcare GmbH - Germany
Siemens Aktiengesellschaft - Germany
Siemens Shenzhen Magnetic Resonance Ltd. - China
Siemens Shanghai Medical Equipment Ltd. - China
Siemens Healthcare Private Ltd. - India
Siemens Medical Solutions USA Inc. - USA
Siemens Healthcare Diagnostics Inc. - USA and/or
Siemens Healthcare Diagnostics Products GmbH - Germany

DICOM MPPS + SC AX41-000.814.01.02.02 Page 2 of 29 © Siemens, 2001


08.01 TD SD 21
Table of Contents 3

1 General Information 4

1.1 Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.1 Applicability and Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.2 System Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.3 Required Aids and Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.4 Additionally Required Documents . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1.5 Room Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1.6 Remark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1.7 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1.8 Product-specific Remarks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2 Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

2 Installation 7

3 Image System Setup 8

3.1 FLUOROSPOT TOP Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8


3.2 Configuring the POLYTRON TOP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.3 Configuring the AXIOM Artis. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4 ExamWorks+ Setup 11

4.1 General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11


4.1.1 Network Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.1.2 Accessing the Network Configurator . . . . . . . . . . . . . . . . . . . . . . 11
4.2 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.2.1 Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.2.2 Service Login . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.2.3 Configuration Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2.4 Application Language Configuration . . . . . . . . . . . . . . . . . . . . . . 17
4.2.5 Network Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.2.6 Config in the System Configuration Time and Date . . . . . . . . . . . 20
4.2.7 Input Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.2.8 Workflow Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.2.9 Output Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.2.10 Reboot ExamWorks+ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.2.11 Configure Time and Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.2.12 Save the Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4.2.13 ExamWorks+ Shutdown from GUI (Graphic User Interface) . . . . . 24
5 Final Work Steps 26

5.1 Function Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26


6 Changes to Previous Version 27

7 List of Hazard IDs 28

© Siemens, 2001 AX41-000.814.01.02.02 Page 3 of 29 DICOM MPPS + SC


08.01 TD SD 21
4 1 General Information

1.1 Requirements

1.1.1 Applicability and Purpose


Installation and startup of the DICOM MPPS + SC (ExamWorks+) option in the FLUORO‐
SPOT T.O.P., POLYTRON T.O.P. systems and systems with AXIOM Artis are described in
this document.
With the DICOM Modality Performed Procedure Step (MPPS), progress of the examination
in the modality is reported to the RIS and data such as the dose area product can be trans‐
ferred to the RIS. With the DICOM Storage Commitment (SC) it is assured that DICOM im‐
ages will be reliably saved in the image archive.
Images to be archived are sent from the image system to ExamWorks+. ExamWorks+ han‐
dles archiving and saves the images until archiving has been successfully completed.
Additional patient data (e.g. dose data) can be entered in the ExamWorks+ via the
touchscreen monitor.
The keyboard that is included in the shipment is connected exclusively for service and
configuration purposes. After completing these work steps, the keyboard must be re‐
moved.

1.1.2 System Requirements


n Systems with FLUOROSPOT T.O.P.
- DICOM Worklist Option
- DICOM Send Option
- Software Version ≥ VD10A
- Service CD ≥ 6
n Systems with POLYTRON T.O.P.
- DICOM Worklist Option
- DICOM Send Option and/or Udig Option
- Software Version ≥ VF00R
n AXIOM Artis Systems
- DICOM Worklist Option

1.1.3 Required Aids and Tools


All tools, test equipment and aids, with the exception of the "standard installation tool
kit", are listed and specified in the Service Tools Catalogue.

n Standard installation tool kit

DICOM MPPS + SC AX41-000.814.01.02.02 Page 4 of 29 © Siemens, 2001


08.01 TD SD 21
General Information 1 5

1.1.4 Additionally Required Documents

n Project Plan
n FLUOROSPOT T.O.P. Installation and Setting Instruction HCC, SIENET, DICOM,
RIS, Paper Print, SW ≥ VD10A
n POLYTRON T.O.P. Configuration Guide
n AXIOM Artis Configuration System

1.1.5 Room Preparations


n A power line outlet must be installed in a radius of 1.5 m from the installation loca‐
tion. Since the ExamWorks+ is switched off only in a service situation, connection to
line power may not be made in the control room distributor.
n Have the RJ45 Ethernet connection cable ready at the setup location.

1.1.6 Remark
This document was originally written in German.

1.1.7 Safety Information


When carrying out the work steps and checks, the general safety information for medical
products must be observed.

1.1.8 Product-specific Remarks


Checking the Cabling and Power Connection
n For safety reasons, the power connection may only be made by a SIEMENS technician/
electrician.
n Cabling must also be checked by the SIEMENS technician prior to switching on the sys‐
tem.

© Siemens, 2001 AX41-000.814.01.02.02 Page 5 of 29 DICOM MPPS + SC


08.01 TD SD 21
6 1 General Information

1.2 Service
n The “Installation and Service Manual” from the Merge Company is included in the ship‐
ment as a service document.
n In an error situation, the complete ExamWorks+ must be replaced.
n Contact with Merge should be made only through HSC.

DICOM MPPS + SC AX41-000.814.01.02.02 Page 6 of 29 © Siemens, 2001


08.01 TD SD 21
Installation 2 7

Fig. 1: ExamWorks+ Fig. 2: ExamWorks+ cable connections

1
2

n Install the desktop stand included in the shipment according to the accompanying in‐
stallation instructions.
n Set up the ExamWorks+ per the Project Plan. As a rule, the ExamWorks+ is set up in
the control room within easy reach of the operator. Make sure that there is sufficient
space left for the ventilation slots (> 15 cm).
n Provide a power supply ( 1/Fig. 2 Page 7).
n Plug in the Ethernet cable ( 2/Fig. 2 Page 7).
n Snap the rear cover panel onto the desktop stand.

© Siemens, 2001 AX41-000.814.01.02.02 Page 7 of 29 DICOM MPPS + SC


08.01 TD SD 21
8 3 Image System Setup

3.1 FLUOROSPOT TOP Configuration


n Configure the ExamWorks+ as a new network node in the image system (Create new
record) and add the ExamWorks+ in the group members in “my nodes” (FLUOROSPOT
T.O.P.)n.
n For this, see:
FLUOROSPOT T.O.P.
Installation and Setting Instructions
HCC, SIENET, DICOM, RIS, Paper Print, SW ≥ VD10A

DICOM MPPS + SC AX41-000.814.01.02.02 Page 8 of 29 © Siemens, 2001


08.01 TD SD 21
Image System Setup 3 9

3.2 Configuring the POLYTRON TOP


n Configure the ExamWorks+ as a new network node in the image system (Create new
record) and add the ExamWorks+ in the group members in “my nodes” (POLYTRON
T.O.P.).
n For this, see:
Configuration Guide
System software ≥ VF0..

© Siemens, 2001 AX41-000.814.01.02.02 Page 9 of 29 DICOM MPPS + SC


08.01 TD SD 21
10 3 Image System Setup

3.3 Configuring the AXIOM Artis


The ExamWorks+ is configured as a new DICOM network node in the AXIOM Artis System.
n For this, see:
AXIOM Artis
Configuration
System

DICOM MPPS + SC AX41-000.814.01.02.02 Page 10 of 29 © Siemens, 2001


08.01 TD SD 21
ExamWorks+ Setup 4 11

4.1 General Remarks


The ExamWorks+ Service Script is a configuration interface with help files to assist you in
understanding the information required for a given field. The Service Script configures
the following:
n User interface language
n Network
n Time and Date
n DICOM inputs and outputs
n Workflow, including MWL and MPPS

4.1.1 Network Configuration


This section explains how to configure the ExamWorks+ network parameters such as stat‐
ic IP address and netmask. Contact the on-site system administrator for IP Address, Net‐
mask, Host name, and related network information.

4.1.2 Accessing the Network Configurator


To access the Network Configurator, select Perform Network Configuration from the Serv‐
ice Script.

© Siemens, 2001 AX41-000.814.01.02.02 Page 11 of 29 DICOM MPPS + SC


08.01 TD SD 21
12 4 ExamWorks+ Setup

4.2 Configuration

4.2.1 Preparations
n The keyboard included in the shipment must be connected for configuration and serv‐
ice purposes.
n The following configuration data must be requested from the customer’s network ad‐
ministrator:

4.2.1.1 Configuration Data: ExamWorks+

ExamWorks+ IP address:
Subnet mask:
Host Name: EWPlus1
Domain Name:
Default Gateway:

1. Suggestion, can be changed.

4.2.1.2 Configuration Data: Storage SCP ExamWorks+

Storage SCP Local AE Title1:


Port No:

1. This data’s also have to be configured in the modality (AX System)

4.2.1.3 Configuration Data: Modality Worklist Server Information

Modality Worklist Server Infor‐ MWL Server IP address:


mation
MWL Server Port No:
Called AE Title 1.:
Local AE Title :
Local AE Title for Modality
Unit (Imaging System2).:

1. The CALLED_AE title is the AE title of the MWL server in the network.
2. The LOCAL_AE Title used for Modality Specific Worklist query. The value is the same as that configured in
the modality (AX System).

DICOM MPPS + SC AX41-000.814.01.02.02 Page 12 of 29 © Siemens, 2001


08.01 TD SD 21
ExamWorks+ Setup 4 13

4.2.1.4 Configuration Data: MPPS Information

MPPS Information MPPS Server IP adress:


MPPS Server Port No:
Called AE Title 1.:
Local AE Title .:

1. The Called_AE Title is the AE Title of the MPPS Server in the Network.

4.2.1.5 Configuration Data: Storage Commitment Information

Storage Commitment Informa‐ Local AE Title1:


tion
Local Listening Port2:

1. The local AE title is the ExamWorks+ Storage Commitment AE title.


2. ExamWorks+ Storage Commitment Port Number

4.2.1.6 Configuration Data: Storage Target Information

Storage Target Information


Target 1 Target Name1:
Target IP address:
Target AE Title.:
Target Port Number:
Storage Commit called
AE Title2:
Storage Commit Target
IP adress:
Storage Commit Target
Port Number:

© Siemens, 2001 AX41-000.814.01.02.02 Page 13 of 29 DICOM MPPS + SC


08.01 TD SD 21
14 4 ExamWorks+ Setup

Target 2 Target Name1:


Target IP address:
Target AE Title.:
Target Port Number:
Storage Commit called
AE Title2:
Storage Commit Target
IP adress:
Storage Commit Target
Port Number:

Target 3 Target Name1:


Target IP address:
Target AE Title.:
Target Port Number:
Storage Commit called
AE Title2:
Storage Commit Target
IP adress:
Storage Commit Target
Port Number:

DICOM MPPS + SC AX41-000.814.01.02.02 Page 14 of 29 © Siemens, 2001


08.01 TD SD 21
ExamWorks+ Setup 4 15

Target 4 Target Name1:


Target IP address:
Target AE Title.:
Target Port Number:
Storage Commit called
AE Title2:
Storage Commit Target
IP adress:
Storage Commit Target
Port Number:

1. The target name is the user readable target name.


2. This is the AE title, the ExamWorks+ uses to send storage commitment request for this target. Only nee‐
ded if the target AE title and storage commit AE title are different.

4.2.2 Service Login


n Apply power to the ExamWorks+ unit.
n Select the <Utils> tab.
n Select the <Exit to Service> button.
n Type the following password.
- - Password: <m0rg01>
n Select the <Done> button.
n Press <Ctrl> + <Alt> + <Backspace> on the keyboard to exit to the LINUX command
line prompt.
n At the prompts, enter:
- Login: service ↵
- Password: m0rg01 ↵

© Siemens, 2001 AX41-000.814.01.02.02 Page 15 of 29 DICOM MPPS + SC


08.01 TD SD 21
16 4 ExamWorks+ Setup

Fig. 3: Main menu for ExamWorks+ service interface

The Main Menu for the ExamWorks service interface displays. The ExamWorks+ is now
ready to be configured.
n From the Main Menu for the ExamWorks Service Interface, select <1> ↵ Configure
System.
n Type the following password.
- - Password: m0rg01 ↵

To navigate the Network Configurator, use the keyboard <Tab> key to move among the
fields and functions. Use the <Up Arrow> and <Down Arrow> keys to move through op‐

 tions in a group or list. Press the <Enter> or <Return> key to select the highlighted field,
function, or option.
Enter a question mark (?) to view help for any given prompt in the Service Script.

4.2.3 Configuration Menu


Fig. 4: Configuration Menu

DICOM MPPS + SC AX41-000.814.01.02.02 Page 16 of 29 © Siemens, 2001


08.01 TD SD 21
ExamWorks+ Setup 4 17

n In the Configuration menu, select <1> ↵ for Perform Full Configuration.

4.2.4 Application Language Configuration


n Select the desired language.

4.2.5 Network Configuration


Fig. 5: Network configurator

4.2.5.1 Basic Host information


n Im Network Configurator unter Client tasks <Basic host information> ↵ auswählen.

© Siemens, 2001 AX41-000.814.01.02.02 Page 17 of 29 DICOM MPPS + SC


08.01 TD SD 21
18 4 ExamWorks+ Setup

Fig. 6: Basic host configuration

n Enter the Host name ( Configuration Data: ExamWorks+ / Page 12).


n Set Adaptor 1 to Enabled. An X in the bracket [] field to the left of Enabled indicates
that the adaptor is enabled.
n Set Config mode to Manual. An o in the bracket (o) field indicates that the Config
mode is Manual.
n Enter the Primary Name + Domain ( Configuration Data: ExamWorks+ / Page 12).
The Primary Name and the Host name must be the same (e.g.: acomtdf77.med.sie‐
mens.de).
n Enter an Aliases name (Option)
n Enter IP adress in the IP address field ( Configuration Data: ExamWorks+ / Page 12).
n Enter a netmask in the Netmask field ( Configuration Data: ExamWorks+ / Page 12).
n The Net Device field is pre-configured. Do not change this field.
n Select <Accept> ↵ to complete the settings.
- Use the keyboard <Tab> key to move among the fields.

4.2.5.2 Name Server Specification (Optional)


This fields needs to be filled in only if a DNS server is configured.

DICOM MPPS + SC AX41-000.814.01.02.02 Page 18 of 29 © Siemens, 2001


08.01 TD SD 21
ExamWorks+ Setup 4 19

Fig. 7: Name server specification

n DNS usage, if a DNS server is available, enter an X in the bracket [] field.


n Enter IP address of name server 1.
n Enter the IP address of name server 2 (optional).
n Enter the IP Adress of name server 3 (optional).

4.2.5.3 Routing and Gateways


n Under Client Tasks, select <Routing and Gatways> ↵ in the Configurator.
n Select <set Defaults> ↵
Fig. 8: Default gateway

n Enter in the field “Default gateway” the IP number of the main gateway ( Configura‐
tion Data: ExamWorks+ / Page 12).

© Siemens, 2001 AX41-000.814.01.02.02 Page 19 of 29 DICOM MPPS + SC


08.01 TD SD 21
20 4 ExamWorks+ Setup

n Leave the field of Enable routing disabled.


n Select <Accept> ↵ to complete the settings.
- Use the keyboard <Tab> key to move among the fields.
n Select <Quit> ↵ to complete the settings

4.2.5.4 Host name search path


n In the Network Configurator, select <Host name search path> under client tasks.
Fig. 9: Name service access

n Enter an <X> ↵ in the bracket [] field of Multiple IPs for one Host.
n Select <Accept> ↵ to complete the settings.
- Use the keyboard <Tab> key to move among the fields.
n Leave the Network Configurator with <Quit> ↵.
n Activate the changes with <Quit> ↵.

4.2.6 Config in the System Configuration Time and Date


n Time Zone Selection.
Select your time Zone.
n Enter <1> ↵ for accept changes in System configuration.

DICOM MPPS + SC AX41-000.814.01.02.02 Page 20 of 29 © Siemens, 2001


08.01 TD SD 21
ExamWorks+ Setup 4 21

4.2.7 Input Configuration


n Enter the ExamWorks+ Local Storage SCP AE Title ( Configuration Data: Storage SCP
ExamWorks+ / Page 12).
n Enter the ExamWorks+Storage Listening Port Number ( Configuration Data: Storage
SCP ExamWorks+ / Page 12).
n Enter <1> ↵ for accept changes in System configuration.

4.2.8 Workflow Configuration


n Siemens Modality Model.
- Enter <1> ↵ for AXIOM Artis Systems.
- Enter <2> ↵ for Other Siemens System.

4.2.8.1 Modality Worklist Server Information (MWL)


n Enter the ExamWorks Local MWL AE title ( Configuration Data: Modality Worklist
Server Information / Page 12).
n Enter the MWL Server AE title ( Configuration Data: Modality Worklist Server Infor‐
mation / Page 12).
n Enter MWL Server IP address ( Configuration Data: Modality Worklist Server Informa‐
tion / Page 12).
n Enter MWL Server Port Number ( Configuration Data: Modality Worklist Server Infor‐
mation / Page 12).
n Modality type to be provide by MWL. Enter <2> ↵ for no.
n Patient/Procedure to be schedule by Specific Modality System? Enter <1> ↵ for yes.
n Enter the Local AE title for the Modality Unit (Image System) ( Configuration Data:
Modality Worklist Server Information / Page 12).
n DICOM field to display in column 1 of the Procedure In-Progress Table. <1> ↵ for Pa‐
tient Name.
n DICOM field to display in column 2 of the Procedure In-Progress Table. <4> ↵ for
Study Date.
n DICOM field to display in column 1 of the MWL Information Table. <1> ↵ for Patient
Name.
n DICOM field to display in column 2 of the MWL Information Table. <6> ↵ for Sched‐
uled Procedure Step ID.
n DICOM field to display in column 3 of the MWL Information Table. <5> ↵ for Sched‐
uled Start Time.
n DICOM field to display in column 4 of the MWL Information Table. <7> ↵ Scheduled
Procedure Description.
n MWL Manual Entry Allowed: <1> ↵ for Yes.

4.2.8.2 Modality Performed Procedure Step (MPPS)


n Enter ExamWorks+ MPPS Local AE Title ( Configuration Data: MPPS Informa‐
tion / Page 13).

© Siemens, 2001 AX41-000.814.01.02.02 Page 21 of 29 DICOM MPPS + SC


08.01 TD SD 21
22 4 ExamWorks+ Setup

n Enter MPPS Server AE title (Called AE Title) ( Configuration Data: MPPS Informa‐
tion / Page 13).
n Enter MPPS Server IP address ( Configuration Data: MPPS Information / Page 13).
n MPPS Server Port Number ( Configuration Data: MPPS Information / Page 13).
n Performing physician. <2> ↵ for No.
n Performing technologist. <2> ↵ for No.
n Protocol name allowed to be edited on ExamWorks. <2> ↵ for No.
n AXIOM Artis and Systems with FLUOROSPOTT.O.P. <2> ↵ for No; Systems with POLY‐
TRON T.O.P.<1> for Yes.
n Dose Data. <1> ↵ for Yes.
n Study ID provided by Modality. <1> for Yes.
n Accept changes in Workflow Configuration? <1> ↵ for Yes.

4.2.9 Output Configuration

4.2.9.1 Storage Commitment (SC)


n Configure with Storage Commitment? <1> ↵ for Yes.

 CAUTION
StorageCommitment Service shall only be configured if the destination really sup‐
ports it.
 A wrong configuration will lead to a malfunction of the ExamWorks+ which
can be solved only by MERGE service. Contact with Merge should be made on‐
ly through HSC.

n Enter ExamWorks+ Storage Commitment AE title (Local AE Title) ( Configuration Da‐


ta: Storage Commitment Information / Page 13).
n Enter ExamWorks+ Storage Commitment Port Number (Local Listening Port) ( Con‐
figuration Data: Storage Commitment Information / Page 13).
n Currently Configured Targets. <n> ↵ for new target.
Enter all targets to which the ExamWorks+ sends images or other data to.
n New target name as displayed to the user ( Configuration Data: Storage Target Infor‐
mation / Page 13).
n Enter Target AE title ( Configuration Data: Storage Target Information / Page 13).
n Enter Target IP address ( Configuration Data: MPPS Information / Page 13).
n Enter Target Port Number ( Configuration Data: MPPS Information / Page 13).

DICOM MPPS + SC AX41-000.814.01.02.02 Page 22 of 29 © Siemens, 2001


08.01 TD SD 21
ExamWorks+ Setup 4 23

n Target Service Selection


1. Storage
2. Storage with Storage Commitment.

If Storage Commit called AE Title und Target AE Title are the same ( Configuration Data:


Storage Target Information / Page 13), the following configuration of the Storage Com‐
mit Target
does not have to be performed.

n Enter AE Title of Storage Commitment Device (Storage Commit called AE title) ( Con‐
figuration Data: Storage Target Information / Page 13).
n Enter Target IP Number of Storage Commitment Device (Storage Commit called IP
adress) ( Configuration Data: Storage Target Information / Page 13).
n Enter Port Number of Storage Commitment Device (Storage Commit called AE Title)
( Configuration Data: Storage Target Information / Page 13).
n Default on this target being selected on startup.
1. Yes (the target selection is automatically selected)
2. No (images can be sent after manual selecting to the selected target)
n Configure the remaining targets in the same way.
n Exit Output Configuration with <quit> ↵ .
n <1> ↵ for accept
n Exit the Configuration Menu with <quit> ↵.

4.2.10 Reboot ExamWorks+


n In Main Menu for Exam Works Service Interface ( Fig. 3 Page 16) enter <6> ↵ Re‐
boot.
n Enter Password: m0rg01 ↵

4.2.11 Configure Time and Date

4.2.11.1 Accessing the Unix Command Prompt


n Log in to ExamWorks+ through the service login ( Service Login / Page 15).
n Select menu item <0 Exit to Command Prompt> ↵ , from the Main Menu for Exam‐
Works+ Service Interface.
n Press <Enter> ↵ to access the shell prompt.
n Enter <su> ↵
n Enter Password: m0rg01 ↵

© Siemens, 2001 AX41-000.814.01.02.02 Page 23 of 29 DICOM MPPS + SC


08.01 TD SD 21
24 4 ExamWorks+ Setup

4.2.11.2 Set Time and Date


The LINUX date command sets the time and date. For more information on the LINUX
date command, enter <man date> ↵ at a LINUX shell prompt.
Display or set the date UCT (universal) time. At command prompt.
n Enter <quit> ↵
n Enter <date MMddhhmmyyyy.ss> ↵

MM = Numeric month, a number from 1 to 12


dd = Day, a number from 01 to 31
hh = Hour, a number from 00 to 23
mm = Minutes, a number from 00 to 59
yyyy = Year, such as 2001
.ss = Seconds, a number from 00 to 61 (59 plus a maximum of two leap sec‐
onds)
n Exit the Time and Date configuration with <quit> ↵ and <exit> ↵.

4.2.11.3 Reboot ExamWorks+


n I the Main Menu for ExamWorks Service Interface ( Fig. 3 Page 16) enter <6> ↵ Re‐
boot.
n Enter the password: m0rg01 ↵

4.2.12 Save the Configuration


n Enter the Configuration menu ( Service Login / Page 15).
n Im Configuration menu, select <8> ↵ für Perform Configuration Backup.
n Insert an empty MS-DOS formatted floppy disk into the floppy drive. It is shipped
along with the system.
n Press enter to backup the configuration file.
n Store the backup configuration file in a safe place.
n Reboot the ExamWorks+.

 You must reboot the ExamWorks+ after making changes to the configuration. Rebooting
the ExamWorks+ ensures proper memory allocation and information distribution.

4.2.13 ExamWorks+ Shutdown from GUI (Graphic User Interface)


This section explains how to shut down ExamWorks+ from the GUI.

DICOM MPPS + SC AX41-000.814.01.02.02 Page 24 of 29 © Siemens, 2001


08.01 TD SD 21
ExamWorks+ Setup 4 25

 CAUTION
Failure to follow shutdown procedures may result in lost or corrupted data and
files.
 Follow these shutdown procedures exactly.

n Select <System Shutdown> ↵ from the Utils Menu.


n Wait for the System confirm that it is safe to turn off ExamWorks+ power.

© Siemens, 2001 AX41-000.814.01.02.02 Page 25 of 29 DICOM MPPS + SC


08.01 TD SD 21
26 5 Final Work Steps

5.1 Function Test


HIS/RIS System
n Create a test patient in the HIS/RIS system.
AX System
n Request the patient list from the HIS/RIS system.
n Select a test patient and make an exposure.
n Send the images to the Examworks+ for archiving.
ExamWorks+

 The name of the buttons depends on the application language selected in the configura‐
tion.

n Select <Acquire> on the touchscreen monitor for the workstation.


n Select the test patient.
n The current worklist will be acquired from the HIS/RIS system by pressing the <done>
button.
n Select <Target>, the archiving destination must be online.
n Select <Acquire>.
n Select <check worklist>.
- The patient will appear under “Procedures in Progress”.
- The status of the patient must be: “Matches found”.
n Press <Use Exist>, the test patient will appear in the “Modality Worklist Information”
field.
n Enter the dose information.
n Press <completed>.
- The images will be sent to the selected target.
- The MPPS will be sent to the HIS/RIS.
n There will be no acknowledgement, and the Examworks+ is now ready for a new
transfer.

DICOM MPPS + SC AX41-000.814.01.02.02 Page 26 of 29 © Siemens, 2001


08.01 TD SD 21
Changes to Previous Version 6 27

all Chapters: Document expanded to include Axiom Artis and Polytron T.O.P.
all Chapters: Rewritten
Chapter 5: Added

© Siemens, 2001 AX41-000.814.01.02.02 Page 27 of 29 DICOM MPPS + SC


08.01 TD SD 21
28 7 List of Hazard IDs

There are no Hazard IDs in this document.

DICOM MPPS + SC AX41-000.814.01.02.02 Page 28 of 29 © Siemens, 2001


08.01 TD SD 21
All documents may only be used for
rendering services on Siemens
Healthcare Products. Any document
in electronic form may be printed
once. Copy and distribution of elec‐
tronic documents and hardcopies is
prohibited. Offenders will be liable
for damages. All other rights are re‐
served.

healthcare.siemens.com/services

Siemens Healthineers Headquarters


Siemens Healthcare GmbH
Henkestr. 127
91052 Erlangen
Germany
Telephone: +49 9131 84-0
siemens.com/healthineers

Print No.: AX41-000.814.01.02.02 | Replaces: AX41-000.814.01.01.02


Doc. Gen. Date: 08.01 | Language: English
Part No.: 5916791
© Siemens Healthcare GmbH, 2001

siemens.com/healthineers