MRC 200 0407 Rot-Gs 1004 (4512 988 00136 Rev Aa)
MRC 200 0407 Rot-Gs 1004 (4512 988 00136 Rev Aa)
Internet: www.medical.philips.com
All rights reserved. Philips Medical Systems reserves the right to make changes in specifications or to
discontinue any product, at any time without notice or obligation, and will not be liable for any
consequences resulting from the use of this publication.
Reproduction in whole or in part is prohibited without the prior written consent of the copyright owner.
The English Service Information is the original version. Published in Germany.
MRC 200 0407 ROT-GS 1004
SERVICE INFORMATION
CONTENTS
INSTRUCTIONS
CONTENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
MANUAL HISTORY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1 BRIEF DESCRIPTION OF THE X-RAY TUBE ASSEMBLY. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2 SAFETY INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2 Accompanying documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.3 General safety precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.4 Safety signs on the X-ray tube assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.5 Cooling / Insulating medium . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.6 Load limit values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.7 Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.8 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.9 Safety messages and other messages used in this manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.9.1 Safety messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.9.2 Other messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3 COMPATIBILITY. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4 TECHNICAL INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.1 Denotation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.2 Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.3 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.4 Filtration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.5 Grid switch GS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.6 Excess temperature switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.7 Cooling unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
5 INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
5.1 Mechanical installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
5.1.1 Installation of the X-ray tube assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
5.1.2 Cooling unit and cooling hoses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
5.1.3 Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.1.4 Fine-adjustment of the X-ray tube assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5.2 Electrical installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5.2.1 Connection of the protective earth conductor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5.2.2 Connection of the high-voltage cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.2.3 Connection of the stator cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
5.2.4 Connection of the X-ray tube can . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.2.5 Connection of the excess temperature switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.2.6 Connection of the grid switch cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
6 SETTING-TO-WORK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
6.1 X-ray tube conditioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
7 MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
7.1 Corrective maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
7.1.1 Exchange of the X-ray tube assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
7.1.2 Exchange of the cooling unit and the cooling hoses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
7.2 Planned maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
7.2.1 Maintenance by a service organization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
7.2.2 Safety checks that refer to national and international standards and to local regulations . . . . . . . . . 42
7.2.3 Cleaning and disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
TECHNICAL DATA
MANUAL HISTORY
This service information manual was prepared by the BU GTC, Service Innovation Department, Hamburg,
Germany.
For any additions, corrections or suggestions please contact your local service organization.
1 2 3 4
CMT_005
Figure 1: X-ray tube assembly
X-ray tube
See 5. Cathode
figure 2 8
6. Metal housing of the X-ray tube
7. Anode disk
8. Rotor
5 6 7
CMT_006
Figure 2: X-ray tube
The high performance metal ceramic X ray tube MRC 200 0407 is especially suitable for
invasive vascular/angiographical applications.
The design of this X-ray tube is based on state of the art X-ray technology. The most rele-
vant design feature is the support of the rotary anode in a spiral groove bearing.
This spiral groove bearing of the anode guarantees an optimum lifetime of the X rays tube
even under high continuous load.
For pulsed fluoroscopy, this X-ray tube assembly offers the most up-to-date X-ray pulsa-
tion technology.
The MRC 200 0407 X-ray tube is installed in the X-ray tube assembly:
MRC 200 0407 ROT-GS 1004
2 SAFETY INFORMATION
IMPORTANT
The conditions for safe operation and environmental conditions are part of the accompany-
ing documents of the X-ray system.
– X-ray tube assemblies are sensitive components! Protect them against mechanical
deformation!
– Protect yourself and others from leakage radiation or scattered radiation! Do not
expose yourself or others directly to the X-ray beam!
– The X-ray tube assembly must not be operated along with combustible anesthesia sub-
stances!
– The oxygen content of the ambient air during operation must be less than 25 %!
– Be careful with falling masses as they can cause bruises, lacerations, etc.!
– Use appropriate lifting devices when you have to lift heavy loads or ask another person
for help!
– Use the carrying handles or auxiliary lifting devices on the X-ray tube if available!
– If necessary, wear appropriate protective clothing, such as safety boots, safety gog-
gles, and gloves!
– When you use cleaning agents such as detergents and disinfectants, make sure that
they do not contain explosive substances as they can create explosive gas mixtures!
– Before you do service procedures, make sure that the temperature of the X-ray tube
housing surfaces decreases to a comfortable value!
– Be careful with sharp edges that some components can have!
– Do all relevant safety checks before you hand the X-ray system over to the customer!
– Never let problems unsolved that can affect the safety of the X-ray system!
Attached This safety sign indicates that the X-ray tube contains the toxic material beryllium.
safety sign
CMT_129/130
Attached This safety sign shows that the oil expansion bellow contains a medium that causes an as-
safety sign piration hazard, Category 1.
The label refers to “Globally Harmonized System of Classification and Labeling of Chemi-
cals (GHS)” with the hazard statements (H) and the precautionary statements (P).
CMT_261
This safety sign in combination with the signal word DANGER includes the following haz-
ard and precautionary statements:
– H304 May be fatal if swallowed and enters airways.
– P331 Do NOT induce vomiting.
– P301 + P310 IF SWALLOWED: Immediately call a POISON CENTER
or doctor/physician.
– P405 Store locked up.
– P501 Dispose of contents/container to an approved waste disposal plant.
Hazardous components:
– Contains Distillates (Fischer - Tropsch), heavy, C18-50 - branched, cyclic and linear.
To prevent an aspiration hazard obey the instructions of the chapter “Cooling / Insulating
medium”.
2.7 Tools
– Use tools and measurement devices, which are suitable for the respective service pro-
cedure!
– Use test and adjustment devices which are calibrated at regular intervals!
– Do not use auxiliary materials past their expiration date!
Safety messages within single steps of a service procedure occur without a preceding sig-
nal panel. The single signal word precedes these steps.
Other messages within single steps of a service procedure occur without a preceding mes-
sage symbol. The single message word precedes these steps.
3 COMPATIBILITY
The X-ray tube assembly MRC 200 0407 ROT-GS 1004 is compatible with:
– the X-ray generators: Velara CV, Velara CVFD, Certeray iX
– the cooling unit: CU 3101
4 TECHNICAL INFORMATION
4.1 Denotation
X-ray tube assembly MRC 200 0407 ROT-GS 1004 means:
MRC Maximus ROTALIX Ceramic X-ray tube,
High continuous load
200 Diameter of the anode disk [mm]
04 Nominal focal spot value, small focus
07 Nominal focal spot value, large focus
ROT-GS 1004 X-ray tube of Philips family 1004, oil cooled
GS Grid Switch, pulsation technology
4.2 Identification
4.4 Filtration
The permanent filtration of the X-ray tube assembly is 2.5 mm Al/75 kV IEC 60522/1999.
An additional filter plate of 1.7 mm Al is part of permanent filtration.
The identification label of the X-ray tube assembly shows the permanent filtration value.
The excess temperature switch of the X-ray tube assembly has a double safety switch.
Two switches are connected in series.
It is activated by oil temperatures in the X-ray tube housing at about 85 °C.
This excess temperature switch interrupts the READY state of the system.
5 INSTALLATION
See
figure 3
20 ± 2 Nm
Expansion bellow
Hazard of aspiration.
• Do not swallow media that spills!
• Do not let media or its vapor to go into your sewerage system!
• Do not breathe the media vapors!
• Do not cause vomiting!
• Make sure to supply closed rooms with fresh air!
If you do not obey this instruction, there is a risk of death or serious injury.
Hazard of aspiration.
• Do not swallow media that spills!
• Do not let media or its vapor to go into your sewerage system!
• Do not breathe the media vapors!
• Do not cause vomiting!
• Make sure to supply closed rooms with fresh air!
If you do not obey this instruction, there is a risk of death or serious injury.
See 2 = Cooling unit 9. Check the oil level in the cooling unit with
figure 5 3 = Cooling hoses the oil measuring stick.
4 = Hose couplings, The oil level must be between the mini-
tube assembly end mum and maximum mark. If necessary,
5 = Hose couplings, refill some oil
cooling unit end .
- +
1. Connect the cooling hoses (3) to the
cooling unit (2).
ROT
2. Interconnect the hose couplings at
the tube assembly end (4).
3. Switch the cooling unit ON. Let the
pump of the cooling unit run evenly.
4
4. During pump operation fill about 6 l
of oil into the cooling unit.
Use a funnel. 3
5. Let the oil circulate in the cooling cir- 3
cuit 20 min. 5
6. Check the oil level in the cooling unit
with the oil measuring stick while the
pump is operating.
The oil level must be between the 2 CU 3xxx
minimum and maximum mark.
7. Deaerate the cooling unit.
CMT_093
8. Switch the cooling unit OFF.
Figure 5: Connection of cooling hoses
1. Clean the filter in the return line of the cooling unit with benzine or alcohol.
2. Reconnect hose couplings at the cooling unit end (5) to the cooling unit (2).
3. Switch the cooling unit ON.
4. Let the oil circulate in the cooling circuit 10 min.
5. Visually check the complete cooling circuit for any leakages.
This means cooling unit, cooling hoses, tube assembly and all couplings.
6. Switch the cooling unit OFF.
1 20 s Switch ON
Horizontally
2 20 s Switch OFF
3 20 s Switch ON
Lift at cathode end 45°
4 20 s Switch OFF
Table 1: De-aerating procedure
For the handling of cooling unit and cooling hoses in case of an exchange of a tube as-
sembly refer to chapter 7.1.1 " Exchange of the tube assembly".
5.1.3 Collimator
The universal spacing plates are located at the X-ray port of the collimator. The universal
spacing plates are already adjusted by the factory.
CMT_075/014
To install the coupling flange of the collimator to the X-ray window flange following fixing
screws are required:
Nature of screws = Countersunk screws
Minimum number = 4 pcs.
Thread = M6
Strength category = 8.8
Minimum
engagement length = 6 mm *
Maximum
engagement length = to be determined *
Tightening torque = 1130 Ncm
The fixing screws must be secured with LOCTITE #243!
For installation of the coupling flange of the collimator to the X-ray window flange the ap-
propriate screw length has to be determined.
2 XA
CMT_122
CMT_120
• Measure the remaining engagement length between the top of the fixing screw and
the top of the countersunk washer = EL.
In case the engagement length falls below the minimum or exceeds the maximum use oth-
er fixing screws. But make sure to maintain the minimum engagement length of 6 mm as
well as the maximum engagement length determined.
3 2
TIP: The collimator can be installed without
further alignment.
CMT_116
T
See The alignment ring (1) is already installed
figure 10 by the factory upon delivery. 1
• Remove the alignment ring and dis-
pose it.
It is a high-precision tool and cannot be
recycled!
CMT_066
Figure 10: Removal of alignment ring
To install the coupling flange of the collimator to the X-ray window flange following fixing
screws are required:
Nature of screws = Countersunk screws
Minimum number = 4 pcs.
Thread = M6
Strength category = 8.8
Minimum
engagement length = 6 mm *
Maximum
engagement length = to be determined *
Tightening torque = 1130 Ncm
The fixing screws must be secured with LOCTITE #243!
For installation of the coupling flange of the collimator to the X-ray window flange the ap-
propriate screw length has to be determined:
X2
CMT_121
2
3
CMT_119
• Disperse locking liquid LOCTITE #243 along the whole threaded shank of the fixing
screws.
• Install the coupling flange of the collimator to the X-ray window flange.
The alignment of the collimator is described in the installation instructions of the system.
XA 1401
CMT_102
Risk of a damage of the X-ray tube assembly due to burned and damaged contact
surfaces.
• Use new silicon disks each time you connect or connect back a HV plug to a HV socket.
• Use new silicon disks each time you install and replace an X-ray tube assembly.
• Do not use silicon disks again even after you clean and grease the silicon disks again.
If you do not obey this instruction, there is a risk of property damage.
IMPORTANT
Silicone paste is not a good insulator.
Use the silicone paste to
- make the HV plug installation into the HV socket easy.
- keep the entrapped air to a minimum.
CMT 075
Figure 15: Silicone paste and two silicone disks
CMT 077
Figure 17: Pin gage to adjust a deformed pin
CMT 079
Figure 19: Silicone plate on the HV plug
CMT 081
Figure 20: Application of silicone paste
CMT 079
Figure 21: Align the feather key and the keyway
1 2 3 4 5 (5 +1 Nm)
CMT_099/100
See 9 = Screen
figure 23 10 = Stator cable with screen Generator XA
11 = Earth clamp
1101 (U)
• Connect the stator cable (10): 1102 (V)
1103 (W)
Generator --- X-ray tube
assembly
XA1101/
9
1102/
1103
1101 (U)
• Earth the screen (9) with the earth 1102 (V)
clamp (11).
1103 (W)
10 11 CMT_095
49 10%
1 U XA 1101 U - V:
225 mH 10%
49 10%
MRC 2 V XA 1102 V - W:
225 mH 10%
49 10%
3 W XA 1103 U - W:
225 mH 10%
Table 2: Stator connection
CMT_098
XA 1201
XA 1202
CMT_104/108
XA 1300 CMT_103
6 SETTING-TO-WORK
Setting-to-work includes:
– Software and hardware programming for the X-ray tube type.
– Adaptation of the filament.
– Conditioning of the X-ray tube.
• Do the conditioning procedure for each new connected X-ray tube assembly.
• Obey the intervals between exposures.
• Use a watch to monitor the intervals.
• Do the conditioning procedure with the large focal spot only.
The first exposure of each day does not require warming up.
For operating breaks of more than 3 months, an exposure-series have to be made with the
specified kV and mAs values as given in the table below.
Every exposure sequence must always be performed at the respective selected kV level
without repeating the start-up procedure of the X-ray tube. The PREP button must remain
to be pressed during the whole exposure sequence.
Conditioning procedure
• Switch ON the system.
• Select the large focal spot of the X-ray tube.
• Select 3-button-technique "kV-mA-s".
Choose kV-mA-s-values at the system corresponding the first exposure sequence.
See table below.
• Release exposures sequences. See table below.
If electrical interferences occur during the conditioning procedure, obey the following
steps:
• Continue the procedure after an interval of 5 minutes.
• Restart the procedure at a lower kV level.
Break time
between Break time
Number of ON-time
X-ray tube X-ray tube two expo- between
Exposure exposures per
voltage current sures two
sequence per exposure
[kV] [mA] within the sequences
sequence [s]
sequence [s]
[s]
1 6 81 250 1 1
20
2 4 102 250 1 1
20
3 3 109 250 1 1
20
4 2 125 250 1 1
60
Table 3: Conditioning of the X-ray tube
7 MAINTENANCE
TIP
Also refer to the service manual "Cooling Unit".
2 CU 3xxx
CMT_093
Figure 27: Connection of cooling hoses
1. Clean the filter in the return line of the cooling unit with benzine or alcohol.
2. Reconnect cooling hoses (3) to the cooling unit (2).
3. Switch the cooling unit ON.
4. Let the oil circulate in the cooling circuit 10 min.
5. Visually check the complete cooling circuit for any leakages.
This means cooling unit, cooling hoses, tube assembly and all couplings.
6. Switch the cooling unit OFF.
1 CMT_097
3. Deaerate the oil circuit five times in four steps according to the table:
1 20 s Switch ON
Horizontally
2 20 s Switch OFF
3 20 s Switch ON
Lift at cathode end 45°
4 20 s Switch OFF
Table 4: De-aerating procedure
For the handling of cooling unit and cooling hoses in case of an exchange refer to chapter
5.1.2 " Cooling unit and cooling hoses".
This X-ray tube assembly contains mechanical components which wear due to operation.
The correct setting of the electromechanical and electronic assemblies makes sure that
the function, image quality, and electrical safety is optimal.
The settings make sure that exposure of the patient and medical personnel to radiation is
at a minimum.
7.2.2 Safety checks that refer to national and international standards and to
local regulations
For the particular product other special technical tests that refer to the generally accepted
standards of engineering practice can be necessary.
Do the following cleaning and disinfection procedures once a year or when the X-ray tube
assembly is soiled:
Cleaning
Hazard of electrical shock or short-circuit caused by liquids inside the X-ray tube as-
sembly.
• Set the mains voltage of the X-ray system to OFF before you start the cleaning proce-
dure of the X-ray tube assembly.
• Do not use caustic, solvent, or abrasive detergents or polishes.
• Obey the manufacturers information and safety instructions of the cleaning solvent.
If you do not obey this instruction, there is a risk of death or serious injury.
Disinfection
The disinfection method used must conform to valid legal requirements and guidelines for
disinfection and protection against explosion.
Wipe disinfection
Disinfect all parts of the X-ray tube assembly, including accessories and connecting ca-
bles, by wiping only with a soft cloth.
Spray disinfection
Spray disinfection is not recommended because the disinfectant could get inside the X-ray
tube assembly.
Disinfection by atomizing
Cover the X-ray tube assembly carefully with a sheeting when it has cooled down. After
the disinfection mist has subsided, you may remove the sheeting and disinfect the X-ray
tube assembly by wiping with a soft cloth.
• Send the Tube Report via E-Mail to Hamburg, X-ray tube factory.
E-Mail address: [email protected]
When the X-ray tube assembly arrives in Hamburg and the Tube Report is completely
filled out the customer receives a credit note for the X-ray tube housing or warranty is is-
sued according to the warranty regulations arranged between BU Generators, Tubes &
Components and Customer Services.
IMPORTANT
Note to the housing credit:
If the OTR is not supplied or the data of the OTR are incomplete, the housing credit is re-
duced respectively!
IMPORTANT
• Do not discard the X-ray tube assembly, the attached
parts, the cable connectors, and the cables together with
industrial or domestic waste!
• Discard the X-ray tube assembly, the attached parts, the
cable connectors, and the cables in a way that refers to
the local environmental laws and regulations!
2.5 mm Al/75kV
7 Permanent filtration
IEC 60522/1999
HV cable sockets
13 O3
IEC 60526
U-V 49 ± 10 %
225 mH ± 10 %
----------------------------------------
V-W 49 ± 10 %
14 Stator data, phase/phase
225 mH ± 10 %
----------------------------------------
U-W 49 ± 10 %
225 mH ± 10 %
Safety classification
15 Class 1, type B
IEC 60601-1
Technical data of
X–ray tube assembly
MRC 200 0407 ROT-GS 1004 01/2016 TD-2.2
2006-2016 Koninklijke Philips N.V.
ALL RIGHTS RESERVED
Legend
Ie = Emission current
Uf = Filament voltage
If = Filament current
Uf / If = Characteristic filament curve
CMT TD 683
CMT TD 684
TD-4: Charts for single load rating Charts for single load rating
CMT TD 685
CMT TD 686
20 30 30 30 29 29 29 29 29 29 29
40 29 29 29 29 29 29 29 29 29 28
60 29 29 29 29 28 28 28 28 28 27
80 29 29 28 28 28 28 28 27 27 27
100 29 28 28 28 27 27 27 27 27 26
120 29 28 28 27 27 27 26 26 26 25
160 29 28 27 27 26 26 25 25 25 24
200 28 27 26 26 25 25 24 24 24 23
240 28 27 26 25 25 24 24 23 23 21
280 28 26 25 24 24 23 23 22 22 20
320 27 26 25 24 23 23 22 22 21 20
360 27 25 24 23 23 22 21 21 20 19
400 27 25 24 23 22 21 21 20 20 18
450 26 24 23 22 21 21 20 19 19 17
500 26 24 22 21 21 20 19 19 18 16
The tables give the maximum anode input power of each radiography in kW based on an equivalent
anode power of 250 W.
The load data refer to IEC 60613.
20 64 64 63 63 63 62 62 62 62 61
40 63 62 62 61 61 60 60 59 59 57
60 63 61 60 59 59 58 57 57 56 54
80 62 60 59 58 57 56 55 54 53 51
100 61 59 57 56 55 54 53 52 51 48
120 61 58 56 55 53 52 51 50 49 46
160 60 56 54 52 50 49 48 46 45 42
200 58 55 52 49 48 46 45 43 42 38
240 57 53 50 47 45 44 42 41 39 35
280 56 51 48 45 43 41 40 38 37 33
320 55 50 46 43 41 39 38 36 35 31
360 54 48 44 42 39 37 36 34 33 29
400 53 47 43 40 38 36 34 32 31 27
450 52 46 41 38 36 34 32 30 29 24
500 51 44 40 36 34 32 30 29 27 22
The tables give the maximum anode input power of each radiography in kW based on an equivalent
anode power of 250 W.
The load data refer to IEC 60613.
20 25 25 25 25 25 25 25 25 25 25
40 25 25 25 25 25 25 25 25 25 25
60 25 25 25 25 25 25 25 25 25 25
80 25 24 24 24 24 24 24 24 24 24
100 24 24 24 24 24 24 24 24 23 23
120 24 24 24 24 23 23 23 23 23 22
160 24 24 23 23 23 22 22 22 22 21
200 24 23 23 22 22 22 21 21 21 19
240 23 23 22 22 21 21 21 20 20 18
280 23 22 22 21 21 20 20 19 19 17
320 23 22 21 21 20 20 19 19 18 16
360 23 22 21 20 20 19 19 18 17 15
400 23 21 20 20 19 18 18 17 17 15
450 22 21 20 19 18 18 17 17 16 14
500 22 20 19 18 18 17 17 16 15 13
The tables give the maximum anode input power of each radiography in kW based on an equivalent
anode power of 1500 W.
The load data refer to IEC 60613.
20 53 53 53 53 53 53 53 53 53 53
40 53 53 53 53 52 52 52 52 52 51
60 53 52 51 51 51 50 50 50 49 47
80 52 51 50 50 49 49 48 47 47 44
100 52 50 49 48 48 47 46 45 44 41
120 51 49 48 47 46 45 44 43 42 39
160 50 48 46 45 44 42 41 40 39 35
200 49 46 44 43 41 40 39 37 36 31
240 48 45 43 41 39 38 36 35 33 29
280 47 44 41 39 37 36 34 33 31 25
320 47 42 39 37 35 34 32 31 29 22
360 46 41 38 36 34 32 31 29 28 20
400 45 40 37 34 32 31 29 28 25 18
450 44 39 35 33 31 29 28 25 22 16
500 43 37 34 31 29 28 25 22 20 14
The tables give the maximum anode input power of each radiography in kW based on an equivalent
anode power of 1500 W.
The load data refer to IEC 60613.
12 7 2 1 - - - - - - - - -
10 7 2 1 1 - - - - - - - -
8 7 2 1 1 - - - - - - - -
6 7 2 1 1 - - - - - - - -
4 7 2 1 1 - - - - - - - -
2 8 2 1 1 - - - - - - - -
1 9 2 1 1 - - - - - - - -
12 100 49 22 - - - - - - - - -
10 100 51 23 18 - - - - - - - -
8 100 54 23 18 15 - - - - - - -
6 100 63 24 19 15 12 - - - - - -
4 100 100 27 20 16 12 10 9 8 7 - -
2 100 100 53 29 22 14 12 10 9 7 7 6
12 100 100 54 - - - - - - - - -
10 100 100 56 43 - - - - - - - -
8 100 100 61 45 36 - - - - - - -
6 100 100 71 50 39 29 - - - - - -
The tables give the max. number of exposures permissible per series based on an equivalent
anode input power of 250 W. The load data refer to IEC 60613.
12 8 4 2 - - - - - - - - -
10 8 4 2 - - - - - - - - -
8 8 4 2 1 1 - - - - - - -
6 8 4 2 1 1 1 - - - - - -
4 8 4 2 1 1 1 - - - - - -
2 8 4 2 1 1 1 - - - - - -
1 8 4 2 1 1 1 - - - - - -
12 52 24 11 - - - - - - - - -
10 53 24 11 8 - - - - - - - -
8 56 24 11 8 7 - - - - - - -
6 62 26 11 8 7 5 - - - - - -
4 79 28 11 9 7 5 5 4 3 3 - -
2 100 43 13 9 8 6 5 4 4 3 3 3
1 100 100 20 13 9 7 5 4 4 3 3 3
12 100 60 27 - - - - - - - - -
10 100 63 28 22 - - - - - - - -
8 100 67 28 22 18 - - - - - - -
6 100 77 30 23 18 14 - - - - - -
4 100 100 33 25 20 15 13 11 9 8 - -
2 100 100 58 35 25 17 14 12 10 9 8 7
The tables give the max. number of exposures permissible per series based on an equivalent
anode input power of 250 W. The load data refer to IEC 60613.
Label of
protective earth
terminal
See accompanying
ring binder.
CMT TD 419
155.9
to 20 2 Nm
6 x M10
180
264
Reference axis
90
132
358
$#% !&
%
! ""#
Reference axis
600
Beam angle concerning the
maximum X-ray field.
300
64 2
+
CMT TD 420