Neal12MRes - Tee Mixer Homogeneous Mixing
Neal12MRes - Tee Mixer Homogeneous Mixing
by
GRACE NEAL
A thesis submitted to
MASTERS OF RESEARCH
University of Birmingham Research Archive
e-theses repository
By adapting the techniques Planar Laser Induced fluorescence (PLIF) and Particle
Image Velocimetry (PIV) for a transient system with a rising free surface the mixing
The PLIF technique has been demonstrated to give good repeatability for the same
flow conditions and qualitative examination of the images has revealed a change in
the mixing mechanism from laminar to turbulent above a critical value of the
Reynolds number (Re~1000), with cases above this reaching a log variance of -2
which equates to 92% mixed. PIV allowed the distribution of momentum and shear
in the tank to be investigated, highlighting dead regions in the bottom corners of the
tank and showing the top half of the tank was relatively quiescent. To improve the
mixing performance, changes to the flow rate and the nozzle design were made to
Sinusoidal variations in the inlet velocity increased the level of mixing from an
unmixed state so that the vessel reached 84% mixed. With a swirl insert in the end of
the nozzle the level of mixing in the tank reached 95% (log variance of -2.6) at one
point in the fill. Combining these improvements may produce a more sustained and
Firstly I would like to thank my supervisor Dr Mark Simmons for his support and
guidance throughout the last 6 years. I'd also like to acknowledge Dr Andreas
Tsoligkas for teaching me to use the PIV and PLIF equipment. Thank you also to
Thank you to all the technical staff who contributed to the building of my
Green and Andrew Tanner from the Biosciences Engineering Workshop. Thanks go
Thanks go to my family and friends, for picking me up when I'd lost hope. To
everyone at Unilever who made me feel so welcome and kept me sane with endless
cups of green tea. Thanks also to my friends at the University for the hours of
Most of all thank you to the EPSRC and to Unilever for funding this research.
Contents
system ......................................................................................25
3.2.2 Moving nozzle set up and control ................................ 26
References ............................................................................................................. 78
Figure 2.1 Illustration of the three different scenarios described by Pan and Meng
(2001) ……………………………………………………………….…. 11
Figure 2.2 Illustration of the regions of an impinging jet, Reungoat et al. (2001) ... 13
Figure 2.3 (Unger and Muzzio 1999) Geometry of the impinging jet mixers ….…. 19
Figure 3.1 Schematic of the experimental rig and PLIF set up ……………………. 26
Figure 3.3 Schematic of a straight pipe and the three nozzles improvements …….. 30
Figure 3.4 Plot of greyscale as a function of tracer concentration at two different laser
intensities ………………………………………………………............ 33
Figure 3.5 Calculation of elements of the matrices M and K from the linear relation
between greyscale and tracer concentration for each pixel, (only two
Figure 3.6 Definition of the location of the fluid surface via the change in the average
zones are shown and also the maximum fill height …............................. 36
Figure 4.1 Evolving concentration fields at different fill heights for the 6 mPa s fluid
at (a) h = 2.2 cm; (b) h = 4.7 cm; (c) h = 7.3 cm; (d) h = 10.3 cm at
Figure 4.2 Evolving concentration fields at different fill heights for the 3 mPa s fluid
at (a) h = 2.2 cm; (b) h = 4.7 cm; (c) h = 7.3 cm; (d) h = 10.3 cm at
Q = 1.67× 10-5 m3 s-1 (a) log variance plotted against dimensionless fluid
height (H/T); Image of the vessel filled with fluid at (b) Re=343
(12 mPa s), (c) Re=661 (6 mPa s), (d) Re=1245 (3 mPa s) ……………. 46
Figure 4.4 Mixing performance for three different fluid viscosities at a fill rate of
Q = 3.33× 10-5 m3 s-1 (a) log variance plotted against dimensionless fluid
height (H/T); Image of the vessel filled with fluid at (b) Re=687
(12 mPa s), (c) Re=1322 (6 mPa s), (d) Re=2490 (3 mPa s) ………..... 48
Figure 4.5 Mixing performance for three different fluid viscosities at a fill rate of
Q = 5.00× 10-5 m3 s-1 (a) log variance plotted against dimensionless fluid
height (H/T); Image of the vessel filled with fluid at (b) Re=1030 (12
mPa s), (c) Re=1983 (6 mPa s), (d) Re=3735 (3 mPa s) ……………..... 50
Figure 4.6 The dimensionless height (h/T) of the centre of the vortex from the base of
the tank, measured for three Reynolds numbers at flow rate of Q = 10-5
m3 s-1…………….………………………………………………..…….. 51
Figure 4.7 The dimensionless distance of the centre (h/T) of the vortex from the edge
of the inlet pipe, measured for three Reynolds numbers at flow rate of Q =
Figure 4.8 Mixing performance for three different fluid viscosities at a fill rate of
Q = 5.00×10-5 m3 s-1 (a) log variance plotted against fluid height; Image
of the vessel filled with the (b) 16 mPa s fluid, (c) 8 mPa s fluid, (d) 4
Figure 4.9 Plot of the minimum log variance reached in each experiment versus the
fluid, Re = 1494 at (a) h = 2.2 cm; (b) h = 4.7 cm; (c) h = 7.3 cm; (d) h =
…………………………………………………………………..……… 58
Figure 4.11 Evolving concentration fields at different fill heights for an 8 mPa s
fluid, Re = 1494 where a bubble has formed at (a) h = 2.2 cm; (b) h = 4.7
cm; (c) h = 7.3 cm; (d) h = 10.3 cm at a flow rate Q = 3.33 × 10-5 m3 s-1
Figure 4.12 The log variance plot of two fills of the 8 mPa.s fluid at a flow rate
Q = 3.33 × 10-5 m3 s-1 where a bubble has formed under the pipe in one
Figure 5.1 An example of the velocity profile in a vessel showing the features
Figure 5.2 Jet impingement zone, shaded to indicate velocity magnitude ….….… 62
Figure 5.3. The change in normalised velocity along the jet centreline for the four
Figure 5.4. Vertical velocity profile across horizontal slices taken at heights h above
Figure 5.5. The centreline velocity profile for low Reynolds number fills when the
Figure 5.6. The centreline velocity profile for high Reynolds number fills when
Figure 5.7. Velocity magnitude contour plots for flow at Re = 517 and Re = 1911
…………………………………………………………………………. 67
Figure 5.8. The shear rate calculated from the velocity field measured at Re = 518
……………………………………………………………………...……68
Figure 6.1 Mixing performance for three nozzles at a viscosity of 18 m Pa s (a) log
variance plotted againsts fluid height (cm); Image of the vessel filled with
(b) constant flow rate of Q = 3.33×10-5 m3 s-1 (c) sinusoidal flow rate
varying between 1.67 x 10-5 m3 s-1 and 5 x 10-5 m3 s-1 (d) flow rate
Figure 6.2 Mixing performance for three nozzles at a fill rate of Q = 3.33×10-5 m3 s-1
(cm); Image of the vessel filled with (b) straight pipe (c) cone 1 nozzle
Figure 6.3 Mixing performance for two nozzles at a fill rate of Q = 3.33×10-5 m3 s-1
(cm); Image of the vessel filled with (b) straight pipe (c) swirl nozzle
……………………………………………………………………..…… 73
List of Tables
ρd j u j
Re = .............................................................................................. 31
µ
Nomenclature
θ Mixing time
Gn Normalised greyscale
DM Molecular Diffusivity
K Intercept
M Gradient
Pixel size (Houcine et. al. 1996)
Re Reynolds number
T Tank width
D Tank Diameter
tD Diffusion time
ε Specific energy
μ: Dynamic viscosity
ν Kinematic viscosity
ρ Density
σ2 Variance
Li Impingement height
We Weber number
γ Shear
x Horizontal positon
y Vertical position
Manufacturers will often produce a range of very similar products. This allows the
increasing sales and expanding the market. There is usually a choice of colour,
fragrance or flavour within each group of products so the consumer has a full range of
options to cover their needs. This increased diversity has been driven by the
The shelf space available in retail outlets for each type of product has stayed constant
one product, retailers now only display a small quantity of each variety. Storing large
amounts of stock is expensive and so retailers want to order small amounts of a wide
range of products regularly. This in turn requires the manufacturer to either store
large amounts of each variety or to make smaller batch sizes, in reality both of these
have occurred.
production in large batches followed by filling into packaging. As the batch size
required has decreased the efficiency of the production decreases due to the increase
in the down time and cleaning which required between batches. The down time
increases costs due to the reduction in output, the increase in cleaning produces both
A further challenge to the industry comes in the form of regulation; the European 7th
fragrances above a specific low level. This is so the potential sensitizers which are
contained within the fragrance are controlled in each product. Since this legislation
was implemented in 2004, the level of cleaning between each batch has increased, this
in turn has caused an increase in the quantity of waste generated and hence the cost
Changes to the manufacturing methods that would allow the flexibility to be increased
would not only alleviate the increased waste and loss of efficiency, but would also
provide the flexibility to react quickly to changes in the market with new products.
To maximise the efficiency and flexibility of production, all products need to share as
much of the process as possible. Products are often made using the same base fluid
products. The aim is to find a method for mixing these variants into the base product
A technology that would allow the products to be mixed on the production line would
reduce the waste and improve efficiency. The increased flexibility would also give
amounts of stock could be released generating cash flow and also reducing storage
costs.
3
The two major technical challenges for achieving this late variant addition are the
accurate dosing of the variant and the effective mixing. In the case of personal care
products the variants particularly the fragrances are usually oil based liquids which
are being mixed into oil in water emulsions. The microstructure of the resulting
product is greatly affected by both the quantity of the variant and the length scale to
which it has been mixed. It is important that the microstructure of the product is
correct as it has a large effect on the efficacy, stability and appearance and therefore
customer satisfaction.
aerosol deodorants. Accurate weighing machines allow the dosage of the variants to
be accurately measured and controlled; it is then mixed during the gasification of the
aerosol. A similar dosing machine could be used for roll on deodorants however the
There are a number of mechanisms that could provide the level of mixing required for
agitation such as a stirrer, mixing in the filling pipe using a static mixer, shaking or
through the use of jet mixing. Of this list jet mixing would provide the most suitable
solution as a stirrer or static mixer would require cleaning and shaking the product
would require a large change to the filling lines which would involve a large capital
expenditure.
4
Extensive studies over many years have provided a very strong understanding of
submerged jets and geometry (e.g. the effect of liquid height). However what has
received less attention is the mixing that occurs as ingredients are added to vessels of
small length scale, where there is no mechanical agitation. The study of such
up effects account for a significant proportion of the mixing time. There can be large
disparities between the properties and ratio of the materials being added, the level of
the free surface changes and the entrainment of air makes visualization difficult.
The aim of this project is to understand the mixing processes that occur during jet
filling, this will characterise the range of products that can be produced in this
manner. This overall goal can be broken down into smaller objectives:
the level of mixing in the vessel to be measured throughout the filling process.
• Determine the parameters that set the limits for which mixing can be achieved
The lengthscales of the bottle to be filled are demonstrated in fig. 1.1, the flow rates
d = 0.6 cm
hmax = 10 cm
hinit= 1 cm
T = 6 cm
The first stage in this work was to investigate the available literature, this is included
standard straight inlet pipe while Chapter 5 goes on to examine the velocity field in
the tank. Chapter 6 investigates whether improvements can be made to the mixing by
using alternative pipe designs. Conclusions have been drawn in Chapter 7 along with
Work from this thesis has also been published in the archival Journal Experiments in
Fluids (Neal et al.,. 2008), and presented at both the International Symposium for
Mixing in Industrial Processes (ISMIP 2008) (Neal et al.,. 2008), and the World
2.1 Introduction
Although there is little published research into the mixing that occurs as a vessel is
filled there is a large body of work into mixing by a range of other mechanisms.
Principles from other mixing techniques and impinging jets can be applied to the
current problem and for this reason a brief outline of these are included here. Also of
importance to this project is the measurement techniques which have been used to
characterise the flow and mixing within the vessel: Particle Image Velocimetry (PIV)
2.2 Mixing
temperature in order to achieve the desired process result (Paul et al. 2004). The
quality of mixing achieved in a process can affect product quality from appearance
The mixing mechanism depends on the flow of the fluids being mixed, turbulent
flows where the velocity is constantly fluctuating reduces the scale of segregation
quickly down to the size of the smallest eddies while diffusion reduces the intensity of
the segregation (Kresta and Brodkey, 2004). Laminar flows do not contain the eddies
seen in turbulent flow and so there is little reorientation of the fluid particles; to create
flow provides the stretching and reorientation required to reduce the scale of
A large body of work has been done on a range of different mixing problems
especially the flow in stirred tanks; jets have been used for mixing in a range of
different situations from large storage tanks to two impinging jets on a smaller scale.
There are a number of ways that the degree of mixing in a vessel can be quantitatively
2
1 C −C
CoV =
N
∑ i C mean (2.1)
mean
concentration and N is the number of sample points. An alternative is to take the log
concentration and determine when it has decreased to an acceptable level. The time
taken to achieve this level of mixedness is called the mixing time. This has been
products from competing reactions (Brown et al.. 2004). Alternatively a method such
as PLIF can be applied which measures changes in the optical properties of the
system.
8
To allow the measurement of these properties a tracer is needed, this is added to the
flow and causes a change that can be monitored. In the case of conductivity a salt can
be used so areas of higher concentration have higher conductivity, for pH an acid will
fulfil the same role. These techniques are usually measured at single points; for
whole field measurements such as the optical technique PLIF a fluorescent tracer such
made from a greyscale image of the tank. This allows a large sample of concentration
values across a whole plane of the vessel, these concentrations have a distribution and
Houcine et al. (1996) and Fall et al. (2001) used a number of different statistics to
analyse the concentration distributions obtained for continuously fed vessels, where a
fluorescent tracer was continuously added into one feed. The first was the field of
reduced mean concentration ξAij, where the value at each pixel represents the
probability of the molecules of tracer being in the volume represented by that pixel.
This is calculated by dividing the mean concentration at the pixel Cij , obtained over a
Cij
ξA = (2.2)
ij
C A0
The second statistic used was the contacting parameter, similar to the reduced
minus itself: this allows the areas of different mixing performance to be more easily
ξ A B = ξ A ⋅ ξ B = ξ A (1 − ξ A )
ij ij ij ij ij
(2.3)
The third statistic was the temporal variance where the concentration at each pixel
was compared with the temporal average. Houcine et al. (1996) proposed that the
contacting parameter and the field of variance are the most important parameters to
characterize, since the former gives the average state of the mixing at each point and
the latter characterizes the mixing dynamics of the concentration fields. However
these fields are defined by the geometry of the system so a local mixing parameter
was defined as the ratio of the contacting parameter at each point to its maximum.
including the pixel size M and a sensitivity exponent k. An overall parameter more
sensitive to the non-homogeneous pixels was also defined and used to compare the
1
1 ξ AB k
ω = 1−
M ∑ (1 − ξ i
) k
(2.4)
i A BMAX
Fall et al. (2001) used a different mixing parameter to compare between systems: this
was based on the difference between the concentrations of two passive tracers at each
point.
C A − CB
α =1− (2.5)
C A + CB
Whilst these statistics have been designed to give a mixing criterion for continuous
by Hall et al. (2004) and Chung et al. (2006). They determined mixing times within
10
analysis. The variance in this case is calculated by comparing each measured value,
Cn, at a specific time with the concentration when the vessel is completely mixed, Cm,
these values are normalised using the concentration before the tracer is added, C0:
Cn' (t ) =
(Cn (t ) − C0 ) (2.6)
(Cm − C0 )
Feng et al. (2005) examined mixing in planar-jet reactors, also by measuring the
variance of the intensity across the image. This statistic is also used as a mixing
parameter in work by Mortensen et al. (2004) who examined a coaxial jet in a pipe
and by Pan and Meng (2001) whose work focused on the behaviour of a tee mixer.
While CoV statistics give a measure of the intensity of the mixing it fails to describe
the degree of segregation. To achieve this a measure of the structure within the
mixture is required. One way to achieve this is to measure the striation thickness, as
this decreases the scale of segregation has decreased, alternative methods such as that
vessel which is analogous to the rate of mass transfer, it is the rate of change of
segregation which they labelled the exposure, which they define as:
Nt Nb
1
E = ∑∑ Kaij (Ci − C j ) (2.7)
1 1 2
points, K is the strength of the interaction and a is the contact area per side, Ci-Cj is
11
particularly important when the problem is dominated by the mixing timescale, such
Jet-impaction regime
Jet-mixing regime
Wall-source regime
Figure 2.1 Illustration of the three different scenarios described by Pan and Meng (2001).
Pan and Meng (2001) identify three different scenarios when a jet enters a turbulent
crossflow; the jet-impaction regime, the jet-mixing regime and the wall-source
regime, these are demonstrated in Figure 2.1. In the jet-impaction regime the jet hits
the opposite wall and the large vortical structures are broken down improving the
12
mixing. The disadvantage of this is the stress it exerts on the opposite wall of the tee
mixer.
The jet-mixing regime is when the jet penetrates the primary flow and bends until it is
aligned. Turbulent entrainment creates macromixing in the pipe and the most
efficient mixing in this regime occurs when the jet stream is placed so it aligns in the
centre of the primary flow (Cozewith and Busko 1989). This flow has very similar
The wall-source regime is not efficient for mixing as the jet does not penetrate the
pipe flow, Forney and Lee (1982) found in these cases the mixing length in the pipe
This mixing length is dependent on the relative diameters of the two streams and the
ratio of their flowrates. They came up with the optimum relation for the volumetric
flow in the pipe (Q) and side stream (q) and their respective pipe diameters (D,d)
1.5
q d
= (2.8)
Q D
Pan and Meng (2001) used a pipe inlet where 6d=D, they tested two velocity ratios
v/V of 3.05 and 5.04, which give a flow rate ratio q/Q of 0.085 and 0.14 respectively.
According to equation 2.8, a centred jet would be produced with a flow rate ratio of
0.068. The lower inlet flow was closer to this ideal ratio and produced an
approximately centred jet, the higher flow rate was well above the ideal ratio which
caused the jet to impinge on the far side of the main pipe.
13
While these rules apply to turbulent flows, for laminar flow, mixing will rarely occur
in a tee mixer or co axial pipe inlet due to the lack of radial mixing, these flow
conditions require a static mixer to enforce this radial flow (Etchells and Meyer,
2004).
The structure of a turbulent jets impinging on a flat surface has been widely
investigated in the literature, largely due to their use in cooling systems (Chen et al.,
2005). A large amount of work has been carried out looking at planar or slot jets
(Law and Wang, 2000, Fondse et al., 1983, Kim et al. 2007) but Reungoat et al.
(2007) used PIV to measure the turbulent mixing in a round impinging jet at different
impinging distances. There are three main regions used when describing an
impinging jet; the free jet region, the impingement zone and the wall jet region, these
Figure 2.2 Illustration of the regions of an impinging jet, Reungoat et al. (2007).
14
The free jet region describes the jet between the injection point and where the
impingement begins to have an effect, in this zone the jet behaves as a free jet. Near
the point of injection, the momentum is maintained in the centre of the jet, this is
referred to as the potential core. It was observed by Reungoat et al. (2007) who
looked at the centreline velocity of the jet, the diameter of this core gradually
decreases as the mixing layer around it expands until about 6 nozzle diameters from
the injection where the jet moves into the developing zone. This agrees with Revill
(1985) who stated the same relation for a free jet in an unbound bulk liquid.
In the region of 6-50 nozzle diameters, Reungoat et al. (2007) found the centreline
velocity decreases linearly with distance, this is the developing zone and is in
agreement with previous work on slot jets such as Law and Wang (2000). Fondse et
al. (1983) describe this region as extending to 70 nozzle diameters away from the
injection point where the zone of established flow or self similar zone begins. Once
self similarity has occurred the cross sectional flow profile of the jet can be shown to
In the impinging region the jet is redirected from an axial into a radial direction, this
region is composed of a stagnation zone and then three wall zones (Chen et al. 2005).
The concentration of the tracer fluid in the Reungoat el al. (2007) investigation
appears to be high in the stagnation zone where the velocity of the fluid is low. They
also observe vortical structures that roll along the wall, these are four times larger
al. (1998) who used the difference between the 25% and 10% loci to determine the
mixing layer. This mixing layer increased linearly away from the inlet until it reached
a point 1.5 nozzle diameters from the impingement wall where the increase became
exponential; this curvature was also found by Reungout et al. The diffusion of the jet
is accelerated due to the deceleration of the jet caused by impingement on the wall.
Ding et al. (2003) found that the concentration core in the jet decays much quicker
than the velocity core of the jet, they attribute this to the faster transport of a scalar
quantity such as concentration than momentum, due to the pressure gradient in the
The studies above are all carried out on turbulent jets; Revill (1985) states that a jet is
fully turbulent at Re above or about 1000-2000 and laminar for Re below 100. Kim et
al. (2007) investigated water jets between Rej = 404 and Rej = 1026 and found that
the jet underwent a transition to unsteady within this range. Pawlak et al. (2007) look
at the dynamics of a starting laminar jet, the defining feature of these jets is the
leading vortex ring, which is followed by a secondary vortex, behind this leading
The effects of the boundary layer on the mixing performance of a jet has been
investigated, Ding et al. (2003) used PLIF with a local injection point at the edge of
the jet entry to look at the mixing layer, the concentration fluctuation at each point
was measured and it’s maximum coincided with the area of maximum shear stress.
Fondse et al. (1983) looked at the influence of the exit conditions on the entrainment
rate and found the most important factor was whether the boundary layer was laminar
16
or turbulent, with laminar jets entraining 15% more than turbulent jets. Placing a grid
across the outlet to promote turbulence in both the jet core and boundary layer
In all the current jet impingement studies the distance between the nozzle and the
impingement wall remained constant throughout the experiment. The fluid height in
the vessel was maintained using a side weir (Reungoat et al. 2007) or similar
recirculation and the vessel was large enough that the effects of the side walls on the
Jet mixing is also found in large storage tanks where fluid from an outlet is circulated
and jetted back in. Initial work on a simple system was carried out by Fossett and
Prosser (1949) who were investigating underground fuel storage tanks during World
War II. They proposed a relationship between the mixing time in the vessel, the
vessel diameter (T), the nozzle diameter (D) and the jet velocity (U):
T2
θ∝ (2.9)
UD
Although other relations have been suggested this was found to be the most accurate
The effect of the jet angle, the vessel geometry, symmetry and aspect ratio, the nozzle
diameter and the jet Reynolds number are all considered in the recent literature. One
of the underpinning papers in this field is Lane and Rice (1982) who compared three
different designs of jet mixing. The first was an inclined side entry jet near the base
17
of a flat bottomed cylinder, this design is refined to maximise the length of the jet, the
second design was an axial vertical jet in a flat based cylinder, and the third was an
axial vertical jet in a hemispherical based cylindrical tank. They found that the
hemispherical base is the best for mixing. This is attributed to reduced stagnant areas
in the bottom corners of the tank and that the liquid is constantly being redirected
Patwardhan and Gaikwad (2003) use a side entry cylindrical tank to investigate the
nozzle angle and diameter. They use the same power input at different angles and
find that the optimum mixing occurs when the jet is angled at 45 degrees to the base,
they believe this is due to maximising the length of the jet so it entrains more of the
surrounding liquid. The mixing is aided in this case by the outlet, which is situated in
the poorly mixed bottom corner opposite the jet. The nozzle diameter is also varied;
they found that increasing the diameter improves the mixing at the same power
consumption.
The effect of jet angle can be modelled using computational fluid dynamics (CFD).
Zughbi and Rakib (2004) used FLUENT with a tetrahedral mesh for their simulations,
they validated their results using the experimental results of Lane and Rice (1982).
They used the temperature rather than the concentration to measure the mixing time, it
was assumed that density and viscosity were constant over the temperature range used
so that the flow properties are unaffected. Their results showed that the optimum
angle is around 30 degrees to the base of the tank for entry at the base. This appears
to disagree with the theory presented originally by Lane and Rice that the optimum
A theory to explain the results was suggested in a following paper (Zughbi and
Ahmad 2005). It states that the jet length does need to be maximised but this does not
necessarily occur at 45 degrees. The jet may have dissipated before reaching the
opposite corner, however if the jet is deflected off the opposite wall a greater jet
length is achieved. This theory agrees with the evidence from tee mixer experiments
Studies have also been carried out on the mixing that occurs between two impinging
jets. The work by Unger and Muzzio (1999) investigates the two different geometries
for mixing jets shown in figure 2.3. Low Reynolds number jets (Re < 80) were better
mixed in the asymmetric geometry this was due to the swirling motion created, little
mixing occurred when these jets were fired directly at each other. For unsteady
laminar flows (80 < Re < 300) the symmetrical geometry created better mixing, this
was thought to be due to the oscillations in the jet streams which are more effective at
mixing than the swirling nature of the flow in the asymmetric case. High Reynolds
number jets were effectively mixed in both geometries although the asymmetric case
Figure 2.3 (Unger and Muzzio 1999) Geometry of the impinging jet mixers.
Mixing and flow diagnostic techniques can be either single point or whole field
concentration at a fixed point at a defined frequency in time, while the whole field
techniques measure over an entire area or volume at an instant in time, examples are
The techniques relevant to this work are the optical techniques of PLIF and PIV, this
2.4.1 PLIF
plane. To relate the intensity of the fluorescence of the tracer to its concentration it is
important that there is a linear relationship between them; this is the case for dilute
tracers and is verified and calibrated before each study. PLIF has been used as a
(Hult et al., 2005; Degardin et al., 2006; Weigand et al., 2006), flows of liquids in
vessels (Hall et al., 2004; Dazin et al., 2006) and impinging liquid jets (Unger et al.,
1999).
Analysis of the data obtained from PLIF analysis requires the creation of a matrix of
greyscale values from the black and white digital images obtained, this matrix is then
used to compute various statistics such as those outlined in section 2.2.1. Due to the
large amount of data to be handled, analysis using computation is essential. The use
of the software package, Matlab® to analyze the results was investigated by Golnabi
(2006) who described how Matlab can be used to import and convert an image into a
two dimensional array with values based on the intensity of the image data at each
pixel.
There are a number of error sources in PLIF measurements, mainly associated with
the optics, camera sensitivity, the Gaussian distribution of the laser intensity and
concentration illuminated from the side by a Nd:YAG laser, the signal was then
21
averaged along each row and column. The signal intensity was found to decrease
linearly across the cell, which was attributed to a decrease in camera sensitivity away
from the centre of the image. The errors that occur in PLIF images were also
analyzed by Law et al. (2000) who state the importance of the perpendicular
positioning of the camera relative to the light sheet to avoid deformation in the
images. The unsteadiness of the laser sheet was also measured in this study by
The above studies show that there is a body of work on the application and validation
unsteady (time-dependent) flows with moving free-surfaces has received very little
attention.
2.4.2 PIV
yields an instantaneous velocity field obtained over a short time interval over a given
particles within a thin laser sheet. PIV is non intrusive, indirect velocity measurement
technique has developed rapidly recently with the improvements in cameras lasers,
The PIV technique is demonstrated in Figure 2.4 This area of interest is illuminated
by a laser sheet and images of this sheet are captured using a camera. Each image is
22
then split into a grid of interrogation windows and statistical techniques are used to
frame 1
Crosscorrelation
Particle
displacement
frame 2
Vector
field
Pawlak et al. (2007) look at the dynamics of a starting laminar jet using a dynamic
particle image velocimetry (DPIV) system and PLIF. The DPIV system uses cross
correlation to produce a coarse grid of velocity vectors, filters this to remove outliers
then a fine grid analysis is performed, the resulting velocity field is filtered again
using a threshold median filter to eliminate outliers. The individual velocity fields are
then presented alongside the PLIF images for this transient process. Six different
integrals are calculated in the region of the leading vortex, the first is proportional to
the axial momentum flux exiting a disc perpendicular to the axial direction, the
second is related to the flux of vertical momentum and the third is responsive to radial
velocity changes. These three integrals are used to determine the exact axial location
23
of the vortex and a further three similar cases are used to determine the radial
position. The vorticity of the vectors was also used to analyse the local recirculation
in the vortex.
Turbulence is where there are temporal fluctuations in the fluid velocity at any point,
this can be measured through the PIV technique. It is assumed that the instantaneous
velocity can be decomposed into an average velocity and a fluctuating part. This
fluctuating part is usually characterised through an RMS value, the turbulent kinetic
energy (TKE) is the portion of the kinetic energy that is due to these fluctuations. As
PIV measures the velocity in a plane the third dimension is estimated from the two
(2001) states that the spatial resolution needed to capture 90% of the turbulent energy
is twice the Kolmogorov scale however Chung et al. (2006) obtain 95% accuracy with
Turbulent Kinetic Energy is a common statistic used in the analysis of turbulent jet
flow (Weisgraber and Liepmann, 1998, Bi et al., 2003,) however it may not be
jets include proper orthogonal decomposition. Bi et al. (2003) used DPIV and proper
The principles behind this involve maximising the correlation between the flow
description and the experimental results from the PIV, this is done by finding the
eigenmodes of the flow. They captured the velocity field at a frequency of 1000Hz,
using this data they plotted the spectral density of the frequency of the velocity
fluctuations in the centre of the shear layer on one side of the jet and found that the
24
preferred frequency of the fluctuations was approximately 11Hz. The PIV images
were then grouped into 80 blocks each containing 256 files giving a bandwidth of
3.9Hz, these were then analysed using POD, they note that the first six spatial
eigenmodes capture over 50% of the energy, lower than in traditional turbulence as
more eigenmodes were needed to capture the higher number of degrees of freedom of
turbulence caused by carrying out the analysis in the streamwise direction and in the
temporal domain. Kim et al., (2007) state that the eigenvalue represents the energy of
each eigenmode, they note that the proportion of energy in the first six modes
increases as the Reynolds number increases, which implies the flow has lower
dimensional structures.
2.5 Summary
Although there is little published research into the mixing that occurs as a vessel is
filled there is a large body of work into mixing by a range of other mechanisms. This
mixing is measured in a variety of ways from full field techniques such as PILF to
of variance statistics. From studies on mixing involving jets it can be seen that the
mixing performance improves if the length of the jet, including any impinged flow, is
maximised.
The techniques PLIF and PIV have been used to measure a variety of flow scenarios,
however in most cases the system is a steady state. There has been no study of
transient systems with a moving free surface, this work will adapt these techniques to
3.1 Introduction
A number of experiments were carried out throughout the project to determine the
flow properties and mixing performance for filling systems. Initially work was done
using a square section tank and a stationary pipe, this was then modified so that the
pipe could retract above the surface of the fluid. In both cases experiments were
carried out using PLIF and PIV to determine the level of mixing and the velocity field
respectively. Parameters such as nozzle design and flow rate were adapted based on
A schematic of the experimental set-up is shown in Figure 3.1. The vessel used was a
T = 0.06 m square section tank; the square shape was chosen to eliminate any errors
due to refraction through curved surfaces. The fluid enters the tank through a 0.006 m
diameter vertical dip pipe aligned with the central axis of the tank. The tank was
free flow. The volumetric flow rate of the pump was calibrated to the pump speed
(controlled digitally) for each viscosity of fluid used using the bucket and timer
method. Additional checks on the flow rate supplied by the pump were made by
Figure 3.1 Schematic of the experimental rig and PLIF set up.
Entrainment of air in the jet causes bubbles to be formed in the tank, when the laser
sheet hits these bubbles it is scattered causing shadows in the plane of interest. Bin
(1988) calculated the impingement height of the jet Lj for which air entrainment
begins in the region where the plunging jet would split into individual droplets region
as:
Lj −1.79
= 1.4 × 10− 5 d 0 We0j .937 (3.1)
d0
Where We is the Weber number which is calculated from the density (ρ) the velocity
ρv 2 l
We = (3.2)
σ
27
This gives an approximate value of Lj as 16 cm for the highest velocity fluid used in
these experiments, however in the continuous jet region entrainment will occur at a
shorter length. Whether the jet is in the continuous or droplet regions is dependent on
the inlet turbulence parameters. Initial test cases demonstrated air entrainment when
the pipe was fixed above the maximum fill level of 10cm, an impingement height of
3cm prevented most entrainment however a pipe fixed in this position would quickly
become submerged. To allow the jet to impinge on the surface of the fluid while
preventing the impingement height from exceeding this limit the nozzle was raised so
To adapt the filling set up to allow the pipe to be moved, a purpose built traverse was
designed (Figure 3.2) which allowed the pipe to move in the vertical direction at a
controlled speed and with minimum vibration. This consisted of a motor which
rotated a vertical screw thread along which the pipe moved guided by two supports.
The speed was controlled using a specifically written control program and was
manually set before filling began. This speed was chosen to be the same as the rate at
which the free surface rose which was calculated from the flow rate. The rig included
cut off switches positioned to limit the pipe’s range of movement to keep it within the
range of the screw and also to prevent it from colliding with the vessel. The traverse
was started manually at the beginning of the filling process to coincide with the
Motor controlling
traverse
Traverse path
Clamp for
nozzle
Limit switches
Pyrex square
section tank
Laser source
supported by
bread board
improvements fell into two groups; varying the flow rate and changing the pipe
geometry.
29
A variable flow rate was created using a computer program to control the pump, this
was written by Andrew Tanner from the Biosciences Workshop at the University of
Birmingham. The program was written so that an excel file read the time and the
flow rate and any function that could be written in excel could be used to describe the
flow, the program then runs until it reaches an empty cell, at this point the pump is
stopped. The actual flow rate achieved is fed back to the computer and is displayed
graphically alongside the prescribed function this can then be exported to excel. Due
to limitations of the pump there was a lead time between the function and the actual
speed achieved and depending on the desired flow rate there was a certain length of
time to start up. This could be up to 2.5 seconds for the highest flows which meant
that there was a limit on the rate of change of the flow. For a sinusoidal function the
period achievable was 4 seconds, the flow rate was varied from 1.67 x 10-5 m3 s-1 to 5
x 10 -5 m3 s-1. For the ramped flow the flow rate started at zero and accelerated up to
the highest rate of 6.67 × 10-5 m3 s-1. In both cases the average flow rate was 3.33 x
10-5 m3 s-1.
Three nozzle designs were made to improve the mixing in the tank, two of these had
cone inserts in the base and one had a ‘swirl’ insert. For the cone insert the pipe was
flared to accommodate it, as shown in Figure 3.3, in practice spot welds were
required to attach the inserts to the outside of the pipe. The swirl nozzle was created
by inserting and welding a twisted piece of metal into the end of the pipe.
30
10mm
8mm 6mm
10mm 8mm
Figure 3.3 Schematic of a straight pipe and the three nozzles improvements.
The experiments were carried out using Newtonian fluids. It was necessary for the
fluids to be transparent to allow the laser beam to pass through for the PLIF and PIV
diagnostic techniques. Aqueous glycerol solutions were chosen for initial studies,
equipped with a cone and plate geometry (TA AR1000, TA instruments). The fluid
Energy dissipated,
number, Re (-)
number, Re (-)
length scale, λ
length scale, λ
ε (W kg-1)
Fill time, t
Kolmogoroff
Kolmogoroff
Flow rate,
Viscosity, μ
Viscosity, μ
Density, ρ
Density, ρ
(m3s-1)
Reynolds
Reynolds
(kg m-3)
(kg m-3)
(Pa s)
(Pa s)
(s)
(μm)
(μm)
Q
1.67 × 10-5 21.6 0.008 6 × 10-3 1121 661 389 8 × 10-3 1126 498 459
3.33 × 10-5 10.8 0.064 6 × 10-3 1121 1322 221 8 × 10-3 1126 996 273
5.00 × 10-5 7.2 0.217 6 × 10-3 1121 1983 158 8 × 10-3 1126 1494 202
ρd j u j
Table 3.1 Experimental parameters, where Reynolds number is defined as Re = .
µ
PLIF images were obtained using a TSI Powerview system (TSI Inc. USA)
camera (TSI PIVCAM 10-30, TSI Inc, USA), synchroniser and a dual head Nd-Yag
laser emitting at 532 nm (New Wave Inc., USA) and equipped with laser sheet optics.
The lenses used were a -15 mm cylindrical lens and a 200 mm spherical lens and the
laser was positioned 150 mm from the base of the vessel; this combination was
chosen so that thickness of the sheet was a minimum in the middle of the vessel. The
32
laser sheet thickness throughout the imaging plane was less than 1 mm. The system
was controlled using a Dell Precision 620 workstation (Dell Corp. USA) running
INSIGHT 6.0 software. The images were captured at a rate of 15 Hz, saved as .tif
The fluorescent tracer used was Rhodamine 6G which emits light at 566 nm when
excited using radiation at 532 nm, Rhodamine was chosen as the relationship between
the concentration and the intensity of the laser light emitted and hence the grayscale
of the images is linear over a certain range of concentration, this is known as the
concentrations, this was carried out at two different laser intensities; the graph
hold up to a concentration of 0.0001 kg m-3 (0.1 mg L-1) and of the two laser
intensities used the higher gave a better resolution over the required range and was
therefore used for the experiments. Before the fluid injection began 1 × 10-6 m3 of the
tracer was spread across the bottom of the vessel, as this tracer is then diluted by the
inflowing fluid the concentration of the tracer was chosen so that the concentrations in
the vessel would be within the linear range when the level of mixing was being
measured. Measurement took place when the fluid level reached approximately 2 cm
until the vessel was full, this meant the tracer would be diluted by a factor of 100
before measurement took place, so the concentration used was 0.008 kg m-3
(8 mg L-1), as this would ensure that the concentration was within the linear region
before measurement began. The camera was fitted with a sharp cut-off high-pass
wavelength filter of 545 nm (ALP545, TSI Inc.) to eliminate the light from the laser at
33
532 nm and allow only the fluorescent light from the tracer fluid at 566 nm to pass
through.
Figure 3.4 Plot of greyscale as a function of tracer concentration at two different laser intensities.
3.4.1 Normalisation
As shown above, the relationship between the measured grayscale values and the
tracer concentration was found to be linear throughout the concentration range found
when the fluid heights were above 2 cm. To determine this linearity precisely a set of
images must be taken to calibrate the grayscale values measured for each pixel in the
There are two main considerations to take into account when determining the pixel by
pixel values of concentration from the grayscale values; the laser sheet intensity has a
Gaussian profile across the plane so a uniformly mixed tank will not emit a uniform
34
intensity of light, and the individual pixels of the CCD array in the camera do not
have a uniform sensitivity. To eliminate these sources of error each pixel is calibrated
concentration with the vessel fully mixed; this process is then repeated with double
the known concentration. The average grayscale value for each pixel over the 50
images is then calculated at each concentration and hence two arrays containing the
pixel by pixel values of grayscale are created. The linear relationships connecting the
matrices are determined and a further two matrices are created containing the gradient
(M) and intercept (K) of these linear equations. This is demonstrated in Figure 3.5
using the first two pixels as examples, the nth element of the M or K matrix is
calculated from the measured greyscale at the nth pixel of the image when the
The zero concentration case is not used in the calibration as it does not include the
error associated with any non-uniformity in the laser sheet. This is because with no
concentration there is no fluorescence is emitted which the only detection method for
the local intensity. However they do allow the level of this variation to be quantified.
back to the zero point there is found to be an error of 2.26 % between the calculated
‘K’ values and the experimental greyscale values for an image of an illuminated
Figure 3.5 Calculation of elements of the matrices M and K from the linear relation between greyscale
and tracer concentration for each pixel, (only two pixels are plotted for demonstration).
Each image is then loaded one at a time into Matlab as an array of greyscale values
(see appendix for sample code). The image is then cropped so that only the area of
the image containing fluid is analyzed. The surface of the fluid is located by
calculating the average of each row and determining when this average is more than
10% higher than the background grayscale, the first point at which this is the case is
taken to be the free surface of the fluid. This was checked and found to be the best
cutoff by viewing cropped images to ensure that the fluid area had been correctly
identified, the calculated height of the fluid is also monitored to ensure it increases
linearly throughout the filling processes. If the height does not increase steadily with
image number then the images are not accurate and are discarded, this occurs when a
droplet forms on the pipe of edge of the tank which reflects the light higher in the
36
vessel. For the purposes of this analysis the image is split into two interrogation
zones, one on each side of the inlet pipe. The change in the average greyscale values
at the surface and the area used for the analysis are shown in Figure 3.6.
Figure 3.6 Definition of the location of the fluid surface via the change in the average row greyscale
values as a function of the vessel height. The interrogation zones are shown and also the maximum fill
height.
The height of the liquid in the vessel is then calculated from the number of pixels that
cover it using a scaling factor calculated using the base of the tank as a standard
distance, and from this the average concentration of the fluid in the vessel at that
height. This is then converted into a grayscale value (Gn) at each pixel using the
each pixel would be at if the vessel were fully mixed; it is dependent on the fluid
height in the vessel, the concentration of the tracer and the volume of tracer used.
37
This concentration decreases as the vessel is filled which means that the degree of
mixing is being measured against a coarser scale when the vessel is nearly full.
analysis is used as described by Brown et al. (2004). The grayscale value at the nth
pixel is normalized at each time t by considering the grayscale values when the
concentration is zero i.e. background (G0) and the value if the vessel was completely
(Gn (t ) − G0 )
Gn' (t ) = , (3.3)
(Gm − G0 )
where G’n (t) is the normalized grayscale value and Gn (t) is the grayscale value of the
nth pixel in an image taken at a time t. The log-variance (σ2) of the concentration in
1
∑ (G )
N
2
log σ 2 = log '
n −1 , (3.4)
N n =1
where N is the total number of pixels used in the image. The log-variance gives a
direct measure of the level of mixing on the scale of a single pixel and as the mixing
improves this number decreases. For example, when the vessel is 95% mixed the log-
The number of pixels included in the variance increases with the height of the fluid
due to the larger area covered, this causes the error in the measured log variance to be
reduced though the process as the sample size increases. However the scale of the
38
mixing being measured does not change throughout the fill, for this to be reduced a
In order to ensure that the equipment has sufficient resolution to resolve mixing
events across all the length-scales, a characteristic length scale equivalent to the
Kolmogoroff scale for turbulent mixing was calculated. Although the mixing
mechanism was shown later to be predominantly laminar for the higher viscosity
fluids, hence the Kolmogoroff scale would not be expected to apply, calculation of
this scale provides an order of magnitude check on the likely accuracy of the
technique. The Kolmogoroff length scale is a function of the energy input per unit
1/ 4
ν 3
λ = (3.5)
ε
In a filling system, if the fill rate is constant, the mass of fluid in the vessel increases
linearly in time. The energy input per unit mass to the fluid from the liquid jet was
calculated on the basis of the dissipation of the kinetic energy into the bulk fluid in the
fully filled vessel; this provides the minimum value of specific energy, ε, input
1
ρQu 2j
ε= 2 , (3.6)
ρT 2 H
where Q is the volume flow rate of liquid, ρ is the density, u is the velocity of the jet
and T and H are the height and width of the tank.. Values of λ and ε, calculated using
(3.5) and (3.6) respectively are given for each experiment in Table 1. The range of λ
is from 94-650 μm, this compares well with a pixel by pixel resolution in the images
of 130 μm.
39
Using the calibration method, the pixel by pixel variations in the laser sheet and in the
camera sensitivity are taken into account, however, this analysis assumes that the
intensities within the laser sheet are constant each time the laser head is triggered.
The fluctuation in the laser sheet with time was examined by observing multiple
calibration images. The concentration and liquid height were kept constant for 50
images so that temporal fluctuations in the signal could be attributed to the laser. The
laser sheet fluctuation had an average standard deviation of 2% for each pixel. This
error propagates through the calculations, there will be a 2% error in the two
intensities used to calculate the values of M and K, giving a 2.8% error in each pixel.
These are in turn used to calculate the fully mixed concentration and the normalized
concentration. Despite this propagation of error the large number of pixels used for
each log-variance value reduces the error. If more than 120 pixels are used it is
reduced to below 1% giving an error due to laser fluctuation in the log value of less
than 0.004, for a full vessel more than 15000 pixels can be used.
A further cause of anomalous concentration results occur when there are pixels which
have very low sensitivity or have been damaged. When the difference between the
grayscale value at zero concentration, Go and at the fully mixed condition, Gm, is too
small, their use in the denominator of Equation 3.2, leads to a very coarse
measurement which is not statistically significant. Any damaged pixels in the camera
appear as permanently saturated with light so the differences between the grayscale
values from these pixel is zero, this means that when the concentration values are
normalized an infinite answer is obtained. To eliminate both these problems all pixels
where the difference between the concentration at zero and the concentration at fully
40
mixed is less than a certain cut off value, usually taken as 10 are discarded prior to
PIV images were obtained using a TSI high-speed PIV system. A high repetition rate
laser (10mJ at 1kHz) was used for illumination (New Wave) and a high frame rate
camera (Photron APX RS) to capture data at 500 Hz with a spatial resolution of 1025
× 1025 pixels2. The lens set up used was identical to that used in the PLIF
The particles used to trace the fluid were silver coated glass particles of diameter 10
μm which reflected the light at the same wavelength as the initial laser plane (532
nm). These particles have a sufficiently small relaxation time that they can be
assumed to be following the fluid motion; they were well mixed into the fluid before
the injection of the fluid into the vessel during the experiment. The instantaneous
spatial displacement of the particles in the vessel was calculated by splitting the
images into small interrogation windows 32x32 pixels. Fast Fourier Transform and
Gaussian statistics were then used to cross correlate between the images from the two
time exposures. The spatial displacement that produces the maximum cross-
interrogation window (Hall et al. 2005). The velocity of the particles can then be
calculated by combining this information with the known pulse separation. The
resulting matrix of velocity vectors defines the instantaneous flow field in the whole
image.
41
The resolution of the velocity field measured using the PIV is dependent on the
magnification of the camera, the size of the interrogation window and the pulse
displacement between the pulses as one quarter of the interrogation window. The
amount of particles in each interrogation window also affects the accuracy of the
cross correlation, the optimum seeding for the fluid is 10-12 particles per
interrogation window.
images, however in a transient flow this is not possible. An assumption that the fluid
height was stationary throughout the capture of 200 images was made so that
averaging could be used. The time needed to capture 200 images was 0.4 sec, for the
worst case scenario, a flow rate of 5.00 × 10-5 m3 s-1 the fluid height changed by
0.0056 m in this time; for the assumption to be valid the flow conditions must not
change significantly due to this height change. Sets of images were taken when the
fluid level reached 3 cm, 5 cm, 7 cm and 9 cm at two different pulse separations, this
was necessary due to the large difference in the velocity magnitude between the bulk
fluid and the jet. The data from both pulse settings was then combined during
analysis to give a complete picture of the velocity fields, an example MATLAB code
As a large amount of the mixing that occurs is laminar the areas of high shear in the
vessel are important, as it is the elongation and shear as well as folding of fluid that
causes it to mix. The magnitude of the shear rate in the x and y direction is calculated
by:
42
∂v x ∂v y
γ = + (3.7)
∂y ∂x
This is then plotted as a colour map with the velocity vector field, to show the regions
of high shear.
43
4.1 Introduction
The results presented in this Chapter provide basic knowledge of the mixing in the
simplified systems described in section 3.2. The first of these had the inlet pipe fixed
1 cm from the base of the tank; later experiments used a traverse to keep the pipe
above the liquid free-surface as the vessel filled. Jets with nine different Reynolds
numbers from 259 to 2867 were investigated by changing both the jet exit velocity
and the viscosity of the fluid, this range allowed the transition from relatively
unmixed laminar flows to better mixed turbulent flows to be observed and identified.
The technique employed in this Chapter to determine the mixing properties of the
flow was Planar Laser Induced Fluorescence (PLIF) as described in section 3.4. The
results obtained from this technique are discussed and where possible conclusions
The simplest system set up investigated using PLIF had a fixed pipe inlet (as
described in section 3.2) this system was used to develop the PLIF measurement
technique over an appropriate range of flow conditions chosen for investigation; these
1
A selection of the PLIF results presented in this Chapter for the fixed pipe inlet have been published
Figure 4.1 Evolving concentration fields at different fill heights for the 6 mPa s fluid at (a) h = 2.2 cm;
(b) h = 4.7 cm; (c) h = 7.3 cm; (d) h = 10.3 cm at Re=661 (Q = 1.67 × 10-5 m3 s-1).
Figure 4.2 Evolving concentration fields at different fill heights for the 3 mPa s fluid at (a) h = 2.2 cm;
(b) h = 4.7 cm; (c) h = 7.3 cm; (d) h = 10.3 cm at Re = 1245 (Q = 1.67 × 10-5 m3 s-1).
45
Figure 4.1 and Figure 4.2 show greyscale images taken by the CCD camera for the 6
and 3mPas fluids at values of Re of 661 and 1245 respectively. Both fluids generate a
similar flow pattern; a toroidal vortex is formed in the bottom quarter of the vessel
(Figure 4.1a, Figure 4.2a) whilst the flow in the upper half of the vessel is comparably
quiescent and does not have a well defined structure (Figure 4.1c-d, Figure 4.2c-d).
The striations created by the fluorescent tracer remain more defined in the low
Reynolds number fluid than for the high Reynolds number fluid when the vessel is
Using the naked eye, the degree of mixedness appears to be higher for the higher
Reynolds number fluid as both the scale of the striations is smaller and the contrast
between them is lower, suggesting qualitatively a reduction in both the scale and
since eddy diffusion would be expected to eliminate them if the flow were turbulent.
The images shown here may form a basis for a mechanistic determination of the
and white, whilst eddy diffusion would be expected to cause regions of ‘grey’.
Comparing Figure 4.1d and Figure 4.2d, regions of the vessel appear more mixed in
Figure 4.3 Mixing performance for three different fluid viscosities at a fill rate of Q = 1.67× 10-5 m3 s-
1
(a) log variance plotted against dimensionless fluid height (H/T); Image of the vessel filled with fluid
at (b) Re=343 (12 mPa s), (c) Re=661 (6 mPa s), (d) Re=1245 (3 mPa s).
dimensionless fluid height (h/T) in the vessel for three fluids of decreasing viscosity
at the same fill rate of Q = 1.67×10-5 m3s-1, giving Reynolds numbers of 343, 661 and
value of Re of 1245 (red line), the log-variance reaches a minimum of ~ -1.8 (87%
mixed) once the vessel is filled beyond 3 cm in height and remains constant until the
47
fill is complete. At values of Re of 661 and 343 (blue and green lines respectively),
the log-variance stays at a value close to zero throughout the fill, indicating that the
There is a high level of noise in the log-variance plot, a small amount of which will be
due to the measurement technique however comparison between the low and higher
Reynolds number flows show that there is an increase in the noise for the higher
Reynolds number, this could be caused by areas of high or low concentration moving
through the measurement plane tangentially causing the average concentration in the
These results are borne out by the images shown in b-d which display an image of the
filled vessel taken at the instant after completion of the filling process at values of Re
of 343, 661 and 1245 respectively. The tracer remains unmixed with the bulk of the
fluid in Figure 4.3b and c and is located either in the toroidal vortices at the bottom of
the vessel or is swept up close to the fill pipe or the free surface. At Re = 1245, as
shown in Figure 4.3d, the mixing performance is much improved and the tracer is
spread over the entirety of the vessel, although striations of black, where no tracer is
present, are still visible. Comparison of the visual images with the log-variance
determine that the vessels are in an unmixed state, at Re = 1245, the degree of
information on the level of mixedness shown in the image is much higher than can be
remarkable in that a threefold increase in fluid viscosity causes the vessel to go from a
relatively well mixed state to almost unmixed at the same flow conditions.
48
Figure 4.4 Mixing performance for three different fluid viscosities at a fill rate of Q = 3.33× 10-5 m3 s-
1
(a) log variance plotted against dimensionless fluid height (H/T); Image of the vessel filled with fluid
at (b) Re=687 (12 mPa s), (c) Re=1322 (6 mPa s), (d) Re=2490 (3 mPa s).
Figure 4.4 shows similar data for fluids with the same three decreasing viscosities at a
higher fill rate, Q of 3.33×10-5 m3s-1, giving values of Re of 687, 1322 and 2490
respectively. Again, at the lowest value of Re (687), the log-variance remains close to
zero (Figure 4.4a) and the tracer remains unmixed through the bulk of the vessel
49
(Figure 4.4b). The mixing performance at the higher values of Re of 1322 and 2490
improves noticeably, with the log variance reaching ~ -1.5 (82% mixed) and -2.5
(94% mixed) respectively. Notably, for the experiment with the value of Re equal to
1322 (6 mPa s fluid, Figure 4.4c), the degree of mixedness is qualitatively similar to
the tracer appears visually to be almost uniformly mixed through the vessel and very
few striations are present; this may be indicative of a change of mixing mechanism
The change in mixing mechanism is even more apparent at the highest flow rate used,
Q = 5 ×10-5 m3s-1. The change in flow is still not sufficient to mix the highest
viscosity fluid at Re = 1030, (Figure 4.5b) but for Re = 1983 and Re = 3735 the
distribution of the tracer in the tank (Figure 4.5c,d respectively) is almost uniform.
The log-variance drops to -2.2 (92% mixed) and -3 (97% mixed) for the lower
viscosity fluids at Re = 1983 and Re =3735 respectively when the dimensionless fill
height (h/T) is of the order 0.5-1. However, upon further filling of the vessel, the
mixing performance actually worsens and the log-variance begins to rise again,
reaching – 1.8 at Re = 1983 (87% mixed) and -2.4 (94% mixed) at Re = 3735. The
worsening in mixing performance can be attributed to the quiescent nature of the flow
in the top half of the vessel and to the linear drop in specific energy input as the vessel
is filled.
50
Figure 4.5 Mixing performance for three different fluid viscosities at a fill rate of Q = 5.00× 10-5 m3 s-
1
(a) log variance plotted against dimensionless fluid height (H/T); Image of the vessel filled with fluid
at (b) Re=1030 (12 mPa s), (c) Re=1983 (6 mPa s), (d) Re=3735 (3 mPa s).
For flows that appear to be dominated by laminar mixing, the toroidal vortices are
visible, Figure 4.6 and Figure 4.7 show the position of this vortex in relation to the
base of the tank and the inlet pipe at values of Re of 343, 661 and 1245 respectively
(Q = 1.67×10-5 m3s-1). It can be seen from Figure 4.6 that as the viscosity increases
the vortex centre moves closer to the base of the tank. For all three viscosities the
51
height of the vortex at the beginning of the fill is less than 50% of the fluid height,
perhaps caused by interactions with the free surface of the fluid. As the fluid height
increases so does the vortex height until it reaches a maximum value which increases
with Re. At Re = 343, the vortex height that does not fluctuate significantly from its
the vortex position; this instability is also evident in Figure 4.7. At Re = 1245, the
fluctuations have a standard deviation of 8%, this may be the first signs of a more
Figure 4.6 The dimensionless height (h/T) of the centre of the vortex from the base of the tank,
measured for three Reynolds numbers at flow rate of Q = 1.67× 10-5 m3 s-1
Figure 4.7 shows that the average horizontal distance of the vortex from the pipe does
not vary significantly across the values of Re used in these experiments which
suggests that it is affected only by the geometry. This implies that the vortices are the
same size for the different viscosities but they have just moved further up the vessel.
52
While this is true for the higher two viscosities which are still within the laminar flow
regime, the 3 mPa s fluid shows a large amount of distortion of the vortex which
manipulating the geometry: clearly this will also alter the ratio of the jet to tank
Reynolds numbers and the distribution of local specific energy dissipation rate.
Figure 4.7 The dimensionless horizontal distance of the centre (h/T) of the vortex from the edge of the
inlet pipe, measured for three Reynolds numbers at flow rate of Q = 1.67× 10-5 m3 s-1
The transition between laminar and turbulent flow is normally defined on the basis of
a critical value of Re determined experimentally for a given system. This value can
vary substantially, since the choice of length and velocity scales depends on the
geometry in question.. As shown in section 3.3.1, the values of Re based upon the dip
53
pipe diameter and jet exit velocity are of O(103). An alternative would be to consider
a tank Reynolds number based upon the flow in the vessel, using the tank width and
the superficial velocity of the fluid (based on the cross-sectional area of the tank
bottom) as the length and velocity scales respectively. This would give Reynolds
numbers of O(10). Clearly, for the constant geometry used in these experiments, the
However, it is not sufficient to use Re as the sole parameter to characterize the mixing
a critical parameter is the energy input per unit mass, ε, which can be evaluated
locally and also averaged over the whole vessel. Indeed Kresta & Wood (1993) stated
that prediction of the local specific energy dissipation rate, ε T , is, in many cases, the
key to successful process modeling. In the filling case, the analysis is complicated
since the mass of liquid in the vessel increases linearly in time. At the start of the
filling process, when the mass in the tank is small, the energy input per unit mass is
high which causes intense mixing. As shown in Figure 4.1a and Figure 4.2a, the
initial charge of tracer is swept off the bottom of the vessel and the initial mixing
performance is commensurate with the high specific energy of the fluid. As the
vessel is filled further, the energy input per unit mass drops linearly with time and the
mixing performance noticeably worsens: Figure 4.1 and Figure 4.2 (b-d) demonstrate
that the tracer is swept upwards near the dip tube as it exits the toroidal vortex and
does not appreciably mix further with the fresh fluid entering the vessel. This effect is
more noticeable with the higher viscosity fluid, which is to be expected since the
value of Re is half that of the lower viscosity fluid. The extra mixing of the low
viscosity fluid may be either due to the creation of eddies via turbulence or due to
54
and molecular diffusivity holds (μ ~ DM-1); mixing due to molecular diffusion would
be twice as rapid in the low viscosity fluid. If one were to consider the remainder of
DM = O(10-8) m2s-1, then if the striations were 10% of the vessel diameter, the
diffusion time,
td =
λ2
=
(0.1T )2
DM DM
This gives tD ~ 3600 s for T = 0.06 m, or about one hour. This would be possible for
single phase bottled products during shipping. Over a fill time of ~ 20 s, assuming the
same value of DM, molecular diffusion would only occur over a length-scale of
Figure 4.8 shows the log variance as a function of fluid height for three Reynolds
numbers using the moving pipe described in section 3.2.2; the mixing improves as the
Reynolds number increases as in the fixed pipe experiments. The lowest Re = 257
shows very little mixing throughout the fill with the log variance staying around zero,
this is comparable to the static pipe results at similar Reynolds numbers where the
mixing does not improve throughout the fill. For the highest Reynolds number
Re = 2867 the log variance decreases to ~ -2.6 indicating that it has reached 95%
mixed. Between these two is the Re = 996 case where some mixing occurs but not
sufficient to reach the 95% level, here the log variance reaches ~ -1.5. For both cases
where mixing occurs once the fill height has reached 5 - 6 cm, there is little further
improvement to the mixedness and in some cases the log variance increases indicating
55
that the ‘fresh’ fluid entering the vessel is no longer mixing as effectively. For the
highest Reynolds number this may be due to the fact that there is always an area of
lower tracer concentration around the jet, so despite the homogeneity in the
Figure 4.8 Mixing performance for three different fluid viscosities at a fill rate of Q = 5.00×10-5 m3 s-
1
(a) log variance plotted against fluid height; Image of the vessel filled with the (b) 16 mPa s fluid, (c)
Figure 4.8b-d show a very clear change in the mixing mechanism, the Re = 257 fluid
(Figure 4.8b), the tracer is distributed around the vessel but there are two toroidal
vortices and obvious striations which indicate laminar flow conditions. The highest
Reynolds number fluid Re = 2867 (Figure 4.8d), is a more turbulent flow and as a
result it shows a more uniform distribution of dye, the striations have been removed
through eddy diffusion. For the Re = 996 (Figure 4.8c) case there are still striations
56
visible but there is more uniformity than the Re = 257 case, this shows the transition
between the two different flow regimes. This agrees with the fixed pipe case
A further indication of the onset of turbulence is the stability of the jet, the more
laminar jet (Figure 4.8b) shows the ‘fresh fluid’ travelling directly to the base of the
tank without undergoing spreading or jet break up and hence little mixing. The
transition jet shows similar travel to the base of the tank of the fresh fluid, however as
the jet becomes longer with the surface rising instabilities appear towards the base
(Figure 4.8c) indictating the jet is no longer purely laminar. The turbulent jet (Figure
4.8d) shows the jet mixing quickly with the surrounding fluid due to the jet spreading
Figure 4.9 Plot of the minimum log variance reached in each experiment versus the jet Reynolds
number.
57
Figure 4.9 shows the minimum log variance reached by each experiment compared
with the Reynolds number of the jet in each case. The log variance in the tank
decreases rapidly as the Reynolds number increases for small values of Re. However
once the log variance decreases to below -2.5 this improvement in mixing becomes
more gradual, this occurs for Reynolds numbers above 1000. The mixing
with the turbulent mechanism becoming dominant above a value of Re of ~1000, the
evidence for this being the disappearance of the toroidal vortices and the appearance
of instabilities in the jet. This agrees with Kim et al. (2007) who observed the
stability of a water jet and found it was steady at Re = 404 but had become unsteady
In comparison with the stationary pipe results (Figure 4.9) using a moving pipe
appears to improve the mixing performance at the same liquid flow rate. From
observing the images the flow appears to be more turbulent with the moving pipe,
possibly because of the increasing length of the free jet which allowed instabilities to
develop. Also mixing will also have been improved by increasing entrainment into
It should be noted that the level of mixing in the tank has not reached the limit of the
resolution of the camera. For very highly mixed conditions the ability of this
technique to measure the fluctuations in concentration will be limited by the pixel size
Figure 4.10 shows the flow pattern at different points in the fill for the Re = 1494
fluid, this clearly shows the striations that form in the tank and the improvement in the
mixing as the jet length increases. However a small number of nominally identical
comparison between Figure 4.10 and Figure 4.11 shows that the tank in Figure 4.11
appears to be better mixed, this is verified by comparing the log variance of the two
Figure 4.10 Evolving concentration fields at different fill heights for an 8 mPa s fluid, Re=1494 at (a)
h = 2.2 cm; (b) h = 4.7 cm; (c) h = 7.3 cm; (d) h = 10.3 cm at a flow rate Q = 3.33 × 10-5 m3 s-1
Figure 4.11 Evolving concentration fields at different fill heights for an 8 mPa s fluid, Re=1494
where a bubble has formed at (a) h = 2.2 cm; (b) h = 4.7 cm; (c) h = 7.3 cm; (d) h = 10.3 cm at a flow
Unlike Figure 4.10, in Figure 4.11a-b the dark jet is not clearly visible down the
centre of the tank although small areas of low concentration can be seen towards the
base, this suggests that the jet has been obstructed in some way and the fluid is now
entering the tank away from the centre plane. Figure 4.11c shows a bubble being
released into the flow, behind which the jet becomes clearly visible; in Figure 4.11d
the jet is now flowing through the centre of the tank. It appears that this bubble was
obstructing the jet, the dark eddies seen towards the base in Figure 4.11a-b suggest
that this obstruction has made the jet instable fortuitously creating a more turbulent
flow and better mixing. This mixing improvement could be created in the process by
putting a permanent obstruction in the end of the dip pipe, this is investigated in
Section 6.3. However, not surprisingly, the random presence of the bubble led to
Figure 4.12 The log variance plot of two fills of the 8 mPa s fluid at a flow rate Q = 3.33 × × 10-5 m3
s- where a bubble has formed under the pipe in one (blue) and the jet is unobstructed in the other (red)
4.4 Conclusions
The PLIF technique has been demonstrated to give good repeatability for the same
flow conditions and qualitative examination of the images has revealed a change in
the mixing mechanism from laminar to turbulent above a critical value of the
Reynolds number (Re~1000) which agrees with previous literature. For the cases
above Re~1000 the log variance reaches -2 which equates to 90% mixed. The use of
a retracting inlet pipe is shown to give a small increase in the mixing performance,
above Re~1000 the log variance reaches -2.2 in each case which equates to 92%
mixed. The retracting pipe also allows the stability of the jet to be examined giving
further evidence of the mixing regime in the tank. A bubble causing an obstruction in
the end of the pipe improved the mixing performance and could be replicated in the
5.1 Introduction
To gain a better understanding of the mechanisms effecting the mixing in the tank the
velocity distribution was investigated using PIV. A comparison between the flow
patterns in the cases where the jet was able to mix the fluid and those where the
mixing was unsuccessful was made. In these cases the dead regions in the tank could
be identified and the areas where the shear rates are higher and more mixing occurred
highlighted. This chapter looks at the velocity profile and structure in the jet, a
comparison between jets at different Reynolds numbers is then made. The links
between the mixing in the tank and the flow patterns are made.
The PIV was set up as described in section 3.5, this allowed the flow field in the
vessel to be measured for the same values of Reynolds numbers as for the PLIF
The general structure of the jet impinging on a pool of liquid is shown in Figure 5.1;
this jet structure in this region agrees with previous works as described by Reungoat
et al. (2007). The fluid enters the tank at the top of the vessel and travels vertically
downwards until it reaches the impingement zone, it is then forced along the base of
the tank towards the walls, at this point the fluid has lost the majority of its
momentum and the subsequent upwards movement is slow. This slow upward
62
velocity is directed slightly towards the jet as it replaces the fluid that has been
Relatively
quiescent, but
with a slow
drift towards
the centre
Figure 5.1 An example of the PIV velocity profile in a vessel showing the features present in all cases.
Figure 5.2. shows the jet impingement zone at the base of the tank, the stagnation
region in the centre of the jet is clearly visible however the low accuracy of the results
next to the glass base means that the flow in the boundary layer cannot be analysed
further. Dead zones can be seen in both the bottom corners of the tank where the
63
velocity is low caused by the abrupt change in direction of the jet along from the base
Figure 5.3. The change in normalised velocity along the jet centreline for the four different points in
the fill
Further insight into the dynamics of the jet can be gained from Figs. 5.3 and 5.4.
Figs 5.3 shows the decrease in the centreline velocity of the jet as it moves from the
point of injection towards the point of impingement. The velocity has been
normalised by dividing though by peak velocity at the end of the pipe to allow
comparison. The deceleration of the jet is caused by the viscous drag of the
surrounding fluid, this viscous drag causes energy to be dissipated and also the
surrounding fluid to become entrained into the jet, the more the surrounding fluid is
entrained the more kinetic energy is transferred to the rest of the tank. The
deceleration of the jet is relatively high near the point of injection and decreases to a
64
steady rate when the normalised velocity reaches ~ 0.8. The effect of the
impingement can be seen when the fluid reaches approximately 1 cm from the base
where the velocity suddenly begins to decrease more rapidly, this point is this same
0.05
-0.05
Vertical velocity (m/s)
-0.1
-0.15
-0.2
h= 0
h= 0.46
-0.25
h= 1.02
h= 1.48
-0.3
h= 2.04
h= 2.50
-0.35
-0.4
3 2 1 0
Distance From Centre
Figure 5.4. Vertical velocity profile across horizontal slices taken at heights h above the base of the
tank
Figure 5.4. shows the velocity at points away from the centreline for the heights in
and near the impingement zone. At 2.5 cm the peak has well defined edges, moving
closer to the wall the peak spreads and there is a larger amount of upward velocity
towards the tank wall. For the data closest to the tank base this upward motion next to
the wall is not present, this is due to the dead zones in the corners of the tank
65
Figure 5.5. The centreline velocity profile for low Reynolds number fills when the injection height has
reached 9 cm
Figure 5.5 shows the centreline velocity of the lower Reynolds number jets where the
flow is mostly laminar, Figure 5.6 shows the same plot for the higher Reynolds
numbers where flow appeared more turbulent, Re=959 appears in both graphs. For
the laminar flows there is a change in the gradient of the jet velocity is as the
Reynolds number is increased; the lower Reynolds numbers show the largest drop in
the velocity along the jet. As the fluid is more viscous the drag on the edges of the
fluid flow is larger and so the jet’s kinetic energy is lost. At high Reynolds numbers
the jet is turbulent, Figure 5.6 shows that once the transition to turbulence has
occurred the gradient of the velocity decrease is approximately constant for all values
of Reynolds number. The case with Re=996 does not fit into either catagory but is in
transition between the two, this agrees with previous observations from the PLIF data
Figure 5.6. The centreline velocity profile for high Reynolds number fills when the injection height has
reached 9 cm
The transition in the flow can be seen from the velocity magnitude plots (fig 5.7). For
low Re there is very little observable detail of the velocity change in the top half of
the vessel and the contours are fairly well defined; this flow showed the dye trapped
along the wall or up the side of the vessel in the PLIF data. For the transitional flow
there is some movement in the top of the vessel and the contours loose their
smoothness, this flow exhibited the definite striations in the PLIF data. For the higher
Re plots there is a larger amount of velocity fluctuation across the top of the tank, as
this becomes more pronounced the corresponding PLIF data shows the jet becoming
more unsteady and the mixing performance is improved. There is an error around the
(a)
(b)
Figure 5.7. Velocity magnitude contour plots for flow at a) Re=517 and b) Re=1911
As a large amount of the mixing that occurs is laminar the areas of high shear in the
vessel are important, as it is the elongation and shear as well as folding of fluid that
causes it to mix. The magnitude of the shear rate in the following plots is calculated
From Figure 5.8 it can be seen that the values of the shear are highest at the outsides
of the jet, where there is a sharp change in the velocity and along the base of the tank
where the jet moves radially away from the centre of the tank. These areas are where
68
the striations seen in the laminar flow originate; this is then folded by the recirculation
in the tank.
Figure 5.8. The shear rate calculated from the velocity field measured at Re=517
5.4 Conclusions
To examine the flow regime in the tank further PIV was used; the resulting vector
fields allowed the distribution of momentum and shear in the tank to be investigated.
For all flows, dead regions were observed in the bottom corners of the tank and the
top half of the tank was relatively quiescent, however this was more prominent for
more laminar flows. For the more turbulent flows (Re > 1000) there was a higher
level of movement in the top half of the tank which corresponds to the increase in the
laminar cases this movement must be created and accentuated by changing parameters
6.1 Introduction
To improve the level of mixing in the tank for lower Reynolds number it is necessary
to alter the flow patterns so that there is more energy being discipated in the top half
of the tank. There are several ways in which this can be achieved, two different
methods have been investigated, varying the flow rate of the jet and changing the
design of the inlet nozzle. Section 6.2 examines the improvements due to ramping the
flow rate or of applying a sinusoidal variation to the flow rate. Section 6.3 describes
As the tank is filled the energy from the jet dissipated per unit volume in the vessel
decreases. This means that the level of mixing in the tank is greater at the start of the
filling process than at the end, an effect that was demonstrated in Chapter 4. To
prevent this drop off in the mixing performance the flow rate of the jet could be
increased throughout the filling process. A case with this ramped flow was set up as
An alternative way of improving the mixing is to increase the perturbations in the top
half of the jet, this has been achieved by varying the flow rate as a sinusoidal, this
case was designed to provide fluctuations in the flow field that would increase the
levels of turbulence.
70
Figure 6.1 Mixing performance for three nozzles at a viscosity of 18 mPa s (a) log variance plotted
against fluid height (cm); Image of the vessel filled with (b) constant flow rate of Q = 3.33×10-5 m3 s-1
(c) sinusoidal flow rate varying between 1.67 x 10-5 m3 s-1 and 5 x 10-5 m3 s-1 (d) flow rate increasing
Figure 6.1 shows the log variance of these two scenarios next to a case with constant
flow rate at the average flow rate for each case (3.33 x 10-5 m3 s-1), the case at
constant flow rate reaches a log variance of -0.6 which is 50% mixed . Both the sine
and the ramped flow result in a marked improvement in the level of mixing within the
71
tank. For the ramped flow the low flow rate at the beginning of the fill means the
decrease in the log variance is more gradual throughout the process, the log variance
reaches approximately -1.4 which is 80% mixed. In contrast the sine function reaches
a log variance of -1.6 which 84% mixed. Figures 6.1c and 6.1d confirm this, the sine
case appears to be more homogeneous than the ramped case, however both are a large
The obstruction at the end of the pipe seen in section 4.3.2 resulted in a great
improvement in the level of mixing seen. To replicate this two nozzles were built
with a cone insert in the base as described in section 3.2.3. A third nozzle with a swirl
insert was also investigated; this nozzle was designed to give the jet angular
Figure 6.2 shows the log variance achieved in the tank for the two cone nozzles in
comparison with the simple straight flow that reaches 50% mixed. The mixing
performance has dramatically improved. The second cone which is the smaller of the
two shows the best improvement in the mixing performance reaching a log variance
of -1.7 (86% mixed), however due to the level of noise in the log variance signal it is
difficult to differentiate between the two cases. Comparison of the images in Fig
6.2.c and Fig 6.2.d does indicate that the smaller nozzle has produced a more
homogeneous vessel.
72
Figure 6.2 Mixing performance for three nozzles at a fill rate of Q = 3.33×10-5 m3 s-1 and a viscosity of
18 mPa s (a) log variance plotted against fluid height (cm); Image of the vessel filled with (b) straight
Figure 6.3 shows the improvement achieved using a swirl nozzle, this nozzle provides
a much greater improvement than the cone nozzles with the log variance reaching -2.6
(95% mixed) at points in the fill. Figure 6.3c shows that the fluid in the tank is much
more homogeneous than in Figures 6.2c and 6.2d. It appears that the angular
momentum applied to the fluid by the swirl insert in the end of the nozzle has
73
increased the level of mixing particularly in the top half of the tank, by directing the
Figure 6.3 Mixing performance for two nozzles at a fill rate of Q = 3.33×10-5 m3 s-1and a viscosity of
18 mPa s (a) log variance plotted against fluid height (cm); Image of the vessel filled with (b) straight
6.4 Conclusions
Improvements to the mixing performance have been made by both changing the inlet
pipe geometry and varying the inlet flow rate. Two flow rate variations were
investigated, ramped flow and a sinusoidal flow, the sinusoidal function variations in
the inlet velocity increased the level of mixing from 50% mixed so that the vessel
reached 84% mixed. Changing the nozzle design had a greater effect on the level of
mixing, three nozzle designs were examined, the most successful of these had a swirl
insert in the bottom of the pipe. With the swirl insert the level of mixing in the tank
reached 95% (log variance of -2.6) at one point in the fill, however this was not
combine these improvements, such as by using a sinusoidal flow rate through a swirl
nozzle. This may produce a more sustained and reliable improvement and should
Chapter 7 Conclusions
There is a large body of work into mixing by a range of mechanisms, however very
little has been carried out on the mixing that occurs as a vessel is filled. This mixing
process has been investigated using the techniques PLIF and PIV. These techniques
have been used to measure a variety of flow scenarios in the literature, however in
most cases the system is a steady state. This work has adapted these techniques to
The PLIF technique has been demonstrated to give good repeatability for the same
flow conditions and qualitative examination of the images has revealed a change in
the mixing mechanism from laminar to turbulent above a critical value of the
Reynolds number (Re~1000) which agrees with previous literature. For the cases
with a retracting pipe above Re~1000 the log variance reaches -2.2 in each case which
PIV was used to measure the vector fields in the tank, allowing the distribution of
momentum and shear in the tank to be investigated. For all flows dead regions were
observed in the bottom corners of the tank and the top half of the tank was relatively
quiescent, however this was more prominent for more laminar flows. For the more
turbulent flows (Re > 1000) there was a higher level of movement in the top half of
the tank. To improve the mixing performance in the laminar cases this movement
must be created and accentuated, changes to the flow rate and the nozzle design were
Sinusoidal function variations in the inlet velocity increased the level of mixing from
an unmixed highly striated case so that the vessel reached 84% mixed, however
changing the nozzle design had a greater effect on the level of mixing. Three nozzle
designs were examined, the most successful of these had a swirl insert in the bottom
of the pipe. With the swirl insert the level of mixing in the tank reached 95% (log
variance of -2.6) at one point in the fill, however this was not maintained throughout.
such as by using a sinusoidal flow rate through a swirl nozzle. This may produce a
more sustained and reliable improvement and should feature in any future work.
PLIF has proven to be a useful tool in measuring the mixing within a filling tank, the
development of the mixing analysis allows this technique to be used for other systems
There are a number of avenues that can be explored to improve the mixing in the tank
further; combining the variable flow rate with the nozzle design changes, changing
the tank geometry, adding internal baffles to the tank, adding the variant and the base
flow patterns into rules that will work with real products.
One option for the screening of different improvements to the flow is the use of
challenges to be overcome. The first is the difficulties in solving a flow with such
high viscous stresses and the accuracy of the results obtained, as the pressure
correction solver in the commercially available code often assumes that the viscous
stresses are much smaller than the inertial stresses. Furthermore while the mixing in
the tank can be modelled using the commercially available solvers the accuracy is
greatly dependent on the mesh. A very fine mesh is required as the lengthscale of the
mixing can only be determined down to the size of the mesh being used. Also the
mesh quality has to be high with predominantly hexahedral elements, and the time
step must be carefully chosen to reduce the numerical diffusion. These factors can
lead to long run times which reduces the effectiveness of CFD as a screening tool.
It would be a useful measure of the mixing improvement, gained through the nozzle
and flow rate changes made in this thesis, to test the techniques on real products.
Limited understanding of the flow structure would be gained due to the opaque nature
of the product but the level of mixing in the bottle can be measured using a dye and
measuring the colour of the resulting liquid. This would allow a correlation between
References
Bi, W., Sugii, Y., Okamoto, K., Madarame, H. (2003) "Time-resolved proper
14 L1-L5
methods." In: Handbook of industrial mixing: science and practice, Paul E.L.,
Atiemo-Obeng V.A., and Kresta S.M., eds., John Wiley & Sons Inc., New
Jersey, 145-256.
Chen, Y. C., Ma, C. F., Qin, M., and Li, Y. (2005). "Theoretical study on
3-D flow field inside miniature stirred vessels using a 2-D PIV technique."
Cozewith, C., and Busko, M. (1989). "Design Correlations for Mixing Tees."
the instability of the vortex ring. Part I: Linear phase." Experiments in Fluids,
40(3), 383-399.
79
Ding, R., Revstedt, J., Fuchs, L. (2003) "LIF Study of Mixing in Circular Impinging
France F4015
Etchells, A.W., Meyer, C.F. (2004). " Mixing in pipelines." In: Handbook of
industrial mixing: science and practice, Paul E.L., Atiemo-Obeng V.A., and
Kresta S.M., eds., John Wiley & Sons Inc., New Jersey, Ch 7, 391 - 478.
Fall, A., Lecoq, O., and David, R. (2001). "Characterization of mixing in a stirred
Feng, H., Olsen, M. G., Liu, Y., Fox, R. O., and Hill, J. C. (2005). "Investigation
2649-2664.
Fondse, H., Leijdens, H., and Ooms, G. (1983). "On the Influence of the Exit
Fosset, H., Prosser, L. E., (1949). “The application of free jets to mixing of fluids in
Forney, L. J., and Lee, H. C. (1982). "Optimum Dimensions for Pipeline Mixing at
Grenville, R.K., Nienow, A.W. (2004). " Blending of miscible liquids." In:
Obeng V.A., and Kresta S.M., eds., John Wiley & Sons Inc., New Jersey, Ch
9, 507 - 542.
Guillard, F., Fritzon, R., Revstedt, J., Tragardh, C., Alden, M., and Fuchs, L. (1998).
Guillard, F., Tragardh, C., and Fuchs, L. (2000). "A study on the instability of
Hall, J. F., Barigou, M., Simmons, M. J. H., and Stitt, E. H. (2004). "Mixing in
Hall, J. F., Barigou, M., Simmons, M. J. H., and Stitt, E. H. (2005). "Comparative
Houcine, I., Vivier, H., Plasari, E., David, R., and Villermaux, J. (1996). "Planar
Hult, J., Meier, U., Meier, W., Harvey, A., and Kaminski, C. F. (2005).
Kim, K. C., Min, Y. U., Oh, S. J., An, N. H., Seoudi, B., Chun, H. H., and Lee, I.
379.
Obeng V.A., and Kresta S.M., eds., John Wiley & Sons Inc., New Jersey, Ch
2, 19 – 87.
Kresta, S. M., Wood, P.E. (1993). “The flow field produced by a pitched blade
Kukukova, A., Aubin, J., and Kresta, S. M. (2009). "A new definition of mixing and
of the 3 Designs for Liquid Jet Mixing." Industrial & Engineering Chemistry
Mortensen, M., Orciuch, W., Bouaifi, M., and Andersson, B. (2004). "Mixing of a
Neal, G., Simmons, M. J. H., Hough, J. A., and Fryer, P. J. (2008). "Criteria for
Neal G.., Simmons M.J.H., Hough J.A., Fryer P.J. (2008) “Mixing in a vessel with a
Neal G., Simmons M.J.H., Hough J.A., Fryer P.J. (2009) “Mixing in a vessel with a
Engineering 2009.
Ottino, J. M., (1989). The kinematics of mixing: stretching, chaos, and transport
Pan, G., and Meng, H. (2001). "Experimental study of turbulent mixing in a tee
Paul, E.L., Atiemo-Obeng, V.A., and Kresta S.M. (2004). "Introduction" In:
Obeng V.A., and Kresta S.M., eds., John Wiley & Sons Inc., New Jersey.
Pawlak, G., Cruz, C. M., Bazan, C. M., and Hrdy, P. G. (2007). "Experimental
730.
Raffel, M., Willert, C. E., Wereley, S. T., and Kompenhans, J. (2007). Particle
Reungoat, D., Riviere, N., and Faure, J. P. (2007). "3C PIV and PLIF measurement
99-110.
Revill, B. K. (1985). "Jet Mixing." In: Mixing in the process industries, Harnby, N.,
practice, Paul E.L., Atiemo-Obeng V.A., and Kresta S.M., eds., John Wiley &
2486.
Weigand, P., Meier, W., Duan, X. R., Stricker, W., and Aigner, M. (2006).
1052-1066.
Zughbi, H. D., and Rakib, M. A. (2004). "Mixing in a fluid jet agitated tank: effects
of jet angle and elevation and number of jets." Chemical Engineering Science,
59(4), 829-842.
85
[M,K]=fcalibration(R,T);
for count=1:length(Irange)
cd 'E:\Experiments\0606PLIF8\4run\Image\Single' %the directory with the images in it
I=zeros(1016,1000);
I=imread (sprintf('4run0%04.4d.tif',Irange(count))); %reads the images
cd 'E:\mfiles' %changes to the directory with the mfiles in it
[A,Mi,Ki]=ffluidarea(I,vbase,vleft,vright,M,K);
clear I
[Cinf,H]=finfconc(A,R,Mi,Ki,T);
[C1,C2]=fnormconc(A,Ki,Cinf,inletleft,inletright);
clear A Cinf
[V,E]=flogsigmaRMS(C1,C2);
heightlogsigma(count,1)=H;
heightlogsigma(count,2)=V(1,1);
heightlogsigma(count,3)=V(1,2);
size(count,1)=E(1,1);
size(count,2)=E(1,2);
clear V Mi Ki C
count=count+1;
end
plot(heightlogsigma(:,1),heightlogsigma(:,2),'r',heightlogsigma(:,1),heightlogsigma(:,3),'b')
86
Associated functions
function [M,K,I]=fcalibration(R,T)
%a function to give the gradient and intercept of graph relating the
%concentration to the greyscale. The output is two mattrices with these
%values for each pixel in the field.
cd 'E:\Experiments\0606PLIF8\cal\Image\Single'
I=imread (sprintf('cal0%04.4d.tif',irange1(1)));
for count=2:length(irange1);
I1=imread (sprintf('cal0%04.4d.tif',irange1(count)));
I=cat(3,I,I1);
clear I1
count=count+1;
end
I1=mean(I,3);
clear I
I=imread (sprintf('cal0%04.4d.tif',irange2(1)));
for count=2:length(irange1);
I2=imread (sprintf('cal0%04.4d.tif',irange2(count)));
I=cat(3,I,I2);
clear I2
count=count+1;
end
I2=mean(I,3);
%clear I
K=2*I1-I2;
Conc=R/(T*T*height);
M=(I2-I1)/Conc;
function [A,Mi,Ki]=fluidarea(I,vbase,vleft,vright,M,K)
%program for focusing on the area that contains fluid in the
%vessel described in image I. This function takes the image determines the
%area with a greyscale 10% higher than the background greyscale and outputs
%a mattrix with greyscale values for the fluid area. It then scales the
%concentration gradient and intercept so that they are mattrices of the
%correct size.
B=mean(I(:,2));
cutoff=mean(B)+0.07*mean(B); %determines greyscale cut off value
%cutoff=77;
F=mean(G');
E=find(F>cutoff);%creates vector containing the rows above the cut off greyscale
A=G(min(E)+30:vbase,:);
Mi=M(min(E)+30:vbase,vleft:vright);%cuts the gradient matrix to size
Ki=K(min(E)+30:vbase,vleft:vright);%cuts the intercept matrix to size
end
function [Cinf,H]=infconc(A,Cinit,M,K,T)
%function to deterime what the final greyscale would be if the vessel was
%well mixed at that fill height. A is the fluid area as defined by the
%fluid area function, Cinit is the concentration of the 1ml of tracer
%intially placed in the bottom of the tank in mg. M is the gradient and K the
%intercept of the linear relationship between concentration(x)and greyscale(y),
%T is the tank width in centimetres.
[B,D]=size(A); % tells us the no of rows (B) and the number of columns (D)
end
function [C1,C2]=normconc(A,Czero,Cinf,inletleft,inletright);
%outputs a normalised concentration - Cinf is the greyscale that the
%vessel would be at if completely mixed at the fill height of that image,
%Czero is the greyscale value when the concentration is zero, A is the area
%of fluid that has been previously determined by the fluid area function
B=double(A);
Dt=(B-Czero);
Du=(Cinf-Czero);
Ct1=Dt(:,1:inletleft);
Ct2=Dt(:,inletright:end);
Cu1=Du(:,1:inletleft);
Cu2=Du(:,inletright:end);
Ct1=Ct1(:);
Cu1=Cu1(:);
y=find(Cu1<2);
Ct1(y)=[];
Cu1(y)=[];
C1=Ct1./Cu1;
Ct2=Ct2(:);
Cu2=Cu2(:);
88
y=find(Cu2<2);
Ct2(y)=[];
Cu2(y)=[];
C2=Ct2./Cu2;
end
function [V,E]=logsigmaRMS(C1,C2)
%function to determine the log of the RMS variance, C is the normalised
%concentration mattrix
A1=C1-1;
A2=C2-1;
B1=A1.^2;
B2=A2.^2;
D1=sum(B1);
D2=sum(B1);
E1=length(B1);
E2=length(B2);
D=[D1,D2];
E=[E1,E2];
V=log(D./E);
end
89
range1=00:299;
A=csvread
(sprintf('3ml02car7J00%04.4d.T000.D000.P001.H001.L.vec',range1(1)),1,0);%front vector
file behind which the others are stacked
l=2;
for l=2:length(range1)
B=csvread (sprintf('3ml02car7J00%04.4d.T000.D000.P001.H001.L.vec',range1(l)),1,0);
A=cat(3,A,B);
l=l+1;
end
clear B
cd ('F:\mfiles');
for l=1:(size(A,3))
B(:,3:4,l)=(A(:,3:4,l)-ubar(:,3:4)).^2;
l=l+1;
end
for l=1:length(D);
B(D(l),3:5,L(l))=0;%resets the velocities and the 5th column to zero for the false vectors
l=l+1;
end
C=sum(B,3);
urms2=C(:,3:4)./(C(:,5)*ones(1,2));%where this is the rms value squared
clear B C l
clear D L
%TKE
k=ones(length(A),3);
k(:,1:2)=A(:,1:2,1);
k(:,3)=0.75*(urms2(:,1).*urms2(:,2));
c=min(ubar(:,1));
b=max(ubar(:,1));
f=min(ubar(:,2));
g=max(ubar(:,2));
lim=[c-0.5 b+0.5 f-0.5 g+0.5];
clear k d e c b f g
%code to create average ubar, velocity magnetude field, and contour plot,
ubarg=ubar9J;
ubarj=ubar9g;
kj=K9g;
K=K9J;
limit=1.4;%speed in m/s of the cutoff point between the two versions of ubar and K;
B=find(U>limit);
ubar=ubarg;
for l=1:length(B);
ubar(B(l),:)=ubarj(B(l),:);
l=l+1;
end
ubar=ubar./10;
U=(ubar(:,3).^2+ubar(:,4).^2).^0.5;%0.5; give column with average speed with jet values for
larger numbers
91
U=reshape(U,size(K'));
B=find(U>limit);
K=K';
for l=1:length(B);
K(B(l))=kj(B(l));
l=l+1;
end
K=K'./100;
U=U';
K=flipud(K);
U=flipud(U);
contour(U)