GE Senograpge Crystal QC Manual
GE Senograpge Crystal QC Manual
Crystal
Quality Control Manual
5573538-1EN Revision6
English
Preface
Applicability of Manual
It describes Quality Control (QC) Tests for the Senographe Crystal, which constitutes the digital image
acquisition sub-system of a full-field digital mammography (FFDM) system.
WARNING
Indicates that there is an immediate danger that leads to death or serious physical injury.
CAUTION
Indicates a risk of danger that if disregarded, it leads or may lead to a potential situation that may result
in an undesirable result or state other than death, physical injury or property damage.
NOTE
Acronyms Glossary
This manual uses the following terminology:
Table of Contents
Preface ...........................................................................................................................................................................3
Applicability of Manual ........................................................................................................................................................................................3
Overview of this Manual ....................................................................................................................................................................................3
Definition of Warnings and Notes ........................................................................................................................................................................... 4
Acronyms Glossary.......................................................................................................................................................................................................... 5
Document Revision History......................................................................................................................................................................................... 6
SECTION 1: INTRODUCTION
1.1 Training
Basic training in digital mammography is required before becoming qualified to use the Senographe Crystal. The required
training is as follows.
Medical Physicist
Before using the digital mammography machine independently, a medical physicist should first meet all requirements such
as behavior regulations on breast quality including training on new mammogram machines.
Radiologic Technologist
The radiologic technologist should receive all necessary training, including the use of the machine, before using the digital
mammography alone.
1.2 Tools
Image Quality
Tools Test
Calibration PMMA Block (40mm)
Bar pattern
- Fluke Bio Medical 07-521
MTM100 Phantom
- CIRS Model 011A
Measuring Meters
Tools Test
X-ray Field Alignment Tool
- Unfors DXR+
Section3.3) Collimation Assessment
X-ray Dosimeter
- Radical9015RM
kVp Dosimeter
- Unfors XI-P (Compatible with W/Rh)
To display any of these pattern images, in the Browser tab, search for the following patient:
5. If you want to send the Raw image to PACS server, Click the Send icon.
6. If you want to print the Raw image, Click the Printer icon.
SETTINGS Mode
1. Start SenoCrystal S/W as administrator mode (Default administrator ID: crystal / PW: 1234)
3. Enter the desired maximum compression force. (The available range of Maximum compression force is 3~20kg).
4. To apply the changes, click the Apply button in the lower-left, and the present setting will be saved.
Capture Mode
3. Click the compression force icon and the popup windows appear.
4. Enter the desired maximum compression force. (The available range of Maximum compression force is 3~20kg).
2. Enter the administrator account of SenoCrystal S/W ID and PW to log into the SenoCrystal QAP S/W.
(Default administrator ID : crystal / PW : 1234)
5. Click the “Start” button and Input exposure condition on the right of CrystalQAP S/W
6. Take an exposure.
7. You can see that the acquired image (Raw image) and exposure condition are displayed.
NOTE
If you want to use raw images, you can confirm images to save in “C:\ImageData\YYYYMMDD
Introduction
QC tests for the Radiologic Technologist are listed below.
Daily
2) Purpose
To create the optimal environment for reading images, wipe the dust, fingerprints or other stains off the monitor.
3) Supplies to prepare
The Soft cloth or other cleaning tools for the monitor. If necessary, the cloth may be moistened with either water or
ethyl alcohol (up to 96%).
Notice: Do not use isopropyl (“rubbing”) alcohol. Do not use cleaning agents which attack the surface, such as
petroleum (mineral) spirits. The front panel is extremely sensitive to mechanical damage. Avoid all scratches,
knocks, etc. Do not apply the cleaning liquid directly to the monitor housing or screen. Do not allow the cleaning
liquid to enter the monitor housing; be sure to dampen the cloth sparingly.
4) Procedure
1. Use cloth to wipe dust, fingerprints or other stains off of the monitor.
2. Check for any other foreign substances on the monitor.
3. Take notes on the checklist of the items checked.
6) Maintenance Standards
The screen must be free from dust, finger prints, and other marks.
If these results are not obtained, the source of the problem must be identified, and corrective action taken, before
any further mammographic images are acquired using the Senographe Crystal.
Weekly
2) Purpose
To ensure optimal quality of the film printer output, follow the QC program developed by the manufacturer of the
device. Refer to the Printer Operator’s Manual.
3) Maintenance Standards
Weekly
2) Purpose
It is intended to guarantee the detector uniformity.
3) Supplies to prepare
Calibration PMMA Block (40mm)
4) Procedure
CAUTION
Recommend that operator need to firstly insert the PMMA block in the face shield rails then
remove the Bucky for protect the detector.
6. For start CrystalQAP S/W, Double-click the CrystalQAP icon on the desktop.
7. Enter the administrator account of SenoCrystal S/W ID and PW to log into the SenoCrystal QAP S/W.
(Default administrator ID: crystal / PW: 1234)
10. Click the “Start” button and then Current focus/Exposure Condition are displayed on the right of CrystalQAP S/W
– Current Focus: Large
- Current Exposure Condition: 28kVp / 100mAs
11. Take an Exposure.
12. Raw image is displayed and then results of the test are displayed.
(The results Pop-up windows contain the each Mean Values of 4 ROIs and Pass/Fail result.)
13. Note in Appendix 1. Mammography QA Checklist (Daily, Weekly) that the test was completed,
6) Maintenance Standards
1. If all tests are not passed, check the test conditions and repeat the test conditions and repeat the test.
-Ensure that the compression paddle and Bucky have been removed.
-Ensure that the C-arm angle is at zero degrees.
-Ensure that the PMMA block is clean and free from scratches.
-Ensure that the surface of the image receptor is clean.
2. If the re-test result does not meet the requirements, contact the technical support of the manufacturer to have the
system correct.
Weekly
2) Purpose
The test is designed to ensure adequate and consistent quality of the images acquired by the detector.
3) Supplies to prepare
ACR Phantom
4) Procedure
5) Analysis
6) Action limit
The score for fibers must be at least 4, the score for masses must be at least 3, and the score for speck groups must
be at least 3.
The ACR phantom is an acrylic block containing 16 cells as below image. Each cell simulates an anatomical
structure which may be found in breast tissue. Starting from the top left hand corner, as shown in the illustration, the
cells are of three types:
8) Maintenance Standards
If it doesn't meet the requirements, try retaking the image. If the re-imaging result does not meet the requirements,
contact your GE Medical System Service Engineer to have the system corrected.
Weekly
2) Purpose
To ensure good image review conditions by keeping the View box free of dust, finger prints, and other marks and
the viewing conditions optimized.
3) Supplies to prepare
Glass cleansing solution
Soft cloth
4) Procedure
1. Remove all the dust, finger prints, and other marks the surface of the reading box by using a glass cleansing
solution and a soft cloth.
2. Check if the brightness of the reading box is uniform.
3. Check if the switches or devices of the reading box works normally.
4. Check the degree of indoor lighting, check if there are any light-producing materials in the room, or any lights
reflecting off of the surface of the reading box.
5. Check the lighting, ventilation, temperature and noise in the image reading room to make sure that no unnecessary
elements exist.
6. Take notes of maintenance results and save them.
6) Action Limit
The viewbox must be free from dust, finger prints, and other marks. Viewing conditions must meet accepted
standards for mammographic image review.
7) Maintenance Standards
If these results are not obtained, the source of the problem must be identified and corrective action taken, before
any further mammographic images are reviewed or interpreted using the view boxes.
Weekly
2) Purpose
The test is designed to ensure that contrast is adequate over the 0-5 lp/mm spatial frequency range by obtaining
an estimate of the MTF (Modulation Transfer Function) values at 2 and 4 lp/mm.
3) Supplies to prepare
Bar pattern
DICOM Viewer including ROI measuring function (ex. ImageJ)
4) Procedure
60mm
<MTF parallel>
5. Make an exposure in Manual Mode (28kV, 80mAs).
6. Place the Bar pattern on the surface of Bucky as picture below.
a. Orient the pattern’s bars perpendicular to the anode-cathode axis (parallel to the chest wall edge of the Bucky)
b. Position the edge of the bar pattern 60 mm from the chest wall surface of Bucky.
60mm
9. Open the (4) Raw image (Refer to section 1.4 How to locate Raw files)
10. Adjust the image size for optimum visibility by using “Magnifying glass” icon ( ).
11. Click the “Elliptical” icon ( ) to make ROI measurement.
<MTF paralle>
13. Open the step(6) Raw image and Repeat the perpendicular MTF measurement in the same way as described above.
<MTF Perpendicular>
6) MTF Calculations
Use the following equations to estimate the MTF for the two frequency points :
𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆2𝑙𝑙𝑙𝑙/𝑚𝑚𝑚𝑚
𝑀𝑀𝑀𝑀𝑀𝑀2𝑙𝑙𝑙𝑙/𝑚𝑚𝑚𝑚 = × 222
𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠 − 𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑏𝑏𝑏𝑏𝑏𝑏
𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆4𝑙𝑙𝑙𝑙/𝑚𝑚𝑚𝑚
𝑀𝑀𝑀𝑀𝑀𝑀4𝑙𝑙𝑙𝑙/𝑚𝑚𝑚𝑚 = × 222
𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠 − 𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑏𝑏𝑏𝑏𝑏𝑏
7) Action Limit
The system passes the MTF Measurement test if all of the following conditions are met:
- MTF Parallel at 2 lp/mm > 50% / MTF Parallel at 4 lp/mm > 15%
- MTF Perpendicular at 2 lp/mm >50% / MTF Perpendicular at 4 lp/mm > 15%
For each condition that is met, the Status will be Passing
8) Maintenance Standards
Failure of MTF Measurement Test: If the system fails the MTF Measurement Test, you should ensure that all test
conditions are met and then repeat the test. If the system still fails, it is recommended that you contact your GE
Medical Systems Service Engineer.
-
Monthly
2) Purpose
It is intended to find out if the system displays, locking devices, and mechanical parts work well and if the system is
mechanically stable.
3) Supplies to prepare
None
4) Procedure
Check the system parts below to determine normal working order and take note of the results in the checklist.
Base display
: Check that all base display values are consistent with C-arm rotation, compression force and compression
thickness.
Locks (all)
: Check if the device stops properly when releasing rotation and lift motion buttons and footswitches.
Smoothing of motion
: Check if the device moves smoothly.
Check all Paddles for cracks.
Paddle movement
: Check Paddle movement by using keypad and pressing Footswitch.
6) Maintenance Standards
If you find any inappropriate results in the checklist, contact your GE Medical System Service Engineer to have the
system corrected.
Quarterly
For the repeat rates to be meaningful, an analysis period that yields a patient volume of at least 250 patients or
1,000 exposures is needed.
2) Purpose
Determining the number and cause of re-imaging and problem images is intended to improve efficiency, reduce
cost, and reduce radiation exposure to patients.
3) Supplies to prepare
None
4) Procedure
1. Collect the re-taken images for the period to be checked.
2. Classify and divide the images based on the reason re-imaging and note their numbers in the checklist.
3. Calculate the total number of re-imaged and rejected images.
4. The percentage of re-taken images can be found as follows:
6) Action limit
The total repeat rate or reject rate must not change by more than 2.0% of the total exposures included in the
analysis from the rate determined for the previous analysis period.
7) Maintenance Standards
The images taken for quality control are not counted in the re-imaging control analysis.
If the total repeat rate or reject rate changes from the rate determined for the previous analysis period by more
than 2.0% of the total exposures included in the analysis, the source of the problem must be identified, and
corrective action taken, within 30 days of the test date. Any corrective actions taken must be recorded, and an
assessment must be made of their effectiveness.
Quarterly
2) Purpose
It is intended to keep the brightness and contrast consistent for the image reading monitor.
3) Supplies to prepare
4) Procedure
1. Start SenoCrystal and click the ”Browser” Tap
2. Enter “TestQC” at the Patient Name blank or “TG18PQT” at Patient ID and click the “Search Patient” button.
3. Open the SMPTE pattern Image.(Refer to Section 1.5 “How to open Pattern Image)
4. Send the SMPTE pattern image to image reading monitor.
6. Artificial Shade
When any artificial shade is shown in the pattern, take note of it and adjust it accordingly.
7. Geometrical Distortion
The pattern has horizontal and vertical white lines. The square should be seen as a regular tetragon, and must
have no distortions. If there is any protruding or concave lines, mark the degree of the greatest distortion in mm.
The lines cannot stray more than 5mm anywhere on the monitor.
8. Spatial Resolution
The four corners and the pair of white and black lines in the light green square in the center of the pattern must be
distinguishable. The six sections must be clearly distinguishable, and not a single one of the five cells can have a
problem.
9. Open the TG18-UNL10 and TG18-UNL80 pattern on the image reading monitor.
Semi-annually
2) Purpose
It is intended to find out if the mammography compression Paddle provides the proper amount of compression and not
excessive compression.
3) Supplies to prepare
Scale
Towels
4) Procedure
1. Install the compression paddle and Bucky.
2. Put a towel on the Bucky to protect the system, and put the scale on top of that. Place the scale over the center of
the compression paddle.
3. Put a towel on the scale to prevent the compression paddle from breaking it.
6) Action limit
The maximum compression force for the initial power drive must be between 11 and 20 kg
Semi-annually
2) Supplies to prepare
Calibration PMMA Block (40mm)
Blank CD or DVD
Flat field test software (available in EUREF website)
3) Procedure
3. Change Small Focus to Large Focus by toggling the focal spot button.
This is because the focal spot is automatically changed to ‘Small Focus’ after the Bucky is removed.
4. The test configuration is as below for finding Uncorrected Defective Pixels (aka, UDP). This test is performed using
Calibration PMMA block in the Manual Exposure Mode (28kV, 100mAs).
5. Take an exposure to get an image.
6. Acquire 4 more images. Note that PMMA blocks should be detached and attached for each image acquisition.
7. Collect 5 raw image files and burn them into CD or DVD. (Refer to Secction1.4 How to locate Raw files)
8. Launch the flat field test software, and set the parameters as below.
9. Click “Open DICOM File” button and select the image to get UDP analysis report.
10. Click “Output File” button, and set the destination directory and file name (e.g. 1.txt) to save analysis report.
11. Click “Calculate” button to start UDP analysis.
12. Repeat 4 more times from step 4 to 6 to analyze each of all 5 images.
13. Open 5 reports and check if there is any defective pixel. A pixel is considered defective if its coordinates are the
same on 5 images.
14. Pass the test if there is no defective pixel. Below is an example of a passing result with zero defective pixel.
Introduction
QC Test for the medical physicist is as follows:
CAUTION
When test procedures are run which do not require the capture of X-rays by the detector and
require the presence of an object in the X-ray beam (e.g., the measurement of dose using a
dosimeter), the detector must be protected by a Lead plate or equivalent attenuator. This will prevent
any possibility of ghost images.
Weekly
2) Purpose
To assure consistency of the digital image receptor by evaluating the signal-to-noise ratio(SDNR).
3) Supplies to prepare
MTM 100 Phantom
4) Procedure
1. Install Bucky, standard compression paddle and normal collimation plate.
2. Start the SenoCrystal S/W and Click the “Manual Entry” to create new study.
3. Click 'Capture' button to enter Capture mode.
4. Place the MTM100 phantom on surface of Bucky.
5. Apply a compression force of 5kg to the MTM100 Phantom.
6. Take an exposure in Auto Mode.
7. Close Capture mode and enter “Viewer mode” to measure the SDNR on acquired image.
8. Open the raw image. (Refer to Section.1.6 How to review the Raw image) and then adjust the windowing level.
9. Click the SDNR icon( )near the middle left of the screen.
10. Click the 2 ROI as picture below and then appear the SDNR value on screen.
11. Record the SDNR value in Appendix 5. SDNR Measurement Test Record (weekly)
5) Action Limit
Semi-annually
2) Purpose
It is intended to keep the brightness and contrast consistent for the imaging monitor.
3) Supplies to prepare
SMPTE Pattern (Off-the-shelf)
4) Procedure
1. Start SenoCrystal and click the “Browser “Tap to open the pattern images.
2. Enter “TestQC” at the Patient Name blank or “TG18PQT” at the Patient ID and click the “Search Patient” button.
(Refer to Section 1.5 “How to open the Pattern Image)
3. Open SMPTE pattern on the imaging monitor.
Annually
2) Purpose
Check the difference between the X-ray field and Light field, and the X-ray field and Image Receptor to determine if
the proper X-ray field is being maintained.
3) Supplies to prepare
1 sheet of white paper
4 coins of the same size
Radiation Protection Plate (in a size that can cover the detector completely)
X-ray Light Field Alignment tool
4) Procedure
Check the difference between the X-ray and Light Fields
1. Install Bucky and remove the compression paddle.
2. Place the Radiation Protection Plate on the surface of the Bucky to prevent detector ghost.
3. Start SenoCrystal and create a random patient as “Manual Entry”.
4. Click 'Capture' button to open 'Study'.
5. Press the footswitch slightly to turn the light on.
6. Fit the X-ray Light Field Alignment Tool to the edge of the collimator shadow. Use the white paper at this time to fit
the tool in more easily.
9. Make an exposure for the remaining three edges and note the results in the checklist.
Check the difference between Light Field and Image Receptor
1. Install Bucky and remove the compression paddle.
2. Start SenoCrystal and create a random patient as “Manual Entry”.
3. Click 'Capture' button to open 'Study'.
4. Press the footswitch slightly to turn the light on.
5. Place four coins inside the collimator. Each coin has to touch one of the edges of the collimator shadow.
6. Use the white paper at this time to fit the tool in more easily.
<Coin Positioning to Check the Difference between Light Field and Image Receptor>
7. Make an exposure as manual mode 28kVp and 50mAs.
8. Close the Capture mode and Enter the viewer mode to display the captured image.
9. Check the difference between Light Field and Image Receptor
10. Measure how many the coins on each edge come off of the image, and note the results in the checklist.
11. Check the difference between the X-ray field and the image receptor
12. Measure how many the coins on each edge come off of the image, and note the results in the checklist.
6) Action Limit
1. Congruence of the light field with the X-ray field must be such that the total misalignment (sum of misalignments
on opposite sides) is within 2% of the SID.
2. The difference of X-ray field and image receptor should not exceed 1% of SID to each edge
7) Maintenance Standards
If the result does not meet the requirements, contact your GE Medical System Service Engineer to have the system
corrected.
Annually
2) Purpose
To evaluate focal spot performance by using a high-contrast resolution pattern
3) Supplies to prepare
Resolution Test Pattern
4) Procedure
6) Action Limit
Measurements made with the bars parallel to the anode-cathode axis must be at least 7 lp/mm; measurements with
the bars perpendicular to the anode-cathode axis must be at least 5 lp/mm.
7) Maintenance Standards
If it doesn't meet the requirements, try retaking the image. If the re-imaging result does not meet the requirements,
contact your GE Medical System Service Engineer to have the system corrected.
Annually
2) Purpose
It is intended to check if the selected tube voltage is accurately measured in the limited range and if it shows
consistent tube voltage in repetitive imaging.
3) Supplies to prepare
kVp meter
Radiation Protection Plate (in a size that can cover the detector completely)
4) Procedure
1. Install Bucky and remove the compression paddle.
2. Turn off ‘Auto Release’ mode.
3. Place the Radiation Protection Plate on the surface of the Bucky to prevent detector ghost.
4. Place the kVp meter probe 6 cm away from the center of the chest wall.
5. Start SenoCrystal and create a random patient as “Manual Entry”.
6. Click 'Capture' button to open 'Study'.
7. Take an exposure in Manual Mode (28kV, 100mAs).
8. Note the measured kVp value in the checklist.
9. Note in the checklist of the kVp values measured by taking three more exposures in the same conditions.
10. Exit SenoCrystal.
6) Action Limit
- The kVp shall be accurate within ± 5 percent of the indicated or selected kVp at each of the above measurement
points.
- At the most commonly used clinical setting of kVp, the coefficient of variation of reproducibility of the kVp shall be
equal to or less than 0.02
7) Maintenance Standards
- If it doesn't meet the requirements, try retaking the image. If the re-imaging result does not meet the requirements,
contact your GE Medical System Service Engineer to have the system corrected.
Annually
2) Purpose
It is intended to check if HVL (Half Value Layer) of the X-ray beam is suitable to minimize the glandular dose of a
patient without reducing contrast
3) Supplies to prepare
X-ray Dosimeter
5 - 6 thin plates with 0.1 mm thickness consisting of 99.9% pure aluminum (1145-Type aluminum alloy) or 99%
pure aluminum (1100-Type aluminum alloy) that can cover the collimator completely. The use of type 1100
aluminum alloy for HVL measurement can give (depending on specific samples) HVL values up to 7.5% lower than
those measured using pure aluminum. If type 1100 aluminum is used, results should be corrected to agree with
those obtained using pure aluminum.
Radiation Protection Plate (in a size that can cover the detector completely)
Tapes to attach the thin aluminum plates
4) Procedure
1. Install Bucky and remove the compression paddle.
2. Turn off ‘Auto Release’ mode.
3. Place the Radiation Protection Plate on the surface of the Bucky to prevent detector ghost.
4. Place the dosimeter 6cm away from the center of the chest wall.
5. Start SenoCrystal and create a random patient as “Manual Entry”.
6. Click 'Capture' button to open 'Study'.
7. Make an exposure as manual mode 26kVp and 80mAs and note the dose in the checklist.
8. Note : If dosimeter can directly measures HVL, measure it directly instead of using the aluminum thin plates and
skip to analysis chapter
9. Attach the thin aluminum plates with the 0.2 mm thickness in front of the collimator.
10. Make an exposure as manual under the same condition and note the dose in the checklist.
11. Add thin aluminum plates with a thickness of 0.1 mm in front of the collimator, and note the value of the dose every
time until it drops to less than half of the original exposure dose (this is the does without the aluminum plates
inserted between the X-ray tube and the ionization chamber).
12. Repeat these 11 step 3times and note the dose in the checklist
13. Exit SenoCrystal.
5) Analysis
To calculate the Half-Value Layer (HVL) by logarithmic interpolation, use the following notation and procedure.
- Denote the direct exposure reading, without any added aluminum, as𝐸𝐸0 .
- Divide this value in half and find the two exposure readings and added aluminum thicknesses that bracket the
𝐸𝐸0 /2 exposure.
- Let 𝐸𝐸𝑎𝑎 be the exposure reading that is just greater than one-half of 𝐸𝐸0 and ta be the corresponding aluminum
thickness.
- Let 𝐸𝐸𝑏𝑏 be the exposure reading that is just less than one-half of 𝐸𝐸0 and 𝑇𝑇𝑏𝑏 the corresponding aluminum
thickness.
7) Action Limit
kV/100 + 0.03 (in mm Al) ≤ Measured HVL ≤ kV/100 + 0.3(in mm Al)
8) Maintenance Standards
If it doesn't meet the requirements, try retaking the image. If the re-imaging result does not meet the requirements,
Contact your GE Medical System Service Engineer to have the system corrected.
Annually
2) Purpose
It is intended to measure the typical incident exposure dose from a normal patient (about 4.2 cm compressed
breast thickness - 50% fatty substance, 50% mammary alveolar system) to calculate the corresponding average
glandular dose.
3) Supplies to prepare
ACR Phantom
X-ray Dosimeter
4) Procedure
1. Install the compression paddle and Bucky.
2. Start SenoCrystal and create a random patient as “Manual Entry”.
3. Click 'Capture' button to open 'Study'.
4. Place the ACR Phantom on the center of the surface of Bucky and align edge with chestwall
5. Apply a compression force of 5kg to the ACR Phantom.
6. Take an exposure in AEC Mode, and note the conditions.
7. Turn off ‘Auto Release’.
8. Set the compression paddle to 4.5 cm high by pressing the two footswitches simultaneously.
9. Place the dosimeter probe 6cm away from the chest wall, with the upper part of the phantom level with the center
of the dosimeter.
10. Select the Manual mode and set the kV, mAs as close as possible to the kV, mAs used in Auto Mode. (Refer to
Section “1.8 How to use QC Exposure tool”)
11. Take four manual exposures.
12. Note the dose values in the checklist.
13. Exit SenoCrystal.
5) Analysis
1. Take note of the average value of the measured dose value in mGy. 1R in the air is 8.73mGy.
2. Note the average glandular dose by using the formula below.
𝑚𝑚𝑚𝑚𝑚𝑚
𝐀𝐀𝐀𝐀𝐀𝐀 (𝐦𝐦𝐦𝐦𝐦𝐦) = 𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴 𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀 (𝑚𝑚𝑚𝑚𝑚𝑚)𝑥𝑥 𝐷𝐷𝐷𝐷𝐷𝐷( )
𝑚𝑚𝑚𝑚𝑚𝑚
3. Refer to the table below.
HVL DgN(mGy/mGy)
(mm) : Average Glandular Dose conversion factor W-Rh (50 μm thick)
0.338 0.184
0.365 0.206
0.392 0.227
HVL DgN(mGy/mGy)
(mm) : Average Glandular Dose conversion factor W-Rh (50 μm thick)
0.420 0.248
0.444 0.267
0.462 0.279
0.477 0.289
0.489 0.296
0.500 0.304
0.509 0.310
0.518 0.315
0.527 0.321
0.535 0.327
0.544 0.332
0.552 0.338
0.560 0.344
0.569 0.350
0.577 0.356
0.585 0.362
0.593 0.369
0.601 0.374
4. When an applicable HVL is unavailable for the Average Glandular Dose conversion factor, use it by interpolation.
(Use the value of HVL(mm Al) calculated in Section 3.6 Beam Quality Assessment(HVL)
7) Action Limit
- The mean glandular dose to the Standard breast must not exceed 3 mGy (0.3 rad) per view.
8) Maintenance Standards
- If it doesn't meet the requirements, try retaking the image. If the re-imaging result does not meet the requirements,
contact your GE Medical System Service Engineer to have the system corrected.
Annually
2) Purpose
To assess the degree and source of artifacts visualized in full field digital mammograms or phantom
3) Supplies to prepare
Calibration PMMA Block(40mm)
4) Procedure
5) Analysis
1. Open the raw file of the image in Viewer mode.(Refer to section 1.6 How to review the raw image)
2. Check for any artifacts in the image by using Zoom and Windowing tools. Examine the image for gridlines, plus or
minus signal variations, streaking in the horizontal or vertical directions, residual images left over from repeated
test or clinical exposures, acrylic sheet artifacts, or bad pixels.
3. Take note of the results in the checklist.
7) Action Limit
- There must be no artifact or non-uniformity that is expected to mimic or obscure clinical information.
8) Maintenance Standards
- If it doesn't meet the requirements, try retaking the image. If the re-imaging result does not meet the requirements,
contact your GE Medical System Service Engineer to have the system corrected.
Annually
2) Purpose
It is intended to check the accuracy and reproducibility of the conditions in AEC (Automatic Exposure Control) Mode.
3) Supplies to prepare
4) Procedure
5) Analysis
1. Compute the average value and the standard deviation of mAs values for each exposure.
2. Compute the average value and the standard deviation of the pixel values, recorded for each image.
3. Note the standard deviation divided by the average value as a % value in the "coefficient of variation" of the
checklist.
𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆 𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑(𝑠𝑠𝑠𝑠)
𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂 𝐨𝐨𝐨𝐨 𝐯𝐯𝐯𝐯𝐯𝐯𝐢𝐢𝐚𝐚𝐚𝐚𝐚𝐚𝐚𝐚𝐚𝐚 (%) = × 100
𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀 (𝑀𝑀)
7) Action Limit
- The coefficient of variation of mAs and average pixel values must not exceed 5%
8) Maintenance Standards
- If it doesn't meet the requirements, try retaking the image. If the re-imaging result does not meet the requirements,
contact your GE Medical System Service Engineer to have the system corrected.
SECTION 4: APPENDIX
Date 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
Monitor cleaning
(Daily)
Viewboxes and
Viewing Conditions
(Weekly)
Flat FieldTest
(Weekly)
Year Month
Date 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
Monitor cleaning
(Daily)
Viewboxes and
Viewing
Conditions
(Weekly)
Flat FieldTest
(Weekly)
Year
Month
Date
Control
Operating Level
Limit
Mid Density
±0.15
Value (40%)
Density
Difference
±0.15
Value
(10%-40%)
Lower Density
±0.15
Value (90%)
PASS / FAIL
Month
Date
Fibers
Speck
groups
Masses
PASS/FAIL
• Action Limit:
If this condition is met, the Result will be Passing
ACR Mammography
Fibers Speck Groups Masses
Accreditation Phantom
Minimum Passing Score 4.0 4.0 3.0
Month
Date
MTF Parallel
Mean Sd. MTF (%) PASS / FAIL
Space
Bar
MTF 2lp/mm
MTF 4lp/mm
MTF Perpendicular
Mean Sd. MTF (%) PASS / FAIL
Space
Bar
MTF 2lp/mm
MTF 4lp/mm
• MTF Calculations:
𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆2𝑙𝑙𝑙𝑙/𝑚𝑚𝑚𝑚
𝑀𝑀𝑀𝑀𝑀𝑀2𝑙𝑙𝑙𝑙/𝑚𝑚𝑚𝑚 = × 222
𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠 −𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑏𝑏𝑏𝑏𝑏𝑏
𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆4𝑙𝑙𝑙𝑙/𝑚𝑚𝑚𝑚
𝑀𝑀𝑀𝑀𝑀𝑀4𝑙𝑙𝑙𝑙/𝑚𝑚𝑚𝑚 = × 222
𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠 − 𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑏𝑏𝑏𝑏𝑏𝑏
• Action Limit:
The system passes the MTF Measurements test if all of the following conditions are met:
- MTF Parallel at 2 lp/mm > 50%/MTF parallel at 4 lp/mm > 15%
- MTF Perpendicular at 2 lp/mm > 50%/MTF Perpendicular at 4 lp/mm > 15%
Month
Weekly
SDNR
Value
PASS/
FAIL
• Action Limit:
The system passes the SDNR Measurement test:
- If the change in SDNR does not exceed 0.2 when computed with on existing SDNR value, such that
change in SDNR = l SDNR – SDNR𝐨𝐨𝐨𝐨 l / SDNR𝐨𝐨𝐨𝐨
- If this condition is met, the Status will be Pass
Month
Date
Gantry
Base display
Locks (all)
Collimator Lamp
Smoothness of
motion
Inspect all paddles
for cracks
Switches
(Keypad, Footswitch)
Image Artifact
Gird Artifact
Positioning
Motion Blur
Under-exposed
Over-exposed
No Image
Detector Artifact
Artifact other than grid or
detector artifact
Mechanical Failure
Electrical Failure
Software Failure
Inappropriate Image
Processing
Unknown Failure
Double Exposure
Others
Total Repeats
Total Rejects
• Calculation:
100 𝑋𝑋 𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇 𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅 + 𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅
𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇 𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅 + 𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅 𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟(%) =
𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇 𝑛𝑛𝑛𝑛𝑛𝑛𝑛𝑛𝑛𝑛𝑛𝑛 𝑜𝑜𝑜𝑜 𝑒𝑒𝑒𝑒𝑒𝑒𝑒𝑒𝑒𝑒𝑒𝑒𝑒𝑒𝑒𝑒
PASS / FAIL
• Calculation:
(𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝑚𝑚𝑚𝑚𝑚𝑚 − 𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝑚𝑚𝑚𝑚𝑚𝑚 )
𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿 𝑈𝑈𝑈𝑈𝑈𝑈𝑈𝑈𝑈𝑈𝑈𝑈𝑈𝑈𝑈𝑈𝑈𝑈𝑈𝑈 = × 100 %
𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐
• Action Limit:
In terms of luminance uniformity, the luminance value measured at the 4 corners should be over 70% of the value at
the center
Year
The half year The first half year The second half year
Date(mm/dd)
PASS / FAIL
• Action Limit:
The maximum compression force for the initial power drive must be between 11 and 20 kg.
The half year The first half year The second half year
Date (mm/dd)
0 ~ 5 % Patch
95 ~ 100 % Patch
0 ~ 10 % Patch
The half year The first half year The second half year
Number of Uncorrected
Defective Pixels
PASS/Fail
• Action Limit:
The number of uncorrected defective pixel must be “0”.
Sum as % of SID
Sum as % of SID
Pass/ Fail
PASS / FAIL
• Action Limit:
o Deviation between X-ray field and Light field
: Sum of left plus right edge deviation or anterior plus chest edge deviation must be less than 2% of
SID.
o Deviation between X-ray field and edges of the image receptor:
: Deviation must be less than 1% of SID
Date (mm/dd)
KVp
mAs
PASS / FAIL
• Action Limit:
The system passes the Focal Spot Performance test if all of the following conditions are met:
≥5lp/mm for line-pair bars perpendicular to anode-cathode axis
≥7lp/mm for line-pair bars parallel to anode-cathode axis.
Date (mm/dd)
% Error
Coefficient of Variation
(Sd/Mean kVp)
PASS / FAIL
• Action Limit
- The mean kVp should not differ from the nominal by more than ±5% of the nominal kVp
- The coefficient of variation should not exceed 0.02
Date (mm/dd)
Calculations
E0=(E0a+E0b)/2
E1/2=E0/2
Al thickness at Ea: ta
Al thickness at Eb: tb
Minimum allowed
PASS / FAIL
• Calculations:
𝑇𝑇𝑏𝑏 *ln[2𝐸𝐸𝑎𝑎 /𝐸𝐸0 ] – 𝑇𝑇𝑎𝑎 *ln[2𝐸𝐸𝑏𝑏 /𝐸𝐸0 ]
x 100
HVL(mmAl)= ln[𝐸𝐸𝑎𝑎 /𝐸𝐸𝑏𝑏 ]
Date (mm/dd)
kVp
mAs
Measured
HVL(mm Al)*
DgN*
(mGy/mGy)
Mean(mGy)
AGD(mGy)*
PASS/FAIL
Date (mm/dd)
KVp
mAs
Artifact (Y/N)
Description
PASS / FAIL
Date (mm/dd)
Acrylic plates
Thickness 20 mm 40 mm 60 mm
(mm)
Exposure 1
Exposure 2
Exposure 3
Exposure 4
Average
Mean(M)
Average
Standard
deviation (Sd)
Coefficient of
variation
(Sd/M, %)
PASS/FAIL
• Calculations:
𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆 𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑(𝑠𝑠𝑠𝑠)
𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂𝐂 𝐨𝐨𝐨𝐨 𝐯𝐯𝐯𝐯𝐯𝐯𝐯𝐯𝐯𝐯𝐯𝐯𝐯𝐯𝐯𝐯𝐯𝐯 (%) = × 100
𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀 (𝑀𝑀)
• Action Limit:
- The coefficient of variation shall not exceed 5%.