Update Instructions Refinement - CT036/09/R: Hotfixes 2008 Function Pack Update

This document provides update instructions for the Emotion (2007)/Spirit system, detailing the installation of various Windows XP Hotfixes. It includes information on affected systems, prerequisites, and step-by-step work instructions for the update process. Additionally, it outlines the necessary materials, customer training, and completion protocol for the update.

Uploaded by

Rogerio Castro
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
3 views11 pages

Update Instructions Refinement - CT036/09/R: Hotfixes 2008 Function Pack Update

This document provides update instructions for the Emotion (2007)/Spirit system, detailing the installation of various Windows XP Hotfixes. It includes information on affected systems, prerequisites, and step-by-step work instructions for the update process. Additionally, it outlines the necessary materials, customer training, and completion protocol for the update.

Uploaded by

Rogerio Castro
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 11

Emotion (2007)/Spirit

Update Instructions
Refinement - CT036/09/R
Hotfixes 2008 Function Pack update

Urgent? Q Yes H No
Time period within 3 months
Intranet download available? Q Yes H No
Remote update handling feasible? Q Yes without CSE Q Yes with CSE H No
(Category 2) (Category 1/3)
Materials for logistic update re‐ H Yes Q No
quired?
Materials free of charge? H Yes Q No
Q No, but credit at re‐
turn
Return of parts? Q Yes H No
Estimated completion time 0,5 hour(s) Number of CSE’s: 1
Customer application training? Q Yes H No

Criteria for system to be affected

Criteria Value
System material number 10045692, 10165566, 10165888, 10165977
System software version VB27, VB35, VB38, VB39
Component material number n.a.

Chg. Ref. No.: 534758


Name: Yan Ge
Dept.: SSME CS

Answers for life.


2 Document Version / Disclaimer / Copyright

Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents cor‐
respond to the version at the time of system delivery and/or printout. Versions to hard‐
copy documentation are not automatically distributed.
Please contact your local Siemens office to order a current version or refer to our website
http://www.healthcare.siemens.com.

Disclaimer
Siemens provides this documentation “as is” without the assumption of any liability under
any theory of law.
The content described herein shall be used by qualified personnel who are employed by
Siemens or one of its affiliates or who are otherwise authorized by Siemens or its affiliates
to use such documents.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are not entitled to use this documen‐
tation without prior written authority.

Copyright
“© Siemens, 2009” refers to the copyright of a Siemens entity such as:
Siemens Healthcare GmbH - Germany
Siemens Aktiengesellschaft - Germany
Siemens Shenzhen Magnetic Resonance Ltd. - China
Siemens Shanghai Medical Equipment Ltd. - China
Siemens Medical Solutions USA Inc. - USA
Siemens Healthcare Diagnostics Inc. - USA and/or
Siemens Healthcare Diagnostics Products GmbH - Germany

Emotion (2007)/Spirit C2-019.897.51.01.02 Page 2 of 11 © Siemens, 2009


07.09 SSME CS Restricted
Table of Contents 3

1 Preparations 4

1.1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4


1.1.1 Systems/products affected . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.2 Reason for the update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.3 Additional information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.4 Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.5 Special tools / documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Information on materials. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2.1 Content of the Update Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2.2 Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2.3 Return of parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

2 Update 6

2.1 Work steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6


2.1.1 Installation on the ICS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.1.2 Installation on the IES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.1.3 Final check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1.4 Backup in the ICS / IES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1.5 Informing the customer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1.6 Final work steps. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1.7 Changes to previous version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

3 Completion protocol 9

4 List of Hazard IDs 10

© Siemens, 2009 C2-019.897.51.01.02 Page 3 of 11 Emotion (2007)/Spirit


Restricted 07.09 SSME CS
4 1 Preparations

1.1 General information

1.1.1 Systems/products affected


Affected system numbers may be subject to change; the UI data published on the Intranet
(update handling tool) always reflect the current status. UI hardcopies are not updated
when the affected serial numbers change.

1.1.2 Reason for the update


This Update Instructions installs various Windows XP Hotfixes.

1.1.3 Additional information


The name of the function pack depends on the SW version on which it is installed.
After a successful installation, it will show up as follows:
VB27: FP3
VB35: FP1
VB38: FP3
VB39: FP3
The update instructions are identical for all the versions; no selection has to be performed
during the installation.
The function pack will install the Microsoft Hotfixes 08-061, 08-062, 08-063, 08-067.

1.1.4 Prerequisites
This function pack does not require any previous packs.

1.1.5 Special tools / documents


n.a.

Emotion (2007)/Spirit C2-019.897.51.01.02 Page 4 of 11 © Siemens, 2009


07.09 SSME CS Restricted
Preparations 1 5

1.2 Information on materials

1.2.1 Content of the Update Kit


Update kit 10355377 contains the following parts:

Pos. Quantity Material No. Name


1 1 10165510 CD: Somaris/5 ICS/IES_Hotfixes_08
2 1 C2-019.897.51.xx.02 UI: Hotfixes 2008 Function Pack Up‐
date

1.2.2 Ordering information


The following update kit has to be ordered from CSML (SAP Distribution Channel, plant
2050/6050) :

Update Kit 10355377 (Material Number)

1.2.3 Return of parts


n.a.

© Siemens, 2009 C2-019.897.51.01.02 Page 5 of 11 Emotion (2007)/Spirit


Restricted 07.09 SSME CS
6 2 Update

2.1 Work steps

2.1.1 Installation on the ICS


n To disconnect the IES from the ICS and to avoid conflicts from an activity running in
parallel to the update:
Login as administrator at the IES.
- Select System > End > Restart Application
- Click Yes and press the left Shift key until the login window appears
- Login as the administrator
n Login as administrator at the ICS.
- Select System > End > Restart Application
- Click Yes and press the left Shift key until the login window appears
- Login as the administrator
n Insert the CD Somaris/5 ICS/IES_Hotfixes_08 into the CD drive (e.g: [-R]) of the ICS
n Open Windows Explorer
n Go to R:\, double-click the file install.bat
n Wait a moment. An automatic system shut-down and restart are performed.
n After restart, the installation continues automatically. Wait until the message window
The Function Pack XX successfully installed appears.
After a successful installation, it will show up as follows:
VB27: FP3
VB35: FP1
VB38: FP3
VB39: FP3
n An automatic system shut-down and restart are performed.
n Remove the CD.

2.1.2 Installation on the IES


n Insert the CD Somaris/5 ICS/IES_Hotfixes_08 into the CD drive (e.g: [-R]) of the IES
n Open Windows Explorer
n Go to R:\, double-click the file install.bat
n Wait a moment. An automatic system shut-down and restart are performed.
n After restart, the installation continues automatically. Wait until the message window
The Function Pack XX successfully installed appears.
n An automatic system shut-down and restart are performed.
n Connect the IES with the ICS again.

Emotion (2007)/Spirit C2-019.897.51.01.02 Page 6 of 11 © Siemens, 2009


07.09 SSME CS Restricted
Update 2 7

n Remove the CD.

2.1.3 Final check


After finishing the installation, select Options > Version and check if the function pack is
listed ( see above for the name of the FP).

2.1.4 Backup in the ICS / IES


n Insert a backup medium into the respective drive (e.g: [-R])
n In the service software, select Backup&Restore.
n From Command ,select Backup, and from Drives, select the CD drive, e.g [-R].
n From Package, select CT-Settings-XXX and click Go
 BackupRestore Running...
 Wait until Ready is displayed.
n From Package, select SW-Setting02-XXX and click Go
 BackupRestore Running...
 Wait until Ready is displayed.
n From Package, select Security-Settings-XXX and click Go
 BackupRestore Running...
 Wait until Ready is displayed.
n Remove the backup medium.
n Click Home and terminate the service software by closing the Browser window.

2.1.5 Informing the customer


n.a.

2.1.6 Final work steps


1. Update the system documentation.
(not required for remote service activities performed without a service engineer).
Fill out, and if needed, make a copy of the attached "Completion Protocol" and file it in
the corresponding binder.
2. The update must be reported by authorized personnel, either via an intranet applica‐
tion or automated data transfer (only for countries connected via an update handling
interface).

© Siemens, 2009 C2-019.897.51.01.02 Page 7 of 11 Emotion (2007)/Spirit


Restricted 07.09 SSME CS
8 2 Update

2.1.7 Changes to previous version


n.a.

Emotion (2007)/Spirit C2-019.897.51.01.02 Page 8 of 11 © Siemens, 2009


07.09 SSME CS Restricted
Completion protocol 3 9

(Not required for "Remote updates without a service technician")

Update number CT036/09/R has been completed.

System material no.: ...................................

System serial no.: ...................................

Customer: ................................... Site: ...................................

Customer no.: ...................................

Name (CSE): ................................... Telephone: ...................................

Country: ................................... Location: ...................................

Date: ................................... Signature: ...................................

Remark: ...................................................................................................
..

...................................................................................................
..

 After completing the update, make a copy of this page, fill it out, and file it in the appro‐
priate binder.

© Siemens, 2009 C2-019.897.51.01.02 Page 9 of 11 Emotion (2007)/Spirit


Restricted 07.09 SSME CS
10 4 List of Hazard IDs

There are no Hazard IDs in this document.

Emotion (2007)/Spirit C2-019.897.51.01.02 Page 10 of 11 © Siemens, 2009


07.09 SSME CS Restricted
- Restricted - All documents may
only be used by authorized person‐
nel for rendering services on Sie‐
mens Healthcare Products. Any
document in electronic form may
be printed once. Copy and distribu‐
tion of electronic documents and
hardcopies is prohibited. Offenders
will be liable for damages. All other
rights are reserved.

healthcare.siemens.com/services

Siemens Healthcare Headquarters


Siemens Healthcare GmbH
Henkestr. 127
91052 Erlangen
Germany
Telephone: +49 9131 84-0
siemens.com/healthcare

Print No.: C2-019.897.51.01.02 | Replaces: n.a.


Doc. Gen. Date: 07.09 | Language: English
© Siemens Healthcare GmbH, 2009

siemens.com/healthcare

You might also like