DOC2107162 - LOGIQ - E10 - E10s - E11 - E20 - Fortis - DICOM Conformance Statement
DOC2107162 - LOGIQ - E10 - E10s - E11 - E20 - Fortis - DICOM Conformance Statement
Technical
Publications
Direction DOC2107162
Revision 6
Do not duplicate
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
2
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
1 Overview
LOGIQ Scanner supports the following DICOM functionalities
• DICOM Worklist Download
• DICOM Query/Retrieve
• DICOM Storage
• DICOM SR Storage
• DICOM Print
• Modality Performed Procedure Step service
• DICOM Export/Import
PACS
Query Storage
DICOM
Retrieve Commitment
Storage
Worklist service
DICOM SR Storage
HIS/RIS LOGIQ Scanner Report Server
Modality Performed
Procedure Step
DICOM DICOM
Media Print
Storage
Print Server
3
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Comprehensive SR 1.2.840.10008.5.1.4.1.1.88.33 NI Y N N N Y Y Y S N N Y
Key Object Selection NI Y N N N N N N S N N Y
1.2.840.10008.5.1.4.1.1.88.59
Document Storage
4
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
5
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
1.2.2 Storage
For details on supported Storage SOP Classes refer to Section 1.1
Implicit Little
1.2.840.10008.1.2 Y N
Modality Worklist Endian
1.2.840.10008.5.1.4.31
Information Model – FIND Explicit Little
1.2.840.10008.1.2.1 Y N
Endian
Implicit Little
1.2.840.10008.1.2 Y N
Modality Performed Endian
1.2.840.10008.3.1.2.3.3
Procedure Step SOP Class Explicit Little
1.2.840.10008.1.2.1 Y N
Endian
Implicit Little
1.2.840.10008.1.2 Y N
Storage Commitment Push Endian
1.2.840.10008.1.20.1
Model SOP Class Explicit Little
1.2.840.10008.1.2.1 Y N
Endian
6
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
1.2.4 Query/Retrieve
Table 1.2-3 Query/Retrieve SOP Classes
DIMSE
SOP Classes Transfer Syntax
SCU SCP
Study Root Q/R - Implicit Little Endian 1.2.840.10008.1.2 Y N
1.2.840.10008.
Information Model
5.1.4.1.2.2.1 Explicit Little Endian 1.2.840.10008.1.2.1
– FIND Y N
1.2.5 Printing
Table 1.2-4 Printing SOP Classes
SOP Classes SOP Class UID Transfer Syntax
SCU SCP
Basic Grayscale Print Implicit Little Endian 1.2.840.10008.1.2 Y N
Management Meta 1.2.840.10008.5.1.1.9
SOP Class Explicit Little Endian 1.2.840.10008.1.2.1 Y N
Basic Color Print Implicit Little Endian 1.2.840.10008.1.2 Y N
Management Meta 1.2.840.10008.5.1.1.18
SOP Class Explicit Little Endian 1.2.840.10008.1.2.1 Y N
7
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
STD-GEN-CD Y Y Y
DVD
STD-GEN-DVD-JPEG Y Y Y
STD-GEN-DVD-J2K Y Y Y
USB
STD-GEN-USB-J2K Y Y Y
STD-GEN-USB-JPEG Y Y Y
8
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
2 Table of Contents
1 Overview .................................................................................................................................................. 3
1.1 Content and Transfer .......................................................................................................................... 3
1.1.1 Structured Reporting Root Template ID ...................................................................................... 5
1.2 DIMSE Services ................................................................................................................................... 6
1.2.1 Verification ................................................................................................................................. 6
1.2.2 Storage........................................................................................................................................ 6
1.2.3 Workflow Management .............................................................................................................. 6
1.2.4 Query/Retrieve ........................................................................................................................... 7
1.2.5 Printing ....................................................................................................................................... 7
1.3 DICOM Web Services - N/A ................................................................................................................. 7
1.3.1 URI Service (WADO-URI) - N/A .................................................................................................... 7
1.3.2 Study Service - N/A ..................................................................................................................... 7
1.3.3 Worklist Service - N/A ................................................................................................................. 7
1.3.4 Non-Patient Instance Service - N/A ............................................................................................. 7
1.4 Media Services ................................................................................................................................... 8
1.5 De-Identification Profiles .................................................................................................................... 8
2 Table of Contents ..................................................................................................................................... 9
3 Introduction ........................................................................................................................................... 17
3.1 Revision History ................................................................................................................................ 17
3.2 Audience .......................................................................................................................................... 17
3.3 Remarks ........................................................................................................................................... 17
3.4 Terms and Definitions ....................................................................................................................... 18
3.5 Abbreviations ................................................................................................................................... 20
3.6 Conventions ..................................................................................................................................... 21
3.7 References........................................................................................................................................ 21
4 Implementation Model .......................................................................................................................... 21
4.1 Application Entities and Data Flow ................................................................................................... 21
4.1.1 Functional Definition of AE’s ..................................................................................................... 23
4.2 File Meta Information Options (See PS3.10) ..................................................................................... 23
5 Service and Interoperability Description ................................................................................................ 25
5.1 Mapping of Services to Application Entities ..................................................................................... 25
5.2 Supported DIMSE Services ............................................................................................................... 25
5.2.1 Basic Worklist Management Service ......................................................................................... 25
5.2.1.1 SCU of the Modality Worklist Information Model – FIND SOP Class .................................. 25
5.2.1.2 SCP of the Modality Worklist Information Model – FIND SOP Class - N/A ......................... 28
5.2.2 Modality Performed Procedure Step Service ............................................................................. 29
5.2.2.1 SCU of the Modality Performed Procedure Step SOP Class ............................................... 29
9
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
5.2.2.2 SCP of the Modality Performed Procedure Step SOP Class - N/A ...................................... 33
5.2.3 Unified Worklist and Procedure Step Service - N/A ................................................................... 33
5.2.4 Instance Availability Notification Service - N/A ......................................................................... 33
5.2.5 Storage Service ......................................................................................................................... 33
5.2.5.1 SCU of Storage SOP Classes .............................................................................................. 33
5.2.5.2 SCP of Storage SOP Classes ............................................................................................... 33
5.2.5.3 Transcoding of Transfer Syntaxes ...................................................................................... 34
5.2.6 Storage Commitment Service .................................................................................................... 34
5.2.6.1 SCU of the Storage Commitment Push Model SOP Class................................................... 34
5.2.6.2 SCP of the Storage Commitment Push Model SOP Class - N/A .......................................... 36
5.2.7 Query/Retrieve Service Class .................................................................................................... 36
5.2.7.1 SCU of the Study Root Q/R - Information Model – FIND SOP Class ................................... 36
5.2.7.2 SCU of the Patient Root Q/R - Information Model – FIND SOP Class - N/A ........................ 39
5.2.7.3 SCU of the Study Root Q/R - Information Model – MOVE SOP Class ................................. 39
5.2.7.4 SCU of the Patient Root Q/R - Information Model – MOVE SOP Class - N/A ..................... 39
5.2.7.5 SCP of the Study Root Q/R - Information Model – FIND SOP Class - N/A .......................... 40
5.2.7.6 SCP of the Patient Root Q/R - Information Model – FIND SOP Class - N/A ........................ 40
5.2.7.7 SCP of the Study Root Q/R - Information Model – MOVE SOP Class - N/A ........................ 40
5.2.7.8 SCP of the Patient Root Q/R - Information Model – MOVE SOP Class - N/A ...................... 40
5.2.8 Print Management Service ........................................................................................................ 40
5.2.8.1 SCU of the Basic Grayscale Print Management Meta SOP Class ........................................ 40
5.2.8.2 SCU of the Basic Color Print Management Meta SOP Class ............................................... 44
5.2.8.3 SCU of the Basic Annotation Box SOP Class ...................................................................... 45
5.2.8.4 SCU of the Print Job SOP Class - N/A ................................................................................. 45
5.2.8.5 SCU of the Presentation LUT SOP Class - N/A .................................................................... 45
5.2.8.6 SCU of the Printer Configuration Retrieval SOP Class - N/A ............................................... 46
5.2.8.7 SCP of the Basic Grayscale Print Management Meta SOP Class - N/A................................ 46
5.3 Supported DICOM® Web Services .................................................................................................... 46
5.3.1 URI Web Service (WADO-URI) - N/A .......................................................................................... 46
5.3.2 Studies Web Service - N/A......................................................................................................... 46
5.3.3 Worklist Web Service - N/A ....................................................................................................... 46
5.3.4 Notification Web Services - N/A ................................................................................................ 46
5.4 Media Storage .................................................................................................................................. 46
5.4.1 FSC ............................................................................................................................................ 46
5.4.2 FSR ............................................................................................................................................ 46
5.4.3 FSU ........................................................................................................................................... 47
5.5 Cross Service consideration - N/A ..................................................................................................... 47
5.6 Specific Charactersets ...................................................................................................................... 47
6 DICOM® Configuration ........................................................................................................................... 47
10
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
11
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
12
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
A.3.3. Ultrasound Multi-frame Image IOD Specific Private Modules ............................................. 123
A.3.4. Ultrasound Multi-frame Image IOD Specific Values and Code Sets ..................................... 123
A.4. Enhanced US Volume IOD ........................................................................................................... 123
A.4.1. Enhanced US Volume IOD Specific Modules ....................................................................... 124
A.4.2. Enhanced US Volume IOD Specific Functional Group Macros ............................................. 134
A.4.3. Enhanced US Volume IOD Specific Private Modules ............................................................ 134
A.4.4. Enhanced US Volume IOD Specific Values and Code Sets .................................................... 135
A.5. Secondary Capture Image IOD .................................................................................................... 135
A.5.1. Secondary Capture Image IOD Specific Modules ................................................................. 135
A.5.2. Secondary Capture Image IOD Specific Functional Group Macros ....................................... 136
A.5.3. Secondary Capture Image IOD Specific Private Modules ..................................................... 136
A.5.4. Secondary Capture Image IOD Specific Values and Code Sets ............................................. 136
A.6. Comprehensive SR IOD................................................................................................................ 136
A.6.1. Comprehensive SR IOD Specific Modules ............................................................................ 137
A.6.2. Comprehensive SR IOD Specific Functional Group Macros ...................................................... 139
A.6.3. Comprehensive SR IOD Specific Private Modules .................................................................... 139
A.6.4. Comprehensive SR IOD Specific Values and Code Sets ............................................................ 139
A.7. Encapsulated PDF IOD ................................................................................................................. 139
A.7.1. Encapsulated PDF IOD Specific Modules ............................................................................. 140
A.7.2. Encapsulated PDF IOD Specific Functional Group Macros ................................................... 141
A.7.3. Encapsulated PDF IOD Specific Private Modules ................................................................. 141
A.7.4. Encapsulated PDF IOD Specific Values and Code Sets ......................................................... 141
A.8. Key Object Selection Document IOD ........................................................................................... 141
A.8.1. Key Object Selection Document IOD Specific Modules ....................................................... 142
A.8.2. Key Object Selection Document IOD Specific Functional Group Macros ............................. 144
A.8.3. Key Object Selection Document IOD Specific Private Modules ............................................ 144
A.8.4. Key Object Selection Document IOD Specific Values and Code Sets .................................... 144
A.9. Basic Directory IOD ..................................................................................................................... 144
B. Structured Report Content Encoding .................................................................................................... 150
B.1. Vascular Ultrasound Procedure Report (TID 5100) ...................................................................... 150
B.1.1. Vascular Template Structure ................................................................................................... 150
B.1.2. Usage and Extension of TID 5100 Vascular Ultrasound Report ................................................ 150
B.1.3. TID 5101 Vascular Patient Characteristics ............................................................................... 151
B.1.4. TID 5102 Vascular Procedure Summary Section ...................................................................... 151
B.1.5. TID 5103 Vascular Ultrasound Section ..................................................................................... 152
B.1.6. TID 5104 Vascular Ultrasound Measurement Group (extended) ............................................. 153
B.1.7. GEU Applications and Extensions ............................................................................................ 154
B.1.8. TID 300 Measurement (extended for Vascular) ....................................................................... 165
B.1.9. GE Ultrasound Modes ............................................................................................................. 166
13
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
14
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
15
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
B.5.1. Thyroid Imaging Report Template Structure supported by LOGIQ scanner .............................. 264
B.5.2. TID GEMS_US_SP_0100 Soft Tissue Neck and Head Imaging Report ....................................... 265
B.5.3. TID GEMS_US_SP_0101 Thyroid Imaging Finding Section ....................................................... 265
B.5.4. TID GEMS_US_SP_0102 Thyroid Lobe Section......................................................................... 267
B.5.5. TID GEMS_US_SP_0103 Lymph Node Section ......................................................................... 268
B.5.6. TID GEMS_US_SP_0104 Parathyroid Section ........................................................................... 270
B.5.7. TID GEMS_US_SP_0105 Isthmus Section ................................................................................ 271
B.5.8. TID GEMS_US_SP_0106 Nodules Section ................................................................................ 271
B.5.9. TID 1400 Linear Measurement Template ................................................................................ 275
B.5.10. TID 1402 Volume Measurement Template .......................................................................... 275
B.5.11. TID 1204 Language of Content Item and Descendants ........................................................ 275
B.5.12. TID 1401 Area Measurement Template ............................................................................... 275
B.6. Key Object Selection Document Template (TID 2010) ................................................................. 276
B.6.1. TID 2010 Template .................................................................................................................. 276
B.7. User Defined Template Structure ................................................................................................ 276
B.7.1. TID G301 Template .................................................................................................................. 276
B.8. GE Ultrasound M&A Report (TID GEMS_US_0100) ...................................................................... 277
B.8.1. TID GEMS_US_0100 Template ................................................................................................. 278
B.9. TID 300 Measurement................................................................................................................. 279
B.9.1. TID 300 Measurement ............................................................................................................ 280
B.9.2. TID 310 Measurement ............................................................................................................ 280
B.9.3. TID 320 Image or Spatial Coordinates ..................................................................................... 281
C. Security Details .................................................................................................................................... 281
C.1. External Network Requirements details ...................................................................................... 281
C.1.1. Basic Time Synchronization ..................................................................................................... 281
C.1.2. Basic Network Address Management ...................................................................................... 281
C.1.3. Application Configuration Management ................................................................................. 281
C.1.4. DNS Service Discovery............................................................................................................. 281
C.2. DICOM® Security Profile Details .................................................................................................. 281
C.2.1. Online Electronic Storage Secure Use...................................................................................... 281
C.2.2. Audit Trail Messages ............................................................................................................... 281
C.2.3. Audit Trail Message Transmission Profile – Syslog – TLS .......................................................... 291
C.2.4. Audit Trail Message Transmission Profile - Syslog-UDP ............................................................ 292
C.2.5. Secure Transport Connection Details ...................................................................................... 292
C.2.6. Attribute Confidentiality Details .............................................................................................. 292
C.2.7. Digital Signature details .......................................................................................................... 292
C.2.8. Additional DICOM® Security Profiles supported ...................................................................... 293
D. Mapping of Attributes .......................................................................................................................... 293
E. Index of Tables ..................................................................................................................................... 294
16
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
3 Introduction
3.1 Revision History
Revision Date Reason for Change
5.0 26-July-2021 Document template change,
Updates for R3
3.2 Audience
This document is intended for hospital staff, health system integrators, Research and Development, sales,
and service. It is assumed that the reader has a working knowledge of the DICOM® Standard.
The document structure was designed for easier access to relevant information for different user groups:
• Clinical Users, who want to get an overview of the implemented interoperability features of the
system can refer to Section Error! Reference source not found. Error! Reference source not found..
• Personnel involved in Sales can use the information in Section Error! Reference source not found.
Error! Reference source not found. to asses the compatibility between different systems involved in
a sales situation.
• System Integrators can use information in Section 6 DICOM® Configuration
LOGIQ Scanner provides a dedicated interface to configure the basic and advanced parameters
17
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Verify Before Query USER Enabled If enabled, the system checks the
DICOM connection status PRIOR to
issuing the Worklist Query/Retrieve
request. This way, the system ensures
that the PACS or HIS/ RIS server is ’alive’
before performing the actual query. If
the server is down, the scanner will not
issue the query request and the system
is accessible to the user within the
configured Verify Timeout
18
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
19
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
There are no limitations on the number of remote servers supported by LOGIQ Scanner
20
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
21
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
22
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
23
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
24
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
• Bottom
• Top
• Per Image
• Other
25
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
• during system installation and also information from Section Error! Reference source not found.
Error! Reference source not found. for details regarding the implemented services.
• Field Service engineers can use the details from Section Error! Reference source not found. Error!
Reference source not found. and from Section 6.5 Audit trail -Syslog configuration
• The following table shows the configuration parameter required for audit trail / Syslog
configuration.
In the Configurable column the following terms are used:
• USER: the parameter is configurable by the USER
• SERVICE: the parameter is configurable by SERVICE
• NO: the parameter is not configurable (it has a fixed value). The value is required for the
configuration of the remote system.
• N/A: the parameter is not applicable
Table 6.5-1: Audit Trail Originator Parameters
Originator Audit Trail Message Transmission-SYSLOG parameters
Parameter Configurable Default Value Comment
Remote Host name/IP USER
Remote Port number (TCP Listener Port) USER 514 Provide secure port, in case
of TLS mode
UDP Listerner Port Number USER 514
Mode of Transfer USER TCP Mode of transferring the
data to the server (UDP,
TCP, or TLS)
Windows Event Log
Type of Logs to transfer: USER Enabled
• Application Logs
• Security Logs
Source of application event log to • Add Source: The user
transfer needs to specify where
26
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
27
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Time-out is
ARTIM Timeout configurable for each
service, see 6.2
28
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Configuration for
DIMSE Services
29
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
4.2 AE Specifications
4.2.1 Logiq Scanner Application Entity
4.2.1.1 Sequencing of Real-World Activities
30
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
31
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
32
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
33
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
34
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
35
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
36
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
37
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
If the Logiq scanner is unable to establish an association with the Retrieve SCP, it will retry for the
configured number of times.
This section describes the transfer syntax preference for this Real World Activity
Table 7.2-1: Transfer Syntax Selection Preference Order - Image SOP Classes for Real World Activity -
Exam Retrieve/ Image Receive
Preference Order Transfer Syntax UID Comment
38
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
39
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Exception Behavior
Timeout Notifies the error to user and logs the error details
Association aborted Notifies the error to user and logs the error details
Network Disconnect Notifies the error to user and logs the error details
Exception Behavior
Association aborted Notifies the error to user and logs the error details
4.3.2.1.1 SCU of the Modality Worklist Information Model Find SOP Class - C-
FIND
Table 7.3-1: Status Codes for C-FIND of the Modality Worklist information Model SOP Class - SCU
Success Matching is complete - No final identifier 0000H Displays the query result to user.
is supplied
40
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Cancel Matching terminated due to cancel FE00H No error dialog is displayed, error cod
e is logged.
Pending Matches are continuing - Current Match FF00H Waits for query completion
is supplied and any Optional Keys were
supported in the same manner as
Required Keys.
Matches are continuing - Warning that FF01H Waits for query completion
one or more Optional Keys were not
supported for existence for this Identifier
4.3.2.1.2 SCP of the Modality Worklist Information Model Find SOP Class - C-
FIND - N/A
Not Applicable
4.3.2.2.1 SCU of the Modality Performed Procedure Step SOP Class - N-CREATE
Table 7.3-2: Status Codes for N-CREATE of the Modality Performed Procedure Step SOP Class - SCU
Warning Attribute Value Out of Range 0116H Job is marked as failed after retrying
configured number of times. Error is
logged.
41
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
42
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
4.3.2.2.2 SCU of the Modality Performed Procedure Step SOP Class - N-SET
Table 7.3-3: Status Codes for N-SET of the Modality Performed Procedure Step SOP Class - SCU
Warning Attribute Value Out of Range 0116H Job is marked as failed after retrying
configured number of times. Error is
logged.
43
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
4.3.2.2.3 SCP of the Modality Performed Procedure Step SOP Class - N-CREATE
- N/A
Not Applicable
4.3.2.2.4 SCP of the Modality Performed Procedure Step SOP Class - N-SET -
N/A
Not Applicable
Not Applicable
SCU of Request UPS Cancel on UPS Push SOP Class - N-ACTION - N/A
Not Applicable
44
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Not Applicable
Not Applicable
Not Applicable
Not Applicable
SCU of the Change UPS State on UPS Pull SOP Class - N-ACTION - N/A
Not Applicable
Not Applicable
Not Applicable
Not Applicable
SCU of the - Request UPS Cancel on UPS Watch SOP Class - N-ACTION - N/A
Not Applicable
Not Applicable
Not Applicable
45
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
SCP of the Request UPS Cancel on UPS Push SOP Class - N-ACTION - N/A
Not Applicable
Not Applicable
Not Applicable
Not Applicable
Not Applicable
SCP of the Change UPS State on UPS Pull SOP Class - N-ACTION - N/A
Not Applicable
Not Applicable
Not Applicable
Not Applicable
SCP of the Request UPS Cancel on UPS Watch SOP Class - N-ACTION - N/A
Not Applicable
Not Applicable
46
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Data Set does not match Job is marked as failed after retrying
A900-A9FF configured number of times. Error is logged.
SOP Class
47
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
2. Scanner is in Battery
mode
4. Scanner is in wireless
mode and Receive via
Wireless is disabled.
Failure
Error: Data Set does not A901 Never
match SOP Class
4.3.2.6.1 SCU of the Storage Commitment Push Model SOP Class - N-ACTION
Table 7.3-6: Status Codes N-ACTION of the Storage Commitment Push Model SOP Class - SCU
48
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
* Any other status code. * Job is marked as failed after retrying configured
number of times. Error is logged.
49
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
50
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Failure Refused: Out of resources A700H Error dialog is displayed and error
code is logged
Error: Identifier does not match SOP Class A900H Error dialog is displayed and error
code is logged
SOP Class Not Supported 0122H Error dialog is displayed and error
code is logged
Pending Matches are continuing - Current Match FF00H Waits for query completion
is supplied and any Optional Keys were
supported in the same manner as
Required Keys.
Matches are continuing - Warning that FF01H Waits for query completion
one or more Optional Keys were not
supported for existence for this Identifier
(0000,1022)
(0000,1023)
Warning Sub-operations B000 (0000,1020) If scanner receives this status after receiving
Complete - One images, the scanner notifies user about the
or more Failures (0000,1022) partial transfer of exams.
(0000,1023)
51
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Failed Out of Resources A701 (0000,0902) Error dialog is displayed, and error code is
- Unable to logged
calculate number
of matches
Out of Resources A702 (0000,1020) Error dialog is displayed, and error code is
- Unable to logged
perform sub- (0000,1021)
operations
(0000,1022)
(0000,1023)
Move Destination A801 (0000,0902) Error dialog is displayed, and error code is
unknown logged
Identifier does A900 (0000,0901) Error dialog is displayed, and error code is
not match SOP logged
Class (0000,0902)
Unable to process Cxxx (0000,0901) Error dialog is displayed, and error code is
logged
(0000,0902)
Cancel Sub-operations FE00 (0000,1020) Error dialog is displayed, and error code is
terminated due logged
to Cancel (0000,1021)
Indication
(0000,1022)
(0000,1023)
(0000,1022)
(0000,1023)
52
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Table 7.3-10: Status CodesN-CREATE of the Basic Film Session SOP Class - SCU
Status Class Further Meaning Status Code Behavior
53
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Not Applicable
Not Applicable
Not Applicable
Table 7.3-11: Status CodesN-CREATE of the Basic Film Box SOP Class - SCU
Warning Attribute List Error 0107 Job is marked as failed after retrying
configured number of times. Error is
logged.
Attribute Value Out Of Range 0116 Job is marked as failed after retrying
configured number of times. Error is
logged.
Requested Min Density or Max Density B605 Scanner shall continue to process the job
outside of printer's operating range
54
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
There is an existing Film Box that has not C616 Job is marked as failed after retrying
been printed and N-ACTION at the Film configured number of times. Error is
Session level is not supported. A new Film logged.
Box will not be created when a previous
Film Box has not been printed
Not Supported
Table 7.3-12: Status CodesN-DELETE of the Basic Film Box SOP Class - SCU
Behavior
Status class Further Meaning Status Code
55
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Table 7.3-13: Status Codes N-ACTION of the Basic Film Box SOP Class - SCU
Film Box SOP Instance hierarchy does not B603H Scanner shall continue to process the
contain Image Box SOP Instances (empty job
page)
Image size is larger than Image Box size. B604H Job is marked as failed after retrying
The image has been demagnified. configured number of times. Error is
logged.
56
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Image size is larger than Image Box size. B609H Scanner shall continue to process the
The image has been cropped to fit. job
Image size or Combined Print Image Size is B60AH Scanner shall continue to process the
larger than Image Box size. The image or job
combined Print Image has been decimated
to fit.
57
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Unable to create Print Job SOP Instance; C602 Job is marked as failed after retrying
print queue is full. configured number of times. Error is
logged.
Image size is larger than Image Box size. C603 Job is marked as failed after retrying
configured number of times. Error is
logged.
Combined Print Image Size is larger than C613 Job is marked as failed after retrying
Image Box size. configured number of times. Error is
logged.
4.3.2.8.3 SCU of the Basic Grayscale Image Box SOP Class - N-SET
Table 7.3-14: Status Codes N-SET of the Grayscale Image Box SOP Class - SCU
Warning Image size is larger than Image Box size. B604H Job is marked as failed after
The image has been demagnified. retrying configured number of
times. Error is logged.
Requested Min Density or Max Density B605H Scanner shall continue to process
outside of printer's operating range. the job
Image size is larger than Image Box size. B609H Scanner shall continue to process
The image has been cropped to fit. the job
Image size or Combined Print Image Size is B60AH Scanner shall continue to process
larger than Image Box size. The image or the job
combined Print Image has been decimated
to fit.
58
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Image size is larger than Image Box size. C603 Job is marked as failed after
retrying configured number of
times. Error is logged.
Insufficient memory in printer to store the C605 Job is marked as failed after
image. retrying configured number of
times. Error is logged.
59
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Combined Print Image Size is larger than C613 Job is marked as failed after
Image Box size. retrying configured number of
times. Error is logged.
4.3.2.8.4 SCU of the Basic Color Image Box SOP Class - N-SET
Table 7.3-15: Status Codes N-SET of the Color Image Box SOP Class - SCU
Warning Image size is larger than Image Box size. B604H Job is marked as failed after
The image has been demagnified. retrying configured number of
times. Error is logged.
Image size is larger than Image Box size. B609H Scanner shall continue to process
The image has been cropped to fit. the job
Image size or Combined Print Image Size is B60AH Scanner shall continue to process
larger than Image Box size. The image or the job
combined Print Image has been decimated
to fit.
60
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Image size is larger than Image Box size. C603 Job is marked as failed after
retrying configured number of
times. Error is logged.
Insufficient memory in printer to store the C605 Job is marked as failed after
image. retrying configured number of
times. Error is logged.
Combined Print Image Size is larger than C613 Job is marked as failed after
Image Box size. retrying configured number of
times. Error is logged.
61
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Table 7.3-16: Status Codes N-EVENT-REPORT of the Printer SOP Class - SCU
62
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Table 7.3-17: Status Codes N-GET of the Printer SOP Class - SCU
Warning Optional Attribute Not Supported 0001 Scanner shall continue to process the
job
63
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
64
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
65
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
4.3.2.8.12 SCP of the Basic Grayscale Image Box SOP Class - N-SET - N/A
4.3.2.8.13 SCP of the Basic Color Image Box SOP Class - N-SET - N/A
4.3.2.8.14 SCP of the SCP of the Printer SOP Class- N/A
SCP of the Printer SOP Class – N-EVENT-REPORT- N/A
4.3.2.8.15 SCP of the Basic Annotation Box SOP Class – N-SET - N/A
4.3.2.8.16 SCP of the Print Job SOP Class- N/A
SCP of the Print Job SOP Class – N-EVENT-REPORT- N/A
4.3.2.8.18 SCP of the Printer Configuration Retrieval SOP Class - N-GET - N/A
66
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
67
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
4.4 Remarks
The scope of this DICOM® Conformance Statement is to facilitate integration between LOGIQ Scanner
and other DICOM® products. The Conformance Statement should be read and understood in conjunction
with the DICOM® Standard [1]. DICOM® by itself does not guarantee interoperability.
The Conformance Statement does, however, facilitate a first-level comparison for interoperability
between different applications supporting compatible DICOM® functionality.
This Conformance Statement is not supposed to replace validation with other DICOM® equipment to
ensure proper exchange of intended information. In fact, it is the user’s responsibility to perform the
following validation activities:
• The comparison of conformance statements from the LOGIQ Scanner and other DICOM®
conformant equipment is the first step towards assessing interconnectivity and interoperability
between those systems.
• Test procedures should be defined and executed to validate the required level of interoperability
with specific DICOM® conformant equipment, as established by the healthcare facility.
68
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
LOGIQ Scanner has participated in an industry-wide testing program sponsored by Integrating the
Healthcare Enterprise (IHE). The IHE Integration Statement for LOGIQ Scanner together with the IHE
Technical Framework, may facilitate the process of validation testing.
69
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Protocol Data Unit (PDU) A packet (piece) of a DICOM® message sent across the network. Devices
must specify the maximum size packet they can receive for DICOM® messages.
Security Profile A set of mechanisms, such as encryption, user authentication, or digital
signatures, used by an Application Entity to ensure confidentiality, integrity,
and/or availability of exchanged DICOM® data.
Service Class Provider (SCP) Role of an Application Entity that provides a DICOM® network service;
typically, a server that performs operations requested by another Application
Entity (Service Class User). Examples: Picture Archiving and Communication
System (image storage SCP, and image query/retrieve SCP), Radiology
Information System (modality worklist SCP).
Service Class User (SCU) Role of an Application Entity that uses a DICOM® network service;
typically, a client. Examples: imaging modality (image storage SCU, and modality
worklist SCU), imaging workstation (image query/retrieve SCU).
Service/Object Pair Class (SOP Class) The specification of the network or media transfer (service) of a
particular type of data (object); the fundamental unit of DICOM® interoperability
specification. Examples: Ultrasound Image Storage Service, Basic Grayscale Print
Management.
Service/Object Pair Instance (SOP Instance) An information object; a specific occurrence of
information exchanged in a SOP Class. Examples: a specific x-ray image.
Specialized SOP Class A SOP class that is derived from the standard, and that is specialized by
additional type 1, 1C, 2, 2C or 3 attributes, by enumeration of specific permitted
values for Attributes, or by enumeration of specific permitted Templates. The
additional Attributes may either be drawn from the Data Dictionary in PS3.6, or
may be Private Attributes.
Standard SOP Class A SOP class defined in the standard, and that is implemented and used without
any modifications.
Standard Extended SOP Class A SOP class that is defined in the standard, and that is extended by
additional type 3 attributes. The additional Attributes may either be drawn from
the DICOM Data Dictionary in PS3.6, or may be Private Attributes.
Tag A 32-bit identifier for a data element, represented as a pair of four-digit
hexadecimal numbers, the "group" and the "element". If the "group" number is
odd, the tag is for a private (manufacturer-specific) data element. Examples:
(0010,0020) [Patient ID], (07FE,0010) [Pixel Data], (0019,0210) [private data
element].
Transfer Syntax The encoding used for exchange of DICOM® information objects and messages.
Examples: JPEG compressed (images), Little Endian Explicit Value Representation.
Unique Identifier (UID) A globally unique "dotted decimal" string that identifies a specific object or a
class of objects; an ISO-8824 Object Identifier. Examples: Study Instance UID, SOP
Class UID, SOP Instance UID.
User Agent A client in a network protocol used in communications within a client–server
distributed computing system. In particular, the Hypertext Transfer Protocol
(HTTP) identifies the client software originating the request, using a user-agent
header, even when the client is not operated by a user
Value Representation (VR) The format type of an individual DICOM® data element, such as text, an
integer, a person's name, or a code. DICOM® information objects can be
transmitted with either explicit identification of the type of each data element
(Explicit VR), or without explicit identification (Implicit VR); with Implicit VR, the
70
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
receiving application must use a DICOM® data dictionary to look up the format of
each data element.
4.6 Abbreviations
Abbreviations that are used in this DICOM conformance statement are listed here.
AE Application Entity
AET Application Entity Title
CD-R Compact Disk Recordable
CT Computed Tomography
DHCP Dynamic Host Configuration Protocol
DICOM® Digital Imaging and Communications in Medicine
DIMSE DICOM Message Service Element
DNS Domain Name System
FSC File-Set Creator
FSU File-Set Updater
FSR File-Set Reader
GSDF Grayscale Standard Display Function
HIS Hospital Information System
HL7 Health Level 7 Standard
IHE Integrating the Healthcare Enterprise
IOD Information Object Definition
IPv4 Internet Protocol version 4
IPv6 Internet Protocol version 6
ISO International Organization for Standardization
JPEG Joint Photographic Experts Group
KOS Key Object Selection
KIN Key Image Note
LUT Look-up Table
MPEG Moving Picture Experts Group
MG Mammography (X-ray)
MPPS Modality Performed Procedure Step
MR Magnetic Resonance Imaging
MWL Modality Worklist
NEMA National Electrical Manufacturers Association
NTP Network Time Protocol
OID Object Identifier
OS Origin Server
71
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
4.7 Conventions
N/A
4.8 References
1. NEMA PS3 Digital Imaging and Communications in Medicine (DICOM®) Standard, available free
at http://www.dicomstandard.org/current
2. IHE Radiology Technical Framework available at
https://www.ihe.net/resources/technical_frameworks/#radiology
5 Implementation Model
5.1 Application Entities and Data Flow
The network and media interchange application model for the LOGIQ Scanner is shown in Figure 5.1-1
72
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
DICOM Standard Interface
Remote
Echo Verification
SCU /SCP
Association
Initiation &
Worklist Acceptance Modality
Query Worklist
SCP
Instance Instance
Store/End Store/End
Exam Exam
LOGIQ
Scanner Remote
Storage
Exam
Save
AE SCP
Storage
Commitment
SCP
Image Remote
Print Print SCP
Exam Query/
Query/ Retrieve
Retrieve SCP
Receive Storage
Image SCU
DICOM
Export
LOGIQ
Scanner Media
Storage
DICOM
Import AE
73
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
This section describes the organization of the supported Services into Application Entities based on the
default configuration of the system. This may change based on the actual setup at the customer side.
Refer to section 6 for details about the configurability of Services into AEs.
74
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Note: The Implementation Version Name and may change in the future without modification of this
document.
75
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Server
Origin
Agent
User
SCU
FSU
SCP
FSR
FSC
Basic Worklist Y N N N N N N
Management
MPPS Y N N N N N N
Storage Y Y Y N N N N
LOGIQ Scanner AE
Storage Commitment Y N N N N N N
Query/Retrieve Y N N N N N N
Print Management Y N N N N N N
Media Storage N N N N Y Y Y
For the Query Value Source column, the following values are supported:
76
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
77
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Query Dis-
Matching
Attribute Name Tag Value Value play Comments
Type
Sources on UI
Requested Procedure
Referenced Study Sequence (0008,1110) RETURN_KEY EMPTY D
>Referenced SOP Class UID (0008,1150) RETURN_KEY D
>Referenced SOP Instance UID (0008,1155) RETURN_KEY D
Study Instance UID UID, CONFIGURATI
(0020,000D) D
RETURN_KEY ON, EMPTY
Requested Procedure Description (0032,1060) CONFIGURATI
SINGLE_VALUE,
ON, USER, D
RETURN_KEY
EMPTY
Requested Procedure Code (0032,1064)
RETURN_KEY EMPTY D
Sequence
Requested Procedure ID (0040,1001) SINGLE_VALUE, CONFIGURATI
D
RETURN_KEY ON, EMPTY
Names of Intended Recipients of (0040,1010) SINGLE_VALUE, CONFIGURATI
D
Results RETURN_KEY ON, EMPTY
Imaging Service Request
Accession Number SINGLE_VALUE, CONFIGURATI
(0008,0050) RETURN_KEY ON, USER, D
EMPTY
Referring Physician's Name (0008,0090) SINGLE_VALUE, CONFIGURATI
D
RETURN_KEY ON, EMPTY
Requesting Physician (0032,1032) SINGLE_VALUE, CONFIGURATI
D
RETURN_KEY ON, EMPTY
Requesting Service (0032,1033) SINGLE_VALUE, CONFIGURATI
D
RETURN_KEY ON, EMPTY
Visit Identification
Admission ID (0038,0010) SINGLE_VALUE, CONFIGURATI
D
RETURN_KEY ON, EMPTY
Visit Status
Current Patient Location (0038,0300) SINGLE_VALUE, CONFIGURATI
D
RETURN_KEY ON, EMPTY
Visit Relationship
Referenced Patient Sequence (0008,1120) RETURN_KEY EMPTY D
>Referenced SOP Class UID (0008,1150) RETURN_KEY D
>Referenced SOP Instance UID (0008,1155) RETURN_KEY D
Patient Identification
Patient's Name CONFIGURATI
SINGLE_VALUE,
(0010,0010) ON, USER, D
RETURN_KEY
EMPTY
78
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Query Dis-
Matching
Attribute Name Tag Value Value play Comments
Type
Sources on UI
The association will be aborted immediately after user triggering Cancel request.
6.2.1.2 SCP of the Modality Worklist Information Model – FIND SOP Class -
N/A
Not Supported
79
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Attribute Name
Tag Source Value Value Comments
N- CREATE N-SET
80
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Attribute Name
Tag Source Value Value Comments
N- CREATE N-SET
Procedure Code Sequence (0008,1032) MWL, EMPTY Taken from Taken from
worklist if it is worklist if it is
there(Procedure there(Procedure
Code Sequence) Code Sequence)
81
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Attribute Name
Tag Source Value Value Comments
N- CREATE N-SET
Performed Procedure Step (0040,0252) GENERATED IN PROGRESS COMPLETED, For an exam with
Status DISCONTINUED saved images or
results, the N-SET
will be sent with
status
COMPLETED.
For an exam
without saved
images or results,
the N-SET will be
sent with status
DISCONTINUED.
Performed Procedure Step (0040,0254) MWL, USER, EMPTY In case of Worklist In case of Worklist
Description exams, copied exams, copied
from Requested from Requested
Procedure Procedure
Description. For Description. For
manually started manually started
exams, uses the exams, uses the
user selected user selected
Exam Description Exam Description
82
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Attribute Name
Tag Source Value Value Comments
N- CREATE N-SET
<YYYYMMDD.HHM
MSS> format
83
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
N-CREATE is sent when starting image acquisition and N-SET is sent when the exam is ended.
6.2.2.2 SCP of the Modality Performed Procedure Step SOP Class - N/A
Not Supported
LOGIQ Scanner saves the images as it is received; no further manipulation is done on the received
images.
If the incoming image(s) already present in the scanner database, the scanner doesn’t overwrite the
images.
If the operation fails while receiving the study, the already received instances will be persistented in the
Database.
84
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Table 6.2-3 defines possible transcodings between transfer syntaxes, when objects received (stored
locally) must be sent out again. Possible values include
• Supported: When transcoding is possible and same SOP Instance UID is re-used.
85
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Comprehensive SR 1.2.840.10008.5.1.4.1.1.88.33
Below Table lists the behavior of LOGIQ Scanner for each possible Event Type ID (0000,1002) upon
receiving an N-EVENT-REPORT request from the SCP.
Table 6.2-5: Behavior based on Event Type ID
Event Type ID Description Behavior
2 Storage Commitment Request Complete – Failures The request for storage commitment is
Exist marked as failed.
Table 6.2-6 lists the behavior of LOGIQ Scanner for each possible Failure Reason (0008,1197) in the
Failed SOP Sequence (0008,1198) upon receiving an N-EVENT-REPORT request from the SCP with an
Event Type ID of 2 (Storage Commitment Request Complete – Failures Exist).
0000H Success
86
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
LOGIQ Scanne keeps the Storage Commitment job until N-EVENT-REPORT is received.
Failed Storage commitment requests shall be available until user manually deletes the request or it
becomes success. Failed requests shall be retried on each system restart. User can also retry the request.
The product shall not delete any instance automatically, irrespective of the Storage commitment success
or failure.
6.2.6.2 SCP of the Storage Commitment Push Model SOP Class - N/A
Not Supported
For the Query Value Source column, the following values are supported:
87
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
CONFIGURAT
SINGLE_VALUE,
ION, USER,
Study Date (0008,0020) RANGE, YES
GENERATED,
RETURN_KEY
EMPTY
SINGLE_VALUE,
CONFIGURAT
Study Time (0008,0030) RANGE, NO
ION, EMPTY
RETURN_KEY
Supported
modalities are US,
SINGLE_VALUE, CONFIGURAT CT, MR, MG, PT
Modalities in Study (0008,0061) NO
RETURN_KEY ION, EMPTY and XA. Multiple
values are
separated by \.
Referring Physician's SINGLE_VALUE, CONFIGURAT
(0008,0090) NO
Name RETURN_KEY ION, EMPTY
CONFIGURAT
SINGLE_VALUE,
Study Description (0008,1030) ION, USER, NO
RETURN_KEY
EMPTY
Procedure Code
(0008,1032) RETURN_KEY EMPTY NO
Sequence
Name of Physician(s) SINGLE_VALUE, CONFIGURAT
(0008,1060) NO
Reading Study RETURN_KEY ION, EMPTY
Referenced Patient
(0008,1120) RETURN_KEY EMPTY NO
Sequence
SINGLE_VALUE,
Patient's Name (0010,0010) WILDCARD, USER, EMPTY YES
RETURN_KEY,
SINGLE_VALUE,
Patient ID (0010,0020) USER, EMPTY YES
RETURN_KEY
88
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
SINGLE_VALUE, CONFIGURAT
Patient's Birth Time (0010,0032) NO
RETURN_KEY ION, EMPTY
SINGLE_VALUE, CONFIGURAT
Patient's Age (0010,1010) NO
RETURN_KEY ION, EMPTY
SINGLE_VALUE, CONFIGURAT
Patient's Size (0010,1020) NO
RETURN_KEY ION, EMPTY
SINGLE_VALUE, CONFIGURAT
Patient's Weight (0010,1030) NO
RETURN_KEY ION, EMPTY
SINGLE_VALUE, CONFIGURAT
Ethnic Group (0010,2160) NO
RETURN_KEY ION, EMPTY
SINGLE_VALUE, CONFIGURAT
Occupation (0010,2180) NO
RETURN_KEY ION, EMPTY
UID, CONFIGURAT
Study Instance UID (0020,000D) NO
RETURN_KEY ION, EMPTY
Series Level
SINGLE_VALUE, CONFIGURAT
Series Date (0008,0021) RANGE, ION, USER, YES
RETURN_KEY EMPTY
SINGLE_VALUE, CONFIGURAT
Series Time (0008,0031) RANGE, ION, USER, NO
RETURN_KEY EMPTY
SINGLE_VALUE, CONFIGURAT
Modality (0008,0060) YES
RETURN_KEY ION, EMPTY
SINGLE_VALUE, CONFIGURAT
Series Description (0008,103E) YES
RETURN_KEY ION, EMPTY
89
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
SINGLE_VALUE, CONFIGURAT
Operator’s Name (0008,1070) NO
RETURN_KEY ION, EMPTY
SINGLE_VALUE, CONFIGURAT
Software Version(s) (0018,1020) NO
RETURN_KEY ION, EMPTY
SINGLE_VALUE, CONFIGURAT
Protocol Name (0018,1030) NO
RETURN_KEY ION, EMPTY
Study Instance
UID is taken from
Study Instance UID (0020,000D) UID GENERATED NO
STUDY level
Query response
SINGLE_VALUE, CONFIGURAT
Series Instance UID (0020,000E) NO
RETURN_KEY ION, EMPTY
SINGLE_VALUE, CONFIGURAT
Series Number (0020,0011) YES
RETURN_KEY ION, EMPTY
Number of Series SINGLE_VALUE, CONFIGURAT
(0020,1209) YES
Related Instances RETURN_KEY ION, EMPTY
Request Attributes
(0040,0275) RETURN_KEY EMPTY NO
Sequence
The association will be aborted immediately after user triggering Cancel request.
6.2.7.2 SCU of the Patient Root Q/R - Information Model – FIND SOP Class -
N/A
Not Supported
6.2.7.3 SCU of the Study Root Q/R - Information Model – MOVE SOP Class
If no C-STORE requests are received within this configurable timeframe it repeats the C-MOVE-Request
for the configured number of retries; and shows proper error to the user if no C-STORE requests received
even after all retries.
6.2.7.4 SCU of the Patient Root Q/R - Information Model – MOVE SOP Class
- N/A
Not Supported
90
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
6.2.7.5 SCP of the Study Root Q/R - Information Model – FIND SOP Class -
N/A
Not Supported
6.2.7.6 SCP of the Patient Root Q/R - Information Model – FIND SOP Class -
N/A
Not Supported
6.2.7.7 SCP of the Study Root Q/R - Information Model – MOVE SOP Class -
N/A
Not Supported
6.2.7.8 SCP of the Patient Root Q/R - Information Model – MOVE SOP Class
- N/A
Not Supported
Printer 1.2.840.10008.5.1.1.16
91
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
92
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
N-SET Set each Image attributes for a previously created film box
93
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
An N-EVENT-REPORT request can be received by the SCU at any time during an association.
Table 5.2-24 summarizes the behavior of the SCU when receiving Event Types within the N-EVENT-
REPORT.
Table 6.2-16: Printer SOP Class N-EVENT-REPORT Behavior
Event Type Name Event Type ID Behavior
Normal 1
Warning 2
94
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
<<NORMAL
WARNING
FAILURE>>
Printer Status (2110,0010) Used to check the status of the printer before a
print operation is started.
If the printer status is FAILURE, the print operation
is aborted, a message is displayed, and the print
files reside in the print buffer.
If the “Printer Status” is "FAILURE" then this status
Printer Status Info (2110,0020) information is displayed, and the print files resides
in the print buffer.
Printer Name (2110,0030)
6.2.8.2 SCU of the Basic Color Print Management Meta SOP Class
The Basic Color Print Management SOP Class is composed of the following mandatory SOP Classes:
Table 6.2-18: Basic Color Print Management SOP Classes
SOP Class Name SOP Class UID
Printer 1.2.840.10008.5.1.1.16
N-SET Set each Image attributes for a previously created film box
95
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
N-SET Set each image attributes for a previously created film box
96
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
6.2.8.7 SCP of the Basic Grayscale Print Management Meta SOP Class - N/A
Not Supported
6.4.2 FSR
As a File Set Reader, the LOGIQ Scanner supports the Media Application Profiles, listed in Error!
Reference source not found. in the Overview. For a list of supported SOP Classes, refer to Table 1.1-1
Storage SOP ClassesError! Reference source not found. in the Overview.
97
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
LOGIQ scanner can read the media by acting as an FSR to read the DICOM File-set on CD/DVD or USB
flash drive / HDD. The images will be read as is on he media.
LOGIQ Scanner supports following media operations as FSR:
• Import images
• Import SR documents
• Import encapsulated PDF documents
This product doesn’t support KOS document objects.
6.4.3 FSU
As a File Set Updater, LOGIQ Scanner supports creating the Basic Directory IOD as defined in Annex
A.9Error! Reference source not found..
For a list of supported Media Application Profiles, refer to Error! Reference source not found. in the
Overview. For a list of supported SOP Classes, refer to Error! Reference source not found. in the
Overview.
LOGIQ scanner can update the media by acting as an FSU to update the DICOM File-set to add new SOP-
instances to the File-set.
7 DICOM® Configuration
LOGIQ Scanner provides a dedicated interface to configure the basic and advanced parameters
98
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
99
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Verify Before Query USER Enabled If enabled, the system checks the
DICOM connection status PRIOR to
issuing the Worklist Query/Retrieve
request. This way, the system ensures
that the PACS or HIS/ RIS server is ’alive’
before performing the actual query. If
the server is down, the scanner will not
issue the query request and the system
is accessible to the user within the
configured Verify Timeout
100
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
101
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
102
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
103
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
104
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
105
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
• Bottom
• Top
• Per Image
• Other
106
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
107
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
108
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
109
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Time-out is
configurable for each
ARTIM Timeout service, see 6.2
Configuration for
DIMSE Services
110
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
8.2 AE Specifications
8.2.1 Logiq Scanner Application Entity
8.2.1.1 Sequencing of Real-World Activities
111
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
112
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
113
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
114
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
115
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
116
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
117
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
118
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
If the Logiq scanner is unable to establish an association with the Retrieve SCP, it will retry for the
configured number of times.
This section describes the transfer syntax preference for this Real World Activity
Table 7.2-1: Transfer Syntax Selection Preference Order - Image SOP Classes for Real World Activity -
Exam Retrieve/ Image Receive
Preference Order Transfer Syntax UID Comment
119
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
120
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Exception Behavior
Timeout Notifies the error to user and logs the error details
Association aborted Notifies the error to user and logs the error details
Network Disconnect Notifies the error to user and logs the error details
Exception Behavior
Association aborted Notifies the error to user and logs the error details
8.3.2.1.1 SCU of the Modality Worklist Information Model Find SOP Class - C-
FIND
Table 7.3-1: Status Codes for C-FIND of the Modality Worklist information Model SOP Class - SCU
Success Matching is complete - No final identifier 0000H Displays the query result to user.
is supplied
121
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Cancel Matching terminated due to cancel FE00H No error dialog is displayed, error cod
e is logged.
Pending Matches are continuing - Current Match FF00H Waits for query completion
is supplied and any Optional Keys were
supported in the same manner as
Required Keys.
Matches are continuing - Warning that FF01H Waits for query completion
one or more Optional Keys were not
supported for existence for this Identifier
8.3.2.1.2 SCP of the Modality Worklist Information Model Find SOP Class - C-
FIND - N/A
Not Applicable
8.3.2.2.1 SCU of the Modality Performed Procedure Step SOP Class - N-CREATE
Table 7.3-2: Status Codes for N-CREATE of the Modality Performed Procedure Step SOP Class - SCU
Warning Attribute Value Out of Range 0116H Job is marked as failed after retrying
configured number of times. Error is
logged.
122
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
123
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
8.3.2.2.2 SCU of the Modality Performed Procedure Step SOP Class - N-SET
Table 7.3-3: Status Codes for N-SET of the Modality Performed Procedure Step SOP Class - SCU
Warning Attribute Value Out of Range 0116H Job is marked as failed after retrying
configured number of times. Error is
logged.
124
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
8.3.2.2.3 SCP of the Modality Performed Procedure Step SOP Class - N-CREATE
- N/A
Not Applicable
8.3.2.2.4 SCP of the Modality Performed Procedure Step SOP Class - N-SET -
N/A
Not Applicable
Not Applicable
125
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
SCU of Request UPS Cancel on UPS Push SOP Class - N-ACTION - N/A
Not Applicable
Not Applicable
Not Applicable
Not Applicable
Not Applicable
SCU of the Change UPS State on UPS Pull SOP Class - N-ACTION - N/A
Not Applicable
Not Applicable
Not Applicable
Not Applicable
SCU of the - Request UPS Cancel on UPS Watch SOP Class - N-ACTION - N/A
Not Applicable
Not Applicable
126
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Not Applicable
SCP of the Request UPS Cancel on UPS Push SOP Class - N-ACTION - N/A
Not Applicable
Not Applicable
Not Applicable
Not Applicable
Not Applicable
SCP of the Change UPS State on UPS Pull SOP Class - N-ACTION - N/A
Not Applicable
Not Applicable
Not Applicable
Not Applicable
SCP of the Request UPS Cancel on UPS Watch SOP Class - N-ACTION - N/A
Not Applicable
127
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Not Applicable
128
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
6. Scanner is in Battery
mode
8. Scanner is in wireless
mode and Receive via
Wireless is disabled.
Failure
Error: Data Set does not A901 Never
match SOP Class
129
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
8.3.2.6.1 SCU of the Storage Commitment Push Model SOP Class - N-ACTION
Table 7.3-6: Status Codes N-ACTION of the Storage Commitment Push Model SOP Class - SCU
* Any other status code. * Job is marked as failed after retrying configured
number of times. Error is logged.
130
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
131
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Failure Refused: Out of resources A700H Error dialog is displayed and error
code is logged
Error: Identifier does not match SOP Class A900H Error dialog is displayed and error
code is logged
SOP Class Not Supported 0122H Error dialog is displayed and error
code is logged
Pending Matches are continuing - Current Match FF00H Waits for query completion
is supplied and any Optional Keys were
supported in the same manner as
Required Keys.
Matches are continuing - Warning that FF01H Waits for query completion
one or more Optional Keys were not
supported for existence for this Identifier
(0000,1022)
(0000,1023)
Warning Sub-operations B000 (0000,1020) If scanner receives this status after receiving
Complete - One images, the scanner notifies user about the
or more Failures (0000,1022) partial transfer of exams.
132
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Failed Out of Resources A701 (0000,0902) Error dialog is displayed, and error code is
- Unable to logged
calculate number
of matches
Out of Resources A702 (0000,1020) Error dialog is displayed, and error code is
- Unable to logged
perform sub- (0000,1021)
operations
(0000,1022)
(0000,1023)
Move Destination A801 (0000,0902) Error dialog is displayed, and error code is
unknown logged
Identifier does A900 (0000,0901) Error dialog is displayed, and error code is
not match SOP logged
Class (0000,0902)
Unable to process Cxxx (0000,0901) Error dialog is displayed, and error code is
logged
(0000,0902)
Cancel Sub-operations FE00 (0000,1020) Error dialog is displayed, and error code is
terminated due logged
to Cancel (0000,1021)
Indication
(0000,1022)
(0000,1023)
(0000,1022)
(0000,1023)
133
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Table 7.3-10: Status CodesN-CREATE of the Basic Film Session SOP Class - SCU
Status Class Further Meaning Status Code Behavior
134
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Not Applicable
Not Applicable
Not Applicable
Table 7.3-11: Status CodesN-CREATE of the Basic Film Box SOP Class - SCU
Warning Attribute List Error 0107 Job is marked as failed after retrying
configured number of times. Error is
logged.
Attribute Value Out Of Range 0116 Job is marked as failed after retrying
configured number of times. Error is
logged.
Requested Min Density or Max Density B605 Scanner shall continue to process the job
outside of printer's operating range
135
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
There is an existing Film Box that has not C616 Job is marked as failed after retrying
been printed and N-ACTION at the Film configured number of times. Error is
Session level is not supported. A new Film logged.
Box will not be created when a previous
Film Box has not been printed
Not Supported
Table 7.3-12: Status CodesN-DELETE of the Basic Film Box SOP Class - SCU
Behavior
Status class Further Meaning Status Code
136
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Table 7.3-13: Status Codes N-ACTION of the Basic Film Box SOP Class - SCU
Film Box SOP Instance hierarchy does not B603H Scanner shall continue to process the
contain Image Box SOP Instances (empty job
page)
137
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Image size is larger than Image Box size. B604H Job is marked as failed after retrying
The image has been demagnified. configured number of times. Error is
logged.
Image size is larger than Image Box size. B609H Scanner shall continue to process the
The image has been cropped to fit. job
Image size or Combined Print Image Size is B60AH Scanner shall continue to process the
larger than Image Box size. The image or job
combined Print Image has been decimated
to fit.
138
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Unable to create Print Job SOP Instance; C602 Job is marked as failed after retrying
print queue is full. configured number of times. Error is
logged.
Image size is larger than Image Box size. C603 Job is marked as failed after retrying
configured number of times. Error is
logged.
Combined Print Image Size is larger than C613 Job is marked as failed after retrying
Image Box size. configured number of times. Error is
logged.
8.3.2.8.3 SCU of the Basic Grayscale Image Box SOP Class - N-SET
Table 7.3-14: Status Codes N-SET of the Grayscale Image Box SOP Class - SCU
Warning Image size is larger than Image Box size. B604H Job is marked as failed after
The image has been demagnified. retrying configured number of
times. Error is logged.
Requested Min Density or Max Density B605H Scanner shall continue to process
outside of printer's operating range. the job
Image size is larger than Image Box size. B609H Scanner shall continue to process
The image has been cropped to fit. the job
Image size or Combined Print Image Size is B60AH Scanner shall continue to process
larger than Image Box size. The image or the job
combined Print Image has been decimated
to fit.
139
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
140
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Image size is larger than Image Box size. C603 Job is marked as failed after
retrying configured number of
times. Error is logged.
Insufficient memory in printer to store the C605 Job is marked as failed after
image. retrying configured number of
times. Error is logged.
Combined Print Image Size is larger than C613 Job is marked as failed after
Image Box size. retrying configured number of
times. Error is logged.
8.3.2.8.4 SCU of the Basic Color Image Box SOP Class - N-SET
Table 7.3-15: Status Codes N-SET of the Color Image Box SOP Class - SCU
Warning Image size is larger than Image Box size. B604H Job is marked as failed after
The image has been demagnified. retrying configured number of
times. Error is logged.
Image size is larger than Image Box size. B609H Scanner shall continue to process
The image has been cropped to fit. the job
Image size or Combined Print Image Size is B60AH Scanner shall continue to process
larger than Image Box size. The image or the job
combined Print Image has been decimated
to fit.
141
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Image size is larger than Image Box size. C603 Job is marked as failed after
retrying configured number of
times. Error is logged.
Insufficient memory in printer to store the C605 Job is marked as failed after
image. retrying configured number of
times. Error is logged.
142
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Combined Print Image Size is larger than C613 Job is marked as failed after
Image Box size. retrying configured number of
times. Error is logged.
Table 7.3-16: Status Codes N-EVENT-REPORT of the Printer SOP Class - SCU
143
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Table 7.3-17: Status Codes N-GET of the Printer SOP Class - SCU
Warning Optional Attribute Not Supported 0001 Scanner shall continue to process the
job
144
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
145
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
146
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
147
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
8.3.2.8.12 SCP of the Basic Grayscale Image Box SOP Class - N-SET - N/A
8.3.2.8.13 SCP of the Basic Color Image Box SOP Class - N-SET - N/A
8.3.2.8.14 SCP of the SCP of the Printer SOP Class- N/A
SCP of the Printer SOP Class – N-EVENT-REPORT- N/A
8.3.2.8.15 SCP of the Basic Annotation Box SOP Class – N-SET - N/A
8.3.2.8.16 SCP of the Print Job SOP Class- N/A
SCP of the Print Job SOP Class – N-EVENT-REPORT- N/A
8.3.2.8.18 SCP of the Printer Configuration Retrieval SOP Class - N-GET - N/A
148
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
149
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
9 Security
9.1 Introduction
The security section describes security features implemented by this product. It includes description of
non-DICOM network protocols, information to configure firewalls and application white-lists, list of
supported DICOM security profiles as well as Web Security features. Additionally, secured media storage,
VPN, etc are also specified in this security section.
150
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
151
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
152
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
153
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Appendices
A. Information Object Definitions (IODs)
This section provides the detailed content of the IODs natively created by LOGIQ Scanner, e.g images
created by an acquisition modality or evidence documents created on a review workstation (e.g. all IODs
that are marked in the Created Column in Table 1.1-1 Storage SOP Classes). Details on attribute coercion
are defined in Section 5.2.5.2.
Throughout the tables listed in Annex A the following codes are used for the Source and Presence
columns.
In the Source Column, the following values are supported:
• FIXED: the value is pre-defined and cannot be modified.
• GENERATED: the value is generated by the system.
• CONFIGURATION: the value is copied from system configuration.
• MWL: the value is copied from modality worklist.
• USER: the value is entered by the user.
• SCANNED: the value is read from a barcode scanner or similar device.
• EMPTY: the attribute is sent without value.
• SRC_INSTANCE: the value is copied from previously created instances.
The Presence columns reflect the usage of the module, functional group macro, attributes or value in the
LOGIQ Scanner implementation and is not necessarily the same as defined in the DICOM standard. For
the Presence column the following values are supported:
• ALWAYS: the module, functional group macro, attributes or value is always present
• CONDITIONAL: the presence of the the module, functional group macro, attributes or value is
dependent on a condition. The condition must be listed in the Conditions column
• EMPTY: The attribute is present but without a value (zero length)
154
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
155
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
If Patient
Patient's Age (0010,1010) GENERATED CONDITIONAL ALWAYS Birth Date is
valid
Patient's Size (0010,1020) USER, MWL CONDITIONAL ALWAYS Present if
worklist
entry
contains the
tag or user
manually
enters it
Patient's Weight (0010,1030) USER, MWL CONDITIONAL ALWAYS Present if
worklist
entry
contains the
tag or user
manually
enters it
Additional (0010,21B0) MWL CONDITIONAL ALWAYS Taken from
Patient History worklist if it
is there
Admitting (0008,1080) USER CONDITIONAL ALWAYS Taken from UI - Present if
Diagnoses Indication user
Description manually
enters it
Visit Comments (0038,4000) USER CONDITIONAL ALWAYS Taken from UI - Present if
Comments user
manually
enters it
General Series
156
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
157
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Procedure
Step is
enabled.
Performed (0040,0254) MWL, USER CONDITIONAL ALWAYS Takes the value Used if
Procedure Step of Requested Modality
Description Procedure Performed
Description or Procedure
user selected Step is
Exam enabled.
Description
Performed (0040,0260) MWL CONDITIONAL ALWAYS For Stess Echo Taken from
Protocol Code Exams, see worklist if it
Sequence is
there(Sched
uled
Protocol
Code
Sequence)
or in case of
Stress test
General Equipment
158
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Value 3 is the
description of
the type of
application
per the DICOM
PS3.3, section
C.8.5.6.1.1;
Supported
enumerated
values are:
ABDOMINAL
GYNECOLOGY
OBSTETRICAL
PEDIATRIC
SMALL PARTS
VASCULAR
SCROTAL
INTRACARDIAC
Value 4 is a
description of
the mode per
the DICOM
PS3.3, section
C.8.5.6.1.1;
Supported
enumerated
values are sum
of values in
hexadecimal
encoded, which
based on:
Value Meaning
0001 2D
Imaging
0002 M-
Mode
0004 CW
Doppler
0008 PW
Doppler
0010 Color
Doppler
0020 Color M-
Mode
0040 3D
Rendering
159
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
0100 Color
Power Mode
Values 5 and 6
may be used for
private data.
Lossy Image (0028,2110) GENERATED ALWAYS ALWAYS Set to 01, if
Compression image is lossy
compressed.
Set to 00,
otherwise
Presentation LUT (2050,0020) FIXED CONDITIONAL ALWAYS IDENTITY When GSDF
Shape enabled
Lossy Image (0028,2112) GENERATED CONDITIONAL ALWAYS Used if lossy
Compression compressed
Ratio
Image Pixel
If contrast
agent is
used, the
value is
taken from
worklist
(Requested
160
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Contrast
Agent).
Palette Color Lookup Table
Value Meaning
0 None or not
applicable
1 2D (tissue or
flow)
2 M -Mode
(tissue or flow)
161
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
3 Spectral (CW
or PW Doppler)
162
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
3 cm
4 seconds
>Physical Units Y (0018,6026) GENERATED ALWAYS ALWAYS Enumerated
Direction Values
supported:
3 cm
4 seconds
7 cm/sec
>Reference Pixel (0018,6028) GENERATED ALWAYS ALWAYS
Physical Value X
>Reference Pixel (0018,602A) GENERATED ALWAYS ALWAYS
Physical Value Y
>Physical Delta X (0018,602C) GENERATED ALWAYS ALWAYS
>Physical Delta Y (0018,602E) GENERATED ALWAYS ALWAYS
>Transducer (0018,6030) GENERATED ALWAYS ALWAYS
Frequency
>Pulse (0018,6032) GENERATED ALWAYS ALWAYS
Repetition
Frequency
US Image
Value 3 is the
description of
the type of
application
163
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Value 4 is a
description of
the mode per the
DICOM PS3.3,
section
C.8.5.6.1.1;
Supported
enumerated
values are sum
of values in
hexadecimal
encoded, which
based on:
Value Meaning
0001 2D
Imaging
0002 M- Mode
0004 CW
Doppler
0008 PW
Doppler
0010 Color
Doppler
0020 Color M-
Mode
0040 3D
Rendering
0100 Color
Power Mode
Values 5 and 6
may be used for
private data.
Lossy Image (0028,2110) GENERATED ALWAYS ALWAYS Set to 01, if
Compression image is lossy
compressed.
Set to 00,
otherwise
164
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
165
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
166
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Con- Presence
Presence Condition
Attribute Name Tag VR VM tains (Attribute Value Description
(Value) s
PHI )
GEMS_Ultrasound_MovieGroup_001
Private Creator 7FE1,00xx LO 1 GEMS_Ultr If Allow If so configured,
asound_M Raw Data the product will
ovieGroup is enabled send ultrasound
_001 raw data
information in
private data
elements
designated by
this Private
Creator element.
GE_GROUP
Private Creator 6005,00xx LO 1 GE_GROUP The product will
send
measurements
raw data
information in
private data
167
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Con- Presence
Presence Condition
Attribute Name Tag VR VM tains (Attribute Value Description
(Value) s
PHI )
elements
designated by
this Private
Creator element.
168
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Con- Presence
Presence Condition
Attribute Name Tag VR VM tains (Attribute Value Description
(Value) s
PHI )
169
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
170
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
LOGIQE10
LOGIQE10s
LOGIQE11
LOGIQE20
LOGIQFortis
LOGIQFortisPro
LOGIQFortisExpre
LOGIQFortisSuper
LOGIQFortisExper
LOGIQFortisPlus
LOGIQFortisPower
For example,
LOGIQE10:R3.0.2S
171
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
172
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
173
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Vector
(0018,1065)
174
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
175
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
176
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
177
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
178
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
179
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
See DICOM
PS3.3, section
C.7.6.23.4 for
details.
>Blending (0028,1406) GENERATED CONDITIONAL CONDITIONAL Used if
Weight Constant (0028,1405
) value is
"CONSTAN
T".
>Blending (0028,1407) GENERATED CONDITIONAL CONDITIONAL Used if
Lookup Table (0028,1405
Descriptor ) value is
"TABLE".
180
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
See DICOM
PS3.3, C.7.6.23.4
for details.
>Blending (0028,1406) GENERATED CONDITIONAL CONDITIONAL Used if
Weight Constant (0028,140
D) is
CONSTANT.
>Blending (0028,1407) GENERATED CONDITIONAL CONDITIONAL Used if
Lookup Table (0028,140
Descriptor D) is
TABLE.
>Blending (0028,1408) GENERATED CONDITIONAL CONDITIONAL Used if
Lookup Table (0028,140
Data D) is
TABLE.
Enhanced (0028,140B) GENERATED CONDITIONAL CONDITIONAL Used if
Palette Color Data Path
Lookup Table Assignmen
Sequence t
(0028,1402
) is present
with a
value
other than
PRIMARY_
PVALUES.
181
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
182
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
See DICOM
PS3.3, C.8.24.3.4
for further
explanation.
>Code Value (0008,0100) FIXED ALWAYS ALWAYS 125241
>Coding Scheme (0008,0102) FIXED ALWAYS ALWAYS DCM
Designator
>Code Meaning (0008,0104) FIXED ALWAYS ALWAYS Plane scan
pattern
Transducer (0018,980D) GENERATED ALWAYS ALWAYS Geometric
Geometry Code structure of the
Sequence transducer.
183
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Only a single
item shall be
included in this
sequence.
See DICOM
PS3.3, C.8.24.3.4
for further
explanation.
>Code Value (0008,0100) GENERATED ALWAYS ALWAYS Set to “125254”
if the volume
acquired using
Mechanical 4D
probes or
“125253” in case
of Tru3D probes.
>Coding Scheme (0008,0102) GENERATED ALWAYS ALWAYS DCM
Designator
>Code Meaning (0008,0104) GENERATED ALWAYS ALWAYS Set to “Sector
ultrasound
transducer
geometry” if the
volume acquired
with Mechanical
4D probes or
“Curved linear
ultrasound
transducer
geometry” in
case of Tru3D
probes.
Transducer (0018,980E) GENERATED ALWAYS ALWAYS Technique used
Beam Steering by the
Code Sequence transducer for
beam steering.
One or more
items shall be
included in this
sequence.
If more than one
item is present,
the order is
significant from
plane-forming
technique to
volume-forming
technique.
See DICOM
PS3.3, C.8.24.3.4
for further
explanation.
>Code Value (0008,0100) FIXED ALWAYS ALWAYS 125258
>Coding Scheme (0008,0102) FIXED ALWAYS ALWAYS DCM
Designator
>Code Meaning (0008,0104) FIXED ALWAYS ALWAYS Mechanical
beam steering
184
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
185
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
186
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
mentioned in
section A.1.4
Secondary (0018,1019) GENERATED ALWAYS ALWAYS Same as the
Capture Device value of
Software Software
Versions Versions
(0018,1020)
mentioned in
section A.1.4
SC Image
187
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
188
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
General Study
Module
>Placer Order (0040,2016) EMPTY ALWAYS EMPTY
Number /
Imaging Service
Request
>Filler Order (0040,2017) EMPTY ALWAYS EMPTY
Number /
Imaging Service
Request
>Requested (0040,1001) MWL ALWAYS CONDITIONAL Taken from
Procedure ID Worklist if
its present
>Requested (0032,1060) MWL ALWAYS CONDITIONAL Taken from
Procedure Worklist if
Description its present
>Requested (0032,1064) EMPTY ALWAYS EMPTY
Procedure Code
Sequence
Performed (0040,A372) EMPTY ALWAYS EMPTY
Procedure Code
Sequence
Current (0040,A375) GENERATED ALWAYS ALWAYS List of all
Requested composite SOP
Procedure Instances of US
Evidence Single, US MF,
Sequence Enhanced US
Volume or
Secondary
Capture
referenced in
the Content
Sequence
(0040,A730).
>Include GENERATED ALWAYS ALWAYS
‘Hierarchical SOP
Instance
Reference
Macro’
Author Observer (0040,A078) GENERATED ALWAYS ALWAYS Filled with
Sequence observer
information.
SR Document Content
189
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
190
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
191
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Con- Presence
Presence Condition
Attribute Name Tag VR VM tains (Attribute Value Description
(Value) s
PHI )
GE Report Data
Private Creator 6007,00xx LO 1 GE Report The product will
Data send
measurements
raw data
information in
private data
elements
designated by
the Private
Creator element:
192
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
193
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
194
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Reference
Macro’
Author Observer (0040,A078) GENERATED ALWAYS ALWAYS Filled with
Sequence observer
information.
195
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
0000H: no
known
inconsistencies,
ignored when
reading.
Directory Record
(0004,1220) GENERATED ALWAYS ALWAYS
Sequence
>Offset of the Next
(0004,1400) GENERATED ALWAYS ALWAYS
Directory Record
FSC/FSU: Is set
to FFFFH
FSR: A value of
>Record In-use Flag (0004,1410) GENERATED ALWAYS ALWAYS
0000H: imply
skipping this
record
>Offset of
Referenced Lower-
(0004,1420) GENERATED ALWAYS ALWAYS
Level Directory
Entity
The values
support by FSC
and FSU are
• PATIENT
>Directory Record • STUDY
(0004,1430) GENERATED ALWAYS ALWAYS • SERIES
Type
• IMAGE
• SR
DOCUMENT
• ENCAP DOC
Contains the file
path consisting
of 5 elements:
1. “GEMS_IM
G” (if
IMAGE)
2. Month of
>Referenced File ID (0004,1500) GENERATED ALWAYS ALWAYS exam
3. Day of
exam
4. Patient
initials and
time of
exam
5. Time stamp
Is set to the
SOP class UID in
>Referenced SOP
(0004,1510) SRC_INSTANCE CONDITIONAL ALWAYS File if Directory
Class UID in File
Record Type is
IMAGE
Is set to the
SOP instance
>Referenced SOP UID in File if
(0004,1511) SRC_INSTANCE CONDITIONAL ALWAYS
Instance UID in File Directory
Record Type is
IMAGE
196
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Is set to the
Transfer Syntax
>Referenced
UID in File if
Transfer Syntax UID (0004,1512) SRC_INSTANCE CONDITIONAL ALWAYS
Directory
in File
Record Type is
IMAGE
>Record Selection
Keys
Patient Keys
SRC_INSTANCE ALWAYS CONDITIONAL Value may
be empty if
Patient's Name (0010,0010)
Emergency
Patient
Patient ID (0010,0020) SRC_INSTANCE ALWAYS ALWAYS
SRC_INSTANCE ALWAYS CONDITIONAL Value empty
if Birth date
Patient's Birth Date (0010,0030)
is not
provided
Patient's Sex (0010,0040) SRC_INSTANCE ALWAYS ALWAYS
Patient's Birth Time (0010,0032) SRC_INSTANCE ALWAYS ALWAYS
If user
manually
Other Patient Ids (0010,1000) SRC_INSTANCE CONDITIONAL ALWAYS entered
Other
Patient ID
Taken from
Ethnic Group (0010,2160) SRC_INSTANCE CONDITIONAL ALWAYS worklist if it
is there
Present if
worklist
entry
Patient Comments (0010,4000) SRC_INSTANCE CONDITIONAL ALWAYS contains the
tag or user
manually
enters it
Study Keys
Study Date (0008,0020) SRC_INSTANCE ALWAYS ALWAYS
Study Time (0008,0030) SRC_INSTANCE ALWAYS ALWAYS
Accession Number (0008,0050) SRC_INSTANCE Present if
worklist
entry
ALWAYS CONDITIONAL contains the
tag or user
manually
enters it
Referring (0008,0090) SRC_INSTANCE Present if
Physician's Name worklist
ALWAYS CONDITIONAL entry
contains the
tag or user
197
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
manually
enters it
Study Description (0008,1030) SRC_INSTANCE Present if
worklist
entry
ALWAYS CONDITIONAL contains the
tag or user
manually
enters it
Present if
Admitting
user
Diagnoses (0008,1080) SRC_INSTANCE CONDITIONAL ALWAYS
manually
Description
enters it
If Patient
Patient's Age (0010,1010) SRC_INSTANCE CONDITIONAL ALWAYS Birth Date is
valid
Present if
worklist
entry
Patient's Size (0010,1020) SRC_INSTANCE CONDITIONAL ALWAYS contains the
tag or user
manually
enters it
Present if
worklist
entry
Patient's Weight (0010,1030) SRC_INSTANCE CONDITIONAL ALWAYS contains the
tag or user
manually
enters it
Taken from
Additional Patient
(0010,21B0) SRC_INSTANCE CONDITIONAL ALWAYS worklist if it
History
is there
Study Instance UID (0020,000D) SRC_INSTANCE ALWAYS ALWAYS
Study ID (0020,0010) SRC_INSTANCE ALWAYS ALWAYS
Series Keys
Series Date (0008,0021) SRC_INSTANCE ALWAYS ALWAYS
Series Time (0008,0031) SRC_INSTANCE ALWAYS ALWAYS
Modality (0008,0060) SRC_INSTANCE ALWAYS ALWAYS
Manufacturer (0008,0070) SRC_INSTANCE ALWAYS ALWAYS
Institution Name (0008,0080) SRC_INSTANCE ALWAYS ALWAYS
Station Name (0008,1010) SRC_INSTANCE ALWAYS ALWAYS
Institutional (0008,1040) SRC_INSTANCE
ALWAYS ALWAYS
Department Name
Performing (0008,1050) SRC_INSTANCE Present if
Physicians’ Name worklist
CONDITIONAL ALWAYS entry
contains the
tag or user
198
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
manually
enters it
Operator’s Name (0008,1070) SRC_INSTANCE ALWAYS ALWAYS
Manufacturer's (0008,1090) SRC_INSTANCE
ALWAYS ALWAYS
Model Name
Software Versions (0018,1020) SRC_INSTANCE ALWAYS ALWAYS
Protocol Name (0018,1030) SRC_INSTANCE Used in
Stress Echo
CONDITIONAL ALWAYS
Exam
Protocol
Series Description (0008,103E) SRC_INSTANCE Taken from
worklist if it
is there
(from
CONDITIONAL ALWAYS
Scheduled
Procedure
Step
Description)
Request Attributes (0040,0275) Taken from
Sequence SRC_INSTANCE CONDITIONAL ALWAYS worklist if it
is there
>Scheduled (0040,0007) Taken from
Procedure Step SRC_INSTANCE CONDITIONAL ALWAYS worklist if it
Description is there
>Scheduled (0040,0009) Taken from
Procedure Step ID SRC_INSTANCE CONDITIONAL ALWAYS worklist if it
is there
>Requested (0040,1001) Taken from
Procedure ID SRC_INSTANCE CONDITIONAL ALWAYS worklist if it
is there
Related Series (0008,1250) If instance is
Sequence SRC_INSTANCE CONDITIONAL ALWAYS SRDOC or
ENCAP DOC
>Study Instance (0020,000D)
SRC_INSTANCE ALWAYS ALWAYS
UID
>Series Instance (0020,000E)
SRC_INSTANCE ALWAYS ALWAYS
UID
Series Instance UID (0020,000E) SRC_INSTANCE ALWAYS ALWAYS
Series Number (0020,0011) SRC_INSTANCE ALWAYS ALWAYS
Image Keys
Specific Character (0008,0005)
GENERATED ALWAYS ALWAYS
Set
Image Type (0008,0008) SRC_INSTANCE ALWAYS ALWAYS
Content Date (0008,0023) SRC_INSTANCE ALWAYS ALWAYS
Content Time (0008,0033) SRC_INSTANCE ALWAYS ALWAYS
Stage Number (0008,2122) SRC_INSTANCE ALWAYS ALWAYS
View Number (0008,2128) SRC_INSTANCE ALWAYS ALWAYS
199
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
200
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
TID 5100
Vascular
Ultrasound
Procedure Report
TID 5101
Vascular Patient
Characteristics
TID 5102
Vascular Procedure
Summary Section TID 5104
Vascular
Measurement
Group
TID 5103
Vascular
Ultrasound
Section TID 300
Measurement
TID G301
GE Ultrasound User
Defined
Measurement
201
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
202
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
May contain
indications
MOD
See - $Anatomy
Ratio
203
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
NL Relation with Value Type Concept Name VM Req Condition Value Set
Parent Type Constraint
B.1.11
$Derivation = See
DCID (3627)
Measurement Type
204
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
NL Relation with Value Type Concept Name VM Req Condition Value Set
Parent Type Constraint
See - $Anatomy
Ratio
DT (121070, EV (G-C171,
DCM, SRT,
‘Findings’) "Laterality")
(T-40501, SRT, (G-A101, SRT, DCID 12105 Anatomy
‘Blood Vessel Left) for Left, GEU Code and Description
of Head’) Intracranial
(G-A100, SRT, Cerebral parameter
Right) for Vessels
Right. Or ICA (T-45300, SRT, ‘Internal Carotid Artery’)
AC (125106,DCM,"Doppler Angle")
205
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
DT (121070, EV (G-C171,
DCM, SRT,
‘Findings’) "Laterality")
(T-45005, SRT, (G-A101, SRT, DCID (12104) DCID Anatomy
‘Artery of Left) for Left, Extracranial (1212 GEU Code and Description
neck) or Arteries 3) parameter
Carot
(G-A100, SRT, id
Right) for VERT (T-45700, SRT, ‘Vertebral Artery’)
Ratio
Right. s
CCA (T-45100, SRT, ‘Common Carotid Artery’)
2D/Stenosi
(T-45300, SRT, Internal Carotid Artery)
s
206
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
DT (121070, EV (G-C171,
DCM, SRT,
‘Findings’) "Laterality")
(T-47040, SRT, (G-A101, SRT, DCID (12109) Anatomy GEU
‘Artery of Left) for Left, Lower Code and Description
parameter
Lower or Extremity
Extremity’) Arteries
(G-A100, SRT, ComIliac (T-46710, SRT, ‘Common Iliac Artery’)
Right) for
Right. ExtIliac (T-46910, SRT, ‘External Iliac Artery’)
207
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
AntAccessory
(GEU-1004-112, 99GEMS, ‘AASV’);
Saph
208
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
DT (121070, EV (G-C171,
DCM, SRT,
‘Findings’) "Laterality")
(T-47020, SRT, (G-A101, SRT, DCID (12107) Section Artery of Upper Extremity
‘Artery of Left) for Left, Upper
Upper or Extremity
Extremity’) Arteries Anatomy GEU
Code and Description
(G-A100, SRT, parameter
Right) for
Right. SUBC (T-46100, SRT, ‘Subclavian artery’)
209
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
210
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
DT (121070, EV (G-C171,
DCM, SRT,
‘Findings’) "Laterality")
(T-71019, SRT, (G-A101, SRT, DCID(12115) DCID( Anatomy GEU
‘Vascular Left) for Left, Renal Vessels 1212 Code and Description
parameter
Structure of or 4)
Kidney’) Renal
(G-A100, SRT, MRenalA (T-46600, SRT, ‘Renal Artery’)
Ratio
Right) for s
Right. RenalV (T-48740, SRT, ‘Renal Vein’)
211
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Anatomy
GEU Code and Description
parameter
Gall
(T-63000, SRT, ‘Gall Bladder’)
Bladder
AutoConto
(F-03FCD, SRT, ‘Lesion Finding’)
urROI
Shearwave
Code and
GEU
Description Remarks
parameter
Shearwave (GEU-1010-01,
Stiffness 99GEMS,
212
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
DT (121070, EV (G-C171,
DCM, SRT,
‘Findings’) "Laterality")
‘Shearwave
Stiffness’)
(GEU-1010-02, Refer
Shearwave 99GEMS,
Velocity ‘Shearwave
Velocity)
(GEU-1010-03,
Shearwave
99GEMS,
Stiffness
‘Shearwave
Velocity
Stiffness Velocity’)
Attenuatio
Code and
n GEU Remarks
Description
parameter
(GEU-1013-01, Refer
99GEMS,
UGAPC
‘Attenuation
Coefficient’)
(GEU-1013-02,
UGAPR 99GEMS,
‘Attenuation Rate’)
(GEU-1013-03,
99GEMS,
UGAPCR
‘Attenuation
Coefficient Rate’)
213
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
DT (121070, EV (G-C171,
DCM, SRT,
‘Findings’) "Laterality")
(T-487A0, (G-A101, SRT, DCID 12113 Anatomy
SRT, ‘Vein of Left) for Left, or (12114) GEU Code and Description
Abdomen) Abdominal parameter
(G-A100, SRT, Veins lateral
Right) for
Right. Or (unilateral) IIV (T-48940, SRT, ‘Internal Iliac Vein’)
214
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
DT (121070, EV (G-C171,
DCM, SRT,
‘Findings’) "Laterality")
(T-D000F, SRT, (G-A101, SRT, DCID 12113 Anatom
'Vascular Left) for Left, or (12114) y GEU
Graft') Abdominal Code and Description
(G-A100, SRT, parame
Veins lateral ter
Right) for
Right. Or (unilateral)
Limb (GEU-1004-46, 99GEMS, 'Limb')
(G-A103, SRT,
Unilateral) Inflow (GEU-1004-63, 99GEMS, 'Inflow')
Ante
(T-49215, SRT, 'Antecubital Vein')
Cub
UserVes
(GEU-1005-6, 99GEMS, 'User Vessel Anatomy')
sel
Pre-
(GEU-1004-71, 99GEMS, 'Pre-Stent')
Stent
Post-
(GEU-1004-72, 99GEMS, 'Post-Stent')
Stent
215
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
- Trans (G-A117.
Transverse, SRT)
216
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Note (*) when there is no Sidedness or Locations, the SR nodes are not populated
217
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
218
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
219
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
1 CONTAINER EV (GEU-1010-01, 1 M
99GEMS, “Shearwave
Stiffness”)
4 CONTAINER EV (GEU-1010-02, 1 M
99GEMS, “Shearwave
Velocity”)
7 CONTAINER EV (GEU-1010-03, 1 M
99GEMS, “Shearwave
Stiffness Velocity”)
220
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
1 CONTAINER EV (GEU-1012-01, 1 M
99GEMS, “Attenuation
Coefficient”)
MOD
6 CONTAINER EV (GEU-1012-02, 1 M
99GEMS, “Attenuation
Rate”)
MOD
221
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
222
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
MD, Md_velocitySD,
(11665-7, LN, ‘Minimum Diastolic Velocity’)
Md_velocityDS
223
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
224
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
RenalVolume
inside (T-71000, SRT, Kidney) container
BladderVolume
Inside (T-74000, SRT, Bladder) container
225
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
(Length, SRT,G-A22A)
CBD
inside (T-60610, SRT, Bile Duct)
AP = (Diameter,SRT, M-02550),
with Measurement Label (Anterior-Posterior, DCM, 122675),
AP = (Diameter,SRT, M-02550),
with Measurement Label (Anterior-Posterior, DCM, 122675),
226
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
AP = (Diameter,SRT, M-02550),
with Measurement Label (Anterior-Posterior, DCM, 122675),
AP = (Diameter,SRT, M-02550),
with Measurement Label (Anterior-Posterior, DCM, 122675),
The General Ultrasound User Defined Measurement Template provides a CONTAINER with a structure for
reporting user-defined measurements and calculations.
The difference between this from TID 5103, TID 300 or 1414 for general measurement is the template
uses labels that specified by users at configuration time rather assigning specific codes to the individual
measurements.
Note: In practice, to assign specific codes to user defined measurements are impractical and cumber-
some since the system/organization has to maintain database/map of codes and labels with
public/private code designator schemes.
227
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
2 > HAS CONCEPT CODE EV (G-C171, SRT, ‘Laterality’) 1 MC IFF anatomy DCID (244) Laterality
MOD has laterality
CONCEPT
MOD
General Anatomy
mainly used in PWD
and B-Mode and when
user added individual
folders and
measurements.
5 >> HAS TEXT DT (GEU-1005-7, 99GEMS, 1 MC IFF The Labels from user which
container uses defined at
CONCEPT ‘Anatomy Label’)
Finding as configuration
MOD Concept Name
When individually
added ‘Add
Measurement’ please
refer to DCID (G5001)
GE Defined
Measurements since
new measurement
codes are determined
228
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
by its measurement
units
$Derivation = DCID
(3627) Measurement
Type
$Label = Measurement
Label created by User
229
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
230
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
TID 5200
Echocardiography
Procedure Report
TID 5203
Echo Measurement
TID 5202
Echo Section
TID 300
Measurement
TID 5204
Wall Motion
Analysis
TID G301
GE Ultrasound User
Defined
Measurement
EV (125200, DCM,
“Adult
1 CONTAINER Echocardiography 1 M
Procedure
Report”)
231
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
232
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
$Procedure = DT
(P5-B3121, SRT,
DTID (5204) Wall "Echocardiography
16 CONTAINS INCLUDE 1-n U
Motion Analysis for Determining
Ventricular
Contraction")
NL Rel with Parent VT Concept Name VM Req Condition Value Set Constraint
Type
233
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
NL Rel with Parent VT Concept Name VM Req Condition Value Set Constraint
Type
2 > HAS CONCEPT MOD CODE EV (G-C048, SRT, 1 U BCID 12221 “Flow
"Flow Direction") Direction”
3 > HAS CONCEPT MOD CODE EV (R-40899, SRT, 1 U DCID 12234 “Respiration
"Respiratory Cycle State”
Point")
4 > HAS CONCEPT MOD CODE EV (R-4089A, SRT, 1 U DCID 12233 “Cardiac
"Cardiac Cycle Point") Phase”
5 > HAS ACQ CONTEXT CODE EV (G-0373, SRT, 1 U DCID 12224 “Ultrasound
"Image Mode") Image Modes”
7 > HAS ACQ CONTEXT CODE EV (18139-6, LN, 1 U BCID 12002 “Ultrasound
"Stage") Protocol Stage Types”
234
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
NL Rel with Parent VT Concept Name VM Req Condition Value Set Constraint
Type
11 >> HAS PROPERTIES CODE EV (F-32050, SRT, 1 MC IF row 12 is DCID 3703 “Wall
> "Cardiac Wall absent Motion”
Motion")
B.2.2. Aorta
Table B.2-5 Aorta
Name of GEU Base Measurement Concept
Concept or Acquisition Context Modifier
parameter Name
(GEU-1005-45, 99GEMS, "Ao
2D/Ao Diam SVals Diam SVals")
235
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
236
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
237
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
AV SV (F-32120, SRT, "Stroke Volume") (G-C0E3, SRT, “Finding Site”) = (T-42000, SRT, "Aorta")
AV SI (F-00078, SRT, "Stroke Index") (G-C0E3, SRT, “Finding Site”) = (T-42000, SRT, "Aorta")
AV CO (F-32100, SRT, "Cardiac Output") (G-C0E3, SRT, “Finding Site”) = (T-42000, SRT, "Aorta")
AV CI (F-32110, SRT, "Cardiac Index") (G-C0E3, SRT, “Finding Site”) = (T-42000, SRT, "Aorta")
238
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
AVA/AV SV (F-32120, SRT, "Stroke Volume") (G-C0E3, SRT, “Finding Site”) = (T-42000, SRT, "Aorta")
AVA/AV SI (F-00078, SRT, "Stroke Index") (G-C0E3, SRT, “Finding Site”) = (T-42000, SRT, "Aorta")
AVA/AV CO (F-32100, SRT, "Cardiac Output") (G-C0E3, SRT, “Finding Site”) = (T-42000, SRT, "Aorta")
AVA/AV CI (F-32110, SRT, "Cardiac Index") (G-C0E3, SRT, “Finding Site”) = (T-42000, SRT, "Aorta")
239
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
240
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
(GEU-1005-200, 99GEMS, “ (G-0373, SRT, “Image Mode”) = (R-409E2, SRT, "Doppler Color
PISA/AR/Radius PISA/AR/Radius”) Flow")
(GEU-1005-201, 99GEMS, “ (G-0373, SRT, “Image Mode”) = (R-409E2, SRT, "Doppler Color
PISA/AR/Velocity PISA/AR/Velocity”) Flow")
241
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
(GEU-1005-101, 99GEMS, "PDA Systolic (G-C0E3, SRT, “Finding Site”) = (D4-32012, SRT,"Patent ductus
PDA Systolic PG PG") arteriosus")
(GEU-1005-102, 99GEMS, "PDA (G-C0E3, SRT, “Finding Site”) = (D4-32012, SRT,"Patent ductus
PDA Diastolic Diastolic") arteriosus")
(GEU-1005-103, 99GEMS, "PDA Diastolic (G-C0E3, SRT, “Finding Site”) = (D4-32012, SRT,"Patent ductus
PDA Diastolic PG PG") arteriosus")
*1: Pre Ductal means the narrowing is proximal to the ductus arteriosus.
*2: Post Ductal means the narrowing is distal to the insertion of the ductus arteriosus.
242
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
2D/LA Major (G-A193, SRT, "Major Axis") (G-0373, SRT, “Image Mode”) = (G-03A2, SRT, "2D mode")
2D/LA Minor (G-A194, SRT, "Minor Axis") (G-0373, SRT, “Image Mode”) = (G-03A2, SRT, "2D mode")
2D/LAA diastole (GEU-1005-69, 99GEMS, "LAA diastole") (G-0373, SRT, “Image Mode”) = (G-03A2, SRT, "2D mode")
2D/LAA systole (17977-0, LN, "Left Atrium Systolic Area") (G-0373, SRT, “Image Mode”) = (G-03A2, SRT, "2D mode")
2D/LA Area (G-A166, SRT, "Area") (G-0373, SRT, “Image Mode”) = (G-03A2, SRT, "2D mode")
(29467-8, LN, "Left Atrium Superior- (R-4089A, SRT, “Cardiac Cycle Point”) = (F-32010, SRT,
LALd(A4C) Inferior Dimension, 4-chamber view") "Diastole")
243
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
244
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
245
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
(18026-5, LN, "Left Ventricular End (G-C036, SRT, “Measurement Method”) = (125207, DCM,
LVEDV(MOD BP)_03 Diastolic Volume") "Method of Disks, Biplane")
(18148-7, LN, "Left Ventricular End (G-C036, SRT, “Measurement Method”) = (125207, DCM,
LVESV(MOD BP)_03 Systolic Volume") "Method of Disks, Biplane")
(18043-0, LN, "Left Ventricular Ejection (G-C036, SRT, “Measurement Method”) = (125207, DCM,
EF(Biplane)_03 Fraction") "Method of Disks, Biplane")
2D/LV Major (G-A193, SRT, "Major Axis") (G-0373, SRT, “Image Mode”) = (G-03A2, SRT, "2D mode")
2D/LV Minor (G-A194, SRT, "Minor Axis") (G-0373, SRT, “Image Mode”) = (G-03A2, SRT, "2D mode")
2D/LVIDd Index (GEU-1005-139, 99GEMS, "LVIDd Index") (G-0373, SRT, “Image Mode”) = (G-03A2, SRT, "2D mode")
246
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
247
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
248
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
(G-0375, SRT, "Left Ventricular Diastolic (111031, DCM, “Image View”) = (G-A19C, SRT, "Apical four
LVAd(A4C) Area") chamber")
(G-0374, SRT, "Left Ventricular Systolic (111031, DCM, “Image View”) = (G-A19C, SRT, "Apical four
LVAs(A4C) Area") chamber")
249
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
(G-0375, SRT, "Left Ventricular Diastolic (111031, DCM, “Image View”) = (G-A19B, SRT, "Apical two
LVAd(A2C) Area") chamber")
250
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
(G-0374, SRT, "Left Ventricular Systolic (111031, DCM, “Image View”) = (G-A19B, SRT, "Apical two
LVAs(A2C) Area") chamber")
(G-0375, SRT, "Left Ventricular Diastolic (111031, DCM, “Image View”) = (G-0395, SRT, "Apical long
LVAd(LAX) Area") axis")
251
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
(G-0374, SRT, "Left Ventricular Systolic (111031, DCM, “Image View”) = (G-0395, SRT, "Apical long
LVAs(LAX) Area") axis")
252
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
(18026-5, LN, "Left Ventricular End (G-C036, SRT, “Measurement Method”) = (125207, DCM,
LVEDV(MOD BP) Diastolic Volume") "Method of Disks, Biplane")
(18148-7, LN, "Left Ventricular End (G-C036, SRT, “Measurement Method”) = (125207, DCM,
LVESV(MOD BP) Systolic Volume") "Method of Disks, Biplane")
(18043-0, LN, "Left Ventricular Ejection (G-C036, SRT, “Measurement Method”) = (125207, DCM,
EF(Biplane) Fraction") "Method of Disks, Biplane")
(G-0375, SRT, "Left Ventricular Diastolic (111031, DCM, “Image View”) = (G-039A, SRT, "Parasternal
LVAd(sax MV) Area") short axis at the Mitral Valve level")
(G-0374, SRT, "Left Ventricular Systolic (111031, DCM, “Image View”) = (G-039A, SRT, "Parasternal
LVAs(sax MV) Area") short axis at the Mitral Valve level")
(G-0375, SRT, "Left Ventricular Diastolic (111031, DCM, “Image View”) = (G-039B, SRT, "Parasternal
LVAd(sax PM) Area") short axis at the Papillary Muscle level")
(G-0374, SRT, "Left Ventricular Systolic (111031, DCM, “Image View”) = (G-039B, SRT, "Parasternal
LVAs(sax PM) Area") short axis at the Papillary Muscle level")
(G-0375, SRT, "Left Ventricular Diastolic (111031, DCM, “Image View”) = (G-0397, SRT, "Parasternal
LVAd(sax) Area") short axis")
(G-0374, SRT, "Left Ventricular Systolic (111031, DCM, “Image View”) = (G-0397, SRT, "Parasternal
LVAs(sax) Area") short axis")
(18026-5, LN, "Left Ventricular End (G-C036, SRT, “Measurement Method”) = (125227, DCM,
EDV(mod sim) Diastolic Volume") "Modified Simpson")
(18148-7, LN, "Left Ventricular End (G-C036, SRT, “Measurement Method”) = (125227, DCM,
ESV(mod sim) Systolic Volume") "Modified Simpson")
(18043-0, LN, "Left Ventricular Ejection (G-C036, SRT, “Measurement Method”) = (125227, DCM,
EF(mod sim) Fraction") "Modified Simpson")
253
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
(18026-5, LN, "Left Ventricular End (G-C036, SRT, “Measurement Method”) = (125228, DCM,
EDV(bullet) Diastolic Volume") "Bullet Method")
(18148-7, LN, "Left Ventricular End (G-C036, SRT, “Measurement Method”) = (125228, DCM,
ESV(bullet) Systolic Volume") "Bullet Method")
(18043-0, LN, "Left Ventricular Ejection (G-C036, SRT, “Measurement Method”) = (125228, DCM,
EF(bullet) Fraction") "Bullet Method")
(18026-5, LN, "Left Ventricular End (G-C036, SRT, “Measurement Method”) = (125211, DCM,
EDV(bp el) Diastolic Volume") "Biplane Ellipse")
(18148-7, LN, "Left Ventricular End (G-C036, SRT, “Measurement Method”) = (125211, DCM,
ESV(bp el) Systolic Volume") "Biplane Ellipse")
(18043-0, LN, "Left Ventricular Ejection (G-C036, SRT, “Measurement Method”) = (125211, DCM,
EF(bp el) Fraction") "Biplane Ellipse")
MM/LVIDd Index (GEU-1005-88, 99GEMS, "LVIDd Index") (G-0373, SRT, “Image Mode”) = (G-0394, SRT, "M mode")
254
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
MM/LVIDs Index (GEU-1005-89, 99GEMS, "LVIDs Index") (G-0373, SRT, “Image Mode”) = (G-0394, SRT, "M mode")
255
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
MM/HeartRate (8867-4, LN, "Heart rate") (G-0373, SRT, “Image Mode”) = (G-0394, SRT, "M mode")
MM/LVET (20222-6, LN, "Ejection Time") (G-0373, SRT, “Image Mode”) = (G-0394, SRT, "M mode")
(G-038F, SRT, "Cardiovascular Orifice (G-C0E3, SRT, “Finding Site”) = (T-32650, SRT, "Left Ventricular
MP/LVOT Diam Diameter") Outflow Tract")
256
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
257
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
(G-038F, SRT, "Cardiovascular Orifice (G-C0E3, SRT, “Finding Site”) = (T-32650, SRT, "Left Ventricular
AP/LVOT Diam Diameter") Outflow Tract")
(18037-2, LN, "Mitral Valve E-Wave Peak (G-C0E3, SRT, “Finding Site”) = (T-35313, SRT, "Mitral
MV Eann Velocity Velocity") Annulus")
(20269-7, LN, "Mitral Valve Acceleration (G-C048, SRT, “Direction of Flow”) = (R-42E61, SRT,
MR Acc Slope Slope") "Regurgitant Flow")
2D/MVL Thck (59110-7, LN, "Leaflet Thickness") (G-0373, SRT, “Image Mode”) = (G-03A2, SRT, "2D mode")
258
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
MM/MV CE Dist (59122-2, LN, "C-E Distance") (G-0373, SRT, “Image Mode”) = (G-0394, SRT, "M mode")
MM/MV E/A Ratio (18038-0, LN, "Mitral Valve E to A Ratio") (G-0373, SRT, “Image Mode”) = (G-0394, SRT, "M mode")
259
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
(GEU-1005-36, 99GEMS,
PISA/MR/Radius "PISA/MR/Radius")
(GEU-1005-37, 99GEMS,
PISA/MR/Velocity "PISA/MR/Velocity")
260
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
MV E/Eprime (59111-5, LN, "E Velocity to Annulus E (G-0373, SRT, “Image Mode”) = (R-409E4, SRT, "Doppler
Ratio/Calc Velocity Ratio") Pulsed")
(GEU-1005-133, 99GEMS, "MV Aprime (G-0373, SRT, “Image Mode”) = (R-409E4, SRT, "Doppler
MV Aprime Velocity Velocity") Pulsed")
(G-0387, SRT, "Mitral Valve Closure to (G-0373, SRT, “Image Mode”) = (R-409E4, SRT, "Doppler
IMP/MCO Opening Time") Pulsed")
MV Aprime Septal (GEU-1005-174, 99GEMS, "TVD- A' Septal (G-0373, SRT, “Image Mode”) = (R-409E4, SRT, "Doppler
Velocity velocity(Tissue Velocity Doppler)") Pulsed")
(GEU-1005-175, 99GEMS, "TVD- E'/A' (G-0373, SRT, “Image Mode”) = (R-409E4, SRT, "Doppler
MV E/A Septal Ratio septal ratio(Tissue Velocity Doppler)") Pulsed")
MV Eprime Lateral (GEU-1005-176, 99GEMS, "TVD- E' lateral (G-0373, SRT, “Image Mode”) = (R-409E4, SRT, "Doppler
Velocity velocity (Tissue Velocity Doppler)") Pulsed")
MV Aprime Lateral (GEU-1005-177, 99GEMS, "TVD- A' lateral (G-0373, SRT, “Image Mode”) = (R-409E4, SRT, "Doppler
Velocity velocity (Tissue Velocity Doppler)") Pulsed")
(GEU-1005-178, 99GEMS, "TVD- E'/A' (G-0373, SRT, “Image Mode”) = (R-409E4, SRT, "Doppler
MV E/A Lateral Ratio lateral ratio (Tissue Velocity Doppler)") Pulsed")
(GEU-1005-181, 99GEMS, "TVD- E'/A' (G-0373, SRT, “Image Mode”) = (R-409E4, SRT, "Doppler
MV E/A Medial Ratio medial ratio (Tissue Velocity Doppler)") Pulsed")
261
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
2D/LAX/RPA area (GEU-1005-73, 99GEMS, "LAX/RPA area") (G-0373, SRT, “Image Mode”) = (G-03A2, SRT, "2D mode")
2D/LAX/LPA area (GEU-1005-74, 99GEMS, "LAX/LPA area") (G-0373, SRT, “Image Mode”) = (G-03A2, SRT, "2D mode")
2D/SAX/RPA area (GEU-1005-76, 99GEMS, "SAX/RPA area") (G-0373, SRT, “Image Mode”) = (G-03A2, SRT, "2D mode")
2D/SAX/LPA area (GEU-1005-75, 99GEMS, "SAX/LPA area") (G-0373, SRT, “Image Mode”) = (G-03A2, SRT, "2D mode")
(G-038A, SRT, "Main Pulmonary Artery (G-C0E3, SRT, “Finding Site”) = (T-44000, SRT, " Pulmonary
MPA Vmax Peak Velocity") Artery")
262
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
263
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
(20295-2, LN, "Time from Q wave to (G-0373, SRT, “Image Mode”) = (R-409E4, SRT, "Doppler
SD/Q-to-PV close Pulmonic Valve Closes") Pulsed")
264
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
(GEU-1005-202, 99GEMS, “ (G-0373, SRT, “Image Mode”) = (R-409E2, SRT, "Doppler Color
PISA/PR/Radius PISA/PR/Radius”) Flow")
(GEU-1005-203, 99GEMS, “ (G-0373, SRT, “Image Mode”) = (R-409E2, SRT, „Doppler Color
PISA/PR/Velocity PISA/PR/Velocity”) Flow“)
265
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
2D/RA (M-02550, SRT, „Diameter“) (G-0373, SRT, “Image Mode”) = (G-03A2, SRT, „2D mode“)
2D/RAD Major (G-A193, SRT, „Major Axis“) (G-0373, SRT, “Image Mode”) = (G-03A2, SRT, „2D mode“)
2D/RAD Minor (G-A194, SRT, „Minor Axis“) (G-0373, SRT, “Image Mode”) = (G-03A2, SRT, „2D mode“)
2D/RAA diastole (GEU-1005-70, 99GEMS, „RAA diastole“) (G-0373, SRT, “Image Mode”) = (G-03A2, SRT, „2D mode“)
2D/RA Area (GEU-1005-134“, 99GEMS, „RA Area“) (G-0373, SRT, “Image Mode”) = (G-03A2, SRT, „2D mode“)
(29466-0, LN, „Right Atrium Superior- (R-4089A, SRT, “Cardiac Cycle Point”) = (F-32010, SRT,
RALd(A4C) Inferior Dimension, 4-chamber view“) „Diastole“)
(29466-0, LN, „Right Atrium Superior- (R-4089A, SRT, “Cardiac Cycle Point”) = (F-32020, SRT,
RALs(A4C) Inferior Dimension, 4-chamber view“) „Systole“)
266
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
2D/RVD Major (G-A193, SRT, „Major Axis“) (G-0373, SRT, “Image Mode”) = (G-03A2, SRT, „2D mode“)
2D/RVD Minor (G-A194, SRT, „Minor Axis“) (G-0373, SRT, “Image Mode”) = (G-03A2, SRT, „2D mode“)
2D/RVA diastole (GEU-1005-71, 99GEMS, „RVA diastole“) (G-0373, SRT, “Image Mode”) = (G-03A2, SRT, „2D mode“)
2D/RVA systole (GEU-1005-72, 99GEMS, „RVA systole“) (G-0373, SRT, “Image Mode”) = (G-03A2, SRT, „2D mode“)
267
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
MM/RVET (20222-6, LN, „Ejection Time“) (G-0373, SRT, “Image Mode”) = (G-0394, SRT, „M mode“)
268
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
(G-038E, SRT, „Cardiovascular Orifice (G-C048, SRT, “Direction of Flow”) = (R-42047, SRT, „Antegrade
TVA Area“) Flow“)
269
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
(20296-0, LN, „Time from Q wave to (G-0373, SRT, “Image Mode”) = (R-409E4, SRT, „Doppler
SD/Q-to-TV open Tricuspid Valve Opens“) Pulsed“)
270
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Name of GEU
Base Measurement Concept Name Concept or Acquisition Context Modifier
parameter
(G-0389, SRT, „Tricuspid Valve Closure to (G-0373, SRT, “Image Mode”) = (R-409E4, SRT, „Doppler
IMP/TCO Opening Time“) Pulsed“)
(GEU-1005-172, 99GEMS, „TVD- S‘ (G-0373, SRT, “Image Mode”) = (R-409E4, SRT, „Doppler
RV Sprime Velocity velocity (Tissue Velocity Doppler)“) Pulsed“)
(GEU-1005-204, 99GEMS, (G-0373, SRT, “Image Mode”) = (R-409E2, SRT, "Doppler Color
PISA/TR/Radius “PISA/TR/Radius”) Flow")
(GEU-1005-205, 99GEMS, (G-0373, SRT, “Image Mode”) = (R-409E2, SRT, „Doppler Color
PISA/TR/Velocity “PISA/TR/Velocity”) Flow“)
TR Env.Ti (GEU-1005-206, 99GEMS, “TR Env.Ti”) (R-409E3, SRT, “Doppler Continuous Wave”)
TV Env.Ti (GEU-1005-207, 99GEMS, “TV Env.Ti”) (R-409E3, SRT, “Doppler Continuous Wave”)
(18006-7, LN, "Inferior Vena Cava (R-40899, SRT, “Respiratory Cycle Point”) = (F-20010, SRT,
2D/IVC Diam Ins Diameter") "Inspiration")
(18006-7, LN, "Inferior Vena Cava (R-40899, SRT, “Respiratory Cycle Point”) = (F-20020, SRT,
2D/IVC Diam Exp Diameter") "Expiration")
(18007-5, LN, "Superior Vena Cava (R-40899, SRT, “Respiratory Cycle Point”) = (F-20010, SRT,
2D/SVC Diam Ins Diameter") "Inspiration")
(18007-5, LN, "Superior Vena Cava (R-40899, SRT, “Respiratory Cycle Point”) = (F-20020, SRT,
2D/SVC Diam Exp Diameter") "Expiration")
271
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
MM/R-R interval (122182, DCM, "R-R interval") (G-0373, SRT, “Image Mode”) = (G-0394, SRT, "M mode")
272
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
TID 5000
OB-GYN
Ultrasound
Procedure
Report
TID 5008
TID 5006 TID 5013
Fetal Biometry
Long Bones Group Follicles Section
Section
273
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
274
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
275
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
276
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
2 Sigma Upper Value of population (GEU-1009- 16,99GEMS, "2 Sigma Upper Value of population")
2 Sigma Lower Value of population (GEU-1009-17,99GEMS, "2 Sigma Lower Value of population")
2 Sigma Upper Value of population (GEU-1009-19,99GEMS, "2 Sigma Upper Value of population")
2 Sigma Lower Value of population (GEU-1009-20,99GEMS, "2 Sigma Lower Value of population")
2 Sigma Upper Value of population (GEU-1009-22,99GEMS, "2 Sigma Upper Value of population")
2 Sigma Lower Value of population (GEU-1009-23,99GEMS, "2 Sigma Lower Value of population")
2 Sigma Upper Value of population (GEU-1009-25,99GEMS, "2 Sigma Upper Value of population")
2 Sigma Lower Value of population (GEU-1009-26,99GEMS, "2 Sigma Lower Value of population")
2 Sigma Upper Value of population (GEU-1009-28,99GEMS, "2 Sigma Upper Value of population")
2 Sigma Lower Value of population (GEU-1009-29,99GEMS, "2 Sigma Lower Value of population")
277
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
1 Sigma Upper Value of population (GEU-1009-32,99GEMS, "1 Sigma Upper Value of population")
1 Sigma Lower Value of population (GEU-1009-33,99GEMS, "1 Sigma Lower Value of population")
1 Sigma Upper Value of population (GEU-1009-34,99GEMS, "1 Sigma Upper Value of population")
1 Sigma Lower Value of population (GEU-1009-35,99GEMS, "1 Sigma Lower Value of population")
1 Sigma Upper Value of population (GEU-1009-36,99GEMS, "1 Sigma Upper Value of population")
1 Sigma Lower Value of population (GEU-1009-37,99GEMS, "1 Sigma Lower Value of population")
1 Sigma Upper Value of population (GEU-1009-38,99GEMS, "1 Sigma Upper Value of population")
1 Sigma Lower Value of population (GEU-1009-39,99GEMS, "1 Sigma Lower Value of population")
1 Sigma Upper Value of population (GEU-1009-43,99GEMS, "1 Sigma Upper Value of population")
1 Sigma Lower Value of population (GEU-1009-44,99GEMS, "1 Sigma Lower Value of population")
1 Sigma Upper Value of population (GEU-1009-46,99GEMS, "1 Sigma Upper Value of population")
1 Sigma Lower Value of population (GEU-1009-47,99GEMS, "1 Sigma Lower Value of population")
1 Sigma Upper Value of population (GEU-1009-49,99GEMS, "1 Sigma Upper Value of population")
1 Sigma Lower Value of population (GEU-1009-50,99GEMS, "1 Sigma Lower Value of population")
1 Sigma Upper Value of population (GEU-1009-53,99GEMS, "1 Sigma Upper Value of population")
1 Sigma Lower Value of population (GEU-1009-54,99GEMS, "1 Sigma Lower Value of population")
278
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
OB/Hadlock/EFW(AC,BPD,FL,HC) (11732-5, LN, "EFW by AC, BPD, FL, HC, Hadlock 1985")
279
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Biometry
Measurements}
280
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
281
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Note: If configured,
the Gestational Age
will be sent in
WeekDay format.
4 >> INFERRED CODE DCID (228) Equation or 1 U DCID (12013)
FROM Table
Gestation Age
Equation and Tables
5 >> R-INFERRED NUM 1-n U
FROM
6 >> HAS NUM DCID (226) Population 1-n U
PROPERTIES Statistical Descriptors
7 > CONTAINS NUM DTID (12017) Growth 1 U
Distribution Rank
8 >> INFERRED CODE DCID (228) Equation or 1 U $Equation = DCID
FROM Table (12015) Fetal
Growth Equations
and Tables
282
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
283
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
284
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
285
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
286
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
287
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
288
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
289
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
290
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
OB/Eriksen/TAD-Min
OB/Jeanty/CRL-Min
OB/Jeanty/AC-Min
OB/Hadlock/CRL-Min
OB/Hansmann/TAD-Min
OB/Hansmann/AC-Min
OB/Hellman/GS-Min
OB/Hill/TCD-Min
OB/Kurtz/BPD-Min
OB/Robinson/CRL-Min
OB/Osaka/CRL-Min
(R-00347, SRT, ‘1 Sigma Lower Value of population’)
OB/Osaka/BPD-Min
OB/Osaka/FL-Min
OB/Osaka/HL-Min
OB/Tokyo/GS-Min
OB/Tokyo/CRL-Min
OB/Tokyo/BPD-Min
OB/Tokyo/FL-Min
OB/Shinozuka/BPD-Min
OB/Shinozuka/AC-Min
OB/Shinozuka/FL-Min
OB/Shinozuka/CRL-Min
OB/ASUM2000/BPD-Min
(R-00388, SRT, ‘1 Sigma Lower Value of population’)
OB/ASUM2000/AC-Min
OB/Eriksen/TAD-Max
OB/Jeanty/CRL-Max
(R-00346, SRT, ‘1 Sigma Upper Value of population’)
OB/Jeanty/AC-Max
OB/Hadlock/CRL-Max
291
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
OB/Hansmann/TAD-Max
OB/Hansmann/AC-Max
OB/Hellman/GS-Max
OB/Hill/TCD-Max
OB/Kurtz/BPD-Max
OB/Robinson/CRL-Max
OB/Osaka/CRL-Max
OB/Osaka/BPD-Max
OB/Osaka/FL-Max
OB/Osaka/HL-Max
OB/Tokyo/GS-Max
OB/Tokyo/CRL-Max
OB/Tokyo/BPD-Max
OB/Tokyo/FL-Max
OB/Shinozuka/BPD-Max
OB/Shinozuka/AC-Max
OB/Shinozuka/FL-Max
OB/Shinozuka/CRL-Max
OB/ASUM2000/BPD-Max
(R-00387, SRT, ‘1 Sigma Upper Value of population’)
OB/ASUM2000/AC-Max
OB/ASUM/BPD-Max
OB/ASUM/AC-Max
OB/Jeanty1984/FL-Max
OB/Hadlock/BPD-Max
OB/Hadlock/AC-Max
OB/Hadlock/HC-Max
OB/Hadlock/FL-Max
OB/Hansmann/BPD-Max
OB/Hansmann/CRL-Max (R-00387, SRT, "2 Sigma Upper Value of population")
OB/Hansmann/OFD-Max
OB/Hansmann/HC-Max
OB/Hansmann/ThD-Max
OB/Hansmann/FL-Max
OB/Rempen/GS-Max
OB/Rempen/CRL-Max
OB/Rempen/BPD-Max
OB/Mercer/Ft-Max
292
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
OB/ASUM/BPD-Min
OB/ASUM/AC-Min
OB/Jeanty1984/FL-Min
OB/Hadlock/BPD-Min
OB/Hadlock/AC-Min
OB/Hadlock/HC-Min
OB/Hadlock/FL-Min
OB/Hansmann/BPD-Min
OB/Hansmann/CRL-Min (R-00388, SRT, "2 Sigma Lower Value of population")
OB/Hansmann/OFD-Min
OB/Hansmann/HC-Min
OB/Hansmann/ThD-Min
OB/Hansmann/FL-Min
OB/Rempen/GS-Min
OB/Rempen/CRL-Min
OB/Rempen/BPD-Min
OB/Mercer/Ft-Min
OB/Jeanty/BPD-Min
OB/Jeanty/HC-Min
OB/Jeanty/FL-Min
OB/Jeanty/ULNA-Min
OB/Jeanty1984/ULNA-Min
OB/Jeanty/HL-Min
(R-00397, SRT, ‘5th Percentile Value of population’)
OB/Jeanty/TIB-Min
OB/Merz/BPD-Min
OB/Merz/AC-Min
OB/Merz/HC-Min
OB/Merz/FL-Min
OB/Rempen2004/CRL-Min
OB/Jeanty/BPD-Max
OB/Jeanty/HC-Max
OB/Jeanty/FL-Max
OB/Jeanty/ULNA-Max
(R-00337, SRT, ‘95th Percentile Value of population’)
OB/Jeanty1984/ULNA-Max
OB/Jeanty/HL-Max
OB/Jeanty/TIB-Max
OB/Merz/BPD-Max
293
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
OB/Merz/AC-Max
OB/Merz/HC-Max
OB/Merz/FL-Max
OB/Rempen2004/CRL-Max
294
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
295
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
296
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
297
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
298
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
299
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
300
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
301
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
302
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
$Measurement =
EV (GEU-1011-08,
99GEMS, “Auto dv”)
$Measurement =
EV (GEU-1011-09,
99GEMS, “Auto dx”)
$Measurement =
EV (GEU-1011-10,
99GEMS, “Auto dy”)
303
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
$Measurement =
EV (GEU-1011-11,
99GEMS, “Auto dz”)
$Measurement =
EV (GEU-1011-12,
99GEMS, “Auto
meand”)
$Measurement =
EV (GEU-1011-13,
99GEMS, “Auto
volume”)
304
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
TID 4200
Breast
Imaging
Report TID 4202
Report
Narrative
TID 1204
TID G301
Language of
Content Item User Defined
and Measurements
Descendants
TID 4208
TID 4201
Supplementar TID 4203
y Data Breast
Breast
Imaging
Imaging
Procedure
Assessment
TID 4206
TID 1400
Breast
Linear
Imaging
Measurement
Finding
Section
TID 4101
Area
Measurement
305
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
1 CONTAINER EV (111412, 1 M
DCM, “Narrative
Summary”
2 > CONTAINS CONTAINER (121058, DCM, 1-n M
“Procedure
reported”)
306
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
- (GEU-1006-01,
99GEMS,
“Lesion”)
307
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
C (G-A106, SRT,
Posterior)
308
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
13 >> HAS CODE EV (111037, DCM, “Margins”) 1-n U GE Value DICOM Map
PROPERTIES "Circumscribed" (F-01741, SRT,
"Circumscribed
lesion")
309
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
“Vascular (GEU-1006-24,
abnormalities 99GEMS, “Foreign
AVMs” body”)
310
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
“Internal (GEU-1006-18,
vascularity” 99GEMS,
“Internal
vascularity”)
“Soft” (GEU-1006-20,
99GEMS, “Soft”)
“Hard” (GEU-1006-21,
99GEMS, “Hard”)
311
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
SP/2D/DiamRatioD1 (GEU-1006-08,
99GEMS, “A/B
Ratio(Diam) D1”)
SP/2D/DiamRatioD2 (GEU-1006-09,
99GEMS, “A/B
Ratio(Diam) D2”)
312
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
313
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
TID G301
User Defined
Measurements
314
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
315
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
"Symmetric" (GEU-1007-05,
99GEMS,”Symmetric”)
316
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
317
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
NL Rel with Parent VT Concept Name VM Req Conditio Value Set Constraint
Type n
1 CONTAI EV (T-C4000, SRT, "Lymph 1 M
NER Node")
2 > HAS CONCEPT CODE EV (G-C171, SRT, 1 M $SectionLaterality
MOD “Laterality”)
3 > CONTAINS CODE EV (121071, DCM, “Finding”) 1-n M GE Value DICOM Map
for
Compositio
n
Solid ( 111462, DCM, “Solid
mass”)
Cystic (GEU-1007-56, 99GEMS,
“Cystic”)
Complex (GEU-1007-61, 99GEMS,
“Complex”)
318
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
NL Rel with Parent VT Concept Name VM Req Conditio Value Set Constraint
Type n
4 >> HAS CONCEPT CODE EV(GEU-1007-48, 1 U GE Value DICOM Map
MOD 99GEMS,”Location”)
"Submental (GEU-1007-38, 99GEMS,
" ”Submental”)
"Middle (GEU-1007-42,
cervical" 99GEMS,”Middle cervical”)
319
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
NL Rel with Parent VT Concept Name VM Req Conditio Value Set Constraint
Type n
8 >> HAS PROPERTIES INCLUD DTID (1400) Linear 1-n U
E Measurement
9 >> HAS PROPERTIES INCLUD DTID (1401) Area 1-n U
E Measurement
10 >> HAS PROPERTIES NUM EV(GEU-1007-13, 99GEMS, 1 U UNITS = DCID(7460) “Units of Linear Measurement”
“Cortical Thickness”)
NL Rel with Parent VT Concept Name VM Req Conditio Value Set Constraint
Type n
1 CONTAI EV (T-B7000, SRT, 1 M
NER “Parathyroid”)
2 > HAS CONCEPT CODE EV (G-C171, SRT, 1 M $SectionLaterality
MOD “Laterality”)
3 > CONTAINS CODE EV (121071, DCM, “Finding”) 1-n M EV (T-B7000, SRT, “Parathyroid”)
4 >> HAS PROPERTIES CODE EV(GEU-1007-51, 1 U GE Value DICOM Map
99GEMS,”Visibility”)
"Visualized" (GEU-1007-10, 99GEMS,”
Visualized”)
320
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
321
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
NL Rel with Parent VT Concept Name VM Req Conditio Value Set Constraint
Type n
“Lateral” (GEU-1007-33,
99GEMS,
"Lateral")
“Midline” (GEU-1007-35,
99GEMS,
"Midline")
“Medial” (GEU-1007-34,
99GEMS,
"Medial")
6 >> HAS CODE EV (GEU-1007-36, 99GEMS, 1 U
PROPERTIES “Composition”) GE Value DICOM Map
for
Composition
Solid (111462, DCM, “Solid
mass”)
Cystic (GEU-1007-56,
99GEMS, “Cystic”)
Mixed cystic (GEU-1007-20,
and Solid 99GEMS, “Mixed cystic
and solid”)
Spongiform (GEU-1007-81,
99GEMS,
“Spongiform”)
322
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
NL Rel with Parent VT Concept Name VM Req Conditio Value Set Constraint
Type n
9 >> HAS CODE EV (111037, DCM, 1 U GE Value DICOM Code
PROPERTIES “Margins”) Smooth (G-A545, SRT,
“Smooth”)
Ill-defined (GEU-1007-86,
99GEMS, “Ill-
defined”)
Lobulated (112135, DCM,
“Lobulated”)
Irregular (G-A402, SRT,
“Irregular”)
Extra-thyroidal (GEU-1007-87,
extension 99GEMS,
“Extra-thyroidal
extension”)
Lobulated or Irregular (GEU-1007-96,
99GEMS,
“Lobulated or
Irregular”)
323
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
NL Rel with Parent VT Concept Name VM Req Conditio Value Set Constraint
Type n
“Coarse EV (GEU-1007-15,
central” 99GEMS, “Coarse
central”)
“Coarse rim” EV (GEU-1007-16,
99GEMS, “Coarse
rim”)
"Punctate EV (GEU-1007-79,
scattered" 99GEMS, “Punctate
scattered”)
“Punctate EV (GEU-1007-80,
clumped” 99GEMS, “Punctate
clumped”)
“Colloid” EV (GEU-1007-19,
99GEMS, “Colloid”)
“Mixed” EV (GEU-1007-20,
99GEMS, “Mixed”)
324
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
NL Rel with Parent VT Concept Name VM Req Conditio Value Set Constraint
Type n
20 >> HAS PROPERTIES TEXT (GEU-1006-27, 99GEMS, 1 U <unique identifer>
“Finding ID”)
21 >> HAS TEXT EV (121106, DCM, 1 U
PROPERTIES “Comments”)
22 >> HAS INCLUDE DTID (1400) Linear 1-n U
PROPERTIES Measurement
23 >> HAS INCLUDE DTID (1402) Volume 1 U
PROPERTIES Measurement
325
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
The difference between this from TID 5103, TID 300 or 1414 for general measurement is the template uses labels
that specified by users at configuration time rather assigning specific codes to the individual measurements.
Note: In practice, to assign specific codes to user defined measurements are impractical and cumber-some since the
system/organization has to maintain database/map of codes and labels with public/private code designator
schemes.
326
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
4 > HAS CONTAINER EV (GEU-1005-6, 99GEMS, 1 MC IFF there exists Note: User Vessel Anatomy is mainly
CONCEPT ‘User Vessel Anatomy’) or user defined used to define vessels and for
MODE (GEU-1005-3, 99GEMS, measurements inserting the Study/Vessel template.
‘General Anatomy’or General Anatomy is mainly used in
(125007, DCM, PWD and B-Mode and when the user
‘Measurement Group’) or has added individual folders and
(125005, DCM, ‘Biometry measurements.
Group’) or (99100, DCM,
‘Doppler Group’)
5 >> HAS TEXT DT (GEU-1005-7, 99GEMS, 1 MC IFF the container Labels from user which defined at
CONCEPT ‘Anatomy Label’) or DT uses Finding as configuration
MOD (GEU-1002-58, 99GEMS, Concept Name
‘Biometry Group Name’)
or DT (GEU-1002-60,
99GEMS, ‘Doppler Group
Name’)
6 >> CONTAINS INCLUDE DTID (300) Measurement 1-n M IFF there exists $Measurement = When ‘Add Folder’
– see section: B.9 user defined insert study measurements by
measurements prepopulated study –Please refer to
section B.1.6
327
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
328
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
NL Relation with Value Type Concept Name VM Req Condition Value Set Constraint
Parent Type
>> CONTAINS TEXT (43144, 99GEMS, 1 M
“ultrasound image
mode”)
>> CONTAINS TEXT (43131, 99GEMS, 1 M
“ultrasound parameter
ID”)
>> CONTAINS TEXT (43133, 99GEMS, 1 M
“ultrasound parameter
preferred display
name”)
>> CONTAINS TEXT (43134, 99GEMS, 1 M
“ultrasound parameter
value source”)
>> CONTAINS TEXT (43142, 99GEMS, 1 M
“Result Number”)
>> CONTAINS TEXT (43145, 99GEMS, 1 M
“ultrasound study
type”)
>> CONTAINS TEXT (43129, 99GEMS, 1 M
“ultrasound parameter
textual value”)
>> CONTAINS NUM (43128, 99GEMS, 1 M
“ultrasound parameter
numeric value”)
TID 300
Measurement
329
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
330
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
C. Security Details
C.1. External Network Requirements details
C.1.1. Basic Time Synchronization
LOGIQ scanner follows the Consistent Time profile specifies the use of the Network Time Protocol (NTP)
defined in RFC 1305.
The following tables specify the DICOM Specific Audit Messages that LOGIQ Scanner can detect and
report. It defines the list of triggers that will cause audit message to be generated, if these triggers can
be configured or not. It also specifies if the content of the Audit message can be configured or not.
TABLE C.2-1 Supported Audit Trail Messages
331
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
332
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
Security Alert Used Any event for which a node NO E.g., A node
needs to report a security authentication
alert failure when
establishing a
secure
communication
s channel
User Authentication Used A user has attempted to log NO Note : The
on or log off. This report is user usually
made regardless of whether has
the attempt was successful UserIsRequesto
or not. r TRUE, but in
the case of a
logout timer,
the Node can
be the
UserIsRequesto
r.
Order Record Not Used NA
Patient Record Not Used NA
Procedure Record Not Used NA
The following table specifies the implementation detail of each audit message supported by this product.
Application Activity
Event EventID M EV (110100, DCM, "Application Activity")
EventActionCode M Enumerated Value E = Execute
System Date/Time, e.g. 2019-04-
EventDateTime M
18T20:15:20.1540000-06:00
”0” Nominal Success
EventOutcomeIndicator M
”8” Serious failure
Active UserID M Empty
Participant:Applicati
AlternativeUserID MC LOGIQE10
on started (1)
UserIsRequestor M true
RoleIDCode M EV (110150, DCM, "Application")
NetworkAccessPointTypeCode U 2
NetworkAccessPointID U Local IP Address
Audit Source AuditEnterpriseSiteID U Empty
AuditSourceID M LOGIQE10
333
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
334
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
335
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
336
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
337
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
NetworkAccessPointTypeCode U 2
NetworkAccessPointID U Local IP Address.
Audit Source AuditEnterpriseSiteID U Empty
AuditSourceID M LOGIQE10
EV (1, DCM, "End-user display device, diagnostic
AuditSourceTypeCode U
device)
Participating Object: Enumerated Value:
ParticipantObjectTypeCode M
2 = system
Studies being Enumerated Value:
ParticipantObjectTypeCodeRole M
transferred (1..N) 3 = report
ParticipantObjectIDTypeCode M EV (110180, DCM, "Study Instance UID")
ParticipantObjectID M The Study Instance UID
Participating Object: Enumerated Value:
ParticipantObjectTypeCode M
Patient (1) 1 = person
Enumerated Value:
ParticipantObjectTypeCodeRole M
1 = person
ParticipantObjectIDTypeCode M EV (2, RFC-3881, "Patient Number")
ParticipantObjectID M The patient ID
DICOM Study Deleted
Event EventID M EV (110105, DCM, "DICOM Study Deleted")
Enumerated Value
EventActionCode M
D = Delete
System Date/Time, e.g. 2019-04-
EventDateTime M
18T22:05:58.7370000-06:00
”0” Nominal Success
EventOutcomeIndicator M
”8” Serious failure
Active UserID M 2
Participant:the
UserIsRequestor M true
person or process
deleting the study NetworkAccessPointTypeCode U 2
(1..2)
NetworkAccessPointID U Local IP Address
Audit Source AuditEnterpriseSiteID U Empty
AuditSourceID M LOGIQE10
EV (1, DCM, "End-user display device, diagnostic
AuditSourceTypeCode U
device)
Enumerated Value:
ParticipantObjectTypeCode M
2 = system
338
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
339
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
340
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
341
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
The following table describes the Secure Transport Connection configuration parameters supported by
this product:
342
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
D. Mapping of Attributes
The table below describes the mapping of attributes between Modality Worklist, Instances and MPPS
messages.
In the Scenarios column the following values are supported:
• SCHEDULED: the image acquisition was scheduled at the RIS and procedure details have been
communicated in the MWL query)
• UNSCHEDULED: the image acquisition was performed without Modality Worklist information
• APPEND: instances acquired are added to an existing study after the initial procedure was finalized
In the Value Source columns, the following values are supported:
• GENERATED: the value is generated by the system.
• SRC_INSTANCE: the value is copied from previously created instances.
• MWL: the value is copied from modality worklist.
• USER: the value is entered by the user.
• SCANNED: the value is read from a barcode scanner or similar device.
• EMPTY: the attribute is sent without value.
The Destination column uses either ROOT, if the attribute is added to the root of the instance or list the
attribute Tag of the Sequence, the attribute will be added to. The comment column can be used to provide
additional information regarding the values added to the IOD.
343
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
E. Index of Tables
344
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
345
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
346
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
347
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
348
LOGIQ E10/E11/E20 R3.x
LOGIQ E10s R3.x
LOGIQ Fortis R3.x
GE HEALTHCARE DICOM CONFORMANCE STATEMENT
DIR DOC2107162 Rev 6
349