0% found this document useful (0 votes)
37 views9 pages

Unit-3 Instrumental Analysis

The document outlines calibration procedures for various laboratory instruments including UV-Vis spectrophotometers, electronic balances, flame photometers, fluorimeters, gas chromatographs, and HPLC systems. Each section details specific calibration steps, acceptance criteria, and the importance of regular calibration for ensuring accurate and reliable measurements. Emphasis is placed on the necessity of using known standards, maintaining records, and adhering to manufacturer guidelines for optimal instrument performance.

Uploaded by

unmeshdalui407
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
37 views9 pages

Unit-3 Instrumental Analysis

The document outlines calibration procedures for various laboratory instruments including UV-Vis spectrophotometers, electronic balances, flame photometers, fluorimeters, gas chromatographs, and HPLC systems. Each section details specific calibration steps, acceptance criteria, and the importance of regular calibration for ensuring accurate and reliable measurements. Emphasis is placed on the necessity of using known standards, maintaining records, and adhering to manufacturer guidelines for optimal instrument performance.

Uploaded by

unmeshdalui407
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 9

Calibrating a UV-Vis spectrophotometer involves verifying its performance

against known standards to ensure accurate measurements. This includes


checking wavelength accuracy using holmium filters, absorbance accuracy
using solutions like potassium dichromate, and stray light levels.

Calibration Procedures:
1. 1. Wavelength Accuracy:
• Use a holmium oxide filter (e.g., holmium perchlorate solution) to verify
wavelength accuracy.
• Scan the filter in the UV-Vis range and compare the observed peak
positions to the known values.
• Acceptance criteria typically involve checking peak positions within a
defined tolerance.
2. 2. Absorbance Accuracy:
• Use solutions with well-defined absorbances at specific wavelengths,
such as potassium dichromate.
• Measure the absorbance of the standard solution at the specified
wavelengths and compare it to the known values.
• Ensure linearity of absorbance over the required range by measuring at
multiple wavelengths and plotting a calibration curve.
3. 3. Stray Light:
• Use a potassium chloride solution or a holmium oxide filter to check for
stray light.
• Measure the absorbance at wavelengths near the cutoff of the filter and
compare it to the expected value.
• Acceptance criteria typically specify a maximum allowable stray light
level.
4. 4. Resolution:
• Use a toluene in hexane solution to assess the resolution of the
instrument.
• Scan the solution and compare the spectral resolution to the acceptance
criteria.
5. 5. Other Important Aspects:
• Baseline correction: Perform baseline correction using a blank solution
(e.g., 0.005 M sulfuric acid) to ensure accurate absorbance readings.
• Photometric repeatability: Measure absorbance values at the same
wavelength multiple times to assess the instrument's ability to return
consistent results.
• Regular maintenance: Follow the manufacturer's recommendations for
instrument maintenance and calibration frequency.
• Data logging: Maintain a record of calibration results and any deviations
from the expected values.

Acceptance Criteria:
• Wavelength accuracy: Peak positions should fall within the specified tolerance
for each wavelength.
• Absorbance accuracy: Absorbance values should fall within the specified
range for the chosen standard solutions.
• Stray light: Stray light should be below the specified limit.
• Resolution: The spectral resolution should meet the required criteria.
Calibrating an electronic balance ensures its accuracy. The process
generally involves using known standard weights to verify its readings and
make adjustments if needed.

Here's a general procedure for calibrating an electronic balance:


1. 1. Preparation:
• Place the balance on a stable, level surface, free from vibrations or air
currents.
• Ensure the balance is clean and free from dust.
• Switch the balance on and allow it to warm up and stabilize.
2. 2. Zeroing:
• After stabilization, check the zero reading.
• If the zero reading fluctuates, use the balance's "zero" or "tare" button to
reset it to zero.
3. 3. Calibration with Standard Weights:
• Place standard weights on the balance, starting with a low weight and
gradually increasing to the balance's maximum capacity.
• Record the displayed readings for each standard weight.
• Compare the displayed readings with the known weights of the standard
weights.
4. 4. Adjustments:
• If the displayed readings are not within an acceptable range (e.g., ±0.5%
of the standard weight), you may need to adjust the calibration using the
balance's calibration buttons or settings.
• Consult the balance's manual for specific instructions on how to adjust
the calibration.
5. 5. Verification:
• After adjusting the calibration, verify the balance's accuracy using the
same standard weights or by weighing known objects.
• Record the calibration date and details in a log book.
6. 6. Repeatability Check:
• For more accurate balances, you might also check for repeatability by
placing and removing the same weight multiple times and analyzing the
variations.

Important Considerations:
• Acceptance Criteria:
The acceptable difference between the displayed weight and the standard
weight should be within the specified tolerance (e.g., ±0.5%).
• Maintenance:
Regular calibration is essential to maintain the balance's accuracy.
• Proper Technique:
Use standard weights that are appropriate for the balance's capacity and follow
the manufacturer's instructions carefully.
• Environmental Factors:
Vibrations, air currents, and temperature fluctuations can affect the balance's
readings.
By following these steps, you can calibrate your electronic balance and
ensure its accuracy for reliable weighing.

Calibrating a flame photometer involves preparing standard solutions,


using a blank (distilled water), measuring the intensity of emitted light from
the standards, and creating a calibration curve to determine unknown
concentrations. This ensures accurate and reliable results by
compensating for instrumental variations and environmental factors.

Here's a more detailed breakdown of the calibration process:


1. 1. Prepare Standard Solutions:
Prepare solutions of known concentrations of the element you're measuring
(e.g., sodium, potassium). These standards are used to create a calibration
curve.
2. 2. Set the Instrument to Zero:
Use a blank (distilled water or a solvent with no analyte) to set the instrument to
zero.
3. 3. Measure Standard Solutions:
Aspirate the standard solutions into the flame photometer and measure the
intensity of the emitted light.
4. 4. Create a Calibration Curve:
Plot the intensity of light against the known concentrations of the
standards. This curve will be used to determine the concentration of unknown
samples.
5. 5. Check Calibration Regularly:
Regularly check the calibration with control samples to ensure the accuracy of
the instrument.

Importance of Calibration:
• Accuracy and Reliability:
Calibration ensures that the instrument is properly adjusted and that
environmental variables are accounted for, leading to more accurate and reliable
results.
• Data Validation:
Proper calibration facilitates data validation, quality control, and compliance
with industry standards.
• Optimal Performance:
Calibration ensures that the flame photometer performs optimally, allowing for
the detection of low concentrations and mitigating matrix effects.
Calibrating a fluorimeter involves adjusting its settings to ensure accurate
readings. This typically includes setting the excitation and emission
wavelengths, and using standards to adjust the concentration
readout. Calibration ensures the fluorimeter produces reliable results for
measuring fluorescence.

Here's a more detailed breakdown:


1. Setting Wavelengths:
• Excitation Wavelength: This is the wavelength of light that excites the analyte,
causing it to emit fluorescence.
• Emission Wavelength: This is the wavelength of light that the analyte emits
after being excited.
• Calibration may involve adjusting these wavelengths to ensure optimal
excitation and emission for the specific analyte being measured .

2. Using Standards:
• Calibration solutions with known concentrations of the analyte are used to
adjust the fluorimeter's readout.
• The fluorimeter's sensitivity is calibrated by comparing the measured
fluorescence intensity to the known concentration of the standard.
• This process allows the fluorimeter to accurately report the concentration of
the analyte in unknown samples.

3. Calibration Procedures:
• The specific calibration procedure will vary depending on the fluorimeter
model and the analyte being measured.
• General steps may include preparing calibration solutions, running the
calibration protocol on the fluorimeter, and recording the results.
• Calibration results are often used to create a calibration curve, which is a
graph of fluorescence intensity versus concentration.

4. Importance of Calibration:
• Accurate calibration is crucial for obtaining reliable and reproducible results
when measuring fluorescence.
• Proper calibration helps to ensure that the fluorimeter provides accurate and
precise measurements of analyte concentration.
• Regular calibration is recommended to maintain the fluorimeter's
performance and accuracy over time.
Calibration in gas chromatography (GC) is the process of ensuring the
instrument provides accurate measurements by comparing it against
standards. This involves calibrating various parameters like column oven
temperature, gas flow rates, and detector response. Calibration is crucial
for accurate quantification and identification of analytes, especially in fields
like environmental monitoring and pharmaceuticals.

Key Steps in GC Calibration:


1. 1. Column Oven Temperature Calibration:
Verify the accuracy of the temperature settings using a calibrated thermometer.
2. 2. Carrier Gas Flow Rate Calibration:
Ensure the carrier gas flow rate is consistent and accurate by comparing
setpoints to observed flows.
3. 3. Detector Calibration:
Verify the detector's response linearity and sensitivity by injecting known
concentrations of analytes.
4. 4. Calibration Curve Construction:
Plot the detector's response against known concentrations of analytes to
establish a relationship for quantification.
5. 5. Internal Standard Calibration:
Use an internal standard to correct for variations in sample injection volume and
instrument performance.
6. 6. Regular Checks and Maintenance:
Regularly check calibration standards and perform maintenance to ensure
instrument accuracy and precision.

Importance of Calibration:
• Accurate Quantification:
Calibration allows for precise determination of analyte concentrations in
samples.
• Reliable Data:
Proper calibration ensures the data generated by the GC is reliable and accurate,
leading to more trustworthy results.
• Method Validation:
Calibration is a crucial part of method validation, ensuring the GC method meets
specific performance criteria.
• Regulatory Compliance:
Calibration is often required to meet regulatory requirements in fields like
environmental monitoring and pharmaceuticals.
Calibration Methods:
• External Calibration:
Use a series of standard solutions with known concentrations of analytes to
create a calibration curve.
• Internal Standard Calibration:
Add a known amount of an internal standard to the sample and the standards,
which helps correct for variations in sample injection and instrument
performance.
In summary, calibrating a GC instrument involves verifying key parameters
like temperature, flow rates, and detector response using standards. This
ensures accurate and reliable quantification of analytes in various
applications..

HPLC (High-Performance Liquid Chromatography) calibration ensures


accurate and reproducible results. This involves checking various
parameters like flow rate, wavelength accuracy, detector linearity, and
injection volume. Calibration is performed using standard solutions and by
comparing the instrument's output to known standards.

Calibration Process:
1. 1. Flow Rate:
• Measure the flow rate by collecting a known volume of mobile phase at
different flow rate settings using a volumetric flask and stopwatch.
• Calculate the actual flow rate and compare it to the set flow rate.
• Use an acceptable error margin to determine if the pump is calibrated
correctly.
2. 2. Wavelength Accuracy:
• Use a reference standard with known absorption peaks at specific
wavelengths.
• Measure the wavelength of the absorption peaks and compare them to
the known wavelengths.
• Ensure the detector's wavelength accuracy is within acceptable limits.
3. 3. Detector Linearity:
• Prepare standard solutions of known concentrations and inject them into
the HPLC system.
• Measure the peak areas and plot them against the corresponding
concentrations.
• Assess the linearity of the detector response based on the correlation
coefficient (r²).
4. 4. Injection Volume:
• Inject a known volume of a standard solution and measure the peak area.
• Ensure the injection volume accuracy and precision are within acceptable
limits.
5. 5. Other Parameters:
• Column Oven: Check the temperature accuracy of the column oven using
a calibrated thermometer.
• Gradient Proportioning Valve: Calibrate the gradient proportioning valve if
your system uses it.
• Autosampler: Check the injection volume accuracy and precision of the
autosampler.
6. 6. Calibration Log:
• Record all calibration data in a calibration log, including the date, time,
results, and any remarks.
• Use the calibration log to track the instrument's performance and identify
any potential issues.

Importance of Calibration:
• Accurate Results:
Regular calibration ensures the HPLC system produces accurate and reliable
results.
• Reproducibility:
Calibration helps maintain consistency in the instrument's performance.
• Method Validation:
A calibrated HPLC system is essential for method validation, which establishes
the method's accuracy, precision, and linearity.
• Regulatory Compliance:
Calibration is often a requirement for regulatory compliance in industries like
pharmaceuticals.

Common Issues:
• Non-Linearity:
If the detector response is not linear, it can lead to inaccurate quantification of
analytes.
• Flow Rate Instability:
Fluctuations in flow rate can affect peak shape and retention time, impacting
method reproducibility.
• Wavelength Drift:
Drift in the detector's wavelength can lead to inaccurate peak detection and
quantification.

Maintenance:
• Regular Calibration:
Calibrate the HPLC system regularly, as per the manufacturer's
recommendations and your laboratory SOPs.
• Troubleshooting:
If you encounter any issues with the HPLC system, refer to the manufacturer's
documentation and consult with your laboratory's experts.

You might also like