DATA SHEET > UCS 500N5 > 20120518
UCS 500N5
MULTIFUNCTIONAL TESTGENERATOR FOR TRANSIENTS
(EFT/BURST, SURGE & POWER FAIL) UP TO 5.5KV
FOR TESTS ACCORDING TO ...
> EN 300329
> EN 300340
> EN 300342-1
> EN 300386 V1.3.2
> EN 301489-1
> EN 301489-17
> EN 301489-24
> EN 301489-7
> EN 50121
> EN 55024
> EN 61000-4-11
> EN 61000-4-29
> EN 61000-4-4
> EN 61000-4-5
> EN 61000-4-8
> EN 61000-4-9
> EN 61000-6-1
> EN 61000-6-2
> FCC 97-270 (part 68)
> IEC 60255-22-5
> IEC 61000-4-11
> ...
UCS 500N5 - COMPACT TESTER FOR EFT/BURST, SURGE AND POWER FAIL
The UCS 500N5 ultra-compact simulator is the most versatile tester to cover transient and power fail requirements according
to international standards (basic and generic standards) and product/product family standards. The UCS 500N5 is the most
economic solution for tests during development as well as for full-compliant immunity tests and CE Marking for single phase
DUT with the ability to be extended for testing three-phase DUTs by means of an automatically controlled external coupling
network up to 100A.
EM TEST supplies a large range of accessories for the various applications such as magnetic field tests.
HIGHLIGHTS APPLICATION AREAS
> Burst module (IEC/EN 61000-4-4) up to 5.5kV INDUSTRY COMPONENTS
> Surge module (IEC/EN 61000-4-5) up to 5kV
> TSurge module (IEC/EN 61000-4-5); optional MEDICAL BROADCAST
> PowerFail module (IEC/EN 61000-4-11)
RESIDENTIAL RENEWABLE ENERGY
> Magnetic field tests with optional accessoires
> Built-in single phase coupler 300V/16A TELECOM
> Manual operation from front panel
www.emtest.com © EM TEST > PAGE 1/8
DATA SHEET > UCS 500N5 > 20120518
TECHNICAL DETAILS
BENEFITS SOFTWARE
ALL IN ONE - ALL WHAT YOU NEED FOR YOUR TESTS IEC.CONTROL SOFTWARE FOR CONTROL AND
The UCS 500N5 includes everything necessary to conduct DOCUMENTATION
fully compliant tests. The power mains supply for the Outstanding user convenience, clearly structured
controls and for the DUT is separate to render it more windows and operation features and the EM TEST
flexible to use with different DUT supply voltages. standards library along with the flexibility to generate
The UCS 500N5 can be operated manually from the front user specific test sequences very easily are the main
panel or by remote via the built-in USB or GPIB interface. features of iec.control software. The software is
Fail inputs allow to control an ongoing test sequence automatically configured according to the connected EM
based on the status of the DUT. Monitoring outputs (BNC) TEST generators. Extensive reporting capabilities help the
are offered for easy signal measurement and verification. user to create test reports that meet international
Safety features such as interlock and warning lamp requirements.
control are available. iec.control is supported by Windows 2000, Windows XP,
Pre-programmed Standard Test routines allow highest Windows Vista and Windows 7. Remote control is
user convenience. Still the UCS 500N5 offers the Quick achieved either via USB or GPIB. iec.control supports a
Start test routine where parameters can be changed wide range of GPIB cards of National Instruments.
on-line during the test to evaluate the susceptibility level
of an individual DUT.
OTHER MODELS
OPERATION UCS 500N SERIES - COMPACT TESTERS UP TO 7KV
The UCS 500N ultra-compact testers for EFT/burst, Surge
EASY TO OPERATE and Power Fail are available in two different models; with
Front panel menu and function keys enable the user to voltage capability up to 5.5kV or up to 7kV.
program his test routines quickly and accurately. The
cursor allows fast control of all test parameters of the
programmed routine, thus test procedures are simplified AUXILIARY DEVICES
and confidence is generated that every step is carried out
CNI 503 - 3PHASE COUPLING/DECOUPLING NETWORKS
correctly.
FOR BURST AND SURGE
EM TEST offers a range of fully automatic 3-phase
coupling/decoupling networks for burst and surge to
extend the test capability for three-phase DUTs. The
networks have a rated current of up to 100A.
MV 2616 - MOTORISED VARIAC FOR VOLTAGE VARIATION
A motorised variac is offered as an alternative to the
tapped autotransformers for voltage dips/interruptions
and voltage variation tests as per IEC 61000-4-11. The
motorised variac can also be used for automated
magnetic field tests.
CNV 504N/508N - SURGE COUPLING/DECOUPLING
NETWORKS FOR SIGNAL/DATA LINES
CNV 504N/508N coupling/decoupling networks are
available to perform surge tests on I/O lines, signal/data
lines and telecom lines as per IEC 61000-4-5.
www.emtest.com © EM TEST > PAGE 2/8
DATA SHEET > UCS 500N5 > 20120518
TECHNICAL DETAILS
ACCESSORIES
V 4780 - TAPPED VOLTAGE TRANSFORMER FOR VOLTAGE
DIPS AND INTERRUPTIONS
The V 4780 tapped autotransformer is designed to supply
the required voltages as per IEC 61000-4-11 Ed.2:2004 to
perform voltage dips and interruptions.
V 4780S2 - TAPPED VOLTAGE TRANSFORMER FOR
VOLTAGE DIPS AND INTERRUPTIONS
The V 4780S2 tapped autotransformer is designed to
supply the required voltages as per IEC 61000-4-11
Ed.2:2004 to perform voltage dips and interruptions.
Compared to the manually operated V 4780 the V 4780S2
model offers automatic change of taps according to the
selected voltage level.
MS 100N - MAGNETIC FIELD COIL FOR POWER-FREQUENCY
AND PULSED MAGNETIC FIELDS
The MS 100N is a 1sqm magnetic field coil as specified in
IEC/EN 61000-4-8 and IEC/EN 61000-4-9. Its design
allows easy moving of the coil. The field coil is adjustable
in height and allows for 360degr rotation.
To generate power-frequency magnetic fields in the lower
range the current transformer MC 2630 is used while
high-field strength above 100A/m requires the MC 26100
current transformer.
HFK - CAPACITIVE COUPLING CLAMP
The HFK is a fully compliant capacitive coupling clamp as
per specification of IEC 61000-4-4.
ITP - IMMUNITY TEST PROBES
ITP is a tool being used for development test. It consists
of a variety of electrical field probes. The probes allow to
locate weak points within a system or on a PCB. The burst
pulse is used to generate the disturbance signal.
CA EFT KIT - VERIFICATION KIT FOR EFT/BURST PULSES
As per IEC 61000-4-4 Ed.2 the characteristic of the burst
generator needs to be verified with two different loads,
50ohm and 1,000ohm. EM TEST offers a calibration kit
consisting of the two loads and an adapter to verify the
pulses at the DUT output.
www.emtest.com © EM TEST > PAGE 3/8
DATA SHEET > UCS 500N5 > 20120518
TECHNICAL DETAILS
OPTIONS
ELECTRICAL FAST TRANSIENTS
HFK Capacitive coupling clamp as per
IEC 61000-4-4
BURST MODULE, EFT/N5 KW50 100:1 divider, 50ohm
As per EN/IEC 61000-4-4 and KW1000 500:1 divider, 1,000ohm
EN 61000-6-1, -6-2
CA EFT kit Kit for burst pulse verification
Test voltage 200V - 5,500V ± 10%; consisting of KW50, KW1000 and
100V - 2,750V ± 10% into 50ohm adapter for DUT port in a plastic case
Pulse shape 5/50ns into 50ohm and 1,000ohm for storage
Rise time tr 5ns ± 30% into 50ohm; A6dB 6dB attenuator, 50ohm
5ns ± 30% into 1,000ohm ITP Immunity test probes (electrical field
Pulse width td 50ns ± 30% into 50ohm; generation)
50ns -15/+100ns into 1,000ohm ITP/H Immunity test probe (magnetic field
Source impedance 50ohm generation)
Polarity Positive/negative
TRIGGER CIRCUIT
Trigger of bursts Automatic, manual, external
Synchronization 0° - 360°, resolution 1° (16 - 500Hz)
Burst duration td = 0.10ms - 999ms
Repetition rate tr = 10ms - 9,999ms
Spike frequency f = 0.1kHz - 1,000kHz
Test duration T = 0:01min - 99:59min
T > 99:59min --> endless
OUTPUTS
Direct Via 50ohm coaxial connector
Coupling mode L, N, PE; all combinations
DUT supply AC: 300V/16A; 50/60Hz
DC: 300V/10A
CRO trigger 5V trigger signal for oscilloscope
TEST ROUTINES
Quick Start On-line adjustable parameters,
easy-to-use
Standard Test As per IEC 61000-4-4, Levels 1 - 4
routines As per EN 61000-6-1, -6-2
Manual Standard Test routine
User Test routines Synchronous burst release
Random burst release
Change voltage after T
Frequency sweep within one burst
Frequency sweep with constant
number of pulses
Frequency sweep with constant
burst duration
Change polarity after T
www.emtest.com © EM TEST > PAGE 4/8
DATA SHEET > UCS 500N5 > 20120518
TECHNICAL DETAILS
TEST ROUTINES
COMBINATION WAVE / SURGE
Quick Start One-line adjustable parameters,
easy-to-use
SURGE MODULE, VCS/N5 Standard Test As per IEC 61000-4-5, Levels 1 - 4
As per EN/IEC 61000-4-5 and routines As per EN 61000-6-1, -6-2
EN 61000-6-1, -6-2 Manual Standard Test routine
Voltage (o.c.) 160V - 5,000V ± 10% User Test routines Change polarity after n pulses
Change coupling after n pulses
Pulse front time 1.2us ± 30% Change voltage after n pulses
Pulse time to half 50us ± 20% Change phase angle after n pulses
value Pulsed Magnetic as per IEC 61000-4-9
Current (s.c.) Max. 2,500A ± 10% Field Test levels 100, 300 and 1,000A/m
Test level steplessly adjustable
Pulse front time 8us ± 20% under Quick Start
Pulse time to half 20us ± 20%
value
OPTIONS
Polarity Positive/negative/alternating
CNV504Nx Coupling network for 4 signal/data
Event counter 1 - 30,000 or endless, selectable lines as per IEC 61000-4-5
CNV508Nx Coupling network for 8 signal/data
TRIGGER CIRCUIT lines as per IEC 61000-4-5
Release of pulses Automatic, manual, external IMN 2 Impedance matching adapter to
match direct output to 2ohm source
Synchronization 0° - 360°, resolution 1° impedance
Repetition rate max. 1Hz (1s - 999s)
OUTPUTS
Direct Via HV connectors for external
coupling networks
(Zi = 2ohm with optional adapter
IMN 2)
Coupling mode Line to line
Line(s) to ground
DUT supply AC: 300V/16A; 50/60Hz
DC: 300V/10A
CRO trigger 5V trigger signal for oscilloscope
MEASUREMENTS
CRO Û-monitor 10Vp at 5,000V
CRO Î-monitor 10Vp at 2,500A
Peak voltage 5,000V in the LCD display
Peak current 2,500A in the LCD display
www.emtest.com © EM TEST > PAGE 5/8
DATA SHEET > UCS 500N5 > 20120518
TECHNICAL DETAILS
TEST ROUTINES
POWER FAIL, DIPS & INTER-
Quick Start On-line adjustable parameters,
RUPTIONS, VOLTAGE VARIATIONS easy-to-use
Standard Test As per EN/IEC 61000-4-11
POWER FAIL MODULE, PFS/N5 routines for AC supplies
As per EN/IEC 61000-4-29 for DC
As per EN/IEC 61000-4-11, supplies
IEC/EN 61000-4-29 and As per EN 61000-6-1, -6-2
EN 61000-6-1, -6-2 Manual Standard Test routine
Channel PF1/PF2 AC voltage: max. 300V User Test routines Voltage variation, control of an
AC current: max. 16A external variac
DC voltage: max. 300V Change phase angle after n events
DC current: max. 10A Change event duration after n
Frequency 16Hz - 500Hz and DC events
Inverse mode
Switching time < 5us into a 100ohm resistive load
50/60Hz magnetic As per EN/IEC 61000-4-8
Inrush current > 500A field Test levels 1, 3, 10 and 30A/m with
Protection Both channels are protected against external current transformer
short-circuit conditions. MC2630
Test levels 100, 300 and 1,000A/m
with external current transformer
TRIGGER CIRCUIT MC26100
Trigger of events Automatic, manual, external
Synchronization 0° - 360°, resolution 1° (16 - 500Hz) OPTIONS
Repetition rate 10ms - 9,999s V4780 Tapped autotransformer as per IEC
61000-4-11 Ed.2
Event duration 20us - 9,999s
V4780 S2 Tapped autotransformer as per IEC
61000-4-11 Ed.2 with automatic
OUTPUTS change of tap
DUT terminals L, N and PE MV2616 Motorised variac (0 - 250V, 16A)
CRO trigger 5V trigger signal for oscilloscope MS100N Magnetic field coil, 1m x 1m
MC2630 Current transformer for magnetic
MEASUREMENTS fields up to 30A/m
DUT voltage In the LCD display MC26100 Current transformer for magnetic
fields up to 1,000A/m
DUT current In the LCD display
CA PFS Calibration box for inrush current
MON V Measurement of the DUT voltage; verification as per IEC 61000-4-11
built-in 100:1 divider
MON I Measurement of the DUT current;
10mV/A; max. 1,000A
www.emtest.com © EM TEST > PAGE 6/8
DATA SHEET > UCS 500N5 > 20120518
TECHNICAL DETAILS
TELECOM SURGE GENERAL DATA
TSURGE MODULE, TSURGE5 (OPTIONAL) INTERFACES
Test voltage 160V - 5,000V ± 10% Serial interface USB
Energy storage 20uF Parallel interface IEEE 488, addresses 1 - 30
capacitor
Analog output 0 - 10VDC to control an external
Polarity Positive, negative, alternating transformer
Counter 1 - 30,000 or endless, selctable CN interface 15pin SubD connector to control an
external coupling network
As per ITU and ETSI
recommendations Fail inputs DUT monitoring via Fail1 and Fail2
input (one each)
Front time 10us ± 30%
Pulse duration 700us ± 20%
DIMENSIONS
As per FCC part 68, Pulse B
Dimensions 19", 3HU,
Front time 9us ± 30% 19", 6HU, (with TSurge5 module)
Pulse duration 720us ± 20% Weight approx. 25kg
Output current 4 - 125A (short-circuit)
Rise time 5us ± 30% MAINS
Pulse duration 320us ± 20% Supply voltage 115V/230VAC +10%/-15%
As per IEC 61000-4-5 Power approx. 75W
Rise time 6.5us ± 30% Frequency 50/60Hz
Pulse duration 700us ± 20% Fuses 2 x T 2AT (230V) or 2 x T 4AT (115V)
Output current 4A - 125A (short-circuit)
Rise time 4us ± 20% SAFETY
Pulse duration 300us ± 20% Safety standard EN/IEC 61010
Security circuit Control input (24VDC)
TRIGGER CIRCUIT Warning lamp Floating contact (max. 230V/6A)
Trigger of events Automatic, manual, external
Repetition rate max. 0.5Hz (2s - 999s) ACCESSORIES INCLUDED
Mains supply Plug depends on the country of use
OUTPUTS DUT supply Plug depends on the country of use
As per ITU For 2-wire T1/T2 with 25ohm each DUT adapter Socket depends on the country of
use
As per FCC part 68 For 2-wire T1/T2 with 25ohm each
Operation manual,
As per IEC For 4-wire T1,T2,T3,T4 with 100ohm
Calibration certificate,
61000-4-5 each
iec.control remote control software
For other requirements special
output configurations are available
OPTIONS
CNI 503Ax 3-phase coupling/decoupling
OPTIONS
networks as per IEC 61000-4-4 and
CNV 504S1 Coupling network for 4 telecom lines -4-5 up to 100A per phase
as per IEC 61000-4-5
iec.control 1 Remote control and documentation
CNV 508S1 Coupling network for 8 telecom lines software, including standard test
as per IEC 61000-4-5 routines and reporting capabilities.
www.emtest.com © EM TEST > PAGE 7/8
DATA SHEET > UCS 500N5 > 20120518
COMPETENCE WHEREVER
YOU ARE
CONTACT EM TEST DIRECTLY
Switzerland USA / Canada
EM TEST (Switzerland) GmbH > Sternenhofstraße 15 > 4153 Reinach > EM TEST USA Inc. > 9250 Brown Deer Road > San Diego > CA 92121
Switzerland Phone +1 (858) 699 1685 > Fax +1 (858) 458 0267
Phone +41 (0)61/7179191 > Fax +41 (0)61/7179199 Internet: www.emtest.com > E-mail:
[email protected] Internet: www.emtest.ch > E-mail:
[email protected] P.R. China
Germany E & S Test Technology Limited > Rm 913, Leftbank >
EM TEST GmbH > Lünener Straße 211 > 59174 Kamen > Deutschland No. 68 Bei Si Huan Xi Lu > Haidian District > Beijing 100080 > P.R. China
Phone +49 (0)2307/26070-0 > Fax +49 (0)2307/17050 Phone +86 (0)10 82 67 60 27 > Fax +86 (0)10 82 67 62 38
Internet: www.emtest.com > E-mail:
[email protected] Internet: www.emtest.com > E-mail:
[email protected] France Malaysia
EM TEST FRANCE > Le Trident - Parc des Collines > Immeuble B1 - Etage 3 > EM TEST (M) SDN BHD > Unit B2-6, Jalan Dataran SD2 > Dataran SD2, PJU9 >
36, rue Paul Cézanne > 68200 Mulhouse > France Bandar Sri Damansara > 52200 Kuala Lumpur > Malaysia
Phone +33 (0)389 31 23 50 > Fax +33 (0)389 31 23 55 Phone +60 (03)62 73 22 01 > Fax +60 (03)62 74 22 01
Internet: www.emtest.fr > E-mail: [email protected] Internet: www.emtest.com > E-mail: [email protected]
Poland
EM TEST Polska > ul. Ogrodowa 31/35, 00-893 Warszawa > Polska
Phone +48 (0)518 64 35 12
Internet: www.emtest.com/pl > E-mail: [email protected]
Information about scope of delivery, visual design and technical data correspond with the state of development at time of release.\nTechnical
data subject to change without further notice.
www.emtest.com © EM TEST > PAGE 8/8