Datasheet CSG60

The document outlines the electrical specifications and testing capabilities of the GSC60 professional safety tester and network analyzer, including accuracy, voltage ranges, and test currents for various electrical safety tests. It details the parameters for insulation resistance, RCD testing, and earth leakage delay testing, along with their respective accuracies and resolutions. Additionally, it provides information on safety protections and operational limits for the device.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
16 views7 pages

Datasheet CSG60

The document outlines the electrical specifications and testing capabilities of the GSC60 professional safety tester and network analyzer, including accuracy, voltage ranges, and test currents for various electrical safety tests. It details the parameters for insulation resistance, RCD testing, and earth leakage delay testing, along with their respective accuracies and resolutions. Additionally, it provides information on safety protections and operational limits for the device.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 7

GSC60 Rel 2.

00 of 03/07/24

Professional safety tester and network analyzer Pag 1 of 7

1. ELECTRICAL SPECIFICATIONS – SAFETY SECTION


Accuracy is indicated as  ( readings + no. of digits*resolution) at 23°C  5°C, <80RH
Voltage (RCD, LOOP, Phase sequence)
Range [V] Resolution [V] Accuracy
15  460 1 (3.0 rdg + 2dgt)

Continuity test on protective and equalizing conductors with 200mA


Range [] Resolution [] Accuracy (*)
0.01  19.99 0.01
(5.0rdg + 3dgt)
20.0  99.9 0.1
(*) calibrate the cables to null their resistance
Test current: > 200mA DC for R5 (calibration included) ; Resolution for DC current :1mA
Open-circuit voltage: 4V  V0  12V
Safety protection: the display shows an error message for input voltage > approx. 10V

Insulation resistance (DC voltage)


Test voltage[V] Range [M] Resolution [M] Accuracy
0.01  9.99 0.01
(2.0rdg + 2dgt)
50 10.0  49.9 0.1
50.0  99.9 0.1 (5.0rdg + 2dgt)
0.01  9.99 0.01
(2.0rdg + 2dgt)
100 10.0  99.9 0.1
100.0  199.9 0.1 (5.0rdg + 2dgt)
0.01  9.99 0.01
(2.0rdg + 2dgt)
250 10.0  99.9 0.1
100  499 1 (5.0rdg + 2dgt)
0.01  9.99 0.01
10.0  199.9 0.1 (2.0rdg + 2dgt)
500
200  499 1
500  999 1 (5.0rdg + 2dgt)
0.01  9.99 0.01
10.0  199.9 0.1 (2.0rdg + 2dgt)
1000
200  999 1
1000  1999 1 (5.0rdg + 2dgt)
Open-circuit voltage: nominal test voltage –0% +10%
Short circuit current: <6.0mA at 500V test voltage
Nominal test current: >1mA if load= 1k*Vnom (Vnom=50V, 100V, 250V, 500V, 1000V)
Safety protection: the display shows an error message for input voltage > approx.10V

Z Line (L-L, L-N, L-PE)


Range [] Resolution [] Accuracy
0.00  199.9 m (*) 0.1 m (*)
(5.0 rdg + 1m) (*)
200  1999 m (*) 1 m (*)
0.01  9.99  0.01 
(5.0 rdg + 3dgt)
10.0  199.9  0.1 
(*) By means of IMP57 optional accessory
Maximum test current: 5.81A (at 265V); 10.10A (at 457V)
Test voltage ranges: 100265V (Line-Neutral) / 173460V (Line-Line); 50/60Hz  5%
Protection type: MCB (B, C, D, K), Fuse (gG, aM)
Insulation materials: PVC, Rubber butyl, EPR, XLPE

First fault current (IT systems)


Range (mA) Resolution (mA) Accuracy
0.1  0.9 0.1 (5.0 rdg + 1dgt)
1  999 1 (5.0 rdg + 3dgt)
Limit contact voltage (ULIM) : 25V, 50V
HT ITALIA SRL
Via della Boaria, 40
48018 – Faenza (RA) – Italy WHERE
T +39 0546 621002 | F +39 0546 621144 WE ARE
M [email protected] | ht-instruments.com
GSC60 Rel 2.00 of 03/07/24

Professional safety tester and network analyzer Pag 2 of 7

RCD test (Molded case type)


RCD type: AC ( ),A/F( ),B/B+( ) – General (G), Selective (S) and Delayed ( )
Rated tripping currents (IN):: 10mA, 30mA, 100mA, 300mA, 500mA, 650mA, 1000mA
Line-PE, Line-N voltage: 100V 265V RCD type AC and A/F, 190V 265V RCD type B/B+
Frequency: 50/60Hz  5%

RCD tripping current (Molded case type – RCD General)


RCD type IN Range IN [mA] Resolution [mA] Accuracy IN
IN = 10mA - 0%, +10%IN
AC, A/F
10mA <IN 650mA (0.3  1.1) IN  0.1 IN
- 0%, +5%IN
B/B+ 30mA IN 100mA

RCD Molded type tripping time range [ms] (TT/TN system)


x 1/2 x1 x2 x5 AUTO
\ G S G S G S G S G S G S
AC 999 999 999 999 999 999 200 250 50 150 ✓ ✓ 310
10mA A/F 999 999 999 999 999 999 200 250 50 150 ✓ ✓ 310
B/B+
AC 999 999 999 999 999 999 200 250 50 150 ✓ ✓ 310
30mA
A/F 999 999 999 999 999 999 200 250 50 150 ✓ ✓ 310
100mA
B/B+ 999 999 999 999 999 999 310
AC 999 999 999 999 999 999 200 250 50 150 ✓ ✓ 310
300mA A/F 999 999 999 999 999 999 200 250 50 150 ✓ ✓ 310
B/B+ 999 999 999 999 999 999
AC 999 999 999 999 999 999 200 250 50 150 ✓ ✓ 310
500mA
A/F 999 999 999 999 999 999 200 250 310
650mA
B/B+
AC 999 999 999 999 999 999 200 250
1000mA A/F 999 999 999 999 999 999
B/B+
Resolution: 1ms, Accuracy: (2.0rdg + 2dgt)

RCD Molded type tripping time range [ms] (IT system)

x 1/2 x1 x2 x5 AUTO
\ G S G S G S G S G S G S
AC 999 999 999 999 999 999 200 250 50 150 ✓ ✓ 310
10mA A/F
B/B+
30mA AC 999 999 999 999 999 999 200 250 50 150 ✓ ✓ 310
100mA A/F
300mA B/B+
AC 999 999 999 999 999 999 200 250 50 150 ✓ ✓ 310
500mA
A/F
650mA
B/B+
AC 999 999 999 999 999 999 200 250
1000mA A/F
B/B+
Resolution: 1ms, Accuracy: (2.0rdg + 2dgt)

HT ITALIA SRL
Via della Boaria, 40
48018 – Faenza (RA) – Italy WHERE
T +39 0546 621002 | F +39 0546 621144 WE ARE
M [email protected] | ht-instruments.com
GSC60 Rel 2.00 of 03/07/24

Professional safety tester and network analyzer Pag 3 of 7

Test on earth leakage delay tester RCDs (with RCDX10 optional accessory)
RCD type: AC ( ),A/F( ),B/B+( ) – General (G), Selective (S) and Delayed ( )
Rated tripping currents (IN):: 0.3A  10A
Line-PE, Line-N voltage: 100V 265V RCD type AC and A/F, 190V 265V RCD type B/B+
Frequency: 50/60Hz  5%
Earth leakage delay tester RCDs tripping current (RCD General)
RCD type IN Range IN [mA] Resolution [mA] Accuracy IN
AC, A/F 300mA IN 6.5A
(0.3  1.1) IN  0.1IN - 0%, +5%IN
B/B+ 300mA IN 1A

Earth leakage delay tester RCDs trip out time range [ms] (TT/TN system)
x 1/2 x1 x2 x5 AUTO
\ G S G S G S G S G S G S
0.3A AC 999 999 999 999 999 999 200 250 50 150 ✓ ✓ 310
 A/F 999 999 999 999 999 999 200 250 50 150 ✓ ✓ 310
1.0A B/B+ 999 999 999 999 999 999 310
1.1A AC 999 999 999 999 999 999 200 250 50 150 ✓ ✓ 310
 A/F 999 999 999 999 999 999 200 250 50 150 ✓ ✓ 310
3.0A B/B+ 999 999 999 999 999 999
3.1A AC 999 999 999 999 999 999 200 250 50 150 ✓ ✓ 310
 A/F 999 999 999 999 999 999 200 250 50 150 ✓ ✓ 310
6.5A B/B+ 999 999 999 999 999 999
AC 999 999 999 999 999 999 200 250
6.6A
 A/F 999 999 999 999 999 999
10.0A
B/B+
Resolution: 1ms, Accuracy: (2.0rdg + 2dgt)

Earth leakage delay tester RCDs trip out time range [ms] (IT system)
x 1/2 x1 x2 x5 AUTO
\ G S G S G S G S G S G S
0.3A AC 999 999 999 999 999 999 200 250 50 150 ✓ ✓ 310
 A/F
3.0A B/B+
3.1A AC 999 999 999 999 999 999 200 250 50 150 ✓ ✓ 310
 A/F
6.5A B/B+
6.6A AC 999 999 999 999 999 999 200 250
 A/F
10.0A B/B+
Resolution: 1ms, Accuracy: (2.0rdg + 2dgt)
RA – Non-trip earth loop impedance
Test voltage: 100265V (Line-PE), 50/60Hz  5%
RA – Systems with Neutral wire
Range [] Resolution [] Accuracy
0.01  9.99 0.01 -0%, +(5.0 rdg + 0.1)
10.0  199.9 0.1 -0%, +(5.0 rdg + 1)
200  1999 1 -0%, +(5.0 rdg + 3)
Test current: ~10mA
RA – Systems without Neutral wire
Range [] Resolution [] Accuracy
1  1999 1 -0%, +(5.0 rdg + 3dgt)
Test current: < ½ IN set
HT ITALIA SRL
Via della Boaria, 40
48018 – Faenza (RA) – Italy WHERE
T +39 0546 621002 | F +39 0546 621144 WE ARE
M [email protected] | ht-instruments.com
GSC60 Rel 2.00 of 03/07/24

Professional safety tester and network analyzer Pag 4 of 7

Contact voltage (RCD and Ra test)


Range [V] Resolution [V] Accuracy
0  Utlim 0.1 -0%, +(5.0 rdg + 3V)
Contact voltage (EARTH test – TT system)
Range [V] Resolution [V] Accuracy
0  99.9 0.1 -0%, +(5.0 rdg + 3V)
Contact voltage (EARTH test – TN system)
Range [V] Resolution [V] Accuracy
0  99.9 0.1
-0%, +(5.0 rdg + 3V)
100  999 1
Ground resistance with 3-wire method
Range [] Resolution [] Accuracy (*)
0.01  9.99 0.01
10.0  99.9 0.1
(5.0 rdg + 3dgt)
100  999 1
1.00k  49.99k 0.01k
Test current: <10mA – 77.5Hz, Open-circuit voltage: < 20Vrms
(*) Add 5% to the accuracy if the probe resistances (Rs or Rh) > 100 x Rmeas

Soil resistivity with 4-wire Wenner method


Range [m] Resolution [m] Accuracy (*)
0.06  9.99 0.01
10.0  99.9 0.1
100  999 1
1.00k  9.99k 0.01k (5.0 rdg + 3dgt)
10.0k  99.9k 0.1k
100k  999k (*) 1k
1.00M  3.14M (*) 0.01M
(*) with distance d=10m, Distance “d” range: 1  10m
Test current: <10mA – 77.5Hz, Open-circuit voltage: < 20Vrms

Phase sequence rotation with 1-wire method


Voltage range P-N, P-PE[V] Frequency range
100  265 50Hz/60Hz  5%
Measurement is only carried out by direct contact with metal live parts (not on insulation sheath)

Voltage drop on main power lines (V%)


Range (%) Resolution (%) Accuracy
0  100 0.1 (10.0 rdg + 4dgt)
Voltage range Phase-PE, Phase-Neutral: 100 265V, Frequency: 50/60Hz  5%

Leakage current (by HT96U optional clamp transducer)


FS clamp AC (A) Resolution Accuracy
1 0.1mA
1 < FS <10 0.01A
(1.0rdg + 20dgt)
10 ≤ FS <100 0.1A
100 ≤ FS <1000 1A

Environmental parameters (AUX function)


Parameter Range Resolution Accuracy
Temperature [°C] -20°C  80°C 0.1 °C
Temperature [°F] -4°F  176°F 0.1 °F
Relative humidity [%HR] 0  100%HR 0.1% UR
DC output voltage 0.1mV  1.0V 0.1mV (2.0rdg+2dgt)
Illuminance [Lux] 0.001Lux  20.00 Lux (*) 0.001  0.02 Lux
(*) Accuracy of HT53 lux probe is 0.1 Lux  2000 Lux (*) 0.1  2 Lux
according to Class AA 1 Lux  20 kLux (*) 1  20 Lux
HT ITALIA SRL
Via della Boaria, 40
48018 – Faenza (RA) – Italy WHERE
T +39 0546 621002 | F +39 0546 621144 WE ARE
M [email protected] | ht-instruments.com
GSC60 Rel 2.00 of 03/07/24

Professional safety tester and network analyzer Pag 5 of 7

2. ELECTRICAL SPECIFICATIONS – PQA SECTION


AC TRMS Voltage (L-N)
Range [V] Resolution [V] Accuracy
15.0  380.0 0.1V (1.0rdg + 1dgt)
Allowed crest factor:  1,5 ; Frequency: 42  69.0 Hz

AC TRMS Voltage (L-L)


Range [V] Resolution [V] Accuracy
15.0  660.0 0.1V (1.0rdg + 1dgt)
Allowed crest factor:  1,5 ; Frequency: 42  69.0 Hz

Frequency
Range [Hz] Resolution [Hz] Accuracy
DC, 42  69.0 0.01 (2.0rdg + 2dgt)
Allowed voltage: 15.0  660V ; Allowed current: 5%FS clamp  FS clamp

DC/ AC TRMS Current (STD clamp)


FS clamp Range [A] Resolution [A] Accuracy
 10A 5% FS  9.99 0.01
10A  FS  300 5% FS  299.9 0.1 (1.0rdg + 3 dgt)
300A  FS  3000 5% FS  2999 1
Range: 5  999.9 mV; Values under 5mV are zeroed
Allowed crest factor:  3; Frequency: 42  69.0 Hz

AC TRMS Current (FLEX clamp – 300A AC)


Range [mV] Frequency [Hz] Resolution Accuracy Overload protection
0.085  85.0 42  69.0 8.5V (0.5rdg+0.17%FS) 10V
Allowed crest factor 3, Values under 1A are zeroed

AC TRMS Current (FLEX clamp – 3000A AC)


Range [mV] Frequency [Hz] Resolution Accuracy Overload protection
0.425  255.0 42  69.0 85V (0.5rdg+0.17%FS) 10V
Allowed crest factor 3, Values under 10A are zeroed

DC Power
FS clamp Range [kW] Resolution [kW] Accuracy
0.000  9.999 0.001
 10A
10.00  99.99 0.01
0.00  99.99 0.01
10A  FS  200 (2.0rdg + 7dgt)
100.0  999.9 0.1
0.0  999.9 0.1
200A  FS  1000
1000  9999 1

Active power (@ 230V, I> 5%FS, cos ≥ 0.5, f=50.0Hz)


FS clamp Range [kW] Resolution [kW] Accuracy
0.000  9.999 0.001
 10A
10.00  99.99 0.01
0.00  99.99 0.01
10A  FS  200
100.0  999.9 0.1 (2.0rdg + 7dgt)
0.0  999.9 0.1
200A  FS  1000
1000  9999 1
1000A  FS  3000 0  9999 1

HT ITALIA SRL
Via della Boaria, 40
48018 – Faenza (RA) – Italy WHERE
T +39 0546 621002 | F +39 0546 621144 WE ARE
M [email protected] | ht-instruments.com
GSC60 Rel 2.00 of 03/07/24

Professional safety tester and network analyzer Pag 6 of 7

Reactive power (@ 230V, I >5%FS, cos<0.9, f=50.0Hz)


FS clamp Range [kVAr] Resolution [kVAr] Accuracy
0.000  9.999 0.001
 10A
10.00  99.99 0.01
0.00  99.99 0.01
10A  FS  200
100.0  999.9 0.1 (2.0rdg + 7dgt)
0.0  999.9 0.1
200A  FS  1000
1000  9999 1
1000A  FS  3000 0  9999 1

Power factor / cos (@ 230V, I >5%FS)


Range Resolution Accuracy

0.70c  1.00  0.70i 0.01 (2.0rdg + 3dgt)

Voltage harmonics (@ 230V in 1Ph systems, 400V in 3Ph systems)


Range [%] Resolution [%] Order Accuracy
0.1  100.0 0.1 DC, 01  49 (5.0rdg + 5dgt)
Frequency of fundamental: 42  69.0 Hz
Harmonics are zeroed at the below conditions:
➢ DC : DC value <0.5% fundamental value or DC value < 1.0V
➢ 1° Harmonic: value of 1° Harmonic < 15V
➢ 2nd ÷ 49th Harmonics: harmonic value <0.5% fundamental value or if value < 1.0V

Current harmonics
Range [%] Resolution [%] Order Accuracy
0.1  100.0 0.1 DC, 01  49 (5.0rdg + 5dgt)
Frequency of fundamental: 42  69.0 Hz
Harmonics are zeroed at the below conditions:
➢ DC : DC value <0.5% fundamental value or DC value < 0.5%FS clamp
➢ 1° Harmonic: value of 1° Harmonic < 0.5%FS clamp
➢ 2nd ÷ 49th Harmonics: harmonic value <0.5% fundamental value or if value < 0.5%FS clamp

Voltage anomalies (L-N, L-PE)


Range [V] Resolution [V] Resolution [ms] Accuracy [V] Accuracy [ms]
15.0  380 0.2 20ms (1.0rdg + 2dgt)  1cycle

Voltage anomalies (L-L)


Range [V] Resolution [V] Resolution [ms] Accuracy [V] Accuracy [ms]
15.0  660 0.2 20ms (1.0rdg + 2dgt)  1cycle

HT ITALIA SRL
Via della Boaria, 40
48018 – Faenza (RA) – Italy WHERE
T +39 0546 621002 | F +39 0546 621144 WE ARE
M [email protected] | ht-instruments.com
GSC60 Rel 2.00 of 03/07/24

Professional safety tester and network analyzer Pag 7 of 7

3. GENERAL SPECIFICATIONS
DISPLAY AND MEMORY:
Features: TFT, touch screen, color graphic LCD, 320x240mm
Memory safety section: 999 locations, 3 marker levels
Memory PQA section: 8MB (not expanded)
Communication: Optical-USB and built-in WiFi
Aggregation time (IP) PQA feature: 2s ÷ 30min selectable
Parameters saved PQA feature: ca 600
POWER SUPPLY:
Batteries: 6 x 1.2V(rechargeable) type AA or 6 x 1.5V type AA
Battery life: > 500 test for each safety functions
Recording autonomy: approx.. 43 days (IP=15min)
approx 2 days (IP=1min)
approx.. 2 hours (IP=2s)
Recharging time: approx. 12 hours
External charger: 100-240VAC, 50/60Hz / 15VDC, CAT IV 300V
Auto Power OFF: after 5 min of idleness (disabled)
MECHANICAL FEATURES:
Dimensions (L x W x H): 225 x 165 x 75mm
Weight (included batteries): 1.2kg
Mechanical protection: IP40
WORKING ENVIRONMENTAL CONDITIONS:
Reference temperature: 23°C  5°C
Working temperature: 0°C  40°C
Allowed relative humidity: <80%RH
Storage temperature: -10°C  60°C
Storage humidity: <80%RH
Max height of use: 2000m
GENERAL REFERENCE STANDARDS:
Safety of measuring instruments: IEC/EN61010-1, IEC/EN61010-031, IEC/EN61010-2-032
Product type standard: IEC/EN61557-1-2-3-4-5-6-7-10
EMC : IEC/EN61326-1
Technical documentation : IEC/EN61187
Insulation: double insulation
Pollution degree: 2
Measurement category: CAT IV 300V to ground, CAT III 350V to ground
max 600V among inputs
TEST VERIFIES REFERENCE STANDARDS:
Continuity test with 200mA: IEC/EN61557-4
Insulation resistance: IEC/EN61557-2
Earth resistance: IEC/EN61557-5
Fault loop impedance: IEC/EN61557-3
RCD test: IEC/EN61557-6 (only Phase-Neutral-Ground systems)
Multifunction: IEC/EN61557-10
Prospective short circuit current: EN60909-0
Earth resistance on TN systems: EN61936-1 + EN50522
Power quality: EN50160
This instrument complies with the requirements of the European Low Voltage Directives 2014/35/EU
(LVD), EMC Directive 2014/30/EU and RED 2014/53/EU Directive
This instrument complies with the requirements of the European 2011/65/EU (RoHS) and with the
requirements of the European 2012/19/EU (WEEE)

HT ITALIA SRL
Via della Boaria, 40
48018 – Faenza (RA) – Italy WHERE
T +39 0546 621002 | F +39 0546 621144 WE ARE