Critisure Meditech Scope CC 4008
Critisure Meditech Scope CC 4008
Laboratory Name :
RAJKOT, GUJARAT, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-4008 Page No 1 of 19
Validity 01/08/2024 to 08/11/2024 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Permanent Facility
MEDICAL
DEVICES- Using Gas Flow
Anesthesia Machine
1 DISCHARGE analyzer by 10 psi to 80 psi 1.11 psi
-PRESSURE
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Vital Sign
10 mmHg to 290
2 DISCHARGE BP Apparatus Simulator by 3.55 mmHg
mmHg
EQUIPMENT/DE Comparison method
VICES
Electrical Safety -
Protective Earth-
MEDICAL
Suction
DEVICES- Using Electrical
Pump,Nebulizer,Syri 0.01 ohm to 0.04
3 DISCHARGE Safety Analyzer by 0.2 Ohm to 1.8 Ohm
ng Pump,Infusion ohm
EQUIPMENT/DE Direct method
Pump, CPAP, BIPAP,
VICES
Anesthesia Machine,
Pulse Oxymeter
Electrical Safety
Analyze-Mains
MEDICAL
Voltage-Suction
DEVICES- Using Electrical
Pump,Nebulizer,Syri
4 DISCHARGE Safety Analyzer by 10 V to 240 V 7.4 % to 2.3 %
ng Pump,Infusion
EQUIPMENT/DE Direct method
Pump, CPAP, BIPAP,
VICES
Anesthesia Machine,
Pulse Oxymeter
This is annexure to 'Certificate of Accreditation' and does not require any signature.
CRITISURE MEDITECH, B 310, WING B, THE ONE WORLD, 150 FT RING ROAD,
Laboratory Name :
RAJKOT, GUJARAT, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-4008 Page No 2 of 19
Validity 01/08/2024 to 08/11/2024 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Electrical Safety-
Insulation
MEDICAL
Resistance-Suction
DEVICES- Using Electrical
Pump,Nebulizer,Syri 10 MegaOhm to 100 0.24 Mohm to 8.77
5 DISCHARGE Safety Analyzer by
ng Pump,Infusion MegaOhm Mohm
EQUIPMENT/DE Direct method
Pump, CPAP, BIPAP,
VICES
Anesthesia Machine,
Pulse Oxymeter
Electrical Safety-
Leakage
MEDICAL Current(Paitent,Encl,
DEVICES- Earth)-Suction Using Electrical
6 DISCHARGE Pump,Nebulizer,Syri Safety Analyzer by 10 µA to 10000 µA 11.6 %
EQUIPMENT/DE ng Pump,Infusion Direct method
VICES Pump, CPAP, BIPAP,
Anesthesia Machine,
Pulse Oxymeter
MEDICAL
DEVICES- Using Gas Flow
FIO2 - Anesthesia
7 DISCHARGE analyzer by 21 % to 100 % 0.78 % to 1.68 %
Machine
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Gas Flow
Flow - Boyles
8 DISCHARGE analyzer by 1 lpm to 30 lpm 0.08 lpm to 1.2 lpm
Apparatus
EQUIPMENT/DE Comparison method
VICES
This is annexure to 'Certificate of Accreditation' and does not require any signature.
CRITISURE MEDITECH, B 310, WING B, THE ONE WORLD, 150 FT RING ROAD,
Laboratory Name :
RAJKOT, GUJARAT, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-4008 Page No 3 of 19
Validity 01/08/2024 to 08/11/2024 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MEDICAL
DEVICES- Using Gas Flow
Flow - Flow meter
9 DISCHARGE analyzer by 1 lpm to 30 lpm 0.08 lpm to 1 lpm
with Humidifier
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Infusion
10 ml/hr to 999 0.15 ml/hr to 12.64
10 DISCHARGE Flow - Infusion Pump device analyzer by
ml/hr ml/hr
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Infusion
10 ml/hr to 999 0.18 ml/hr to 12.66
11 DISCHARGE Flow - Syringe Pump device analyzer by
ml/hr ml/hr
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Gas Flow
12 DISCHARGE Flow- Nebulizer analyzer by 1 lpm to 11 lpm 0.30 lpm to 0.57 lpm
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Infusion
Occlusion Pressure -
13 DISCHARGE device analyzer by 0 psi to 21 psi 0.71 psi
Infusion Pump
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Infusion
Occlusion Pressure -
14 DISCHARGE device analyzer by 0 psi to 21 psi 0.71 psi
Syringe Pump
EQUIPMENT/DE Comparison method
VICES
This is annexure to 'Certificate of Accreditation' and does not require any signature.
CRITISURE MEDITECH, B 310, WING B, THE ONE WORLD, 150 FT RING ROAD,
Laboratory Name :
RAJKOT, GUJARAT, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-4008 Page No 4 of 19
Validity 01/08/2024 to 08/11/2024 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MEDICAL
DEVICES- Using Gas Flow
15 DISCHARGE Pressure - CPAP analyzer by 1 mbar to 10 mbar 0.16 mbar
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Gas Flow
Pressure(IPAP/EPAP)
16 DISCHARGE analyzer by 1 mbar to 10 mbar 0.16 mbar
- BiPAP
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Gas Flow
Pressure- Boyles
17 DISCHARGE analyzer by 10 psi to 80 psi 1.11 psi
Apparatus
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Gas Flow
Pressure- Oxygen
18 DISCHARGE analyzer by 10 psi to 80 psi 1.11 psi
Cylinder
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Vital Sign
Pulse Rate - Pulse
19 DISCHARGE Simulator by 30 bpm to 240 bpm 1%
Oxymeter
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Gas Flow
Respiration Rate - 30 brpm to 240 0.8 brpm to 9.1
20 DISCHARGE analyzer by
BiPAP brpm brpm
EQUIPMENT/DE Comparison method
VICES
This is annexure to 'Certificate of Accreditation' and does not require any signature.
CRITISURE MEDITECH, B 310, WING B, THE ONE WORLD, 150 FT RING ROAD,
Laboratory Name :
RAJKOT, GUJARAT, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-4008 Page No 5 of 19
Validity 01/08/2024 to 08/11/2024 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MEDICAL
DEVICES- Using Gas Flow
Respiratory Rate - 30 brpm to 240 0.8 brpm to 9.1
21 DISCHARGE analyzer by
Anesthesia Machine brpm brpm
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using SPO2 analyzer
SPO2 - Pulse
22 DISCHARGE by comparison 80 % to 100 % 1%
Oxymeter
EQUIPMENT/DE method
VICES
MEDICAL
DEVICES- Using Gas Flow
Suction Pump -
23 DISCHARGE analyzer by 0 psi to 12 psi 1%
Vacuum
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Infusion
Volume - Syringe
24 DISCHARGE device analyzer by 5 ml to 400 ml 0.20 ml to 9.31 ml
Pump
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Infusion
Volume- Infusion
25 DISCHARGE device analyzer by 5 ml to 400 ml 0.71 ml
Pump
EQUIPMENT/DE Comparison method
VICES
MEDICAL
Using Vital Sign
DEVICES-
26 ECG Output - ECG patient simulator by 0.48 mV to 2.24 mV 0.12 mV to 0.30 mV
IMAGING/PLOT
Comaprison method
TERS
This is annexure to 'Certificate of Accreditation' and does not require any signature.
CRITISURE MEDITECH, B 310, WING B, THE ONE WORLD, 150 FT RING ROAD,
Laboratory Name :
RAJKOT, GUJARAT, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-4008 Page No 6 of 19
Validity 01/08/2024 to 08/11/2024 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
This is annexure to 'Certificate of Accreditation' and does not require any signature.
CRITISURE MEDITECH, B 310, WING B, THE ONE WORLD, 150 FT RING ROAD,
Laboratory Name :
RAJKOT, GUJARAT, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-4008 Page No 7 of 19
Validity 01/08/2024 to 08/11/2024 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MEDICAL
Using Vital Sign
DEVICES- NIBP - Patient 50 mmHg to 390
34 patient simulator by 3.05 mmHg
MONITORING Monitors mmHg
Comaprison method
UNIT
MEDICAL
Using SPO2 analyzer
DEVICES- PULSE RATE-Patient
35 by comparison 30 BrPM to 240 BrPM 1.63 BrPM
MONITORING Monitors
method
UNIT
MEDICAL
DEVICES-
Using Defibrillator
PATIENT Charge Time -
36 analyzer by 1.8 s to 7.5 s 0.3 s to 0.25 s
CONDITIONING Defibrilator
Comparison method
/
MAINTENANCE
Electrical Safety -
Insulation
MEDICAL
Resistance-
DEVICES-
Autoclave, Using Electrical
PATIENT 10 MegaOhm to 100 0.24 Mohm to 8.77
37 Defibrillator,Ventilat Safety Analyzer by
CONDITIONING MegOhm Mohm
or, Electronic Direct method
/
Tourniquet, OT
MAINTENANCE
Table, Electronic/
Mechanical Bed
Electrical Safety -
MEDICAL Protective Earth-
DEVICES- Autoclave,
Using Electrical
PATIENT Defibrillator,Ventilat 0.01 ohm to 0.04
38 Safety Analyzer by 0.2 Ohm to 1.8 Ohm
CONDITIONING or, Electronic ohm
Direct method
/ Tourniquet, OT
MAINTENANCE Table, Electronic/
Mechanical Bed
This is annexure to 'Certificate of Accreditation' and does not require any signature.
CRITISURE MEDITECH, B 310, WING B, THE ONE WORLD, 150 FT RING ROAD,
Laboratory Name :
RAJKOT, GUJARAT, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-4008 Page No 8 of 19
Validity 01/08/2024 to 08/11/2024 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Electrical Safety-
MEDICAL Mains Voltage -
DEVICES- Autoclave,
Using Electrical
PATIENT Defibrillator,Ventilat
39 Safety Analyzer by 10 V to 240 V 7.4 % to 2.3 %
CONDITIONING or, Electronic
Direct method
/ Tourniquet, OT
MAINTENANCE Table, Electronic/
Mechanical Bed
MEDICAL
DEVICES-
Using Defibrillator
PATIENT
40 Energy - Defibrilator analyzer by 10.50 J to 300 J 0.68 J to 3.75 J
CONDITIONING
Comparison method
/
MAINTENANCE
MEDICAL
DEVICES-
Using Gas Flow
PATIENT
41 FIO2 - Ventilator analyzer by 21% to 100% 0.78 % to 1.68 %
CONDITIONING
Comparison method
/
MAINTENANCE
MEDICAL
DEVICES-
Using Gas Flow
PATIENT
42 Flow - Ventilator analyzer by 1 LPM to 80 LPM 0.08 lpm to 1.11 lpm
CONDITIONING
Comparison method
/
MAINTENANCE
This is annexure to 'Certificate of Accreditation' and does not require any signature.
CRITISURE MEDITECH, B 310, WING B, THE ONE WORLD, 150 FT RING ROAD,
Laboratory Name :
RAJKOT, GUJARAT, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-4008 Page No 9 of 19
Validity 01/08/2024 to 08/11/2024 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MEDICAL
DEVICES-
Using Defibrillator
PATIENT Heart Rate -
43 analyzer by 30 BPM to 300 BPM 0.76 bpm to 9.1 bpm
CONDITIONING Defibrillator
Comparison method
/
MAINTENANCE
MEDICAL
DEVICES-
Using Gas Flow
PATIENT Pressure (PEEP,
44 analyzer by 1 mbar to 10 mbar 0.16 mbar
CONDITIONING PIP,MAP) - Ventilator
Comparison method
/
MAINTENANCE
MEDICAL
DEVICES-
Using Vital Sign
PATIENT Pressure - Electronic 50 mmHg to 360
45 Simulator by 4.47 mmHg
CONDITIONING Tourniquet mmHg
Comparison method
/
MAINTENANCE
MEDICAL
DEVICES-
Using Gas Flow
PATIENT Respiration Rate - 30 brpm to 240 0.8 brpm to 3.8
46 analyzer by
CONDITIONING Ventilator brpm brpm
Comparison method
/
MAINTENANCE
MEDICAL
DEVICES-
Volume (Tidal Using Gas Flow
PATIENT
47 Volume, Minute analyzer by 0.22 ltr to 25 ltr 0.02 ltr to 0.7 ltr
CONDITIONING
Volume) - Ventilator Comparison method
/
MAINTENANCE
This is annexure to 'Certificate of Accreditation' and does not require any signature.
CRITISURE MEDITECH, B 310, WING B, THE ONE WORLD, 150 FT RING ROAD,
Laboratory Name :
RAJKOT, GUJARAT, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-4008 Page No 10 of 19
Validity 01/08/2024 to 08/11/2024 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Site Facility
MEDICAL
DEVICES- Using Gas Flow
Anesthesia Machine
1 DISCHARGE analyzer by 10 psi to 80 psi 1.11 psi
-PRESSURE
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Vital Sign
10 mmHg to 290
2 DISCHARGE BP Apparatus Simulator by 3.55 mmHg
mmHg
EQUIPMENT/DE Comparison method
VICES
Electrical Safety -
Protective Earth-
MEDICAL
Suction
DEVICES- Using Electrical
Pump,Nebulizer,Syri 0.01 ohm to 0.04
3 DISCHARGE Safety Analyzer by 0.2 Ohm to 1.8 Ohm
ng Pump,Infusion ohm
EQUIPMENT/DE Direct method
Pump, CPAP, BIPAP,
VICES
Anesthesia Machine,
Pulse Oxymeter
Electrical Safety
Analyze-Mains
MEDICAL
Voltage-Suction
DEVICES- Using Electrical
Pump,Nebulizer,Syri
4 DISCHARGE Safety Analyzer by 10 V to 240 V 7.4 % to 2.3 %
ng Pump,Infusion
EQUIPMENT/DE Direct method
Pump, CPAP, BIPAP,
VICES
Anesthesia Machine,
Pulse Oxymeter
This is annexure to 'Certificate of Accreditation' and does not require any signature.
CRITISURE MEDITECH, B 310, WING B, THE ONE WORLD, 150 FT RING ROAD,
Laboratory Name :
RAJKOT, GUJARAT, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-4008 Page No 11 of 19
Validity 01/08/2024 to 08/11/2024 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Electrical Safety-
Insulation
MEDICAL
Resistance-Suction
DEVICES- Using Electrical
Pump,Nebulizer,Syri 10 MegaOhm to 100 0.24 Mohm to 8.77
5 DISCHARGE Safety Analyzer by
ng Pump,Infusion MegaOhm Mohm
EQUIPMENT/DE Direct method
Pump, CPAP, BIPAP,
VICES
Anesthesia Machine,
Pulse Oxymeter
Electrical Safety-
Leakage
MEDICAL Current(Paitent,Encl,
DEVICES- Earth)-Suction Using Electrical
6 DISCHARGE Pump,Nebulizer,Syri Safety Analyzer by 10 µA to 10000 µA 11.6 %
EQUIPMENT/DE ng Pump,Infusion Direct method
VICES Pump, CPAP, BIPAP,
Anesthesia Machine,
Pulse Oxymeter
MEDICAL
DEVICES- Using Gas Flow
FIO2 - Anesthesia
7 DISCHARGE analyzer by 21 % to 100 % 0.78 % to 1.68 %
Machine
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Gas Flow
Flow - Boyles
8 DISCHARGE analyzer by 1 lpm to 30 lpm 0.08 lpm to 1.2 lpm
Apparatus
EQUIPMENT/DE Comparison method
VICES
This is annexure to 'Certificate of Accreditation' and does not require any signature.
CRITISURE MEDITECH, B 310, WING B, THE ONE WORLD, 150 FT RING ROAD,
Laboratory Name :
RAJKOT, GUJARAT, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-4008 Page No 12 of 19
Validity 01/08/2024 to 08/11/2024 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MEDICAL
DEVICES- Using Gas Flow
Flow - Flow meter
9 DISCHARGE analyzer by 1 lpm to 30 lpm 0.08 lpm to 1 lpm
with Humidifier
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Infusion
10 ml/hr to 999 0.15 ml/hr to 12.64
10 DISCHARGE Flow - Infusion Pump device analyzer by
ml/hr ml/hr
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Infusion
10 ml/hr to 999 0.18 ml/hr to 12.66
11 DISCHARGE Flow - Syringe Pump device analyzer by
ml/hr ml/hr
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Gas Flow
12 DISCHARGE Flow- Nebulizer analyzer by 1 lpm to 11 lpm 0.30 lpm to 0.57 lpm
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Infusion
Occlusion Pressure -
13 DISCHARGE device analyzer by 0 psi to 21 psi 0.71 psi
Infusion Pump
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Infusion
Occlusion Pressure -
14 DISCHARGE device analyzer by 0 psi to 21 psi 0.71 psi
Syringe Pump
EQUIPMENT/DE Comparison method
VICES
This is annexure to 'Certificate of Accreditation' and does not require any signature.
CRITISURE MEDITECH, B 310, WING B, THE ONE WORLD, 150 FT RING ROAD,
Laboratory Name :
RAJKOT, GUJARAT, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-4008 Page No 13 of 19
Validity 01/08/2024 to 08/11/2024 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MEDICAL
DEVICES- Using Gas Flow
15 DISCHARGE Pressure - CPAP analyzer by 1 mbar to 10 mbar 0.16 mbar
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Gas Flow
Pressure(IPAP/EPAP)
16 DISCHARGE analyzer by 1 mbar to 10 mbar 0.16 mbar
- BiPAP
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Gas Flow
Pressure- Boyles
17 DISCHARGE analyzer by 10 psi to 80 psi 1.11 psi
Apparatus
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Gas Flow
Pressure- Oxygen
18 DISCHARGE analyzer by 10 psi to 80 psi 1.11 psi
Cylinder
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Vital Sign
Pulse Rate - Pulse
19 DISCHARGE Simulator by 30 bpm to 240 bpm 1%
Oxymeter
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Gas Flow
Respiration Rate - 30 brpm to 240 0.8 brpm to 9.1
20 DISCHARGE analyzer by
BiPAP brpm brpm
EQUIPMENT/DE Comparison method
VICES
This is annexure to 'Certificate of Accreditation' and does not require any signature.
CRITISURE MEDITECH, B 310, WING B, THE ONE WORLD, 150 FT RING ROAD,
Laboratory Name :
RAJKOT, GUJARAT, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-4008 Page No 14 of 19
Validity 01/08/2024 to 08/11/2024 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MEDICAL
DEVICES- Using Gas Flow
Respiratory Rate - 30 brpm to 240 0.8 brpm to 9.1
21 DISCHARGE analyzer by
Anesthesia Machine brpm brpm
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using SPO2 analyzer
SPO2 - Pulse
22 DISCHARGE by comparison 80 % to 100 % 1%
Oxymeter
EQUIPMENT/DE method
VICES
MEDICAL
DEVICES- Using Gas Flow
Suction Pump -
23 DISCHARGE analyzer by 0 psi to 12 psi 1%
Vacuum
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Infusion
Volume - Syringe
24 DISCHARGE device analyzer by 5 ml to 400 ml 0.20 ml to 9.31 ml
Pump
EQUIPMENT/DE Comparison method
VICES
MEDICAL
DEVICES- Using Infusion
Volume- Infusion
25 DISCHARGE device analyzer by 5 ml to 400 ml 0.71 ml
Pump
EQUIPMENT/DE Comparison method
VICES
MEDICAL
Using Vital Sign
DEVICES-
26 ECG Output - ECG patient simulator by 0.48 mV to 2.24 mV 0.12 mV to 0.30 mV
IMAGING/PLOT
Comaprison method
TERS
This is annexure to 'Certificate of Accreditation' and does not require any signature.
CRITISURE MEDITECH, B 310, WING B, THE ONE WORLD, 150 FT RING ROAD,
Laboratory Name :
RAJKOT, GUJARAT, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-4008 Page No 15 of 19
Validity 01/08/2024 to 08/11/2024 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
This is annexure to 'Certificate of Accreditation' and does not require any signature.
CRITISURE MEDITECH, B 310, WING B, THE ONE WORLD, 150 FT RING ROAD,
Laboratory Name :
RAJKOT, GUJARAT, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-4008 Page No 16 of 19
Validity 01/08/2024 to 08/11/2024 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MEDICAL
Using Vital Sign
DEVICES- NIBP - Patient 50 mmHg to 390
34 patient simulator by 3.05 mmHg
MONITORING Monitors mmHg
Comaprison method
UNIT
MEDICAL
Using SPO2 analyzer
DEVICES- PULSE RATE-Patient
35 by comparison 30 BrPM to 240 BrPM 1.63 BrPM
MONITORING Monitors
method
UNIT
MEDICAL
DEVICES-
Using Defibrillator
PATIENT Charge Time -
36 analyzer by 1.8 s to 7.5 s 0.3 s to 0.25 s
CONDITIONING Defibrilator
Comparison method
/
MAINTENANCE
Electrical Safety -
Insulation
MEDICAL
Resistance-
DEVICES-
Autoclave, Using Electrical
PATIENT 10 MegaOhm to 100 0.24 Mohm to 8.77
37 Defibrillator,Ventilat Safety Analyzer by
CONDITIONING MegOhm Mohm
or, Electronic Direct method
/
Tourniquet, OT
MAINTENANCE
Table, Electronic/
Mechanical Bed
Electrical Safety -
MEDICAL Protective Earth-
DEVICES- Autoclave,
Using Electrical
PATIENT Defibrillator,Ventilat 0.01 ohm to 0.04
38 Safety Analyzer by 0.2 Ohm to 1.8 Ohm
CONDITIONING or, Electronic ohm
Direct method
/ Tourniquet, OT
MAINTENANCE Table, Electronic/
Mechanical Bed
This is annexure to 'Certificate of Accreditation' and does not require any signature.
CRITISURE MEDITECH, B 310, WING B, THE ONE WORLD, 150 FT RING ROAD,
Laboratory Name :
RAJKOT, GUJARAT, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-4008 Page No 17 of 19
Validity 01/08/2024 to 08/11/2024 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Electrical Safety-
MEDICAL Mains Voltage -
DEVICES- Autoclave,
Using Electrical
PATIENT Defibrillator,Ventilat
39 Safety Analyzer by 10 V to 240 V 7.4 % to 2.3 %
CONDITIONING or, Electronic
Direct method
/ Tourniquet, OT
MAINTENANCE Table, Electronic/
Mechanical Bed
MEDICAL
DEVICES-
Using Defibrillator
PATIENT
40 Energy - Defibrilator analyzer by 10.50 J to 300 J 0.68 J to 3.75 J
CONDITIONING
Comparison method
/
MAINTENANCE
MEDICAL
DEVICES-
Using Gas Flow
PATIENT
41 FIO2 - Ventilator analyzer by 21% to 100% 0.78 % to 1.68 %
CONDITIONING
Comparison method
/
MAINTENANCE
MEDICAL
DEVICES-
Using Gas Flow
PATIENT
42 Flow - Ventilator analyzer by 1 LPM to 80 LPM 0.08 lpm to 1.11 lpm
CONDITIONING
Comparison method
/
MAINTENANCE
This is annexure to 'Certificate of Accreditation' and does not require any signature.
CRITISURE MEDITECH, B 310, WING B, THE ONE WORLD, 150 FT RING ROAD,
Laboratory Name :
RAJKOT, GUJARAT, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-4008 Page No 18 of 19
Validity 01/08/2024 to 08/11/2024 Last Amended on -
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MEDICAL
DEVICES-
Using Defibrillator
PATIENT Heart Rate -
43 analyzer by 30 BPM to 300 BPM 0.76 bpm to 9.1 bpm
CONDITIONING Defibrillator
Comparison method
/
MAINTENANCE
MEDICAL
DEVICES-
Using Gas Flow
PATIENT Pressure (PEEP,
44 analyzer by 1 mbar to 10 mbar 0.16 mbar
CONDITIONING PIP,MAP) - Ventilator
Comparison method
/
MAINTENANCE
MEDICAL
DEVICES-
Using Vital Sign
PATIENT Pressure - Electronic 50 mmHg to 360
45 Simulator by 4.47 mmHg
CONDITIONING Tourniquet mmHg
Comparison method
/
MAINTENANCE
MEDICAL
DEVICES-
Using Gas Flow
PATIENT Respiration Rate - 30 brpm to 240 0.8 brpm to 3.8
46 analyzer by
CONDITIONING Ventilator brpm brpm
Comparison method
/
MAINTENANCE
MEDICAL
DEVICES-
Volume (Tidal Using Gas Flow
PATIENT
47 Volume, Minute analyzer by 0.22 ltr to 25 ltr 0.02 ltr to 0.7 ltr
CONDITIONING
Volume) - Ventilator Comparison method
/
MAINTENANCE
This is annexure to 'Certificate of Accreditation' and does not require any signature.
CRITISURE MEDITECH, B 310, WING B, THE ONE WORLD, 150 FT RING ROAD,
Laboratory Name :
RAJKOT, GUJARAT, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-4008 Page No 19 of 19
Validity 01/08/2024 to 08/11/2024 Last Amended on -
* CMCs represent expanded uncertainties expressed at approximately the 95% level of confidence, using a coverage factor of k = 2.
This is annexure to 'Certificate of Accreditation' and does not require any signature.