Gs11g02m01-01e 04B
Gs11g02m01-01e 04B
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02M01-01E 4th Edition June 16, 2017-09
3
are open. analyzer error No. 1, 3 or 10.
Auto calibration: Calibration error contact output:
Auto calibration is carried out Contacts close at occurrence of manual
periodically at the preset cycle. or auto calibration error (any of errors
When a standard gas cylinder for No.4 to 9).
calibration and a solenoid valve for Auto calibration status contact output:
opening/ closing the gas flow line are Contacts close during auto calibration.
prepared externally by the customer, Pump ON/OFF contact output:
calibration will be carried out with the During measurement, contacts close.
solenoid valve drive contacts for zero
While calibration gas is flowing, contacts
calibration and each span calibration
open. Contacts are connected in power
turned on/off sequentially at the set auto
calibration timing. supply of pump, and stop the sample
Auto calibration cycle setting: gas while calibration gas is flowing.
Auto calibration cycle is set. Optional Functions
Setting is variable within 1 to 99 hours
(in increments of 1 hour) or 1 to 40 days O2 correction: Conversion of measured CO, SO2 and
(in increments of 1 day). NO gas concentrations into values at
Gas flow time setting: reference O2 concentration.
The time for flowing each calibration gas 21 – On
Correction formula: C = 3 Cs
in auto calibration is set. 21 – Os
Settable within 60 to 599 seconds (in Where:
increments of 1 second) C: Sample gas concentration after
Auto calibration remote start: O2 correction
Auto calibration is carried out only once Cs: Measured concentration of
according to an external input signal. sample gas
Calibration sequence is settable in the Os: Measured O2 concentration
same way as the cyclic auto calibration. On: Reference O2 concentration
Calibration starts by opening the auto (value changeable by setting)
calibration remote start input terminals * The upper limit value of the fractional
after short-circuiting for 1.5 seconds or part in this calculation is 4.
longer. Auto calibration is started when The result of calculation is indicated
the contacts open. and output in an analog output signal.
Auto zero calibration: Average value after O2 correction and O2 average value
Auto zero calibration is carried out calculation: The result of O2 correction or instanta-
periodically at the preset cycle. neous O2 value can be output as an
This cycle is independent of “Auto average value in the determined period
calibration” cycle. of time.
When zero calibration gas and solenoid Used for averaging is the moving
valve for opening/closing the calibration average method in which sampling is
gas flow line are prepared externally by carried out at intervals of 30 seconds.
the customer, zero calibration will be (Output is updated every 30 seconds. It
carried out with the solenoid valve drive is the average value in the determined
contact for zero calibration turned on/off period of time just before the latest
at the set auto zero calibration timing. updating.)
Auto zero calibration cycle setting: Averaging time is settable within 1 to 59
Auto zero calibration cycle is set. minutes (in increments of 1 minute) or 1
Setting is variable within 1 to 99 hours to 4 hours (in increments of 1 hour).
(in increments of 1 hour) or 1 to 40 days Average value reset:
(in increments of 1 day). The above-mentioned output of average
Gas flow time setting: value is started from the initial state by
The time for flowing zero gas in auto opening the average value resetting
zero calibration is set. input terminals after short-circuiting for
Settable 60 to 599 seconds (in 1.5 seconds or longer.
increments of 1 second) Output is reset by short-circuiting and
Upper/lower limit alarm: restarted by opening.
Alarm contact output turns on when the CO concentration peak alarm:
preset upper or lower limit alarm value is (available only for CO + O2
reached. measurement)
Contacts close when the instantaneous Alarm output turns on according to the
value of each component becomes preset concentration and count.
larger than the upper alarm limit value or Whenever the instantaneous value of
smaller than the lower alarm limit value. CO exceeds the preset concentration
Instrument error contact output: value, count increments. If the count
Contacts close at occurrence of exceeds the preset value in one hour,
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02M01-01E 4th Edition Jun.01, 2006-00
4
the alarm contacts close. Corrosive component: HCl 1 ppm or less
Communication function: Standard gas for calibration:
RS-232C (9 pins D-sub) Zero gas; Dry N2
Half-duplex bit serial Span gas; Each sample gas having
Start-stop synchronization concentration 90 to 100%
ModbusTM protocol of its measuring range
Contents: Read/write parameters (recommended).
Read measurement Gas beyond concentration
concentration and instrument 100% is unusable.
In case a zirconia O2 analyzer is
status
installed externally and calibration is
Remark: When connecting via RS-485
carried out on the same calibration gas
interface, a RS-232C ↔ RS- line:
485 converter should be used. Zero gas; Dry air or atmospheric air
(provided without CO2
Performance
sensor)
Repeatability:±0.5% of full scale Span gas; For other than O2
Linearity: ±1% of full scale measurement, each sample
Zero drift: ±2% of full scale/week gas having concentration 90
Span drift: ±2% of full scale/week to 100% of its measuring
Response time (for 90%FS response): range
1 or 2 component measurement; For O2 measurement, O2 gas
Within 15 seconds including of 1 to 2 vol%
replacement time of sample gas It is understood that a large quantity
More than 2 component measurement; of hydrogen, helium, or argon in sample
Within 30 seconds including gas affects the calibration model of
replacement time of sample gas an infrared gas analyzer (pressure
broadening). When measuring such
sample gas, use a gas which has
Interference from other gases: similar composition to the sample gas
as a span gas.
NO Built-in
CO2 CO CH4 SO2
Interference
component analyzer analyzer analyzer analyzer analyzer paramagnetic
O2 analyzer
Installation Requirements
CO 1000 ppm ≤ 1%FS − ≤ 1%FS ≤ 1%FS ≤ 1%FS − • Indoor use: Avoid exposure to direct sunlight,
≤ 1%FS weather, and radiant heat from hot substances.
CO2 15% − for 200 ppm
analyzer,
≤ 1%FS ≤ 1%FS ≤ 1%FS ≤ 2% FS Where exposure to such conditions are unavoidable,
≤ 2.5%FS a protective hood or cover should be prepared.
≤ 1%FS • Minimal vibration
H2O saturation
at 20°C ≤ 1%FS
for 500 ppm
analyzer,
≤ 1%FS − − − • A clean atmosphere
≤ 2.5%FS
≤ 50ppm ≤ 60ppm
≤ 2.5%FS * ≤ 2%FS * ≤ 2%FS
H2O saturation
at 2°C − for 200 ppm − with inter- with inter- −
analyzer ference ference
compen- compen-
sation sation
CH4 1000 ppm ≤ 1%FS ≤ 1%FS − ≤ 50ppm − −
T02.EPS
*: The H2O interference of NO and SO2 analyzer can be reduced by the
interference compensation.
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02M01-01E 4th Edition Jun.01, 2006-00
5
Diagram of measurement principle of infrared gas analyzer (CO2, CO, CH4, SO2, NO)
Front expansion chamber
Detector
Chopper Preamplifier
Signal
Indication processing
and calculation
block
Output F01.EPS
Magnetic field
Gas outlet
Preamplifier
Permanent
magnets
LED
Photodiode
Signal processing
Indication
and calculation block
Output
F02.EPS
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02M01-01E 4th Edition Jun.01, 2006-00
6
Examples of sampling system configuration
Typical examples are shown below. The system configuration may vary depending on sample gas. Consult with
Yokogawa.
Measurement of sample gas with low moisture content (room-temperature saturation level or below):
CO, CO2 and CH4 measurement
10 Secondary
filter Gas External Exhaust
1 Primary filter analyzer (atmospheric
Sample O2 analyzer
7 4-way selector (IR200) pressure)
gas inlet 3 Pump cock
9 Flowmeter
p A dehumidifier must be used for CO (0-200 ppm)
p 4 Electronic
and SO2 measurements (to prevent saturation at
dehumidifier
28°C). Even in any other gas measurements, if
2 Safety sample gas has a high moisture content or if
drain condensation is likely to occur, a safety drain trap
trap 5 Drain 8 and a humidifier should be used.
pot
14 Solenoid 11 NO2/NO
∗ 3 Pump valve converter
Atmosphere ∗ 14
4 Electronic
2 Safety duhumidifier 14 Solenoid
dran trap 14 Solenoid
13 valve
valve Bubbler Gas analyzer
(IR200)
5 Drain
pot
∗ External O2
8 Standard gas analyzer
for zero calibration
6 8 Standard gas
Drain for span calibration Exhaust
Ball valve (atmospheric pressure)
∗ A dehumidifier must be used for NO, SO2, and CO (0-200 ppm) measurements (to prevent
saturation at 28°C). Use either atmospheric air or cylinder gas as a zero calibration gas and
supply it through a bubbler (humidifying) to reduce interference of water.
7 4-way selector cock Used for switching sampling and calibration lines.
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02M01-01E 4th Edition Mar.05, 2014-04
7
MODEL AND SUFFIX CODE
[Style: S3]
Model Suffix code Option code Description
IR200 Infrared gas analyzer 19-inch rack mounting type with slide rail
-A SO2
Measurable -B CO
components -C CO2
-D CH4
(note 8)
-E NO
-F CO2 + CO
-G CH4 + CO
-H CO2 + CH4
-J CO2 + CO + CH4
-K NO + SO2
-L NO + CO
-M NO + SO2 + CO
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02M01-01E 4th Edition Mar.05, 2014-04
8
MODEL AND SUFFIX CODE (Continued)
O2 Analyzer 1 0−5%
1st Range 2 0−10%
(note 2) 3 0−25%
4 0−50%
5 0−100%
N Not available
O2 Analyzer 2 0−10%
2nd Range 3 0−25%
(note 2) 4 0−50%
5 0−100%
N Not available
Output -4 4−20 mA DC, non-isolation
-1 0−1 V DC, non-isolation
Piping R Rc1/4
T 1/4 NPT
Indication, Power Cable (note 6) J Japanese, Power Cable; rated voltage 125 V AC
E English, Power Cable; rated voltage 125 V AC (UL)
U English, Power Cable; rated voltage 250 V AC (CEE)
Option O2 Correction and O2 Average (note 5) /K With O2 correction and O2 average value
Peak alarm (note 5) /A With peak alarm (CO gas Only)
Communication /C RS−232C (note 7)
Internal Purge /P Analyzer internal purging
T04-2.EPS
Footnotes:
1: A signal from the external O2 analyzer should be 0-1 V DC linear to full scale.
2: Possible combinations of ranges are specified in separate tables.
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02M01-01E 4th Edition Mar.05, 2014-04
9
3: Only available for CO measurement.
4: Only available for CO2, CO, SO2 or NO measurements.
5: O2 correction is available only for CO, SO2, and NO. Both average value output after O2
correction and average O2 value output are provided at the same time.
A peak alarm can be provided only for CO measurement.
6: Suffix Codes “E” and “U” are power cables with different voltage rating and plug type. Select
appropriate code according to the operating power supply voltage to be used in the field.
Suffix Code “E” is of the North American plug type and “U” of the European type.
7: Should be specified when using ModbusTM communication.
8: For NOx measurement, a NO2/NO converter (P/N K9350LE or K9350LF) should be purchased separately.
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02M01-01E 4th Edition Sep.11, 2014-07
10
Table 4. Two-component analyzer (CO2 and CH4)
2nd component (CH4), 1st range CH4
1st component C D E F G H K L M P R T
(CO2),1st range 0 – 1000ppm 0 – 2000ppm 0 – 2500ppm 0 – 5000ppm 0 – 1% 0– 2% 0 – 5% 0 – 10% 0 – 20% 0 – 25% 0 – 50% 0 – 100%
D 0 – 2000ppm – – – – – – –
E 0 – 2500ppm – – – – – – –
F 0 to 5000ppm – – – – – –
G 0 – 1% – –
CO2 H 0 – 2% –
K 0 – 5%
L 0 – 10%
M 0 – 20%
R 0 – 50% –
T 0 – 100% –
T08.EPS
: Available as single range, : 2 ranges of 2 or 2.5 times each range available
See Table 4 for CO2 + CH4 measurement of three-component analyzer (CO2 + CO + CH4) and Table 5 for NO +
SO2 measurement of three-component analyzer (NO + SO2 + CO). See Table 1 for CO measurement.
Table 8. O2 analyzer
2nd range 2 3 4 5
1st range 0-10% 0-25% 0-50% 0-100%
1 0-5% −
2 0-10% −
3 0-25% − −
4 0-50% − − −
5 0-100% − − −
T09.EPS
: Built-in O2 analyzer measurable range,
: External zirconia type O2 analyzer (in this case, Yokogawa’s ZX8D)
measurable range
* O2 analyzer is selectable indifferently to combination with other components.
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02M01-01E 4th Edition Dec. 27, 2017-11
11
Table 9. Measurable Components and Their Corresponding Channel Numbers
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02M01-01E 4th Edition Jun.01, 2006-00
12
EXTERNAL DIMENSIONS Unit: mm
500 max
355.6
365
336.5
254
123.8
111.1
101.5
(24)
38
88.5
M4 (for slide rail)
<Front View>
Power Switch
M5
INFRARED GAS ANALYZER
7.5
ENT SPAN
465
483
<Rear View>
Purge gas inlet, Rc1/4 or 1/4NPT
<REAR>
RS232C Terminal Terminal Terminal AC inlet (100 to 240 V AC, 50/60 Hz)
connector block 1 block 3 block 5 F04.EPS
Terminal Terminal
block 2 block 4
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02M01-01E 4th Edition Oct.30, 2009-02
13
EXTERNAL DIMENSIONS OF ACCESSORY SLIDE RAIL Unit: mm
Model: 305A-20/Accuride International Inc.
Closed
(24.4)
365.1±0.3
15.9±0.5
19.1±0.3 505.0±0.8 F05.EPS
Reinforcement plate
101.6
Guide rails
For the guide rail mounted type, a maintenance
space (200 mm or more) should be provided on
top of the main unit.
F06.EPS
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02M01-01E 4th Edition Jun.01, 2006-00
14
F07.EPS
Terminal block 5
<TN5>
Unassigned 1 11
*O2 sensor input is used when an external O2 analyzer is selected.
2 12
CH3 alarm output Unassigned Note: Unassigned terminals are used for internal connection.
(ALM_CH3) 3 13 So they should not be used as repeating terminals either.
4 14
5 15
CH2 alarm output Peak alarm
(ALM_CH2) 6 16 output (PEAK_ALM)
7 17
8 18
CH1 alarm output CH4 alarm output
(ALM_CH1) 9 19 (ALM_CH4)
10 20
(M3.5 screw)
STANDARD ACCESSORIES
Item Part No. Description Qty
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02M01-01E 4th Edition Jun.01, 2006-00
15
Dedicated Zirconia O2 Sensor (to be purchased Dimensions (H x W x D): 140 × 170 ×190 mm
separately) Weight: Approx. 3 kg
For O2 correction, the IR200 can accept linealized 0 Finish color: Munsell 5Y 7/1
to 1 V DC signal coming from an analyzer calibrated
to 0 to 25%O2 of full scale. Dedicated zirconia O2
sensor, Model ZX8D, is available from Yokogawa.e Model Suffix Option Description
code code
Measuring method: Zirconia system ZX8D - - - - - - - - - - - - - - - - Zirconia Oxygen Analyzer
Measurable component and measuring range:
Power -5 -------- 100-115 V AC, 50/60 Hz
supply -3 -------- 200-240 V AC, 50/60 Hz
Measurable component Minimum range Maximum range
Style code *C - - - - - - - - For IR202, IR200, IR400
Oxygen (O2) 0−5 vol% 0−25 vol% *D - - - - - - - - For IR202, IR200, IR400
T12.EPS
(CE conformity)
CAUTION
This instrument is a Class A product, and it is
designed for use in the industrial environment. Please
use this instrument in the industrial environment only.
RoHS ; EN50581
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02M01-01E 4th Edition Dec. 27, 2017-11
16
External Dimensions of ZX8D
(Unit: mm)
150 or
more
4-M4
OXYGEN ANALYZER
TEMP.
150 or more
150 or
141
110
more
131
AC250V 1 2 3 4 5 6 7
T3.15A
US
F
E
L AC N OUT ALM
INPUT
F
E
L N
AC OUT ALM
INPUT
150 or more
Maintenance space F0401.ai
1 2 3 4 5 6 7
+ -
Caution on Safety
Before using this product, be sure to read its instruction manual in advance.
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02M01-01E 4th Edition Aug.22, 2014-06
17
NO2/NO Converter Safety, EMC and RoHS conformity standards
Part number: K9350LE (K9350LF only):
K9350LF Installation altitude; 2000 m or less
Mounting: Indoor surface mounting Pollution degree ; 2 (Note)
Target gases: General boiler exhaust gas, Installation category; II (Note)
Note: Installation category, called over
atmosphere voltage category, specifies impulse
Catalyst: Amount; 2 cm3 withstanding voltage. Category II is
Replacement cycle; Approx. 12 months for electrical equipment.
(at flow rate of 0.3 L/min with 5%O2,10 Pollution degree indicates the degree
ppm NO) of existence of solid, liquid, gas or other
Temperature setpoint; 210 ±10°C inclusions which may reduce dielectric
(Sensing tip: K thermocouple) strength. Degree 2 is the normal in door
Wetted materials: environment.
Ceramic, Viton, glass filter, SUS316 Safety; EN61010-1
Conversion efficiency: 90% or higher, conforms to JIS EMC ; EN61326-1 Class A, Table 2 (For use
Gas flow rate: 0.5 L/min in industrial locations), EN61326-2-3,
Ambient temperature: -5 to 45°C EN61000-3-2, EN61000-3-3
Power supply: 100 V AC, 50/60 Hz (K9350LE) EMC Regulatory Arrangement in
100 to 240 V AC, 50/60 Hz (K9350LF) Australia and New Zealand
Power consumption: Approx. 85 VA
Weight: Approx. 1.1 kg (K9350LE)
Approx. 1.2 kg (K9350LF) CAUTION
Sample gas requirements:
Dust/drain removed, gas This instrument is a Class A product, and it is designed
temperature at 150°C or for use in the industrial environment. Please use this
less instrument in the industrial environment only.
External Dimensions
Unit: mm
(150)
135
Wiring Diagram
(120)
(210)
105
7 8
4-∅5
100 V AC
(Shared with internal wiring)
Approx.130
110
K9350LE K9350LF
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02M01-01E 4th Edition Dec. 27, 2017-11
18
Inquiry Sheet for IR200 Infrared Gas Analyzer
Place a checkmark in the appropriate box and fill in the specific information in the blanks for your reference.
1. General Information
Company : Delivery destination :
Contact person : Section : (Phone No. )
Plant name : Measurement location:
Purpose : Indication reading, Recording, Telemeter transmission, Alarm, Control, Other
2. Requirements
Measurable component: O2 Analyzer:
SO2 Without O2 analyzer
CO External zirconia type sensor (use ZX8D)
CO2 Style C (Non-CE conformity)
CH4 Style D (CE conformity)
NO External O2 analyzer
CO2 + CO Built-in paramagnetic type O2 sensor
CH4 + CO NO2/NO Converter:
CO2 + CH4
NO + SO2 With NO2/NO Converter
NO + CO K9350LE (Non-CE conformity)
CO2 + CO + CH4 K9350LF (CE conformity)
NO + SO2 + CO Without NO2/NO Converter
Range:
1st component, 1st range 1st component, 2nd range 2nd component, 1st range 2nd component, 2nd
range
0 – 200 ppm 0 – 500 ppm 0 – 500ppm 0 – 1000 ppm
0 – 500 ppm 0 – 1000 ppm 0 – 1000 ppm 0 – 2000 ppm
0 – 1000 ppm 0 – 2000 ppm 0 – 2000 ppm 0 – 2500 ppm
0 – 2000 ppm 0 – 2500 ppm 0 – 2500 ppm 0 – 5000 ppm
0 – 2500 ppm 0 – 5000 ppm 0 – 5000 ppm 0 – 1%
0 – 5000 ppm 0 – 1% 0 – 1% 0 – 2%
0 – 1% 0 – 2% 0 – 2% 0 – 5%
0 – 2% 0 – 5% 0 – 3% 0 – 10%
0 – 3% 0 – 10% 0 – 5% 0 – 20%
0 – 5% 0 – 20% 0 – 10% 0 – 25%
0 – 10% 0 – 25% 0 – 20% 0 – 50%
0 – 20% 0 – 50% 0 – 25% 0 – 100%
0 – 25% 0 – 100% 0 – 40% Not available
0 – 40% Not available 0 – 50%
0 – 50% 0 – 70%
0 – 70% 0 – 100%
0 – 100% Not available
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02M01-01E 4th Edition Aug.04, 2014-06
Output : 4 – 20 mA DC 0 – 1 V DC RS-232C
O2 Correction and O2 Average : Yes No
Peak alarm : Yes No
Composition (Detailed composition of sample gas should be provided. This is important for the purpose of knowing
the effect of interference gases.)