EMEC SCL Data Sheet R11 19 1
EMEC SCL Data Sheet R11 19 1
They are equipped with a special membrane system, except for SCL17 and
SCL 18.
SCL models work in chlorine water system.
SCL 3/S SCL2 SCL 9 SCL 11 SCL 17
The probe has to be installed into a probe holder and connected to a measuring SCL 3/N SCL 10 SCL 18
and control instrument. SCL 8
SBR
SCT
MODELS
• SCL3S for free chlorine (organic and inorganic)
• SCL3N for free chlorine (inorganic)
• SCL8 for total chlorine
• SCL2 for chlorine dioxide
• SCL9 for hydrogen peroxyde
• SCL10 for ozone
• SCL11 for peracetic acid
• SBR for bromine
• SCT for Chlorite
• SCL17 for chlorine dioxide - with automatic cleaning system
• SCL18 for free chlorine (inorganic) - with automatic cleaning system
ISO 9001:2008 Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 14001:2004 Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emecpumps.com EN R1-01-18
OHSAS 18001:2007
SCL Data Sheet
Closed amperometric cells
SCL3S
SCL3S/20
Parameter FREE CHLORINE (ORGANIC)
Measuring range 0,1-20 mg/l (0,1-20 ppm)
Voltage supply 0/5 VDC (±10%) - 10 mA
Connection 5-pole screw connector
Measuring system membrane covered, amperometric potentiostatic 2-electrode system
Ph working range 5,5-9,5 pH
Sample water 30-10.000 µS/cm
conductivity
First commissioning: 6 h approx.
Run-in-time
Recommissioning: 3 h approx.
Response time T90: 2 min. approx.
Zero point see chapter “Probe alignment”
adjustment
Slope calibration see chapter “Probe alignment” DPD1 method
Alcalinity 100 ppm
Working 5-45 °C with temperature compensation
temperature
1 bar - 10 mwc [mH2O]
Max pressure
no pressure impulses and/or vibration, no depressure
Power supply 4 wires (green -485; white +485; red +5VDC; black GND)
Cable (standard) 1 m (3.28 ft)
Electrolyte mod. ELESCL3
Membrane mod. MESCL3
Working flow 30 l/h
Suitable for probe PEF1, PEF1/E, PEF5, PEF23
holder mod.
Membrane cap: PPE
Material
Shaft: PVC
probe: frost-protected, dry and without electrolyte ( (5-40° C)
membrane cap: 1 year, depending on water quality
Storage
electrolyte: in original bottle, protect from sunlight (5-25°C). Expiration
date on the label.
regular control of measuring signal
change of the membrane cap: every 3-6 months
Maintenance change electrolyte: every 3-6 months
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.
Diam.: 25 mm
Dimension
Length: 241 mm
ISO 9001:2008 Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 14001:2004 Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emecpumps.com EN R1-01-18
OHSAS 18001:2007
SCL Data Sheet
Closed amperometric cells
SCL3N
SCL3N
Parameter FREE CHLORINE (INORGANIC) FOR FRESH WATER
SCL3N/0.5: 0-0.5 mg/l (0-0.5 ppm) resolution: ± 0.001
Measuring range SCL3N/2: 0-2 mg/l (0-2 ppm) resolution: ± 0.001
SCL3N/20: 0-20 mg/l (0-20 ppm) resolution: ± 0.01
Voltage supply 0/5 VDC (±10%) (10 mA)
Connection 5-pole screw connector
Measuring system membrane covered, amperometric 2-electrode system
5-9.5 pH, reduced dependance on pH value. When the pH value increases,
Ph working range
the measured signal decreases at about 10% per pH unit.
Water sample 500 µS/cm - 10000 µS/cm
conductivity
Run-in-time first start up: 1-24 h (6 hours usually) depending on water quality
Response time T90: 2 min. approx.
Zero point See Operating manual: “Probe alignment”
adjustment
Slope calibration See Operating manual: “Probe alignment” DPD1 method
Alcalinity 80 ppm
Working 1-40° C (34-104°F)
temperature
Temperature automatically, by an integrated temperature sensor
compensation
0.5 bar (7 PSI) - 5 mwc [mH2O]
Max pressure
no pressure impulses and/or vibration, no depressure
Power supply 4 wires (green -485; white +485; red +5VDC; black GND)
Cable (standard) 1 m (3.28 ft)
Electrolyte mod. ELESCL3N
Membrane mod. MESCL3
Working flow 30 l/h
Suitable for probe PEF1, PEF1/E, PEF5, PEF23
holder mod.
Material Shaft: PP/PMMA; Membrane cap: PPE
probe: frost-protected, dry and without electrolyte (5-40° C)
Storage membrane cap: used membrane cap can not be stored
electrolyte: in original bottle, protect from sunlight (5-25°C)
regular control of measuring signal
change of the membrane cap: every 3-6 months
Maintenance change electrolyte: every 3-6 months
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.
Diam.: 25 mm
Dimension
Length: 241 mm
ISO 9001:2008 Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 14001:2004 Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emecpumps.com EN R1-01-18
OHSAS 18001:2007
SCL Data Sheet
Closed amperometric cells
SCL8
SCL8
Parameter TOTAL CHLORINE
SCL8/2: 0-2 mg/l (0-2 ppm) resolution: ± 0.001
Measuring range
SCL8/20: 0-20 mg/l (0-20 ppm) resolution: ± 0.01
Voltage supply 0/5 VDC (±10%) (10 mA)
Connection 5-pole screw connector
Measuring system membrane covered, amperometric 2-electrode system
6.5-9.5 pH, reduced dependance on pH value. When the pH value increases,
Ph working range
the measured signal decreases at about 10% per pH unit.
Water sample 0.03 - 40 mS/cm
conductivity
Run-in-time first start up: 24 h approx.
Response time T90: 60 s approx.
Zero point
See Operating manual: “Probe alignment”
adjustment
Slope calibration See Operating manual: “Probe alignment” DPD4 method
Alcalinity 80 ppm
Working
1-40° C (34-104°F)
temperature
Temperature
automatically, by an integrated temperature sensor
compensation
1 bar (14.5 PSI) - 10 mwc [mH2O]
Max pressure
no pressure impulses and/or vibration, no depressure
Power supply 4 wires (green -485; white +485; red +5VDC; black GND)
Cable (standard) 1 m (3.28 ft)
Electrolyte mod. ELESCL8
Membrane mod. MESCL8/2 or MESCL8/20
Working flow 30 l/h
Suitable for probe
PEF1, PEF1/E, PEF5, PEF23
holder mod.
Material Shaft: PVC; membrane cap: PPE.
probe: frost-protected, dry and without electrolyte (5-40° C)
Storage membrane cap: used membrane cap can not be stored
electrolyte: in original bottle, protect from sunlight (5-25°C)
regular control of measuring signal
change of the membrane cap: every 3-6 months
Maintenance change electrolyte: every 3-6 months
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.
Diam.: 25 mm
Dimension
Length: 241 mm
ISO 9001:2008 Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 14001:2004 Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emecpumps.com EN R1-01-18
OHSAS 18001:2007
SCL Data Sheet
Closed amperometric cells
SCL2
Parameter CHLORINE DIOXIDE
SCL2/0.5: 0-0.5 mg/l (0-0.5 ppm) resolution: ± 0.001
Measuring range SCL2/2: 0-2 mg/l (0-2 ppm) resolution: ± 0.001
SCL2/20: 0-20 mg/l (0-20 ppm) resolution: ± 0.01
Voltage supply 0/5 VDC (±10%) (10 mA)
Connection 5-pole screw connector
Measuring system membrane covered, 2-electrode system
Ph working range 4-11 pH
Run-in-time first start up: 1-24 h approx. (usually 6 h)
Response time T90: 15 sec.. approx.
Zero point adjustment See Operating manual: “Probe alignment”
Slope calibration See Operating manual: “Probe alignment”
Alcalinity 80 ppm
Working temperature 1-40° C (34-104°F)
Temperature
automatically, by an integrated temperature sensor
compensation
1 bar (14.5 PSI) - 10 mwc [mH2O]
Max pressure
no pressure impulses and/or vibration, no depressure
Power supply 4 wires (green -485; white +485; red +5VDC; black GND)
Cable (standard) 1 m (3.28 ft)
Electrolyte mod. ELESCL2
Membrane mod. MESCL2
Working flow 30 l/h
Suitable for probe
PEF1, PEF1/E, PEF5, PEF23
holder mod.
Material Shaft: PVC; membrane cap: PPE.
probe: frost-protected, dry and without electrolyte (5-40° C)
Storage membrane cap: used membrane cap can not be stored
electrolyte: in original bottle, protect from sunlight (5-25°C)
regular control of measuring signal
change of the membrane cap: every 3-6 months
Maintenance change electrolyte: every 3-6 months
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY
DEPENDING ON WATER QUALITY.
Diam.: 25 mm
Dimension
Length: 241 mm
ISO 9001:2008 Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 14001:2004 Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emecpumps.com EN R1-01-18
OHSAS 18001:2007
SCL Data Sheet
Closed amperometric cells
SCL9
Parameter HYDROGEN PEROXIDE
Measuring range SCL9/200: 0-200 mg/l (0-200 ppm) resolution: ± 0.1
Voltage supply 0/5 VDC (±10%) - 25 mA
Connection 5-pole screw connector
Measuring system membrane covered, amperometric electrode system
Ph working range 1-11 pH
Water sample 0.05 - 5.00 mS/cm
conductivity
Run-in-time first start up: 2-6 h approx.
Response time T90: 10 min. approx.
Zero point adjustment See Operating manual: “Probe alignment”
Slope calibration See Operating manual: “Probe alignment”
Alcalinity 80 ppm
Working temperature 1-40° C (34-104°F)
Temperature
automatically, by an integrated temperature sensor
compensation
1 bar (14.5 PSI) - 10 mwc [mH2O]
Max pressure
no pressure impulses and/or vibration, no depressure
Power supply 4 wires (green -485; white +485; red +5VDC; black GND)
Cable (standard) 1 m (3.28 ft)
Electrolyte mod. ELESCL9
Membrane mod. MESCL9
Working flow 30 l/h
Suitable for probe
PEF1, PEF1/E, PEF5, PEF23
holder mod.
Material Shaft: PVC-C; membrane cap: PVDF, PVC
probe: frost-protected, dry and without electrolyte (5-40° C)
Storage membrane cap: used membrane cap can not be stored
electrolyte: in original bottle, protect from sunlight (5-25°C)
regular control of measuring signal
change of the membrane cap: every 3-6 months
Maintenance change electrolyte: every 3-6 months
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY
DEPENDING ON WATER QUALITY.
Diam.: 25 mm
Dimension
Length: 241 mm
ISO 9001:2008 Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 14001:2004 Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emecpumps.com EN R1-01-18
OHSAS 18001:2007
SCL Data Sheet
Closed amperometric cells
SCL10
Parameter OZONE
SCL10/2: 0-2 mg/l (0-2 ppm) resolution: ± 0.001
Measuring range
SCL10/20: 0-20 mg/l (0-20 ppm) resolution: ± 0.01
0/5 VDC (±10%)
Voltage supply
10 mA
Connection 5-pole screw connector
Measuring system membrane covered, amperometric 2-electrode system
Ph working range 2-11 pH
Run-in-time first start up: 1 h approx.
Response time T90: 50 sec. approx.
Zero point adjustment See Operating manual: “Probe alignment”
Slope calibration See Operating manual: “Probe alignment”
Alcalinity 80 ppm
Working temperature 1-40° C (34-104°F)
Temperature
automatically, by an integrated temperature sensor
compensation
1 bar (14.5 PSI) - 10 mwc [mH2O]
Max pressure
no pressure impulses and/or vibration, no depressure
Power supply 4 wires (green -485; white +485; red +5VDC; black GND)
Cable (standard) 1 m (3.28 ft)
Electrolyte mod. ELSCL10
Membrane mod. MSCL10
Working flow 30 l/h
Suitable for probe
PEF1, PEF1/E, PEF5, PEF23
holder mod.
Material PVC-U, stainless steel 1.4571
probe: frost-protected, dry and without electrolyte (5-40° C)
Storage membrane cap: used membrane cap can not be stored
electrolyte: in original bottle, protect from sunlight (5-25°C)
regular control of measuring signal
change of the membrane cap: every 3-6 months
Maintenance change electrolyte: every 3-6 months
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY
DEPENDING ON WATER QUALITY.
Diam.: 25 mm
Dimension
Length: 241 mm
ISO 9001:2008 Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 14001:2004 Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emecpumps.com EN R1-01-18
OHSAS 18001:2007
SCL Data Sheet
Closed amperometric cells
SCL11
Parameter PERACETIC ACID
SCL11/200: 0-200mg/l (0-200 ppm) resolution: ± 0.1
Measuring range
SCL11/2000: 0-2000 mg/l (0-2000 ppm) resolution: ± 1
0/5 VDC (±10%)
Voltage supply
10 mA
Connection 5-pole screw connector
Measuring system membrane covered, amperometric 2-electrode system
Ph working range 1-9 pH
Run-in-time first start up: 1 h approx.
Response time T90: 3 min. approx.
Zero point adjustment See Operating manual: “Probe alignment”
Slope calibration See Operating manual: “Probe alignment”
Alcalinity 80 ppm
Working temperature 1-40° C (34-104°F)
Temperature
automatically, by an integrated temperature sensor
compensation
1 bar (14.5 PSI) - 10 mwc [mH2O]
Max pressure
no pressure impulses and/or vibration, no depressure
Power supply 4 wires (green -485; white +485; red +5VDC; black GND)
Cable (standard) 1 m (3.28 ft)
Electrolyte mod. ELESCL11
Membrane mod. MESCL11
Working flow 30 l/h
Suitable for probe
PEF1, PEF1/E, PEF5, PEF23
holder mod.
Material PVC-U, stainless steel 1.4571
probe: frost-protected, dry and without electrolyte (5-40° C)
Storage membrane cap: used membrane cap can not be stored
electrolyte: in original bottle, protect from sunlight (5-25°C)
regular control of measuring signal
change of the membrane cap: every 3-6 months
Maintenance change electrolyte: every 3-6 months
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY
DEPENDING ON WATER QUALITY.
Diam.: 25 mm
Dimension
Length: 241 mm
ISO 9001:2008 Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 14001:2004 Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emecpumps.com EN R1-01-18
OHSAS 18001:2007
SCL Data Sheet
Closed amperometric cells
SBR
Parameter BROMINE
Measuring range SBR/20: 0-20mg/l (0-20 ppm) resolution: ± 0.01
0/5 VDC (±10%)
Voltage supply
10 mA
Connection 5-pole screw connector
Measuring system membrane covered, amperometric 2-electrode system
Ph working range 6.5-9.5 pH
Water sample 500 µS/cm - 10000 µS/cm
conductivity
Run-in-time first start up: 1-24 h approx. (usually 6 h)
Response time T90: 2 min. approx.
Zero point adjustment See Operating manual: “Probe alignment”
Slope calibration See Operating manual: “Probe alignment”
Alcalinity 80 ppm
Working temperature 1-40° C (34-104°F)
Temperature
automatically, by an integrated temperature sensor
compensation
0.5 bar (7 PSI) - 5 mwc [mH2O]
Max pressure
no pressure impulses and/or vibration, no depressure
Power supply 4 wires (green -485; white +485; red +5VDC; black GND)
Cable (standard) 1 m (3.28 ft)
Electrolyte mod. ELESBR
Membrane mod. MESBR
Working flow 30 l/h
Suitable for probe
PEF1, PEF1/E, PEF5, PEF23
holder mod.
Material Shaft: PVC; membrane cap: PPE
probe: frost-protected, dry and without electrolyte (5-40° C)
Storage membrane cap: used membrane cap can not be stored
electrolyte: in original bottle, protect from sunlight (5-25°C)
regular control of measuring signal
change of the membrane cap: every 3-6 months
Maintenance change electrolyte: every 3-6 months
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY
DEPENDING ON WATER QUALITY.
Diam.: 25 mm
Dimension
Length: 241 mm
ISO 9001:2008 Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 14001:2004 Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emecpumps.com EN R1-01-18
OHSAS 18001:2007
SCL Data Sheet
Closed amperometric cells
SCT
Parameter CHLORITE
Measuring range SCT: 0-2mg/l (0-2 ppm) resolution: ± 0.001
0/5 VDC (±10%)
Voltage supply
10 mA
Connection 5-pole screw connector
Measuring system membrane covered, amperometric 2-electrode system
pH range 5.5-9.5 pH
Conductivity range 0.05-5 mS/cm
Run-in-time first start up: 1-24 h approx. (usually 6 h)
Response time T90: 60 s approx.
Zero point adjustment See Operating manual: “Probe alignment”
Slope calibration See Operating manual: “Probe alignment”
Alcalinity 100 ppm
Working temperature 1-40° C (34-104°F)
Temperature
automatically, by an integrated temperature sensor
compensation
1 bar (14.5 PSI) - 10 mwc [mH2O]
Max pressure
no sbalzi di pressione, no depressione
Power supply 4 wires (green -485; white +485; red +5VDC; black GND)
Cable (standard) 1 m (3.28 ft)
Electrolyte mod. ELESCT
Membrane mod. MESCT
Working flow 30 l/h
Suitable for probe
PEF1, PEF1/E, PEF5, PEF23
holder mod.
Material Shaft: PVC; membrane cap: PPE
probe: frost-protected, dry and without electrolyte (5-40° C)
Storage membrane cap: used membrane cap can not be stored
electrolyte: in original bottle, protect from sunlight (5-25°C)
regular control of measuring signal
change of the membrane cap: every 3-6 months
Maintenance change electrolyte: every 3-6 months
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY
DEPENDING ON WATER QUALITY.
Diam.: 25 mm
Dimension
Length: 241 mm
ISO 9001:2008 Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 14001:2004 Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emecpumps.com EN R1-01-18
OHSAS 18001:2007
SCL Data Sheet
Closed amperometric cells
SCL17
SCL17
Parameter CHLORINE DIOXIDE
SCL17/0.5: 0-0.5 mg/l (0-0.5 ppm) resolution: ± 0.001
Measuring range SCL17/2: 0-2 mg/l (0-2 ppm) resolution: ± 0.001
SCL17/20: 0-20 mg/l (0-20 ppm) resolution: ± 0.001
Voltage supply 0/5 VDC (±10%) (10 mA)
Connection 5-pole screw connector
Measuring system amperometric 3-electrode system
Ph working range 5-9 pH (ref. HOCl dissociation curve)
Conductivity 0.05 µS/cm - 10 mS/cm
water sample
Run-in-time first start up: 1-24 h (6 hours usually) depending on water quality
Response time T90: 60 sec. approx.
Zero point See Operating manual: “Probe alignment”
adjustment
Slope calibration See Operating manual: “Probe alignment” - DPD1
Alcalinity 80 ppm
Working 5-70° C (41-158°F)
temperature
Temperature automatically, by an integrated temperature sensor
compensation
8 bar (116 PSI) - 80 mwc [mH2O]
Max pressure
no pressure impulses and/or vibration, no depressure
Power supply 4 wires (green -485; white +485; red +5VDC; black GND)
Cable (standard) 1 m (3.28 ft)
Electrolyte mod. ELESCL17
Working flow 40 l/h
Cleaning system selfcleaning with 3-balls systems
Suitable for probe PEF1, PEF1/E, PEF5, PEF23
holder mod.
Material shaft: PEEK; clamp ring:PPE; electrode: gold; oring: EPDM
probe: frost-protected, dry and without electrolyte (5-40° C)
Storage
electrolyte: in original bottle, protect from sunlight (5-25°C)
regular control of measuring signal
change electrolyte: every 3-6 months
Maintenance
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.
ISO 9001:2008 Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 14001:2004 Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emecpumps.com EN R1-01-18
OHSAS 18001:2007
SCL Data Sheet
Closed amperometric cells
SCL18
SCL18
Parameter FREE CHLORINE (INORGANIC)
SCL18/0.5: 0-0.5 mg/l (0-0.5 ppm) resolution: ± 0.001
Measuring range SCL18/2: 0-2 mg/l (0-2 ppm) resolution: ± 0.001
SCL18/20: 0-20 mg/l (0-20 ppm) resolution: ± 0.001
Voltage supply 0/5 VDC (±10%) (10 mA)
Connection 5-pole screw connector
Measuring system amperometric 3-electrode system
Ph working range 5-9 pH (ref. HOCl dissociation curve)
Conductivity 0.05 µS/cm - 10 mS/cm
water sample
Run-in-time first start up: 1-24 h (6 hours usually) depending on water quality
Response time T90: 60 sec. approx
Zero point See Operating manual: “Probe alignment”
adjustment
Slope calibration See Operating manual: “Probe alignment” - DPD1
Alcalinity 80 ppm
Working 5-70° C (41-158°F)
temperature
Temperature automatically, by an integrated temperature sensor
compensation
8 bar (116 PSI) - 80 mwc [mH2O]
Max pressure
no pressure impulses and/or vibration, no depressure
Power supply 4 wires (green -485; white +485; red +5VDC; black GND)
Cable (standard) 1 m (3.28 ft)
Electrolyte mod. ELESCL18
Working flow 40 l/h
Cleaning system selfcleaning with 3-balls systems
Material shaft: PEEK; clamp ring:PPE; electrode: gold; oring: EPDM
probe: frost-protected, dry and without electrolyte (5-40° C)
Storage
electrolyte: in original bottle, protect from sunlight (5-25°C)
regular control of measuring signal
change electrolyte: every 3-6 months
Maintenance
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.