EMEC SCL Data Sheet R11 19 1

The document provides a data sheet for closed amperometric cells used for measuring various chlorine compounds, including free chlorine, total chlorine, and chlorine dioxide. It details the specifications, measuring ranges, and maintenance requirements for different models such as SCL3S, SCL3N, SCL8, and SCL2. Each model is designed for specific applications and has unique features, including pH dependency and response times.

Uploaded by

kyalundziga
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
72 views12 pages

EMEC SCL Data Sheet R11 19 1

The document provides a data sheet for closed amperometric cells used for measuring various chlorine compounds, including free chlorine, total chlorine, and chlorine dioxide. It details the specifications, measuring ranges, and maintenance requirements for different models such as SCL3S, SCL3N, SCL8, and SCL2. Each model is designed for specific applications and has unique features, including pH dependency and response times.

Uploaded by

kyalundziga
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 12

SCL Data Sheet

Closed amperometric cells

• Chlorine probes for chlorine and chlorine bioxide measurement.


• Stables and reliables measurement even for low chlorine concentration.
• Low pH dependency for chlorine bioxide measurement

They are equipped with a special membrane system, except for SCL17 and
SCL 18.
SCL models work in chlorine water system.
SCL 3/S SCL2 SCL 9 SCL 11 SCL 17
The probe has to be installed into a probe holder and connected to a measuring SCL 3/N SCL 10 SCL 18
and control instrument. SCL 8
SBR
SCT

Dissolved chlorine into water may have many features:


FREE CHLORINE (ACTIVE): HOCl (hypoclorus acid).
COMBINED CHLORINE: Monochloramine, dichloramine, trichloramine (DPD4-DPD1 analysis system)
FREE ORGANIC CHLORINE: Free cholorine with isocianuric/isocianurate (DPD1 analysis system)
FREE CHLORINE INORGANIC: Free chlorine Use fo SCL3N amperometric cells is recommended (DPD1 analysis system)
TOTAL CHLORINE: Free chlorine and combined chlorine. Use of SCL8 amperometric cells is recommended (DPD4 analysis system)

MODELS
• SCL3S for free chlorine (organic and inorganic)
• SCL3N for free chlorine (inorganic)
• SCL8 for total chlorine
• SCL2 for chlorine dioxide
• SCL9 for hydrogen peroxyde
• SCL10 for ozone
• SCL11 for peracetic acid
• SBR for bromine
• SCT for Chlorite
• SCL17 for chlorine dioxide - with automatic cleaning system
• SCL18 for free chlorine (inorganic) - with automatic cleaning system

ISO 9001:2008 Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 14001:2004 Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emecpumps.com EN R1-01-18
OHSAS 18001:2007
SCL Data Sheet
Closed amperometric cells
SCL3S
SCL3S/20
Parameter FREE CHLORINE (ORGANIC)
Measuring range 0,1-20 mg/l (0,1-20 ppm)
Voltage supply 0/5 VDC (±10%) - 10 mA
Connection 5-pole screw connector
Measuring system membrane covered, amperometric potentiostatic 2-electrode system
Ph working range 5,5-9,5 pH
Sample water 30-10.000 µS/cm
conductivity
First commissioning: 6 h approx.
Run-in-time
Recommissioning: 3 h approx.
Response time T90: 2 min. approx.
Zero point see chapter “Probe alignment”
adjustment
Slope calibration see chapter “Probe alignment” DPD1 method
Alcalinity 100 ppm
Working 5-45 °C with temperature compensation
temperature
1 bar - 10 mwc [mH2O]
Max pressure
no pressure impulses and/or vibration, no depressure
Power supply 4 wires (green -485; white +485; red +5VDC; black GND)
Cable (standard) 1 m (3.28 ft)
Electrolyte mod. ELESCL3
Membrane mod. MESCL3
Working flow 30 l/h
Suitable for probe PEF1, PEF1/E, PEF5, PEF23
holder mod.
Membrane cap: PPE
Material
Shaft: PVC
probe: frost-protected, dry and without electrolyte ( (5-40° C)
membrane cap: 1 year, depending on water quality
Storage
electrolyte: in original bottle, protect from sunlight (5-25°C). Expiration
date on the label.
regular control of measuring signal
change of the membrane cap: every 3-6 months
Maintenance change electrolyte: every 3-6 months
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.
Diam.: 25 mm
Dimension
Length: 241 mm

ISO 9001:2008 Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 14001:2004 Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emecpumps.com EN R1-01-18
OHSAS 18001:2007
SCL Data Sheet
Closed amperometric cells
SCL3N
SCL3N
Parameter FREE CHLORINE (INORGANIC) FOR FRESH WATER
SCL3N/0.5: 0-0.5 mg/l (0-0.5 ppm) resolution: ± 0.001
Measuring range SCL3N/2: 0-2 mg/l (0-2 ppm) resolution: ± 0.001
SCL3N/20: 0-20 mg/l (0-20 ppm) resolution: ± 0.01
Voltage supply 0/5 VDC (±10%) (10 mA)
Connection 5-pole screw connector
Measuring system membrane covered, amperometric 2-electrode system
5-9.5 pH, reduced dependance on pH value. When the pH value increases,
Ph working range
the measured signal decreases at about 10% per pH unit.
Water sample 500 µS/cm - 10000 µS/cm
conductivity
Run-in-time first start up: 1-24 h (6 hours usually) depending on water quality
Response time T90: 2 min. approx.
Zero point See Operating manual: “Probe alignment”
adjustment
Slope calibration See Operating manual: “Probe alignment” DPD1 method
Alcalinity 80 ppm
Working 1-40° C (34-104°F)
temperature
Temperature automatically, by an integrated temperature sensor
compensation
0.5 bar (7 PSI) - 5 mwc [mH2O]
Max pressure
no pressure impulses and/or vibration, no depressure
Power supply 4 wires (green -485; white +485; red +5VDC; black GND)
Cable (standard) 1 m (3.28 ft)
Electrolyte mod. ELESCL3N
Membrane mod. MESCL3
Working flow 30 l/h
Suitable for probe PEF1, PEF1/E, PEF5, PEF23
holder mod.
Material Shaft: PP/PMMA; Membrane cap: PPE
probe: frost-protected, dry and without electrolyte (5-40° C)
Storage membrane cap: used membrane cap can not be stored
electrolyte: in original bottle, protect from sunlight (5-25°C)
regular control of measuring signal
change of the membrane cap: every 3-6 months
Maintenance change electrolyte: every 3-6 months
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.
Diam.: 25 mm
Dimension
Length: 241 mm

ISO 9001:2008 Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 14001:2004 Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emecpumps.com EN R1-01-18
OHSAS 18001:2007
SCL Data Sheet
Closed amperometric cells
SCL8
SCL8
Parameter TOTAL CHLORINE
SCL8/2: 0-2 mg/l (0-2 ppm) resolution: ± 0.001
Measuring range
SCL8/20: 0-20 mg/l (0-20 ppm) resolution: ± 0.01
Voltage supply 0/5 VDC (±10%) (10 mA)
Connection 5-pole screw connector
Measuring system membrane covered, amperometric 2-electrode system
6.5-9.5 pH, reduced dependance on pH value. When the pH value increases,
Ph working range
the measured signal decreases at about 10% per pH unit.
Water sample 0.03 - 40 mS/cm
conductivity
Run-in-time first start up: 24 h approx.
Response time T90: 60 s approx.
Zero point
See Operating manual: “Probe alignment”
adjustment
Slope calibration See Operating manual: “Probe alignment” DPD4 method
Alcalinity 80 ppm
Working
1-40° C (34-104°F)
temperature
Temperature
automatically, by an integrated temperature sensor
compensation
1 bar (14.5 PSI) - 10 mwc [mH2O]
Max pressure
no pressure impulses and/or vibration, no depressure
Power supply 4 wires (green -485; white +485; red +5VDC; black GND)
Cable (standard) 1 m (3.28 ft)
Electrolyte mod. ELESCL8
Membrane mod. MESCL8/2 or MESCL8/20
Working flow 30 l/h
Suitable for probe
PEF1, PEF1/E, PEF5, PEF23
holder mod.
Material Shaft: PVC; membrane cap: PPE.
probe: frost-protected, dry and without electrolyte (5-40° C)
Storage membrane cap: used membrane cap can not be stored
electrolyte: in original bottle, protect from sunlight (5-25°C)
regular control of measuring signal
change of the membrane cap: every 3-6 months
Maintenance change electrolyte: every 3-6 months
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.
Diam.: 25 mm
Dimension
Length: 241 mm

ISO 9001:2008 Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 14001:2004 Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emecpumps.com EN R1-01-18
OHSAS 18001:2007
SCL Data Sheet
Closed amperometric cells
SCL2
Parameter CHLORINE DIOXIDE
SCL2/0.5: 0-0.5 mg/l (0-0.5 ppm) resolution: ± 0.001
Measuring range SCL2/2: 0-2 mg/l (0-2 ppm) resolution: ± 0.001
SCL2/20: 0-20 mg/l (0-20 ppm) resolution: ± 0.01
Voltage supply 0/5 VDC (±10%) (10 mA)
Connection 5-pole screw connector
Measuring system membrane covered, 2-electrode system
Ph working range 4-11 pH
Run-in-time first start up: 1-24 h approx. (usually 6 h)
Response time T90: 15 sec.. approx.
Zero point adjustment See Operating manual: “Probe alignment”
Slope calibration See Operating manual: “Probe alignment”
Alcalinity 80 ppm
Working temperature 1-40° C (34-104°F)
Temperature
automatically, by an integrated temperature sensor
compensation
1 bar (14.5 PSI) - 10 mwc [mH2O]
Max pressure
no pressure impulses and/or vibration, no depressure
Power supply 4 wires (green -485; white +485; red +5VDC; black GND)
Cable (standard) 1 m (3.28 ft)
Electrolyte mod. ELESCL2
Membrane mod. MESCL2
Working flow 30 l/h
Suitable for probe
PEF1, PEF1/E, PEF5, PEF23
holder mod.
Material Shaft: PVC; membrane cap: PPE.
probe: frost-protected, dry and without electrolyte (5-40° C)
Storage membrane cap: used membrane cap can not be stored
electrolyte: in original bottle, protect from sunlight (5-25°C)
regular control of measuring signal
change of the membrane cap: every 3-6 months
Maintenance change electrolyte: every 3-6 months
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY
DEPENDING ON WATER QUALITY.
Diam.: 25 mm
Dimension
Length: 241 mm

ISO 9001:2008 Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 14001:2004 Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emecpumps.com EN R1-01-18
OHSAS 18001:2007
SCL Data Sheet
Closed amperometric cells
SCL9
Parameter HYDROGEN PEROXIDE
Measuring range SCL9/200: 0-200 mg/l (0-200 ppm) resolution: ± 0.1
Voltage supply 0/5 VDC (±10%) - 25 mA
Connection 5-pole screw connector
Measuring system membrane covered, amperometric electrode system
Ph working range 1-11 pH
Water sample 0.05 - 5.00 mS/cm
conductivity
Run-in-time first start up: 2-6 h approx.
Response time T90: 10 min. approx.
Zero point adjustment See Operating manual: “Probe alignment”
Slope calibration See Operating manual: “Probe alignment”
Alcalinity 80 ppm
Working temperature 1-40° C (34-104°F)
Temperature
automatically, by an integrated temperature sensor
compensation
1 bar (14.5 PSI) - 10 mwc [mH2O]
Max pressure
no pressure impulses and/or vibration, no depressure
Power supply 4 wires (green -485; white +485; red +5VDC; black GND)
Cable (standard) 1 m (3.28 ft)
Electrolyte mod. ELESCL9
Membrane mod. MESCL9
Working flow 30 l/h
Suitable for probe
PEF1, PEF1/E, PEF5, PEF23
holder mod.
Material Shaft: PVC-C; membrane cap: PVDF, PVC
probe: frost-protected, dry and without electrolyte (5-40° C)
Storage membrane cap: used membrane cap can not be stored
electrolyte: in original bottle, protect from sunlight (5-25°C)
regular control of measuring signal
change of the membrane cap: every 3-6 months
Maintenance change electrolyte: every 3-6 months
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY
DEPENDING ON WATER QUALITY.
Diam.: 25 mm
Dimension
Length: 241 mm

ISO 9001:2008 Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 14001:2004 Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emecpumps.com EN R1-01-18
OHSAS 18001:2007
SCL Data Sheet
Closed amperometric cells
SCL10
Parameter OZONE
SCL10/2: 0-2 mg/l (0-2 ppm) resolution: ± 0.001
Measuring range
SCL10/20: 0-20 mg/l (0-20 ppm) resolution: ± 0.01
0/5 VDC (±10%)
Voltage supply
10 mA
Connection 5-pole screw connector
Measuring system membrane covered, amperometric 2-electrode system
Ph working range 2-11 pH
Run-in-time first start up: 1 h approx.
Response time T90: 50 sec. approx.
Zero point adjustment See Operating manual: “Probe alignment”
Slope calibration See Operating manual: “Probe alignment”
Alcalinity 80 ppm
Working temperature 1-40° C (34-104°F)
Temperature
automatically, by an integrated temperature sensor
compensation
1 bar (14.5 PSI) - 10 mwc [mH2O]
Max pressure
no pressure impulses and/or vibration, no depressure
Power supply 4 wires (green -485; white +485; red +5VDC; black GND)
Cable (standard) 1 m (3.28 ft)
Electrolyte mod. ELSCL10
Membrane mod. MSCL10
Working flow 30 l/h
Suitable for probe
PEF1, PEF1/E, PEF5, PEF23
holder mod.
Material PVC-U, stainless steel 1.4571
probe: frost-protected, dry and without electrolyte (5-40° C)
Storage membrane cap: used membrane cap can not be stored
electrolyte: in original bottle, protect from sunlight (5-25°C)
regular control of measuring signal
change of the membrane cap: every 3-6 months
Maintenance change electrolyte: every 3-6 months
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY
DEPENDING ON WATER QUALITY.
Diam.: 25 mm
Dimension
Length: 241 mm

ISO 9001:2008 Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 14001:2004 Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emecpumps.com EN R1-01-18
OHSAS 18001:2007
SCL Data Sheet
Closed amperometric cells
SCL11
Parameter PERACETIC ACID
SCL11/200: 0-200mg/l (0-200 ppm) resolution: ± 0.1
Measuring range
SCL11/2000: 0-2000 mg/l (0-2000 ppm) resolution: ± 1
0/5 VDC (±10%)
Voltage supply
10 mA
Connection 5-pole screw connector
Measuring system membrane covered, amperometric 2-electrode system
Ph working range 1-9 pH
Run-in-time first start up: 1 h approx.
Response time T90: 3 min. approx.
Zero point adjustment See Operating manual: “Probe alignment”
Slope calibration See Operating manual: “Probe alignment”
Alcalinity 80 ppm
Working temperature 1-40° C (34-104°F)
Temperature
automatically, by an integrated temperature sensor
compensation
1 bar (14.5 PSI) - 10 mwc [mH2O]
Max pressure
no pressure impulses and/or vibration, no depressure
Power supply 4 wires (green -485; white +485; red +5VDC; black GND)
Cable (standard) 1 m (3.28 ft)
Electrolyte mod. ELESCL11
Membrane mod. MESCL11
Working flow 30 l/h
Suitable for probe
PEF1, PEF1/E, PEF5, PEF23
holder mod.
Material PVC-U, stainless steel 1.4571
probe: frost-protected, dry and without electrolyte (5-40° C)
Storage membrane cap: used membrane cap can not be stored
electrolyte: in original bottle, protect from sunlight (5-25°C)
regular control of measuring signal
change of the membrane cap: every 3-6 months
Maintenance change electrolyte: every 3-6 months
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY
DEPENDING ON WATER QUALITY.
Diam.: 25 mm
Dimension
Length: 241 mm

ISO 9001:2008 Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 14001:2004 Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emecpumps.com EN R1-01-18
OHSAS 18001:2007
SCL Data Sheet
Closed amperometric cells
SBR
Parameter BROMINE
Measuring range SBR/20: 0-20mg/l (0-20 ppm) resolution: ± 0.01
0/5 VDC (±10%)
Voltage supply
10 mA
Connection 5-pole screw connector
Measuring system membrane covered, amperometric 2-electrode system
Ph working range 6.5-9.5 pH
Water sample 500 µS/cm - 10000 µS/cm
conductivity
Run-in-time first start up: 1-24 h approx. (usually 6 h)
Response time T90: 2 min. approx.
Zero point adjustment See Operating manual: “Probe alignment”
Slope calibration See Operating manual: “Probe alignment”
Alcalinity 80 ppm
Working temperature 1-40° C (34-104°F)
Temperature
automatically, by an integrated temperature sensor
compensation
0.5 bar (7 PSI) - 5 mwc [mH2O]
Max pressure
no pressure impulses and/or vibration, no depressure
Power supply 4 wires (green -485; white +485; red +5VDC; black GND)
Cable (standard) 1 m (3.28 ft)
Electrolyte mod. ELESBR
Membrane mod. MESBR
Working flow 30 l/h
Suitable for probe
PEF1, PEF1/E, PEF5, PEF23
holder mod.
Material Shaft: PVC; membrane cap: PPE
probe: frost-protected, dry and without electrolyte (5-40° C)
Storage membrane cap: used membrane cap can not be stored
electrolyte: in original bottle, protect from sunlight (5-25°C)
regular control of measuring signal
change of the membrane cap: every 3-6 months
Maintenance change electrolyte: every 3-6 months
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY
DEPENDING ON WATER QUALITY.
Diam.: 25 mm
Dimension
Length: 241 mm

ISO 9001:2008 Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 14001:2004 Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emecpumps.com EN R1-01-18
OHSAS 18001:2007
SCL Data Sheet
Closed amperometric cells
SCT
Parameter CHLORITE
Measuring range SCT: 0-2mg/l (0-2 ppm) resolution: ± 0.001
0/5 VDC (±10%)
Voltage supply
10 mA
Connection 5-pole screw connector
Measuring system membrane covered, amperometric 2-electrode system
pH range 5.5-9.5 pH
Conductivity range 0.05-5 mS/cm
Run-in-time first start up: 1-24 h approx. (usually 6 h)
Response time T90: 60 s approx.
Zero point adjustment See Operating manual: “Probe alignment”
Slope calibration See Operating manual: “Probe alignment”
Alcalinity 100 ppm
Working temperature 1-40° C (34-104°F)
Temperature
automatically, by an integrated temperature sensor
compensation
1 bar (14.5 PSI) - 10 mwc [mH2O]
Max pressure
no sbalzi di pressione, no depressione
Power supply 4 wires (green -485; white +485; red +5VDC; black GND)
Cable (standard) 1 m (3.28 ft)
Electrolyte mod. ELESCT
Membrane mod. MESCT
Working flow 30 l/h
Suitable for probe
PEF1, PEF1/E, PEF5, PEF23
holder mod.
Material Shaft: PVC; membrane cap: PPE
probe: frost-protected, dry and without electrolyte (5-40° C)
Storage membrane cap: used membrane cap can not be stored
electrolyte: in original bottle, protect from sunlight (5-25°C)
regular control of measuring signal
change of the membrane cap: every 3-6 months
Maintenance change electrolyte: every 3-6 months
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY
DEPENDING ON WATER QUALITY.
Diam.: 25 mm
Dimension
Length: 241 mm

ISO 9001:2008 Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 14001:2004 Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emecpumps.com EN R1-01-18
OHSAS 18001:2007
SCL Data Sheet
Closed amperometric cells
SCL17
SCL17
Parameter CHLORINE DIOXIDE
SCL17/0.5: 0-0.5 mg/l (0-0.5 ppm) resolution: ± 0.001
Measuring range SCL17/2: 0-2 mg/l (0-2 ppm) resolution: ± 0.001
SCL17/20: 0-20 mg/l (0-20 ppm) resolution: ± 0.001
Voltage supply 0/5 VDC (±10%) (10 mA)
Connection 5-pole screw connector
Measuring system amperometric 3-electrode system
Ph working range 5-9 pH (ref. HOCl dissociation curve)
Conductivity 0.05 µS/cm - 10 mS/cm
water sample
Run-in-time first start up: 1-24 h (6 hours usually) depending on water quality
Response time T90: 60 sec. approx.
Zero point See Operating manual: “Probe alignment”
adjustment
Slope calibration See Operating manual: “Probe alignment” - DPD1
Alcalinity 80 ppm
Working 5-70° C (41-158°F)
temperature
Temperature automatically, by an integrated temperature sensor
compensation
8 bar (116 PSI) - 80 mwc [mH2O]
Max pressure
no pressure impulses and/or vibration, no depressure
Power supply 4 wires (green -485; white +485; red +5VDC; black GND)
Cable (standard) 1 m (3.28 ft)
Electrolyte mod. ELESCL17
Working flow 40 l/h
Cleaning system selfcleaning with 3-balls systems
Suitable for probe PEF1, PEF1/E, PEF5, PEF23
holder mod.
Material shaft: PEEK; clamp ring:PPE; electrode: gold; oring: EPDM
probe: frost-protected, dry and without electrolyte (5-40° C)
Storage
electrolyte: in original bottle, protect from sunlight (5-25°C)
regular control of measuring signal
change electrolyte: every 3-6 months
Maintenance
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.

ISO 9001:2008 Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 14001:2004 Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emecpumps.com EN R1-01-18
OHSAS 18001:2007
SCL Data Sheet
Closed amperometric cells
SCL18
SCL18
Parameter FREE CHLORINE (INORGANIC)
SCL18/0.5: 0-0.5 mg/l (0-0.5 ppm) resolution: ± 0.001
Measuring range SCL18/2: 0-2 mg/l (0-2 ppm) resolution: ± 0.001
SCL18/20: 0-20 mg/l (0-20 ppm) resolution: ± 0.001
Voltage supply 0/5 VDC (±10%) (10 mA)
Connection 5-pole screw connector
Measuring system amperometric 3-electrode system
Ph working range 5-9 pH (ref. HOCl dissociation curve)
Conductivity 0.05 µS/cm - 10 mS/cm
water sample
Run-in-time first start up: 1-24 h (6 hours usually) depending on water quality
Response time T90: 60 sec. approx
Zero point See Operating manual: “Probe alignment”
adjustment
Slope calibration See Operating manual: “Probe alignment” - DPD1
Alcalinity 80 ppm
Working 5-70° C (41-158°F)
temperature
Temperature automatically, by an integrated temperature sensor
compensation
8 bar (116 PSI) - 80 mwc [mH2O]
Max pressure
no pressure impulses and/or vibration, no depressure
Power supply 4 wires (green -485; white +485; red +5VDC; black GND)
Cable (standard) 1 m (3.28 ft)
Electrolyte mod. ELESCL18
Working flow 40 l/h
Cleaning system selfcleaning with 3-balls systems
Material shaft: PEEK; clamp ring:PPE; electrode: gold; oring: EPDM
probe: frost-protected, dry and without electrolyte (5-40° C)
Storage
electrolyte: in original bottle, protect from sunlight (5-25°C)
regular control of measuring signal
change electrolyte: every 3-6 months
Maintenance
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.

ISO 9001:2008 Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 14001:2004 Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emecpumps.com EN R1-01-18
OHSAS 18001:2007