Operation Manuals
Large Capacity Refrigerated
Centrifuge
Model:DL-6M
Thanks for using Drawell centrifuge! In order to get quick and sincere
service, please read the operation manuals carefully before you operate.
Contents
Safety Notices ................................................................................................................ 1
1. General Introduction ................................................................................................. 5
1.1 Product Name and Classify ................................................................................5
1.2 Application and Scope ........................................................................................5
1.3 Centrifugal Theory .............................................................................................. 5
2. Main Technical Parameters ...................................................................................... 5
3. Structure Characteristics ..........................................................................................7
3.1 Structure ............................................................................................................. 7
3.2 Main Frame Shap ............................................................................................ 7
3.3 Drive System ...................................................................................................... 8
3.4 Refrigeration System ..........................................................................................8
3.5 Electric Control Part ........................................................................................... 8
4. Installation and Adjustment ..................................................................................... 9
4.1 Unpacking of package ........................................................................................9
4.2 Installation Requirements ...................................................................................9
4.3 Adjusting Requirements ................................................................................... 10
5. operation ...................................................................................................................12
5.1 Safety Notice .................................................................................................... 12
5.2 Preparation and Installation of the Rotor ......................................................... 14
5.3 Operation Procedures & Directions ................................................................ 15
6.Maintenance & Trouble Shooting ........................................................................... 16
6.1 maintenance ..................................................................................................... 16
6.2 Trouble Shooting .............................................................................................. 17
7. Transportation and Storage ................................................................................... 18
7.1 Transportation ...................................................................................................18
7.2 Storage ............................................................................................................. 19
Warranty ........................................................................................................................20
Quality Certificate ........................................................................................................ 21
Packing List .................................................................................................................. 22
Warranty Card .............................................................................................................. 23
Service Records ...........................................................................................................23
Safety Notices
We appreciate your business with Drawell. To prevent any potential accident,
please operate centrifuges according to the following safety protocol.
Safety in installation and maintenance
* The electricity should be cut off before repairing. Please take care if
instrument must be repaired with electricity on. Before opening any lids,
please cut off the electricity and wait for 4 minutes.
* Make sure for placing centrifuge on a steady floor or table and adjusting the
level for normal operation.
* The power supply of centrifuge is the type of nameplate. If you have any
problem of the type of power, please consult the dealer and power supply
bureau in your place.
* As safety measures, the centrifuge equipped with a single-phase
three-phase cable, and the longer cable connects with casing as the protection
grounding. If your sockets hasn’t protection grounding, please contact your
electrician to replace the original socket without allowing security role of
single-phase with three-wire power cable to be invalid.
* When replacing parts, the serviceman should use parts manufacturer
required with the same function of original parts. The unrecognized alternative
parts many cause fire, electric shock or other hazards.
* The room should be installed an air-conditioner for good refrigerating effect
when temperature is higher than 35.℃.
* Please keep cleaning of the installation site for the centrifuge. Pay attention
to the blockage of condenser, it should be cleaned once in three months for
guarantee the refrigerating effect.
* Don’t unpack and repair instrument by oneself. Please contact with our
technicians for necessary matter.
Attention for fireproofing
* Small breaker (Air switch) can protect the circuit when centrifuge
1
over-currents. Please use breaker (air switch) with same specification and
model when replacing it.
* The centrifuge is not designed for the flammable and explosive matters. You
cannot make centrifugal for the matter (chloroform, acetaldehyde), and cannot
put the matters in the centrifuge or store them around centrifuge within 30cm.
Attention for the safe operation
* Please use the rotor and accessories which are designed for this centrifuge.
* Please check rotor, bucket and other accessories carefully before operation.
* Input the right rotor number before operation for not exceeding of its max
speed.
Attention: The max speed of rotor, buckets and accessories may be
different.
* Make sure that rotor can be installed on main spindle correctly for mot
damaging the spindle part.
* Put all buckets (the same model) in rotor before operation for not exceeding
its max imbalance volume, close the rotor lid then.
* Make sure is the centrifugal chamber is cleaned before operation.
* Don’t reduce the speed by hand when it is rotating.
* Don’t open door lid when rotor is rotating
* Don’t move centrifuge lid when rotor is rotating.
* Anything is not allowed to enter into centrifuge during operation.
* If tube broken in centrifugal chamber, please clean out the fragments before
next operation.
* If there is water condensing in centrifugal chamber, please wipe up it with
clean cloth.
* The container with liquid should not be put around centrifuge; the electric
parts may be damaged for water spraying in.
* If centrifuge is irradiated by ultraviolet radiation for long time, the outer skin
may be faded. Clothing should be covered for avoiding irradiation.
* This centrifuge must not be used to separate the sample which producing
flammable and explosive gas. The stock around centrifuge is also forbidden.
* It is forbidden to operate the rotor exceed of its max speed. If user has not
2
set rotor type according with the number on rotor lid, all damages caused by
above misoperation should be borne by user.
* please wipe up the rotor with neutral cleaning mixture after taking out for
avoiding corrosion and then stock it at dry, ventilated place. Central hole in
rotor should be smeared some lubricating grease for protection.
* All action on rotor should be soft, and rotor should be taken out vertically for
protecting the shaft.
* Please wipe up the water in centrifugal chamber if it is leaving unused. Some
lubricating grease should be smeared on drive shaft for protection.
* The interval of between two operations is no less than 5 minutes for avoiding
damage of compressor.
* If rotor damage occurring, allowable space to move for centrifuge is 30cm. *
Anything should not be put around centrifuge within 30cm, especially for
flammable and explosive matter.
* It is not allowed to lean on the centrifuge when it is running. Non-staff are not
allowed to stay in the operation area.
* When unpacking rotor and replacing rotor accessory and sample under
low-temperature, the operator must adapt protection actions for avoiding
tissue frostbiting.
* Centrifuge must be stopped when ant abnormal phenomena occurring or
error code displaying, then you should inform our company or service agent
soon.
Attention for safety in chemistry and biology
The routine manipulation may include all kinds of liquid and test sample which
may be the disease-causing and poisonous matters. The centrifuge can not be
manipulated with the matters unless you have taken the protection measures.
* Please pay attention to the liquid explanation in container before operation.
* Be careful of the infectious liquid. If not understanding the provided sample.
Please at least make sure that they aren’t microbe. You must make further
aerosol protection (the suspended speck aren’t in the air, such as smoke and
fog). For the easier disease-causing hepatitis virus (B or C), HIV, atypical
myco-bacteria and the fungus of physical structure.
3
* The infectious samples must be manipulated according to the manipulating
program and method in lab for avoiding expanding disease.
* The spilt matters may produce aerosol (the suspended speck in the air, such
as smoke and fog.), so please pay attention to the warning about aerosol ( the
suspended speck in the air, such as smoke and fog.)
* Only use the suitable rotor and adapter. The centrifuge can not be
manipulated with disease-causing, poisonous or radioactive matters unless
you substance (according to the “Laboratory Biosafety Manual” of the World
Health Organization.) is manipulated, the air safety level must be monitored.
The superior substances require a higher level of safe protection
measurements.
* You should follow the environment security and protection requirements
when destroying the discarded liquid.
* Users have responsibility to clean ans sterilize centrifuge and accessory if
you need maintenance service from our company.
In a word, centrifuge is a high-risk common equipment in the lab which
refers to safety of people. Correct operation can avoid and reduce
accidents.
4
1. General Introduction
1.1 Product Name and Classify
Floor-type Large capacity Refrigerated Centrifuge DL-6M
It belongs to Class I of medical laboratory equipment according to the
management of “The Classified Catalog of Medical Devices”.
It belongs to Class I of ordinary equipment with protective grounding
according to the protection against electric shock.
1.2 Application and Scope
The centrifuge makes use of centrifugal theory to separate, purify and
condense various mixture liquid or emulsion with tiny particle and similar
density. It can be widely used in hospitals, blood banks, pharmaceutical and
biology, medicine, agricultural science laboratory and production units. Under
the cooling condition, it makes separation, sedimentation and concentration
for blood products, biological products, chemical products, pharmaceutical
products. This machine is with large centrifugal force,large capacity. It is the
routine equipment of separation and production.The machine used AC
induction motor, frequency control, microprocessor control, touch keys and
digital display panel is a new generation of models.
1.3 Centrifugal Theory
Centrifuge makes use of the strong RCF that centrifuge rotates with high
speed in order to accelerate the setting speed of the granule in the liquid, and
separate, condense and purify the matters with different settling coefficient
and density.
2. Main Technical Parameters
5
Table I
Model DL-6M
Max Speed 6000r/min
Max RCF 5120xg
Max Capacity 6x500ml
Speed Accuracy ±20r/min
Temp Accuracy ±1°C
Temp Range -20°C~+40°C
Acc/Dec Rate Grade 0-9
Timer Range 0~99min
Refrigeration System High-performance Compressor, CFC-Free Refrigeration
Control & Drive Frequency Motor/Microprocessor Control/Direct Drive
Noise ≤ 65dB (A)
Power Supply AC220V 50HZ 30A
Dimensions 740×620×780mm(L*W*H)
Net Weight 180 kg
Gross Weight 220kg
Rotors for option (According to the customer's request to process various
rotor)
Table 2
Max Speed Max RCF
Rotor No. Capacity
(r/min) (×g)
No.1 Angle Rotor 6000 5120 12×10/15ml
4390 4×750ml
No.2 Swing Rotor 4000
3780 4×500ml
No.3 Swing Rotor 4000 3600 4×250ml
No.4 Swing Rotor 4000 3940 4×500ml (blood bag)
No.5 Swing Rotor 4000 3780 6×500ml
4000 3440 72×5ml
No.6 Swing Rotor
4000 2300 148/108×7ml
No.7 Micro Plates Rotor 4000 2300 2×2×96 holes
8×50ml
No.8 Swing Rotor 4000 2810 32×10/15ml
16×15ml
No.9 Swing Rotor 5000 4390 4×100ml
No.10 Swing Rotor 4000 3360 8×100ml
Basic formula for centrifugal technology theory :
6
2 r = Rotor radius
rpm
Relative Centrifugal Force RCF=1.12r
1000 rpm = Speed
1 δr ω = Acceleration
Sedimentation Coefficient S
ω2 r δt k = Rotor k-factor
Test Time t = k/S S = Sedimentation
Coefficient
3. Structure Characteristics
3.1 Structure
This centrifuge is composed of main machine, rotor and other accessories
(by user’s option). And main machine is composed of machine frame, door lid,
control panel, drive system, refrigerating system and electric control system.
3.2 Main Frame Shap
7
When you need check motor, control board or other parts’ working status,
please loosen screws below the front end plate firstly and then remove it.
3.3 Drive System
Direct drive system. Frequency conversion motor with high torque is
equipped with special shocking absorber for reducing vibration and noise.
3.4 Refrigeration System
This centrifuge is equipped High-performance compressor, CFC-free
refrigeration system. And with two circuit control and start/stop solenoid valve
for good efficiency of heating and refrigerating.
3.5 Electric Control Part
System control is adopted 8-digital CPU. Through sensors CPU detects the
temperature in centrifugal chamber, speed and fault signal etc., and compare
them with the setup parameters such as temperature and speed, then deal
with them correspondingly.
8
4. Installation and Adjustment
4.1 Unpacking of package
User should check the appearance of packing case after receiving, and the
goods should not be happened with sharp hitting and upending during delivery.
Please contact with forwarding agent and inform our company immediately if
occurring above accidents.
Attention : The idler wheels of this centrifuge can be only moved a short
distance on smooth ground for avoiding damage. Be sure to check
accessories and packing list after unpacking the case.
4.2 Installation Requirements
Warning:This centrifuge should be installed indoor with steady ground, no
conducting dust, no corrosive gas, no strong shocking source and no direct
sunlight.
Space requirement:
The distance between centrifuge and wall should be kept 3cm at least for air
9
circulation. The room temperature should not exceed 35 for the best.
Power requirements:
Warning: The ground wire should be connected properly in case of
electric shock.
As for measures, this centrifuge is adopted with single phase and three
wires plug to make sure of grounding. If your plug can not plug in the mains
jack, please contact the electrician to replay your main jack. Make sure that
the centrifuge is grounded properly.
Installation requirements:
Two adjusting screws are located on 2 corners of base board, four feet can
be adjusted on floor and universal wheels can be raised 3-5mm by spinning
them. Two truckles and two screwing feet should be adjusted until stressed
equably.
4.3 Adjusting Requirements
Checking the goods:
Check the packing list the and make sure if damage is occurred during
delivery.
Installation , adjustment and maintenance should all be guided or operated by
professional technicians.
Door lid opening:
After installing the main machine, check the connection of all connectors
and fuses first, and turn on the main switch then. When the screen display
model/Logo, and then the main specifications will be showed. Pull the latch on
the door lid lock device by manually, and the door lid gradually opens at the
holding of gas springs polar and your hands,
Rotor installation:
Rotor should be checked carefully before using. It is forbidden to use rotor
10
and buckets with crack and corrosion spots (especially the buckets bottom of
swing rotor). All accessories should not be used if it is exceeding of its service
life. The conical surface of driving shaft should be smeared with some
lubricating grease if the main spindle is leaving unused.
Warning: The conical surface of rotor and driving shaft must be connected
closely.
As shown in the following figure: Firstly clasp the symmetric spoke of rotor,
and put it on the rotor base vertically and placidly while the rotor bottom’s
location cone aims at the location cone of drive shaft; then put the flat gasket
on the lead screw and screw down the cap nut. At last, hold up the rotor
steadily (not shacking) by one hand, and put the solid wrench on the cap nut
by another hand. In the mean while turn the wrench for 3 to 6 rounds in
clockwise direction until the rotor connects the drive shaft closely.
Rotor disassembly: (After finishing adjust or centrifugal)
First clasp the spoke of rotor with one hand, and put the solid wrench on the
cap nut by another hand. In the mean while turn the wrench for 3 to 6 round in
counter clock wise direction to loose the cap nut until the rotor can be
separated from driving shaft. If the rotor is not taken out from the drive shaft for
a lone time and blocked, you can beat the surrounding spoke of rotor several
times lightly with wooden or rubber hammer. Then clasp the symmetric spoke
11
of rotor and lift slightly by both hands, and put the rotor on the flat table and
ground covered with soft pad.
Warning: All actions on the rotor must be soft for avoiding the damage of
driving shafting.
Adjusting level
Take out the gradienter after rotor installation, open lid of centrifugal
chamber and then put it on the rotor bracket. Adjust 2 leveling screws (on the
front of baseboard) until the bubbles of any directions are all located in center.
Warning: Adjust the screws for instrument leveling; reliable installation
can reduce rotor damage.
Attention: Make sure that 2back wheels and 2 adjusting screws are all
stressed, it is forbidden for hanging. Installation, adjustment and
maintenance must be worked by professional technician..
5. operation
5.1 Safety Notice
When the centrifuge is under high-speed operation, a small mistake of
operation will cause strong damage of the centrifuge. In order to operate the
12
centrifuge safely, operators should read the instruction manual carefully, and be
careful with the following things:
Performance index for main machine
1.Pls make sure the machine and rotor balance before operation.
2. Please pre-cooling for more than half hour (Just turn the power on)before
operation so that could gain your condition.
3. Heating the centrifuge chamber to 25℃ to 30℃ for 10 minutes after using (if
under low temp.) to keep the chamber and Motor shaft DRY so as to extend the
life of the machine.
1. Forbidden for Artificial over speed
Artificial over speed is a serious violation of operating procedures, very
easy to occur personal equipment accidents. Please don’t run the centrifuge
over the max speed of the rotor.
2. Don’t start the centrifuge machine under these conditions:
a. The rotor is unbalance.
b. You have not adjust the level of the centrifuge
3. Please don’t change or set the fixed value of the inverter.
4. The service life of the rotor: The service life of the rotor is 5 years. The use
of cumulative number is 3000 times, the accumulated use time is 2000 hours.
One of the three conditions is reached; the duration of rotor ‘use is invalid.
Prohibit using the rotor with exceeding service life.
5. Please check the rotors if had hair line before use every time. The rotors will
crack if using cracked and corrosion sign rotors.
Check rotor carefully before each operation.
Never use rotor with cracking, corrosion sign, and never use expired rotor.
When maintenance, please take off the plug, wait at least 3mins, then open
the front cover, otherwise, very easy to occur electric shock.
The rotor should be equipped on the rotor block correctly, make sure transfer
pin gather in upon each other.
6. Instrument repair, you should remove the power plug, wait at least 3
minutes, and then open the front cover, or prone to electric shock.
13
Note:Separation of the radioactive, toxic or viral substances,
response turned around, try cup, hanging baskets, test adapters and
other steam sterilization and purification methods in Table 3
Table 3 Time- Temperature Conditions
Absolute Corresponding steam temperature Minimum hold
Pressure Rated temperature Range/℃ time/
kpa /℃ min
225 136.0 134-138 3
150 127.5 126-129 10
115 122.5 121-124 15
75 116.5 115-118 30
Note: The minimum hold time is purification time under temperatures
5.2 Preparation and Installation of the Rotor
1). Choose the rotor and centrifuge tubes as the sample.
2). Use the double pan balance to balance after the tubes with sample. The
difference cant be over 0.5g. Then put it into the centrifugal cup of the rotor in
symmetry. The rotor lid should be locked closely.
3). The pipe rack of the swing rotor should be put on the rotor block according
to the number and tubes should be put in in symmetry. Do not allow the weight
of the pipe rack in symmetry not the same with the other pipe rack in
symmetry.( if the sample is in low, use other liquid instead).
4). Open the centrifuge chamber,clean it.
5). Put the rotor into the rotor block reposefully.Pay attention to make transfer
pin in the middle of the rotor put into the transfer pin' blank of the rotor block.
Make sure the rotor completely fall,or it is dangerous. Then close the lid of the
centrifuge chamber.
14
5.3 Operation Procedures & Directions
Touch Panel
1-Accelerate grade 2- Deceleration Grade 3-Running speed 4-Setting speed
5-Setting RCF 6-running time 7-Setting Time 8-Running RCF 9-Run/stop state
10-Rotor No. 11-Program No. 12-Temperature 13-Set 14-increase button
15-Decrease Grade 16-Running indicate light 17-Stop Indicate Light 19-Stop Key
20-Door open button
a) Door Open: Press the “open door” button, when the it stop running. When
door opening , indicator lightening.
b) Running: Press “Start/Stop” button, when the door is closed, centrifuge
running and indicator light turns green.
c) Stop: Press “Start/Stop” button, centrifuge will be stop and indicator light turns
red.
d) Speed set: Press”Set” and touch the “speed set” place on the screen, then
adjust the speed by pressing “▲”、“▼”button, it will confirmed itself if no operate
more than 5 seconds.
e) RCF Set: Press”Set” and touch the “RCF Set” place on the screen, then adjust
the RCF by pressing “▲”、“▼”button, it will confirmed itself if no operate more
than 5 seconds.
15
f) Time Set: Press”Set” and touch the “Time set” place on the screen, then adjust
the speed by pressing “▲”、“▼”button, it will confirmed itself if no operate more
than 5 seconds.
g) Temperature Set:Press”Set” and touch the “Temperature set” place on the
screen, then adjust the speed by pressing “▲”、“▼”button, it will confirmed itself
if no operate more than 5 seconds
h) Acc/Dec Set: Press”Set” and touch the “Acc/Dec set” place on the screen,
then adjust the speed by pressing “▲”、“▼”button, it will confirmed itself if no
operate more than 5 seconds
i) Rotor Number Set: Press”Set” and touch the “Rotor Number” place on the
screen, then adjust the speed by pressing “▲”、“▼”button, it will confirmed itself
if no operate more than 5 seconds
j) Parameters Set: Click the parameters that you need to set, and then set as the
above instruction.
6.Maintenance & Trouble Shooting
We try to reduce the complex work of maintenance and checking in the
centrifuge’s design. But operator should also maintain daily according to the
requirements of “Instruction Manual” for a long-term, safe and efficient work.
6.1 maintenance
1) Maintenance of the centrifuge chamber
Clean the centrifugal chamber time, take out rotor and then scrubbed by
cleanly cloth. Put the silica gel bag into chamber if the centrifuge will not be
used for a long time to absorb humidity and avoid rust.
2) Maintenance of the rotor
Rotor block can not bump. Do not damage the cone. Clean carefully rotor
block and rotor cone with soft cloth before the rotor put in and coated with a
little Vaseline or other grease medical.
3) Maintenance of the control box and work surface
Control box and work surfaces must be kept clean. Rotor centrifuge tube
16
and other tools of centrifuge may not be placed on the panel to prevent
scratches. Control box and work surface can be only used with a soft cloth and
mild detergent.
4) Maintenance of drive shaft
The drive shaft must not be hit for conical surface/cone protection. Use soft
cloth to clean the drive shaft and taper hole before rotor installation, and then
smear some Vaseline or other lubrication grease.
6.2 Trouble Shooting
1) Common troubleshooting
A. The rotor does not work.
a. If the power light is off, check if the fuse is burned out, whether the power
contact is good.
b. Whether the value set is correct.
c. Whether it is under thermal protection after a long time working.
B. Cooling effect is not good
a. The temperature in the centrifuge chamber is too high.
b. The centrifuge is too close to the wall.
c. Condenser heat sink is blocked by dust or condenser fan is not running.
d. Refrigeration system refrigerant leaks or is mixed with air. Cooling system is
blocked or broken.
e. The re-poured amount of refrigerant is too little or too much after repair or
refrigerant is not a match.
f. The vibration and noise of the centrifuge increases or sound abnormal.
g. Whether the ground is solid, and whether there are strong availability of
source effects.
h. Whether rotor balancing is destroyed.
Whether centrifuge tube is qualified for balance, placing is symmetrical.
i. Whether level did not adjust well, it is with uniform stress.
j. Whether shaft bending jumps too large.
k. In the centrifugal midway,the centrifuge tubes or 0-type seal ring broke.
17
C. The rotor can not be put into the rotor block reposefully.
a. Check rotor and rotor block if there are scars or dirty objects.
b. Whether transfer pin is out of shape.
D. The rotor can not be got out of rotor block.
a. With using the rotor handle screwing into center of the rotor screw,the rotor
can be removed.
b. The rotor block cone and the rotor cone coated with a little grease can easily
remove the rotor.
c. Do not run for a long time when there is no refrigeration.
Table 4 Trouble reminder and solution
Trouble Reminder Solution
Please check whether the power switch
The display window not working is lighten, and if the connect line of
display panel is loose.
Please check whether the electric
No Speed connector line of motor is loose, and if
the speed sensor is normal.
Turn on again, check whether the speed
Over Speed
sensor is normal
Door Unclosed Please close the door tightly before start
Please check whether the frequency c
onverter is damaged, whether the co
Transducer Communication Trouble
mmunication line is connect good
Module Trouble Please change the control board
7. Transportation and Storage
7.1 Transportation
You should use the wooden case and carton during long-distance
transportation. Put centrifuge with dust cap into the case, and fill with foam or
plastic shock absorption materials. It is forbidden to be hit, inverted, rolled and
soaked by rain and snow.
You can directly move the centrifuge in room, but avoid big vibration, hitting
and inversion
18
7.2 Storage
If the centrifuge is leaving unused for long time, you should open the door lid
ans store it in ventilated, dry and clean room where is no corrosive, flammable
and explosive matters.
The ambient temperature is -10℃~45℃, and relative humidity is not over
75%. No corrosive air and well-ventilated indoor.
As the technology continuously renews, please consult us if there is
something unconformable with this instruction manual.
19
Warranty
Respected customers,
Thank you for purchase of our company’s product. To protect your legal
rights and avoid your worries, we provide warranty service for one year with
the main machines. If it is broken under warranty with un-artificial reason, we
will supply free maintenance service. The customer needs to pay the prime
cost of the product parts if the product has exceeded the warranty period. We
will not supply the free maintenance if any following matter is happend:
1.Exceed of the warranty period or without the warranty card.
2.The trouble is caused by wrong installation, operation and maintenance.
3.The trouble is caused by the unauthorized uninstalling and repairing by our
company.
4.The trouble is caused by using the rotor not equipped with this centrifuge.
5.The trouble is caused by war or other irresistible factors.
This centrifuge has been subjected to thorough testing and quality control
before delivery. We promise to do as below:
1. One year warranty for this centrifuge (from the date of delivery).
2. Our company after-service center will build up files in time and follow up the
user's information.
3. We will offer the replacement parts freely if the centrifuge within warranty
20
Quality Certificate
Floor-type Large Capacity Refrigerated Centrifuge DL-6M
Max Capacity : 6x500ml
Max Speed : 6000r/min
Max RCF : 5120xg
Product series number:
This product is tested and qualified for delivery.
Test Department:
QC:
Date:
21
Packing List
Ser.No Description Quantity Remark
1 Centrifuge 1set √
2 Rotor 1sets √
3 Quality Certificate 1pcs √
4 Instruction Manual 1pcs √
5 Packing List 1pcs √
6 Rotor Handle 1pcs √
7 Compression Screw 1pcs √
Rotor Option
Max Speed Max RCF
Rotor No. Capacity Marks
(r/min) (×g)
No.1 Angle Rotor 6000 5120 12×10/15ml
4390 4×750ml
No.2 Swing Rotor 4000
3780 4×500ml
No.3 Swing Rotor 4000 3600 4×250ml
No.4 Swing Rotor 4000 3940 4×500ml (blood bag)
No.5 Swing Rotor 4000 3780 6×500ml √
4000 3440 72×5ml
No.6 Swing Rotor
4000 2300 148/108×7ml
No.7 Micro Plates 4000 2300 2×2×96 holes
Rotor
8×50ml
No.8 Swing Rotor 4000 2810 32×10/15ml
16×15ml
No.9 Swing Rotor 5000 4390 4×100ml
No.10 Swing 4000 3360 8×100ml
Rotor
22
Warranty Card
Floor-type Large Capacity Refrigerated Centrifuge
Name & Model
DL-6M
Specifications 6*500ml swing rotor
Date of Delivery
Address
Service Records
Date Contents Signature
23
Drawell International Technology Limited
Chongqing Drawell Instrument Co., Ltd.
Shanghai Drawell Scientific Instrument Co.,Ltd.
Chongqing Center : Suite 2705,Building No.12,Shiyou Road No.1,
Yuzhong District, Chongqing, China.
Shanghai Office : Suite 1117,Lane561 XiuChuan Rd.,PuDong New
Area,Shanghai,China
Homepage : www.drawell.com.cn
Tel : 0086-023-63268643
Email :
[email protected]