传感器说明书 - English (2025-04-08 18 - 40 - 35)
传感器说明书 - English (2025-04-08 18 - 40 - 35)
User Manual
1
Summary
1. General Information .................................................................................................................................. 2
1.1. Indication for use .................................................................................................................................... 2
1.2. Contra Indication .................................................................................................................................... 2
1.3. Safety Instructions .................................................................................................................................. 2
2. Product .......................................................................................................................................................3
2.1 Accessories .............................................................................................................................................. 3
2.2 symbology ............................................................................................................................................... 5
3. Software installation .................................................................................................................................. 6
3.1. Software requirements ............................................................................................................................6
3.2. Hardware requirements ...........................................................................................................................6
3.3. Cyber Security ........................................................................................................................................ 6
3.4 Sensor installation ....................................................................................................................................7
3.4.1 Connecting the sensor ...........................................................................................................................7
3.4.2 Fixation .................................................................................................................................................7
3.5 Software Installation ................................................................................................................................7
4. Software operation ...................................................................................................................................12
4.1. Open the software .................................................................................................................................12
4.2. Activate the software ............................................................................................................................ 12
4.3. Login .....................................................................................................................................................13
4.4. Patient search ........................................................................................................................................14
4.5. Create and edit patients .........................................................................................................................17
4.6. Acquire and edit exams ........................................................................................................................ 18
4.7. Capturing templates ..............................................................................................................................27
4.8. Exam printing ....................................................................................................................................... 28
4.9. Report ................................................................................................................................................... 29
4.10. Manage Users ..................................................................................................................................... 30
4.11. Manage Dentists ................................................................................................................................. 32
4.12. Database backup and restore .............................................................................................................. 33
4.13. Image calibration ................................................................................................................................ 34
4.14. Software information ..........................................................................................................................34
2
1. General Information
1.1. Indication for use
Intraoral Software is indicated to acquire radiological images to be stored, visualized and manipulated for
evaluation and analysis by dentists. The software should be used only in hospital environments, radiology
clinics, dental offices and handled by qualified and trained professionals.
● To avoid accidental changes of the patient's examinations, the top bar shows the patient's
information, where the patient's full name will always remain visible to the operator. Always
check if the running exam belongs to the patient indicated.
● The software uses the date and time of the operating system installed on the computer as a date
and time reference. If the operating system date and time are incorrect, the date and time of scan
scans saved at the time of the exam is performed are incorrect. For this reason, the date and time
of your computer's operating system should be checked periodically.
● The diagnosis and calculations generated by the software depend on the external capture devices,
and the operator is responsible for proving their information, performing the procedures correctly,
validating calculations and diagnosing based on the data generated by the software.
2. Product
2.1 Accessories
Open the outer packaging to access and check the available accessories.
ACCESSORY DESCRIPTION:
Lightning sensor
Be USB
4
Support
2.2 symbology
Instructions for operations
Maker
Date of manufacture
Reference number
Serial number
Fragile
This side up
Temperature limits
6
3. Software installation
3.1. Software requirements
For the correct use of Intraoral Software, the following software must be installed on your computer.
● .Net Framework 4.7.2 Runtime: Is a managed execution environment for Windows that provides
a variety of services and features for your running applications.
● Visual C++ Redistributable Packages 2015 - 2019: Install software libraries that are required to
run C++ applications that are built using Visual Studio 2015-2019.
● PostgreSQL 10.5: is a POSTGRES-based object-relational database management system. It
supports a large part of the SQL standard and offers many modern features, including complex
queries, foreign keys, encryption, upgradeable views, transactional integrity, and multiple version
concurrency control.
3.2. Hardware requirements
As this software is not computer equipped, during the installation of this product, customers need to
provide a laptop or computer. The specific performance requirements of the computer are as follows:
RAM 4 GB or higher
Monitor 1024x768
○ Software access: The software requires login credentials to start, operators can open the
software only after entering the correct username and password.
○ Encryption: The software is encrypted and linked with Intraoral Digital Image Sensor.
Only when the license file and sensor calibration file are included, the user can use the
software normally.
3.4 Sensor installation
3.4.1 Connecting the sensor
Insert the sensor into the USB input port of a desktop or laptop device.
3.4.2 Fixation
Secure the stand in a convenient location near the computer that will use the sensor for housing and
protection.
Read and understand the end-user license agreement and if accepted, select the "I accept agreement "
option and click "Next".
9
Wait for the full installation to end. This operation may take few minutes.
11
4. Software operation
4.1. Open the software
Double-click the software shortcut on the desktop.
If the license is activated, the message "License was installed successfully" appears. Click OK.
4.3. Login
The login interface below is displayed every time the software starts.
Sign in with your credentials (user and password) and click "Log in".
14
2 3
1
15
4
8
5
7 6
Item Description
Patient Information: Basic information of the current patient selected by the user:
● Patient photo, when there is.
1 ● Patient's first and last name.
● Gender and age (Years and months).
New patient.
2 Edit patient.
Delete patient.
Log out.
4
Import exams.
Export exams.
5
Delete tests.
List of examinations of the selected patient organized in chronological order from the
6 most recent to the least recent.
Click the button to create a patient. The following interface will appear:
5 2
Item Description
Click the button to edit a patient. The form fields and the patient's photo will be filled in with the
information recorded in the database.
2 3
19
1 4
5
8 6
Item Description
Simple measurement
Performs a measurement between two points in the exam.
Multiple measurement
Performs distance measurement of multiple points in the
image.
Angular measurement
Performs three-point angular measurement in the image.
Rectangle
Inserts a rectangle into the image.
Ellipse
Inserts an ellipse into the image.
Straight
Inserts a line into the image.
22
Arrow
Inserts an arrow into the image.
Freehand drawing
Inserts a freehand drawing into the image.
Asterisk
Inserts an asterisk into the image.
Memo
Inserts a memo with predefined texts in the image.
Distance calibration
Performs distance calibration on the image.
Mark two points from a known distance in the image and type
the value in millimeters of that distance. Enter the calibration
name and click "Ok".
List of measurement and annotation tools inserted, containing the type, color and
description.
Brightness
23
Gamma
Adjust the exam gamma correction from -99 to 99.
Sharpen
Adjust the sharpness of the exam with intensity 0 to 7.
Denoise
Remove noise on the exam with intensity 0 to 3.
Emboss
Apply self-relief filter with intensity from 0 to 3.
Histogram.
Histogram Graph
The number of tones occurring throughout
the image. The horizontal axis represents
the values of the tones, where the left side
represents pixels with low values (black
areas or shadows), the right side
represents those with high values
(highlights or light areas).
Pseudo colorization
Grayscale images are mapped to a new
map color to improve the viewing of
regions with the same contrast.
Histogram equalization
Equalizes the grayscale scans in the
24
image.
Historic
Filter list
List of filters applied in the exam
containing type and intensity.
Presets
List of predefined filters to be applied to
the exam.
Report edition:
7
Click on the button to expand the text box for report editing.
Add bookmark.
Center text.
Justify text.
Exam acquisition area: site intended to show the exam purchased and being worked.
8
Exam manipulation tools.
Zoom in.
Zoom out.
Magnifying glass.
Hide/show notes.
Click the button and select one of the existing options: FMX 18, FMX 14, FMX 10 and
bitewing.
Connect the intraoral sensor to the COMPUTER's USB port.
Select the exposure parameters (time and kV) on the X-ray emitter and shoot to perform the image
acquisition. The following interface should appear.
Item Description
28
Template area: Select the acquisition sequence by clicking on the image boxes. The
1 software will follow the automatically established sequence.
2
1
Printing tools.
2
29
Viewing options:
● Patient photo.
● Patient name.
● Patient gender.
● Name of the dentist.
● Right from the clinic.
● Name of the clinic.
● Clinic info.
Select magnification:
● Fit in the area.
● 1:1 Printing
Print type:
● Default: Printer printing installed
on the operating system.
● DICOM: PRINTING ON DICOM
printer registered in the settings.
4.9. Report
Click the button to show the report view of the report.
Click the Print, Export, Cloud, and Email buttons to send the report to the desired option.
The following interface should appear.
30
Viewing options:
● Patient photo.
● Patient name.
● Patient gender.
2 ● Name of the dentist.
● Right from the clinic.
● Signature.
● Clinic info.
● Odontogram.
Item Description
List of users.
1
User details:
● Login.
3 ● Email.
● Password.
● Confirm password.
Available permissions:
● Manage patients.
● Manage users.
● Manage dentists.
● Manage exams.
5 ● Export exams.
● Create annotations.
● Create reports.
● Change settings.
Save changes.
6
4
33
Item Description
List of dentists.
1
Dentist details:
● Title.
● Name.
3 ● Surname.
● Email.
● Specialty: Dentist, endodontist, periodontist, implantodontist, stomatologist, oral
pathologist, radiologist.
Save changes.
4
4.15. Settings
Click the button to open the Settings window.
35
1
2
4 3
Item Description
Export Select the default export format: Bitmap, Dicom, Jpeg, Png and
Tiff.
Select the default export directory by clicking the .
37
Email Enter the Outlook Email account and the default message.
Clinic Edit clinic information: name, address, phone number, email and
website.
39
DICOM Printing Options for creating, editing, deceiting, and testing connections to
DICOM printers.
● Network configuration: AE Title, IP and Port.
● Media: Blue Film, Clear Film or Paper.
● Magnification: Bilinear, Replicate, or Cubic.
● Priority: Low, Med or High.
● Destination: Magazine or Preprocessor.
● Cut: Yes or No.
● Empty part fill: dark or light.
DICOM Server Options for creating, editing, deceding, and testing connection to
DICOM servers.
● Network configuration: AE Title, IP and Port.
5. Maintenance
Note:
Only the use of 70% alcohol is indicated for disinfection.
Do not use other disinfectants and do not do so with the sensor connected to a desktop or laptop
workstation.
Equipment that cannot be sterilized in a steam autoclave or any other sterilizer.
Do not use hard brushes or scrapers of any kind to clean or strong acids or alkaloides.
5.3 Troubleshooting
If you have problems with the operation of the software, check the following:
Symptoms Diagnosis and Solution
42
The sensor does not connect to the The software does not recognize the sensor.
software Make sure the USB cable is properly connected to
your computer.
Unplug and reconnect the sensor to the USB port.
Make sure the intraoral sensor drivers are installed.
The sensor is not calibrated The software did not find the sensor calibration files.
Follow the steps in section 4.13
Sensor status is "Connecting" The software cannot connect to the sensor.
Make sure the USB cable is properly connected to
your computer.
Unplug and reconnect the sensor to the USB port.
Check that the intraoral sensor drivers are installed.
The image does not appear after X-ray The software did not acquire the sensor image.
exposure Make sure the USB cable is properly connected to
your computer.
Check the operation of the X-ray transmitter.
Check the X-ray exposure time.
Make sure that the active area of the sensor is
oriented towards the X-ray emitter and that they are
properly aligned.
Recalibrate the sensor by section 4.13.
The examination appears incomplete Loss of communication with the sensor during exam
after the acquisition transfer.
Make sure the USB cable is properly connected to
your computer.
Unplug and reconnect the sensor to the USB port.
Very clear acquired examination Insufficient exposure time.
The voltage of the generator is insufficient.
Adjustments to your monitor. (Brightness and
contrast).
Check to see if there is a filter preset being applied.
Recalibrate the sensor by section 4.13.
Examination acquired too dark Exposure time too high.
Generator voltage's high.
Adjustments to your monitor. (Brightness and
contrast).
Check to see if there is a filter preset being applied.
Existence of rows and columns in the It is necessary to recalibrate the image.
exam Follow the steps in section 4.13
43
When you open the software, the The software is not activated.
activation screen appears Follow the instructions in section 3.2
The software is not enabled for the Follow the instructions in section 3.2
intraoral sensor serial number
When activating the software, activation Sensor not registered.
does not recognize the sensor serial Contact technical support.
number
User and password are invalid Reset the user's password.
The software does not find the database The software does not connect to the database.
Add the software to the antivirus and firewall
exception list.
Contact technical support.
Error sending scan to DICOM printer The software does not connect with the printer.
Add the Printer TCP port to the antivirus and
firewall exception list.
Check the connection to the printer.
Check the printer configuration in the software
according to section 3.15.
Error sending scan to DICOM server The software does not connect to the server.
Add the TCP port from the server to the antivirus
and firewall exception list.
Check the connection to the server.
Check the server configuration in the software
according to section 3.15.
Incisors 0.1-0.2
Canines 0.1-0.2
Premolars 0,15-0,25
44
Molars 0,15-0,25
Recommended table for an adult patient using a 60kV/2mA emitter beam
NOTE:
Adjustments of ±25% of the recommended table are recommended for clinical cases of edentulous
patients, patients up to 21 years old or x-ray sources with different parameters of 60kV/2mA to obtain a
clear picture or according to the expectations of each user.
6. Technical data
The images depicted in this manual are for technical reference only and the manufacturer reserves the
right to make changes without notice.
This manual may not be reproduced and/or altered without the prior permission of the manufacturer or
importer.
6.1 General
MODELS Z-sensor Z-sensor Max
Packing size 30,0x11,0x3,2 cm 30,0x11,0x3,2 cm
sensor size 25 0x38,0x4,5mm 32,0x54,0x4,5mm
Active area 20x30mm² 26x36mm²
pixel matrix (1,5M) 1000x1500 (2,34M) 1300x1800
Data interface USB Direct, USB 2.0/3.0 USB Direct, USB 2.0/3.0
Sensor technology APS CMOS APS CMOS
centelleador CsI:TL CsI:TL
Connector cable size 3m 3m
pixel size 20 μm 20 μm
Resolution (theoretical) 25lp/mm 25lp/mm
grayscale 16 bits 16 bits
Exposure mode Smart DEA Smart DEA
Protection index IP68 (disinfection) IP68 (disinfection)
Protocol of dos Yes Yes
Energy range (X-rays) 55-100kV 55-100kV
Electrical safety class Class II Class II
Recommended separation distances between portable or mobile RF communication device and detector:
Portable RF communications equipment, including antennas, may affect medical electrical equipment. A
wear distance of at least 30 cm (12 inches) from any part of the sensor, including cables specified by the
manufacturer, shall be considered."
Web:wukongdent.com
.