The document is a comprehensive manual for the T7 Automatic Moisture Analyzer, detailing its introduction, installation, specifications, and operational instructions. It includes safety precautions, functional descriptions, calibration procedures, and menu settings for effective use. Additionally, it outlines the automatic cover opening feature and provides tips for optimal operation and maintenance.

Uploaded by

ELERmano
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
37 views8 pages

Automatic Moisture Analyzer Xy Xing Yun T7

The document is a comprehensive manual for the T7 Automatic Moisture Analyzer, detailing its introduction, installation, specifications, and operational instructions. It includes safety precautions, functional descriptions, calibration procedures, and menu settings for effective use. Additionally, it outlines the automatic cover opening feature and provides tips for optimal operation and maintenance.

Uploaded by

ELERmano
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 8

T7

AU TO M AT IC MO I S T U R E ANA LY Z E R
Seven. Automatic cover opening (optional function)
CATALOGUE
ONE.INTRODUCTION ...................................................................................1

TWO.INSTALLATION.....................................................................................2
2.1 Standard packing list
2.2 Water placement requirements 6.1 Functional description
2.3 Installation of the accessories
THREE. Specification Parameters...............................................................3 This machine has optional automatic open heating cabin
function, can open the door through the home page. Through the biaxis
Four. Assembly diagram............................................................................4
without mechanical direct drive cabin door open, effectively reduce the
Five. Home page..........................................................................................5 mechanical noise and mechanical vibration.
5.1 Open machine interface 6.2 Intelligent sensing
5.2 Test interface When the door is closed, the door will automatically return to the starting
position to avoid clip. At the same time, when the door is forced, the door will
be automatically opened to avoid artificial damage to the mechanical
Six. Menu....................................................................................................7 structure.
6.1 Parameter settings 6.3 Precautions
1. The automatic opening mechanism should operate under human
6.2 Historical Records monitoring to avoid errors affecting the experimental results.
6.3 Instrument calibration 2. Forced opening or closing will affect the life of the transmission.
6.4 Language settings Please avoid this operation as far as possible.
3. In case of failure, do not dismantle the machine by yourself. Please
6.5 System
consult the technical support or the factory.
6.6 Print Settings Eight, Tips
6.7 System information 1. Please start the first calibration after 30 minutes;
6.8 step-up heating 2. A moisture experiment can be performed as a preheat after the
calibration.(See 3.6 for details);
3. Each switch, please interval of 5 seconds;
SEVEN. Automatic Open Cover................................................................14 4. If a continuous experiment, do a new experiment with the temperature
6.1 Functional description below 40℃.
5. When covering the heater, do not leave the weight on the scale,
6.2 Intelligent sensing otherwise the halogen lamp will be crushed;
6.3 Precautions 6. Leave the machine, be sure to cut off the power supply;
Eight. Tips...................................................................................................14 7. Do not allow the moisture analyzer to work under no one monitoring
condition.
- 13 -
ONE. INTRODUCTION
6.5.6 Label selection 1. Safety precautions
For the safe and reliable use of the water analyzer, be sure to observe the
following terms and conditions:
Choose the adhesive size, and choose the specified adhesive printer to print
a. This instrument is suitable for the determination of the sample
the label. moisture content. All improper operations may result in personal injury
and instrument damage.
b. Please use the AC power supply as specified in the specification. The
instrument uses 3pin earth plug and do not disconnect the earth plug.
6.5.7 Data Parameters c. The power cord connection shall not hinder the normal activities of
the human body. Keep the human body from touching or tripping.
Functional settings with different data parameters. d. Do not operate this instrument in dangerous, wet and unstable conditions.
e. Unplug the power supply when cleaning the water meter.
f. There is sufficient space around the instrument in use. At least 1m of the above space.
g. The instrument must be operated by professional personnel and wear the
6.5.8, administrator authority corresponding protective equipment. Such as safety eyes, protective clothing, gloves,
and face masks.
Administrator permission function, which can set the login password for
h. Please do not arbitrarily change the parts and other aspects of the instrument.
administrators and operators. Maintenance of this instrument is performed by professionals authorized by the
factory.
The moisture meter is operated by heating
i. No flammable articles shall be placed around the instrument.
6.6 Print Settings j. During use, the sample, heating halogen lamp and surrounding parts will
be very hot. Special care should be taken to prevent and cure burns.
Special attention was paid to the hazards of some samples.
k. For some sample substances with potential safety risks, please do a
careful risk analysis and prevention before the test.work.
Fire / explosion: flammable and explosive and solvent containing sample
You can choose the parameter settings for printing or connecting to the heating will produce flammable and explosive gas or steam. Therefore, it
should be done in a dry and low-temperature environment. In case of a
computer according to the communication requirements.
fire or an explosion.
Toxic / corrosion: Conduct samples of toxic or corrosive ingredients in a
6.7 System information well-ventilated environment. Such substances will release toxic and
corrosive gases during heating. It is recommended to test the samples as
little as possible.
6.8 Step test: Above the sample test will be particularly careful, for any damage caused
by your own risk
-1
- 12 -
TWO.INSTALLATION 6.5.2 Stop conditions:
2.1 Standard Packing List Set shutdown conditions for special requirements, if the standard reference
Please check whether the packaging and accessories is complete when you open it value is not a modification.
Standard Quantity Note 6.5.3 System Parameters:
Moisture Analyzer Mainframe 1 set This item is used for system parameter adjustment and requires a password
Wind Cover 1 piece to turn on.
The sample pan holder 1 piece 6.5.4 Information Settings:
Tray handle 1 piece
Power cord 1 piece
Aluminum sample pan 1 box 80pcs
100g standard weight 1 piece
Product Manual 1 piece
Product certification 1 piece
You can click on the corresponding item to input information for test report
Warranty card 1 piece
printing and retention.
2.2 Placing Requirement.

6.5.5 Environmental calibration:


Used to deduct the effect of moisture on the scales in humid environments on
the results of experiments.
When the air humidity is too high, it will affect the moisture test, then we need
Operating table Temperature stable, good ventilation to select this function
In order to test the effect of the environment on the moisture of the product,
should be stable dry, without strong vibration, of the environment the procedure is as follows:
without strong magnetic field. 1.Select the environment calibration to open, press to confirm, and return to
2.3 Installation of accessories the test page;
2.Remove the sample weighing pan, tare, and then place the empty pan to
show the weight;
3. Press the START icon to start the test, when the page status area .
4. Indicates that the function is already turned on. Note that this item is only
suitable for the testing of special samples.
Place the wind cover, Place the sample pan Put the aluminum sample 5. If you need to close, please enter the environment calibration category,
aligned holes. holder, rotate until pan on the bracket. select the environment calibration function off, and press confirm.
positioning.
2 - 11 -
6.3 Instrument calibration Three:SPECIFICATION PARAMETERS

Model T7 Pro

Specification 110g/0.0001g

Readability 0.0001g

When the instrument weighing and temperature deviate, the instrument can Pan size Φ 100
be calibrated with standard weights as indicated. Heating lamp Halogen lamp
One point calibration: touch icon enter the one point calibration, flash
Display 7 inch Touch panel
"100.000" and put on 100g standard weight. When "100.0000" is stable,
remove the weight and wait until the display returns to "0.000", the calibration Temperature Setting 40℃~230℃
will end. Time Setting 1~99 minutes
Liner calibration: touch icon enter the calibration, flash "100.0000", put 100g
History 50 groups
standard weight, when "100.000" stable, remove the weight, then flash
"50.0000", put the 50g standard weight, when "50.000" stable, remove the Calibration External Calibration
weight, then flash "20.0000", put 20g standard weight, when "20.000" stable, User Administration Administrators & Operators
take the weight to "0.0000", calibration ends.
Heating Mode Standard/Fast/Soft/Customize
Temperature calibration: Temperature calibration is required with the
specified thermometer. Stop mode Automatic/Manual/Timed/Step

6.4 Language settings Information input User information/Sample name


Choose simplified Chinese and English. Label printing Available sizes
6.5 System
Baud rate setting 1200/2400/4800/9600
This item includes:
TIME/DATE,STOP,MODE,SYSTEM,SETTING,INFORMATION SETTING, Languages Chinese/English
HUMIDITY CALIBRATION,NTERNETDATA BASE,ADMIN Other functions Grain rate setting、Environmental calibration

6.5.1 Time and Date:


This item sets the current time and date, and can be updated in real time and
recorded on a test report via a printer.

-3-
- 10 -
4
a
d
.
ly
b
m
e
s
g
in
rw
Four.Assembly Drawing 6.2 Historical Records

1. The results of the last 50 groups of tests will be directly displayed on this
function page;

2. You can read and retrieve the history records through the
function keys. Convenient comparison and analysis of sample properties;

3. At the same time, this page can directly print the test report
through the printer;
4.Note that the history can be viewed under the test state, but it cannot be
printed.
Please follow the installation in the following order
Installation sequence from top to bottom: 1 ——2—— 3 —— 4
1. Aluminum Pan
2. The Triangle Scale Frame
3. Scale Pan Brackets
4. Wind Cover
5. Stainless Steel Plate

-4- -9-
Enter the required test temperature directly, press click input, touch
CONFIRM and confirm.

TEST MODE: 5.1 Open machine interface


Power on then enter the weighingt page, this operation page is the operation
Standard: This mode is suitable for the moisture measurement of most
home page;
substances; MENU: the moisture analyzer parameter can be adjusted and set;
Soft: The temperature will slowly reach the set temperature, this mode is TARE: remove the excess weight on the aluminium pan, and then weigh the
more better for protecting the subject from pyrolysis, measured objects;
START: when the weighing is stable, the water measurement can be started;
Fast: The heating temperature will reach the set temperature even faster, CAL: calibrate the moisture analyzer, and enter the calibration page after
Fast mode can effectively improve the experimental efficiency and shorten touch the button;
the experimental period, But it may also cause the pyrolysis of the tested REPORT: enter the history test record page, you can retrieve the historical
test records;
object. PRINT: print the current test record;
TIME SETTING: WEIGHT: sample weighing value;
This item is combined with the shutdown mode, set the heating cycle, and USER : click to edit the user information, and you can print it;
SAMPLE: Click to enter the name of the current test sample to print it;
make reasonable use
TEMP: close the door display: set the temperature of the test, open the door
Time setting, you can truly simulate the oven to accurately measure the display the current ambient temperature;
moisture percentage. Other feature icons
①: Test time; ②: Test mode; ③: Stop mode; ④: Door opening state; ⑤: Clock

-8-
Five. WEIGHING PAGE(HOME PAGE) -5-
5.2 Test page Six, MENU
After starting the test, enter the test home page and view it through the
conversion icon:
Moisture content, dry weight, and wet weight.

6.1 Parameter settings


Conversion: View the different test results Select ② to enter the parameter setting, and select the corresponding mode:
Return: under the stop state, can return to the weighing home page, the test STOP MODE:
process back, the test end; Automatic: the system automatically determines the downtime (the time
Pause / stop: End of the test; setting under this function is invalid);
①: TEST TIME; ②: TEST MODE; ③:STOP MODE; Manual: the water meter will not stop the heating test by itself until the
④: Heating state temperature display manual stop;
⑤: Sample heating test curve, X axis time, Y axis moisture percentage Timing: set the test time, the test will stop with time
⑥: Test result of grain yield (shutdown) (This item works with the timing setting).
⑦: Moisture test results, converted to solid content / dry weight TEMPERATURE SETTING:
⑧: Water test process indication Different peak temperatures can be set according to the subject item,
⑨: Clock display The temperature set range is: 40℃ ~230℃
Temperature-adjustable ladder: 1℃

-6- - 7-