2830 / 2831
Precision Solid and Liquid
Dielectric Analyzer
Datasheet
General Description
The HAEFELY 2830/2831 is the result of extensive The system consists of 2830, the measuring bridge
research and years of experience testing dielectric and control unit, and 2831, the extension unit. This
properties of liquid and solid insulating materials. It system is a complete replacement of the Tettex
incorporates a fast and highly advanced procedure 2821 and 2822.
to measure Capacitance, tan , resistivity and
The 2830 contains the controller and the
relative permittivity r (dielectric constant) of liquid or
measurement part of the system.
solid insulating materials. A simple one-time-
connection system together with resistivity The 2831 extension unit contains a standard capacitor,
measurement according to pre-selected standards 2.5 kV AC & DC supplies and a temperature control
drastically reduces measuring time. unit.
Features Advantages
▪ Capacitance, tan 𝛿, resistivity 𝜌 and relative Simple and one-time connection
permittivity 𝜀𝑟 on liquid and solid insulating
materials.
▪ Large 12’’ touch display Full graphical test visualization
▪ Customized programmable sequences and Guaranteed test repeatability and additional
manual mode flexibility
▪ Integrated AC power supply 5 – 2500 V and 40 – Adjustable control
65 Hz
▪ Two independent integrated temperature control Robust
units
▪ Integrated ambient temperature and humidity Reliable measurement
sensor
▪ Place to connect two liquid test cell heater 2903 Reduced testing time
or one solid test cell 2914
▪ Existing Tettex test cells 2901 (from 1990 and Compatible
later) and 2914 (from 1985 and later) can be used
▪ Data transfer over USB memory stick Convenient memory storage and test recall
Applications
C and tan measurement and resistance measurement on liquid and solid insulating materials.
▪ Liquids like mineral, natural or synthetic oils.
▪ Solid materials like insulation paper, silicone, rubber, polymeric plastics, etc.
The 2830/2831 is a valuable tool for factory tests, routine tests, R&D tests and regular maintenance.
Scope of Supply
▪ Solid and liquid dielectric analyzer 2830 and 2831
▪ Test certificate
▪ Set of connection cables
▪ Operating Manual
Technical Data
Measurement 2830 Range Max. Resolution Accuracy
Dissipation factor (tan 𝛿)1 0 … 100 1 x 10-6 ±0.5 % rdg ± 1 x 10-5
Capacitance2 ≥ 10 pF 0.001 pF ±0.2 % rdg ± 0.01 pF
Relative permittivity 𝜀𝑟 1 … 30 1 x 10-3
Resistance 120 kΩ … 5 TΩ3 1 kΩ <1 TΩ ± 5 % rdg +3 digits
≥1 TΩ ± 15 % rdg +3 digits
Resistivity (Liquid)4 900 kΩm … 27 TΩm5
Resistivity (Solid)6 2.4 MΩm … 80 TΩm7
Test current @ Input Cx 10 µA … 10 mA 0.01 µA ±0.1 % rdg ± 0.1 µA
Test current @ Input Cn 10 µA … 10 mA 0.01 µA ±0.1 % rdg ± 0.1 µA
Test Frequency 45 – 65 Hz 0.1 Hz ±0.1 % rdg ± 0.1 Hz
AC test voltage 5 … 2500 V 1V ±0.3 % rdg ± 1 V
AC frequency 40 – 65 Hz 0.1 Hz
AC current max. 5 mA
DC test voltage 250 – 2500 V 25 V ±10 % rdg +20 V
Hardware
Heater controller Ambient – 200 oC 0.1 oC ±0.5 oC
Internal standard 1x 10-5 ±1 x 10-5
capacitor2 Dissipation
factor (tan 𝛿)
Internal standard capacitor 1 nF 5 % @ 25 °C 20 ppm/°C
Capacitance
Display 12” TFT, 800 x 600, integrated Touch-Screen
Software
Operating System Windows 7 embedded
Interfaces 3 x USB
Data Format XML, CSV
Mains Power Supply
Voltage 90 … 264 V AC
Power Max. 1.7 kVA
Frequency 50 / 60 Hz
Environmental
Operating temperature 10 °C +40 °C
Storage temperature -20 °C +70 °C
Humidity 10 .. 60 % r.h. , non-condensing
Mechanical
Safety specifications VDE 0411/part 1a , IEC/EN 61010-1:2002
Dimensions (W x D x H) 2 pcs 480 x 440 x 270 mm (19 x 17.3 x 10.6 in)
Weight 21 kg (46.3 lb) (2830), 19 kg (41.9 lb) (2831)
1. Accuracy values @ 50/60 Hz
2. Range limit is given by test current and voltage
3. @ 2.5 kV (Rmax = 2 G x Utest [V])
4. Resistivity range is given by the resistance range multiplied with the cell factor of the test cell (2903 = 0.113 x Cair [in pF])
5. Typical range (calculated with Cair = 60.0 pF of 2903)
6. Resistivity range is given by the resistance range multiplied with the ratio: surface area of the measurement electrode / distance between
the HV and the measurement electrode (2914 = 0.002m2 / distance in m)
7. Typical range (calculated with a distance of 0.1mm between the electrodes and 2 kV test voltage)
Global Presence
Europe China
HAEFELY AG HAEFELY AG Representative Office
Birsstrasse 300 8-1-602, Fortune Street, No. 67
4052 Basel Chaoyang Road, Beijing 100025
Switzerland China
+ 41 61 373 4111 + 86 10 8578 8099
[email protected]