Tdls8000 Yokogawa Laser Analyzer-12-13

The document provides detailed specifications for the TDLS8000 Tunable Diode Laser Spectrometer, including various model suffix codes, option codes, and gas parameters. It outlines different configurations for general purpose, explosion-proof, and specific gas measurements. Additionally, it includes notes on compatibility and selection criteria for various options and accessories.

Uploaded by

mjhmeh2003
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
58 views2 pages

Tdls8000 Yokogawa Laser Analyzer-12-13

The document provides detailed specifications for the TDLS8000 Tunable Diode Laser Spectrometer, including various model suffix codes, option codes, and gas parameters. It outlines different configurations for general purpose, explosion-proof, and specific gas measurements. Additionally, it includes notes on compatibility and selection criteria for various options and accessories.

Uploaded by

mjhmeh2003
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

12

 MODEL AND CODES


● TDLS8000 Tunable Diode Laser Spectrometer

Model Suffix Code Option Code Description


TDLS8000 ..................................... ....................... Tunable Diode Laser Spectrometer
Type -G1 ....................... General Purpose, cable entry/piping:NPT
-G2 ....................... General Purpose, cable entry:Metric thread, piping:Rc
-GQ ....................... EAC with PA General Purpose, cable entry:Metric thread, piping:Rc
-GR ....................... EAC General Purpose, cable entry:Metric thread, piping:Rc
-D2 ....................... FM (US) Class I Div 2, Zone2, cable entry/piping:NPT
-C2 ....................... FM (Canada) Class I Zone2, cable entry/piping:NPT
-S2 ....................... ATEX Type of protection “n”, cable entry:Metric thread, piping:Rc
-E2 ....................... IECEx Type of protection “n”, cable entry:Metric thread, piping:Rc
-K2 ....................... Korea Ex Type of protection “n”, cable entry:Metric thread, piping:Rc (*15)
-N2 ....................... NEPSI Type of protection “n”, cable entry:Metric thread, piping:Rc (*15)
-Q2 ....................... EAC with PA Type of protection “n”, cable entry:Metric thread, piping:Rc (*15)
-R2 ....................... EAC Type of protection “n”, cable entry:Metric thread, piping:Rc (*15)
-U2 ...................... INMETRO Type of protection “n”, cable entry: Metric thread, piping: Rc (*15)
-D1 ...................... FM (US) Class I Div 1, Zone1, cable entry/piping:NPT (*1)
-C1 ...................... FM (Canada) Class I Zone1, cable entry/piping:NPT (*1)
-S1 ....................... ATEX Flameproof ”d”, cable entry:Metric thread, piping:Rc (*1)
-E1 ....................... IECEx Flameproof “d”, cable entry:Metric thread, piping:Rc (*1)
-K1 ...................... Korea Ex Flameproof “d”, cable entry: Metric thread, piping: Rc (*1)
-J1 ...................... Japan Ex / Zone 1, cable entry:Metric thread, piping:Rc (*1) (*11)
-J2 ...................... Japan Ex / Zone 2, cable entry:Metric thread, piping:Rc (*11) (*15)
-N1 ...................... NEPSI Flameproof “d”, cable entry:Metric thread, piping:Rc (*1)
-Q1 ...................... EAC with PA Flameproof “d”, cable entry:Metric thread, piping:Rc (*1)
-R1 ...................... EAC Flameproof “d”, cable entry:Metric thread, piping:Rc (*1)
-U1 ...................... INMETRO Flameproof “d”, cable entry:Metric thread, piping:Rc (*1)
Gas Parameter -X1 ....................... O2 < 600°C, 0-25% (*2)
-X2 ....................... O2 < 1500°C, 0-25% Combustion
-C1 ....................... CO (%) 0-20%/0-50% <500°C
-C2 ....................... CO ppm 0-200ppm/0-10,000ppm <500°C (*3)
-C3 ....................... CO ppm <1500°C Combustion (*3)
-C4 ....................... CO ppm <1500°C or CH4 0-5% Combustion (*3)
-C5 ....................... CO (%) + CO2(%) 0-30%/0-100% <150°C
-A1 ....................... NH3 up to 0-5,000ppm <450°C DeNOX
-A3 ....................... NH3 (ppm) + H2O (%), QAL1 (*1) (*14) (*15) (*16)
-S1 ....................... H2S 0-5%/0-100% <100°C (*2)
-D1 ....................... CO2 High Range 0-1%/0-5% <100°C
-D5 ....................... CO2 Extend. Range 0-30/0-50% <150°C
-H1 ....................... H2O ppm non-Hydrocarbon Background (*1)
-H3 ....................... H2O ppm Hydrocarbon Background (*1)
-H4 ....................... H2O 0-10%/0-100% <500°C (*1)(*2)
-L1 ....................... HCl 0-50ppm/0-5,000ppm <500°C
Optics Accessory -NN ....................... Without Alignment Flanges (*4)
-LA ....................... Large Aperture Optics, ANSI CLASS150-4-RF (*5) (*6)
-U2 ....................... ANSI CLASS150-2-RF(Eq.) Alignment Flange, piping: NPT
-U3 ....................... ANSI CLASS150-3-RF(Eq.) Alignment Flange, piping: NPT
-U4 ....................... ANSI CLASS150-4-RF(Eq.) Alignment Flange, piping: NPT
-D5 ....................... DIN PN16-DN50-D(Eq.) Alignment Flange, piping: Rc
-D8 ....................... DIN PN16-DN80-D(Eq.) Alignment Flange, piping: Rc
-J5 ....................... JIS 10K-50-FF(Eq.) Alignment Flange, piping: Rc
-J8 ....................... JIS 10K-80-FF(Eq.) Alignment Flange, piping: Rc
-FC ....................... Flow Cell Alignment Flange (*6)
I/O Interface -A1 ....................... Analog with HART+Modbus Ethernet
SI Unit -N ....................... SI Unit or non SI Unit (*7)
-J ....................... Only SI Unit
— -N ....................... Always -N

All Rights Reserved. Copyright © 2015, Yokogawa Electric Corporation GS 11Y01D01-01EN Apr. 10, 2025-01
13
Option /D Diverging Beam without LAO (*8)
/RX Reference Cell for O2 (*9)
/RC Reference Cell for CO (*10)
/SCT Stainless Steel Tag Plate
/JA1 Cable gland for Japan Ex “-J1” (Cable O.D. 8-12mm, G1/2) 1pc, for local HMI
/JB3 Cable gland for Japan Ex “-J1” (Cable O.D. 10-16mm, G3/4) 3 pcs
/JB4 Cable gland for Japan Ex “-J1” (Cable O.D. 10-16mm, G3/4) 4 pcs
/JC1 Cable gland for Japan Ex “-J2” (Cable O.D. 6-9.5mm, M20) 1pc, for local HMI
/JD1 Cable gland for Japan Ex “-J2” (Cable O.D. 8.5-13.4mm, M20) 1pc, for local HMI
/JE3 Cable gland for Japan Ex “-J2” (Cable O.D. 9.5-15.4mm, M25) 3 pcs
/JE4 Cable gland for Japan Ex “-J2” (Cable O.D. 9.5-15.4mm, M25) 4 pcs
/P2 130<P<400kPa (abs.) (*12)
/P3 P>400kPa (abs.) (*12)
/WCD DNV Type Approval (*13)

*1: Type “-D1”, “-C1”, “- S1”, “-E1”, “-K1”,“-J1”, “-N1”,“-Q1”, “-R1”, “-U1” cannot be selected with “-H1”, “-H3”, “-H4” or “-A3”.
*2: When the process gas pressure is out of 90 to 130 kPa (abs.) or the process gas contains CO2 ≥ 40 % or H2 ≥ 20 % as
coexisting gas components, please contact YOKOGAWA.
*3: When CO or CH4 ingredient coexist, please contact YOKOGAWA.
*4: When “-NN” is selected, Zone2/Div2/Type of protection “n”, FM (Canada) Zone1 is not available.
*5: For applications whose optical path length is 6 m or longer, select the “-LA”. A condensing lens unit (LAO unit) is added to
the SCU side.
“-LA” can be selected when Oxygen or CO (-C2, -C3, -C4) analyzer is selected.
“-LA” can be selected with Zone 1/Div.1/Flameproof “d” when Gas Parameter “-X2”, “-C3”, “-C4” is selected.
*6: When FM (US) or FM (Canada) is specified, the connecting port for window purge is 1/4NPT.
When ATEX, IECEx, Korea Ex, NEPSI, EAC or Japan Ex is specified, the connecting port for window purge is Rc1/4.
*7: Available only to an end user located outside of Japan.
*8: The Option “/D” can be selected when Large Aperture Optics “-LA” of the Optic Accessory is not specified and Oxygen or
CO (-C2, -C3, -C4) analyzer is selected.
*9: The Option “/RX” can be used when Oxygen analyzer is selected. When both “-X2” of the Gas Parameter and “-LA” of the
Optics Accessory are selected, “/RX”must be specified.
*10: The Option “/RC” can be used when CO analyzer is selected. When both “-C3” or “-C4”of the Gas Parameter is selected,
“/RC” must be specified.
*11: For Japan Ex model (TDLS8000-J1, TDLS8000-J2), specified cable glands shall be attached to each cable entry for wiring.
Select one cable gland out of two types: (/JB3 or /JB4 for “-J1”, /JE3 or /JE4 for “-J2”). If you need, specify (/JA1 for “-J1”, /
JC1 or /JD1 for “-J2”) as well. For detailed information, refer to Japanese General Specifications.
The Option “/JA1”, “/JB3” and “/JB4” can be used only when Japan Ex/Zone 1 model (TDLS8000-J1) is selected.
The Option “/JC1”, “/JD1”, “/JE3” and “/JE4” can be used only when Japan Ex/Zone 2 (TDLS8000-J2) model is selected.
If “/JA1”, “/JB3”, “/JB4”, “/JC1”, “/JD1”, “/JE3” or “/JE4” is necessary for other model, please contact Yokogawa.
*12: P is the process gas pressure. The option “/P2” and “/P3” can be selected with the gas parameter “-X1”. When the gas
parameter is not “-X1”, please contact YOKOGAWA.
*13: TDLS8000-G2-aa-bb-A1-c-N (Option)
-aa: -X1, -X2, -D1, -D5
-bb: -D5, -D8, -J5, -J8
-c: -N, -J
(Option): Only /D, /RX, /SCT can be selected.
A noise filter must be prepared. Refer to “ WIRING” on page 23.
Cable glands (M25, 3 pcs, Applicable cable diameter 9-17mm) are attached.
*14: Only when “-S2”, “-G1” or “-G2” for Type and “-A3” for Gas Parameter are selected, TDLS8000 is considered as QAL1
certified analyzer.
*15: Type “-GQ”, “-GR”, “-J2”, “-K2”, “-N2”, “-Q2”, “-R2”, “-U2” cannot be selected with Gas Parameter “-A3”.
*16: When “-A3” for Gas Parameter is selected, TDLS8000 cannot be used in combination with YC8000 or IF8000.

All Rights Reserved. Copyright © 2015, Yokogawa Electric Corporation GS 11Y01D01-01EN Apr. 10, 2025-01