1. The measure range of the loadcell is 10kg.

the weigh to calibrate the loadcell is over 1/10 of


the measure range.
2. Here is a 1KG weigh is used to calibrate the loadcell.
3. Short press K1 3 times to unlock.
4.Then long press K2, then shows 5 nos 0.
5.Press K2 to do 0 calibration, then press K4 to clear when the loadcell is empty.
6.Put the 1 kg weigh on the loadcell,it shows 840, then we should modify 840 to be 1000.
7.Short press K2,press K4 to shift digit, press K3 to increase and decrease the value.
8. press K2 to confirm,Press K1 to return. Calibration done.
9. put one more weigh of 200g,shows 1200. then the calbibration is right

the calibration select 1kg as a example

There are 5 weights to be calibrated, 1 pulverized coal bin and 4 injection tanks;
1. Verify the calibrated weight. The pulverized coal bin should have 4 weighing
sensors and the injection tank should have 3 weighing sensors. The weighing sensor
ranges should be determined on site. For example, if the pulverized coal bin has a 5t
range, the total weight of the pulverized coal bin is 5x4=20t. At least 2t of weight
standard weights should be prepared;
2. Prepare steel plates and standard weights. The steel plates are laid in the
pulverized coal bin and the injection tank (prepare 2 according to the size of the
tank). After zeroing, calibrate according to the video. Remove the steel plates and
then clear them before use. Alternatively, weigh the steel plates and zero them
before laying them, and then calibrate according to the video.

模拟量电流 4-20mA 设置。


在传感器跟仪表校准准确的情况下,下面我们来调一下模拟量 4-20mA 电流输出,万用表的黑表笔 COM 接仪表的 AO-,
我这边用的是黑线代替的,AO+我用的是红线代替的,接的是万用表的红表笔毫安 mA 档位,然后解锁状态下找到 27#,额定
量程,27#就是传感器的实际量程,传感器是 3kg 的,我们把他改成 3000,然后在找到 52#参数变送零点,注意看下万用
表显示,把他调到 4mA。再找到 53#参数,是变送满度,我们把它改为 20.00mA,燃烧后按第一个按钮保存,第二个按
钮退出,空称为 4mA,放一个 1kg 的砝码验证一下,1.5kg,12.00mA。准确的。
When the sensor and the meter are calibrated accurately, let's adjust the analog 4-20mA current
output. The black probe COM of the multimeter is connected to the AO- of the meter. I use a
black wire instead. I use a red wire instead of AO+, which is connected to the red probe mA
position of the multimeter. Then find 27# ( the rated range ) in the unlocked state, 27# is the
actual range of the sensor. The sensor is 3kg , We change it to 3000, and then find the 52#
parameter (transmission zero point ). Pay attention to the multimeter display and adjust it to
4mA. Find the 53# parameter, which is the transmission full scale. We change it to 20.00mA. Then
press the first button to save and the second button to exit. The empty is 4mA. Put a 1kg weight
to verify it. 1.5kg, 12.00mA. Accurate.