Siremobil Iso-C 3D Option

SP
Service Instructions SP034/03/R
Title: Option 3D, SW Update VB00D / VC00B

Reason for Update:

X Reliability

Urgency: Immediate X Within 12 months


Update material required? X Yes No
Materials free of charge? X Yes No
Return of parts? Yes X No
Estimated completion time: 2.5 hours Number of CSE’s: 1
Customer application training? Yes X No
Required for all systems under warranty or covered by service contract.
X Required for all affected systems.

Systems/Products affected/System identifying IVK

Name Material No. Serial No.


SIREMOBIL Iso-C 28 29 450 G5464 n.a.

Components affected/to be modified/IVK

Name Material No. Serial No. Component status


Affected
Option 3D Image 56 46 950 G5486 01001 - 01193, n.a.
01195-01197,
01200

Remark: n.a.

© Siemens AG 2003
Chg. Ref. No.: 130 013 The reproduction, transmission or
use of this document or its contents
Name: Arnold is not permitted without express
written authority. Offenders will be
Dept.: CS PS 24 liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.

Print No.: SPR2-230.898.03.01.02 Doc. Gen. Date: 12.03


Replaces: SPR2-230.898.01.01.02 Version 2.1; January 14, 2003
0-2 Revision

Chapter Page Revision


all all 01

Document Revision Level


This document corresponds to the version/revision level effective at the time of system delivery. Revi-
sions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.

Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such service.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho-
rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.

Siremobil Iso-C 3D Option SPR2-230.898.03 Page 2 of 4 Siemens AG


Rev. 01 12.03 CS PS 24 Medical Solutions
Contents 0-3

Page
1 _______General Information ____________________________________________ 1 - 1

Systems/Products Affected . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1


Reason for the Update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Special Tools / Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Contents of the Update Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Return of Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2

2 _______Work Steps ____________________________________________________ 2 - 1

Work flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1


Backup * . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
MEMOSKOP Software Update . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Flow Chart MEMOSKOP software update. . . . . . . . . . . . . . . . . . . . . .2-5
Prerequisites. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Saving the MEMOSKOP parameters to diskette . . . . . . . . . . . . . . . . . .2-7
Download MEMOSKOP software . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
Download MEMOSKOP with empty Flash Proms . . . . . . . . . . . . . . . . . 2 - 11
SIREMOBIL Iso-C 3D Software * . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 13
Installing the SIREMOBIL Iso-C 3D Software . . . . . . . . . . . . . . . . . . . 2 - 13
Start of the Syngo service software . . . . . . . . . . . . . . . . . . . . . . . . 2 - 15
Restore . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 17
Monitor settings * . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 18
Connectors *. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 22
Control software * . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 22
Manual Settings * . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 23
User Interface Language (not necessary for English) * . . . . . . . . . . . . . . 2 - 23
Country-specific Keyboard Layout (necessary for non US-Keyboard) * . . . . . . 2 - 24
Installation of the MSBLAST Hotfix . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 25
Final work steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 25
Covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 25
System and safety tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 25
Customer Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 25

3 _______Update Completion Form ________________________________________ 3 - 1

Completion Protocol / Update Completion Form . . . . . . . . . . . . . . . . . . . . 3 - 1

Siemens AG SPR2-230.898.03 Page 3 of 4 Siremobil Iso-C 3D Option


Medical Solutions Rev. 01 12.03 CS PS 24
0-4 Contents

Page

Siremobil Iso-C 3D Option SPR2-230.898.03 Page 4 of 4 Siemens AG


Rev. 01 12.03 CS PS 24 Medical Solutions
General Information 1 1-1

Systems/Products Affected 1

All SIREMOBIL ISO-C systems, delivered or upgraded with the Option 3D Reconstruc-
tion.

Reason for the Update 1

Improvement of the software, Bugfix MEMOSKOP 3D selection.

Prerequisites 1

NOTE This Service Instruction SP034/03/R replaces the withdrawn Ser-


vice Instruction SP029/03/R.
If the work, described in the withdrawn Service Instruction
SP029/03/R has not been performed for the existing system, do
not install it. Perform the work steps of this SI SP034/03/R.
If the work, described in the withdrawn Service Instruction
SP029/03/R were performed for the existing system, all work
steps, described in this SI have to be performed.

NOTE After the 3D reconstruction software was reloaded, the MSBLAST


Hotfix installation must be performed again.
Use the SI SP023/03/R and install the Hotfix.
Because the 3D reconstruction software is new installed, the
section "Checking for Virus / Worm" (page 4 of the SI SP023/03/R)
can be skipped.

Special Tools / Documents 1

Service PC Operating system Windows 95 / 98


Service cable for PC e.g. 99 00 440
1 set of X-ray filters,10 x 0.3 mm e.g. 44 06 120
17 mm Cu filter e.g. 11 67 662
Resolution test e.g. 28 71 820
Dynamic range test kit e.g. 37 90 156
Protective conductor meter e.g. 44 15 899
or Bender Safety Tester 97 06 979
SIREMOBIL ISO-C Service Instructions SPR2-230.061.01.03.
Speed Info Password list SIREMOBIL Iso-C R141-97
Service Instructions SP023/03/R SPR2-230.898.02.01.02
MSBLAST Hotfix on Floppy disk or CD-R (downloaded from Intranet,
see SP023/03/R)
1 blank CD-ROM
Image Quality Quick Test SPR2-230.037.01.05
(refer to the SIREMOBIL log book)
Calibration service case 75 51 620

Siemens Medical Solutions SPR2-230.898.03 Page 1 of 2 Siremobil Iso-C 3D Option


Medical Solutions Rev. 01 12.03 CS PS 24
1-2 General Information

Ordering Information 1

The following update kit has to be ordered from CSML (SAP Distribution Channel, factory
2050) or BU Logistics (factory BU) (only with an order via BU Logistics):
Update Kit 71 40 978

List the system serial number on the order!

Contents of the Update Kit 1

NOTE The Service Instructions SP034/03/R replaces the withdrawn Ser-


vice Instructions SP029/03/R.
The new Update Kits 71 40 978 includes the MEMOSKOP C-3D
and MEMOSKOP C-SUB-3D software version VC01H and two 3D
Software versions for the 3D reconstructions PC Celsius 650 and
Celsius 444.
The MEMOSKOP SW version VB00D works in combination with
both types of 3D PC (Celsius 650 and Celsius 444).

The update kit 71 40 978 contains the following parts:

Item Quan. Part No. Name


1 1 31 00 828 SW Memoskop C-3D VC01H
2 1 31 00 802 SW Memoskop C-SUB-3D VC01H
3 1 71 39 814 3D Software VB00D for 3D PC
Celsius 650, part no. 71 29 013
4 1 75 38 999 3D Software VC00B for 3D PC
Celsius 444, part no. 77 14 145
5 1 71 39 913 Control-SW VB00B
6 1 SPR2-230.898.03.01 Update instructions
Tab. 1

Based on the above description, the CSE should be able to check the update kit for com-
pleteness prior to beginning the update.

Return of Parts 1

n.a.

Siremobil Iso-C 3D Option SPR2-230.898.03 Page 2 of 2 Siemens Medical Solutions


Rev. 01 12.03 CS PS 24 Medical Solutions
Work Steps 2 2-1

Work flow 2

W o r k flo w

P e rfo rm B a c k u p *

S a v e P a tie n t *
Im a g e s o n M O D
o r C D -R O M

D o w n lo a d
MEMOSKOP
S o ftw a re

In s ta ll 3 D *
S o ftw a re

P e rfo rm R e s to re *

In s ta ll*
C a lib r a tio n
S o ftw a re

P e rfo rm *
M a n u a l s e ttin g s

End

* See note in section "Prerequisites", page 1-1.

Siemens AG SPR2-230.898.03 Page 1 of 26 Siremobil Iso-C 3D Option


Medical Solutions Rev. 01 12.03 CS PS 24
2-2 Work Steps

Backup * 2

NOTE A backup is always necessary after software updates or system


adjustment steps.

• From the Syngo User Interface, please select “Options>>Local Service” in the window
menu header.

• Enter the service key (6 characters in 2nd mask).


• Confirm with <OK>.

* See note in section "Prerequisites", page 1-1.

Siremobil Iso-C 3D Option SPR2-230.898.03 Page 2 of 26 Siemens AG


Rev. 01 12.03 CS PS 24 Medical Solutions
Work Steps 2-3
• The Service Home menu appears.

• Select “Backup & Restore”.


• Insert the BACKUP CD-ROM into the CD-R drive.

Siemens AG SPR2-230.898.03 Page 3 of 26 Siremobil Iso-C 3D Option


Medical Solutions Rev. 01 12.03 CS PS 24
2-4 Work Steps
¹ The following window appears:

• In the “Command” selection menu, select “Backup”.


• For “Drives”, select the CD-R storage medium.
• The corresponding package is selected in “Packages”.
Select the Package "SW Settings02" and save with "GO".
• Exit the window with “Home”.
• Remove the backup CD-ROM from the CD-R drive.

NOTE A backup of the database (patient images) does not function!


Only the following is possible:
Saving the patient images to a MOD or CD-R.

Siremobil Iso-C 3D Option SPR2-230.898.03 Page 4 of 26 Siemens AG


Rev. 01 12.03 CS PS 24 Medical Solutions
Work Steps 2-5

MEMOSKOP Software Update 2

Software 2

for MEMOSKOP C-Sub 31 00 802 G5437


for MEMOSKOP C 31 00 828 G5437

Flow Chart MEMOSKOP software update 2

The flow chart SW update MEMOSKOP shows only the main working steps. In the case
of problems during the MEMOSKOP software update, please follow the path of the flow
chart.

SW Update
MEMOSKOP

Install
MEMOSKOP
Service SW to the
Service PC
Check Service PC
settings. See ,
prerequisites
Saving the
MEMOSKOP
paramers to
diskette

NO
worked well?

Yes Tried all NO


prerequisites ?

Yes
Read out all
parameter settings
(User and
technical setup)
and note them.

Download
MEMOSKOP
Software

NO
worked well?

Check Service PC
Yes settings. See
Prerequisites.

Download
according to
"Download
Memoskop with Please try several
empty Flash times
Proms

NO
worked well?

Yes

END

Siemens AG SPR2-230.898.03 Page 5 of 26 Siremobil Iso-C 3D Option


Medical Solutions Rev. 01 12.03 CS PS 24
2-6 Work Steps
Prerequisites 2

NOTE Please pay attention to the service PC settings.

- Do not use the incompatible service PC "SCENIC MOBILE 350".


Before starting with the software updates, please observe the points mentioned below:
- The used serial interface cable must be fully wired between SIREMOBIL Iso-C and the
service PC.
- Switch off the screen saver and end any programs running in the background.
- Check the jumpers D1.X95 / D1.X96 / D1.X 99. These should be set to position 2-3.
- Check the serial port FIFO settings of your service PC. The FIFO of the serial port must
be enabled (default settings).
- In the case of correct host communication, but poor MEMOSKOP communication,
temporarily switch D1.S3.2 to position ON. Ignore the error messages displayed at the
SIREMOBIL user console. Ignore the MEMOSKOP service software error message
"No communication with the X-ray system" and close Error message by clicking the
"Ignore" button. Select the correct type of MEMOSKOP and start the necessary
service work. After completion of the MEMOSKOP service work, switch D1.S3.2 back
to OFF.

Covers 2

• Remove the back cover from the monitor trolley (important for installing MEMOSKOP
software).

Siremobil Iso-C 3D Option SPR2-230.898.03 Page 6 of 26 Siemens AG


Rev. 01 12.03 CS PS 24 Medical Solutions
Work Steps 2-7
Saving the MEMOSKOP parameters to diskette 2

NOTE Please pay attention to the prerequisites.

NOTE Before starting the download of the MEMOSKOP, save the para-
meters to diskette. In the case of transfer problems, saving them,
read the prerequisites .
Follow the flow chart in section "Flow Chart MEMOSKOP soft-
ware update".
If the parameter set can’t be stored on the diskette, read out the
MEMOSKOP parameters (User setup and Technical setup) and
note them manually.

NOTE If a transfer error or error messages with "ComPort", "Download",


"Get from Memoskop", "Put to Memoskop" and when loading the
language file still occur, please repeat this function several times.
If the problem persists, then exit the service software.
Switch SIREMOBIL off and then on again, wait for power-up and
start the service software again. Perform the last work step again.

NOTE Before using the host service software or MEMOSKOP service


software, the write protection of the corresponding parameter
diskettes have to be switched off

• Insert the MEMOSKOP diskette "User Programs" into the service PC, drive A.
• Start the MEMOSKOP service software.
• Enter the password and press the Return key on the service PC.
• Select the required serial interface on the service PC.

Siemens AG SPR2-230.898.03 Page 7 of 26 Siremobil Iso-C 3D Option


Medical Solutions Rev. 01 12.03 CS PS 24
2-8 Work Steps
• The program window "Memoskop Service Program" (Fig. 3) is then displayed.

Fig. 1

• After the main menu appears, click on the "Get from Memoskop" button and wait until the
parameters have been transferred from the MEMOSKOP to the PC. After the transfer,
the status line shows "....Receiving configuration done"
• Then save the parameters to the diskette (menu "File", submenu "Save").
• For the file name, enter the actual date. The file extension should be "mem".
Example:
The current date is: (DD_MM_YY) 24_12_03
Enter the file name+file extension: 24_12_03.mem
• Wait for the parameters to be saved on diskette.

NOTE If errors occur during the transfer of the MEMOSKOP parameters


to the service PC (e.g. communication errors), read the prerequi-
sites.

NOTE If the MEMOSKOP service program must be started again, the


unit must also be switched off and on again.

• If the storage of the MEMOSKOP parameter set to diskette is not possible, all para-
meters (User setup and technical setup) must be read and noted manually.

Siremobil Iso-C 3D Option SPR2-230.898.03 Page 8 of 26 Siemens AG


Rev. 01 12.03 CS PS 24 Medical Solutions
Work Steps 2-9
Download MEMOSKOP software 2

NOTICE If errors occur during download of the MEMOSKOP software


(e.g. communication errors) read the prerequisites.
In this case, stop the download procedure and try download as
described in section "Download MEMOSKOP with empty Flash
Proms" if needed again.

• For the MEMOSKOP displayed insert the corresponding diskette supplied into drive A of
the service PC (MEMOSKOP software).
• Select the "Download" menu.
• Select "Select file for Download:".
• The program window "Open File" opens.
• Select drive A.
• Select the file xVC0xxx.mot and confirm the selection with "OK".

(first x = 3 or 4, depending on existing MEMOSKOP. See note below.)


3 MEMOSKOP C-E 100
MEMOSKOP C Select 3VC01H0.mot
MEMOSKOP C-MOD
4 MEMOSKOP CSUB
MEMOSKOP C-SUB &MOD Select 4VC01H0.mot

NOTE If the service program is not able to recognize the MEMOSKOP


installed automatically, select the type of MEMOSKOP manually.
For MEMOSKOP C-E 100, MEMOSKOP C and
MEMOSKOP C-MOD, select model 3.
For MEMOSKOP C-SUB and MEMOSKOP C-SUB & MOD,
select model 4

Fig. 2

Siemens AG SPR2-230.898.03 Page 9 of 26 Siremobil Iso-C 3D Option


Medical Solutions Rev. 01 12.03 CS PS 24
2 - 10 Work Steps
• Select "Erase FLASHes" .
¹ While the Flash Proms are being erased, dots will be displayed in the status line.
¹ After erasing, the "...Ready" message will be displayed in the status line.
¹ The DOWNLOAD button is now selectable.

NOTE The "Download File" button must be pressed (clicked) within


120 seconds maximum!

• Select "Download File" .

NOTE It will take approximately 20 minutes to load the new


MEMOSKOP software.

• After successful completion of the Download, the "Download complete" message will be
displayed in the status line.

NOTE If the Download is not performed completely (communication


errors or undefined symbols in the status line), stop the down-
load procedure and follow the description in section "Download
MEMOSKOP with empty Flash Proms".

• Exit the Service program.


• Switch the unit off .

Siremobil Iso-C 3D Option SPR2-230.898.03 Page 10 of 26 Siemens AG


Rev. 01 12.03 CS PS 24 Medical Solutions
Work Steps 2 - 11
Download MEMOSKOP with empty Flash Proms 2

NOTE This section describes the downloading procedure for the


MEMOSKOP with empty Flash Proms. If downloading as
described in the previous section could be carried out success-
fully, skip this section.

• Switch the unit off and exit the Memoskop service program.
• The cover of the basic unit and the rear cover of the monitor trolley are removed.
• Set the service switch D1.S3.2 to the ON position.
• Switch on the unit. Ignore the error messages displayed at the user console.
• Insert the corresponding diskette supplied into drive A of the service PC (MEMOSKOP
software ).
• Start the MEMOSKOP service program and enter the password.
• Select the service PC serial interface.
• Ignore the message "No communication with the X-Ray system by clicking the "Ignore"
button.
• For the error message "The Program didn’t receive the version message from the
MEMOSKOP", manually select the MEMOSKOP model.
3 MEMOSKOP C-E 100
MEMOSKOP C
MEMOSKOP C-MOD
4 MEMOSKOP C-SUB
MEMOSKOP C-SUB & MOD

Fig. 3

• Select the "Download" menu.


• Select "Select file for Download:".
• The program window "Open File" opens.
• Select drive A.

Siemens AG SPR2-230.898.03 Page 11 of 26 Siremobil Iso-C 3D Option


Medical Solutions Rev. 01 12.03 CS PS 24
2 - 12 Work Steps
• Select the file xVC0xxx.mot (Fig.and confirm selection with "OK".
(first x =3 or 4, depending on existing MEMOSKOP. See below.)
3 MEMOSKOP C-E 100
MEMOSKOP C Select 3VC01H0.mot
MEMOSKOP C-MOD
4 MEMOSKOP C-SUB
MEMOSKOP C-SUB & MOD Select 4VC01H0.mot

• Select "Erase FLASHes" .


• Press the reset button on the back of the MEMOSKOP for 4 seconds as soon as
possible.
¹ While the Flash Proms are being erased, dots are displayed in the status line.
¹ After erasing, the "...Ready" message is displayed in the status line.
¹ The DOWNLOAD button is now selectable.

NOTE The "Download File" button must be pressed (clicked) within


120 seconds maximum!

• Select "Download File" .

NOTE It will take approximately 20 minutes to load the new


MEMOSKOP software.

• After successful completion of the download, the "Download complete" message is


displayed in the status line.
• Wait until the MEMOSKOP has rebooted.
• Exit the Service program.
• Switch off the unit.
• Switch the service switch D1.S3.2 back to the OFF position.
• Switch the unit on (wait for entire power-up sequence) .

NOTE If the download is not performed completely (communication


errors or undefined symbols in the status line), read the pre-
requisites.
Stop the download procedure, end the Memoskop service pro-
gram, switch off the unit and repeat the download procedure as
described in section "Download with empty Flash Proms".

Siremobil Iso-C 3D Option SPR2-230.898.03 Page 12 of 26 Siemens AG


Rev. 01 12.03 CS PS 24 Medical Solutions
Work Steps 2 - 13

SIREMOBIL Iso-C 3D Software * 2

Installing the SIREMOBIL Iso-C 3D Software 2

• The SIREMOBIL Iso-C is still running.


• With 3D reconstructions PC Celsius 650, part no. 71 29 013,
insert the CD with part no.71 39 814 ("3D Software") into the CD-R drive.
• With 3D reconstructions PC Celsius 444, part no. 77 14 145,
insert the CD with part no. 75 38 999 ("3D Software") into the CD-R drive.
• Shut down the SIREMOBIL Iso-C and wait until power is off.
Video switch 1 button

"Memoskop SCSI Adapter"

"power plug"

• Disconnect the "Memoskop SCSI Adapter" plug from the PC.


• Disconnect the "power plug" from the MEMOSKOP.
• Release the video switch 1 button (to the left as viewed from the back of the monitor),
the switch should not be pressed.

* See note in section "Prerequisites", page 1-1.

Siemens AG SPR2-230.898.03 Page 13 of 26 Siremobil Iso-C 3D Option


Medical Solutions Rev. 01 12.03 CS PS 24
2 - 14 Work Steps

"Memoskop SCSI Adapter"

"power plug"

DICOM Network Plug

• Disconnect the "DICOM Network Plug" from the 3D PC.


• Connect the Dongle, if not already connected.
• Restart the SIREMOBIL Iso-C.
¹ The PC will start automatically. The start-up can be viewed on the right monitor.
¹ After approx. 15 minutes, the following message will appear on the monitor.

= Installation complete …===========================


|| System and image partition were set up.||
|| Please make sure there is no disk in the floppy drive||
|| (Press any key to reboot the system now)||
|| IMPORTANT : While the system is rebooting, eject||
|| the CD-ROM and remove it from the drive !||
===============================================
• Press the space key and remove the CD. The 3D PC will reboot.
• Wait till the PC bios messages are visible at the monitor, open the CD-ROM drive and
remove the CD-ROM.

Siremobil Iso-C 3D Option SPR2-230.898.03 Page 14 of 26 Siemens AG


Rev. 01 12.03 CS PS 24 Medical Solutions
Work Steps 2 - 15
• Wait until the following screen appears.
Ignore the message concerning monitor settings.

• Enter the 3D PC name.


• Press the "Enter" key 2x.
¹ The computer will shut down and restart automatically.
The Internet Explorer will start; the following screen will appear

Start of the Syngo service software 2

• Select the line "Click here to start the Local Service Configuration".

Siemens AG SPR2-230.898.03 Page 15 of 26 Siremobil Iso-C 3D Option


Medical Solutions Rev. 01 12.03 CS PS 24
2 - 16 Work Steps
¹ The "Service Software" screen will appear.

• Enter the service key (6 characters in 2nd mask).


• Select "set as default" and confirm with <OK>.
¹ The Service Home menu appears.

Siremobil Iso-C 3D Option SPR2-230.898.03 Page 16 of 26 Siemens AG


Rev. 01 12.03 CS PS 24 Medical Solutions
Work Steps 2 - 17
Restore 2

• Select the “Backup & Restore” button in the Service Home menu.
• Insert the BACKUP CD-ROM into the CD-R drive.

• Select "Restore" in the “Command” list box.


• Select "CD-R" in the "Drives" lsit box.
• Select the correct (latest) "SW Settings02" backup file in "Archives".
• Mark all files in "Groups" i.e. (with Ctrl + left mouse button).
• The restore routine is started by clicking on the "Go" button.
¹ Wait until the "Ready" message appears in the footer. This will take a few
minutes.
• Exit the window with “Home”.
• The PC initiates a restart automatically or you must restart it manually.
• Remove the backup CD-ROM from the CD-R drive when the system has shut down.

Siemens AG SPR2-230.898.03 Page 17 of 26 Siremobil Iso-C 3D Option


Medical Solutions Rev. 01 12.03 CS PS 24
2 - 18 Work Steps

Monitor settings * 2

from the Syngo User Interface.


• Select "Option".
• Select "Local Service.
¹ The Service Home menu appear.

• Select "Utilities".

• Under "Source," select the item "Unlock system key".


• Select "OK".
* See note in section "Prerequisites", page 1-1.

Siremobil Iso-C 3D Option SPR2-230.898.03 Page 18 of 26 Siemens AG


Rev. 01 12.03 CS PS 24 Medical Solutions
Work Steps 2 - 19

NOTE After exiting the program, the Windows task bar is closed auto-
matically. To display the task bar again, press the key combina-
tion "CTRL" + "ESC".
The "Matrox Display Properties" program is closed automatically
after each setting and selection of the OK button. The program
must be restarted for the next settings (Start - Programs - Matrox
Power desk NT - Matrox Display Properties).

• Press the "CTRL" + "ESC" keys simultaneously.


• Click "Enter".
¹ The WINDOWS task bar will appear.
• Select "Start".
• Select "Programs".
• Select "Matrox Power Desk NT".
• Select "Matrox Display Properties".
(If it is not displayed in front, move the other displayed window(s) to the side.)

• Select the "Monitor" button.


• Select "Matrox Monitor".
• In the selection window, select "Siemens" by double-clicking.
• Select the "Medical Simomed H @ 72 Hz" monitor.
• Select "Apply".

Siemens AG SPR2-230.898.03 Page 19 of 26 Siremobil Iso-C 3D Option


Medical Solutions Rev. 01 12.03 CS PS 24
2 - 20 Work Steps

• Confirm by clicking "OK".

• Confirm the message by selecting "YES".


¹ The monitor resolution has been changed over.

• Under "Current display", select 2.


• Select "Output to TV (PAL)".
• Confirm by clicking "Apply".

Siremobil Iso-C 3D Option SPR2-230.898.03 Page 20 of 26 Siemens AG


Rev. 01 12.03 CS PS 24 Medical Solutions
Work Steps 2 - 21

• Confirm by clicking "OK".

• Confirm by clicking "Yes".

• Select "Settings".
• Select "True Color (32-bit)".
• Select "72 Hz" refresh frequency.
• Confirm by clicking "OK", even if the other settings are correct.
• Select "Start".
• Select "Shutdown".
• Select "OK" for shutdown and wait until the WINDOWS message "It is now save to..."
appears.
• Shut down the SIREMOBIL Iso C 3D and wait until the SIREMOBIL Iso C 3D has shut
down completely.

Siemens AG SPR2-230.898.03 Page 21 of 26 Siremobil Iso-C 3D Option


Medical Solutions Rev. 01 12.03 CS PS 24
2 - 22 Work Steps

Connectors * 2

• Connect the MEMOSKOP power plug.


• Connect the SCSI plug to the MEMOSKOP.
• Connect the DICOM Network connector at the 3D PC
• Press the video switch 1 button (to the left as viewed from the back of the monitor),
the button is now depressed and hit up.
• Switch on the SIREMOBIL Iso-C 3D.
¹ The MEMOSKOP image will appear on the right monitor.
• Wait approx. 2 - 3 minutes until the PC has rebooted.
• Press "CTRL" + "D" to display the PC image on the right monitor.

Control software * 2

NOTE Check the installed Control SW version (Syngo menu- "Help" -


"About")
If the Control SW version VB00B is installed, it is not more
needed, installing the Control SW VB00B again and you can skip
this section.

• The System is still running.


• In Syngo, select the menu "Help" / "About". Read the Control SW version. If the Contol
SW version VB00B is installed, skip this section. Continue with section "Manual
settings".
• In Syngo, start the local service (menu "Option" / "Local Service").
• Enter the password.
• Select the "Upgrade Ctr.SW" button.
• Insert the CD-ROM "Control Software VB00B" into the CD-R drive.
• Select the "CD-ROM" drive.
• Click on the "Go" button.
• Wait until the message "download successful" appears.
• Remove the CD-ROM from the CD-ROM drive.
¹ The Reconstruction PC will shut down automatically.
• Switch off the SIREMOBIL Iso-C 3D.
• Restart the SIREMOBIL Iso-C 3D.

* See note in section "Prerequisites", page 1-1.

Siremobil Iso-C 3D Option SPR2-230.898.03 Page 22 of 26 Siemens AG


Rev. 01 12.03 CS PS 24 Medical Solutions
Work Steps 2 - 23

Manual Settings * 2

NOTE The changes become effective only after a reboot is performed!

• The system is still running.


• The regional settings should be performed manually.

User Interface Language (not necessary for English) * 2

NOTE A selection can be made between English, German, Spanish and


French

• From the Viewer Interface select "<Options> <Configuration...>".


¹ The following window appears:

• Double-click on "Regional Settings".

• Select the desired language here and confirm with "OK".


• Confirm the prompt for a restart with "OK" to accept the settings.
* See note in section "Prerequisites", page 1-1.

Siemens AG SPR2-230.898.03 Page 23 of 26 Siremobil Iso-C 3D Option


Medical Solutions Rev. 01 12.03 CS PS 24
2 - 24 Work Steps
Country-specific Keyboard Layout (necessary for non US-Keyboard) * 2

NOTE It is possible to choose between an English, German and French


keyboard layout.

The following window appears:

• Double-click on "Regional Settings".


• Select "Input Locales".

• Double-click on one selection.

* See note in section "Prerequisites", page 1-1.

Siremobil Iso-C 3D Option SPR2-230.898.03 Page 24 of 26 Siemens AG


Rev. 01 12.03 CS PS 24 Medical Solutions
Work Steps 2 - 25

• Select the layout that corresponds to the keyboard that has been shipped in
"Keyboard layout".
• Click "OK".
• Click "OK".

Installation of the MSBLAST Hotfix 2

NOTE Because the 3D reconstruction software is new installed, the


section "Checking for Virus / Worm" (page 4 of the SI SP019/03/R)
can be skipped.

• Reinstall the MSBLAST worm blaster security update as described in the


Service Instructions SP023/03/R.

Final work steps 2

Covers 2

• Close all covers.

System and safety tests 2

• Perform a final check. Check of the overall function of the system.


• Perform the Image Quality Quick Test (refer to the SIREMOBIL logbook).
• Perform the protective conductor test in accordance with ARTD - 002.731.17.
The protective conductor resistance must not exceed 0.2 ohms.
• File the update instructions in the System Binder or the System Logbook.
• Fill out, and if needed, copy the attached "Update Completion Form" and insert it into
Register 10 of the System Binder or System Logbook.

Customer Information 2

n.a.

Siemens AG SPR2-230.898.03 Page 25 of 26 Siremobil Iso-C 3D Option


Medical Solutions Rev. 01 12.03 CS PS 24
2 - 26 Work Steps
This page intentionally left blank.

Siremobil Iso-C 3D Option SPR2-230.898.03 Page 26 of 26 Siemens AG


Rev. 01 12.03 CS PS 24 Medical Solutions
Update Completion Form 3 3-1

Completion Protocol / Update Completion Form 3

The update with the number SP034/03/R has been completed.

Material number: ................................

Serial number: ................................

Customer: ................................ Functional Location:................................

Customer No.: ................................

Name ( CSE ): ................................ Telephone: ................................

Country: ................................ Location: ................................

Date: ................................ Signature: ................................

Remark: ....................................................................................................

....................................................................................................

NOTE After completing the update, make a copy of this page, fill it out
and file it in the corresponding System Binder/User Handbook.

Siemens AG SPR2-230.898.03 Page 1 of 2 Siremobil Iso-C 3D Option


Medical Solutions Rev. 01 12.03 CS PS 24
3-2 Update Completion Form
This page intentionally left blank.

Siremobil Iso-C 3D Option SPR2-230.898.03 Page 2 of 2 Siemens AG


Rev. 01 12.03 CS PS 24 Medical Solutions