Fluorospot H Mai - Ins.
Fluorospot H Mai - Ins.
AX
Maintenance Instructions
FLUOROSPOT H/HK with VID. SX
Register 8 English
Print No.: RX41-020.101.04.01.02 Doc. Gen. Date: 05.97
Replaces: RX41-020.101.02.01.02
0-2 Revision
Page
1 _______Requirements__________________________________________________ 1 - 1
2 _______VIDEOMED SX _________________________________________________ 2 - 1
Safety Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
1 Requirements
2 VIDEOMED SX
2.2 Maintenance
PM Camera Optics
• Loosen the knurled screws on both sides of the camera mount and remove the camera
with the mount (Caution: do not expose the camera optics to bright light).
CAUTION Check all optically active surfaces (I.I. lens, mirror, TV camera
lens and multiplier lens) for dust and dirt and, if needed, clean.
Remove dust with the brush; when doing so, do not exert any pressure on the brush.
Carefully wipe away any residue with a suitable cloth.
Breathing on the surface can be helpful; do not get any saliva on optical surfaces.
For further information, see Service Note RA0-000.076.17...
PM CCU
• Remove any dust that has collected in the CCU.
3 FLUOROSPOT H/HK
3.2 Maintenance
PM Opening the Unit
PM Remove dust that has collected on the inside of the unit
3.4 Monitors
PM Cleaning the Monitor Screens
- To clean the non-glaring monitor screen, which is recognizable by the mat surface, use
a cloth dampened with plain water .
Clean any stubborn dirt with pure alcohol or a mixture of 1/3 alcohol with 2/3 distilled
water.
- Immediately after cleaning with the items described above, wipe the screen dry with a
soft cotton cloth.
- Under no circumstances should acidic or caustic agents be used. These can damage
the picture tube.
- LIST FILES displays a list of the error files in the Service PC.
- A file can be selected for view or transferred to a floppy disk using SELECT FILES.
- TYPE FILE displays the selected error protocol on the Service PC screen.
- TRANSFER FILE TO FLOPPY transfers a selected file to a floppy disk.
- A floppy disk can be formatted from this menu by using FORMAT 1.44Mb FLOPPY.
• Disconnect the Service PC.
Test Conditions
• Select the operating conditions documented in the IQ Test Certificate:
- Focus-I.I. distance (SID)
OT units = 115 cm
UT units = max. distance (without Distator)
C-Arm units = 100 cm
- Insert the 2.1 mm precision radiation filter into the collimator; with UT units, place it on
the tabletop.
- Open the collimator all the way.
Procedure
• Trigger fluoro : With full format for all automatic steps
With the other I.I. formats only with Fluoro 1 / Step 1
NOTICE Step 1/Step 2 = 1st/2nd footswitch pressure point for each auto-
matic control step (currently configured only in the POLYSTAR)
• Document the kV and mA values that stabilize (display on the generator console). The
same values as those listed in the column “Initial Values” must result (tolerance: ± 1kV/
±10% mA). If there are differences that are outside of the admissible tolerance,
remeasure the dose rates as described under Point 4.1 of the Test Procedure,
RXD0-000.074.01...
If the dose rate values are within tolerance (IQ Reference Value List), determine new
initial values ("-"/"Act"/"+") for indirect dose rate control; this also applies if the dose rate
has been set again.
- Insert the 2.1 mm precision radiation filter into the collimator; with UT units, place it on
the tabletop.
- Move in the grid.
Procedure
• Select the ADC measuring field; with the 40 cm I.I., select the central measuring
field
(POLYSTAR )
• Open the collimator all the way.
• Connect the mAs meter to the generator (high voltage transformer).
• Prior to the DR exposures, briefly trigger fluoro so that the defined exposure data can be
acquired.
• To have the reproducible exposure parameters defined in the DSA scenes acquired,
proceed as follows:
1. Select a scene using the desired organ program.
2. Switch on fluoro for approx. 5 sec. to allow the ADC to stabilize and to allow the
water value calculator to provide the correct exposure data.
3. Trigger an initial brief scene (approx. 5 images) to allow the dose to stabilize in the
constant time mode.
4. Then trigger the actual test scene without first triggering fluoroscopy and evaluate
the scene. Trigger the first brief scene.
CAUTION The 5% fields must always be visible even if the tolerance ranges
are slightly exceeded.
• ADC measuring field; with the 40 cm I.I., select the central measuring field
(POLYSTAR
• I.I. format=28 cm, 27 cm, 23 cm or 22 cm
• Move the Dynamics Test close to the I.I.
(without heart contour collimator, without capillary displacement test)
OT units on the tabletop
UT units on the back wall of the spotfilm device
C-Arm units on the grid
NOTICE The reference values for the B/B amplitudes can be reached only
if the AGC measuring field is covered over completely by the 1st
field in the dynamics test. To make the measuring field visible,
jumper between the TP17/KR board and the TP8/VV board.
If the AGC measuring field is larger than the 1st field, move the
dynamics test so that the 2L field is completely outside the AGC
measuring field and only the 5R field projects into the ACG mea-
suring field.
The 5R field adversely affects the measurement much less than
the 2L field
Essentially, the same applies for the rectangular ADC measuring
field for the POLYSTAR
Collimation of
collimator
Fig. 1
Procedure
• Switch fluoroscopy ON/OFF.
Measure one line in the middle of the screen:
- B-Signal (TP8) in fields 2L / 1/ 5R without bias light
- BA-Signal (TP BAS1) in fields 2L / 1 / 5R
• Read the fluoroscopy kV during fluoro at the generator console and document the value..
2L
200mV/Div. 500mV/Div.
2L
1 1
5R
5R
Field 2L
f=
Field 1
• Compare the calculated dynamics factor with the Fig. on Page 10 (Maintenance
Protocol).
The dynamics factor must be above the illustrated curve, relative to the read-out
fluoroscopy kV.
CAUTION No additional filter may be selected in the collimator for the mea-
surements with the dynamics test! The reference values are rela-
tive to a tube current of 70 kV.
Amplitudes
The BA signal amplitudes of the various attenuation steps (fields 1, 2L, 5R) of the dynam-
ics test are measured at a tube current of 70 kV at the FLUOROSPOT output. The
required BA-levels are reached by correctly adjusting the iris diaphragm.
Test Conditions and Preparations
• Focus-I.I. distance (SID):
OT units = 115 cm
UT units = max. distance (without Distator)
C-Arm units= 100 cm
2L
100mV/Div.
5R
Line in middle of
CAUTION The iris correction must be =0 at the generator console and in the
organ program.
CAUTIION The iris correction must be =0 at the generator console and in the
organ program.
Procedure:
Carry out a check in the following modes with each of the I.I. formats.
1 Dig.
Hardcopy
Dig. Fluoro Exposure
• At the FLUOROSPOT, select the test programs that are described in the text below while
taking the above-listed test conditions into consideration.
• Determine the maximum visible resolution and document the result (darken the room,
avoid glare).
• Check whether or not the 17 µm Cu strip (minimum contrast) is visible; documentation is
made using either a yes or no answer.
The minimum contrast is more easily detectable at a greater viewing position.
Q Checking the DSA Device (IQ Chap. 8.9)
• Using the test exposures with the dynamics test, check the most important image
properties from the DSA sensitivity – contrast sensitivity, logarithming and subtraction.
For this, evaluate the DSA scenes that were made.
Evaluation Conditions
• For evaluation, it is necessary to set the mask on an image (if needed, using REMASK),
on which it is certain that the stabilization procedure for dose has been completed:
- at 6 F/s on Image 8
• Window values: brightness W1 = 80; contrast W2 = 220
• No edge enhancement; E (%) = 0
• Always evaluate the next to last DSA image in the series.
Determining the Initial Values
• If the contrast sensitivity has already been documented in the IQ Test Certificate using
the new capillary test (Part No. 37 80 180), transcribe the values into the Maintenance
Protocol, column ”Initial Values”.
If the contrast sensitivity has been documented in the IQ Test Certificate using the old
iodone capillary test (Part No. 97 50 019), the initial values for the Maintenance Protocol
must be redetermined using the new capillary test (Part No. 37 80 180) per the IQ Test
Procedure (Chapter 8.9).
• Connect the Service PC to the FLUOROSPOT and start the ST 320 program.
• Select the DSA mode at the FLUOROSPOT (not native image).
• At the FLUOROSPOT, select PAT-DIR, TOOLS, SERVICE.
Enter the following at the Service PC:
Enter the password
- "IMAGE QUALITY ASSURANCE"
- "SELECT FOLDER FOR ANALYSIS",
Enter the folder name of the DSA scene.
- "SELECT IMAGE FOR ANALYSIS",
The split image is selected. Use the PC arrow keys (left, right) to select an image.
- "SELECT IMAGE FOR ANALYSIS",
The selected image is displayed.
- "DYNAMICS BOX",
The rectangular measuring field appears on the FLUOROSPOT monitor.
- "MOVE BOX"
Use the PC arrow keys to move the rectangle to the middle of the 3 mm capillary.
Make sure that the full width of the capillary is always inside the DYNAMICS BOX
measuring field.
1
8 8 66
2
-1 0 -1
Fig. 2
Maximum difference between the gray step of the 1st field and the gray step of the 2L field
or of the 5R field = 3 gray steps.
Maximum tolerance to the initial values ± 1 gray step.
Subtraction:
Subtraction is okay when all Cu steps are displayed with the same gray value (2/Fig.2).
• Service PC: use the arrow keys to move the rectangular measuring field between two
capillaries.
• Measure the basic gray value of fields 2L, 1 and 5R and document the values. The basic
gray values must be 0, ± 1; in the recurring tests, it must be 0, ± 2.
Description of Errors:
- Pixel Errors are image pixels which do not contain image information. They are visible
as dark, pixel-size dots in the basic image. In the DSA image, they can occur as black
or white dots, depending on whether they are missing in the mask or in the fill image.
The cause is memory dropouts or addressing errors which occur in the RAM or in the
mass storage device (Post Processing).
Tolerable pixel dropouts are sporadic, do not occur in one location, are always 1 pixel
at one location, maximum of 3 pixels/image and occur at most in every 20th image.
- Ghost Images: are object shapes which are usually offset to one side, and appear
double. They are usually caused by reflections in poorly adapted, long video cables.
Clearly visible ghost images are not tolerable.
- Contouring are horizontal or vertical ”stripes” which mimic linear structures which do
not exist. They can be best detected on an image generated with a sawtooth signal by
continuously changing the windowing. Slight contouring in the above-described test is
tolerable.
- Horizontal Jitter: vertical edges of objects are not displayed on the monitor as “sharp”
borders but as slightly jagged, “frayed” lines. Jitter is caused by slight irregularities in
the horizontal deflection from image line to image line or in image storage and to a
certain extent are unavoidable. In the DSA image it leads to slight, tolerable coverage
errors.
To carry out the check, tape a soldering wire cross on the middle of the I.I., as exactly
vertical and horizontal as possible and make a DR exposure at dose step 100 and as a
single image at the smallest zoom format.
- Island Formation: We speak of island formation when visibly delineated areas of the
image have an even gray value and generally appear without noise. They often occur
because of insufficient memory, etc. and accordingly contain incorrect image
information. What are called ”burned out” image locations also fall into this category.
They are tolerable only to a very limited extent.
- Vignetting always appears as a decrease in brightness from the center of the image to
the edge of the image; it is caused by the heel effect of X-radiation, by the curve of the
I.I. input screen, by the electron optics in the I.I., by the tandem optics between the I.I.
and the TV camera, by the TV camera tube and by the monitor picture tube. It is
considered to be an image disturbance only if it occurs too strongly asymmetrical (see
also Test Point 5.4 in the IQ Test Certificate), or when greater differences in brightness
appears irregularly distributed over the image area.
If it is below the tolerance limit listed under Point 5.3 or Point 8.4.3 with fluoroscopy, it
should be evaluated as an ”image disturbance” with evaluation numeral 1.
- Artifacts caused by exposure occur only in DSA images and are caused by
fluctuations in exposure from image to image during the DSA scene.
If two completely equal images (mask and fill image) are subtracted from each other,
an evenly bright, completely structureless image results in which only noise is visible. If
the exposure changes between mask and basic image, an apparent image structure
caused by the difference in exposure is created in the DSA image, even when the
object is unchanged. It is acceptable only when it is slightly apparent because a similar
effect caused by patient movement can occur in medical exposures.
To carry out the check, make DSA test scenes at dose step 200; 6 F/s and dose step
500; 1 F/s with the soldering wire cross (see the check for horizontal jitter) and always
evaluate the entire scene. When doing this, place the mask on an image for which the
dose control has been completed with certainty.
- Noise Patterns: All noise patterns which were not described above and whose cause
is not known should be entered here.
Disturbances which are caused by dirt on optically active surfaces should be entered
here. They are limited to one location and are not temporary.
As already mentioned, these disturbances are scarcely tolerable.
Line test
The individual elements of the line test make it possible
to make the following statements about the quality of
the tomo device or about the tomo exposures:
Line groups (LG)
- Visual resolution
- Continuity of blurring
- Selectivity (“tomo thickness” depending on object
size)
- Degree of blur (angle, tomo profile)
Holes, Notch
- Tomo height (tomo plane)
- Tomo profile
- Evenness of the tomo movement
(synchronous movement, smooth movement)
- Dose distribution over the exposure
- Parallelism of focus and film stage
Edge enhancement
- Continuity of blurring
(even or uneven tomo plane)
Procedure:
• Position the tabletop so it is horizontal.
• Center the foam plastic block with the focus test on the tabletop in the beam path. The
focus test must be positioned at the anode side transversely to the table longitudinal
axis.
The notch of the focus test must be in the central beam (see Fig.3).
Anode side
20°
2mm
approx. 115 mm
SHa
TV Monitor
Fig. 3
• Measure exactly at which height the notch in the focus test is located above the tabletop
(approx. 115 mm, depending on the foam plastic bock size).
Evaluation:
• The hole of the line light test appears sharply focused, it is in the tomo plane.
If the hole next to the notch appears sharply focused, the numerically displayed tomo
height agrees with the actual tomo height, i.e. the deviation = 0.
• If there is a deviation, determine the difference in height in mm between the sharply
focused hole and the notch (there is a difference in height of 2 mm from hole to hole).
• Calculate the actual tomo height and document the value (transverse test):
CAUTION When doing this, take into consideration whether the tomo height
is above (+ deviation) or below (– deviation) the selected tomo
height (notch)!
The displayed tomo height or the line light height may not deviate by more than 5 mm
from the actual tomo height..
Test Conditions:
Evaluation of the resolution of a tomo exposure is made by comparing it to an exposure
without tomo which has been prepared under comparable conditions. In this way, all fac-
tors which influence the resolution of an exposure (focal point size, geometric ratios,
exposure data, ...) are cancelled out and it is possible to make a clear statement about the
quality of the tomo device.
Procedure:
• Position the foam plastic cube with the focus test on the tabletop so it is centered in the
beam path.
• Position the focus test parallel to the table longitudinal axis, see Fig.4.
TV Monitor
Anode side
Fig. 4
NOTICE If there are larger differences, check the tomo device for mechan-
ical errors (jamming movement, play, etc.).
If the blur patterns of the holes are uneven (jagged, wavy lines), there are unsynchronous
vibrations in the tomo system (jamming locations, ...).
This aslo results in reduced resolution.
The blur pattern of the holes should have an even density along the entire length. If there
is a dotted blur pattern, the tomo movement is uneven. If there is a continuous change in
density of the blur pattern, the speed of movement is uneven or there is a discontinuous
dose rate.
4.2.2 SIMOMED H
• Set the ambient light sensor S476.4 to ON (down).
TD AX 5 / Rödder
TD AX 1 / Biedermann