Shenzhen Drawray Biotech Co., Ltd.
No: Drawray/CS-iF05-22009
Document No: Drawray/CS-iF05-22009 Document Version A2
Applicable Products iStar 500
File Name
iStar 500 LIS Configuration Guide
References
Document Code Document Name Version Information
Adress:Room 301, Kingsignal Building 1, No.50, Baolong 2nd Road, Baolong Subdistrict, Longgang District, Shenzhen, China
Tel: +86 755 26601910 Email: [email protected]
Shenzhen Drawray Biotech Co., Ltd.
No: Drawray/CS-iF05-22009
Revision Record
Entry-into-force
Version Change Content Modification Person
Time
A0 Newly increased Hu Feng 2021.05.25
A1 Update relevant instructions Huang Rui 2022.07
Update instrument IP address
A2 Huang Rui 2022.12
modification method
Adress:Room 301, Kingsignal Building 1, No.50, Baolong 2nd Road, Baolong Subdistrict, Longgang District, Shenzhen, China
Tel: +86 755 26601910 Email: [email protected]
Shenzhen Drawray Biotech Co., Ltd.
No: Drawray/CS-iF05-22009
iStar 500 LIS Configuration Guide
A2
Adress:Room 301, Kingsignal Building 1, No.50, Baolong 2nd Road, Baolong Subdistrict, Longgang District, Shenzhen, China
Tel: +86 755 26601910 Email: [email protected]
Shenzhen Drawray Biotech Co., Ltd.
No: Drawray/CS-iF05-22009
Catalogue
Chapter 1 Overview........................................................................................................5
1. Background......................................................................................................... 5
2. Purpose................................................................................................................ 5
3. Document Scope..................................................................................................5
Chapter 2 Preparation.....................................................................................................6
Chapter 3 Connection Configuration of LIS.................................................................. 7
1. Connect with network cable, and the transmission mode is [TCP/IP]................7
2. Connect with serial port line, and the transmission mode is [Serial Port].......... 9
Chapter 4 LIS Settings..................................................................................................11
1. LIS setting interface description........................................................................11
2. Setting method...................................................................................................12
3. Set channel number........................................................................................... 13
4. LIS connection.................................................................................................. 14
5. Matters needing attention.................................................................................. 14
Adress:Room 301, Kingsignal Building 1, No.50, Baolong 2nd Road, Baolong Subdistrict, Longgang District, Shenzhen, China
Tel: +86 755 26601910 Email: [email protected]
Shenzhen Drawray Biotech Co., Ltd.
No: Drawray/CS-iF05-22009
Chapter 1 Overview
1. Background
Through the configuration of LIS, the instrument can automatically download the sample
application information from the LIS server, and automatically send samples and quality control
results in real time; You can also manually send sample results and quality control results to LIS
server individually or in batches.
2. Purpose
The LIS configuration described in this guide is used for data exchange between the iStar 500
automatic chemiluminescence analyzer and the laboratory information system (LIS).
3. Document Scope
This document is applicable to iStar 500 Automatic Chemiluminescence Analyzer.
Adress:Room 301, Kingsignal Building 1, No.50, Baolong 2nd Road, Baolong Subdistrict, Longgang District, Shenzhen, China
Tel: +86 755 26601910 Email: [email protected]
Shenzhen Drawray Biotech Co., Ltd.
No: Drawray/CS-iF05-22009
Chapter 2 Preparation
Order Name Description Number
1. A certain storage space needs to be reserved to
install LIS software and store result data.
2. Meet the installation conditions of LIS software.
3. According to LIS connection mode (network
cable/serial cable connection), at least one idle
1 LIS computer 1 pcs
network port/serial port should be reserved for each
iStar 500 connection, or use switches and other
devices to assist connection.
4. Meet the requirements of printer connection and
printing.
1、Network cable: at present, a 5m network cable
has been added to the installed accessory package,
but the instrument with earlier delivery does not
have a standard network cable, which needs to be
Network cable /
2 purchased separately. 1 pcs
Serial port cable
2、Serial port line: the instrument is not equipped
with standard serial port line. If necessary, RS232
cross serial port line (female to female) can be
prepared.
Adress:Room 301, Kingsignal Building 1, No.50, Baolong 2nd Road, Baolong Subdistrict, Longgang District, Shenzhen, China
Tel: +86 755 26601910 Email: [email protected]
Shenzhen Drawray Biotech Co., Ltd.
No: Drawray/CS-iF05-22009
Chapter 3 Connection Configuration of LIS
1. Connect with network cable, and the transmission mode is [TCP/IP]
(1) Power on the iStar 500 instrument and log in the instrument software. Connect the network
port of LIS computer and the network port on the left side of the iStar 500 instrument with the
network cable;
(2) On the LIS computer, modify the IP address of the network port used for LIS communication
as required. The following figure is an example, which can be set according to the actual situation;
Figure 1 LIS computer IP address setting
(3) If the IP address used for LIS communication is not in the 192.168.3.XX network segment,
the instrument IP address needs to be modified. The default IP address of the instrument is
192.168.3.31. The instrument IP address can only be modified after upgrading to V1.00.10
and above software (see iStar 500 Software Upgrade Guide for software upgrade method);
Figure 2 Software version required by instrument IP address modification function
(4) If you need to modify the IP address of the instrument, you need to enter the "Utility - System
Settings - LIS Settings" interface, click the "Local Setting" button to enter the setting interface,
Adress:Room 301, Kingsignal Building 1, No.50, Baolong 2nd Road, Baolong Subdistrict, Longgang District, Shenzhen, China
Tel: +86 755 26601910 Email: [email protected]
Shenzhen Drawray Biotech Co., Ltd.
No: Drawray/CS-iF05-22009
check [Use The Following IP], fill in the IP address, Subnet Mask and Default Gateway below,
and click the "Utility" button to complete the IP modification. After the IP is modified, click the
"Back" button, and then click the "Local Setting" button again to enter the setting interface.
Confirm that the IP address just modified is displayed in the [Obtain IP Address] column;
Figure 3 Modification of instrument IP address
Figure 4 After the instrument IP address is modified
(5) Confirm the network cable connection status. Click the "Start" menu of the computer on the
LIS computer, enter "cmd" in the search box, right click to select to run as administrator, enter
Adress:Room 301, Kingsignal Building 1, No.50, Baolong 2nd Road, Baolong Subdistrict, Longgang District, Shenzhen, China
Tel: +86 755 26601910 Email: [email protected]
Shenzhen Drawray Biotech Co., Ltd.
No: Drawray/CS-iF05-22009
ping [instrument IP] in the pop-up command prompt interface (for example, ping 192.168.3.31),
and click Enter to confirm whether you can ping;
Figure 5 Confirm the network cable connection status
(6) Set server IP, port number and other information on LIS software;
(7) On the iStar 500 instrument software, enter the "Utility - System Settings - LIS Settings"
interface, select [Transmission Mode] as [TCP/IP], enter the LIS server IP address set on the LIS
computer at [LIS Host IP], enter the LIS server listening port number set on the LIS computer at
[Port], click [Save], and then click [Connection] to confirm whether the connection is successful;
Figure 6 LIS settings of network cable connection
1. Connect with serial port line, and the transmission mode is [serial port]
(1) Power on the iStar 500 instrument and log in the instrument software. Connect the serial port
of the LIS computer and the serial port on the left side of the iStar 500 instrument with the
serial port cable;
(2) Set serial port, data bit, baud rate and other information on LIS software;
(3) On the iStar 500 instrument software, enter the "Settings - System Settings - LIS Settings"
interface, select [Transmission Mode] as [Serial Port], select the corresponding [Serial Port],
Adress:Room 301, Kingsignal Building 1, No.50, Baolong 2nd Road, Baolong Subdistrict, Longgang District, Shenzhen, China
Tel: +86 755 26601910 Email: [email protected]
Shenzhen Drawray Biotech Co., Ltd.
No: Drawray/CS-iF05-22009
[Data Bit], [Stop Bit], [Parity Check Bit], [Baud Rate] information according to the LIS
server interface, click the [Save] button, and then click the [Connect] button to confirm
whether the connection is successful; (You can use the serial port debugging tool to confirm
the connection status).
Figure 7 LIS settings for serial cable connection
Adress:Room 301, Kingsignal Building 1, No.50, Baolong 2nd Road, Baolong Subdistrict, Longgang District, Shenzhen, China
Tel: +86 755 26601910 Email: [email protected]
Shenzhen Drawray Biotech Co., Ltd.
No: Drawray/CS-iF05-22009
Chapter 4 LIS Settings
1. LIS Setting Interface Description
Figure 8 LIS setting interface
(1) [Code]: currently supports two coding types [ANSI] and [UTF-8]. The default is [ANSI].
The coding method can be consulted with LIS engineers;
(2) [Protocol Type]: select the corresponding protocol mode according to the LIS agreement of
the LIS engineer. At present, the software supports [HL7] and [ASTM], and the main
protocol mode for domestic customers is [HL7]; (See the LIS interface manual for details,
which is consistent with the LIS protocol of iFlash 3000 series instruments).
(3) [Communication Mode]: select [two-way] or [one-way] according to the actual type;
[Two-way]: LIS server can actively send sample application information to the instrument.
After the instrument test, it can automatically transmit sample results or quality control results to
LIS server; (Note: The iStar 500 instrument cannot automatically scan the sample barcode at
present, and the instrument still needs to manually complete the test application information after
downloading the application from the LIS terminal.)
[One-way]: The device cannot download the test from the LIS server. It supports manual
application for sample test or quality control results, which are automatically or manually sent to
the LIS server after completion;
(4) [Communicated Timeout]: set the timeout limit for the connection between the instrument
and LIS, and the range is [10~60s];
Adress:Room 301, Kingsignal Building 1, No.50, Baolong 2nd Road, Baolong Subdistrict, Longgang District, Shenzhen, China
Tel: +86 755 26601910 Email: [email protected]
Shenzhen Drawray Biotech Co., Ltd.
No: Drawray/CS-iF05-22009
(5) [Interval]: enter the interval time for automatic reconnection of the system after
disconnecting the LIS link, the range is [30~60s];
(6) [Connect to LIS server automatically after startup]: When this option is selected, the
system will actively establish a connection with LIS server when starting;
(7) [Auto Reconnection After Disconnection]: When this option is selected, the system will
check the LIS connection status every set [Interval]. If the LIS is disconnected, the LIS will
be automatically reconnected. If the connection is normal, do not reconnect;
(8) [Send test results of completed samples in real time]: after selecting this item, when the
status of a sample changes from [Under test] to [Completed], the system will automatically
send the test results to LIS. This function is only applicable to the sample test results of the
current day. Historical sample results cannot be automatically sent to LIS;
(9) [Send test results of uncompleted samples in real time]: When this option is selected,
when the status of a sample changes from [Under test] to [Incomplete], the system will
automatically send the test results to LIS. This function is only applicable to the sample test
results of the current day. Historical sample results cannot be automatically sent to LIS;
(10) [Send test results of completed items in real time]: after selecting this item, when the test
item status changes from [Under test] to [After completion], the system will automatically
send the test results to LIS. This function is only applicable to the sample test results of the
current day. Historical sample results cannot be automatically sent to LIS;
1. Setting Method
(1) Transmit by test sample progress, and the setting method is as follows:
Figure 9 Transmission by test sample progress
(2) Transmit by test project progress, and the setting method is as follows:
Adress:Room 301, Kingsignal Building 1, No.50, Baolong 2nd Road, Baolong Subdistrict, Longgang District, Shenzhen, China
Tel: +86 755 26601910 Email: [email protected]
Shenzhen Drawray Biotech Co., Ltd.
No: Drawray/CS-iF05-22009
Figure 10 Transmission by test project schedule
1. Set Channel Number
Channel number: it is to ensure the correspondence between the instrument item and the LIS
server identification item. The channel number of the LIS server identification item ensures the
transmission of the corresponding item data.
The project channel number is the project ID number. The project channel number can be
displayed by adding an item.
Add method: click the [Add] button on the "Utility - Assay Settings - Assay" interface of the
software, and scan the corresponding project methodology QR code to complete the addition.
Note: channel number cannot be set for [Calc. Assay] at present.
Adress:Room 301, Kingsignal Building 1, No.50, Baolong 2nd Road, Baolong Subdistrict, Longgang District, Shenzhen, China
Tel: +86 755 26601910 Email: [email protected]
Shenzhen Drawray Biotech Co., Ltd.
No: Drawray/CS-iF05-22009
Figure 11 Add Item
1. LIS Connection
Click the [Save] button first, and then click the [Connect] button.
Figure 12 Saving LIS settings
2. Matters Needing Attention
(1) Before setting LIS, first click the [Disconnect] button to disconnect the LIS connection, and
then click the [Save] button after setting;
(2) After setting LIS, click the [Save] button before clicking the [Connect] button.
Adress:Room 301, Kingsignal Building 1, No.50, Baolong 2nd Road, Baolong Subdistrict, Longgang District, Shenzhen, China
Tel: +86 755 26601910 Email: [email protected]