Sensors
Sensors
Communication
Detection of Triacetone Triperoxide (TATP) and Hexamethylene
Triperoxide Diamine (HMTD) from the Gas Phase with
Differential Ion Mobility Spectrometry (DMS)
Mirosław Maziejuk 1 , Monika Szyposzyńska 1, * , Aleksandra Spławska 1 , Monika Wiśnik-Sawka 1 and
Michał Ceremuga 2
1 Military Institute of Chemistry and Radiometry, al. gen. A. Chruściela “Montera”, 105,
00-910 Warsaw, Poland; [email protected] (M.M.); [email protected] (A.S.);
[email protected] (M.W.-S.)
2 Military Institute of Armament Technology, ul. Prym. St. Wyszyńskiego, 7, 05-220 Zielonka, Poland;
[email protected]
* Correspondence: [email protected]
Abstract: One of the significant problems in the modern world is the detection of improvised
explosives made of materials synthesized at home. Such compounds include triacetone triperoxide
(TATP) and hexamethylene triperoxide diamine (HMTD). An attempt was made to construct an
instrument allowing for the simultaneous detection of both compounds despite the large difference of
vapor pressure: very high for TATP and very low for HMTD. The developed system uses differential
ion mobility spectrometry (DMS) in combination with a specially designed gas sample injection
Citation: Maziejuk, M.; system. The created system of detectors allowed for the detection of a high concentration of TATP and
Szyposzyńska, M.; Spławska, A.; a very low concentration of HMTD. TATP detection was possible despite the presence of impurities—
Wiśnik-Sawka, M.; Ceremuga, M. acetone remaining from the technological process and formed as a coproduct of diacetone diperoxide
Detection of Triacetone Triperoxide (DADP) synthesis. Ammonia added to the carrier gas improved the possibility of detecting the
(TATP) and Hexamethylene
abovementioned explosives, reducing the intensity of the acetone signal. The obtained results were
Triperoxide Diamine (HMTD) from
then compared with the detection capabilities of drift tube ion mobility spectrometer (DT-IMS), which
the Gas Phase with Differential Ion
has not made possible such detection as DMS.
Mobility Spectrometry (DMS). Sensors
2021, 21, 4545. https://doi.org/
Keywords: differential ion mobility spectrometry (DMS); ion mobility spectrometry (IMS); explosives;
10.3390/s21134545
triacetone triperoxide (TATP); hexamethylene triperoxide diamine (HMTD)
Academic Editor: Santiago Marco
Figure
Figure 1. Vapor
1. Vapor pressure
pressure values
values forfor various
various groups
groups of of explosive
explosive compounds
compounds (Reprinted
(Reprinted from
from [12]).
[12]).
DADP
Since TATPis aand
peroxide
HMTD with
do the
nothighest
containVP, which
nitro is 24.7 Pa;
or aromatic for TATP,groups,
functional this value is 6.39 Pa.
analytical
The vapor pressure of these substances is compared to that of liquid
devices for conventional explosives are unsuitable for detecting them. In recent years,nitro derivatives such
as nitroglycerin (NG). HMTD has a vapor pressure 150 times lower than that of TATP,
many techniques of−2TATP and HMTD analysis have been developed, including infrared
which is 3.9 × 10 Pa. These values are significantly higher than those of commonly used
(IR) [13,14], liquid chromatography (LC) [15–17] and IMS detectors, which play a signifi-
strong grinding materials such as octogen (HMX) or hexogen (RDX). The vapor pressures
cant role in detecting these materials [18–21].
of the described acetone peroxides are higher than that of TNT, 9.27 × 10−4 Pa.
Since TATP and HMTD do not contain nitro or aromatic functional groups, analytical
1.1. IMS Technique
devices for conventional explosives are unsuitable for detecting them. In recent years,
Ion techniques
many mobility spectrometry
of TATP and (IMS)
HMTD usesanalysis
differences
haveinbeen
ion mobility
developed, in carrier gas infrared
including under
the(IR)
influence of an electric field [22–26] (Figure 2). A classical IMS (drift tube
[13,14], liquid chromatography (LC) [15–17] and IMS detectors, which play a significant ion mobility
spectrometer, DT-IMS)
role in detecting thesespectrometer has a weak linear electric field with values from 150
materials [18–21].
to 400 V/cm. Within these fields, ion mobility is constant.
1.1.Ion
IMSseparation
Techniquein a homogeneous electric field takes place based on differences in
their mobility. The results
Ion mobility obtained
spectrometry from
(IMS) usesthedifferences
detector are presented
in ion mobility ininthe form gas
carrier of drift
under
time spectra. An IMS detector with a membrane was used in the laboratory research.
the influence of an electric field [22–26] (Figure 2). A classical IMS (drift tube ion mobility
spectrometer, DT-IMS) spectrometer has a weak linear electric field with values from 150 to
400 V/cm. Within these fields, ion mobility is constant.
Ion separation in a homogeneous electric field takes place based on differences in their
mobility. The results obtained from the detector are presented in the form of drift time
spectra. An IMS detector with a membrane was used in the laboratory research.
Sensors 2021, 21, x FOR PEER REVIEW 3 of 12
Sensors 2021, 21, 4545 3 of 12
Figure 3. Principle of differential mobility spectrometry. (a) Schematic diagram of DMS spectrometers; (b) CV waveform;
Figure 3. Principle of differential mobility spectrometry.
(c) SV waveform.
The analyte
The analyteintroduced
introducedinto
intothe
theIMS/DMS
IMS/DMS detector is ionized
detector by charge
is ionized transfer
by charge (pro-
transfer
(proton) between the reaction ions and the neutral molecule of the analyte. The changethe
ton) between the reaction ions and the neutral molecule of the analyte. The change in in
enthalpy
the of this
enthalpy reaction
of this taking
reaction takingplace
placein in
the gas
the gasphase
phaseisiscalled
calledproton
protonaffinity
affinity (PA). For
(PA). For
the ionization
the ionization of
of each
each substance,
substance, the
the values
values are
are different for TATP
different for TATP and
and HMTD.
HMTD. They They are
are
860 kJ/mol [37] and 940 kJ/mol [11], respectively.
860 kJ/mol [37] and 940 kJ/mol [11], respectively.
In the membrane mode, the Z1 valve is switched to gas flow through the membrane,
which is forced by the P1 pump at this time. The Z2 valve at the P2 pump outlet allows
gas to flow through the carbon filter, and then the sample is taken from the membrane
exchanger (sample introduction system), and goes to the DMS.
In the membrane-free mode, the gas to be analyzed is charged directly to the DMS.
Setting the valve Z1 to take air allows direct sampling from the test area.
The P2 pump allows you to regulate the flow with the required amount of gas for
analysis (external circuit). In this mode, the valve Z2 at this time is switched over to allow
gas to be purged after the analysis.
The P3 pump is used to regulate the carrier gas flow in the DMS chamber at 3.2 L/min
(internal circuit). The applied molecular sieves clean the gas in the internal circuit of the
DMS chamber.
Due to the high vapor pressure of TATP, most measurements for this substance are
made in the membrane mode, while for HMTD all measurements are made in the mem-
brane-free mode. In the membrane mode, the appearance of the HMTD peaks was not rec-
orded as the intensity of peaks was below the limit of detection (LOD). TATP, HMTD, and
DADP were obtained synthesized at the Military Institute of Armament Technology
Figure 4. Diagram of the measurement system.
(Zielonka).
Figure 4. Diagram of the measurement system.
In the membrane mode, the Z1 valve is switched to gas flow through the membrane,
which is forced by the P1 pump at this time. The Z2 valve at the P2 pump outlet allows
gas to flow through the carbon filter, and then the sample is taken from the membrane
exchanger (sample introduction system), and goes to the DMS.
In the membrane-free mode, the gas to be analyzed is charged directly to the DMS.
Setting the valve Z1 to take air allows direct sampling from the test area.
The P2 pump allows you to regulate the flow with the required amount of gas for
analysis (external circuit). In this mode, the valve Z2 at this time is switched over to allow
gas to be purged after the analysis.
The P3 pump is used to regulate the carrier gas flow in the DMS chamber at 3.2 L/min
(internal circuit). The applied molecular sieves clean the gas in the internal circuit of the
DMS chamber.
Due
Figure5.5.The
to the high vapor
Themeasurement
measurementsystem.
pressure of TATP, most measurements for this substance are
system.
Figure
made in the membrane mode, while for HMTD all measurements are made in the mem-
brane-free mode. Inmode,
In the membrane the membrane mode,
the Z1 valve the appearance
is switched of thethrough
to gas flow HMTD thepeaks was not rec-
membrane,
which is forced by the P1 pump at this time. The Z2 valve at the P2 pump outlet allowsand
orded as the intensity of peaks was below the limit of detection (LOD). TATP, HMTD,
gas DADP
to flowwere
throughobtained synthesized
the carbon at then
filter, and the Military
the sampleInstitute
is takenof from
Armament Technology
the membrane
(Zielonka).
exchanger (sample introduction system), and goes to the DMS.
In the membrane-free mode, the gas to be analyzed is charged directly to the DMS.
Setting the valve Z1 to take air allows direct sampling from the test area.
The P2 pump allows you to regulate the flow with the required amount of gas for
analysis (external circuit). In this mode, the valve Z2 at this time is switched over to allow
gas to be purged after the analysis.
The P3 pump is used to regulate the carrier gas flow in the DMS chamber at 3.2 L/min
(internal circuit). The applied molecular sieves clean the gas in the internal circuit of the
DMS chamber.
Due to the high vapor pressure of TATP, most measurements for this substance
are made in the membrane mode, while for HMTD all measurements are made in the
membrane-free mode. In the membrane mode, the appearance of the HMTD peaks was
not recorded as the intensity of peaks was below the limit of detection (LOD). TATP,
HMTD, and DADP were obtained synthesized at the Military Institute of Armament
Technology (Zielonka).
Figure 5. The measurement system.
Sensors2021,
Sensors 2021,21,
21,4545
x FOR PEER REVIEW 66ofof12
12
3.3.Results
Results
The
Thegas
gassystem
systempresented
presentedin inFigure
Figure44 was
was used
used to analyze pure TATP,
TATP,DADP,
DADP,HMTD
HMTD
and
andaamixture
mixtureofofTATP
TATP and
andDADP.
DADP.Scattering spectra
Scattering spectrawere recorded
were for for
recorded a separation voltage
a separation volt-
of
age118
ofTd.
118 Td.
For
ForTATP
TATP(99%(99%purity),
purity),monomer
monomerand anddimer
dimerionsionswere
wereobtained
obtainedininpositive
positivemode
mode
(Figure 6). The monomer ion peak was issued at − 0.6 Td and the dimer
(Figure 6). The monomer ion peak was issued at −0.6 Td and the dimer ion peakion peak at +0.1 Td.
at +0.1
No peaks for negative ions were observed.
Td. No peaks for negative ions were observed.
3 and drift
Figure6.6.Dispersion
Figure Dispersionplots
plots for
for positive
positive ions
ions of
of triacetone
triacetone triperoxide (TATP) (500
triperoxide (TATP) µg/m3))and
(500 µg/m drift
spectra
spectrafor
for118
118Td.
Td.
Sensors 2021, 21, x FOR PEER REVIEW 7 of 12
For
ForTATP
TATP(99%
(99%purity),
purity),monomer
monomerand
anddimer
dimerions
ionswere
wereobtained
obtainedininpositive
positivemode
mode
(Figure
(Figure7).
7).
Traces of dimer ions were observed for DADP (low intensity compared to monomer
ions). When increasing the DADP concentration, the RIP and NH3 peaks disappeared.
This means that the peak intensity of the DADP monomer increased but the peak of the
DADP dimer did not increase.
In further studies, a mixture of TATP and DADP (approximately 78% w/w TATP)
was analyzed. This research doped NH3 at two concentrations (50 ppb and 200 ppb) in the
internal circuit. The results for 118 Td are shown in Figures 8 and 9.
In the case of a TATP + DADP mixture for positive ions, there were generally four
peaks: one from the TATP + DADP monomer, the other from the acetone monomer, and
two peaks from the dimer ion, from acetone and TATP. Due to the rather low resolution
of DMS, the peaks derived from the TATP and DADP monomers were not separated and
appeared as one peak. In the presence of a low concentration of ammonium, the dominant
peaks were those of acetone (Figure 9).
For a system with a 200 ppb ammonium concentration, two reaction peaks were visi-
ble: one for ammonium and the other for H+(H2O)n, both of which showed small ampli-
Figure 7. Dispersion plots for positive ions of diacetone
diacetone diperoxide
diperoxide (DADP)
(DADP) (600 µg/m3)3 and
(600 µg/m ) anddrift
drift
tudes. The ammonium
spectra for 118 Td. peak was only twice as high as the noise.
From the spectra presented in Figure 9, we can see that acetone significantly disturbed
the accuracy
Traces ofofdimer
the measurement. In particular,
ions were observed for DADPthe acetone dimer peak
(low intensity interfered
compared with the
to monomer
analysis.
ions). Theincreasing
When acetone peaks were higher
the DADP than thethe
concentration, TATP
RIP peaks.
and NH Doping
3 peaks the carrier
disappeared.gas with
This
means that the peak intensity of the DADP monomer increased but the peak of
200 ppb NH3 caused a decrease in the intensity of the acetone peak (Figure 10). Therefore, the DADP
dimer didconcluded
it can be not increase.
that the minimum value of ammonia concentration in the carrier gas,
whichInisfurther studies,
necessary a mixture
for the of TATP
separation and
of the DADP
peaks from (approximately 78% w/w TATP)
TATP, was approximately 200 was
ppb.
analyzed. This research doped NH3 at two concentrations (50 ppb and 200 ppb) in the
internal circuit. The results for 118 Td are shown in Figures 8 and 9.
Sensors 2021, 21, 4545 7 of 12
Figure 7. Dispersion plots for positive ions of diacetone diperoxide (DADP) (600 µg/m3) and drift
Figure 7. Dispersion plots for positive ions of diacetone diperoxide (DADP) (600 µg/m3) and drift
spectra for 118 Td.
spectra for 118 Td.
Figure8.8. Differential
Figure Differentialion
ion mobility
mobility spectrometry
spectrometry (DMS)
(DMS) spectrum
spectrum for
for 118
118 Td
Td for
for TATP
TATPpositive
positiveions,
ions,
Figure 8. Differential ion mobility spectrometry (DMS) spectrum for 118 Td for TATP positive ions,
withNH
with NH33 atat aa concentration
concentrationof
of50
50ppb
ppbin
inthe
theinternal
internalcircuit.
circuit.
with NH3 at a concentration of 50 ppb in the internal circuit.
Figure 9. DMS spectrum for TATP positive ions, with NH3 at a concentration of 200 ppb in the
internal circuit.
In the case of a TATP + DADP mixture for positive ions, there were generally four
peaks: one from the TATP + DADP monomer, the other from the acetone monomer, and
two peaks from the dimer ion, from acetone and TATP. Due to the rather low resolution
of DMS, the peaks derived from the TATP and DADP monomers were not separated and
appeared as one peak. In the presence of a low concentration of ammonium, the dominant
peaks were those of acetone (Figure 9).
For a system with a 200 ppb ammonium concentration, two reaction peaks were
visible: one for ammonium and the other for H+ (H2 O)n , both of which showed small
amplitudes. The ammonium peak was only twice as high as the noise.
From the spectra presented in Figure 9, we can see that acetone significantly disturbed
the accuracy of the measurement. In particular, the acetone dimer peak interfered with the
analysis. The acetone peaks were higher than the TATP peaks. Doping the carrier gas with
200 ppb NH3 caused a decrease in the intensity of the acetone peak (Figure 10). Therefore,
Sensors2021,
Sensors 2021,21,
21,4545
x FOR PEER REVIEW 88of
of12
12
itFigure
can be9. concluded that for
DMS spectrum theTATP
minimum value
positive ions,ofwith
ammonia
NH3 atconcentration
a concentrationinofthe
200carrier
ppb ingas,
the
which is necessary
internal circuit. for the separation of the peaks from TATP, was approximately 200 ppb.
Figure 10.
Figure 10. DMS
DMS spectra
spectra for
for HMTD
HMTD positive
positive ions.
ions.
Figure 10 shows
Figure 10 showsthethespectrum
spectrumfor
for HMTD
HMTD positive
positive ions.
ions. For For this substance,
this substance, both
both mon-
monomer and dimer ions were formed. Table 1 presents the compensating voltage
omer and dimer ions were formed. Table 1 presents the compensating voltage for all pos- for all
positive ions.
itive ions.
Table1.1.Compensating
Table Compensatingvoltage
voltagevalues
valuesfor
forpositive
positiveions.
ions.
Substance
Substance Positive
Positive IonIon CV(Td)
CV (Td)
RIPRIP −1.3
−1.3
TATP
TATP99% 99%purity
purity monomer
monomer −−0.6
0.6
Dimer
Dimer 0.1
0.1
RIPRIP − 1.3
−1.3
DADP 99% purity monomer −0.6
DADP 99% purity monomer −0.6
dimer 0.1
dimer 0.1
RIP −1.3
RIP −1.3
acetone monomer −0.9
TATP + DADP acetone monomer −0.9
TATP + DADP acetone dimer −0.2
(50 ppb NH3 )
acetone
TATP dimer
+ DADP −−0.2
0.7
(50 ppb NH3) TATP
TATPdimer
+ DADP 0.2
−0.7
TATP
RIP dimer 0.2
−1.3
acetone monomer
RIP − 1.0
−1.3
TATP + DADP
acetone dimer −0.2
(200 ppb NH
TATP + DADP 3 ) acetone monomer −1.0
TATP + DADP monomer −0.6
(200 ppb NH3) acetone
TATP dimer
dimer −0.2
0.1
TATP + DADP monomer −0.6
RIP −1.3
HMTD TATP
HMTD monomerdimer 0.1
−1.0
HMTD RIPdimer −−1.3
0.5
HMTD HMTD monomer −1.0
To determine the possibility of detecting HMTD dimer
the mixture of TATP and HMTD,−0.5
tests were
carried out with the DT-IMS detector. The tests were carried out under the same conditions
as with a DMS detector, with concentrations of TATP totaling 250 µg/m3 and 5 ppm NH3
in the internal cycle.
Sensors 2021, 21, x FOR PEER REVIEW 9 of 12
Figure 11.
Figure 11. Drift time
time spectrum
spectrum of
of TATP.
TATP.
Table2.2.Drift
Table Drifttime
timefor
forTATP
TATPand
andDADP
DADPions.
ions.
PositivePositive
Ion Ion Time (ms)
Time (ms)
RIP (NH3) 9.3
RIP (NH3 ) 9.3
DADP
DADP monomer monomer
ion ion 9.7 9.7
TATP monomer
TATP monomer ion ion 11.2 11.2
TATP +TATP
DADP+dimer
DADP ion
dimer ion 13.2 13.2
In the
In the measurement
measurement system
system shown
shown in
in Figure
Figure 2,
2, no
no recorded
recorded peaks
peaks from
from HMTD
HMTD were
were
observed. After removing the inner membrane, the spectrum showed a peak with
observed. After removing the inner membrane, the spectrum showed a peak with reduced reduced
mobility of
mobility of approximately
approximately1.51.5cm
cm22/Vs.
/Vs. HMTD
HMTD detection
detection using
using DT-IMS was possible
DT-IMS was possible only
only
under laboratory conditions (for dried air).
under laboratory conditions (for dried air).
4.
4. Discussion
The
The acetone residues
residues reduced
reducedthe thedetection
detectioncapability
capabilityofofthethe DMS
DMS detector.
detector. TheThePA
PA for acetone is 812 kJ/mol [38], which is lower than the PA of the
for acetone is 812 kJ/mol [38], which is lower than the PA of the added ammonia (853.6 added ammonia
(853.6
kJ/molkJ/mol [39]); therefore,
[39]); therefore, the use the use of ammonium
of ammonium as an admixture
as an admixture in thegas
in the carrier carrier gas
will help
will help to
to reduce reduce
the the appearance
appearance of acetoneofpeaks.
acetone
Whenpeaks. When improvised
detecting detecting improvised
explosivesexplo-
using
sives using differential
differential ion mobilityion mobility spectrometry,
spectrometry, the carrier the
gas carrier
shouldgas be should be doped
doped with with a
a minimum
minimum of 200 ppb NH to reduce the effect of acetone (TATP contamination).
of 200 ppb NH3 to reduce the effect of acetone (TATP contamination). During the analysis
3 During
the analysis
of TATP of TATP DADP
containing containing
as anDADP as anan
impurity, impurity,
increasean inincrease in the amplitude
the amplitude of the
of the monomer
monomer was observed, the peak of which was at the same location
was observed, the peak of which was at the same location as that of the TATP monomer, as that of the TATP
monomer, without a significant increase in the dimer amplitude. This may indicate a lack
of separation of the dimer for DADP.
In the case of TATP measurements with the DT-IMS detector, the measurement system
used a carrier gas with NH3 content at the level of several ppm. The appearance of a peak
close to the reaction ions (NH3 ) was observed in both the TATP and DADP studies.
Sensors 2021, 21, 4545 10 of 12
Differential ion mobility spectrometry was more effective than DT-IMS. For DT-IMS,
the peaks derived from TATP and DADP were located close to the peaks of the reaction
ions; in the case of DMS, the peaks were spaced apart and it was possible to separate them
more efficiently. Moreover, for the gas system with the membrane, no HMTD peaks were
observed in the case of IMS. Furthermore, it is impossible to detect HMTD in generally
used DT-IMS systems with membranes. In this case, there was no peak observed from
the analyte.
In the DMS studies, it was observed that the peaks moved towards higher compen-
satory voltages in membrane-free mode, due to a humidity effect. When adjusting the
effect of humidity on the location of peaks, the identification of HMTD and TATP was
maintained at the same level—no shift of adjusted peaks was observed.
The response time for instruments with membranes for TATP is a few seconds, while
that for HMTD is from several dozen seconds to two minutes. The response time for HMTD
results is the time required for membrane permeation, whereas for a membrane-free system,
TATP and HMTD are detected in seconds.
5. Conclusions
Detection of triacetone trioxide and hexamethylene triperoxide diamine can provide
early warning of a terrorist threat. DMS is currently one of the leading new technologies
for the separation and detection of chemicals in the gas phase.
When we used the membrane-free mode, the differential ion mobility spectrometry
detected TATP and HMTD correctly.
The use of ammonia as the dopant carrier gas in system with and without membrane
allows for the detection of inaccurately purified TATP containing the by-product DADP
and residual acetone. The use of ammonia in the carrier gas as a dopant helps in TATP
analysis for a membrane or membrane-free system.
Due to the differences resulting from the vapor pressure, it is necessary to use both
a membrane and membrane-free system at the same time. In the case of the membrane-
free mode, it is necessary to apply a peak position correction due to the variable value
of wetness.
Regarding laboratory analysis, the detection times of TATP and HMTD are short and
allow for quick scanning tests. Such a system allows the detection of HMTD at a level
lower than that of the gas chromatography used in WIChiR.
This study is the first to report detection of TATP and HMTD by differential ion
mobility spectrometry.
Author Contributions: Conceptualization, M.M.; data curation, M.S. and A.S.; formal analysis, M.M.,
M.S. and A.S.; funding acquisition, M.C.; investigation, M.M.; methodology, M.M. and M.C.; project
administration, M.M. and M.C.; validation, M.M. and M.W.-S.; visualization, M.S. and M.W.-S.;
writing—original draft, M.M. and M.S.; writing—review and editing, M.M. and M.W.-S. All authors
have read and agreed to the published version of the manuscript.
Funding: The work was carried out as part of a project financed by the National Centre for Research
and Development as part of the program for Programmers and Projects Defence, Security No. 9/2018,
project No. DOB-BIO9/28/01/2018.
Institutional Review Board Statement: Not applicable.
Informed Consent Statement: Not applicable.
Data Availability Statement: Data sharing not applicable.
Conflicts of Interest: The authors declare no conflict of interest.
Sensors 2021, 21, 4545 11 of 12
References
1. Oxley, J.C.; Smith, J.L.; Brady, J.; Naik, S. Determination of urea nitrate and guanidine nitrate vapor pressures by isothermal
thermogravimetry. Propellants Explos. Pyrotech. 2010, 35, 278–283. [CrossRef]
2. Damour, P.; Freedman, A.; Wormhoudt, J. Knudsen effusion measurement of organic peroxide vapor pressures. Propellants Explos.
Pyrotech. 2010, 35, 514–520. [CrossRef]
3. Menzies, A.W.C. The explanation of an apparent anomaly outstanding in the results of measurement of dissociation pressures. J.
Am. Chem. Soc. 1920, 42, 1951–1956. [CrossRef]
4. Oxley, J.C.; Smith, J.L.; Luo, W.; Brady, J. Determining the vapor pressures of diacetone diperoxide (DADP) and hexamethylene
triperoxide diamine (HMTD). Propellants Explos. Pyrotech. 2009, 34, 539–543. [CrossRef]
5. Oxley, J.C.; Smith, J.L.; Shinde, K.; Moran, J. Determination of the vapor density of triacetone triperoxide (TATP) using a gas
chromatography headspace technique. Propellants Explos. Pyrotech. 2005, 30, 127–130. [CrossRef]
6. Takada, Y.; Nagano, H.; Suzuki, Y.; Sugiyama, M.; Nakajima, E.; Hashimoto, Y.; Sakairi, M. High-throughput walk through
detection portal for counter terrorism: Detection of triacetone triperoxide (TATP) vapor by atmospheric-pressure chemical
ionization ion trap mass spectrometry. Rapid Commun. Mass Spectrom. 2011, 25, 2448–2452. [CrossRef] [PubMed]
7. Giordano, C.B.; Lubrano, A.L.; Field, C.R.; Collins, G.E. Dynamic headspace generation and quantitation of triacetone triperoxide
vapor. J. Chromatogr. A 2014, 1331, 38–43. [CrossRef] [PubMed]
8. Mbah, J.; Knott, D.; Steward, S. Thermogravimetric study of vapor pressure of TATP synthesized without recrystallization. Talanta
2014, 129, 586–593. [CrossRef]
9. Egorshev, V.Y.; Sinditskii, V.P.; Smirnov, S.P. A comparative study on two explosive acetone peroxides. Thermochim. Acta 2013,
574, 154–161. [CrossRef]
10. DeGreeff, L.E.; Cerreta, M.M.; Katilie, C.J. Variation in the headspace of bulk hexamethylene triperoxide diamine (HMTD) with
time; environment; and formulation. Forensic Chem. 2017, 4, 41–50. [CrossRef]
11. Aernecke, M.J.; Mendum, T.; Geurtsen, G.; Ostrinskaya, A.; Kunz, R.R. Vapor pressure of hexamethylene triperoxide diamine
(HMTD) estimated using secondary electrospray ionization mass spectrometry. J. Phys. Chem. A 2015, 119, 11514–11522.
[CrossRef]
12. Ewing, R.G.; Waltman, M.J.; Atkinson, D.A.; Grate, J.W.; Hotchkiss, P.J. The vapor pressure of explosives. Trends Anal. Chem. 2013,
42, 35–48. [CrossRef]
13. Schulte-Ladbeck, R.; Vogel, M.; Karst, U. Recent methods for the determination of peroxide-based explosives. Anal. Bioanal. Chem.
2006, 386, 559–565. [CrossRef]
14. Vodochodský, O.; Jalový, Z.; Matyáš, R.; Novotná, M. Determination of triacetone triperoxide and hexamethylene triperoxide
diamine in various matrices using infrared spectroscopy. Appl. Spectrosc. 2019, 73, 195–202. [CrossRef] [PubMed]
15. Xu, X.; Craats, A.M.; Kok, E.M.; Bruyn, P.C. Trace analysis of peroxide explosives by high performance liquid chromatography-
atmospheric pressure chemical ionization-tandem mass spectrometry (HPLC-APCI-MS/MS) for forensic applications. J. Forensic
Sci. 2004, 49, 1230–1236. [CrossRef] [PubMed]
16. Schulte-Ladbeck, R.; Kolla, P.; Karst, U. Trace analysis of peroxide-based explosives. Anal. Chem. 2003, 75, 731–735. [CrossRef]
[PubMed]
17. Dunn, L.; Obaidly, H.S.A.A.; Khalil, S.E. Development and validation of fast liquid chromatography high-resolution mass
spectrometric (LC-APCI-Q ToF-MS) methods for the analysis of hexamethylene triperoxide diamine (HMTD)and triacetone
triperoxide (TATP). Forensic Chem. 2018, 10, 5–14. [CrossRef]
18. Buttigieg, G.A.; Knight, A.K.; Denson, S.; Pommier, C.; Denton, M.B. Characterization of the explosive triacetone triperoxide and
detection by ion mobility spectrometry. Forensic Sci. Int. 2003, 135, 53–59. [CrossRef]
19. Rasanen, R.M.; Nousiainen, M.; Perakorpi, K.; Sillanpaa, M.; Polari, L.; Anttalainen, O.; Utriainen, M. Determination of gas phase
triacetone triperoxide with aspiration ion mobility spectrometry and gas chromatography–mass spectrometry. Anal. Chim. Acta
2008, 623, 59–65. [CrossRef] [PubMed]
20. Marr, A.J.; Groves, D.M. Ion mobility spectrometry of peroxide explosives TATP and HMTD. Int. J. Ion Mobil. Spectrom. 2003, 6,
59–61.
21. Oxley, J.C.; Smith, J.L.; Kirschenbaum, L.J.; Marimganti, S.; Vadlamannati, S. Detection of explosives in hair using ion mobility
spectrometry. J. Forensic Sci. 2008, 53, 690–693. [CrossRef] [PubMed]
22. Borsdorf, H.; Eiceman, G.A. Ion mobility spectrometry: Principles and applications. Appl. Spectrosc. Rev. 2006, 41, 323–375.
[CrossRef]
23. Stach, J.; Baumbach, J.I. Ion mobility spectrometry—basic elements and applications. Int. J. Ion Mobil. Spectrom. 2002, 5, 1–21.
24. Eiceman, G.A. Ion-mobility spectrometry as a fast monitor of chemical composition. Trends Anal. Chem. 2002, 21, 259–275.
[CrossRef]
25. Vautz, W.; Franzke, J.; Zampolli, S.; Elmi, I.; Liedtke, S. On the potential of ion mobility spectrometry coupled to GC pre-
separation–A tutorial. Anal. Chim. Acta 2018, 1024, 52–64. [CrossRef] [PubMed]
26. Sorribes-Soriano, A.; de la Guardia, M.; Esteve-Turrillas, F.A.; Armenta, S. Trace analysis by ion mobility spectrometry: From
conventional to smart sample preconcentration methods. A review. Anal. Chim. Acta 2018, 1026, 37–50. [CrossRef]
27. Kaur-Atwal, G.; O’Connor, G.; Aksenov, A.A.; Bocos-Bintintan, V.; Paul Thomas, C.L.; Creaser, C.S. Chemical standards for ion
mobility spectrometry: A review. Int. J. Mass Spectrom. 2009, 12, 1–14. [CrossRef]
Sensors 2021, 21, 4545 12 of 12
28. Kolakowski, B.M.; Mester, Z. Review of applications of high-field asymmetric waveform ion mobility spectrometry (FAIMS) and
differential mobility spectrometry (DMS). Analyst 2007, 132, 842–864. [CrossRef]
29. Jakubowska, M.; Maziejuk, M.; Ceremuga, M.; Siczek, J.; Gallewicz, W. Ceramic DMS—type detector. Int. J. Ion Mobil. Spectrom.
2012, 15, 99–108. [CrossRef]
30. Pathak, P.; Shvartsburg, A.A. Low-Field Differential Ion Mobility Spectrometry of Dipole-Aligned Macromolecules. Anal. Chem.
2020, 92, 13855–13863. [CrossRef]
31. Winter, D.L.; Wilkins, M.R.; Donald, W.A. Differential ion mobility–mass spectrometry for detailed analysis of the proteome.
Trends Biotechnol. 2019, 37, 198–213. [CrossRef] [PubMed]
32. Maziejuk, M.; Szczurek, A.; Maciejewska, M.; Pietrucha, T.; Szyposzyńska, M. Determination of benzene; toluene and xylene
concentration in humid air using differential ion mobility spectrometry and partial least squares regression. Talanta 2016, 152,
137–146. [CrossRef] [PubMed]
33. Szczurek, A.; Maziejuk, M.; Maciejewska, M.; Pietrucha, T.; Sikora, T. BTX compounds recognition in humid air using differential
ion mobility spectrometry combined with a classifier. Sens. Actuators B Chem. 2017, 240, 1237–1244. [CrossRef]
34. Krylov, E.V.; Coy, S.L.; Nazarov, E.G. Temperature effects in differential mobility spectrometry. Int. J. Mass Spectrom. 2009, 279,
119–125. [CrossRef]
35. Guevremont, R. High-field asymmetric waveform ion mobility spectrometry: A new tool for mass spectrometry. J. Chromatogr. A
2004, 1058, 3–19. [CrossRef]
36. Miller, R.A.; Nazarov, E.G.; Eiceman, G.A.; King, A.T. A MEMS radio-frequency ion mobility spectrometer for chemical vapour
detection. Sens. Actuator A Phys. 2001, 91, 307–318. [CrossRef]
37. Shen, C.; Li, J.-Q.; Xu, G.; Wang, H.-M.; Zheng, P.-C.; Li, H.; Wang, Y.-J.; Chu, Y.-N. Proton Transfer Reaction Mass Spectrometry
of Triacetone Triperoxide. Chem. J. Chin. Univ. 2009, 30, 274–278.
38. Wróblewski, T.; Ziemczonek, L.; Szerement, K.; Karwasz, G. Proton affinities of simple organic compounds. Czechoslov. J. Phys.
2006, 56, B1110–B1115. [CrossRef]
39. Canaval, E.; Hyttinen, N.; Schmidbauer, B.; Fischer, L.; Hansel, A. NH4+ Association and Proton Transfer Reactions with a Series
of Organic Molecules. Front. Chem. 2019, 7, 191. [CrossRef]