FCO70500004 - Possible Falling Down of The Tube Arm - PDF Nodeid 540326&vernum - 2

Philips Medical Systems issued a Field Change Order (FCO70500004) regarding the Mobil 300 CP product due to a potential issue with the tube arm falling down, caused by material fatigue. The remedy involves installing a retaining clamp to prevent this issue, requiring specific tools and a modification kit. The installation procedure is detailed, emphasizing safety precautions and the need for proper documentation of the action taken.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
8 views11 pages

FCO70500004 - Possible Falling Down of The Tube Arm - PDF Nodeid 540326&vernum - 2

Philips Medical Systems issued a Field Change Order (FCO70500004) regarding the Mobil 300 CP product due to a potential issue with the tube arm falling down, caused by material fatigue. The remedy involves installing a retaining clamp to prevent this issue, requiring specific tools and a modification kit. The installation procedure is detailed, emphasizing safety precautions and the need for proper documentation of the action taken.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 11

Philips Medical Systems

FIELD CHANGE ORDER Service


Issued by : PMS DMC Hamburg Reference No. : FCO70500004
Released by : H. Schmidt signed Date : November 2004
Publication No. : 4512 980 60291 Product Group : 705

tory
d a
man tion
ac
Mobil 300 CP

APPLIES TO:
Mobil 300 CP all serial numbers
9804 410 80004
9890 010 19021

TITLE:
Possible falling down of the tube arm

LIST OF PAGES & DRAWINGS:


1 ... 10 (04.1)
ANR (04.1)

INTRODUCTION:
Symptom : Tube arm can fall down
Cause : Fatigue of material
Remedy : Installation of a retaining clamp to prevent the tube arm from falling down

FCO70500004 (04.1) 1
Copyright © 2003 Philips Medical Systems
ALL RIGHTS RESERVED
MANPOWER / TIME TO COMPLETE:
1 service engineer, 4 hours

TOOLS & TEST EQUIPMENT:


Standard tool set
Drilling machine
4.6 mm and 8.6 mm metal drill bits
Center punch
Loctite 607 or similar

MODIFICATION KIT / PARTS REQUIRED:

4512 104 67241


Containing:
4512 104 67251 Retaining clamp Mobil 300CP
4512 104 67261 Retaining belt Mobil 300CP
2522 006 42076 Screw M8x40 (only used for premounting of the clamp)
2522 600 86041 Washer M8 3d (only used for premounting of the clamp)
2522 401 64015 Nut M8 2x
Countersunk screw M8 x 35 DIN 7991-8.8 ZN
Short open-end wrench 13 across flats
2512 200 17113 Screw M5 2x
2522 600 86032 Washer 2x
2622 335 01075 Drill bit 4.6 mm
2622 335 01121 Drill bit 8.6 mm
2522 187 44088 Countersunk screw 2x M4x8 (screws for covers)
4512 980 60291 This document

Ordering SSDs are notified of the quantity by Service Logistics International, Hamburg.
information: Kits are delivered free of charge until the end of March, 2004.

PROCEDURE:
Due to possible fatigue of material of the tube arm a retaining clamp must installed.

Installation of the retaining clamp

• Disconnect the system from mains.


• Unscrew four screws and remove the cover plate (a) for the
wall Bucky connectors.
• Unscrew four screws and remove the rear cover (b), make
sure not to damage the protective ground wire.

2 (04.1) FCO70500004
Copyright © 2003 Philips Medical Systems
ALL RIGHTS RESERVED
• Loosen two screws and slightly lift the top cover (c).
• Unscrew three screws and remove the left side cover (d),
make sure not to damage the protective ground wire.

WARNING

Dangerous capacitor voltage is still present after the system


has been disconnected from mains.
For safe operation on internal circuitry strictly follow the
specified procedure!
_____________________

• Set the discharge switch to “discharge”.


• Wait for 5 minutes.
• Make sure that the yellow indicator lights are off.
• Use a voltmeter to check that the “+” and “-“ voltages are
lower than 10 V DC with respect to ground.
• After installation work is finished set the discharge switch
back to “ normal”.
• Restore normal operating conditions.

• Remove the left and right tube arm cover.

FCO70500004 (04.1) 3
Copyright © 2003 Philips Medical Systems
ALL RIGHTS RESERVED
• Remove the cover and the parking interlock plate above
the swivel joint.

• To place the arm into a position that the retaining clamp


can be installed over the swivel joint, push, for instance, a
screwdriver grip under the bar as indicated in the image so
that the arm is fixed for further work.
The bar is a part of the system. It is located at the
right-hand inner side of the tower.

4 (04.1) FCO70500004
Copyright © 2003 Philips Medical Systems
ALL RIGHTS RESERVED
• Push the clamp over the joint.

• Determine the fixing point of the clamp and mark it with the
center punch.

• Predrill a hole in the metal tube with a diameter of 4.6 mm.


Take care to drill exactly vertically with respect to the tube
arm.

• Enlarge the hole to 8.6 mm and debur its edges.

FCO70500004 (04.1) 5
Copyright © 2003 Philips Medical Systems
ALL RIGHTS RESERVED
• Remove all metal swarf from the system. For that reason
move the arm into the lowest position and eject the swarf
with soft hammer hits and a vacuum cleaner.

• Take care that the countersink of the clamp is at the upper


side.
• Fix the clamp with the M8x40 screw, washer M8 3d and the
normal M8 nut.
(This procedure ensures that the clamp sits closely to the
tube).
• To hold the nut use the delivered short open-end wrench
13 across flats.

NOTE

To make installation easier first position the nut under the hole
on the lower side of the clamp and then insert the screw with
the washer and bolt them together.
_________________

• Tighten the screw so that the clamp fits closely to the metal
tube. This is very important for further installation.

• Remove the screw and the nut.


• Take care that the countersink of the clamp is at the upper
side.

6 (04.1) FCO70500004
Copyright © 2003 Philips Medical Systems
ALL RIGHTS RESERVED
• Install the countersunk screw M8x35 with Loctite.
• Tighten the screw.

• Check the up and down movement of the arm, it must not


be obstructed anywhere. Check that the arm can be moved
up and down as far as before.

• If the screw head or the clamp scrapes at the metal clip in


front of the metal tube, bend the clip a little bit so that the
screw is not restricted by the metal clip.
This can be performed with a ring spanner as indicated in
the images.

FCO70500004 (04.1) 7
Copyright © 2003 Philips Medical Systems
ALL RIGHTS RESERVED
• Install the cover over the swivel joint without installing the
parking interlock plate. Use only the small screws as shown
in the image. The screws from the parking interlock could
be to long and might scrape the metal clip.

NOTE

Reason why the parking interlock mechanism needs to


removed:
Due to possible improper of the parking interlock it may
happen that the parking interlock damages parts of the bearing
bushing of the arm of the unit. Therefore the parking interlock
must be removed and replaced by a clip band after the
retaining clamp is installed.
_________________

• Install the front side and the top cover


• Remember to set the discharge switch to “normal“
operation.

• Install the left-hand and right-hand arm cover. Use the


delivered screws M4x8 as shown in the image.

8 (04.1) FCO70500004
Copyright © 2003 Philips Medical Systems
ALL RIGHTS RESERVED
Installation of the new parking interlock

• Determine the holes for the new interlock at the system


base. The holes must be drilled 35 cm away from the base
plate.

• Drill the holes with the 4.6m drill bit.


• Fix the new interlock clip belt with the 5 mm screws with
the large washer at both sides. Fix the belt in such a way
that the lock is on the front side of the tube arm.

NOTE

Use the holes with the wider distance.


_________________

• Check all movements and the functions of the system.

FCO70500004 (04.1) 9
Copyright © 2003 Philips Medical Systems
ALL RIGHTS RESERVED
PARTS DISPOSAL:
Parts are to be disposed locally

DOCUMENTATION:
• Log this action in the section "History Record" of the System Reference Manual.
• File this FCO in the section "Service Information" of the System Reference Manual.
• Fill out the attached Action Notification Report and send it to your SSD Customer Support Manager.

10 (04.1) FCO70500004
Copyright © 2003 Philips Medical Systems
ALL RIGHTS RESERVED
FCO ACTION NOTIFICATION REPORT
For local SSD use only; do not return to PMG.
TITLE : Possible falling down of the tube arm

CLASSIFICATION : mandatory action FCO REF. NO.: FCO70500004


APPLIES TO : Mobil 300 CP - all serial numbers, 9804 410 80004, 9890 010 19021

HOSPITAL / ADDRESS :

LOCATION / FW SITE NO.: SALES ORDER NO. / OA NO.:

PRODUCT NUMBER :

UNIT SERIAL NUMBER :

ACTION ON THIS UNIT WAS: (select one) JOB NO. / SERVICE INCIDENT NO.:

Completed per instruction on


DATE

Completed by the factory prior to delivery.

Not completed as this unit is not affected per instruction because: (state reason)

Not completed because customer has unit in storage.


Required parts & instructions received by the customer.

CUSTOMER ACKNOWLEDGEMENT (Required for MANDATORY ACTIONS only).


The REASON and PURPOSE of this modification have been explained to me.

CUSTOMER NAME (PLEASE PRINT) TITLE

CUSTOMER SIGNATURE DATE

BRANCH
REGION / SERVICE UNIT /
DEALER : SERVICE AREA NO.:

SIGNATURE CUSTOMER SUPPORT ENGINEER DATE

SIGNATURE CUSTOMER SUPPORT MANAGER MAIL TO : SSD Customer Support Manager

FCO70500004 (04.1) ANR


Copyright © 2003 Philips Medical Systems
ALL RIGHTS RESERVED