ld
o
rn
A
SIREMOBIL
SP
Maintenance Instructions
SIREMOBIL
DICOM Bridge
The protocol SPR2-130.832.01.02.02 is required for
these instructions
04407651
© Siemens AG 2001
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
Print No.: SPR2-130.831.01.02.02 English
Replaces: SPR2-130.831.01.01.02 Doc. Gen. Date: 08.05
2 Revision / Disclaimer
1Revision / Disclaimer
Document revision level
The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.
SIREMOBIL SPR2-130.831.01.02.02 Page 2 of 18 Siemens AG
08.05 CS PS 24 Medical Solutions
Table of Contents 3
0 Table of Contents
1 _______ General information______________________________________________ 4
requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Required Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Required Tools, Test Equipment and Aids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Text designations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Safety information and protective measures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Maintenance interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2 _______ Inspection of exterior and surroundings____________________________ 11
Inspection of exterior . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Inspection of environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3 _______ Safety Inspection _______________________________________________ 13
Mechanical Safety Inspection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Electrical Safety Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4 _______ Final results/quality inspection/general maintenance _________________ 15
Final Work Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
5 _______ Changes since previous version __________________________________ 17
Siemens AG SPR2-130.831.01.02.02 Page 3 of 18 SIREMOBIL
Medical Solutions 08.05 CS PS 24
4 General information
1-
1 General information
1.1 requirements
• The prerequisites described in Chapter 1 of the Service Instructions apply to mainte-
nance as well.
SIREMOBIL SPR2-130.831.01.02.02 Page 4 of 18 Siemens AG
08.05 CS PS 24 Medical Solutions
General information 5
1.2 Required Documents
• The safety information in accordance with ARTD Part
2.
• Service protocol SPR2-130.105.02..
• DICOM Bridge Installation and Setting Instructions SPR2-130.033.04..
• Service instructions for the system
Siemens AG SPR2-130.831.01.02.02 Page 5 of 18 SIREMOBIL
Medical Solutions 08.05 CS PS 24
6 General information
1.3 Required Tools, Test Equipment and Aids
NOTE The indicated articles are listed in the STC (Service Tools Catalog)
unless otherwise stated (the STC is a component of the Spare
Parts Catalog), except for those items identified with "*".
Tool Material no.:
• Standard tool kit*
• Protective conductor meter 44 15 899 RV090
• 25 mm measurement setup AL type 26765 acc. to
DIN 6868 part 50 or
- 1.2 mm Cu from the X-ray filter set 97 98 596 G5321
- 17 µm strips 11 67 662 G5247
• Dynamic test kit • 37 90 156 X 1963
or 97 50 001 X 1963
- TV dynamic test - 37 90 164 X 1963
- Bracket - 87 13 901 X 1963
- Veiling glare test - 87 13 901 X 1963
- Capillary Test - 37 90 180 X 1963
- Heart contour diaphragm - 37 90 172 X 1963
• Service PC
• Service PC serial interface cable 99 00 440 RE999
or 96 60 978 RE999
• Set of resolution tests 28 71 820 RE999
• Precision X-ray filter 2.1 mm Cu 99 00 598 XE999
• X-ray filters 10 x 0.3 mm Cu, set 44 06 120 RV090
SIREMOBIL SPR2-130.831.01.02.02 Page 6 of 18 Siemens AG
08.05 CS PS 24 Medical Solutions
General information 7
1.4 Text designations
! DANGER DANGER indicates when there is an immediate danger that
l e a d s to death or serious physical injury.
! WARNING WARNING indicates a risk of danger that m a y l e a d to death
or serious physical injury.
! CAUTION CAUTION used with the safety alert symbol indicates a risk of
danger that leads to slight or moderate physical injury and/ or
damage to property.
NOTICE NOTICE used without the safety alert symbol indicates a risk of
danger that if disregarded leads or may lead to a potential
situation which may result in an undesirable result or state other
than death, physical injury or property damage.
Fig. 1: Safety Notes
Siemens AG SPR2-130.831.01.02.02 Page 7 of 18 SIREMOBIL
Medical Solutions 08.05 CS PS 24
8 General information
1.5 Safety information and protective measures
NOTE • When performing service work and tests, observe:
- the product-specific safety instructions in the documenta-
tion,
- the safety guidelines TD00-000.860.01... ,
- the general safety guidelines contained in the ARTD part 2.
• Always unplug the line power connection when working on the
system.
• Follow all general safety rules when working with the system
under power.
• Observe ESD regulations!
• Perform the protective conductor test according to
ARTD-002.731.17 upon completion of the work and after all cov-
ers have been reattached. The ground wire resistance must not
exceed 0.2 ohms.
SIREMOBIL SPR2-130.831.01.02.02 Page 8 of 18 Siemens AG
08.05 CS PS 24 Medical Solutions
General information 9
1.6 Abbreviations
Abbrev. Description
SI Safety Inspection
SIE Electrical Safety Inspection
SIM Mechanical Safety Inspection
PM Preventive Maintenance
PMP Periodic Preventive Maintenance
PMA Preventive Maintenance Adjustments
PMF Preventive Maintenance, Operating Value Check, Function Check
Q Quality Check
QIQ Image Quality Check
QSQ System Quality Check
SW Software Maintenance
• The steps identified by these abbreviations appear in the maintenance report as items
which should be checked off upon completion.
NOTE The sequence for complete maintenance and inspection is de-
scribed on the following pages.
Each work step must be performed on an annual basis, if not oth-
erwise specified.
Siemens AG SPR2-130.831.01.02.02 Page 9 of 18 SIREMOBIL
Medical Solutions 08.05 CS PS 24
10 General information
1.7 Maintenance interval
12 months
SIREMOBIL SPR2-130.831.01.02.02 Page 10 of 18 Siemens AG
08.05 CS PS 24 Medical Solutions
Inspection of exterior and surroundings 11
2-
2 Inspection of exterior and surroundings
2.1 Inspection of exterior
PMP Damage
• Inspect all power cables, plugs, socket panels and sockets for damage.
Siemens AG SPR2-130.831.01.02.02 Page 11 of 18 SIREMOBIL
Medical Solutions 08.05 CS PS 24
12 Inspection of exterior and surroundings
2.2 Inspection of environment
n.a.
SIREMOBIL SPR2-130.831.01.02.02 Page 12 of 18 Siemens AG
08.05 CS PS 24 Medical Solutions
Safety Inspection 13
3-
3 Safety Inspection
3.1 Mechanical Safety Inspection
SIM Mounting
• Check the mounting of the DICOM bridge in the monitor carriage.
Siemens AG SPR2-130.831.01.02.02 Page 13 of 18 SIREMOBIL
Medical Solutions 08.05 CS PS 24
14 Safety Inspection
3.2 Electrical Safety Inspection
SIE Ventilation
• Clean the ventilation slots on the DICOM bridge and inspect the fan for proper opera-
tion.
SIE Protective Conductor Test
• Measure the protective conductor with the protective conductor testing device accord-
ing to ARTD-002.731.17.
NOTE The protective conductor test is part of the safety inspection. It can
only be performed, however, when maintenance has been com-
pleted and the covers are closed or reattached. Refer to the "Final
work steps" section.
SIREMOBIL SPR2-130.831.01.02.02 Page 14 of 18 Siemens AG
08.05 CS PS 24 Medical Solutions
Final results/quality inspection/general maintenance 15
4-
4 Final results/quality inspection/general mainte-
nance
SIE Image quality (IQ) quick test
• As part of the system IQ quick test, perform the IQ Quicktest as described in Chapter 6
of the Installation and Adjustment Instructions, SPR2-130.033.04.
Siemens AG SPR2-130.831.01.02.02 Page 15 of 18 SIREMOBIL
Medical Solutions 08.05 CS PS 24
16 Final results/quality inspection/general maintenance
4.1 Final Work Steps
SIE Protective Conductor Test
• Perform the protective conductor test with the system's covers closed and panels at-
tached, according ARTD-002.731.17. The protective conductor resistance may not ex-
ceed 0.2 Ohms.
SIREMOBIL SPR2-130.831.01.02.02 Page 16 of 18 Siemens AG
08.05 CS PS 24 Medical Solutions
Changes since previous version 17
5-
5 Changes since previous version
Document was converted to DMS
Siemens AG SPR2-130.831.01.02.02 Page 17 of 18 SIREMOBIL
Medical Solutions 08.05 CS PS 24
18 Changes since previous version
SIREMOBIL SPR2-130.831.01.02.02 Page 18 of 18 Siemens AG
08.05 CS PS 24 Medical Solutions