LiquiPhysics Meters
One Click® Density Measurement
with Built-in Quality Control
One Click® operation
Measure your products with One Click!
The color touchscreen allows quick,
simple and direct access to all routine
™
tasks utilizing shortcut buttons to start
routine tasks immediately. Every
operator has a personalized home
screen with shortcuts and his own
operation language.
Immediate QC results
For each product lower and upper result
limits can be defined. Measured products
are automatically verified with the limits,
and the color of the result(s) indicates if
specifications are met: black = passed,
red = failed. Results can be displayed in
density, specific gravity, %Alcohol, Brix,
API or user-defined concentrations.
Smart error detection
Most measuring errors due to air
bubbles, remaining rinsing liquids and
faulty adjustments are eliminated due to
comprehensive test and error detection
capabilities, such as the Bubble Check™,
automatic multiple measurement with
evaluation and automatic cell test.
LiquiPhysics™ Excellence
The straight way to reliable results
DryPal™ drying pump Product database: no methods to select and no results
Used with DM density meters in to judge – just One Click on the product shortcut
order to dry the measuring cell prior Automatic error detection
to adjustments.
Tailored measuring procedures even for very particular
needs thanks to flexible method structure, conditions,
formula editor and user tables
Full SOP user guidance
Advanced user management including biometric
instrument protection
LabX® PC Software
LiquiPhysics™ Excellence
Product Order Number
DM40 10376200
DM45 DeltaRange™ 10376201
DM50 10376202
Specifications
DM40 DM45 DeltaRange™ DM50
Measuring range 0–3 g/cm3 0–3 g/cm3 0–3 g/cm3
Resolution 0.0001 g/cm3 0.00001 g/cm3 0.00001 g/cm3
Repeatability (S.D. n = 10) 0.00005 g/cm3 0.000005 g/cm3 0.000005 g/cm3
0.03 °Brix 0.015 °Brix 0.003 °Brix
0.1% v/v Alcohol 0.01% v/v Alcohol 0.01% v/v Alcohol
Limit of error 0.0001 g/cm3 0.00002 g/cm3 (0.7–1 g/cm3 Measuring Range) 0.00002 g/cm3 (0–1 g/cm3 Measuring Range)
0.03 °Brix 0.00005 g/cm3 (0–3 g/cm3 Measuring Range) 0.00003 g/cm3 (1–2 g/cm3 Measuring Range)
0.1% v/v Alcohol 0.02 °Brix 0.006 °Brix
0.02% v/v Alcohol 0.02% v/v Alcohol
Temperature range 0–91 °C 0–91 °C 0–91 °C
Limit of error temperature 0.05 °C (10–30 °C) 0.02 °C (15–20 °C) 0.02 °C (10–30 °C)
0.10 °C (0–91 °C) 0.05 °C (0–91 °C) 0.05 °C (0–91 °C)
Thermostat 2 built-in Peltier thermostats
Viscosity correction Yes
Automatic error detection Bubble Check™, verification of stored product limits, multiple measurement with definable max. S.D.
Built-in result units / Specific gravity, light and heavy Baumé, Twaddell
concentration tables Sugar: Plato, Brix (Emmerich, NBS 113), HFCS 42/55, Invert Sugar, KMW, Oechsle, Babo
Alcohol: OIML, AOAC, Proof Degree, HM C&E, Gay Lussac
Petro: API degrees and gravity tables for crude oils, refined products and lubricants
Up to 30 user-defined concentration tables (can be entered as tables or formulas)
Minimal sample volume Approx. 1.2 mL
Materials In contact with sample: borosilicate glass, PTFE, PP (syringe adaptor)
Dimensions and weight 272 x 385 x 215 mm / 10.70 x 15.16 x 8.46 in (W x L x H). 13.53 kg / 29.83 lbs
User Management
Number of users 30 (to each user a role with different rights can be assigned)
Languages en, de, fr, it, es, pt, zh, ru (can be set individually for each user)
Shortcuts 8 (individually for each user) for One Click® start of measurements, adjustments, tests etc.
Password protection Yes. Also biometric protection possible with LogStraight™ fingerprint reader
Data Handling
Printer Compact strip printer USB P-25, lightfast paper (normal or adhesive)
Results export Automatic or manual export to USB stick (CSV), export to RS-232 (XML),
LabX® LiquiPhysics: automatic storage in database and optional LIMS / ERP export
Import / export Share between instruments
Pumps
Drying pump DryPal™ (to dry measuring cell if no SC1 or SC30 is used)
Sampling pump FillPal™ Food (aqueous products), FillPal™ Chem (acids, bases, solvents)
Other Options
Certified standards Water (0.99... g/cm3), dodecane (0.75...), dichlorotoluene (1.25...), bromonaphthalene (1.48...)
www.fishersci.com/mettlertoledo
For more information
INS-2812-LIT
BN0620137