0% found this document useful (0 votes)
27 views30 pages

Siemens Mammomat Novation DR WH AWS VA11B Software

The document provides a detailed guide for the AEC Calibration Tool used with the MAMMOMAT Novation DR system, including calibration procedures and software reinstallation instructions. It outlines the steps for performing AEC calibration, determining the Half Value Layer (HVL), and generating goal tables, emphasizing that only qualified personnel should perform installation and service. Additionally, it includes a disclaimer about the document's usage and revision information.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
27 views30 pages

Siemens Mammomat Novation DR WH AWS VA11B Software

The document provides a detailed guide for the AEC Calibration Tool used with the MAMMOMAT Novation DR system, including calibration procedures and software reinstallation instructions. It outlines the steps for performing AEC calibration, determining the Half Value Layer (HVL), and generating goal tables, emphasizing that only qualified personnel should perform installation and service. Additionally, it includes a disclaimer about the document's usage and revision information.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 30

t

e
n
a
rd
U

MAMMOMAT Novation DR
SP

Software
System

WH AWS, VA11B

08380490

© Siemens AG 2005
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

Print No.: SPB7-250.816.06.01.02 English


Replaces: n.a. Doc. Gen. Date: 04.05
2 Revision / Disclaimer
1Revision / Disclaimer

Document revision level


The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.

Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.

MAMMOMAT Novation DR SPB7-250.816.06.01.02 Page 2 of 30 Siemens AG


04.05 CS SD 2 Medical Solutions
Table of Contents 3
1- 0Table of Contents

1 _______ AEC Calibration Tool_____________________________________________ 4

Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
When is AEC calibration required? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Calibration-related procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Starting the AEC Calibration Tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Exiting the tool and restarting the syngo application. . . . . . . . . . . . . . . . . . . . . . . . . . 5
Main window functions and options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Dark offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Setting the AEC dark offset values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Determining the Half Value Layer (HVL) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Determining the Entrance Skin Exposure (ESE) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Generating the goal tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

2 _______ WH AWS software reinstallation __________________________________ 21

Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Loading the software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Final steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

3 _______ Changes to the previous version __________________________________ 30

Siemens AG SPB7-250.816.06.01.02 Page 3 of 30 MAMMOMAT Novation DR


Medical Solutions 04.05 CS SD 2
4 AEC Calibration Tool
2-
Introduction
1AEC Calibration Tool

This chapter describes how to calibrate the AEC by generating the dark offset, a goal
table for a desired target/filter combination, and loading it into the DirectRay Power and
X-Ray Control Module (PXCM/Brick).
Use the AEC Calibration Tool to perform the calibration.

When is AEC calibration required? 0

AEC calibration should be performed:


• Once a year.
• If changes are made to any of the DirectRay components or to the generator (for exam-
ple, if the PXCM is replaced).

Calibration-related procedures 0

The purpose of AEC calibration is to determine a valid mAs value and to use that value to
take a series of exposures to generate goal images. The AEC Calibration Tool averages
the goal images and generates and loads a goal table into the PXCM.
There are various calibration procedures which are performed in sequence:
• Dark offset; described in (Dark offset / p. 9).
• Determining the Half Value Layer (HVL) of the X-ray generator for a given target/filter
combination; described in (Determining the Half Value Layer (HVL) / p. 10).
• The HVL is used to determine the Entrance Skin Exposure (ESE); see the descrip-
tion in . The resulting mAs value is used during goal image exposure, in the next proce-
dure.(Determining the Entrance Skin Exposure (ESE) / p. 13)
• The mAs value, determined in the ESE procedure, is used to generate the goal image
(see the description in (Generating the goal tables / p. 17)). This results in a goal table
which is loaded into the PXCM.

NOTE If you know the correct HVL value, as determined in the preceding
"Determine HVL" procedure, you can skip this procedure and go
right to the "Determine ESE" procedure.

Starting the AEC Calibration Tool 0

1. Log onto the service software. Select Options-> Service>Local service.


- Enter the service key (6 characters in the 2nd field) in the syngo software screen.
- Confirm the screen with OK.
- Select Acquisition System.
- Select AEC Calibration Tool.
- The syngo application software is automatically stopped.

MAMMOMAT Novation DR SPB7-250.816.06.01.02 Page 4 of 30 Siemens AG


04.05 CS SD 2 Medical Solutions
AEC Calibration Tool 5

- The AEC Calibration Tool main window the opens.


The main window is shown in the image below in (Fig. 1 / p. 5).

Fig. 1: Start the AecCalTool


The screenshot above (Fig. 1 / p. 5) shows the default configuration of the tool:
- The target/filter setting is Mo/Mo.
- kV is set to 28.

Exiting the tool and restarting the syngo application 0

To restart the syngo application software:

1. Select Exit in the AEC Calibration Tool.


2. Go to Utilities in the service software and select Source -> System.
3. Select Restart Application and press Go to restart the syngo application.

Siemens AG SPB7-250.816.06.01.02 Page 5 of 30 MAMMOMAT Novation DR


Medical Solutions 04.05 CS SD 2
6 AEC Calibration Tool

Fig. 2: Restart the syngo application

Main window functions and options 0

The AEC Calibration Tool window's functions and options are described in
(Tab. 1 / p. 7).

MAMMOMAT Novation DR SPB7-250.816.06.01.02 Page 6 of 30 Siemens AG


04.05 CS SD 2 Medical Solutions
AEC Calibration Tool 7
Tab. 1 Main window functions and options

Function/option Description
Procedure options Select the procedures you want to perform from the following
three options. Only one option can be selected at a time. A proce-
dure is not selected automatically when the tool starts. You must
select one of the three procedures to continue.
Determine HVL. Select this procedure to determine the Half
Value Layer (HVL) of the X-ray generator for a given target/filter
combination. For the procedure for determining the HVL, refer to
(Determining the Half Value Layer (HVL) / p. 10).
Determine ESE. Select this procedure to determine the Entrance
Skin Exposure (ESE). This step may be performed without deter-
mining the HVL of the X-ray generator. However, a valid HVL
value must be known. For the procedure for determining the ESE,
refer to (Determining the Entrance Skin Exposure (ESE) / p. 13).
Generate Goal. Select this procedure to generate the goal table
for the selected target/filter combination. This step may be per-
formed without determining the HVL of the X-ray generator or the
ESE. However, a valid mAs value must be known. For the proce-
dure for generating the goal tables, refer to (Generating the goal
tables / p. 17).
Target/filter The desired target/filter combination. The goal table is named to
match (Mo/Mo is goaltab1, Mo/Rh is goaltab2, W/Rh is goaltab3).
Make sure that the generator setting matches this selection.
Since this selection determines future instructions and calcula-
tions, it is disabled once the first HVL exposure is taken.
AGD Use this option to select the desired Average Glandular Dose to
be used for calculating the mAs value for calibration. In a future
version, this value will be logged at the end of the goal table file.
The default is 1.5 mGy (150 mR).
kV Use this option to indicate the kV to be used during calibration.
This selection in the AEC Calibration Tool window does not actu-
ally change the generator settings, but is intended to remind the
user of the current generator settings. The default is 28 kV.
mAs Use this option to indicate the mAs to be used during calibration.
This selection in the AEC Calibration Tool window does not actu-
ally change the generator settings, but is intended to remind the
user of the current generator settings.
Enable Capture Use this option to enable the DRAPI for image capture.
button
Units Use this option to select the radiation units to be used. You may
select either R/mR or Gy/mGy. When this selection is changed,
the AGD setting also changes to convert the value to the current
units. The selected unit is also displayed in the instructions field
after each exposure.

Siemens AG SPB7-250.816.06.01.02 Page 7 of 30 MAMMOMAT Novation DR


Medical Solutions 04.05 CS SD 2
8 AEC Calibration Tool

Function/option Description
Reading Use this field to enter the dosimeter reading, as instructed after
each exposure for HVL or ESE determination. Enter the numeri-
cal characters in the field and click the Submit button. The last
entry is cleared when the Enable Capture button is pressed.
Submit button Use this button to submit a radiation entry.
Status • Detector NOT Ready
The DRAPI is processing the previous exposure.
• Detector Ready
The DRAPI is ready to be enabled for exposure.
• Detector Armed
The DRAPI is ready for exposure. Press the generator expo-
sure button to start the exposure.
• Detector Exposed
The detector has been exposed to X-rays.

MAMMOMAT Novation DR SPB7-250.816.06.01.02 Page 8 of 30 Siemens AG


04.05 CS SD 2 Medical Solutions
AEC Calibration Tool 9

Dark offset 0

The dark offset has to be set to 0.

Setting the AEC dark offset values 0

1. Log onto the service software. Select Options-> Service>Local service.


- Enter the service key (6 characters in the 2nd field) in the syngo software screen.
- Confirm the screen with OK.
- Select Acquisition System.
- Select Brick Configuration.
For security reasons, you may be prompted to enter a login name and password before
executing this function.
User name: root
Password:*****
2. Select Brick and then Brick AEC Configuration.
3. Select Modify Settings and set Modify Array Dark Offset and Modify Array Mag Off-
set to 0 in the Dark Offset Value field.
Enter 0 for both.
- Short-term (Load) and for
- Long-term (Save)
by choosing the Duration of Change button.
4. After the changes have been made, scroll to the bottom of the page and click Back to
the Brick Control Panel....
5. Check the setting with the Status button.

Siemens AG SPB7-250.816.06.01.02 Page 9 of 30 MAMMOMAT Novation DR


Medical Solutions 04.05 CS SD 2
10 AEC Calibration Tool

Determining the Half Value Layer (HVL) 0

Select the Determine HVL procedure option for determining the HVL of the X-ray genera-
tor for a given target/filter combination.
This procedure requires taking four exposures.
• The first and the last exposure are "control" exposures with no attenuating material in
place.
• The second exposure is taken with aluminum in the path of the X-ray beam. For the mo-
lybdenum target, use aluminum with a thickness of 0.3 mm. For the tungsten target,
use aluminum with a thickness of 0.4 mm.
• The third exposure is taken with a larger aluminum thickness in the path of the X-ray
beam. For Mo/Mo, use aluminum with a thickness of 0.4 mm. For Mo/Rh, use a thick-
ness of 0.5 mm and for W/Rh, use 0.6 mm.
The result of the HVL procedure is a fractional number representing the HVL in millime-
ters of aluminum.

NOTE If you already know the HVL and want to proceed with determining
the ESE value, go to (Determining the Entrance Skin Exposure
(ESE) / p. 13).

Procedure
In the main AEC Calibration Tool window, select the Determine HVL procedure option.

1. Mount the 24x30 compression plate.


2. Start the AecCalTool:
startcalibration

MAMMOMAT Novation DR SPB7-250.816.06.01.02 Page 10 of 30 Siemens AG


04.05 CS SD 2 Medical Solutions
AEC Calibration Tool 11

3. In the main window, the settings change and instructions appear on the display. An ex-
ample of the main window with the selected Determine HVL procedure option is shown
below.

Fig. 3: Determining the Half Value Layer


Start with the following settings:
- You have to determine the HVL for Mo/Mo, Mo/Rh and W/Rh. The appropriate tar-
get/filter combination has to be set.
- kV has to be set to 28.
- Instructions for generator and dosimeter configuration are displayed in the text win-
dow.
4. Be sure to select the correct values. Then press the Start button.
X 5. Once the setup is complete, click the Enable Capture button.
Make sure the generator is configured properly and then take the exposure. The Status
indicator will change from Detector Ready to Detector Armed, Detector Exposed,
Detector NOT Ready and then to Detector Ready again.
6. Select the appropriate units (mGy/RmR).

NOTE Select "mGy" for "units", except in the USA, where the units have
to be set to "R/mR".

Siemens AG SPB7-250.816.06.01.02 Page 11 of 30 MAMMOMAT Novation DR


Medical Solutions 04.05 CS SD 2
12 AEC Calibration Tool

In the Readings field, enter the dosimeter value with the appropriate units: mR
(milliRoentgen) or mGy (milliGray).

NOTE If you enter values, be sure to use a decimal point (e.g., 4.5) and
NOT a comma.

Make sure the entry is correct, then click Submit.


The text window includes instructions for the next step.
7. Perform the displayed instructions.
The amount of aluminum added in this step depends on the target/filter selection.
(Mo/Mo = 0.3 mm; Mo/Rh = 0.3 mm; W/Rh = 0.5 mm)
8. Click the Enable Capture button.
X 9. Ensure that the generator is configured properly and take the exposure.
10. In the Readings field, enter the dosimeter value with the appropriate units. Make sure
that the entry is correct, then click Submit. The text window includes instructions for the
next step. The total thickness of aluminum in this step depends on the target/filter selec-
tion (Mo/Mo = 0.4 mm; Mo/Rh = 0.5 mm; W/Rh = 0.6 mm).
11. Perform the displayed instructions and click the Enable Capture button.
X 12. Ensure that the generator is configured properly and take the exposure.
13. In the Readings field, enter the dosimeter value with the appropriate units. Make sure
that the entry is correct, then click Submit.
The text window includes instructions for performing the final HVL step.
14. Perform the displayed instructions and click the Enable Capture button.
X 15. Ensure that the generator is configured properly and take the exposure.
16. In the Readings field, enter the dosimeter value with the appropriate units. Make sure
that the entry is correct, then click Submit.
The calculated Half Value Layer value is computed and displayed. The tool internally
stores this value, which is then used for the ESE procedure.
P 17. You have to determine the Half Value Layer for Mo/Mo, Mo/Rh and W/Rh. Note the
measured HVL values in Table 2.
Tab. 2 HVL values

Target/filter Min. HVL value Max. HVL Measured HVL


at 28 kV value at 28kV value
Mo/Mo 0.32 0.40
Mo/Rh 0.38 0.46
W/Rh 0.49 0.59

MAMMOMAT Novation DR SPB7-250.816.06.01.02 Page 12 of 30 Siemens AG


04.05 CS SD 2 Medical Solutions
AEC Calibration Tool 13

Determining the Entrance Skin Exposure (ESE) 0

Select the Determine ESE procedure option if you need to determine a valid mAs for
generating a goal image.

NOTE If you already know the valid mAs to use and want to proceed with
generating the goal image, go to (Generating the goal
tables / p. 17).

The ESE procedure requires the HVL. If the HVL procedure has already been performed,
the tool already has the HVL value. If you are skipping the HVL procedure to start the ESE
procedure, enter the appropriate HVL value.
The ESE procedure involves taking three exposures. The dosimeter readings for each
exposure are averaged to determine the ESE for the selected Average Glandular Dose
(AGD). This, in turn, suggests the mAs that should be used to generate the goal table.
To determine the mAs value:

1. In the tool's main window, select the desired AGD value before starting to determine the
ESE.
Once the first ESE exposure has been taken, the AGD value cannot be changed.

Siemens AG SPB7-250.816.06.01.02 Page 13 of 30 MAMMOMAT Novation DR


Medical Solutions 04.05 CS SD 2
14 AEC Calibration Tool

2. Select the Determine ESE procedure option.


If you know the HVL value, enter the fractional value, for example, 0.35.
If you do not know the HVL value, refer to (Determining the Half Value Layer
(HVL) / p. 10).

Fig. 4: Determine ESE


Start with the following settings:
- You have to determine the mAs for Mo/Mo, Mo/Rh and W/Rh. The appropriate Tar-
get/Filter combination has to be set. Start with Mo/Mo.
- kV has to be set to 28.
- AGD has to be set to 1.5 mGy.
- HVL is filled in automatically with the Determine HVL test result.
- Instructions for generator and dosimeter configuration are displayed in the text win-
dow.
3. Make sure that you have selected the correct values. Then press the Start button.
X 4. Click Enable Capture, then take the exposure.
The Status indicator will change from Detector Ready to Detector Armed, Detector
Exposed, Detector NOT Ready and then to Detector Ready again.
5. Select the appropriate units (mGy/RmR).

MAMMOMAT Novation DR SPB7-250.816.06.01.02 Page 14 of 30 Siemens AG


04.05 CS SD 2 Medical Solutions
AEC Calibration Tool 15

NOTE Set "mGy" for "Units", except in the USA, where the units have to
be set to "R/mR".

In the Readings field, enter the dosimeter value with the appropriate units: mR
(milliRoentgen) or mGy (milliGray).

NOTE If you enter values, be sure to use a decimal point (e.g., 4.5) and
NOT a comma.

X 6. Repeat steps 4 and 5 for the second and third ESE exposures. After the third exposure,
the entered average of the three dosimeter values is displayed. In addition, the sug-
gested mAs value to be used for goal table generation is displayed.

Fig. 5: Average mAs value

NOTE In most cases, the suggested mAs value does not exactly match a
mAs value that can be used to set the generator. In this case, it is
recommended that the next lower mAs setting be used to avoid ex-
ceeding the desired AGD during the following AEC exposures.

P 7. You have to determine the mAs for Mo/Mo, Mo/Rh and W/Rh. Note the measured
mAs values in (Tab. 3 / p. 16).

Siemens AG SPB7-250.816.06.01.02 Page 15 of 30 MAMMOMAT Novation DR


Medical Solutions 04.05 CS SD 2
16 AEC Calibration Tool
Tab. 3 mAs values

Target/filter mAs value


Mo/Mo
Mo/Rh
W/Rh

MAMMOMAT Novation DR SPB7-250.816.06.01.02 Page 16 of 30 Siemens AG


04.05 CS SD 2 Medical Solutions
AEC Calibration Tool 17

Generating the goal tables 0

Use the Generate goal procedure to generate the goal table for the selected target/filter
combination. The goal table is created for the mAs value used by the generator to create
the goal images.
Before performing the Generate goal procedure, you must have a valid mAs value. If you
do not know a valid mAs value, you must perform the HVL and ESE procedures to deter-
mine it. Refer to (Determining the Half Value Layer (HVL) / p. 10)and (Determining the
Entrance Skin Exposure (ESE) / p. 13).
During the procedure, 4 goal images are taken, and the average of these is used for the
final result. The tool instructions prompt you to enable capture and exposure for each of
the images. When the final image exposure is complete,
• the tool automatically generates the goal table with a name derived from the target/filter
combination used and
• then loads it into the DirectRay PXCM.

To generate the goal image:

1. Select the Generate goal procedure option in the tool's main window.

NOTE You can select the "Generate Goal" option without performing the
"Determine HVL" and "Determine ESE" procedures. In this case,
the current mAs value is used to configure the generator.

Siemens AG SPB7-250.816.06.01.02 Page 17 of 30 MAMMOMAT Novation DR


Medical Solutions 04.05 CS SD 2
18 AEC Calibration Tool

In the main window, the settings change and instructions are displayed.

Fig. 6: Generate the goal image


Start with the following settings:
- You have to determine the goal tables for Mo/Mo, Mo/Rh and W/Rh. The appropri-
ate target/filter combination has to be set.
- kV has to be set to 28.
- AGD has to be set to 1.5 mGy.
- Instructions for generator and dosimeter configuration are displayed in the text win-
dow.
2. Follow the instructions in the text field for generator configuration.

NOTE After determining the ESE, the suggested mAs value appears in
the text field. This merely serves as a reminder; the user must
manually set the proper mAs on the generator.

3. Make sure that you have selected the correct values. Then press the Start button.
X 4. Click the Enable Capture button, then take the exposure(s).

MAMMOMAT Novation DR SPB7-250.816.06.01.02 Page 18 of 30 Siemens AG


04.05 CS SD 2 Medical Solutions
AEC Calibration Tool 19

5. Once the final image has been written to a file, the images are averaged and the goal
table file is generated and installed on the BRICK.

Fig. 7: Generate the goal image


6. Simply reboot the BRICK before attempting to take AEC exposures.
- Log onto the service software. Select Options-> Service>Local service.
- Select Acquisition System.
- Select Brick Configuration.

Siemens AG SPB7-250.816.06.01.02 Page 19 of 30 MAMMOMAT Novation DR


Medical Solutions 04.05 CS SD 2
20 AEC Calibration Tool

7. Select Brick -> Reboot Brick.


The BRICK software switches to a higher user level, requiring you to enter a user name
and password.
User name: root
Password: ******

Fig. 8: Reboot the BRICK


8. You have to create the goal tables for Mo/Mo, Mo/Rh and W/Rh.
P After all goal tables have been created, the system is ready to be used in the AEC
mode.

MAMMOMAT Novation DR SPB7-250.816.06.01.02 Page 20 of 30 Siemens AG


04.05 CS SD 2 Medical Solutions
WH AWS software reinstallation 21
3-
Introduction
2WH AWS software reinstallation

The WH AWS installation is mainly automated and requires little user interaction. Not all
of the following steps require user interaction. Nonetheless, the steps have been included
to permit a better understanding of the overall process.
If the system does not respond right away, it typically does not mean an error has
occurred. In most cases, it takes a few moments for the system to react.
This chapter describes the complete WH AWS system software installation, including the
Windows XP installation.

Siemens AG SPB7-250.816.06.01.02 Page 21 of 30 MAMMOMAT Novation DR


Medical Solutions 04.05 CS SD 2
22 WH AWS software reinstallation

Preparation 0

Do the following before you start to reinstall the software:


• Back up all packages, if no up-to-date backup is available, using the service software
Backup & Restore function.
• Archive all patient images residing on the WH AWS.
• Write down the following data:

Regional Settings, Lan- Options -> Configura-


guage tion -> Regional Set-
tings
Service Key

MAMMOMAT Novation DR SPB7-250.816.06.01.02 Page 22 of 30 Siemens AG


04.05 CS SD 2 Medical Solutions
WH AWS software reinstallation 23

Loading the software 0

Prerequisites
• The PC is turned OFF at the beginning of the installation.
• No CD is placed in the CD-ROM drive.

Loading the WH AWS BIOS CD


1. Turn the PC ON and press F2, to enter the BIOS menu. Enter the BIOS password.
2. Press F9 in the BIOS menu to load the default BIOS settings. Confirm the switch with
YES.
3. Insert the BIOS CD into the CD-ROM.
Press the ESC button and leave the BIOS via Save Changes & Exit. Confirm with
YES. The PC reboots.
4. The PC boots from the CD-ROM and loads the BIOS version, if required. The follow-
ing messages are displayed.
BIOS Installation Done - System Ready for Software Installation
Please remove BIOS CD!
Please turn off or reset system!
5. Remove the BIOS CD.

Loading the WH AWS software


1. Insert the first WH AWS installation CD.
2. Reboot the PC.
MS-DOS (as part of Windows 98) is started from the CD-ROM.
Starting Win98
After a short time, a start screen is displayed in text mode. In the top right corner, a
message indicates that the EIDE, SCSI and CD-ROM drivers are loaded.
If any drivers cannot be found with this hardware configuration, an error message is
generated and can be ignored.

Siemens AG SPB7-250.816.06.01.02 Page 23 of 30 MAMMOMAT Novation DR


Medical Solutions 04.05 CS SD 2
24 WH AWS software reinstallation

3. After a short time, the screen changes and the PC starts to beep.
The following screen is shown:

Siemens AG B Med AWS VA11B installation (Int. WXP)


SELECT WINDOW INFORMATION WINDOW
Turn off computer to interrupt Creating a new primary partition
or press any key to confirm on first harddisk. This parti-
tion will be converted to NTFS
operation and to during Windows XP installation.
Confirm to DESTROY DATA on the
first partition of the first
CONTINUE INSTALLATION
harddrive.
HARDWARE DETECTION IN PROGRESS
Checking first harddrive...

4. Press Enter to continue with the installation. A new primary partition is created and the
PC is restarted.
5. Leave the installation CD in the CD-ROM drive. After a short time, the following screen
is displayed:

Siemens AG B Med AWS VA11B installation (Int. WXP)


SELECT WINDOW INFORMATION WINDOW
Distribution type: CUSTOMER Press the ENTER key to toggle
UDB-file: on CD-ROM the distribution type to be
installed.
Machine Name: MedPC
Keep data on other partitions
CUSTOMER: install release files
Country: Other
DEVELOPER: debug files, as well.
CONTINUE INSTALLATION
Use ARROW-KEYS to navigate
HARDWARE DETECTION RESULTS
Matrox Graphic adapter (G400 or G450) found

6. The following settings need to be made:


- Check the PCI card
Press ALT O. This checks the other hardware. The entry DRC PCI card found must be
shown. In case it is not shown, check if the card is plugged in correctly. After error
correction, reboot the PC and continue with the installation.
- Machine name
Enter the appropriate machine name.

MAMMOMAT Novation DR SPB7-250.816.06.01.02 Page 24 of 30 Siemens AG


04.05 CS SD 2 Medical Solutions
WH AWS software reinstallation 25

NOTE The machine name must be changed.


If you do not change the machine name ’MedPC’, the syngo instal-
lation cannot be properly completed.

- Partitioning
Move the cursor to Keep data on other partitions and toggle to create partitions on
hard drive 2.
Under Country:
USA must be selected for installations in the USA
or
Other must be selected for installations in ROW (Rest of World).
- Continue with the installation
Move the cursor to CONTINUE INSTALLATION and press Enter.
7. The final text on the screen changes. Press Enter again to continue with the installa-
tion. It is normal for the PC to start beeping again.

Siemens AG B Med AWS VA11B installation (Int. WXP)


SELECT WINDOW INFORMATION WINDOW
Distribution type: CUSTOMER Now destroying all data and cre-
UDB-file: on CD-ROM ating further partitions on hard
drive 1-2.
Machine Name: YBHH007154
Create partitions on harddrive
2 Please confirm to continue.

Country: Other
Press any key to continue.
HARDWARE DETECTION RESULTS
Matrox Graphic adapter (G400 or G450) found

The first and second hard drives are partitioned.

8. The installation files are copied from the CD-ROM to the hard drive. The screen chang-
es as follows:

Siemens AG B Med AWS VA11B installation (Int. WXP)


SELECT MENU INFORMATION WINDOW

Siemens AG SPB7-250.816.06.01.02 Page 25 of 30 MAMMOMAT Novation DR


Medical Solutions 04.05 CS SD 2
26 WH AWS software reinstallation

Distribution type: CUSTOMER Copying OEM setup files


UDB-file: on CD-ROM Operating System setup files
Machine Name: YBHH007154 Operating System fixes
Country: Other Additional device drives
Create further partitions on
harddrive 1-2
Installation in progress...
HARDWARE DETECTION RESULTS
Matrox Graphic adapter (G400 or G450) found

After the installation, files are prepared and the CD-ROM is unloaded. The PC is then
rebooted.

NOTE The CD-ROM must be removed immediately, otherwise the PC au-


tomatically reloads the CD after rebooting and the installation re-
starts.
If this happens, remove the CD-ROM without striking any key and
restart the PC.

9. The Windows XP setup copies the files to the hard drive. A scroll bar appears on the
screen. After that, the PC is restarted.
10. Windows XP continues preparing the files. The PC is restarted.
11. The file system is converted from FAT32 to NTFS. The PC is restarted.
12. The Windows XP setup starts for the first time in graphical mode (VGA mode, 640 x 480
pixels). The setup continues automatically, and several windows and dialogs appear.
All dialogs are automatically answered as part of the automatic installation scripts.

NOTE Do not manually interrupt the setup.

The PC restarts automatically.


13. At this point the WH AWS installation stops and displays the following DOS window.
=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=
Siemens AG B Med AWS VA11B installation (Int. WXP)
=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=
The installation has been configured to perform a partially
unattended installation. Now you have the opportunity to check
(and replace) system services, drivers and system settings.
Warning: No package has been found. Please insert the package to
be installed in a CD-Drive or press any key to skip installation
of Internet Explorer and the application.
Press any key to continue. . .
14. Insert the second WH AWS CD. Wait approximately 15 seconds until Windows XP rec-
ognizes the CD. Click the title of the DOS window and press Enter.

MAMMOMAT Novation DR SPB7-250.816.06.01.02 Page 26 of 30 Siemens AG


04.05 CS SD 2 Medical Solutions
WH AWS software reinstallation 27

15. Additional applications are installed (Internet Explorer, Acrobat Reader, etc.). After that,
the PC restarts.
16. A DOS window with the following text is displayed:
Installing Med Software
This installation step takes some time, but does not require any user interaction.
17. After another reboot of the system, a message appears asking if Internet Explorer is to
be the default browser. Accept this. The system recognizes that no application licenses
are installed.
- Click on Click here to start local service configuration to access the service
software.
- Enter the appropriate service license and select the Set as Default checkbox and
click OK.
- The service software home menu is displayed.
18. Remove the WH AWS software CD.

Restoring the system settings

1. Select Backup & Restore -> Restore in the service software home menu.

NOTE Before the restore operation is started, all patient registrations


must be closed.

2. Insert your backup medium and select it in the Backup menu under Drives. Select the
archives SW-Settings_02 and Security-Settings . For SW-Settings_02-Restore, se-
lect all groups on the right side.
3. Restore Security-Settings and SW-Settings_02.
4. Restart the system, i.e., select Local service-> Utilities-> System-> Restart system.
5. After the restart, the syngo login screen is displayed. The first login takes some time,
since the patient database is being created. The AWS login is complete when the ex-
amination task card is displayed.
6. If “Initialization of the Mammomat System failed” appears, ignore it and press OK.
7. Language and keyboard setup
The basic installation always uses English (United States) as the default for both the
user interface and the keyboard layout.
The languages French, English, Spanish and German are supported. Proceed as
follows.
- Select Options->Configuration->Regional and language options in the main
menu.
- Adjust the language and keyboard setup.
- Press OK.
8. Restart the system.
9. Select Backup & Restore->Restore.

Siemens AG SPB7-250.816.06.01.02 Page 27 of 30 MAMMOMAT Novation DR


Medical Solutions 04.05 CS SD 2
28 WH AWS software reinstallation

10. Select the files DXMG Mammomat and WH_AWS_backup_restore and perform a re-
store operation.
11. Remove the backup CD.
12. Restart the system.

Loading the BIOS CD


1. Load the BIOS CD. Reboot the PC.
2. The following message is displayed:
Installation Done - HIPAA installed.
The boot procedure from CD-ROM is now disabled again.
3. After the restart, the syngo login screen is displayed. The AWS login is complete when
the examination task card is displayed.

MAMMOMAT Novation DR SPB7-250.816.06.01.02 Page 28 of 30 Siemens AG


04.05 CS SD 2 Medical Solutions
WH AWS software reinstallation 29

Final steps 0

• Gain calibration has to be performed. Please see the MAMMOMAT Novation DR


Start-up manual for more details on how to calibrate the system.
• Take a normal image and visually check it.(see Startup Instructions,
SPB7-250.815.05..,Chapter Image Quality)

Siemens AG SPB7-250.816.06.01.02 Page 29 of 30 MAMMOMAT Novation DR


Medical Solutions 04.05 CS SD 2
30 Changes to the previous version
4-
3Changes to the previous version

Chapter Page Changes


All All Initial version

MAMMOMAT Novation DR SPB7-250.816.06.01.02 Page 30 of 30 Siemens AG