Manual Indicador NT200A
Manual Indicador NT200A
Manuals / Brands / CAS Manuals / Measuring Instruments / NT-200 Series / Operating manual / PDF
05.04.13
-1-
Table of Contents
0. Precaution ------------------ 3
1. Introduction ------------------ 4
3. Specification ------------------ 6
4. Dimensions ------------------ 7
7. Installation ------------------ 12
-2-
0. Precaution
Our Dealers: CAS feels that each of its valued customers should get the
best service available. Whether it’s the initial installation of our product,
maintenance/repair work, or simply answering questions about our
products, CAS Corporation and all of its Authorized Dealers are highly
trained to assist you with any need regarding CAS products.
-3-
1. Introduction
-4-
2. Features & Main Function
Features
Appropriate for Weight and Measurement System
Easy Operation
Full Digital Calibration
Weight Back-Up [ Weight Memory at Sudden Power Failure ]
4 Multi-Point Calibration
Command Mode Function (Control by PC – Request and Set the Data)
6 Wires / 4 Wires Load Cell (Select the Dip Switch inside)
Kg/lb Conversion & kg/lb Calibration
Lock Function of Front Key
User Message Print Function
Output the signal of Hi & Low limit, Zero, OK with serial comm.
Count, Limit, Accumulation and Percent Function
Set Point ( 20 of tare ), (Hi-limit, Low-limit, Count unit each )
Preset Tare
Gravity Compensation
Main Function
Various Printer Connection [ Roll DEP & Label DLP Printer Support ]
Users can Set the Desirous Max. weight and a Division Freely
Independent Zero Calibration
Self Hard Ware Test
* State Mark
- Only LED Version [ NT-200A / NT-200S ]
- Only LCD Version [ NT-201A / NT-201S ]
Nothing – Common Function
-5-
3. Specification
Digital Part
NT-200A, NT-200S LED (6 digit)
Display
NT-201A, NT-201S LCD (6 digit + Sign)
NT-200A, NT-200S 25 mm (Height)
Character Size
NT-201A, NT-201S 24 mm (height)
Display below zero "-" minus sign
Lamp ZERO, TARE, GROSS, NET, STABLE, HOLD,
UNIT(kg, lb)
AC Adapter AC 230 V (DC 9V, 600mA)
Operating Temperature -10℃ ~ 40℃
Product Size 96mm(H) x 200mm(L) x 52mm(W)
Product Weight Approx. 0.5Kg
Communication Part
Standard COM1 (RS-232 Printer & PC Interface )
Option RTC (Real Time Clock)
-6-
COM2 (RS-422 Multi Drop Interface)
* You can use COM2 for connecting printer (RS-232)
4. Dimension
NT-200A, NT-201A
NT-200S, NT-201S
-7-
5. Front Panel
NT-201A
Display Lamp
STABLE Measured Weight is stable
-8-
NET On : Net OFF : Gross
NT-200A
Display Lamp
STABLE Measured Weight is stable
ZERO Current Weight is 0 kg
TARE Used to weigh an item by using the container
-9-
Key Board
Used to enter setting value in TEST, CAL, and SET mode instead of
Numeric keys
▲ Increase setting value of first digit by one
▼ decrease setting value of first digit by one
◀ Moves one digit to left
Reset the Current Value
ZERO KEY
Used to remove small variations in the indicator’s zero
If pressed for 2~3 sec, you can go to SYSTEM Mode
TARE KEY
Used to weigh an item by using the container
Current weight is memorized as tare weight
If pressed in unload condition, tare setting is released
If pressed for 2~3 sec, you can go to KeyTare Mode
When you know the tare weight, you can enter the tare weight
PRINT KEY
If you print, current weight is added to total weight
- 10 -
If pressed for 2~3 sec, you can see "SUB = Z GrAn = T".
You can print out the sub total weight with ZERO key and
total sum weight with TARE key
After the sub total weight and grand total weight is printed
out, it is set to zero(0)
UNIT KEY
You can change the unit of weight ( kg/lb ) (USA Version)
All functions are used with the unit of weight is selected
in weight calibration in CAL MODE.
If you press UNIT key for 2~3 sec, power is turned on or off
6. Rear Panel
NT-200A, NT-201A
NT-200S, NT-201S
- 11 -
.CAL S/W Calibration Switch Bolt
.POWER Port for Power
.LOAD CELL Port for connecting Load Cell
.RS-232C Serial Interface COM1, COM2 Port
[connect to PC, Printer]
.F/G Ground Terminal [ Frame Ground ]
If it is not connected, it may cause trouble
7. lnstallation
- 12 -
Note: Wire color can be different depending on the manufacturer or
Load cell’s model
4 (SIG-) BLUE
8. Test Mode
- 13 -
Test Menu(TEST1 – TEST6)
Test 1 : Key test
Test 2 : A/D conversion test
Test 3 : Serial interface(Com1) test (RS-232)
Test 4 : Serial interface(Com2) test (RS-422/485: Option)
Test 5 : Printer test
TEST 1
KEY TEST
KEY FUNCTION DISPLAY DESCRIPTION
UNIT Go to TEST2 tESt 1 TEST 1 Condition
Testing 1 Press any key to test then the
OTHERS
Key Codes Code of that key is shown below
KEY
TEST 2
A/D Conversion TEST (Load Cell Test)
KEY FUNCTION DISPLAY DESCRIPTION
UNIT Go to TEST3 tESt 2 TEST 2 Condition
TEST 2 is performed automatically
24750
This value can be changed
Note 1. Check the numeric by loading and unloading a weight.
- 14 -
If the numeric is not changed or it is o, check the connection of the load cell.
TEST 3
RS - 232 TEST with Computer (RS - 232C)
KEY FUNCTION DISPLAY DESCRIPTION
ZERO Transmit ‘0
TARE Transmit ‘1 tESt 3 TEST 3 Condition
NET/GRS Transmit ‘2
PRINT Transmit ‘4 ----- Waiting for Transmission & reception
ENTER Transmit ‘8’ 0 ----- 1 Receive : 1 Transmit : 0
UNIT Go to TEST4 2 ----- 1 Receive : 1 Transmit : 2
Note 1. Before testing, connect RS-232C of PC to COM1 port of Indicator.
Note 2. If you send ASCII code 0 ~ 9 in PC keyboard, Indicator receives this data.
If you press "1(ZERO key) " of indicator, you can check it in PC.
- 15 -
TEST 4
RS - 422/485 TEST with Computer (RS - 422/485 : Option)
KEY FUNCTION DISPLAY DESCRIPTION
ZERO Transmit ‘0’
TARE Transmit ‘1’ tESt 4 TEST 4 Condition
NET/GRS Transmit ‘2’
PRINT Transmit ‘4’ ----- Waiting for Transmission & reception
ENTER Transmit ‘8’ 0 ----- 1 Receive : 1 Transmit : 0
UNIT Go to TEST5 2 ----- 1 Receive : 1 Transmit : 2
Note 1. Before testing, connect RS-422/485 of PC to COM2 port of Indicator and
run the communication program of PC.
Note 2. Before testing, connect COM2 port of Indicator.
Note 3. If you send ASCII code 0 ~ 9, Indicator receives this data.
If you press "1(ZERO key) " of indicator, you can check it in connected
device.
TEST 5
PRINTER TEST
KEY FUNCTION DISPLAY DESCRIPTION
Go to
UNIT
Normal Mode tESt 5
TEST 5 Condition
Testing the
ENTER
Printer
Note 1. You should set the use of printer and printer in Device mode.
Note 2. If printer connection and setting is done successfully,
the display shows "good" and if not, the displays shows error message.
Note 3. Test Output Form of Printer is as follows.
And see " TEST OK " in the Hyper Terminal also.
TEST OK
- 16 -
9. System Mode
SYSTEM MODE
MODE [G/N] key DISPLAY [ENTER] key ACTION
Weight MODE (Ⅰ) ↓ WEiGht END Weighing Mode
Count MODE (Ⅱ) ↓ CoUnt END Count Mode
Percent MODE (Ⅲ) ↓ Circulation PEr END Percent Mode
Key
Accumulation Accumulation
MODE (Ⅳ) ↓ ACC END
Mode
ENTER WEiGht END Weighing Mode
WEIGHT MODE(Ⅰ)
INITIAL DISPLAY DESCRIPTION
Weighing Mode
- 17 -
COUNT MODE(Ⅱ)
INITIAL DISPLAY DESCRIPTION
Counting Mode
(Ⅱ) Press the [TARE] key long to set the Count Mode in detail
DISPLAY KEY DESCRIPTION
Input the weight of unit with sample,
on ZERO and go to count weighing mode
- 18 -
PERCENT MODE(Ⅲ)
INITIAL DISPLAY DESCRIPTION
Percent Mode
(Ⅲ) Press the [TARE] key long to set the Percent Mode in detail
DISPLAY KEY DESCRIPTION
After weighing the weight of 100%,
on ZERO
go to the percent weighing mode
- 19 -
▌The use of key in the Counter & Percent Weighing Mode
KEY DESCRIPTION
The current value of counter or percent
Short
ZERO set to zero
Long Move to System mode
Short Save the value of tare
TARE
Long Enter the count or percent mode to set value
Short Display the weight of Gross or Net in turn
GROSS/NET
Long Move to Code mode (Product ID)
Short Print the current value
PRINT
Long Print the total value
Short Display the current weight for 3 seconds and
ENTER Return to the counter or percent mode
Long Move to the Function mode
Display the weight of 10-pcs(count) or the weight
Short of 100 % for 3 seconds and return to the
UNIT
counter or percent mode
Long Power off
Ref. User can set by 2.5times of external resolution in count and percent mode(unit:0.1%).
- 20 -
ACCUMULATION MODE(Ⅳ)
INITIAL DISPLAY DESCRIPTION
Accumulation Mode
- 21 -
10. Calibration Mode
Multi-point
rEAdY
CALMod
- 22 -
PRINT Key Long UNIT KEY
UNIT : 0 : kg 1 : lb
CAL 0 : Multi Calibration Setting
CAL 1 : Maximum Capacity Setting
CAL 2 : Minimum Division Setting
CAL 3 : Zero Calibration
CAL 4/6/8/10 : A Weight Setting (First~Fourth)
CAL 5/7/9/11 : Span Calibration (First~Fourth)
CAL 0
MULTI - CALIBRATION SETTING
KEY FUNCTION DISPLAY DESCRIPTION
CAL 0 CAL 0 Condition
P=1 1 Point Multi-Calibration
Increase or
P=2 2 Point Multi-Calibration
decrease
P=3 3 Point Multi-Calibration
set value
P=4 4 Point Multi-Calibration
ENTER
Next Menu Save the value and go to CAL 1
UNIT
1 (OFF) : You can calibrate indicator with zero calibration and span calibration.
2 ~ 4 (ON): You can set the weight what you want in the zero calibration and
the setting weight (CAL 4,6,8,10) and span calibration (CAL 5,7,9,11)
- 23 -
CAL 1
MAXIMUM CAPACITY SETTING
Available Setting Range : 1kg ~ 999,999kg
KEY FUNCTION DISPLAY DESCRIPTION
CAL 1 CAL 1 Condition
Increase or
decrease 5000 5000kg
set value
Move one
digit to left 10000 10000kg
CAL 2
MINIMUM DIVISION SETTING
Available Setting Range : 0.001kg ~ 100kg
KEY FUNCTION DISPLAY DESCRIPTION
CAL 2 CAL 2 Condition
Increase or 1 1kg
decrease
0.01 0.01kg
set value
ENTER Next Menu Save the value and go to CAL 3
Note 1. The minimum division means one division
Note 2. External resolution should be within 1/20,000.
(External resolution= a division/maximum weight).
- 24 -
CAL 3
ZERO CALIBRATION
KEY FUNCTION DISPLAY DESCRIPTION
CAL 3 CAL 3 Condition
Remove weight from the platter
ULoAd
and Press UNIT KEY
--------- Performing Zero Calibration
Good Zero Calibration is finished
ENTER Next Menu Save the value and go to CAL 4
Note 1. If zero calibration is finished successfully, the display shows “Good" and
then you will go to CAL 4 automatically.
CAL 4
WEIGHT SETTING
Available Setting Range : 1kg ~ 999,999kg
KEY FUNCTION DISPLAY DESCRIPTION
CAL 4 CAL 4 Condition
Increase or
decrease 1000 1000kg
set value
Move one
digit to left 2000 2000kg
- 25 -
CAL 5
SPAN CALIBRATION
KEY FUNCTION DISPLAY DESCRIPTION
Put the weight that is set in CAL 4
CAL 5
and then press the UNIT KEY
Repeat this step according to multi
LoAd1
calibration setting
--------- Performing Span Calibration
Good Span calibration is finished
Save Remove the weight,
press ENTER KEY Move to the
initial CALMOd
※ Example
Max. capacity: 5000 kg, Min. division: 1 kg, Multi calibration(CAL 0) : 4 point
A weight: 2000 kg 1EA, 1000 kg 2EA, 500 kg 1EA
CAL 0 : 4 CAL 1 : 5000 CAL 2 : 1 CAL 3 : Zero calibration
CAL 4 : 2000 CAL 5 : LOAD1 CAL 6 : 1000 CAL 7 : LOAD2
CAL 8 : 1000 CAL 9 : LOAD3 CAL 10 : 500 CAL 11 : LOAD4
- 26 -
Gravity Compensation :: ( G/N Key Long - )
If there is a different of gravity acceleration between the setting and calibration
place, it can be compensated to use this function.
GRAVITY COMPENSATION
KEY FUNCTION DISPLAY DESCRIPTION
G-CAL Gravity Compensation
Increase or
Gr CAL Gravity of calibration place
decrease
98000 Set the value
set value
ENTER Next Menu Save and go to Gr Set
Move one
Gr Set Gravity of using place
digit to left
98000 Set the value
- 27 -
Sealing Mode :: ( G/N KEY - )
STABLE CONDITION
KEY FUNCTION DISPLAY VAL. DESCRIPTION
Increase or 11 21: Judge if the weight is changed
decrease Within 1 division for 1 sec
set value S01 = ~ 45: Judge if the weight is changed
Within 2 division for 5 sec
ENTER Save &
99 99: Judrge if the weight is changed
UNIT Go to Next
Within 4.5division for 9 sec
Kg/lb CONVERSION
KEY FUNCTION DISPLAY VAL. DESCRIPTION
set value 0 Not used kg/lb conversion
↓↑ S03 =
ENTER Go to Next 1 Use kg/lb conversion
- 28 -
Zero Calibration :: ( PRINT KEY - )
ZERO CALIBRATION
KEY FUNCTION DISPLAY DESCRIPTION
After setting ULoAd Remove weight from the platter
the A/D value, and Press ENTER KEY
ENTER
go to initial --------- Performing Zero Calibration
CALMOd Good Zero Calibration is finished
Note 1. Only Zero Calibration is performed When the zero value exceeds the initial
Zero range in this mode.
Sealing Method
After calibration, you have to seal as follows.
1. Fasten the CAL Switch Bolt.
2. Connect the sealing wire as figure.
3. Press the sealing wax as figure.
- 29 -
11. Function Mode
How to Go to Set Mode
If you press the [ENTER] key during 2~3 seconds in normal mode,
you can go to Function Mode after displaying
"on" and " dEUI=Z " " SEt=t "
WEIGHT BACKUP
KEY FUNCTION DISPLAY VAL. DESCRIPTION
SET
Select 0 or 1 0 Not use weight backup
Save & F01
ENTER 1 Use weight backup (Zero, Tare….)
Go to F02
UNIT Save & Exit Save & go to Normal Mode
- 30 -
SET HOLD TYPE
KEY FUNCTION DISPLAY VAL. DESCRIPTION
Compute the average value of
Select 0 ~ 3 0
oscillating weight
Compute the maximum value of
1
oscillating weight
F02
Save & Compute the current value of
ENTER 2
Go to F03 oscillating weight
Compute the average value of
3
weight automatically (Live-stock)
UNIT Save & Exit Save & go to Normal Mode
BACK-LIGHT
KEY FUNCTION DISPLAY VAL. DESCRIPTION
0 Always off
Select 0 ~ 3 If you press any key, back-light is
1
F04 on for 5 seconds
If there is a weight change, back-
Save & 2
ENTER light is on for 5 seconds
Go to F05
3 Always on
UNIT Save & Exit Save & go to Normal Mode
- 31 -
SELECT LIMIT MODE
KEY FUNCTION DISPLAY VAL. DESCRIPTION
Select 0 or 1 0 Not use
Save & F06 1 Checker mode
ENTER
Go to F07 2 Limit mode
UNIT Save & Exit Save & go to Normal Mode
[ CHECKER MODE]
Weight
(Low Limit) (High Limit) OUT
Comm
0 kg 50 kg 100 kg PUT
Signal
1
ZERO 0
1
LOW 0
1
HIGH 0
1
OK 0
Note 1. All outputs are generated regardless of stable state.
[ LIMIT MODE]
Weight
(Low Limit) (High Limit) OUT
Comm
0 kg 50 kg 100 kg PUT
Signal
1
ZERO 0
1
LOW 0
1
HIGH 0
1
OK 0
Note 1. OK signal is output when the state is stable only.
- 32 -
KEY FUNCTION DISPLAY VAL. DESCRIPTION
Select 0 ~ 2 0 When the PRINT Key is pressed
Automatic accumulation - only
1
Save & F07 stable state
ENTER
Go to F08 Automatic accumulation - when the
2
state is OK in the limit mode
UNIT Save & Exit Save & go to Normal Mode
Note 1. If you want to modify the value which is input in the F01~F07 previously,
reset to zero with [ZERO] key and then input the value what you want to
with key.
KEY LOCK
KEY FUNCTION DISPLAY VAL. DESCRIPTION
Select 0 or 1 0 Not change the key lock
Save & F08
ENTER 1 Set the key lock item (L01 ~ L10)
Go to F09
F08 = 1 (Key Lock)
0 : Not L01 Zero Key
lock the key L02 Tare Key
1 : Lock L03 Gross Key
0
that key L04 Print Key
Sub : L05 Enter Key
or
Press for L06 Unit Key
3 seconds L07 Zero Sub Key
1
L08 Tare Sub Key
Select 0 ~ 1 L09 Gross Sub Key
ENTER Save & Next L10 Print Sub Key
UNIT Save & Exit Save & go to Normal Mode
Note 1. Although key lock is set after setting the value of F08 to 1,
The value of F08(1) is not saved. Always the value of F08 is started to 0.
Same to F09 and F10.
- 33 -
KEY FUNCTION DISPLAY VAL. DESCRIPTION
Select 0 or 1 0 Not set
Save & F09
ENTER 1 Set the range of password
Go to F10
F09 = 1 (Set the range of password)
ENTER PASS Display
___0 Using the key, Input the password
ENTER Input password
Good 4 digits
PASS - 0 Not set the Enter key password
Select 0 or 1
PASS - 1 Set the Enter key password
UNIT Save & Exit Save & go to Normal Mode
Note 1. If the password is not correct, the message of "AgAIn" is displayed.
If you failed three times continuously, the message of "Fail" is displayed
and exit this mode automatically.
DEVICE Mode
- 34 -
Menu Description Menu Description
D01 Device ID D08 Automatic print
D02 COM1(RS-232) use D09 Line feed
D03 COM1 transmission method D10 Select print format
D04 COM1/2 Baud rate D11 Select date print format
D05 COM2 (RS-422) use D12 Select Product ID print
D06 COM2 transmission method D13 Select user’s message print
D07 Printer type D14 Set the current clock
DEVICE ID
KEY FUNCTION DISPLAY VAL. DESCRIPTION
Increase or 01: ID=0, 99: ID=99
decrease 00 If you use a system, you can iden-
set value d01 ~ tify each indicator by this device ID
99 If you set it to 00, there is no trans-
Save &
ENTER mission response cause 00 means
Go to d02
that is no device
UNIT Save & Exit Save & go to Normal Mode
- 35 -
Transmit in stable & unstable
1
d03 condition
Save & 2 Transmit only in stable condition
ENTER
Go to d04 3 Command mode
UNIT Save & Exit Save & go to Normal Mode
Ref) Command mode can’t apply com1, com2 at once.
PRINTER TYPE
KEY FUNCTION DISPLAY VAL. DESCRIPTION
- 36 -
Select 0 ~ 2 0 Do not use
Save & d07 1 DLP (Label printer)
ENTER
Go to d08 2 DEP (Roll printer)
UNIT Save & Exit Save & go to Normal Mode
AUTOMATIC PRINT
KEY FUNCTION DISPLAY VAL. DESCRIPTION
Select 0 ~ 2 0 Do not use
Save & d08 1 When the weight is stable
ENTER
Go to d09 2 When the weight is OK in limit mode. only 201
LINE FEED
KEY FUNCTION DISPLAY VAL. DESCRIPTION
Select 0,1,5 1 1 line feed
Save & d09 |
ENTER
Go to d10 5 5 line feed
UNIT Save & Exit Save & go to Normal Mode
- 37 -
001 ID_01 120.52 kg ID_01 TOTAL
002 ID_01 100.50 kg --------------------------------------------------
003 ID_01 50.28 kg 2004.06.24 14:32:54
COUNT 22
Format 2 WEIGHT 4234.48kg
2004.06.24 14:32:54 --------------------------------------------------
No. 001 ID_01 GRAND TOTAL
Gross : 120.52kg --------------------------------------------------
Tare : 50.00kg 2004.06.24 14:32:58
Net : 72.52kg COUNT 123
WEIGHT 23423.42kg
Note 1. The No which is able to be printed is from 1 to 999.
Note 2. The unit of weight is changed by the mode (kg/lb)
- 38 -
Select 0 ~ 1 0 Do not print the user’s message
Save & d13
ENTER 1 Print the user’s message
Go to d14
UNIT Save & Exit Save & go to Normal Mode
Note 1. How to input the format of user’s message print is explained chapter 14
CAS & Command Mode Protocol in detail.
SELECT CLOCK
KEY FUNCTION DISPLAY VAL. DESCRIPTION
Select 0 ~ 1 0 Do not use clock
d14
ENTER Go to next 1 Use clock
Change date / time (Ex. 2004.08.31 14:28:47) [d14 = 1]
KEY FUNCTION DISPLAY DESCRIPTION
Increase or C1 : 04 Year : 2004
decrease C2 : 08 Month : 08
Shift to one C3 : 31 Day : 31
digit to left C4 : 14 Hour : 14
Save & go to C5 : 28 Minute : 28
ENTER
weighing mode C6 : 47 Second : 47
UNIT Save & Exit Save & go to Normal Mode
Note 1. Although key clock is set after setting the value of d14 to 1,
the value of d14(1) is not saved. Always the value of d14 is started to 0.
- 39 -
12 - 1. HOW to SET the Original Number (Product ID) of Goods
STEP VFD DISPLAY and KEY INPUT PLATFORM DESCRIPTION
1 0kg Empty
2 1500kg Goods(Iron) Weight(1500kg)
3 Press [G/N] key for 2~3 seconds. “COdE”
Display : “CodE=z LiM=t tArE=g”
201
Display : “On” -> “ID” 200
4 Press [ZERO] key and input 10 Input ID(=10) of iron
201
Input product ID 200
Show the ID of goods
5 Press [ENTER] key Goods(Iron)
Display the weight
Note 1. The ID of goods can be from 0 to 19.
- 40 -
It’s displayed “tArE = z” “t-id = t”,
2 Empty
press [ZERO] key
3 Input the value of tare which you know
already with key
7 Press [ENTER] key and exit
- 41 -
It’s displayed “Sys”, Enter the counter
2
mode with [G / N] key
3 Press [TARE] key for 2~3 seconds
It’s displayed “SAMP=Z in=t”,
4
press [ZERO] key
After displaying “SAMPLE” “UnLoAd”
“A/D value”, press ENTER Key and then
5 after displaying “LoAd” “A/D value” Sample
put the sample on the platform and
press [ENTER] key
6 Display : “Good” “nUMbEr” “0 pcs”
7 Using the key, input the number Sample
8 Press [ENTER] key Sample
- 42 -
It’s displayed “Sys”, Enter the percent
2 Empty
mode with [G / N] key
3 Press [TARE] key for 2~3 seconds
It’s displayed “SAMP=Z in=t”,
4
press [ZERO] key
After displaying “SAMPLE” “UnLoAd”
“A/D value”, press ENTER Key and then
5 after displaying “LoAd” “A/D value” Sample
put the sample on the platform and
press [ENTER] key
Display : “Good” “Per” “0 %”
6 Using the key, Sample
input the percent of one sample
7 Press [ENTER] key Sample
- 43 -
13. Communication
How to connect PC (COM1)
- 44 -
OPTION ( RS-485/422 :: COM2 & CLOCK )
- 45 -
14. CAS & Command Mode Protocol
- 46 -
Command Mode Protocol
Command (ASCII code) Description State
HI High Limit Value Read / Write
LO Low Limit Value Read / Write
KT Key Tare Value Read / Write
CO Code Value Read / Write
WT Current Weight Read
ZE Operating like the ZERO key Read
TR Operating like the TARE key Read
GN Operating like the Gross/Net key Read
ID Change Device ID Read
HD Operating like the ENTER key Read
PR Operating like the PRINT key Read
TP Operating like the Total Print key Read
PW Power off Read
Read
1 2 3 4 5
Device ID Command CR LF
Note. 1 Device ID is hex and Command is ASCII
[Ex] Device ID is 13 when user want to know the current weight.
-> 0x0d 0x57 0x54 0x0d 0x0a
Write
1 2 3 4 5 6 7 8 9 10
Device ID Command DATA (Not include decimal point) CR LF
Note. 1 Device ID is hex and DATA is ASCII
[Ex] When user want to change hi limit weight (10.50kg).
-> 0x02 0x48 0x49 0x30 0x31 0x30 0x35 0x30 0x0d 0x0a
Note. 1 When Device ID and Code is changed, Data value is HEX and 1byte.
P.S.
1. To input the command, set the value D02=1 or 2, D03=3 when com1 uses
and D05=1 or 2, D06=3 when com2 uses.
- 47 -
CAS DLP Protocol
VARIABLE DESCRIPTION
V00 Gross Weight (8 bytes)
V01 Tare Weight (8 bytes)
V02 Net Weight (8 bytes)
V03 Barcode (net weight) (8 bytes)
V04 Number of count when count mode (8 bytes)
V05 Percentage when percent mode (8 bytes)
It is impossible to print the weight, count and percentage at the same time.
If you do, only one data which is met the current mode can be printed correctly
- 48 -
ASCII CODE
Letter Hex Letter Hex Letter Hex Letter Hex Letter Hex
20 4 34 H 48 \ 5C p 70
! 21 5 35 I 49 ] 5D q 71
“ 22 6 36 J 4A ^ 5E r 72
# 23 7 37 K 4B _ 5F s 73
$ 24 8 38 L 4C ‘ 60 t 74
% 25 9 39 M 4D a 61 u 75
& 26 : 3A N 4E b 62 v 76
‘ 27 ; 3B O 4F c 63 w 77
( 28 < 3C P 50 d 64 x 78
) 29 = 3D Q 51 e 65 y 79
* 2A > 3E R 52 f 66 z 7A
+ 2B ? 3F S 53 g 67 { 7B
, 2C @ 40 T 54 h 68 | 7C
- 2D A 41 U 55 i 69 } 7D
. 2E B 42 V 56 j 6A ~ 7E
/. 2F C 43 W 57 k 6B END FF
0 30 D 44 X 58 l 6C
1 31 E 45 Y 59 m 6D
2 32 F 46 Z 5A n 6E
3 33 G 47 [ 5B o 6F
- 49 -
15.Error Message & Trouble Shooting
- 50 -