0% found this document useful (0 votes)
112 views11 pages

XY-WM Series Moisture-analyzer-Manual

xy

Uploaded by

kminina2014
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
112 views11 pages

XY-WM Series Moisture-analyzer-Manual

xy

Uploaded by

kminina2014
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 11

Chanzhou Xingyun Elecetronic Equipment Co., Ltd.

Halogen Moisture Analyzer


About the Halogen Moisture Analyzer

Mode XY-105MW XY-102MW XY-100MW

Specification 110g/5mg 110g/2mg 110g/1mg

Readability 5mg 2mg 1mg

Tray size Φ 90

Heating mode Halogen Lamp

Display LCD

Temperature Setting 40℃~199℃

Time Setting 1~99 minutes

ATRO 100%~999%

ATRO 0%~999%

History store 15 set

1
Chanzhou Xingyun Elecetronic Equipment Co., Ltd.

I Instruction
1. Security
Please reading the following before you use the Moisture analyzer
a. The instrument can use for test the moisture of the sample,each
improper operation may result in personal injury and equipment
damage
b. Please access the AC power accordance with the manual;
c. Please make sure the power cord does not interfere with people
action;
d. Do not use the instrument under the hazardous,wet or unstable
enviroment;
e. Unplug the power before you clean the analyzer;
f. Please keep enough space around the instrument,and keep the top of the space
at least 1m;
g. During a special experiment, please wear protective equipment by youself;
h. Please do not change the instrument's components and any other aspects.
Maintenance of this equipment by the factory authorized professionals。
Moisture analyzer use Halogen lamp.
i. Do not put the flammable materials around the instrument;
j. Do touch the halogen lamp when it on working.
Special attention to the dangers of some sample;
k. Before you test the moisture of dangerous good,pleaes take Full
investigation
Inflammable and explosive materials:Heating the inflammable and
explosive materials will produce flammable gases or vapors, so need
to lower the temperature in a dry environment. In case of fire or
explosion. Toxic and corrosive substance: Keep the samples of toxic or
corrosive ingredients in well-ventilated environment.
Above the sample test will be particularly careful, for any damage caused by your
own risk
2
Chanzhou Xingyun Elecetronic Equipment Co., Ltd.

II.Installation
2.1 Standard Packing List Please check the packaging and accessories.
Standard Quantity Note
Moisture Analyzer Mainframe 1 set

Wind Cover 1 piece


The sample pan holder 1 piece
Tray handle 1 piece
Power cord 1 piece
Aluminum sample tray 1 set 50 pcs of aluminum pan

100g standard weight 1 piece


Product Manual 1 piece
Product certification 1 piece
Warranty card 1 piece

2.2 Moisture meter placement requirements.

Operating table Temperature stable,


good ventilation
should be stable dry,without strong vibration
of the environment
without strong magnetic field

2.3 Installation of accessories

Place the wind cover, Place the sample tray Put the aluminum sample
aligned holes. holder, rotate until tray on the bracket.
positioning.

3
Chanzhou Xingyun Elecetronic Equipment Co., Ltd.

2.4 Adjust the level and connect the power

There are two feet in front of the instrument,adjust it and keep the
instrument level.

Connect the instrument with power,put one side of power cord

Please warm up more than 30minutes for keep it stable and accurate.

III Operation

3.1.Display

3.2.Icon Introdution

----The percentage of dry weight ----The percentage of moisture

----Gram ----Temperature

---- Time set and heat time

----Heated mode

----Standard mode ----Rapid heating mode

---- Gentle heating mode ---- Stop heated mode

---- Automatic -----Manual stop mode

----Timing stop mode

---- Heated state

4
Chanzhou Xingyun Elecetronic Equipment Co., Ltd.

3.3. Keypad

NAME Key Features

ON/OFF Power on / off, Just off the internal DC power supply

START / STOP Start heated/ stop heated

TARE Weighing peeled, Settings exit

MENU Enter the setting menu

CONVERSION Switching measurement results

ENTER Confirm the current setting, go to next


Return to the weighing mode (short press), calibration
RESET
balance (long press)
DOWN Reduce

UP Add

5
Chanzhou Xingyun Elecetronic Equipment Co., Ltd.

3.4 First testing


1. First test, if not change the temperature and time settings, the moisture meter defaults
to factory defaults.
2. Put empty sample pan on the holder (see section 2.3).
3. Press the “TARE” to clean weight.
4.Put the sample on the pan (Sample weight should be heavy than 1g, the general 3-5g).
5. Close the heating cover, press the "Start / Stop" button to start the test. At this point,
heated tube lights, fans turn on, heated symbol in the upper left flashing. While
rolling on the screen displays the current temperature, has been heating time,
moisture content of the current sample.
6. End of the test, heated lamp automatic shut off, the buzzer alarm. Then the sample is
displayed on the moisture content, heating time, current temperature and the parameter
setting mode.Note: Don't touch the sample plate,the temperature is too high now.
7 .Press the "Reset" button, the instrument returns to weighing mode.
3.5 Parameters Setting
3.5.1 Heated parameter storage and recall
a. The instrument can store a total of 16 0-F group of heated parameters. Each
group can set the heating method, stop mode, time and time test parameters.Press
the"MENU" button to enter the parameter setting mode, press the "ENTER" key
until the shutdown mode.When flashing .Press"UP"or"Down"to choose
the stop mode.and press "ENTER" confirm,Parameters are defined, and go to
the next parameter (temperature) setting.
b. In weighing mode, press the "Settings" button to enter the parameter setting mode,
then store the address number (the number show the highest bit) is flashing. Press
"down" or "up" key to select the desired address number, the 0 is the factory
setting,1-F for the user store.
c. The memory address number 0 means factory settings, then click "ENTER"
button to exit the settings back to weighing mode. At this point heated mode is
the factory defaults.
d. When the memory address range for the 1-F any one number, press "ENTER"
6
Chanzhou Xingyun Elecetronic Equipment Co., Ltd.

key, and enter the parameter setting after the completion of a parameter (heated
mode, stop mode, time mode) setting.Press the "TARE" key, the address
number of the test parameters have been stored.
3.5.2 Heated mode and setting( )
a. Standard heated mode:( )
This mode is the factory setting, suitable for most samples. Heating
temperature can be set, the sample is heated to a temperature, and keep to the
test ends automatically.
b. Rapid heating mode: ( )
This mode is suitable for samples with higher moisture content. You can set
the heating temperature, heating the temperature will exceeds the set value,
the future will be reduced to the set temperature and hold until the test ends
automatically.
c. Gentle heating mode: ( )
This mode is suitable for low moisture content, surface casing is easy to form,
easy to burn the sample. From the heat began to rise to the long time required
to set temperature, heating up even slowly. Heating time user selectable.
d. Press the "MENU" to enter the parameter setting mode, press the "ENTER"
until the heating mode, the icon flashes 。Press "UP"or"Down"choose
the mode you need.and press "ENTER"confirm.
3.5.3 Stop mode setting:( )
a. Automatic stop mode:( )
Heating the sample within the specified time during the weight loss of less
than 2mg will automatically stop when the test. This mode after 15 seconds
in the heating effect.
b. Manual stop mode:( )
Sample heating process is the continuing until the press the "start/stop"
button to end testing.

c. Timing Stop Mode:( )

7
Chanzhou Xingyun Elecetronic Equipment Co., Ltd.

Stop time set by the user, the sample heating process is continued until the
end of the heating time to set the value of the test.
3.5.4 Heating Temperature Setting
The instrument temperature setting range is from 0-199 degrees. Press the
"MENU" to enter the parameter setting mode, press the "ENTER" key until
the flashing,Press"UP"or"Down"to choose the temperature.and press
"ENTER" to go to next setting(time setting)
Icon will transform other temperature when its heating.
3.5.5 Heating time setting
1.Just choose timing stop mode need set the stop time.
Press the "MENU" to enter the parameter setting mode, press the "ENTER"

key until the flashing , Press"UP"or"Down"to choose the

temperature.and press "ENTER"


2, Heating cycle:Just gentle heating mode need choose heating cycle
3.6 For example:
a. sample:water,
b. Press the "MENU" to enter the parameter setting mode,choose the (

),temperature:120℃,Press "ENTER" go to weighing state

c. Put a dry paper on the pan,and press"Tare" clean the weight.


Put on the 3g water on the paper,until the readability stable.
Press" START / STOP " begin test.
Waiting for it sop.look the result.
The m% will be 100%.
Weighing error or sample weight deficiency can lead to the result of the test a
deviation

8
Chanzhou Xingyun Elecetronic Equipment Co., Ltd.

3.7 Calibration
When the instrument error or improper operate. Calibration weights can be
adjusted with accuracy.
Press "Reset" key and does not release until it display“-cal-”,
Then flashing “100.000”, put on the calibration weight, then it will display "= = =
= = =", until it display “100.000” remove the weight, it will display "= = = = =
= "again, Waiting it show" 0.000 ", the calibration is completed.

IV RS232 Setting

Baud Rate:1200

Date Bits:7

Odd-even check:No

Flow control:Continuous

Rs232 Commands

Commands Function

On/Off Power on /Power off

g % Conversion results

9
Chanzhou Xingyun Elecetronic Equipment Co., Ltd.

Schedule: Assembly drawing

Please installation the instrument according the picture


Installation order from top to bottom :1 --2--3--4--5
1. Aluminum scale weighing tray
2. The triangle scale frame
3. Scale tray brackets
4. Wind cover
5. Stainless steel plate

10
Chanzhou Xingyun Elecetronic Equipment Co., Ltd.

Tips:
1. Warm-up the analyzer 30 minutes and take the first time calibration;
2. After the end of a water calibration experiments (see 3.6);
3.Each time the switch machine, please interval of 5 seconds;
4 .Cover the heater, do not leave the weight on the pan, or halogen lamps will be
crushed;

11

You might also like