Neofuge13R Service Manual 20130704
Neofuge13R Service Manual 20130704
Important!
Before maintenance of Neofuge 13R centrifuge, please read the manual carefully.
A maintainer must pass certified qualification for maintenance of the centrifuge.
The centrifuge and its accessories shall be prevented from being damaged by
maloperation.
During rotation of rotor, the cover mustn’t be opened by any mean.
During maintenance of the centrifuge, please work strictly as the manual.
During maintenance of the centrifuge, damaged parts can be replaced only with
fittings from us.
I
Maintenance Manual of Neofuge 13R High Speed Refrigerated Centrifuge
II
Maintenance Manual of Neofuge 13R High Speed Refrigerated Centrifuge
Contents
1. Manual opening ........................................................................................................................... 1
2. Precautions for centrifuge installation ...................................................................................... 2
3. Installation and replacement of rotor........................................................................................ 2
4. Rotor loading ............................................................................................................................... 4
5. Alarm message code and troubleshooting ................................................................................. 5
6. Troubleshooting ........................................................................................................................... 6
6.1 Start and no display .................................................................................................... 6
6.2 E30 MOTOR OVER HEAT ....................................................................................... 6
6.3 E34 BRAKE RESISTANCE OVER HEAT .............................................................. 7
6.4 E37 CONVERTER OVER LOAD............................................................................. 7
6.5 E40 ROTOR NOT SECURED ................................................................................... 8
6.6 E41 LID NOT SECURED .......................................................................................... 8
6.7 E42 LID OPEN FAILED ............................................................................................ 9
6.8 E51 OVER ROTOR MAX SPEED............................................................................ 9
6.9 E52 OVER ROTOR SETTING SPEED.................................................................. 10
6.10 E53 SPEED CONTROL FAILED ........................................................................... 10
6.11 E54 TEMP CONTROL FAILED ............................................................................. 11
6.12 E55 TEMP OVER 40℃ ............................................................................................ 11
6.13 E56 SPEED SENSOR ERROR ................................................................................ 12
6.14 E57 TEMP SENSOR ERROR ................................................................................. 12
6.15 E58 ROTOR ID SENSOR ERROR ......................................................................... 13
6.16 E60 COMMUNICATION ERROR ......................................................................... 13
7. Dismounting and installation of the machine ......................................................................... 14
7.1 Preparatory work ...................................................................................................... 15
7.2 Dismounting of front cover....................................................................................... 16
7.3 Disassembly of front cover ....................................................................................... 17
7.4 Dismounting of door lock assembly ......................................................................... 18
7.5 Dismounting of fan assembly ................................................................................... 19
7.6 Dismounting of rear cover board ............................................................................. 20
7.7 Dismounting of lid seal ring and upper cover......................................................... 21
7.8 Dismounting of brake resistor .................................................................................. 22
7.9 Dismounting of motor sheath and temperature sensor Pt100 ............................... 23
7.10 Dismounting of positioning disc and rotor ID board ............................................. 24
7.11 Dismounting of motor assembly............................................................................... 25
7.12 Disassembly of motor assembly ............................................................................... 26
7.13 Dismounting of lid ..................................................................................................... 27
8. Description of Neofuge13R circuit ........................................................................................... 29
8.1 Functional block diagram description of Neofuge13R circuit............................... 29
8.2 Description of circuit boards and devices ............................................................... 31
8.3 Circuit functional block diagram 错误!未定义书签。
9. Neofuge 13R explosion diagram and description ................................................................... 37
III
1. Manual opening
When opening the door of the centrifuge fails, or the door can not be electrically opened arising
from sudden power failure in the running process of the centrifuge, but the sample in the
centrifuge chamber must be taken out at once, here the lid can be opened by manual method and
opening steps are descried as below:
Make sure the rotor stops completely (observe from the observation hole at the top of lid);
In the case of power failure, there is no brake function when the rotor stops running, so it will
take a long time to make the rotor stop completely, please wait with patience!
An emergency door-opening hole is available on the left top side of the machine (shown as 2 in the
above figure) and you can use a screwdriver with diameter of less than 4mm or any other hard tool
(shown as 1 in the above figure) to vertically insert into the hole and push hard, door lock will be
released and here the lid can be opened. After the lid is opened, manual opener will be automatically
reset under the action of spring force. Here the state of door lock is the same as that after opened by
pressing “Open” button and it can be normally used when electrified next time.
Attention: the method is used only in an emergency and it can not be used at
will.
1 2
2. Precautions for centrifuge installation
Desk centrifuge must be installed on the rigid enough horizontal table (the area required for
Neofuge 13R is about 30cm × 52cm; at least 15cm space shall be available around the machine.
After installed, the machine shall be horizontally adjusted to make the four standing legs at its
bottom uniformly supported on the table.
After a new machine is unpacked, it must not be started immediately; the lid must
be opened first to take out the cushion for fixing rotor and the rotor; the chamber
and rotor shall be cleaned; then the rotor shall be installed correctly for starting of
the machine.
Electrical connection
Before the centrifuge is made power on, make sure the power supply voltage is in conformity with
the nominal voltage indicated on the nameplate of the centrifuge (Neofuge 13R applied power
supply is AC230V, 50/60Hz); an independent power socket shall be used; power wire must be
reliably grounded.
Power switch
Main power switch is on the right side at the back of the centrifuge. After the centrifuge is made
power on, LCD screen will display the logo icon of Heal Force and version number of software
several seconds (the machine is normally initialized) and subsequently enter into main interface.
Prompt: after the power switch is turned off, setting parameters of final running of the machine
will be saved automatically.
E37 CONVERTER OVER LOAD Detailed judgment and handling procedures are stated as page 8
E40 ROTOR NOT SECURED Detailed judgment and handling procedures are stated as page 9
E41 LID NOT SECURED Detailed judgment and handling procedures are stated as page 10
E42 LID OPEN FAILED Detailed judgment and handling procedures are stated as page 11
E51 OVER ROTOR MAX SPEED Detailed judgment and handling procedures are stated as page 12
E52 OVER ROTOR SETTING SPEED Detailed judgment and handling procedures are stated as page 13
E53 SPEED CONTROL FAILED Detailed judgment and handling procedures are stated as page 14
E54 TEMP CONTROL FAILED Detailed judgment and handling procedures are stated as page 15
E55 TEMP OVER 40℃ Detailed judgment and handling procedures are stated as page 16
E56 SPEED SENSOR ERROR Detailed judgment and handling procedures are stated as page 17
E57 TEMP SENSOR ERROR Detailed judgment and handling procedures are stated as page 18
E58 ROTOR ID SENSOR ERROR Detailed judgment and handling procedures are stated as page 19
E60 COMMUNICATION ERROR Detailed judgment and handling procedures are stated as page 20
6. Troubleshooting
placed horizontally and then power switch is turned on or the machine is started, and here it is under power-on operation,
personal protection measures are basically ineffective. Therefore, please confirm the qualification of the operator and
Note 2: The positions of such electronic devices as F1 and F2 have been indicated on the above-mentioned layout
Troubleshooting method:
Electromagnet controlled by the CPU on main board; open signal is transmitted from CPU;
electromagnet action is driven by terminal X3 on main board and lid position is judged by lid
switch.
The “Lid Open Failed” alarm is given arising from lid switch acting improperly within a certain
time after CPU received the door open key signal and outputted the electromagnet control signal. In
such a case, the false alarm arising from external interference shall be executed first and then
consideration is given to the following reasons: electromagnet failing to act, lid switch being not in
the appropriate position, or electromagnet control failed, etc.
Troubleshooting method:
The “OVER ROTOR SETTING SPEED” alarm is similar to the “OVER ROTOR MAX SPEED”
alarm and the overshooting problem needs also to be considered, and their troubleshooting methods
are also similar. Firstly exclude overshooting and interference; then check the circuit.
Check the speed detection circuit of main board related to speed and speed detection board and
replace it if any problem.
Troubleshooting method:
When the machine starts, CPU receives the control signal and then controls converter module to
drive main motor, and at this time, if the speed signal can not be received from speed detection
board, or the speed does not reach the preset value, the “SPEED CONTROL FAILED” alarm will
be given.
So the reason for the alarm shall be judged in the maintenance whether it is caused by external
interference or the fault of speed detection system or motor.
6.11 E54 TEMP CONTROL FAILED
Troubleshooting method:
The “TEMP CONTROL FAILED” alarm is usually given when refrigeration is required. At this
time, CPU receives the temperature control command and then starts the compressor to refrigerate
the chamber; real-time temperature in the chamber is detected by temperature sensor. If it does not
reach the set temperature within 30min, the alarm will be given.
In the maintenance, firstly exclude the false alarm arising from external interference. Secondly
check the temperature probe whether it is loose or the plug is in loose contact; check the running
state of compressor whether it is normal and whether CPU outputs the refrigeration command but
the compressor does not run; check the running condition of each fan whether there is blockage and
dust deposit causing bad heat emission; check the temperature control part of main board, etc. And
then check the set temperature value whether it can not reach the set value because set temperature
is excessively low.
When inspecting the temperature detection circuit on main board, measure the voltage of
TP18testing point to SGND with a multimeter and the voltage value shall be within 3.3±0.1V under
normal temperature.
Troubleshooting method:
The “Temp OVER 40℃” alarm is to protect the separated samples or parts in the chamber from
high temperature damage.
Firstly exclude external interference. Secondly check the temperature probe; check whether
refrigeration measure has been executed in high speed run but the compressor is not started as a
result that the temperature in the chamber is excessively high; check the temperature detection
circuit on main board (the method same as E54) and if any unit is faulty, it shall be replaced.
6.13 E56 SPEED SENSOR ERROR
Troubleshooting method:
The “SPEED SENSOR ERROR” alarm is similar to the “SPEED CONTROL FAILED” alarm.
When the machine starts, CPU receives the control signal and then controls converter module to
drive main motor, and at this time, if the speed signal can not be received from speed and imbalance
detection board, or the voltage value of speed signal is not correct, the “SPEED SENSOR ERROR”
alarm will be given.
Check the speed detection board whether there is a fault by referring to E40.
If the speed detection board or main board is damaged, it shall be replaced.
Troubleshooting method:
The “TEMP SENSOR ERROR” alarm arises from abnormal temperature signal that temperature
sensor transmits through detection by CPU on main board.
Firstly exclude external interference factor; then check the temperature probe whether it is in good
condition and check the connection between temperature probe and main board whether it is normal;
finally check the speed detection circuit whether it is normal.
The temperature detection circuit can be detected by referring to E54.
If temperature probe or main board is faulty, it shall be replaced.
6.15 E58 ROTOR ID SENSOR ERROR
Troubleshooting method:
When the “ROTOR ID SENSOR ERROR” alarm arises, it is because CPU does not detect the rotor
ID signal input by rotor ID board or the voltage value of signal is incorrect.
Firstly exclude external interference factor; then check the connection between rotor ID board and
main board whether it is in loose contact, or whether the rotor ID board has a fault and so it can not
normally output rotor ID signal (detection method of rotor ID board is the same as E40), or the rotor
ID signal processing part on main board is out of order.
If the circuit itself is faulty, circuit board assembly shall be replaced.
Troubleshooting method:
When the “COMMUNICATION ERROR” alarm arises, it is because the communication between
two CPU on main board is poor or signal connection between main board and display screen goes
wrong.
Firstly exclude external interference factor; then check the ten-core fixable flat cable between main
board and display screen whether its connection is good, and whether there is breakage and damage.
In addition, check main board and display screen whether it has a fault; if any fault, replace it.
7. Dismounting and installation of the machine
Maintenance tools:
Big and small cross A set of spanner Sharp nose Converter control
screwdrivers pliers panel
Big and small M4 socket spanner Diagonal pliers Digital thermometer
straight screwdrivers
Tweezers 5mm inner hexagon Knife Tachometer
spanner
Soldering iron and Insulating tape and Level meter Vibroscope
tin solder ties
Attention:
Installation shall be in reverse sequence, unless otherwise specified in the following dismounting
illustrations.
Parts must be replaced with original accessories and maintenance personnel must do the operation
in accordance with the illustrations in this manual and must not change the operation sequence at
will.
When the accessories replaced are reinstalled, all screws shall be coated with thread fastening agent
and then tightened, unless otherwise specified. After each electrical connector is reinserted into
circuit board, it shall be fixed with hot-melt adhesive to keep firm.
Attention:
1. Earthing screws need not to be coated with thread fastening agent to avoid affecting electric
conduction, but spring washer must be installed;
2. The screws used to fasten plastic parts must not be coated with thread fastening agent to
avoid the spilled thread fastening agent causing corrosion to plastic parts.
7.1 Preparatory work
Before the centrifuge is disassembled, the lid shall be opened to take the rotors out. If door lock is
abnormal and it can not be opened by Door Open key, the door can be manually opened through
emergency door-opening device; for details, see Page 6.
Power switch shall be turned off and power wire shall be unplugged.
Condensate in the inner chamber shall be discharged.
Before related maintenance work is conducted, the parts that need to be dismounted and replaced
shall be determined first; related parts shall be dismounted according to actual maintenance need.
7.2 Dismounting of front cover
Remove the two screws (used to fix the front cover) on the front cover.
Raise the front cover to make the clamp disconnected from upper cover, and then move the front
cover forward. Here the circuit relating to main board can be inspected and repaired and replaced.
Unplug the connecting wires between main board and machine body and remove the front cover.
4 5
2 3
6
1
7
10
Unplug the keyboard surface-mounted flat cable inserted into display board.
Remove each mounting screw.
Remove the two mounting screws 1 used to fix the door lock assembly and then move the door lock
assembly out from front side
1
2
3
4
5
Remove the four screws used to fix the fan and then dismount the fan.
1 2 3
Remove the screws used to fix the brake resistor and then dismount and replace the brake resistor.
3
2
4 5
1
2
3
4
5
Remove the three mounting screws used to fix the motor with an inner hexagon spanner and take
the motor out along with the patch panel and speed detection board under motor.
5
4
3
2
1
Remove the screws used to fix inner and outer lids shown as Fig.4 and then dismount outer lid 7
and glass cock 6. Remove the screw 9 in the figure and then dismount and replace inner lid.
Neofuge 13R
P1 P2 P3 6
Esc Menu 7
8
13 12 11 10 9
⑥
①
④ ⑦
⑤
②
①
⑤
① Shell
② Fixing screw hole, total 2 ones
③ PCB, installed in the shell, encapsulated with epoxy resin
④ X2, welding point of leading wire
⑤ U3, Hall sensor SS413A
9. Neofuge 13R explosion diagram and description
Neofuge 13R Explosion Diagram
17
16
15
14 18
13
12
19
20
21
22
11 23
10
24
9
8 25
26
1 2 3 4 5 6 7 27
35 34 33 32 31 30 29 28
Description of Explosion Diagram
Manufacturer
Shanghai Lishen Scientific Equipment Co., Ltd.
Address:
6788 Songze Avenue, Shanghai Qingpu Industrial Zone,
Shanghai 201706, P. R. China