Beckman Coulter Airfuge Centrifuge - Service Manual
Beckman Coulter Airfuge Centrifuge - Service Manual
Airfuge ®
Air-Driven Ultracentrifuge
Instruction Manual
Symbol Simbolo
Symbol
Title / Titel / Titre / Titulo / Titolo / /
Symbole
Símbolo
Dangerous voltage
Gefährliche elektrische Spannung
Courant haute tension
Voltaje peligroso
Pericolo: alta tensione
On (power)
Ein (Netzverbindung)
Marche (mise sous tension)
Encendido
Acceso (sotto tensione)
Off (power)
Aus (Netzverbindung)
Arrêt (mise hors tension)
Apagado
Spento (fuori tensione)
Earth (ground)
Erde
Terre
Tierra
Scarica a terra
! SAFETY NOTICE
This safety notice summarizes information basic to the safe operation of the equipment
described in this manual. The international symbol displayed above is a reminder that all
safety instructions should be read and understood before installation, operation, mainte-
nance, or repair of this instrument. When you see the symbol on other pages, pay special
attention to the safety information presented. Observance of safety precautions will also
help to avoid actions that could damage or adversely affect the performance of the
equipment.
Any servicing of this equipment that requires removal of any covers can expose parts
which involve the risk of electric shock or personal injury. Make sure that the power
switch is off and the ultracentrifuge is disconnected from the main power source, and refer
such servicing to qualified personnel.
Do not replace any ultracentrifuge components with parts not specified for use on this
instrument.
Electrical Safety
To reduce the risk of electrical shock, this instrument uses a three-wire electrical cord and
plug to connect this equipment to earth-ground. To preserve this safety feature:
• Make sure that the matching wall outlet receptacle is properly wired and earth-
grounded. Check that the line voltage agrees with the voltage listed on the name-rating
plate affixed to the instrument.
Do not place containers holding liquid on or near the chamber door. If they spill, liquid
may get into the instrument and damage electrical or mechanical components.
Fuses protect certain electrical circuits within this instrument against overcurrent condi-
tions. For continued protection against the risk of fire, replace only with the same type and
rating specified.
This instrument is not designed for use with materials capable of developing flammable or
explosive vapors. Do not centrifuge such materials (such as chloroform or ethyl alcohol)
in this instrument nor handle or store them within the required 30-cm (1-ft) area
surrounding the ultracentrifuge.
Mechanical Safety
• Use only the rotors and accessories designed for use in this ultracentrifuge.
• Never operate the instrument without a rotor cap or lid installed on the rotor.
• NEVER open the chamber door while the rotor is spinning. The instrument has a
protective window to allow safe observation of rotor motion.
Normal operation may involve the use of solutions and test samples that are pathogenic,
toxic, or radioactive. Such materials should not be used in this ultracentrifuge, however,
unless all necessary safety precautions are taken.
• Observe all cautionary information printed on the original solution containers prior to
their use.
• Handle body fluids with care because they can transmit disease. No known test offers
complete assurance that they are free of micro-organisms. Some of the most virulent—
Hepatitis (B and C) and HIV (I–V) viruses, atypical mycobacteria, and certain systemic
fungi—further emphasize the need for aerosol protection. Handle other infectious
samples according to good laboratory procedures and methods to prevent spread of
disease. Because spills may generate aerosols, observe proper safety precautions for
aerosol containment. Do not run toxic, pathogenic, or radioactive materials in this
instrument without taking appropriate safety precautions. Biosafe containment should
be used when Risk Group II materials (as identified in the World Health Organization
Laboratory Biosafety Manual) are handled; materials of a higher group require more
than one level of protection.
• Dispose of all waste solutions according to appropriate environmental health and safety
guidelines.
Airfuge®
Air-Driven Ultracentrifuge
Instruction Manual
Page
INTRODUCTION
Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Scope of Manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . viii
Notes, Cautions, and Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . viii
CFC-Free Centrifugation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ix
Radio Interference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ix
Canadian Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . x
Recycling Label . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . x
SECTION 1: DESCRIPTION
Functional Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Safety Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Controls and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Pressure Regulator Knob . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Pressure Gauge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Run Light . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Time-Delay Knob . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Brake Tension Screw. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Levitation Air Screw . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Chamber Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Stator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Stator Pad. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Brake Pin . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Door Latch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Name Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Available Rotors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
iii
Contents
Page
SECTION 2: INSTALLATION
Space and Location Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Electrical Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Installation Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Filter Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Time-Delay Setting Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Leveling the Airfuge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Installing the Brake Pin and Stator Pad . . . . . . . . . . . . . . . . . . . . . 2-5
Trial Run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Cold Room Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
SECTION 3: OPERATION
Starting the Run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Ending the Run. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
SECTION 4: TROUBLESHOOTING
SECTION 5: MAINTENANCE
Instrument Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
General Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Instrument Interior and Exterior . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Stator Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Lubrication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Filter Element Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Brake Spring Wire Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Brake Spring Wire Tension Adjustment . . . . . . . . . . . . . . . . . . . . . . . 5-7
Levitation Air Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Time-Delay Period Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Decontamination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
Sterilization and Disinfection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
Storage and Transportation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Supply List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Warranty
iv
Illustrations and Tables
Page
v
Introduction
CERTIFICATION
To ensure full system quality, Beckman Coulter Airfuge® ultracentri-
fuges are manufactured in a registered ISO 9001 or 13485 facility.
They have been designed and tested to be compliant (when used with
Beckman Coulter rotors) with the laboratory equipment requirements
of applicable regulatory agencies. Declarations of conformity and
certificates of compliance are available at www.beckmancoulter.com.
SCOPE OF MANUAL
This manual is designed to familiarize you with the Airfuge ultracen-
trifuge, its functions, specifications, operation, and routine operator
care and maintenance. We recommend that you read this entire
manual, especially the SAFETY NOTICE and all safety-related infor-
mation, before operating the ultracentrifuge or performing instrument
maintenance.
vii
Introduction
➠ NOTE
If the ultracentrifuge is used in a manner other
than specified in this manual, the safety and
performance of this equipment could be
impaired. Further, the use of any equipment
other than that recommended by Beckman
Coulter has not been evaluated for safety. Use of
any equipment not specifically recommended in
this manual and/or the appropriate rotor manual
is the sole responsibility of the user.
CONVENTIONS
Certain symbols are used in this manual to call out safety-related and
other important information. These international symbols may also be
displayed on the ultracentrifuge and are reproduced and described
below and on the inside of the front cover.
➠ NOTE
Used to call attention to important information
that should be followed during installation, use,
or servicing of this equipment.
! CAUTION
Used to indicate a potentially hazardous situa-
tion which, if not avoided, may result in minor or
moderate injury and/or mechanical damage. It
is also used to alert against unsafe practices.
viii
Introduction
! WARNING
Used whenever an action or condition may
potentially cause personal injury or loss of
life. Mechanical damage may also result.
WARNING
Indicates high voltage or risk of electric
shock. Refer servicing of all areas displaying
either symbol to service personnel.
CFC-FREE CENTRIFUGATION
To ensure minimal environmental impact, no CFCs are used in the
manufacture or operation of Airfuge ultracentrifuges.
CFC
RADIO INTERFERENCE
This equipment has been tested and found to comply with the limits
for a Class A digital device, pursuant to Part 15 of FCC Rules. These
limits are designed to provide reasonable protection against harmful
interference when the equipment is operated in a commercial environ-
ment. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with
this instruction manual, may cause interference to radio communica-
tions. Operation of this equipment in a residential area is likely to
cause interference, in which case the user will be required to correct
the interference at his own expense.
ix
Introduction
CANADIAN REGULATIONS
This equipment does not exceed the Class A limits for radio noise
emissions from digital apparatus as set out in the radio interference
regulations of the Canadian Department of Communications.
RECYCLING LABEL
This symbol is required in accordance with the Waste Electrical and
Electronic Equipment (WEEE) Directive of the European Union. The
presence of this marking on the product indicates:
A28219-AA
1) the device was put on the European market after August 13, 2005
and
Note: On the instrument, the
triangle background is yellow 2) the device is not to be disposed via the municipal waste collection
rather than gray. system of any member state of the European Union.
x
1
Description
FUNCTIONAL DESCRIPTION
The Beckman Coulter Airfuge® (see Figure 1-1) is a benchtop
air-driven ultracentrifuge capable of accelerating rotors up to
110 000 rpm (revolutions per minute) in as little as 30 seconds. The
instrument uses no vacuum or high-speed bearings; the rotor is
supported and turned by streams of air. Rotor speed can be deter-
mined by conversion of the applied air pressure, as displayed on the
PRESSURE gauge, to rpm.1
1 This conversion is different for each type of rotor (see the graph in the applicable rotor manual).
1-1
Description
SAFETY FEATURES
The Airfuge ultracentrifuge has been designed and tested to operate
safely indoors at altitudes up to 2 000 m (6 562 ft).
The pressure regulator knob controls the driving air pressure. Pressing
down on the knob and turning to the right (clockwise) closes the door.
Turning the knob further brings up the air pressure and accelerates the
rotor to speed.
PRESSURE GAUGE
RUN LIGHT
The run light comes on when the timer is set and remains lit until the
braking system is engaged.
TIME-DELAY KNOB
This knob, located on the ultracentrifuge bottom panel (see Figure 1-2),
is used to set a time-delay period that allows the rotor to coast after
the timer reaches zero at the end of a run. The run light goes out and
a brake pin is released at the end of the set delay period. A setting of
120 seconds on the knob (or about 3.5 on older instruments) represents
the 2-minute delay period.
1-2
Description
The brake spring wire (see Figure 1-3) forces the brake pin up into
the rotor bottom during braking. Occasional adjustment of the spring
tension, using the BRAKE TENSION adjusting screw located on the
back panel (see Figure 1-4), may be necessary.
When the timer reaches zero at the end of a run, the driving air is
turned off and levitation air supports the rotor during its deceleration.
If the flow of levitation air is too high and the rotor fails to stop
completely, or too low and the rotor stops too fast, adjustment of the
LEVITATION AIR screw located on the rear panel may be necessary.
1-3
Description
Instrument
Door Protective Window
Stator Turbine
Pad Flutes
Stator
Driving
Levitation
Air Passage
Air Passage
1-4
Description
CHAMBER COMPONENTS
STATOR
The brass stator (see Figure 1-3) contains the driving air jets
that control the rotor speed and the air jets that levitate the rotor
during deceleration. Levitation air operates automatically during
deceleration.
STATOR PAD
The black stator pad (see Figure 1-5) rests on the channel of the stator
and cushions the rotor during a run and when the rotor is at rest.
1-5
Description
BRAKE PIN
DOOR LATCH
FILTER
A filter is supplied with the ultracentrifuge for water and oil removal
from the air supply. The filter element turns red when it is saturated
and needs to be replaced.
1-6
Description
SPECIFICATIONS
Only values with tolerances or limits are guaranteed data. Values
without tolerances are informative data, without guarantee.
* Relative Centrifugal Field (RCF) is the ratio of the centrifugal acceleration at a specified radius
and speed (rω 2) to the standard acceleration of gravity (g) according to the following formula:
rω 2
RCF = --------
-
g
where r is the radius in millimeters, ω is the angular velocity in radians per second
(2 π RPM /60), and g is the standard acceleration of gravity (9807 mm/s2). After substitution:
RPM 2
RCF = 1.12 r ⎛⎝ ------------⎞⎠
1000
† Instrument operation at gauge pressures below 20 psig (138 kPa) may cause rotor instability.
‡ Normally only nonconductive pollution occurs; occasionally, however, a temporary conduc-
tivity caused by condensation must be expected.
1-7
Description
AVAILABLE ROTORS
Airfuge rotors are made of anodized aluminum and have turbine
flutes on the bottom that provide the driving surface for the jets of air.
Pressurized air impinges on the rotor bottom and lifts and turns the
rotor. A white plastic bushing fitted in the rotor bottom is engaged by
the brake pin during braking.
• The A-110 fixed angle rotor holds up to six 180-µL tubes at a tube
angle of 18 degrees from the axis of rotation, and is ideal for the
rapid separation of small sample volumes in a very short time.
Maximum operating speed is 110 000 rpm at 30 psig (207 kPa) air
pressure.
• The A-95 fixed angle rotor holds up to four 450-μL tubes at a tube
angle of 30 degrees from the axis of rotation, and is ideal for the
rapid separation of small sample volumes in a very short time or
to isolate microsomal fractions and plasma membranes. It can also
be used to clarify small volumes of lipemic serum by flotation of
the chylomicrons. Maximum operating speed is 95 000 rpm at
30 psig (207 kPa) air pressure.
• The Batch rotor can be used to pellet particles from a single large
sample. Sample volumes of 2 to 7 mL can be run. The rotor is
rated for 90 000 rpm at 30 psig (207 kPa) air pressure.
1-8
2
Installation
! WARNING
Do not place the ultracentrifuge near areas
containing flammable reagents or combus-
tible fluids. Vapors from these materials
could enter the ultracentrifuge air system and
be ignited by the motor. Maintain a 30-cm
(1-ft) clearance envelope around the ultracen-
trifuge while it is running. No persons or any
hazardous materials should be within this
clearance boundary while the ultracentrifuge
is operating except to change operating
controls, if required.
2-1
Preinstallation Requirements
ELECTRICAL REQUIREMENTS
60-Hz instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120 VAC
50-Hz instrument . . . . . . . . . . . . . . . . . . . . . . . 100 or 200–240 VAC
INSTALLATION PROCEDURES
Installation of the ultracentrifuge requires the following steps—in
order:
2-2
Preinstallation Requirements
FILTER INSTALLATION
➠ NOTE
If the air source is especially wet or oily, install a
second filter as a prefilter. A second filter will
cause an additional pressure drop of approxi-
mately 5 psig (35 kPa) in the air pressure that
reaches the instrument.
➠ NOTE
Do not install the filter upside down. The filter
bottom has a valve that distinguishes it from the
top. Escaping air can be heard if the filter is
Filter Bottom installed upside down.
Arrow indicates
direction of airflow Arrow indicates
Remove barb fitting
direction of airflow
before installing 1/4"
piping directly into filter
Mounting
Screws (4)
2-3
Preinstallation Requirements
2. If non-rigid piping leads from the air source, mount the filter on a
ring stand or on the wall. Clamp the filter to a ring stand, or use the
supplied mounting brackets (see Figure 2-1) to mount the filter
to the wall. Attach the brackets to the assembled filter with the
mounting screws, then attach the brackets to the wall. The air flow
arrow must point toward the ultracentrifuge. Add hose clamps to
the plain high-pressure hose. Clamp one end of the hose to the
barb fitting of the filter inlet and one end to the air source.
If rigid piping (standard 1/4-in. pipe thread) leads from the air
source, the filter can be free-standing. Remove the barb fitting
from the filter inlet to the air source. The air flow arrow must point
toward the ultracentrifuge.
If the air source does not supply sufficient air pressure to close the
filter’s automatic drain valve you will hear air escaping from the
filter. Routine operation at 30 psig (207 kPa) on the instrument
PRESSURE gauge requires 290 kPa (42 psig) air pressure supplied to
the filter.
2. Adjust the instrument front feet until the bubble of the level is
centered.
2-4
Preinstallation Requirements
1. Place the brake pin in the hole at the center of the stator (see
Figure 2-2). Make sure that the slot in the pin bottom straddles
the brake spring wire as shown in the figure.
3. Install the black ring-shaped stator pad in the channel of the stator.
Brake Pin
Stator Pad
Brake
Spring Wire
WRONG
Figure 2-2. Aligning the Brake Pin in the Stator. The slot on the end of the pin straddles the brake spring wire.
The broken line shows the wire direction (aligned with the screws). If the tip of the pin slot stands
on the wire instead of straddling it, the pin will be too high above the stator.
2-5
Preinstallation Requirements
TRIAL RUN
1. After the rotor runs for a few minutes, turn the TIME dial to zero
and begin timing. The run light should go out approximately
2 minutes after the TIME dial is turned off.
➠ NOTE
If a fixed angle rotor stops in less than
45 seconds, sample stirring may result; adjust
the brake or levitation air to provide a longer
stopping time. If a rotor takes too long to stop,
the brake, pin tension, or levitation air needs
adjusting. See Section 5 for adjustment
procedures.
2-6
Preinstallation Requirements
Compressor or
Pressure
Laboratory Air Supply
Gauge
Compressor Capacity
3.5 to 4 cfm at 47
psig (330 kPa)
Oil Removal
Filter
Regulator
(Norgren Part
No. 339620)
COLD ROOM
Water Removal Filter
(Norgren F11-200A3T)
0.6 m
(2 ft)
or less
3 m (10 ft) of
3/8 in. ID finned Airfuge Ultracentrifuge
copper tubing
2-7
3
Operation
! WARNING
Normal operation may involve the use of solu-
tions and test samples that are pathogenic,
toxic, or radioactive. Handle body fluids
with care because they can transmit disease.
No known test offers complete assurance that
they are free of micro-organisms. Some of the
most virulent—Hepatitis (B and C) and HIV
(I–V) viruses, atypical mycobacteria, and
certain systemic fungi—further emphasize
the need for aerosol protection. Handle other
infectious samples according to good labora-
tory procedures and methods to prevent
spread of disease. Because spills may generate
aerosols, observe proper safety precautions
for aerosol containment. Do not run toxic,
pathogenic, or radioactive materials in this
ultracentrifuge without taking appropriate
safety precautions. Biosafe containment
should be used when Risk Group II materials
(as identified in the World Health Organiza-
tion Laboratory Biosafety Manual) are
handled; materials of a higher group require
more than one level of protection.
3-1
Operation
! WARNING
Do not use the ultracentrifuge in the vicinity
of flammable liquids or vapors, and do not
run such materials in the instrument. Do not
lean on the instrument or place items on it
while it is operating.
! CAUTION
Never operate the ultracentrifuge without a
rotor cap or lid installed on the rotor.
2. Load the rotor and secure the rotor lid or cap as described in the
applicable rotor manual. Tubes should be symmetrically loaded in
a fixed angle rotor; rotor liners should be full and evenly seated in
their rotor base.
➠ NOTE
If you are using an Airfuge with the digital
speed readout accessory installed, use a flat
black marker to color half of the rotor lid or cap
flat black. If the black surface is shiny, inaccu-
rate speed readings will be displayed.
➠ NOTE
Stator pads—especially the white circles on the
bottom—should be kept clean with alcohol.
3-2
Operation
Worn
Stator Pads
New
Stator Pad
5. Place the rotor on the stator pad and shut the instrument door.
6. Set the TIME dial for the required run time. The dial indicates up
to 5 hours of run time in 5-minute increments. (In some models,
the dial indicates 5 minutes of run time in 5-second increments.)
For longer runs, set the timer to HOLD. For runs shorter than
30 minutes, turn the dial past the 30-minute point, then back to the
required time. (ACR-90 rotor—set the run time for 10 minutes to
prevent significant evaporative loss.)
3-3
Operation
! CAUTION
If the rotor makes an unusual noise, turn the
timer to zero. When the rotor stops, open the
chamber door and remove the rotor. Check to
make sure the rotor is loaded correctly. Correct
the problem before restarting the run (see
Table 4-1).
➠ NOTE
If the coasting time is too short—45 seconds or
less—check the time-delay relay setting and/or
adjust the flow of levitation air as described in
Section 5.
2. After the rotor has stopped, turn the pressure regulator knob to
the left (counterclockwise) until the PRESSURE gauge reading
is zero.
3. Open the chamber door. Refer to the applicable rotor manual for
instructions on unloading and cleaning the rotor.
➠ NOTE
Do not tighten the pressure regulator knob
between runs.
3-4
4
Troubleshooting
➠ NOTE
It is your responsibility to decontaminate the
instrument, as well as any rotors and/or accesso-
ries, before requesting service by Beckman
Coulter Field Service representatives.
• Note the operating situation when the error occurred (rotor in use,
speed, load type, and so forth).
4-1
Troubleshooting
Table 4-1. Troubleshooting. If the problem persists call Beckman Coulter Field Service.
Brake pin falls flat on stator Brake spring wire not centered Adjust the brake spring wire.
surface
Brake spring tension too low Adjust the brake spring tension.
Brake pin not actuating Brake spring tension too low Adjust the brake spring tension.
Rotor will not spin Air supply or power disconnected Check air and power connections.
Rotor runs erratically or Stator pad damaged Replace the stator pad.
makes unusual noise
Dirty air jets Clean the stator assembly.
Rotor fails to stop Brake pin shaft oily Clean the pin and shaft.
Brake spring tension too low Adjust the brake spring tension.
Rotor stops too fast Levitation air flow too low Adjust levitation air.
Brake pin shaft oily Clean the stator assembly and shaft.
Oil visibly collecting in Filter element saturated Replace the filter element.
instrument
Filter element turns red Filter element saturated Replace the filter element.
Pressure at instrument will Filter element saturated Replace the filter element.
no longer read 30 psig
Continued —
4-2
Troubleshooting
Run light fails to come on Light burned out Replace the lamp.
! CAUTION
Repeated tripping of the circuit breaker may cause
instrument damage.
4-3
5
Maintenance
INSTRUMENT CARE
➠ NOTE
It is your responsibility to decontaminate the
instrument, as well as any rotors and/or accesso-
ries, before requesting service by Beckman
Coulter Field Service representatives.
WARNING
Any maintenance procedure requiring
removal of a panel exposes the operator to
the possibility of electrical shock and/or
mechanical injury. Therefore, turn the power
OFF and disconnect the instrument from the
main power source, and refer such mainte-
nance to service personnel.
5-1
Maintenance
GENERAL MAINTENANCE
Perform the following procedures regularly to ensure satisfactory
performance and long service life of the ultracentrifuge.
• Regularly check the stator pad for ease of rotation. Rotate it on the
stator, using light finger pressure, to ensure it moves easily. Also
check for dirt or oil; clean with alcohol* (especially the white
circles on the bottom).
• Regularly check the stator pad for wear; a worn pad is smooth and
shiny on the upper half of the inside surface. A shiny pad has lost
its cushioning effect and will eventually cause rotor failure.
Replace a worn pad.
➠ NOTE
Proper deceleration of the rotor (that is, allowing
the rotor to come to a complete stop before
opening the door) will increase pad life.
CLEANING
➠ NOTE
Before using any cleaning or decontamination
methods except those recommended by the
manufacturer, users should check with the
manufacturer that the proposed method will not
damage the equipment.
Frequent cleaning will ensure proper operation and prolong the life
of the ultracentrifuge. Always clean up spills when they occur to
prevent corrosives or contaminants from drying on component
surfaces. (Be careful not to spill liquid on the instrument where
electrical or mechanical components could become damaged.)
Refer to Chemical Resistances (publication IN-175) for chemical
compatibilities of cleaning materials.
5-2
Maintenance
STATOR ASSEMBLY
Clean the stator assembly when the filter element is replaced, or if the
rotor runs erratically.
! WARNING
Disconnect the power cord and the air supply.
2. Remove the stator pad and the brake pin. (Lift the pin out with a
forceps.)
3. Unscrew the two stator screws, then pry the stator off, using the
tools supplied. See Figures 5-1 and 5-2. Note that the slots on the
stator outside rim are used to provide a leverage point for the
stator tools.
5. Clean the stator, including the air jets and shaft, with acetone.*
5-3
Maintenance
7. Clean the brake pin and install it in the shaft at the center of the
stator. The pin should move freely up and down in the stator
shaft.
8. To reassemble the stator and pin, align the screw holes then lower
the stator and pin into place. The slot on the end of the stator
must straddle the brake spring in the hole as shown in Figure 5-4.
(The wire direction is indicated by an imaginary line between the
stator screws.) Press the stator down firmly. The brake pin will
bounce up about 3 mm (1/8 in.) above the stator. Lightly touch the
pin to test its motion.
12. To reconnect the air supply, pull back the spring-loaded sleeve of
the quick-disconnect hose and push the sleeve over the fitting at
the rear of the instrument.
LUBRICATION
• Periodically lubricate the threads of the pressure regulator knob
with Spinkote lubricant (306812).
• Lightly coat the three stator O-rings with silicone vacuum grease
(335148) when removed for stator cleaning.
5-4
Maintenance
Figure 5-1. Prying Up the Stator Figure 5-2. Lifting Out the Stator.
Keep the stator level to prevent binding.
Brake Pin
Stator Screw
Stator Screw
Stator
Brake
Spring Wire
Stator Slot
Brake Spring
Tension Adjust
Figure 5-3. Stator O-Ring Positions Figure 5-4. Reassembling the Stator and Pin
5-5
Maintenance
! WARNING
Disconnect the power cord and the air supply.
1. Remove the stator pad and brake pin. Lift the pin out with a
forceps.
3. Unscrew the two stator screws, then pry the stator off, using the
tools supplied. See Figures 5-1 and 5-2. Note that the slots on
the stator outside rim are used to provide a leverage point for the
stator tools.
WARNING
Any maintenance procedure requiring
removal of a panel exposes the operator to
the possibility of electrical shock and/or
mechanical injury. Therefore, disconnect the
power cord and the air supply, and refer such
maintenance to trained and qualified service
personnel.
a. Ensure that the power cord and air supply are disconnected,
then turn the instrument on its side and remove the bottom
plate.
5-6
Maintenance
1. Remove the brake pin and clean it with alcohol, then reinstall it.
5-7
Maintenance
➠ NOTE
Refer to publication AF-TB-013 for levitation
air adjustment procedures on an instrument
equipped with a digital speed readout accessory.
1. Remove the stator pad and clean it with alcohol,* especially the
white circles on the bottom. Place it on the channel of the stator.
2. Turn the timer on, then back to zero. This will engage the braking
system and hold the brake pin down for 2 minutes while you
complete this adjustment.
3. Place a rotor on the stator pad. Be sure rotor caps or lids are
properly installed. (The rotor does not need to be loaded.)
4. Close the Airfuge door and turn the pressure regulator knob on
the door clockwise until the PRESSURE gauge indicates 30 psig
(207 kPa). The rotor will wobble.
5-8
Maintenance
At the end of the 2-minute period, the run light will go out and a
clicking sound can be heard when the brake pin is released.
! WARNING
Disconnect the power cord and the air supply.
5-9
Maintenance
1. Close the Airfuge door and turn the instrument on its side.
3. Reconnect the power cord and the air supply and perform a trial
run to check the new setting.
DECONTAMINATION
If the instrument and/or accessories are contaminated with radioactive
or pathogenic solutions, follow appropriate decontamination
procedures as determined by your laboratory safety officer. Refer
to Chemical Resistances (publication IN-175) or contact Beckman
Coulter Field Service to ensure that the decontamination method
does not damage any part of the instrument (or accessories).
While Beckman Coulter has tested these methods and found that they
do not damage the instrument, no guarantee of sterility or disinfection
is expressed or implied. When sterilization or disinfection is a
concern, consult your laboratory safety officer regarding proper
methods to use.
5-10
Maintenance
SUPPLY LIST
➠ NOTE
Publications referenced in this manual can be
obtained by calling Beckman Coulter at 1-800-
742-2345 in the United States, or by contacting
your local Beckman Coulter office.
SUPPLIES
5-11
WARRANTY FOR THE AIRFUGE® ULTRACENTRIFUGE
DISCLAIMER
REPLACEMENT
IT IS EXPRESSLY AGREED THAT THE ABOVE WAR-
Any product claimed to be defective must, if requested by RANTY SHALL BE IN LIEU OF ALL WARRANTIES OF
Beckman Coulter, be returned to the factory, transportation FITNESS AND OF THE WARRANTY OF MERCHANT-
charges prepaid, and will be returned to Buyer with the trans- ABILITY AND THAT BECKMAN COULTER, INC.,
portation charges collect unless the product is found to be SHALL HAVE NO LIABILITY FOR SPECIAL OR CONSE-
defective, in which case Beckman Coulter will pay all transpor- QUENTIAL DAMAGES OF ANY KIND WHATSOEVER
tation charges. ARISING OUT OF THE MANUFACTURE, USE, SALE,
HANDLING, REPAIR, MAINTENANCE, OR REPLACE-
MENT OF THE PRODUCT.
Beckman Coulter, Inc. • 250 S. Kraemer Blvd. • Brea, California 92821
Sales and Service: 1-800-742-2345 • Internet: www.beckmancoulter.com