N2
DESIGN AND PRODUCTION
OF LABORATORY
GAS GENERATORS
ZEFIRO series
NITROGEN GENERATORS
WITH INTEGRATED COMPRESSOR
The constant and completely
autonomous supply of nitrogen flux
Nitrogen for LC-MS
application
ZEFIRO LC/MS 20-25-35-40
ZEFIRO Nitrogen Generator is designed specifically as a stand-alone
system to provide nitrogen to single LC/MS applications which require a
constant and autonomous supply of nitrogen flux.
The Nitrogen generator uses pressure swing adsorption technology (PSA)
to produce nitrogen gas. This system uses carbon molecular sieve (CMS)
which selectively adsorbs oxygen and water vapor molecules under high Application
pressure, while allowing Nitrogen to pass through. The alternation between
LC/MS
purification and regeneration of the sieve bed pressure Vessels ensure
continuous nitrogen flow production.
Technical data
ZEFIRO LC/MS 20-25 ZEFIRO LC/MS 35-40
STANDARD NITROGEN FLOW RATE 20 l/min at STP 25 l/min at STP 35 l/min at STP 40 l/min at STP
OUTLET PRESSURE up to 8 bar (116 psi) up to 8 bar (116 psi)
STANDARD PURITY 99,0% at STP 99,0% at STP
INPUT VOLTAGE 230 V - 50 or 60 Hz 230 V - 50 or 60 Hz
WEIGHT 140 Kg 140 Kg
POWER CONSUMPTION 1800 W 1800 W
PRESSURE ACCURACY 0.1 bar (± 0.5 %) 0.1 bar (± 0.5 %)
MICROPROCESSOR CONTROLLED DISPLAY Graphic display, 128 x 64 px Graphic display, 128 x 64 px
INDEX OF PROTECTION IP20 IP20
TEMPERATURE 5°C to +30°C 5°C to +30°C
RELATIVE HUMIDITY 0-80%, non condensing 0-80%, non condensing
SOUND PRESSURE LEVEL 50 dB(A) 50 dB(A)
OUTPUT PORT G 1/4 G 1/4
width 48 cm, height 70 cm, width 48 cm, height
CASE DIMENSIONS length 84 cm 70 cm, length 84 cm
Nitrogen for LC-MS
application
ZEFIRO LC/MS 65
ZEFIRO Nitrogen Generator is designed specifically as a stand-alone
system to provide nitrogen flux to single LC/MS applications which require
a high nitrogen consumption or allows two standard LC/MS to be supplied
contemporaneously.
The Nitrogen generator uses membrane Technology to produce nitrogen
gas at specific purity required. Application
LC/MS
Technical data
STANDARD NITROGEN FLOW RATE up to 65 l/min at STP
OUTLET PRESSURE up to 7.0 bar (102 psi)
STANDARD PURITY 98,5% at STP
INPUT VOLTAGE 230 V - 50 or 60 Hz
WEIGHT 120 Kg
POWER CONSUMPTION 1900 W
PRESSURE ACCURACY 0.1 bar (± 0.5 %)
MICROPROCESSOR CONTROLLED DISPLAY Graphic display, 128 x 64 px
INDEX OF PROTECTION IP20
TEMPERATURE 5°C to +30°C
RELATIVE HUMIDITY 0-80%, non condensing
SOUND PRESSURE LEVEL 55 dB(A)
OUTPUT PORT G1/4
width 48 cm, height
CASE DIMENSIONS 70 cm, length 84 cm
STP: Standard Temperature and Pressure (20°C, 1 bar)
Nitrogen for LC-MS
application
ZEFIRO LC/MS DUAL-PURITY
ZEFIRO Nitrogen Generator LC/MS DUAL PURITY is specifically designed to
meet the gas flow, purity and pressure requirements of AGILENT LCMS
instruments.
The system supplies simultaneously Nitrogen for LC/MS & ultra high purity
nitrogen for the collision cell.
Application
AGILENT
LC/MS
Technical data
ZEFIRO LC/MS DUAL PURITY ZEFIRO LC/MS DUAL PURITY S
LC/MS N2: 35 or 65 l/min at STP LC/MS N2: up to 35 l/min at STP
STANDARD NITROGEN FLOW RATE Collision Cell N2: 200 ml/min at STP Collision Cell N2: 1100 ml/min at STP
7 bar at STP (102 psi) 7,2 bar at STP (105 psi)
OUTLET PRESSURE 2,5 bar at STP (36 psi) 6,5 bar at STP (95 psi)
99.0% at STP 98.0% at STP
STANDARD PURITY 99,999% at STP 99,999% at STP
INPUT VOLTAGE 230 V - 50 or 60 Hz 230 V - 50 or 60 Hz
WEIGHT 130 Kg 125 Kg
POWER CONSUMPTION 1800 or 1900 W 1900 W
PRESSURE ACCURACY 0.1 bar (± 0.5 %) 0.1 bar (± 0.5 %)
MICROPROCESSOR CONTROLLED DISPLAY Graphic display, 128 x 64 px Graphic display, 128 x 64 px
INDEX OF PROTECTION IP20 IP20
TEMPERATURE 5°C to +30°C 5°C to +30°C
RELATIVE HUMIDITY 0-80%, non condensing 0-80%, non condensing
SOUND PRESSURE LEVEL 50 or 55 dB(A) 55 dB(A)
G 1/4 + quick connection for 1/4’’ LC/MS G 1/4 + quick connection for 1/4’’ LC/MS
OUTPUT PORT G 1/4 + 1/8’’ swagelock for Collision Cell G 1/4 + 1/8’’ swagelock for Collision Cell
width 48 cm, height 70 cm, width 48 cm, height 70 cm,
CASE DIMENSIONS length 84 cm length 84 cm
STP: Standard Temperature and Pressure (20°C, 1 bar)
Nitrogen and Air
for LC-MS application
ZEFIRO COMBINED
ZEFIRO COMBINED Nitrogen-Air Generator is specifically designed to
supply Nitrogen & Dry Air for all ABI SCIEX LC/MS.
The Nitrogen generator uses membrane Technology to produce nitrogen
gas at specific purity required.
The internal oil free air compressor which is firstly pre filtered provide
also to the exhaust air flow which is treated by a drying filter.
Application
ABI SCIEX LC/MS
Technical data
ZEFIRO COMBINED ZEFIRO COMBINED S
STANDARD NITROGEN FLOW RATE up to 15 l/min at STP* up to 20 l/min at STP*
STANDARD DRY AIR FLOW RATE up to 35 l/min at STP* up to 55 l/min at STP*
OUTLET NITROGEN PRESSURE 4,2 bar (60 psi) at STP* 4,2 bar (60 psi) at STP*
OUTLET DRY AIR PRESSURE up to 7,2 bar (105 psi) at STP* up to 7,2 bar (105 psi) at STP*
up to 99,0% at STP* up to 99,0% at STP*
STANDARD NITROGEN PURITY
INPUT VOLTAGE 230 V – 50/60 Hz 230 V – 50/60 Hz
WEIGHT 130 Kg 130 Kg
POWER CONSUMPTION 1800 W 1900 W
PRESSURE ACCURACY 0.1 bar (± 0.5 %) 0.1 bar (± 0.5 %)
MICROPROCESSOR CONTROLLED DISPLAY Graphic display, 128 x 64 px Graphic display, 128 x 64 px
INDEX OF PROTECTION IP20 IP20
TEMPERATURE 5°C to +30°C 5°C to +30°C
RELATIVE HUMIDITY 0-80%, non condensing 0-80%, non condensing
SOUND PRESSURE LEVEL 55 dB(A) 55 dB(A)
OUTPUT PORT N° 02 G 1/4 N° 02 G 1/4
width 48 cm, height 70 cm, width 48 cm, height 70 cm,
CASE DIMENSIONS length 84 cm length 84 cm
Nitrogen and Air
for LC-MS application
ZEFIRO Q
ZEFIRO Q Nitrogen-Air Generator is specifically designed to supply Nitrogen &
Dry Air for Bruker EVOQ Triple Quadrupole Mass Spectrometer. The Nitrogen Application
generator uses membrane Technology to produce nitrogen gas at specific
purity required.
The internal oil free air compressor which is firstly pre filtered provide also Bruker LC/MS
to the exhaust air flow which is treated by a drying filter.
ZEFIRO QS
ZEFIRO QS Nitrogen-Air Generator is specifically designed to supply Nitrogen Application
& Dry Air for Perkin Elmer QSight triple quad LC/MS/MS systems. The Nitrogen
generator uses membrane Technology to produce nitrogen gas at specific
purity required. Perkin Elmer LC/MS
The internal oil free air compressor which is firstly pre filtered provide also
to the exhaust air flow which is treated by a drying filter.
Technical data
ZEFIRO Q ZEFIRO QS
STANDARD NITROGEN FLOW RATE up to 32 l/min at STP* up to 18 l/min at STP*
STANDARD DRY AIR FLOW RATE up to 50 l/min at STP* up to 70 l/min at STP*
OUTLET NITROGEN PRESSURE 5,5 bar (80 psi) at STP* 5,5 bar (80 psi) at STP*
OUTLET DRY AIR PRESSURE 5,5 bar (80 psi) at STP* up to 7,2 bar (105 psi) at STP*
STANDARD NITROGEN PURITY up to 99,0% at STP* up to 99,0% at STP*
INPUT VOLTAGE 230 V – 50/60 Hz 230 V – 50/60 Hz
WEIGHT 130 Kg 130 Kg
POWER CONSUMPTION 1800 or 1900 W 1900 W
PRESSURE ACCURACY 0.1 bar (± 0.5 %) 0.1 bar (± 0.5 %)
MICROPROCESSOR CONTROLLED DISPLAY Graphic display, 128 x 64 px Graphic display, 128 x 64 px
INDEX OF PROTECTION IP20 IP20
TEMPERATURE 5°C to +30°C 5°C to +30°C
RELATIVE HUMIDITY 0-80%, non condensing 0-80%, non condensing
SOUND PRESSURE LEVEL 55 dB(A) 55 dB(A)
OUTPUT PORT N° 02 G 1/4 N° 02 G 1/4
width 48 cm, height 70 cm, width 48 cm, height 70 cm,
CASE DIMENSIONS length 84 cm length 84 cm
STP: Standard Temperature and Pressure (20°C, 1 bar)
Nitrogen and Air
for LC-MS application
ZEFIRO 8050
ZEFIRO 8050 Nitrogen-Air Generator is specifically designed to supply Nitrogen
& Dry Air for Shimadzu LC/MS 8045, 8050, 8060, 9030.
The Nitrogen generator uses membrane Technology to produce nitrogen gas at
specific purity required.
The internal oil free air compressor which is firstly pre filtered provide also
to the exhaust air flow which is treated by a drying filter. Application
Shimadzu LC/MS
Technical data
STANDARD NITROGEN FLOW RATE up to 30 l/min at STP*
STANDARD DRY AIR FLOW RATE up to 25 l/min at STP*
SET NITROGEN PRESSURE up to 7,2 bar (105 psi) at STP*
OUTLET DRY AIR PRESSURE up to 7,2 bar (105 psi) at STP*
STANDARD NITROGEN PURITY up to 99,0% at STP*
INPUT VOLTAGE 230 V – 50/60 Hz
WEIGHT 130 Kg
POWER CONSUMPTION 1900 W
PRESSURE ACCURACY 0.1 bar (± 0.5 %)
MICROPROCESSOR CONTROLLED DISPLAY Graphic display, 128 x 64 px
INDEX OF PROTECTION IP20
TEMPERATURE 5°C to +30°C
RELATIVE HUMIDITY 0-80%, non condensing
SOUND PRESSURE LEVEL 55 dB(A)
OUTPUT PORT N° 02 G 1/4
width 48 cm, height 70 cm,
CASE DIMENSIONS length 84 cm
High Purity
Nitrogen
ZEFIRO 5 HP
& mini-ZEFIRO 1 HP
The model HP (High Purity) Nitrogen Generator is designed specifically for
use as make up and carrier gas for GC Applications, which require ultra
high purity nitrogen for operation.
The HP nitrogen generator series is also suitable for ICP.
The Nitrogen generator use pressure swing adsorption technology (PSA)
combined with a pressure management in two steps (DP PSA) to produce Application
pure nitrogen gas with ultra high purity.
Carrier gas for GC
ICP
Technical data
mini-ZEFIRO 1 HP ZEFIRO 5 HP
STANDARD NITROGEN 1 l/min at STP 5 l/min at STP
FLOW RATE
OUTLET PRESSURE 5 bar (72.5 psi) up to 6 bar (86 psi)
STANDARD PURITY 99.999% (O2 < 10 ppm) at STP 99.999% (O2 < 10 ppm) at STP
INPUT VOLTAGE 230 V - 50 or 60 Hz 230 V - 50 or 60 Hz
WEIGHT 60 Kg 140 Kg
POWER CONSUMPTION 600 W 1800 W
PRESSURE ACCURACY 0.1 bar (± 0.5 %) 0.1 bar (± 0.5 %)
MICROPROCESSOR Graphic display, 128 x 64 px Graphic display, 128 x 64 px
CONTROLLED DISPLAY
INDEX OF PROTECTION IP2 IP2
TEMPERATURE 5°C to +30°C 5°C to +30°C
RELATIVE HUMIDITY 0-80%, non condensing 0-80%, non condensing
SOUND PRESSURE LEVEL 50 dB(A) 50 dB(A)
STANDARD OUTPUT PORT Swagelok 1/8” G1/4
width 55 cm, height 42 cm, width 48 cm, height 70 cm,
CASE DIMENSIONS
length 51 cm length 84 cm
STP: Standard Temperature and Pressure (20°C, 1 bar)
Special Unit
ZEFIRO 10 ELSD
ZEFIRO Nitrogen Generator ELSD series is specifically designed to meet
the gas flow, purity and pressure requirements of Evaporative Light
Scattering Detectors. The Nitrogen generator uses pressure swing
adsorption technology (PSA) to produce nitrogen gas.
This system uses carbon molecular sieve (CMS) which selectively
adsorbs oxygen and water vapor molecules under high pressure, while
allowing Nitrogen to pass through.
Application
ELSD
Technical data
STANDARD NITROGEN FLOW RATE 10 l/min at STP
OUTLET PRESSURE up to 5.5 bar at STP (80 psi)
STANDARD PURITY 99,9% at STP
INPUT VOLTAGE 230 V - 50 or 60 Hz
WEIGHT 110 Kg
POWER CONSUMPTION 800 W
PRESSURE ACCURACY 0.1 bar (± 0.5 %)
MICROPROCESSOR CONTROLLED DISPLAY Graphic display, 128 x 64 px
INDEX OF PROTECTION IP20
TEMPERATURE 5°C to +30°C
RELATIVE HUMIDITY 0-80%, non condensing
SOUND PRESSURE LEVEL 50 dB(A)
OUTPUT PORT G 1/4
CASE DIMENSIONS width 48 cm, height 70 cm, length 84 cm
Special Unit
ZEFIRO HIGH PRESSURE
The HIGH PRESSURE nitrogen generator has been specifically designed
to meet the gas flows, purity and pressure requirements of the Accelerated
Solvent Extraction System (ASE®), the technique for the extraction of solid
and semisolid sample matrices using common solvents at elevated
temperatures and pressures. The Nitrogen reach the maximum pressure
of 11 bars.
Application
ASE
Sample Preparation
Technical data
STANDARD NITROGEN FLOW RATE Up to 40 l/min at STP
OUTLET PRESSURE Up to 11 bar (160 psi)
STANDARD PURITY 99,0% at STP
INPUT VOLTAGE 230 V - 50 or 60 Hz
WEIGHT 140 Kg
POWER CONSUMPTION 1800 W
PRESSURE ACCURACY 0.1 bar (± 0.5 %)
MICROPROCESSOR CONTROLLED DISPLAY Graphic display, 128 x 64 px
INDEX OF PROTECTION IP20
TEMPERATURE 5°C to +30°C
RELATIVE HUMIDITY 0-80%, non condensing
SOUND PRESSURE LEVEL 50 dB(A)
OUTPUT PORT G1/4
CASE DIMENSIONS width 48 cm, height 70 cm, length 84 cm
* All LC/MS brands can be suppliend combining max 30 l/min of N2 or max 70 l/min of AIR
Special Unit
ZEFIRO MP
ZEFIRO Nitrogen Generator “MP” series is specifically designed to supply
Nitrogen & Exhaust Air for the AGILENT MP-AES.
The Nitrogen generator uses pressure swing adsorption technology (PSA)
to produce nitrogen gas at specific purity required > 99,5%.
This system uses carbon molecular sieve (CMS) which selectively adsorbs
oxygen and residual water vapor molecules under high pressure, while
allowing Nitrogen to pass through. The internal oil free air compressor
Application
which is firstly pre filtered provide also to the exhaust air flow which is
treated by a drying filter.
AGILENT
MP-AES
Technical data
STANDARD NITROGEN FLOW RATE 26,5 l/min at STP
STANDARD DRY AIR FLOW RATE 25 l/min at STP
OUTLET NITROGEN PRESSURE up to 5.5 bar (80 psi)
OUTLET DRY AIR PRESSURE up to 5.5 bar (80 psi)
STANDARD NITROGEN PURITY ≥ 99,5% at STP
INPUT VOLTAGE 230 V - 50 or 60 Hz
WEIGHT 150 Kg
POWER CONSUMPTION 1900 W
PRESSURE ACCURACY 0.1 bar (± 0.5 %)
MICROPROCESSOR CONTROLLED DISPLAY Graphic display, 128 x 64 px
INDEX OF PROTECTION IP2
TEMPERATURE 5°C to +30°C
RELATIVE HUMIDITY 0-80%, non condensing
SOUND PRESSURE LEVEL 55 dB(A)
OUTPUT PORT N. 02 G 1/4
width 48 cm, height 70 cm,
CASE DIMENSIONS length 84 cm
STP: Standard Temperature and Pressure (20°C, 1 bar)
Feb. 2018
CINEL GROUP
HYSTORY
Cinel Strumenti Scientifici was founded in Padua in the 70’s with a technical partnership of INFN LNL
Legnaro Laboratory on particle accelerator projects and since then has been involved in some of the most
challenging projects all over Europe.
Nowadays, CINEL has reached a long experience on mechanical design and manufacture of apparatuses
in several scientific and research fields such as Synchrotron Light Sources (monochromators, fully integra-
ted front ends and beam lines, experimental chambers), as well as accelerator components (vacuum
chambers, accelerating cavities, radiofrequency quadrupole cavities). Cinel has acquired skilled experien-
ce in the field of cryogenics, superconductivity, astrophysics and bio-mechanics collaborating with
well-known institutions as a qualified partner in the mechanical, thermal and control system design and it
can now propose turnkey solutions with high level standardization.
A new activity branch has been started in 2005 profiting from the expertise gained in years of R&D
programs finalized to high precision mechanics in several research fields; a complete range of analytical
gas supply generators has been devolved (nitrogen, hydrogen and zero air) widening the business area to
the laboratory analytical instrumentation. Since the beginning the new division has been devoted to the
design and production of analytical gas generators which have been rapidly renowned in Italy first than in
Europe and outside Europe.
Thanks to the positive impact and to the very promising results on the market, the analytical gas genera-
tors branch activity in the end of 2014 was separated from the main company “Strumenti Scientifici Cinel”
and the new company “CINEL SRL Gas Generators Technology” has been founded.
PATENT
ITALIAN PATENT
Number 0001397254
INTERNATIONAL PATENT
Application Number EP10814714.1
CERTIFICATIONS
Azienda con sistema di qualità certificato ISO 9001:2008
“Design and production
of laboratory N2 H2 AIR
gas generators”
CINEL S.r.l. Gas Generators Technology
Via Andreon, 5
35010 Vigonza (Padova) - Italy
tel. +39 049 725022
fax +39 049 8931881
e-Mail
[email protected]VAT IT04866570288 www.cinelsrl.com