Camera Installation
Product: MAMMOMAT Novation DR (DROC)
Title: Specific Hardcopy Camera Information for
Agfa LR 5200 / MG 3000
Affected systems: MAMMOMAT Novation DR with Image Processing, V1.0 and Hardcopy Cameras
Replaces: n.a.
1 General Information
1.1 Block Diagram
SCP HCC
SCU / Modality
Agfa Agfa
MAMMOMAT MG 3000 LR 5200
Novation DR (DROC)
Network
1.2 Available Film Size
Application Film Type Medium Type
14"x17" Portrait n.a. n.a. n.a.
Landscape n.a.
14"x14" Portrait n.a. n.a. n.a.
Landscape n.a.
11"x14" Portrait n.a. n.a. n.a.
Landscape n.a.
10"x12" Portrait n.a. n.a. n.a.
Landscape n.a.
8"x10" Portrait yes LT3B Blue-based
Landscape yes
others n.a. n.a. n.a.
yes - Film size is available and released
no - Film size is available but not released
n.a. - Film size is not available
1.2.1 Supported Format
Supported page format is always 1x1
Siemens AG, Medical Solutions Print No.: SPB7-250.814.13.01.02 © Siemens AG, 2004
CS SD 24 05.2004 Page 1 of 5
MAMMOMAT Novation DR (DROC) Camera Installation
Specific Agfa LR 5200 / MG 3000 Information
1.3 Restrictions
For HCC-independent restrictions, see also the general restrictions in the MAMMOMAT Novation DR
document.
The linear look up table is out of tolerance within the minimum density range.
LT3B Calibration Film must be measured with the high reflecting side downwards (Agfa Text looks
mirrored).
LT3B Film is only available for film size 8"x10".
2 Modality SW Configurations / Parameters
Refer to: MAMMOMAT Novation DR/CB-DOC/ Installation/Installations instructions.
File name Content Remarks
lr5200.cf Configuration DROC This parameter
g Software Configuration can be changed
Edit ./installed/raw/vdevlr5200_001 within Netscape
Edit ./installed/raw/vdevlr5200_001's configuration file: Service Login.
/linx_mp/resources/installed/raw/vdevlr5200_001/lr5200.cfg Only importend
-------------------------------------------------------------------------------- parts of the file
Group: Server shown.
factory object name =
RMS://siemens06/installed/services/jss_server:PJF_OREF
spooler name = DicomPrintSpooler
priority = normal
max retries = 20
--------------------------------------------------------------------------------
Group: Output Device
device type = DRYSTAR
--------------------------------------------------------------------------------
Group: Print Job Description
device name = LR5200
device address = 146.254.124.11:104
request annotation SOP class = false
request print job SOP class = false
print job poll period = 15
print job timeout = 300
printer poll period = 15
association timeout = 0
collate = false
custom lut file =
photometric interpretation = automatic
--------------------------------------------------------------------------------
Group: Film Session
copies = 1
medium type = BLUE FILM
film destination = PROCESSOR
film session label = DROC film
memory allocation = 0
print priority = MED
--------------------------------------------------------------------------------
Siemens AG, Medical Solutions Print No.: SPB7-250.814.13.01.02 © Siemens AG, 2004
CS SD 24 05.2004 Page 2 of 5
MAMMOMAT Novation DR (DROC) Camera Installation
Specific Agfa LR 5200 / MG 3000 Information
Group: Film Box
available image display formats = PORTRAIT\1,1
image display format = PORTRAIT\1,1
annotation display format id = ANNOTATION
number of annotation boxes = 6
annotation boxes per text line = 3
max annotation box text string size = 63
annotation string =
max text line size = 189
number of image text lines = 5
image text position = left
image text overlap = overwrite
image text reorient = reorient_0
film size id = 8INX10IN
film magnification type = REPLICATE
film smoothing type = NONE
border density = BLACK
empty image density = BLACK
full pixel range = true
min density = 20
max density = 350
configuration info = PERCEPTION_LUT=LINEAR
trim = NO
--------------------------------------------------------------------------------
Group: Image Box
polarity = NORMAL
available image magnification types = REPLICATE BILINEAR CUBIC
image magnification type = REPLICATE
image smoothing type = NONE
true size capable = true
enable selective true size = true
pixel size = 70
pad = false
image text line 1 = <Off>
image text line 2 = <Off>
image text line 3 = <Off>
image text line 4 = <Off>
image text line 5 = <Off>
----------------------------------------------------------------
Group: PORTRAIT\1,1
width = 3337
height = 4114
Note: If Film Size ID is “automatic”, the appropriate film size (Large Patient Matrixsize=large
Filmsize) will be printed.
After a change of Film Size ID in Netscape you have to press the Update button and restart
the Sun!!!
Siemens AG, Medical Solutions Print No.: SPB7-250.814.13.01.02 © Siemens AG, 2004
CS SD 24 05.2004 Page 3 of 5
MAMMOMAT Novation DR (DROC) Camera Installation
Specific Agfa LR 5200 / MG 3000 Information
3 Hardcopy Camera, Configuration
3.1 Identification
Device Type: DICOM Print Server(SCP) Printer
Manufacturer / Device Name: Agfa MG 3000 LR 5200
HW Revision: Board GEMINI Revision 2 Type 8394 / 200
SW Revision: MeGa 3.32, LTP 01806.240698
aos 32946,
cad 32946,
ecu 32948,
dcm 32956.
OEM 43969.
PNPDCM 1001 echo,
Image Memory: 128 MB n.a.
Hard disk: 4.1 GB None
Number of Operating Panels: None 1 built-in keypad
No./Type of Input: 1 Ethernet Board FOI Optic
No./Type of Output: FOI Optic n.a.
Physical Network Interface: 10Base-TP Ethernet None
Network Protocol: TCP / IP V.4 None
DICOM Conformance State- Rev.1.4, April 20, 2001 None
ment:
3.2 SCP / Printer Settings
File name Content Remarks
PMS.INI [DIRECT_DIGXRAY] PMS2000.IDF
IDF=C:\PMS2000.IDF default STANDARD\1,1
CFG=D:\SIEMGLIN.PMS Agfa layout ID:
8x10P ID2011
8x10L ID2111
SIEMGLIN.PMS 00
treshold_scale 200 Enable HighRes Mode
utl_id 1 LINEAR LUT
AGFA default confrmLevel is 0 (warnings)
3.3 Error and Warning Messages
n.a.
4 Image Quality and Optical Density
4.1 Reference Film and LUT
Released Film Type LT3B BLUE FILM
Refer to: MAMMOMAT Novation DR/CB-DOC/ Installation/Installations instructions.
Siemens AG, Medical Solutions Print No.: SPB7-250.814.13.01.02 © Siemens AG, 2004
CS SD 24 05.2004 Page 4 of 5
MAMMOMAT Novation DR (DROC) Camera Installation
Specific Agfa LR 5200 / MG 3000 Information
4.2 Real Size Scaling
n.a.
5 Abbreviations
AE Title Application Entity Title
DCS DICOM Conformance Statement
DICOM Digital Imaging and Communication in Medicine
HW / SW Hardware / Software
HCC Hardcopy Camera
LUT Look Up Table
OEM Original Equipment Manufacturer
SCP Service Class Provider
SCU Service Class User
TCP/IP Transfer Control Protocol / Internet Protocol
PLA Pacsnet Logical Address
6 Changes to Previous Version
n.a.
© Siemens AG 2003
The reproduction, transmission or use of this document or its contents is not permitted without ex-
press written authority. Offenders will be liable for damages. All rights, including rights created by pat-
ent grant or registration of a utility model or design, are reserved.
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery. Revi-
sions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho-
rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.
Siemens AG, Medical Solutions Print No.: SPB7-250.814.13.01.02 © Siemens AG, 2004
CS SD 24 05.2004 Page 5 of 5