3.

7the use of the printer and Settings


A. the realization of the function of moisture meter print must choose my
company XY - 08 p type with special printer
B. moisture test over, if you want to print the test report according to the
following:
1. Long press \"set\" button, when there is a \"print\" let go, to print function
setting state. When a T for the RS232 serial port output.
2. L - C (Chinese) or L - E (English) select print language, according to the \"on\"
or \"down\" button to print the language. Press the \"confirm\" to enter the next LUOYANG YUBO TESTING TECHNOLOGY
level menu.
3. Showing 1-5 select print in the recent 5 times water test records at any time, CO., LTD
choose to end press \"confirm\" button to print the selected the water test report.
Continue to choose 1-5, then press \"confirm\" button to print the selected
another water test report. Halogen Moisture Analyzer
4. Press the \"reset\" button to exit the Settings back to the state of weighing

Tips:
1. Warm-up the analyzer 30 minutes and take the first time calibration;
2. After the end of a water calibration experiments (see 3.6);
3.Each time the switch machine, please interval of 5 seconds;
4 .Cover the heater, do not leave the weight on the pan, or halogen lamps will be
crushed;
ATRO 100%~999%
ATRO 0%~999%
History store 15 set
1
About the Halogen Moisture Analyzer Schedule: Assembly drawing

Mode MS-100-XY
Specification 110g/5mg
Readability 5mg
Tray size Φ 90
Heating mode Halogen Lamp
Display LCD Please installation the instrument according the picture
Temperature Setting 40℃~199℃ Installation order from top to bottom :1 --2--3--4--5
Time Setting 1~99 minutes 1. Aluminum scale weighing tray
2. The triangle scale frame
3. Scale tray brackets
4. Wind cover
5. Stainless steel plate
10 9
3.5 Calibration I Instruction
When the instrument error or improper operate. Calibration weights can be adjusted 1. Security
with accuracy. Please reading the following before you use the Moisture analyzer
Press "Reset" key and does not release until it display“-cal-”, a. The instrument can use for test the moisture of the sample,each
Then flashing “100.000”, put on the calibration weight, then it will display "= = = = = improper operation may result in personal injury and equipment
=", damage
until it display “100.000” remove the weight, it will display "= = = = = = "again, b. Please access the AC power accordance with the manual;
Waiting it show" 0.000 ", the calibration is completed. c. Please make sure the power cord does not interfere with people
IV RS232 Setting action;
d. Do not use the instrument under the hazardous,wet or unstable
Baud Rate:1200
enviroment;
Date Bits:7 e. Unplug the power before you clean the analyzer;
f. Please keep enough space around the instrument,and keep the top of the space at
Odd-even check:No
least 1m;
Flow control:Continuous g. During a special experiment, please wear protective equipment by youself;
h. Please do not change the instrument's components and any other aspects.
Rs232 Commands
Maintenance of this equipment by the factory authorized professionals。
Commands Function Moisture analyzer use Halogen lamp.
i. Do not put the flammable materials around the instrument;
On/Off Power on /Power off
j. Do touch the halogen lamp when it on working.
g % Conversion results Special attention to the dangers of some sample;
k. Before you test the moisture of dangerous good,pleaes take Full
investigation
Inflammable and explosive materials:Heating the inflammable and
explosive materials will produce flammable gases or vapors, so need
to lower the temperature in a dry environment. In case of fire or
explosion. Toxic and corrosive substance: Keep the samples of toxic or
corrosive ingredients in well-ventilated environment.
Above the sample test will be particularly careful, for any damage caused
by your own risk
2
Place the wind cover, Place the sample tray Put the aluminum sample
II.Installation aligned holes. holder, rotate until tray on the bracket.
2.1 Standard Packing List Please check the packaging and accessories. positioning.
Standard Quantity Note 3
Moisture Analyzer Mainframe 1 set when flashing .Press"UP"or"Down"to choose the stop mode.and press
Wind Cover 1 piece "ENTER" confirm , Parameters are defined, and go to the next parameter
The sample pan holder 1 piece (temperature) setting.
Tray handle 1 piece 3.5.4 Heating Temperature Setting
Power cord 1 piece The instrument temperature setting range is from 0-199 degrees. Press the
"MENU" to enter the parameter setting mode, press the "ENTER" key until
Aluminum sample tray 1 set 50 pcs of aluminum pan
the flashing,Press"UP"or"Down"to choose the temperature.and press
100g standard weight 1 piece
"ENTER" to go to next setting(time setting)
Product Manual 1 piece
Icon will transform other temperature when its heating.
Product certification 1 piece
3.5.5 Heating time setting
Warranty card 1 piece
1.Just choose timing stop mode need set the stop time.
2.2 Moisture meter placement requirements.
Press the "MENU" to enter the parameter setting mode, press the "ENTER"

key until the flashing , Press"UP"or"Down"to choose the

temperature.and press "ENTER"


Operating table Temperature stable, good ventilation should
2, Heating cycle:Just gentle heating mode need choose heating cycle
be stable dry, without strong vibration, of the environment
3.6 For example:
without strong magnetic field.
a. sample:water,
2.3 Installation of accessories
b. Press the "MENU" to enter the parameter setting mode,choose the (

),temperature:120℃,Press "ENTER" go to weighing state


c. Put a dry paper on the pan,and press"Tare" clean the weight. b. Manual stop mode: ( )
Put on the 3g water on the paper,until the readability stable. Sample heating process is the continuing until the press the "start/stop"
Press" START / STOP " begin test. button to end testing.
Waiting for it sop.look the result. c. Timing Stop Mode: ( )
The m% will be 100%. Stop time set by the user, the sample heating process is continued until the end
Weighing error or sample weight deficiency can lead to the result of the test of the heating time to set the value of the test.
a deviation Press the "MENU" button to enter the parameter setting mode, press the
"ENTER" key until the shutdown mode
8
a. Standard heated mode: ( ) 7
This mode is the factory setting, suitable for most samples. Heating 2.4 Adjust the level and connect the power
temperature can be set, the sample is heated to a temperature, and keep to the There are two feet in front of the instrument,adjust it and keep the
test ends automatically. instrument level.
b. Rapid heating mode: ( )
This mode is suitable for samples with higher moisture content. You can set
the heating temperature, heating the temperature will exceeds the set value, the Connect the instrument with power,put one side of power cord
future will be reduced to the set temperature and hold until the test ends
automatically. Please warm up more than 30minutes for keep it stable and accurate.
c. Gentle heating mode: ( )
This mode is suitable for low moisture content, surface casing is easy to form, III Operation
easy to burn the sample. From the heat began to rise to the long time required
3.1.Display
to set temperature, heating up even slowly. Heating time user selectable.
d. Press the "MENU" to enter the parameter setting mode, press the "ENTER"
until the heating mode, the icon flashes 。Press "UP"or"Down"choose
the mode you need.and press "ENTER"confirm.
3.5.3 Stop mode setting:( )
a. Automatic stop mode:( )
3.2.Icon Introdution
Heating the sample within the specified time during the weight loss of less
than 2mg will automatically stop when the test. This mode after 15 seconds in ----The percentage of dry weight ----The percentage of moisture
the heating effect.
START / STOP Start heated/ stop heated
----Gram ----Temperature
TARE Weighing peeled, Settings exit
---- Time set and heat time MENU Enter the setting menu
CONVERSION Switching measurement results
----Heated mode
ENTER Confirm the current setting, go to next
----Standard mode ----Rapid heating mode Return to the weighing mode (short press), calibration
RESET
balance (long press)
---- Gentle heating mode ---- Stop heated mode DOWN Reduce
4 UP Add
---- Automatic -----Manual stop mode 5
----Timing stop mode 3.4 First testing
1. First test, if not change the temperature and time settings, the moisture meter defaults
---- Heated state to factory defaults.
3.3. Keypad 2. Put empty sample pan on the holder (see section 2.3).
3. Press the “TARE” to clean weight.
4.Put the sample on the pan (Sample weight should be heavy than 1g, the general 3-5g).
5. Close the heating cover, press the "Start / Stop" button to start the test. At this point,
heated tube lights, fans turn on, heated symbol in the upper left flashing. While rolling
on the screen displays the current temperature, has been heating time, moisture content
of the current sample.
6. End of the test, heated lamp automatic shut off, the buzzer alarm. Then the sample is
displayed on the moisture content, heating time, current temperature and the parameter
setting mode.Note: Don't touch the sample plate,the temperature is too high now.
7 Press the "Reset" button, the instrument returns to weighing mode.
3.5 Parameters Setting
NAME Key Features 3.5.1 Heated parameter storage and recall
a. The instrument can store a total of 16 0-F group of heated parameters. Each group
ON/OFF Power on / off, Just off the internal DC power supply
can set the heating method, stop mode, time and time test parameters.
b. In weighing mode, press the "Settings" button to enter the parameter setting mode,
then store the address number (the number show the highest bit) is flashing. Press
"down" or "up" key to select the desired address number, the 0 is the factory
setting,1-F for the user store.
c. The memory address number 0 means factory settings, then click "ENTER"
button to exit the settings back to weighing mode. At this point heated mode is
the factory defaults.
d. When the memory address range for the 1-F any one number, press "ENTER" key,
and enter the parameter setting after the completion of a parameter (heated mode,
stop mode, time mode) setting.Press the "TARE" key, the address number of the
test parameters have been stored.
3.5.2 Heated mode and setting( )