Informe de dispositivos
Etiqueta:
PC Host, COM3 [HART Communication], HARTCH, [Cerabar M
Ruta de comunicación:
5x / PMx 5x / V1.00.xx]:0
Creado por: Aux instrumentacion
Creado en: viernes, 26 de agosto de 2022 8:16:24
Estado DTM: En línea
Dirección del dispositivo: 0
Número de serie: S7000B15BC0
Tipo de dispositivo: Cerabar M 5x / PMx 5x / V1.00.xx
Proveedor: Endress+Hauser
Clase: Pressure
Versión: 1.4.188.39
Página 1 de 8
Versión DTM: 1.4.188
Página 2 de 8
OnlineParameterize
Online Parameterize
Language
Parameter Value
Language English
Display/Operat.
Parameter Value
Display mode Main value only
Add. disp. value None
Format 1st value Auto
Setup
Parameter Value
Measuring mode Pressure
Press. eng. unit psi
Corrected press. 0,0psi
Pos. zero adjust Abort
Set LRV 0,0000psi
Set URV 300,0000psi
Damping switch On
Damping value 2,0s
Pressure af.damp 0,0psi
Extended setup
Parameter Value
Code definition 0
Device tag
Operator code 0
Current output
Parameter Value
Alarm behav. P Warning
Alarm cur.switch SW setting
Output fail mode Max
High alarm curr. 22,000mA
Set min. current 3.8 mA
Output current 22,000mA
Get LRV Abort
Set LRV 0,0000psi
Get URV Abort
Set URV 300,0000psi
Diagnosis
Parameter Value
Status Failure
Diagnostic code F062 Sensor connect.
Instructions 1.Check sensorcable 2.Change electronics 3.Contact EH service
Last diag. code F062 Sensor connect.
Min. meas.press. -0,1307714psi
Max. meas.press. 1,3998660psi
Página 3 de 8
Diagnostic list
Parameter Value
Diagnostic 1 F062 Sensor connect.
Instructions 1 1.Check sensorcable 2.Change electronics 3.Contact EH service
Diagnostic 2 F081 Initialization
Instructions 2 1.Reset 2.Chk sensorcable 3.Contact EH service
Diagnostic 3 No error
Instructions 3 No action needed
Diagnostic 4 No error
Instructions 4 No action needed
Diagnostic 5 No error
Instructions 5 No action needed
Diagnostic 6 No error
Instructions 6 No action needed
Diagnostic 7 No error
Instructions 7 No action needed
Diagnostic 8 No error
Instructions 8 No action needed
Diagnostic 9 No error
Instructions 9 No action needed
Diagnostic 10 No error
Instructions 10 No action needed
Event logbook
Parameter Value
Last diag. 1 F062 Sensor connect.
Last diag. 2 F081 Initialization
Last diag. 3 No error
Last diag. 4 No error
Last diag. 5 No error
Last diag. 6 No error
Last diag. 7 No error
Last diag. 8 No error
Last diag. 9 No error
Last diag. 10 No error
Instrument info
Parameter Value
Firmware version 01.00.03
Serial number S7000B15BC0
Ext. order code XPP0001-S2+AD
Order identifier XPP0001-10M7/0
Cust. tag number ________
Device tag
ENP version 02.01.01
Config. counter 6
LRL sensor -14,50psi
URL sensor 580,13psi
Manufacturer ID 17
Device type code 25
Device revision 1
Página 4 de 8
Measured values
Parameter Value
Meas. pressure 0,0psi
Sensor pressure 0,0psi
Corrected press. 0,0psi
Pressure af.damp 0,0psi
Sensor temp. 25,0°C
Simulation
Parameter Value
Simulation mode None
Reset
Parameter Value
Enter reset code 0
Expert
System
Parameter Value
Code definition 0
Lock switch Off
Operator code 0
Instrument info
Parameter Value
Cust. tag number ________
Device tag
Serial number S7000B15BC0
Firmware version 01.00.03
Ext. order code XPP0001-S2+AD
Order identifier XPP0001-10M7/0
ENP version 02.01.01
Electr.serial no S614B101BCC
Sensor serial no S61D9A01C22
Display
Parameter Value
Language English
Display mode Main value only
Add. disp. value None
Format 1st value Auto
Management
Parameter Value
Download select. Configuration copy
Enter reset code 0
Página 5 de 8
Measurement
Parameter Value
Measuring mode Pressure
Basic setup
Parameter Value
Pos. zero adjust Abort
Calib. offset 0,0000psi
Damping switch On
Damping value 2,0s
Press. eng. unit psi
Temp. eng. unit °C
Sensor temp. 25,0°C
Pressure
Parameter Value
Set LRV 0,0000psi
Set URV 300,0000psi
Meas. pressure 0,0psi
Sensor pressure 0,0psi
Corrected press. 0,0psi
Pressure af.damp 0,0psi
Sensor limits
Parameter Value
LRL sensor -14,50psi
URL sensor 580,13psi
Sensor trim
Parameter Value
Lo trim measured -14,5032597psi
Hi trim measured 580,1303711psi
Lo trim sensor -14,5033psi
Hi trim sensor 580,1304psi
Output
Current output
Parameter Value
Output current 22,000mA
Alarm behav. P Warning
Alarm cur.switch SW setting
Output fail mode Max
High alarm curr. 22,000mA
Set min. current 3.8 mA
Get LRV Abort
Set LRV 0,0000psi
Get URV Abort
Set URV 300,0000psi
Startcurrent 12mA
Curr. trim 4mA 4,000mA
Curr. trim 20mA 20,000mA
Offset trim 4mA 0,000mA
Offset trim 20mA 0,000mA
Página 6 de 8
Communication
HART config
Parameter Value
Burst mode Off
Burst option Command 1
Current mode Signaling
Bus address 0
Preamble number 5
HART info
Parameter Value
Device type code 25
Device revision 1
Manufacturer ID 17
Hart version 6
Descriptor ________________
HART message ________________________________
HART date __/__/__
HART output
Parameter Value
Primary value is Meas. pressure
Primary value 0,0000000psi
Secondary val.is Corrected press.
Secondary value 0,0000000psi
Third value is Sensor pressure
Third value 0,0000000psi
4th value is Temp. sensor
4th value 25,0000000°C
HART input
Parameter Value
HART input value 0,0Unknown
HART input stat. Uncertain
HART input unit Unknown
HART input form. x.x
Application
Parameter Value
Electr. delta P Off
Página 7 de 8
Diagnosis
Parameter Value
Status Failure
Diagnostic code F062 Sensor connect.
Instructions 1.Check sensorcable 2.Change electronics 3.Contact EH service
Last diag. code F062 Sensor connect.
Reset logbook Abort
Min. meas.press. -0,1307714psi
Max. meas.press. 1,3998660psi
Reset peakhold Abort
Operating hours 0
Config. counter 6
Diagnostic list
Parameter Value
Diagnostic 1 F062 Sensor connect.
Instructions 1 1.Check sensorcable 2.Change electronics 3.Contact EH service
Diagnostic 2 F081 Initialization
Instructions 2 1.Reset 2.Chk sensorcable 3.Contact EH service
Diagnostic 3 No error
Instructions 3 No action needed
Diagnostic 4 No error
Instructions 4 No action needed
Diagnostic 5 No error
Instructions 5 No action needed
Diagnostic 6 No error
Instructions 6 No action needed
Diagnostic 7 No error
Instructions 7 No action needed
Diagnostic 8 No error
Instructions 8 No action needed
Diagnostic 9 No error
Instructions 9 No action needed
Diagnostic 10 No error
Instructions 10 No action needed
Event logbook
Parameter Value
Last diag. 1 F062 Sensor connect.
Last diag. 2 F081 Initialization
Last diag. 3 No error
Last diag. 4 No error
Last diag. 5 No error
Last diag. 6 No error
Last diag. 7 No error
Last diag. 8 No error
Last diag. 9 No error
Last diag. 10 No error
Simulation
Parameter Value
Simulation mode None
Página 8 de 8