0% found this document useful (0 votes)
13 views72 pages

Rev L

Uploaded by

Mohand Alqubati
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
13 views72 pages

Rev L

Uploaded by

Mohand Alqubati
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 72

1 2 3

2008K
HEMODIALYSIS
4 5 6

7 8 9 0

SYSTEM

PREVENTIVE
MAINTENANCE
PROCEDURES

Part Number 507297 Rev. L


FRESENIUS MEDICAL CARE
NORTH AMERICA
800-227-2572

Fresenius Medical Care North America


920 Winter St.
Waltham, MA 02451

Manufactured by:
Fresenius USA, Inc.
4040 Nelson Avenue
Concord, CA 94520

REGIONAL EQUIPMENT SPECIALIST: ___________________________________


2008K
HEMODIALYSIS SYSTEM

PREVENTIVE MAINTENANCE
PROCEDURES
Part Number 507297 Rev. L

INCLUDING
PREVENTIVE MAINTENANCE CHECKLISTS
QUARTERLY/1000 HOUR AND ANNUAL/4000 HOUR

http://www.fmcna.com

Copyright  2000 – 2017 Fresenius Medical Care, All Rights Reserved

Page i
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
2008K Preventive Maintenance Procedures
This document contains proprietary information of Fresenius USA, Inc. d/b/a Fresenius
Medical Care North America and its affiliates (“Fresenius Medical Care”). The contents of
this document may not be disclosed to third parties, copied, or duplicated in any form, in
whole or in part, without the prior written permission of Fresenius Medical Care.
Fresenius Medical Care, the triangle logo, 2008 and DIASAFE are trademarks of Fresenius
Medical Care Holdings, Inc., or its affiliated companies. All other trademarks are the
property of their respective owners.
Caution: US Federal law restricts this device to sale only by or on the order of a physician.
Frequency, duration, and parameters of treatment are to be determined by the prescribing
physician.
Installation, maintenance, calibration and other technical information may be found in the
2008K Technician’s Manual, P/N 490049.
Contact Fresenius Medical Care Technical Support for applicable Field Service Bulletins.
The spare parts manual for the model 2008K and other information may be found on our web
site at www.fmcna.com
Indications for Use: The 2008K hemodialysis machine is indicated for acute and chronic
dialysis therapy.

Conventions
Symbol Description
Warning! A warning is a statement that identifies conditions or
actions that could result in personal injury or loss of life.
Warnings found in this manual outside of this section are
designated with the warning symbol.
Shock Hazard: A shock hazard warning refers to a risk of a
possibly severe electrical shock due to improper use or handling
of the equipment.

Caution: A caution is a statement that identifies conditions or


actions that could result in damage to the machine.

Note: Notes are advisory comments or recommendations


regarding practices or procedures.

Page ii
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
PREVENTIVE MAINTENANCE PROCEDURES
TABLE OF CONTENTS
1.0 INTRODUCTION ............................................................................................................................................ 1
1.1 TEST EQUIPMENT AND SUPPLIES NEEDED .......................................................................... 1
1.2 OPERATING MODES ................................................................................................................ 3
1.3 FRONT PANEL CONTROLS ...................................................................................................... 4
1.4 MEASURING FLUID VOLUMES ................................................................................................ 6
2.0 QUARTERLY (1000 HOUR) PREVENTIVE MAINTENANCE ....................................................................... 7
2.1 FILTERS AND O-RINGS ............................................................................................................ 9
2.2 PRE-UF PUMP FILTER ............................................................................................................ 10
2.3 DIASAFE FILTER (IF APPLICABLE) ...................................................................................... 10
2.4 UF PUMP CHECK VALVES ..................................................................................................... 10
2.5 HIGH VOLTAGE AC CONNECTIONS ...................................................................................... 11
2.6 UF PUMP ................................................................................................................................. 12
2.7 CONDUCTIVITY ...................................................................................................................... 12
2.8 TEMPERATURE ...................................................................................................................... 13
2.9 LEVEL DETECTOR .................................................................................................................. 14
2.10 ALARM OPERATION AND PRESSURE HOLDING TESTS ..................................................... 18
2.11 VERIFY PH .............................................................................................................................. 20
2.12 POWER FAILURE ALARM ....................................................................................................... 20
2.13 BLOOD PUMP ROTOR INSPECTION ..................................................................................... 21
2.14 FINAL CHECKS ....................................................................................................................... 23
3.0 ANNUAL (4000 HOUR) PREVENTIVE MAINTENANCE ............................................................................ 24
3.1 FILTERS AND O-RINGS .......................................................................................................... 24
3.2 PRE-UF PUMP FILTER ............................................................................................................ 24
3.3 UF PUMP CHECK VALVES ..................................................................................................... 24
3.4 DEAERATION RESTRICTOR .................................................................................................. 25
3.5 DIASAFE FILTER (IF APPLICABLE) ...................................................................................... 25
3.6 DIAPHRAGM PUMPS .............................................................................................................. 25
3.7 HEATER ELEMENT ................................................................................................................. 26
3.8 HIGH VOLTAGE AC CONNECTIONS ...................................................................................... 26
3.9 DEAERATION MOTOR BRUSHES .......................................................................................... 27
3.10 INLET WATER PRESSURE REGULATOR .............................................................................. 28
3.11 ONLINE CLEARANCE TEST (IF APPLICABLE)....................................................................... 29
3.12 DEAERATION AND LOADING PRESSURE ............................................................................ 32
3.13 FLOW RELIEF PRESSURE ..................................................................................................... 33
3.14 CONCENTRATE AND BICARB PUMPS .................................................................................. 34
3.15 UF PUMP ................................................................................................................................. 34
3.16 CONDUCTIVITY ...................................................................................................................... 34
3.17 TEMPERATURE ...................................................................................................................... 34
3.18 VOLT HI LO DETECT ............................................................................................................... 34
3.19 BLOOD LEAK AND DIMNESS ................................................................................................. 35
3.20 ARTERIAL, VENOUS AND TRANSMEMBRANE PRESSURE ................................................. 36
3.21 DIALYSATE FLOW .................................................................................................................. 39
3.22 HEPARIN PUMP ...................................................................................................................... 39
3.23 BLOOD PUMP .......................................................................................................................... 42
3.24 LEVEL DETECTOR .................................................................................................................. 44
3.25 ALARM AND PRESSURE HOLDING TESTS ........................................................................... 44
3.26 VERIFY PH .............................................................................................................................. 44
3.27 RINSE CHECKS ....................................................................................................................... 45
3.28 POWER FAILURE ALARM AND BATTERY REPLACEMENT .................................................. 46
3.29 BLOOD PRESSURE MODULE ................................................................................................ 47
3.30 FINAL CHECKS ....................................................................................................................... 50
4.0 REBUILDING THE DIAPHRAGM PUMPS .................................................................................................. 51
4.1 REBUILDING THE ULTRAFILTRATION PUMP ....................................................................... 51
4.2 REBUILDING THE CONCENTRATE AND BICARBONATE PUMPS ....................................... 54
4.3 TESTING CONCENTRATE AND BICARBONATE PUMPS ...................................................... 56
PREVENTIVE MAINTENANCE CHECKLIST QUARTERLY/1000 HOUR ........................................................... 61
PREVENTIVE MAINTENANCE CHECKLIST ANNUAL/4000 HOUR .................................................................. 63

Page iii
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
General Warnings
Warning! Never perform maintenance when a patient is
connected to the machine. If possible, remove the machine from
the treatment area when it is being serviced. Label the machine to
ensure it is not accidentally returned to clinical use before the
service work is completed. Always fully test the machine when
maintenance is completed. Confirm dialysate conductivity and pH
level before returning the machine to clinical use.

Shock Hazard: Refer servicing to qualified personnel.


The electrical source must be 120 volts, 60 Hz, single phase. The
outlet must be a three-conductor type with a hospital-grade
receptacle and a ground fault interrupter. Test the polarity and
ground integrity before installation and ensure it is maintained
thereafter. Failure to do so may result in electrical shock to the
operator or patient

Warning! Only Original Equipment Manufacturer (OEM)


Fresenius Medical Care parts should be used in the repair or
upgrade of the Fresenius Medical Care 2008K Hemodialysis
System. Although, parts may look similar to parts in various
vendor catalogs or brick and mortar stores the 2008K Hemodialysis
System uses parts that have been specified and tested in accordance
to ANSI/AAMI/ISO guidelines. The use of non-OEM parts will
void your warranty and may cause patient harm.

Note: This document is written for the 2008K Hemodialysis


Systems using software versions 2.16 or later.

Page iv
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
1.0 INTRODUCTION
Preventive Maintenance for the 2008K Hemodialysis System is simple and
straightforward. Maintenance is performed in only two intervals: Quarterly or after
1000 hours of operation, and annually or after 4000 hours of operation. The
maintenance procedures have been devised to require a minimum of time while
ensuring that the machine is maintained in optimum operating condition.

Included in the Preventive Maintenance procedures are tests to verify normal machine
operation. Should the machine fail to pass any of these tests, repair or re-calibrate as
needed, then repeat the tests until the specifications are met before returning the
machine to service.

Checklists are provided in the back of this manual to record the work done. Make
copies of these checklists as needed. Your initials on the checklist certifies that each
procedure has been completed and that the machine is performing according to the
specifications given.

1.1 TEST EQUIPMENT AND SUPPLIES NEEDED


A number of small parts must be available to perform the Preventive Maintenance.
Part number 190098 is a kit of the parts needed, except for the 9-Volt battery that
must be replaced during the annual preventive maintenance. An NEDA 1604AC
heavy-duty (alkaline type) battery is required. In addition, the following test
equipment is needed:

Warning! Test equipment used must be maintained and/or calibrated per the
test equipment manufacturer’s requirements. In particular, the dialysate meter
must meet the specifications listed below. Refer to the test equipment's'
operator's manual, or contact the manufacturer for calibration and maintenance
requirements. Failure to properly maintain and calibrate test equipment could
lead to improper calibration and/or failure of the device to meet its
specifications.

Warning! Disinfect the machine internally and externally and check all
pressure transducer protectors for contamination before working on the
machine.

Page 1
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
• 2008K Calibration Procedures (part number 507296).
• Test Kit (part number 150034), which contains two pressure gauges with
fittings and hoses for measuring loading pressure and deaeration pressure.
• Dialysate meter to measure dialysate pressure, temperature and conductivity at
the ends of the dialysate lines. The meter must be capable of making pressure
measurements from -250mmHg to +400mmHg with an accuracy of at least
±3mmHg. The temperature function of this meter must be accurate within
0.2°C from 20°C to 45°C and must be capable of measuring dialysate
temperatures up to 85°C with an accuracy of at least ±4.0°C. The conductivity
function of this meter must be accurate to within 0.1mS over a range of 12mS to
17mS at a temperature of 25°C.
• Stopwatch with a resolution to 0.01 second and an accuracy of 0.01% or better.
• Buret, 25ml capacity with 0.1ml graduations (part number 290104).
• Graduated cylinder: 1000ml capacity with a tolerance of 5.0ml at 1000ml or
better.
• Syringe, 60cc capacity. Tolerance is not important; the syringe is not used for
volume measurements.
• Tubing, 24” long (part number 545325-10). Use on the tip of the Buret.
• Resistor Plug Set for OLC Testing (part number 190168).
The following equipment is also required to test the blood pressure module:
• Dummy Cuff (part number 370090). The Dummy Cuff contains two air
chambers with calibrated volumes.
• Mercury manometer or equivalent pressure meter accurate to within 1mmHg at
pressures up to 330mmHg.

Page 2
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
1.2 OPERATING MODES
The following preventive maintenance procedures contain instructions to place the
2008K into Dialysis Mode and Service Mode.

To place the machine in Service Mode, turn the machine power On and wait for the
message Press CONFIRM for Service Mode to appear. Once it appears, press the
[CONFIRM] key and the message will change to Machine in Service Mode. After
the System Initializing process is complete, the machine will be in Service Mode.

If the [CONFIRM] key is not pressed when the Press CONFIRM for Service Mode
message is on the screen, the screen will change and the message Machine in
Dialysis Mode will appear. After the System Initializing process is complete, the
machine will be in Dialysis Mode.

Page 3
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
1.3 FRONT PANEL CONTROLS
The front panel consists of two areas, the touch screen and the control panel. The
touch screen is the area under the glass in the center of the front panel. The control
panel surrounds the touch screen and it contains the membrane keys.
Touch Screen

Control Panel

Figure 1 – 2008K Front Panel

Control Panel Operation


Throughout the preventive maintenance procedures, whenever a control panel key is
to be pressed, the appropriate key name is surrounded by square brackets as in the
following example:

Press the [CONFIRM] key and the screen will change.

In this example, the [CONFIRM] key on the control panel should be pressed.

Page 4
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
Touch Screen Operation
The touch screen is designed to display information and is used to enter data. To
select a button during a procedure, locate the button on the screen. Select a
parameter for which data will be entered using the keys below the screen.
Depending on the type of button, the screen will change. Data boxes are also
displayed on the touch screen. The following describes the type of buttons and data
boxes that will be encountered during the preventive maintenance process.
Data Button
A yellow data button is used to enter a measured volume or
value. When the yellow area of the data button is touched,
it will change to a darker yellow. The data can be changed
using the [▲] and [▼] keys or the value can be entered using
the number keys both on the control panel. Once the data is
entered, press the [CONFIRM] key and the data button
changes back to light yellow. The [Escape] key can be
pressed when the data button is dark yellow to abort the data
entry and return it to light yellow. The entered data does not
get stored until the [CONFIRM] key is pressed.
Some data buttons will change the screen and the data entry
will be performed on the new screen.
A gray data button means the button is not active and
touching it will have no effect.
Screen Button
Blue rectangles on the touch screen are screen buttons. By
touching the blue area of the screen button the display will
either change to another screen or the selection of an option
will change. A screen button is not active if it is gray.

Data Box
This type of box shows selected data or data the machine is
measuring. During the preventive maintenance process this
type of box is used to verify a value or selection.

Page 5
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
1.4 MEASURING FLUID VOLUMES
Several of the following procedures require measuring fluid volumes using
graduated cylinders and laboratory burets. When making these measurements do the
following:

• Make certain the container is clean and dry before collecting the fluid to be
measured. Two drops of fluid are approximately 0.1ml, which is enough to
affect the accuracy of critical measurements.

• Ensure that no items such as thermometers or tubing are allowed to come in


contact with the fluid in the graduate. Such items will change the calibration of
the graduate and affect the accuracy of measurements. Both the total volume
indicated and the amount of fluid indicated by each increment on the graduated
scale will be incorrect. For example, if a graduate is calibrated in 1ml
increments, a piece of tubing in contact with the fluid will cause each increment
to be less than 1ml, depending upon the total volume of the tubing that penetrates
into the fluid.

• Surface tension causes the fluid to curve into a meniscus (See Figure 2).
Measure the volume at the bottom of the meniscus curve as shown.

BOTTOM OF
MENISCUS
CURVE

Figure 2. Meniscus Curve.

Page 6
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
2.0 QUARTERLY (1000 HOUR) PREVENTIVE MAINTENANCE
Perform the following Preventive Maintenance procedures every 3 months or
1000 hours of machine operation, whichever comes first.

Make copies of the Quarterly Preventive Maintenance Checklist provided in the


back of this manual and use them to record the maintenance done.

Note: When performing an Annual Preventive Maintenance, do not perform


the quarterly procedures below first. Go directly to Section 3 and perform the
annual procedures described there.

While performing the following procedures, check the floor of the hydraulic unit and
all surfaces for moisture that might indicate a leak. Locate and correct any leaks
detected. Clean the floor of the hydraulic unit so that future leaks will be readily
apparent. Also, check all electrical connectors that can be reached to be sure they
are fully seated and there is no strain on the electrical cables.

Page 7
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
9-VOLT BATTERY

BLOOD PRESSURE
DIALYSATE LINE MODULE
CONN. O-RINGS
P/N 579097

STRAIN
RELIEF

DIALYSATE
INLINE FILTER HEATER
P/N 650113 CONNECTIONS

INLET WATER FILTER


P/N 330636
REDUNDANT
GROUND
UF PUMP
ADJUSTMENT DRAIN PORT

BICARBONATE PUMP
HEATER ELEMENT
INLET PRESSURE P/N 250169
REGULATOR

ACID PUMP BEHIND DEAERATION RESTRICTOR


INLET PRESSURE O-RINGS P/N F40007100 or
REGULATOR 579070

DEAERATION PUMP
FLOW PUMP INLET
OUTLET

Figure 3. 2008K Hemodialysis System, Rear View.

Page 8
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
2.1 FILTERS AND O-RINGS
Clean filters and replace O-rings as follows. Replace any filters that appear
damaged or corroded.

1. Inlet Water Filter in the inlet connector of the water supply (See Figure 3,
pg. 8).

Warning! After cleaning or replacing the inlet filter screen, disinfect the
water inlet line as described in the Operator’s manual and in accordance with
your Unit Policy.

Warning! Excessive silicone-gel O-ring lubricant can damage the hydraulic


pressure transducers (P-DIAL and CFS). Only apply a minimal thin layer of
O-ring lubricant when needed.

2. Filters and O-rings in the Concentrate and Bicarbonate Connectors


(See Figure 4). Replace all three O-rings in each connector.
FILTER INSERT P/N 566307 O-RING P/N F40007100 OR 579070

O-RING P/N 579092


O-RING
CONCENTRATE CONNECTOR (RED): P/N F40007107 or 641759
BICARBONATE CONNECTOR (BLUE): P/N F40007101 or 579072

Figure 4. Concentrate and Bicarbonate Connector Assemblies.

3. Clean the Dialysate Inline filter (See Figure 3, pg. 8).


4. Replace the O-rings in the Dialysate Line Connectors (See Figure 3, pg. 8).

Page 9
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
2.2 PRE-UF PUMP FILTER
Inspect the Pre-UF Pump filter for leaks or distortion. Replace the Pre-UF Pump Filter
if leakage or distortion is found.

Caution: Do not attempt to disassemble the Pre-UF Pump Filter. If not


properly reassembled, the Pre-UF Pump filter may leak. A leak in the
hydraulic system at this location may affect the operation of the machine or
cause fluid loss from the patient.

2.3 DIASAFE FILTER (IF APPLICABLE)


If the machine is equipped with a DIASAFE or DIASAFE PLUS filter system,
replace the filter every 90 days (quarterly). After replacing the filter, perform either
the automated test function or a manual pressure holding test AND then perform the
DIASAFE filter integrity test.

Note: If the machine is equipped with a DIASAFE filter, replace its quick
connector O-rings (part number 579097) when replacing the filter.

2.4 UF PUMP CHECK VALVES

Caution: If a check valve is replaced, ensure it is oriented correctly to allow


fluid flow in the proper direction.

Inspect the UF Pump Output check valves (one at the UF pump and the other
downstream at the UF Sample Port). Replace any that show signs of wear, damage
or leaking.

Page 10
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
2.5 HIGH VOLTAGE AC CONNECTIONS

Shock Hazard: Dangerous high voltage is present at the connections


accessed in this procedure when the machine is operating. Ensure the
machine's power plug is disconnected from the wall outlet before proceeding.

1. Remove power from the machine then check and tighten the 8-pin heater
connections next to the distribution board (See Figure 3, pg. 8). Check
heater block AC connections for signs of arcing or melting.
2. If applicable, inspect the power plug for loose or frayed wires. Ensure the
strain relief is securely fastened.

3. Inspect the entire length of the power cord (from plug to strain relief) for
nicks or cuts in the insulation and replace if necessary (part number
150425).

4. Confirm that the strain relief is tightly secured to the power supply chassis.

5. At the strain relief, locate the black, white and green wires from the power
cord (inside the power supply chassis). Follow the black and white wires to
the main power switch. Attached to the main power switch are four (4)
wires (2 black and 2 white). Look for loose connections, cracked insulation,
and signs of overheating, such as discolored or melted insulation. Replace
wires with power supply wire kit (part number 190411).

6. Inspect the main power switch and verify that its operation is smooth (no
grinding or sticky operation) and that the wires are not crossed.

Shock Hazard: Do not operate the machine if the resistance is greater than
0.2 ohm. A shock hazard to operators and patients could exist.

7. With a digital multimeter, measure the resistance between the round


(ground) pin on the power plug and the redundant ground terminal on the
machine (See Figure 3, pg. 8). Verify that the resistance is less than 0.2
ohm. If the value is above 0.2 ohm, measure the internal resistance of your
meter by shorting the leads together, then subtract this value from the
resistance measured between the power plug ground pin and the redundant
ground terminal on the machine to obtain the true ground resistance.

Page 11
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
8. Record the measured value on either the Quarterly Preventive Maintenance
Checklist or the Annual/4000 Hour Preventive Maintenance Checklist
depending upon which is being performed.

9. Perform the electrical safety checks required by local codes, facility


procedure and the Joint Commission on Accreditation of Healthcare
Organizations.

2.6 UF PUMP
Calibrate the UF Pump Volume (Refer to the 2008K Calibration Procedures –part
number 507296.)

Record the measured value on the Six (6) Month Preventive Maintenance Checklist.

2.7 CONDUCTIVITY
Verify that the dialysate conductivity measured by the internal cell in the machine
agrees with an external conductivity meter within 0.1mS/cm as follows:

1. Connect an external conductivity meter to the dialysate lines.

2. With the machine in Dialysis Mode and flow ON, compare the value shown
on the external conductivity meter with the conductivity shown on the
Display screen. They must be within 0.1mS/cm of each other.

3. Record the measured value on either the Quarterly Preventive Maintenance


Checklist or the Annual/4000 Hour Preventive Maintenance Checklist
depending upon which is being performed.

Page 12
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
2.8 TEMPERATURE
Verify that the actual dialysate temperature measured by an external temperature
meter agrees with the display screen within 0.3°C at 37°C and 39°C as follows:
1. Connect the dialysate lines to an external temperature meter.

2. Place the machine in Dialysis Mode with concentrate in the system. Clear
any alarms.

3. Select the Temperature button. The button label will change to Temp.
Setting. The value now displayed on this button is the temperature set
point. Adjust the temperature set point to exactly 37.0 then press the
[CONFIRM] key. The button will change back and now reads the actual
temperature of the dialysate again. Wait until this value stabilizes. It will
settle very close to the value set, depending upon inlet water temperature
and other conditions.

4. After the temperature of the dialysate stabilizes, compare the temperature


shown on the Temperature button with the temperature shown on the
external temperature meter connected to the dialysate lines. The two
readings must be within 0.3°C of each other.

5. Record the measured value on either the Quarterly Preventive Maintenance


Checklist or the Annual/4000 Hour Preventive Maintenance Checklist
depending upon which is being performed.

6. Repeat steps 3 and 4 with the temperature set to 39.0°C. Verify that the
actual temperature reported on the front panel display screen and the
external temperature meter is within 0.3°C of each other after the
temperatures stabilize again.

7. Record the measured value on either the Quarterly Preventive Maintenance


Checklist or the Annual/4000 Hour Preventive Maintenance Checklist
depending upon which is being performed.

Page 13
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
2.9 LEVEL DETECTOR

Caution: In the following steps, the level detector must be removed from the
machine cabinet. Before returning it to the cabinet, wipe the lower edge of the
module and the level detector cabinet opening to remove any residual
disinfecting agent. When installing the module screws do not use a power
screwdriver.

ALARM TEST
1. Perform the Level Detector calibration per the procedure outlined in the
2008K Calibration Procedures (part number 507296).
2. Place the machine in Dialysis Mode and turn the blood pump ON. Insert a
water filled venous chamber into the level detector and clear all alarms.
3. Position the Level Detector so you can watch the Channel 1 and Channel 2
LED's on the circuit board (See Figure 6 on page 16).
4. While watching the Channel 1 and Channel 2 LED's, remove the venous
chamber to create a blood alarm and close the occlusion clamp. Verify that
the Channel 1 LED lights first followed quickly by Channel 2.
OCCLUSION CLAMP TEST
1. Place a venous line into the closed occlusion clamp. Do not connect this
line to the venous chamber in the level detector.
2. Place the lower end of the venous line below the occlusion clamp in a
container of water positioned so that air escaping from the end of the line is
easily seen.
3. Connect a syringe and a pressure gauge to the venous line above the
occlusion clamp.
4. With the syringe, apply a pressure of at least 30psi (1550mmHg) to the
venous line while watching the end of the venous line in the water.
5. Verify that no air escapes from the venous line, indicating that the clamp is
fully occluding the line.

Page 14
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
RELAY CONTACT TEST
A relay prevents the level up and level down membrane keys on the face of the
level detector to function if they are pressed at the same time.
Various module configurations are available. Use the following information to
select the correct method to use to check the relay contacts:
• Use Method 1 if LP450 has LED’s D6 & D19 (See Figure 5)
• Use Method 2 if LP450 does not have relay test LED’s then locate a separate
relay board LP1026 with test LED’s D3 & D4 (See Figure 5).
• Use Method 3 if LP1026 is present and does not have LED’s D3 & D4
(See Figure 6).

Method 1:
1. With the machine in Dialysis Mode and the blood pump turned ON, insert a
water filled venous chamber into the level detector and clear all alarms
2. Locate the relay test LED’s D6 and D19 on the LP450 board.
3. To test the LED’s, press the level down switch on the face of the level
detector and verify that both LED’s D6 and D19 light. The on-board air
pump will also run. If both LED’s light, proceed to step 4.

Note: If either LED doesn’t light during the LED test (step 3), then the
results of the Relay Contact Test (step 4) are invalid. In this case use the
appropriate step below instead of step 4:
While in alarm condition (clamp closed), attach the ground lead of a voltmeter
to TP3 (ground) on the LP450 board. Measure the voltage on the solder side
of pins 1 and 2 of X153 on the LP450 board. Verify that both pins are 0 volts.
If voltage is present, the relay contacts are bad.

4. To test the relay contacts, remove the venous chamber to create a blood
alarm and verify that both LED’s D6 and D19 do not light when the level
down switch is pressed. If either LED lights, the relay contacts are bad.

Page 15
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
RELAY CONTACT TEST RELAY CONTACT
LED’S D19 & D6 ON LP450 TEST LED’S D3 & D4
BOARD WHEN LP1026 IS ON LP1026 BOARD.
NOT PRESENT

X153 ON LP450

Figure 5 - Level Detector with Relay Contact Test LED’s.

CHANNEL 1 LED
LIGHTS FIRST

CHANNEL 2 LED
LIGHTS SECOND

TP3 (GROUND)

X4 ON RELAY
BOARD LP1026

Figure 6 - Level Detector without Relay Contact Test LED’s.

Page 16
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
Method 2:
1. With the machine in Dialysis Mode and the blood pump turned ON, insert a
water filled venous chamber into the level detector and clear all alarms.
2. Locate the relay test LED’s D3 and D4 on the LP1026 board.
3. To test the LED’s, press the level down switch on the face of the level
detector and verify that both LED’s D3 and D4 light. The on-board air
pump will also run. If both LED’s light, proceed to step 4.

Note: If either LED doesn’t light during the LED test (step 3), then the
results of the Relay Contact Test (step 4) would be invalid. In this case use
Method 3 instead.

4. To test the relay contacts, remove the venous chamber to create a blood
alarm and verify that both LED’s D3 and D4 do not light when the level
down switch is pressed. If either LED lights, the relay contacts are bad

Method 3:
1. With the machine in Dialysis Mode and the blood pump turned ON, insert a
water filled venous chamber into the level detector and clear all alarms
2. Locate the solder side of X4 on the LP1026 (or X153 on the LP450).
3. Attach the ground lead of a voltmeter to TP3 (ground) on the LP450 board.
4. Test for 24 volts at X4 (X153) by pressing the level down switch on the face
of the level detector and verify that both pins 1 and 2 on the solder side of
X4 (X153) are 24 volts.
5. To test the relay contacts, remove the venous chamber to create a blood
alarm and measure the voltage on the solder side of pins 1 and 2 of X4
(X153). Verify that both pins are 0 volts. If voltage is present, the relay
contacts are bad.

Page 17
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
2.10 ALARM OPERATION AND PRESSURE HOLDING TESTS
Verify the automatic alarms produce the responses shown in Table 1 and the
machine passes the automatic pressure holding test as follows:

1. Place the dialysate lines in the shunt and close the door.

Warning! The use of a “test drip chamber” or “dummy drip chamber” must
never be used on the treatment floor. It must only be used in a controlled
technical environment.

2. Place a venous chamber filled with water in the holder on the level detector
module.
3. Place the machine in Dialyze mode and start the blood pump. Clear all
blood and water alarms.
4. Select the Test & Options button. On this screen press the Both Tests
button. Press the [CONFIRM] key to start.
5. Observe the machine stepping through the following alarm tests and ensure
that each alarm produces all of the responses shown in Table 1.

Page 18
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
Table 1. Alarm Tests.

ALARM TEST RESPONSE

1. Red visual alarm


Air Detector
2. Venous clamp closes
3. Blood pump stops

Blood Leak 1 Red visual alarm


2. Venous clamp closes
3. Blood pump stops

Arterial Pressure 1. Red visual alarm


2. Venous clamp closes
3. Blood pump stops

Venous Pressure 1. Red visual alarm


2. Venous clamp closes
3. Blood pump stops

TMP 1. Red visual alarm


2. Venous clamp closes
3. Blood pump stops

Passes if battery voltage is greater than 7.0


9-Volt Battery
volts under a load of 22Ω.

1. Lower venous alarm limit rises to 10mmHg


Optical Detector causing a venous pressure alarm
2. Venous clamp closes
3. Blood pump stops

1. Red visual alarm


Temperature 2. Yellow visual bypass
3. Machine in Bypass mode (no flow through the
dialysate flow indicator).

1. Red visual alarm


Conductivity 2. Yellow visual bypass
3. Machine in Bypass mode (no flow through the
dialysate flow indicator).

Page 19
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
6. When the tests shown in Table 1 are complete, the machine conducts a
pressure holding test. When the test ends verify that the display screen
reports TEST COMPLETE, indicating that all the tests were passed
successfully. Press the [RESET] key.
7. Test the UF pump integrity as follows:
• Pull the hydraulics out and remove the output tube from the UF pump.
• Install a 24” tube (part number 545325-10) to the empty output port.
Route this tubing out the back of the machine so that it will not be
kinked or pinched when the hydraulics is closed. Close the hydraulics
and place the other end of tube into a collection container.
• Conduct steps 1-6 again.
- If the pressure holding test fails, refer to Section 4.1.
- If the pressure holding test passes, disconnect the 24” tube and
reconnect the original output tube to the UF pump.

8. Test the audible alarm as follows:

• Place a piece of opaque paper inside the housing of the optical detector
to simulate a line containing blood. Close the door of the optical
detector.
• Open the shunt door. Verify that the machine responds with an
audible alarm. Press the [Mute] key and verify that the Mute lamp
lights, and the audible tone stops.

2.11 VERIFY pH
1. Place the machine into dialysis mode and allow it 10 minutes to stabilize.
2. Use a pH test strip (part number 335130-01 or equivalent) to verify the
dialysate fluid is between the ANSI/AAMI RD52 standard of 6.9 and 7.6.
3. Record the measured value on the Quarterly Preventive Maintenance
Checklist.

2.12 POWER FAILURE ALARM


Test the Power Failure alarm by turning the main power switch off on the back of
the power supply with the machine powered on. Verify that the audible alarm
sounds. If no audible alarm occurs, check/replace the 9-Volt battery.

Page 20
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
2.13 BLOOD PUMP ROTOR INSPECTION
Clean and inspect the arterial blood pump rotor assembly as follows:

1. Remove the blood pump rotor by opening the door, pulling out the crank
lever and turning the rotor 90 degrees.
2. Clean the rollers with a cloth dampened only with water.
3. With the crank lever pulled out, locate and lubricate the crank lever retainer
ball. See Figure 7 for assistance. If the crank lever is hard to pull out,
replace the crank lever retainer assembly (part number 564301).
4. Clean the inside of the blood pump housing with the damp cloth.
Compressed air may be used if it is available.
5. Locate the four (4) tubing guides. See Figure 7 for assistance. If any of the
four (4) tubing guides are missing, the rotor assembly needs to be replaced
(part number M30990).

Tubing Guides

Tubing Guide

Crank Lever Crank Lever


Retainer

Tubing Guide

SIDE VIEW TOP VIEW

Figure 7. Blood Pump Rotor Assembly

Page 21
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
Note: On newer blood pump rotors (M30990) the plastic sleeves are not
replaceable. If sleeves are damaged, replace the rotor assembly.

6. On each of the tubing guides there is a plastic sleeve that should rotate
smoothly. Inspect each plastic sleeve for smooth operation. If any of the
plastic sleeves are missing, damaged, or do not rotate smoothly, replace with
a new plastic sleeve. See Figure 8 for an exploded view and part numbers.

Solid Guide Sleeve Retaining


Post Screw P/N 552255

Plastic Sleeve
P/N 641256

Tubing Guide
(Exploded View)
Figure 8. Tubing Guide (Exploded View)

7. Without removing the plastic sleeve, inspect the solid guide post on all four
(4) tubing guides. The solid guide post should not be loose or bent. If any
of the solid guide posts are loose or bent, the rotor assembly needs to be
replaced. (part number M30990).

Note: The plastic sleeve will make it feel like the solid guide post is loose.
When inspecting, pay close attention to the movement of the metal shaft.

Page 22
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
2.14 FINAL CHECKS
Before returning the machine to clinical use after successful completion of all of the
Preventive Maintenance procedures listed above, complete the following:

• Verify that the machine label with serial number is in place, usually on the back
of the cabinet near the Monitor Control Unit or above the quick connectors of the
open shunt door assembly. Record this serial number on the Preventive
Maintenance Checklist form.

• Verify that no dialysate spills or leaks are visible in the hydraulics or on the
bottom of the cabinet. Clean and dry any spills found and correct the source.

• Verify that all cables are properly routed to prevent pinching or chaffing.

• Verify that all covers are replaced and that all cover screws and mounting
hardware has been replaced.

Caution: Reliable operation of the machine requires that all screws and
covers be properly installed. Ensure that all screws and covers are in place
before returning the machine to clinical use.

• Clean the exterior surfaces of the machine and remove all traces of dirt, oil or
other contaminants.

Caution: Do not use a cleaner containing Dimethyl Benzylammonium


Chloride. This ingredient will damage many plastic surfaces. Certain brands
of cleaners specifically marketed to clinics and hospitals contain this
ingredient. Check the contents of any unknown cleaner before using it.

Page 23
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
3.0 ANNUAL (4000 HOUR) PREVENTIVE MAINTENANCE
Perform the following Preventive Maintenance procedures every 12 months or
4000 hours of machine operation, whichever comes first. Perform the procedures in
the order given below to complete the Annual Preventive.
Make copies of the Annual Preventive Maintenance Checklist provided at the back of
this manual and use them to record the maintenance done.

Note: When performing an Annual Preventive Maintenance do not perform


the quarterly procedures first. The annual procedures will refer back to a
Section 2 quarterly procedure when necessary.

3.1 FILTERS AND O-RINGS


Perform the Quarterly Preventive Maintenance procedures in Sections 2.1.

3.2 PRE-UF PUMP FILTER


Replace the Pre-UF Pump Filter (part number 672574).

Caution: Do not attempt to disassemble the Pre-UF Pump Filter. If not


properly reassembled, the Pre-UF Pump filter may leak. A leak in the
hydraulic system at this location may affect the operation of the machine or
cause fluid loss from the patient.

3.3 UF PUMP CHECK VALVES

Caution: Ensure each check valve is oriented correctly to allow fluid flow in
the proper direction.

Replace the UF Pump Output check valves. There are two check valves at the UF
pump output. One is at the UF pump itself. The other is downstream at the UF
Sample Port.

Page 24
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
3.4 DEAERATION RESTRICTOR

Note: On newer machines with hydrochamber assemblies the deaeration


restrictor is a glass bead with a fixed orifice built into one of the hydrochamber
interconnecting tubes. Since there are no o-rings in this type of setup,
disregard this step. If a screw is present in the back side of the hydroblock
proceed with the following check.

With the machine off, replace the deaeration restrictor O-rings by clamping the
hydroblock vent tube and removing the deaeration restrictor from the hydroblock (See
Figure 3, pg. 8). Replace the two O-rings on the deaeration restrictor (part number
F40007100 or 579070). Clean any debris that may be present in the angled hole at the
tip of the deaeration restrictor. Reinstall deaeration restrictor.

3.5 DIASAFE FILTER (IF APPLICABLE)


Perform the Quarterly Preventive Maintenance procedures in Sections 2.3.

3.6 DIAPHRAGM PUMPS


Rebuild the UF, Bicarbonate and Concentrate diaphragm pumps as described in
Section 4.1 and 4.2.

Page 25
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
3.7 HEATER ELEMENT

Shock Hazard: Dangerous high voltage is present at these connections


when the machine is operating. Ensure the machine is disconnected from the
wall outlet.

Caution: The heater element is hot enough to inflict serious injury if it is


touched while power is applied or shortly after power is removed. Ensure the
machine is disconnected from the wall outlet.

1. Remove the rear access panels and distribution board cover so that the wires of
the heater are accessible.

Note: If the heater element is replaced, replace the heater O-ring as well
(part number F40007103 or 579075).

2. Remove the heater element from the hydroblock on the machine (See Figure 3,
pg. 8). Inspect the heater element for signs of corrosion. If corrosion exists,
replace the heater element.
3. Attach one lead of the voltmeter to the ground (yellow / green wire) and the
other lead to the brown or blue wire. Measure the resistance. Resistance should
be “ OL “ or greater than 19.9 meg-ohms. If this is not the case, replace the
heater element.
4. After completing this check, reinstall the heater element and O-ring.

3.8 HIGH VOLTAGE AC CONNECTIONS


Perform the Quarterly Preventive Maintenance procedures in Sections 2.4.

Page 26
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
3.9 DEAERATION MOTOR BRUSHES
Replace the deaeration motor brushes every 8000 hours using the following steps:
1. Remove the complete deaeration motor/pump assembly from the machine.
2. Using a marker, make marks on the motor housing and motor cap as
illustrated below. These marks will ensure the correct alignment when
reassembling the motor case.

Use Marker
Use marker to make hatch
marks
marks to ensure
to ensure proper
proper
reassembling
reassembling of motor
of motor

Motor Cap

Figure 9. Deaeration Motor/Pump Assembly with marks.

3. Using a T-25 Torx screwdriver, remove the two screws holding the motor
cap to the motor housing.
4. Remove the motor cap to gain access to the motor brushes. Replace the
motor brushes.
5. Reassemble the motor, aligning the marks.
6. While holding the pieces together, install and tighten the two T-25 screws.

Note: If the motor case and motor cap are not installed using the marks, the
motor will run in reverse and flow errors will result.

Page 27
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
3.10 INLET WATER PRESSURE REGULATOR
Verify that the regulator provides the proper water pressure as follows:

1. Shut off the water supply to the machine.


2. Install a pressure gauge with a T-fitting to monitor the pressure at the outlet
side of the Inlet Pressure Regulator (See Figure 3, pg. 8).

Caution: Use tie wraps or tubing clamps to secure the connections. The
water pressure may be sufficient to blow the lines off the fittings if they are not
secured.

3. Turn the water supply to the machine ON.


4. With the dialysate lines in the shunt, select Dialysis Mode. The pressure gauge
will cycle between two readings as the water inlet valve opens and closes.
5. Allow the water inlet valve to open and close for approximately 30 seconds to
clear any air.
6. Once the air is gone, verify the pressure gauge reads between 18 and 20psi
when the pressure is at its highest value, and reads greater than 8psi when the
pressure is at its lowest value.

Note: Readings below 8psi at the lowest level indicates inadequate inlet
water flow into the machine. This may be caused by a dirty inlet filter screen
or problems with the treated water supply.

7. Record the measured value on the Annual/4000 Hour Preventive


Maintenance Checklist.
8. Turn the dialysate flow off and the pressure gauge will stop cycling and will
show only a single pressure.
9. Note this single pressure reading on the pressure gauge.
10. After 15 minutes, check for any pressure increase on the pressure gauge.
11. If the pressure reading on the pressure gauge increases, the Inlet Pressure
Regulator is leaking internally and needs to be rebuilt or replaced
Part Number
Complete Inlet Pressure Regulator: 565048
Rebuild Kit for the Inlet Pressure Regulator: 190934

12. If the Inlet Pressure Regulator passes the internal leakage test, turn the
machine OFF, turn OFF the water supply, remove the pressure gauge and
reconnect the tubing using clamps to prevent leaks.
13. Turn the treated water supply source ON, turn the machine ON and select
Dialysis Mode. Start dialysate flow and inspect all hoses and connections.
Ensure that there are no leaks.

Page 28
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
3.11 ONLINE CLEARANCE TEST (IF APPLICABLE)
If the machine is equipped with Online Clearance (OLC), conduct the following test:
OLC Self-Test
1. With stable temperature and conductivity, start an OLC Self-Test by
pressing the OLC Self-Test screen button on the Kt/V screen.

Note: Switch to the debug screens by pressing and holding the [▲] and [▼]
keys at the same time for approximately 1 sec. The main display will change
indicating that the machine is in debug mode.

2. When the OLC Self-Test is complete, go to debug screen #5 and confirm that
0 Test (0TST prior to functional software version 3.02) is in the range ±20.

Note: If 0 Test (0TST) is out of range, the OLC Self-Test will fail. If this
happens, conduct temperature (PRE & POST) and conductivity calibrations,
then conduct OLC Self-Test again.

OLC Self-Test Troubleshooting


If the OLC Self-Test continues to fail after conducting the temperature (PRE &
POST) and the conductivity calibrations, conduct the following troubleshooting steps:
1. Turn the machine ON and allow the machine to run in Dialysis Mode at
500ml/min flow rate for approximately 10 minutes. This allows the machine to
come up to temperature and conductivity.

Note: On functional software prior to version 3.02, an OLC Self Test will
start automatically as soon as temperature and conductivity are stable. Once
the test starts cancel it by opening the shunt door. The cancel message will be
hidden by the Shunt Door Open banner. Close the shunt door and confirm
that OLC Self Test banner is gone.

2. Once the machine is warmed up, remove X3 (MON-NTC) and X7 (COND)


from the distribution board.
3. Using the Resistor Plug Set for OLC Testing (part number 190168), plug the
6.04KΩ plug into X3 (MON-NTC) and the 274Ω plug into X7 (COND) on the
distribution board.

Note: Due to the resistor placement inside the OLC test plugs, the orientation
of the plug when inserted onto the distribution board is irrelevant.

Page 29
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
4. Switch to the debug screens by pressing and holding the [▲] and [▼] keys at
the same time for approximately 1 sec. The main display will change
indicating that the machine is in debug mode.
5. Use the [▲] and [▼] keys to go to debug screen #5.
6. On debug screen #5, locate TPRE and observe its value for 1 minute. During
this time, record the highest and lowest values observed in the table below.

Highest Value Lowest Value Difference

TPRE - =

7. Calculate the difference between the highest and lowest TPRE value and
record it in the table above.
8. Verify that the difference for TPRE < 2. If the difference for TPRE > 2, then
possible causes are sensor board, motor noise, or -12V problems.
9. Again, on debug screen #5, locate CPRE and observe its value for 1 minute.
During this time, record the highest and lowest values observed in the table
below.

Highest Value Lowest Value Difference

CPRE - =

10. Calculate the difference between the highest and lowest CPRE value and
record it in the table above.
11. Verify that the difference for CPRE < 8. If the difference for CPRE > 8,
then possible causes are sensor board, motor noise, or -12V problems.
12. Remove the 6.04KΩ and 274Ω plug from the distribution board and
replace the original connectors.
13. Remove X44 (NTC-POST) and X13 (COND-POS) from the distribution
board.
14. Using the Resistor Plug Set for OLC Testing (part number 190168),
plug the 6.04KΩ plug into X44 (NTC-POST) and the 274Ω plug into
X13 (COND-POS) on the distribution board.

Page 30
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
15. On debug screen #5, locate TPOS and observe its value for 1 minute.
During this time, record the highest and lowest values observed in the
table below.

Highest Value Lowest Value Difference

TPOS - =

16. Calculate the difference between the highest and lowest TPOS value and
record it in the table above.
17. Verify that the difference for TPOS < 2. If the difference for TPOS > 2, then
possible causes are sensor board, motor noise, or -12V problems.
18. Again, on debug screen #5, locate CPOS and observe its value for 1 minute.
During this time, record the highest and lowest values observed in the table
below.

Highest Value Lowest Value Difference

CPOS - =

19. Calculate the difference between the highest and lowest CPOS value and
record it in the table above.
20. Verify that the difference for CPOS < 8. If the difference for CPOS > 8,
then possible causes are sensor board, motor noise, or -12V problems.

Note: In the following step, conduct the test for the dialysate flow ON to
OFF transition.

21. With the 6.04KΩ plug on X44 (NTC-POST) and the 274Ω plug on X13
(COND-POS) of the distribution board, verify that CPOS does not change
more than 10 increments when the dialysate flow is turned OFF. (Note:
Conduct this test for the dialysate flow ON to OFF transition.)
22. Remove the 6.04KΩ and 274Ω plug from the distribution board and
replace the original connectors.
23. With stable temperature and conductivity, start an OLC Self Test by
pressing the OLC Self-Test screen button on the Kt/V screen.
24. When the OLC Self Test is complete, go to debug screen #5 and confirm that
0 Test (0TST prior to functional software version 3.02) is in the range ±20.

Page 31
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
3.12 DEAERATION AND LOADING PRESSURE
Verify that the deaeration pressure is between –24 and -25 inHg and the loading
pressure is between 18 and 20psi (between 23 and 25psi if a DIASAFE filter
system is installed) as follows:

1. With dialysate flow OFF, install a pressure gauge with a T-fitting to monitor the
pressure on the inlet (suction) side of the deaeration pump (See Figure 3, pg. 8).
2. Connect a pressure gauge equipped with a yellow connector into the red
ACETATE/ACID port.
3. In Dialysis Mode, turn dialysate flow ON and verify that the pressure gauge
on the deaeration pump indicates between -24 and -25 inHg. The needle
will be vibrating somewhat. Verify that it does not go higher than -24 inHg
or lower than -25 Hg. Verify that the pressure gauge in the
ACETATE/ACID port indicates between 18 and 20psi (between 23 and
25psi if a DIASAFE filter system is installed).
4. Record the measured values on the Annual/4000 Hour Preventive
Maintenance Checklist.

Note: When the machine is at a different elevation above sea level, it may be
difficult or impossible to achieve -24inHg. The following table will help in
determining the appropriate deaeration pressure calibration point at different
elevations:
Table 2

Approx. Minimum target deaeration


Elevation atmospheric pressure relative to
pressure atmospheric pressure
feet mmHg inches of Hg
0 760 -24.0
1000 728 -23.0
2000 697 -22.0
3000 667 -21.0
4000 639 -20.0
5000 612 -19.0
6000 585 -18.5
7000 561 -17.5
8000 537 -16.9
9000 514 -16.2
10000 492 -15.5

Page 32
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
3.13 FLOW RELIEF PRESSURE
Verify that the flow relief pressure is between 29 and 30psi (between 35 and 36psi if a
DIASAFE filter system is installed) as follows:

1. Turn the machine on in Service mode. From the Calibrate Hydraulics


screen, select the Flow Pressure screen button. The screen will change to
the following:

2. Connect a pressure gauge in line at the output of the flow pump (See Figure 3,
pg. 8).

Note: The output side of the flow pump is the side with the white reinforced
jacket over the line. The input side has clear plastic line.

3. Press the [CONFIRM] key to start the flow pump.


4. Verify the pressure gauge indicates a pressure between 29 and 30psi
(between 35 and 36psi if a DIASAFE filter system is installed).
5. Record the measured value on the Annual/4000 Hour Preventive
Maintenance Checklist

Page 33
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
3.14 CONCENTRATE AND BICARB PUMPS
Test the Concentrate and Bicarbonate diaphragm pumps as described in Section 4.3.

3.15 UF PUMP
Calibrate the UF Pump Volume (Refer to the 2008K Calibration Procedures –part
number 507296.)

Record the measured value on the Annual/4000 Hour Preventive Maintenance


Checklist.

3.16 CONDUCTIVITY
Verify that the dialysate conductivity measured by the internal cell in the machine
agrees with an external conductivity meter within 0.1mS/cm as described in Section
2.7.

3.17 TEMPERATURE
Verify that the dialysate temperature shown on the front panel Display screen agrees
with an external temperature meter within 0.3°C at 37°C and 39°C as described in
Section 2.8

3.18 VOLT HI LO DETECT


Verify that the 5-volt supply is operating within 0.2 volts as reported on the debug
screen as follows:

1. Switch to the debug screens by pressing and holding the [▲] and [▼] keys at
the same time for approximately 1 sec. The main display will change
indicating that the machine is in debug mode.
2. Use the [▲] and [▼] keys to go to debug screen #4.
3. Locate the 5V value on the display screen. This value must be between
4.8V and 5.2V.

Note: If the 5V value is not within limits, the problem is most likely the
12volt supply out of tolerance. Perform the Volt Hi Lo Detect calibration
procedure. Refer to the 2008K Calibration Procedures –part number 507296.

4. Record the measured value on the Annual/4000 Hour Preventive


Maintenance Checklist

Page 34
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
3.19 BLOOD LEAK AND DIMNESS
Verify that the blood leak level is between 4.5 and 5.2 volts and the blood dimness
level is within 5.0 ±1.0 volts as follows:

1. Switch to the debug screens by pressing and holding the [▲] and [▼] keys at
the same time for approximately 1 sec. The main display will change
indicating that the machine is in debug mode.
2. Use the [▲] and [▼] keys to go to debug screen #4.
3. Locate the LEAK value on the display screen. This value must be between
4.5V and 5.2V.
4. Locate the DIMN value on the display screen. This value must be between
4.0V and 6.0V.

Note: If these values are outside the limits given, perform a bleach rinse to
clean the glass detector tube inside the machine before considering other
repairs.

5. Record the measured values on the Annual/4000 Hour Preventive


Maintenance Checklist.

Page 35
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
3.20 ARTERIAL, VENOUS AND TRANSMEMBRANE PRESSURE
ARTERIAL PRESSURE
1. Inspect the internal pressure transducer protector for contamination. If
contamination is found, replace the pressure transducer protector (part
number M30971) and remove and disinfect the pressure port with 1:100
bleach for a minimum of 15 minutes.

2. Open the arterial transducer port PART. On the blood pump module to air
(atmospheric pressure). Verify that the Arterial Pressure bargraph
indicates 0.

3. Attach a syringe and a calibrated pressure meter to the PART. Port using a
T-fitting.

4. Push the syringe in to show a pressure of 200mmHg on the external pressure


meter. Verify that the Arterial Pressure bargraph indicates 200.

5. Switch to the debug screens by pressing and holding the [▲] and [▼] keys at
the same time for approximately 1 sec. The main display will change
indicating that the machine is in debug mode.

6. Use the [▲] and [▼] keys to go to debug screen #1 and locate the ART value.

7. Increase the pressure to a range of 310 – 315mmHg. Clamp off the tubing
at the PART. Port and monitor the debug ART value for 30 seconds. The
maximum allowable leakage is 2mmHg in 30 seconds.

Page 36
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
VENOUS PRESSURE
1. Inspect the internal pressure transducer protector for contamination. If
contamination is found, replace the pressure transducer protector (part
number 650158) and remove and disinfect the pressure port with 1:100
bleach for a minimum of 15 minutes.

2. Open the venous transducer port PVEN. On the level detector module to air
(atmospheric pressure). Verify that the Venous Pressure bargraph
indicates 0.

3. Attach a syringe and a calibrated pressure meter to the PVEN. Port using a
T-fitting.

4. Push the syringe plunger in to show 400mmHg on the external pressure


meter. Verify that the Venous Pressure bargraph indicates 400.

5. Switch to the debug screens by pressing and holding the [▲] and [▼] keys at
the same time for approximately 1 sec. The main display will change
indicating that the machine is in debug mode.

6. Use the [▲] and [▼] keys to go to debug screen #1 and locate the VEN value.

7. Decrease the pressure to a range of 310 – 315mmHg. Clamp off the tubing
at the PVEN. Port and monitor the debug VEN value for 30 seconds. The
maximum allowable leakage is 2mmHg in 30 seconds.

Page 37
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
TRANSMEMBRANE PRESSURE
1. With dialysate flow ON, ensure the dialysate lines are full of fluid and no
air is visible passing through the flow indicator.
2. Hang a four-way connector to the I.V. pole at normal dialyzer height.
3. Set dialysate flow to 500ml/min, then press the [CONFIRM] key.
4. Open the arterial (PART.) and venous (PVEN.) transducer ports to air
(atmospheric pressure).
5. Turn the blood pump off.
6. Turn dialysate flow OFF, remove the dialysate lines from the shunt and
attach them to the connector. Close the shunt door.

Note: Filling the lines before removing them from the shunt will avoid
wetting the pressure meter transducer during the test.

7. Connect a 30cc syringe to one of the four-way connector outlets and a


calibrated pressure meter to the remaining outlet.
8. Switch to the debug screens by pressing and holding the [▲] and [▼] keys
at the same time for approximately 1 sec. The main display will change
indicating that the machine is in debug mode.
9. Use the [▲] and [▼] keys to go to debug screen #1.
10. Use the syringe to set the pressure on the external pressure meter to
0mmHg. Calculate the dialysate pressure measured by the machine as
follows:
• Note the value shown for TMP on the display screen.
• Subtract the value shown for VEN on the display screen.
11. Verify that the calculated value is between +20 and -10mmHg.
12. Use the syringe to create a pressure of -250mmHg on the external pressure
meter.
13. Calculate the dialysate pressure measured by the machine again as
described in step 10. Verify that the calculated value is between -
230 and -260mmHg.

Page 38
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
3.21 DIALYSATE FLOW
Verify the dialysate flow is within ±3% of the stated rate at 500ml/min as follows:

Note: All flow rates are controlled by the software. Testing the rate at
500ml/min verifies the accuracy of all rates.

1. In Dialysis Mode, turn the dialysate flow ON at 500ml/min. Verify that the
UF pump is OFF and the machine is out of bypass. Allow the dialysate flow
to run for 2 minutes, minimum, before continuing.

Note: Do not collect the spent dialysate from the drain hose in the next step.
The accuracy of the collection will be affected if not collected directly from
the drain port with the drain hose removed.

2. Collect spent dialysate from the drain port on the back of the machine (See
Figure 3, pg. 8) for exactly 1 minute. Verify that the amount collected is
between 485 and 515ml.
3. Record the measured value on the Annual/4000 Hour Preventive
Maintenance Checklist

3.22 HEPARIN PUMP


If the machine is equipped with a heparin pump module, clean and test the pump as
follows:
1. Remove the heparin pump module from the machine and wipe the lower
edge of the module and the heparin pump cabinet opening to remove any
residual disinfecting agent.
2. Reinstall the heparin pump into the machine cabinet. When installing the
module screws do not use a power screwdriver
3. Place the machine in Dialysis Mode and clear all alarms.
4. Select the Heparin screen button. On this screen press the Syringe button.
Use the [▲] and [▼] keys to select the syringe being used for this test and press
the [CONFIRM] key.
5. Press the Load Syringe button. Press the [CONFIRM] key and the carriage
will fully retract downward.
6. When the carriage has stopped, press the [Escape] key and clean the
carriage bar (See Figure 10).

Page 39
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
7. Extend the syringe plunger past the 10ml mark. Install the syringe into the
pump and latch the plunger handle into the carriage. If needed, squeeze the
carriage/plunger release lever to slide the carriage up to meet the plunger
handle. The syringe plunger should be at a position greater than 10ml.

SYRINGE BODY HOLDER

CARRIAGE BAR
PLUNGER HOLDER
CARRIAGE

CARRIAGE / PLUNGER
RELEASE LEVER

Figure 10. Heparin Pump Module.

8. Press the Heparin Prime button. Press and hold the [CONFIRM] key. The
plunger will start moving towards the 10ml mark. Release the [CONFIRM]
key as soon as the syringe plunger is aligned exactly with the 10ml mark on
the side of the syringe body.
9. Select the Bolus data button. Use the [▲] and [▼] keys to set a value of
5.0ml then press the [CONFIRM] key.
10. Select the Total Infused data button. Press the 0 (zero) key and then the
[CONFIRM] key.
11. Press the Infuse Bolus button. Press the [CONFIRM] key and start timing
the interval with the stopwatch.

Page 40
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
12. When the pump stops, use the table below to verify the time to run a 5ml
Bolus is correct for the selected syringe type.

Type Syringe Size 5ml Bolus Time


B-D 10ml 29.0 – 30.0 seconds
Terumo 10ml (12ml) 24.0 – 25.0 seconds
Monoject 10ml (12ml) 24.5 – 25.5 seconds

13. Also, verify that the Total Infused data button reads 5.0 and that the
syringe plunger has moved to between 4.8 to 5.2 on the syringe scale.
14. Move the carriage all the way up to the end of its travel.

Note: Use the Heparin Pump Prime function to move the carriage to the end
of travel to ensure the screw gear stays fully engaged for the end of travel
alarm test.

15. Select the RATE data button. Use the [▲] and [▼] keys to set a value of
9.9ml/hr. then press the [CONFIRM] key.
16. Press the [Heparin on/off] key to start the pump. The green lamp above
the key should light.
17. Verify that a HEPARIN PUMP ALARM is displayed on the status line
within 2 minutes.

Note: If the machine is not alarm free, another message of higher priority
may be displayed on the status line.

Page 41
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
3.23 BLOOD PUMP
Clean and test the arterial blood pump module as follows:
1. Remove the blood pump rotor by opening the door, pulling out the handle
and turning the rotor 90 degrees
2. Clean the rollers with a cloth dampened only with water.
3. With the crank lever pulled out, locate and lubricate the crank lever retainer
ball. If the crank lever is hard to pull out, replace the crank lever retainer
assembly (part number 564301).
4. Without removing the plastic sleeve, inspect the solid guide post on all four
(4) tubing guides. The solid guide post should not be loose or bent. If any
of the solid guide posts are loose or bent, the rotor assembly needs to be
replaced. (Part number M30990).

Note: The plastic sleeve will make it feel like the solid guide post is loose.
When inspecting, pay close attention to the movement of the metal shaft.

5. Clean the inside of the blood pump housing with the damp cloth.
Compressed air may be used if it is available.
6. Remove the blood pump module from the machine and inspect the motor
gearbox. In heavy use, some oil may accumulate on the gearbox housing.
Wipe the housing clean.
7. Wipe the lower edge of the module and the blood pump cabinet opening to
remove any residual disinfecting agent.
8. Reassemble the blood pump module and reinstall it in the machine cabinet.
When installing the module screws do not use a power screwdriver.
9. Place the machine in Dialysis Mode and clear all blood alarms.
10. Start the blood pump if it is not already running.
11. Open the blood pump door. Verify that the red alarm light on the blood
pump module lights within 15 or 30 seconds.

Note: Delay time before the blood pump alarm lights is set by dipswitch 4 on
the LP955 blood pump board.

Page 42
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
12. Close the blood pump door and press the [RESET] key to clear the alarm.
13. Test the pump as follows:
• Set the Tubing Size to 8mm.
• Insert a bloodline in the pump. Do not change the setting of the
tubing size selector, even if you are not using 8mm tubing.
• Connect a pressure gauge to the bloodline in the pump (See
Figure 11) and allow the rotor to pull up 37°C ± 1.5°C water. Let
this fluid flow past the pressure gauge long enough to clear out
air.

Clamp

Figure 11. Blood Pump Rotor Occlusion Test

• With the pump running at 600ml/min, clamp the outlet so that the
pressure gauge is between the output of the pump and the clamp
(See Figure 11). The peak pressure on the gauge must be between
25 and 35psi. If the peak pressure is out of range, replace the
blood pump rotor springs (part number 650174).
• Set the Tubing Size Selector to agree with the size of bloodline
used in the pump.

Page 43
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
3.24 LEVEL DETECTOR
Perform the Quarterly Preventive Maintenance procedures in Section 2.9

3.25 ALARM AND PRESSURE HOLDING TESTS


Perform the Quarterly Preventive Maintenance procedures in Section 2.10

3.26 VERIFY pH
1. Place the machine into dialysis mode and allow it 10 minutes to stabilize.
2. Use a pH test strip (part number 335130-01 or equivalent) to verify the
dialysate fluid is between the ANSI/AAMI RD52 standard of 6.9 and 7.6.
3. Record the measured value on the Annual/4000 Hour Preventive
Maintenance Checklist.

Page 44
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
3.27 RINSE CHECKS
Perform the Rinse Checks as follows:
1. Place the machine in rinse mode, then turn the water supply off. Verify that
the display screen shows NO WATER.
2. Turn water supply on.
3. Start rinse mode, and watch the flow from the drain line. The water from
the drain line will stop at one point for 15 seconds. At this time put the
drain line in a 1000ml graduated cylinder. Start timing when the flow starts
again. At 30 seconds, remove the line from the graduated cylinder. Verify
that a minimum of 310ml is collected.
4. Record the measured value on the Annual/4000 Hour Preventive
Maintenance Checklist.
5. To avoid problems with pre-rinse, allow the rinse to complete prior to
proceeding to the next step.
6. Open the shunt door and remove the dialysate lines.
7. Connect the dialysate lines to an external temperature meter capable of
reading at least 85°C.
8. Close the shunt door.
9. Start the machine in heat disinfect mode with the dialysate lines on the
external temperature meter as follows:
• With the shunt door closed, reach up into the opening at the bottom of the
shunt and find the two spring-loaded switches that protrude through the
holes next to the connectors. Push these switches in with your fingers,
simulating dialysate lines connected to the shunt.
• While holding these switches pushed in, press the Heat Disinfect
button on the startup screen.
• While holding the switches, open the shunt door. Verify the heat
disinfect does not get interrupted.
• Leave the shunt door open; closing it will stop the heat disinfect mode.

Page 45
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
Warning! Scalding hot water is passing through the dialysate lines and the
external meter manifold. Allow the dialysate lines and the external
temperature meter manifold to cool before disconnecting the lines.

10. The heat disinfect timer will not be operating. It is not needed. Monitor the
external temperature meter and verify that the temperature rises to between
80°C and 90°C.

Note: Some external temperature meters may not display correctly at higher
temperatures. Consult the appropriate product literature from the meter
manufacturer for applicable conversion charts.

11. Record the measured value on the Annual/4000 Hour Preventive


Maintenance Checklist.

3.28 POWER FAILURE ALARM AND BATTERY REPLACEMENT


Remove the 9-volt battery located behind the monitor unit (See Figure 3, pg. 8). The
battery is inside the plastic housing. Press the cover in and to the left to release it,
then pull it out.

Note: If the 9-volt battery is swollen, the power logic board is bad and should
be replaced.

With the battery removed, perform an Alarm Test and confirm the Battery Test fails.

Note: If the Battery Test passes with the 9-Volt battery removed, IC9 is bad
on the Power Logic Board and should be replaced.

Replace the 9-volt battery with a new one observing the correct polarity marked
inside the holder. Install the battery/holder assembly into the monitor unit.

With the new battery in place, perform an Alarm Test and confirm the Battery Test
passes.

Test the power failure alarm by removing the machine power plug from the wall
outlet with the machine powered on. Verify that the audible alarm sounds.

Page 46
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
3.29 BLOOD PRESSURE MODULE

Note: There are seven types of blood pressure modules that can be connected
to the 2008K. Since each module type has different test criteria and the 2008K
is unable to detect which module is connected, disregard the displayed
Pass/Fail results if present and use the appropriate table to determine pass or
fail.

Test the blood pressure module (See Figure 3, pg. 8), as follows:
Place the machine in Service Mode.
Select the Test BP Module button and the screen will change.
INFLATION SPEED TEST
From the Test BP Module screen, select the Inflation Speed button.
1. Connect the 10 foot blood pressure tubing from the module to the 700cc
port on the Dummy Cuff. Ensure the tubing fits snugly on the ports.
2. Press the [CONFIRM] key. The screen will change and the test will start.
3. The blood pressure module will pressurize the Dummy Cuff and report the
time. When the test is complete, use the table below to verify that the Time
to 250 value is within range dependent upon the module type installed.

Module Type Time to 250

M400 ≤ 7.3 seconds

M2000, M2600 or M3600 ≤ 10.0 seconds

TM-2910 or TM-2915 ≤ 10.0 seconds


SunTech ≤ 11.0 seconds

4. Press the [CONFIRM] key.

Page 47
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
DEFLATION SPEED TEST

Note: The deflation speed test is not applicable to the SunTech module.

From the Test BP Module screen, select the Deflation Speed button.
1. Connect the 10 foot blood pressure tubing from the module to the 220cc
port on the Dummy Cuff. Ensure the tubing fits snugly on the ports.
2. Press the [CONFIRM] key. The screen will change and the test will start.
3. The blood pressure module will depressurize (deflate) the Dummy Cuff. The
screen will change and report the deflation speed at various pressures in mmHg
per second.
4. Use the table below to verify that the values reported are within the
following ranges dependent upon the module type installed.

Pressure RATE AT 240 RATE AT 160 RATE AT 80 RATE AT 40


(mmHg/sec) (mmHg/sec) (mmHg/sec) (mmHg/sec)
Model Type
Min Max Min Max Min Max Min Max
M400 4.3 6.2 4.6 6.2 3.3 5.2 2.6 4.4

M2000 or M2600 4.4 6.7 4.4 6.7 3.0 5.2 2.4 4.6

M3600 3.8 6.6 3.8 6.6 2.8 5.5 2.3 4.6

TM2910 or TM-2915 3.5 7.5 3.5 7.5 3.5 7.5 3.5 7.5

SunTech N/A N/A N/A N/A N/A N/A N/A N/A

5. Press the [CONFIRM] key.

AIR LEAKAGE TEST


From the Test BP Module screen, select the Air Leakage button.
1. Connect the 10 foot blood pressure tubing from the module to the
appropriate module specific port on the Dummy Cuff (see table at the top of
the next page). Ensure the tubing fits snugly on the ports.
2. Press the [CONFIRM] key. The screen will change and the test will start.
3. The blood pressure module will pressurize the Dummy Cuff as indicated by
a rise in the Cuff Pressure displayed to approximately 290mmHg. An
internal timer is then activated automatically. Any change in the pressure
being held will show in the Cuff Pressure display.
4. At the end of the Air Leakage test, the pressure is automatically released,
the screen will change and the Leak Rate in mmHg is displayed.

Page 48
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
5. Use the table below to verify that the Leak Rate value is within range
dependent upon the module type installed.
Module Type Dummy Cuff Port Leak Rate
M400, M2000 or M2600 220cc ≤15mmHg
M3600 220cc ≤12mmHg
TM-2910 or TM-2915 700cc ≤30mmHg
SunTech 700cc ≤22mmHg

6. Press the [CONFIRM] key.

CALIBRATION CHECK

For the M400, TM-2910 & TM-2915


1. From the Test BP Module screen, select the Calibrate Mode button.
2. Disconnect the 10 foot blood pressure tubing from the Dummy Cuff and
connect it to a calibrated pressure meter accurate to within ±1mmHg.
3. Press the [CONFIRM] key. The screen will change and the blood pressure
module will pressurize the line.
4. When the pressure shown on the display screen and on the external pressure
meter stabilizes, verify that they agree within ±3mmHg.

Caution: Do not exceed 330mmHg in the following step. The blood


pressure module may be damaged if this pressure is exceeded.

5. Over Pressure Relief Check:


Remove the hose from the external pressure meter and connect it to a large
syringe (60cc) that has its plunger pulled back. Use the syringe to over-
pressurize the blood pressure module while watching the display carefully.
Use the table below to verify that the pressure is automatically released
within range dependent upon the module type installed.

Module Type Over Pressure Release Range


M400 between 320 and 330mmHg
TM-2910 and TM-2915 between 310 and 330mmHg

Page 49
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
For the M2000, M2600, M3600 & SunTech
1. From the Test BP Module screen, select the Air Leakage button.
2. Disconnect the 10 foot blood pressure tubing from the blood pressure module.
3. Using a T-fitting, connect the blood pressure module, calibrated pressure
meter accurate to within ±1mmHg and to the appropriate module specific port
on the Test Device. (See the table below). Ensure the tubing fits snugly at all
connections.
Module Type Dummy Cuff Port
M2000, M2600 or M3600 220cc
SunTech 700cc

4. Press the [CONFIRM] key. The screen will change and the blood pressure
module will pressurize the line.
5. The blood pressure module will pressurize as indicated by a rise in the Cuff
Pressure displayed to approximately 290mmHg. An internal timer is then
activated automatically.
6. When the pressure shown on the display screen and on the external pressure
meter stabilizes, verify that they agree within ±3mmHg.
7. At the end of the Air Leakage test, the pressure is automatically released.
8. Disconnect the tubing from the blood pressure module and reconnect the
10 foot blood pressure tubing
9. Over Pressure Relief Check:
Caution: Due to the microprocessor control of the M2000, M2600, M3600
and SunTech module, the over-pressure relief check does not need to be
conducted. An over-pressure relief test could permanently damage the
module.

3.30 FINAL CHECKS


1. Check the machine for any optional components that are not included here.
Perform the preventive maintenance tests provided with those units.
2. Perform the Quarterly Preventive Maintenance procedures in Section 2.14.

Page 50
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
4.0 REBUILDING THE DIAPHRAGM PUMPS
The machine contains two types of diaphragm pumps. The ultrafiltration (UF) pump
is a solenoid coil type. The concentrate and bicarbonate pumps are stepper motor
types.

4.1 REBUILDING THE ULTRAFILTRATION PUMP


Figure 12 shows an exploded view of the ultrafiltration pump. When working on the
pump, be especially careful to do the following:

• Do not lose the wear button or the shim washers inside the pump solenoid.

• Count the shim washers when you disassemble the pump. The number of
washers varies as needed to mate the solenoid to the pump properly.

Caution: Replace exactly the same number of shims in the ultrafiltration


pump as were removed. The pump will not operate correctly if the same
number of shims are not replaced.
Examine all components for signs of corrosion. Replace any components that
show signs of excessive rust.

Note: If the UF pump does not provide output even though the green light is
on and not flashing, inspect the check valves. If a check valve is installed
backwards, the pump will not work.

Page 51
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
WEAR BUTTON OUTLET SPRING
P/N 565219 P/N 563025
(STRONGER)
SHIM
WASHERS SEALS O-RINGS
P/N 562343 P/N 565279 P/N F40007099 or 579058

SPRING RETAINING CLIP INLET SPRING SPACER, P/N 566149


IS POSITIONED FOR P/N 565278 (ON SOME PUMPS THE SPACER
MAXIMUM COMPRESSION (WEAKER) MAY BE MOULDED INTO THE
OUTLET NOZZLE.)

Figure 12. Ultrafiltration Pump Exploded View.

Rebuild the ultrafiltration pump as follows:

1. Remove the UF pump from the machine (See Figure 3, pg. 8). Make a note
of which line goes to which port on the pump. When looking at the end
plate, the arrow pointing into the pump nozzle is the input side. The arrow
pointing away from the pump nozzle is the output side.
2. Hold the two sections of the pump housing and end plate together with one
hand and remove the four long screws from the end plate.
3. Carefully remove the end plate to reveal the input and output nozzles,
springs, seals and O-rings.
4. Replace the inlet and outlet springs.

Caution: Be certain the correct springs are used. The weaker spring goes on
the inlet side. The stronger spring goes on the outlet side.

5. Inspect the seals. Replace them if there is any sign of swelling or wear.
6. Inspect the O-rings. Replace them if there is any sign of wear.

Page 52
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
Note: Before reassembling the UF pump, verify nothing has fallen into the
solenoid opening.

7. Reassemble the pump as follows:


• The outlet seal goes toward the pump membrane.
• The inlet seal faces away from the pump membrane.
• Reinstall the spacer in the outlet nozzle if one was removed from the
pump.
• Ensure the outlet nozzle is oriented so it will be at the top of the pump
solenoid to help avoid air locks in the pump.
8. Replace the long screws into the end plate and tighten.
9. Replace the pump in the machine and reconnect the electrical and hydraulic
lines. Orient the pump so the outlet nozzle is above the inlet nozzle to help
avoid air locks.
10. To displace air and to check for leaks, start a rinse and allow the machine to
complete the entire rinse cycle.
11. Verify proper function of the UF pump by performing the Quarterly
Preventive Maintenance procedure in Section 2.6.

Page 53
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
4.2 REBUILDING THE CONCENTRATE AND BICARBONATE PUMPS
Figure 13 is an exploded view of the concentrate and bicarbonate pumps.

OUTLET SPRING IF THE PUMP CONTAINS A PLASTIC SPACER


SEE BELOW * IN THE OUTLET PORT, BE SURE TO REPLACE
SEALS IT WHEN REASSEMLING THE PUMP. NEWER
P/N 565279 PUMPS HAVE THIS SPACER BUILT INTO THE PORT

OUTLET

INLET

INLET SPRING RETAINING PLATE


P/N 565278 O-RINGS
P/N F40007099 or 579058
* ON CONCENTRATE (ACID) PUMPS THIS SPRING
IS STRONGER THAN THE INLET SPRING. USE P/N 563025.
ON BICARBONATE PUMPS USE WEAK SPRINGS
FOR BOTH INLET AND OUTLITS (P/N 565278).

ON MACHINE WITH ONLINE CLEARANCE (OLC). THE


CONCENTRATE (ACID) PUMP USES WEAK SPRINGS
FOR BOTH INLET AND OUTLET (P/N 565278).

Figure 13. Concentrate and Bicarbonate Pump Exploded View.

Rebuild the concentrate pumps as follows:


1. Remove the pump from the machine (See Figure 3, pg. 8). Make a note of
which line goes to which port on the pump. When looking at the end plate,
the arrow pointing into the pump nozzle is the input side. The arrow
pointing away from the pump nozzle is the output side.
2. Hold the two sections of the pump housing and end plate together with one
hand and remove the long screws that hold the end plate. On some models
of the pump, all four screws hold the end plate, and on others only two
screws hold the end plate.
If you removed all four screws to free the end plate, replace one or two of
them to avoid having the pump come completely apart. You only need
access to the inlet seals and springs.

Page 54
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
3. Carefully remove the end plate to reveal the input and output nozzles,
springs, seals and O-rings.
4. Replace the inlet and outlet springs.

Caution: Be certain the correct springs are used on concentrate pumps. See
Figure 13 on previous page.

5. Inspect the seals. Replace them if there is any sign of swelling or wear.
6. Inspect the O-rings. Replace them if there is any sign of wear.
7. Reassemble the pump as follows:
• The outlet seal goes toward the pump membrane.
• The inlet seal faces away from the pump membrane.
• Reinstall the spacer in the outlet nozzle if one was removed from the
pump.
• Ensure the outlet nozzle is oriented so it will be at the top of the pump
solenoid to help avoid air locks in the pump.
8. Replace the long screws into the end plate and tighten.
9. Reinstall the pump in the machine and reconnect the electrical and hydraulic
lines. Orient the pump so the outlet nozzle is above the inlet nozzle to help
avoid air locks.

Note: To avoid discoloration of the pump wiring, route the pump wiring
away from heat sources such as the pump housing and valve coils when
reinstalling the pump.

10. To displace air and to check for leaks, start a rinse and allow the machine to
complete the entire rinse cycle.
11. Test the rebuilt diaphragm pump by performing the procedure described in
Section 4.3.

Page 55
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
4.3 TESTING CONCENTRATE AND BICARBONATE PUMPS
The concentrate (acid) and bicarbonate pumps are tested by measuring the volume of
water they pump and comparing that measurement with the expected value
determined by the machine. The measured pump volume must be within 2% of the
expected value.
Test each pump as follows:
1. Place the machine in Dialysis Mode and open the shunt door.
2. Switch to the debug screens by pressing and holding the [▲] and [▼] keys at
the same time for approximately 1 sec. The main display will change
indicating that the machine is in debug mode.
3. Use the [▲] and [▼] keys to go to debug screen #1.
• Locate the debug value for FILACT. In order to turn off dialysate flow
later, FILACT must be 0.
• Locate the debug value for DOUBLE. Certain types of Bicarbonate or
higher Bicarbonate levels selected will require the pump to double
stroke. If the debug value for DOUBLE is 1, the pump is double
stroking. Before continuing the test, adjust the Bicarbonate level until
the debug value for DOUBLE is 0.
• If the debug values for the Bicarbonate Pump Volume (BMIN and
BMAX) are showing 0 on debug screen #1, the machine is set up for a
concentrate family that does not use bicarbonate. Change the
concentrate selection to one requiring bicarbonate.

Note: The AMIN, AMAX, BMIN, BMAX are the expected limits for twenty
pulses of the acid (concentrate) pump volume and the bicarbonate pump
volume, respectively. These volumes are based on the measured volume of the
balancing chamber to produce the correct mixing ratio.

4. Press the [Dialysate Flow on/off] key to stop dialysate flow.


5. Connect a 25ml buret filled with water to the concentrate connector for the
pump to be tested. Use a short length of tubing forced over the end of the
connector, and make sure there are no leaks in the fitting to the connector or
to the buret.

Page 56
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
6. With the machine in Dialysis Mode, press the [Dialysate Flow on/off] key
to start the flow. Observe that liquid is drawn from the buret in discrete
steps. Allow the machine to run and pump fluid from the buret for about 20
strokes, to prime the pump. Press the [Dialysate Flow on/off] key to stop
the flow.

Note: Do not let the buret run dry so that air can enter the system.

7. Refill the buret exactly to the full mark. Press the [Dialysate Flow on/off] key
to start the flow and count exactly 20 pulses of water drawn from the buret,
then stop the flow.

Note: If DOUBLE=1 on debug screen #1 then count each pulse of the double
stroking bicarbonate pump.

8. Measure the amount of water drawn from the buret carefully to within
0.05ml (within 1/2 division on a buret calibrated in 0.1ml increments).
Verify that this value is between the displayed limits of AMIN and AMAX
for the acid pump volume and BMIN and BMAX for the bicarbonate pump
volume.
9. If a pump volume falls outside of the permitted error, repeat the above
procedure to ensure that you are getting consistent results before re-
calibrating the machine.

Page 57
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
NOTES:

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

Page 58
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
NOTES:

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

Page 59
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
NOTES:

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

Page 60
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
PREVENTIVE MAINTENANCE CHECKLIST
QUARTERLY/1000 HOUR
SERIAL NO. _________________________________________________DATE __________________________
MACHINE I.D. NO. ___________________________________________MACHINE HOURS ______________
TECHNICIAN(S) _____________________________________________LABOR HOURS _________________
See the Referenced Section of the 2008K Preventive Maintenance Procedures (part number
507297 Rev. L) for detailed instructions to perform the following:

Reference Section Procedure Value Completed By:

2.1 Clean Filters (replace if necessary) and replace O-Rings: ................... ________
Water Inlet ............................................................................................... ________
Acid and Bicarbonate Connectors ......................................................... ________
Dialysate Inline Filter .............................................................................. ________
Dialysate Line Connectors ...................................................................... ________
2.2 Inspect Pre-UF Pump Filter ............................................ (Replace If Necessary) ________
2.3 Replace DIASAFE Filter ....................................................... (If Applicable) ________
2.4 Inspect UF Pump Check Valves ...................................... (Replace If Necessary) ________
2.5 Check High Voltage AC Connections ........................................ (<0.2 ohms) _________ ________
2.6 Calibrate UF Pump ...................................... (Range: 24 ± 0.1cc at 24 UF Pump strokes) _________ ________
2.7 Check Conductivity Calibration ............... (Range: Δ ± 0.1mS/cm) _________ _________ ________
Machine Display External Meter

2.8 Check Temperature Calibration ........................................ (Range: 37 ± 0.3°C) _________ ________


(Range: 39 ± 0.3°C) _________ ________
2.9 Check and Calibrate Level Detector ...................................................... ________
2.10 Check Alarm Operation and Pressure Holding Test ........................... ________
2.11 Verify pH ................................................................................ (Range 6.9 to 7.6) _________ ________
2.12 Check Power Failure Alarm ................................................................... ________
2.13 Inspect Blood Pump Rotor ..................................................................... ________
2.14 Final Checks ............................................................................................. ________

NOTES:
___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

Page 61
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
NOTES:

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

Page 62
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
PREVENTIVE MAINTENANCE CHECKLIST
ANNUAL/4000 HOUR
SERIAL NO. _________________________________________________DATE __________________________
MACHINE I.D. NO. ___________________________________________MACHINE HOURS ______________
TECHNICIAN(S) _____________________________________________LABOR HOURS _________________
See the Referenced Section of the 2008K Preventive Maintenance Procedures (part number
507297 Rev. L) for detailed instructions to perform the following:
Reference Section Procedure Value Completed By:
3.1 Clean Filters and Replace O-Rings................................................................... __________
3.2 Replace Pre-UF Pump Filter ............................................................................. __________
3.3 Replace UF Pump Check Valves....................................................................... __________
3.4 Check Deaeration Restrictor and Replace O-Rings ........................ (If Applicable) __________
3.5 Replace DIASAFE filter .................................................................. (If Applicable) __________
3.6 Rebuild UF, Concentrate and Bicarbonate Pumps ......................................... __________
3.7 Check Ground Heater Element ........................................................................ __________
3.8 Check High Voltage AC Connections.................................................. (<0.2 ohms) __________ __________
3.9 Replace Deaeration Motor Brushes .............................................. (every 8000 hours) __________
3.10 Check Inlet Water Pressure Regulator ...................................... (Range: 18 to 20psi) __________ __________
3.11 Conduct Online Clearance Test ........................................................ (If Applicable) __________
3.12 Check Deaeration Pressure ............................................ (Range: See Table 2 on page 32) __________ __________
Check Loading Pressure ........................ (Range: 18 to 20psi / with DIASAFE filter 23 to 25psi)
®
__________ __________
3.13 Check Flow Relief Pressure................... (Range: 29 to 30psi / with DIASAFE filter 35 to 36psi)
®
__________ __________
3.14 Test Concentrate and Bicarbonate Pumps ...................................................... __________
3.15 Calibrate UF Pump ............................................. (Range: 24 ± 0.1cc at 24 UF Pump strokes) __________ __________
3.16 Check Conductivity Calibration ....................... (Range: Δ ± 0.1mS/cm) _________ _________ ________
Machine Display External Meter

3.17 Check Temperature Calibration.................................................. (Range: 37 ± 0.3°C) __________ __________


(Range: 39 ± 0.3°C) __________ __________
3.18 Check Volt Hi Lo Detect ........................................................ (Range: 4.8 to 5.2 ± 0.2V) __________ __________
3.19 Check Blood Leak Calibration..................................................... (Range 4.5 to 5.2V) __________ __________
Check Dimness Calibration .......................................................... (Range 4.0 to 6.0V) __________ __________
3.20 Check Arterial, Venous and TMP Calibration................................................ __________
3.21 Check Dialysate Flow.................................................. (Range @ 500ml/min: 485 to 515ml) __________ __________
3.22 Check Heparin Pump ........................................................................ (If Applicable) __________
3.23 Check Blood Pump............................................................................................. __________
3.24 Check and Calibrate Level Detector ................................................................ __________
3.25 Check Alarm Operation and Pressure Holding Tests..................................... __________
3.26 Verify pH ......................................................................................... (Range 6.9 to 7.6) __________ __________
3.27 Rinse Check ........................................................................... (Range: ≥ 310ml collected) __________ __________
Heat Disinfect Check .................................................................... (Range: 80 to 90°C) __________ __________
3.28 Check Power Failure Alarm ............................................................................. __________
Replace 9-Volt Battery....................................................................................... __________
3.29 Check Blood Pressure Module .......................................................................... __________
3.30 Final Checks ....................................................................................................... __________

Page 63
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
NOTES:

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

___________________________________________________________________________________

______________________________________________________________________________

______________________________________________________________________________

______________________________________________________________________________

Page 64
2008K Preventive Maintenance Procedures
P/N 507297 Rev. L
COMMON CONVERSIONS

PRESSURE

1 Bar 29.53 inHg


1 inHg 25.4 mmHg
1 Psi 51.72 mmHg

VOLUME

1 FLUID OUNCE 29.6 MILLILITERS


1 U.S QUART 0.946 LITERS
1 U.S. GALLON 3.8 LITERS
0.034 FLUID OUNCES 1 MILLILITER
1.057 QUARTS 1 LITER
0.26 U.S. GALLON 1 LITER

MASS

1 OUNCE (avdp.) 28.35 GRAMS


1 POUND (avdp.) 0.45 KILOGRAM
0.035 OUNCE (avdp.) 1 GRAM
Fresenius Medical Care North America

Manufactured by:
Fresenius USA, Inc.
4040 Nelson Avenue
Concord, CA 94520
800 227-2572
http://www.fmcna.com

You might also like