CERTIFICATE OF CALIBRATION
ISSUED BY OPTICAL TEST and CALIBRATION LTD.
DATE OF ISSUE: 21 February 2024 Certificate Number 150226/ACU
0286
19-23 Campus Road
Listerhills Science Park Page 1 of 2 Pages
Bradford BD7 1HR
Tel +44(0)1274 393857 Approved Signatory
Fax +44(0)1274 393336
Name Jenny Harris
Email:
[email protected] Web: www.otc.co.uk Signature QC028
Customer: Bentham Instruments Ltd for Quantum Design GmbH
2 Boulton Road Breitwieserweg 9
READING 64319 Pfungstadt
RG2 0NH Germany
Equipment Information
Make: BENTHAM
Model: Description: Serial No: Reference No:
ORM400 PHOTOMETER/RADIOMETER 41812/1
VL-3701-1-V06 LUX SENSOR 67436
Date of Calibration: 21 February 2024
Calibrated By: J Grabowski
This instrument has been calibrated by direct intercomparison to traceable standards in accordance with approved procedure
RAD02 which follows the general principles of BS 667:2005.
The calibration was undertaken with the instrument being operated in accordance with the manufacturer's handbook
where appropriate.
The uncertainties shown are consistent with the requirements of M3003 and are inclusive of the uncertainty contribution of the
instrument under test. The results, statements and uncertainties relate only to the measured values as displayed by the instrument,
and do not carry any implication regarding the stability of the instrument or its performance outside the measured range.
Laboratory Conditions: Temperature 21 °C ± 2 °C
Relative Humidity 50 %rh ± 20 %rh
The reported expanded uncertainty is based on a standard uncertainty multiplied by a coverage factor of k = 2 unless
otherwise shown, providing a coverage probability of approximately 95 %. The uncertainty evaluation has been carried out
in accordance with UKAS requirements.
This certificate is issued in accordance with the laboratory accreditation requirements of the United Kingdom Accreditation Service. It provides
traceability of measurement to the SI system of units and/or to units of measurement realised at the National Physical Laboratory or other
recognised national metrology institutes. This certificate may not be reproduced other than in full, except with the prior written approval of
the issuing laboratory.
Certificate number
CERTIFICATE OF CALIBRATION 150226/ACU
UKAS ACCREDITED CALIBRATION LABORATORY No 0286 Page 2 of 2 Pages
Calibration Undertaken
Illuminance in the range 1 lux to 20000 lux using a tungsten lamp at a colour temperature of 2856 K.
The reference plane of the sensor was taken to be the front face of the diffuser, with a diameter of 7 mm.
The zero-correction function was used prior to calibration. The zero offset has not been taken into account.
Results shown as despatched.
No specification available
This model requires zero correction prior to every use.
Results as despatched on channel 1 using calibration factor 1 set to 1.980Glx/A
The unit had no programed calibration factors as received.
Calibration Results #
Reference Meter Unit Under Test
CAL1108 & CAL390A As Despatched As Despatched
Range
Applied Displayed pc Displayed pc
0.00 lux Auto 0.000 lux * N/A N/A
2.00 lux Auto 1.963 lux N/A N/A
20.0 lux Auto 20.00 lux N/A N/A
200.0 lux Auto 0.201k lux N/A N/A
2000 lux Auto 1.954k lux N/A N/A
20000 lux Auto 20.59k lux N/A N/A
Calibrations marked '*' in this certificate are not UKAS accredited and have been included for completeness.
Any measurements indicated as 'OOS ' are outside the specification used.
'pc ' indicates probability of conformance to the specification used.
Results indicated as 'PASS** ' are within the stated specification but have a probability of conformance below 50%
Calibrated by: J. Grabowski Date: 21 February 2024
UNCERTAINTY OF MEASUREMENT
The estimated uncertainty in the measured value of illuminance from 1 lux to 20000 lux when the instrument is used
under the same conditions as when calibrated is ± 1 incremental digit in the least significant figure of the displayed
value due to the resolution of the meter plus:
a) ± 3.1 % in the 1.0 lux to 10.0 lux range ( = 2.00) b) ± 3.0 % in the 10 lux to 20 lux range ( = 2.00)
c) ± 2.4 % in the 20 lux to 200 lux range ( = 1.99) d) ± 2.3 % in the 200 lux to 1000 lux range ( = 1.98)
e) ± 2.3 % in the 1000 lux to 2000 lux range ( = 1.98) f) ± 3.4 % in the 2000 lux to 10000 lux range ( = 1.89)
g) ± 5.9 % in the 10000 lux to 20000 lux range ( = 1.77) h) ± 0.8 % at a colour temperature of 2856 K ( = 2.00)