D

FINE
igital Weighing Indicator

Instruction Manual
( FS-2000C )

FINE INTER KOREA CO,. LTD


780-2, Noonsa -dong, Kwangmyeong-City, Gyeonggi Do,Korea.
TEL : +82 2 899 4472 FAX : +82 2 899 4479
http://www.fineweigher.co.kr Contact : [email protected]

Copyright ® 2009 by FINE Inter Korea Co., Ltd All Rights Reserved. Version : 2009, July.
Any form or Any other means can not be copied without the permission from the publisher.
If you will get this install manuals without the permission then all responsibility will give you.
CHAPTER 1. PREFACE
1-1. INTRODUCE
Thank you very much for your purchasing FINE Digital Weighing Indicator of FS-2000C
This Instruction Manual will lead you to use FS-2000C with top reliability,
High speed, high accuracy.

FS-2000C is Digital Weighing Indicator amplifying the analog output from a load
Cell, converting the analog signal to digital data and then displaying this data
As a weight reading and is designed for flawless performance in your demanding
Application of input-weighing, Discharger, accumulating-weighing, 3step control.
Also, an additional option will make Modern Industry demand equipment that both versatile
And available to easily connect to other devices

※ Example for application :


1. PACKING EQUIPMENTS FOR MANUAL WEIGHING
2. EQUIPMENTS FOR AUTO-FILLER WEIGHING
3. EQUIPMENTS FOR OUTPUT WEIGHING
4. EQUIPMNETS FOR ACCURACY WEIGHING
5. RECORD-MANAGEMENT FOR PRODUCT WEIGHT

☞ REMARK
- Specification subject to change for improvement without prior notice.
- If changing ,the Version No can be increased ,but keeps a former version
As far as possible

1-2. SAFTY CONDITIONS


Please keep the following using conditions certainly

◆ EARTH
To avoid an electric error such as a noises in your production line
It should be earthed before installation certainly.
Specially it will be safe to divide the power of Indicator into a load cell.

◆ SAFTY CONDITIONS
Do not use it closed to a explosive gas and an inflammable dust environments

◆ POWER
Use the power under 110/220V 50/60HZ ±10% and divide it into the power line

◆ TEMPERTURE CONDITIONS
OPERATING TEMPERTURE : -10o C +40o C ( +14o to 104o F )
CUSTODY TEMPERTURE : -40o C +80o C ( -40o to 176o F )

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 2
◆ INSTALLATION LOAD CELL
- Available to use the same load cell of 8pcs ( 300Ω standard )
- A ground should be installed horizontal
- Installing over 2pcs of load cell and connect each line in parallel and
Insert a variable resistor under 50Ω in EX + line and minimize a output
Accuracy of load cell.
It may occur a weight error by each accuracy of load cell.
- It may occur a weight error in case of a temperature variation of load cell
- Please weld(electric spark) at the place installed with load cell and equipments,
Divide the power into a connector of load cell in inevitable case
- Please connect the below construction of load cell with the above ones using
The Ground to the weighing part weighing a material or a electric spark.

1-3. FEATURES
- A compact Appearance by DIN regulations ( DIN 192 × 96 Panel system )
- Easy to preset, change, confirm the weight value by the numeral key.
- Improved a convenience and precision of operating with Message Function.
- Can display a various information using F1,F2,F3 key for the end-user.
- Can use or disuse each key function. (SETUP F10 Reference)
- Weight Memory function even in electric spark case (SETUP F02 Reference)
- The permit or prohibition function of Calibration (ADJUST 10 NO Switch)
- Watch-Dog timer guards for self-diagnostics.
- Set up to 1/20,000 display resolution
- Function available to change the unit value such as kg, ton, lb ,g
( In case of Serial Communication & Printer )
- Available to change the function of the external input terminal (SETUP F16 Reference)
- Various option and addition for customer satisfaction such as serial communication,
RS-422/485, Current Loop, Printer, Analog out, BCD Input/Output and so on.

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 3
1-4. FRONT PANEL DESCRIPTION

1-4-1. LAMP
▼ STEADY : This Lamp will be turned on the stable weight
When presetting and working a auto-function,
This STEADY Lamp will be the weighing standard. .

▼ ZERO : This Lamp will be turned when the weighing device is empty.
When presetting and working a auto-function, this
ZERO Lamp will be the weighing standard.

▼ TARE : This Lamp will be switched to Net mode


(SET-UP F12 REFERENCE)

▼ G/W : This Lamp will be turned on the gross weight


When presetting a TARE, It will be a Lamp Function.

▼ EXT. : This Lamp will be displayed with external input

▼ LOW : This Lamp will be displayed when Weight is in LOW weight range.

▼ HIGH : This Lamp will be displayed when Weight is in HIGH weight range.

▼ RUN : This Lamp is being running by the start of Seque ce Board.

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 4
1-4-2. HOW TO USE KEY
* The Key operating can be permitted or prohibited by SETUP-F10
* When pushing the key, it sounds "OK".
* Each Key works either a single function or compound functions.
A compound function key will be a command key when it push first and
According to the command key, the fixed value works its function,
The key to finish a input data is SET Key.
* The input time of compound key is limited with 5sec and automatically will
Be removed without the next key input.

☞ ZERO Key : This key returns the display to the center of ZERO when the weighing device
is empty(the end-user selected within 2%, 10%, 50%, 90% by SET-UP F07)

☞ TARE Key : The way to set-up the tare weight is two way as follows.
- Manual Way
1.Set-up of TARE Key
① Put a TARE on the weighing plate
② TARE Key →SET Key OR TARE Key → Numeral Key → SET Key
2.Remove of TARE Key
① Remove TARE on the weighing plate
② Push TARE Key and push SET Key.
- Automatic Way
1.Auto-TARE setting if TARE was on the weighing plate
2.Auto-TARE setting after putting TARE and Auto-TARE Remove
After Taking away TARE on the weighing plate.
※ Please refer to SETUP F12

☞ G/W Key : After setting TARE, Net Weight converts Gross weight,
OR Gross weight converts Net Weight
* Available to convert setting TARE only.

☞ PART Key : Usable to confirm or change the product part


* Can set up the data of each product from 1 No to 20 No.
- Confirmation PART : PART Key → CLR Key
- Changing PART : PART Key → Numeral Key →SET key

☞ LOW key : This key to input the low weight


LOW key → LO-SET → Numeral Key →SET key(Reference F40 Control System)

☞ HIGH key : This key to input the High weight


HIGH key → LO-SET → Numeral Key →SET key(Reference F40 Control System)

☞ Fall Key : Can preset a Fall weight


Preset + ( FALL Number) + SET

☞ F1,F2,Key : This keys appear a various data as the end-user demands.


Available to use the end-user demanding by SETUP F21,F22,F23
( SET UP F21 REFERENCE )

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 5
☞ TOTAL Key : The function to delete and print SUB-total and TOTAL
* Delete : CLR + TOTAL + SET (SUB-total Delete)
CLR + TOTAL + TOTAL + SET(TOTAL Delete)
* Deleting TOTAL,SUB-total delete automatically.
* Print: TOTAL + PRINT (SUB-total PRINT)
TOTAL + TOTAL + PRINT(TOTAL PRINT)
* Possible to delete automatically when printing (SETUP F18 REFERENCE)

☞ Start / Stop Key (refer : F40)


Can use start & stop a weighing work.
* Please use CLR + Start Key to stop a weighing work.

☞ PRINT Key : This Key is to Transmit, Totalize, Print a DATA


* Unavailable to work it while Auto Mode
* Please push CLR + Print when deleting the last TOTAL date.
Only Unavailable to re-power, change the PART, Available 1time only
(The last total data will be deleted also on Auto-total)

☞ CLR Key : This have 4way to use as follows .


1) When cancelling it with input the setting value
2) CLR + TOTAL(+TOTAL) +SET When setting the total data.
3) CLR + Print when deleting the last TOTAL date
4) When using SETUP or CALIBRATION ( 3Chapter, 4Chapter REFERENCE)
* After CLR Key, If no a additional data, it will be deleted automatically. .

☞ SET/CAL Key : SET key have 2way to use as follows


1) When recording each set data
2) When using SETUP or CALIBRATION( 3Chapter, 4Chapter REFERENCE)

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 6
1-4. REAR-SIDE PANEL

1. F.G. : Please earth it for safe.

2. AC IN : Available to change AC110/220V with multiple.


Before Installation then check out the power voltage.
Change 220V terminal into 110V when it change the power voltage.
( The first power supply was made by 220VAC.)

3. FUSE : please use the standard approved when it replace.


(FUSE) AC250V, 0.3A (a glass tube with small type)

4. POWER S/W) ON/OFF : It will be safe to use it after 10minuate warming time.

5. DATA OUT (Option Board.) :


Serial Communication. RS422, BCD In/Out, Analog Out 0~10V, 4~20 mA
Print Out( Serial Print / Centronics Parallel)

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 7
6. OUT-PUT :
It can connect between COM and each Output Terminal by no voltage contact.
Each Output Terminal Function can be selected by SET UP F40.
Also use this output by control signal but do not use it by driving.
Max earth capacity : AC250V / 0.5A

7. IN-PUT :
It will be used for controlling this indicator from the external equipment.
The functions of input terminal is to choose it by SETUP F16
Because the power supply of input terminal is done by 12V inside,
Do not power in from the external
z An electric current is about 10mA.
z Make Min. input time by more than 50mSEC.

8. RS-232C (25P D-type Female) : (OP-01)


9. Load cell Connector(N-16)
c EX+ (+5V) d EX- (-5V) e SIG+
f SIG- g SHIELD
10. ADJUST : It has DIP Switch which can adjust ZERO and SPAN
z ( No 1 - No 6 : ZERO Adjust , No 7 - 8: SPAN Adjust , No 10 : Calibration Lock
z The Functions of each input terminal can be selected by SETUP F16

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 8
1-6. SPECIFICATION
1. Analog Input & A/D Conversion
Input Sensitivity 0.2 /D
ZERO adjustment Range -4mV 42.0mV
Load cell excitation DC 10V (± 5 V)
Max Input voltage 32mV
Temperature Coefficient ± 20 ppm / ℃
INPUT Noise ± 0.5 P.P
INPUT Impedance 10 (MAX)
A/D Converter 130,000 Count
Non-Linearity 0.005 F.S

2. DIGITAL SECTION
MAX.DISPLAY "1000000"
MIN.DIVISION x1, x2, x5, x10, x20, x50
DISPLAY UNIT 7-Segment, 7digit Highly bright fluorescent tube
KEY BOARD Numerical Key and Function Key(0-9,CLR,SET/CLR)
Data Back-up APPR.10 YEAR

3. GENERAL
POWER AC110 / 220V (±10%), 50 / 60Hz, 10VA
PRODUCT WEIGHT NET 2.3kg BOX 3.3kg
Operating Temperature -10℃ 40℃
Operating Humidity 85%RH MAX (Non-Condensing)
Physical Dimension 193.6 x 98 x 166 (mm)

4. OPTION
OP-01 STANDARD
OP-02 Serial I/F : CURRENT LOOP
OP-03 Parallel I/F : BCD Out
OP-04 Serial I/F : RS422, RS485
OP-05 Analog Output : V out (0-10V / 10V-0V)
OP-06 Analog Output : I out (4-20mA / 20-4mA)
OP-07 Print I/F : CENTRONICS Parallel
OP-10 Parallel I/F : BCD In PART

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 9
1-7. The example for the connecting to external devices

EXTERAL
Centronics PRALLEL DISPLAY

PRINTER

Summing
BOX
P.L.C ANALOG OUT 0 ~ 10V
4 ~ 20mA
ANALOG
RECODER Load cell

Load cell

Load cell
※ Printer : FS-7000D, FS-7000P
FS-7024, FS-7040P
※ External DISPLAY : FS-4200, FS-4400 Load cell

Serial interface (RS232C, RS422)

COMPUTER

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 10
CHAPTER 2. INSTALLATION
☞ Installation Caution.

- Be careful for avoid from a strong impact, vibration. temperature. water, wind.
- Be careful for avoid Installation from a high moisture around.
- Be careful for avoid Installation from a high temperature fluctuation.(±10℃/h).
- Be careful that the power should be isolated from the main power box.
- Be careful that the power should be done by the standard voltage
- ( 110V/220V ± 10% 50/60Hz – First Power voltage 220VAC)
- Be careful that the main switch should be off for connecting to the external device.
- Be careful that it should ground with the external device.
- Note that it should calibrate and set up for the first installation.

☞ Necessary Part for installation.


- Power Code Connector : 1EA
- FUSE : 2EA (PIPE TYPE 250V 0.3A SMALL TYPE)
- LOAD CELL Connector : 1EA (N16-05)
- Instruction Manual : 1Copy
- Adaptable Connector for Option Connection.

LIVE
NEUTRAL

Chassis
Ground

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 11
2-1.Out-Dimmension & CUTTING SIZE

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 12
2-2. ASSEMBLE DRAWING

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 13
2-3.HOW TO CONNECT TO LOAD CELL

1. Recommend Load cell


The Rated Output(R.O) of a load cell should be 1mV/V ∼ 3mV/V
Specially FINE Cell is 2 mV/V ±0.005 and can be supplied together.

▣ The Rated Output voltage of load cell is not absolute value but relative value.
Ex) The Rated output voltage is the same with 3mV/V for 10kg Load cell and 10ton.

2. Load cell Connector


z Please connect the indicator to a load cell according to the wire color.
z Available to connect the load cell Until Max. 8pcs Load cell in parallel.( Max 300Ω)

3. The wire color depending on the load cell manufacturer

Manufacturer. 1 2 3 4 5 Reference.
EXC + EXC - SIG + SIG - SHLD
FINE CELL Red White Green Blue Shield
CAS Red White Green Blue Shield
Interface Red Black Green White Shield
Tedia Green Black Red White Shield

z Load Cell connector Standard : N16-05.

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 14
2-4. ERROR & CLEAR

Error Source Clear Reference

Unstable Weight c Load cell Broken. cMeasure the rated output of c Input
d Load cell Insulation load cell. resistance : about
Resistance Error dMeasure load cell Insulation 400 Ω
③ Touching to a moving frame. resistance doutput resistance
④ Touching to a moving frame about 350 Ω
③ Insulation
resistance : 100

Increasing c Load cell broken. c Measure load cell insulation


weight d Load cell not connecting resistance
or (normal : 100 or OL display)
Not return to d check load cell connector
ZERO
Minus Weight(-) c Rated output of load cell c check load cell connector. ERR-55.
(SIG +, SIG -) exchanged

“bAd” c Load cell Broken. c check load cell.


d Connect Condition. dcheck load cell connector.
③ Out of basic ZERO ③adjust ZERO(5000~50000)

“UL” c Load cell Broken. c check load cell.


(Under Load) d Connect Condition. dcheck load cell connector.
③ Out of basic ZERO ③adjust ZERO(5000~50000)

“OL” c Load cell Broken. c check load cell.


(Over Load) d Connect Condition. dcheck load cell connector.
③ Excessive Weight. ③ remove excessive weight

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 15
CHAPTER 3.CALIBRATION
▣ What is Calibration?

Calibration is to make Max. Weight, Min. Division, Decimal point which Digital Indicator displays be
consistent to the actual weight loaded by a load cell on the platform.

☞ It should calibrate certainly when a load cell or indicator will be changed.

3-1. ZERO ADJUSTMENT

▣ What is zero adjustment.?

ZERO is the standard point which can indicate the weight..


Digital Indicator displays the weight as much as amplified analog value of ZERO from load cell.
It displays “UL” if the actual weight of display was less than ZERO which FINE indicator demands.
Otherwise, it displays “bAd” if the actual weight of display was more than ZERO.
Then FINE indicator will not run when it displays “ UL “or “bAd”

☞ ZERO RANGE Which FINE Indicator demands.


Adjust the value displayed to “test1” closed to 1000 - 20000 (Recommend 5000) by
( Dip-Switch No 1- No 6 )

1. HOW TO ADJUST ZERO POINT

If you power in Digital Indicator while pressing No 1 Key of keypad

Then it displays “ test “


Then if pressing No 1key again then it displays “test1” and displays “ ZERO POINT “.
Then if it does not display “ ZERO POINT” then turn “ ON “ of all No 1 ~ No 6 Dip-Switch on real
Size panel of Digital indicator and adjust ZERO POINT closed to around 5000 by Dip-Switch.

For example,
z Power In while pressing No 1key of Keypad
z Press No 1key of keypad again.;
z Adjust ZERO POINT closed to around 5000 by Dip-switch on the real panel.

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 16
2. How to adjust a Dip-switch
Small Amplified Å-----------------------------Æ Large Amplified
1 2 3 4 5 6 Zero point Amplified Amplified Example.
1 ON ON ON ON ON ON 0 0
2 OFF ON ON ON ON ON 1 -980 Amplified Volume
3 ON OFF ON ON ON ON 2 -1960 Amplified Volume
4 OFF OFF ON ON ON ON 3 -2940 Amplified Volume
5 ON ON OFF ON ON ON 4 -3920 Amplified Volume

62 OFF ON OFF OFF OFF OFF 61 -59980 Amplified


Volume
63 ON OFF OFF OFF OFF OFF 62 -60760 Amplified
Volume
64 OFF OFF OFF OFF OFF OFF 63 -61740 Amplified
Volume

z Adjust ZERO POINT by NO 1 ~ NO 6 Only of Dip-Switch.


z Do not adjust ZERO POINT by NO 7 and NO 8 which can adjust for SPAN adjustment.
z NO 9 of Dip-Switch is to Self- Test Digital Indicator through the input signal( SIG +, SIG-) of
Load cell and always turn OFF NO 9 of Dip-Switch.
z NO 10 of Dip-Switch is to prohibit or permit “Calibration” which can not be access to any
persons.

For example,

(1) A current display of digital indicator = 317720.


(2) No 1 ~ No 6 of Dip-Switch = ON.
(3) Then how ZERO Point can be closed to around 30000?
(4) If Amplified Volume was 980 if NO1 key was OFF then The Amplified Volume of Each Dip-Switch
was as follows.
Dip-Switch 1 2 3 4 5 6
Amplified/V 12540 25080(12540X2) 50160 100320 200640 401280

Then if NO 1,NO2,NO3,NO5 of Dip-Switch was OFF


(12540+25080+50160+100320+200640+401280=288420)
and ZERO POINT will be 29300(317720-288420) closed to around 30000.

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 17
3-2. SPAN ADJUSTMENT

▣ what is span adjustment.

Span adjustment is to adjust the Linearity which makes the display value from "0" to Max. Weight
Consistent to the actual weight

z Please do OFF NO 10 of Dip-switch For Calibration Access.

▶ How to access the SPAN ADJUSTMENT.


There are 2ways to access the span adjustment

☞ The first way


If you power in Digital Indicator while pressing No 3 Key of keypad

Then it displays “ test “


Then if pressing No 3 Key again then it displays “St. CAL”
And if pressing “ St. CAL” key it displays “ d xx(01,02,05,10,20,50) “.

For example,
z Power In while pressing No 3key of Keypad >>> “test”
z Press No 3key of keypad again. >>> “St. CAL”
z Press SET/CAL >>>”d 02”

☞ The second way


c If Pressing SET/CAL key for 3second, It will display "St. CAL"
Press SET/CAL again
d "St. CAL" means SETUP & CALIBRATION mode

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 18
▶ HOW TO ADJUST SPAN.

S&C MODE have 7way to adjust span. Each step will be advanced with SET/CAL key.
Also, CLR key was used to return to the previous display.
z For the next Step : Press SET/CAL key
z For the previous Step : Press CLR key

☞ 1STEP.
A step to set up a division ( Digit) and decimal point.
"d" means "Division" and "xx" means a division capable of displaying.
Also this “xx” will be displayed as 01-02-05-10-20-50 whenever pressing 0(zero) Key
In case decimal point is "0.0" then press 2 Key.
In case decimal point is "0.00" then press 3 Key
In case decimal point is "0.000" then press 4 Key.
In case decimal Point is not then press 1 Key.
And press SET/CAL key, then it will be go to the next step recording a division and the position
Of decimal point.

☞ 2Step
A step to set up Max. Weight.
The display will appear "CAPA"(Capacity) and discretion number(Max.6Digit)
It can input the Max. Weight as the end-user demands instead of discretion number.

How to input is to press SET/CAL key after inputting discretion number.

♣ Do not excess (A division ÷ Max. Weight) with over 1/20,000(0.00005)


If exceeding over 1/20,000, it will display "Err 01".

☞ 3Step
A step to check the zero conditions of Indicator.
After appearing "dEAd", the discretion number(Max.5digits) will appear.
If the present number is closed by 30,000(5,000~50,000) please press SET/CAL key.
If a discretion number did not display or was over 20000,
Do it as the zero adjustment instruction again.

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 19
☞ 4Step
A step to set up test weighter for SPAN Adjustment.
Indicator will display the capacity at weight column which was set at 2 step
After being displayed “SPAn”.

Please input the value of Test Weighter for span adjustment by numeric key.
This value of span Test Weighter must be equal to full capacity, or over 10% of full capacity.

z In case of less 1/5,000 resolution , the value of Test Weighter must be over 20% of full
capacity )
z If span capacity is set less 5% or over Max. Weight , Indicator will display error message.
As “Err 03”

☞ 5Step
A step to load test weighter on the platform .
If it input 1000kg on the above 4STEP. Then it displays “Load”.
Put the actual test weighter or the test weighter, 10% of full capacity. And SET Key.

z (Notice)
If indicator is unmatched with load cell capacity or span standard weight,
The Indicator will display Error message “Err04”

☞ 6Step
A step to display Span Constant Value counted.
If the range of this constant value is between 0.5000 -- 3.50000, All procedure of span adjustment
Is normal.

z ( Notice)
This span constant value can not be adjusted by the numeral key or other way.

☞ 7Step
A END Step.
If it display "END" then Span Adjustment was finished completely.
Then unload the test weighter from the platform and press SET Key.

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 20
▣ For Example of SPAN ADJUSTMENT

z Max. Display Division = 50.00kg


z Display Setting Interval = 10g
z Test Weighter = 10.00kg

STEP S&C Select Mode St. CAL


1 STEP Press SET/CAL d 50

Adjust a division by pressing 0 key d 01

Setting Decimal point by 3key d 0.01

2 STEP Press SET/CAL c 8 0.00 after displaying CAPA

Press 5000 by numeral Key c 5 0.00

3 STEP Press SET/CAL d 4879 after displaying dead

* Adjust ZERO POINT if this value was not closed to 10000 ~ 30000
4 STEP Press SET/CAL s 50.00 after displaying SPAn

Press 5000 by numeral Key


5 STEP Press SET/CAL Load

Load test weighter on the platform

6 STEP Press SET/CAL after 3second 0. 97482


1.
7 STEP Press SET/CAL End

Press SET/CAL after unloading Test TEST “ 7segment “ in display.


Weighter. After “FinE”

In the display 0.0 then it will be normal


1.0

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 21
3-3. ERROR MESSAGES & ADJUST
※ tESt or FS-XXXX : If indicator display only " tESt " or FS-XXXX (Model number)
without any operation ,first of all you must adjust "Dip-Switch"
of back side panel for span and zero value.
※ ERR--01
①Cause : In case resolution (A Interval/Max. Display weight) was set over 1/20,000 resolution.
②Adjust : Set under 1/20,000 resolution(A Interval/Max. Display weight)

※ ERR--02
①Cause : In case Standard Test weight was more than Max CAPACITY
②Adjust : Make Set Standard Test weight equal or less than Max CAPACITY

※ ERR--03
①Cause : In case Standard Test weight for span adjust was set less than 5% of Max CAPACITY
②Adjust : Set Standard Test weight for span adjust into less than 5% of Max CAPACITY

※ ERR--04
①Cause : In case the weight was not Steady when it account the value of a span constant
②Adjust : Adjust a span again after removing a cause to be unstable or to have vibration.

※ ERR--05
①Cause : In case the actual weight was more than Standard Test weight
Or the amplification Volume of Analog circuit inside was more than.
②Adjust :
- Please check it if the actual weight was more than Standard Test weight or not
If it did so, please adjust the standard Test weight into the value set up.
- If it continue to display ERR--05, Adjust NO 7,8 of Dip-Switch on the rear panel.
For a reference,
NO 1 ~6 of Dip-Switch is to adjust ZERO.
NO 7 ~8 of Dip-Switch is to adjust SPAN
Also because ZERO was changed according to NO 7 ~8 of Dip-Switch,
Please adjust ZERO again as 3-1 ZERO Adjustment
OFF
NO7 NO8 Amplified Volume
ON ON Small 1times
OFF ON Normal 2times
ON OFF Big 3times
OFF OFF Very Big 4times ON

- The way to use the Dip-Switch.


- Please adjust SPAN again after adjusting less than the present adjusted value.
※ If it continues to display ERR--05 in spite of adjusting the Dip-Switch as the above,
Please check it if the cable wire of a Load cell was normal or nor.

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 22
※ ERR--55
①Cause : In case a cable wire of a Load cell was connected on reverse.
②Adjust : Please check the connection of a Load cell as a reference of 2.3 CHAPTER

※ ERR--06
①Cause : In case the actual weight was loaded less than standard balance weight
Or was less than Analog Circuit Amplification.
②Adjust :
- Please adjust a standard balance weight into the weight set up.
- If continue to display ERR--06,Adjust NO 7,8 of the Dip-Switch on the rear panel.
For a reference,
NO 1 ~6 of Dip-Switch is to adjust ZERO.
NO 7 ~8 of Dip-Switch is to adjust SPAN
Also because ZERO was changed according to NO 7 ~8 of Dip-Switch,
Please adjust ZERO again as 3-1 ZERO Adjustment
The way to use the Dip-Switch.
OFF
NO7 NO8 Amplified Volume
ON ON Small 1times
OFF ON Normal 2times
ON OFF Big 3times
ON
OFF OFF Very Big 4times

- Please adjust SPAN again after adjusting less than the present adjusted value.
※ If it continue to display ERR--06 in spite of adjusting the Dip-Switch as the above,
Please check it if the cable wire of a Load cell was normal or nor.
※ ERR--07
①Cause : In case it was deviated from a range of value which can be set by SET UP,
②Adjust : Please input the contents of SET UP again.

※ ERR--10
①Cause : In case the record device of Memory or Hardware was not normal
②Adjust : It can be worked by a voluntary key, but it was temporary way.
So, please try to send this Indicator to the head office for A/S.

※ " UL" (UNDER LOAD)


①Cause : In case the connection of a Load cell was not normal or a Load cell was broken.
②Adjust : Refer to the part related with a Load cell or CHAPTER 3 ZERO ADJUSTMENT.

※ " OL" (OVER LOAD)


①Cause : In case the connection of a Load cell was not normal or a Load cell was broken.
②Adjust : Refer to the part related with a Load cell or Remove a excess weight.

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 23
CHAPTER 4. SET-UP
4-1. PREFACE
" SET-UP " is to choose each proper functions for matching the indicator with the appliances of field.

▣ How to enter into set-up mode


This set-up mode is required for proper weighing operation when Indicator connects
With other appliance. It can enter into sep-up mode by the below two steps.

☞ The first Step


Depress key "ekey" first and power on at the same time.
At that time, "tESt" word will be displayed on indicator.
Depress key "ekey" again, and indicator will display as following :

S t, C A L. ; S & C Mode
At this time, press CLR key.
Indicator will display to " F01-xx " from above test message.
* For example
The power was OFF
1. Power "ON" while pushing ekey --------- "tESt"
2. Pushing ekey again --------- "St. CAL"
3. Push CLR key --------- "F01 - xx"

☞ The second Step


If you depress key " SET/CAL" for 3 seconds at the normal weighing mode,
Indicator will also display "St. CAL" as the above.

4-2.SET-UP
① If it press CLR key at S&C Mode, Indicator will display "F01-xx"
The F of "F01-xx" means Function and 01 means Function number
And the last 2figure "-xx" means each functional setting number
* For example
Pushing CLR key in "St. CAL" mode then displays as “ F01-01 “
Function number will be increased to the next Function whenever it presses .

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 24
② If you proceed to next function, press CLR key or,

If you want to see your desirous any function number,


Press "CLR" key after input any function number by numeric key.
Indicator will display function number directly from present function number.

(EXAMPLE)
* Present display : F01-01
Press CLR key ----> "F02-00" display ----> Press CLR key.
----> "F03-01 display ----> Continuously press CLR key ---->
"F04-XX" ----> "F05-XX" ----> "F06-XX" ---->
Press CLR key in streams, the next function number will be displayed.

* Present display : F01-01


If you want to see function number 12,
Press numeral key "1" and "2" ----> Press CLR key ----> "F12-XX" display

③ If you want to change each functional setting number newly,


Press SET key after input the functional setting number by numeral key.

(EXAMPLE)

If "F01-01" is changed to "F01-03",


Press 3 key ----> F01-03 display ----> Press SET key.
A new function number will be memorized.

(Remarks) When you want to change " S & C MODE " from Set-up mode,
Please press key " 0 " + " CLR " consecutively.

※ ERR--07
①Cause : In case it was deviated from a range of value which can be set by SET UP,
②Adjust : Please input the contents of SET UP again.

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 25
4-3.F-FUNCTON LIST
F-NO FUNCTION CONTENTS
F-00 GROUP-SETTING A BASIC WEIGHIG
F 00 S & C MODE Convert SETUP & CALIBRATION
F 01 weight unit choice kg, ton, lb
F 02 weight BACK-UP NORMAL, BACK-UP
F 03 Set ZERO tracking Range 0, 0.5, 1, 2
F 04 Set Safety Motion Band 0.5, 1, 2, 4, 8
F 05 Set AUTO ZERO Range 0-99 (Auto Zero Range)
F 06 Digital Filter 0-9 (anomalous decrease)
F 07 Set ZERO Range Max. weight 2, 10, 50, 90%
F 08 Set Delay time of Safety adjustment 0-99 (1count = 0.1sec)
F-10 GROUP-SETTING A BASIC DEVICE
F 10 Selecting a Key Lock Prohibition & Permit for KEY
F 11 ZERO,TARE,OPERATE MODE Safety, Un-safety
F 12 TARE weight INPUT MODE Set Numeral, Actual, Auto TARE
F 13 EMPTY Signal MODE Output Choice when it is ZERO or Empty
F 14 SET EMPTY Range Set Set Empty Range Weight
F 15 SET EMPTY Standard Set Display weight, basic ZERO,TARE ZERO
F 16 External INPUT MODE Input terminal function
F 18 DELETE Totalization information Delete in Manual/Auto for totalization
F 19
F-20 GROUP-SETTING CONTROL SYSTEM
F 21 User key definition No definition or Set
F 22 User key definition No definition or Set
F-30 GROUP-SETTING Serial Interface Specification
F 30 BRUD RATE 300, 600, ........ 38.4 kbps
F 31 Set Parity Bit EVEN, ODD, NO PARITY
F 32 Set Transmit MODE Continue, Safety, Totalization, Command.
F 33 Set Format Transmit DATA weight, weight + time, CAS transmit format
F 34 Insert Transmit DATA (STX) No, Insert
F 35 Control Interface wire / RS,CS No use(RS422/485), USE
F 36 Select Interface weight Weighing weight, G.Weight, N.Weight
F 37 Set Transmit Comma(,) Select “,” or not.
F-40 GROUP – CONTROL MODE
F 40 Set - up For Controlling Weight Simple / Sequence / Check Mode Control
F 41 Delay time for Finish Signal 0 ~9.9 sec set
F 42 Maintaining time for finish signal Continuous or 0.1 ~ 9.9 sec set
F 43 Set for FALL Compensation No use or 5,10,20,50 times
F 44 Set for Prohibited Time No use or 0.1 ~ 9.9 sec set
F 45 Set for a weighing Range + range, Absolute,- range
F 46 Set for Finish Signal way Manual, steady, steady + time, time
F 47 Set for display weight way No use ,fixed display weight
F 48 Set for zero or tare of start No use for a simple control mode

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 26
F-50 GROUP-SETTING BCD Output Specification
F 50 Weight choice for output Display, Gross, Net weight
F 51 BCD OUT Parity Positive / Negative OUT
F-60 GROUP-SETTING Analog Out Specification
F 60 Weight choice for output Display, Gross, Net weight
F 61 Standard weight choice of Analog Out Max, display weight, Standard weight
F 62 Analog Out Parity Positive / Negative OUT
F 63 Standard weight Set of Analog Out Set standard value of Max. Analog Out
F-70 GROUP-SETTING Printer Specification
F 71 Set Printer system Continue / Each
F 72 Set Line Feed after printer finished 1 Count = 1 Line (0∼99)

F-90 GROUP-SETTING
F 90 Set Device Identification No 00 ∼ 99
F 91 Set Display Color 00 ~ 08
F 92 Set language 0~1
F 93 Set Display LCD 0~7
F 95 DATE Modification MODE Yy – mm - dd in Printer option only
F 96 TIME Modification MODE hh- mm - ss in Printer option only
F 98 Check A/D Count of Basic ZERO Check load cell if it is normal or not

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 27
CHAPTER 5. SET-UP ILLUSTRATION
5-1. BASIC FUNCTION FOR WEIGHING

F-00 S&C MODE converting

Dimension unit set

▶ 0 kg
F-01
1 ton

2 lb

Weight Back-Up
▶ 0 NORMAL
F-02
1 BACK-UP

※ The indicator will not be done back-up the weight when power is off or down.

※ The indicator will be done back-up the weight when power is off or down.
This weight back-up must be used after span calibration.

Zero tracking range


0 No zero tracking
▶ 1 0.5DIGIT / 0.25sec
F-03 2 1DIGIT / 0.25sec
3 2DIGIT / 0.25sec

※ The meaning of zero tracking ??

In case of increasing the weight minutely by certain external factor


(like dust), this minute weight will be compensated to zero if it is within
a fixed division and time.

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 28
Motion Band range

0 0.2DIGIT
※ Motion Band ;
▶ 1 2DIGIT This is to compensate for the momentary
F-04 2 5DIGIT vibration value.
3 10DIGIT

If indicator is used in vibration area, please set enough motion band range

Auto zero range

0
F-05 ~ The remaining weight is automatically to be "Zero"
99

※ Setting way
<Example> If the indicator is set to 3kg * 1g and F05-10 ,
the range of auto zero will be to 1--10g.

At this time, F05 function is available up to 10 % of full capacity.

Digital filter

0 Low More sensitive


F-06 ~ ↕
9 High Less sensitive

If you use the indicator with conveyer belt system or any other vibrating appliance,
this f06 function will be applied for filtering or absorbing the Vibrating or oscillating
weighing value.

Set for zero range


0 Within 2 % range of Max. capacity
1 Within 10 % range of Max. capacity
F-07
2 Within 50 % range of Max. capacity
3 Within 90 % range of Max. capacity

Notice : If the zero range is set to 50% by 100kg capacity scale, the load cell of
scale is loaded up to 150kg. So, when the zero range is set, the load cell
capacity must be considered.

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 29
Set for stable judgment delay time

If the weight is within the setting value at F04,


0
the indicator display "steady" and make a judgment after
F-08 ~
setting delay time.
99
(1 count = 0.1 sec delay)

Keyboard lock

▶ 0 All keys can be used.


F-10 1 1st, 2nd, 3rd, 4th, START, STOP keys are locked.
2 All keys except Zero are locked.
This function can protect the wrong key-operation of worker what may happen.

Operation mode of Zero & Tare key


▶ 0 Zero & Tare key can be used when the scale is stable only.
F-11
1 Zero & Tare key can be used even if the scale is in unstable.

Tare weight setting mode


Keyboard input of known tare weight.
0 After input the tare weight, press SET/CAL key to enter into
F-12 memory.
1 Preset tare weight on weighing plate

Empty signal mode


0 Empty signal will be out when the zero signal is out.
Empty signal will be out when the scale is in Zero & -
1
(minus) weight.
F-13 Empty signal will be out on the absolute weight within
2
empty range
Empty signal will be out when the scale is in "+" range &
3
"-" weight of empty range.

Set of Empty range

Empty range
F-14 This function is used when F-13 is set to 2 or 3.
Weight set

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 30
External input mode
COM 1 2 3 4
0 ZERO TARE NET GROSS
F-15 1 ZERO TARE PRINT NET/GROSS
2 ZERO TARE START STOP
3 BCD 10^0 BCD 10^1 START STOP

External output mode

COM 1 2 3 4 5 6
F-16 0 1st 2nd 3rd 4th End Empty
1 No use No use No use No use No use No use

Remove total Data

0 Remove with CLR + Sub total, CLR + Gross key


F-18
1 Auto Remove with Sub, Gross weight printing key

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 31
F1 KEY FUNCTION SETTING BY USER
0
□ No AVAILABLE
1 DATE
2 TIME
3 DATE & SETTING
4 TIME & SETTING.
5 S/T Working Number
6 G/T Working Number
7 S/T Weight.
F21- 8 G/T Weight
9 P.N Working Start DATE
10 P.N Working Finish DATE
11 P.N Working Start TIME
12 P.N Working Finish TIME
13 Display CODE No and Setting
14 Setting Underweight Accuracy (F28 Reference)
15 Setting Overweight Accuracy (F29 Reference)

F2 KEY FUNCTION SETTING BY USER.

The above F1 FUNCTION SETTING is the same


F22- 0 No available

SET DEVICE INDENTIFICATION NO

00 Unavailable for ID NO in case of '00' setting.


F90- ~ Available for ID NO in case of INPUTTING ID NO
99 * First Setting 00

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 32
SET DISPLAY COLOR

0 RED

1 GREAN

2 YELLOW

ZERO WEIGHING STEADY

3 RED YELLOW GREEN


F91-
4 GREEN RED YELLOW

5 YELLOW RED GREEN

6 RED GREEN YELLOW

GREEN YELLOW RED

8 YELLOW GREEN RED

SET LANGUAGE

SET LCD DISPLAY LANGUAGE

F92- KOREAN

1 ENGLISH

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 33
SET LCD DISPPLAY

Display the Setting Value for High, Low


(High: upside, Low: down)

Display the Setting Value for High, Low


1
(Low: left, High: right)

2 Display the Setting Value High, Low as bar graph

Display Low, High, Part, Total Value


3 (Low: Upside-left, High : Upside-right)
(Part : Down - left, Total Weight : Down - right)

Display Low, High, Fall Value


4 (High: Upside-left, Fall : Upside-right)

F93- (Low : Down-right)

Display Low, High, Fall , Part Value


5 (High: Upside-left, Fall : Upside-right)
(Low : Down – left, Part : Down-right)

Display Low, High, High Fall , Low Fall Value


6 (High: Upside- l`eft, High Fall : Upside-right)
(Low: Down-left, Low Fall : Down-right)

It will be applied after setting F40-6,7 (Refer to function F40)

Display Low, High, High Fall , Low Fall Value


7 (Low : Upside - left, High: High -right)
(Low Fall : Down - left, High Fall : Down - right)

It will be applied after setting F40-6,7 (Refer to function F40)

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 34
DATE MODIFICATION MODE

Example)

Display “F01-00”
F95- k key → g key→ CLR key

In case of 97year 09month 30day →2001year02month14day


b key → c key →b key → d key → c key → f key
→ SET/CAL

* DATA & TIME was subjected to the OPTION setting.

TIME MODIFICATION MODE

Example)
Display “F01-00”
k key → h key → CLR key
F96-
17hour 25min 30sec → 21hour 55min 56sec
d key → c key→ g key → g key → g key →
h key → SET/CAL key

* DATA & TIME was subjected to the OPTION setting.

CHECK A/D COUNT OF BASIC ZERO

F98- * This key was used to check a load cell Error.

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 35
5-3 SERIAL INTERFACE
( RS-232C,CURRENT LOOP,RS-422/485)
SET BAUD RATE
0 300 bps 5
□ 9600 bps
1 600 bps 6 14.4k bps
F30- 2 1200 bps 7 19.2k bps
3 2400 bps 8 28.8k bps
4 4800 bps 9 38.4k bps

SET PARITY BIT


0
□ EVEN
F31- 1 ODD
2 NO PARITY

SET TRANSMIT MODE


0
□ Stream (Output in streams )
1 Auto print ( Output only when stable over 1% of full capacity)
2 Print-key ( Output only by pressing " Print " key )
F32-
3 Output data when weight was finished
4 Transmit COMMAND MODE
5 Serial Printer mode only

SET FORMAT TRANSMIT DATA

S T N T k g (CR) (LF)
0

Header1 Header2 WEIGHT(8) UNIT

F33- S T , N T , k g , (CR) (LF)


1
Header1 Header2 WEIGHT(8) UNIT TIME(6)

2 S T , N T , , k g (CR (LF)

* ID NO will be displayed automatically when setting of ID NO of F-90


* No available in case of F33 - 02

INSERT TRANSMIT DATA(STX)


b NO STX
F34-
1 Transmit of STX(ASCII=02)

CONTROL INTERFACE WIRE/ RS422 (485)


b NO USE for CS, RS / in case of RS422,485
F35-
1 USE FOR CS, RS

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 36
SET INTERFACE WEIGHT
b Transmit Weighing Weight
F36-
1 Transmit Display Weight
2 Transmit G.Weight
3 Transmit N.Weight

SET COMMA “,”


b NO USE “,” AFTER ID
F37-
1 “,” AFTER ID
This function can use if ID is setting with F90.

5-3-1. RS-232C SERIAL INTERFACE


Because RS-232C Interface is the system that transmit the signal by Voltage Volume
It should install AC Power Cable or Electric Wire separately and the cable must be used with shield Coax
Cable. And the suitable Interface Distance should be in 10M.

▩ SIGNAL FORMAT

(1) Signal Type : EIA RS-232C


(2) Transmit Method : FULL-Duplex, Asynchronous
(3) Baud rate : 300, 600, 1200, 2400, 4800, 9600, 14.4k, 19.2k, 28.8k, 38.4kbps
(4) Bit Format ⓐ Data bit : 7 or 8 (No parity)
ⓑ Stop bit : 1 Bit
ⓒ Parity bit : EVEN,ODD,NONE
ⓓ Code : ASCII

0 +10V
LSB MSB
0 1 2 3 4 5 6
1 -10V

↑ ↘ Data Bit ↙ ↑ ↘Stop 1 Bit


Start bit Parity bit
or MSB 7

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 37
▩ STREAM MODE
It is doing Data Output in Stream Mode whenever A/D Conversion
( Reference)
z A/D Conversion : Appr.25times/sec in lower weight.
z A/D Conversion : Appr.15times/sec in Heavy Duty weight.

▩ DATA FORMAT

(STX) 1 2 , S T , N T , + k g (CR) (LF)


t
FC ID. No Header1 Header2 weight (8) l b

Unit

▶ FC(First Character)
- Insert in case of SETUP F34-01
▶ ID. No
- Insert in case of except SETUP F90- “00”
▶ Header 1
- OL : OVER LOAD
- UL : UNDER LOAD
- ST : WEIGHT STEADY
- US : WEIGHT WAVE
▶ Header 2
- NT : (NET WEIGHT MODE)
- GS : (GROSS WEIGHT MODE)
▶ WEIGHT (8)
- SIGN Signal ( + or - )
- Weight ( Decimal Point Included)
▶ DATA For Number
- 2B(H) “ ” : PLUS
- 2D(H) “ ” : MINUS
- 2O(H) “ ” : SPACE
- 2E(H) “ . ” : Decimal Point
▶ Unit
- k : Unit of Kg
- t : Unit of TON
- lb : Unit of Pound

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 38
▩ RS-232C Circuit (25PIN -Type Female Connector)

1 FG
RXD
2 RXD (from TXD)
TXD
3 TXD (to RXD)
CTS
RTS 4 CTS (from RTS)
5 RTS (to CTS)
6
7
8
20
.

25
▣ Receive Program example (Personal Computer)
In case of setting of F30-00, F31-00, F32-00, F33-00, F34-00

Basic Program

10 OPEN "COM1: 300, E, 7, 1, DS, CS" AS # 1


20 INPUT #1, A$, B$, C$
30 PRINT A$, B$, C$
40 GOTO 20

5-3-1. RS-232C SERIAL INTERFACE

▣ SIGNAL FORMAT

- Type : EIA RS-232C


- Method : Half-duplex, Non-synchronize, Bi-direction
- Format : ⓐ Baud rate : 300,600,1200,4800,9600,14.4k,19.2k,28.8k,38.4k,bps(Baud-Rate)
ⓑ Data bit : 7 or 8 (No parity)
ⓒ Stop bit :1
ⓓ Parity bit : Even, Odd, No parity
ⓔ Code : ASCII

0 +10V
LSB MSB
1 -10V
0 1 2 3 4 5 6

↑ └→ Data Bit ←┘ ↑ └→ Stop 1 Bit


Start 1 bit Parity bit
or MSB 7

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 39
▣ STREAM MODE
Stream Mode output a data by A/D conversion.

▩ DATA FORMAT

(STX) 1 2 , S T , N T , + k g (CR) (LF)


t
FC ID. No Header1 Header2 weight (8) l b
Unit

▶ FC(First Character)
- Insert in case of SETUP F34-01
▶ ID. o
- Insert in case of except SETUP F90- “00”
▶ Header 1
- OL : OVER LOAD
- UL : UNDER LOAD
- ST : WEIGHT STEADY
- US : WEIGHT WAVE
▶ Header 2
- NT : (NET WEIGHT MODE)
- GS : (GROSS WEIGHT MODE)

▶ WEIGHT (8)
- SIGNAL ( + or - )
- Weight ( Included Decimal point)
▶ DATA For Number
- 2B(H) “ ” : PLUS
- 2D(H) “ ” : MINUS
- 2O(H) “ ” : SPACE
- 2E(H) “ . ” : Decimal Point
▶ Unit
- k : Unit of Kg
- t : Unit of TON
- l b : Unit of Pound

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 40
▣ RS-232C Circuit (25P D-Type Female Connector)


Receive Program example (Personal Computer)
F30-00, F31-00, F32-00, F33-00, F34-00
Basic Program

10 OPEN "COM1: 300, E, 7, 1, DS, CS" AS # 1


20 INPUT #1, A$, B$, C$
30 PRINT A$, B$, C$
40 GOTO 20

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 41
5-3-2. CURRENT LOOP (OP-02)

▣ SINGAL FORMAT

0 20mA

1 0mA

Same as 5-1 RS-232 option


It must use the interface speed with 4800bps.
The distance recommend is 100 M and the resistance is 500 Ω

▣ DATA FORMAT

Same as 5-1 RS-232 option

▣ 25P D-Type Female Connector

* The transmission terminal was made with Non-polarity.


* The receive terminal was powered with 12V.

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 42
COMMAND MODE FORMAT

RESPONSE
COMMAND FUNCTION COMMAND MODE Transfer
( F32-04 ) Mode
R CR LF demand current weight Standard DATA FORMAT No receive
T CR LF Same as [TARE] Key ACK CR LF NO
KT weight(6) CR LF Same as [TARE] Key ACK CR LF NO
G CR LF Change to ‘Gross Weight ACK CR LF NO
N CR LF Change to ‘Net Weight ACK CR LF NO
Z CR LF Same as [ZERO] Key ACK CR LF NO
P CR LF Same as [PRINT] Key ACK CR LF NO
A CR LF Same as [Sub-total] Key ACK CR LF NO
M CR LF Remove Auto setting ACK CR LF NO
ST CR LF Same as [Sub-total] Key ACK CR LF NO
GT CR LF Same as [TOTAL] Key ACK CR LF NO
STC CR LF Deleting Sub-total ACK CR LF NO
GTC CR LF Deleting TOTAL ACK CR LF NO
HON CR LF Setting HOLD ACK CR LF NO
HOF CR LF Remove HOLD ACK CR LF NO
PN Part(2) CR LF Change PART ACK CR LF NO
CD 코드(6) CR LF Setting CODE 6numerial ACK CR LF NO
DT YYMMDD CR LF Setting DATE ACK CR LF NO
TI HHMMSS CR LF Setting TIME ACK CR LF NO
RDT CR LF Amend DATE YY MM DD CR LF No receive
RTI CR LF Amend TIME HH MM SS CR LF No receive
RPN CR LF Amend PART PART(2) CR LF No receive
RCD CR LF Amend CODE NO CODE (6) CR LF No receive
PART(2),
FREQUENCEY(6),WEIGHT(1
RST CR LF Amend Sub-total data No receive
1)
CR LF
FREQUENCEY(8) , WEITHT
RGT CR LF Amend Total data No receive
(13) CR LF
REN CR LF Amend The last Weight WEIGHT(7) CR LF No receive

* F90- (01-99) : It should add to 2numerial of device ID no.


* F34- 01 : Starting will come to STX(ASCII=02).

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 43
5-4. SET-UP FOR CONTROLLING WEIGHT.

SET WEIGHING SYSTEM

0 Control Simple Comparison 1 ( Limit Mode )


1 Control Simple Comparison 2 (FALL)
2 Control Simple Comparison 3 (LO-FALL,HI-FALL)
F40- 3 Control Simple Comparison 4 (Accumulation of 2 Materials)
4 Sequence Control 1 ( Packer Mode )
5 Sequence Control 2 ( Fall )
6 Sequence Control 3 ( LO-FALL,HI-FALL)
7 Sequence Control 4 (Accumulation of 2 Materials)
8 Control Weight Judgment1 ( Checker Mode )
9 Control Weight Judgment 2
10 Control Weight Judgment 3

* HIGH,LOW Judging Prohibition Time


This Function is to delay Judging weight for the time set by F44 in case of a Vibration.
* Weighing Range : The range to weigh a final weight after prohibition time for
HIGH weight judging.
a) Manual Weighing : Total weight in inputting printer key F46-00.
b) Steady : Total weight when it was steady F46-01
c) Steady & Time : Total weight after a general time for the above (b) F46-02.
d) Time : Total weight after time set by F41
* ZERO/TARE INPUT: Function to keep ZERO the present weight before weighing.

*EMPTY Signal: It occurs a empty signal when the weight of a weighing part was less than.
- F13 : SET EMPTY MODE
- F14 : SET EMPTY SIZE when F13-02,F13-03.
- F15 : SET weight Comparison (Display Weight, Basic ZERO)
Recommend F15-01(Basic ZERO).
* FINISH Signal : ON in weighing and OFF after Time set by F42

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 44
5-4-1 Control Simple Comparison 1 SET
Control Simple Comparison 1
F40-00
weight

high

Low

Empty

remain

time
1,2step no weigh time
weighing range
weight Fixed range

zero/tare input
empty signal

low signal

high signal
(Refer1)
finish signal
▣ High, Low Control signal according to the weight was changed.
When Each weight was same with or grater than set weight
Then, it is signal is ON, less than is OFF
▣ The weighing range according to F45 occur the high, low signal from
Input-weighing(+range),discharge(-range),absolute value
▣ The FALL compensation was unavailable
▣ It keeps the finish signal till the empty range by (refer1)
F42-00 setting.

OUT 1 : LOW OUT 2 : HIGH OUT 3 : FINISH OUT 4 :EMPTY


F40 0 F44 Set prohibited time of weighing
Set Finish weighing time after high
F41 F45 Set weighing range(input, output)
signal
Finish Standard
F42 Set time of finish signal F46
(Manual, Time, Steady)
F43 Set FALL & No Compensation F47 Weight Hold in weighing finish

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 45
5-4-2 Control Simple Comparison 2 SET
Control Simple Comparison 2
F40-01
weight

fall

high
Low

Empty

remain

time
1,2step no weigh time
weighing time
Weight Fixed range

zero/tare input
empty signal

low signal

high signal
(Refer1)
fini▣
sh sThis
ignaisl same as simple comparison control 1,
High control signal was different by FALL set
HIGH SIGNAL = HIGH SET - FALL
HIGH SET is very more convenient by setting a final weighing target
Also, auto fall compensation(F43) made the weight value precise
▣ High, Low Control signal according to the weight was changed.
When Each weight was same with or grater than set weight
Then, it is signal is ON, less than is OFF
▣ The weighing range according to F45 occur the high, low signal from
Input-weighing(+range), discharge (-range),absolute value
▣ The FALL compensation was unavailable
▣ It keeps the finish signal till the empty range by (refer 1) F42-00 setting.
OUT 1 : LOW OUT 2 : HIGH OUT 3 : FINISH OUT 4 :EMPTY
F40 1 F44 Set prohibited time of weighing
Set Finish weighing time after high
F41 F45 Set weighing range(input, output)
signal
Finish Standard
F42 Set time of finish signal F46
(Manual, Time, Steady)
F43 Set FALL & No Compensation F47 Weight Hold in weighing finish

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 46
5-4-3 Control Simple Comparison 3 SET
Control Simple Comparison 3
F40-02
weight
fall
high
fall

low

Empty

remain

time
1,2step no weigh time
weighing time
Fixed weight range

zero/tare input
empty signal

low signal

high signal

finish signal (Refer1)

▣ This is same as simple comparison control 1,


High, Low control signal was different by High, Low Fall set
LOW SIGNAL = LOW SET - LOW FALL
HIGH SIGNAL = HIGH SET - HIGH FALL
By setting Several Fall to HIGH,LOW Set,
HIGH,LOW can be controlled by 1step control.
Only, auto fall compensation(F43) was unavailable
▣ High, Low Control signal according to the weight was changed.
When Each weight was same with or grater than set weight
Then, it is signal is ON, less than is OFF
▣ The weighing range according to F45 occur the high, low signal from
Input-weighing(+range), discharge (-range),absolute value
▣ It keeps the finish signal till the empty range by (refer1)
F42-00 setting.
OUT 1 : LOW OUT 2 : HIGH OUT 3 : FINISH OUT 4 :EMPTY
F40 2 F44 Set prohibited time of weighing
Set Finish weighing time after high
F41 F45 Set weighing range(input, output)
signal
Finish Standard
F42 Set time of finish signal F46
(Manual, Time, Steady)
F43 Set FALL & No Compensation F47 Weight Hold in weighing finish

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 47
5-4-4 Control Simple Comparison 4 SET
Control Simple Comparison 4
F40-03
weight
fall
high
fall

low

Empty

remain

time
1,2step no weigh time
weighing time
Fixed weight range

zero/tare input
empty signal

low signal

high signal

finish signal (Refer1)

▣ This is same as simple comparison control 3,


High, Low was set different by each weight
LOW SIGNAL = LOW SET - LOW FALL
HIGH SIGNAL = LOW SET+(HIGH SET - HIGH FALL)
By setting Several weight to HIGH,LOW Set,
it is very convenient to manage each different weight
▣ Only, auto fall compensation(F43) was unavailable
▣ High, Low Control signal according to the weight was changed.
When Each weight was same with or grater than set weight
Then, it is signal is ON, less than is OFF
▣ The weighing range according to F45 occur the high ,low signal from
Input-weighing(+range), discharge (-range),absolute value
▣ It keeps the finish signal till the empty range by (refer1)
F42-00 setting.
OUT 1 : LOW OUT 2 : HIGH OUT 3 : FINISH OUT 4 :EMPTY
F40 3 F44 Set prohibited time of weighing
Set Finish weighing time after high
F41 F45 Set weighing range(input, output)
signal
Finish Standard
F42 Set time of finish signal F46
(Manual, Time, Steady)
F43 Set FALL & No Compensation F47 Weight Hold in weighing finish

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 48
5-4-5 Sequence Control 1 set
Sequence Control 1
F40-04
weight
h ig h

lo w

Empty

remain

time
1st,2nd Judge prohibited time
Judge count range
Weight fixed range

Zero/Tare input
Start signal input

Empty signal output

Low output
High output
Finish output (Ref1)

▣ High, Low signal will output by Start Input and will be OFF when it will be the same
or more with Setting weight.
FINISH Signal after High signal OFF will be transmitted by F46 Setting and OFF
Signal will not be ON even if the weight was less than Setting Value.
Also it had better use STOP Signal in No normal case
and please don’t use it normally.
▣ the weight range became input weighing(+range),
discharging (- range), or High, Low in the absolute value.
▣ Auto Fall Compensation(F43) will not be available
▣ Finish signal will be available until empty range in setting F42-00
OUT 1 : LOW OUT 2 : HIGH OUT 3 : FINISH OUT 4 :EMPTY
F40 4 F44 Set prohibited time of weighing
Set Finish weighing time after high
F41 F45 Set weighing range(input, output)
signal
Finish Standard
F42 Set time of finish signal F46
(Manual, Time, Steady)
F43 Set FALL & No Compensation F47 Weight Hold in weighing finish
F48 Start Input + ZERO or TARE SET

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 49
5-4-6 Sequence Control 2 set
Sequence Control 2
F40-05
weight
h ig h

lo w

Empty

remain

time
1st,2nd Judge prohibited time
Judge count range
Weight fixed range

Zero/Tare input
Start signal input

Empty signal output

Low output
High output
Finish output (Ref1)

▣ this function was similar to F40-04 but it can be output High signal by Fall Set
HIGH SIGNAL = HIGH SET - FALL
High Set had better use a final target weight also can weigh it correctly by F43
▣ High, Low signal will output by Start Input and will be OFF when it will be
the same or more with Setting weight.
FINISH Signal after High signal OFF will be transmitted by F46 Setting and OFF
Signal will not be ON even if the weight was less than Setting Value.
Also it had better use STOP Signal in No normal case and don’t use it normally.
▣ the weight range became input weighing(+range),
Discharging (- range), or High, Low in the absolute value depending on F45
▣ Auto Fall Compensation(F43) will not be available
▣ Finish signal will be available until empty range in setting F42-00

OUT 1 : LOW OUT 2 : HIGH OUT 3 : FINISH OUT 4 :EMPTY


F40 5 F44 Set prohibited time of weighing
F41 Set Finish weighing time after high Signal F45 Set weighing range(input, output)
Finish Standard
F42 Set time of finish signal F46
(Manual, Time, Steady)
F43 Set FALL & No Compensation F47 Weight Hold in weighing finish
F48 Start Input + ZERO or TARE SET

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 50
5-4-7 Sequence Control 3 set
Sequence Control 3
F40-06
weight
fall
high
fall

low

Empty

remain
time
1,2step prohibited time
Fixed weight range

zero/tare input
empty signal

low signal

high signal
(Refer1)
finish signal

▣ this function was similar to F40-04 but it can be output High, Low signal by Fall
Set of High, Low
LOW SIGNAL = LOW SET – LOW FALL
HIGH SIGNAL = HIGH SET – HIGH FALL
A different Kinds of products can be controlled by 1step signal by an individual
FALL Set in HIGH,LOW Set.
▣ Auto Fall Compensation was not available.
▣ High, Low signal will output by Start Input and will be OFF when it will be
the same or more with Setting weight.
FINISH Signal after High signal OFF will be transmitted by F46 Setting and OFF
Signal will not be ON even if the weight was less than Setting Value.
Also it had better use STOP Signal in No normal case and don’t use it normally.
▣ the weight range became input weighing(+range),
Discharging (- range), or High, Low in the absolute value depending on F45
▣ Auto Fall Compensation(F43) will not be available
▣ Finish signal will be available until empty range in setting F42-00
OUT 1 : LOW OUT 2 : HIGH OUT 3 : FINISH OUT 4 :EMPTY
F40 6 F44 Set prohibited time of weighing
F41 Set Finish weighing time after high Signal F45 Set weighing range(input, output)
Finish Standard
F42 Set time of finish signal F46
(Manual, Time, Steady)
F43 Set FALL & No Compensation F47 Weight Hold in weighing finish
F48 Start Input + ZERO or TARE SET

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 51
5-4-8 Sequence Control 4 set
Sequence Control 4
F40-07
weight
fall
high
fall

low

Empty

remain

time
1,2step prohibited time
Fixed weight range

zero/tare input
empty signal

low signal

high signal

finish signal (Refer1)

▣ this function was similar to F40-06 but High, Low signal Can be set by Individual
weight each other
LOW SIGNAL = LOW SET – LOW FALL
HIGH SIGNAL = HIGH SET +( HIGH SET – HIGH FALL)
A different Kinds of products can be controlled by 1step signal by an individual
FALL Set in HIGH,LOW Set
▣ Auto Fall Compensation was not available.
▣ High, Low signal will output by Start Input and will be OFF when it will be
the same or more with Setting weight.
FINISH Signal after High signal OFF will be transmitted by F46 Setting and OFF
Signal will not be ON even if the weight was less than Setting Value.
Also it had better use STOP Signal in No normal case and don’t use it normally.
▣ the weight range became input weighing(+range),
Discharging (- range), or High, Low in the absolute value depending on F45
▣ Auto Fall Compensation(F43) will not be available
▣ Finish signal will be available until empty range in setting F42-00
OUT 1 : LOW OUT 2 : HIGH OUT 3 : FINISH OUT 4 :EMPTY
F40 6 F44 Set prohibited time of weighing
F41 Set Finish weighing time after high Signal F45 Set weighing range(input, output)
Finish Standard
F42 Set time of finish signal F46
(Manual, Time, Steady)
F43 Set FALL & No Compensation F47 Weight Hold in weighing finish
F48 Start Input + ZERO or TARE SET

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 52
5-4-8 Check Mode Control 1 set
Check Mode Control 1
F40-08

▣ It occurs signal included in range of present weight by high, low set


And signal according to the weight changing
LOW SIGNAL = EMPTY ~ LOW WEIGHT
NORMAL SIGNAL = LOW WEIGHT ~ HIGH WEIGHT
HIGH SIGNAL = FROM WEIGHT MORE THAN
HIGH WEIGHT
▣ auto fall compensation(F43) was unavailable
▣ the weight range became (+range), (- range), or in the absolute value depending
on F45

OUT 1 : LOW OUT 2 : HIGH OUT 3 : NORMAL OUT 4 :EMPTY


F40 8 F44 NO Available.
Set weighing range
F41 NO Available . F45
(+, -, absolute value range)
Available for manual input
F42 NO Available. F46
(Printer KEY)
F43 NO Available F47 No available
F48 No available

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 53
5-4-10 Check Mode Control 2 set
Check Mode Control 2
F40-09

Fixed Weight Range

▣ It occurs signal included in range of present weight by high, low set


the next Judgment can be judged under empty
LOW SIGNAL = EMPTY ~ LOW WEIGHT
NORMAL SIGNAL = LOW WEIGHT ~ HIGH WEIGHT
HIGH SIGNAL = FROM WEIGHT MORE THAN
HIGH WEIGHT
▣ the weight range became (+range), (- range), or in the absolute value depending
on F45
▣ Finish Signal keeps it until Empty range by t = F42-00

OUT 1 : LOW OUT 2 : HIGH OUT 3 : NORMAL OUT 4 :EMPTY


F40 9 F44 NO Available.
Set weighing range
F41 NO Available . F45
(+, -, absolute value range)
F42 Set the range of Finish Signal F46 Manual or Steady
F43 NO Available F47 Hold Function in weighing
F08 Steady Finish Delay Time SET F48 No available

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 54
5-4-10 Check Mode Control 3 set
Check Mode Control 2
F40-10
weight

high

low

Empty

Fixed weight range time

Start Signal
Prohibited Time
Finish Delay Time
Empty Signal
Low Signal (Refer1)
(T)
High Signal (Refer2)
(T)
Normal signal (T)

▣ Finish Delay Signal will be output by Start Signal and the range signal will
output depending on the HIGH,LOW Set after Finish Delay Time.

OW SIGNAL = EMPTY ~ LOW WEIGHT


NORMAL SIGNAL = LOW WEIGHT ~ HIGH WEIGHT
HIGH SIGNAL = FROM WEIGHT MORE THAN
HIGH WEIGHT
▣ the weight range became (+range), (- range), or in the absolute value depending
on F45
▣ Finish Signal keeps it until Empty range by t = F42-00
▣ Refer 1) Finish Signal keeps it until Empty by F42-00
▣ Refer 2) Finish Signal keeps it until Start Input by F42-00

OUT 1 : LOW OUT 2 : HIGH OUT 3 : NORMAL OUT 4 :EMPTY


F40 10 F44 NO Available.
Set weighing range
F41 NO Available . F45
(+, -, absolute value range)
F42 Set the range of Finish Signal F46 Manual or Steady
F43 NO Available F47 Hold Function in weighing
F48 No available

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 55
Delay time for End Judgment

This is the delay time for the weighing judgement after the
final setting value is operated.
0
0 = 0 sec
F-41
1 = 0.1 sec
99
99 = 9.9 sec
*First Setting : 10(1sec)

Maintaining time for End judgment

Set the maintaining time of End judgment signal.


0 0 = Continuous maintaining
F-42 1 = 0.1 sec
99 99 = 9.9 sec
*First Setting : “00”(Continuous)

Function SETTING For FALL COMPANSTION


0
□ NO FALL COMPENSATION
1 80% FALL COMPENSATION WITH 5Times
F-43 2 90% FALL COMPENSATION WITH 10Tmes

3 90% FALL COMPENSATION WITH 20Times

4 100% FALL COMPENSATION WITH 50Times


Fall Compensation will not be available if setting weight was over ± 10% range

Setting for Judge Prohibited Time


0 Judgment will not be done for this function setting time after
F-44 1st,2nd Relay.
99 *First Setting : “03”(sec)

SETTING For a weighing Range


0
□ " + " only Control Signal possible
1 Absolute Value Control Signal Possible
F-45 2 " + " only Control Signal possible

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 56
SETTING For Finish Judgment system
0 Printer key
1
□ Finished Signal by Safe Signal
F-46 2 Finished Signal by Safe Signal or F41

3 Finished Signal by F41

Automatic Histogram when Finish Relay work

SETTING For Finish Display system


0
□ Display for weight change
F-47 1 Empty Signal or Start Signal

SETTING ZERO ,TARE WHEN START TO IINPUT


0
□ No Function
1 ZERO setting by F07 only
F-48
2 Setting TARE

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 57
5-5. Additional Set up Function

5-5-1 OP - 03 BCD OUTPUT

Parallel BCD OUT is a device to output after make the displayed weight into BCD CODE.
Also, this device is to control, display, record as connected with PLC(Programmed Logic Control).

* A recommend distance is in 10 M
* BCD code makes a denary into 4figure of a binary number

(Example)
In case of BCD 1987, it display 0001 1001 1000 0111
BCD OUTPUT Weight Selecting

b Displayed Weight Value

F50- 1 GROSS Weight

2 NET Weight

BCD OUTPUT POLARITY


b Positive Logic
F51-
1 Negative Logic

▩ Connected Pin Diagram

PIN NO SIGNAL PIN NO SIGNAL

1 GROUND (GND) 26 Hi : Net LOW : Gross


2 1×100 27
3 2×100 28
4 4×100 29
5 8×100 30
6 1×101 31
7 2×101 32
8 4×101 33
9 8×101 34
10 1×102 35
11 2×102 36
12 4×102 37 EX. Vcc
13 8×102 38
14 1×103 39 EX. Vcc
15 2×103 40
16 4×103 41
17 8×103 42 Hi : Positive Polarity
18 1×104 43 Decimal Point 101
19 2×104 44 " 102
20 4×104 45 " 103
21 8×104 46
22 1×105 47 OVER LOAD
23 2×105 48
24 4×105 49 BUSY
25 8×105 50 HOLD (INPUT)

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 58
► 50 PIN CONNECTOR: CHAMP 57-40500(Ampheonol) (Female)
► TTL OPEN-COLLECTOR OUTPUT
► HOLD INPUT should be connected with OPEN COLLECTOR TYPE or Switch Earth.
► And OUTPUT DATA will hold while HOLD INPUT

Weight DATA

BUSY

▪ Signal Logic ① BCD DATA Out → Positive logic


② POLARITY →“―” =H
③ OVER → “ OVER ” = H
④ BUSY → “ BUSY ” = H
⑤ BCD HOLD → “ BUSY ” = L

▩ BCD OUTPUT CIRCUIT

Inner Circuit Other equipment

COMMON(GND) +V
FULLUP RESISTANCE
1

OUTPUT
3

+V
1

Voltage 30V max.


1

Current 30mA max.


3

2
1

vcc
+12V
1

1
42

12

INPUT
3
3

► OUTPUT CIRCUIT IS OPEN COLLECTOR TYPE


► If output demand TTL LEVEL ,insert full up - resistance to a board of BCD OPTION
► When inserting a full up resistance ,please change 5v ∼ 30V in 37,39 NO
Resistance and Voltage .
5V = 1 kΩ , 10V = 2 kΩ , 15V = 2.7kΩ , 24V = 5 kΩ

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 59
5-5-2 OP-04 RS-422 / 485 Serial Interface

z RS-422/485 is to transmit signal by voltage deviation and more stable than others.
z RS- 485 should be connected as follows.
RXD(+) + TXD(+), RXD(-) + TXD(-)
z Please Specially connect them Separately disconnecting AC Power Cable or Other Wire
z Also Cable should be surely connected with Shield Twist Cable.
z Recommend Distance is in 1.2 km .
z It should connect Termination Resistance of 300Ωon the end side of wire.

▩ SINGAL FORMAT
c TYPE : RS-422/485
d FORMAT : ⓐ Baud-Rate : 300 ∼ 38.4k . Selection
ⓑ Data Bit : 7 or 8 (NO Parity)
ⓒ Stop :1
ⓓ Parity Bit : Even, Odd, NO Parity Selection
ⓔ Code : ASCII

0
LSB MSB
1 0 1 2 3 4 5 6
↑ ↘ Data Bit ↙ ↑ ↘ Stop
1.5V Potential Difference 1 Bit
Start 1 bit Parity bit
▩ DATA FORMAT
z It is the Same with RS - 232C

▩ RS-422 / 485 Circuit (9P D-Type Female Connector)

1 FG

2 RXD(+)
3 RXD(-)

6 TXD(+)
7 TXD(-)

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 60
5-4-3 OPTION-05/06 ANALOG OUT
z This option is a device to output and convert the weight value to External device(Recorder
P.L.C Center control so) controlled by Analog Signal.
z The voltage output occurs the voltage according to the size of weight in 0V ~10V.
z The current output occurs the current according to the size of weight in 4mA~20Ma
z The precision of Analog output is Max.1/3000
z It will not be suitable for a high precision over 1/3000

Select Weight for Analog Out

b Displayed Weight value

F60- 1 GROSS Weight

2 NET Weight

Gross or Net Weight can be different with weight value displayed

Select Analog Out Standard

b Max. Weight Standard


F61-
1 Standard value setup by F-63

Select Analog Out Polarity

b Positive out : 4mA, 0V while weight is 0


F62-
1 Negative out : 20mA, 5V, 10V while weight is 0

SET Standard Weight For Analog Out

Analog max out value when weight setup.


F63- * first Setting 000000

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 61
5-4-3V OP-05 voltage (0 ~ 10V) ANALOG OUT
z Voltage output occurs proportionally according to the weight range In 0V ~10V.
z Type of voltage output can be adjusted according to SET UP F60
▩ SPECIFICATION

Output Voltage 0~ 10V DC output


Precision Max 1/3000
Min Load Impedance More than 1 kΩ

▩ Connector Pin Diagram( 9P D-TYPE Female )& Voltage out circuit

1
3 1
6 (+) 1
2
2
7
3
8
4
9
5 (-)5

※ Caution
Because NO 5(-) Terminal is not GND, It should not connect with GND line or BODY GND
▩ Adjustment

1 FG
2 RXD(+)
3 RXD(-)

6 TXD(+)
7 TXD(-)

① If the weight was 0 then 0V and was Max. Weight then 10V.
※Caution
Because display weight and output voltage may be different in case of Gross/Net Weight.
Be careful that the weight setting should be checked out before.
② When output voltage was measured by digital multi-meter then if it was different then adjust
VR1(ZERO), VR2(SPAN) located in Analog output PCB inside Digital Indicator.
▩ How to calibrate for output rate between 0v and 10v.
① Adjust voltage with 0V when display weight is 0(ZERO) by VR1(ZERO)
② Adjust voltage difference with 10V when display weight is 0(ZERO) and Max. weight by
VR2(SPAN)
③ Adjust voltage with 0V when display weight is 0(ZERO) by VR 1(ZERO).

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 62
5-4-3I OP -06 Electric current (4 -20mA) ANALOG OUT
z Voltage output occurs proportionally according to the weight range In 4 ~20 mA
z Type of voltage output can be adjusted according to SET UP F60

▩ Specification

output Voltage 4 ~ 20 mA DC Current out


Precision Max 1/3000
Min. Load Impedance More than 500 Ω

▩ Connector Pin Diagram(9P D-TYPE Female & Current out circuit)

No 1 +

No 5 -

※ Caution
Because NO 5(-) Terminal is not GND, It should not connect with GND line or BODY GND
▩ Adjustment
① If the weight was 0 then 4mA and was Max. Weight then 20mA.
※Caution
Because display weight and output voltage may be different in case of Gross/Net Weight.
Be careful that the weight setting should be checked out before.
② When output voltage was measured by digital multi-meter then if it was different then adjust
VR1(ZERO), VR2(SPAN) located in Analog output PCB inside Digital Indicator.

▩ How to calibrate for output rate between 0v and 10v.


① Adjust voltage with 4mA when display weight is 0(ZERO) by VR1(ZERO)
② Adjust voltage difference with 16mA when display weight is 0(ZERO) and Max.
weight by VR2(SPAN)
③ Adjust voltage with 4mA when display weight is 0(ZERO) by VR 1(ZERO).

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 63
5-4-4. OP-07 PRINTER
z This printer Interface have Centronics Parallel and Serial system.
z SERIAL is from 1 to 999999
z CODE figure is 6(six) and set according each PART.
z SUB TOTAL can be recorded Until 10figures.
z GRD TOTAL can be recorded Until 12figures.
z The data can be kept in spite of stoppage of the power

PRINTER SHEET SELECT

b PRINT SHEET 0
1 PRINT SHEET 1

F71- 2 PRINT SHEET 2


3 PRINT SHEET 3
4 PRINT SHEET 4

PRINT SHEET 0 PRINT SHEET 1

============
DATE : 1999-01-01
============
DATE : 1999-01-01 TIME : 12:35:07
TIME : 12:35:07 CODE : 123456
CODE : 123456 SERIAL PART WEIGHT
SERIAL PART WEIGHT 1 1 1.000
1 1 1.000 kg kg
2 1 1.100 kg ============
DATE : 1999-01-01
3 1 1.200 kg
TIME : 12:35:07
4 1 0.900 kg
CODE : 123456
5 1 1.000 kg
============ SERIAL PART WEIGHT
SUB-TOTAL 2 1 1.000
START : 1998-12-30 8:12 kg
END : 1999-01-01 14:26 ============
PART : 01 SUB-TOTAL
CODE : 123456 START : 1998-12-30 8:12
END : 1999-01-01 14:26
COUNT = 5
PART : 01
WEIGHT = 5.200 kg
CODE : 123456
COUNT = 2
WEIGHT = 2.000 kg

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 64
PRINT SHEET 2 PRINT SHEET 3
===================== =====================
DATE : 2001-01-01 DATE : 2001-01-01
TIME : 12:35:07 TIME : 12:35:07
CODE : 123456 CODE : 123456
SERIAL PART WEIGHT SERIAL PART WEIGHT
GROSS NET WEIGHT GROSS NET WEIGHT
1 1 20.00 kg 1 1 20.00 kg
520.00kg 500.00 kg 520.00kg 500.00 kg
2 1 20.00 kg 2 1 20.00 kg
530.00kg 510.00 kg 530.00kg 510.00 kg
===================== =====================
SUB-TOTAL DATE : 2001-01-01
START : 1998-12-30 12:35 TIME :12 : 38 : 07
END : 1999-01-01 12:38 CODE : 123656
PART : 01 SERIAL PART WEIGHT
CODE : 123456 GROSS NET WEIGHT
COUNT = 5 1 1 20.00 kg
WEIGHT= 110.00kg 520.00kg 500.00 kg
2 1 20.00 kg
530.00kg 510.00 kg
=====================
SUB-TOTAL
START : 1998-12-30 12:35
END : 1999-01-01 12:38
PART : 01
CODE : 123456
COUNT = 5
WEIGHT= 110.00kg

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 65
PRINT SHEET 4
=====================
DATE : 2001-01-01
SERIAL : 123456
CODE : 765432
GROSS : 24.560 Kg.
TARE : 5.670 Kg
NET : 18.890 Kg

=====================
D/T : 2002-05-21 12:38
SERIAL : 123456
CODE : 765432
GROSS : 24.560 Kg.
TARE : 5.670 Kg
NET : 18.890 Kg

=====================
D/T : 2002-05-21 12:45
SERIAL : 123456
CODE : 765432
GROSS : 24.570 Kg.
TARE : 5.670 Kg
NET : 18.900 Kg

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 66
SET LINE FEED FOR PRINTING
0
1 LINE PRINT OUT PER 1COUNT(LINE FEED)
F72 ~
* FIRST SET-UP 00
99

SET SUB TOTAL PRINTER MODE


0 SUB TOTAL PRINT SHEET 0
F73 1 SUB TOTAL PRINT SHERT 1

Sub-total PRINT SHEET 1 Sub-total PRINT SHEET 0

===================
SUB-TOTAL ====================
START : 2000-03-28 12:34 SUB-TOTAL
END : 2000-03-29 9:50 START : 2000-03-28 12:34
PART : 1 END : 2000-03-29 9:50
CODE : 123456 PART : 1
COUNT : 10 CODE : 123456
MIN : 9.998 kg COUNT : 10
MAX : 10.002 kg WEIGHT : 100.000 kg
AVG : 10.000 kg
=====================
===================

▩ PRINTER
PIN NO. Contents PIN NO. Contents

1 STROBE 14 N.C
2 D0 15 N.C
3 D1 16 N.C
4 D2 17 N.C
5 D3 18 GND
6 D4 19 N.C
7 D5 20 N.C
8 D6 21 N.C
9 D7 22 N.C
10 ACK 23 N.C
11 BUSY 24 N.C
12 N.C 25 N.C
13 N.C

CONNECTOR PIN ( 25P D-Type Female Connector)

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 67
5-4-5. OP-10 BCD INPUT.
Parallel BCD input is used to change the PART to the external device.
This device make it effective to weigh a various works changing the PART with a connection of
Computer, PLC, Digital Switch.

The inside circuit of Input & Output circuit use a photo-coupler and was isolated from the external

* Recommend distance is under 10 M


* BCD code makes a denary into 4figure of a binary number
* In case PART 19 displayed with BCE CODE such as 0001 10001
0 = OFF, 1 = ON

▩ 15P D-Type Female Connector


When a additional Input needs, Additional Input will be used except of the external 4EA.

▩ BCD INPUT CIRCUIT

PIN NO SIGNAL PIN NO SIGNAL

1 1×100 9 EARTH (GND)


2 2×100 10

3 4×100 11 AID INPUT 1

4 8×100 12 AID INPUT 2

5 1×101 13 AID INPUT 3

6 2×101 14 AID INPUT 4

7 4×101 15 EARTH (GND)


8 8×101

-----------------------------------------------------------------------------------------------------------------------
FINE DIGITAL INDICATOR FS-2000C 68
FINE INTER KOREA CO,. LTD
780-2 Noonsa Dong, Kwangmyeong City, Gyeonggi-Do KOREA.
TEL : 82 2 899 4472 FAX : 82 2 899 4479
http://www.fineweigher.co.kr Contact : [email protected]

Copyright ® 2009 by FINE Inter Korea Co., Ltd All Rights Reserved. Version : 2009,July.
Any form or Any other means can not be copied without the permission from the publisher.