Lm-900 Users' Manual

The LM-900 is an auto lens blocker designed for precise measurement of various lens types, featuring a user-friendly LCD touch screen. The manual includes safety notices, operational instructions, and maintenance guidelines to ensure proper use and longevity of the device. Key functionalities include lens measurement, PD measurement, and data printing, all adhering to national metrological standards.

Uploaded by

crisbalseca
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
48 views36 pages

Lm-900 Users' Manual

The LM-900 is an auto lens blocker designed for precise measurement of various lens types, featuring a user-friendly LCD touch screen. The manual includes safety notices, operational instructions, and maintenance guidelines to ensure proper use and longevity of the device. Key functionalities include lens measurement, PD measurement, and data printing, all adhering to national metrological standards.

Uploaded by

crisbalseca
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 36

LENS BLOCKER

LM-900 Operation Manual

V.202401
Lens Blocker Operation Manual

CONTENT
1 Product introduction ............................................................................................................................ 2
2 Safe Notice .......................................................................................................................................... 3
2.1 Notice in operation ................................................................................................................... 3
2.2 Notice in storage .......................................................................................................................4
2.3 Notice in moving ...................................................................................................................... 4
2.4 Notice in assembling ................................................................................................................ 4
2.5 Notice in wires connection .......................................................................................................5
2.6 Notice after operation ...............................................................................................................6
2.7 Notice in maintenance ..............................................................................................................6
2.8 Notice in accident .....................................................................................................................7
2.9 Marks on machine .................................................................................................................... 7
3 Introduction of machine ...................................................................................................................... 8
3.1 Lens blocker................................................................................................................................
8
3.2 Operation panel ...................................................................................................................... 11
4. Operation .......................................................................................................................................... 16
4.1 Installation .............................................................................................................................. 16
4.2 Test ..........................................................................................................................................16
4.3 Operation ................................................................................................................................ 17
4.3.1 Lens settings ........................................................................................................................ 17
4.3.2 Lens measurement ...............................................................................................................17
5 Maintenance ................................................................................................................................... 30
5.1 Daily maintenance ..................................................................................................................30
5.2 Replacement of materials .......................................................................................................30
5.3 Machine clean .........................................................................................................................31
5.4 Notice in moving ....................................................................................................................32
5.5 Service information ................................................................................................................ 32
6 Simple fault phenomenon and solution ..........................................................................................33
6.1 Solution to general problem ................................................................................................... 33
7 Performance parameter .................................................................................................................. 34
8 Attached list (packing parts) .................................................................................................... 35

1
1 Product introduction
LM-900 can measure the degree of single vision lens, double vision lens, three visions lens
and progressive lens (PPL)quickly and precisely.
This machine has LCD touch screen, on which can show the measured data and PD of left
and right eyes. And it also can show the centering by diagram marks on the screen and it can
help to adjust the optical center of lens quickly. The 7-inch high-resolution TFT touch screen
can show everything smooth and fine.
There is brief menu. Just pressing the keys on the screen, the graphs will show their
functions accurately.
LM-900 is the auto lens blocker which is specialized for lens measurement. The
Implementation of the enterprise standard Q/ZZGX1-2023, which is in the line with the
national metrological verification procedures JJG 580—2005.
The max degree of this machine:±25.00D.

The function direction of LM-900

Model LM-900

Sphere √

Cylinder √

Prism √

Progressive √

Color lens √

PD, PH √

Print √

Remarks With PD,PH and Print

2
Lens Blocker Operation Manual

2 Safe Notice

In this manual, the following marks mean the safe notice, please refer to the follow:

Notice:there will be dangerous. If not avoid, it can cause damage.

Notice:there will be dangerous, if not avoid, it can cause damage to the function of machine.

2.1 Notice in operation

Notice:

● Don’t open or touch the inner part of machine if not necessary, or it may cause electric
shock or make problem to machine.

● To use the stipulated voltage, if too high or too low, it will influence the operation or even

cause damage to machine or result in electric shock.

● Don’t put the wires under machine, or it will break the wires and cause cutting-out, fire

disaster or electric shock.

● If the metal core of wires is exposed, it should be exchanged right away,otherwise, it will

cause fire disaster or electric shock.

3
2.2 Notice in storage

Notice:

● Don’t put the machine at some place humid or poisonous or with corrosive liquid or air.

● Don’t put machine under the sunshine. To keep the suitable temperature and humidness.

2.3 搬动时应注意的事项
● Notice: if the change of the outside temperature is too big, it will influence the operation of

machine. Then it just need to put machine under the stable temperature for 3-10 hours, then

it can be operated normally.

2.3 Notice in moving

● Don’t pull the wires or cable, which will damage the machine or hurt people.

● Don’t take machine by seizing the LCD screen when moving machine. You should hold

the base by hands. Or it will hurt people or damage machine.

2.4 Notice in assembling

Notice:

● Don’t put machine in humid place. Any water into machine will cause electric shock or

other machine problem.

● To put machine at smooth place. Any shake during operation will make machine fall

down and may hurt people or damage machine.

4
Lens Blocker Operation Manual

Notice:

● Don’t operate machine under sunshine or near filament lamp. Don’t install machine on the

place that the surface with strong light reflect, like glass show box, mirror or polishing

desk. Otherwise, these situation will influence the operation or make error.

● Don’t operate machine at the place against electric fans or air conditioner, otherwise the

dust will be flowed into machine which will influence the preciseness of measured data.

● Don’t operate the machine at the place with dust or humid and hot, which will influence

operation.

2.5 Notice in wires connection

Notice:

● Don’t put too many plugs on one power socket, which will make it too hot and cause fire

disaster.

● There must be ground connection to the power supply to protect the security of people and

machine.

5
2.6 Notice after operation

Notice:

● To cut off power supply when not run the machine, and put on the dust cover. Long time

connection to power supply will short the usage life of machine. And without dust for long

time, the falling dust will influence the preciseness of machine.

● Not operate machine for long time, please pull the plug out from power socket, in case to

prevent from fire disaster.

2.7 Notice in maintenance

Notice:

● If there is any problem with machine and need to take down the machine for check, please

contact with our agent or our branch office. You need to take all the responsible to any

damage to machine from your personal action to it.

● Don’t scrape the protecting lens which under the lens base, otherwise it will very influence

the preciseness of measured data.

6
Lens Blocker Operation Manual

2.8 Notice in accident

Notice:

● If the LCD screen broken, to knock it into pieces and wash and sweep them away by

alcohol, then to burn them down. If hands touch the liquid crystal, then to wash hands with

water right away.

2.9 Marks on machine

Please read the manual carefully.

○ It means that the power supply be cut off.

| It means that the power supply closed.

7
3 Introduction of machine

3.1 Lens Blocker

WIFI pilot Point button


lamp

Sliding nose pad


Lens table knob
frame

lens fixing
Lens table
frame

Lens base Memory switch

Figure 3.1 Outlook of main body

8
Lens Blocker Operation Manual

 lens fixing frame: when fixing the lens, to up the frame and down it with gentle. When take

down lens, to up the frame till there is a sound like “ka”, then the frame will be locked.

 Sliding nose frame: which used for measuring PD. To put the nose pad on the frame and move

the frame, the PD will be measured out automatically.

Port

Main power port

Fuse cover

USB FUSE:F2AL250V

Figure 3.2 Port on the back of machine

 Lens base: which used for putting lens and as the basic point of measurement. Dust cover:

which used for protecting machine from dust. Before operating machine, please take down the

dust cover.

9
 Switch :short-time press to turn on machine, and longer press to turn off machine.

 Point button: to mark the center of lens and the optical axis of astigmatism glasses.

 Lens table button: to dolly moves lens and measure the diameter of lens (the diameter can be

read out by the degree scale under the button.)

 lens table: to fix the direction of lens frame. To put the bottom of lens frame on the lens table

when make the measurement. Don’t move the lens table directly, but just moving it by lens

table knob.

 Printer: to print the measured data.

Note: no operation for some time, machine will turn to screen saver mode. The time of this mode can

be set in Menu. Back to measure mode, just gentle touching on the screen.

10
Lens Blocker Operation Manual

3.2 Operation panel

 Measurement interface Drone Time

L/ R

PD
PH

Figure 3.3 measurement interface

Showed degree: 0.01、0.12、0.25D

Astigmatism mode: +、+/—、—

 Drone

To show the measured point when measure lens.

The shape of drone will change as follow when focusing:

十:Off center

十:close to center( within 0.5△) and it will be showed in screen.

十:focused on center (mark point) and it will be showed on screen.

 L/ R

When the left side or right side (L/ R) has been set and the mark showed up, the degree of the

side will be showed under the mark.

11
S means to measure one side lens.

 PD

To show PD(LPD:left PD;RPD:right PD) and the PD of glasses.

Marks:functions are showed as marks


Setting: to enter into
Cylinder:for positive
and modify the
and negative
system parameter.
conversions

Mode: to change the Print: to print out the


measurement data
mode( single or
progressive
measurement mode)

L/ R: to choose left Clearance: to clear


side or right side the saved data.

Figure 3.4 function of marks on measurement interface

Menu shows:
After entering into Menu, to press the corresponding button to change the menu options, then back
by pressing “exist”.

12
Lens Blocker Operation Manual

Figure3.5 mark setting menu

Figure3.6 display setting menu

13
Figure3.7 system setting menu

Figure3.8 network setting menu

14
Lens Blocker Operation Manual

Figure3.9 user setting menu

User setting menu can set user information and time,like company name and company address

Figure3.10 about setting menu

This menu shows the hardware version and software version,“reset”menu is for factory reset.

Mode choose:

Below icons will be displayed after pressing mode menu

15
Figure3.11 about setting menu

From the left:(1)Dioptric profile mode(2)UV and blue light measurement mode,(3)PD meter

measurement mode,(4)auto identification mode,(5)progressive lens measurement mode(6)

normal lens measurement mode;

4. Operation

4.1 Installation

( 1 ) To check the attached parts: after opening the package, to check all the attached parts

(according to the attached list).

(2)To find a smooth place without direct sunshine and put down machine well.

( 3 ) To connect with power supply: to plug the power output injection of adapter to the power

injection at the back of machine.

(4)To put the grounding terminal well.

4.2 Test

(1)to turn on the power switch, just gentle touching on switch.

(2)the test data should be 0D without any lens. To turn on the power so check if everything is OK.

Note: if all the data showed on measurement menu are not 0, then it means that the protecting glass

is dirty, which is in lens base. To blow off the dust by the balloon. If it is still dirty, then to wipe it

16
Lens Blocker Operation Manual

with lens cloth. It should be cleaned every short time. Be sure to move lens base up while the

protecting lens is in cleaning.

4.3 Operation

4.3.1 Lens settings

1、To clarify the left side and right side of lens:

 No PD measurement, the sliding nose pad frame should be move to left side. L/ R keys

will be workable.

 In PD measuring, the sliding nose pad frame will direct the left or right side and it will

measure out the PD of left or right while the L/ R keys are not workable.

 If no need for choosing left or right side, there will be S showed on the right up corner of

measurement interface.

2、Lens settings

To put the lens on the lens base, and the convexity upward.

Figure 4.1 to put the measured lens

3、Lens fixing

To put the lens fixing frame on lens.

4.3.2 Lens measurement

17
1、Single vision lens measurement

(1) to focus the drone on the center

To move lens to make the drone(十 ) to the center. When the distance between the drone and

center less than 0.5△,the shape of the drone will be as a cross (十);When the drone at the very

place at center, the shape of the drone will be a thicker cross(十). There are two ways to confirm the

measured data:

A:to press the memory key and the data will be saved. When data be saved, the direction type of

astigmatism +/- can be changed. To press clear key, then it will restart the measurement.

B:automatical memory settings, the data will be saved automatically according to the settings,

then there is no need to press memory key anymore, and the direction type of astigmatism +/- can be

changed. To press clear key, then it will restart the measurement.

Far from the centre near the centre

Figure 4.2 measurement of cursor move on lens

18
Lens Blocker Operation Manual

Figure 4.3 measurement of center focusing

2、measurement to lens with frame


Normal measurement

(1)to press L/ R key to move the drone (十) to the center. When the distance between the two

less than 0.5△, the shape of the drone will be as a cross(十);When the drone is just at the very point

of center, the shape will be a thicker cross (十). To press memory key (or automatical memory set

before) , the measured data will be saved.

(2)to press L/ R key to measure the other lens, and repeat action(1).

19
Figure 4.4 measurement of finished lens and save the data of left side

PD measurement: (no such function in Model A) The change of left or right side should be done by

the move of sliding nose pad frame.

(3)to rotate the lens table knob to move the lens table till touch the bottom of lens frame, and

move sliding nose pad frame to the center of lens frame. To move the lens to make the drone(十 )

close to the center,when the distance between the two less than 0.5△, the shape of the drone will be

as a cross(十);When the drone is just at the very point of center, the shape will be a thicker cross

(十). To press memory key (or automatical memory set before) , the measured data will be saved.

20
Lens Blocker Operation Manual

Figure 4.5 measurement of finished lens and save the data of left and right side

(4)to check the other lens, to move the sliding nose pad frame along with the lens frame. Action

as (1).

3、measurement of high index lens:

This machine is manufactured with standard green light source. There is no need to set the Abbe

number when measure the high index lens. The measured data showed on the screen will be the

correct one.

Figure 4.6 PD of lens with frame

4、to press print, then the machine will print out the data.

21
One side Two side

Figure 4.7 print of measured data

5、Measurement of double vision lens:

There are two keys used in this measurement, a. Memory; b.Mode

(1) To put the far vision part(far vision range) of lens on the lens base.

Figure 4.8 measurement of far vision range and measured data saving

(2) Measurement of far vision degree

When the shape of the drone as (十)or (十), to press the Memory key, then the measured data of

far vision degree will be confirmed.

22
Lens Blocker Operation Manual

(3)ADD of double vision lens

To move the lens toward yourself till the near vision part(near vision range) of the lens on the lens

base. To press Mode, there will be AD1 on the screen. When the notice showed under the drone as

“close to the center”(there is no need to focus the center perfectly), to press the Memory , then the

measured data of near vision part will be confirmed. And the measurement finished.

Figure 4.9 to save the measured data of near vision range

6、Measurement of progressive lens (PPL)

To press the lens mode exchange key to enter into the PPL measuring mode. During this

measurement, the Memory key should be used. There are two PPL measurement: non-edgered lens

and lens with frame.

23
Figure 4.10 diagram of progressive lens

●Measurement of non-edgered lens

To mark the lens with the special eyes ports supplied by manufacturer and make the lens

convexity upward. In the near vision range, to keep the horizontal reference line in horizontal and

lock the lens. To measure the far vision range first, and to move the lens up and down and from left

to right to make the drone just on the center of the cross line. When the drone as the thicker cross

(十), and to press the Memory key, the data of far vision range be confirmed. Then to move lens till

the near eyes ports on the lens base, there will be ADD and measured data showed on screen, just to

press Memory to save it.

Figure 4.11 center focusing of far vision range

24
Lens Blocker Operation Manual

Figure 4.12 the biggest data, ADD of near vision range

● measurement of lens with frame

There is no marks for range on lens when measure with the progressive screen. So be careful during

the operation.

(1)to put the lens with frame well

(2)to measure the parameter of the far vision range. The center of far vision range is close to the

center of the frame. The measuring way is same as that of item 6 above.

(3)then to measure the parameter in near vision range.

To move the lens with frame by moving the knob and dam-board outward. The optical center of

near vision range is close to the center of frame, to find the optical center of near vision range by

moving it from left to right. Watching the ADD on the screen, to press the Memory key when it

shows the maximum.

7、The Prism Lens measurement

25
The prism lens measurement

(1) mark the center of the pupil: ask the customer to look away with glasses, and mark the

center of the pupil on each lens.

Note: the diameter of the mark should be less than 1mm. If the mark is too large, it will interfere

with the measurement.

(2)Determine left and right lenses: press the left and right keys to determine the left or right lenses.

(3)Select the prism representation type

Rectangular coordinate system (X-Y):I 5.00cm/m

Polar coordinates system(P-B):1.00cm/m

D 0.03cm/m 0°

(4)Place the lens

Place the lens on the lens holder as shown. Convexity side up.

(5)Fix lens on holder

(6)Place an eye mark on the center of the lens holder (unnecessary to point to the center).

(7)Press the memory key. (note: if the data was wrong, indicate that the mark is just on the

measuring light path, and slightly move the lens to get the measured data).

8、Optical center Marking

Determine the optical center and horizontal direction of the lens to fit into the frame.

(1)Moving the lens make the target to align with the center intersection line, and the shape of

the target changes from (+) to (十).Move the lens supporter so it can near the bottom of the frame.

(2)Set axis value as prescription: through the display of the axis (A XIS), rotate the lens until

the value on the prescription appears. If (十) changes to (十)], the center needs to be realigned.

26
Lens Blocker Operation Manual

(3)Set the prescription value of lens with prism degree: the prism's selection in the menu

enables the prism display to allocate mirror prescription in the same way. When rectangular

coordinates are selected, the letters "I" are displayed on the screen to indicate the base is inward, "O"

means the base is outward, "u" means the base is upward, and "D" means the base is downward.

Place the pending lens on the lens holder. Rotate properly. Make the prism base the same direction as

the prescription.

(4)Lens marking: rotate the dot knob to mark three points on the lens that are level with the lens

platform. Remove the lens, lift the stationary holder of the lens until you hear a "click" and remove

the lens. Don't touch the mark, otherwise the mark will be erased. The mark is not clear.

9、Dioptric profile measurement

press“ ”button , enter the menu“ ” ,

select“ ”to enter dioptric profile measurement mode,When measuring the progressive

film, click the real memory key and the following picture will appear

Near vision
part sign

Far vision
part sign

Figure 4.13 Dioptric profile measurement

This figure can be used to judge the far and near area of the progressive lens. When the color of the

27
figure below is consistent with the color of the left figure, it indicates that the direction of the lens

measurement is correct when the progressive sheet is measured. Conversely, turn the lens 180

direction and measure again

Figure 4.14 Diopter distribution color chart

9、UV and blue light measurement

Press“ ”button, ,enter the menu“ ”,

select“ ”to enter the UV and blue light measurement mode,like the following picture:

Figure 4.15 Initial UV and blue light measurement interface

If the bar like this picture not full when no lens was tested,click to reset and then

test.

28
Lens Blocker Operation Manual

Figure 4.16 Interface of UV and blue light measurement done

10、PD measurement

Click“ ”button,enter the menu“ ”,

select“ ”the PD PH measurement like the following picture::

Figure 4.17 PD measurement model

Sign ,can choose white or black lens measurement model.Take an example

of white lens

29
PD measurement:Press or to adjust PD left and right, can lock the

result.Then click or to adjust.

PH measurement:press to adjust PH, can

lock the result.Then click or to adjust.

5 Maintenance

5.1 Daily maintenance

(1)Sun exposure is strictly prohibited;

( 2) It is strictly forbidden to put heavy objects on the instrument to avoid damage to the

machine cover;

(3)Please keep the equipment and surrounding air clean;

(4)Please keep the equipment clean and do not use volatile solutions, diluents or organic

solutions such as benzene.

(5)If the machine will be stocked for a long time, please turn off the power switch and pull

out the power cord;

( 6 ) When not in use, please cover the instrument with a dust cover to prevent dust

accumulation.

5.2 Replacement of materials

5.2.1 Exchange the printing paper

30
Lens Blocker Operation Manual

(1)Open the printing cover,put into the printing paper and close the cover.

5.3 machine clean

(1)when machine is dirty, please wipe machine with soft dry cloth. If there is any dirty spot, to

wipe by cloth with neutral detergent then clean with soft dry cloth.

(2)to clean the protecting glass by balloon to blow away the dust. If it is still dirty, to wipe it with

clean dry cloth. To clean the protecting glass regularly and move up the lens base when clean it.

Note: don’t scratch the glass, it will influence the preciseness.

(3) Don’t use alcohol, gas, ether or acetone to wipe machine.

(4) to clean the light transfer glass regularly.

The lens base should be taken down while turning off machine and clean the light transfer glass,

then to put on the lens base and turn on machine to calibrate again (Below is the clean to the light

transfer glass after taking down the lens base.)

Figure 5.2 the protecting glass

31
5.4 Notice in moving

(1) to cut off power switch;

(2) to pull out the plug;

(3) to hold machine by hands from the base of machine and move machine;

(4) handle with care, and to put on a smooth table.

5.5 Service information

(1)Repair

Anything problem with machine can not be settle according to the manual, please contact with

the after sales services office

Please offer us with following information about machine:

Model No.: LM-900

Serials number: the number on the sticker

Problem: please try to be detailed(picture and video are the better choice )

(2)The range of maintenance:

We will offer maintenance to machine for the whole usage life to maintain the function

(3) About the useless machine

● Don’t throw it everywhere, the machine and the parts with pollution;

● Before your action with the useless machine, please contact with the professional waste

disposal company or the agent.

32
Lens Blocker Operation Manual

6 Simple fault phenomenon and solution

6.1 solution to general problem

No. fault problem Solution

To check the wires if


There power wires are not
No reaction from screen when in good condition, and
1 connected well, or turn off
turn on machine cut off the power then
machine not correctly
turn on it after a while.

To touch screen and

The screen doesn’t show The screen protection has enter into Menu to set
2
anything suddenly. been set. a reasonable time for

screen protection.

There data is not 0 when there is The light is too strong or the To avoid the strong
3
no lens. protecting glass is dirty light or clean the glass

The data should be

4 Not reaction when press the print No measured data saved. saved then can be

printed.

There is problem with the


5 The marks on lens not clear To change the pen
pen.

Enter into the


Need to calibrate the data
6 Measured data not correct interface for
again
calibration

33
7 Performance parameter
The range of measurement

Sphere 0m-1to±25m-1

Cylinder 0m-1to±10.00m-1

Add 0m-1to±10.00m-1

Astigmatism 0° to 180°

Prism basal Angle 0° to 360°


measurement range

Prism Horizontal 10cm/m,Vertical 15cm/m

The accuracy of measurement

Dioptric (0.01/0.12/0.25)m-1

Prism (0.01/0.12/0.25)cm/m

Mode of measurement

Cylinder +,±,-

Prism X-Y,P-B

Speed of measurement 0.2 second

Diameter od measurable lens 18 to 110mm

Other parameters

Display 1024*600 TFT and touch screen

Printer Thermal printer

Size 255(W)x 212(L)x 473(H)mm

Weight About 5.1kg

Power supply AC100~240V 50/60Hz

Power 25VA

34
Lens Blocker Operation Manual

8 Attached list (packing parts)

(1) Lens Blocker 1 unit

(2) Print paper 1 rol

(3) Inkpad oil 1 box

(4) Power cable 1 piece

(5) Operation manual 1 piece

(6) Certificate of quality 1 piece

(7) Maintenance card 1 piece

35